Dwyer DPGAB Manual

Page 1
Series DPGAB & DPGWB Digital Pressure Gages
®
Specications - Installation and Operating Instructions
DPGWB DPGAB
3-3/32 [78.58]
3-59/64
[99.62]
Bulletin A-34D
1-25/32
[45.24]
1/4 NPT
The Series DPGAB & DPGWB Digital Pressure Gages have a precise ±0.5% full scale accuracy. The 4 digit digital display will reduce the potential for errors in readings by eliminating parallax error commonly produced with analog gages.
The DPGAB & DPGWB are packaged in an ABS plastic case. The unit is battery powered and has an auto-shut off to conserve battery life. A two button key pad allows easy access to features without the need to work through complex menus. These features include, auto zero and conversion of the pressure units. Readings may be converted to various engineering units. See reverse page for available units.
OPERATING INSTRUCTIONS
For complete instructions on the two button operation see the ow chart provided on
back. On/Off: Press ON/OFF to toggle on and off. Device will automatically shut off after 20 minutes if not used. Zero: Zeros the display - Push & hold for 2 seconds to zero the display. If after pressing
zero the display does not re-zero, the unit is no longer meeting specied accuracy and
should be sent back to the factory for calibration. Do not use the zero button when pressure is applied.
Units: To change units, press both On/Off and Zero buttons simultaneously. Overpressure Indicator: The LCD will ash “OFL”, if the pressure applied is over
105% F.S. “OFL” may also be displayed when the device subjected to electromagnetic
interference between 0.9 to 1 GHz.
INSTALLATION
When installing gage always use 1˝ hex at the base of the housing to tighten the gage to a mating tting. Do not apply wrench to housing.
MAINTENANCE
Battery Removal: Remove the screw on backplate. Lift backplate off by hand. To reassemble replace backplate and screw.
Replace batteries per polarity indicators.
SPECIFICATIONS Service: DPGAB: Air and compatible gases; DPGWB: Liquids and compatible
gases.
Wetted Materials: DPGAB: 316L SS, silicone sensor; DPGWB: Type 316L SS. Housing Materials: ABS plastic. Accuracy: ±0.5% FS (Includes linearity, hysteresis, repeatability). Pressure Limits: 2X pressure range. Vacuum range max pressure is 30 psig. Temperature Limits: 30 to 120°F (-1 to 49°C). Thermal Effect: 0.05% FS/°F. Size: 2.62˝ O.D. x 1.52˝ deep. Process Connections: 1/4˝ male NPT. Display: 4-digit LCD (.425˝ H x .234˝ W digits). Power Requirements: 9 V alkaline battery. Battery included but not connected. Auto Shut-off: 20 minute auto shut-off. Weight: 5.6 oz (160 g). Agency Approvals: CE.
A “LOW BAT” descriptor indication will appear on the display when batteries need to
be replaced.
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
www.dwyer-inst.com
e-mail: info@dwyermail.com
Page 2
DPGAB & DPGWB BUTTON OPERATION
To Turn On & Off:
Press to turn unit on. Press again to turn unit off.
To Zero Display:
To zero display, disconnect gage from any pressure source. Press
and hold zero button until the LCD displays “- - -”, then release.
To Change Units:
To change units, disconnect gage from any pressure source. Press both & buttons simultaneously until desired units are displayed.
DPGAB MODELS
Model Range DPGAB-00 DPGAB-01 DPGAB-02 DPGAB-03 DPGAB-04 DPGAB-05 DPGAB-06 DPGAB-07 DPGAB-08 DPGAB-09 DPGAB-10 DPGAB-11
DPGWB MODELS
Model Range DPGWB-00 DPGWB-04 DPGWB-05 DPGWB-06 DPGWB-07 DPGWB-08 DPGWB-09 DPGWB-10 DPGWB-11
30˝ Hg to 0 (vac) 0 to 20˝ w.c.
0 to 1 psi 0 to 2 psi 0 to 5 psi 0 to 15 psi 0 to 30 psi 0 to 50 psi 0 to 100 psi 0 to 200 psi 0 to 300 psi 0 to 500 psi
30˝ Hg to 0 (vac)
0 to 5 psi 0 to 15 psi 0 to 30 psi 0 to 50 psi 0 to 100 psi 0 to 200 psi 0 to 300 psi 0 to 500 psi
Pressure Ranges Resolution psi kg/cm2bar in Hg ft wc kPa oz/in2in wc mbar cm wc mm Hg psi
-14.70
-1.033
-1.013
-29.93
-33.94
-101.4
-235.2
-407.3
-1013
-1034
-761
0.722
.0508
.0498
1.471
1.667
4.980
11.55
20.00
49.80
1.000
.0703
.0689
2.036
2.307
6.89
16.00
2.000
.1406
.1379
4.072
4.614
5.000
.3515
.3447
15.00
1.055
30.00
2.109
50.00
3.515
100.0
7.03
200.0
14.06
300.0
21.09
500.0
35.15
Pressure Ranges Resolution psi kg/cm2bar in Hg ft wc kPa oz/in2in wc mbar cm wc mm Hg psi
-14.70
-1.033
5.000
.3515
15.00
1.055
30.00
2.109
50.00
3.515
100.0
7.03
200.0
14.06
300.0
21.09
500.0
35.15
1.034
2.068
3.447
6.89
13.79
20.68
34.47
-1.013 .3447
1.034
2.068
3.447
6.89
13.79
20.68
34.47
10.18
30.54
61.1
101.8
203.6
407.2 611 1018
-29.93
10.18
30.54
61.1
101.8
203.6
407.2 611 1018
11.53
34.60
69.2
115.3
230.7
461.3 692 1153
-33.94
11.53
34.60
69.2
115.3
230.7
461.3 692 1153
13.79
34.47
103.4
206.8
344.7 689 1379 2068 3447
-101.4
34.47
103.4
206.8
344.7 689 1379 2068 3447
32.00
80.0
240.0
480.0 800 1600 3200 4800
-235.2
80.0
240.0
480.0 800 1600 3200 4800
27.68
55.4
138.4
415.2 830 1384 2768
-407.3
138.4
415.2 830 1384 2768
68.9
137.9
344.7 1034 2068 3447
-1013
344.7 1034 2068 3447
50.8
70.3
140.6
351.5 1055 2109 3515
-1034
351.5 1055 2109 3515
37.37
51.7
103.4
258.6 776 1551 2586
-761
258.6 776 1551 2586
0.01
0.001
0.001
0.001
0.002
0.01
0.01
0.02
0.1
0.1
0.1
0.2
0.01
0.002
0.01
0.01
0.02
0.1
0.1
0.1
0.2