Series DPG-100 Digital Pressure Gage
3
[76.20]
3-5/8
[91.39]
1
/4 NPT
3
-3/4
[
95.25]
S
TANDARD RANGES
C
ERAMIC RANGES
1
-5/8
[40.39]
(3) MOUNTING HOLES FOR #6 FLAT HEAD SCREW
EQUALLY SPACED ON A 3-5/8 (92.08) B.C.
4-1/16
(103.19)
3-1/4 (82.55)
HOLE IN PANEL
5/64
(1.98)
9/32 (7.14) MAX
PANEL THICKNESS
Specifications - Installation and Operating Instructions
he Series DPG-100 Digital Pressure Gage has a precise +/- 0.25% full scale
T
ccuracy. The 4 digit digital display will reduce the potential for errors in readings
a
y eliminating parallax error commonly produced with analog gages.
b
The DPG-100 is packaged in a durable extruded aluminum case designed to meet
NEMA 4/4X. The unit is battery powered and has an auto-shut off to conserve battery life. A four button key pad allows easy access to features without the need to
ork through complex menus or difficult key combinations. These features include
w
acklight, peak and valley, auto zero and conversion of the pressure units.
b
eadings may be converted to various engineering units. See reverse page for
R
available units.
DPG Quick Start Button Operation
For complete detailed instructions on the four button operation see the Operation
Chart provided on back.
On/Backlight: Push to turn on; Press again for backlight. Press and
hold for 5 seconds to turn unit off.
Zero: Zeros the display - Push & hold for 2 seconds to zero the display.
If after pressing zero the display does not re-zero, the unit is no longer
meeting specified accuracy and should be sent back to the factory for
calibration. Do not use the zero button when pressure is applied.
Units: Scroll (see next page for Model/Range Chart).
P/V: Peak recorded reading retrieved by initial push of button; push
again retrieves the valley or lowest value recorded.
INSTALLATION
When installing gage always use 1˝ hex at the base of the housing to tighten the
gage to a mating fitting. Do not apply wrench to housing.
MAINTENANCE
A “LOW BAT” descriptor indication will appear on the display when batteries need
to be replaced.
Battery Replacement: Remove the three screws on backplate. Lift backplate off
by hand. Replace batteries per polarity indicators. To reassemble, ensure O-Ring
seats into housing evenly and press until seated completely. Replace screws.
Bulletin A-34
Service: Liquids and combustible
ompatible gases (for FM listings see
c
gency approvals below).
a
etted Materials: Type 316L SS.
W
Housing Materials: Black polycarbon-
ate front & back cover, anodized aluminum extruded housing with
recessed grooves, polycarbonate
verlay,Buna-N O-Rings, 316L SS
o
sensor construction.
Accuracy: 0.25% full scale; Ceramic
Ranges: 0.5% full scale; ± 1 least significant digit @ 70°F (21°C) . (Includes
linearity, hysteresis, repeatability).
Pressure Limit: 2x pressure range for
models ≤1000 psi;5000 psi for 3000
psi range;7500 psi for 5000 psi range.
Enclosure Rating: Designed to meet
NEMA 4/4X (IP65).
Temperature Limits: 0 to130°F
(-18 to 55°C).
Thermal Effect: Between 70 to 130°F
is 0.016%/F; Between 32 to 70°F
is0.026%/F; Between 10 to 32°Fis
0.09%/F; Between 0 to 10°Fis
0.50%/F.
*Note: Gages with ceramic sensor are not CE and FM approved.
WARNING: DO NOT pressurize DPG over maximum allowable pressure limits.
Extreme damage can occur if limits are exceeded (see Pressure Limit under
Specifications). An overflow “OFL” designator will flash to acknowledge that the
gage’s pressure range has been exceeded.
PANEL MOUNT OPTION
Drill a 3-1/4˝ (82.55) diameter hole in a panel, locate three mounting holes for a #6
screw 120° apart on a 3-58˝ (92.08) bolt circle.
Size: 3.00˝ OD x 1.90 deep (max).
rocess Connection: 1/4˝ male NPT.
P
eight: 8.84 oz (275 g).
W
isplay: 4 digit (.425 H x .234W dig-
D
its).
Sample Rate: 2 sec.
Power Requirements: (2) AAA alka-
line batteries, included, user replace-
ble.
a
Battery Life: 2000 hours typi-cal; low
battery indicator.
Auto Shut-Off: Gage: 60 minute auto
shut-off. Auto shut-off may be
disengaged; Backlight: 2 minute auto
shut-off.
Agency Approvals: FM approved to
be intrinsically safe for Class I,
Division I, Groups A,B, C and D, for
ranges 0-15 to0-3000 psi. CE EMC
Directive:
EN 61000-4-2
EN 61000-4-3 & ENV50204
EN 61000-6-2
EN 61000-6-4
EN 55011.
WARNING: Substitution of components may impair intrinsic safety. Batteries must
be changed in a non-hazardous location only. DO NOT mix batteries from different
manufacturers. DO NOT mix old batteries with new batteries.
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. Box 373 • Michigan City, IN 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyermail.com
LOW BATTERY INDICATION
U
NITS
PRESSURE READING
AUTO SHUT-OFF
I
NDICATOR
PG-100 DISPLAY
D
DPG Button Operation
Button
plus
O
T
T
Turn Backlight Off
Turn Unit Off
Zero Gage
U
V
View Valley
Reset Peak & Valley
Exit Peak & Valley
Disengage Auto Shut-Off
peration
urn Unit On
urn Backlight On (Auto Shut-Off: 2 Min)
nit Selection
iew Peak
Action
ress and hold ON button
P
hile gage is on press ON button
W
hile backlight is on press ON button
W
While gage is on, press and hold ON button 5 seconds for unit to turn off.
With no pressure applied to the gage, press and hold ZERO button until display flashes
If gage will not zero, then gage is no longer meeting specified accuracy and should be
sent back to the factory for recalibration
Note: Gage is not to be zeroed while pressure is applied
ress UNIT button to change units on gage (Not all units are available for each range,
P
lease refer to model/range chart located below for available units)
p
ress P/V button to read peak (highest pressure) measurement
P
ress P/V button again to read valley (lowest pressure) measurement
P
Press and hold P/V button to reset peak and valley
Press ON button to exit peak/valley
Auto shut-off will turn off an idle gage after 60 minutes. To disengage auto shut-off,
press and hold ON button, then press and hold P/V button. A decimal will blink in the
lower left hand corner of the display to indicate the auto-shut off is turned off. To turn on
auto shut-off, repeat above procedure until decimal point stops blinking
Model/Range Chart
Range
Pressure Ranges
Model
DPG-100*
DPG-102
DPG-103
DPG-104
DPG-105
DPG-106
DPG-107
DPG-108
DPG-109
DPG-110
DPG-111*
Compound Ranges available: DPG-120* Range: 30˝ Hg-0-15 psi; DPG-121* Range: 30˝ Hg-0-
30 psi; DPG-122* Range: 30˝ Hg-0-45 psi; DPG-123*: Range 30˝ Hg-0-60 psi; DPG-124*: 30˝
Hg-0-100 psi.
*Optional ceramic sensor available. Add “C” to the range (Ex. DPG-100C
psi
-14.70-0
15.00
30.00
50.00
100.0
200.0
300.0
500.0
1000
3000
5000
kg/cm
-1.033
1.055
2.109
3.515
7.03
14.06
21.09
35.15
70.3
210.9
351.5
2
bar
-1.013
1.034
2.069
3.448
6.895
13.79
20.69
34.48
68.98
206.9
344.8
in Hg
-29.93
30.54
61.08
101.8
203.6
407.2
610.8
1018
2036
6108
ft wc
-33.94
34.61
69.21
115.4
230.7
461.4
692.1
1154
2307
6921
kPa
-101.4
103.4
206.9
344.8
689.5
1379
2069
3448
6895
oz/in
-235.2
240
480
800
1600
3200
4800
8000
2
in wc
-407.3
415.2
830.4
1384
2768
5536
8304
mbar
-1013
1034
2069
3448
6895
).
cm wc
-1034
1055
2109
3515
7031
mm Hg
-760.7
775.7
1551
2586
5172
©Copyright 2014 Dwyer Instruments, Inc. Printed in U.S.A. 3/14 FR# 18-443202-04 Rev. 11