Series DPG-100 Digital Pressure Gage
(3) MOUNTING HOLES FOR #6 FLAT HEAD SCREW
Specications - Installation and Operating Instructions
The SERIES DPG-100 Digital Pressure Gage has a precise ±0.25% full scale
accuracy. The 4 digit digital display will reduce the potential for errors in readings by
eliminating parallax error commonly produced with analog gages.
The DPG-100 is packaged in a durable extruded aluminum case designed to meet
NEMA 4/4X. The unit is battery powered and has an auto-shut off to conserve
battery life. A four button key pad allows easy access to features without the need
to work through complex menus or difcult key combinations. These features include
backlight, peak and valley, auto zero and conversion of the pressure units. Readings
may be converted to various engineering units. See reverse page for available units.
DPG Quick Start Button Operation
For complete detailed instructions on the four button operation see the Operation
Chart provided on back.
On/Backlight: Push to turn on; Press again for backlight. Press and hold
for 5 seconds to turn unit off.
Zero: Zeros the display - Press the ZERO button until dashes “---”
are seen on the LCD, then release the button. If after pressing zero the
display does not re-zero, the unit is no longer meeting specied accuracy
and should be sent back to the factory for calibration. Do not use the zero
button when pressure is applied.
Units: Scroll (see next page for Model/Range Chart).
P/V: Peak recorded reading retrieved by initial push of button; push again
retrieves the valley or lowest value recorded.
INSTALLATION
When installing gage always use 1˝ hex at the base of the housing to tighten the gage
to a mating tting. Do not apply wrench to housing.
MAINTENANCE
A “LOW BAT” descriptor indication will appear on the display when batteries need to
be replaced.
Battery Replacement: Remove the three screws on backplate. Lift backplate off by
hand. Replace batteries per polarity indicators. To reassemble, ensure O-ring seats
into housing evenly and press until seated completely. Replace screws.
WARNING
with new batteries. Use 1.5 volt AAA alkaline batteries Duracell
Substitution of components may impair intrinsic safety. Batteries
must be changed in a non-hazardous location only. DO NOT mix
batteries from different manufacturers. DO NOT mix old batteries
®
MN2400.
Bulletin A-34
CERAMIC RANGES
3
[76.20]
3-5/8
3-3/4
[91.39]
[95.25]
1/4 NPT
SPECIFICATIONS
Service: Liquids and combustible compatible gases (for FM listings see agency
approvals below).
Wetted Materials: Type 316L SS.
Housing Materials: Black polycarbonate front & back cover, anodized aluminum
extruded housing with recessed grooves, polycarbonate overlay, Buna-N O-rings,
316L SS sensor construction.
Accuracy: 0.25% full scale; Ceramic Ranges: 0.5% full scale; ±1 least signicant
digit @ 70°F (21°C) . (Includes linearity, hysteresis, repeatability).
Pressure Limit: 2x pressure range for models ≤1000 psi;5000 psi for 3000 psi
range;7500 psi for 5000 psi range.
Enclosure Rating: Designed to meet NEMA 4/4X (IP65).
Temperature Limits: 0 to130°F
(-18 to 55°C).
Thermal Effect: Between 70 to 130°F is 0.016%/F; Between 32 to 70°F is
0.026%/F; Between 10 to 32°Fis 0.09%/F; Between 0 to 10°Fis 0.50%/F.
Size: 3.00˝ OD x 1.90 deep (max).
Process Connection: 1/4˝ male NPT.
Weight: 8.84 oz (275 g).
Display: 4 digit (.425 H x .234W digits).
Sample Rate: 2 sec.
Power Requirements: (2) AAA alkaline batteries, included, user replaceable.
Battery Life: 2000 hours typical; low battery indicator (60 hours in continuous use).
Auto Shut-Off: Gage: 60 minute auto shut-off. Auto shut-off may be
disengaged; Backlight: 2 minute auto shut-off.
Agency Approvals: FM, CE.
*Note: Gages with ceramic sensor are not CE and FM approved.
WARNING: DO NOT pressurize DPG over maximum allowable pressure limits.
Extreme damage can occur if limits are exceeded (see Pressure Limit under
Specications). An overow “OFL” designator will ash to acknowledge that the gage’s
pressure range has been exceeded.
PANEL MOUNT OPTION
Drill a 3-1/4˝ (82.55) diameter hole in a panel, locate three mounting holes for a #6
screw 120° apart on a 3-58˝ (92.08) bolt circle.
EQUALLY SPACED ON A 3-5/8 [92.08] B.C.
3-1/4 [82.55]
HOLE IN PANEL
4-1/16
[103.19]
1-5/8
[40.39]
WARNING
WARNING
Enclosure parts are constructed of plastic. When cleaning the
exterior surfaces use a damp cloth to avoid electrostatic sparking.
Enclosure parts are constructed of aluminum. Enclosure must be
protected from mechanical friction and impact with iron/steel to
prevent ignition capable sparks.
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
5/64
[1.98]
9/32 [7.14] MAX
www.dwyer-inst.com
e-mail: info@dwyermail.com

LOW BATTERY INDICATION
UNITS
PRESSURE READING
A
INDICA
DPG-100 DISPLAY
UTO SHUT-OFF
TOR
DPG Button Operation
Button Operation Action
Turn Unit On
Turn Backlight On (Auto Shut-Off: 2 Min)
Turn Backlight Off
Turn Unit Off
Zero Gage With no pressure applied to the gage, press the ZERO button until dashes “---”
Unit Selection Press UNIT button to change units on gage (Not all units are available for each
Press and hold ON button
While gage is on press ON button
While backlight is on press ON button
While gage is on, press and hold ON button 5 seconds for unit to turn off.
are seen on the LCD, then release
If gage will not zero, then gage is no longer meeting specied accuracy and
should be sent back to the factory for recalibration
Note: Gage is not to be zeroed while pressure is applied
range, please refer to model/range chart located below for available units)
View Peak
View Valley
Reset Peak & Valley
Exit Peak & Valley
Disengage Auto Shut-Off Auto shut-off will turn off an idle gage after 60 minutes. To disengage auto shut-
plus
Model/Range Chart
Model Range psi
DPG-100*
DPG-102
DPG-103
DPG-104
DPG-105
DPG-106
DPG-107
DPG-108
DPG-109
DPG-110
DPG-111*
Compound Ranges available: DPG-120* Range: 30˝ Hg-0-15 psi; DPG-121* Range:
30˝ Hg-0-30 psi; DPG-122* Range: 30˝ Hg-0-45 psi; DPG-123*: Range 30˝ Hg-0-60
psi; DPG-124*: 30˝ Hg-0-100 psi.
*Optional ceramic sensor available. Add “C” to the range (Ex. DPG-100C).
-14.70-0
15.00
30.00
50.00
100.0
200.0
300.0
500.0
1000
3000
5000
Pressure Ranges
2
bar in Hg ft wc kPa oz/in2in wc mbar cm wc mm Hg
kg/cm
-1.033
-1.013
-29.93
1.055
2.109
3.515
7.03
14.06
21.09
35.15
70.3
210.9
351.5
1.034
2.069
3.448
6.895
13.79
20.69
34.48
68.98
206.9
344.8
30.54
61.08
101.8
203.6
407.2
610.8
1018
2036
6108
-33.94
34.61
69.21
115.4
230.7
461.4
692.1
1154
2307
6921
Press P/V button to read peak (highest pressure) measurement
Press P/V button again to read valley (lowest pressure) measurement
Press and hold P/V button to reset peak and valley
Press ON button to exit peak/valley
off, press and hold ON button, then press and hold P/V button. A decimal will
blink in the lower left hand corner of the display to indicate the auto-shut off is
turned off. To turn on auto shut-off, repeat above procedure until decimal point
stops blinking
-101.4
-235.2
240
480
800
1600
3200
4800
8000
-407.3
415.2
830.4
1384
2768
5536
8304
103.4
206.9
344.8
689.5
1379
2069
3448
6895
-1013
1034
2069
3448
6895
-1034
1055
2109
3515
7031
-760.7
775.7
1551
2586
5172