Dwyer DMF User Manual

Page 1
Technical Data Sheet No. TD9805M Rev. L Date of issue: September, 2006
OPERATING MANUAL
FOR DMF
DIGITAL MASS FLOW CONTROLLERS
P.O. Box 373 Michigan City, IN 46361 USA Phone: (219) 879 8000 FAX: (219) 879 9057
E mail: info@dwyer.com
Internet: http://www.dwyer-inst.com
Page 2
TABLE OF CONTENTS
1. UNPACKING THE DMF MASS FLOW CONTROLLER........................
1.1 Inspect Package for External Damage..............................................
1.2 Unpack the Mass Flow Controller.......................................................
1.3 Returning Merchandise for Repair.....................................................
2. INSTALLATION....................................................................
2.1 Primary Gas Connections.................................................................
2.2 Electrical Connections......................................................................
2.3 Communication Parameters and Connections................................
3. PRINCIPLE OF OPERATION..................................................
4. SPECIFICATIONS..................................................................
4.1 DMF 41 Mass Flow Controllers.........................................................
4.2 CE Compliance................................................................................
5. OPERATING INSTRUCTIONS.....................................................
5.1 Preparation and Warm Up................................................................
5.2 Flow Signal Output Readings..............................................................
5.3 Swamping Condition...........................................................................
5.4 Set Point Reference Signal .............................................................
5.5 Valve OFF Control ..........................................................................
5.6 Valve Open/Purge ............................................................................
5.7 Analog Interface Configuration...........................................................
11 11 11 11 12 12 12 13
1 1 1 1
1 1 2 2
6
6 7 8
6. MAINTENANCE...................................................................3
6.1 Introduction........................................................................................
6.2 Flow Path Cleaning..............................................................................
6.2.1 Restrictor Flow Element (RFE).................................................
6.2.2 DMF 41401-41411 models..........................................................
6.2.3 DMF 41431-41842 models..........................................................
6.2.4 Valve Maintenance ...................................................................
7. CALIBRATION PROCEDURES....................................................
7.1 Flow Calibration...................................................................................
7.2 Calibration of DMF Mass Flow Controllers.......................................
8. TROUBLESHOOTING.............................................................
8.1 Common Conditions...........................................................................
8.2 Technical Assistance............................................................................
8.3 Troubleshooting Guide....................................................................
9. CALIBRATION CONVERSIONS FROM REFERENCE GASES................
13 13 14 14 14 14 15
15 15 16
17 17 17 18
20
Page 3
APPENDIX 1 COMPONENT DIAGRAM......................................................
21
APPENDIX 2 GAS FACTOR TABLE ("K" FACTORS).....................................
APPENDIX 3 DIMENSIONAL DRAWINGS..................................................
APPENDIX 4 SENDING COMMANDS TO THE DMF........................................
APPENDIX 5 SDPROC TABLES: GAS DEPENDENT VARIABLES....................
GAS INDEPENDENT VARIABLES...............
APPENDIX 6 WARRANTY...........................................................................
25
29
31
37
39
41
Page 4
1. UNPACKING THE DMF MASS FLOW CONTROLLER
1.1 Inspect Package for External Damage
Your DMF Mass Flow Controller was carefully packed in a sturdy cardboard car­ton, with anti-static cushioning materials to withstand shipping shock. Upon receipt, inspect the package for possible external damage. In case of external damage to the package contact the shipping company immediately.
1.2 Unpack the Mass Flow Controller
Open the carton carefully from the top and inspect for any sign of concealed ship­ping damage. In addition to contacting the shipping carrier please forward a copy of any damage report to your distributor or Dwyer directly.
When unpacking the instrument please make sure that you hav e all the items indi­cated on the Packing List.Please report any shortages promptly.
1.3 Returning Merchandise for Repair
Please contact the customer service representative of your distributor or Dwyer if you purchased your Mass Flow Controller directly, and request a Return
Authorization Number (RAN). Equipment returned without an RAN will not be accepted. Dwyer reserves the right to charge a fee to the customer for equip-
ment returned under warranty claims if the instruments are tested to be free from warrantied defects.
Shipping charges are borne by the customer.Meters returned "collect" will not be accepted!
It is mandatory that any equipment returned for servicing be purged and neutral­ized of any dangerous contents including but not limited to toxic, bacterially infec­tious, corrosive or radioactive substances. No work shall be performed on a returned product unless the customer submits a fully executed, signed SAFETY CERTIFICATE. Please request form from the Service Manager.
2. INSTALLATION
2.1 Primary Gas Connections
Please note that the DMF Mass Flow Controller will not operate with liquids.Only clean gases are allowed to be introduced into the instrument. If gases are con­taminated they must be filtered to prev ent the introduction of impediments into the sensor.
Caution: DMF transducers should not be used for monitoring OXYGEN gas unless specifically cleaned and prepared for such
application. For more information, contact your distributor or Dwyer.
1
Page 5
Attitude sensitivity of the Mass Flow Controller is +15F. This means that the gas flow path of the Flow Controller must be horizontal within those stated limits. Should there be need for a different orientation of the meter, re-calibration may be neces­sary.It is also preferable to install the DMF transducer in a stable environment, free of frequent and sudden temperature changes, high moisture, and drafts.
Prior to connecting gas lines inspect all parts of the piping system including fer­rules and fittings for dust or other contaminants.
Be sure to observe the direction of gas flow as indicated by the arrow on the front of the meter when connecting the gas system to be monitored.
Insert tubing into the compression fittings until the ends of the properly sized tub­ings home flush against the shoulders of the fittings. Compression fittings are to be tightened according to the manufacturer's instructions to one and one quarter turns. Avoid over tightening which will seriously damage the Restrictor Flow Elements (RFE's)!
DMF transducers are supplied with standard 1/4 inch (DMF 41401-41433) or 3/8 inch (DMF 41842), or optional 1/8 inch inlet and outlet compression fittings which should not be removed unless the meter is being cleaned or calibrated for a new flow range.
Using a Helium Leak Detector or other equivalent method perform a thorough leak test of the entire system. age within stated limits. See specifications in this manual.)
(All DMF's are checked prior to shipment for leak-
2.2 Electrical Connections
DMF transducers require a +15VDC and -15VDC power supply to operate. Additionally, a readout panel meter, digital multimeter, or other equivalent device is required to observe the flow signal in analog mode. A variable analog 0-5VDC reference input is required for DMF models to operate in analog mode. The Dwyer SDPROC accessory Command Modules offer a convenient and compact means to fulfill these needs.
DMF is supplied with a 25 pin "D" connector. Pin diagram is presented in figure b-2.
2.3 Communication Parameters and Connections
Baud rate: 9600 baud Stop bit: 1 Data bits: 8 Parity: NON
RS-232 option: Crossover connection has to be established:
Pin 11 (TX) of the “D” connector has to be connected to RX (pin 2 on the DB9 connector). Pin 24 (RX) of the “D” connector has to be connected to TX (pin 3 on the DB9 connector). Pin 20 (Common) of the “D” connector has to be connected to GND (pin 5 on the DB9 connector).
2
Page 6
RS-485 option:
The RS485 converter/adapter has to be configured for:multidrop, 2 wire, half duplex mode.The transmitter circuit has to be enabled by TD or RTS (depending on which is available on the con v erter/adapter).Settings for the receiver circuit usually should follow the selection made for the transmitter circuit in order to eliminate Echo.
Pin 11 (-) of the “D” connector has to be connected to T- or R- on the RS-485 converter/adapter. Pin 24 (+) of the “D” connector has to be connected to T+ or R+ on the RS-485 converter/adapter. Pin 20 (Common) of the “D” connector has to be connected to GND on the RS-485 converter/adapter.
3
Page 7
FIGURE b-1, WIRING DIAGRAM FOR DMF TRANSDUCERS.
4
Page 8
PIN FUNCTION
1 +15 VDC Power Supply 2 0-5 VDC Flow Signal (4-20mA Option) 3 0-5 VDC Set Point Input (4-20mA Option) 4 Force Valve Open Control 5 Force Valve Closed Control 6 (Reserved) 7 (Reserved) 8 Relay No. 1 - Common Contact 9 Relay No. 1 - Nor mally Open Contact 10 Relay No.2 - Normally Closed Contact 11 RS485 (-) (Optional RS232 TX) 12 (No Connection) 13 Chassis Ground 14 -15 VDC Power Supply 15 Common, Signal Ground For Pin 2 16 Common, Signal Ground For Pin 3 17 (Optional) RS232 Common 18 Common, Power Supply 19 Common 20 Common 21 Relay No.1 - Normally Closed Contact 22 Relay No.2 - Common Contact 23 Relay No.2 - Normally Open Contact 24 RS485 (+) (Optional RS232 RX) 25 Return for Pin 2 (Optional 4-20 mA Only)
FIGURE b-2, DMF 25 PIN "D" CONNECTOR CONFIGURATION
Important notes:
In general, "D" Connector numbering patterns are standardized.There are, how­ever , some connectors with nonconf orming patterns and the numbering sequence on your mating connector may or may not coincide with the numbering sequence shown in our pin configuration table above. It is imperative that you match the appropriate wires in accordance with the correct sequence regardless of the par­ticular numbers displayed on your mating connector.
Make sure power is OFF when connecting or disconnecting any cab les in the system. The (+) and (-) power inputs are each protected by a 500mA M (medium time-lag)
resettable fuse. If a shorting condition or polarity reversal occurs, the fuse will cut power to the flow transducer circuit. Disconnect the power to the unit, remove the faulty condition, and reconnect the power. The fuse will reset once the faulty con­dition has been removed.
Cable length may not exceed 9.5 feet (3 meters).
5
Page 9
Use of the DMF flow transducer in a manner other than that specified in this man-
F
ual or in writing from Dwyer, ma y impair the protection provided b y the equipment.
3. PRINCIPLE OF OPERATION
The stream of gas entering the Mass Flow transducer is split by shunting a small portion of the flow through a capillary stainless steel sensor tube. The remainder of the gas flows through the primary flow conduit. The geometr y of the primary conduit and the sensor tube are designed to ensure laminar flow in each branch. According to principles of fluid dynamics the flow rates of a gas in the two lami­nar flow conduits are proportional to one another.Therefore, the flow rates meas­ured in the sensor tube are directly proportional to the total flow through the trans­ducer.
In order to sense the flow in the sensor tube, heat flux is introduced at two sec­tions of the sensor tube by means of precision wound heater sensor coils.Heat is transferred through the thin wall of the sensor tube to the gas flowing inside. As gas flow takes place heat is carried by the gas stream from the upstream coil to the downstream coil windings. The resultant temperature dependent resistance differential is detected by the electronic control circuit.The measured gradient at the sensor windings is linearly proportional to the instantaneous rate of flow tak­ing place.
An output signal is generated that is a function of the amount of heat carried by the gases to indicate mass molecular based flow rates.
Additionally , DMF model Mass Flo w Controllers incorporate a microprocessor and non-volatile memory that stores all calibration factors and directly controls a pro­portionating solenoid valve. The digital closed loop control system of the DMF continuously compares the mass flow output with the selected flow rate. Deviations from the set point are corrected by compensating valve adjustments, thus maintaining the desired flow parameters with a high degree of accuracy.
4. SPECIFICATIONS
FLOW MEDIUM: Please note that DMF Mass Flow Controllers are
designed to work with clean gases only. Never try to meter or control flow rates of liquids with any DMF.
CALIBRATIONS: Performed at standard conditions [14.7 psia
(1.01 bars) and 70 requested or stated.
ENVIRONMENTAL (PER IEC 664):
Installation Level II;Pollution Degree II.
F (21.1FC)] unless otherwise
6
Page 10
4.1 DMF Mass Flow Controllers
F
F
F
ACCURACY: + 59
F to 77FF (15FC to 25FC) and pressures of 10 to 60 psia (0.7 to 4.1 bars).
REPEATABILITY: +
TEMPERATURE COEFFICIENT: 0.1% of full scale/
1% of full scale, including linearity for gas temperatures ranging from
0.15% of full scale.
C.
PRESSURE COEFFICIENT: 0.01% of full scale/psi (0.07 bar). RESPONSE TIME: DMF 41401-41411: 300ms time constant; approximately 1
second to within +
2% of set flow rate for 25% to 100% of full scale flow.
DMF41431-41842: 600ms time constant; approximately 2
seconds to within +
2% of set flow rate for 25% to 100% of full scale flow.
GAS PRESSURE: 500 psig (34.5 bars) maximum; optimum pressure is 20 psig (1.4 bars); 25 psig (1.7 bars gauge) for DMF 41842.
DIFFERENTIAL PRESSURES REQUIRED: 5 to 50 psig (0.35 to 3.34 bars) differential pressures. Optimum differential pressure is 25 psid (1.7 bars). See Table IV for pressure drops associated with various models and flow rates.
MAXIMUM PRESSURE DIFFERENTIAL: 50 psid for DMF 41401-41433, 40 psid for DMF 41842.
GAS AND AMBIENT TEMPERATURE: 41
F to 122FF (5FC to 50FC).
RELATIVE GAS HUMIDITY: Up to 70%. LEAK INTEGRITY: 1 x 10
-9
sccs He maximum to the outside environment.
ATTITUDE SENSITIVITY: 1% shift for a 90 degree rotation from horizontal to verti­cal; standard calibration is in horizontal position.
OUTPUT SIGNALS: Linear 0-5 VDC (2000 : minimum load impedance); 4-20 mA optional (50-500 :loop resistance); 20 mV peak to peak max noise.
Contact your distributor or Dwyer for optional RS232 or IEEE488 interfaces.
COMMAND SIGNAL: 0-5 VDC (200K :input impedance); 4-20 mA optional. TRANSDUCER INPUT POWER: DMF - +15 +
max; -15 +
5% VDC, 450 mA max; 6.75 watts max; Power inputs are each pro-
5% VDC, 450 mA max, 6.75 watts
tected by a 500mA M (medium time-lag) resettable fuse, and an inverse shunt rectifier diode for polarity protection.
WETTED MATERIALS: 316 stainless steel, 416 stainless steel, VITON7 O-rings; BUNA-N7, NEOPRENE7 or KALREZ7 O-rings are optional.
7
Page 11
Dwyer makes no expressed or implied guarantees of corrosion resistance of mass flow meters as pertains to different flow media reacting with components of meters. It is the customers' sole responsibility to select the model suitable for a particular gas based on the fluid contacting (wetted) materials offered in the dif­ferent models.
INLET AND OUTLET CONNECTIONS: 1/4" (DMF 41401-41433) (DMF 41842) or 3/8" compression fittings standard; 1/8" or 3/8" compression fittings and 1/4" VCR7 fit-
tings are optional. TRANSDUCER INTERFACE CABLE: Flat cable with 25-pin "D" connectors on the
ends is standard.Optional shielded cable is available with male/female 25-pin "D" connector ends. [Cable length may not exceed 9.5 feet (3 meters)]
4.2 CE Compliance
Any model DMF bearing a CE marking on it, is in compliance with the below stated test standards currently accepted.
EMC Compliance with 89/336/EEC as amended; Emission Standard: EN 55011:1991, Group 1, Class A Immunity Standard: EN 55082-1:1992
8
Page 12
FLOW RANGES
TABLE I DMF 41 LOW FLOW MASS FLOW CONTROLLERS*
CODE
401 0 to 10 07 0 to 1
402 0 to 20 08 0 to 2
403 0 to 50 09 0 to 5
404 0 to 100 10 0 to 10
405 0 to 200
406 0 to 500
scc/min [N2]
CODE
std liters/min [N
TABLE II DMF 36 MEDIUM FLOW MASS FLOW CONTROLLERS*
CODE standard liters/min [N2]
11 0 to 15
30 20
31 30
32 40
33 50
]
2
TABLE III DMF 46 HIGH FLOW MASS FLOW CONTROLLERS*
CODE standard liters/min [N2]
40 60
41 80
42 100
* Flow rates are stated for Nitrogen at STP conditions [i.e. 70FF (21.1FC) at 1 atm].
For other gases use the K factor as a multiplier from APPENDIX 2.
9
Page 13
TABLE IV PRESSURE DROPS
MODEL
DMF 41411 up to 10 720 1.06 75
DMF 41431
DMF 41433 50 5440 8.00 551
DMF 41842 100 12850 18.89 1302
FLOW RATE
[std liters/min]
15 2630 3.87 266
30 2380 3.50 241
MAXIMUM PRESSURE DROP
[mm H2O] [psid] [mbar]
TABLE V APPROXIMATE WEIGHTS
MODEL WEIGHT SHIPPING WEIGHT
DMF 41401-41411 controller 2.20 lbs (1.00 kg) 3.70 lbs (1.68 kg)
DMF 41431-41842 controller 2.84 lbs (1.29 kg) 4.34 lbs (1.97 kg)
10
Page 14
5. OPERATING INSTRUCTIONS
5.1 Preparation and Warm Up
It is assumed that the Mass Flow Controller or Controller has been correctly installed and thoroughly leak tested as described in section (2). Make sure the flow source is OFF. Power up the transducer using your own power supply (or switch the POWER switch to the ON position at the front panel of your SDPROC Command Module). Allow the Mass Flow Meter or Controller to warm-up for a minimum of 15 minutes.
During initial powering of the DMF transducer, the flow output signal will be indi­cating a higher than usual output. This is indication that the DMF transducer has not yet attained it's minimum operating temperature. This condition will automat­ically cancel within a few minutes and the transducer should eventually zero.
Caution: If the valve is left in the AUTO (control) or OPEN mode for an extended period of time, it may become warm or even hot to the touch.
Use care in avoiding direct contact with the valve during operation.
5.2 Flow Signal Output Readings
The flow signal output can be viewed on the panel meter, digital multimeter, or other display device used as shown in figure b-1.
When using the accessory SDPROC Command Module the flow rate will appear on the display at the front panel.The obser ved reading is a 0 to 100% indication (direct engineering units are optional). [If using a multichannel readout, be sure that the CHANNEL selector switch is set to the correct channel.]
Analog output flow signals of 0 to 5 VDC or optional 4 to 20 mA are attained at the appropriate pins the 25-pin "D" connector (see Figure b-2) on the side of the DMF transducer.The output flow signal is also a v ailab le at the DATA connector on the rear panel of the SDPROC Command Module.
Meter signal output is linearly proportional to the mass molecular flow rate of the gas being metered. The full scale range and gas for which your meter has been calibrated are shown on the flow transducer's front label.
For information on the RS485 or optional RS232 interfaces please contact your distributor or Dwyer.
5.3 Swamping Condition
If a flow of more than 10% above the maximum flow rate of the Mass Flow Controller is taking place, a condition known as "swamping" may occur.Readings of a "swamped" meter cannot be assumed to be either accurate or linear. Flow must be restored to below 110% of maximum meter range. Once flow rates are lowered to within calibrated range, the swamping condition will end. Operation of the meter above 110% of maximum calibrated flow may increase recovery time.
11
Page 15
5.4 Set Point Reference Signal
DMF flow controllers have a built-in solenoid valve and allow the user to set the flow to any desired flow rate within the range of the particular model installed.This valve is normally closed when no power is applied.
The set point input in analog mode responds to an analog 0 to 5 VDC reference voltage or 4-20mA reference current.This voltage is a linear representation of 0 to 100% of the full scale mass flow rate.Response time to set point changes are 1 second to within 2% of the final flow over 25 to 100% of full scale.
A variable 0 to 5VDC analog signal may be applied directly to the SET POINT and COMMON connections of the DMF transducer (see Figure b-1).
If a potentiometer is used to adjust the set point reference signal its value should be between 5K to 100K ohm and it should be capable of at least 10-turns or more for adjustment.
5.5 Valve OFF Control
It may, at times, be desirable to set the flow and maintain that setting while being able to turn the flow control valve off and on again.This can be accomplished via pin 5 on the 25-pin "D" connector.When 0 VDC (LO W) signal is applied (connection via a relay, switch or NPN open collector transistor is permissible), the solenoid valve is not powered and therefore will remain normally closed. Conversely, when the pin is disconnected from 0 VDC ("floating”) the solenoid v alv e will remain activ e .
The simplest means for utilizing the VALVE OFF control feature, is to connect a toggle switch between the COMMON and FORCE VALVE CLOSED pins of the DMF transducer.Toggling the switch on and off will allow for activating and deac­tivating the solenoid valve.
5.6 Valve Open /Purge
At times, it may be necessary to purge the flow system with a neutralizing gas such as pure dry nitrogen.The DMF transducer is capable of a full open condition for the solenoid valve, regardless of set point conditions. Connecting the FORCE VALVE OPEN pin (pin 4 on 25-pin "D" connector) to ground will fully open the valve. This connection can be made with a relay, switch or NPN open collector transistor. Conversely, when the pin is disconnected from 0 VDC ("floating”) the solenoid valve will remain active. (Note: in digital mode hardware I/O overrides software command)
The simplest means for utilizing the V ALVE OPEN control f eature, is to connect a tog­gle switch between the COMMON and FORCE VALVE OPEN pins of the DMF trans­ducer.Toggling the switch on will cause the v alv e to open fully and purge the system. Toggling the switch off will allow the solenoid valve to resume normal activity.
Caution: If the valve is left in the AUTO (control) or OPEN mode for an extended period of time, it may become warm or even hot to the touch.
Use care in avoiding direct contact with the valve during operation.
12
Page 16
5.7 Analog Interface Configuration
The DMF can be configured for the desired range and scaling by selection of analog board (see APPENDIX 1 on page 21) jumpers as follows:
0 to 5 V output: Jumper pins 2 and 3 of JP6.
Jumper pins 2 and 3 of JP3. Jumper pins 2 and 3 of JP5. Jumper pins 1 and 2 of JP12.
0 to 5 V input: Jumper pins 2 and 3 of JP2.
Jumper pins 2 and 3 of JP4. Jumper pins 1 and 2 of JP11.
0 to 10 V output: As for 0 to 5V, but jumper pins 2 and 3 of JP12. 4 to 20 mA output: Jumper pins 1 and 2 of JP6.
Jumper pins 1 and 2 of JP3. Jumper pins 1 and 2 of JP5. Jumper pins 1 and 2 of JP12.
4 to 20 mA input: Jumper pins 1 and 2 of JP2.
Jumper pins 1 and 2 of JP4. Jumper pins 1 and 2 of JP11.
By default the DMF is configured for analog input output ranges set to 0-5V (unless ordered with special configuration).
6. MAINTENANCE
6.1 Introduction
It is important that the Mass Flow Controller is used with clean, filtered gases only . Liquids may not be metered. Since the RTD sensor consists, in part, of a small capillary stainless steel tube, it is prone to occlusion due to impediments or gas crystallization. Other flow passages are also easily obstructed. Therefore, great care must be exercised to avoid the introduction of any potential flow impediment. To protect the instrument a 50 micron (DMF41401-41411) or 60 micron (DMF41431-41842) filter is built into the inlet of the flow transducer. The filter screen and the flow paths may require occasional cleaning as described below. There is no other recommended maintenance required. It is good practice, how­ever, to keep the meter away from vibration, hot or corrosive environments and excessive RF or magnetic interference.
If periodic calibrations are required they should be performed by qualified per­sonnel and calibrating instruments, as described in section (7). It is recommend­ed that units are returned to Dwyer for repair service and calibration.
CAUTION: TO PROTECT SERVICING PERSONNEL IT IS MANDATORY THAT ANY INSTRUMENT BEING SERVICED IS COMPLETELY
PURGED AND NEUTRALIZED OF TOXIC, BACTERIOLOGICALLY INFECTED, CORROSIVE OR RADIOACTIVE CONTENTS.
13
Page 17
6.2 Flow Path Cleaning
Before attempting any disassembly of the unit for cleaning, try inspecting the flow paths by looking into the inlet and outlet ends of the meter for any debris that ma y be clogging the flow through the meter. Remove debris as necessary. If the flow path is not unclogged, then proceed with steps below.
Do not attempt to disassemble the sensor. If blockage of the sensor tube is not alleviated by flushing through with cleaning fluids, please return meter to Dwyer for servicing.
6.2.1 Restrictor Flow Element (RFE)
The Restrictor Flow Element (RFE) is a precision flow divider inside the trans­ducer, which splits the inlet gas flow by a preset amount to the sensor and main flow paths.The particular RFE used in a given Mass Flow Controller depends on the gas and flow range of the instrument.
6.2.2 DMF 41401-41411 models
Unscrew the inlet compression fitting of meter.Note that the Restrictor Flow Element (RFE) is connected to the inlet fitting.
Carefully disassemble the RFE from the inlet connection. The 50 micron filter screen will now become visible.Push the screen out through the inlet fitting.Clean or replace each of the removed parts as necessary. If alcohol is used for clean­ing, allow time for drying.
Inspect the flow path inside the transducer for any visible signs of contaminants. If necessary, flush the flow path through with alcohol. Thoroughly dr y the flow paths by flowing clean dry gas through.
Carefully re-install the RFE and inlet fitting, avoiding any twisting and deforming the RFE. Be sure that no dust has collected on the O-ring seal.
Note: Overtightening will deform and render the RFE defective.
It is advisable that at least one calibration point be checked after re installing the inlet fitting - see section (7).
6.2.3 DMF 41431-41842 models
Unscrew the four socket head cap screws (two 10-24 and two 6-32) at the inlet side of the meter.This will release the short square block containing the inlet com­pression fitting.
The 60 micron filter screen will now become visible.Remove the screen.DO NOT remove the RFE inside the flow transducer! Clean or replace each of the remo v ed parts as necessary. If alcohol is used for cleaning, allow time for drying.
14
Page 18
Inspect the flow path inside the transducer for any visible signs of contaminants. If necessary, flush the flow path through with alcohol. Thoroughly dr y the flow paths by flowing clean dry gas through.
Re-install the inlet parts and filter screen. Be sure that no dust has collected on the O-ring seal.
It is advisable that at least one calibration point be checked after re installing the inlet fitting - see section (7).
6.2.4 Valve Maintenance (DMF)
The solenoid valve consists of 316 and 416 stainless steel, and VITON7 (or optional NEOPRENE7 or KALREZ7) O-rings and seals. No regular maintenance is required except for periodic cleaning.
Various corrosive gases may demand more frequent replacement of VITON7 O- rings and seals inside the valve.Be sure to use an elastomer material, appro­priate for your specific gas application. Contact your distributor or Dwyer for optional sealing materials available.
Set the DMF into PURGE mode, and attempt to flush through with a clean, fil­tered, and neutral gas such as nitrogen.[Another option for fully opening the valve is to remove the plastic cap on top of the valve, and turn the set screw counter­clockwise until it stops. Set valve for the closed position. Apply an inlet pressure of 5 psig and atmospheric pressure at the outlet. If a small flow occurs, turn the set screw on top of the solenoid valve clockwise until the flow through the DMF just stops.
Note: Removal of the factory installed calibration seals and/or any adjustments made to the meter, as described in this section, will void
any calibration warranty applicable.
7. CALIBRATION PROCEDURES
7.1 Flow Calibration
Dwyer Instruments' Flow Calibration Laboratory offers professional calibration support for Mass Flow Meters and Controllers, using precision calibrators under strictly controlled conditions. NIST traceable calibrations are available. Calibrations can also be performed at customers' site using available standards.
Factory calibrations are performed using NIST traceable precision volumetric cal­ibrators incorporating liquid sealed frictionless actuators.
Generally, calibrations are performed using dry nitrogen gas. The calibration can then be corrected to the appropriate gas desired based on relative correction [K] factors shown in the gas fa ctor table see Appendix 2.A reference gas, other than nitrogen, may be used to closer approximate the flow characteristics of cer tain gases. This practice is recommended when a reference gas is found with ther­modynamic properties similar to the actual gas under consideration. The appro-
15
Page 19
priate relative correction factor should be recalculated see section (9). It is standard practice to calibrate Mass Flow Meters/Controllers with dry nitrogen
gas at 70FF (21.1EC), 20 psig (1.4 bars) [25 psig (1.7 bars) for DMF 41842] inlet pressure and 0 psig (0 bar) outlet pressure. It is best to calibrate the DMF trans­ducers to actual operating conditions. Specific gas calibrations of non-toxic and non-corrosive gases are available at specific conditions. Please contact your dis­tributor or Dwyer for a price quotation.
It is recommended that a flow calibrator of at least four times better collectiv e accu­racy than that of the Mass Flow Controller to be calibrated be used. Equipment required for calibration includes a flow calibration standard and a certified high sensitivity multimeter (which together have a collective accuracy of + ter), an insulated (plastic) screwdriver, a flow regulator (example: metering needle valve) installed upstream from the Mass Flow Controller and a pressure regulated source of dry filtered nitrogen gas (or other suitable reference gas).
The gas and ambient temperature, as well as inlet and outlet pressure conditions should be set up in accordance with actual operating conditions.
7.2 Calibration of DMF Mass Flow Controllers
All adjustments to the DMF calibration and control loop tuning are accomplished using the RS485 (or optional RS232) interface in conjunction with setup and cal­ibration software available from Dwyer. The sensor zero is automatically adjust­ed internally whenever the control valve is fully closed (set point less than 2% of full scale) and the unit is warmed up.
0.25% or bet-
DMF Mass Flow Meters may be field recalibrated/checked using the setup and calibration program for the same range they were originally factory calibrated for. Flow range changes may require a different Restrictor Flow Element (RFE). Additionally, a different Solenoid Valve Orifice for the DMF Mass Flow Controller (see Table VI) may also be required. Consult your distributor or Dwyer for more information.
TABLE VI DMF SOLENOID VALVE ORIFICE SELECTION TABLE
ORIFICE PART NUMBER FLOW RATE [N2]
OR.010
OR.020
OR.040 1 to 5 slpm
OR.055 5 to 10 slpm
OR.063 10 to 15 slpm
OR.073 15 to 20 slpm
OR.094 20 to 50 slpm
OR.125 50 to 100 slpm
Under 10 sccm
10 to 1000 sccm
16
Page 20
8. TROUBLESHOOTING
8.1 Common Conditions
Your Mass Flow Controller was thoroughly checked at numerous quality control points during and after manufacturing and assembly operations.It was calibrat­ed in accordance to your desired flow and pressure conditions for a given gas or a mixture of gases.
It was carefully packed to prevent damage during shipment. Should you feel that the instrument is not functioning properly please check for the following common conditions first:
Are all cables connected correctly? Are there any leaks in the installation? Is the power supply correctly selected according to requirements? When several
meters are used a power supply with appropriate current rating should be selected. Were the connector pinouts matched properly? When interchanging with other
manufacturers' equipment, cables and connectors must be carefully wired for cor­rect pin configurations.
Is the pressure differential across the instrument sufficient? For best results it is recommended that instruments are returned to the factory
for servicing. See section 1.3 for return procedures.
8.2 Technical Assistance
Dwyer Instruments will provide technical assistance over the phone to qualified repair personnel. Please call our Technical Assistance at (219) 879-8000. Please have your Serial Number and Model Number ready when you call.
17
Page 21
8.3 Troubleshooting Guide
Indication
lack of reading or output
output reads at (+) or (-) saturation only
Likely Reason
power supply off
fuse blown (DMF)
fuse blown (SDPROC)
filter screen obstructed at inlet
occluded sensor tube
pc board defect
valve adjustment wrong
fuse blown (DMF)
Remedy
check connection of power supply
disconnect DMF transducer from power supply; remove the shorting condition or check polarities; fuse resets automatically
disconnect power cord from AC sup­ply; remove and inspect fuses at AC power input connector of SDPROC; replace as necessary
REMOVE CAUSE OF SHORT CIRCUIT!
flush clean or disassemble to remove impediments or replace
flush clean or disassemble to remove impediments or return to factory for replacement
return to factory for replacement
re-adjust valve (section 6.2.4)
disconnect DMF transducer from power supply; remove the shorting condition or check polarities; fuse resets automatically
flow reading does not coincide with the set point (DMF models only)
no response to set point (DMF models only)
unstable or no zero reading
inadequate gas pressure
filter screen obstructed at inlet
ground loop
inadequate gas pressure
cable or connector malfunction
set point is too low (<2% of full scale)
valve adjustment wrong
gas leak
pc board defective
18
REMOVE CAUSE OF SHORT CIRCUIT!
apply appropriate gas pressure
flush clean or disassemble to remove impediments or replace
signal and power supply commons are different
apply appropriate gas pressure
check cables and all connections or replace
re-adjust set point
re-adjust valve (section 6.2.4)
locate and correct
return to factory for replacement
Page 22
Indication
full scale output at "no flow" condition or with valve closed
Likely Reason
defective sensor
gas Leak
Remedy
return to factory for replacement
locate and repair
calibration off
DMF valve does not work in open position
gas metered is not the same as what meter was calibrated for
composition of gas changed
gas leak
pc board defective
RFE dirty
occluded sensor tube
filter screen obstructed at inlet
transducer is not mounted properly
incorrect valve adjustment
pc board defect
cable or connectors malfunction
use matched calibration
see K factor tables in APPENDIX 2
locate and correct
return to factory for replacement
flush clean or disassemble to remove impediments
flush clean or disassemble to remove impediments or return to factory for replacement
flush clean or disassemble to remove impediments or replace
check for any tilt or change in the mounting of the transducer; generally, units are calibrated for horizontal installation (relative to the sensor tube)
re-adjust valve (section 6.2.4)
return to factory for replacement
check cable and connectors or replace
DMF valve does not work in close position
differential pressure too high
insufficient inlet pressure
incorrect valve adjustment
pc board defect
cable or connectors malfunction
orifice obstructed
19
decrease pressure to correct level
adjust appropriately
re-adjust valve (section 6.2.4)
return to factory for replacement
check cable and connectors or replace
disassemble to remove impediments or return to factory
Page 23
9. CALIBRATION CONVERSIONS FROM REFERENCE
F
GASES
The calibration conversion incorporates the K factor.The K factor is derived from gas density and coefficient of specific heat. For diatomic gases:
gas
d X C
p
1
=
K
where d = gas density (gram/liter) C
= coefficient of specific heat (cal/gram)
p
Note in the above relationship that d and Cp are usually chosen at standard con­ditions of one atmosphere and 25
C.
If the flow range of a Mass Flow Controller or Controller remains unchanged, a relative K factor is used to relate the calibration of the actual gas to the reference gas.
Q
a
=
K
Q
r
K
a
=
K
r
where Qa= mass flow rate of an actual gas (sccm) Q
r
K
a
K
r
= mass flow rate of a reference gas (sccm) = K factor of an actual gas = K factor of a reference gas
For example, if we want to know the flow rate of oxygen and wish to calibrate with nitrogen at 1000 SCCM, the flow rate of oxygen is:
Q
= Qa= Qrx K = 1000 X 0.9926 = 992.6 sccm
O
2
where K = relative K factor to reference gas (oxygen to nitrogen)
20
Page 24
APPENDIX 1
COMPONENTS DIAGRAMS
DMF Digital PC Board
(Primary Side)
21
Page 25
APPENDIX 1
(CONTINUED)
DMF Digital PC Board
(Secondary Side)
22
Page 26
APPENDIX 1
(CONTINUED)
DMF Analog PC Board
(Primary Side)
23
Page 27
APPENDIX 1
(CONTINUED)
DMF Analog PC Board
(Secondary Side)
24
Page 28
APPENDIX 2
GAS FACTOR TABLES (“K” FACTORS)
Actual Gas
AcetyleneC
2H2
Air Allene (Propadiene) C Ammonia NH
3H4
3
Argon Ar Arsine AsH Boron Trichloride BCl Boron Trifluoride BF Bromine Br Boron Tribromide Br Bromine Pentafluoride BrF Bromine Trifluoride BrF
3
3
3
2
3
5
3
Bromotrifluoromethane (Freon-13 B1) CBrF 1,3-Butadiene C4H Butane C4H 1-Butane C4H
6
10
8
2-Butane C4H8CIS 2-Butane C Carbon Dioxide CO Carbon Disulfide CS
4H8
TRANS
2
2
Carbon Monoxide C0 Carbon Tetrachloride CCl Carbon Tetrafluoride (Freon-14)CF Carbonyl Fluoride COF
4
4
2
Carbonyl Sulfide COS Chlorine Cl Chlorine Trifluoride ClF
2
3
Chlorodifluoromethane (Freon-22)CHClF Chloroform CHCl
3
Chloropentafluoroethane(Freon-115)C2ClF Chlorotrifluromethane (Freon-13) CClF CyanogenC2N
2
3
CyanogenChloride CICN Cyclopropane C Deuterium D Diborane B2H
3H5
2
6
3
2
5
K Factor
Relative to N
.5829
1.0000 .4346 .7310
1.4573 .6735 .4089 .5082 .8083 .38 .26 .3855 .3697 .3224 .2631 .2994 .324 .291 .7382 .6026
1.00 .31 .42 .5428 .6606 .86 .4016 .4589 .3912 .2418 .3834 .61 .6130 .4584
1.00 .4357
Cp
[Cal/g]
2
.4036 .240 .352 .492 .1244 .1167 .1279 .1778 .0539 .0647 .1369 .1161 .1113 .3514 .4007 .3648 .336 .374 .2016 .1428 .2488 .1655 .1654 .1710 .1651 .114 .1650 .1544 .1309 .164 .153 .2613 .1739 .3177
1.722 .508
Density
[g/I]
1.162
1.293
1.787 .760
1.782
3.478
5.227
3.025
7.130
11.18
7.803
6.108
6.644
2.413
2.593
2.503
2.503
2.503
1.964
3.397
1.250
6.860
3.926
2.945
2.680
3.163
4.125
3.858
5.326
6.892
4.660
2.322
2.742
1.877
1.799
1.235
25
Page 29
Actual Gas
K Factor
Relative to N
Cp
[Cal/g]
2
Density
[g/I]
Dibromodifluoromethane CBr
2F2
Dichlorodifluoromethane (Freon-12) CCl2F Dichlofluoromethane (Freon-21) CHCl2F Dichloromethylsilane (CH Dichlorosilane SiH2Cl
SiCl
3)2
2
2
Dichlorotetrafluoroethane (Freon-114) C2Cl2F 1,1-Difluoroethylene (Freon-1132A) C2H2F Dimethylamine (CH3)2NH Dimethyl Ether (CH 2,2-Dimethylpropane C Ethane C2H
6
3)2
O
3H12
Ethanol C2H6O Ethyl Acetylene C
4H6
Ethyl Chloride C2H5Cl Ethylene C
2H4
Ethylene Oxide C2H4O Fluorine F Fluoroform (Freon-23) CHF
2
3
Freon-11 CCl3F Freon-12 CCl Freon-13 CClF Freon-13B1 CBrF Freon-14 CF
2F2
3
3
4
Freon-21 CHCl2F Freon-22 CHClF Freon-113 CCl2FCClF Freon-114 C2Cl2F Freon-115 C2ClF Freon-C318 C4F Germane GeH Germanium Tetrachloride GeCl
2
2
4
5
8
4
4
Helium He Hexafluoroethane C Hexane C Hydrogen H
6H14
2
(Freon-116)
2F6
Hydrogen Bromide HBr Hydrogen Chloride HCl Hydrogen Cyanide HCN
.1947
2
.3538 .4252 .2522 .4044 .2235
4
.4271
2
.3714 .3896 .2170 .50 .3918 .3225 .3891 .60 .5191 .9784 .4967 .3287 .3538 .3834 .3697 .4210 .4252 .4589 .2031 .2240 .2418 .1760 .5696 .2668
1.454 .2421 .1792
1.0106
1.000
1.000
1.070
.15 .1432 .140 .1882 .150 .1604 .224 .366 .3414 .3914 .420 .3395 .3513 .244 .365 .268 .1873 .176 .1357 .1432 .153 .1113 .1654 .140 .1544 .161 .160 .164 .185 .1404 .1071
1.241 .1834 .3968
3.419 .0861 .1912 .3171
9.362
5.395
4.592
5.758
4.506
7.626
2.857
2.011
2.055
3.219
1.342
2.055
2.413
2.879
1.251
1.965
1.695
3.127
6.129
5.395
4.660
6.644
3.926
4.592
3.858
8.360
7.626
6.892
8.397
3.418
9.565 .1786
6.157
3.845 .0899
3.610
1.627
1.206
26
Page 30
Actual Gas
Hydrogen Fluoride HF Hydrogen Iodide HI Hydrogen Selenide H Hydrogen Sulfide H Iodine Pentafluoride IF Isobutane CH(CH3) Isobutylene C4H
Se
2
S
2
5
3
6
Krypton Kr Methane CH Methanol CH Methyl Acetylene C3H
4
3
4
Methyl Bromide CH2Br Methyl Chloride CH Methyl Fluoride CH Methyl Mercaptan CH Methyl Trichlorosilane (CH Molybdenum Hexafluoride MoF Monoethylamine C2H5NH Monomethylamine CH3NH
Cl
3
F
3
SH
3
)SiCl
3
3
6
2
2
Neon NE Nitric Oxide NO Nitrogen N Nitrogen Dioxide NO Nitrogen Trifluoride NF
2
2
3
Nitrosyl Chloride NOCl Nitrous Oxide N
O
2
Octafluorocyclobutane (Freon-C318) C Oxygen O Oxygen Difluoride OF
2
2
Ozone Pentaborane B Pentane C5H
5H9
12
Perchloryl Fluoride ClO3F Perfluoropropane C Phosgene COCl Phosphine PH Phosphorous Oxychloride POCl Phosphorous Pentafluoride PH
3F8
2
3
3
5
4F8
K Factor
Relative to N
.9998 .9987
.7893
.80
.2492
.27 .2951
1.453 .7175 .5843 .4313 .5835 .6299 .68 .5180 .2499 .2126 .3512 .51
1.46 .990
1.000 .737 .4802 .6134 .7128 .176 .9926 .6337 .446 .2554 .2134 .3950 .174 .4438
1.070 .36 .3021
Cp
[Cal/g]
2
.3479 .0545
.1025
.2397 .1108 .3872 .3701 .0593 .5328 .3274 .3547 .1106 .1926 .3221 .2459 .164 .1373 .387 .4343 .246 .2328 .2485 .1933 .1797 .1632 .2088 .185 .2193 .1917 .195 .38 .398 .1514 .197 .1394 .2374 .1324 .1610
Density
[g/I]
.893
5.707
3.613
1.520
9.90
3.593
2.503
3.739 .715
1.429
1.787
4.236
2.253
1.518
2.146
6.669
9.366
2.011
1.386 .900
1.339
1.25
2.052
3.168
2.920
1.964
8.397
1.427
2.406
2.144
2.816
3.219
4.571
8.388
4.418
1.517
6.843
5.620
27
Page 31
Actual Gas
Phosphorous Trichloride PCl Propane C3H Propylene C3H Silane SiH
8
6
4
Silicon Tetrachloride SiCl Silicon Tetrafluoride SiF Sulfur Dioxide SO
2
Sulfur Hexafluoride SF Sulfuryl Fluoride SO2F
3
4
4
6
2
Tetrafluoroethane (Forane 134A) CF3CH2F Tetrafluorohydrazine N
2F4
Trichlorofluoromethane (Freon-11) CCl3F Trichlorosilane SiHCl
3
1,1,2-Trichloro-1,2,2 Trifluoroethane (Freon-113) CCl
FCClF
2
2
Triisobutyl Aluminum (C4H9)AL Titanium Tetrachloride TiCl Trichloro Ethylene C2HCl
4
3
Trimethylamine (CH3)3N Tungsten Hexafluoride WF Uranium Hexafluoride UF
6
6
Vinyl Bromide CH2CHBr Vinyl Chloride CH
CHCl
2
Xenon Xe
K Factor
Relative to N
.30 .35 .40 .5982 .284 .3482 .69 .2635 .3883 .5096 .3237 .3287 .3278
.2031
.0608 .2691 .32 .2792 .2541 .1961 .4616 .48
1.44
Cp
[Cal/g]
2
.1250 .399 .366 .3189 .1270 .1691 .1488 .1592 .1543 .127 .182 .1357 .1380
.161
.508 .120 .163 .3710 .0810 .0888 .1241 .12054 .0378
Density
[g/I]
6.127
1.967
1.877
1.433
7.580
4.643
2.858
6.516
4.562
4.224
4.64
6.129
6.043
8.36
8.848
8.465
5.95
2.639
13.28
15.70
4.772
2.788
5.858
28
Page 32
APPENDIX 3
DIMENSIONAL DRAWINGS
DMF 41401-41411 Mass Flow Controller
NOTES: Dwyer reserves the right to change designs and dimensions at its sole discretion at any time without notice. For certified dimensions please contact Dwyer.
29
Page 33
DMF 41431-41842 Mass Flow Controller
NOTES: Dwyer reserves the right to change designs and dimensions at its sole discretion at any time without notice. For certified dimensions please contact Dwyer.
30
Page 34
APPENDIX 4
SENDING COMMANDS TO THE DMF
RS485
The standard DMF comes with an RS485 interface. The protocol described below allows for the unit using either a custom software program or a “dumb terminal”. All values are sent as printable ASCII characters. The start character is always !and the command string is terminated with a carriage return (line feeds are automatically stripped out by the DMF:
!<Addr>, <Cmd>,Arg1,Arg2,Arg3,Arg4<CR>
WHERE:
! Start character
Addr RS485 device address in the ASCII representation of
hexadecimal (00 through FF are valid).**
Cmd The one or two character command from the table above.
Arg1 to Arg4 The command arguments from the table above.
Multiple arguments are comma delimited.
CR Carriage return character.
** Default address for all units is 11.
Several examples of commands follow. All assume that the DMF has been configured for address 15 (0F hex) on the RS485 bus:
1. To put the unit in digital mode: !0F,M,D<CR> The DMF will reply: !0FMD<CR>
2. To set the flow of 50% of FS: !0F,S,50.0<CR> The DMF will reply: !0FS50.0<CR>
3. To get a flow reading: !0F,F<CR> The DMF will reply: !0F50.0<CR>
(Assuming the flow is at 50% FS)
4. Set the high alarm limit to 5% above Set point: !0F,A,H,5.0<CR> The DMF will reply: !0FA5.0<CR>
31 32
Page 35
Page 36
333435
Page 37
Page 38
Page 39
36
Page 40
APPENDIX 5
CALIBRATION TABLE: GAS DEPENDENT VARIABLES
INDEX
0 BlankSDPROC char[10] Do not modify. For internal use only.
1 SerialNumber char[20]
2 ModelNumber char[20]
3 SoftwareVer char[10]
4 TimeSinceCalHr float Time since last calibration in hours.
5 Options uint Misc. Options.
6 AOutOffset_mA int
7 AddressRS485 char[3] Two character address for RS485 only.
8 AInScaleV float
9 AInOffsetV float
10 AInScale_mA float
11 AInOffset_mA float
12 AoutScaleV float
13 AoutScale_mA float
14 SensorZero uint
15 Klag[0] float
16 Klag[1] float
17 Klag[2] float
18 Klag[3] float
19 Klag[4] float
20 Klag[5] float
21 Reserved float
22 Reserved float
23 Reserved float
24 Reserved float
25 Reserved float
26 Reserved float
27 Kgain[0] float
28 Kgain[1] float
NAME
DATA TYPE
NOTES
37
Page 41
INDEX NAME
29 Kgain[2] float
30 Kgain[3] float
31 Kgain[4] float
32 Kgain[5] float
33 Reserved float
34 Reserved float
35 Reserved float
36 Reserved float
37 Reserved float
38 Reserved float
39 ValveTbl[0][open] float Index 0: Valve actuation. Must be 0.0.- Do Not Alter
40 ValveTbl[0][valve value] uint Index 0: Valve: D/A value - Do Not Alter
41 ValveTbl[1][flow] float Index 1: Actual valve opening in % FS. Do Not Alter
42 ValveTbl[1][valve value] uint Index 1: Valve D/A counts corresponding to flow. Do Not Alter 43 ValveTbl[2][flow] float Do Not Alter
44 ValveTbl[2][valve value] uint Do Not Alter
45 ValveTbl[3][flow] float Do Not Alter
46 ValveTbl[3][valve value] uint Do Not Alter
47 ValveTbl[4][flow] float Do Not Alter
48 ValveTbl[4][valve value] uint Do Not Alter
49 ValveTbl[5][flow] float Do Not Alter
50 ValveTbl[5][valve value] uint Do Not Alter
51 ValveTbl[6][flow] float Do Not Alter
52 ValveTbl[6][valve value] uint Do Not Alter
53 ValveTbl[7][flow] float Do Not Alter
54 ValveTbl[7][valve value] uint Do Not Alter
55 ValveTbl[8][flow] float Do Not Alter
56 ValveTbl[8][valve value] uint Do Not Alter
57 ValveTbl[9][flow] float Index 9: Valve fully open. Must be 1.0- Do Not Alter
58 ValveTbl[9][valve value] uint Index 9: D/A count for a fully open valve. Must be 4095.- Do Not Alter
59 AutoTune Time Constant uint Do Not Alter
DATA TYPE
NOTES
38
Page 42
CALIBRATION TABLE: GAS INDEPENDENT VARIABLES
INDEX NAME
100 GasIdentifer char[27]
101 FullScaleRange float
102 StdTemp float
103 StdPressure float
104 StdDensity float
105 CalibrationGas char[27]
106 CalibratedBy char[20]
107 CalibratedAt char[20]
108 DateCalibrated char[10]
109 DateCalibrationDue char[10]
110 PID_Kp float
111 PID_Ki float
112 PID_Kd float
113 SensorTbl[0][Sensor Value] uint Index 0: Must be 120 (zero value)
114 SensorTbl[0][Flow] float Index 0: Must be 0.0 (zero PFS)
115 SensorTbl[1][Sensor Value] uint A/D value from sensor.
DATA TYPE
NOTES
116 SensorTbl[1][Flow] float Actual Flow in PFS.
117 SensorTbl[2][Sensor Value] uint
118 SensorTbl[2][Flow] float
119 SensorTbl[3][Sensor Value] uint
120 SensorTbl[3][Flow] float
121 SensorTbl[4][Sensor Value] uint
122 SensorTbl[4][Flow] float
123 SensorTbl[5][Sensor Value] uint
124 SensorTbl[5][Flow] float
125 SensorTbl[6][Sensor Value] uint
126 SensorTbl[6][Flow] float
127 SensorTbl[7][Sensor Value] unit
128 SensorTbl[7][Flow] float
39
Page 43
INDEX NAME
129 SensorTbl[8][Sensor Value] uint
130 SensorTbl[8][Flow] float
131 SensorTbl[9][Sensor Value] uint
132 SensorTbl[9][Flow] float
133 SensorTbl[10][Sensor Value] uint
134 SensorTbl[10][Flow] float Flow in PFS. Should be 1.0
135
136
137
DATA TYPE
Note:Values will be available for selected gas only.
NOTES
40
Page 44
APPENDIX 6
WARRANTY
Dwyer Mass Flow Systems are warranted against parts and workmanship for a period of one year from the date of purchase. Calibrations are warranted for up to six months after date of purchase, provided calibration seals have not been tampered with. It is assumed that equipment selected by the cus­tomer is constructed of materials compatible with gases used. Proper selec­tion is the responsibility of the customer. It is understood that gases under pressure present inherent hazards to the user and to equipment, and it is deemed the responsibility of the customer that only operators with basic knowledge of the equipment and its limitations are permitted to control and operate the equipment covered by this warranty. Anything to the contrary will automatically void the liability of Dwyer and the provisions of this warranty. Defective products will be repaired or replaced solely at the discretion of Dwyer at no charge. Shipping charges are borne by the customer. This war­ranty is void if the equipment is damaged by accident or misuse, or has been repaired or modified by anyone other than Dwyer or factory authorized serv­ice facility. This warranty defines the obligation of Dwyer and no other war­ranties expressed or implied are recognized.
NOTE: Follow Return Procedures In Section 1.3.
TRADEMARKS
Dwyer7 is a registered trademark of Dwyer Instruments. Buna7 is a registered trademark of DuPont Dow Elastometers. Kalrez7 is a registered trademark of DuPont Dow Elastomers.
Neoprene7 is a registered trademark of DuPont. Viton7 is a registered trademark of Dupont Dow Elastometers L.L.C.
41
Loading...