CAUTION: DEVICE MUST BE USED ONLY IN A MANNER CONSISTENT
WITH THIS MANUAL.
NOTICE: WHEN 230VAC SUPPLY USED, DWYER SPECIFIES USE OF
AC SUPPLY PART # T35-9-100R-3 MANUFACTURED BY ENG
ELECTRIC. THIS SUPPLY IS AVAILABLE FROM DWYER.
1.1 Model information
• Model: DLI-R
• Description: Single channel RTD logger with LCD display
• Manufactured in the USA
1.2 Specifi cations
External Probe
Range: -328 to 1562ºF (-200 to +850ºC)
Resolution: 0.01ºF (0.01ºC)
Calibrated accuracy (case between 59 to 95ºF (15 to 35ºC))
• ±0.05ºF (±0.05ºC) Probe between -328 to 500ºF (-200 to 260ºC)
• ±0.1ºF (±0.1ºC) Probe between 500 to 1562ºF (260 to 850ºC)
Accuracy de-rating:
• 0.05ºF (0.05ºC) per ºF (ºC) outside 59 to 95ºF (15 to 35ºC ) case
temperature, probe between -328 to 500 (-200 to 260ºC)
• 0.1ºF (0.1ºC) per ºF (ºC) outside 59 to 95ºF (15 to 35ºC) case
temperature probe between 500 to 1562ºF (260 to 850ºC)
Dot-Matrix LCDDimensions: 2.5” x 1.375” (63mm x 35mm)
Text: Confi gurable channel text size
Indicators: Power, status, memory
Backlight: Confi gurable w/auto shut-off and contrast adjustment
Start/Stop Time:
tart/Stop Time
Software programmable start time and date, up to six months in
Software programmable start time and date, up to six months in
advance; programmable stop time
advance; programmable stop tim
Page 3
Dwyer DLI-R
Memory:
174,762 total readings; software confi gurable memory wrap
Reading Rate: 1 reading every 2 seconds to 1 every 24 hours
Calibration:
Digital calibration through software
Calibration Date:
Automatically recorded within device
Battery Type:
9V lithium battery included, user replaceable; optional AC adapter,
7-24 VDC, 100mA. For 230 VAC operation use AC supply
T35-9-100 R-3 made by ENG Electric (Available from Dwyer).
Battery Life:
30 days typical @ 1 min reading rate with continuous LCD and no
backlight usage.
Data Format:
Date and time stamped ºC, ºF, ºK, ºR
Time Accuracy:
±1 minute/month at 68 to 86°F (20 to 30°C)
Computer Interface:
PC serial or USB (interface cable required); 115,200 baud
Software:
Windows 95/98/ME/NT/2000/XP/Vista based software
Operating Environment:
-4 to 140ºF (-20 to +60ºC), 0 to 95%RH non-condensing
Dimensions:
4.8” x 3.3” x 1.25” (122mm x 84mm x 32mm)
Weight:
16 oz (440 g)
Enclosure:
Black anodized aluminum
Page 4
Data Logger Manual
1.3 Warranty
Products manufactured by Dwyer Instruments, Inc., are warranted against
defective material and workmanship for a period of one year, starting from the
date of shipment. In the event that a Dwyer product is found to be defective,
Dwyer will repair or replace the product at its sole discretion. Such repair or
replacement shall be the sole remedy of this warranty.
This warranty extends only to the original purchasing customer and does
not apply to any unit, which in our sole judgment, has been subjected to:
a) Operating or environmental conditions in excess of our written
specifi cations or recommendations;
b) Damage, misuse or neglect;
c) Improper installation, repair or alteration.
This warranty excludes batteries.
Except as to title, this is our only warranty for the products. Dwyer
Instruments, Inc. expressly disclaims all other warranties, guarantees or
remedies—whether expressed or implied or statutory—including any implied
warranty of merchantability or fi tness for a particular purpose. We also disclaim
any implied warranty arising out of trade usage or out of a course dealing or
course of performance. We do not guarantee the integrity of data or warranty
that the products will operate uninterrupted or error-free.
Dwyer data loggers and their associated software have been thoroughly
tested and the documentation reviewed. However, Dwyer does not warrant
the performance of its products, or that the products or their associated
software will operate as described in this manual.
Please refer to page 34 for Dwyer’s complete Terms and Conditions.
Page 5
Dwyer DLI-R
Section 2: Device Overview
LED
Indicators
Power Jack
LCD Screen
Black
anodized
aluminum
enclosure
Function Keys
Communication
Jack
Up Key
Okay Key
Down Key
Cancel Key
Page 6
Data Logger Manual
Section 3: Important Status Icons
3.1 Battery Status
Full Empty
Battery Battery
Battery status icons indicate the state of charge of the batteries.
When the battery empty indicator is displayed, the battery should be replaced.
3.2 Memory Status
Memory Memory
Empty Full
The memory status icons indicate the amount of memory left for data storage.
If the DLI-R is in wraparound mode, the memory status icon will always indicate
empty.
3.3 Running Indicator
Indicates that the DLI-R is taking readings. The icons cycle periodically.
Page 7
Dwyer DLI-R
3.4 Delay Start Indicator
The delay start icon indicates that a logging operation is
scheduled for the future.
3.5 Wait Icon
The wait icon indicates that the device is busy.
3.6 Stop Icon
Indicates that the DLI-R is not currently taking readings.
3.7 External Power Icon
Indicates that the DLI-R is powered by an external source.
3.8 Reset Icon
Indicates that the device has been reset.
Page 8
Data Logger Manual
Section 4: Front Panel Overview
Section 4: Front Panel Overview
4.1 Changing the display units
The DLI-R comes with factory default display units of °C for the external
probe channel. These units can be easily changed by pressing the F3 button
in the main screen and then selecting F1 for external probe. After selecting
F1, the available units can be scrolled through by either pressing the F1 key
repeatedly or using the UP and DOWN keys.
Button pressing chain:
Main Screen -> F3 -> F1 ->F1 key repeatedly or UP and DOWN
4.2 Changing the number, type, and size of channels viewed
By default the DLI-R displays recently measured values from the external
probe channel on its Main Screen with the readings displayed in a large size
font. These recently measured values can, however, be hidden or viewed on
a smaller or larger scale.
To change the view of recently measured values:
From the Main Screen, press the F4 key to enter the Setup Menu and from
this menu press the F1 key to enter the Display screen. On this screen, function
key 1 (F1) corresponds to the external probe. Pressing this function key will
cause the channel to scroll between “show” or “hide”. If “show” is selected
recently recorded values from the external probe will appear on the main
screen and if “hide” is displayed they will not.
Button pressing chain:
Main Screen -> F4 -> F1 -> F1
To change the size of displayed channels:
From the Main Screen, press the F4 key to enter the Setup Menu and from
this menu press the F1 key to enter the Display screen, then F4 to scroll to
the next screen. Here the F2 key will change the size of the channels viewed.
By pressing F2 repeatedly the size parameter will scroll between 3 sizes: small,
medium, and large:
Button pressing chain:
Main Screen -> F4 -> F1 -> F4 -> F2 repeatedly to scroll
Page 9
Dwyer DLI-R
4.3 Checking the memory Status
A status icon appears on all screens representing memory, but further
information including percent memory left and number of readings taken can
also be viewed. From the Main Screen press the F1 key to enter the Status
screens then press F2 to view memory status information.
Button pressing chain:
Main Screen -> F1 -> F2
4.4 Checking power status
A battery status and external power status (if available) icon appear on all
screens, but percent battery power remaining and external power presence as
well as battery type, current battery voltage, and current external voltage can
also be viewed. From the Main Screen press F4 to view the Device Confi guration
Menu, F2 to access the power options, then F4 twice to view the Power Status
screen, including battery power percent remaining and the presence of
external power. Battery type and battery voltage are also displayed, as well as
external power voltage (if connected).
Button pressing chain:
Main Screen -> F4 -> F2 -> F4 -> F4
4.5 Changing the Contrast
The DLI-R’s LCD screen contrast values can be changed in two ways. One
method is outlined in the Function Reference Guide. A faster, simpler way
involves simultaneously pressing the CANCEL and UP or DOWN button
repeatedly in any screen.
Button pressing chain:
CANCEL + UP repeatedly (to increase) or DOWN repeatedly (to decrease)
Page 10
Data Logger Manual
Section 5: DLI-R Function Reference
Section 5: DLI-R Function Reference
5.1 Main Screen
The main screen of the DLI-R features a real-time display of most recently
measured external probe channel data. At the bottom of the main screen are
tabs corresponding to each of the four function keys. These tabs are used to
access the four main function menus of the DLI-R: status, statistics, units, and
setup. The left side of the main screen and all subsequent screens of the device
is where important status information icons can be found (detailed in Section
3: Important Status Icons – page 7) including recording status, memory status,
busy status, external power status, and battery power status.
5.2 Status Menu
Pressing F1 on the main screen brings up the Status menu. The fi rst screen
that appears in the Status Menu is Run Parameters, but the Memory Status
and Time and Date screens can also be viewed by pressing the F2 and F3
keys respectively.
5.2.i Run Parameters
The Run Parameters screen displays important information regarding the
device’s current recording session. These parameters include the time and
date the recording session started (start time and start date), the time and
date the recording session will end (stop date and stop time) due to either
lack of memory or preprogramming in the Dwyer software. The rate at which
the DLI-R is recording (rate) is also displayed. The device’s current status
(either running or stopped) is the last parameter on the Run Parameters
screen.
5.2.ii Memory Status
The Memory Status screen is where all information regarding the DLI-R’s
memory. This screen displays the percent of memory space currently available
(memory left), the number of readings currently stored on the device
(readings), the maximum number of readings the device can record (max
readings), as well as information about the wrap feature displaying either
“disabled” or the number of wrap readings currently stored in memory.
5.2.iii Time
The time screen displays current time data including the current time and
date, time and date of last measured data, and current time zone.
Page 11