Dwyer DLI2 User Manual

DLI2
Data Logger Manual
Table of Contents
Section 1: Device Safety............................................................................3
1.1 Model information ...................................................................3
1.2 Specifi cations ...........................................................................3
1.3 Warranty ..................................................................................5
Section 2: Device Overview ......................................................................6
Section 3: Status Indicator Icons ..............................................................7
3.1 Battery Status ...........................................................................7
3.2 Memory Status .........................................................................7
3.3 Running Indicator .....................................................................7
3.4 Delay Start Indicator .................................................................8
3.5 Wait Icon ..................................................................................8
3.6 Stop Icon ..................................................................................8
3.7 External Power Icon ..................................................................8
3.8 Reset Icon .................................................................................8
Section 4: Front Panel Overview ..............................................................9
4.1 Changing the display units .......................................................9
4.2 Changing the number, type, and size of channels viewed .........9
4.3 Checking the memory status ..................................................10
4.4 Checking power status ...........................................................10
4.5 Changing the Contrast ...........................................................10
Data Logger Manual
Section 5: DLI2 Function Reference ........................................................11
5.1 Main Screen ...........................................................................11
5.2 Status Menu ...........................................................................11
5.3 Statistics Menu .......................................................................12
5.4 Units.......................................................................................13
5.5 Setup Menu ...........................................................................13
Section 6: DLI2 Screen Descriptions .......................................................18
6.1 Main Screen: ..........................................................................18
6.2 Status Screens (Run Parameters): ............................................18
6.3 Status Screens (Memory Status): .............................................19
6.4 Status Screens (Time): ............................................................19
6.5 ID Parameters .........................................................................20
6.6 Statistics Menu Screen: ...........................................................20
Revised 08/13/08
Page 1
Dwyer DLI2
6.7 Channel Statistics: ..................................................................21
6.8 Type Statistics: ........................................................................21
6.9 Statistics Information Screen: ..................................................22
6.10 Unit Selection Menu: ............................................................22
6.11 Device Confi guration Menu: .................................................23
6.12 Adjust Visibility: ....................................................................23
6.13 Display Confi guration: ..........................................................24
6.14 Power Modes Screen: ...........................................................25
6.15 Display Update Mode Screen: ...............................................25
6.16 Power Status Screen: ............................................................26
6.17 Device Information Screens (Minimum Device Range): .........26
6.18 Device Information Screens (Maximum Device Range): ........27
6.19 Device Information Screens (Device Version): .......................27
6.20 Device Information Screens (Firmware Version): ...................28
6.21 Calibration Information Screens (Calibration Date): ..............28
6.22 Calibration Information Screens (Pressure Calibration): .........29
6.23 Device Reset Screen (Hardware Reset): .................................29
6.24 Device Reset Screen (Power Interruption): ............................30
Section 7: Computer Interface ...............................................................31
Section 9: Maintenance ..........................................................................32
Section 10: Terms and Conditions .......................................................... 33
10.1 Limited Warranty: .................................................................33
10.2 Limitations:...........................................................................35
10.3 Software License: ..................................................................35
Page 2
Data Logger Manual
Section 1: Device Safety
CAUTION: DEVICE MUST BE USED ONLY IN A MANNER
CONSISTENT WITH THIS MANUAL.
NOTICE: WHEN 230VAC SUPPLY USED, DWYER SPECIFIES USE OF AC SUPPLY PART # T35-9-100R-3 MANUFACTURED BY ENG ELECTRIC. THIS SUPPLY IS AVAILABLE FROM DWYER.
1.1 Model information
• Model: DLI2
Description: Measure and record pressure with LCD display featuring real-time on screen graphing capabilities and a 1/4” NPT fi tting.
Manufactured in the USA
1.2 Specifi cations
Pressure
Range: 0 to 30, 100, 300 and 500 psia/g 0 to 1000 and 5000 psia
Resolution: 0.002, 0.005, 0.02, 0.05 psia/g
Calibrated Accuracy: 2% FSR, 0.25% @ 77°F (25°C) typical Dot-Matrix LCD
Dimensions: 2.5” x 1.375” (63mm x 35mm)
Text: Confi gurable channel text size
Indicators: Power, status, memory
Backlight: Confi gurable w/auto shut-off and contrast adjustment Start/Stop Time:
Software programmable start time and date, up to six months in advance; programmable stop time
Memory: 262,143 total readings; software confi gurable memory wrap
Reading Rate: 1 reading every 2 seconds to 1 every 24 hours
Calibration:
Digital calibration through software
Page 3
Dwyer DLI2
Calibration Date:
Automatically recorded within device
Battery Type:
6 alkaline AA batteries, user replaceable; optional AC adapter, 7-24 VDC, 100mA. For 230 VAC operation use AC supply T35-9-100 R-3 made by ENG Electric.
Battery Life:
60 days typical with continuous screen use; 360 days with display off.
Data Format:
Date and time stamped PSI, inHg, mmHg, bar, atm, Torr, Pa, kPa, MPa
Time Accuracy:
±1 minute/month at 68 to 86ºF (20 to 30°C)
Computer Interface:
PC serial or USB (interface cable required); 115,200 baud
Software:
Windows 95/98/ME/NT/2000/XP based software
Operating Environment:
68 to 140ºF (-20 to +60ºC), 0 to 95%RH non-condensing
Dimensions:
5.1” x 4.8” x 1.78” (130mm x 122mm x 45mm)
Weight:
40 oz (1134 g)
Enclosure:
Black anodized aluminum
Page 4
Data Logger Manual
1.3 Warranty
Products manufactured by Dwyer Instruments, Inc., are warranted against defective material and workmanship for a period of one year, starting from the date of shipment. In the event that a Dwyer product is found to be defective, Dwyer will repair or replace the product at its sole discretion. Such repair or replacement shall be the sole remedy of this warranty.
This warranty extends only to the original purchasing customer and does
not apply to any unit, which in our sole judgment, has been subjected to:
a) Operating or environmental conditions in excess of our written
specifi cations or recommendations;
b) Damage, misuse or neglect;
c) Improper installation, repair or alteration.
This warranty excludes batteries.
Except as to title, this is our only warranty for the products. Dwyer Instruments, Inc. expressly disclaims all other warranties, guarantees or remedies—whether expressed or implied or statutory—including any implied warranty of merchantability or fitness for a particular purpose. We also disclaim any implied warranty arising out of trade usage or out of a course dealing or course of performance. We do not guarantee the integrity of data or warranty that the products will operate uninterrupted or error-free.
Dwyer data loggers and their associated software have been thoroughly tested and the documentation reviewed. However, Dwyer does not warrant the performance of its products, or that the products or their associated software will operate as described in this manual.
Please refer to page 33 for Dwyer’s complete Terms and Conditions.
Page 5
Dwyer DLI2
Section 2: Device Overview
LED Indicators
LCD Screen
Black anodized aluminum enclosure
Pressure Sensor
Function Keys
Power Jack
Communication Jack
Up Key
Okay Key
Down Key
Cancel Key
Page 6
Data Logger Manual
Section 3: Important Status Icons
3.1 Battery Status
Full Empty Battery Battery
Battery status icons indicate the state of charge of the batteries. When the battery empty indicator is displayed, the battery should be replaced.
3.2 Memory Status
Memory Memory Empty Full
The memory status icons indicate the amount of memory left for data storage. If the DLI2 is in wraparound mode, the memory status icon will always indicate empty.
3.3 Running Indicator
Indicates that the DLI2 is taking readings. The icons cycle periodically.
Page 7
Dwyer DLI2
3.4 Delay Start Indicator
The delay start icon indicates that a logging operation is scheduled for the future.
3.5 Wait Icon
The wait icon indicates that the device is busy.
3.6 Stop Icon
Indicates that the DLI2 is not currently taking readings.
3.7 External Power Icon
Indicates that the DLI2 is powered by an external source.
3.8 Reset Icon
Indicates that the device has been reset.
Page 8
Data Logger Manual
Section 4: Front Panel Overview
Section 4: Front Panel Overview
4.1 Changing the display units
The DLI2 comes with factory default display units of PSI for pressure and the real-time pressure graphing feature. These units can be easily changed by pressing the F3 button in the main screen and then selecting F1 for pressure or F2 for the pressure graph. After selecting the channel, the available units can be scrolled through by either pressing the channel’s function key repeatedly or using the UP and DOWN keys.
Button pressing chain:
Main Screen -> F3 -> F1(pressure) or F2 (pressure graph) -> function key repeatedly or UP and DOWN
4.2 Changing the number, type, and size of channels viewed
By default the DLI2 displays recently measured values of the pressure channel and the real-time pressure graph on its Main Screen in a large size font. The channel and graph can, however, be hidden or viewed on a smaller or larger scale.
To change the number and type of displayed channels:
From the Main Screen, press the F4 key to enter the Setup Menu and from this menu press the F1 key to enter the Display screen. On this screen, F1 and F2 correspond to the Pressure channel and Pressure graph respectively. Pressing these function keys repeatedly will cause the channels to scroll between “show” or “hide” channels displaying “show” will appear on the main screen and channels displaying “hide” will not.
Button pressing chain:
Main Screen -> F4 -> F1 -> F1(pressure) or F2 (pressure graph)
To change the size of displayed channels:
From the Main Screen, press the F4 key to enter the Setup Menu and from this menu press the F1 key to enter the Display screen, then F4 to scroll to the next screen. Here the F2 key will change the size of the channels viewed. By pressing F2 repeatedly the size parameter will scroll between 3 sizes: small, medium, and large.
Button pressing chain:
Main Screen -> F4 -> F1 -> F4 -> F2 repeatedly to scroll
Page 9
Dwyer DLI2
4.3 Checking the memory status
A status icon appears on all screens representing memory, but further information including percent memory left and number of readings taken can also be viewed. From the Main Screen press the F1 key to enter the Status screens then press F2 to view memory status information.
Button pressing chain:
Main Screen -> F1 -> F2
4.4 Checking power status
A battery status and external power status (if available) icon appear on all screens, but percent battery power remaining and external power presence as well as battery type, current battery voltage, and current external voltage can also be viewed. From the Main Screen press F4 to view the Device Confi guration Menu, F2 to access the power options, then F4 twice to view the Power Status screen, including battery power percent remaining and the presence of external power. Battery type and battery voltage are also displayed, as well as external power voltage (if connected).
Button pressing chain:
Main Screen -> F4 -> F2 -> F4 -> F4
4.5 Changing the contrast
The DLI2’s LCD screen contrast values can be changed in two ways. One method is outlined in the Function Reference Guide. A faster, simpler way involves simultaneously pressing the CANCEL and UP or DOWN button repeatedly in any screen.
Button pressing chain:
CANCEL + UP repeatedly (to increase) or DOWN repeatedly (to decrease)
Page 10
Data Logger Manual
Section 5: DLI2 Function Reference
Section 5: DLI2 Function Reference
5.1 Main Screen
The main screen of the DLI2 features a real-time display of most recently measured pressure data and a real-time graph of the 100 most recently recorded data points. At the bottom of the main screen are tabs corresponding to each of the four function keys. These tabs are used to access the four main function menus of the DLI2: status, statistics, units, and setup. The left side of the main screen and all subsequent screens of the device is where important status information icons can be found (detailed in Section 3: Important Status Icons – page 5) including recording status, memory status, busy status, external power status, and battery power status.
5.2 Status Menu
Pressing F1 on the main screen brings up the Status menu. The fi rst screen that appears in the Status Menu is Run Parameters, but the Memory Status and Time screens can also be viewed by pressing the F2 and F3 keys respectively.
5.2.i Run Parameters
The Run Parameters screen displays important information regarding the device’s current recording session. These parameters include the time and date the recording session started (start time and start date), the time and date the recording session will end (stop date and stop time) due to either a full memory or preprogramming in the Dwyer software. The rate at which the DLI2 is recording (rate) is also displayed. The device’s current status (either running or stopped) is the last parameter on the Run Parameters screen.
5.2.ii Memory Status
The Memory Status screen is where all information regarding the DLI2’s memory. This screen displays the percent of memory space currently available (memory left), the number of readings currently stored on the device (readings), the maximum number of readings the device can record (max readings), as well as information about the wrap feature displaying either “disabled” or the number of wrap readings currently stored in memory.
5.2.iii Time and Date
The time and date screen displays current time data including the current time and date, time and date of last measured data, and current time zone.
Page 11