Dwyer DL8 User Manual

Page 1
Bulletin E-90-DL8
Series DL8 Process Data Logger
Specifications - Installation and Operating Instructions
Temperature Sensor
GENERAL
Eliminate traditional paper chart recorders with the Series DL8 Process Data Logger. This versatile, mulit-channel data logger monitors and stores up to 32,768 process readings for later download to IBM compatible computers. Seven external input channels easily interface with commom transducer and transmitter outputs. Internal thermistor measures ambient temperature conditions. The DL200 Windows (sold separately) makes programming the data logger sim­ple and easy. The Series DL8 Process Data Logger is ideal for monitoring temperature, relative humidity, pressure, wind speed, current, voltage and power.
GENERAL INSTALLATION (for all DL Loggers)
The Trendreader software must be installed on your com­puter prior to the use of any data logger described in this reference. NOTE: For additional information on graph manipulation, custom equations, modern setup, and sample rate tables, consult the Trendreader software manual.
(Trendreader) software and connecting cable
Input Jack for Interface Cable
PHYSICAL DATA
No. of Channels: Eight; One
for internal thermistor and seven for external analog inputs.
Accuracy: ±1% of full scale. Internal Thermistor Range:
-40 to 158°F (-40 to 70°C). Memory Size: 32,768 read­ings.
Sampling Methods:
Continuous (First-in, First-out) or Stop when full (Fill-then­stop). Sampling Rates: Selectable from 8 seconds to once every 5 days. Resolution: 8 bits (1 in 256). Power: Built-in 3.6V Lithium battery. Power Consumption: 5-10 µA. Clock Accuracy: ±8 sec­onds per day plus one sam­pling interval.
Security Eyelet
2.95 [75]
0.656 [17]
Input Impedance:
0 to 200 mV: > 20MΩ. 0 to 2.5 V: > 20M 0 to 5.0 V: 20K 0 to 10.0 V: 40.9K 0 to 25 mA: 100 Maximum Input Voltage: All Voltage channels: ±40 V, reverse polarity protected. All Current channels: ±7 V, reverse polarity protected.
Ambient Operating Temp:
-50 to 160°F (-45 to 70°C), 0 to 95% RH, non-condensing. Connection: Removable screw terminal.
Computer Requirements:
IBM compatible 386 or above and Windows with 2 mB RAM and 2 mB hard drive disk space, one ser­ial port.
Housing: Noryl Weight: 5 oz. (110 g).
3.375 [9.5]
3.1 or later
Magnetic Strip
0.45 [12]
Setup
To setup your datalogger you must first have Trendreader software installed and running on your computer. You can then configure your logger with various options by plugging into your computer via the interface cable. From the Communicate menu, choose Preferred Logger SR+. To access the logger window, double click on the logger icon or click Communicate on the menu bar and choose the highlighted Open COM from the menu.
DWYER INSTRUMENTS, INC.
P.O. Box 373 • Michigan City, IN 46361-0373, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
SERIES DL8 PROCESS DATA LOGGER MODELS & EQNS
Model No.
DL8
DL81
Input Type
0 to 2.5 VDC
0 to5 VDC (2 channels)
0 to 10 VDC
0 to 200 mV DC
0 to 25 mA DC (2 channels)
All channels 0 to 25 mA.
Equation #
7 18 17 19
6
6
Table 1
Phone: 219/879-8000 www.dwyer-inst.com
Page 2
If more than one logger icon appears on the screen or more than one Open COM is highlighted on the Communicate menu, this means you have more than one COM port avail­able on your computer. Choose the port that the particular logger is connected to.
The logger window will display the serial number of the log­ger Trendreader software is communicating with. See Figure 1. To contact another logger connected to the com­pute. 1. Double-click the serial number displayed to high­light it. 2. Type the serial number of the logger you want to contact. 3. Click the Contact button.
Enabling and Disabling Channels
Enabling a channel activates that channel and it will contin­ually record and store readings in memory. In the channel ON column, Yes means the channel is active, No means it is not. Click on the channel being changed to toggle between yes and no. Disable any channels not required to conserve memory space.
Equation Numbers
Equations are calculations the software uses to convert raw logger data to understandable units. Refer to page one (Table 1) for the appropriate equation number.
To see a list of standard equations available, double click one of the equation numbers in the Eqn column. The Select an Equation dialog box will appear, enabling you to scroll through the standard equations. If you want to choose one of the equations on the list, click it to highlight, and click Accept (or Enter). You can also choose an equation by typ­ing the number in the selected Equation Number box and clicking Accept (or Enter).
Custom equations can be created and assigned to specific logger channels. For additional information on creating cus­tom equations, please refer to the instruction manual includ­ed with the Trendreader software.
Figure 1
Filename
You cannot name the file during setup. Name the file when you are backing up the logger.
Current Time
Make sure the current time is correct. To change the time, access the Program manager on your computer and follow instructions in your computer setup manual.
Stop When Full and Continuous Mode
The logger can be set to stop taking readings when the memory is full, or to continue to take readings when the memory is full. To choose the first option, click the box in front of Stop When Full so that there is an “X” in it. To choose the continuous mode, click the box in front of Stop When Full so there is no “X” in it. The logger will continue to take readings when the memory is full, discarding the old­est reading as it records new readings.
Stopping Realtime
When the logger window is opened, the logger is operating in realtime mode, which means current readings are being displayed in the window. Before changing any settings, click Edit Setup.
Clearing Stored Data
To empty stored data, click Clear Stored Data.
Calibrations
The calibration values displayed in the low, med, high columns are factory set and match the sensor applied.
Viewing Realtime Readings
The realtime function allows the user to view measurements as they are being recorded. Viewing current readings can be useful when monitoring remote loggers by modem. In realtime mode, the logger reads and displays the variable every 8 seconds. If the sample rate is greater than 8-sec­onds, the average value for the sample interval is stored (based on 8-second readings).
Do not leave realtime running if you want to perform other tasks in Trendreader or if you want to switch to another Windows™application. The operation of Trendreader of other Windows™program will be slowed considerably with realtime running. To stop or start realtime, click the Edit Setup button. Stopping realtime has no effect on the log­ger’s information gathering.
Units of Measure
The units of measure can be changed to fit your particular needs. To change the units displayed in Realtime and on the logged data choose the Options menu, highlight the para­meter being measured. The units appropriate for that par­ticular parameter will be displayed in the units window. Select the box corresponding to the desired units of mea­sure and press Enter.
Page 3
When viewing in realtime, the new units will be displayed at the next update interval. The standard unit of measure will be designated with a 1 in the Scale column. When select­ing a different unit of measure, make sure the conversion from the standard is correct.
Choosing a Sample Rate
The sample rate is the frequency that the logger takes read­ings. The current sample rate is always displayed. to change the sample rate: 1. Click the current rate. The sam­ple rate dialog box appears. The dialog box displays current sample rate, the number of active channels, and the length of time it will take to fill the memory using the number of active channels and rate. (To determine the loggers capac­ity, enable the number of channels required before choosing a sample rate). 2. Click the down arrow to see available sample rates from 8 seconds to 8 hours in Continuous and Stop When Full modes. Scroll if necessary, and click on the desired rate. In Fast Read mode, sample rates less than 8 seconds are available. 3. Click Accept. The new sample rate will apply to all active channels. If you do not want to change the sample rate, click cancel.
Delaying Starting Time
Make sure the correct time is displayed in the logger win­dow before using this function. Once the logger has been setup, click Save Setup, and the logger will start gathering data. To delay the start time: 1. Double click in the Start Delay box to open the Select a Time dialog box. 2. Set the time and/or date for the logger to start taking readings. 3. Click Accept. The logger can be set for a delayed start time of up to two years.
Number of Readings
The maximum number of readings the DL8 logger can store is 32,768. If fewer readings are desired, change the number (in increments of 500).
Fast Read Mode Choose Fast Read Mode when:
1. You need a sample rate of less than 8-seconds, and /or
2. You want to delay the logger’s start time but do not want to specify the start time until after the logger has been setup.
Fast Read Mode allows you to choose sample rates from
0.04 seconds (25 readings per second) to 2 minutes. The sample rate must be in increments of 0.04 seconds. NOTE: A separate power source is required when using Fast Read Mode. Power can be supplied from the computer the log­ger is connected to or a separate battery pack (BP-100). If using the battery pack, the logger will start taking reading as soon as the battery is connected.
The logger cannot work in continuous Mode when using Fast Read Mode. The logger will stop taking readings as soon as the memory is full. The logger will also interrupt readings when it is connected to Trendreader software.
To delay start in Fast Read Mode:
1. Set the delay time in seconds in the Start Delay box. Use this method when the logger will remain connected to a computer or battery pack.
2. If the battery pack will be connected at a later time, fol­low this procedure before disconnecting the logger from the computer: Before clicking Save Setup, set that Delay for long enough that you can save the setup and disconnect the logger while it is asleep. The logger must be asleep (on start delay) when it is disconnected to be able to start tak­ing readings when the battery is connected to it. NOTE: You must set a minimum start delay of 16 seconds before you disconnect the logger from the computer. the logger must be disconnected from the computer within 8-seconds of setting the delay. If these steps are omitted, the logger will not start when the battery is connected.
To calculate the number of readings a particular time span will need, divide the time span by the sample rate (making sure both are in seconds or minutes). If you are taking read­ings at 8-second intervals for 24-hours, and need to know the number of readings, convert 24-hours to seconds (86400) and divide by the sample rate of 8-seconds to get 10,800 readings. When backing up the logger, it will stop at the specified number of readings.
Saving Changes
When you have finished setting up your logger, click Save Setup. A dialog box will appear asking to clear the logger and save setup. (Setting up a logger erases all the previ­ously gathered data). Click OK.
Do not contact a logger in Fast Read Mode until you are going to back it up. Contacting the logger will stop it from taking readings. You can contact a logger by clicking Contact or by opening the window once it has been closed.
BACKING UP A LOGGER
Copying data from a logger to your computer requires an IC-100 or IC-200 Interface cable. to backup a single logger use the IC-100 Interface Module. The IC-200 Interface Module can be used to backup several loggers. Backing up a logger can take a number of minutes depending on the speed of your computer and the amount of data being transferred.
Page 4
The process can be speed up by selecting the Set Baud rate, Fastest Available from the Communicate menu. (If you have problems backing up a logger, try a slower baud rate. On some computers, choosing the fastest available baud rate causes timing conflicts. NOTE: The IC-200 module automatically transmits at 1200 baud and cannot be adjust­ed.
Backing Up a Logger
From the Communicate menu, choose Preferred Logger SR+. To access the logger window, double click on the log­ger icon or click Communicate on the menu bar and choose the highlighted Open COM from the menu. To backup a log­ger, click Backup in the logger window. A dialog box appears showing the baud rate, the serial number of the logger, and the progress of the backup. (You may cancel the backup procedure by clicking Cancel).
Figure 2
Opening a Graph after Backup
If you want to view the transferred information as a graph immediately after backup, open the Options menu in the main window and make sure a check mark is in front of Open Graph after Backup. NOTE: See the Trendreader soft­ware manual for details on manipulating graphs.
Naming a File
When backup is completed, the Save As dialog box appears to enable you to name the file. The default name is always the serial number of the logger. If you want to keep the logger’s serial number as the filename, click OK. If you want to give the file a different name, type the new name and click OK. NOTE: You cannot type anything after the dot in the Trendreader filename. The three letter extension, is automatically assigned.
Backing Up Additional Loggers
You can daisy-chain the DL8 loggers together and back them up one by one. You may want to disable the opening a graph after backup function when you are backing up several loggers. If this function is not disabled, a graph will be opened after backing up each logger, increasing the total backup time. To disable this function, open the Options menu and make sure there is no check mark in front of Open Graph After Backup.
Alarm Dialout is enabled by setting high and low alarm thresholds, entering the number that is to receive the alarm call, and sending the setup to the logger. When an alarm condition occurs, the logger will dial out and leave the remote site number on the receiver’s telephone caller ID or on the pager’s display. (Alarm Dialout does not leave a ver­bal message). To activate Alarm dialout, click Alarm Dialout in the logger window so that there is a check in the Alarm Dialout box. A dialog box will appear at the bottom of the window. See Figure 2.
Setting High and Low Alarm Thresholds
Set high and low alarm thresholds for each active channel on the logger. Any reading lower than the low alarm thresh­old or higher than the high alarm threshold will cause the logger to dial out from the remote site. Set the high alarm to the highest possible value and low alarm to the lowest pos­sible value if you do not want the alarm activated on that channel.
Choosing a Baud Rate
Baud rate is a measure of the transmission speed of the dial commands that the logger sends to the modem. The default baud rate is 1200, which is reliable and accurate for most situations. Higher baud rates are available in the baud rate box.
ALARM DIALOUT
Alarm Dialout enables you to setup the logger at a remote site so that it will dial a phone number when readings have exceeded preset alarm thresholds. It can be setup to dial a telephone or pager. The receiving telephone requires caller ID to use this feature.
If sending the alarm to a pager, the pager system must accept touch-tone codes that can be displayed on the pager.
Entering the Number to Dial A. Calling a telephone: Remember the telephone must
have caller ID to receive the alarm.
1. Type a dial command in the Number to Dial box. If the phone line at the remote site has tone dial, type ATDT. If pulse dial, type ATDP.
2. After the dial command, type the number of the telephone that is to receive the alarm call.
Page 5
B. Calling a pager: The pager system must accept touch-tone codes that can be displayed on the pager.
1. Type the dial command ATDT in the Number to
Dial box.
2. After the dial command, type the number of the
pager that is to receive the alarm call.
3. After the pager number and before the identify­ing code there must be a pause to allow the receiving sta­tion to answer. Enter this pause by typing commas. (For most modems, each comma causes a pause of two sec­onds).
4. Type a code that will identify the logger. For example: ATDT 555-2222,,,,,40167, would instruct the log­ger to dial the pager number and pause for ten seconds, and then send the serial number of the logger.
Setting Call Delay
Call delay is the length of time that a logger will wait before dialing out once an alarm condition has occurred. If you do not want the remote logger to dial out immediately when one or more of the channels goes into alarm condition, you can set the Call delay to designate the number of seconds the logger will wait before dialing.
If the alarm condition does not exist for the entire delay peri­od, the logger will not dial out. If the logger comes out of an alarm condition during the call delay period, the delay timer will stop and will start again from zero at the next alarm event.
Setting Retry Delay
If a remote logger dials out an alarm but cannot get through (line is busy), it will continue to dial out until connection is made. Retry delay is the length of time the logger will wait between unsuccessful attempts. It is recommended to set the retry dial to 5 minutes. The logger will continue to dial out after every retry delay time and as long as the alarm condition exists and has not been acknowledged.
Setting Reset Delay
Reset Delay begins when you acknowledge the alarm by backing up the logger that has dialed out. Reset Delay deactivates Alarm Dialout to give time to fix the problem that caused the alarm. When Reset Delay period has elapsed, Alarm Dialout will reactivate, and if an alarm condition exists, the logger will dial out again. IMPORTANT: If an alarm condition occurs and Reset Delay has been initiated by the last backup, the logger will wait until the Reset Delay period is over before dialing out. To bypass the Reset Delay period (if you have fixed the alarm problem but time remains on the Reset Delay), click the Clear Stored Data button and follow the subsequent prompts.
Sending the Setup to the Logger
After choosing Alarm Dialout settings, setup logger by click­ing the Save Setup/Clear Stored Data button and follow the subsequent prompts.
Responding to Alarm Call
1. When the logger dials out to warn of alarm condition, dial into the site and backup the logger.
2. Fix the problem that caused the alarm.
3. Reactivate Alarm Dialout by contacting the logger and clicking Clear Stored Data. NOTE: Test Alarm Dialout regularly to ensure proper opera­tion.
Fast Read Mode Disabled
Alarm Dialout is not designed to operate in Fast Read Mode. When Alarm Dialout is enabled, fast Read Mode is disabled; when Fast Read Mode is enabled, Alarm Dialout is disabled.
Cold or Humid Environments
The environment the logger will be placed in must be suit­able, please refer to the physical data section. If conditions are not acceptable, consider using a protective enclosure. For humid environments, the logger can be protected by placing it in a ziplock plastic bag.
If the logger is used in a cold environment, make sure con­densation will not settle on the logger when it is brought back into a warmer environment. The best way to prevent condensation is to place the logger in a plastic ziplock bag and include a dessicant. When you bring the logger back into the warmer climate, leave the logger in the bag with the dessicant until the logger has come to equilibrium with the environment.
Label
If working with more than one logger, label each, identifying the task and location before you distribute them throughout a building or system. To do this, simply use shipping tags. Later, when you retrieve them to graph their data, you will know what each graph refers to.
Mounting
Use the magnetic backing to conveniently mount the logger on metal surfaces like ductwork or electrical control cabi­nets. If concerned about theft, make sure to lock the logger to a permanent fixture using the special locking tab.
Special mounting methods (using Velcro®fasteners) to secure the loggers to other surfaces may be used. NOTE: Do not rely on the logger’s magnetic strip for secure mount­ing if the surface is uneven, unstable, or above 150°F (65°C).
Page 6
Keep Track
Be sure to keep record of the locations of each logger. This will save time in looking for them when the data-gathering is completed. Also, keep track of when the loggers were put into service. This will help when producing graphs.
Retrieval
After sufficient time has passed to obtain a representative profile of data, retrieve the logger and bring it back immedi­ately for analysis. Make sure the logger has a label so it can be properly identified and differentiated.
Analysis
To analyze the logger, you must first transfer a copy of its data into your computer. To do this, plug the logger into the Trendreader interface cable and choose Logger, Communicate, Backup. After describing the information to the computer, the data is automatically copied to disk, time and date stamped, and converted into the appropriate measurement units. A portion or the entire data set can be copied.
Each file will initially have the same descriptive title, but you can use Files, Revise Logger to alter these accordingly. To view graphs, choose Draw, New and select the appropriate file. A detailed description of all software functions can be found in the Trendreader Reference Guide.
4. Connect the power supply adapter to the gender chang­er.
5. Connect the 6VDC power supply to the adapter. NOTE: The cable length between the IC-200 and the com­puter or modem can be extended. Maximum length should be 500 feet (150 m). Use 22-gauge stranded wire, with two conductors and a shield.
Identifying Loggers on a Network
Before you can contact a network group of loggers at a site for the first time, you must name the site and list serial num­bers so they can be identified. 1 From the Communicate menu in the main Trendreader window, choose SR+ Network so that a check mark exists in front of it.
2. Choose the COM port that the loggers are connected to. The Logger List dialog box will appear. Site name and log­ger serial numbers will be entered in this dialog box. See Figure 3 below.
OPERATING A LOGGER NETWORK
The Series DL8 Dataloggers can be networked to enhance the monitoring and backup capabilities. Use model IC-200 Interface Module to connect as many as 8 loggers to your computer simultaneously. The IC-200 Module will allow backup of a 8 loggers, one after the other, with one com­mand. Realtime data from up to 64 channels (8 loggers with 8 channels) can be displayed on a single screen.
Connecting an IC-200 to the Computer
1. Connect the serial port extension cable’s female end to the computer’s 25-pin serial port. Use the 9-pin connector if the computer has a 9-pin port).
2. Connect the cable’s other end to the IC-200 connection maked “PC”.
3. Connect the power supply adapter to the other end of the IC-200 (the end marked “modem”).
4. Connect the 6VDC power supply to the adapter.
Connecting an IC-200 to a Modem at a Remote Site
1. Connect the serial port extension cable to the modem.
2. Connect the cable’s other end to the IC-200 connection marked “modem”.
3. Connect the gender changer to the other end of the IC­200 (marked “PC”).
Figure 3
3. In the Site Name box type a unique name for the site and press Enter.
4. Click in the Serial # Column, type the serial number of one of the loggers, and press Enter. (The order does not matter, however, if you click Show All to view realtime read­ings, the readings will be displayed in the order they were entered.)
5. If desired, add a description in the Description column and press Enter.
6. Repeat steps 4 and 5 until all the loggers at the site are listed.
7. Click Save Changes. You can now contact any of the loggers listed.
Deleting a Site and its Loggers
1. Highlight the site to be deleted in the Site Name box. To scroll between sites, click the down arrow.
2. Click Delete. The site name serial numbers and descrip­tions will be deleted.
3. Click Save Changes.
Deleting One Logger from a List
1. Click in the loggers serial number box to highlight the number.
2. Click Delete. The serial number and description will be deleted.
3. Click Save Changes.
Page 7
Viewing Realtime Readings of All Loggers
1. From the Communicate menu in the main Trendreader window, click SR+ Network until a check mark appears in the box.
2. Choose the COM port the network is connected to. The Logger List dialog box will appear.
3. Click Show All. The Network Realtime window will open displaying each logger’s realtime readings. See figure 4 below. When finished viewing the realtime readings, either click Close to close the window, or double-click a logger’s serial number to open that logger’s window.
Figure 4
Contacting a Network Logger
There are three ways to contact a logger on a network:
1. From the Logger List dialog box
2. From the Network Realtime window
3. From the Logger window
SETUP (specifically for the DL8)
This section provides guidelines to follow to get you started with your DL8.
1. The first step in using your DL8 is to decide what infor­mation, or parameters, you want to log. For example, you may simply wish to replace an existing hard-wired 4-20 mA chart recorder or, instead, monitor from a special multiple­transducer circuit that you have assembled.
2. Determine what sensors, transducers, or transmitters you need to do the job (if not already present).
3. Modify the channels, externally (using resistors), if required for maximum resolution (see Customizing Input ranges).
4. Write an equation to convert the raw readings from the transducer or circuit you want to monitor into the proper engineering units you require (for example: converting 4-20 mA signal from a relative humidity transmitter to 0 to 100% R.H.). For help in writing these equations refer to your Trendreader Software manual.
Contacting a Logger from the Logger List
Double-click the logger’s serial number (or click the logger’s serial number and click Contact Logger). The loggers main window will open.
Contacting a Logger from the Network Realtime
When you have contacted a logger by one of the previous two methods and you have that logger’s window open, you can contact another logger on the network by taking the fol­lowing steps:
1. Click Edit Setup. (See figure 5).
2. Double-click the serial number in the Serial Number box.
3. Type the number of the logger you want to contact.
4. Click the Contact button to open that logger’s window. NOTE: DO NOT setup networked loggers in the Fast Read Mode. Any networked logger that is setup in Fast Read Mode will stop logging as soon as it or any other logger on the network is contacted.
Figure 5
Figure 6: Using Transducers with Power Supply
5. Decide how you are going to power your transducers (if required). You can either use a standard power supply and create a circuit like that shown in Figure 6 or use batteries (see External Battery Operation) and produce your own self­contained logging assembly (Figure 7). For this application, you must activate the internal battery switch.
Figure 7: Using Transducers with a Battery
6. Connect your transducers or transmitters and check operation through the Logger, Communicate, Realtime function in the Trendreader Software.
Page 8
Avoiding Ground Loop Problems
Take special care to avoid ground loop problems when you use your DL8 Process Data Logger. A ground loop can occur when there is more than one path to ground in your logger-transducer circuit. Ground loops can damage your DL8 Process Data Logger as well as your transducers. To avoid ground-loop problems your DL8 circuit:
• Do not use more than one grounded power sup ply to power your transducers.
• Do not connect your logger to your computer in real time readings unless:
• your computer is battery-operated (i.e. not
grounded); or
• your transducer power supply is not grounded. If you must use more than one grounded power supply in your circuit, each transducer you use must be isolated. If you must use non-isolated transducers, then you must use a Signal Repeater/Loop Isolator between the transducer output and the DL8 input.
Wiring Notes and Considerations
1. The Series DL8 Process Data Logger has two common inputs (labeled “-Common”). These are not isolated from each other and can be used as the “-” connection for the power supply (or battery) and any of the transducers being logged by the Series DL8.
2. For permanent applications, wiring connections can first be made to the Series DL8 removable terminal block. The terminal block can then be attached (with glue, for example) to a permanent fixture. When it comes time to analyze data, the data logger can simply be unplugged from the terminal block and brought back to your computer. After backing up the data, the logger can be returned and plugged back into the terminal block. Logging will then automatically continue.
3. The 2.5 V and 200 mV channels “float” when no con­nections are made to them. They can thus be expected to read a positive voltage when left both active and uncon­nected. When connected, however, they will read the cor­rect input voltage.
4. The “25 mA” channels read positive current only. Make sure to observe polarity. If you are using both channels simultaneously, ensure that both 4-20 mA transmitters source a positive current through the logger to a common (-).
CHOOSING INPUT CHANNELS
To choose which of the DL8’s seven external input channels to use for your particular applications and understanding of resolution is required.
Maximizing Resolution
The resolution of your DL8 Logger is eight bits. This means that it can resolve analog signals, with a defined range, to 256 discrete steps (2 to the power of 8). When logging from the 2.5V channel, for example, the DL8 will record digital values in increments of 2.5 volts divided by 256, or approx­imately 10 millivolts.
Resolution is usually not a significant factor until you have logging applications that will produce input signals far less than that of the range of the logger’s channel itself.
The 2.5V channel will have a resolution of 0.4% over a range of 0 to 2.5 volts (.01/2.5 x 100) which should be more than adequate for most applications. If, however, you are intending to measure signals ranging from 50 to 150 milli­volts (with the same 2.5V channel), resolution will be much coarser (10%). By switching the 50 to 150 millivolt input to the 200mV channel, resolution can be improved to approx­imately 1%.
Choosing Input Channels
It is important to maximize resolution when measuring and recording from the DL8’s current and voltage channels. Usually you can do this simply by choosing the channels that match or approximate the input signals you will be using. The table below lists several possible input signal ranges and the recommended channels to use. It also lists alternative channels that can be used to add additional simultaneous monitoring capability. These alternative chan­nels will, however, require the addition of external resistors (see Customizing Input Ranges).
Range
of Input
0 to 500 mV
0 to 1.0 V 0 to 2.0 V 0 to 3.0 V 0 to 6.0 V
0 to 20 V
0 to 2.0 mA
4 to 20 mA 0 to 50 mA
0 to 100 mA
In the above table, the Range of Input column lists exam­ples of input signals that you may wish to monitor using your logger. The Standard Solution column lists the channel you would normally choose to monitor that particular signal.
In this table, 200 refers to the 200mV channel, 2.5 the 0 to
2.5V channel, 5 to either of the two 0 to 5V channels, 10 to the 0 to 10V channel, and 25 to either of the two 0 to 25mA channels. The Alternative Solution column lists other chan­nels that can be used to monitor the same input signals. In most cases, these channels will need to be fitted with exter­nal resistors.
Standard
Solution
2.5
2.5
2.5 5
10
N/A
25
25 N/A N/A
Alternative Solution(s)
200 200 200 200
5, 2.5, 200
10, 5, 2.5, 200
200
2.5, 200 25, 5
25, 10
Page 9
Non-Standard Ranges
You may want to use a special transducer or tie in an exist­ing process signal loop that does not match the DL8’s stan­dard input ranges. The following step-by-step procedure will help you choose which channel is best.
1. Determine the maximum output signal (M) and the zero offset signal (Z) of the transmitter or circuit you wish to tie into. For example, a transducer with a specified range of 1 to 6 volts DC will have a maximum output signal of 6 volts and a zero offset signal of 1 volt. A 4 to 20 milliamp trans­mitter will have a maximum output signal of 20 mA and a zero offset signal of 4 mA.
5. To find out in what actual steps (S) your logger will collect data, first determine the input (I) of your transducer or trans­mitter. You can do this by subtracting the lowest level input from the high. A thermocouple transmitter, for example, with a range of -20° to 600° will have an input span of 620° (600­(-20)). A pressure transducer with a range of 0 to 100 psi will have an input span of 100 psi.
6. To calculate S, measured in the same units as the input span (I) above, use this equations:
R x 1
S =
100
2. Determine the transducer’s full scale output (F). You can do this simply by subtracting Z from M. The full scale out­put of the 1 to 6 volt transducer is 5 volts. The full scale out­put of the 4-20 mA transmitter is 16 mA.
3. Compare M with the channels available on the logger and choose a channel with an input equal or greater to this value (if available). If no such channel exists, then modify one to suit by referring to the Customizing Input Ranges section.
4. Determine the resolution (R) of the channel you chose in Step # 2 will be adequate. You can determine this, in per­cent, by using the following equation:
C x 0.4
R =
Where:
C = the input range of the channel to use (in either millivolts, volts or milliamps). F = the full scale output of your transducer (M-Z) measured in the same units as for C. R = the resolution expressed in a percentage (%) of full scale output.
For example, a 0.5 to 2.5 volt transducer connected to the
2.5V channel will be logged with a resolution of 0.5%. A 4 to 20 mA transducer hooked up to the 25 mA channel will log with a resolution of 0.6%.
F
For example, a 4 to 20 mA Relative Humidity transmitter with an input span (I) of 90% R.H. and a calculated resolu­tion (R) of 0.6% will record data in steps of 0.54% R.H.
CUSTOMIZING INPUT RANGES
The standard input ranges on the DL8 should be suitable for most process signal applications, but occasionally you may require measurement of non-standard input levels where resolution is a primary concern (see the Choosing Input Channels section). Ranges can be adapted easily by using either one or two external resistors. Trendreader Software’s Expert, Equations function can then be used to scale the data to the new input levels.
The resistor value that you will require to modify each log­ger’s input channel is selected by inserting your special input requirements into a simple equation. The following sections detail these equations and the method of tying in these resistors.
10 Volt Channel
The 10 volt channel can be modified to span higher ranges (greater than 0 volts DC) by adding a single resistor to the “10V” terminal shown in Figure 8. The value of the resistor R must be calculated by the following formula based on the desired voltage input range Ein:
R = (4,090 x Ein) - 40,900
Generally, it is good practice to keep resolution to within 1% for most applications. However, you may be able to accept far coarser resolution in some applications and, in others, require far better. You’ll be able to determine your own requirements after you’d be become familiar with the results you can expect with different arrangements. If your require­ments demand greater resolution than what you calculated, then you should set up your inputs with resistors as described in the Customizing Input Ranges section.
For example, if you wish to measure up to 20 volts, the resistor value should be:
(4,090 x 20) - 40,900 = 40,900 ohms.
Since you won’t always be able to get the exact resistor value that you calculated, you can determine the logger’s actual input range by plugging the value of the resistor (R) you obtain back into the equation. For example, if the clos­est resistor you could obtain was 42K ohms, then the actu­al voltage input range would be:
42,000 + 40,900
Ein=
4,090
= 20.27 volts
Page 10
This value will be needed when it comes time to create your own custom equation using Trendreader Software’s Expert, Equations function.
Figure 8: Modifying the 10 volt channel for extended range
You can also easily convert the 10V channel to measure 0 to 100 mA current simply by putting a 100 ohm resistor (minimum rating: 2 watts) between the “10V” and “-Common” terminals.
5 Volt Channels
The two 5 volt channels can be modified to span higher ranges (greater than 5V DC) by adding a single resistor, in the same way as shown for the 10V channel in Figure 8, to either, or both, of the two “5V” terminals. The value of the resistor R must be calculated by the following equation based on the desired voltage input range Ein:
R = (4,000 x Ein) - 20,000
For example, if you wish to measure up to 20 Volts, the resistor value should be:
(4,000 x 20) - 10,000 = 70,000 ohms.
Figure 9: Modifying the 2.5 volt channel
Since you won’t always be able to get the exact resistor value that you calculated, you can determine the actual input range you’ll have by plugging the value of the resistor back into the formula:
R + 10,000
Ein=
4,000
This value will be needed when it comes time to create your own custom equation using Trendreader software’s Expert, Equations function.
For example, if you wish to measure up to 20 volts, the resistor value should be:
(4,000 x 20) - 20,000 = 60,000 ohms.
Since you won’t be able to get the exact resistor value that you calculated, you can determine the actual input range you’ll have by plugging the value of the resistor back into the formula:
R + 20,000
Ein =
4,000
You can easily convert either of both 5V channels to mea­sure 0 to 50 mA current by wiring a 100 ohm resistor (min­imum rating: 0.5 watts) from the “5V” terminals to common.
2.5 Volt Channel
The 2.5 volt channel can be modified to span higher ranges (greater than 2.5V DC) by adding two resistors to the input terminals as shown in Figure 9. The value of the resistor marked “10K” must be 10,000 ohms ±1%. The value of the resistor marked “R” must be calculated by the following equation based on the desired voltage input range (E
in
):
R = (4,000 x Ein) - 10,000
You can easily convert the 2.5V channel to measure 0 to 25 mA current by wiring a 100 ohm resistor from the “2.5V” ter­minal to common.
200 millivolt Channel
The 200 millivolt channel can be modified to span higher ranges (greater than 200 mV DC) by also adding two resis- tors, in the same way as shown for the 2.5V channel in Figure 9. The value of the resistor marked “10K” must be 10,000 ohms ±1%. The value of the resistor marked “R” must be calculated by the following formula based on the desired voltage input range (Ein):
R = (50,000 x Ein) - 10,000
You can also modify the 200mV channel to log current loop signals with greater sensitivity than the two 25 mA chan­nels. To do this, put a 100 ohm resistor across the “200mV” and “-Common” terminals and you will be able to log cur­rents from 0 to 2.0 mA with approximately ten times the res­olution of the 25mA channels. The equation to use to determine other current (Iin) ranges, in milliamps is:
200
S =
in
I
Page 11
25 milliamp Channel
The 25 milliamp channel can be modified to span higher ranges (greater than 25 mA DC) by also adding a single resistor across the input terminals as shown in Figure 10. The value of the resistor “R” must be calculated by the fol­lowing formula based on the desired current input range (Iin):
The internal switch can be activated during exit from setup session by choosing “Yes” as answer to “Activate the Logger’s external battery switch?” The battery switch sta­tus can be checked in setup menu. If the battery switch is activated Hi Cal for channel 1 (25mA) reads -5. If the switch is not activated, this value is 0.
2,500
R =
Iin- 25
Make sure the resistor can withstand at least twice the power you will be sending it. To calculate this, use the fol­lowing equation:
2
P = 2 * I
in
* R
Where:
P = Minimum Power Rating of resistor (in watts) Iin= Maximum amperage expected through resistor (in Amps). R = Resistance of Resistor (in ohms).
For example, a 50 ohm resistor calculated to measure cur­rents up to 75mA will require a power rating greater or equal to 0.56 watts. Therefore, a one watt resistor will be fine.
The switch works by closing eight seconds prior to a read­ing, remaining closed until the reading is taken, then imme­diately opening again before repeating the cycle.
It extends battery life by only drawing power, when neces­sary, to meet the input requirements of the transducer or transmitter circuit that you want to monitor.
To use the battery switch, the battery’s positive (+) terminal must be wired to the “+ Battery In” terminal on the data log­ger (refer to Figure 11). The power to your external trans­ducers will then be available from the “+ Power Out” termi­nal. The battery’s negative (-) terminal can then be wired to either of the “-Common” terminals.
Figure 10: Modifying a 25 mA channel
For increased sensitivity at lower current ranges down to
2.5 milliamps, refer to the section for modifying the 200 mV channel.
External Battery Operation
You can activate a special function on the DL8 designed to optimize the use of power from external batteries while out in the field. The battery-saving contacts can provide switch­ing capability to draw power from the battery only when needed, thus maximizing its life for extended field applica­tions.
The internal battery switch allow you to power external transducers with a separate battery pack. This capability makes it easy for you to assemble your own self-contained data logging kit for long-term in-field use where there is no convenient power supply receptacle.
Figure 11: Using the Battery Switch
Maximum current that the battery switch can source is 100mA. Battery supply voltage can be from 9 to 25 volts. Short circuit protection is continuous at 9 volts, but only one second at higher voltages.
Things you should know about batteries
When using batteries for power your external sensors and transducers, you should keep in mind a few key points.
1. The rated voltage of a battery only applies to initial use. Once under load, the voltage will decrease gradually until completely drained. By knowing how the battery voltage is affected through use you can make sure that the voltage requirements of your external transducers will be met. A good rule of thumb to follow is to exceed the minimum required input voltage for your transducers by at least two times (where acceptable).
2. Nickel-Cadmium batteries (rechargeable) self-drain them­selves at a rate approximately 1% of their remaining capac­ity per day. This should be taken into account for extended logging sessions.
Page 12
Determining how long your batteries will last
To estimate how long your batteries will successfully power your external sensors and transducers, you will need to know both the capacity of the batteries as well as the pro­posed resistance presented to them by the load (your own transducers).
Battery capacity, usually expressed in milliamperes hours (mAh), is the total amount of electrical charge a cell can store or deliver. Unfortunately, most commercially-available batteries are sold without this information.
As a general guide, most good quality nine volt batteries have a capacity of approximately 500 mAh. This means, at an average continuous current draw of 50 mA it will deliver approximately ten hours of service.
The load you impose on your battery will depend on how many transducers you intend to have in your logging circuit. The greater the number of transducers, the greater the power draw from the battery. If you are logging from a 4 to 20mA transducer, the maximum current it will draw will be approximately 20mA. If you power it continuously, your 500 mAh battery should not run out until you’d had at least twenty-five hours of service. Since your transducer proba­bly will not always draw maximum current (12mA might be a more reasonable figure), you can probably expect up to forty-two hours of service.
Regulated Voltages
The unstable voltage characteristics of batteries makes them, by themselves, incompatible for use with transducers that required a regulated voltage supply. You can, however, add your own voltage regulator to your transducer circuit to compensate for this instability (see Figure 12). Voltage reg­ulators are readily available at most electronic parts supply stores (such as Radio Shack).
Figure 12: Regulating Battery Supply Voltage
Make sure that your supply voltage to your regulator is at least two volts higher than the level to which you want to regulate to. To do this, you can easily increase the voltage of your batteries by hooking them up in series as shown in Figure 12.
The battery-saving switch contacts can increase the lifetime of your external batteries significantly. Since the batteries will be powered only for eight seconds out of every logging interval, the power draw on the battery will be reduced con­siderably.
Example Problem: You want to log pressure once every thirty min­utes from a 0 to 300 psig transducer. The output of the transducer is 4 to 20mA. The excitation voltage is listed at between 9 and 40 volts DC.
Solution: From this information we know that a 4 mA sig­nal will represent 0 psig and a 20 mA signal will represent 300 psig. Since the minimum excitation voltage is 9 volts, we should supply at least 18 volts initially from our batteries. We can do this by simply connecting two nine volt 500 mAh batteries in series.
Without the battery-saving switch enabled, our minimum expected battery life would be approximately twenty-five hours. With the switch enabled, and thus closing only eight out of every 1,800 seconds (or thirty minutes), the life will be extended by a factor of 225 (or 1800/8). This means that we should be able to log from this transducer for over 5,625
If the wires between the voltage regulator, batteries, and transducers are long (more than a few inches) the regulator may oscillate and produce an unstable voltage. You can correct this by connecting one 0.1 microfarad capacitor to each of the regulator’s outside pins then wiring them to the regulator’s center pin. Make sure to keep the capacitor leads as short as possible.
MAINTENANCE
No routine maintenance is required on the Series DL data loggers. Periodic checks of connections and mounting is recommended. Please contact Dwyer Instruments, Inc. before returning unit for repair to review information relative to your application and obtain a return authorization num­ber. When returning a product to the factory, carefully pack­age and ship freight prepaid. Be sure to include a complete description of the application and problem and identify any hazardous material used with the product.
©Copyright 1999 Dwyer Instruments, Inc Printed in U.S.A. 2/99 FR R1-443087-00
DWYER INSTRUMENTS, INC.
P.O. Box 373 • Michigan City, IN 46361-0373, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
Phone: 219/879-8000 www.dwyer-inst.com
Loading...