Dwyer DL7 User Manual

Bulletin E-90-DL7
INPUT JACK FOR INTERFACE CABLE
MAGNETIC STRIP
11/32
[8.730]
2-1/2
[62.29]
TEMPERATURE SENSOR
1/2
[12.70]
2-7/32
[56.36]
SECURITY
EYELET
Series DL7 Differential Pressure Logger
Specifications - Installation and Operating Instructions
GENERAL
The Series DL7 Dataloggers are easy-to use battery-powered data loggers that can monitor pressure and temperature. Each unit can reliably record time-based data for later analysis, by Trendreader software and any IBM PC or 100% compatible com­puter.
GENERAL INSTALLATION
The Trendreader software must be installed on your computer prior to the use of any data logger described in this reference. NOTE: The Series DL7 Differential Pressure Data Logger has additional features not found on some of the other DL Series log­gers. This manual describes some of the unique features of the DL7. For additional information on graph manipulation, custom equations, modem setup, and sample rate tables, consult the Trendreader software manual.
SETUP
To setup your datalogger you must first have Trendreader software installed and running on your computer. You can then configure your logger with various options by plugging into your computer via the interface cable. From the Communicate menu, choose Preferred Logger SR+. To access the logger window, double click on the logger icon or click Communicate on the menu bar and choose the highlighted Open COM from the menu.
PHYSICAL DATA
No. of Channels: Two; internal
thermistor, and a differential pressure sensor. Media: Dry air and noncorrosive gases. Max. Pressure Rating: 5x rated pressure
Accuracy: ±0.05” w.c. Resolution: ±0.01”w.c. Nonlinearity: ±0.05% FS
(BFSL)
Hysteresis and Repeatability:
±0.05% FS. Internal Thermistor Range: ­40 to 158°F (-40 to 70°C)
Internal Thermistor Resolution: 0.7°F (0.4°C), R25
value equal to 10,000[10kΩ @ 25°C (77°F)]
Compensated Temp Range:
32 to 158°F (0 to 70°C)
Thermal Accuracy: ±0.5% FS. Memory Size: 21,500 readings Resolution: 12 bits (1 in 4096). Stability: ±0.5% FS/yr.
Sampling Methods:
Continuous (First-out: not avail­able from 40 ms to 8 sec.), Stop when full (Fill-then-stop), or Delayed start. Sampling Rates: Selectable from 0.04 seconds to 8 hours. Power: Built-in 3.6V Lithium battery, 1 Amp-hour. Power Consumption: 5-10 µA, continuous. Clock Accuracy: ±2 sec/day plus one sampling interval. Ambient Operating Temp: -40 to 158°F (-40 to 70°C), 0 to 95% RH, non-condensing. Connection: Two 1/8’ I.D. per­manent protective tubing. Computer Requirements: IBM compatible 386 or above and Windows™ 3.1 or later with 2mB RAM and 2mB hard drive disk space, one serial port.
Housing: Noryl™ Weight: 4 oz (110 g).
EQUATION #
Model
Number
DL70 ±0.5˝ w.c. (1.2 mbar) 13 DL71 ±1˝ w.c. (2.5 mbar) 13 DL72 ±2.5˝ w.c. (6.2 mbar) 88 DL75 ±5˝ w.c. (12.4 mbar) 37
DL710 ±10˝ w.c.(24.9 mbar) 37
Pressure Range Equation #
TABLE 1
DWYER INSTRUMENTS, INC.
P.O. Box 373 • Michigan City, IN 46361-0373, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
Phone: 219/879-8000 www.dwyer-inst.com
If more than one logger icon appears on the screen or more than one Open COM is highlighted on the Communicate menu, this means you have more than one COM port available on your computer. Choose the port that the particular logger is connect­ed to.
The logger window will display the serial number of the logger Trendreader software is communicating with. See Figure 1. To contact another logger connected to the compuet. 1. Double­click the serial number displayed to highlight it. 2. Type the serial number of the logger you want to contact. 3. Click the Contact button.
FIGURE 1
Filename
You cannot name the file during setup. Name the file when you are backing up the logger.
Current Time
Make sure the current time is correct. To change the time, access the Program manager on your computer and follow instructions in your computer setup manual.
Stop When Full Continuous Mode
The logger can be set to stop taking readings when the memory is full, or to continue to take readings when the memory is full. To choose the first option, click the box in front of Stop When Full so that there is an “X” in it. To choose the continuous mode, click the box in front of Stop When Full so there is no “X” in it. The logger will continue to take readings when the memory is full, discarding the oldest reading as it records new readings.
Stopping Realtime
When the logger window is opened, the logger is operating in realtime mode, which means current readings are being dis­played in the window. Before changing any settings, click Edit Setup.
Enabling and Disabling Channels
Enabling a channel activates that channel and it will continually record and store readings in memory. In the channel ON col­umn, Yes means the channel is active, No means it is not. Click on the channel being changed to toggle between yes and no. Disable any channels not required to conserve memory space.
Equation Numbers
Equations are calculations the software uses to convert raw log­ger data to understandable units. Refer to page one (Table 1) for the appropriate equation number.
To see a list of standard equations available, double click one of the equation numbers in the Eqn column. The Select an Equation dialog box will appear, enabling you to scroll through the standard equations. If you want to choose one of the equa­tions on the list, click it to highlight, and click Accept (or Enter). You can also choose an equation by typing the number in the selected Equation Number box and clicking Accept (or Enter).
Custom equations can be created and assigned to specific log­ger channels. For additional information on creating custom equations, please refer to the instruction manual included with the Trendreader software.
Calibrations
The calibration values displayed in the low, med, high columns are factory set and match the sensor supplied.
Viewing Realtime Readings
The realtime function allows the user to view measurements as they are being recorded. Viewing current readings can be useful when monitoring remote loggers by modem. In realtime mode, the logger reads and displays the variable every 8 seconds. If the sample rate is greater than 8-seconds, the average value for the sample interval is stored (based on 8-second readings).
Do not leave realtime running if you want to perform other tasks in Trendreader or if you want to switch to another Windows application. The operation of Trendreader of other Windows pro­gram will be slowed considerably with realtime funning. To stop or start realtime, click the Edit Setup button. Stopping realtime has no effect on the logger’s information gathering.
Units of Measure
The units of measure can be changed to fit your particular needs. To change the units displayed in Realtime and on the logged data choose the Options menu, hightlight Units, and highlight the parameter being measured. The units appropriate for that particular parameter will be displayed in the units win­dow. Select the box corresponding to the desired units of mea­sure and press Enter.
Clearing Stored Data
To empty stored data, click Clear Stored Data.
When viewing in realtime, the new units will be displayed at the next update interval. The standard unit of measure will be desig­nated with a 1 in the Scale column. When selecting a different unit of measure, make sure the conversion from the standard is correct.
FAST READ MODE
Choose Fast Read Mode when:
1. You need a sample rate of less than 8-seconds, and /or
2. You want to delay the logger’s start time but do not want to specify the start time until after the logger has been setup.
Choosing a Sample Rate
The sample rate is the frequency that the logger takes readings. The current sample rate is always displayed. to change the sam­ple rate: 1. Click the current rate. The sample rate dialog box appears. The dialog box displays current sample rate, the num­ber of active channels, and the length of time it will take to fill the memory using the number of active channels and rate. (To determine the loggers capacity, enable the number of channels required before choosing a sample rate). 2. Click the down arrow to see available sample rates from 8 seconds to 8 hours in Continuous and Stop When Full modes. Scroll if necessary, and click on the desired rate. In Fast Read mode, sample rates less than 8 seconds are available. 3. Click Accept. The new sample rate will apply to all active channels. If you do not want to change the sample rate, click cancel.
Delaying Starting Time
Make sure the correct time is displayed in the logger window before using this function. Once the logger has been setup, click Save Setup, and the logger will start gathering data. To delay the start time: 1. Double click in the Start Delay box to open the Select a Time dialog box. 2. Set the time and/or date for the logger to start taking readings. 3. Click Accept. The logger can be set for a delayed start time of up to two years.
Number of Readings
The maximum number of readings the DL7 logger can store is 21,500. If fewer readings are desired, change the number (in increments of 500).
Fast Read Mode allows you to choose sample rates from 0.04 seconds (25 readings per second) to 2 minutes. The sample rate must be in increments of 0.04 seconds. NOTE: A separate power source is required when using Fast Read Mode. Power can be supplied from the computer the logger is connected to or a separate battery pack (BP-100). If using the battery pack, the logger will start taking reading as soon as the battery is connect­ed.
The logger cannot work in continuous Mode when using Fast Read Mode. The logger will stop taking readings as soon as the memory is full. The logger will also interrupt readings when it is connected to Trendreader software.
To delay start in Fast Read Mode:
1. Set the delay time in seconds in the start Delay box. Use this method when the logger will remain connected to a computer or battery pack.
2. If the battery pack will be connected at a later time, follow this procedure before disconnecting the logger from the computer: Before clicking Save Setup, set Start Delay for long enough that you can save the setup and disconnect the logger while it is asleep. The logger must be asleep (on start delay) when it is dis­connected to be able to start taking readings when the battery is connected to it. NOTE: You must set a minimum start delay of 16 seconds before you disconnect the logger from the computer. the logger must be disconnected from the computer within 8­seconds of setting the delay. If these steps are omitted, the log­ger will not start when the battery is connected.
To calculate the number of readings a particular time span will need, divide the time span by the sample rate (making sure both are in seconds or minutes). If you are taking readings at 8-sec­ond intervals for 24-hours, and need to know the number of readings, convert 24-hours to seconds (86400) and divide by the sample rate of 8-seconds to get 10,800 readings. When back­ing up the logger, it will stop at the specified number of readings.
Saving Changes
When you have finished setting up your logger, click Save Setup. A dialog box will appear asking to clear the logger and save setup. (Setting up a logger erases all the previously gathered data). Click OK.
Do not contact a logger in Fast Read Mode until you are going to back it up. Contacting the logger will stop it from taking read­ings. You can contact a logger by clicking Contact or by open­ing the window once it has been closed.
BACKING UP A LOGGER
Copying data from a logger to your computer requires an IC-100 or IC-200 Interface cable. to backup a single logger use the IC­100 Interface Module. The IC-200 Interface Module can be used to backup several loggers. Backing up a logger can take a num­ber of minutes depending on the speed of your computer and the amount of data being transferred.
Loading...
+ 5 hidden pages