DIGIWEIGH DWP-98 User Manual

Page 1
DWP-98 Series
Counting scale
Users Guide
PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT
Specifications subject to change without prior notice
Page 2
CONTENTS
1. INSTALLATION
2. SPECIFICATIONS
3. KEYS, DISPLAY AND CONNECTIONS
4. GETTING STARTED
5. INITIAL SETUP
6. INSTRUCTION FOR USE
7. RS232 DATA OUTPUT
8. BATTERY POWER AND RECHARGING
9. ERROR CODES
10. DAILY CARE AND MAINTENANCE
Page 3
Appendix A: - About Bi-Directional Communication
INSTALLATION
Because of metrological legislation, installation/some metrological parameter settings are limited to be done by authorized personnel only. Do not attempt to change any of the built-in metrological parameters. Contact your dealer for more information and technical assistance.
To ensure performance accuracy, do not use the instrument in where or when the environment condition falls beyond as those listed on SPECIFICATIONS.
Do not attempt to open the instrument, no user serviceable parts inside.
Page 4
Interface
RS232
Stabilisation
time
2 second typical
Operating
temperature
0°C - 40°C / 32°F - 104°F
Power supply
220~240V AC (110V optional) 50/60Hz
built-in rechargeable battery 6V4AH.
Display
3 windows 6 digits 0.8’LCD display.
and 11 status indication
Zero range
0mV~8mV
Housing
ABS housing and Stainless steel pan(230*290mm)
ADC
Σ-Δ
External
Resolution
1/30000; 1/60000
Interface
RS-232 Output
Other function
Real-time clock and units conversion (kg,lb)
Model#
DWP-98CDH
DWP-98CBH
DWP-98CAH
Max Capacity
3kg
15kg
30kg
Readability
0.05g
0.2g
0.5g
1. SPECIFICATIONS
2.1 GENERAL SPECIFICATIONS
DWP-98 Series(resolution 1:60000; unit :kg/lb)
Common Specification
Page 5
2. KEYS, DISPLAY AND CONNECTIONS
1. UNIT PIECE WEIGHT SET KEY Press this key to confirm the unit piece weight entered by numeric keys.
2. WEIGHT UNIT KEY
Press this key to shift among various weight units.
3. M+ KEY
Press this key to accumulate the current quantity to memory or to recall the accumulated total quantity when unloaded.
4. TARE KEY
Press this key to tare off the weight of a container. Press this key to enter parameter
5. SAMPLE QUANTITY SET KEY
Press this key to confirm the sample size (pieces) entered by numeric keys.
6. CHECK KEY
Press this key to set upper & lower quantity check limit and to trigger the
Page 6
quantity check mode. Press this key as next parameter when setting
7. PRINT KEY
Press this key to send print data to a printer. Press this key as last parameter when setting
8. ZERO KEY
Press this key to set weight displayed to zero when an empty scale has drifted away from a true zero reading.
9. CE KEY Press this key to clear value entered.
10. NUMERIC KEYS
Numeric keys 0 ~ 9 and decimal key.
11. BATTERY POWER / LEVEL INDICATOR
This indicator appears to show:-
The instrument is running on the built-in rechargeable battery, It shows the remaining battery level.
12. SPARE
No assignment.
Page 7
13. ZERO INDICATOR
Visible when instrument at true zero weight status.
14. NET INDICATOR Visible when tare function is in effect. The weight being shown on Weight Panel is net weight.
15. STABLE INDICATOR
Visible when weight reading is stable.
16. WEIGHT UNIT INDICATOR
kg = kilogram, g = gram, lb = pound
17. UNIT PIECE WEIGHT INDICATOR
The value left to it is the unit piece weight. This indicator remains visible during normal operation status.
18. INSUFFICIENT SAMPLE WEIGHT CARNING INDICATOR
Visible when total weight of the sample size is below satisfactory level. If possible, increase the sample quantity unit this indicator disappears.
19. INSUFFICIENT UNIT PIECE WEIGHT CARNING INDICATOR
Visible when the unit piece weight of the sample is below satisfactory level. If possible, increase the sample quantity.
20. TOTAL INDICATOR
Visible when total accumulated quantity is being displayed.
21. QUANTITY INDICATOR
The value left to it is the quantity of the current transaction. This indicator remains visible during normal operation status.
22. QUANTITY CHECK RESULT INDICATORS
HI = Quantity is higher than the HI limit set. OK = Quantity is in between than the low and HI limits set. LO = Quantity is lower than the LO limit set.
Page 8
A. WEIGHT PANEL
Weight value is shown here.
B. UNIT PIECE WEIGHT PANEL
Unit piece weight is shown here.
C. QUANTITY PANEL
Quantity value is shown here.
D. BUBBLE LEVEL
Refer to this bubble to obtain a horizontal position for the instrument.
E. CHARGE STATUS INDICATOR
Red = Recharging battery; Green = Charging completed.
F. ON/OFF KEY Press this key to turn the instrument on or off.
G. DC JACK INPUT
External power adaptor is plugged in here.
H. RS232 COMPORT
9 pin (DB9) RS232 interface output for connection with computer and other peripherals.
Page 9
4. GETTING STARTED
In order to obtain accurate weighing and counting result, the instrument must be placed on a strong and level surface horizontally.
Avoid using the instrument in environment where excessive wind flow, vibration and extreme temperature change exist.
The instrument should be installed CA from any sources of excessive electrical noise. For full EMC or for RFI immunity, termination of cable shields and correct grounding of the instrument is essential.
General warning: -
The instrument is not an explosion proof device. The instrument is not a water proof device. Do not open the instrument, no user serviceable parts inside.
Always contact your dealer for service.
The instrument not to be subject to shock, excessive vibration
or extremes of temperature (before or after installation).
4.1 BUILT-IN RECHARGEABLE BATTERY
The instrument is equipped with a built-in rechargeable battery. Before first time use, recharge it for at least 8 hours to ensure the best battery performance.
4.2 POWER ADAPTOR
Before plugging in the power adaptor, check and make sure the input voltage of the adaptor matches with output voltage of the electricity outlet. If not, contact your dealer immediately.
4.3 CONNECTING OTHER DEVICES
AlCAys turn the instrument off before making any connections or disconnections with external devices.
Page 10
4.3.1. Connecting with a Computer
RS232 COMPORT
ON INSTRUMENT
COM PORT ON COMPUTER
(DB9)
(DB25)
2 = RXD
3 = TXD
3 = TXD
3 = TXD
2 = RXD
2 = RXD
5 = GND
5 = GND
7 = GND
RS232 COMPORT ON INSTRUMENT
DB25 COMPORT ON
PRINTER
2 = RXD
3 = TXD
3 = TXD
2 = RXD
5 = GND
7 = GND
4.3.2 Connecting RS232 to Printer (DB25)
5. INITIAL SETUP
5.1 INTERNAL SETTINGS
Preferred application parameters can be checked or set through internal functions. Refer to 5.4 for internal function description and setting procedures.
5.2 HOW TO ENTER AND SELECT INTERNAL FUNCTION
Follow the below steps to enter and select desired parameter of an internal function.
a. Turn the instrument off and on again, b. Press [TARE] during countdown,
Page 11
c. The instrument displays F1, d. The instrument is now in internal function,
5.3 KEY FUNCTION DURING INTERNAL FUNCTION SETTING
[TARE] = Enter, save and return, [ZERO] = Quit without saving,  [CHECK] = Go next,  [PRINT] = Go previous,  [CE] = Clear,
Page 12
Page 13
Page 14
Page 15
6. INSTRUCTION FOR USE
6.1 POWER ON
Powered on the instrument, instrument displays: -
a. softCAre and revision number and capacity of instrument, b. calibration count value, c. parameter set count value, d. all LCD segments,
Then countdown process starts. After that, the instrument is ready for operation.
6.2 START WEIGHING
a. If zero weight cannot be obtained when unloaded, press [ZERO]. After
[ZERO] is pressed, Zero Indicator appears. Refer to SPECIFICATIONS for maximum zero range,
b. AlCAys place an object onto platter gently. Excessive force applied to
platter may cause damages to the weight sensor inside instrument, c. Weight of the object is displayed on thus unit automatically, d. It is a good practice to remove all loads from platter after weighing. It
will prolong the life of the weight sensor inside instrument.
6.3 ABOUT WEIGHT UNIT CONVERSION
The instrument supports conversion among weight units. Press [UNIT] to
shift between kg, g and lb.
To enable or disable a particular weight unit, set it ON or OFF respectively
in internal function F9.
The weight unit being employed before power off will be employed when turned on again.
Page 16
6.4 TARE OFF THE WEIGHT OF A CONTAINER1
Tare function is used to temporarily set the instrument to zero (such as cancelling the weight of a box or a container) in order to get the net weight result
6.4.1 Enable / Disable Repeat (Multiple) Tare
Depends on internal function setting, repeated (multiple) tare operation may be enable (Mode 1) or disabled (Mode 2). Contact your dealer for more information.
When repeat (multiple) tare is enabled: -
1. the instrument will permit multiple tare operations provided that
both of the below requirements are fulfilled: - The tare operation does not permit a reduction of the value of
the tare;
The tare effect can only be cancelled when there is no load on
the platter.
2. tare effect can only be cancelled when container is removed and gross weight is zero.
6.4.2 Manual Tare2
When a container is used, follow the below steps to tare the weight of the container off to get a net weight result.
a. Remove all loads from platter, b. Make sure that the Zero Indicator is on. If not, press [ZERO] to set
weight reading to zero, c. Place container on platter, d. Press [TARE], e. Net Indicator appears to indicate tare is in effect and weight reading
is net weight. Refer to SPECIFICATIONS for maximum tare range,
The tare weight is deducted from the weighing capacity (Max), reducing the
maximum weight that can be displayed.
a. The tare operation does not permit a reduction of the value of the tare;
b. The tare effect can only be cancelled when there is no load on the platter.
Page 17
f. To cancel tare effect, remove all loads and container from platter and
press [TARE] , g. Net Indicator disappears. Tare effect has been removed and weight
reading is gross weight.
6.4.3 Preset Tare
This feature enables the user to manually enter the tare weight. During normal operation, enter the pre-determined tare weight through numeric keys then press [TARE]. This pre-determined tare value will be deducted.
To cancel the preset tare effect, remove all loads from platter then press
[TARE].
NOTE: -
1. Refer to SPECIFICATIONS for maximum tare range.
2. Manual tare is possible when preset tare is in effect. Manual tare
will cancel the effect of preset tare.
6.5 QUANTITY COUNTING
Quantity counting starts with determining the unit piece weight either by keyboard entry or sampling process.
6.5.1 Entering Unit Piece Weight3 4
This method is used where unit piece weight is known and is highly standardized.
a. Refer to 6.2 to 6.4 for zero, selecting preferred weight unit and tare
operation, b. Enter the unit piece weight and press [@WT]. The unit piece weight
is now displayed on the Unit Piece Weight Panel. c. Place a load on the platter. The weight of the load is displayed on the
Weight Panel and the quantity is displayed on the Quantity Panel.
For best counting result, refer to 2.2 for recommended minimum unit piece weight
and weight applied.
When the individual unit piece weight is not standardized, it is strongly
recommended that counting procedures as described in 6.6 should be employed.
Page 18
d. To enter new unit piece weight, repeat step b above.
6.5.2 Sampling Process This method is used where unit piece weight is unknown or the weight of individual piece is not highly standardized.
a. Refer to 6.2 and 6.4 for zero, selecting preferred weight unit and tare
operation, b. Place a sample size with known quantity on platter c. Enter the quantity of the sample through the numeric keypad and
confirm by pressing [# SET]. d. The instrument will automatically determine the unit piece weight and
display it on the Unit Piece Weight Panel5. e. Step by step add more load onto the platter (or remove part of the
load from the platter), f. The latest Weight, unit piece weight and total quantity would be
displayed on the corresponding panels, g. To start a new sampling, repeated step a to d above.
6.6 ABOUT AUTO PIECE WEIGHT ENHANCEMENT FUNCTION
In order to obtain the best counting result accuracy, the instrument is equipped with Auto Unit Piece Weight Enhancement Function. This function will automatically come into effect when an unit piece weight is obtained through the sampling method as described in 6.5.2.
6.6.1 How Auto Unit Piece Weight Enhancement Function Works
After a unit piece weight is obtained by methods as described in 6.5.2, then place more loads onto the platter. The new quantity will be shown on the
Total Count Panel. And at the same time, Auto Unit Piece Weight Enhancement Function will update the unit piece weight if both below
requirements are fulfilled: -
a. The quantity added to platter is more than 1 piece, and
At this point, sampling process is completed and Auto Piece Weight
Enhancement Function comes into effect.
Page 19
b. The quantity added to platter is less than 100% of the maximum
counts previously attained from the same transaction.
If the above requirements are fulfilled, a new unit piece weight will be determined and displayed on the Unit Piece Weight Panel and confirmed by 2 audio beeps.
Auto Unit Piece Weight Enhancement Function will terminate when: -
a. a zero weight is detected or during the transaction process, or b. quantity added is more than 100% of the maximum counts previously
attained from the same transaction, or c. quantity removed is not more than 50% of the maximum counts
previously attained from the same transaction.
To start Auto Piece Weight Enhancement Function again after terminated, repeat 6.5.2.
6.7 MEMORY ACCUMULATION FUNCTION
6.7.1 To Accumulate a Transaction to Memory6 7 8
a. When quantity is being displayed on Quantity Panel, press [M+]9 to
and accumulate data of current transaction to memory,
b. The instrument: -
displays “ACC” on Weight Panel. ACC denotes data has been
accumulated to memory,
displays “n” on Unit Piece Weight Panel. n donates the number
of transactions which have been accumulated to memory.
displays Total Indicator and total accumulated value on
Quantity Panel.
Transaction with zero quantity will not be accumulated. Transaction with weight less than 20d will not be accumulated. All data stored will be erased when the instrument is powered off or changing
weight unit.
Unstable weight will not be accumulated to memory. If M+ is pressed when weight
is unstable, the instrument will reject this command and response with 2 beeps.
Page 20
output transaction printout(s)10 through the RS232 comport. The
10
11
12
instrument returns to normal operation status after 2 seconds,
c. Repeat a for subsequent transactions11,
6.7.2 Memory Recall and Clearance a. Remove all loads from platter, b. Press [M+] to recall total accumulated weight from memory, c. After [M+] is pressed, the instrument displays: - d. The instrument displays: -
ACC” on Weight Panel. n” on Unit Piece Weight Panel. n donates the number of
transactions which have been accumulated to memory.
Total Indicator and total accumulated value on Quantity Panel.
e. At this point: -
Press [ZERO] to quit or, Press [CE] to clear memory, then [ZERO] to return to normal
operation status.
6.8 QUANTITY CHECK FUNCTION12
6.8.1 To Enter High and Lo Quantity Limits
The instrument is equipped with quantity check function. Follow the below steps to trigger checkweighing mode: -
a. During normal operation, press [CHECK]. b. The instrument displays High Limit and current HI limit value is
displayed on Quantity Panel.
c. Press [TARE] to accept or enter a new HI limit through the numeric
keys and then press [TARE] to confirm,
d. The instrument then displays Lo Limit and current LO limit value is
displayed on Quantity Panel.
e. Press [TARE] to accept or enter a new LO limit through the numeric
keys and then press [TARE] to confirm,
The number of copies sent = the number of copy set in internal function F8. Refer
to 5. INITIAL SETUP for more information.
Weight reading must return to or below zero to enable another weight
accumulation.
Quantity check function will not function when weight reading is less than 20d.
Page 21
Hints: -
1. For normal comparison, set both HI and LO limits,
2. To check only if result is lower or equal to LO (result LO), set HI limit = 0,
3. To check only if result is higher or equal to HI (result HI), set LO limit = 0,
4. To check if result is equal to a specified value, set both HI limit and LO limit = the specified value
6.8.2 Visual and Audio Quantity Check Signals
6.8.2.1 Visual signal: - Quantity check result is displayed by one of the Check Result Indicators located on Quantity Panel.
6.8.2.2. Audio signal: - depends on setting of internal function F5, audio
signal can be triggered when result is = within range, out of range or turned off.
6.8.3 To Quit Checkweighing Mode
To quit checkweighing mode, set both HI and LO limits to zero.
Page 22
7. RS232 DATA OUTPUT
DATA BIT
DESCRIPTION
1~2
MOTION STATUS
US = UNSTABLE ST = STABLE
3
COMMA SEPARATION
4~5
NET/GROSS
NT = NET WEIGHT GS = GROSS WEIGHT
6
SIGN (Sign of weight reading) Positive = space. Negative = minus (-)
7~13
WEIGHT VALUE
7-character string containing the current weight including location of decimal point.
If there is no decimal point, then the first character is a
13
Data output parameters are fixed as below: -
Data Bit = 8  Parity = None  Stop Bit = 1
There are 2 data output modes (PC and Prt) are available13. PC is for communication with computer and other peripherals which accepts and processes continuous data communication. Prt (printer) is for transmission to printer or other peripherals which accept only single or manual data transmission.
Baud rate has to be set before proceeding to other settings.
7.1 PC (COMPUTER) MODE
Weight status and data are sent under the PC (Computer) Mode). Data is transmitted in ASCII code. Data format is listed on below table.
Refer to 5. INITIAL SETUP for setting information
Page 23
space.
14
COMMA SEPARATION
15~16
UNIT kg = kilogram g_= gram (_ = blank space) lb = pound
17
Cr
18
LF
14
15
7.2 PRT (PRINTER) MODE
7.2.1 Printing Current Transaction
A standard printout as will be sent through the RS232 Comport when [PRINT]14 is pressed. Refer to Standard Printout Sample on next page
for description.
Number of copies sent = the number of copy set in internal function F8. Refer to 5 INITIAL SETUP for more information. If extra copies are needed, press [PRINT] again.
7.2.2 Printing the Totalized Accumulated Result
Follow the below steps for to print the totalized accumulated result: -
a. Remove all loads from platter, b. If zero weight reading cannot be obtained when unloaded, press
[ZERO]
c. Press [M+] then followed by [PRINT], The below data and format
will be sent15,
Refer to Totalized Accumulated Printout Sample on next page for description.
Pressing [PRINT] will not accumulate transaction to memory. Press [M+] to
trigger memory accumulation and print function simultaneously
The number of copies sent = the number of copy set in internal function F8. Refer to 5. INITIAL SETUP for more information. If extra copies are needed, press [PRINT] again.
Page 24
Standard Printout Sample Description
16
TIME 12:08:57 DATE 07.08.2010 No. 0 TARE 2.000kg GROSS 6.000kg NET 4.000kg UNIT WT. 0.5 g Q'ty PCS
HIGH 5000PCS LOW 4500PCS HIGHER THAN LIMIT
TIME 13:48:23 DATE 07.08.2010 No. 4 TOTAL 14997PCS
Totalized Accumulated Printout Sample Description
8. BATTERY POWER AND RECHARGING
Remaining battery power of the built-in rechargeable battery is displayed on the Battery Power / Level Indicator.
When this transaction is accumulated to memory.
**
Visible when quantity check function is in effect.
1. Time of printing,
2. Date of printing,
3. M+ sequent number16
4. Tare Weight,
5. Gross Weight,
6. Net weight,
7. Unit piece weight,
8. Quantity,
9. One blank line,
10. One blank line,
11. HI limit set**
12. LO limit set**
13. Quantity Check result**
1. Time of printing,
2. Date of printing,
3. Total No. of transaction
4. Total accumulated quantity
Page 25
8.1 SYMBOLS AND REMAINING POWER: -
Error
Code No.
Description
Full Battery: 6.3V 2 Blocks: 6.0V 1Block: 5.7V Frame only: 5.7V
8.2 BATTERY OPERATION TIME
Depends on the battery condition, a fully charge new rechargeable battery will usually give: -
Around 200 hours of continuous operation without backlight, or Around 70 hours of continuous operation with backlight on, or
8.3 RECHARGE BATTERY
When appears, (when battery is less than 5.7V), it means the built-in rechargeable battery is at low voltage status. It is recommended to recharge immediately.
To protect the built-in rechargeable battery, the instrument will power off automatically when battery voltage is at extremely low level (when battery is at around 5.4V). Do not attempt to power the instrument on. Recharge the instrument immediately. Fail in doing so may cause unrecoverable damage to the built-in rechargeable battery.
Battery charging status is shown on the dual color Charge Status Indicator:
-
Red: - Recharging in process, Green: - Charging is completed.
Battery recharge is possible while instrument is in operation. Overcharge protection circuit is built inside to prevent battery damages caused by overcharge.
Heat will be generated during recharging and it is normal to feel minor heat at front panel of the instrument.
9. ERROR CODES
Page 26
Err 1
Time value error
Err 2
Date value error
Err 3
Exceed maximum manual zero range
Err 4
Offset out of range / unstable during power on
Err 5
No load cell signal detected
Err 6
Tare operation error
Err 7
Logic error. HI limit set is lower than LO limit (and HI is not = 0)
Err 8
Logic error. LO limit is higher than HI limit (and HI is not = 0)
--oL--
Overload (Gross weight is more than Max plus 9d)
UndEr
Under load (Gross weight is less than minus 20d)
10. DAILY CARE AND MAINTENANCE
Clean the instrument with a soft, damp cloth. If necessary, use a mild
detergent in CAter,
Do not use any harsh, abrasive material, acetone, volatile solvent,
thinner or alcohol for cleaning,
Verify the accuracy of this unit periodically. Re-calibrate this unit if
necessary. In some countries, calibration requires authorized / qualified agent. Contact your dealer for more information,
Store this unit in a dry and clean place, Recharge battery before and every 2 months during long time
storage.
Appendix A: -
Page 27
About Bi-Directional Communication
Key on Instrument Panel
Equivalent to
Computer Command
1
1
2
2
3 3 4 4 5 5 6
6
7 7 8
8
9 9 0
0
.
(Decimal)
.
(Decimal)
ZERO
A
TARE
B
# SET
C
CHECK
D
CE
E
@WT
F
PRINT
G
UNIT
H
M+
I
Computer commands can be sent to this instrument through any standard communication program like Hyper Terminal which comes with Windows 2000 and Windows XP.
Equivalent commands from computer are illustrated on below table for each of the key on panel.
Appendix B: -
Page 28
Calibration
Turn on the scale and press/hold the tare key for 4 sec then Let go and it will display F1, with the “Print” button go to F10 Cal Press Tare to enter ,show P----; the initial pass ward is 1234 . (Alarm for next parameter; Print for last parameter ) Press Tare to enter to
“P1 Cap” (capacity : 3kg,5kg,6kg,10kg,15kg,20kg,30kg) “”P2 Inc” (1,2) “P3 Zero track ” (off, 0.25, 0.5, 1, 1.5 , 2.0) “P4 Auto zero” (off,1,2,4,5,10,20)” “P5 key zero” (off,1,2,4,5,10,20)” “P6 Filter” (1,2,3,4) “P7 G1” gravity “P8 G2” gravity “P9 User Cal” user calibration, Press “tare” to enter “Cal unit “ : choose kg, lb as calibration unit
Press tare to enter, it show unload. Clear platform ,pressTare Show 000000, enter weight value with numeric keys. Add weight Press “tare to show Ld1 ; when stab on, press Tare to succeed calibration
P9 linear Cal Press tare to enter Cal unit :
choose kg, lb as calibration unit Press tare to enter, it show unload. Clear platform , press Tare Show 000000, enter weight value with numeric keys. Add weight Press “tare to show Ld1 ; when stab on, enter, it show unload. Clear platform , press Tare Show 000000, enter weight value with numeric keys. Add weight
Press “tare to show Ld2” ;
P11 Pin” pass ward setting ; press” tare ” to show “P1----’;
Enter pass ward with numeric keys Press tare to show P2---- Enter pass ward again with numeric keys Press Tare to done the linear calibration
P12 tare cont ,press tare to enter mode 1 or mode 2” “P13 Cal count, press tare to choose pin or Jumper way to enter
calibration mode