PATHOLOGY
Hybridizer Quick Start Guide
Hybridizer (120 V), Code S2450,
Hybridizer (240 V), Code S2451 &
Humidity Control Strips, Code S2452
Main Menu |
|
Use arrow keys to move cursor up and down
Use arrow keys to scroll through character set
A-Z when naming program
Press“Enter” button to accept
|
Enter numeric |
|
values for time and |
||
|
||
|
temperature |
|
|
Enter numeric |
|
|
values in program |
|
|
name |
|
Press “STOP” |
|
button to abort a |
||
|
||
|
program in process |
Press “Backspace” button to return to previous field or previous screen
Installation of the Hybridizer
nPlace the Hybridizer on a level surface suitable for laboratory instrumentation.
nThe Hybridizer has an intake fan located at the bottom and an outlet fan at the rear; assure no obstructions exist on intake and outlet.
Note: Avoid loose paper under the Hybridizer to prevent obstruction.
nPosition the Hybridizer away from direct sunlight and sources of heat or cold.
nVerify voltage requirements located on serial No. label, rear of instrument.
Connect Power and Turn on the Hybridizer
nPlug the Hybridizer into the appropriate outlet supplying the voltage and frequency indicated on the serial No. label.
nMove switch to “ON” position. The main power switch is located on the rear panel (I = On,
O = Off).
nInstrument will beep on start-up and the initialization screen will appear.
nThe Main Menu will be displayed.
Please use the system as intended. Improper use of the Hybridizer may cause damage to the system, inaccurate results, potentially nullify warranties, or cause personal injury.
Display Abbreviations
Humidity Control Strips
Slide Location
Select Program Type
S e l e c t P G M Ty p e
D e n a t & H y b
H y b O n l y
F i x e d Te m p
PGM: |
Program |
Denat & Hyb: |
Denaturation & Hybridization |
Denat Temp: |
Denaturation Temperature |
Denat Time: |
Denaturation Time |
Hyb Temp: |
Hybridization Temperature |
Hyb Time: |
Hybridization Time |
Hyb Only: |
Hybridization Only |
Fixed Temp: |
Fixed Temperature |
Heating to Fxd: |
Heating to Fixed (-- °C) |
|
|
nMinimize evaporation of sample from prepared slides.
nTo install strips, lift lid and insert strips into slot positions.
nSaturate strips with distilled water or equivalent.
nStrips should be replaced frequently as they will deteriorate over time and with use.
nAccommodates up to 12 slides per run.
nThe Hybridizer prompts to add slides.
nLift lid and load slides onto plate using slide locator for easy positioning.
nFrosted and/or label end of slides should be away from slide locator.
nMove slides across plate until they rest against slide locator.
Note: If slides are not resting against slide locator, closing of lid may break slides.
nThe Hybridizer is capable of storing 40 programs. Each program can be one of three program types:
-Denaturation and Hybridization
-Hybridization Only
-Fixed Temperature
nSelect program type and follow screen prompts to enter run times and set temperatures.
Abort Program in Process
A B O RT I N G ! !
A re Yo u S u re ?
N o
Ye s - M a i n M e n u
nPress “STOP” button to abort a program in process.
nUse arrows to move cursor to “Yes” line.
nPress “Enter” button to accept.