Hybridizer  Quick Start Guide
Hybridizer (120 V), Code S2450, Hybridizer (240 V), Code S2451 & Humidity Control Strips, Code S2452
Main Menu
Enter numeric
values for time and temperature
Enter numeric values in program name
Use arrow keys to move cursor up and down
Use arrow keys to scroll through character set A-Z when naming program
Press “Enter” button to accept
Installation of the Hybridizer
n Place the Hybridizer on a level surface suitable
for laboratory instrumentation.
n The Hybridizer has an intake fan located at the
bottom and an outlet fan at the rear; assure no obstructions exist on intake and outlet. Note: Avoid loose paper under the Hybridizer to prevent obstruction.
n Position the Hybridizer away from direct sunlight
and sources of heat or cold.
n Verify voltage requirements located on serial
No. label, rear of instrument.
Press “STOP”
button to abort a program in process
Press “Backspace” button to return to previous eld or previous screen
Connect Power and Turn on the Hybridizer
n Plug the Hybridizer into the appropriate outlet
supplying the voltage and frequency indicated on the serial No. label.
n Move switch to “ON” position. The main power
switch is located on the rear panel (I = On, O = Off).
n Instrument will beep on start-up and the
initialization screen will appear.
n The Main Menu will be displayed.
Please use the system as intended. Improper use of the Hybridizer may cause damage to the system, inaccurate results, potentially nullify warranties, or cause personal injury.
Page 2
Display Abbreviations
Humidity Control Strips
PGM: Program
Denat & Hyb: Denaturation & Hybridization
Denat Temp: Denaturation Temperature
Denat Time: Denaturation Time
Hyb Temp: Hybridization Temperature
Hyb Time: Hybridization Time
Hyb Only: Hybridization Only
Fixed Temp: Fixed Temperature
Heating to Fxd: Heating to Fixed (-- °C)
n Minimize evaporation of sample from prepared slides. n To install strips, lift lid and insert strips into slot positions. n Saturate strips with distilled water or equivalent. n Strips should be replaced frequently as they will deteriorate
over time and with use.
Slide Location
Select Program Type
S e l ec t   PG M  Ty p e
D e n at   &  H yb
H y b  O n l y
F i x ed   Te m p
Abort Program in Process
n Accommodates up to 12 slides per run. n The Hybridizer prompts to add slides. n Lift lid and load slides onto plate using slide locator for
easy positioning.
n Frosted and/or label end of slides should be away from
slide locator.
n Move slides across plate until they rest against slide locator.
Note: If slides are not resting against slide locator, closing of lid may break slides.
n The Hybridizer is capable of storing 40 programs. Each
program can be one of three program types:
- Denaturation and Hybridization
- Hybridization Only
- Fixed Temperature
n Select program type and follow screen prompts to enter
run times and set temperatures.
A B O RT IN G ! !
A r e  Y o u  S u r e ?
N o
Ye s  -  M a in   M en u
n Press “STOP” button to abort a program in process. n Use arrows to move cursor to “Yes” line. n Press “Enter” button to accept.
Page 3
Create a Denaturation and Hybridization Program (Denat & Hyb)
1. Select Create a PGM 2. Select Program Type
Ru n a  PG M
Ed it  a P GM
Cr eat e a  PG M
Pr ese nt  Tem p:  37  °C
n Use the arrow keys to move
cursor to “Create a PGM”
n Press “Enter” button to accept
Se lec t P GM  Type
De nat  &  Hyb
Hy b O nly
Fi xed  Temp
n Cursor highlights “Denat &
Hyb” line
n Press “Enter” button to accept
3. Create a Name 4. Denaturation Temp & Time
Cursor highlights the first
PG M 0 1  Nam e
En ter :  Temp  Ti me
De nat :  °C   :0 0
Hy b:  °C  00:0 0
Note: All 10 character positions must be filled.
6. Save a Program 7. Create a Hyb Only or Fixed Temp Program
De nat :  82  °C   :0 5
Hy b:  45  °C  2 0:0 0
EN TER  to  Ac cep t
ST OP  to  Abo r t
n
name character position
n Use the arrow keys to scroll
through character set Press “Enter” button to accept
n
characters one by one
n Press “Enter” button to accept
blank characters
n
Use keys 0-9 to enter numeric characters
n Display will now show entered
program values
n Cursor highlights “Enter to
Accept” line
n Press “Enter” button to accept
program values
n Display indicates “Program
Saved”
n Cursor will advance to
“Denat Temp”
n Enter a two-digit temperature
value (50-99 °C) Cursor will advance to
n
“Denat Time”
n Enter a two-digit time
value in minutes (0-30)
n Repeat procedure to create “Hyb Only” or “Fixed Temp” program n Follow screen prompts to enter run times and/or set temperatures n See Handbook for more detailed instructions
5. Hybridization Temp & Time
n Cursor advances to “Hyb
Temp”
n Enter a two-digit temperature
value (30-70 °C). Enter “00” for room temperature hybridization
n Cursor advances to “Hyb Time” n Enter a two-digit time value in
hours (0-99) followed by a two-digit value in minutes (0-59)
Edit a Program
1. Select Edit a PGM 2. Select a PGM
Ru n a  PG M
Ed it  a P GM
Cr eat e a  PG M
Pr ese nt  Tem p:  37  °C
3. Select Program Type 4. Enter New Values
Se lec t P GM  Type
De nat  &  Hyb
Hy b O nly
Fi xed  Temp
n Use the arrow keys to move
cursor to “Edit a PGM”
n Press “Enter” button to accept
n Cursor highlights existing
program type
n Move cursor with arrow keys
and press “Enter” to accept program type
En ter  PG M N o.
or  sc rol l ( arrow s)
PG M 0 1   HE R2 
PG M 0 4  Nam e H PV
En ter :  Temp  Ti me
De nat :  92  °C   :0 5
Hy b:  37  °C   0 1:0 0
n Scroll through program
Nos. 1-40/program names
n Press “Enter” button to accept
n Use numeric keys
to enter new values for temperatures and/or times
Page 4
Run a Denaturation and Hybridization Program (Denat & Hyb)
1. Select Run a PGM 2. Select a PGM
Ru n a  PG M
Ed it  a P GM
Cr eat e a  PG M
Pr ese nt  Tem p:  37  °C
3. Insert Humidity Control Strips and Add Slides 4. Heating to Denat Temp
PG M 0 1    HE R2
Ad d S lid es- Clo se  Lid
St art
Ma in  Men u
5. Denaturation 6. Cooling to Hyb Temp
PG M 0 1    HE R2
De nat  in  Pr oce ss
De nat : 8 2 ° C    02 :28
Pr ese nt  Tem p:  82  °C
7. Hybridization 8. Process Completed
PG M 0 1     H ER2
Hy b i n P roc ess
Hy b:  45  °C  1 2 :4 8
Pr ese nt  Tem p:  45 °C
n Cursor highlights “Run a
PGM” line
n Press “Enter” button to accept n Scroll through program
Nos. 1-40/program names
n Press “Enter” button to accept
n Insert two Humidity Control
Strips into the inside slide lid
n Saturate strips with distilled
water or equivalent
n Add slides and close lid n Cursor highlights “Start” line
n The Hybridizer will beep twice
once temperature reaches denaturation set point
n Denaturation time will count
down from the set time
n Hybridization time will count
down from the set time once temperature reaches hybridization set point
PG M 0 1   HER 2
82  °C  :0 5;  45  °C  20: 00
Ru n P GM 
Ma in  Men u
PG M 0 1   HER 2
** Hea tin g**
De nat : 8 2 ° C  :0 5
Pr ese nt  Tem p:  45  °C
Pl eas e Wai t
Co oli ng  to  Hyb  45  °C
Pr ese nt  Tem p:   58  °C
PG M 0 1    HE R2
PR OCE SS  COM PLE TE
Total  Hy b T ime  20 :00
En d P GM/ Mai n M enu   
n Display will conrm PGM No./
name and Denat & Hyb times and temperatures
n Cursor highlights “Run PGM”
line
n Press “Enter” button to accept
n Press “Enter” button to run the
program
n Display indicates present
temperature of the slides
n The Hybridizer will
automatically cool to hybridization set temperature once denaturation is completed
n The Hybridizer will beep
ve times upon program completion
n Cursor highlights “End PGM/
Main Menu”
n Press “Enter” to end program n See Handbook for further
details and for running a “Hyb Only” or “Fixed Temp” program