Cosmed does not assume the liability for interpretation mistakes of this documentation or for casual or
consequential damages in connection with the provision, representation or use of this documentation.
No parts of this manual may be reproduced or transmitted in any form without the express permission of
COSMED Srl.
COSMED Software can be installed only in one device.
Excel is a registered trademark of Microsoft Corporation.
DBIII is a registered trademark of Bordland International Inc.
Lotus 123 is a registered trademark of Lotus Development Corporation .
Quark PFT is an electrical medical device designed to perform pulmonary function
tests. It is to be used by physicians or by trained personnel on a physician responsibility.
Caution: Federal law restricts this device to sale by or on the order of a physician.
This equipment has been conceived with the aim of providing an auxiliary instrument
allowing:
• the formulation of lung pathology diagnosis;
• important studies concerning human physiology;
• the collection of important information in sport medicine.
No responsibility attaches COSMED Srl for any accident happened after a wrong use of
the device, such as:
• use by non qualified people;
• non respect of the device intended use;
• non respect of the hereunder reported precautions and instructions.
Warnings
The device, the program algorithms and the presentation of measured data have been
developed according to the specifications of ATS (American Thoracic Society) and
ERS (European Respiratory Society). Other international references have been followed
when these were not available. All bibliography references are reported in Appendix.
The present handbook has been developed with respect of the European Medical Device
Directive requirements which sort Quark PFT within Class II a.
It is recommended to read carefully the following precautions before putting the device
into operation.
The precautions reported below are of fundamental importance to assure the safety of
all COSMED equipment users.
1. This user manual is to be considered as a part of the medical device and should
2. Safety, measure accuracy and precision can be assured only:
• using the accessories described in the manual or given with the device. Actually
• ordinary equipment maintenance, inspections, disinfection and cleaning are
• any modification or fixing is carried out by qualified personnel;
• the environmental conditions and the electrical plants where the device operates
3. Before powering the system, check the power cables and the plugs. Damaged
4. Large gas cylinders, which may be given by the manufacturer or purchased by the
5. When removing the protective cap, inspect the cylinder valve for damaged threads,
6. Be certain that the materials of the pressure regulators are chemically compatible
always be kept on hand.
non recommended accessories can affect safety unfavourable. Before using non
recommended accessories it is necessary to get in touch with the manufacturer;
performed in the way and with the frequency described;
are in compliance with the specifications of the manual and the present regulations
concerning electrical plants. In particular grounding reliability and leakage current
suppression can only be assured when the device three – wire receptacle is
connected to a yellow - green return connected to earth ground. Attempting to
defeat the proper connection of the ground wire is dangerous for users and
equipment.
electrical parts must be replaced immediately by authorised personnel.
customer, should be secured with cylinder safety chains or safety stands.
dirt, oil or grease. Remove any dust or dirt with a clean cloth. If oil or grease is
present on the valve of a cylinder which contains oxygen, do not attempt to use.
Such combustible substances in contact with oxygen are explosive.
with the intended gas service before installation. Inspect the regulator for the proper
12 - Quark PFT User Manual
connection and note the ranges of the pressure gauges. Also examine the physical
condition of the regulator including threads and fittings. Remove any dust or dirt
from the regulator or cylinder valve with a clean cloth. Do not install a regulator on
a cylinder valve containing oxygen if grease or oil is present on either. Such
substances in contact with oxygen are explosive.
7. Cleaning residue, particulates, and other contaminates (including pieces of torn or
broken components) in the breathing circuit pose a safety risk to the patient during
testing procedures. Aspiration of contaminates can potentially be life-threatening.
You must follow all the cleaning procedures in System Maintenance, and you must
thoroughly inspect the components after cleaning and before each patient test.
8. This device is not suitable for use in presence of flammable anaesthetics. It is not
an AP nor an APG device (according to the EN 60 601-1 definitions).
9. Keep the device away from heat and flame source, flammable or inflammable
liquids or gases and explosive atmospheres.
10. In accordance with their intended use Quark PFT is not to be handled together with
other medical devices unless it is clearly declared by the manufacturer itself.
11. It is recommended to use a computer with electromagnetic compatibility CE
marking and with low radiation emission displays.
12. It is necessary to make the PC, connected to the
Quark PFT, compliant with EN
60601-1 by means of an isolation transformer.
13. Graphical symbols used in accordance to present specifications are described here
below:
Equipment type B (EN60601-1)
Danger: high temperature
OFF
ON
Protective earth ground
Alternating current
Chapter 1 - Getting started - 13
Contraindication
The physical strain to execute the respiratory manoeuvre is contraindicated in case of
some symptoms or pathology. The following list is not complete and must be
considered as a piece of mere information.
Contraindications for the Spirometer tests
Absolute contraindications
For FVC, VC and MVV tests:
• Post-operating state from thoracic surgery
For FVC tests:
• Severe instability of the airways (such as a destructive bronchial emphysema)
• Bronchial non-specific marked hypersensitivity
• Serious problems for the gas exchange (total or partial respiratory insufficiency)
Relative contraindications
For FVC tests:
• spontaneous post-pneumothorax state
• arterial-venous aneurysm
• strong arterial hypertension
• pregnancy with complications at the 3
For MVV test:
• hyperventilation syndrome
rd
month.
Contraindications for Bronchial provocation tests
The bronchial provocation tests must be executed according to the doctor’s discretion.
There are not data that reveal specific contraindication for the bronchial provocation test
through inhalation.
The modern standard processes have been revealing secure in several clinical studies.
However it is recommendable to respect the following contraindications:
Absolute contraindications
• Serious bronchial obstruction (FEV1 in adults)
• Recent myocardium infarct
• Recent vascular-cerebral accident
• Known arterial aneurysm
• Incapacity for understanding the provocation test procedures and its implications.
Relative contraindications
• Bronchial obstruction caused by the respiratory manoeuvre.
• Moderate or serious bronchial obstruction. For ex. Predicted value FEV1 less than
1.51 in men and predicted value FEV1 in women less than 1.21.
• Recent infection in the superior air tracts
• During the asthmatic re-acuting
• Hypertension
• Pregnancy
• A pharmacology treatment epilepsy
Contraindications for Exercise testing
Read carefully the exercise testing chapter.
14 - Quark PFT User Manual
Environmental condition of use
COSMED units have been conceived for operating in medically utilised rooms without
potential explosion hazards.
The units should not be installed in vicinity of x-ray equipment, motors or transformers
with high installed power rating since electric or magnetic interferences may falsify the
result of measurements or make them impossible. Due to this the vicinity of power lines
is to be avoided as well.
Cosmed equipment are not AP not APG devices (according to EN 60601-1): they are
not suitable for use in presence of flammable anaesthetic mixtures with air, oxygen or
nitrogen protoxide.
If not otherwise stated in the shipping documents, Cosmed equipment have been
conceived for operating under normal environmental temperatures and conditions [IEC
601-1(1988)/EN 60 601-1 (1990)].
• Temperature range 10°C (50°F) and 40°C (104°F).
• Relative humidity range 20% to 80%
• Atmospheric Pressure range 700 to 1060 mBar
• Avoid to use it in presence of noxious fumes or dusty environment and near heat
• Adequate floor space to assure access to the patient during exercise testing.
• Adequate ventilation in the room.
Chapter 1 - Getting started - 15
Safety and conformity
Safety
IEC 601-1 (1988) /EN 60 601-1 (1990);
Find reported below the complete classification of the device:
• Class I type B device
• Protection against water penetration: IP00, ordinary equipment unprotected against
water penetration
• Non sterile device
• Device not suitable in the presence of flammable anaesthetics;
• Continuous functioning equipment;
EMC
The system meets the EMC Directive 89/336
EN 60601-1-2
EN 55011 Class B (emission), IEC 1000-4-2, IEC 1000-4-3, IEC 1000-4-4
Quality Assurance
UNI ISO 9001 (Registration n° 387 Cermet)
Medical Device Directive (CE mark)
MDD 93/42/EEC (Notified Body 0476).
Class IIa
16 - Quark PFT User Manual
Keynotes
Here are the keynotes used to make the manual easier to read.
Typographic keynotes
These are the typographic keynotes used in the manual.
Style Description
Boldindicates a control or a key to be pressed.
“Italic” indicates a messages shown by the firmware.
Graphic keynotes
These are the graphic keynotes used in the manual.
Illustration Description
shows the button to click in the software to activate the related feature.
Chapter 1 - Getting started - 17
Systems Overview
Quark PFT is a stationary and modular system designed for Pulmonary Function
Testing. This system has been designed to meet the most accredited criteria and
recommendations of the latest scientific publications.
The system incorporates “plug and play” circuitry, allowing the customisation of
features and instant upgrades.
Quark PFT comes in the following configurations:
• PFT1; Spirometry, (FVC, SVC, MMV and bronchial-challenge tests).
• PFT2; adds Functional Residual Capacity testing via Nitrogen Washout and
Closing Volume techniques.
• PFT3; combines Spirometry plus Lung Diffusing Capacity testing by adding CO
and CH
• PFT4; integrates PFT2 and PFT3 features to become the Complete Pulmonary
Function Testing instrument.
• ERGO Option: The PFT 2 and PFT 4 can be enhanced with the Pulmonary Gas
Exchange module, allowing “breath by breath” analysis, indirect calorimetry and
RS232 ergometer control, PFT3 can be upgraded to PFT4 or PFT4 ERGO.
Tests/Products table
Test PFT 1 PFT 2 PFT 3 PFT 4 PFTErgoPFT2ergoPFT4ergo
Before using the system, please take a moment to fill in the registration form and the
warranty and return them to COSMED, by doing this you are eligible to the customers
assistance service.
For further information, please refer to the enclosed registration and warranty form. If
the form is not enclosed in the packaging, please contact directly COSMED.
Register the product via software
Together with the PC software, a registration software is supplied. With this software it
is possible to fill in an electronic form with the customer information.
1. To run the software, double click on the icon Registration or select Registration…
from ? menu.
2. Type the requested information and click Send… to send the form via e-mail to
COSMED.
How to contact COSMED
For any information you may need, please contact the manufacturer directly at the
following address:
COSMED S.r.l.
Via dei Piani di Monte Savello, 37
P.O. Box n. 3
00040 - Pavona di Albano
Rome - ITALY
With the auxiliary RS 232 port and the digital (TTL) input for external ECG signal,
Quark PFT offers the possibility to integrate data coming from many equipment such as
ECGs and ergometers.
26 - Quark PFT User Manual
Measurements
Pulmonary function tests and measured parameters
Spirometry Tests
FVC - Forced Vital Capacity
Symbol UM Parameter
FVC l Forced Expiratory Vital Capacity
FEV1 l Forced Expiratory Volume in 1 sec
FEV1/FVC% % FEV1 as a percentage of FVC
PEF l/sec Peak Expiratory Flow
FEV0.5 l Forced Expiratory Volume in 0.5 sec
FEV6 l Forced Expiratory Volume in 6 sec
FEV1/FEV6 % FEV1 as a percentage of FEV6
FEV6/FVC% % FEV6 as a percentage of FVC
Best FVC l Best Forced Expiratory Vital Capacity
Best FEV1 l Best Forced Expiratory Volume in 1 sec
Best PEF l/sec Best Peak Expiratory Flow
Vmax25% l/sec Expiratory Flow when 75% of the FVC remains to be exhaled
Vmax50% l/sec Expiratory Flow when 50% of the FVC remains to be exhaled
Vmax75% l/sec Expiratory Flow when 25% of the FVC remains to be exhaled
FEF25-75% l/sec Mid-exp flow between 25-75%FVC
FET100% sec Forced expiratory time
FEV2 l Forced Expiratory Volume in 2 sec
FEV3 l Forced Expiratory Volume in 3 sec
FEV2/FVC% % FEV2 as a percentage of FVC
FEV3/FVC% % FEV3 as a percentage of FVC
FEV1/VC% % Tiffenau index
FEF50-75% l/sec Mid-exp flow between 50-75%FVC
FEF75-85% l/sec Mid-exp flow between 75-85%FVC
FEF0.2-1.2% l/sec Mid-exp flow between 0.2 l - 1.2 l
FiVC L Inspiratory Forced Vital Capacity
FiF25-75% l/sec Forced mid-inspiratory flow
FiV1 l/sec Forced Inspiratory Volume in 1 sec
PIF l/sec Peak Inspiratory Flow
VEXT ml Extrapolated Volume (back extrapolation)
PEFT msec Time to PEF (10% - 90%)
VC/IVC - Slow Vital Capacity and Ventilatory pattern
Symbol UM Parameter
EVC l Expiratory Vital Capacity
IVC l Inspiratory Vital Capacity
ERV l Expiratory Reserve Volume
IRV l Inspiratory Reserve Volume
IC l Inspiratory Capacity
VE l/min Expiratory Minute Ventilation
Vt l Tidal Volume
Rf 1/min Respiratory Frequency
Ti sec Duration of Inspiration
Te sec Duration of Expiration
28 - Quark PFT User Manual
Ttot sec Duration of Total breathing cycle
Ti/Ttot —- Ti/Ttot ratio
Vt/ti l/sec Vt/ti ratio
MVV - Maximum Voluntary Ventilation
Symbol UM Parameter
MVV l/min Maximum Voluntary Ventilation
MVt l Tidal Volume (during MVV)
MRf 1/min Maximum Respiratory frequency
MVVt sec MVV duration time
Bronchoprovocation Response
Symbol UM Parameter
FallFEV1 % Fall in FEV1 from baseline or post diluent
FallVmax50% % Fall in Vmax50% from baseline or post diluent
P10 —- Provocative dose causing FEV1 to fall 10% from baseline
P15 —- Provocative dose causing FEV1 to fall 15% from baseline
P20 —- Provocative dose causing FEV1 to fall 20% from baseline
Multi-breath Nitrogen wash-out
Symbol UM Parameter
FRC l Functional Residual Capacity
RV l Residual Volume
TLC l Total Lung Capacity
Vtot l Total volume during the test
FetN2 % Final Nitrogen Concentration
Wash-out time m Duration of wash-out
VDA L Anatomic dead space
LCI mm:ss Lung Clearance Index
AMDN mm:ss Mean Dilution Number
FRC/TLC % FRC/TLC ratio
CEV l Cumulated Volume of inspired Oxygen
RV/TLC % RV/TLC ratio (Motley index)
Closing Volume test
Symbol UM Parameter
CV l Closing Volume
VC(c.v.) l Expired volume during wash-out
V_I l Related Volume at the end of phase I
V_II l Related Volume at the end of phase II
V_III l Related Volume at the end of phase III
m, q, R^2 l Parameters of the linear fitting on the alveolar plateau
CV/VC% % CV as a percentage of VC
VDf ml Dead volume (Fowler)
CO diffusion Test
Symbol UM Parameter
DLco ml/min/mmHg CO Diffusion Capacity
DLco 3eq ml/min/mmHg DLco computed with the Three Equations method.
Chapter 2 - Measurements - 29
DLco corr ml/min/mmHg CO Diffusion Capacity corrected for Hb, COHb, PB.
DLco/VA ml/min/mmHg Krogh factor
Dm ml/min/mmHg Membrane Diffusion capacity
Vc ml Capillary Volume
FaCO % Alveolar concentration of CO
FACH4 % Alveolar concentration of CH4
ta sec Apnoea time
FiCO % Inspiratory concentration of CO
FiCH4 % Expiratory concentration of CH4
wash-out vol. % Rejection Volume
Vsample vol. % Sampling volume
IVC (DLCO) l IVC calculated during test DLCO
TLC (DLCO) l Total Lung Capacity
Respiratory Drive - P0.1 Test
Symbol UM Parameter
P0.1 cmH2O Respiratory drive
FiCO2 % Inspiratory CO2 (PFT2 and PFT4 only)
RF 1/min Respiratory Frequency
Vt l Tidal Volume
MIP/MEP
Symbol UM Parameter
MIP cm H2O Maximum inspiratory pressure
MEP cm H2O Maximum expiratory pressure
Cardio pulmonary Exercise Testing (CPET)
The Ergo option makes the Quark PFT a complete Cardiopulmonary Exercise Testing
system able to measure the following parameters:
Breath by Breath exercise testing
Symbol UM Parameter
VO2 ml/min Oxygen Uptake
VCO2
Vt l Tidal Volume
FetO2 % End Tidal O2
FetCO2 % End Tidal CO2
R --- Respiratory Quotient
VE l/min Ventilation
HR 1/min Heart Rate
Qt l Cardiac output
AT --- Anaerobic Threshold
VE l/min Ventilation
SV l/min Stroke volume
RF 1/min Respiratory Frequency
FeO2, FeCO2 % Averaged expiratory concentration of O2 e CO
VE/VO2 --- ventilatory equivalent for O2
VE/VCO2 --- ventilatory equivalent for CO2
VO2/HR ml/beat Oxygen pulse
VO2/Kg ml/min/Kg VO2 per Kg
ml/min
Carbon Dioxide production
30 - Quark PFT User Manual
Ti, Te, Ti/Ttot sec time breaths
Vd/Vt --- Vd/Vt ratio
PaCO2 mmHg arterial PCO2 (estimated)
P(a-et)CO2 mmHg Delta PaCO2 – PetCO2
Indirect Calorimetry
Symbol UM Parameter
EE Kcal/day Energy Expenditure
EE/BSA Kcal/day/m2 Energy Expenditure/Body surface area
EE/Kg Kcal/day/Kg Energy Expenditure pro Kg
FAT Kcal/day Fats
CHO Kcal/day Carbohydrate
PRO Kcal/day Protein
FAT% % % Fat
CHO% % % Carbohydrate
PRO% % % Protein
npRQ —- Respiratory quotient not protein
Lactate Threshold (V-Slope)
Symbol UM Description
VO2 @ LT l/m Lactate (Anaerobic) Threshold STPD
R @ LT -- Respiratory Quotient @ LT
Time @ LT hh:mm:ss Time @ LT
VCO2 ml/min CO2 output @ LT STPD
VE l/min Ventilation @ LT BTPS
HR bpm Heart Rate @ LT
O2 Kinetics
Parameter UM Calculation
O2 deficit l/m VO2@work*tau
O2 debt l/m VO2'@work*tau
Chapter 2 - Measurements - 31
32 - Quark PFT User Manual
Installation
Unpacking the system
On receipt of your system, you should immediately inspect your package for shipping
damages, in case damage is suspected please contact the reseller immediately. Your
system has to be installed by COSMED or by an authorised reseller.
Excessive amount of dust and miscellaneous clutter around the instrument can
eventually cause malfunctions due to overheating of components, we strongly suggest
to keep the unit in a clean environment and well ventilated as possible.
Quark PFT and accessories can have place on the trolley as shown in the following
illustration.
34 - Quark PFT User Manual
Setting up the system
Before starting operating with the system make sure to meet the environmental and
operational conditions reported in Chapter 1.
Installing the Calibration Gas Cylinders
In order to calibrate the sensors you need to have available calibration cylinders with
the following gas concentrations:
Cylinder Recommended Gas mixture Test
Calibration O2 16%, CO2 5%, N2 Bal Washout, CV, P0.1, Ergo, MIP-MEP
DLCO CO 0.3%, CH4 0.3%, O2 21%, N2 Bal DLCO
Oxygen O2 100 % Washout, CV
Gas pressure adjustment procedure
The gas regulator has an adjustable second stage that must be adjusted only when used
for the first time. This is necessary to protect the internal demand valve from the high
pressure surge that can be generated when the cylinder is opened.
1. Make sure that the regulator is turned off before opening the cylinder valve.
2. Open the cylinder valve by turning the valve counter-clockwise.
Notice: The cylinders must be
provided with a calibration
certificate, indicating the gas
composition.
3. Adjust the regulator pressure by turning the adjustment knob clockwise in order to
reach a value between 5 and 6 atmospheres (bar).
Notice: The gas cylinder must be replaced when the primary pressure gauge falls below
10 bar.
Chapter 3 - Installation - 35
Connecting cables and tubing
The assembling procedure will be shown in the following illustrations.
A Heart Rate Probe connector
D Flowmeter connector
E Calibration plug
F Power switch
G Sampling tube connector
H Breathing valve connector
I DLCO mix gas connector
L Cal gas (16% O
M O
O RS232 Serial Port for PC connection
P Ground snap connector
Q Power cable connector
R Switch 110V / 220V
36 - Quark PFT User Manual
gas connector
2
, 5% CO2) connector
2
Cables and Tubing’s installation sequence
1. Connect the power cable of the Quark PFT to the trolley
2. Connect the power cable of the PC and the printer to the trolley
3. Connect the RS232 cable from the Quark PFT (PC RS232) to the PC (COM1 or
COM2). If the PC does not have a RS232 port, please use the USB-RS232 adapter
(A 388 410 001), available as an accessory.
4. Connect the DLCO cylinder to the Quark PFT (DLCO) and adjust its pressure
between 5 and 6 bar.
5. Connect the O2 cylinder to the Quark PFT (O
and 6 bar.
6. Connect the O
-CO2 calibration cylinder to the Quark PFT (CAL) and adjust its
2
pressure between 5 and 6 bar.
7. Connect the Flowmeter to the front panel of the Quark
Assemble the breathing valve
) and adjust its pressure between 5
2
Chapter 3 - Installation - 37
Assemble the turbine
Assemble the mask and the flowmeter
1. Insert the turbine in the optoelectronic reader, in the way indicated by the arrow in
the turbine.
2. Insert the turbine plug on the front panel.
Notice: It is advisable to
lubricate periodically the Orings inside the
optoelectronic reader with
sylicone grease for its good
maintenance.
3. Connect the external anti moisture filter to the sampling plug and joint the
Using the "Ultimate Seal"
The "ultimate seal" is a moulded of Elasto-Gel, a glycerine based hydrogel. This
product is a unique polymer gel that forms an intimate seal between the face and the
mask. It has to be used for mask applications on hard to seal faces and where leaks are
not tolerated.
• Will not irritate the skin
• Contains no adhesives.
• Has no odour
• Will not dry out
• Single patient use
Permapure sampling line to the filter. It's strongly suggest to use the filter always
and we recommend to replace it every 3 months.
38 - Quark PFT User Manual
Notice: Avoid the exposure
to the sun. Do not put the
seal into the water.
Connecting PC
Apply the seal to the mask
Apply seal to clean, residue-free mask only and follow the instructions below:
1. Remove the plastic tray from the bag. Peel off clear film and retain for later use.
2. While holding tray align the nose area of mask to nose area of Ultimate Seal™ gel.
Press together and roll mask down over the surface of the gel seal attaching it to the
mask and releasing it from the tray.
3. If needed, adjust the position of the seal, aligning it with the outer perimeter of the
mask sealing surface.
4. The mask is now ready to be placed on the subject's face.
To remove seal on mask
• The Ultimate Seal™ have been conceived for a single patient use only, it can not be
cleaned or sterilised.
• If mask requires cleaning for a new patient application then pull off and dispose of
the Ultimate Seal™.
• To keep the seal clean between use, keep it attached to the mask and place the clear
film against the Ultimate Seal™ gel on the mask. When the seal becomes
discoloured or opaque (approximately two weeks) dispose of the current seal and
replace it with a new one.
Before operating the system make sure the unit is connected to the PC as described in
the following illustration:
Chapter 3 - Installation - 39
Software installation
The software is made of two programs: one for the spirometry and the other one for the
ergometry (only for PFTergo, PFT2ergo and PFT4 ergo). The two programs share the
same archive and use the same program for the system calibration, even if they are used
for performing completely different tests.
Installing the software
1. Select Run... from Windows Start menu.
Notice: The software is
copy-protected. Install the
software from the original
disk.
Run the software
2. In the Command line box, type a:\install (assuming the disk is in drive A:).
3. Click on OK (or press ENTER key).
4. The program will load up a dialog box and ask for a directory where to be installed.
5. When the installation is over, the program will advise you with a message
indicating that the installation has been successfully completed, click on End.
Note: the directory for the Ergo software must be the same of the Quark PFT (spiro).
1. In the Windows Program Manager, open the Program Group in which the software
was installed.
2. Click the Quark PFT icon.
PC port configuration
The first time the software is used, it is necessary to configure the communication port
with the PC (USB, COM1, COM2,…).
For further details, see the chapter Database management.
40 - Quark PFT User Manual
Software main features
Display
The program may contain several windows. The active window is highlighted with a
different colour of the caption. Some functions of the program are "active window"
sensitive (Print, right key of the mouse).
Tool bar
Many of the functions that may be selected from the menu can be activated more
rapidly by clicking with the mouse on the corresponding icon in the tool bar.
Positioning the mouse cursor on one of the buttons of the toolbar (if the option Hints is
enabled), the description of the corresponding function is shown in a label.
Show/hide the toolbar
Select Toolbar from Options menu in order to show or hide the toolbar.
Dialog windows
The typical operating environment of Microsoft Windows is the Dialog box. This
window is provided with a series of fields in which input the information.
Scroll bars
On-line help
Use of the keyboard
• To move the cursor among fields, press the Tab key until you reach the desired
field.
• Press the Enter key to confirm the information input on the dialog box or press the
Esc key to cancel changes.
Use of the mouse
• To move the cursor among fields, move the mouse on the desired field and leftclick.
• Click on the OK button with the Left button of the mouse to confirm the
information input on the dialog box or click on Cancel button to cancel changes.
Some windows are provided with scroll bars that help to see data exceeding the window
space available.
• To move the scroll bar row by row click the scroll arrows at the end of the scroll
bars
• To move the scroll bar page by page click on the grey area at both sides of the
scroll fields
COSMED Quark Help is a complete on-line reference tool that you can use at any time.
Help is especially useful when you need information quickly or when the Quark user
manual is not available. Help contains a description of each command and dialog box,
and explains many procedures for accomplishing common tasks.
To get the Help on line, press the F1 key.
Software version
To know the software version and the serial number of the software, select About
Quark PFT… from Help menu.
Chapter 3 - Installation - 41
42 - Quark PFT User Manual
Calibration
Calibrating Sensors
The software allows to automatically calibrate zero, gain and delay of gas sensors. Even
if the program doesn't force you to carry out the calibration we strongly recommend to
execute it before each test.
Running the Calibration program
Start the program and choose Calibration from the Test Menu. The software runs the
Calibration software and the main menu changes accordingly.
Log file
The program creates and updates as default the calibration log file, containing the
conditions and the results of all the calibrations performed by the user.
To access the file select File/Report File... from the calibration program.
Setting reference values
Before starting calibrating make sure that the system has been configured correctly by
setting the right values of gas concentration of: room air (i.e. 20.95% O
CO
), of gas mixture contained in the cylinders and the volume of the calibration
2
syringe (i.e. 3 litres).
and 0.03%
2
To set the reference values
This operation must be performed only the first time. The next times, the system keeps
stored the reference values entered in this step.
1. Select Reference Values from the Calibration menu.
2. Type the correct values for the O
and CO2 room air concentration (i.e. 2093 for
2
20.93%), and do the same for the gas concentration of the calibration cylinder.
3. Type the volume of the calibration syringe (i.e. 3000 for a 3 litres calibration
syringe).
4. Press OK button to confirm changes.
44 - Quark PFT User Manual
Calibrating analysers
Note: After turning on the
unit, wait 30 minutes warm
up time before starting the
calibration procedure.
ERGO Calibration
This calibration must be repeated at least daily. A new calibration before any
ergospirometry, P0.1 or MIP-MEP test is strongly recommended.
1. Remove the sampling line from the flowmeter.
2. Ensure that the O
3. Check if the reference values for the calibration gas specified in File/Reference
4. Connect the sampling line to the CAL port placed on the front panel of the Quark
5. Select Calibration/Gas…/ERGO… and wait until the procedure is completed.
The software performs automatically the calibration procedure. After 90 seconds the
graph will be displayed. In this way, the user can check the calibration procedure both
graphically and numerically.
At the end of the procedure, the software displays the new calibration factors vs. the old
ones.
-CO2 cylinder is properly connected to the rear panel of the
2
Quark PFT and check that the pressure of the gas is in the range 5-6 bar.
values… correspond to the certified composition of the mixture.
PFT.
FRC Calibration
This calibration must be repeated at least daily. A new calibration before any FRC or
CV test is strongly recommended.
1. Remove the sampling line from the flowmeter.
2. Ensure that the O
cylinder is properly connected to the rear panel of the Quark PFT
2
and check that the pressure of the gas is in the range 5-6 bar.
3. Ensure that the O
-CO2 cylinder is properly connected to the rear panel of the
2
Quark PFT and check that the pressure of the gas is in the range 5-6 bar.
4. Check if the reference values for the calibration gas specified in File/Reference values… correspond to the certified composition of the mixture.
5. Connect the sampling line to the CAL port placed on the front panel of the Quark
PFT.
6. Select Calibration/Gas…/FRC… and wait until the procedure is completed.
DLCO Calibration
This calibration must be repeated at least daily. A new calibration before any DLCO test
is strongly recommended.
1. Remove the sampling line from the flowmeter.
Chapter 4 - Calibration - 45
2. Ensure that the DLCO cylinder is properly connected to the rear panel of the Quark
PFT and check that the pressure of the gas is in the range 5-6 bar.
3. Check if the reference values for the calibration gas specified in File/Reference values… correspond to the certified composition of the mixture.
4. Connect the sampling line to the CAL port placed on the front panel of the Quark
PFT.
5. Select Calibration/Gas…/DLCO… and wait until the procedure is completed.
Print the calibration report
In the Calibration program choose Print from the File menu.
Edit the calibration factors
The last sensors calibration factors can be either edited or viewed. To do this choose
Gas Results… from the File menu.
To view or edit the last Turbine calibration factor choose Turbine results… from the
File menu.
Note: To restore factory setting press Default button in the dialog box. Once you press
the default button you must run a new calibration before testing.
O2 delay check
Selecting Calibration/Gas/O
250ms, please contact the Technical Support.
An high value of the O
• Chemical sensor exhausted.
• Old or obstructed Permapure tube.
• Dust or dirt in the sampling line.
delay, check if this value is below 250ms. If it is above
2
delay can be due to:
2
46 - Quark PFT User Manual
Turbine calibration
The system uses a turbine flowmeter. It opposes a very low resistance to flow (<0,7
cmH
motion which causes the rotation of the turbine rotor.
The rolling blade interrupts the infrared light beamed by the three diodes of the
optoelectronic reader. Every interruption represents 1/6 turn of the rotor, this allows to
measure the number of turn in the time. There is a constant ratio between air passing
through the turbine and number of turns. This allows an accurate measure of flows and
volume. The turbine flowmeter doesn't need daily calibrations as it is not affected by
pressure, humidity and temperature.
To work properly, the turbine only requires the rotor to rotate freely without any friction
that might be caused by dust that can be easily avoided with an ordinary cleaning
procedure (see Maintenance).
However in order to ensure accuracy it's recommended to run periodically the
calibration procedure. Calibration has to be carried out with a calibration syringe of 3
litres volume, the calibration procedure is totally managed by software.
A measurement system should be calibrated daily in order to ensure maximum accuracy
and reliable test results. If a correct maintenance is provided it’s possible to check the
calibration of the turbine flowmeter even at relatively long intervals (i.e. 1 month). The
calibration procedure assures valid and verifiable results within a ±3% accuracy.
O/l/s to 12 l/s). The air passing through the helical conveyors, takes a spiral
2
The calibration syringe
The 3 litres calibrated syringe is included in all the Quark PFT line with the exclusion
of the PFT 1 model.
3 litres calibration syringe: P/N C00600-01-11.
Assembling the flowmeter for calibration
1. Connect the Opto-reader to the calibration syringe through the adapter. Do not use
the breathing valve.
2. Connect the flowmeter to the syringe with the rubber cylinder supplied in the
standard packaging.
Note: If a bacterial filter is used for the tests, do use it also during the turbine
calibration.
Chapter 4 - Calibration - 47
g
Calibrating the turbine
Note: if you are using a slow
PC, we recommend to set
her refresh time.
an hi
After having run the calibration program:
1. Select Reference Values from the File menu. If your syringe has a different value
2. Select Calibration/Turbine….
3. When the Calibration Turbine dialog box appears with the syringe piston initially
from the default one (3 litres), please enter the correct value.
pushed all the way in, move the piston in and out for 5 inspiratory strokes and 5
expiratory strokes in order to get the first values appearing on the display. Then
move the syringe piston for other 10 strokes (IN and EX).
4. At each of the 10 steps the software displays the results of the manoeuvre and the
percentage error in the reading.
5. At the end of this operation, the software displays the new calibration factors. Press
OK to store the new value.
48 - Quark PFT User Manual
Checking the system signals
The control panel
The Control Panel, which can be activated from the Calibration/Control panel…
menu item, is a useful tool to check the main hardware functions of Quark PFT.
By using the controls on Control Panel you are able to do the following:
1. Reading the signals acquired by the system both as voltages and processed data;
2. Activating/Disactivating the valves, the sampling pump and other installed
components (for example, oxymeter).
Using the control panel
Sampling
flow rate
Sampling
pump
mV / real values
display
valve air valve
Analog
input
Apnoea valve
Breathing
Calibration
valve
Deselect all
channels
Oxymeter
Air valve for
DLCO
On/off cylinders
On/Off infrared
sources
HW zero trimmer
adjystment
sensor
O
2
range
Signal refresh
time
Select all
channels
Chapter 4 - Calibration - 49
50 - Quark PFT User Manual
Database
Management
Spirometry patient's database
The Patients database consists of a Patient Card, a Visit Card and a Test Card in which
are listed all tests performed by the patient.
Select Archive Navigator from the File menu or press the button by side.
Note: after having deleted a
record (patient, visit or test),
it is recommended to
reorganize the archive in
order to free disk space.
Patient Card
It collects all the information of a patient (first name, last name, date of birth) which
remain the same for each visit. For each patient there is only one Patient Card, which is
created the first time the Patient performs a test.
To move within the database use the following buttons:
Move to the first patient in the archive
Move to the previous patient in the archive
Move to the next patient in the archive
Move to the last patient in the archive
Find a patient in the archive
Enter a new patient in the archive
Delete current patient from the archive
Edit the current patient card
Visit Card
It collects all information relative to the visit (diagnosis, visit description...) and to the
patient information subject to change between one visit and another (height, weight,
smoke). Each patient can be related to several Visit Cards provided they have been
created in different days. Before carrying out any spirometric test it is necessary to
create a new Visit Card or to open the today’s Visit Card.
To move within the database use the following buttons:
52 - Quark PFT User Manual
Move to the first visit in the archive
Move to the previous visit in the archive
Move to the next visit in the archive
Move to the last visit in the archive
Find a visit in the archive
Enter a new visit card in the archive
Delete current visit card from the archive
Edit the current visit card
Test Card
It contains all the information about the test.
To move within the database use the following buttons:
Delete current test from the archive.
Edit the current test
Import/export a Tests card
This function allows to import /export a test card with the respective visit and patient
card.
1. Select the patient.
2. Choose the test and press the key by side. All data will be imported/exported in the
XPO file format (Cosmed proprietary).
Diagnosis Database
The program allows to manage a diagnosis database, whose records are composed by a
diagnosis ID code and a string of text.
The report of the visits can be done either by typing the desired text in the field
“Diagnosis” of the Visit Card or, more quickly, retrieving from the diagnosis database
the desired one.
If you want to insert, modify or delete a diagnosis from the database select Database Diagnosis... from the File menu.
Chapter 5 - Database Management - 53
Exercise testing patient's database
The exercise testing software uses a different interface for presenting patient
information. The patient database allows to:
• Enter a new patient
• Find patient data
• Edit patient data
• Delete patient data.
Select Patients from the File menu.
Enter a new patient
1. Press New to show the Patient dialog box.
2. Enter data of a new patient and press OK button to confirm.
Find a patient
Edit patient data
Delete a patient
Enter a search string into the Find field and press Find to view the data concerning a
subject already in the database. You can search for “Last name”, “ID code” or
“Progressive”.
Or:
Press List to view the list of patients in the database. Press Next or Previous to view
the data corresponding to the next or to the last patient. Press OK to confirm.
The Next and Previous buttons allow to move to the next or the previous patient in the
database.
1. Select the patient.
2. Press Modify… in order to edit the patient’s data.
3. Edit the desired values and press OK to confirm.
1. Select the patient to be deleted.
2. Press Delete.
54 - Quark PFT User Manual
Archive maintenance
The software allows to manage files selecting Archive from the File menu.
It is advisable to perform the archive reorganisation every month, in order to free space
on the hard disk and/or to correct possible errors present within the database.
It is possible also that your have no more hard disk space. So, you have to delete all the
data. In this case, it is useful to perform the initialising.
Reorganise the archive
1. Select Reorganize archive from the File menu.
2. Wait for the end of the operation before performing any other function.
Delete the archive
1. Select Initialize Archive from the File menu.
2. Wait for the end of the operation before performing any other function.
Delete a test
To delete an ergometry test, select Test/Delete test.
To delete a spirometry test, press the proper button in the Test Card.
Backup and restore
It is strongly recommended to backup files, a warning message will be displayed
monthly. This function allows the user to restore the data if the PC or the HD will not
work anymore.
Backup
1. Select Backup archive from the File menu.
2. Selecting the destination path with the Browse key or press New to create a new
directory. Press OK to confirm.
3. In the dialog box it will appear an estimate of the number of floppy disks you need
in order to back up the archives. Press OK.
Restore
1. Select Restore archive from the File menu.
2. On the Restore dialog box specify the drive source and press OK, a dialog box
will appear indicating all data of the backup processed.
Chapter 5 - Database Management - 55
Spirometry Settings
Quark PFT software allows to configure some options selecting Configure from the
Option menu.
Graphs
All the graphs visualised and/or printed can be customised in colours and appearance.
1. Select the desired colours of the curves (5 curves max can be overlapped on the
same graph).
2. Enable or disable the Grid option.
3. Enable or disable the Show Info Test option.
Serial port
You must select the serial port RS 232 that will be used to connect the Quark PFT with
the PC.
To select the serial port, click on the proper COM button (the selected port must be
different from the mouse one).
Units of measurements
It is possible to configure the units of measurements, weight and height, for printing and
viewing.
To select the units of measurement click on cm/Kg or in/lb according to the desired
format.
Using extra fields
The Patient’s database is organised in 3 different cards (Patient card, Visit Card and
Test card.) where it is possible to store the information about patients and visits .
Besides the standard information, it is possible to customise some fields (user free
fields), entering and labelling measurements coming from other devices.
The customisable free fields are:
• 3 fields in the Patient Card (Patient’s information)
• 3 fields in the Visit Card (information about the visits)
• 3 fields (2 numeric) in the Test card information about Test)
Customise the fields
In the group User free fields type the desired text in the 9 fields available.
56 - Quark PFT User Manual
Spirometry
Setting spirometry options
The software allows to configure some options selecting Configure from the Option
menu.
Spirometry
Automatic Interpretation
Quark PFT has the function of interpreting each test performed by a patient visualising
an automatic diagnosis. The algorithm has been calculated basing on “Lung Function
Testing: selection of reference values and interpretative strategies, A.R.R.D. 144/
1991:1202-1218”.
The automatic diagnosis is calculated at the end of the FVC Test if:
• the automatic diagnosis option is enabled.
• the patient’s anthropometric data allow the calculation of the LLN (Lower Limit of
Normal range).
• at least one FVC test has been performed.
To enable/disable the automatic diagnosis:
1. Click on Enable Automatic Interpretation checkbox to enable or disable the
calculation and the visualisation of the automatic interpretation.
2. Select the LLN (Lower Limit of Normal Range) criteria among the ATS
(LLN=Pred-0.674*SD), ERS (LLN=Pred-1.647*SD) or 80%Pred (LLN=Pred*0.8)
specifications.
Quality control
Quark PFT allows a quality test control. The calculation has been carried out referring
to “Spirometry in the Lung Health Study: Methods and Quality Control, A.R.R.D. 1991;
143:1215-1223”. The messages concerning the quality control are shown at the end of
the test.
To enable/disable the quality control, click on Enable Quality Control checkbox.
58 - Quark PFT User Manual
Parameters manager
The program allows to calculate a huge number of parameters; it is advisable, in order
to simplify the analysis of the results, to view, to print and to sort the desired parameters
only. Select the menu item Options/Parameters...
View
Move the parameters to view into the Selected parameters list.
Print
Move the parameters to print into the Selected parameters list.
Sort
Drag the parameter up or down with the mouse.
Customise
Add, modify and delete custom parameters.
If it is necessary to restore the default parameters press the button in the left corner of
the window to initialise the parameters database.
Predicted values manager
The program contains a preset of predicted equations, but the user is allowed to
customise its own predicted sets. Select Predicteds... from Options menu.
The window is divided into two forms: Predicteds set and Formula definition.
Predicteds set
This form allows the user to manage the set of predicteds. The following information
define a set:
Chapter 6 - Spirometry - 59
Name: identifies the set and cannot be duplicated;
Description: free field;
Age: the adult predicteds start since this age.
To enter a new set of predicteds click on the New button. The field Name must be filled
and must be unique. To stop without saving click on the Cancel button. To save the set,
click on the Save button.
To delete a set of predicteds click on the Delete button. If a set is deleted, also the
associated formulae are deleted.
It is possible to generate a new set of predicteds with the same attributes and the same
formulae of the selected one. To do this click on the Copy... button and specify a new
Name.
To import a set of predicteds click on the Import... button and select a file of Predicteds
files type.
To export a set of predicteds click on the Export... button.
In the list Set current predicteds choose the current predicteds for printing and
viewing.
Set the current predicted
Quark PFT allows to calculate the predicted values according to the following
configurable sets:
Adults Paediatrics
ERS 93 Zapletal
Knudson83 Knudson83
ITS white ITS white
ITS black ITS black
LAM LAM
MC Barcelona MC Barcelona
Nhanes III Nhanes III
Select the desired choice in the group Predicted.
Formula definition
This form allows the user to manage the formulae associated to a set of predicteds.
Select the set of predicteds from the list Predicteds set.
To insert a new parameter click on the New... button.
The parameter formulae can be:
60 - Quark PFT User Manual
Page set-up
• calculated according to the predicteds in the list Use the predicteds formulae;
• customised by the user with the option ...or the customised formulae.
The Delete button deletes the selected parameter.
The Copy button stores the selected parameter in memory.
The Paste button inserts a new parameter from the one copied. If the name is not
unique, the user is asked whether to specify a new name or to replace the existing
parameter.
Select Page Setup... from the File menu.
HeaderAll the printouts carried out by the program are preceded by 3
rows of customisable header (usually they contain the name
and the address of the Hospital using the spirometer).
Data Patient and visit information are printed below the header.
These data are reported on 3 columns and 5 rows. the user may
configure the disposition, change and eventually cancel the
fields, as he prefers.
MarginsConfigures the print margins from the borders of the paper.
The unit of measure is decided in Units of measurements.
Footer Configures information at the bottom of the page.
Printed file name Defines the automatic name to be asssigned to the pdf file, if
the report will be printed in this format.
In the example it has been set to create a filename composed by
<date of the test> followed by <last name> and <first name>.
Chapter 6 - Spirometry - 61
Spirometry tests
Note: Read carefully the
contraindications in Chapter 1.
Once completed the phases of the introduction of the patient’s data and the visit data, it
is possible to carry out the spirometric tests.
Quark PFT allows to perform the following tests:
Key Test
FVC pre Forced Vital Capacity
FVC post Forced Vital Capacity after bronchial stimulation
SVC Slow Vital Capacity
MVV Maximum Voluntary Ventilation
FRC Nitrogen Wash-out
CV Closing Volume
DLCO CO Diffusing capacity
P0,1 Respiratory Drive
Note: The FVC, SVC and MVV tests must be performed with the flowmeter
disconnected from the breathing valve.
Before performing any test make sure that:
1. Quark PFT is properly connected to your PC and the selected serial port (COM1,
COM2) corresponds to the one effectively use.
2. The name shown on the status bar corresponds to the patient who is to carrying out
the tests.
3. The today’s visit card exists.
62 - Quark PFT User Manual
Forced Vital Capacity (pre)
FVC is a reference test to verify obstructive (airflow limitations) and restrictive
disorders (lung volume limitations). To achieve good test results it is fundamental a
good manoeuvre (quality control messages, real time plots …)
The main parameters measured during FVC tests are:
FVC Forced Vital Capacity
FEV1 Forced Expiratory Volume in 1 second
FEV1/FVC% FEV1 as a percentage of FVC
PEF Peak Expiratory Flow
FEF25-75% Forced mid-Expiratory Flow
The two representative plots are the Flow/Volume and Volume/Time loops.
By comparing FVC, FEV1 and FEV1/FVC% values the software allows an automatic
interpretation concerning the levels of obstructive and/or restrictive disorders.
Recommendations
• The flowmeter has to be disconnected from the breathing valve
• The patient should wear the nose clips
• The turbine has been recently calibrated (ATS recommends a daily calibration)
• The paper mouthpiece or the antibacterial filter is properly connected to the
flowmeter through the corresponding adapter
For hygienic reasons, we strongly recommend the use of a bacterial filter.
If a kid must perform the test it is recommended to enable the encouragement function
which shows exactly the manoeuvre of the FVC test.
Perform a FVC (pre) test
1. Select Forced Vital Capacity pre from the Test menu and wait for the green led is
2. Explain the manoeuvre to the patient and press the F2 key.
3. Wait some seconds and perform the test.
4. After having performed the test, press F3 or wait for the automatic end (5 seconds
5. In order to visualise the F/V and V/t graph and the main parameters press the
6. Press Alt+F3 to stop the acquisition discarding the results.
7. Repeat the test until it is correctly performed (ATS recommends 3 times).
8. Press Exit to visualise the test list carried out during current session together with
9. Select the test that you want to save (the Best Test according to the ATS criteria is
prompted on the right side of the screen.
without flow), so that the software displays the F/V and V/t graphs, the main
parameters, and the predicteds values.
following buttons:
view Flow Volume graph
view Volume Time graph
view data of the test
the results of the main parameters.
highlighted as default) and press OK.
Test encouragement
During FVC manoeuvre you might experience some lack of collaboration with kids or
with other patients. In this case you may find a good help in using the encouragement
software tool.
Chapter 6 - Spirometry - 63
Perform the FVC test with the encouragement
1. Select Encouragement from View menu.
2. Perform the test as explained in the previous paragraph.
64 - Quark PFT User Manual
Slow Vital Capacity
Important test for assessing COPD (chronic obstructive pulmonary disease) patients
affected by this disease might present a the Slow Vital Capacity could be higher than
the Forced one (FVC).
The main parameters measured during SVC tests are:
EVC Expiratory Slow Vital Capacity
IVC Inspiratory Slow Vital Capacity
ERV Expiratory Reserve Volume
IRV Inspiratory Reserve Volume
If the inspiratory/expiratory maximal manoeuvre is preceded by a some breaths at tidal
volume the software allows to measure the Respiratory Pattern, represented by the
following parameters:
VE Ventilation per minute
Vt Tidal volume
Rf Respiratory frequency
Ttot Breath time
Ti/Ttot Inspiratory time/Ttot
Vt/Ti Vt/Ti
Perform a SVC test
1. Select Slow Vital Capacity from the Test menu and wait for the green led is
prompted on the right side of the screen.
2. Press F2 and instruct the Patient to breath normally until the message “carry out…”
is prompted; then ask to perform a Slow Vital Capacity (deep inhalation, maximal
slow expiration and deep inhalation again).
3. Press F3 or wait for automatic interruption (5 seconds without flow) in order to
visualise the V/t graph together with the main parameters compared to the predicted
values
4. To visualise the V/t graph and the main parameters press the follow buttons:
view Volume Time graph
view data of the test
5. Press Alt+F3 to stop the acquisition discarding the results.
6. Repeat the test until it is correctly performed (ATS recommends 3 times).
7. Press Exit to visualise the test list carried out during current session together with
the results of the main parameters.
8. Select the test that you want to save (the Best Test according to the ATS criteria is
highlighted by default) and press OK.
The reference for the ERV calculation is displayed on the V/T graph.
Chapter 6 - Spirometry - 65
Maximum Voluntary Ventilation
Test for assessing the maximum ventilatory capacity. In the past, it was commonly
performed during routine PF tests, however its clinical use declined over the years.
Today MVV test is most commonly performed as part of the exercise tolerance tests,
where it is used as an index of maximum ventilatory capacity. Test consists in breathing
in and out deeply and rapidly for 12, 15 seconds. The expired volume during this short
period is then extrapolated
The most important measured parameter is the following:
MVV Maximum Voluntary Ventilation
Perform a MVV test
1. Select Maximum Voluntary Ventilation from the test menu and wait for the
green led is prompted on the right side of the screen.
2. Press F2 and make the Patient breath as much deeply and rapidly as possible for at
least 12 seconds.
3. Press F3 or wait for automatic interruption (5 seconds without flow) in order to
visualise the V/t graph together with the main parameters compared to the predicted
values
4. To visualise the V/t graph and the main parameters press the follow buttons:
view Volume Time graph
view data of the test
5. Press Alt+F3 to stop the acquisition discarding the results.
6. Repeat the test until it is correctly performed (ATS recommends 3 times).
7. Press Exit to visualise the test list carried out during current session together with
the results of the main parameters.
8. Select the test that you want to save (the Best Test according to the ATS criteria is
highlighted as default) and press OK.
66 - Quark PFT User Manual
Bronchial Provocation Test
Bronchodilator test
Note: Read carefully the
contraindications in Chapter 1.
Bronchodilators are administered routinely in the PFT laboratory to determine whether
airflow obstruction is reversible. Bronchodilators increase airway calibre by relaxing
airway smooth muscle.
The test consists of comparing results between the reference FVC (FVC PRE) and the
FVC POST performed after the administration of the drug. Increasing value of 13-15%
of FEV1, respect to the basal value (FVC Pre) is considered as a reversible condition.
Main parameters are the following:
DFEV1%pre Change of FEV1 as a percentage of test PRE
DFVC%pre Change of FVC as a percentage of test PRE
DPEF%pre Change of PEF as a percentage of test PRE
Some authors states that the above mentioned parameters are too dependent from the
FVC Pre, hence latest reference (ERS93, [A comparison of six different ways of
expressing the bronchodilating response in asthma and COPD; reproducibility and
dependence of pre bronchodilator FEV1: E. Dompeling, C.P. van Schayck et Al; ERJ
1992, 5, 975-981]) recommend the following parameters:
DFEV1%pred Change of FVC as a percentage of predicted value
DFEV1%poss Change of FEV1 as a percentage of “possible value”
Methacholine and Histamine Bronchial provocation Tests
The most common indication for performing methacholine and histamine bronchial
challenges is to diagnose hyperresponsive airways. Some patients demonstrate normal
baseline pulmonary function despite complaints of “tightness” wheezing, cough, and a
little or not response to bronchodilator. Other patients demonstrate spirometric
improvement after use of bronchodilator have diurnal variation in peak flows. In this
groups aerosolised bronchial challenges are used to confirm a diagnosis of Asthma.
We can summarise the use of the test as follows:
1. Diagnose asthma
2. Confirm a diagnosis of asthma
3. Document the severity of hyperresponsivness
4. Follow changes in hyperresponsivness
When patients with hyperresponsive airways inhale certain pharmacologic agents (i.e.
Methacholine or histamine) the airways respond by constricting.
Test consists of executing repeated FVC following the pharmacologic agents inhalation
according to an established protocol. The fall of the FEV1 parameter is used to calculate
the bronchial hyperresponsivness. The most important parameter is the PD20 that is
amount of drug (mg/ml) that causes a reduction of 20% of the FEV1 respect the basal
value (without drug).
Main parameters are:
P10 Dose that causes a 10% fall of FEV1.
P15 Dose that causes a 15% fall of FEV1.
P20 Dose that causes a 20% fall of FEV1.
The representative plot is the Dose/response curve, showing the percentage variation of
FEV1 versus the Drug dose in logarithmic scale.
The program assumes as the baseline test the best FVC pre carried out during the
today’s visit. You can change the reference pre test editing the Post test.
The name of the drug, its quantity and its unit of measurement, can be typed
immediately before any FVC post manoeuvre (manual protocol) or can be stored in a
database of bronchoprovocation (File/Bronchial Provocation protocols Database…).
Chapter 6 - Spirometry - 67
Perform the test
(During 1st step only) select Protocol... from the Test menu and choose the name of the
bronchoprovocation protocol that you are going to use (manual protocol if you want to
type the information about the agent before any manoeuvre)
1. Select FVC post from the Test menu.
2. Select an existing protocol or click on “manual protocol”, and wait the green leds
turned on.
3. Press F2, or the button by side, to start the test.
4. Press F3, or the button by side, to achieve the test.
5. In order to visualise the V/t graph and the main parameters press the follow
buttons:
view Flow Volume graph
view data of the test
view bronchial provocation response
6. Press Alt+F3 to stop the acquisition discarding the results.
7. Repeat the test until it is correctly performed (ATS recommends 3 times).
8. Press Exit to visualise the test list carried out during current session together with
the results of the main parameters.
9. Select the test that you want to save (the Best Test according to the ATS criteria is
highlighted as default) and press OK.
Bronchial Provocation protocols Database
The response to a bronchoprovocator is usually assessed in terms of change in the
FEV1, vital capacity or airways resistance on the basis of serial measurements (FVC
manoeuvres) in which the results of the initial test constitute the reference values. The
international literature proposes several standardised protocols in order to address the
methodological issues of the various available techniques.
The possibility to store a bronchoprovocation protocol in a database is useful to simplify
and automate the sequence of operations that the Physician need to execute during the
bronchoprovocation tests.
The typical sequence of activities to carry out a bronchoprovocation test are:
1. Typing and storing a bronchoprovocation protocol in the database (usually only
once).
2. Selection of protocol among the list of the ones already present in the database
before carrying out the FVC post tests (the selection of “manual protocol” allows to
execute the test fully manually).
3. Performing the Post tests.
Enter a new Bronchial provocation protocol in the archive
1. Select Bronchoprov. protocols database from the File menu.
2. Type the Protocol name, the Bronchoprovocator name and the unit of measurement
in the proper input fields.
3. If the bronchoprovocator has a cumulative effect select the cumulative check
button.
4. Enter the quantities for each step and press the button
68 - Quark PFT User Manual
.
Viewing results
All the visualisation functions refer to the test carried out by the Current Patient, whose
name is indicated on the left-side of the status bar.
To view tests results:
1. Select the Patients from the File menu
2. Select the patient corresponding to the test you want to view.
3. Select in the list box of the tests up to 5 tests of the kind (FVC, VC/IVC, or MVV)
and press OK.
To switch between graph and or data use the following buttons on the toolbar:
view Flow Volume graph (F5)
view Volume Time graph (F6)
view data of the test (F7)
view bronchial provocation response.
If you need more than one visualisation meantime use the New Window function from
the Window menu.
If you need to display a list of visits:
• Select Visits list... from the File menu.
• Type the name of the Company and/or the time interval desired or simply confirm
for the complete list.
Tests of the current patient
If a current patient has been selected you can quickly view his tests selecting Test
current patient... from the View menu.
Delete a test
1. Select Patients from the File menu or press the button by side.
2. Select the test that you want to eliminate from the list of the tests referred to the
Current Patient and press Delete.
Chapter 6 - Spirometry - 69
Printing results
Printing Reports
You can print out in three different ways:
• printing the Report
• printing the Active Window
• printing a series of reports
To print a report of the current visit, select Print report… from File menu. The
software will choose automatically the best performed test.
The standard Report is composed by 1, 2 or 3 pages depending if you wish to printout
the FVC data and the graphs together on the first page or if you wish to printout the
bronchoprovocation response.
• Selecting the option One page (no ATS) the report will contain, on one page, the
F/V and V/t graphs of the best test, overlapped on the FVC Post, the patient data,
the notes, the diagnosis and the test results.
• Otherwise the report will contain two pages, the first with the patient data, the
graphs and the diagnosis, and the second one with the measured parameters,
according to the ATS recommendations.
• The 3
rd
page will contain the bronchoprovocation response.
Select the desired options:
FVC graph Prints the F/V and V/t curves for the best FVC test.
One page (no ATS) Prints data and graphs on the first page.
Response Prints the bronchoprovocator response.
FVC post Prints data and graphs for the Post FVC test (the test can be
selected among the test performed in the current visit).
Multi breath N
Prints the best FRC test of the current visit.
2
wash out
Single breath O
Prints the best CV test of the current visit.
2
Single breath CO Prints the best DLCO with apnoea test of the current visit.
diffusing capacity
Single breath CO Prints the best DLCO without apnoea test of the current visit
(no breath hold)
CO diffusing capacity Prints the best DLCO Steady State test of the current visit
Steady State
Respiratory drive Prints the best P0,1 test of the current visit.
Preview Views a report preview on the screen.
70 - Quark PFT User Manual
Printing the active window
This printout function is only enabled when the active window (title bar highlighted) is
one of the following objects:
• Any kind of Graph.
• Numeric data
• List of visit
To print the active window
1. Ensure that the active window is one of the preceding objects.
2. Select Print Active window from File menu.
Printing a series of reports
Sometimes it is useful to printout automatically a series of reports (all tests carried out
with the employees, all tests carried out in the today's session).
To print out proceed as follows:
1. Select Visit List from the File menu
2. Set the criteria of the visits to be added in the list (from, to,...)
3. Select Print Report from the File menu.
Electronic reports (*.pdf)
If an Adobe PDF writer “Printer Driver” is installed and set as the default printer, it is
possible to store the printout report automatically in any location of the HD or
eventually LAN paths according to a customizable filename format.
It is possible to define the created filename format selecting File/Page Set up... (see
Page set-up).
Export data
With this function you can export the test data in 4 different formats:
• *.txt (ASCII)
• *.xls (Microsoft Excel)
• *.wk1 (Lotus 123)
• *.xpo (Cosmed)
Export a test
1. Select Export tests from the File menu.
2. Select the test to export from the list box and press OK.
3. Type the name and the format of the file in the dialog Save as. If the ASCII format
is selected, the Text button in the dialog box Save as allows you to configure the
separators for character based files.
With the *.xpo Cosmed file format it is possible to import data from another Quark
archive. Press OK to confirm.
4. Select the folder for the export and type the file name. Press OK to confirm. A
status bar will show the file creation.
Chapter 6 - Spirometry - 71
72 - Quark PFT User Manual
Lung Volumes
Lung volumes tests (N2 washout)
In the PFT lab the measurement of “Lung Volumes” usually refers to the measurement
of Total Lung Capacity (TLC), residual volume (RV), functional residual capacity
(FRC), and Vital Capacity (VC). These measurements are essential to assess lung
function, they are important for the diagnosis of restrictive disorders.
FRC is most commonly determined with one of three basic techniques:
1. Multiple-breath open circuit Nitrogen washout
2. Body Plethysmography
3. Multiple-breath closed circuit Helium washout
COSMED Quark PFT uses the nitrogen wash-out technique to measure Lung volumes.
This method involves removing or “washing-out” the N2 gas present in a patient’s lung
by having the patient breathe 100% O2 for several minutes (usually 7 minutes or until
the final concentration of N2 is below 1.5-2.5%). Since the whole expired nitrogen
come from the initial lung volume, a simple equation allows to get FRC value.
The main parameters measured during this test are:
FRC Functional Residual Capacity
CEV Total ventilated volume during the test
Wash-out time Test duration
RV Residual Volume
TLC Total Lung Capacity
RV/TLC Motley index
Recommendations
• The system must be calibrated before each test (See Calibration chapter)
• The flowmeter must be connected properly to the breathing valve and the soft
mouthpiece must be properly fixed to the corresponding adapter.
• Check that the O
6 bar.
Setting N2 wash-out options
1. Select Settings... from Options menu.
2. Select Multi-breath N
3. Set the desired value in all the fields described in the next table.
cylinder is open and its output pressure is adjusted in the range 5-
2
Wash-out
2
74 - Quark PFT User Manual
Perform the test
Option Description
End-of-test criterion FetN
value under which the wash-out test is concluded
2
automatically
FRC Extrapolating Enable the extrapolation of FetN
to the end-of-test
2
criterion
Vt stability threshold Recognition criterion of regular respiration
Breathing Depth Range of respiration control bar during FRC (tidal
volume)
Breathing Frequency Range of respiration control bar during FRC
(respiratory frequency).
Pressing the System… button, you can set the following options:
Option Description
System dead space Add the volume of the viral filter eventually used
O2 cylinder impurities Refer to the used O2 cylinder analysis certification
1. Select Multi-Breath N
wash-out from the Test menu and wait for the green led is
2
prompted on the right side of the screen.
2. Connect the patient to the breathing valve and invite him/her to breathe normally.
3. Wait for the message “carry out slow vital capacity” is prompted on the screen of
the PC and let the patient to perform a slow vital capacity.
4. Wait for a normal breathing pattern is again established and press Start (F2) to
begin the wash-out with oxygen.
5. Wait for the final Nitrogen concentration (FetN
) reaches a value lower than 2.5%
2
and press Stop (F3) to conclude the test.
6. In order to visualise the graph and the main parameters press the follow buttons:
view FRC graph
view Volume Time graph
view data of the test
view breath by breath data
At the end of the test, you can edit the test:
• Display the test to be modified
• Right click with the mouse and select Edit...to change the test properties (wash-out
volume, sample volume, effective breath hold time calculation...).
Chapter 7 - Lung volumes - 75
Closing Volume Test (CV)
The Patient perform one single inspiration of 100% O2 following a maximal expiration
(RV level) and then expires smoothly through a breathing system able to measure both
the instantaneous N
The expired nitrogen is plotted on a N
obtained. The graph presents 4 different phases:
Phase I: The gas expired during Phase I represents the oxygen coming from the
dead volume of the measurement system and the upper airways not
interested by any distribution phenomenon.
Phase II: Shows the N
airways and the alveolar volume where the initial N
the inspired oxygen. In this phase the Anatomic Dead Space is calculated
(Fowler method).
Phase III: It is called Alveolar Plateau. It shows the average N
resulting from the dilution of the air originally contained in the Residual
Volume with the inspired Oxygen. The slope of this plateau (about 1.5%
for normal subjects) is an useful index for assessing distribution disorders
(Patients affected by severe emphysema can reach 10% values).
Phase IV: The end of Phase III corresponds to the beginning of the so called
Closing Volume which represents the last section of the Vital Capacity
before the Residual Volume. The quick increase of the N2% belongs to
the emptying of the latest alveoli, richest of N
inspired O
The main parameters calculated during this test are:
CV Closing Volume
Delta N
% Delta N
2
concentration and expired flow.
2
/Volume graph from which test results are
2
% rising belonging to the spaces connecting the upper
2
has be diluted with
2
because not reached by the
. Increased values of CV can be related to lung pathologies.
2
% between 750ml and 1250ml of expiration.
2
2
concentration
2
Perform the test
Recommendations
• The system must be calibrated before each test (See Calibration chapter)
• The flowmeter must be connected properly to the breathing valve and the soft
mouthpiece must be properly fixed to the corresponding adapter.
• Check that the O
cylinder is open and its output pressure is adjusted in the range 5-
2
6 bar.
1. Select Single Breath O
from the Test menu and wait for the green led is prompted
2
on the right side of the screen.
2. Connect the patient to the breathing valve and invite him/her to breathe normally.
3. Wait for the message “carry out Slow Vital Capacity” is prompted on the screen of
the PC and let the patient to perform a slow vital capacity.
4. Wait for a normal breathing pattern is again established and press Start (F2) during
a maximal exhalation.
5. Ask him/her to inspire up to TLC level and then to expire slowly maintaining the
flow rate as much constant as possible (two lines corresponding to 250ml/min and
500ml/min can be used as a target interval).
On the right side of the graph it is displayed a colored column, representing the
instantaneous flow. This column is green if the flow is within the acceptable range,
otherwise it is red.
6. The test stops automatically.
7. In order to visualise the graph and the main parameters press the follow buttons:
76 - Quark PFT User Manual
View CV graph.
view Volume Time graph
view data of the test
At the end of the test, you can edit the test:
• Display the test to be modified
• Right click with the mouse and select Edit...to change the test properties (wash-out
volume, sample volume, effective breath hold time calculation...).
Chapter 7 - Lung volumes - 77
78 - Quark PFT User Manual
P0.1 Respiratory
drive
P0,1 (Respiratory drive)
Aim of the test
The test named P0,1 or Respiratory Drive consists of evaluating the reduction of
pressure caused by the patient's inhalation.
In order to evaluate the P0,1 the subjects breaths in a mouthpiece provided with a
shutter and a pressure detector, by determining a shut down of the respiratory ways for
100 msec at the beginning of the inhalation, its possible to measure the reduction of
pressure. The shut down is so short that the patient keeps breathing normally without
realising it. Many other measurement can be carried out making the patient inhale
increasing concentration of CO
Setting respiratory drive options
1. Select Settings... from Options menu.
2. Select Respiratory drive
3. Set the desired value in all the fields described in the next table.
and verifying the results.
2
Option Description
Vt Stability threshold Recognition criterion of regular respiration
Recommendations
• Connect the P0.1 valve on the “breathing valve” connector on the front panel of the
• Verify that the cylinder driving the P0.1 valve (O
Perform the test
1. Select Respiratory Drive from the Test menu and wait for the green led is
2. Breath at rest until the message “Press F2...” is prompted.
3. Wait for the Patient establish a regular breathing pattern and press F2 whenever
4. Repeat the above mentioned manoeuvre several time both in room air or with
5. Press F3 to end the test.
80 - Quark PFT User Manual
Quark system.
-CO2) is open.
2
prompted on the right side of the screen.
you want to close the inspiratory valve for measuring the P0.1 parameter.
enhanced CO
inspiratory gas.
2
6. To visualise the V/t graph and the main parameters press the follow buttons:
view P0.1 graph
view Volume Time graph
view breath by breath data
7. Press the Abort button to stop the acquisition discarding the results.
Chapter 8 - P0.1 Respiratory drive - 81
82 - Quark PFT User Manual
Lung diffusing
capacity
CO Diffusing Capacity (DLCO) Test
The process of diffusion is defined as the flow of particles from an area of higher
concentration to an area of lower concentration. The measurement of diffusion, as
performed in pulmonary function laboratories, provides information about the transfer
of gas between the alveoli and the pulmonary capillary blood. The two major gases
involved in lung diffusion (O
capillary membrane and the blood plasma-red blood cell barrier.
The rate of diffusion across these primarily liquid barriers is limited by the surface area
for diffusion, the distance the gas molecules must travel, the solubility coefficient of the
gases in the liquid, the partial pressure difference (gradient) between air and blood and
the density of each gas.
Measurements of the diffusing capacity is usually performed using Carbon Monoxide
(CO) because it has a great affinity for Haemoglobin (210 times that of O
in blood and its concentration in venous blood is insignificant.
The main techniques are:
1. Single Breath with Apnoea (ATS/ERS standard)
2. Single Breath without Apnoea (requires a fast response CO analyser)
3. Multiple Breath in Steady-State
Setting DLCO options
1. Select Settings... from Options menu.
2. Select Single Breath CO diffusing capacity
3. Set the desired value in all the fields described in the next table.
2
and CO2) must move through two barriers: the alveolar-
), it is soluble
2
Option Description
Apnoea time ATS and ERS standard: 10 s
Effective time Ogilvie
Vt Stability threshold Recognition criterion of regular respiration
1
2
3
84 - Quark PFT User Manual
1
, Jones & Meade2, ESP3
from the beginning of the inspiration till the beginning of sampling
from 70% of inspiratory time, to 50% of sampling
from 50% of Inspired Volume to the beginning of sampling
Pressing the System… button, you can set the following options:
Option Description
System dead space Add the volume of the viral filter eventually used
concentration O2 content of the used mixture (ATS 21%, ERS
O
2
Pressing the Patient… button, you can set the following options:
Option Description
Physiologic dead space calculation criterion (ATS=2.2*weight, manual)
• The system must be calibrated before each test (See Calibration chapter)
• The flowmeter must be connected properly to the breathing valve and the soft
mouthpiece must be properly fixed to the corresponding adapter.
• Check that the DLCO cylinder is open and its output pressure is adjusted in the
range 5-6 bar.
DLCO with breath-hold (standard technique)
The Patient carry out a maximal inspiration of a mixture containing a low concentration
of CO (0.3%) and a tracer gas which is not metabolised (CH
asked to breath hold for 10 seconds (the expiratory valve is kept closed) and finally he
is allowed to expire slowly up to the Residual Volume (the expiratory valve is opened).
Analysing the concentrations of CO and the Tracer gas (CH
the expired volume (collected volume) obtained by discarding the beginning of the
expiration (wash-out volume), the test indexes are determined.
The “wash-out volume” (0.75-1 L for normal subjects, 0.25-0.5 L with reduced Vital
Capacity ), the “collected volume” (0.5-1 L) and the calculation of the “breath hold
time” (ESP, JONES, OGILVIE) have been standardised by both ATS and ERS.
If fast response CO and CH
the “wash-out” and “collected” volumes according to the Vital Capacity of the subject.
This is usually achieved operating graphically on the CO/CH
Repeating the DLCO test with different mixtures of inhaled gas it is possible to
separates the contributes of the alveolar membrane (Dm) and the Capillary Blood
Volume (Vc)
The main parameters of the DLCO test are:
DLCO Lung Diffusing Capacity
DLCOcorr Lung Diffusing Capacity corrected for Hb, COHb and PB
VA Alveolar Volume
DLCO/VA Krogh Index
Dm Alveolar Membrane
Vc Capillary Blood Volume
analysers are used it is possible to optimise the selection of
4
17%)
women and children)
, Methane); then he is
4
) belonging to a portion of
4
graphs.
4
DLCO with apnoea (standard method)
1. Remove the sampling line from the flowmeter
2. Select Single Breath CO diffusing capacity from the Test menu
3. Confirm or in case change the test options:
- Breath hold time
- Simulated test (it allows to carry out the manoeuvre without inhaling gas, useful
for teaching purposes)
- Hb (haemoglobin)
- Anatomic dead space
Chapter 9 - Lung diffusing capacity - 85
4. Wait for the zeroing of the analyser is completed and plug in again the sampling
line in the flowmeter
5. Connect the Patient to the mouthpiece and let her/him breathing at rest until the
message “Carry out Vital Capacity” is prompted
6. Carry out a Slow Vital Capacity if desired (this is not necessary for completing the
DLCO test)
7. Invite the Patient to exhale up to Residual Volume and press F2 during the
maximal expiration
8. Coach the Patient to inhale up to Total Lung Capacity and to hold the breath until
the expiratory valve will be automatically open; a quick full expiration shall follow
and the test will be completed.
A dashed line indicates the 90% of the Best Vital Capacity (if a FVC or SVC test is
available in the archive, otherwise it indicates the 90% of the Predicted VC), in order to
verify the correctness of the manoeuvre.
Double-click with the left mouse button in order to modify the test options (wash-out
volume, sampling volume, calculation of the real apnoea time…). Checking the box
ATS95 Criteria, it is possible to set automatically 1 litre for wash-out and sampling
values, according to the ATS95 criteria.
86 - Quark PFT User Manual
DLCO without apnoea (“intrabreath” method)
1. Remove the sampling line from the flowmeter
2. Select Single Breath CO diffusing capacity from the Test menu
3. Confirm or in case change the test options:
- Breath hold time (deselect the check box)
- Simulated test (it allows to carry out the manoeuvre without inhaling gas, useful
for teaching purposes)
- Hb (haemoglobin)
- Anatomic dead space
4. Wait for the zeroing of the analyser is completed and plug in again the sampling
line in the flowmeter
5. Connect the Patient to the mouthpiece and let her/him breathing at rest until the
message “Carry out Vital Capacity” is prompted
6. Carry out a Slow Vital Capacity if desired (this is not necessary for completing the
DLCO test)
7. Invite the Patient to exhale up to Residual Volume and press F2 during the
maximal expiration
8. Coach the Patient to inhale up to Total Lung Capacity and to expire slowly
maintaining the flow rates within the two target lines shown on the real-time
window until the Vital Capacity has been exhaled.
A dashed line indicates the 90% of the Best Vital Capacity (if a FVC or SVC test is
available in the archive, otherwise it indicates the 90% of the Predicted VC), in order to
verify the correctness of the manoeuvre.
Double-click with the left mouse button in order to modify the test options (wash-out
volume, sampling volume, calculation of the real apnoea time…).
Multiple breath DLCO test (“Steady-State” method)
In this test, the patient breathes regularly during the whole test (about 8 min): graphs
and results will be displayed breath by breath.
Normally, during the first 2 minutes the patient inhales ambient air, then such a mixture:
0.1% CO, 0.1% CH4, 21% O2, N2 balance.
1. Remove the sampling line from the flowmeter
2. Select CO diffusing capacity steady state from the Test menu
4. Wait for the zeroing of the analyser is completed and plug in again the sampling
line in the flowmeter
5. Connect the Patient to the mouthpiece and let her/him breathing at rest until the
message “Carry out Vital Capacity” is prompted
6. Carry out a Slow Vital Capacity if desired (this is not necessary for completing the
DLCO test)
7. While the patient is exhaling, press F2 in order to start the mixture
8. When the steady state is reached, press F3 to end the test.
Right-clicking you can generate new graphs and parameters to be displayed, both during
the test and at the end.
Chapter 9 - Lung diffusing capacity - 87
88 - Quark PFT User Manual
MIP/MEP (option)
MIP/MEP test
The MIP/MEP test measures the maximum inspiratory/expiratory pressure of the
patient. He/She breathes normally until the start of the test, perform a maximal
expiration/inspiration and then inspires/expires with the maximum force against a valve
which does not allow the air flow.
The measured parameters are:
MIP Maximum inspiratory pressure
MEP Maximum expiratory pressure
How to perform the test
1. Disconnect the breathing valve from the front panel of the Quark
2. Connect the MIP/MEP valve to the breathing valve connector on the front panel of
3. Select MIP/MEP from Test menu and wait for the green led is prompted on the
4. Breathe normally for some time.
5. Carry out a maximal expiration (for the MIP) or a maximal inspiration (for the
6. Press F2 during the expiration/inspiration.
7. Inspire/Expire with the maximum force against the shutter until it opens
8. Repeat the same procedure for the expiratory (inspiratory) maneuver. It is not
9. Press F3 or wait the end of the test, so that the software displays the graphs, the
the Quark and insert the flowmeter into the valve. Use a rubber mouthpiece
connected to the antibacterial filter and nose clip.
right side of the screen.
MEP).
automatically.
necessary to repeat the test since both MEP and MIP can be conducted during the
same acquisition.
measured parameters, and the predicteds values.
90 - Quark PFT User Manual
Pulse oximetry
(option)
The oximetry test
The oximetry test measures the haemoglobin saturation, i.e. the percentage of the blood
haemoglobin bearing oxygen. The test can be performed at rest or during/after a light
exercise phase (cycling, jogging).
This test is not intended for measuring the oximetry during exercise tests. For this use,
see chapter Exercise testing.
The test is completely automatic.
The measured parameters are:
SpO
HR Heart rate
How to perform the test
1. Connect the pulse oximeter to auxiliary RS232 port on the back of the Quark PFT.
2. Connect the probe to finger or ear of the Patient (different probes are available).
3. Select Oximetry from the Test menu and wait for the green led is prompted on the
4. In the first part of the test the HR and SpO
Haemoglobin saturation
2
right side of the screen.
values, together with a
2
plethysmographic graph, are displayed. The graph monitors the quality of the
signal.
5. Wait for an acceptable quality of the real-time trace and press F2 or the Start
button to start the data acquisition.
6. Every 5 seconds, the HR and SpO
values are stored.
2
7. After the test, press F3, so that the software displays the graphs and the measured
parameters.
92 - Quark PFT User Manual
Exercise testing
Recommendations for the exercise testing
The evaluation of the cardiorespiratory function
The physical training requires the interaction of physiological mechanisms that allow
the cardiovascular and respiratory systems to supply the increasing demand of energy
due to the contraction of the muscles.
During the training the systems are both engaged, an adequate answer to the effort is the
measure of theirs health state.
The increase of the metabolic rate, during the exercise, needs an appropriate increase of
oxygen in the muscles. At the same time, the CO
in order to avoid the lactic acid making.
To satisfy the increase in the gas exchange, necessary to the muscles during the
exercise, is requested the intervention of many physiological mechanisms. This process
involves lungs, the pulmonary circulation, the heart and the peripheral circulation.
Precautions
The physician has the responsibility that the patient subjects to the test is a suitable
person able to execute an effort test.
Laboratory
The room, in which the test is performed, must be big enough to contain the whole
necessary equipment, allowing an easier accessibility to the patient in case of
emergency.
In the room should be placed a thermometer and a hygrometer; the heart frequency and
the perceived values of the effort rise as much as the ambient temperature increases, and
the variability of the cardiovascular response grows for humidity values higher of 60%.
Generally it is considered 22°C the temperature adequate for the test execution, even for
short efforts, values till 26°C can be considered acceptable in presence of an efficient air
ventilation.
muscles production must be removed
2
Ending the test
The patient should be monitored with ECG for at least 8 minutes, in resting conditions
or until he returns to the pre-exercise conditions.
Preparing the patient
To enhance the value of a diagnostic test it’s very important patient collaboration. In
most cases a well-informed patient will make a better effort (in relation to his
conditions) and will allow a reliable interpretation of the test. For this reason every
ergometric test must be preceded from a precise training of the patient.
Before testing
The physician applying the exam must be provided with a written request including a
brief description of the diagnosis (confirmed or suspected ), the request’s reason and the
patient therapy carried out showing the dose and time of the drug assumption.
To standardise the response to the test and reduce the patient’s anxiety it’s suggested to
provide him either written (before the exam) or oral (at the same time of the test)
information. At the scheduling time detailed instructions should be delivered to the
patient, consisting in smoke and food abstinence three hours before an ergometric test,
or eight hours before a scintigraphic test.
Test are usually executed supporting the therapeutic outline in progress, but sometimes
it could be necessary to stop some drugs, such as b-block or calcium antagonist, which
could impair the effort response reducing the diagnostic accuracy of the exam.
The patient must wear comfortable suit and gymnastic shoes and two hours before test
stop any kind of drugs, eat light and avoid coffee and smoke.
94 - Quark PFT User Manual
It’s very important acquire information on the patient’s clinical past before performing
the test. Keep attention in particular way to the use of drugs, tobacco, to the physical
fitness and symptoms produced with the exercise.
Patient assent
The patient is informed that he will be submitted to a maximum effort, which could be
stopped at any moment, and of the risks of the test execution.
Ending the test
Test may end when the maximum value of the oxygen consumption has been reached
and the patient’s response established.
Chapter 12 - Exercise Testing - 95
Start Testing
Note: Selecting “Simulated
test”, the software allows to
use the gas calibration in order
to perform the test. This is
useful if the user would check
the accuracy of gas
measurements.
Before starting exercise test type a new patient information or choose one from the list
of patient in the file. As soon as a patient has been entered the software is ready to start
a test. The name of the active patient is shown on the status bar of the program window.
Start a test
1. Calibrate the analysers as described in the Calibration chapter.
2. Choose Execute Test from the Test menu.
3. Enter or modify the antropometric data of the patient and select the options you
4. Select the ergometer you need to control
5. The software environment will change showing a new Menu bar and toolbar while
6. At this point the software starts showing data on the monitor but without saving
Abort the test without saving data
Choose Abort from the Test menu or press Alt+F3.
End the test saving data
1. Choose End from the Test menu or press F3.
2. Choose Yes to end the test or No to continue.
View data in real-time
The visualisation features and capabilities of the data and graphs are identical to the
ones described in the Data management chapter. Starting the test a small window will
appear on the right corner displaying time, bmp and, if selected before, the ergo
protocol and trainer.
need to carry out the test . To use a specific protocol choose it in the list box and
press OK to confirm.
the first data will be displayed in a table format.
them, this in order to monitor the patient before starting the test. To start storing
data press F2.
96 - Quark PFT User Manual
Parameters to view
View graphs in real-time
1. Choose Graph from the View menu.
2. Follow the instructions described in data management section to edit the graphs.
While the test is running, it is possible to choose the parameters to view.
1. Select Parameters to view/Test execution… from the Options menu.
2. Select the parameters and confirm.
Manual protocol
Tip: pressing the Shift key
while choosing the marker
option will allow you to enter
the label for that marker.
Automatic protocol
Set the BPM alarm
If you are using the Quark with a treadmill without serial interface, it is possible to enter
manually from the PC the event, the phase and the marker.
Enter Load and Phase
1. During the test select Load from the Events menu.
2. Select the phase and/or type the value of the load and press OK to confirm.
Set the markers
Select Marker from the Events menu.
The software allows to automatically control the ergometer according to the protocol
previously selected. Anyway it is allowed to change it even after the test is started.
Modify the load during the test
1. During the test choose Ergometric protocol from the Events menu.
2. Select the row corresponding to the desired load and press OK to confirm.
The software allows the user to set the alarm level for the heart rate, in order to monitor
the patient response.
Enter the BPM
1. Choose BPM alarm from the Events menu.
2. Set the alarm by moving the scroll bar and press OK to confirm.
It also allows to enable or disable the acoustic alarm by the option "Acoustic alarm".
Chapter 12 - Exercise Testing - 97
Data management
As soon as the test has been completed, all data stored can be retrieved for a complete
management.
Viewing data
Data can be viewed in the following formats:
Table form numeric values of the various parameters (columns) corresponding to
Graphic form graphical presentation on Y1, Y2, X charts.
Summary results of the test and statistical analysis of the blocks.
Predicted predicted values, maximum value measured.
View data in table form
1. Select Data… from the View menu.
2. Select the test to visualise in the list box and press OK
Note: Double-click in the
window to open the edit test.
each step (rows).
Creating graphs
The software is provided with powerful functions for creating charts. You can add
custom graphs to create exactly what you need.
View data in graph form
1. Choose Graph… from the View menu.
2. Select the tests to visualise from the list and press OK
3. Choose the parameters you require on the X, Y1 and eventually for Y2; select if
necessary some of the following options by pressing the more button and press OK
to confirm.
It is possible to access quickly 5 common graphs from the View/Graph… dialog box.
98 - Quark PFT User Manual
Note: Double-click in the graph
window to open the edit test
Right-clicking, the graph can be exported in bmp file format.
Customise the graphs
1. With a graph on the screen, choose Customise graph from the View menu.
2. On the Customise graph dialog box, select options to obtain the wished graph.
Option Function
Grid on X, Y axesshow the grid lines in correspondence with x or y axes that
make the graph easier for you to analyse data.
Autoscale maximum and minimum values of the graph will be measured
automatically.
Ignore 0 points with 0 value measured won't be shown.
Not interpolated make the graph scattered.
Marker highlight with a symbol all steps of the test in which the marker
button was pressed.
Squared makes the graph a square
Without recovery exclude from the graph all points of the recovery phase.
Mark points marks each point with a symbol
Min. Max. allows to set manually the axes values.
Step Set the axes' scale step.
Switch from graph to data and vice versa
When the active window is a graph (or a report in data form), it is possible to view very
quickly the data (or the graph) corresponding to that test.
Choose Current test data (if the active window is a graph) or Current test graph…
(if the active window is a data report) from the View menu.
Chapter 12 - Exercise Testing - 99
Viewing predicted values
For some parameter it is possible to compare the maximum value measured during the
test with its predicted value and the LT value both in percentage and absolute.
View predicted values
Choose Predicted from the View menu.
Anaerobic (Lactate) Threshold detection
The software allows to detect the Lactate Threshold (Anaerobic Threshold) according to
the "Modified V-slope method" reference. The LT can be detected both manually and
automatically.
Note: Double-click in the
window to open the
corresponding dialog box.
View the Lactate Threshold
Choose Lactate Threshold from the View menu.
Detect the Lactate Threshold
1. Choose Calculate LT from the Test menu.
2. For calculating it automatically on the "Lactate Threshold" dialog box click on the
Auto detect button.
3. For adjusting manually the point you want to detect, move the scroll bar on the
dialog box by pressing the arrow buttons. Data and graph of the LT will be
automatically redrawn.
4. Press Ok button to save your choices.
Customise graphs for the LT viewing
The software allows to customise two of the three graphs for the LT visualisation.
1. Choose Lactate Threshold from the Options menu.
2. Choose the parameters you want to be shown on the LT window and press OK to
100 - Quark PFT User Manual
confirm your choices.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.