COSMED does not assume any liability for end user interpretation of this user manual or for casual or
consequential damages in connection with the provision, representation or use of this documentation.
No parts of this manual may be reproduced or transmitted in any form without the expressed written
permission of COSMED Srl.
Each copy of the COSMED Software can only be installed on one computer.
Excel is a registered trademark of Microsoft Corporation.
DBIII is a registered trademark of Bordland International Inc.
Lotus 123 is a registered trademark of Lotus Development Corporation .
Quark b2 is an electrical medical device designed to perform pulmonary function tests.
It is to be used by physicians or by trained personnel on a physician responsibility.
Caution: Federal law restricts this device to sale by or on the order of a physician.
This equipment has been conceived with the aim of providing an auxiliary instrument
allowing:
• the formulation of lung pathology diagnosis;
• important studies concerning human physiology;
• the collection of important information in sport medicine.
No responsibility attaches COSMED Srl for any accident happened after a wrong use of
the device, such as:
• use by non qualified people;
• non respect of the device intended use;
• non respect of the hereunder reported precautions and instructions.
Warnings
The device, the programme algorithms and the presentation of measured data have been
developed according to the specifications of ATS (American Thoracic Society) and
ERS (European Respiratory Society). Other international references have been followed
when these were not available. All bibliography references are reported in Appendix.
The present handbook has been developed with respect of the European Medical Device
Directive requirements which sort Quark b
It is recommended to read carefully the following precautions before putting the device
into operation.
The precautions reported below are of fundamental importance to assure the safety of
all COSMED equipment users.
1. This user manual is to be considered as a part of the medical device and should
2. Safety, measure accuracy and precision can be assured only:
• using the accessories described in the manual or given with the device. Actually
• ordinary equipment maintenance, inspections, disinfection and cleaning are
• any modification or fixing is carried out by qualified personnel;
• the environmental conditions and the electrical plants where the device operates
3. Before powering the system, check the power cables and the plugs. Damaged
4. Large gas cylinders, which may be given by the manufacturer or purchased by the
5. When removing the protective cap, inspect the cylinder valve for damaged threads,
6. Be certain that the materials of the pressure regulators are chemically compatible
2
within Class II a.
always be kept on hand.
non recommended accessories can affect safety unfavourable. Before using non
recommended accessories it is necessary to get in touch with the manufacturer;
performed in the way and with the frequency described;
are in compliance with the specifications of the manual and the present regulations
concerning electrical plants. In particular grounding reliability and leakage current
suppression can only be assured when the device three – wire receptacle is
connected to a yellow - green return connected to earth ground. Attempting to
defeat the proper connection of the ground wire is dangerous for users and
equipment.
electrical parts must be replaced immediately by authorised personnel.
customer, should be secured with cylinder safety chains or safety stands.
dirt, oil or grease. Remove any dust or dirt with a clean cloth. If oil or grease is
present on the valve of a cylinder which contains oxygen, do not attempt to use.
Such combustible substances in contact with oxygen are explosive.
with the intended gas service before installation. Inspect the regulator for the proper
Chapter 1 - Getting started - 13
connection and note the ranges of the pressure gauges. Also examine the physical
condition of the regulator including threads and fittings. Remove any dust or dirt
from the regulator or cylinder valve with a clean cloth. Do not install a regulator on
a cylinder valve containing oxygen if grease or oil is present on either. Such
substances in contact with oxygen are explosive.
7. Cleaning residue, particulates, and other contaminates (including pieces of torn or
broken components) in the breathing circuit pose a safety risk to the patient during
testing procedures. Aspiration of contaminates can potentially be life-threatening.
Use disposable anti-bacterial filters or disinfect each part in contact with the patient
before each test.
8. You must follow all the cleaning procedures in System Maintenance, and you must
thoroughly inspect the components after cleaning and before each patient test.
9. This device is not suitable for use in presence of flammable anaesthetics. It is not
an AP nor an APG device (according to the EN 60 601-1 definitions).
10. Keep the device away from heat and flame source, flammable or inflammable
liquids or gases and explosive atmospheres.
11. In accordance with their intended use Quark b
2
is not to be handled together with
other medical devices unless it is clearly declared by the manufacturer itself.
12. It is recommended to use a computer with electromagnetic compatibility CE
marking and with low radiation emission displays.
13. It is necessary to make the PC, connected to the
Quark b2, compliant with EN
60601-1 by means of an isolation transformer.
14. The Quark b
2
needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in the section
EMC.
15. Portable and mobile RF communications equipment can affect the Quark b
2
.
16. Use only the cable and accessories supplied with the equipment. The use of
accessories and/or cables other than those supplied may result in increased
emissions or decreased immunity of the equipment.
17. The Quark b
adjacent or stacked use is necessary, the Quark b
2
should not be used adjacent to or stacked with other equipment. If
2
should be observed to verify
normal operation in the configuration in which it will be used.
18. Graphical symbols used in accordance to present specifications are described here
below:
Equipment type B (EN60601-1)
Equipment type BF (EN60601-1)
Danger: high temperature
OFF
ON
Protective earth ground
Alternating current
14 - Quark b2 User Manual
Contraindication
The physical strain to execute the respiratory manoeuvre is contraindicated in case of
some symptoms or pathology. The following list is not complete and must be
considered as a piece of mere information.
Contraindications for the Spirometer tests
Absolute contraindications
For FVC, VC and MVV tests:
•Post-operating state from thoracic surgery
For FVC tests:
• Severe instability of the airways (such as a destructive bronchial emphysema)
• Bronchial non-specific marked hypersensitivity
• Serious problems for the gas exchange (total or partial respiratory insufficiency)
Relative contraindications
For FVC tests:
• spontaneous post-pneumothorax state
• arterial-venous aneurysm
• strong arterial hypertension
• pregnancy with complications at the 3
For MVV test:
•hyperventilation syndrome
rd
month.
Contraindications for Bronchial provocation tests
The bronchial provocation tests must be executed according to the doctor’s discretion.
There are not data that reveal specific contraindication for the bronchial provocation test
through inhalation.
The modern standard processes have been revealing secure in several clinical studies.
However it is recommendable to respect the following contraindications:
Absolute contraindications
• Serious bronchial obstruction (FEV1 in adults)
• Recent myocardium infarct
• Recent vascular-cerebral accident
• Known arterial aneurysm
• Incapacity for understanding the provocation test procedures and its implications.
Relative contraindications
• Bronchial obstruction caused by the respiratory manoeuvre.
• Moderate or serious bronchial obstruction. For ex. FEV1 < 1.51 in men and FEV1
in women < than 1.21.
• Recent infection in the superior air tracts
• During the asthmatic re-acuting
• Hypertension
• Pregnancy
• A pharmacology treatment epilepsy
Contraindications for Exercise testing
Read carefully the exercise testing chapter.
Chapter 1 - Getting started - 15
Environmental condition of use
COSMED units have been conceived for operating in medically utilised rooms without
potential explosion hazards.
The units should not be installed in vicinity of x-ray equipment, motors or transformers
with high installed power rating since electric or magnetic interferences may falsify the
result of measurements or make them impossible. Due to this the vicinity of power lines
is to be avoided as well.
Cosmed equipment are not AP not APG devices (according to EN 60601-1): they are
not suitable for use in presence of flammable anaesthetic mixtures with air, oxygen or
nitrogen protoxide.
If not otherwise stated in the shipping documents, Cosmed equipment have been
conceived for operating under normal environmental temperatures and conditions [IEC
601-1(1988)/EN 60 601-1 (1990)].
• Temperature range 10°C (50°F) and 40°C (104°F).
• Relative humidity range 20% to 80%
• Atmospheric Pressure range 700 to 1060 mBar
• Avoid to use it in presence of noxious fumes or dusty environment and near heat
• Adequate floor space to assure access to the patient during exercise testing.
• Adequate ventilation in the room.
16 - Quark b2 User Manual
EMC
Guidance and manufacturer’s declaration - electromagnetic emissions
The Quark b2 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Quark b
2
should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Quark b
function. Therefore, its RF emissions are very low and
2
uses RF energy only for its internal
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Harmonic Emission
IEC 61000-3-2
Voltage Fluctuations /
Class B
Class A
Complies
The Quark b
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
2
is suitable for use in all establishments,
Flicker Emission
IEC 61000-3-3
Guidance and manufacturer’s declaration - electromagnetic immunity
The Quark b2 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Quark b
Immunity test Test level Compliance level
2
should assure that it is used in such an environment.
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
<5% UT
(>95% dip in U
for 0.5 cycles
40% U
T
(60% dip in U
5 cycles
70% U
T
(30% dip in U
25 cycles
<5% UT
(>95% dip in U
for 5 sec
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
<5% U
)
(>95% dip in U
T
for 0.5 cycles
40% U
) for
(60% dip in U
T
5 cycles
70% U
) for
(30% dip in U
T
25 cycles
<5% UT
(>95% dip in U
)
T
for 5 sec
3 A/m 3 A/m
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
T
Mains power quality should
)
be that of a typical
T
commercial or hospital
T
T
environment. If the user of the
Quark b
) for
T
operation during power mains
2
requires continued
interruptions, it is
recommended that the Quark
) for
T
2
b
be powered from an
uninterruptible power supply
or a battery.
)
T
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
Chapter 1 - Getting started - 17
Guidance and manufacturer’s declaration - electromagnetic immunity
The Quark b2 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Quark b
Immunity test Test level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Veff
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
2
should assure that it is used in such an environment.
Compliance
level
3 V
3 V/m
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the Quark b
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter
Recommended separation distance
d=1.17
d=1.17
d=2.33
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site surveya, should be
less than the compliance level in each
frequency rangeb.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
P
P 80 MHz to 800 MHz
P 800 MHz to 2.5 GHz
2
, including cables,
Notes:
(1) At 80 MHz, the higher frequency range applies.
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Quark b
exceeds the applicable RF compliance level above, the Quark b
normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the Quark b
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
2
2
should be observed to verify
.
2
is used
18 - Quark b2 User Manual
Recommended separation distances between portable and mobile RF communications
equipment and the Quark b
The Quark b2 is intended for use in an environment in which radiated RF disturbances are
controlled. The customer or the user of the Quark b
2
2
can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Quark b2 as recommended below, according to
the maximum output power of the communications equipment..
Rated maximum
output power of
transmitter (W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
d=1.17
P
80 MHz to 800 MHz
d=1.17P
800 MHz to 2.5 GHz
d=2.33P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.70 3.70 7.38
100 11.70 11.70 23.33
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Notes:
(1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Chapter 1 - Getting started - 19
Safety and conformity
Safety
IEC 601-1 (1988) /EN 60 601-1 (1990);
Find reported below the complete classification of the device:
• Class I type B device
• Protection against water penetration: IP00, ordinary equipment unprotected against
water penetration
• Non sterile device
• Device not suitable in the presence of flammable anaesthetics;
• Continuous functioning equipment;
EMC
The system meets the EMC Directive 89/336
EN 60601-1-2
EN 55011 Class B (emission), IEC 1000-4-2, IEC 1000-4-3, IEC 1000-4-4
Quality Assurance
UNI EN ISO 9001:2000 (Registration n° 387-A Cermet)
UNI EN ISO 13485:2003 (Registration n° 387-M Cermet)
Medical Device Directive (CE mark)
MDD 93/42/EEC (Notified Body 0476).
Class IIa
20 - Quark b2 User Manual
Keynotes
Here are the keynotes used to make the manual easier to read.
Typographic keynotes
These are the typographic keynotes used in the manual.
Style Description
Boldindicates a control or a key to be pressed.
“Italic” indicates a messages shown by the firmware.
Graphic keynotes
These are the graphic keynotes used in the manual.
Illustration Description
shows the button to click in the software to activate the related feature.
Chapter 1 - Getting started - 21
System overview
Main features
The Quark b2 unit contains the main items (analysers, electronic controls, valves...). It is
connected to the PC through an RS232 interface. The modular architecture allows to
replace the boards without any other technical service.
The unit can be used on a desktop or on the optional trolley, ideal for the Quark b
cylinders, the PC, the printer…
Flowmeter
The system uses a bi-directional digital turbine. It opposes a very low resistance to flow
(<0,7 cmH
motion which causes the rotation of the turbine rotor. The rolling blade interrupts the
infrared light beamed by the three diodes of the optoelectronic reader. Every
interruption represents 1/6 turn of the rotor, this allows to measure the number of turn in
the time.
Gas analysers
The O2 and CO2 analysers are temperature-controlled and the internal pressure and
expired flow are monitored for an higher reliability if the measurements.
The Quark b² uses Nafion Permapure ® which is a semipermeable capillary tube
capable of removing the humidity in excess without altering the gas concentrations..
The analysers calibration is automatic and shows both graphically and numerically the
flow and concentration signals and the accuracy of the baseline/gain.
2
, the
O/l/s to 12 l/s). The air passing through the helical conveyors, takes a spiral
2
Environmental conditions measurements
The Barometric pressure is measured by an electronic pressure sensor inside the unit. It
has a measuring range between 400 mmHg and 800 mmHg. Ambient temperature and
relative humidity are measured by two sensors contained in the RH/TA probe. The
above measurements are used to calculate BTPS and STPD correction factors.
22 - Quark b2 User Manual
Before starting
Before operating the Quark b2 system we strongly recommend to check the equipment
and register you as a customer.
Checking the packing contents
Make sure that the package contains the items listed below. In case of missing or
damaged parts, please contact Cosmed technical assistance.
Before using the system, please take a moment to fill in the registration form and the
warranty and return them to COSMED, by doing this you are eligible to the customers
assistance service.
For further information, please refer to the enclosed registration and warranty form. If
the form is not enclosed in the packaging, please contact directly COSMED.
Register the product via software
Together with the PC software, a registration software is supplied. With this software it
is possible to fill in an electronic form with the customer information.
1. To run the software, double click on the icon Registration or select Registration…
from ? menu.
2. Type the requested information and click Send… to send the form via e-mail to
COSMED.
How to contact COSMED
For any information you may need, please contact the manufacturer directly at the
following address:
COSMED S.r.l.
Via dei Piani di Monte Savello, 37
P.O. Box n. 3
00040 - Pavona di Albano
Rome - ITALY
Eight analog inputs (0-5, 0-10 Volts) for external measurement (HR, Pressure, ...).
Digital (TTL) input for external ECG signal. Auxiliary RS 232 port for ergometers
control (cicloergometers, treadmill…).
Chapter 1 - Getting started - 27
Measurements
Pulmonary function tests and measured parameters
Breath by Breath exercise testing
Symbol UM Parameter
VO2 ml/min Oxygen Uptake
VCO2
Vt l Tidal Volume
FetO2 % End Tidal O2
FetCO2 % End Tidal CO2
R --- Respiratory Quotient
VE l/min Ventilation
HR 1/min Heart Rate
Qt l Cardiac output
AT --- Anaerobic Threshold
VE l/min Ventilation
SV l/min Stroke volume
RF 1/min Respiratory Frequency
FeO2, FeCO2 % Averaged expiratory concentration of O2 e CO
VE/VO2 --- ventilatory equivalent for O2
VE/VCO2 --- ventilatory equivalent for CO2
VO2/HR ml/beat Oxygen pulse
VO2/Kg ml/min/Kg VO2 per Kg
Ti, Te, Ti/Ttot sec time breaths
Vd/Vt --- Vd/Vt ratio
PaCO2 mmHg arterial PCO2 (estimated)
P(a-et)CO2 mmHg Delta PaCO2 – PetCO2
ml/min
Carbon Dioxide production
Indirect Calorimetry
Symbol UM Parameter
EE Kcal/day Energy Expenditure
EE/BSA Kcal/day/m2 Energy Expenditure/Body surface area
EE/Kg Kcal/day/Kg Energy Expenditure pro Kg
FAT Kcal/day Fats
CHO Kcal/day Carbohydrate
PRO Kcal/day Protein
FAT% % % Fat
CHO% % % Carbohydrate
PRO% % % Protein
npRQ —- Respiratory quotient not protein
Lactate Threshold (V-Slope)
Symbol UM Description
VO2 @ LT l/m Lactate (Anaerobic) Threshold STPD
R @ LT -- Respiratory Quotient @ LT
Time @ LT hh:mm:ss Time @ LT
VCO2 ml/min CO2 output @ LT STPD
VE l/min Ventilation @ LT BTPS
Chapter 2 - Measurements - 29
HR bpm Heart Rate @ LT
O2 Kinetics
Parameter UM Calculation
O2 deficit l/m VO2@work*tau
O2 debt l/m VO2'@work*tau
Spirometry Tests (option)
FVC - Forced Vital Capacity
Symbol UM Parameter
FVC l Forced Expiratory Vital Capacity
FEV1 l Forced Expiratory Volume in 1 sec
FEV1/FVC% % FEV1 as a percentage of FVC
PEF l/sec Peak Expiratory Flow
FEV0.5 l Forced Expiratory Volume in 0.5 sec
FEV6 l Forced Expiratory Volume in 6 sec
FEV1/FEV6 % FEV1 as a percentage of FEV6
FEV6/FVC% % FEV6 as a percentage of FVC
Best FVC l Best Forced Expiratory Vital Capacity
Best FEV1 l Best Forced Expiratory Volume in 1 sec
Best PEF l/sec Best Peak Expiratory Flow
Vmax25% l/sec Expiratory Flow when 75% of the FVC remains to be exhaled
Vmax50% l/sec Expiratory Flow when 50% of the FVC remains to be exhaled
Vmax75% l/sec Expiratory Flow when 25% of the FVC remains to be exhaled
FEF25-75% l/sec Mid-exp flow between 25-75%FVC
FET100% sec Forced expiratory time
FEV2 l Forced Expiratory Volume in 2 sec
FEV3 l Forced Expiratory Volume in 3 sec
FEV2/FVC% % FEV2 as a percentage of FVC
FEV3/FVC% % FEV3 as a percentage of FVC
FEV1/VC% % Tiffenau index
FEF50-75% l/sec Mid-exp flow between 50-75%FVC
FEF75-85% l/sec Mid-exp flow between 75-85%FVC
FEF0.2-1.2% l/sec Mid-exp flow between 0.2 l - 1.2 l
FiVC L Inspiratory Forced Vital Capacity
FiF25-75% l/sec Forced mid-inspiratory flow
FiV1 l/sec Forced Inspiratory Volume in 1 sec
PIF l/sec Peak Inspiratory Flow
VEXT ml Extrapolated Volume (back extrapolation)
PEFT msec Time to PEF (10% - 90%)
30 - Quark b2 User Manual
VC/IVC - Slow Vital Capacity and Ventilatory pattern
Symbol UM Parameter
EVC l Expiratory Vital Capacity
IVC l Inspiratory Vital Capacity
ERV l Expiratory Reserve Volume
IRV l Inspiratory Reserve Volume
IC l Inspiratory Capacity
VE l/min Expiratory Minute Ventilation
Vt l Tidal Volume
Rf 1/min Respiratory Frequency
Ti sec Duration of Inspiration
Te sec Duration of Expiration
Ttot sec Duration of Total breathing cycle
Ti/Ttot —- Ti/Ttot ratio
Vt/ti l/sec Vt/ti ratio
MVV - Maximum Voluntary Ventilation
Symbol UM Parameter
MVV l/min Maximum Voluntary Ventilation
MVt l Tidal Volume (during MVV)
MRf 1/min Maximum Respiratory frequency
MVVt sec MVV duration time
Bronchoprovocation Response
Symbol UM Parameter
FallFEV1 % Fall in FEV1 from baseline or post diluent
FallVmax50% % Fall in Vmax50% from baseline or post diluent
P10 —- Provocative dose causing FEV1 to fall 10% from baseline
P15 —- Provocative dose causing FEV1 to fall 15% from baseline
P20 —- Provocative dose causing FEV1 to fall 20% from baseline
Chapter 2 - Measurements - 31
Installation
Unpacking the system
On receipt of your system, you should immediately inspect your package for shipping
damages, in case damage is suspected please contact the reseller immediately. Your
system has to be installed by COSMED or by an authorised reseller.
Excessive amount of dust and miscellaneous clutter around the instrument can
eventually cause malfunctions due to overheating of components, we strongly suggest
to keep the unit in a clean environment and well ventilated as possible. Quark b
accessories can have place on the trolley as shown in the following illustration.
2
and
1. Quark unit
2. Printer
3. Calibration cylinder
4. Keyboard
5. Mouse pad
6. Table arm
7. Breathing valve
8. PC
9. Trolley
10. Bag
Chapter 3 - Installation - 33
Setting up the system
Before starting operating with the system make sure to meet the environmental and
operational conditions reported in Chapter 1.
Calibration Gas Cylinder
In order to calibrate the sensors you need to have available calibration cylinder with the
following gas concentration:
Cylinder Recommended Gas mixture
Calibration O2 16%, CO2 5%, N2 Bal
Gas pressure adjustment procedure
The gas regulator has an adjustable second stage that must be adjusted only when used
for the first time. This is necessary to protect the internal demand valve from the high
pressure surge that can be generated when the cylinder is opened.
1. Make sure that the regulator is turned off before opening the cylinder valve.
2. Open the cylinder valve by turning the valve counter-clockwise.
Notice: The cylinders must be
provided with a calibration
certificate, indicating the gas
3. Adjust the regulator pressure by turning the adjustment knob clockwise in order to
reach a value between 5 and 6 atmospheres (bar).
Notice: The gas cylinder must be replaced when the primary pressure gauge falls below
10 bar.
34 - Quark b2 User Manual
Connecting cables and tubing
The assembling procedure will be shown in the following illustrations.
A Flowmeter connector
B Power switch
C Sampling tube connector
D Calibration plug
E Oxymeter plug
F Cardiac Output plug
G, H, I, N Auxiliary analog inputs
L Calibration gas (16% O
, 5% CO2) connector
2
M Cardiac Output gas connector
O RH/TA probe connector
P RS232 Serial Port for PC connection
Q RS232 (for future extensions)
R HR probe connector
S RS232 Serial Port for ergometer connection
T Sampling pump input
U Power cable connector
V Switch 110V / 220V
Chapter 3 - Installation - 35
Cables and Tubing installation sequence
1. Connect the power cable of the Quark b2 to the trolley
2. Connect the power cable of the PC and the printer to the trolley
3. Connect the RS232 cable from the Quark b
COM2). If the PC does not have a RS232 port, please use the USB-RS232 adapter
(A 388 410 001), available as an accessory.
4. Connect the O
between 5 and 6 bar.
5. Connect the Flowmeter to the front panel of the Quark.
-CO2 calibration cylinder to the Quark b2 and adjust its pressure
2
2
(PC RS232) to the PC (COM1 or
36 - Quark b2 User Manual
Assemble the turbine
Assemble the mask and the flowmeter
1. Insert the turbine in the optoelectronic reader, in the way indicated by the arrow in
the turbine.
2. Insert the turbine plug on the front panel.
Notice: It is advisable to
lubricate periodically the Orings inside the
optoelectronic reader with
sylicone grease for its good
3. Connect the external anti moisture filter to the sampling plug and joint the
Using the "Ultimate Seal"
The "ultimate seal" is a moulded of Elasto-Gel, a glycerine based hydrogel. This
product is a unique polymer gel that forms an intimate seal between the face and the
mask. It has to be used for mask applications on hard to seal faces and where leaks are
not tolerated.
• Will not irritate the skin
• Contains no adhesives.
• Has no odour
• Will not dry out
• Single patient use
Permapure sampling line to the filter. It's strongly suggest to use the filter always
and we recommend to replace it every 3 months.
1. Headcap
2. Mask
3. Inspiratory valve
4. Turbine
5. Optoelectronic reader
6. Sampling adapter
7. Sampling plug
Chapter 3 - Installation - 37
Apply the seal to the mask
Apply seal to clean, residue-free mask only and follow the instructions below:
1. Remove the plastic tray from the bag. Peel off clear film and retain for later use.
2. While holding tray align the nose area of mask to nose area of Ultimate Seal™ gel.
Press together and roll mask down over the surface of the gel seal attaching it to the
mask and releasing it from the tray.
3. If needed, adjust the position of the seal, aligning it with the outer perimeter of the
mask sealing surface.
4. The mask is now ready to be placed on the subject's face.
To remove seal on mask
•The Ultimate Seal™ have been conceived for a single patient use only, it can not be
cleaned or sterilised.
•If mask requires cleaning for a new patient application then pull off and dispose of
the Ultimate Seal™.
•To keep the seal clean between use, keep it attached to the mask and place the clear
Notice: Avoid the exposure
to the sun. Do not put the
seal into the water.
film against the Ultimate Seal™ gel on the mask. When the seal becomes
discoloured or opaque (approximately two weeks) dispose of the current seal and
replace it with a new one.
Assembling the flowmeter for spirometry tests
In case the spirometry kit option is purchased assemble the turbine as shown in the
illustration below.
38 - Quark b2 User Manual
Cardiac Output Module
The Cardiac Output Module is provided with the Cardiac Output option and it consists
of a special valve together with a bag that has to filled in with a different gas
concentration (see Chapter 6 for further information).
Connecting PC
Before operating the system make sure the unit is connected to the PC as described in
the following illustration:
Chapter 3 - Installation - 39
Software installation
Note: The software can be installed on Windows XP or Vista It will not work on any
previous versions of Windows.
The software consists of two programs: a spirometry program (uses a green CD,
labelled PFT, option) and the program for ergometry (uses a blue CD, labelled CPET).
The programs share the same archive and system calibration program.
One or both software programs may need to be installed depending on the device
configuration.
Installing the software
Note: The software is copyright protected and should be installed only from the
original disk.
1. Insert the installation CD into the CD-ROM drive.
2. The installation will begin automatically. If the disc does not start automatically
you will need to run Setup.exe.
3. Follow the instructions given by the dialog boxes to complete the installation.
4. When the installation is finished, the program will alert you that the installation has
been successfully completed.
Run the software
PC port configuration
Note: If both programs are installed, the directory for the CPET software should be the
same as the PFT directory (default is C:\PFTSUITE).
1. In the Windows Start menu, open the Program Group in which the software was
installed.
2. Click the CPET or the PFTSuite icon.
The first time the software is used, it is necessary to configure the communication port
with the PC (USB, COM1, COM2,…).
For further details, see the chapter Database management.
40 - Quark b2 User Manual
Software main features
Display
The program may contain several windows. The active window is highlighted with a
different colour of the caption. Some functions of the program are "active window"
sensitive (Print, right key of the mouse).
Tool bar
Many of the functions that may be selected from the menu can be activated more
rapidly by clicking with the mouse on the corresponding icon in the tool bar.
Positioning the mouse cursor on one of the buttons of the toolbar (if the option Hints is
enabled), the description of the corresponding function is shown in a label.
Show/hide the toolbar
Select Toolbar from Options menu in order to show or hide the toolbar.
Dialog windows
The typical operating environment of Microsoft Windows is the Dialog box. This
window is provided with a series of fields in which input the information.
Scroll bars
On line help
Use of the keyboard
•To move the cursor among fields, press the Tab key until you reach the desired
field.
•Press the Enter key to confirm the information input on the dialog box or press the
Esc key to cancel changes.
Use of the mouse
•To move the cursor among fields, move the mouse on the desired field and left-
click.
•Click on the OK button with the Left button of the mouse to confirm the
information input on the dialog box or click on Cancel button to cancel changes.
Some windows are provided with scroll bars that help to see data exceeding the window
space available.
•To move the scroll bar row by row click the scroll arrows at the end of the scroll
bars
•To move the scroll bar page by page click on the grey area at both sides of the
scroll fields
COSMED Quark Help is a complete on-line reference tool that you can use at any time.
Help is especially useful when you need information quickly or when the Quark user
manual is not available. Help contains a description of each command and dialog box,
and explains many procedures for accomplishing common tasks.
To get the Help on line, press the F1 key.
Software version
To know the software version and the serial number of the software, select About
Quark b
Chapter 3 - Installation - 41
2
… from Help menu.
Calibration
Calibrating sensors
The software allows to automatically calibrate zero, gain and delay of gas sensors. Even
if the program doesn't force you to carry out the calibration we strongly recommend to
execute it before each test.
Running the Calibration program
Start the program and choose Calibration from the Test Menu. The software runs the
Calibration software and the main menu changes accordingly.
Log file
The program creates and updates as default the calibration log file, containing the
conditions and the results of all the calibrations performed by the user.
To access the file select File/Report File... from the calibration program.
Setting reference values
Before starting calibrating make sure that the system has been configured correctly by
setting the right values of gas concentration of: room air (i.e. 20.95% O
CO
), of gas mixture contained in the cylinders and the volume of the calibration
2
syringe (i.e. 3 litres).
and 0.03%
2
To set the reference values
This operation must be performed only the first time. The next times, the system keeps
stored the reference values entered in this step.
1. Select Reference Values from the Calibration menu.
2. Type the correct values for the O
and CO2 room air concentration (i.e. 2093 for
2
20.93%), and do the same for the gas concentration of the calibration cylinder.
3. Type the volume of the calibration syringe (i.e. 3000 for a 3 litres calibration
syringe).
4. Press OK button to confirm changes.
Chapter 4 - Calibration - 43
p
Calibrating analysers
Note: After turning on the
unit, wait 30 minutes warm
up time before starting the
calibration
rocedure.
This calibration must be repeated at least daily. A new calibration before any
ergospirometry test is strongly recommended.
1. Remove the sampling line from the flowmeter.
2. Ensure that the O
3. Check if the reference values for the calibration gas specified in File/Reference
4. Connect the sampling line to the CAL port placed on the front panel of the Quark
5. Select Calibration/Gas…/Gas… and wait until the procedure is completed.
The software performs automatically the calibration procedure. After 90 seconds the
graph will be displayed. In this way, the user can check the calibration procedure both
graphically and numerically.
At the end of the procedure, the software displays the new calibration factors vs. the old
ones.
-CO2 cylinder is properly connected to the rear panel of the
2
Quark b
2
and check that the pressure of the gas is in the range 5-6 bar.
values… correspond to the certified composition of the mixture.
2
b
.
Print the calibration report
In the Calibration program choose Print from the File menu.
Edit the calibration factors
The last sensors calibration factors can be either edited or viewed. To do this choose
Gas Results… from the File menu.
To view or edit the last Turbine calibration factor choose Turbine results… from the
File menu.
Note: To restore factory setting press Default button in the dialog box. Once you press
the default button you must run a new calibration before testing.
44 - Quark b2 User Manual
Turbine calibration
The system uses a turbine flowmeter. It opposes a very low resistance to flow (<0,7
cmH
motion which causes the rotation of the turbine rotor.
The rolling blade interrupts the infrared light beamed by the three diodes of the
optoelectronic reader. Every interruption represents 1/6 turn of the rotor, this allows to
measure the number of turn in the time. There is a constant ratio between air passing
through the turbine and number of turns. This allows an accurate measure of flows and
volume. The turbine flowmeter doesn't need daily calibrations as it is not affected by
pressure, humidity and temperature.
To work properly, the turbine only requires the rotor to rotate freely without any friction
that might be caused by dust that can be easily avoided with an ordinary cleaning
procedure (see Maintenance).
However in order to ensure accuracy it's recommended to run periodically the
calibration procedure. Calibration has to be carried out with a calibration syringe of 3
litres volume, the calibration procedure is totally managed by software.
A measurement system should be calibrated daily in order to ensure maximum accuracy
and reliable test results. If a correct maintenance is provided it’s possible to check the
calibration of the turbine flowmeter even at relatively long intervals (i.e. 1 month). The
calibration procedure assures valid and verifiable results within a ±3% accuracy.
O/l/s to 12 l/s). The air passing through the helical conveyors, takes a spiral
2
The calibration syringe
The 3 litres calibrated syringe is included in all the Quark PFT line with the exclusion
of the PFT 1 model.
3 litres calibration syringe: P/N C00600-01-11.
Turbine calibration for ergospirometry tests
Before starting the calibration procedure, be sure that the turbine type is properly
selected (in the calibration program, select File/Turbine results..., in the field Type (mm) must be entered 28).
Assembling the flowmeter
1. Connect the Opto-reader to the calibration syringe through the adapter.
2. Connect the flowmeter to the syringe with the rubber cylinder supplied in the
standard packaging.
Chapter 4 - Calibration - 45
1. Syringe
2. Silicone tube
3. Adaptor for calibration syringe.
Note: if you are using a slow
PC, we recommend to set
Note: If a bacterial filter is used for the tests, do use it also during the turbine
calibration.
Calibrating the turbine
After having run the calibration program:
1. Select Reference Values from the File menu. If your syringe has a different value
from the default one (3 litres), please enter the correct value.
2. Select Calibration/Turbine….
3. When the Calibration Turbine dialog box appears with the syringe piston initially
pushed all the way in, move the piston in and out for 5 inspiratory strokes and 5
expiratory strokes in order to get the first values appearing on the display. Then
move the syringe piston for other 10 strokes (IN and EX).
46 - Quark b2 User Manual
4. At each of the 10 steps the software displays the results of the manoeuvre and the
percentage error in the reading.
5. At the end of this operation, the software displays the new calibration factors. Press
OK to store the new value.
Turbine calibration for the RMR test
The turbine used for resting metabolic rate tests is different from the standard one (used
for ergo-spirometric tests). Since the correction factors for the two turbines are different,
before using this turbine, it is necessary to select and calibrate the turbine used.
In the calibration program, select File/Turbine results..., and enter 18 in the field Type (mm).
At the end of the RMR tests, before starting using the standard turbine, set 28 in the
Type(mm) field and perform a turbine calibration with the standard turbine.
Assembling the flowmeter
Connect the RMR reader to the syringe by means of the proper adaptor.
Calibrate the turbine
Perform a turbine calibration according to the procedure described above. Since the
ventilation is very low (normally <10 litres/min), the turbine calibration has to be
performed with very slow manoeuvres (each complete manoeuvre in about 10-15
seconds), to obtain the best accuracy.
Chapter 4 - Calibration - 47
r
Checking the system signals
The control panel
The Control Panel, which can be activated from the Calibration/Control panel…
menu item, is a useful tool to check the main hardware functions of Quark b
By using the controls on Control Panel you are able to do the following:
1. Reading the signals acquired by the system both as voltages and processed data;
2. Activating/Disactivating the valves, the sampling pump and other installed
components (for example, oxymeter).
2
.
mV / real values
display
Using the control panel
Sampling
pump on/off
RAW switch
on/off
Analog
input
Calibration
cylinder On/Off
Bag filling
Air valve
Bag valve
Oxymete
Deselect all
channels
Select all
channels
Signal refresh
time
48 - Quark b2 User Manual
Chapter 4 - Calibration - 49
Database
Management
Exercise testing patient's database
The exercise testing software uses a different interface for presenting patient
information. The patient database allows to:
• Enter a new patient
• Find patient data
• Edit patient data
• Delete patient data.
Select Patients from the File menu.
Enter a new patient
1. Press New to show the Patient dialog box.
2. Enter data of a new patient and press OK button to confirm.
Find a patient
Edit patient data
Delete a patient
Enter a search string into the Find field and press Find to view the data concerning a
subject already in the database. You can search for “Last name”, “ID code” or
“Progressive”.
Or:
Press List to view the list of patients in the database. Press Next or Previous to view
the data corresponding to the next or to the last patient. Press OK to confirm.
The Next and Previous buttons allow to move to the next or the previous patient in the
database.
1. Select the patient.
2. Press Modify… in order to edit the patient’s data.
3. Edit the desired values and press OK to confirm.
1. Select the patient to be deleted.
2. Press Delete.
Chapter 5 - Database Management - 51
Archive maintenance
The software allows to manage files selecting Archive from the File menu.
It is advisable to perform the archive reorganisation every month, in order to free space
on the hard disk and/or to correct possible errors present within the database.
It is possible also that your have no more hard disk space. So, you have to delete all the
data. In this case, it is useful to perform the initialising.
Reorganise the archive
1. Select Reorganize archive from the File menu.
2. Wait for the end of the operation before performing any other function.
Delete the archive
1. Select Initialize Archive from the File menu.
2. Wait for the end of the operation before performing any other function.
Delete a test
To delete an ergometry test, select Test/Delete test.
To delete a spirometry test, press the proper button in the Test Card.
Backup and restore
It is strongly recommended to backup files, a warning message will be displayed
monthly. This function allows the user to restore the data if the PC or the HD will not
work anymore.
Backup
1. Select Backup archive from the File menu.
2. Selecting the destination path with the Browse key or press New to create a new
directory. Press OK to confirm.
3. In the dialog box it will appear an estimate of the number of floppy disks you need
in order to back up the archives. Press OK.
Restore
1. Select Restore archive from the File menu.
2. On the Restore dialog box specify the drive source and press OK, a dialog box
will appear indicating all data of the backup processed.
52 - Quark b2 User Manual
Spirometry patient's database
The Patients database consists of a Patient Card, a Visit Card and a Test Card in which
are listed all tests performed by the patient.
Select Archive Navigator from the File menu or press the button by side.
Note: after having deleted a
record (patient, visit or test),
it is recommended to
reorganize the archive in
order to free disk space.
Patient Card
It collects all the information of a patient (first name, last name, date of birth) which
remain the same for each visit. For each patient there is only one Patient Card, which is
created the first time the Patient performs a test.
To move within the database use the following buttons:
Move to the first patient in the archive
Move to the previous patient in the archive
Move to the next patient in the archive
Move to the last patient in the archive
Find a patient in the archive
Enter a new patient in the archive
Delete current patient from the archive
Edit the current patient card
Visit Card
It collects all information relative to the visit (diagnosis, visit description...) and to the
patient information subject to change between one visit and another (height, weight,
smoke). Each patient can be related to several Visit Cards provided they have been
created in different days. Before carrying out any spirometric test it is necessary to
create a new Visit Card or to open the today’s Visit Card.
To move within the database use the following buttons:
Move to the first visit in the archive
Move to the previous visit in the archive
Move to the next visit in the archive
Move to the last visit in the archive
Find a visit in the archive
Chapter 5 - Database Management - 53
Enter a new visit card in the archive
Delete current visit card from the archive
Edit the current visit card
Test Card
It contains all the information about the test.
To move within the database use the following buttons:
Delete current test from the archive.
Edit the current test
Import/export a Tests card
This function allows to import /export a test card with the respective visit and patient
card.
1. Select the patient.
2. Choose the test and press the key by side. All data will be imported/exported in the
XPO file format (Cosmed proprietary).
Diagnosis Database
The program allows to manage a diagnosis database, whose records are composed by a
diagnosis ID code and a string of text.
The report of the visits can be done either by typing the desired text in the field
“Diagnosis” of the Visit Card or, more quickly, retrieving from the diagnosis database
the desired one.
If you want to insert, modify or delete a diagnosis from the database select Database Diagnosis... from the File menu.
54 - Quark b2 User Manual
Spirometry program settings
The software allows to configure some options selecting Configure from the Option
menu.
Graphs
All the graphs visualised and/or printed can be customised in colours and appearance.
1. Select the desired colours of the curves (5 curves max can be overlapped on the
same graph).
2. Enable or disable the Grid option.
3. Enable or disable the Show Info Test option.
Serial port
You must select the serial port RS 232 that will be used to connect the Quark b2 with the
PC.
To select the serial port, click on the proper COM button (the selected port must be
different from the mouse one).
Units of measurements
It is possible to configure the units of measurements, weight and height, for printing and
viewing.
To select the units of measurement click on cm/Kg or in/lb according to the desired
format.
Using extra fields
The Patient’s database is organised in 3 different cards (Patient card, Visit Card and
Test card.) where it is possible to store the information about patients and visits .
Besides the standard information, it is possible to customise some fields (user free
fields), entering and labelling measurements coming from other devices.
The customisable free fields are:
• 3 fields in the Patient Card (Patient’s information)
• 3 fields in the Visit Card (information about the visits)
• 3 fields (2 numeric) in the Test card information about Test)
Customise the fields
In the group User free fields type the desired text in the 9 fields available.
Chapter 5 - Database Management - 55
Exercise testing
Recommendations for the exercise testing
The evaluation of the cardiorespiratory function
The physical training requires the interaction of physiological mechanisms that allow
the cardiovascular and respiratory systems to supply the increasing demand of energy
due to the contraction of the muscles.
During the training the systems are both engaged, an adequate answer to the effort is the
measure of theirs health state.
The increase of the metabolic rate, during the exercise, needs an appropriate increase of
oxygen in the muscles. At the same time, the CO
in order to avoid the lactic acid making.
To satisfy the increase in the gas exchange, necessary to the muscles during the
exercise, is requested the intervention of many physiological mechanisms. This process
involves lungs, the pulmonary circulation, the heart and the peripheral circulation.
Precautions
The physician has the responsibility that the patient subjects to the test is a suitable
person able to execute an effort test.
Laboratory
The room, in which the test is performed, must be big enough to contain the whole
necessary equipment, allowing an easier accessibility to the patient in case of
emergency.
In the room should be placed a thermometer and a hygrometer; the heart frequency and
the perceived values of the effort rise as much as the ambient temperature increases, and
the variability of the cardiovascular response grows for humidity values higher of 60%.
Generally it is considered 22°C the temperature adequate for the test execution, even for
short efforts, values till 26°C can be considered acceptable in presence of an efficient air
ventilation.
muscles production must be removed
2
Ending the test
The patient should be monitored with ECG for at least 8 minutes, in resting conditions
or until he returns to the pre-exercise conditions.
Preparing the patient
To enhance the value of a diagnostic test it’s very important patient collaboration. In
most cases a well-informed patient will make a better effort (in relation to his
conditions) and will allow a reliable interpretation of the test. For this reason every
ergometric test must be preceded from a precise training of the patient.
Before testing
The physician applying the exam must be provided with a written request including a
brief description of the diagnosis (confirmed or suspected ), the request’s reason and the
patient therapy carried out showing the dose and time of the drug assumption.
To standardise the response to the test and reduce the patient’s anxiety it’s suggested to
provide him either written (before the exam) or oral (at the same time of the test)
information. At the scheduling time detailed instructions should be delivered to the
patient, consisting in smoke and food abstinence three hours before an ergometric test,
or eight hours before a scintigraphic test.
Test are usually executed supporting the therapeutic outline in progress, but sometimes
it could be necessary to stop some drugs, such as b-block or calcium antagonist, which
could impair the effort response reducing the diagnostic accuracy of the exam.
The patient must wear comfortable suit and gymnastic shoes and two hours before test
stop any kind of drugs, eat light and avoid coffee and smoke.
Chapter 6 - Exercise Testing - 57
It’s very important acquire information on the patient’s clinical past before performing
the test. Keep attention in particular way to the use of drugs, tobacco, to the physical
fitness and symptoms produced with the exercise.
Patient assent
The patient is informed that he will be submitted to a maximum effort, which could be
stopped at any moment, and of the risks of the test execution.
Ending the test
Test may end when the maximum value of the oxygen consumption has been reached
and the patient’s response established.
58 - Quark b2 User Manual
Start Testing
Note: Selecting “Simulated
test”, the software allows to
use the gas calibration in order
to perform the test. This is
useful if the user would check
the accuracy of gas
Before starting exercise test type a new patient information or choose one from the list
of patient in the file. As soon as a patient has been entered the software is ready to start
a test. The name of the active patient is shown on the status bar of the program window.
Start a test
1. Calibrate the analysers as described in the Calibration chapter.
2. Choose Execute Test from the Test menu.
3. Enter or modify the antropometric data of the patient and select the options you
4. Select the ergometer you need to control
5. The software environment will change showing a new Menu bar and toolbar while
6. At this point the software starts showing data on the monitor but without saving
Abort the test without saving data
Choose Abort from the Test menu or press Alt+F3.
End the test saving data
1. Choose End from the Test menu or press F3.
2. Choose Yes to end the test or No to continue.
View data in real-time
The visualisation features and capabilities of the data and graphs are identical to the
ones described in the Data management chapter. Starting the test a small window will
appear on the right corner displaying time, bmp and, if selected before, the ergo
protocol and trainer.
need to carry out the test . To use a specific protocol choose it in the list box and
press OK to confirm.
the first data will be displayed in a table format.
them, this in order to monitor the patient before starting the test. To start storing
data press F2.
Chapter 6 - Exercise Testing - 59
Parameters to view
the labe
View graphs in real-time
1. Choose Graph from the View menu.
2. Follow the instructions described in data management section to edit the graphs.
While the test is running, it is possible to choose the parameters to view.
1. Select Parameters to view/Test execution… from the Options menu.
2. Select the parameters and confirm.
Manual protocol
Tip: pressing the Shift key
while choosing the marker
option will allow you to enter
l for that marker.
Automatic protocol
Set the BPM alarm
If you are using the Quark with a treadmill without serial interface, it is possible to enter
manually from the PC the event, the phase and the marker.
Enter Load and Phase
1. During the test select Load from the Events menu.
2. Select the phase and/or type the value of the load and press OK to confirm.
Set the markers
Select Marker from the Events menu.
The software allows to automatically control the ergometer according to the protocol
previously selected. Anyway it is allowed to change it even after the test is started.
Modify the load during the test
1. During the test choose Ergometric protocol from the Events menu.
2. Select the row corresponding to the desired load and press OK to confirm.
The software allows the user to set the alarm level for the heart rate, in order to monitor
the patient response.
Enter the BPM
1. Choose BPM alarm from the Events menu.
2. Set the alarm by moving the scroll bar and press OK to confirm.
It also allows to enable or disable the acoustic alarm by the option "Acoustic alarm".
60 - Quark b2 User Manual
Data management
As soon as the test has been completed, all data stored can be retrieved for a complete
management.
Viewing data
Data can be viewed in the following formats:
Table form numeric values of the various parameters (columns) corresponding to
Graphic form graphical presentation on Y1, Y2, X charts.
Summary results of the test and statistical analysis of the blocks.
Predicted predicted values, maximum value measured.
View data in table form
1. Select Data… from the View menu.
2. Select the test to visualise in the list box and press OK
Note: Double-click in the
window to open the edit test.
each step (rows).
Creating graphs
The software is provided with powerful functions for creating charts. You can add
custom graphs to create exactly what you need.
View data in graph form
1. Choose Graph… from the View menu.
2. Select the tests to visualise from the list and press OK
3. Choose the parameters you require on the X, Y1 and eventually for Y2; select if
necessary some of the following options by pressing the more button and press OK
to confirm.
It is possible to access quickly 5 common graphs from the View/Graph… dialog box.
Chapter 6 - Exercise Testing - 61
p
Note: Double-click in the graph
window to o
en the edit test
Right-clicking, the graph can be exported in bmp file format.
Customise the graphs
1. With a graph on the screen, choose Customise graph from the View menu.
2. On the Customise graph dialog box, select options to obtain the wished graph.
Option Function
Grid on X, Y axesshow the grid lines in correspondence with x or y axes that
make the graph easier for you to analyse data.
Autoscale maximum and minimum values of the graph will be measured
automatically.
Ignore 0 points with 0 value measured won't be shown.
Not interpolated make the graph scattered.
Marker highlight with a symbol all steps of the test in which the marker
button was pressed.
Squared makes the graph a square
Without recovery exclude from the graph all points of the recovery phase.
Mark points marks each point with a symbol
Min. Max. allows to set manually the axes values.
Step Set the axes' scale step.
Switch from graph to data and vice versa
When the active window is a graph (or a report in data form), it is possible to view very
quickly the data (or the graph) corresponding to that test.
Choose Current test data (if the active window is a graph) or Current test graph…
(if the active window is a data report) from the View menu.
62 - Quark b2 User Manual
Viewing predicted values
For some parameter it is possible to compare the maximum value measured during the
test with its predicted value and the LT value both in percentage and absolute.
View predicted values
Choose Predicted from the View menu.
Anaerobic (Lactate) Threshold detection
The software allows to detect the Lactate Threshold (Anaerobic Threshold) according to
the "Modified V-slope method" reference. The LT can be detected both manually and
automatically.
Note: Double-click in the
window to open the
corresponding dialog box.
View the Lactate Threshold
Choose Lactate Threshold from the View menu.
Detect the Lactate Threshold
1. Choose Calculate LT from the Test menu.
2. For calculating it automatically on the "Lactate Threshold" dialog box click on the
Auto detect button.
3. For adjusting manually the point you want to detect, move the scroll bar on the
dialog box by pressing the arrow buttons. Data and graph of the LT will be
automatically redrawn.
4. Press Ok button to save your choices.
Customise graphs for the LT viewing
The software allows to customise two of the three graphs for the LT visualisation.
1. Choose Lactate Threshold from the Options menu.
2. Choose the parameters you want to be shown on the LT window and press OK to
confirm your choices.
Chapter 6 - Exercise Testing - 63
Fittings
The purpose of the fitting function is to find the function that fits as better as possible
the measured data.
The software allows to fit data according to 3 different functions:
Model Function Algorithm Available for
Linear Y=A*X+B Least squares Any Y vs any X graph
Mono-Exp Y=A+B*exp[(t-t0)/tau] Levemberg Marquardt Any Y vs Time Graph
Mean value
This function is available only if the active window is a single Y graph (i.e. VO2 vs time
or VO
vs Load).
2
Fit a graph with a linear regression
1. Make active the graph window (any Y vs any X graph).
2. Right-click and select Fitting.
3. Choose Linear in the type combo box
4. Select the first point (X1) moving the mouse on the desired place in the graph
pressing the Left key or using the + and – keys.
5. Select the second point (X2) moving the mouse on the desired place in the graph
pressing the Right key or using the + and – keys.
6. Press OK to confirm.
Fit a graph with a Mono-exponential regression
1. Make active the graph window (any Y vs any X graph).
2. Right-click and select Fitting.
3. Choose Mono-exponential in the type combo box
4. Select the first point (X1) moving the mouse on the desired place in the graph
pressing the Left key or using the + and – keys.
5. Select the second point (X2) moving the mouse on the desired place in the graph
pressing the Right key or using the + and – keys.
6. Enter (if necessary) the initial values of A, B and TAU (these are the values from
which the iterative algorithm starts in order to reach the best values; the closer are
the initial coefficients to the best ones the higher is the possibility to reach the best
fit).
7. Press OK to confirm.
64 - Quark b2 User Manual
Oxygen Kinetic
Calculate the "Mean Value"
1. Make active the graph window (any Y vs any X graph).
2. Right-click and select Fitting.
3. Choose Mean value in the type combo box
4. Select the first point (X1) moving the mouse on the desired place in the graph
pressing the Left key or using the + and – keys.
5. Select the second point (X2) moving the mouse on the desired place in the graph
pressing the Right key or using the + and – keys.
6. Press OK to confirm.
Note: The results of the O2 Fittings function are not stored therefore, in order to keep
the information, print the page using File/print Active Window.
This function is available only if the active window is a VO2 vs time graph and it has a
sense only with Constant Load Exercise Tests.
The aim of this function is to find the dynamic response of the rising and falling edges
of the VO2 together with the Oxygen Deficit and Oxygen Debt.
Run the O2 Kinetic function
1. Make active a VO2 vs Time graph window.
2. Press the right key of the mouse and select O2 Kinetic.
3. Select the beginning of the exercise phase (t1) moving the mouse on the desired
place in the graph pressing the Left key or using the + and - keys.
4. Select the beginning of the exercise phase (t2) moving the mouse on the desired
place in the graph pressing the Right key or using the + and - keys.
5. Enter (if necessary) the initial values for A, B and Tau (these are the values from
which the iterative algorithm starts in order to reach the best values; the closer are
the initial coefficients to the best ones the higher is the possibility to reach the best
fit) and press OK. You can lock data checking the relative field
Chapter 6 - Exercise Testing - 65
Information about the Test
The Test Information dialog box shows all the information concerning environmental
data, patient data and some data about the test
View the Information
Choose Information from the View menu.
Summary
Tip: double-clicking on the
Summary window the
function Options/Summary
is activated by which you
may configure the structure
Modify the information
1. Press the Modify button on the Information dialog box.
2. Change the values you want to modify and press OK to confirm.
The software allows to assess the energy expenditure and metabolism substratum. In
order to measure FAT and CHO, type the UN (Ureic Nitrogen) value into the field. All
the nutritional parameter will be recalculate considering the UN value.
The summary feature allows to summarise test results according to the workload and
phase during the test.
View the summary
1. Choose Summary from the View menu.
2. The summary of the current test (active window) will be displayed.
Print the data
66 - Quark b2 User Manual
It is possible to print graphs and data by means of two functions: Print report and
Print current window. The last one is active only if the active window is a graph or a
data report.
View the report
Print the current window
1. Be sure that the current active window is the graph or the report you desire to print.
2. Select Print current window from File menu.
3. Press OK to print, or Setup to customise the print.
Print the report
1. Select the report to be printed from File menu.
2. Press OK to print, or Setup to customise the print.
3. To only view the report, without printing it, press Shift during the selection.
This function allows to show a preview of a selected report.
1. Select Report from the View menu.
2. Select the test and the report to visualise and press OK to confirm.
Chapter 6 - Exercise Testing - 67
Data Editing
g
Nota: In Data view, doubleclick in the window to enter
in “Data Editin
”.
The software allows the user to edit the data acquired during the test in the following
ways:
• deleting one or more steps
• editing row data
• input new parameters
• data filtering (averaging or smoothing)
• advanced data elaboration
After data elaboration it is always possible to restore the original data file by pressing
the Restore button .
If you want to save permanently all the changes, press Overwrite; being aware this
function replaces the original test definitely.
Editing values and input numerical values
1. Choose Edit test from the Test menu.
2. Select the cell containing the value you want to replace with others values and press
OK to confirm the editing.
The software will recompute all the parameters. Both the tables and the graphs will be
automatically updated.
68 - Quark b2 User Manual
Data filtering
1. Choose Edit test from the Test menu.
2. Press the button Filtering and choose the option Discard invalid steps to
automatically eliminate all the invalid steps
3. Press the button Filtering and choose the option Averaging and type the desired
value for points Ave/smooth to perform an averaging of the all acquired steps. This
feature reduces the size of the original test.
4. Press the button Filtering, select the option Smoothing and type the desired value
for points. This feature doesn't reduce the size of the original test, although it
smoothes the fluctuation of data.
Using the User fields
The software is provided with three free fields in which the user may enter values
coming from others devices such as lactate, blood pressure etc.
To define the user fields:
1. Choose User Fields from the Options menu
2. Type the desired text in the input fields and press OK.
To enter values in the user fields:
1. Choose Edit test from the Test menu.
2. Scroll horizontally until the fields USER 1, 2 and 3.
3. Enter the desired values and press OK to confirm.
Deleting steps
This feature is useful whenever some steps acquired during the test are to be discarded
(steps acquired before the start of the test, patient disconnected from the face mask....).
1. Choose Edit tests from the Test menu.
2. Position the cursor on the step you want to delete and press the button Delete step.
Advanced Editing
This feature allows to perform some advanced editing of the data stored in the software.
1. Choose Edit test from the Test menu.
2. Press the Advanced button and select from the following options:
Option Function
Delete steps deletes the steps meeting the selection criteria
Smoothing applies a moving average to the selected parameter
Edit parameter edits a parameter according to the selected criteria
Edit parameter Specifications
Value replaces the value of the selected parameter with a new one.
Correction % applies a percentage correction to the value of the selected
parameter.
Offset adds an offset to the value of the selected parameter.
Formula... replaces the value of the selected parameter with a
mathematical function.
Time range Specifications
From, To specifies the time range.
All steps applies the editing from the beginning to the end of the test.
Chapter 6 - Exercise Testing - 69
Apply to Specifications
Parameter specifies the reference parameter
>, >=, =, <, <=, <> higher than, higher or equal, equal to, lower than, lower or
Value/Formula… specifies the value (mathematical expression) compared
All steps do not use any selection criteria.
Restore the original test
To cancel all the editing, in the "Edit Test" dialog box press the Restore button, confirm
your choice by pressing yes.
Overwrite the original test
To save all the editing, replacing the original test with the modified one, in the "Edit
Test" dialog box press the Overwrite button, confirm your choice by pressing yes.
Customise the desktop
Customise the display colours
1. Select Colors from Options menu.
2. Select the item to be modified.
3. Press Change and select the desired colour.
equal, different
with the value of the specified parameter.
Smart edit
This function is useful to correct data from artefacts; the noise affecting the measured
data can be reduced in 2 different ways:
Graphical noise suppression using the mouse
Threshold noise suppression applying a filter to the measured data
Apply the graphical noise suppression
1. Make active a graph or a data window corresponding to the test that you want to
modify.
2. Press the right key of the mouse and select Smart Edit.
3. Select the parameter that you want to modify.
4. Point the mouse on the position where the graph presents the artefacts, click the
Right key and, keeping pressed the key, drag the point on the desired place.
5. If you want to cancel the edit press the Left key of the mouse.
6. Repeat the above mentioned procedure for the all parameters and press OK.
70 - Quark b2 User Manual
Apply the threshold noise suppression
1. Make active a graph or data window corresponding to the test that you want to
modify.
2. Press the Right key of the mouse and select Smart Edit.
3. Select the parameter that you want to modify.
4. Set a Noise(%) Threshold (as a percentage of the parameter value) above which
any peak will be considered an artefact.
5. Press Execute and eventually Undo if you are not satisfied.
6. Press OK to confirm.
Customise the parameters
The software allows the user to create customised parameters and predicted values,
derived from the standard parameters (the ones that are calculated by default) through
any mathematical formula.
All the customised parameters can be used freely for viewing and printing purposes.
Create a new parameter
1. Choose Customise parameters from the Options menu.
2. Press the New button if you want to create a new parameter or Modify if you want
to modify an existing one
3. Type the desired value in the fields "Name", "unit of meas", "integers", "decimals"
and "summary" (to present the parameter in the summary) and press the Formula
button.
4. Insert the mathematical formula by using the appropriate tools and press OK twice
to confirm.
Create
a new predicted parameter
1. Choose Customise predicted from the Options menu
2. Press the New button if you want to create a new parameter or Modify if you
intend to modify an existing one
3. Type the desired value in the fields "Name", "unit of meas", "integers", "decimals".
4. Select the group of the predicted values from the options boxes.
5. Select the reference parameter in the "Compared to" list box.
6. Press the buttons in the calculation group and insert the mathematical formulas for
men and women, adults and paediatrics. Press OK twice to confirm.
Once you create the new predicted you can see it in the predicted window.
Chapter 6 - Exercise Testing - 71
Exporting data
stalled
With this function you can export the tests data and parameters in different file formats:
• *.xpo (Cosmed proprietary file format)
• *.txt (ASCII)
• *.xls (Microsoft Excel)
• *.wk1 (Lotus 123)
Export a test
1. Choose Export Tests from the Test menu.
2. Select the test to export from the list box and press OK to confirm.
3. Select the file output format from the list box, click on *.xpo, *.txt, *.xls or *. wk1.
If you selected ASCII format, by clicking on Text button you can then select the
Thousandssep. and Columnsep. according to the program you want to use. With
the xpo Cosmed format you can import/export the tests performed on another
Quark equipment.
4. Select the folder, type the file name and press OK to confirm.
Note: The DDE function is
available only if the user PC
has Microsoft Excel
.
in
DDE with Excel
If Microsoft Excel is installed on your PC, you can export a test simply pressing a
button on the toolbar.
To send a test to Excel, select Send to Excel from the Test menu.
The program will show a status bar indicating the data transmission to Excel. At the end
of the process a new sheet with all test data will be opened ready to be edited with the
powerful functions of Microsoft Excel.
72 - Quark b2 User Manual
Creating Test Protocols
The software allows to create different exercise protocols to use during the test. The
load of the ergometer is automatically controlled by the software that change it
according to the defined protocol.
Create a new protocol
1. Choose Real Time > Ergom. Tests Protocols from the Options menu.
2. Press New and enter a name for the protocol.
3. In the field "Message Time" type a number that means to get a message to advise
when switching to the next load.
4. Enable the "Drive Ergometer" check box to let the software control the ergometer.
Select the "Initial Command" if the ergometer need it.
5. Enabling the option "Relative Increments", the loads refer to the previous step.
6. Press Generate and enter the values to generate a protocol from only one load (i.e.
30 Watt each minute for a total of 20 steps) and press OK to confirm.
7. Press Add if you want to add a new step.
8. To edit a step, select it from the list and change the relative values in the Edit boxes
below the list. Press the Tab button to save changes.
9. To delete a step, highlight the step and press Delete.
Chapter 6 - Exercise Testing - 73
Software configuration
The software can be customised as you wish. Most of the feature are easily editable to
be tailored according to different purposes.
Data viewing
The software allows to calculate a huge number of parameters; it is advisable, in order
to simplify the analysis of the results, view in the data window the only desired
parameters.
Select the parameters to view
1. Choose Parameters to view/Test visualisation… from the Options menu.
2. Select the parameters you require to view.
3. Press OK to confirm the selected configuration.
Select the parameters to view during the test
1. Choose Parameters to view/Test execution… from the Options menu.
2. Select the parameters you require to view.
3. Press OK to confirm the selected configuration.
Sort the parameters
It is possible to sort the parameters (both for viewing and printing purposes) according
to the desired order.
1. Select Sorting parameters from the Options menu.
2. Move the parameters in the order you want by pressing and holding the left mouse
button.
3. Press OK to maintain the current configuration.
Steady State
The program has an algorithm to tag sets of steps as Steady State.
The algorithm considers belonging to the Steady State the only consecutive steps that
meet the following conditions:
•The value of VO2, VE and R do not vary from their mean values more than
Threshold (%);
•The number of consecutive steps that met the preceding criteria are at least
Minimum number of steps.
Customise the Steady State detection criteria
1. Choose Steady State from the Options menu
2. Type the desired values for Minimum number of steps, VO
threshold (%), VE
2
threshold (%) and R threshold.
The steps which satisfy these conditions will be highlighted with a yellow bar.
74 - Quark b2 User Manual
Printout reports
The software allows the user to printout data and graphics according to 4 customisable
reports. Further it allows the user to customise a printout header that will be printed in
each page.
Set up the printout
1. Choose Reports from the Options menu.
2. Define the desired features of the report and confirm. Enabling the option "Graphs
in one page" all the graphs selected in the report will be printed in one page.
3. Type the name you want apply to the report and press OK to save changes.
Select parameters to be printed
Quark allows to print a large number of parameters; it is advisable, in order to simplify
the analysis of the results, to printout desired parameters only.
1. In the report configuration window select Parameters.
2. Select the parameters you require to be printed in the data printout. The number of
parameters which can be printed depends upon the size of the paper in use (see
Printer Layout) and from the orientation of the sheet.
3. Press OK to confirm the selected configuration.
Customise the printout header
1. Choose Printout header from the Options menu.
2. On the "Report Header" dialog box type the text of the header.
3. To insert an image click the Logo button. An image editor will be opened, draw the
own logo and close the image editor to save changes.
4. Press OK to save the Printout header.
Chapter 6 - Exercise Testing - 75
Electronic reports (*.pdf)
If an Adobe PDF writer “Printer Driver” is installed and set as the default printer, it is
possible to store the printout report automatically in any location of the HD or
eventually LAN paths according to a customizable filename format.
It is possible to define the created filename format selecting Options/Printout header..., and then Name format…
Print the current window
The print current window function is enabled when the active window is a graph or a
data report.
1. Select Print current window from File menu.
2. Press OK to print, or Setup to customise the print.
Print the customised report
This function is enabled only after having customised a report.
1. Select the customised report from File menu.
3. Set the sheet format and press OK.
76 - Quark b2 User Manual
Events management during exercise testing
Cardiac Output (option)
The test is a non invasive method based on the measurement of the carbon dioxide
during an exercise test for the indirect measurement of the Cardiac Output; the
manoeuvre is composed by two different phases:
•Measurement of the metabolic parameters during a steady state condition (both at
rest or during exercise, usually at the end of the 3
•Rebreathing in a bag previously filled with a high CO
the CO
The method is base on the hypothesis that the PCO
equal to the one of the lung capillaries.
The rebreathing phase consists in making the patient breathing into a bag (typically 5
litres) containing an high concentration of CO
expired breath, brings to constant average value.
The optimum value of the CO
workload, and can be desumed from the following table:
Load (W) CO2 concentration (%)
50 10.5/11.5
100 11.0/12.0
150 12.0/13.0
200 13.0/14.0
Balance O2.
The volume of the mixture filled in the bag must be 1.5 - 2 times the Vt of the patient.
concentration reached at the equilibrium.
2
concentration of the cylinder and the bag depends on the
2
rd
minute of a 5 minutes step);
mixture and measurement of
2
in the alveolar gas is approximately
2
in O2 balance that, mixing with the
2
Note: The Cardiac Output must be carried out only with the mask without inspiratory
valves or with the mouthpiece otherwise the results are not reliable.
Setting up the Quark b2 for the cardiac output
1. Connect and open the Cardiac Output cylinder to the corresponding input port on
the rear panel (the output pressure of the gas must be between 4 and 5 bar).
2. Connect the valve to the corresponding connector of the front panel of the system.
3. Connect the rebreathing bag to the valve assembly.
Chapter 6 - Exercise Testing - 77
Performing a Cardiac Output test
1. Select Execute Test from the Test menu or press F2 and start the test as for an
ergometry test.
2. During the desired phase (rest or exercise), in a steady state, select Cardiac Output
from Test/Event… menu.
3. By using the corresponding buttons of the Toolbar prepare the bag by refilling it
starting from completely collapsed (otherwise the initial concentration of CO
2
will
be lower than expected value and the results could be unreliable).
4. Connect the valve assembly to the output port of the turbine flowmeter trying not to
affect too much the performance of the patient.
5. At the end of the expiration press F2 to start the rebreathing.
6. Stop the test by pressing F3 when the equilibrium is reached or wait for the
automatic interruption after 20 seconds from the beginning of the rebreathing.
7. Close the Cardiac output event with the corresponding button on the Toolbar.
Flow Volume loops
Note: The values of VO
, VCO2 and R during the rebreathing and in the following 2
2
minutes are not reliable because of the high concentration of CO2.
This test is useful during exercise to detect abnormalities in the mechanics of ventilation
in patients with pulmonary/ventilatory limitations to exercise.
The test consists in acquiring some flow/volume loops during exercise at different
workloads and overlapping them on the rest maximal flow/volume loop of a Forced
Vital Capacity test.
The majors information that you can get from this manoeuvres are the flow reserve
(flow distance from the peak flow of the F/V loop during exercise to the corresponding
flow on the superimposed F/V loop at rest) and the volume reserve (volume distance
from the maximum volume of the F/V loop during exercise to the corresponding
volume on the superimposed F/V loop at rest).
The manoeuvre consists in the following phases:
• Acquiring some Flow/Volume loops during the exercise
• Making the patient inspire completely up to TLC level (this is necessary to place
the loop correctly into the rest F\V loop of the forced Vital Capacity test)
•Overlapping the F/V loop acquired during exercise and the F/V loop performed at
rest.
78 - Quark b2 User Manual
Flow Volume loop during the test
1. Start with normal Exercise test and begin the memorisation of breath values (F2)
2. During a steady state select F/V loops form Test/Event...
3. As soon as 2 or 3 complete loops have been acquired ask the patient to inspire
completely up to TLC level and press F3 to stop the acquisition.
O2, CO2 vs Time
The O2, CO2 event is useful to check the real-time readings of the O2 and CO2 signals
during the test.
O2, CO2 vs Time during the test
1. Start with normal Exercise test and begin the memorisation of breath values (F2)
2. During a steady state select O2, CO2 vs Time from Test/Event...
3. As soon as 5 or 6 complete breaths have been acquired press F3 to stop the
acquisition.
O2 Saturation (optional)
The O2 Saturation event is useful to check the quality of SpO2 signal acquired by the
on-board Oxymeter (if available) during the test.
O2 Saturation during the test
1. Start with normal Exercise test and begin the memorisation of breath values (F2).
2. Select O2 Saturation from Test/Event...
3. As soon as 5 or 6 complete pulses have been acquired press F3 to stop the
Spirogram
The spirogram event allows to display and store the volume/time plot.
acquisition.
Spirogram during the test
1. Start with normal Exercise test and begin the memorisation of breath values (F2).
2. During a steady state select Spirogram form Test/Event...
3. Acquire volume/time plot until the window is filled and press F3 to stop the
acquisition.
Chapter 6 - Exercise Testing - 79
View the events after the test
1. Select Data... from the View menu
2. Select the test during which spirogram event has been carried out in the list box and
press OK
3. Select View... from the Events menu, choose the desired event and press OK.
4. Select Print Current Window... from the File menu to print the F/V curve page.
It is possible to edit the F/V loops event in the following way:
5. Select Edit... from the Event menu to change the F/V loop at rest (the list contains
all the FVC test carried out by the same Patient with the Spirometry software) and
press OK.
It is possible to edit the Cardiac Output event in the following way:
5. Select Edit... from the Event menu to change the parameters measured during the
steady state before the rebreathing, the CO
calculation method.
Raw data
It's a particular feature with which the user can check and save some parameters (CO2
output, O
sampling rate of 25 Hz.
Save Raw data
1. During the test choose Event from Events menu.
2. Select Raw Data from the list.
3. On the save data box give a name to the file and select the destination folder.
4. To stop saving Raw data press the stop icon or press F3 on the keyboard.
concentration and volumes) in Ascii file format in a archive apart at a
2
concentration at the equilibrium and the
2
80 - Quark b2 User Manual
Chapter 6 - Exercise Testing - 81
Resting Metabolic
Rate Test
Metabolism
Metabolism can be understood as the conversion by the human body between food and
accumulated fat into energy. The energy is used by the body to maintain constant
temperature , to move and to make all the organ function. Measure of metabolism is:
calories (cal).
Total Metabolic Rate
The total metabolic rate are the total calories that the human body needs in order to
actuate the daily functional activities.
Resting Metabolic Rate (RMR)
Resting Metabolic Rate represents the calories that the vital organs need to properly
operate at rest
% and 75 % of the human ‘s total metabolism.
Importance to measure RMR
A knowledge of the RMR is very helpful in order to understand the nutritional needs
and to properly manage it.
Measure of the rest metabolic rate with indirect calorimetry
Energy expenditure can be measured directly by putting a person in a calorimeter and
measuring the amount of heat produced by the body mass.
This is expensive and very impractical in the clinical setting. Energy expenditure can be
measured indirectly with a metabolic cart by analysis of respired gases (usually expired)
to derive volume of air passing through the lungs, the amount of oxygen extracted from
it (i.e., oxygen uptake VO
metabolism, expelled to atmosphere (CO
the resting energy expenditure (RMR) and respiratory quotient (RQ) can be calculated.
The RQ represents the ratio of carbon dioxide exhaled to the amount of oxygen
consumed by the individual. RQ is useful in interpreting the results of the RMR. The
abbreviated Weir equation is probably the most common calculation of RMR.
) and the amount of carbon dioxide, as a by-product of
2
output – VCO2). With these measurements
2
How to perform a RMR test
For best results, when having a REE done, there are certain conditions that need to be
controlled and others that just require documenting at the time of the test. During the
test the individual is interfaced with a metabolic measurement system by means of a
facemask.
A mouthpiece with a nose clip is also sometimes used, but it may create overly stressful
conditions to a subject (patient).
Important considerations or conditions to improve the RMR measurement:
• No food for at least 12 hours and no smoke for at least 2 hours before the test.
• Maintain quiet surroundings when the test is in progress and normal temperature.
The individual should not move arms or legs during the test.
• Medications taken should be noted, such as stimulants or depressants.
• The first 5 minutes of acquisition should be discarded by the computation of RMR
• Steady state should be achieved, which would be identified clinically by the
following criteria: 5 minute period when average minute VO
by less than 10%, average RQ changes by less than 5%
• Stable interpretable measurements should be obtained in a 15 to 20 minute test.
• Renal failure patients requiring hemodialysis should not be tested during dialysis
therapy.
Chapter 7 - Resting Metabolic Rate Test - 83
and VCO2 changes
2
Recommendations
Before starting an RMR test, it is necessary to select and calibrate the turbine used.
Read carefully the calibration procedure, in the Calibration chapter.
Resting metabolic rate test using the face mask
Use the following correction for the dead space (VD):
- 50 ml for the small mask
- 60 ml for the medium mask
- 70 ml for the large mask
Resting metabolic rate test using the canopy option
1. Verify (before and during the test) that the FeCO2 falls within the range 0.7%-1.3%
and adjust the flow rate of the pump as necessary. If the FeCO
should decrease the flow rate and if the FeCO
flow rate. A low FeCO
could be dangerous for the patient.
2. Do not place Canopy hood over a patient’s head before the tube is properly
connected and a continuous flow is applied from Canopy Blower.
may result in unreliable measurements, while a high FeCO2
2
is too low you
2
is too high you should increase the
2
84 - Quark b2 User Manual
Performing a test using the face mask
Calibrations
Before the test, it is necessary to perform an ergo calibration (see Calibration chapter)
and it is advisable to perform also a turbine calibration (see Recommendations in this
chapter).
How to prepare a patient
The patient interfaces with the equipment by means of a face mask, as depicted in the
following image. The mask has to be tight to the face, in order to avoid any air leakage.
C
A
Start the test
B
RM R
Make sure that the subject health status is acceptable according to what stated in the
guidelines.
Make the subject sit or lay on a comfortable chair or bed.
Fix the mask to the subject, as illustrated in the above picture, pull the elastic strings
(Point A and B) accurately in order to eliminate possible leaks. The mask must be
perfectly sealed to the face of the subject, especially in correspondence with the nose
(point C).
The mask adapt differently according to the face shape of the subject. The perfect
position is therefore to be determined from subject to subject.
1. Enter in the ergometry program
2. Select a patient or add a new one (File/Patients...)
3. Select Start test from Test menu
Chapter 7 - Resting Metabolic Rate Test - 85
4. Enter the patient’s data and select the RMR mode (1st picture).
5. Press Other Data… and enter the dead space value (50ml Small mask, 60ml
Medium mask and 70ml Large mask). It is possible to enter the Ureic Nitrogen
value NU (2
nd
picture).
6. Confirm and start the test by pressing OK.
Selecting RMR the system set automatically the following options:
• Data acquisition with a 30 seconds average
• RMR protocol, which is:
- 5 minutes discarded;
- 10 minutes with data acquisition, of which the software will make an average at
the end of the test;
- automatic end of the test after the 16
th
minute.
•Selection of the RMR workspace (windows placement);
The test is fully automatic, the software will stop it and save the data at the end of the
th
minute.
16
The real time view is as shown in the following picture:
86 - Quark b2 User Manual
Viewing the test
At the rend of the test, it will be opened automatically a window with the test results.
At the end of the test, or if it is selected View/RMR, the main results are shown:
•The average time interval (default: 10 minutes)
Nota: The percentage of used
Proteins (PRO%) is calculated
assuming 12 grams of Ureic
Nitrogen in 24 hours. You can
modify this value selecting
View/Information… ->
Modify…
•Average values of VO
predicted values if available.
• Body Mass Index (BMI) and interpretation
• Graph of the energetic expenditure for all the data acquisition interval, highlighting
the selected average interval.
In order to verify the goodness of the test, check that the ventilation and respiratory
frequency are similar to the predicted ones (12 breaths/min for the respiratory frequency
and 6 litres/min for the ventilation), and the heart rate is the rest heart rate of the patient.
How to modify the average interval
If the average interval (automatically identified by the software) is not satisfying, for
example because the patient was speaking in the first minutes, it is possible to modify
the interval of the average.
Right-click and select Edit RMR…. It is possible to move the start and the end lines.
, VCO2, R, RMR, RF, VE, HR, FAT% and CHO% and
2
To move the start line, left-click on the exact time in which you want to start the
calculations, for the end line, right-click.
Chapter 7 - Resting Metabolic Rate Test - 87
Print
The print of the current window generates a report similar to the one in the following
page.
88 - Quark b2 User Manual
Performing a test using the canopy option
The principle of a ventilated bubblehood system is that a stream of air is forced to pass
across the face of a subject and mixes with the air which is collected by a transparent
hood, placed over the subject’s head. A measurement system, knowing the flow rate,
calulates the oxygen consumption and the CO
values, the energy expenditure.
Calibrations
Before the test, it is necessary to perform an ergo calibration (see Calibration chapter)
and it is advisable to perform also a turbine calibration (see Recommendations in this
chapter).
How to prepare the canopy and the patient
Replacement of the power plug
If the power plug does not fit into the mains socket, replace it with the one in the
packaging.
In order to replace the plug:
1. Extract the plug from the battery charger
2. Insert the proper plug in the battery charger.
production and, starting from these
2
Connecting the Canopy
1. Connect the Canopy unit to the mains by means of the medical grade AC/DC
adapter provided.
2. Fix the vail to the bubblehood through the velcro strips.
3. Insert the bubblehood adapter into the bubblehood from the outside and fix it
screwing the ring from the inside, being careful to insert it in the proper hole, as
shown in the following picture.
4. Connect the bubblehood to the wrinkled tube, interposing a bacterial filter.
5. Connect the wrinkled tube to the unit through the Flow in connector.
2
6. Connect the optoelectronic reader of the K4 b
to the Flow out connector of the
Canopy unit.
Chapter 7 - Resting Metabolic Rate Test - 89
How to prepare the patient
1. Switch on the Canopy unit. If there are no problems, the red led on the front panel
of the unit flashes for few seconds and the alarm beeps. If the led does not flash
and/or the alarm does not beep, the test cannot be performed, because the backup
battery is exhausted or there is no backup battery.
2. When the green led turns on, the test can start. If the green led does not turn on, the
red led flashes and the alarm beeps, the test cannot be performed because the pump
does not work or the mains does not power the system.
3. After these checks, put the patient in a supine position.
4. Place the bubblehood with the vail on the patient’s head. The tube has to be placed
near the patient’s mouth.
Warning! Do not place Canopy hood over a patient’s head before the tube is properly
connected and a continuous flow is applied from Canopy Blower.
Performing the test
1. Enter in the ergometry program
2. Select a patient or add a new one (File/Patients...)
3. Select Start test from Test menu.
4. Enter the patient’s data and select the Canopy mode.
5. Confirm and start the test by pressing OK.
6. In the first part of the test the flow rate of the pump has to be adjusted by means of
the Flow adjustment handle on the front panel of the Canopy unit, in order to
measure an FeCO
between 0.7% and 1.3%. FeCO2 values can be read on the right
2
side of the PC monitor.
7. When the FeCO
acquisition. Verify, also during the test, that the measured FeCO
remains within the acceptability range, press F2 to start the data
2
is within the
2
0.7%-1.3% range. Otherwise, adjust it by means of the Flow adjustment handle.
90 - Quark b2 User Manual
Warning: If the green led turns off during the test, the red led flashes and the alarm
beeps, abort the test, because the pump does not work or the mains does not power the
system. In the last case, the pump works only because of the backup battery.
The test is fully automatic, the software will stop it and save the data at the end.
Viewing the test
At the rend of the test, it will be opened automatically a window with the test results.
At the end of the test, or if it is selected View/RMR, the main results are shown:
•The average time interval (default: 10 minutes)
Note: The percentage of used
Proteins (PRO%) is calculated
assuming 12 grams of Ureic
Nitrogen in 24 hours. You can
modify this value selecting
View/Information… ->
Modify…
•Average values of VO
values if available.
• Body Mass Index (BMI) and interpretation
• Graph of the energetic expenditure for all the data acquisition interval, highlighting
the selected average interval.
In order to verify the goodness of the test, check that the FeO
within the acceptability ranges (20.2%-20.8% and 0.5%-0.8% respectively), and the
heart rate is the rest heart rate of the patient.
How to modify the average interval
If the average interval (automatically identified by the software) is not satisfying, for
example because the patient was speaking in the first minutes, it is possible to modify
the interval of the average.
Right-click and select Edit RMR…. It is possible to move the start and the end lines.
To move the start line, left-click on the exact time in which you want to start the
calculations, for the end line, right-click.
, VCO2, R, RMR, VE, HR, FAT% and CHO% and predicted
2
and FeCO2 values are
2
Print
The print of the current window generates a report similar to the one of the RMR test
using the face mask.
Chapter 7 - Resting Metabolic Rate Test - 91
Sub-maximal
Exercise Testing
Introduction
Pre-test screening
Several physiological responses to exercise are used to evaluate cardiorespiratory
fitness, including oxygen consumption, heart rate, and blood pressure. Measuring these
variables during exercise, particularly maximum exercise, increase the chance of
detecting any coronary artery disease or pulmonary disease.
Unfortunately, maximum exercise tests are impractical because they are expensive,
require extensive clinical supervision, and subject individuals to levels of physical stress
that may be unnecessary depending on the objectives of the test. Consequently,
maximal testing is reserved for clinical assessments, athletic evaluation, and research.
A sub-maximal exercise test costs less and carries a lower risk for the individual.
Although less sensitive and specific for detecting disease or estimating maximal oxygen
consumption, correctly performed sub-maximal tests can provide a valid estimate of
cardiorespiratory fitness.
Pre-test health screening is essential for risk stratification and for determining the type
of test that should be performed and the need for an exercise test prior to exercise
training. A thorough pretest health screening includes the following:
• Complete medical history
• Medical contraindications to exercise
• Symptoms suggesting cardiac or pulmonary disease
• Angina or other forms of discomfort at rest or during exercise
• Unusual shortness of breath at rest or during exercise
• Dizziness or light-headedness
• Orthopaedic complications that may prevent adequate effort or compromise the
validity of test results
• Other unusual signs or symptoms that may preclude testing
• Risk factors for coronary heart disease
• History of major cardiorespiratory events
• Current medications
• Activity patterns
• Nutritional habits
• Reading and signing an informed consent form
Chapter 8 - Sub-maximal Exercise Testing - 93
Sub-maximal exercise testing
Heart rate varies linearly with VO2 to the point of maximum exertion; thus, VO
be estimated using the relation between heart rate and VO
without subjecting the
2
2max
may
individual to maximum levels of physical stress. During sub-maximal exercise testing,
predetermined workloads are used to elicit a steady state of exertion (plateau of heart
rate and VO
and extrapolated to the VO
). The steady-state heart rate at each work level is displayed graphically
2
at the age-predicted maximal heart rate (HR = 220-age). A
2
variety of protocols for different exercise modalities (i.e., treadmill, stationary cycle,
and step increments) can be used as long as the VO
requirements of each selected
2
workload can be estimated with accuracy.
The objectives of cardiorespiratory fitness assessments in the apparently healthy
population are as follows:
•Determine the level of cardiorespiratory fitness and establish fitness program goals
and objectives.
•Develop a safe, effective exercise prescription for the improvement of
cardiorespiratory fitness.
•Document improvements in cardiorespiratory fitness as a result of exercise training
or other interventions.
•Motivate individuals to initiate an exercise program or comply with an established
program.
•Provide information concerning health status.
A few assumptions regarding testing are necessary to ensure the highest degree of
accuracy when using sub-maximal exercise testing to estimate VO
2max
:
•Selected workloads are reproducible. A steady-state heart rate is obtained during
each stage of the test. Usually, workload durations of 3 minutes or more are used to
ensure steady state.
• The maximal heart rate for a given age is uniform (HR = 220-age).
• Heart rate and VO
slope of HR/VO
have a linear relation over a wide range of values; thus, the
2
regression can be extrapolated to an assumed maximum heart
2
rate.
•Mechanical efficiency (i.e., VO
at a given work rate) is consistent.
2
Although if done correctly, sub-maximal exercise tests provide valuable information
concerning cardiorespiratory fitness, they have extremely limited diagnostic capabilities
and should not be used as a replacement for clinical exercise tests or other clinical
treatment or management modalities. Health care professionals should avoid detailed
interpretation beyond the scope of the information obtained.
Considerations with sub-maximal exercise testing
Considerations for selection of protocol and equipment include any physical or clinical
limitations that may preclude certain types of exercise (i.e., age, weight, arthritis,
orthopaedic complications, individual comfort, level of fitness, type of exercise training
that will be performed, and individual preference).
For example, some individuals may perform better on a non-weight-bearing modality
(cycle versus treadmill), while others may not have the required range of motion in the
hip or knee to pedal and may perform better walking. Deconditioned, weak, or elderly
persons may have to start the test at a low work level and increase the workload in small
increments. Also, field tests may not be appropriate for those who require strict
supervision during testing, who do not understand the concept of pacing, or who cannot
be expected to put forth a good effort. More consistent results may be obtained by
testing in a controlled environment such as a laboratory setting. Creativity when
selecting protocols may allow adaptations of commonly used protocols to accommodate
athletes competing in specific sports. Regardless of the type of exercise and protocol
selected, the same type of exercise and protocol should be used for repeat testing if
between-test comparisons are important.
94 - Quark b2 User Manual
Staffing
Test termination
Staff members should be able to do the following:
1. Establish rapport with the subject and make him or her feel comfortable.
2. Recognize normal acute and chronic responses to exercise.
3. Recognize abnormal signs and symptoms during exercise.
4. Provide basic life support measures competently.
5. Adhere to established procedures and protocols.
6. Clearly explain test results to the individual.
Sub-maximal tests should be terminated according to ACSM or other accepted
guidelines (see table in the following). In the event of an abnormal response, the test
should be terminated, the medical director of the facility and the individual’s primary
care physician notified, and all specified follow-up procedures performed. In the event
of mechanical or electrical failure that may compromise the accuracy of the test results
or monitoring capabilities, the test should be terminated until the problem is corrected.
General Indications for Stopping an Exercise Test in Apparently Healthy Adults
Onset of angina or angina-like symptoms
Significant drop (20 mmHg) in systolic blood pressure or a failure of the systolic blood
pressure to rise with an increase in exercise intensity
Excessive rise in blood pressure: systolic pressure >260 mmHg or diastolic pressure
>115 mmHg
Signs of poor perfusion: tight-headedness, confusion, ataxia, pallor, cyanosis, nausea, or
cold and clammy skin
Failure of heart rate to increase with increased exercise intensity
Noticeable change in heart rhythm
Subject requests to stop
Physical or verbal manifestations of severe fatigue
Failure of the testing equipment
Assuming that testing is non-diagnostic and is being performed without direct physician
involvement or electrocardiographic monitoring.
Chapter 8 - Sub-maximal Exercise Testing - 95
Considerations for accuracy
The ability to obtain valid and reproducible results is essential to ensure that any
differences between pre-treatment and post-treatment test results are due to exercise
training rather than variations in testing procedures. Some inconsistencies that are
inherent may increase variability:
•Sub-maximal heart rate is influenced by time of day, eating, smoking, and
familiarization with test procedures.
•Prediction equations for estimating VO
and underestimate untrained individuals.
• The efficiency of motion during walking, running, and cycling varies.
• Cardiac output and VO
Psychological factors, such as pre-test anxiety, may influence the heart rate, especially
at rates below 120 beats per minute and at low workloads. It is not unusual for the heart
rate and/or blood pressure to be higher at rest than during the initial stages of exercise in
these cases. Having the subject repeat the first test may improve reliability, particularly
if the subject has never previously performed such a test.
Factors that can cause variation in the heart rate response to testing:
• Use of alcohol, tobacco, or caffeine 2 to 3 hours prior to testing
Because of these inherent inconsistencies, standard procedures for each test must be
strictly followed to ensure the greatest accuracy and reproducibility possible:
• Standard testing protocol
• The same testing modality and protocol for repeat testing
• A constant pedal speed throughout cycle ergometry testing
• Cycle seat height properly adjusted, recorded, and standard for each test
• The time of day for repeat testing consistent
• All data collection procedures standardized and consistent
• Test conditions standard
• Subjects free of infection and in normal sinus rhythm
• Prior to the test, no intense or prolonged exercise for 24 hours, smoking for 2-3
hours, caffeine for 3 hours, or heavy meal for 3 hours
•Room temperature 18-20°C (64-68°F) with air movement provided
may overestimate trained individuals
2max
have a test-retest variability of 3-4%.
2
96 - Quark b2 User Manual
Performing the test
In this chapter it is supposed that the user is able to:
• perform an exercise test
• create exercise protocols
• view, edit and print tests
If this is not the case, please read the Exercise testing chapter.
To perform a sub-maximal test, follow these instructions:
1. Create a proper protocol (procedural guidelines for several sub-maximal testing
protocols are provided in [ACSM’s Guidelines for Exercise Testing and
Prescription, 6
2. Start an exercise test.
3. Perform the test as it were a maximal exercise test, ending it when the heart rate
reaches the 85% of the Hrmax, or it happens an event listed in the section Test termination.
If the predicted HR max (calculated as 220-age) is not suitable for the patient tested, it
is possible to edit the HR max value from the View/Information… page.
An example of testing protocol
An example of protocol is reported here. The YMCA cycle ergometry protocol is
defined as follows.
st
1
step: workload 150 kgm/min
nd
2
step: if the HR at the end of the 1st step is: <80 80-89 90-100 >100
set the workload at (kgm/min) 750 600 450 300
rd
3
step: if the HR at the end of the 2nd step is: <80 80-89 90-100 >100
set the workload at (kgm/min) 900 750 600 450
th
4
step: if the HR at the end of the 3rd step is: <80 80-89 90-100 >100
set the workload at (kgm/min) 1050 900 750 600
Chapter 8 - Sub-maximal Exercise Testing - 97
The mixing chamber
The mixing chamber
Overview
The mixing chamber is a 7-litres plexiglas box, for exercise or resting (VE<40 l/min)
tests. For resting tests only a part (about 2.3 litres) of the mixing chamber is used.
The mixing chamber is shown in the following picture:
6
3
1
4
7
5
2
7
1. Connector for the sampling line, for resting tests or tests with VE<40 l/min.
2. Connector for the sampling line, for exercise tests or tests with VE>40 l/min.
3. Inlet for patient’s exhaled air.
4. Connector to be closed with the proper plug supplied with the equipment.
5. Outlet for patient’s exhaled air.
6. Little plug for closing the connectors #1 or #2.
7. Fixing holes.
Preparing the mixing chamber for a test
1. Connect the wrinkled tube to the inlet #3 of the mixing chamber.
2. Connect the turbine to the outlet #5 of the mixing chamber.
3. Disconnect the sampling line from the turbine and connect it to the connector #1
(for resting tests) or #2 (for exercise tests) of the mixing chamber.
4. Close the connector #2 (for resting tests) or #1 (for exercise tests) of the mixing
chamber with the little plug #6.
5. Close the connector #4 with the proper plug supplied with the equipment.
Two-way non rebreathing valve description
The two-way non rebreathing valve is very important in order to perform the test
properly. It is shown in the following picture:
The mixing chamber - 99
13
1. Valve inlet
2. Connector for the mask
3. Valve outlet
Note: be very careful in order to differentiate inlet from outlet. These two are not
interchangeable, to guarantee proper functionality.
Patient’s preparation
1. Screw the mask to the connector #2 of the valve.
2. Connect the wrinkled tube to the outlet #3 of the valve.
3. Fix the mask as illustrated in the picture below. Adjust the elastic bands on the head
2
cap as necessary to eliminate possible leaks and create a tight seal around the
subject’s face.
Performing the test
100 - Quark b2 User Manual
4. Complete the patient preparation as indicated in the chapter Exercise testing.
1. Calibrate the analyzers as described in the Calibration chapter
2. Select Test/Execute Test or press the icon
3. Select Mixing chamber in the Execute test window and perform the
exercise/resting test as illustrated in the chapter Exercise testing.
or press F2.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.