ENTERPRISE
93025 SysCal™ Calibration System
WIRELESS
MONITORING
OVERVIEW
The Syscal calibration process was designed to monitor the accuracy of a previously installed TempTrak® transmitter and its modified*
Solid Simulator Probe. This process will be used with the transmitter and solid simulator probe connected, measured as a unit, and in
place within the use environment.
Calibration of the transmitter and solid simulator is achieved by comparing the documented readings from the transmitter and solid
simulator with the reading from the SysCal™ 35025 Thermocouple instrument and the 50219-K Probe. This reading is referred to as the
reference standard.
The Total System Accuracy (TSA) of SysCal is ±0.5°F (±0.3°C). A certificate of calibration from an ISO 17025 accredited laboratory is
also supplied. This accuracy will provide a Test Uncertainty Ratio (TUR) of 4:1 when used as the reference standard for the TempTrak
transmitter and Solid Product Simulator.
*The modified solid simulator incorporates a secondary hole in the opposite end from where the thermistor sensor is inserted into the solid simulator. This secondary hole is used to place an
insertion probe in the vicinity of the simulator’s thermistor sensor.
FEATURES
• Modified Solid Simulator
• On-site testing/certification
• Simple and easy to use
SPECIFICATIONS
BENEFITS
• Provides a way to insert the probe to allow temperature stabilization
• Saves time and labor in the recertification process
• Training times reduced
• System Components: (1) 35025 thermocouple instrument and
(2) 50219-K probes
• System Range: -25°F to 180°F (-32°C to 82°C)
• System Accuracy: ±0.5°F (±0.3°C) when probe/instrument are
calibrated together and ambient temperature is between 68°F and 86°F
(20°C and 30°C)
• Ambient Temperature Range: Operating: 32°F to 122°F (0°C to 50°C)
• Ambient Accuracy: Operating: At ambient temperatures between 32°F
to 68°F (0°C to 20°C) and 86°F to 122°F (30°C to 50°C), add ±0.1°
error per °F or °C deviation from 68°F and 86°F (20°C and 30°C)
• Resolution: 0.1°/1° user selectable
SysCal™ Calibration System
#93025
(1) 35025 Thermocouple instrument
(2) 50219-K probes
(1) ISO 17025 certificate of calibration
(1) Calibration log sheet
(1) Hard carrying case
(48) Calibration labels
(48) Blank labels
The reading from the TempTrak transmitter is
compared to the reference standard recorded
by the SysCal thermocouple. If that reading
is within ±2.0°F (±1.1°C) the transmitter and
solid simulator probe are considered to be
calibrated within the stated accuracy.
• Display: LCD with 0.5” (13 mm) digit height
• Battery Type: (2) 1.5V (AAA) batteries (ANSI/NEDA-24F, IEC-R03)
• Battery Life: 1800 hours typical
• Compliance/Regulatory: Calibrated in an ISO 17025 accredited lab,
CE certified, NSF listed, WEEE marked and RoHS compliant.
We certify that this instrument has been calibrated to temperature and/or
electrical standards traceable to the U.S. Department of Commerce National
Institute of Standards and Technology (NIST).
Cooper-Atkins Corporation | Accuracy to the Highest Degree
11353 Reed Hartman Hwy, Suite 110 | Cincinnati, OH 45241-2929 U.S.A.
888-533-6900 | 513-793-5366 | Fax: 513-793-4895 | www.cooper-atkins.com
67-1489
V1014