WIRELESS
#10208 Oxygen Depletion Monitor
OVERVIEW
Escaping liquid nitrogen gas can quickly deplete oxygen levels and cause asphyxiation. Used in conjunction with the Cooper-Atkins TempTrak Wireless
Monitoring System, the O2 sensor is easy to use, easy to calibrate and accurate. The O2 sensor can be attached to the 0-5V analog transmitter to log
oxygen levels from 0 to 25%. When oxygen levels become dangerous, an alarm at the sensor in the liquid nitrogen tank room (as well as outside the
room) warns people nearby via an ethernet-connected device. Available for 900MHz and Wi-Fi (802.11 b/g/n).
MONITORING
FEATURES
• Analog attachment measures the percentage of oxygen in a room
• Offers multiple modes of customizable alarming – email, visual and
audible (local), remote smart devices, alarm escalation and repeater
device (placed outside of room)
• Rugged construction yields 10 years of use. 2-year extended warranty
• O2 sensor is rated IP65 and Repeater IP54
• Sensor technology utilizes an electrochemical cell with atmospheric
pressure compensation
• Due to size, this device is not practical for monitoring oxygen levels in
incubators
SPECIFICATIONS
Oxygen Depletion Monitor
#10208-802 – Wi-Fi (802.11 b/g/n)
#10208-900 – 900 MHz
• Range: 0 to 25% Oxygen
• Accuracy: ±0.75% O2 over 5.0 to 25.0 % O
±1.00% O2 over 0.1 to 5.0 % O
• Alarm: Two presets - 19.5% alarm &18% alarm
• Alarm Decibel Level: 85 dbA
• Power: 110V AC power supply with battery back up
• Battery Back Up: Minimum of 4 hours (unit must be powered for a
minimum of 30 hours to fully charge battery back up)
• Repeater Modes: Sounder and LED Visual Strobe
• Display: 4 digit LCD - 0.5” x 1.625” (13 mm)
• IP65 Waterproof Rated
• 2-year Warranty
2
2
BENEFITS
• Monitors the percentage of oxygen to prevent asphyxiation.
• Multiple mode of alarming provides redundancy of alarms for
maximum safety
• Long lasting and cost effective
• O2 sensor is capable of operating in a wash down environment.
• Highly accurate readings
• Can be easily transferred to a new location
• Dimensions:
-Analyzer: 6.89” x 2.95” x 3.46” (175 mm x 75 mm x 88 mm)
-Repeater: 2.84” x 1.77” x 5.78” (72 mm x 45 mm x 147 mm)
-Battery Back Up: 2.84” x 1.77” x 5.78” (72 mm x 45 mm x 147 mm)
• Cord Lengths:
-Repeater: 12” (0.31 m)
-Power: 3’ (0.91 m)
-Output: 3’ (0.91 m)
-Battery Back Up: 7’ (2.13 m)
• Weight: 3 lbs (48 oz) (Full System)
Detailed Installation Guide provided.
Cooper-Atkins does not offer installation of this unit.
Oxygen Depletion Monitor
#10208-900 (900MHz)
#10208-802 (Wi-Fi)
Cooper-Atkins Corporation | Accuracy to the Highest Degree
11353 Reed Hartman Hwy, Suite 110 | Cincinnati, OH 45241-2929 U.S.A.
888-533-6900 | 513-793-5366 | Fax: 513-793-4895 | www.cooper-atkins.com
67-1525
V0915