ENTERPRISE
Retrofit Solid Simulator & Air Probe
WIRELESS
MONITORING
OVERVIEW
Air temperatures fluctuate when refrigerator doors are opened and closed, but the actual temperature of the item stored in the cabinet is not affected
unless the door is left open for a significant period of time. In response to the need to represent the actual product temperature, Cooper-Atkins has
developed a series of product simulators. Now you can turn your existing air probe into a product simulator probe with the Retrofit Solid Simulator.
The #10185 is designed to fit tightly over the Cooper-Atkins #2033 Air Probe, thus insulating/buffering the probe tip from the frequent air temperature
changes. The material composition and thickness of the wall provide accurate simulated product temperature during changing ambient conditions.
By using the Retrofit Solid Simulator to monitor refrigeration or freezer cabinets, the ambient temperature of the product inside will be registered.
Assembly is simple and directions are included with each Retrofit Solid Simulator.
*See back for test results.
FEATURES
• Can be installed on existing equipment
• Simulates true product temperature
• FDA approved solid simulator material
• Durable casing
SPECIFICATIONS
#10185 Retrofit Solid Simulator
• Temperature Range: -25° to 180°F (-32° to 82°C)
• Stabilization: Up to 2 hours
• Material: FDA-approved Acetal
• Dimensions: 1.5” x 1.5” (38 mm x 38 mm)
• Weight: 2 oz (59 g)
• Patent pending
• 1-year warranty
Retrofit Solid Simulator
#10185
Hole Diameter
0.213” (5.4 mm)
Hole Depth
0.875” (22 mm)
BENEFITS
• Simple and affordable
• Higher accuracy than air temperature
• No potential of leakage or evaporation
• Impact resistant, ideal for high traffic areas
#2033 Air Probe
• Temperature Range: -25° to 180°F (-32° to 82°C)
• Accuracy: ±0.4°F (±0.2°C) from 32° to 158°F (0° to 70°C)
• Response Time: 40 sec in 5 m/sec airstream
• Shaft Length: 0.5” (13 mm)
• Cord Length: 6’ (1.8 m)
• Weight: 0.5 oz (14 g)
• Traceable to Standards of NIST
• 1-year warranty
1.5” (38 mm)
1.5” (38 mm)
ACTUAL
SIZE
Air Probe
#2033
Cooper-Atkins Corporation | Accuracy to the Highest Degree
888-533-6900 | 513-793-5366 | Fax: 513-793-4895 | www.cooper-atkins.com
Air Probe
Set Screw
Retrofi t Simulator Probe
ENTERPRISE
Retrofit Solid Simulator & Air Probe
WIRELESS
MONITORING
SIMULATOR TEST
A test was conducted using an environmental chamber over a two hour time span in order to illustrate graphically how the #2033 Air
Probe reacted to the changing ambient air temperature in comparison to the same Air Probe with the #10185 Retrofit Solid Simulator
attached. Once both probes were allowed to stabilize at 41˚F, the chamber door was opened for 5 minutes, then shut for an hour.
This process was repeated once more and the probe readings were recorded every 5 minutes. The graph below shows how the Air
Probe reacted immediately when the door was opened, rising 7-10 degrees, while the Retrofit Solid Simulator remained stable,
rising only 4 degrees after 5 minutes.
Test: Air Probe vs. Retrofit Solid Simulator
Door: Open / Closed
Retrofi t Solid SimulatorAir Probe
Cooper-Atkins Corporation | Accuracy to the Highest Degree
888-533-6900 | 513-793-5366 | Fax: 513-793-4895 | www.cooper-atkins.com
Door: Open / Closed
67-1460
V0914