Comark KM820 User Manual

Calibration Procedure - KM820/V
Verifying Infra-Red Non contact thermometers
Insert the reference thermometer probe (e.g.
KM20Ref) fully into one of the holes in the Validator.
Allow the instruments to stabilise to environment
conditions.
Switch on both instruments after stabilisation
period.
Point the Infra-Red thermometer at the base of the
Validator while ensuring the nozzle is as far inside the Validator cup as is possible. Note the variation. (acceptable error 2.5°C)
Verifying Contact Thermometers
Insert the reference thermometer probe (e.g.
KM20Ref) fully into one of the holes in the Validator.
Insert the probe of the Instrument being veried
fully into the other hole in the Validator.
Allow the instruments to stabilise to environment
conditions.
Switch on both instruments after stabilisation
period.
Note the variation. The acceptance error will be
dependent on the specication of the unit being veried.
Note: The time for the instruments to stabilise will be dependent on the validator reaching its environment temperature, which at freezer temperatures could be up to 2 hours.
Care must be exercised when handling the validator after exposure to low temperatures.