CME T-34 Nikki Syring Pump Field service manual

CAESAREA MEDICAL ELECTRONICS
NIKKI T34
AMBULATORY INFUSION PUMP
Please Note: This is a simple guide to assist with SPT and EPP while in the field. Any works required outside the guide will require unit to be returned to workshop Guide use is restricted to Chemtronics trained personnel.
CME NIKKI T34 FIELD TEST GUIDE
1
INTRODUCTION
This guide is designed to assist in with SPT and EPP testing. A basic understanding of pump operation and functionality is required. Unit Operational and Service manual can be found in Bendigo Workshop or online through Citrix at http://warney/ServiceManuals/ (CME/NIKI T34 SYRINGE PUMP/)
Test result shown in this guide have been achieved with unit set to 24:00 hr mode with Terumo (5, 10,
20, 30 & 50/60ml) syringes activated, Occlusion level set to 540mmHg (10.44psi) and procedures found in Service Manual (Ref: 100-090SS), Instruction Manual (Ref: 100-090SA) and online at
http://www.cme-infusion.com/documents/pub/index.html
Test Equipment required
Terumo 20 and 50/60ml syringes
Infusion Analyser (IDA)
Vernier callipers (not essential)
Brand new 9V alkaline battery
Access Codes
Technician Code – 123. To access Tech Mode Hold while turning unit on
Hold for 3 seconds, then access code prompt will appear.
Setup Code – 99. To access Setup Mode turn unit on, allow unit to finish Self
Test and Pre-loading. Press Scroll or until Change Set up is active.
1. VISUAL INSPECTION
Inspect unit for damage.
I. Confirm Barrel Clamp Arm moves freely and rotates only when fully extended
II. Confirm Syringe Collar and Sensor are free from obstruction with no damage.
III. Confirm Plunger Sensor/Actuator has no damage and free from obstruction.
CME NIKKI T34 FIELD TEST GUIDE
2
2. BASIC TEST.
Fit new 9v alkaline battery and turn on by pressing for more than 1 sec.
I. Confirm Operation LED blinks RED and Green and backlighting illuminates.
II. Confirm Model, Version and ID are all displayed at start up
III. Confirm Plunger travels to far right during Pre-Loading
IV. During Start up sequence confirm battery level is above 95%
V. Confirm unit finishes start up sequence with Load Syringe displayed & syringe flashing
TEST COMPLETE
CME NIKKI T34 FIELD TEST GUIDE
3