CME BodyGuard 323 User manual

Page 1
.
BodyGuard 323
Infusion Pump System
Manufacturer:
Caesarea Medical Electronics Ltd. European address: Staufenburgstr. 23 Lichtenstein P.O.B. 1248 Lichtenstein 72805 Germany International Toll Free: +800-323-575-00
www.cme-infusion.com
Contact e-mail: Sales@cme-infusion.com
EDITION July 2009
0344
Ref.: 100-090X
Page 2

NOTE:
all users to read this manual carefully, become familiar with the procedures and system functions and follow all recommendations.
To assure proper use of the BodyGuard, Caesarea Medical Electronics Ltd. urges
Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary. Please contact your technician in order to enable the air sensor if disabled with no need.
U.S.A. federal law restricts this device to sale by or on the order of a physician.
BodyGuard Operator Manual
2
Page 3
Table of Contents
1- Introduction................................................................................................5
Overview.................................................................................................5
Button Descriptions and Functions .....................................................5
BodyGuard Pump with Door Open.......................................................8
The Charger............................................................................................ 9
System Safety Checks......................................................................... 10
Pump Accuracy.................................................................................... 11
2 – Symbols, Warnings and Cautions ........................................................ 15
System Symbols .................................................................................. 15
Terms Used In Manual......................................................................... 16
Intended Use ........................................................................................16
Warnings .............................................................................................. 16
Cautions ............................................................................................... 18
3 – Installation and Set up...........................................................................19
Unpacking ............................................................................................ 19
Charging the Pump.............................................................................. 20
4 – Operation ................................................................................................ 21
Before Operating the Pump ................................................................21
Pump Operation Test...........................................................................21
Air in Line – Alarm Test....................................................................... 22
Down Occlusion – Alarm Test ............................................................22
Door Open – Alarm Test...................................................................... 23
Battery Test .......................................................................................... 24
Charger Indicator Test.........................................................................24
Dedicated BodyGuard Administration Sets ......................................25
The BodyValve ..................................................................................... 25
Loading the Administration Set.......................................................... 29
Priming the Line................................................................................... 30
Prime – In Continuous, Intermittent, TPN , and 25 Steps Modes.....30
Prime – In PCA Mode........................................................................... 31
Prime Operation – After Air In Line Alarm .........................................32
Manual Prime Operation......................................................................32
Operation steps.................................................................................... 33
Pump Mounted on IV Pole...................................................................33
Portable Connection............................................................................ 33
Programming........................................................................................34
Changing Current Program – Using Level 1 Code............................35
Changing Current Program – Short Way ...........................................35
Continuous Infusion ............................................................................36
Intermittent Infusion ............................................................................ 41
Program Delay...................................................................................... 46
25 Steps Protocol................................................................................ 48
TPN Infusion......................................................................................... 50
PCA Infusion ........................................................................................53
Setting Pressure Default For Down Occlusion..................................64
BodyGuard Operator Manual
3
Page 4
Lock Mode ............................................................................................ 65
How To Lock.........................................................................................65
Info Mode..............................................................................................67
Review Set up....................................................................................... 69
Change Set up...................................................................................... 70
5 – Alarm Conditions and Troubleshooting............................................... 72
Alarm Condition ...................................................................................72
Troubleshooting...................................................................................73
6 – Specifications.........................................................................................75
7 – Service and Maintenance ......................................................................77
Cleaning................................................................................................ 77
Storage ................................................................................................. 77
Battery Operation.................................................................................78
LIMITED WARRANTY...................................................................................79
BodyGuard Operator Manual
4
Page 5
1-
Introduction
TPN
Intermittent
Continuous
PCA
25
Steps
1-Introduction
Overview
The BodyGuard System provides the following features:
Small, light and compact pump.
Multiple Programs:
Infusion Rate: 0.1 to 1200 ml/hr Volume: 0.1 to 99.9 ml in 0.1ml increments or 1 to 9999 ml in 1ml increments Secondary Infusion (Piggy) programmable with the same infusion rate and volume capabilities as the primary continuous infusion.
A set dose delivered in set intervals. Between Intervals the pump will keep vain open (KVO Mode).
Total Parenteral Nutrition Rate Taper program with programmable up and down times.
Patient Control Analgesia: Continuous delivery plus /or programmed boluses. Bolus Volume: 0-25.5 ml Clinician Bolus Volume: 0.1-100ml Bolus Rate: 0.1-1200 ml/h
Highly accurate fluid delivery
Operated with a custom safe administration set
Rechargeable internal lithium-ion battery or alkaline 2X9V batteries
Battery charged in a pole mount charging cradle or with an A/C adaptor
Free flow protection valve standard on every CME administration set
Quiet operation
Used for ambulatory care in a carrying pouch or
mounted on an IV pole
AEA – Suitable to be used in ambulances.
Operator can design a specific protocol for drug delivery which contains up to 25 steps. Infusion Rate: 0.1 to 1200 ml/hr Volume: up to 9999ml, for each step Accumulated Volume for all steps – up to 10 liters.
BodyGuard Operator Manual
5
Page 6
1-Introduction
1. Display Screen
2. Up Arrow
3. START/ OK
4. S
TOP/NO
3
4
8
2
10
9
Front of
BodyGuard
Infusion Pump
1
5
6
7
Displays pump and infusion status
Displays programming choices and instructions
Scrolls up through options
Starts infusion
Confirms selection and setting
Stops infusion
Silences an alarm condition
Pause priming
Zeroes the displayed value during programming
Erases the last digit during programming
Returns to previous screen
BodyGuard Operator Manual
6
Page 7
8. INFO
1-Introduction
5. PRIME/BOLUS
Pressing Prime/Bolus key during data setting: Enables Priming procedure. A graph appears on the display screen showing the priming volume with the current value increasing until reaching the set volume.
Pressing Prime/Bolus key during Continuous program operation: Enables Secondary (Piggy) Infusion.
Pressing Prime/Bolus key during PCA program operation: Enables bolus procedure.
6. Power ON/OFF
Turn the system ON by pressing and holding the button until the Self-Test screen appears.
Turn the system OFF by pressing and holding the button until the graph is black and a beep is generated.
7. Operation LED
Green Indicator
Lights during system Self-Test
An intermittent green light indicates infusion delivery
Red Indicator
Indicates an alarm state with a continuous red light
Lights when the pump is in a stand-by mode during programming
Supplies information about the pump and its programs (see Info Mode
chapter).
Pressing continuously locks and unlocks the keypad to prevent accidental or deliberate change to pump operation (If needed).
9. Down Arrow
Scrolls down through options
10. Numeric Keys
Enters numeric parameters during programming
BodyGuard Operator Manual
7
Page 8
1-Introduction
3
5
BodyGuard Pump with Door Open
7 8 9
# Area Function
1.
2.
3. Keyway Guides the IV line
Latch Holder
Pressure Sensor Detects downstream tubing restriction and occlusion. Alarm
Holds the door closed when latch is in vertical position.
level can be adjusted to suit patient needs.
1
6
2
4
4.
5.
6. Flow Direction Shows direction of pump operation
7. Anti-Stretch locator (Optional) Locates the tube in the correct length from the key
8. Air sensor – Front Ultrasonic air detector, which is mounted on the front
9.
Pressing Plate
Pump Door Covers the pressing plate
Air sensor - Door
Connected to the door by two springs
housing.
Mounted on the door
BodyGuard Operator Manual
8
Page 9
1-Introduction
The Charger
# Area Function
1.
Charger Display Displays major pump messages and operation values.
1
2
2.
Release Handle Push to release the pump from the Charger.
BodyGuard Operator Manual
9
Page 10
1-Introduction
System Safety Checks
The following details outline the safety checks designed into the BodyGuard infusion system, to minimize the possibility of under or over infusions.
Free Flow Protection
The system’s customized IV lines are equipped with a check valve that prevents free flow towards the patient when the IV line is not attached to the pump. When the pump is attached to the IV line and delivering fluid, the pressure delivered by the pump opens the valve. The valve is also one way valve preventing fluids flow from the patient to the administration line.
Air-in-Line Accumulation
To improve the detection of air in the IV line, the BodyGuard 323 Infusion system utilizes an air-in-line accumulation system in addition to the standard single bubble detection. This feature monitors the volume of air that passes through the IV line by accumulating the volume of individual bubbles over any 15 minutes period. This limit is not configurable. Although an individual bubble may not exceed the pre­programmed threshold, if the cumulative volume of smaller bubbles exceeds 1ml over a period of 15 minutes an air-in-line alarm is initiated. This accumulation feature is particularly useful with infusions for patients who are highly sensitive to air (i.e. infants, neonates, children) or when infusing products that create significant volumes of small air bubbles.
Anti-Bolus Function
The anti-bolus function is designed to reduce the bolus that may occur upon the release of an occlusion following a downstream occlusion alarm. Upon the detection of a downstream occlusion, the alarm is activated and the pump returns the IV line pressure to neutral within 15 seconds. Neutral line pressure is achieved by the reverse operation of the pumping mechanism, and measurement of the IV line pressure through the in-line pressure detection system.
BodyGuard Operator Manual
10
Page 11
1-Introduction
Pump Accuracy
The following graphs and curves were derived from testing described in IEC60601-2-24. Testing was performed under normal conditions at room temperature (720 F). Any deviations from normal conditions and room temperature may cause changes in the accuracy of the pump.
Start-up Curves
The Start-up curves represent continuous flow versus operating time for two hours from the start of the infusion. They exhibit the delay in onset of delivery due to mechanical compliance and provide a visual representation of uniformity. Trumpet curves are derived from the second hour of this data. Tests performed according to IEC 60601-2-24 standard.
Trumpet and Flow Rate
With the BodyGuard 323, as with all infusion systems, the action of the pumping mechanism and variations cause short term fluctuations in rate accuracy.
The following curves show typical performance of the system in two ways:
1. The accuracy of fluid delivery over various time periods is measured (trumpet curves).
2. The delay in onset of fluid flow when infusion commences (start up curves).
Trumpet curves are named for their characteristic shape. They display discrete data averaged over particular time periods or “Observation windows”, not continuous data versus operating time. Over long observation windows, short-term fluctuation has little effect on accuracy as represented by the flat part of the curve. As the observation window is reduced, short-term fluctuations have greater effects as represented by the “mouth” of the trumpet.
BodyGuard Operator Manual
11
Page 12
1-Introduction
Knowledge of system accuracy over various observation windows may be of interest when certain drugs are being administered. Short-term fluctuations in rate accuracy may have clinical impact depending on the shelf life of the drug being infused and the degree of inter-vascular integration. The clinical effect cannot be determined from the trumpet curves alone.
BodyGuard Operator Manual
12
Page 13
1-Introduction
BodyGuard Operator Manual
13
Page 14
1-Introduction
BodyGuard Operator Manual
14
Page 15
2-Symbols, Warnings and Cautions
2 – Symbols, Warnings and Cautions
System Symbols
The following symbols are used on the BodyGuard Infusion System and components. Labels on the system or statements in this manual proceeded by any of the following words and/or symbols are of special significance and are intended to help you to operate the pump in a safe and successful manner.
Attention, consult accompanying Instructions
CE mark indicates conformance to Medical Device Directive 93/42/EEC
0344
Do not dispose of in municipal waste. Symbol indicates separate collection for electrical and electronic equipment. (WEEE Directive 2002/96/EEC). NOTE: Does not apply to the battery.
Do not dispose of battery in municipal waste. Symbol indicates separate collection for battery is required.
LOT
STERILE EO
The use of single-use disposable components on more than one patient is a biological hazard. Do not reuse single-use disposable components.
Type CF applied part
Date of Manufacture
Serial Number
Expiry Date of disposable
Lot Number
Sterilized with Ethylene Oxide
BodyGuard Operator Manual
15
Page 16
Terms Used In Manual

Warning:
circumstances that could result in injury or death to the patient or operator. Read and understand this manual and all warnings completely before operating the
BodyGuard Infusion System.
Caution:
that could result in damage to the device. Read and understand this manual and all cautions completely before operating the BodyGuard Infusion System.
NOTE:
that will help you operating the BodyGuard Infusion System.
Indicates that the information is a warning. Warnings advise you of
Indicates that the information is a caution. Cautions advise you of circumstances
Indicates that the information that follows is additional important information or a tip
2-Symbols, Warnings and Cautions
Intended Use
The BodyGuard infusion Pump system is designed for infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, enteral, in close proximity to nerves, and into an intraoperative site (soft tissue/body cavity/surgical wound site). The system is intended for patients who require maintenance medications, analgesics, PCA therapy, parenteral and enteral nutrition fluids, chemotherapeutic agents and general fluids therapy in hospital and home care environments.
Warnings
To avoid possible personal injury or loss of life, observe the following:
Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary. Please contact your technician in order to enable the air sensor if disabled with no need.
Read the entire Operation Manual before using the pump, since the text includes important precautions.
The maximum volume that may be infused under SINGLE FAULT CONDITION is 0.1 ml
Voltage present on internal components may cause severe shock or death upon contact. Disconnect the Charger from the mains, prior to opening the casing. Only trained service personnel should open the pump cover.
Blown fuses could cause a fire hazard. Replace blown fuses on the Charger only with fuses of the same type and rating (see fuse values on the Charger PCB).
BodyGuard Operator Manual
16
Page 17
2-Symbols, Warnings and Cautions
Boluses will be infused up to the full volume that is requested unless the user stops the bolus
manually by pressing . This is true even when the volume infused is less than the specified Hour Limit for Volume (e.g. 4 Hr Limit) at the beginning of a bolus and goes over the limit during the bolus (to complete it).
Make sure the pump is attached securely to the Charger, which is connected firmly to an IV pole.
A kinked or occluded IV line may impair the operation of the pump and the accuracy of the infusion. Before operation, verify that the IV line is not kinked or occluded.
The BodyGuard should be operated only with the Bodyset IV lines. Use of IV lines other than the Bodyset may impair the operation of the pump and the accuracy of an infusion.
Drugs must not be administered to the epidural space unless the drugs are indicated for this purpose and are administered in accordance with the indications included in the manufacture’s package. Epidural administration of drugs other than those indicated for epidural use could result in serious injury to the patient. For epidural administration of drugs use MicroSet only.
Any adjustments, maintenance, or repair of the uncovered pump may impair the operation of the BodyGuard Infusion System and/or the accuracy of the infusion. Any adjustments, maintenance, or repair of the uncovered pump or charger should be performed by authorized skilled technicians. Any adjustments, maintenance, or repair of the uncovered pump or charger while connected to the power should be avoided.
The BodyGuard Infusion System should be operated within a temperature range of 18°C (59°F) to 45°C (113°F) and up to 85% humidity. Operating the pump at temperatures and/or humidity other than within that range may affect accuracy.
Unsafe operation may result from using improper accessories. Use only accessories and options designed for this system.
Disposables must be compatible with the medicine delivered.
Battery charging is enabled as long as the charger cord is connected to the mains and the pump is in the charger. Switching the pump off does not disconnect it from the mains. To disconnect from mains, remove the charger cord from mains. To disconnect pump from mains, remove it from the charger.
Dropping the BodyGuard Infusion System could cause damage to components. If the pump is dropped, return the pump for inspection by qualified service personnel.
Use aseptic technique. Patient infection may result from the use of non-sterile components. Maintain sterility of all disposable components and do not re-use single use IV sets.
Watch your fingers and nails when opening the pump door.
When operating the pump on PCA program with a rate of 0.0 ml/h there is a hazard of blood clot forming. Connect saline infusion in parallel to avoid this problem.
When operating the pump on PCA program, the program will not enter KVO, even if the limit volume is acceded. The program will complete the bolus and than enter KVO.
Do not operate the pump near high-energy radio-frequency emitting equipment, such as electro­surgical cauterizing equipment. False alarm signals may occur.
BodyGuard Operator Manual
17
Page 18
2-Symbols, Warnings and Cautions
Do not let the pump operate when battery is fully depleted. Pump may turn off during operation on fully depleted battery without alarm.
Never operate the pump without performing the battery test as described in chapter 4 in this manual.
Do not use the BodyGuard 323 for infusion of Insulin or for infusion of blood.
Cautions
To avoid possible damage to the equipment, observe the following:
Leaving the battery in a depleted state for a long period of time may damage the battery. Connect the pump to the mains via the Charger whenever possible to recharge the battery.
Do not store the pump with the battery fully depleted.
Xylene, Acetone or similar solvents could cause damage to components. Do not clean the pump with these chemicals. Clean solution spills on the pump immediately. Use a damp cloth or sponge. A mild detergent may be used. Wipe thoroughly with a dry cloth.
Immersing the BodyGuard Infusion pump into liquid could cause damage to components. Do not immerse the pump into any type of liquid.
Battery damage could occur if left in a temperature warmer than 500C (122°F).
BodyGuard Operator Manual
18
Page 19
3- Installation and Set up
3 – Installation and Set
up
Unpacking
1. Carefully remove the pump and Charger from the box.
2. Make sure no items were damaged during shipment
3. Make sure you have the following items:
BodyGuard Infusion Pump
Charging Cradle with Power Cord or Battery Charging Cable
Rechargeable Battery Pack
Operation Manual
PCA bolus cable (optional)
9V Alkaline Battery compartment (optional)
Disposable Battery Pack (optional)
If any items are missing or damaged, contact your BodyGuard supplier.
External charger (optional)
Extended life battery Li-Ion Polymer 3600 mAh (optional)
External wall charger (Optional)
BodyGuard Operator Manual
19
Page 20
3- Installation and Set up
Charging the Pump

NOTE:
charged, connect the pump to the mains via the Charger whenever possible. Be sure that the battery is fully charged at all times.
Warning:
charger connected to mains, the pump will turn off.
1. Connect the charger unit to AC power and verify that the AC indicator is lit.
2. Put the pump into the charger (top first) until it clicks into place.
The pump is protected against overcharging. In order to keep your battery fully
If the battery is removed during operation while pump is not connected to a
.
3. The battery is charging when the red battery LED on the front of the charger is lit. The battery is fully charged when the battery LED symbol turns green.
If the led does not light or does not
turn green after more than 8 hours, pls. replace battery.
4. Remove the pump from the charger by pressing the release button (located next to the LED indicators on the charger) and lift the pump out of the charger bottom first.
Caution:
battery.
Leaving the battery fully depleted for a long period of time may damage the
BodyGuard Operator Manual
20
Page 21
4 - Operation
4 – Operation
Before Operating the Pump
Before attaching the system to a patient, run the following tests to verify that all indicators and alarms
work properly.
When an alarm is activated the following occurs:
An alarm message appears on the display
An audible alarm sounds
Infusion ceases
The LED changes from green to red

NOTE:
Pump Operation Test
1. Click the pump into a Charger connected to AC power. Verify that the CHARGE indicator is on, Red or Green light (on charger).
2. Insert a BodySet administration set and close the door. Press the door until a click is heard.

NOTE:
“Loading the Administration Set” & “Priming the Line”.
All tests should be performed in the Continuous program.
For instructions on how to load and prime an administration set refer to sections
3. Press until the Self-Test screen appears. If air sensor is off, press to confirm.
4. The program selected and its status are displayed. During the Self-Test a long beep will be heard. This procedure verifies that the pump’s acoustic and visual features are working properly. After self test is completed, the display will show the last rate setting.
Press to prime the IV line.
5.
Warning:
6. Enter the Volume of fluid required to prime the set and press
to begin priming procedure. You may stop priming
at any time by pressing .
7. The display screen will show a graph indicating the prime operation.
Ensure the set is not connected to a patient!!
PRIME
Disconnect Patient Press OK to Start
Prime Vol __ ml
Press OK to Continue
PRIME 5 ml
0 5
BodyGuard Operator Manual
21
Page 22
4 - Operation
Volume
ml/h
ml/h
8. When priming is completed, wait 2 minutes without pressing any key. After two minutes the screen displays “Pump
Unattended” and an alarm will sound. Press
silence the alarm again or to exit alarm condition.
to
Pump Unattended
Press OK to Start
Air in Line – Alarm Test
1. Insert a BodySet administration set and close the pump door. Press the door until a click is heard.

2.
3. The pump displays the program selected and its status. The display will automatically change to
4. Change the infusion mode to Continuous (See “Change
5
.
6. Insert air into the IV line to allow an air bubble greater than 10 mm to enter into the administration
7. Air in line alarm will be activated as soon as the air bubble
NOTE:
“Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test Screen appears. If air sensor is off, press to confirm. Please contact your technician in order to enable the air sensor if disabled with no need.
display the last rate setting.
Set up” Section for instructions). Set the infusion rate at
500ml/hr, using the Numeric keypad and press to confirm setting.
Set Volume at 20 ml and press to start operation.
set.
enters the set segment located behind the pump door.
Press to turn the alarm off.
For instructions on how to load and prime an administration set refer to sections
Rate 500
Press OK to Continue
20
Press OK to Continue
Up Occlusion /Air
Press STOP to Mute
Down Occlusion – Alarm Test
1. Insert a BodySet into the pump and close the door. Press the door until a click is heard.
2. Make sure room temperature is 21-23 oC

3.
4. The pump displays the program selected and its status. The display will automatically change to
BodyGuard Operator Manual
NOTE:
“Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test screen appears. If air sensor is off, press to confirm. Please contact your technician in order to enable the air sensor if disabled with no need.
display the last rate setting.
For instructions on how to load and prime an administration set refer to sections
22
Page 23
5. Change the infusion mode to Continuous (See “Change Set
Door Open
Volume
ml
ml/h
Volume
ml
ml/h
up” Section for instructions). Set the infusion rate at 500
4 - Operation
Rate 500
ml/hr using the Numeric keypad and press .
6.
Set Volume at 20 ml and press to start operation.
7. Wait 5 minutes and than connect a pressure gauge at a distance of 1 meter from the exit of the pump. Place both the pump and the pressure gauge at the same level (on a table)
8. Wait another 5 minutes and occlude the IV line downstream of the pump (between the pump and the pressure gauge).
9. The Down Occlusion alarm occurs at the time and rates listed below (time is measured from actual occlusion time to alarm, with accuracy of +30% -20%):
500 ml/h up to 20 seconds
60 ml/hr up to 1 minute
1 ml/hr up to 80 minutes
0.1 ml/hr up to 11 hours

10.
11.
NOTE:
changing the current default Pressure Limit).
Depending upon the set flow rate and pressure threshold, the Down Occlusion alarm will be activated.
Press to mute the alarm.
The pressure threshold can be changed to suit patient need (refer to section on
Press OK to Continue
20
Press OK to Continue
Down Occlusion
Press STOP to Mute
Door Open – Alarm Test
1. Insert a BodySet in the pump and close the door. Press the door until a click is heard.

2.
3. The pump displays the program selected and its status. The display will automatically change to
4. Change the infusion mode to Continuous (See “Change Set
5.
6. Open the pump door.
7. Door Open alarm will be activated at once.
NOTE:
“Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test screen appears. If air sensor is off, press to confirm. Please contact your technician in order to enable the air sensor if disabled with no need.
display the last rate setting.
up” Section for instructions). Set the infusion rate at
500ml/hr using the Numeric keypad and press .
Set Volume at 20 ml press to start operation.
For instructions on how to load and prime an administration set refer to sections
Rate 500
Press OK to Continue
20
Press OK to Continue
Close Door
8.
Close the door and press
BodyGuard Operator Manual
.
23
Page 24
Battery Test
Low Battery
Volume
9999
ml
ml/h
1. Insert a BodySet in the pump and close the door. Press the door until a click is heard.
4 - Operation

2. Take the pump out of the charger.
3.
4. The pump displays the program selected and its status. The display will automatically change to
5. Change the infusion mode to Continuous (See “Change Set
6.
7. Wait untill pump displays “low battery” message
NOTE:
“Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test screen appears. If air sensor is off, press to confirm. Please contact your technician in order to enable the air sensor if disabled with no need.
display the last rate setting.
up” Section for instructions). Set the infusion rate at 1200
ml/hr using the Numeric keypad and press .
Set Volume at 9999 ml and press to start operation.
For instructions on how to load and prime an administration set refer to sections
Rate 1200
Press OK to Continue
Press OK to Continue
Connect to Charger
8. Press to turn the pump off.
9. Connect the pump to a charger.
10. Verify the red icon of battery is lit (on the charger).
Verify the red battery turns green after a minimum of 2 hours.
11.
WARNING: Do not let the pump operate when battery is fully depleted. Pump may turn off during operation on fully depleted battery without alarm.
Charger Indicator Test
1. Click the pump into the charger and connect the Charger to mains. Disconnect the power cord from the AC power outlet.
2.
Verify that the CHARGE indicator is OFF.
BodyGuard Operator Manual
3. Connect the power cord to the AC power outlet. Verify that the CHARGE indicator is ON.
24
Page 25
4 - Operation
Dedicated BodyGuard Administration Sets
The BodyGuard Infusion System should be operated only with BodySet dedicated Infusion sets.
Warning:
other than the BodySet. The BodySet Infusion Sets are equipped with a free flow protection valve, to prevent a gravity free flow hazard. Using a different set may essentially change the delivering rate and expose the patient to a free flow, which can cause serious injury or even death.
Warning:
death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary. Please contact your technician in order to enable the air sensor if disabled with no need.
Do not operate the BodyGuard Infusion System with any administration set,
Using the pump with air in line detector off, may cause an embolism resulting
The BodyValve
The Anti-Free Flow valve, BodyValve, enhances pump functioning by:
Preventing free-flow in the event the set is detached from the pump.
Preventing back-flow (reflux) in the event several infusion pumps are connected simultaneously
to the same patient.
Preventing free-flow in the event of a mechanical malfunction.
Preventing pump operation if the set has been loaded incorrectly.
Warning:
formation.
Replace the administration set every 72 hours to lessen the incidence of bacteria
Warning:
Warning:
damaged components are used. Visually inspect contents and package before each use.
Warning:
a patient to the BodyGuard Infusion Pump until all trapped air has been cleared from the fluid path. Carefully read the instructions for loading. Puncturing set components may cause air embolism.
Warning:
indicated for this purpose and are administered in accordance with the indications included in the manufacture’s package insert accompanying the drugs. Epidural administration of drugs other than those indicated for epidural use could result in serious injury to the patient.
Warning:
on the tubing.
Caution:
are secure; do not overtighten. This will help minimize leaks, disconnection and component damage.
Disposables must be compatible with the medicine delivered.
Patient or operator injury may result if package is opened or damaged, or if
Air embolization can cause death or serious injury to the patient. Do not connect
Drugs must not be administered to the epidural space unless the drugs are
For epidural administration of drugs use BodySet microbore with yellow stripes
Component damage may occur if not installed properly. Assure all connections
BodyGuard Operator Manual
25
Page 26
The BodySet Macrobore
4 - Operation
#
1 Luer Female Cap Keeps the set sterile
2 Luer Female
3. Slide Clamp Can be used to occlude the tube
4. Key Locates the tube in the correct flow direction
5. Administration Tubing PVC 3 x 4.1 mm.
6. Y Connector Used to add drugs by syringe
7.
8. Valve Cap Keeps the set sterile
BodyValve With Cap
Area Function
Connects the administration set to solution bag
Transfers the solution
Anti free-flow valve
BodyGuard Operator Manual
26
Page 27
BodySet Macrobore with Drip Chamber
bag
4 - Operation
#
1 Spike With Cap Connects the administration set to solution
2 Drip Chamber
3. Slide Clamp Can be used to occlude the tube
4. Key Locates the tube in the correct flow direction
5. Administration Tubing 3 x 4.1 mm
6. Y Connector Used to add drugs by syringe
7.
BodyValve With Cap
Area Function
Displays infused drops from the administration bag to the patient
Transfers the solution
Anti free-flow valve
* The set can be configured without the Y connector.
BodyGuard Operator Manual
27
Page 28
4 - Operation
The BodySet Microbore
#
1. Spike Cap Keeps the set sterile
2. BodySet Spike Connects the administration set to solution
3. Administration Tubing 1.5 x 3 mm
4. Filter 1.2 micron Filters the fluid path from air and particles
5. Male luer lock
6. Male luer lock Cap Keeps the set sterile
Area Function
Transfers the solution
Connects the administration set to the patient access device
7. Administration Tubing 3x 4.1 mm
8. Slide clamp Can be used to occlude the tube
9. Prime valve Anti free- flow valve with manual priming
10. Key Locates the tube in the correct flow direction
11. One way valve Prevents free flow and back flow
12. Anti Stretch locator Prevents incorrect tube stretching and locates
Transfers the solution (pumping segement)
button
the administration set correctly
BodyGuard Operator Manual
28
Page 29
Loading the Administration Set

NOTE:
Warning:
of the pump and the accuracy of the Infusion.
Follow the instructions supplied with the individual administration set.
Use of any administration set, other than the BodySet, may impair the operation
4 - Operation
1. Connect the administration set to the medication bag/ Lift the door latch (on the right side of the door) to open the BodyGuard pump door and remove
2. the administration set from the sterile packaging leaving the end caps on the line.
Connect the administration set to the medication bag.
3. Hold the section of tubing with the key (small plastic block) and make sure the flow direction is in line with the flow direction arrows inside the pump door.
4. Insert the IV tubing into the pump by placing the key into the keyway as shown by the arrow in diagram 1. Insert the tubing from left to right with the shorter end connected to the medication bag on the left , and avoid stretching or pulling the tubing. Check that the key located on the tube is located in its correct position in the tubing guide.

5. Locate the Anti-Stretch locator in the round space on the left side.
6. Verify that the distal flow valve is on the right hand side of the pump.
7. Close the pump door until the catch clicks.

NOTE:
fitting it do not force it in. Try to turn the line around to ensure you have correctly lined up with the direction of flow.
NOTE:
The IV key can only be fitted into the key space one way. If you are having trouble
Ensure that the tubing is inserted completely into the pumping channel.
Warning:
components. Maintain sterility of all disposable components and use only once single use
Use aseptic technique. Patient infection may result from the use of non-sterile
consumables marked
BodyGuard Operator Manual
.
29
Page 30
Priming the Line

Prime function is used:
1. Before starting a program
2. After air in line alarm
NOTE:
cannot be primed using gravity and must be primed either using the BodyGuard pump priming function or by pressing on the manual priming valve. In case manual priming is required, use an administration set with special manual priming valve.
Warning:

NOTE:
disabled.
The BodySet administration set contains a one-way check valve. Therefore, the lines
The default priming flow rate is 600 ml/hr. During priming, the air in line alarm is
Always disconnect IV tubing from patient before starting priming procedure.
4 - Operation
Prime – In Continuous, Intermittent, TPN , and 25 Steps Modes

1.
2. When the display screen shows the existing programmed
NOTE:
detailed instructions, refer to PCA Operation section.
Turn on the BodyGuard Pump, by pressing until the Self-Test screen appears. If air
sensor is off, press to confirm.
rate press once.
Warning:
3. Enter the Volume of fluid required to prime the set and press
Access to the priming procedure in PCA mode is different from all other programs. For
PRIME
Disconnect Patient Press OK to Start
Ensure the set is not connected to the patient!!
Prime Vol __ ml
to begin priming procedure. You may stop priming
Press OK to Continue
at any time by pressing .
4. The display screen will show a graph indicating the prime operation.
5. Once the priming is completed, the display will change back to the previous screen the pump was in before priming. If more priming is needed, repeat procedure.
BodyGuard Operator Manual
Make sure that all air is removed from the IV line before connecting to the patient.
PRIME 5 ml
0 5
30
Page 31
Prime – In PCA Mode
1. Turn on the BodyGuard Pump, by pressing until the Self-Test screen appears. If air
sensor is off, press to confirm.
4 - Operation
2. When the PCA menu displays, select “Prime” and press
to confirm.
3. Press OK to program priming procedure.
Warning:
4. Enter the volume to prime the set and press to begin priming procedure. You may stop priming at any time
by pressing .
5. The display screen will show a graph indicating the prime operation.
6. Once the priming is completed, the display will change back to the previous screen the pump was in before priming. If more priming is needed, repeat procedure.
Ensure the set is not connected to the patient!!
P C A
PRIME
Disconnect Patient
Prime Vol __ ml
Press OK to Continue
PRIME 5 ml
0 5
Prime
Program Infusion Clinician Bolus
Press OK to Start
BodyGuard Operator Manual
Make sure that all air is removed from the IV line before connecting to the patient.
31
Page 32
Prime Operation – After Air In Line Alarm
1.
Press to mute the alarm.
2. Disconnect IV tubing from patient.
4 - Operation
3. Press once.
Warning:
4. Enter the volume to prime the set and press to begin priming procedure. You may stop priming at any time
by pressing .
5. The display screen will show a graph indicating the prime operation.
6. When priming is completed make certain that all air is removed from the IV line.
7.
Connect IV line to the patient and press to resume program.
Ensure the set is not connected to the patient!!
PRIME
Disconnect Patient Press OK to Start
Prime Vol __ ml
Press OK to Continue
PRIME 4 ml
0 5
Manual Prime Operation
1. In any case (before operation or after air in line alarm) a prime operation is needed, the user can prime the set manually by using the manual priming valve, if assembled in the administration set in use at that moment.
2. Disconnect IV tubing from patient.
3. Hold the tubing in a vertical position to maintain gravitation pressure. Press the priming valve button continuously until all trapped air is primed. Real ease the priming valve button to continue the anti free flow protection while infusing.
BodyGuard Operator Manual
32
Page 33
4 - Operation
Operation steps
The following are the steps for starting the infusion procedure. For detailed instructions regarding each step, refer to the relevant sections in the Operation chapter.
Pump Mounted on IV Pole
Connect the Bodyset to the medication bag. Hang the bag on an IV pole.
1.
2. Click the pump into the Charger. Mount the Charger on an IV pole.
3. Connect the Charger via a standard power cord to a grounded AC outlet.
4. Open the door of the pump and load the administration set into the pump’s tubing guide.
5. Turn on the BodyGuard Pump, by pressing until the Self-Test screen appears.
Warning:
sensor. Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary.
6. The program and status will appear on the display followed by the last rate setting.
7. Prime the administration set.
Warning:
8.
Set the required program and continue as described for every program setting.
Warning:
Portable Connection
Connect the Bodyset/MicroSet to the medication bag.
1.
2. Open the door of the pump and load the administration set into the pump’s tubing guide.
3.
Turn on the BodyGuard Pump, by pressing until the Self-Test screen appears.
Warning:
sensor. Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary.
4. Prime the administration set.
If air sensor is OFF, please contact your technician in order to enable the air
Make sure patient is not connected to the pump during priming!!
Verify that infusion is proceeding normally before leaving the pump unattended.
If air sensor is OFF, please contact your technician in order to enable the air
Warning:
5. Set the required program and continue as described for every program setting.
Warning:
the carrying bag.
BodyGuard Operator Manual
Make sure patient is not connected to the pump during priming!!
Verify that infusion is proceeding normally before inserting the bag and pump into
33
Page 34
Programming
The BodyGuard infusion System features five different programming options:
Continuous Program
Rate: 0.1 to 100 ml/hr in 0.1 ml increments, and 100 to 1200 ml/hr in 1 ml increments
Total Volume to be Infused: 0.1 to 9999 ml
Program as Rate Over Volume or Volume Over Time
Secondary Infusion (Piggy) Rate: 0.1 to 100 ml/hr in 0.1 ml increments, and 100 to 1200 ml/hr
in 1 ml increments
Secondary Infusion (Piggy) Volume: 0.1 to 9999 ml
Intermittent Program
Used for setting a dose protocol at set intervals from 0.1 to 100 ml/hr in 0.1 ml increments, and 100 to 1200 ml/hr in 1 ml increments
Volume: 0.1 to 9999 ml
Interval Time: 0:01 minute to 20:00 hours
25 Steps Program
Used for setting specific protocol that contains up to 25 steps
Rate: 0.1 to 100 ml/hr in 0.1 ml increments, and 100 to 1200 ml/hr in 1 ml increments
Volume: 0.1 to 9999 ml for each step
Accumulated volume for all steps is up to ten liters
TPN Program
Used for Total Parenteral Nutrition
Volume: 1 to 9999 ml
The program has a taper up/taper down pattern. The parameters to be set are: Total Volume,
PCA Program
Time Up, Time Down and Total Time. The pump will calculate the ramping infusion rates.
Used for PCA applications. Combines basal rate and pre-programmed boluses
Rate: 0.0 to 100 ml/hr in 0.1 ml increments
Volume: 0.1 to 1000 ml
Bolus Volume: 0 to 100 ml
Bolus Rate: 0.1 to 100 ml/hr in 0.1 ml increments, and 100 to 1200 ml/hr in 1 ml increments
4 - Operation
Keep Vein Open Mode (KVO)
The BodyGuard Infusion System uses a Keep Vein Open (KVO) mode that can run at the end of every program, if desired. KVO also runs during a delay before an Intermittent program and during Interval Time in the Intermittent program. The KVO rate can be set from 0.1 to 5 ml/hr. If the programmed infusion rate is lower than the KVO rate, the KVO will run at the programmed infusion rate. KVO during End Program is limited to a volume of 5 ml.

NOTE:
contains overfill beyond the volume to be infused.
To use the automatic KVO at the end of a protocol, make sure that the IV bag
End Program
When a protocol has completed, an audible alarm will beep 4 times. To stop the alarm and enter a new
program, press . If the stop key is not pressed before the 4 beep have concluded, KVO will begin when it is programmed to do so.
BodyGuard Operator Manual
34
Page 35
4 - Operation
Code
Changing Current Program – Using Level 1 Code
1. Turn on the BodyGuard Pump, by pressing until the Self-Test screen appears.
If air sensor is OFF, please contact your technician in order to enable the air sensor if disabled. Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary.
2. When a programming screen or infusion summary is displayed, press twice.
3.
Scroll to "Change Set up" option. Press .
4.
Scroll to "More …" and press .
Enter Change Set up Level 1 code, using the numeric
5.
keyboard and press .
6.
Scroll down to "Select Program" option and press .
7.
Scroll to the required program and press .
8.
Press to restart the pump.
Volume Infused Review Set up
Change Set up
Buzzer Level
More
Change Set up:
Technician
Pressure Default
XXX
Select Program
Priming Rate
CONTINUOUS
TPN
History
Restart Pump
Pressure Default
: Select Program

NOTE:
operator.
The selected program will remain in the pump’s memory until changed by the
Changing Current Program – Short Way
1.
2.
3.

BodyGuard Operator Manual
Turn on the BodyGuard Pump by pressing until the Self-Test screen appears. If air
sensor is OFF, press to confirm.
Once the Self-Test is completed, immediately press and hold for two seconds. The Program selection mode will be displayed.
Scroll to the required program and press .
CONTINUOUS
TPN
NOTE:
If the current selected program is PCA, the Short Cut option is not available.
:Select Program
35
Page 36
Continuous Infusion
• Press
to go back to R
ate screen
.
Ra
te
25
ml/h
Volume
ml
ml/h
Press NO to Change
Rate Over Volume
1. Turn the pump ON by pressing and holding the until the Self-Test screen appears.
If air sensor is OFF, please contact your technician in order to enable the air sensor if disabled. Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary.
4-Operation – Continuous Infusion
2. Pump displays program name and status. The status is either "Program" or "Locked" (for instruction regarding locking and unlocking the pump – refer to Lock Mode section).
When unlocked, press . Wait for the rate screen to display.

NOTE
out mode, parameters can be viewed, but cannot be changed.



3. Rate Screen Do one of the following:
Press to confirm the existing parameters.
Enter the desired Rate and press .
Volume Screen
4.
Do one of the following:
:
When pump is in Program mode, all parameters can be adjusted. When pump is in lock
NOTE:
NOTE:
NOTE:
display the last set program.
If pump is not in Continuous Program, refer to changing program section for instructions.
Whenever you decide to set a new program, follow instructions at the bottom of screen.
The screen will be blank upon initial setup. When returning to a program, the screen will
C O N
T
C
O
N T
Continuous
Program
Continuous
LOCKED
Rate __
Press OK to Continue
__
or
5. Accumulation Screen
6.
BodyGuard Operator Manual
Press to confirm the existing parameters.
Enter the desired Volume and press .
Do one of the following:
Press to confirm the existing parameters and
start infusion.
Press to Start Infusion.
Press OK to Continue
C O
Volume 200 ml
N
Press OK to Start
T
C O
Start Infusion?
N
T
36
Page 37
4-Operation – Continuous Infusion
3.
Adjust the volume and press
three times to restart infusion
.
ml/h
7. During program operation, the pump displays the rate in which the program is operating.

NOTE:
the new value unless changes are confirmed by pressing .
NOTE:
(previous) screen once priming is completed.
If one of the values is changed during the programming mode, the pump will not recall
If prime procedure is activated during data setting, the pump will return to the original
Rate Titration
The pump allows rate titration during the Rate Over Volume Program.
To change the rate during the infusion:
1. Use numeric keypad to enter the new desired rate during
operation and press within five seconds. The pump will briefly alarm and the infusion will resume at the adjusted rate.
C
Rate
O
N
T
Rate 250
Adjust & Press OK
25
ml/h

2.
NOTE:
operation at the original rate and the display will return to the previous setting.
To stop the infusion press .
If the rate change is not confirmed within ten seconds, the pump will continue
To change the rate while in Stop mode:
1. Press any number on the keypad twice to return to the rate screen.
2.
Adjust the rate and press .

NOTE:
Pressing once clears the last entered digit.
Pressing again clears the next number.
Pressing when the value is empty displays the original value or the previous screen.
During programming you may backspace by using :
BodyGuard Operator Manual
37
Page 38
4-Operation – Continuous Infusion
Rate
125
ml/h
Total Time
02
:00
Bag Vol
:
250
ml
Code
Volume Over Time

NOTE:
Continuous Program and that the IV bag contains an additional 5 ml of volume to be infused. This ensures adequate volume for KVO during End Program.
To set Volume Over Time, enable the time function by following the steps below:
1. Press twice.
2.
Scroll to "Change Set up" option. Press .
3.
Scroll to "More …" and press .
Enter Change Set up Level 1 code, using the numeric
4.
Before operating the pump in Volume Over Time, make sure that you are in the
Volume Infused Review Set up
Change Set up
Buzzer Level
More
Change Set up:
keyboard and press .
5.
Scroll to "Time Option" option. Press .
6.
Press or , to turn Time Option ON.
7.
Press twice.
8. Volume Screen Do one of the following:
Press to confirm the existing parameters.
Enter the desired volume using the numeric keypad and
press .

9. Infusion Time Screen

10. Rate Screen
NOTE:
Enter the desired total time using the numeric keypad (for
example enter 0-2-0-0 for two hours) and press .
NOTE:
The pump calculates the rate based on the time and volume entered, and displays the rate on the screen. Do one of the following:
Press to confirm the existing parameters.
Press to go back to volume setting screen.
The screen will be blank upon initial setup.
First two digits represent hours and last two digits represent minutes.
Technician
Operation LED
XXX
Time Option
Delay Before Start
Press OK to Continue
Press OK to Continue
Press OK to Continue
C
Volume 250 ml
O
Time 2:00
N
T
Press OK to Start
h/ml 125 Rate
BodyGuard Operator Manual
38
Page 39
4-Operation – Continuous Infusion
Resume Press OK
T
11. During program operation, pump displays the rate at which the program is operating.


NOTE:
the new value unless changes are confirmed by pressing .
NOTE:
(previous) screen once priming is completed.
If one of the values is changed during the programming mode, the pump will not recall
If prime procedure is activated during data setting, the pump will return to the original

NOTE:
Stop mode is limited for two minutes. After two minutes, alarm will sound. Press to mute
or to continue.
When required to stop infusion, press . Stop will be displayed on the screen.
Resume or Repeat Infusion
Unlocked
When the pump is restarted in Continuous mode and is not locked, the pump will allow the user to confirm each of the parameters of the infusion. The Volume will be adjusted for the amount infused prior to turning off the pump. To reset the volume, use the numeric keypad to enter new volume.
Locked
1. When the pump is turned OFF in the middle of an infusion, this screen will appear when the pump is turned back ON.
C O N
Repeat Press NO
2. Do one of the following:
Resume: To resume this program exactly where it stopped, press . The accumulation
screen is displayed. The volume remaining will be infused.
Repeat: To restart the program at the beginning and or to set new parameters, press
and follow the instructions on the pump
BodyGuard Operator Manual
39
Page 40
4-Operation – Continuous Infusion
Y
Secondary Infusion (PIGGY) During Continuous Operation
To perform a secondary infusion during Continuous operation, complete the following steps.
1. Press once.
2.
Enter Secondary Infusion Rate and press .
3.
Enter Secondary Infusion Volume and press .
4. Pump will beep once and resume the Primary Infusion when the Secondary Infusion is completed.
P
Secondary Infusion
I
Piggy Rate 0 ml/h
G
Press OK key
P
Secondary Infusion
I
Piggy Vol. __ ml
G
Press OK to Start
Y


NOTE:
pump off. When the pump is turned back on, the pump will resume the Primary Infusion.
NOTE:
primary continuous infusion.
User can stop the Secondary Infusion at any time by pressing and turning the
User can titrate the Secondary Infusion with the same method that is used for a
BodyGuard Operator Manual
40
Page 41
4-Operation – Intermittent Infusion
Dose Rate
:
ml/h
Dose
Vol
:
ml
Bag Vol
:
ml
Intermittent Infusion
The Intermittent Infusion program is a special program which enables the design of a dose delivery protocol at set intervals. Each dose is given at a set time. The pump will Keep Vein Open between the doses. The infusion program continues until the total volume to be infused is delivered.
Rate Over Volume
1.
Turn the pump ON by pressing and holding the until the Self-Test screen appears.
If air sensor is OFF, please contact your technician in order to enable the air sensor if disabled. Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary.
2. Pump displays program name and status. The status is either "Program" or "Locked" (for instruction regarding locking and unlocking the pump – refer to lock mode section).
If pump is unocked, the Bag Volume screen will display after 2 seconds.




3. Bag Volume Screen
NOTE:
out mode, parameters can be viewed, but cannot be changed.
NOTE:
instructions.
NOTE:
NOTE:
Do one of the following:
Press to confirm the existing parameters.
Enter the Bag Volume and press .
When pump is in Program mode all parameters can be adjusted; W hen pump is in lock
If pump is not already in Intermittent Program, refer to Changing Program section for
Whenever you decide to set a new program, follow instructions at the bottom of screen.
When returning to a program, the screen will display the last set program.
I
N
T
R
N
T
R
Intermittent
Program
Press NO to Change
I
or
Intermittent
LOCKED
___
Press OK to Continue
4. Dose Volume Screen Do one of the following:
Press to confirm the existing parameters.
Enter the Dose Volume and press .
5. Dose Rate Screen Do one of the following:
Press to confirm the existing parameters
Enter the Dose Rate and press .
BodyGuard Operator Manual
___
Press OK to Continue
___
Press OK to Continue
41
Page 42
6.
mode is limited for two minutes. After two minutes an audible alarm is sounded. Press
Interval Time
:
__:__
KVO Rate
:
ml/h
Interval Time Screen
Do one of the following:
Press to confirm the existing parameters
Enter the time between the start of each dose and press
NOTE: Interval time is counted from starting time of

one bolus until starting time of the following bolus.
7. KVO Rate Screen Do one of the following:
.
4-Operation – Intermittent Infusion
Press OK to Continue
__
Press to confirm the existing parameters
Enter the desired KVO rate (from 0.1 to 5 ml/h) and press
.
8. Accumulation Screen Do one of the following:
Press to confirm the existing parameters and
start infusion.
Press to go back and adjust program parameters.

9.
10. During program operation, the pump displays the rate in


NOTE:
information regarding the Delay Option, refer to Delay Program section.
Press to Start Infusion.
which the program is operating.
NOTE:
the new value unless changes are confirmed by pressing .
NOTE
(previous) screen once priming is completed.
If pump is in Delay Mode, enter Start Time (24 Hour Clock) and Start Date. For further
If one of the values is changed during the programming mode, the pump will not recall
: If prime procedure is activated during data setting, the pump will return to the original

NOTE:
Whenever required to stop infusion, press . Stop displays on the screen. Stop
Press OK to Continue
I
Dose Rate 200 ml/h
N
Dose Vol: 200 ml/h
T
Interval 1:00
R
Press OK to Start
I
N
Start Infusion?
T
R
I
Rate
N
T
R
25
ml/h
again to mute the alarm.
BodyGuard Operator Manual
42
Page 43
2.
ml/h
Rate Titration
The pump allows rate titration during the Intermittent Program.
To change the rate during the infusion:
1. Use numeric keypad to enter the new desired rate during
operation and press within five seconds. The pump will briefly alarm and the infusion will resume at the adjusted rate.
4-Operation – Intermittent Infusion
Rate 250
Adjust & Press OK



2.
NOTE:
change.
NOTE:
NOTE:
at the original rate and the display will return to the previous setting.
To stop the infusion press .
If pump is LOCKED, the Level 1 technician code must be entered to confirm the rate
Rate titration is allowed only during dose delivery and not during KVO between doses.
If the rate change is not confirmed within ten seconds, the pump will continue operation
To change the rate while in Stop mode:
1. Press any number on the keypad twice to return to the rate screen.
3.
Adjust all other parameters and press to restart infusion.

NOTE:
Pressing once clears the last entered digit.
During programming, you may backspace by using :
Pressing again clears the next number.
Pressing when the value is empty displays the original value or the previous screen.
BodyGuard Operator Manual
43
Page 44
4-Operation – Intermittent Infusion
.
Interval Time
:
__:__
Dose
Time
:
__
:__
Dose
Vol
:
ml
Bag Vol
:
ml
Code
Volume Over Time
To program with Time and Volume instead of Rate and Volume, enable the Time Option function by following the steps below:
1. Press twice.
2.
Scroll to "Change Set up" option. Press .
3.
Scroll to "More …" and press .
Enter Change Set up Level 1 code, using the numeric
4.
Buzzer Level
More
Change Set up:
Volume Infused Review Set up
Change Set up
keyboard and press .
5.
Scroll to "Time Option" option. Press .
6.
Press or , to turn Time Option ON.
7.
Press twice.
8. Bag Volume Screen Do one of the following:
Press to confirm the existing parameters.
Enter the Bag Volume and press .
9. Dose Volume Screen Do one of the following:
Press to confirm the existing parameters.
Enter the Dose Volume and press .
10. Dose Time Screen Do one of the following:
Technician
Press OK to Continue
Press OK to Continue
XXX
Operation LED
Time Option
Delay Before Start
___
___
BodyGuard Operator Manual
Enter the Dose Time and press .
(To enter 30 minutes, press 0-0-3-0)
11. Interval Time Screen Do one of the following:
Press to confirm the existing parameters
Enter the time between the start of each dose and press
Press OK to Continue
Press OK to Continue
44
Page 45
12. KVO Rate Screen
R
U
KVO Rate
:
ml/h
Do one of the following:
4-Operation – Intermittent Infusion
__
Press to confirm the existing parameters
Enter the KVO Rate (from 0.1 to 5 ml/h) and press
.
13. Accumulation Screen Do one of the following:
Press to confirm the existing parameters and
start infusion.
Press to go back and adjust program parameters.

14.
15. During program operation, the pump displays the rate in
NOTE:
information regarding the Delay Option, refer to Delay Program section.
Press to Start Infusion.
which the program is operating.
If pump is in Delay Mode, enter Start Time (24 Hour Clock) and Start Date. For further
Press OK to Continue
Dose Rate 200 ml/h
I
Dose Vol: 200 ml/h
N
Interval 1:00
T
Press OK to Start
R
I
N
Start Infusion?
T
R
I
Rate
N
T
R
25
ml/h
Resume Infusion, Start New Bag, or Restart Infusion
When an Intermittent infusion is unlocked and stopped and the pump is re-started, the user has the following options:
(1) Resume: Resume the original infusion exactly where it left off (2) New Bag: Resume the original infusion with a new bag (3) Repeat: Repeat or re-program the infusion
1. Intermittent Restart Screen
Press or to scroll to the appropriate selection
and press .

2. Resume – New Bag – Restart Screen
NOTE:
available.
Display shows a summary of the current infusion to be started. User can scroll through the details by pressing
If pump is Locked, the user will only have the “Resume” and “New Bag” options
or . Press to start infusion.
N
T
R
E S
Resume
I
>=
New Bag Repeat Press OK key
Dose Rate 100 ml/h Dose Vol 100 ml Pause 5:00 Press OK to start
BodyGuard Operator Manual
45
Page 46
4-Operation – Intermittent Infusion
Code
Program Delay
The option to Delay Program Before Start can only be used in the Intermittent Mode.
Turning the Delay Option ON
1.
3.
4.
5.
Press twice from STOP or setting mode.
Warning: If air sensor is off, please contact your technician in order to enable the air sensor if disabled. Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary.
Scroll to "Change Set up" option and press .
Scroll to "More …" and press .
Enter Change Set up level 1 code, using the numeric
Volume Infused Review Set up
Change Set up
Buzzer Level
More
Change Set up:
XXX
6.
6.
7.
8.
keyboard and press .
Scroll to "Delay Before Start" option and press .
Turn delay option ON by pressing or .
Press and the screen displays Restart Pump
Press to continue.
Technician
Time Option
Delay Before Start
Language
Setting Delay Before Start
After entering the program data, the pump enables delay setting (when Delay Before Start is turned ON).


NOTE:
as follows 10:00, 11:00, 12:00, 13:00, 14:00, 15:00 etc.
NOTE:
The delay time is set according to a 24 hour clock. This means that hours are viewed
The pump automatically displays the current time and date. To start pump
immediately, press to confirm the existing time and date and start infusion.

BodyGuard Operator Manual
NOTE:
Settings Menu and select “Set Time & Date” to reset the clock in the pump.
If the time and/or date automatically displayed by the pump are not correct, go to the
46
Page 47
4-Operation – Intermittent Infusion
immediately, do not change the time and press
.
Start Date
Start Time
1. Start Time Screen To change the Start Time, type the desired start time using
the numeric keypad and press . To start pump

2. Start Date Screen
3. KVO During Delay Screen

NOTE:
To change the Start Date, type the desired start date using
the numeric keypad and press . To start pump on
the same day, do not change the date and press .
During delay time, pump will operate in KVO mode. Time left will count down until 00:00 and will then automatically start the progrmamed infusion. During the delay, this screen will be displayed showing when the infusion will start.
NOTE:
To start pump at 9:30 AM, enter 0-9-3-0. To start the pump at 9:30 PM, enter 2-1-3-0.
After delay time, pump will beep once and infusion will start automatically.
D E
L
Press OK key
A
D E
L
Press OK key
A
4/ 8/2005 12:05
D
Start Time 18:30
E
Start Date 04/08/05
L
Press OK to Start
A
__:__
__/__/__
Turning Pump Off In The Middle Of An Intermittent Infusion
When the pump is turned OFF during an intermittent infusion, the internal clock in the pump will continue to monitor the timing of the infusion. If the pump is stopped during a dose, the user will be able to resume the dose if the infusion is resumed before one third of the interval time has elapsed. For example, if the interval is 4 hours, then the user can restart the infusion within 1 hour and 20 minutes of the interval time to receive the remainder of the dose. In this case, the pump will still resume the following dose at the scheduled time. If the user restarts the infusion after one third of the interval time has elapsed, then the user will not be receive the remainder of the dose. The pump will resume in the middle of the interval and receive KVO until the next dose is due. The purpose of this “One Third” rule is to allow the user to turn their pump off during a dose without changing the scheduled dose times.
If the pump is turned off during a dose and not restarted until after the next scheduled does was scheduled, then the pump will start the next dose immediately followed by a complete interval. For example, if the patient restarts the infusion an hour past the scheduled dose time, then will receive a complete dose followed by a complete interval.
If the pump is turned off during the interval period, the patient can restart the pump anytime before the next scheduled dose without impacting the timing of the delivery of the next dose. If the pump is not restarted until after the next scheduled does was scheduled, then the pump will start the next dose immediately followed by a complete interval.
BodyGuard Operator Manual
47
Page 48
4-Operation – 25 Steps Protocol
Press NO to Change
25 Steps Protocol
This protocol permits programming of up to twenty-five different steps. Each step has its own rate and volume. Step accumulation is limited to a Volume of ten (10) liters. Protocol steps are delivered sequentially until all steps are completed.

1.
NOTE:
additional 5 ml of total volume to be infused. This ensures adequate volume for KVO during End Program. If Delay Before Start is set, you must include enough volume for the delay period, in addition to the 5 ml KVO volume.
Turn the pump on by pressing and holding the until the Self-Test screen appears.
Warning: If air sensor is OFF, please contact your technician in order to enable the air sensor if disabled. Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary.
Before operating the pump in 25 Steps Protocol, make sure that the IV bag contains an
2. Pump displays program name and status. The status is either "Lock out" or "program" (for instruction regarding locking and unlocking the pump, refer to locking section).
When the pump is unlocked, press to continue.




3. Rate Screen – Step 1
NOTE
out mode, parameters can be viewed, but cannot be changed.
NOTE:
instructions.
NOTE:
NOTE:
Do one of the following:
Press to confirm the existing parameters.
Enter the desired rate and press .
:
When pump is in Program mode, all parameters can be adjusted; When pump is in lock
If pump is not already in 25 Steps Protocol, refer to changing program section for
Whenever you decide to set a new program, follow instructions at the bottom of screen.
The screen will be blank upon initial setup.
S T E P
S T E P
25 - Steps
25 - Steps
Step 1 Rate
___ ml/h
Press OK for Volume
Program
LOCKED
4. Volume Screen – Step 1
5. Rate Screen – Step 2
BodyGuard Operator Manual
Do one of the following:
Press to confirm the existing parameters.
Enter the desired volume and press .
Pump displays the rate for step 2. Repeat directions as for step 1 and repeat until the required number of steps are entered.
Step 1 Volume
___ ml
Press OK to Continue
Step 2 Rate
___ ml/h
Press OK for Volume
48
Page 49
4-Operation – 25 Steps Protocol
automatically cleared.
6. After entering the last required step, press without
entering a rate.
This indicates that all data required for the protocol has been set. All programs following the last set program will be

7. Accumulation Screen

8.
9 During program operation, the display identifies which step

NOTE:
Do one of the following:
Press to confirm the existing parameters.
Press to go back to rate setting screen.
NOTE:
information regarding the Delay Option, refer to Delay Program section.
Press to Start Infusion.
is currently running, and the rate.
NOTE:
If pump is in Delay Mode, enter KVO rate.
If pump is in Delay Mode, enter Start Time (24 Hour Clock) and Start Date. For further
If one of the values is changed during the programming mode, the pump will not recall
Step 5 Rate
___ ml/h
Press OK for Volume
Steps Total 5
Volume Total 500 ml
Press OK to Start
S T
Start Infusion?
E P
S
Rate Step 1 of 5
T E P
25
ml/h


the new value unless changes are confirmed by pressing .
If prime procedure is activated during data setting, the pump will return to the original (previous) screen once priming is completed.
NOTE:
mode is limited for two minutes. After two minutes an audible alarm is sounded. Press
again to mute the alarm.
NOTE:
Pressing once clears the last entered digit.
Pressing again clears the next number.
Pressing when the value is empty displays the original value or the previous screen.
Whenever required to stop infusion, press . Stop displays on the screen. Stop
During programming you may backspace by using .
BodyGuard Operator Manual
49
Page 50
4-Operation - PCA infusion
Bag
Vol
:
ml
T1 T
2
T
3
Press NO to Change
TPN
TPN Infusion
Total Parenteral Nutrition
The pattern of the program is a trapezoid.
Rate TPN Trapezoid Program
Max Rate
On a TPN Program the operator may set the total volume which corresponds to the size of the bag to be infused, the total time, and the up and down time. Program operation will start from 2 ml/h and will increase to the maximal rate. During the flat session (T2) the rate will not change. During the last part of the program (T3), the program will start to decrease the infusion rate according to the set taper down time.
Up and down time can be set to zero. A TPN program with up and down time set to zero, is a continuous program.

1.
NOTE:
additional 5 ml of total volume to be infused. This ensures adequate volume for KVO during End Program. If Delay Before Start is set, you must include enough volume for the delay period in addition to the 5 ml KVO volume.
Turn the pump on by pressing and holding the until the Self-Test screen appears.
Warning: If air sensor is off, please contact your technician in order to enable the air sensor if disabled. Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary.
Before operating the pump in TPN program, make sure that the IV bag contains an
- This mode delivers an infusion while automatically tapering up and down.
T1 = Time UP T2= Time FLAT T3= Time DOWN
Time
2. Pump displays program name and status. The status is either "Lock out" or "Program" (for instruction regarding locking and unlocking the pump – refer to Lock Out section).
When pump unlocked, press to continue.



3. Bag Volume Screen
BodyGuard Operator Manual
NOTE:
out mode parameters can be viewed, but cannot be changed.
NOTE:
instructions.
NOTE:
Do one of the following:
Press to confirm the existing parameters.
Enter the desired volume and press
When pump is in Program mode, all parameters can be adjusted; When pump is in lock
If pump is not already in TPN Program, refer to changing program section for
Whenever you decide to set a new program, follow instructions at the bottom of screen.
.
T P N
T P
N
TPN
Program
or
TPN
LOCKED
___
Press OK to Continue
50
Page 51
4. Volume To Be Infused (TBI) Screen
KVO Rate
:
ml/h
Down Time
Up Time
VolumeTBI
ml
Total Time
TPN
Do one of the following:
4-Operation - PCA infusion
___
Press to confirm the existing parameters.
Enter the desired volume and press .
5. Total Time Screen The screen displays the last set time (total time for the
program to run). Press to confirm the existing
parameters or enter the desired time and press .

6. Up Time Screen
7. Down Time Screen
NOTE:
enter 0-2-0-0 for 2 hours.). Use a point to move from hours to minutes.
The screen displays the time for rate to taper up. Press
desired value (from 00:00 to 04:15 hours). Press .
The screen displays the time for rate to taper down. Press
desired value (from 00:00 to 04:15 hours). Press .
First two digits represent hours and last two digits represent minutes (for example,
to confirm the existing parameters or enter the
to confirm the existing parameters or enter the
Press OK to Continue
__:__
Press OK to Continue
__:__
Press OK to Continue
__:__
Press OK to Continue
8. KVO Rate Screen Do one of the following:
Press to confirm the existing parameters
Enter the desired Interval Time and press .
9. Accumulation Screen The pump calculates the set data and displays the accumulation screen with the maximal rate for the program (to be infused during flat period). Do one of the following:
Press to confirm parameters and start infusion or
enter delay time, if applicable. For further information regarding delay option, refer to delay program section.
Press to go back to volume setting screen.
10. Press to start infusion.
11. During program operation, pump displays the rate in which the program is operating and program direction (Up/Flat/Down).
Press OK to Continue
Volume TBI 2000 ml/h
T
P
Max Rate 55.1 ml/h
N
Press OK to Start
T
P
Start Infusion?
N
U
Rate
P
4.1
__
ml/h
BodyGuard Operator Manual
51
Page 52
4-Operation - PCA infusion
Resume Press OK
T
PN
Early Taper Down
If the program must be interrupted early and the planned taper down time must commence immediately, do as follows:
1.
Press continually (2 seconds) to stop the pump operation and display Taper Down screen.
2. Do one of the following:
Press to start the taper down process immediately with the existing parameters.
Enter time for taper down, using the numeric keys, and press to start the taper down
process. Taper down time should be lower than the ongoing taper down time.


NOTE:
the new value unless changes are confirmed by pressing .
NOTE:
(previous) screen once priming is completed.
If one of the values is changed during the programming mode, the pump will not recall
If prime procedure is activated during data setting, the pump will return to the original

NOTE:
mode is limited for two minutes. After two minutes, an audible alarm is sounded.
Whenever required to stop infusion, press . Stop displays on the screen. Stop
T
Taper Down 1:00
P N
Press OK key
Press again to mute the alarm.

NOTE:
Pressing once clears the last entered digit.
Pressing again clears the next number.
Pressing when the value is empty displays the original value or the previous screen.
During programming you may backspace by using :
Resume or Repeat Infusion
1. When the pump is turned OFF during infusion, this screen will appear when the pump is turned back ON.
2. Do one of the following:
Resume: To resume this program exactly where it stopped, press . The accumulation
screen is displayed. The volume remaining will be infused.
T P
N
Repeat Press NO
BodyGuard Operator Manual
Repeat: To restart the program at the beginning and or to set new parameters, press
and follow the instructions on the pump
52
Page 53
4-Operation - PCA infusion
PCA Infusion
Patient Control Analgesia is a program used for patients who require IV pain management or epidural analgesia. Pain management is undertaken by delivering a continuous basal rate, in combination with pre-programmed boluses. The pump will deliver only the boluses corresponding to the preset time intervals. All attempts to obtain a bolus are stored in the pump memory. The physician can read these statistics and adjust the settings as required.
Warning:
PCA Menu
Due to the sensitive nature of PCA infusions, the BodyGuard 323 has a specific menu to enhance the user experience while programming PCA infusions. The menu organizes the infusion information centrally while maintaining the same programming format as the other infusion modes on the pump.
1.
Turn the pump on by pressing the until the Self-Test screen appears.
When operating the pump on PCA program with a rate of 0.0 ml/h there is a hazard
of blood clot forming. Connect saline infusion in parallel to avoid this problem.
Warning:
Using the pump with air in line detector off may cause an embolism resulting death or paralysis. Switching the air detector off is not recommended and should be used only when absolutely necessary.
2. Pump displays program name and status. The status is either "PROGRAM" or "LOCKED" (for instruction regarding locking and unlocking the pump – refer to Unlocking PCA section).
Press . If the pump is unlocked, the PCA Menu will display. If the pump is locked, the PCA Restart Menu will display (See Restart Infusion section if pump is locked)



3. PCA Menu Screen
NOTE:
NOTE:
NOTE:
The user interface in PCA centers around the PCA menu. From the PCA menu, the user can select:
i. Prime – Prime Administration Set
ii. Program Infusion – Enter infusion details
iii. Clinician Bolus – Infuse extra bolus
iv. Patient History – View infusion history of patient
v. Review Set up – View pump settings
vi. Change Set up – Enables user to change pump
settings and switch to other infusion modes.
If air sensor is OFF, please contact your technician in order to enable the air sensor.
The PCA program automatically locks upon completion of programming.
If pump is not already in PCA Program refer to Changing Program section for instructions.
Whenever you decide to set a new program, follow instructions at the bottom of screen.
P C A
P C A
P
C
PCA
PROGRAM
or
PCA
LOCKED
Prime
Program Infusion Clinician Bolus
Patient History Review Set up Change Set up
Use or to scroll through options. Select the
BodyGuard Operator Manual
required option by pressing
.
53
Page 54
Priming
New Patient
?
View History

1.
NOTE:
program.
Warning:
Selecting the Prime option from the PCA menu, the screen will show the priming screen.
4-Operation - PCA infusion
Priming is disabled during PCA operation. Always prime the set before starting a
Always disconnect IV tubing from patient before starting priming procedure.
P
C
Prime
Program Infusion Clinician Bolus
2. Press OK to program priming procedure.
Warning:
3. Enter the volume to prime the set and press to begin priming procedure. You may stop priming at any time
by pressing .
4. The display screen will show a graph indicating the prime operation.
5. Once the priming is completed, the display will change back to the previous screen the pump was in before priming. If more priming is needed, repeat procedure.
Ensure the set is not connected to the patient!!
PRIME
Disconnect Patient Press OK to Start
Prime Vol __ ml
Press OK to Continue
PRIME 5 ml
0 5
Program Infusion
1. PCA Menu Screen
Select Program Infusion and press .
P C
Prime
Program Infusion
Clinician Bolus
Review Set up Change Set up
BodyGuard Operator Manual
2. New Patient Screen If programming a protocol for a new patient, select New
Patient by pressing . If the protocol is not for a new
patient, press Pressing will clear the volume that has been infused to the current patient.
P C
Press OK or NO
54
Page 55
3. Rate/Concentration Selection Screen
Rate
mg/h
Concentration
Bolus Vol
.
Concentration
Rate
ml/h
ml/hr
To program Rate in Milliliters per Hour (ml/hr), press
.
To program Concentration in Milligrams per Milliliter
(mg/ml), press once.
To program Concentration in Micrograms per Milliliter
(mµ/ml), press twice.
4-Operation - PCA infusion
Press OK to Continue NO for Concentration
4. Rate or Concentration Screen Do one of the following:
Press to confirm the existing parameters.
Enter the desired rate or concentration and press .

5. Concentration Rate Screen
6. Rate Titration Limit Screen
7. Bolus Screen
NOTE:
press .
(Only When Programming by Concentration)
Do one of the following:
Press to confirm the existing parameters.
Enter the desired rate or concentration and press .
Enter the maximum rate of titration that can be used in the
current protocol and press . Rate can be ml/h, mg/h or mcg/h depending on program selected.
Do one of the following (Bolus can be ml, mg, or mcg depending on the program selected):
Press to confirm the existing parameters.
To change from Concentration back to Rate, enter 0 (zero) for Concentration and
P C
Adjust & press OK
___
or
P C
___
mg/ml
or
P C
___ micrograms/ml
P C
Adjust & press OK
Max Titration Rate
P C
Rate ___ ml/h
Press OK to Continue
P C
Press OK to Continue
___
__
ml
8. Bolus Titration Limit Screen

BodyGuard Operator Manual
Enter the desired bolus dose and press .
Enter the maximum bolus dose that can be used in the
current protocol and press . Bolus dose can be ml/h, mg/h or mcg/h depending on program selected.
NOTE:
go to the “Change Set up” menu to activate “Titration Bolus”
If the “Maximum Bolus Dose” screen does not appear on the pump during programming,
Maximum Bolus Dose
P C
___ ml
Press OK key
55
Page 56
4-Operation - PCA infusion
The default setting is OFF, meaning there is no limit.
to scroll through the programming again and
Lock
:
Press OK
Bag Vol
.
ml
Hour Limit OFF
Lock Time
min
4
H Limit
ml
9. Lock Time Screen The lock time is the amount of time between patient boluses. Do one of the following:
Press to confirm the existing parameters.
Enter the desired lock time in minutes and press .
10. Hour Limit Screen #1 As an additional safety precaution, user can program a volume limit over a certain amount of hours. User can set this limit for any amount of time between 1 (One) and 24 (Twenty-Four) hours, or can continue without setting a limit.
Do one of the following:
Press to confirm the existing parameters.
Enter the desired hours for the limit and press .
11. Hour Limit Screen #2 If an hour limit is entered in previous step, the next step will propt user to enter the maximum amount of volume to be infused over that timeframe. In the screen to the right, 4 (four) was entered in the previous step for the hour limit.
Do one of the following (Limit can be ml, mg, or µg depending on the program selected):
P C
Press OK to Continue
P C
Press OK to Continue
P C
Press OK to Continue
__
___
Press to confirm the existing parameters.
Enter the desired hours for the limit and press .
12. Bag Volume Screen Do one of the following:
Press to confirm the existing parameters.
Enter the desired bolus volume and press .
13. Locking Screen The pump automatically locks before a PCA infusion program can be completed. Do one of the following:
Press to lock the PCA program and proceed.
Press
review or adjust the program.
14. Accumulation Screen The pump displays the accumulation screen with a summary the program entered. User can scroll through the entire
program by pressing or .
Do one of the following:
__
Press OK to Continue
P C
Review: Press NO
Bag Vol: 200 ml Volume Left 200 ml Infused 0 ml Press OK to Start
Press to confirm parameters and proceed..
Press to go back and adjust parameters.
BodyGuard Operator Manual
56
Page 57
4-Operation - PCA infusion
15. Press to start infusion.

NOTE:
right side of the screen and that the LED light is blinking green.
User can confirm that pump is infusing by checking that the drips are displayed on the

 

NOTE:
for 3 seconds.
NOTE:
NOTE:
However you may use the pump for the delivery of only one drug protocol and the maximum safe ceiling for this regime could be, for example, 30mlh. Users can fix a maximum rate that the pump can be set to in ‘Change Set Up’ using the Technician Access Code to ensure other users cannot accidentally program the pump to deliver above this safe ceiling. See later section for ‘Change Set Up’.
NOTE:
setting is 600ml/h designed to deliver any bolus as quickly as possible. Users can adjust this if they want the bolus to be delivered over a longer period in ‘Change Set Up’ using the Technician Access code.
To access the PCA menu during operation, press once and then hold
LED indicator can be turned off in the main settings menu to save battery power.
The BodyGuard is capable of continuous infusion (basal) rates of up to 100 ml/h.
The BodyGuard is capable of delivering the bolus dose at up to 1000ml/h. The default
P C
Start Infusion?
A
Interpreting the Display Screen During PCA Infusion
Infusion Mode
Rate 2.5
Patient
Controlled
Bolus Volume
Infusion Mode
The first line of the display indicates that the pump is in the PCA infusion mode.
Infusion Rate
The second line of the display shows the current infusion rate. Depending on whether the user programmed the current protocol in mls or mgs, this will display the rate in either ml/h, mg/h or mcg/h.
Bolus Volume
The third line of the display shows the bolus volume for the current protocol. As with infusion rate, this will be displayed in mls, mgs or mcgs depending on how the current program is set up.
Battery Status
In the right hand margin, you can see the battery and infusion status indicator. The battery symbol will be visible when the pump is not connected to the mains charger. User can check the charge level by monitoring the black fill inside the symbol (i.e. completely black symbol indicates a fully charged battery).
Infusion Status
In the right hand margin, infusion status indicator shows drips moving down the screen when the pump is in operation.
Bolus 5 ml
PCA
ml/h
Infusion Status
Programmed Infusion Rate
(ml/hr or mg/ml)
Battery Status
Indicator
Indicator
BodyGuard Operator Manual
57
Page 58
4-Operation - PCA infusion
A
Patient Activated Bolus
A PCA infusion can be programmed to allow the patient to request boluses of a specified volume within predetermined time intervals. The patient can request the bolus by one of the following two methods:
1. Using Bolus Cable Press the button on the end of the bolus cable. The bolus cable plugs into the side of the pump where a label reads “Bolus”
2. Using Keypad
Press the button on the keypad.


NOTE:
NOTE:
and will track the amount of boluses have been requested. The bolus history can be reviewed by scrolling the Info Menu. See “Using History Functions” section of the manual
User can stop bolus at any time by pressing
If patient requests more boluses than are allowed, the pump will display “Bolus Locked”
Clinician Activated Bolus
A Clinician Activated Bolus can be performed before starting the PCA infusion or during the PCA infusion.
To infuse a Clinician Activated Bolus
Selecting the Prime option from the PCA menu, the screen
1. will show the priming screen. Select Clinician bolus option.
2. Enter Level 2 Technician Code and press .
before
starting the PCA infusion, perform the following steps:
P
C
B O
L
U
Program Infusion
Clinician Bolus
Patient History
Bolus – Enter Code
XXX
Press OK key
3. Enter the volume of the Clinician Bolus and press
to begin priming procedure. You may stop the bolus at any
time by pressing .
The Volume for the Bolus will be in ml, mg or µg depending on the settings of the existing protocol entered for the PCA infusion. To change, go to Program Infusion in the PCA menu and follow the steps until the selection can be made for ml, mg or µg.
4. The display screen will show a graph indicating the Bolus is being infused. The graphical bar displays the amount of the bolus to be infused (clear) and the amount infused (dark).

BodyGuard Operator Manual
Once the bolus is completed, the display will change back to the PCA menu.
Clinician Bolus
P C
Volume ___ ml
Press OK to Start
Clinician Bolus
P C
Volume ___ mg
Press OK to Start
Clinician Bolus
P C
Volume ___ mcg
Press OK to Start
B
Volume 5 ml
O
L
0 4.1 5
U
or
or
58
Page 59
4-Operation - PCA infusion

User can stop bolus at any time by pressing
Warning:
stops the bolus manually by pressing . This is true even when the volume infused is less than the specified Hour Limit for Volume (e.g. 4 Hr Limit) at the beginning of a bolus and goes over the limit during the bolus.
To infuse a Clinician Activated Bolus
1.
During operation, press and then press to stop the current infusion and initiate the Clinician Bolus.
2. Enter Level 2 Technician Code and press .
3. Enter the volume of the Clinician Bolus and press to begin priming procedure. You may stop the bolus at any
time by pressing .
The Volume for the Bolus will be in ml, mg or µg depending on the settings of the existing protocol entered for the PCA infusion.
Boluses will be infused up to the full volume that is requested unless the user
during
a PCA infusion, perform the following steps:
B
Bolus – Enter Code
O
XXX
L
Press OK key
U
Clinician Bolus
P C
Volume ___ ml
Press OK to Start
Clinician Bolus
P C
Volume ___ mg
Press OK to Start
Clinician Bolus
P C
Volume ___ mcg
Press OK to Start
or
or
4. The display screen will show a graph indicating the Bolus is being infused. The graphical bar displays the amount of the bolus to be infused (clear) and the amount infused (dark).

Once the bolus is completed, the pump will resume the original PCA infusion.

User can stop bolus at any time by pressing
B
Volume 5 ml
O
L
0 4.1 5
U
BodyGuard Operator Manual
59
Page 60
4-Operation - PCA infusion
End of Infusion, Restart Infusion, or Bag Change
1. End Infusion Screen Once the infusion had reached an end (pre-set volume was infused), an alarm sounds and display will show:
2. KVO Display The alarm will stop after 4 beeps. The pump will show the pre-set KVO rat. The alarm will sound again every 4 minutes until the user changes the bag or stops the pump.
3. PCA Restart Menu This screen displays when: (1) The pump is turned OFF and back ON in PCA mode.
(2) Infusion is stopped by pressing and holding .
User can scroll through the following options by pressing
or .
Restart – Starts infusion exactly where it left off previously.
New Bag – Restarts infusion exactly where it left off previously and updates bag volume to be equal to the bag volume originally programmed.
Menu – Brings user back to the PCA Menu.When asked for a code, enter the Unlock (L2) Code.
End Infusion
Press STOP for menu
KVO 5 ml/h
Press STOP for menu
=> Resume
P
New Bag
C
Menu
A
Press OK key

NOTE:
that the drip chamber is half filled and there is no air in the administration set. If using an ambulatory IV, PCA or Epidural set without a drip chamber, priming the line may be needed. Disconnect the set from patient before priming.
If using an infusion set with drip chamber, change to a new infusion bag, but make sure
Return to PCA Menu
1. PCA Restart Menu
Scroll to “Menu” using and press .
2. Enter Level 1 Technician Code and press .
Resume
P
New Bag
C
=> Menu
A
P C A
Press OK key
Enter Code
XXX
Press OK key
BodyGuard Operator Manual
60
Page 61
4-Operation - PCA infusion
Titrating Basal Rate During The Infusion
To change Rate during the infusion, enter the new rate using the keypad and confirm the new rate by entering the Level 1 Code. The Rate can be increased up to the “Max Titration Rate” programmed into the specific infusion.
Titrating Bolus During The Infusion
To change the Bolus during the infusion, press the INFO button seven times until you reach a screen showing the currently programmed Bolus in ml or mg. Use the keypad to enter the new Bolus and confirm the new Bolus by entering the Level 1 Code. The Bolus can be increased up to the “Maximum Bolus
Dose” programmed into the specific infusion.
Changing Lock Out Time During The Infusion
1. Stop the infusion by pressing .
2. Then press and hold until the screen appears allowing you to access the PCA Menu.
3. Select “Menu” and enter the Level 2 code.
4. Once in the PCA Menu, select “Program Infusion” and follow the programming instructions to adjust the program. Be sure to tell the pump that there is NOT a New Patient and the Bag Volume should NOT be reset.
BodyGuard Operator Manual
61
Page 62
4-Operation
Using History Functions
The BodyGuard keeps a rolling 24 hour history of the current patient’s infusion delivery. This section addresses how this can be used. There is also a 500 event history log keeping track of the last 500 user actions in a date and time stamped record.
While the pump is delivering medication, pressing the button repeatedly will scroll through:
Volume Infused vs. Volume to be Infused
Battery Level
Boluses Attempted, Boluses Given & Lock Time Review for the duration of the current delivery
Program Review Screen - Allows user to check current program parameters without having to
stop the infusion
Date & Time
Actual Pressure
Press to stop infusion delivery and then press the button repeatedly to scroll through:
Volume and Boluses Given in the Last 24 hours
Bolus Attempts, Given and Volume Infused Hour-by-Hour for the Last 24 Hours (use the arrow
keys to navigate through the hours)
Chart of Bolus Delivery During the Last 24 Hours, Hour-by-Hour
Chart of Volume (Including Boluses) Delivered During the Last 24 Hours, Hour-by-Hour

NOTE:
Press to resume infusion when you return to the ‘STOP’ screen.
Events Log
To access the history of events, you need to access Change Set Up using the Level 1 Technician Access Code. The events history carries the last 500 user events (i.e. changes to programs, pump started/ stopped, etc) and the status of the pump (i.e. battery charge, pressure levels, etc). It is not patient specific (i.e. the 500 events are likely to span multiple patients recently treated with that particular pump).
To access the events history:
1. From the main menu scroll to “Change Set up” and press .
2. From the main menu scroll to “More …” and press .
3. Enter the Level 1 Technician Access Code and press .
4. Scroll down to “History” and press .
5. The screen will now show the most current user event, date, time, etc. For example:
Event Num: 450 10/15/2005 15:59 Stop State

BodyGuard Operator Manual
NOTE:
some time, the first event to appear when you enter the events history may be number 754. This means there have been 754 events in this pumps life and events 255-754 are stored in history. The pump deletes the oldest event in the log each time a new event occurs.
Each event is assigned a new number and the pump stores the last 500 in memory. After
Press INFO - Details
this
62
Page 63
4-Operation
Press and to scroll through events. When you find one of interest, press to display further data relevant to this event on the status of the program and pump at the time of this event. The user can scroll through the information shown here to check what the pump was set up for, whether it was operating on battery or mains, whether it was locked out and what the pressure settings and readings were at that time.
Rate 1ml/h Volume Left 99.4 Infused 0.1ml
PCA Program Battery Operation Normal – Level Battery Pressure Level Taken Pump Unlocked Vol. Cal. 540 Pressure Setting High Pressure DELTA 15 Pressure CAP 255 Pressure Baseline 88 Actual Pressure 88
BodyGuard Operator Manual
63
Page 64
4-Operation
Setting Pressure Default For Down Occlusion
1. Turn the pump on by pressing and holding the until the Self-Test screen appears. If air
sensor is OFF, press to confirm.
2. Pump displays program name and status.
Press . Wait for the Rate screen to display.
3. Press twice.
4.
Scroll to "Change set up" and press .
5.
Scroll to "Pressure Default" and press .
6.
Press or to adjust the level (high, normal, low).
7.
Press to store the parameter in memory.

NOTE:
The following pressure parameters can be set:
Low = 7 psi
Normal = 14 psi
High = 21 psi
The selected pressure level will remain in memory until it is changed.
The actual pressure can be viewed at any time by scrolling with the info button during
operation.
After a down stream occlusion, infusion will restart from the stop point.
Volume Infused Review Set up
Change Set up
Restart Pump
Pressure Default
Select Program
Pressure setting
Normal
Adjust ↓↑ & press OK
14 psi
BodyGuard Operator Manual
64
Page 65
4-Operation
Code
Lock Mode
The BodyGuard provides three different level of locking:
A Keypad Locking:
During operation, all keys are disabled excluding the STOP/NO and INFO key.
B Program Locking
This feature enables the operator to lock out the setting keys, so that entered parameters and set programs cannot be changed once they are set. This option is important, for example, with children as users, when there is a danger that the child might play with the pump and unknowingly change the program; or for home-care patients who repeat the same program every day. Locking the program prevents mistakes in setting.
C Maximal Rate Locking
This feature enables the operator to set a maximal rate for the pump to accept. The program rate then can be adjusted only up to a preset limit.
How To Lock
A – Keypad Locking
To Lock Keypad
LOCK Mode
Press and hold until the entire graph is black and a beep is heard. The beep indicates that the locking program is turned ON.

To Unlock Keypad
Press and hold until the entire graph is clear and a beep is heard. The beep indicates that the locking program is turned OFF.
NOTE:
mode.
The keys are locked in memory until they are unlocked or until entering Change setup
B – Program Locking
1. Press twice from STOP or setting mode.
2.
Scroll to "Change Set up" option. Press .
3.
Scroll to "More …" and press .
Enter Change Set up Level 2 code, using the numeric
4.
OFF ON
LOCK Mode
OFF ON
Volume Infused Review Set up
Change Set up
Buzzer Level
More
Change Set up:
keyboard and press .
BodyGuard Operator Manual
Technician
XXX
65
Page 66
4-Operation
Code
5.
Press or to change from OFF to ON and
press to approve.
 
NOTE:
NOTE:
The default setting is OFF.
Unlock using the same procedure.
C – Maximal Rate locking
1. Press twice from STOP or setting mode.
2.
Scroll to "Change Set up" option. Press .
3.
Scroll to "More …" and press .
Enter Change Set up Level 1 code, using the numeric
4.
Program Lock
OFF
Adjust ↑↓ & press OK
Volume Infused Review Set up
Change Set up
Buzzer Level
More
Change Set up:
keyboard and press .
5. Scroll to "Maximal Rate" option and press .
Use numeric keys to set desired rate and press .
Technician
Bolus Rate
XXX
Maximal Rate
KVO Rate
BodyGuard Operator Manual
66
Page 67
4-Operation
0 6:30 12
Info Mode
Access the info mode by pressing during operation, data setting or while in the Stop mode.
Pressing during program operation will display the following:
Number of
Button Presses
One All programs Infused Volume
Two All programs Battery Status
Three Continuous,
Infusion Mode Information Screen
Total Volume to Be Infused
Tubing Temperature Intermittent, TPN
Infused
500 ml 1000
Battery Level
Battery time 5:48
t + 25 C
Three PCA Boluses History for current
program
Four Continuous,
Intermittent,
Four TPN Program Status in a graphic
Four PCA Data of current protocol
Five All programs Time and Date
Six All programs Actual Pressure
Time remaining for the end
of the program
way
Boluses
P
Attempted: 3
C
Given: 1
A
Delay 60Min
Time Left
1:00
Time
Rate 50 ml/h
P
Bolus 5 ML Lock Time 60Min
A
4H Limit 100 ml
Time 12:55
Date 15.11.02
Actual Pressure
0 7 21
BodyGuard Operator Manual
67
Page 68
Pressing during data setting or in the stop mode, will display the following:
Number of
button presses
Information Screen
4-Operation
One Infused Volume
Total Volume to Be Infused
Two Option Selection screen

NOTE:
Infused
500 ml 1000
Pressing STOP/NO will clear Volume Infused.
Volume Infused
Review setup
> Change setup
To change parameters, select Change setup, press Start/OK anf follow instructions in the Change setup section.
To view current pump settings, select Review setup, press Start/OK and follow instruction in Review setup section.
If no selection is made within ten seconds after pressing , the screen will exit the information mode and display the previous screen.
To exit the information mode after making a selection in Review Set up, press once to return to the original screen.
To exit the information mode after making a selection in Change Set up, press .
BodyGuard Operator Manual
68
Page 69
4-Operation
Volumetri
Review Set up
The Review Set up feature is designed to allow you to view the programmable options and other important information about the pump. In Review Set up, settings can be viewed, but not changed.
1. Press twice.
Press to scroll to "Review Set up" option. Press
2.
Volume Infused
Review Set up
and the current settings are displayed.
3.
To view the setting for a particular option, select the desired
option and press .
Screen No.
1 Battery Level Current Battery Level (Empty >>> Full)
2 Program Lock Option Currently ON or OFF
3 Time, Date Time and Date
4 Tube Temperature Tube Temperature
Screen Display Information
Change Set up
Battery Level
Program Lock
Time , Date
5 Buzzer Level Current Buzzer Volume Setting
6 Pressure Level Current Pressure Level Setting
7 Pressure Default Current Pressure Default Setting
8 Select Program Current Selected Program
9 Priming Rate Current Priming Rate Setting
10 Maximal Bolus Volume Maximal Bolus Volume
11 Bolus Rate Current Bolus Rate Setting
12 Maximal Rate Current Maximal Rate Setting
13 KVO Rate Current KVO Rate Setting
14 Flow Control Flow Control
15 Air Bubble Size Current Air Bubble Size Setting
16 Store Last Program Option Currently ON or OFF
17 Operation LED Option Currently ON or OFF
18 Time Option Option Currently ON or OFF
19 Delay Before Start Option Currently ON or OFF
20 Language Current Pump Messages Language
21 Serial Number Pump Serial Number
22 Production Date Pump Production Date
23 Operation Time Hours Since Last Service Calibration
24 Software Version Pump Software Version
25 Volume Calibration
26 Pressure Delta Pressure Calibration – Delta Value
27 Pressure CAP Pressure Calibration – Cap Value
28 EXIT EXIT
BodyGuard Operator Manual
c Calibration Value
69
Page 70
4-Operation
Code
Code
Change Set up
The Change setup mode allows you to make changes to the adjustable parameters.
To Access Change Set up:
1. Press twice.
2.
Scroll to "Change Set up" option. Press .
To Adjust Pressure Level or Buzzer Level:
1. Press or to select an option and press
Volume Infused Review Set up
Change Set up
Change Set up:
Pressure Level
.
2. Make adjustment by pressing or and press .
To Adjust Lock Out Option:
1.
Scroll to "More …" and press .
Enter Level 2 Technician Code, using the numeric keyboard
2.
Buzzer Level
Buzzer Level
More
Change Set up:
and press .
3.
Press or to change from OFF to ON and
press to approve.
To Adjust Other Parameters:
1.
Scroll to "More …" and press .
Enter Level 1 Technician Code, using the numeric keyboard
2.
and press .
3. Press or to select an option and press
.
Adjust the parameter by pressing or , or by using the numeric keypad.
4.
Technician
Program Lock
OFF
Adjust ↑↓ & press OK
Buzzer Level
XXX
More
Change Set up:
Technician
Bolus Rate
XXX
Maximal Rate
KVO Rate
BodyGuard Operator Manual
70
Page 71
off. Turn off to save all
Volume over Time.
5.
Press to confirm selection and the screen displays “Restart Pump”.
6.
Press to continue.
The table below indicates the adjustable parameters:
Screen No. Screen Information Action
4-Operation
1 Pressure Level
2 Buzzer Level Adjust the buzzer level.
3 Technician code level two Lock Out Turn lock out ON/OFF
4 Technician code level one Restart pump Restart pump
5 Pressure Default
6 Select Program Select the infusion program
7 Priming Volume Set the priming volume from 0-60 ml
8 Max. Bolus Volume
9 Bolus Rate Set the bolus rate from 0-1200 ml/hr
10 Maximal Rate Set the max rate from 0.1-1200 ml/hr
11 KVO rate
12 Flow Control 0-20 drops per ml
13 Air bubble size
14 Store last program
Choose pressure level (High 21 psi, Normal 14 psi, Low 7 psi).
Select pressure default (High 21 psi, Normal 14 psi, Low 7 psi).
Set the Max. bolus volume from 0-100
Set the KVO rate from 0.1-5.0 ml/hr
Set detectable size from OFF to 1.0 ml
Turn on to clear all program parameters upon system shut­program parameters upon system shut­off.
15 Operation LED Turn operation LED ON/OFF
Turn time option ON/OFF. While time
16 Time option
17 Delay before start Turn option ON/OFF
18 Language
19 Time, Date Set actual time and date
20 History Indicates injection history
option is ON, Continuous, Intermittent and 25 Steps program will be set to
Enable changing the language of pump messages (only for models in which the software includes more than one language).

NOTE:
used in last program can be confirmed. The memory retention capability is available for the life of the internal battery.
While the pump is in programmable mode, all parameters can be set, and parameters
BodyGuard Operator Manual
71
Page 72
5-Alarm conditions and troubleshooting
5 – Alarm Conditions and Troubleshooting
Alarm Condition
When the infusion pump detects a problem, four things occur:
The infusion stops
An audible alarm is activated
A message appears on the display screen indicating the cause of the alarm, and
The LED indicator will change from green to red

NOTE:
Pressing
during an alarm mutes the alarm for two minutes.
BodyGuard Operator Manual
72
Page 73
5-Alarm conditions and troubleshooting
Place the pump in charger and
Place the pump in charger and
connect to mains. Wait 2 minutes
Connect To Charger
Connect To Charger
Troubleshooting
Description Result Possible Cause Required Action
Air in Line
Air in Line
Press Stop to mute
Down Occlusion Infusion stops
Down Occlusion
Press Stop to mute
Infusion stops and an alarm is activated
and an alarm is activated
Air is present in administration Set.
The roller or clip on the administration set is closed upstream of the pump.
The line was not loaded correctly.
The roller or clip on the administration set is closed downstream of the pump.
Administration set is kinked.
Cannula is blocked.
Administration set loaded incorrectly.
NOTE: Operator may change pressure setting.
Disconnect line from patient, press STOP/NO. Remove the air, as described on priming section.
Open clamp/clip.
Re-load the IV line.
Open clamp/clip.
Straighten the set.
Change the Cannula.
Re-load the IV line.
Refer to “Setting Down Occlusion” section.
Pump Unattended An alarm is
Pump Unattended
Press Start key
Door Open Infusion stops
Door Open
Close Door
Low Battery Pump will only
Low Battery
End Battery Pump operation
End Battery
activated
and an alarm is activated
run for another 30 minutes if it is not connected to mains.
stops. The pump cannot be used before being connected to mains.
Two minutes has elapsed without a button press during programming.
The door of the pump was not closed prior to operation
The door has accidentally opened during operation.
30 Minutes of battery life are remaining.
Battery is depleted.
Press START/OK to resume.
Close the door of the pump.
Send the pump for service.
connect to mains.
before operating.
BodyGuard Operator Manual
73
Page 74
5-Alarm conditions and troubleshooting
to restart a new
the set into the pump as per
Change the set to a BodySet/
Microset and load it according to
ot be
, unlock the pump
he set parameters are
Description Result Possible Cause Required Action
Fatal Error Infusion stops Fatal internal error has occurred Restart the pump. If the alarm does
Error xx(number)
Send for service
not stop pack the pump properly, in its original packaging, and send it for service.
End Program Program ends,
End Program
Press stop to mute
Missing Key Pump will not
Missing Key
Set key & Close Door
Lock Out Setting cann
pump will turn to KVO mode
start
changed
Locked
Lock mode Setting keys do
not function
Lock Mode
Current infusion program has been completed.
Administration set loaded incorrectly (the key was not located in its place correctly).
A set different than the Bodyset/ Microset was loaded in the pump.
Lock Out mode is turned on. No action required. If changing
Lock Out mode is turned on. No action required. If changing
Press STOP/NO program or turn the pump OFF.
Reload the instructions.
instructions.
program is required and refer to lock section for further instructions.
parameters is required, unlock the keys and refer to lock section for further instructions.
Wrong TPN data
Wrong TPN data
Check Data
BodyGuard Operator Manual
Program will not start operating
T impossible to perform. Volume or time parameters are incorrect
Check data and change accordingly.
74
Page 75
6-Specifications
6 – Specifications
Pumping Mechanism: Piston Pump
Flow Rate: PCA: 0 to 100 ml/hr in 0.1ml increments
All other programs: 0.1 to 100 ml/hr in 0.1ml, 100 to 1200 ml/hr in 1 ml increments.
Priming Rate: 600 ml/hr
Total Infused Volume: PCA: 0.1 to 1,000 ml.
Continuous, TPN, Intermittent: 0.1 to 9999 ml 25 Steps: 0.1 to 9999 ml for each step Accumulated volume for all steps: up to 10 liters
Total Time Setting: 100 hours
Accuracy: ±3%
KVO rate: 0.1 to 5 ml/hr
Air Sensor: Ultrasonic, adjustable air bubble size
Maximum Pressure: 21 psi Adjustable (High = 21, Normal = 14, Low = 7)
Power Supply: 110-240 VAC, 50/60 Hz. 0.3A max.
Battery: Rechargeable, Li-Ion 7.2V, 1800mAh (Typical)
Battery Operation at 125 ml/hr:
Battery Charging:
Alarms: When a problem is detected, the BodyGuard displays the following alarms:

NOTE:
20 hours (Rechargeable)
Automatic when clicked into the Charger that is connected to an AC power source. Six hours needed to charge a fully-depleted battery.
Air in line
Down Occlusion
Pump Unattended
End Program
Low Battery
End Battery
Door Open
Fatal Error
Lock Mode
Lock Out Mode
Missing Key
When this symbol is seen, consult accompanying documents.
BodyGuard Operator Manual
75
Page 76
6-Specifications
BodyGuard Dimensions:
Classification Type CF Equipment (degree of protection against electrical shock)
Housing: ABS (fire retardant)
Weight
Electrical Safety
Standards
EMC
Environmental Specifications
112 x 89 x 32mm. (L x W x H).
280 grams without battery
390 grams with battery
Complies with: EN 60601-1 (Medical Electrical Equipment Safety), IEC 60601-2-24 (Infusion pumps and controllers), IEC 60601-1-4
(Programmable Electrical Medical System)
Manufactured in accordance to ISO 9001:2000 and ISO 13485:2003 CE marked (In accordance with the Medical Devices Directive 93/42/EEC)
The BodyGuard Infusion System is designed to be in compliance with EN 60601-1 (safety) and IEC 601-1-2 (EMC).
Non Operating Conditions (Transportation and Storage):
Temperature: -25oC to 50 oC (-13 oF to + 122 oF)
Humidity: 5 % to 100% R.H., non-condensing
Air pressure: 48kPa to 110kPa
Operating Conditions:
(The system may not meet all performance specifications if operated outside of the following conditions)
Temperature: +18oC to +45 oC (+59 oF to + 113 oF)
Humidity: 20 % to 90% R.H. at +40 oC, non-condensing
Air pressure: 70kPa to 110kPa
IV Administration set
BodyGuard Operator Manual
Dedicated IV administration sets with anti-siphon valve.
76
Page 77
7 – Service and Maintenance
7-Service and Maintenance
Cleaning
Before connecting the pump to a patient, and periodically during use, clean the unit using a lint-free cloth lightly dampened with warm water and a mild detergent or disinfectant.
Warning:
Warning:
Caution:
similar solvents. These chemicals can cause damage to plastic components and paint. Use a lint-free cloth dampened with warm water and a mild detergent or disinfectant.
Caution:
Always turn the pump off and remove the battery before cleaning.
Always unplug the charger from AC power before cleaning.
Do not clean the pump or charger with chemicals such as Xylene, Acetone or
Do not soak or immerse any part of the pump or charger in water.
Storage
If the pump is to be stored for an extended period, it should be cleaned and the battery fully charged. Store in a clean, dry atmosphere at room temperature and, if available, employ the original packaging for protection.
Perform functional tests and ensure that battery is fully charged once every three months.
BodyGuard Operator Manual
77
Page 78
7-Service and Maintenance
Battery Operation
The BodyGuard pump can operate on battery power which enables operation when the patient is being moved or during electrical power failure. When the pump operates on battery power, the AC icon is off. At full charge, the standard battery provides 15-17 hours of operation at an infusion rate of 125 ml/hr.




Warning:
for back up.
Note:
into an AC wall outlet (if possible) to charge battery.
Note:
wait 2 minutes after the pump has been connected to an AC power supply before operating.
Caution:
the battery.
Note:
unit. This preserves the battery power supply for emergency use or for situations where the AC power is not available.
Note:
Do not operate the pump on AC power if the battery is not loaded in the pump
When the pump is not in use, click the pump into the Charger and plug the system
After the “End Battery“ signal has been activated or following long periods of storage,
Leaving the battery in an uncharged state for a long period of time may damage
Whenever possible, use the pump connected to an AC power supply via the charging
Replace the battery once every two years.
BodyGuard Operator Manual
78
Page 79
7-Service and Maintenance
LIMITED WARRANTY
The BodyGuard Infusion Pump has been carefully manufactured from the highest quality components.
Caesarea Medical Electronics Ltd. (CME) guarantees the pump against defects in material and workmanship for twelve (12) months from date of purchase by the original purchaser.
CME’s obligation, or that of its designated representative under this Limited Warranty, shall be limited, at CME’s option, or that of its designated representative, to repairing or replacing pumps, which upon examination, are found to be defective in material or workmanship. The repair or replacement of any product under this Limited Warranty shall not extend the above­mentioned Warranty period.
All repairs under this Limited Warranty should be undertaken only by qualified, trained service personnel. In the event that a pump is found to be defective during the warranty period, the purchaser shall notify CME or its designated representative within thirty (30) days after such defect is discovered.
The defective pump should be sent immediately to CME or its designated representative for inspection, repair or replacement. Shipping costs are the purchaser’s responsibility.
Material returned to CME or its designated representative should be properly packaged using CME shipping cartons and inserts. Inadequate packaging may result in severe pump damage.
This Limited Warranty shall not apply to defects or damage caused, wholly or in part, by negligence, spilt fluids, dropping of the pump, misuse, abuse, improper installation or alteration by anyone other than qualified, trained personnel; or to damage resulting from inadequate packaging in shipping the pump to CME or its designated representative.
If, after inspection, CME or its designated representative is unable to identify a problem, CME or its designated representative reserves the right to invoice purchaser for such inspection.
This Limited Warranty is the sole and entire warranty pertaining to CME’s products and is in lieu of and excludes all other warranties of any nature whatsoever, whether stated, or implied or arising by operation of law, trade, usage or course of dealing, including but not limited to, warranties of merchantability and warranties of fitness for a particular purpose. Purchaser expressly agrees that the remedies granted to it under this limited warranty are purchaser’s sole and exclusive remedies with respect to any claim of purchaser arising under this Limited Warranty.
Managing Director
BodyGuard Operator Manual
79
Page 80
INDEX
A
Accuracy 11,12,17,23,29,75. Administration Sets 2,5,16,17,21-30, 33,35,36,41,46,48,50,53,60,73,74,76. Air in Line 2,16,22,25,30,32,33,35, 36,41,46,48,50,53,75. Air Sensor 2,8,16,21,22-25,30,31,33, 35,36,41,46,48,50,53,64. Alarm 6,7,8,10,17,21-23,30,32,34,37, 39,42,43,49,52,60,72-75.
B
Battery 5,17-20,24,57,62,63,67, 69, 71,73,75-78 Bolus 5,7,17,19,31,34,53-59,61,62, 66,67, 69-71
C
Cautions 15,16,18 Changing the Program 35 Changing fluid container 45, 60 Charger 9,16-21,24,33,57,72, 74,76,77 Charging the Pump 20
D
Delay Program 42,45,46 ,49,51 Display screen 6,7,21,30,32,54, 57-59,72 Down Occlusion 22,23,64,73,75
E
Early Taper Down 52 End Program 34,38,48,50,75,76
F
Fatal Error 75,76 Flow Direction 8 ,28,29 Flow Rate 11,23,25,30,75 Free Flow Protection 5,10,25
H
History 53,58,62,67,71
I
Info 7,16,42,45,49,51,53,58,61, 63,64,65,67,68,69,71 Infused Volume 67, 68, 75 Intended Use 16 Intermittent Program 34,41,43 IV pole 5,17,33
K
Keep Vein Open Mode 34 Keypad Locking 8,29 Keyway 8,29 KVO rate 34, 42, 45,49,51,70,72,76
BodyGuard Operator Manual
L
Lock Mode 36,41,65,73,74 Lock Out 36,41,48,50,61,65,70,71,74,75
M
Maximal Rate Locking 65,66 Missing Key 74,75
N Numeric Keys 7,22-24,35,37-39,43,44,46,47, 52,65,66,70
0
Operating Conditions 76 Operation LED 7,38,44,69,71
P
PCA Program 7,17,34,53,56,63 Power Supply 75,78 Pressure Level 62-64,69-71 Priming 6,7,21-24,30-33,35,37,39,42,49,52 54,58-60,69,71,73,75,76
R
Rate Over Volume 34,36,37,41 Rate Titration 37,43,55
S
Safe Container 58 Select Program 35,54,64,70,74 Service 1,16,17,70,75,78,80 Set up 19,22-24,35,38,44,46,53,55,57,62-66, 68,70,71 Start new Bag 45 STOP/NO 6,65,68,74,75
T
Time Delay 46 Time option 38,44,46,70,72 Total Infused Volume 76 TPN Program 34,50
U
Up Arrow 6
V
Valve 5,10,25, 28, 29, 30,76 Volume Over Time34,38,44,72
W
Warnings 15,16
80
Loading...