Caesarea Medical Electronics Ltd.
European address:
Staufenburgstr. 23 Lichtenstein P.O.B. 1248
Lichtenstein 72805 Germany
International Toll Free: +800-323-575-00
www.cme-infusion.com
Contact e-mail: Sales@cme-infusion.com
EDITION July 2009
0344
Ref.: 100-090X
NOTE:
all users to read this manual carefully, become familiar with the procedures and system
functions and follow all recommendations.
To assure proper use of the BodyGuard, Caesarea Medical Electronics Ltd. urges
Using the pump with air in line detector off may cause an
embolism resulting death or paralysis. Switching the air detector
off is not recommended and should be used only when absolutely
necessary. Please contact your technician in order to enable the
air sensor if disabled with no need.
U.S.A. federal law restricts this device to sale by or on the order of a physician.
The BodyGuard System provides the following features:
Small, light and compact pump.
Multiple Programs:
Infusion Rate: 0.1 to 1200 ml/hr
Volume: 0.1 to 99.9 ml in 0.1ml increments or
1 to 9999 ml in 1ml increments
Secondary Infusion (Piggy) programmable with
the same infusion rate and volume capabilities
as the primary continuous infusion.
A set dose delivered in set intervals. Between
Intervals the pump will keep vain open (KVO
Mode).
Total Parenteral Nutrition Rate Taper program
with programmable up and down times.
Patient Control Analgesia: Continuous delivery
plus /or programmed boluses.
Bolus Volume: 0-25.5 ml
Clinician Bolus Volume: 0.1-100ml
Bolus Rate: 0.1-1200 ml/h
Highly accurate fluid delivery
Operated with a custom safe administration set
Rechargeable internal lithium-ion battery or alkaline 2X9V batteries
Battery charged in a pole mount charging cradle or with an A/C adaptor
Free flow protection valve standard on every CME administration set
Quiet operation
Used for ambulatory care in a carrying pouch or
mounted on an IV pole
AEA – Suitable to be used in ambulances.
Operator can design a specific protocol for
drug delivery which contains up to 25 steps.
Infusion Rate: 0.1 to 1200 ml/hr
Volume: up to 9999ml, for each step
Accumulated Volume for all steps – up to 10
liters.
BodyGuard Operator Manual
5
1-Introduction
1. Display Screen
2. Up Arrow
3. START/ OK
4. S
TOP/NO
3
4
8
2
10
9
Front of
BodyGuard
Infusion Pump
1
5
6
7
• Displays pump and infusion status
• Displays programming choices and instructions
• Scrolls up through options
• Starts infusion
• Confirms selection and setting
• Stops infusion
• Silences an alarm condition
• Pause priming
• Zeroes the displayed value during programming
• Erases the last digit during programming
• Returns to previous screen
BodyGuard Operator Manual
6
8. INFO
1-Introduction
5. PRIME/BOLUS
•Pressing Prime/Bolus key during data setting: Enables Priming procedure. A
graph appears on the display screen showing the priming volume with the
current value increasing until reaching the set volume.
•Pressing Prime/Bolus key during Continuous program operation: Enables
Secondary (Piggy) Infusion.
•Pressing Prime/Bolus key during PCA program operation: Enables bolus
procedure.
6. Power ON/OFF
•Turn the system ON by pressing and holding the button until the Self-Test
screen appears.
•Turn the system OFF by pressing and holding the button until the graph is
black and a beep is generated.
7. Operation LED
• Green Indicator
• Lights during system Self-Test
• An intermittent green light indicates infusion delivery
• Red Indicator
• Indicates an alarm state with a continuous red light
• Lights when the pump is in a stand-by mode during programming
• Supplies information about the pump and its programs (see Info Mode
chapter).
•Pressing continuously locks and unlocks the keypad to prevent accidental or
deliberate change to pump operation (If needed).
9. Down Arrow
• Scrolls down through options
10. Numeric Keys
•Enters numeric parameters during programming
BodyGuard Operator Manual
7
1-Introduction
3
5
BodyGuard Pump with Door Open
7
8
9
# Area Function
1.
2.
3. Keyway Guides the IV line
Latch Holder
Pressure Sensor Detects downstream tubing restriction and occlusion. Alarm
Holds the door closed when latch is in vertical position.
level can be adjusted to suit patient needs.
1
6
2
4
4.
5.
6. Flow Direction Shows direction of pump operation
7. Anti-Stretch locator (Optional) Locates the tube in the correct length from the key
8. Air sensor – Front Ultrasonic air detector, which is mounted on the front
9.
Pressing Plate
Pump Door Covers the pressing plate
Air sensor - Door
Connected to the door by two springs
housing.
Mounted on the door
BodyGuard Operator Manual
8
1-Introduction
The Charger
# Area Function
1.
Charger Display Displays major pump messages and operation values.
1
2
2.
Release Handle Push to release the pump from the Charger.
BodyGuard Operator Manual
9
1-Introduction
System Safety Checks
The following details outline the safety checks designed into the BodyGuard infusion system, to
minimize the possibility of under or over infusions.
Free Flow Protection
The system’s customized IV lines are equipped with a check valve that prevents free flow towards the
patient when the IV line is not attached to the pump. When the pump is attached to the IV line and
delivering fluid, the pressure delivered by the pump opens the valve. The valve is also one way valve
preventing fluids flow from the patient to the administration line.
Air-in-Line Accumulation
To improve the detection of air in the IV line, the BodyGuard 323 Infusion system utilizes an air-in-line
accumulation system in addition to the standard single bubble detection. This feature monitors the
volume of air that passes through the IV line by accumulating the volume of individual bubbles over any
15 minutes period. This limit is not configurable. Although an individual bubble may not exceed the preprogrammed threshold, if the cumulative volume of smaller bubbles exceeds 1ml over a period of 15
minutes an air-in-line alarm is initiated. This accumulation feature is particularly useful with infusions for
patients who are highly sensitive to air (i.e. infants, neonates, children) or when infusing products that
create significant volumes of small air bubbles.
Anti-Bolus Function
The anti-bolus function is designed to reduce the bolus that may occur upon the release of an occlusion
following a downstream occlusion alarm. Upon the detection of a downstream occlusion, the alarm is
activated and the pump returns the IV line pressure to neutral within 15 seconds. Neutral line pressure is
achieved by the reverse operation of the pumping mechanism, and measurement of the IV line pressure
through the in-line pressure detection system.
BodyGuard Operator Manual
10
1-Introduction
Pump Accuracy
The following graphs and curves were derived from testing described in IEC60601-2-24. Testing was
performed under normal conditions at room temperature (720 F). Any deviations from normal conditions
and room temperature may cause changes in the accuracy of the pump.
Start-up Curves
The Start-up curves represent continuous flow versus operating time for two hours from the start of the
infusion. They exhibit the delay in onset of delivery due to mechanical compliance and provide a visual
representation of uniformity. Trumpet curves are derived from the second hour of this data. Tests
performed according to IEC 60601-2-24 standard.
Trumpet and Flow Rate
With the BodyGuard 323, as with all infusion systems, the action of the pumping mechanism and
variations cause short term fluctuations in rate accuracy.
The following curves show typical performance of the system in two ways:
1. The accuracy of fluid delivery over various time periods is measured (trumpet curves).
2. The delay in onset of fluid flow when infusion commences (start up curves).
Trumpet curves are named for their characteristic shape. They display discrete data averaged over
particular time periods or “Observation windows”, not continuous data versus operating time. Over long
observation windows, short-term fluctuation has little effect on accuracy as represented by the flat part of
the curve. As the observation window is reduced, short-term fluctuations have greater effects as
represented by the “mouth” of the trumpet.
BodyGuard Operator Manual
11
1-Introduction
Knowledge of system accuracy over various observation windows may be of interest when certain drugs
are being administered. Short-term fluctuations in rate accuracy may have clinical impact depending on
the shelf life of the drug being infused and the degree of inter-vascular integration. The clinical effect
cannot be determined from the trumpet curves alone.
BodyGuard Operator Manual
12
1-Introduction
BodyGuard Operator Manual
13
1-Introduction
BodyGuard Operator Manual
14
2-Symbols, Warnings and Cautions
2 – Symbols, Warnings and Cautions
System Symbols
The following symbols are used on the BodyGuard Infusion System and components. Labels on the
system or statements in this manual proceeded by any of the following words and/or symbols are of
special significance and are intended to help you to operate the pump in a safe and successful manner.
Attention, consult accompanying Instructions
CE mark indicates conformance to Medical Device Directive 93/42/EEC
0344
Do not dispose of in municipal waste. Symbol indicates separate collection for
electrical and electronic equipment. (WEEE Directive 2002/96/EEC). NOTE: Does
not apply to the battery.
Do not dispose of battery in municipal waste. Symbol indicates separate collection
for battery is required.
LOT
STERILE EO
The use of single-use disposable components on more than one patient is a
biological hazard. Do not reuse single-use disposable components.
Type CF applied part
Date of Manufacture
Serial Number
Expiry Date of disposable
Lot Number
Sterilized with Ethylene Oxide
BodyGuard Operator Manual
15
Terms Used In Manual
Warning:
circumstances that could result in injury or death to the patient or operator.
Read and understand this manual and all warnings completely before operating the
BodyGuard Infusion System.
Caution:
that could result in damage to the device.
Read and understand this manual and all cautions completely before operating the BodyGuard
Infusion System.
NOTE:
that will help you operating the BodyGuard Infusion System.
Indicates that the information is a warning. Warnings advise you of
Indicates that the information is a caution. Cautions advise you of circumstances
Indicates that the information that follows is additional important information or a tip
2-Symbols, Warnings and Cautions
Intended Use
The BodyGuard infusion Pump system is designed for infusion of medications or fluids requiring
continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable
routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural,
enteral, in close proximity to nerves, and into an intraoperative site (soft tissue/body cavity/surgical wound
site). The system is intended for patients who require maintenance medications, analgesics, PCA
therapy, parenteral and enteral nutrition fluids, chemotherapeutic agents and general fluids therapy in
hospital and home care environments.
Warnings
To avoid possible personal injury or loss of life, observe the following:
Using the pump with air in line detector off may cause an embolism resulting death or paralysis.
Switching the air detector off is not recommended and should be used only when absolutely
necessary. Please contact your technician in order to enable the air sensor if disabled with no
need.
Read the entire Operation Manual before using the pump, since the text includes important
precautions.
The maximum volume that may be infused under SINGLE FAULT CONDITION is 0.1 ml
Voltage present on internal components may cause severe shock or death upon contact.
Disconnect the Charger from the mains, prior to opening the casing. Only trained service
personnel should open the pump cover.
Blown fuses could cause a fire hazard. Replace blown fuses on the Charger only with fuses of
the same type and rating (see fuse values on the Charger PCB).
BodyGuard Operator Manual
16
2-Symbols, Warnings and Cautions
Boluses will be infused up to the full volume that is requested unless the user stops the bolus
manually by pressing . This is true even when the volume infused is less than the
specified Hour Limit for Volume (e.g. 4 Hr Limit) at the beginning of a bolus and goes over the
limit during the bolus (to complete it).
Make sure the pump is attached securely to the Charger, which is connected firmly to an IV pole.
A kinked or occluded IV line may impair the operation of the pump and the accuracy of the
infusion. Before operation, verify that the IV line is not kinked or occluded.
The BodyGuard should be operated only with the Bodyset IV lines. Use of IV lines other than the
Bodyset may impair the operation of the pump and the accuracy of an infusion.
Drugs must not be administered to the epidural space unless the drugs are indicated for this
purpose and are administered in accordance with the indications included in the manufacture’s
package. Epidural administration of drugs other than those indicated for epidural use could result
in serious injury to the patient. For epidural administration of drugs use MicroSet only.
Any adjustments, maintenance, or repair of the uncovered pump may impair the operation of the
BodyGuard Infusion System and/or the accuracy of the infusion. Any adjustments, maintenance,
or repair of the uncovered pump or charger should be performed by authorized skilled
technicians. Any adjustments, maintenance, or repair of the uncovered pump or charger while
connected to the power should be avoided.
The BodyGuard Infusion System should be operated within a temperature range of 18°C (59°F)
to 45°C (113°F) and up to 85% humidity. Operating the pump at temperatures and/or humidity
other than within that range may affect accuracy.
Unsafe operation may result from using improper accessories. Use only accessories and options
designed for this system.
Disposables must be compatible with the medicine delivered.
Battery charging is enabled as long as the charger cord is connected to the mains and the pump
is in the charger. Switching the pump off does not disconnect it from the mains. To disconnect
from mains, remove the charger cord from mains. To disconnect pump from mains, remove it
from the charger.
Dropping the BodyGuard Infusion System could cause damage to components. If the pump is
dropped, return the pump for inspection by qualified service personnel.
Use aseptic technique. Patient infection may result from the use of non-sterile components.
Maintain sterility of all disposable components and do not re-use single use IV sets.
Watch your fingers and nails when opening the pump door.
When operating the pump on PCA program with a rate of 0.0 ml/h there is a hazard of blood clot
forming. Connect saline infusion in parallel to avoid this problem.
When operating the pump on PCA program, the program will not enter KVO, even if the limit
volume is acceded. The program will complete the bolus and than enter KVO.
Do not operate the pump near high-energy radio-frequency emitting equipment, such as electrosurgical cauterizing equipment. False alarm signals may occur.
BodyGuard Operator Manual
17
2-Symbols, Warnings and Cautions
Do not let the pump operate when battery is fully depleted. Pump may turn off during operation
on fully depleted battery without alarm.
Never operate the pump without performing the battery test as described in chapter 4 in this
manual.
Do not use the BodyGuard 323 for infusion of Insulin or for infusion of blood.
Cautions
To avoid possible damage to the equipment, observe the following:
Leaving the battery in a depleted state for a long period of time may damage the battery.
Connect the pump to the mains via the Charger whenever possible to recharge the battery.
Do not store the pump with the battery fully depleted.
Xylene, Acetone or similar solvents could cause damage to components. Do not clean the
pump with these chemicals. Clean solution spills on the pump immediately. Use a damp cloth
or sponge. A mild detergent may be used. Wipe thoroughly with a dry cloth.
Immersing the BodyGuard Infusion pump into liquid could cause damage to components. Do
not immerse the pump into any type of liquid.
Battery damage could occur if left in a temperature warmer than 500C (122°F).
BodyGuard Operator Manual
18
3- Installation and Set up
3 – Installation and Set
up
Unpacking
1. Carefully remove the pump and Charger from the box.
2. Make sure no items were damaged during shipment
3. Make sure you have the following items:
BodyGuard Infusion Pump
•
Charging Cradle with Power Cord or Battery Charging Cable
•
Rechargeable Battery Pack
•
Operation Manual
•
PCA bolus cable (optional)
•
9V Alkaline Battery compartment (optional)
•
Disposable Battery Pack (optional)
•
•
If any items are missing or damaged, contact your BodyGuard supplier.
External charger (optional)
•
Extended life battery Li-Ion Polymer 3600 mAh (optional)
•
External wall charger (Optional)
BodyGuard Operator Manual
19
3- Installation and Set up
Charging the Pump
NOTE:
charged, connect the pump to the mains via the Charger whenever possible. Be sure that the
battery is fully charged at all times.
Warning:
charger connected to mains, the pump will turn off.
1. Connect the charger unit to AC power and verify that the AC indicator is lit.
2. Put the pump into the charger (top first) until it clicks into place.
The pump is protected against overcharging. In order to keep your battery fully
If the battery is removed during operation while pump is not connected to a
.
3. The battery is charging when the red battery LED on the front of the charger is lit. The battery is
fully charged when the battery LED symbol turns green.
If the led does not light or does not
turn green after more than 8 hours, pls. replace battery.
4. Remove the pump from the charger by pressing the release button (located next to the LED
indicators on the charger) and lift the pump out of the charger bottom first.
Caution:
battery.
Leaving the battery fully depleted for a long period of time may damage the
BodyGuard Operator Manual
20
4 - Operation
4 – Operation
Before Operating the Pump
Before attaching the system to a patient, run the following tests to verify that all indicators and alarms
work properly.
When an alarm is activated the following occurs:
An alarm message appears on the display
•
An audible alarm sounds
•
Infusion ceases
•
The LED changes from green to red
•
NOTE:
Pump Operation Test
1. Click the pump into a Charger connected to AC power. Verify that the CHARGE
indicator is on, Red or Green light (on charger).
2. Insert a BodySet administration set and close the door. Press the door until a click is heard.
NOTE:
“Loading the Administration Set” & “Priming the Line”.
All tests should be performed in the Continuous program.
For instructions on how to load and prime an administration set refer to sections
3.
Press until the Self-Test screen appears. If air sensor is off, press to confirm.
4. The program selected and its status are displayed. During the Self-Test a long beep will be
heard. This procedure verifies that the pump’s acoustic and visual features are working properly.
After self test is completed, the display will show the last rate setting.
Press to prime the IV line.
5.
Warning:
6. Enter the Volume of fluid required to prime the set and press
to begin priming procedure. You may stop priming
at any time by pressing .
7. The display screen will show a graph indicating the prime
operation.
Ensure the set is not connected to a patient!!
PRIME
Disconnect Patient
Press OK to Start
Prime Vol __ ml
Press OK to Continue
PRIME 5 ml
0 5
BodyGuard Operator Manual
21
4 - Operation
Volume
ml/h
ml/h
8. When priming is completed, wait 2 minutes without pressing
any key. After two minutes the screen displays “Pump
Unattended” and an alarm will sound. Press
silence the alarm again or to exit alarm condition.
to
Pump Unattended
Press OK to Start
Air in Line – Alarm Test
1. Insert a BodySet administration set and close the pump door. Press the door until a click is
heard.
2.
3. The pump displays the program selected and its status. The display will automatically change to
4. Change the infusion mode to Continuous (See “Change
5
.
6. Insert air into the IV line to allow an air bubble greater than 10 mm to enter into the administration
7. Air in line alarm will be activated as soon as the air bubble
NOTE:
“Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test Screen appears. If air sensor is off, press to confirm.
Please contact your technician in order to enable the air sensor if disabled with no need.
display the last rate setting.
Set up” Section for instructions). Set the infusion rate at
500ml/hr, using the Numeric keypad and press to
confirm setting.
Set Volume at 20 ml and press to start operation.
set.
enters the set segment located behind the pump door.
Press to turn the alarm off.
For instructions on how to load and prime an administration set refer to sections
Rate 500
Press OK to Continue
20
Press OK to Continue
Up Occlusion /Air
Press STOP to Mute
Down Occlusion – Alarm Test
1. Insert a BodySet into the pump and close the door. Press the door until a click is heard.
2. Make sure room temperature is 21-23 oC
3.
4. The pump displays the program selected and its status. The display will automatically change to
BodyGuard Operator Manual
NOTE:
“Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test screen appears. If air sensor is off, press to confirm.
Please contact your technician in order to enable the air sensor if disabled with no need.
display the last rate setting.
For instructions on how to load and prime an administration set refer to sections
22
5. Change the infusion mode to Continuous (See “Change Set
Door Open
Volume
ml
ml/h
Volume
ml
ml/h
up” Section for instructions). Set the infusion rate at 500
4 - Operation
Rate 500
ml/hr using the Numeric keypad and press .
6.
Set Volume at 20 ml and press to start operation.
7. Wait 5 minutes and than connect a pressure gauge at a distance of 1 meter from the exit of the
pump. Place both the pump and the pressure gauge at the same level (on a table)
8. Wait another 5 minutes and occlude the IV line downstream of the pump (between the pump and
the pressure gauge).
9. The Down Occlusion alarm occurs at the time and rates listed below (time is measured from
actual occlusion time to alarm, with accuracy of +30% -20%):
• 500 ml/h up to 20 seconds
• 60 ml/hr up to 1 minute
• 1 ml/hr up to 80 minutes
• 0.1 ml/hr up to 11 hours
10.
11.
NOTE:
changing the current default Pressure Limit).
Depending upon the set flow rate and pressure threshold,
the Down Occlusion alarm will be activated.
Press to mute the alarm.
The pressure threshold can be changed to suit patient need (refer to section on
Press OK to Continue
20
Press OK to Continue
Down Occlusion
Press STOP to Mute
Door Open – Alarm Test
1. Insert a BodySet in the pump and close the door. Press the door until a click is heard.
2.
3. The pump displays the program selected and its status. The display will automatically change to
4. Change the infusion mode to Continuous (See “Change Set
5.
6. Open the pump door.
7. Door Open alarm will be activated at once.
NOTE:
“Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test screen appears. If air sensor is off, press to confirm.
Please contact your technician in order to enable the air sensor if disabled with no need.
display the last rate setting.
up” Section for instructions). Set the infusion rate at
500ml/hr using the Numeric keypad and press .
Set Volume at 20 ml press to start operation.
For instructions on how to load and prime an administration set refer to sections
Rate 500
Press OK to Continue
20
Press OK to Continue
Close Door
8.
Close the door and press
BodyGuard Operator Manual
.
23
Battery Test
Low Battery
Volume
9999
ml
ml/h
1. Insert a BodySet in the pump and close the door. Press the door until a click is heard.
4 - Operation
2. Take the pump out of the charger.
3.
4. The pump displays the program selected and its status. The display will automatically change to
5. Change the infusion mode to Continuous (See “Change Set
“Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test screen appears. If air sensor is off, press to confirm.
Please contact your technician in order to enable the air sensor if disabled with no need.
display the last rate setting.
up” Section for instructions). Set the infusion rate at 1200
ml/hr using the Numeric keypad and press .
Set Volume at 9999 ml and press to start operation.
For instructions on how to load and prime an administration set refer to sections
Rate 1200
Press OK to Continue
Press OK to Continue
Connect to Charger
8.
Press to turn the pump off.
9. Connect the pump to a charger.
10. Verify the red icon of battery is lit (on the charger).
Verify the red battery turns green after a minimum of 2 hours.
11.
WARNING: Do not let the pump operate when battery is fully depleted. Pump may turn off during
operation on fully depleted battery without alarm.
Charger Indicator Test
1. Click the pump into the charger and connect the Charger to mains. Disconnect the power cord
from the AC power outlet.
2.
Verify that the CHARGE indicator is OFF.
BodyGuard Operator Manual
3. Connect the power cord to the AC power outlet. Verify that the CHARGE indicator is ON.
24
Loading...
+ 56 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.