CIRCUTOR CVM96 User Manual

SUPPLY NETWORK ANALYZER
CVM 96 SERIES
(V. 6.12 and higher)
INSTRUCTION MANUAL
( M98132601-03 / 03C)
(c)
CIRCUTOR S.A.
----- Supply network analyzer CVM 96 ---------- ------- Page No. 1
CONTENTS page
1.- DELIVERY SPOT CHECK .......................................................................................................... 2
2.- MAIN FEATURES ....................................................................................................................... 3
3.- INSTALLATION AND START-UP............................................................................................... 5
3.1.- Installation. .................................................................................................................... 5
3.2.- CVM 96 (RS-485) Connection terninal (see lable on the rear part) .............................. 7
3.3.- CVM 96 (RS-232) Connection terninal (see lable on the rear part) .............................. 8
3.4.- Connection drawing for the CVM 96 ............................................................................. 9
4.- OPERATION MODE ................................................................................................................. 13
5.- SETUP PROCEDURE .............................................................................................................. 15
5.1.- Phase-to-Phase or Phase-to-Neutral voltages ............................................................ 16
5.2.- Voltage display mode .................................................................................................. 16
5.3.- Voltage transformation ratio ........................................................................................ 17
5.3.1.- Voltage Transformer Primary ............................................................................... 17
5.3.2.- Voltage Transformer Secondary........................................................................... 18
5.4.- Current Transformer Primary ...................................................................................... 18
5.5.- Setting power demand utility screens.......................................................................... 19
5.6.- Initial screen setting..................................................................................................... 20
5.7.- Setting of the display shutdown time........................................................................... 20
5.8.- Clearing energy counters ............................................................................................ 21
5.9.- THD or D setting..........................................................................................................22
5.10.- Additional screens when RELAY OUTPUTS (2 relays) are equipped ...................... 23
6.- SPECIFICATIONS .................................................................................................................... 30
7.- SAFETY CONSIDERATIONS................................................................................................... 32
8.- MAINTENANCE ........................................................................................................................ 32
9.- TECHNICAL SERVICE ............................................................................................................. 32
10.- CVM 96 COMMUNICATIONS................................................................................................... 33
10.1.- ! To take into account:............................................................................................... 33
10.2.- RS-485 TYPE CONNECTION TO A RS-232 TYPE INPUT OF A PC ...................... 34
10.3.- RS-232 TYPE CONNECTION TO A RS-232 TYPE INPUT OF A PC ...................... 35
10.4.- MODBUS © protocol ................................................................................................ 36
11.- APPENDIX A: Second SETUP of the CVM 96 ........................................................................ 42
----- Supply network analyzer CVM 96 ---------- ------- Page No. 2
1.- DELIVERY SPOT CHECK
This manual is issued to help all the CVM 96 users to install and use it in order
get the best from it. After receiving the unit, please check the following points:
(a) Does this device correspond to your order specifications? (b) Check if any damage was done during the shipment process. (c) Verify that it includes the correct instruction manual.
The manual you hold in your hands contains information and warnings
about the CVM 96 that the user should respect in order to guarantee a proper operation of all the instrument functions and keep its safety conditions.
Before powering the instrument for the first time, verify following points:
(a) Power supply: see rear part of your CVM 96
 Standard: 230 V a.c. Single-phase, Frequency: 50 ... 60 Hz
 SDC Model : 24 .... 120 V c.c.
 Other supply voltages, on request.
(b) Maximum measuring voltage:
 Standard: 300 V a.c. phase-neutral / 520 V a.c. between phases  Other models: on request
CVM 96 – measuring 110V: 110 V a.c. phase-to-neutral / 190 V a.c. phase-to-phase CVM 96 – measuring 500V: 500 V a.c. phase-to-neutral / 866 V a.c. phase-to-phase
(c) Maximum measurable current: Current transformer of In / 5 A a.c.
----- Supply network analyzer CVM 96 ---------- ------- Page No. 3
2.- MAIN FEATURES
The CVM 96 power meter is a programmable measuring instrument, offering
several operation possibilities selectable in its SETUP option. Before power supplying the instrument, read the CONNECTIONS and SETUP sections and choose the most suitable operation mode for getting your desired data.
The CVM 96 is an instrument which measures, calculates and displays all the
main electrical parameters at any electrical network (balanced or not). The measuring is true RMS value, through three a.c. voltage inputs and three a.c. current inputs (from Current Transformers .../ 5A).
----- Supply network analyzer CVM 96 ---------- ------- Page No. 4
By means of an internal microprocessor it simultaneously measures:
Parameter Symbol L1 L2 L3 Average
Voltage (phase-neutral) V x x x Voltage (phase-phase) V x x x Current A x x x xx Frequency Hz x Active power kW x x x x Reactive power L kvarL x x x x Reactive power C kvarC x x x x Apparent power xx Power factor PF x x x x Cos ϕ Power demand Pd x kW h energy x kvarhL energy x kvarhC energy x
Neutral current IN xx Voltage THD % THD- V x x x Current THD % THD- A x x x Individual harmonic current content (Up to the 15
Available: x: Display and communications xx: Communications
th
)
xx
xx xx xx
The CVM 96 delivers the visualization of above listed parameters by means of three 4-digit, led type, displays. Up to 3 parameters can be simultaneously read in the screen.
---------------------------------------------------------------------------------------------
OTHER FEATURES
- Low-size (96 x 96 mm), panel-mounting analyzer.
- True R.M.S. measuring system.
- Instantaneous, maximum and minimum values of each measured parameter.
- Energy measurement (indication through a lighting led)
- RS-485 or RS-232 type communication to a PC (optional)
----- Supply network analyzer CVM 96 ---------- ------- Page No. 5
3.- INSTALLATION AND START-UP
The manual you hold in your hands contains information and warnings that the user should respect in order to guarantee a proper operation of all the instrument functions and keep its safety conditions. The instrument must not be powered and used until its definitive assembly on the cabinet’s door.
Whether the instrument is not used as manufacturer’s specifications, the protection of the instrument can be damaged.
When any protection failure is suspected to exist (for example, it presents external visible damages), the instrument must be immediately powered off. In this case contact a qualified service representative.
3.1.- Installation.
Before powering the instrument for the first time, verify following points:
a.- Power supply: see rear part of your CVM 96
 Standard supply : Single-phase 230 V ∼ (a.c.)
- Frequency : 50 - 60 Hz
- Supply voltage tolerance : - 10 % / + 15
- Connection terminal : Terminals 1 - 2 (Power supply)
- Burden : 5 VA
 SDC Model : 0.24.... 120 V c.c. ( continua )
- Supply voltage tolerance : - 20 % / + 15
- Connection terminal : Terminals 1 - 2 (Power supply)
- Burden : 2,2 W
 On request: other voltages
----- Supply network analyzer CVM 96 ---------- ------- Page No. 6
b.-Maximum voltage at the voltage measuring circuit:
 Standard : 300 V a.c. phase-to-neutral / 520 V a.c. phase-to-phase 45 to 65 Hz  On request:
500 V a.c. phase-to-neutral / 866 V a.c. phase-to-phase 110 V a.c. phase-to-neutral / 190 V a.c. phase-to-phase
c.- Maximum allowable current : Current Transformer of In / 5 A a.c.
d.- Operation conditions :
- Operation temperature range: -10 to +50 ºC
- Humidity : 5 to 95 % R.H. non-condensing
e.- Safety : Designed to meet protection class III- 300 V a.c. as (EN 61010).
Mounting:
Instrument is to be mounted on panel (cut-out 92
+0.8
x 92
+0.8
mm, as per DIN
43 700). All connections keep inside the cabinet.
Note that with the instrument powered on, the terminals could be dangerous to touching and cover opening actions or elements removal may allow accessing dangerous parts. Therefore, the instrument must not be used until this is completely installed.
The instrument must be connected to a power supply circuit protected with gl type (IEC 269 ) or M type fuses rated between 0.5 and 2 A. This circuit should be provided with a circuit breaker or any equivalent element to connect (ON) or disconnect (OFF) the instrument from the power supply network. The supply and measuring voltage circuits will be both connected through a wire with a minimum cross-section of 1 mm2.
The line of the current transformer secondary will have a minimum cross­section of 2,5 mm2.
----- Supply network analyzer CVM 96 ---------- ------- Page No. 7
3.2.- CVM 96 (RS-485) Connection terninal (see lable on the rear part)
No.
Terminal description
Upper connection terminal
Relay RL1 output
10
Relay RL2 output
11
Relay common
12
Current AL3 - S2 input
13
Current AL3 - S1 input
14
Current AL2 - S2 input
15
Current AL2 - S1 input
16
Current AL1 - S2 input
17
Current AL1 - S2 input
18
Lower connection terminal
*Supply voltage input: 0 V
1
*Supply voltage input:230 Va.c.
2
RS-485 ( + )
3
RS-485 ( - )
4
RS-485 ( GND )
5
Neutral measurement
6
VL3 measurement
7
VL2 measurement
8
VL1 measurement
9
*Lower connection terminal (SDC Model)
Supply voltage (+) d.c.
1
Supply voltage (-) d.c.
2
NOTE: Terminals 13, 15, 17 are internally connected to the 6 terminal (Neutral) ... / 5 A current inputs are isolated for the ITF model.
----- Supply network analyzer CVM 96 ---------- ------- Page No. 8
3.3.- CVM 96 (RS-232) Connection terninal (see lable on the rear part)
RELAY ALARM
10 11 12 13 14 15 16 17 18
0
RL2
COMRL1
RS-232
230 V
(Tx) (Rx) GND
S2
S2
L3
S1
S1
L2
S1
S2
S2
S1
N
VL3 VL2 VL1
S2
S2
L1
S1
S1
No.
Terminal description
Upper connection terminal
10
Relay RL1 output
11
Relay RL2 output
12
Relay common
13
Current AL3 - S2 input
14
Current AL3 - S1 input
15
Current AL2 - S2 input
16
Current AL2 - S1 input
17
Current AL1 - S2 input
18
Current AL1 - S1 input
Lower connection terminal
1
*Supply voltage input: 0 V
2
*Supply voltage input:230 Va.c.
3
RS-232 (Tx )
4
RS-232 (Rx )
5
RS-232 ( GND )
6
Neutral measurement
7
VL3 measurement
8
VL2 measurement
9
VL1 measurement
*Lower connection terminal (SDC Model)
1
Supply voltage (+) d.c.
2
Supply voltage (-) d.c.
NOTE: Terminals 13, 15, 17 are internally connected to the 6 terminal (Neutral) ... / 5 A current inputs are isolated for the ITF model.
A
----- Supply network analyzer CVM 96 ---------- ------- Page No. 9
3.4.- Connection drawing for the CVM 96
a.- Three-phase network.- 4 wires (low voltage) :
P2
P1
L1
S1
L2
L3
N
Vl3
S1
P1
VL2
S2
Vl1
P2
S2
P2
S2
Power Supply
C.
1 2 3 4 5 6 7 8 9
Power Supply (Model SDC)
Terminal Description
1 2
+ V c.c.
P1
S1
10 11 12 13 14 15 16 17 18
N
- V c.c.
A
----- Supply network analyzer CVM 96 ---------- ------- Page No. 10
b.-Three-phase network – 3 wires (low voltage):
P2
P1
L1
Power Supply
C.
P2
S2
P2
S2
1 2 3 4 5 6 7 8 9
P1
S1
10 11 12 13 14 15 16 17 18
S1
P1
S2
S1
L2
L3
Vl1
VL2
Power Supply (Model SDC)
Terminal Description
1 2
+ V c.c.
- V c.c.
Vl3
A
A
A
----- Supply network analyzer CVM 96 ---------- ------- Page No. 11
c.- Three-phase network - 3 wires (2 P.T. and 3 C.T.):
Power Supply
C.
P2
S2
P2
S2
1 2 3 4 5 6 7 8 9
P1
S1
10 11 12 13 14 15 16 17 18
S1
P1
P2
S2
S1
P1
L1
L2
L3
B
a
a
b
B
b
Vl1
VL2
Power Supply (Model SDC)
Terminal Description
1 2
+ V c.c.
- V c.c.
Vl3
A
A
A
----- Supply network analyzer CVM 96 ---------- ------- Page No. 12
d.- Three-phase network - 3 wires (2 P.T. and 2 C.T.):
P2
P1
L1
S2
P2
S2
Power Supply
C.
Power Supply (Model SDC)
Terminal Description
1 2
P1
S1
10 11 12 13 14 15 16 17 18
1 2 3 4 5 6 7 8 9
Vl1
Vl2
Vl3
+ V c.c.
- V c.c.
S1
L2
L3
B
a
a
b
B
b
IMPORTANT REMARK! If power = -0.01 is shown for any of the phases and voltage
and current are not zero for this phase, check out following points:
- Assure that L1, L2 and L3 phases coincide in voltage and current.
- Correct polarity? Reverse the current transformer placed at this phase.
----- Supply network analyzer CVM 96 ---------- ------- Page No. 13
4.- OPERATION MODE
The instrument has three displays. Each one has also a LED type indicators (red colour). Every led will be on according to the parameter presently shown in screen.
When the CVM 96 is powered up, the display shows " xxxx ", so informing about the program version and the SETUP. After some seconds, the analyzer is ready for operation and shows one of the available screens. A led next to the shown parameter is on.
Parameters on display can be switch by pressing the key . Leds on the right or left indicate the parameters shown on screen at any moment.
When the first led (red) is on, it means that VOLTAGE values are shown on screen, that is, the user can read the voltage of the phase L1 (V1), voltage of the phase L2 (V2) and voltage of the phase L3 (V3).
Whether the key " " is press, the next led will be on, so indicating that for this screen CURRENT values of each phase are now shown.
Pressing again the key " " , the next led will be on, and the screen will shown following three parameters, and so successively.
Loading...
+ 30 hidden pages