Carlo Gavazzi WM5-96 User Manual

Page 1
WM5-96
Smart Power Quality Analyzer
Instruction manual
Page 2
Page 3
ARMPowered
PRG
measures
SS
!
R
A
L
WM5-96
SmartPowerQuality Analyzer
Class0.2 A/V
Thank you for choosing our products
WM5-96
Smart Power Quality Analyzer
WM5 96:
• High accuracy (class 0.2 A/V);
• High calculation performances (ARM®technology) for a fast analysis of the signal (FFT up to the 63
• high connection capabilities (RS485 115.2 kbps, RS232, front optical port).
rd
harmonics);
WM5 96 is the state-of-the-art tecnological answer to your needs of power quality analysis.
eover, you can count on a ISO9001/VISION 2000
Mor certified company structure, an experience of many years
and a wide-spread presence both in Europe and all over
the world. All this in order to guarantee the customer with
top-quality service and the best products.
a
Welcome in Carlo Gavazzi and our compliments for your
choice. You can evaluate the complete range of our prod-
ucts on the CARLO GAVAZZI web-site:
www.carlogavazzi.com/ac
CARLO GAVAZZI
Automation Components
Page 4
2
ntroduction
I
Index
WM5-96 Instruction Manual
M5 96 philosophy, ARM tecnology . . . . . . . . . . . . . . . . . . . . . . . .4
W
Optical and communication port . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Becoming aquainted with WM5
Programming WM5 96
Front panel description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
ain measuring page “00” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
M
Adjustment of display contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
nstantaneous variables measurements . . . . . . . . . . . . . . . . . . . . . .6
I
THD measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Digital input status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Digital output status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Energy meters pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Logged events page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Alarm status page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Info pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Programming keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Resets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Access to programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Change password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
CT ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
VT ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
DMD/AVG calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Installed power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Meters menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Programming example of the tariff parameters . . . . . . . . . . . . . .18
Display page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Digital inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Digital outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Logic and alarm parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Examples of alarm parameters pr
Events selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Analog outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Programming example of the analog outputs . . . . . . . . . . . . . . . .30
Serial outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Optical port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Ethernet port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Digital filter
Digital filter programming examples . . . . . . . . . . . . . . . . . . . . . . .34
Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Language
Serial number and Meter Info. . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Keypad lock, lock of acces to programming and reset menu . . . . . . . . .37
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
ogramming
. . . . . . . . . . . . . . . .
26
Layout and list of modules
Lay-out and list of modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
Page 5
WM5-96 Instruction Manual
Index
3
Installation
Wiring diagrams
Technical specifications
Panel cut-out and dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Fixing, removing the modules and mounting the instrument to the
panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Sealing the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
-phase, 2-wire input connections (1P) . . . . . . . . . . . . . . . . . . . . .40
1
2-phase, 3-wire input connections (2P) . . . . . . . . . . . . . . . . . . . . .40
3-phase, 3 and 4-wire input connections - balanced load (3P) . .40
3-phase, 3-wire input connections - unbalanced load (3P) . . . . . .41
3-phase, 3-wire ARON input connections (3P) . . . . . . . . . . . . . . .41
3-phase, 4-wire input connections, unbalanced load (3P+N) . . . .41
Power supply, analogue and digital output modules . . . . . . . . . . .42
Digital inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
RS485 serial port (AR1034 9600bps, AR2040 115200bps)
Input specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Output specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Software functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Wm5Soft parameter programming and variable reading software . . .45
Time period management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Harmonic distortion analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
General specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Supply specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Revenue approval Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Function description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Insulation between inputs and outputs . . . . . . . . . . . . . . . . . . . . .48
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
Used calculation formulas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
List of the variables that can be connected to . . . . . . . . . . . . . . .50
. . . . . .42
Page 6
4
We suggest you to keep the original packing
n case it is necessary to return the
i instrument to our Technical Service Department. In order to achieve the best results with your instrument, we
recommend you to read this instruction man-
al carefully. If the instrument is used in a
u manner not specified by the producer, the
rotection provided by the instrument may be
p impaired. Maintenance: to keep the instru­ment clean, use a slightly damp cloth; do not use any abrasives or solvents. We recom­mend to disconnect the instrument before cleaning it.
WM5 96 philosophy, ARM®technology
Introduction
WM5-96 Instruction Manual
his symbol indicates a particularly important sub-
T ject or information.
This symbol indicates that more details are given
n the current subject.
o
This symbol indicates a suggestion for the user.
WM5 96 is a brand new instrument with a high level of performances and connection capability. Actually, WM5 96 is a synergy of digital
components that, coordinated by an ARM to perform class 0.2 measurements for current and voltage, the man­agement of 3 serial ports, 8 analogue outputs, 12 digital inputs, 16 alarms and a complete and functional management of the energy
meters tariffs. The ARM®-based microprocessors are used in the up­to-date technology such as the palmtop computers: this makes of WM596 a real computer at the service of the electrical parameters analysis and of the electrical tariff management, even the more com­plex one.
®
processor, allows the user
Optical port and communication
WM5 thanks to its great flexibility allows to set in the menu (pls see
the relevant section) also the mod­ules being installed in the base (display mod­ule). In order to make the programming of the modules easier, it is advisable to note the identification code (ex. AO2050) and the rele­vant installation slots (A, B, etc.): we suggest to fill in the special module on the last page of this manual.
The front optical port allows a fast and practical communication between PC and instrument. Thanks to the WM5Soft software it’ll be possible to read the measurements and programm the instrument even if the latter is already installed on the electrical board.
Basic instrument,
display.
Modules
Page 7
WM5-96 Instruction Manual
ARMPowered
P
RG
m
easures
SS
R
!
R
A
L
WM5-96
SmartPowerQuality Analyzer
Class0.2 A/V
Front Panel Description
2
4
3
Becoming acquainted with WM5 96
1
When switched on, the alarm LED will be active (blinking for a vir-
ual alarm, fixed light for a real alarm).
t
2
1
5
Graphic display.
3
This key allows the user to access programming (only page 00) or the details of the measurement.
4
Keys allowing the user to scroll the measuring pages, display the details relating to the measurements: dmM=dmd maximum value, dmd=avg value, max=maximum value, min=minimum value. In the page relating to the meters it’s possible to read the value of the energy counted per month (JAN...DEC), per tariff (T01...T12) and per phase (L1, L2, L3). The contemporaneous pressure of these keys also allows the adjustment of the display
5
Optical communication port (standard ANSI).
5
Main measuring page “00”
a b
c
2
1
2a
3 4
When the instrument is switched on, the operating system will be loaded; this implies a short waiting time (during this phase the display shows a sand glass which will stop moving only when the loading phase is completed). Then, the instrument will start measur-
d
d1
ing and will display the window on the left. This page, called "P00" page, is the only page which can be configured by the user who can choose, by means of the programming, which variables are to be dis­played in the 3 sections
(a, b, c) (see “display page” in the program-
ming menu). The (d) section indicates the consumption of the meas­ured energy (kWh and kvarh), a graphical indicator (d1) allows a quick reading of the % value relating to the active system power being used with reference to the installed power value previously set in the instru­ment. In the other measuring pages, the variable type displayed in the four sections (a, b, c, d) is pre-established and cannot be modified. The display of the lower part of the display is common to all measur­ing pages.
1
Shows the displayed measuring page, the displayed number will
incr
ease or decr
2
Shows the sequence of the phase rotation L1-L2-L3 or L3-L2-L1
ease depending on the displayed page.
(2a).
3
Shows the current tariff (T01, T02, T03, T04...T12) of the energy meters.
4
Shows the current time (only if this function has been enabled).
Adjustment of display contrast
To have a clear reading in every condition of visibility, properly adjust the display contrast. Press the “ ” keys contemporaneously: the contrast starts increasing; release the keys when the display contrast has been properly adjusted.
Page 8
6
easurements of instantaneous variables
M
a b
c
1
2
3
Becoming acquainted with WM5 96
d
WM5-96 Instruction Manual
The various measuring pages of WM5 96 display the necessary infor-
ation for the quality analysis of the network. To scroll (1) the measur-
m ing pages, use the “ ” keys; to display the details of the measures, press the “S” key; to scroll the available details, use the “ ” keys
(3); to exit the details displaying phase, press the “S” key.
he details for the instantaneous variables are the following:
T
dmM= maximum dmd value, dmd= dmd value, max= maximum value, min= minimum value.
For any information on the meaning of “dmd”, see page 14. Reset of details. Keeping the “S” key pressed for approx. 2 seconds,
it’s possible to reset each variable of the selected detail (dmM, dmd, max, min). The instrument displays a message confirming the user he has entered the reset mode: choose “YES” by means of the “ ” keys , confirm with the “S” key and then select the variable you want to reset by means of the “ ” keys, then press “S” to proceed with the reset; to exit the reset mode, keep the “S” key pressed for at least 2 seconds.
THD Measurements
a
b
c
1
d
e
2
Thanks to the powerful calculation capabilities of the ARM® technolo­gies, WM5 96 allows an accurate analysis of the harmonics up to the 63rd.
a
Voltage or current value referred to the relevant measured phase. (V1, V2, V3, A1, A2, A3).
b
Total harmonic distorsion expressed as a percentage value.
c
Histogram of the harmonics. To display the details of the harmon­ics, press the “S” key
(2).
An arrow is displayed on the horizontal axis of the graph to identify the harmonics being examined. To scroll the harmonics one by one, use the “ ” keys (2).
d
Harmonic order (from h1 to h 63), for single phase and the relevant absolute voltage or current value.
e
Harmonic order (from h1 to h63), the detected conventional sign (-,o,+) and its relevant value expressed as a percentage.
f
f
Phase angle between the fundamental and the voltage and current harmonic of the same order: angle between 0° and 90° and between 270° and 360° corresponds to a generated harmonic, an angle between 90° and 180° and between 180° and 270° corre­sponds to an imported harmonic.
The phase angle (f) between the fun­damental and the voltage and current
harmonic of the same or
der will be displayed only if the measurements are taken in a three­phase system with neutral.
To exit the display of measurement details, press the “S” key.
t
Note: The display of histograms is available only up to the 50
h
har­monic (h50). For the other harmonics from h51 to h63 only the numer­ic values are available.
Page 9
WM5-96 Instruction Manual
igital input status
D
a
b
Becoming acquainted with WM5 96
This page displays the status of the digital inputs (b) (0, 1, 2) corre­sponding to each slot (a). The “ ” symbol will be shown on the right
f a close contact, while the “ ” symbol will be shown on the right of
o an open contact, if the digital input is set as “Totalizer” the symbol “ “ will be displayed.
It’s possible to display the values of the cofigurated totalizer (GAS, Water...) by keeping the “S” key pressed, and scroll the details by keeping the “ ” keys pressed. To exit the display of the measurement details press the “s” key.
7
Digital output status
b
a
Energy meter page
a
b
c
d
1
This page displays the status of the digital outputs (b) (0, 1, 2, 3) corrsponding to each slot (a). If the contact or contacts are ON, the “ ” symbol will be displayed. If the contact or contacts are OFF, the “ ” symbol will be displayed. If the digital output is set as “pulse output”, then the symbol “
“ will be displayed. The detail function is not avail-
able in this page.
This page displays the imported kWh (a), exported kWh (b), imported kvarh (c) and exported kvarh (d) energy meters. It’s possible to enter the details of the meters by keeping the “S” key pressed (2), and scroll the details by keeping the “ ” keys pressed (3).
The countings of the energies can be scrolled divided by:
month (JAN...DEC), tariff (T01...T12), phase (L1, L2, L3).
2
3
o enter the details of the meters
T starting fr
om the division by phase,
press the “ ” key first, to start
om the division by month, pr
fr key first.
To exit the display of the measurement details press the “s” key.
ess the “
Page 10
8
ogged events page
L
d
1
Becoming acquainted with WM5 96
WM5-96 Instruction Manual
hanks to the great logging capability of WM5 96, it’s possible to log
a
T a great number of events and manage their displaying, one by one. The page always displays the last logged event which is given a pro-
b
c
e
f
gressive number (a). The page is complete with all the important infor­mation relevant to the logged event: date (b), time (c), type of event
alarm, min, max, etc.) (d), variable relevant to the event (e). Should the
( alarm refer to an alarm, a window (f) summing up all the set-point
nformation (see also following paragraph “alarm status page”).
i
To display all the logged events one by one, press the “S” (2) key, and
a
scroll the events with the “ ”.keys.
d
2
To display the first logged event, press, first press the “ “ key.
Alarm status page
a
b
c
e
To exit the detailed display of the events, press the “S” key.
Thanks to this page, it’s possible to have the complete control of the set alarms: ON if enabled, OFF if disabled. For the alarms which are not enabled, the display shows two horizontal hiphens. The column under the “ALARM” indication identifies the alarm group, for example
esents the gr
b
01 to 04 r their status indicated on the right.
epr
oup of the first 4 alarms (from 1 to 4) and
1
Press the “S” key to display the details (2) of each set alarm (c) (from 1 to 16): the values of the programmed set points (e), the alarm type (d): UP alarm if the “AL” symbol is in the (d1) position as shwon in the figur
c
d
g
h
2
d1
d2
f
d3
e
is in the (d2) position or DOWN alarm if the “AL” symbol is in the (d3) position. Moreover, the display shows the alarm status (g) which can be ON or OFF and the value of the alarm variable
To go from one alarm to the other, use the “ ” keys. To exit the display of the alarm status press the “S” key.
e on the right, WINDOW alarm if the “AL” symbol
(h).
Further information on the alarms and the set-point adjustment is given in the “alarms programming” section of this manual.
To display the details of the 01 alarm, press the “ ” key first.
Page 11
WM5-96 Instruction Manual
nfo pages
I
METER INFORMATION
TANDARDS:
S
kWh: EN62053-22 CL.0,5 kvarh: EN62053-23 CL.2
P37
1
METER INFORMATION
TEST OUTPUT CONSTANT: OPTICAL: kWh+, TOT
1.000 Wh/PULSE
OUT C0: kWh+ TOT
1.000Wh/PULSE
Becoming acquainted with WM5 96
he last pages in the “display” mode show the main information relat-
T ing to the programming of the instrument. These data are also avail­able when the instrument is completely sealed.
a
b
The first of the 4 pages (1) only the standards to which the instrument refers (a, b) are displayed.
n the 2nd page (2) the output specifications of the (infrared light) opti-
I cal port and open-collector output (C0) are displayed if these are used
o retransmit the energy.
t In the 3rd page (3) the following data are displayed: system measure­ment (a) the example figure shows the 3 phase 4 wire symbol, CT and PT ratio selected (c, d) and the DMD time selected (e). In the 4th page (4) the following data are displayed: serial number of the instrument (a), production year (b), firmware release (c), measuring input specifications, (d).
9
P38
2
METER INFORMATION
SYSTEM CT RA
TIO PT RATIO 1.000 DMD TIME 15 min
P39
3
METER INFORMA
SN YEAR 2008
BI3230033001P
b
REV 2.1.5.0 (AV5) INPUT 3x230 (400)V
5 (10)A, 50/60Hz
a
1.000
c
d
e
TION
a
c
d
P40
4
Page 12
Page 13
WM5-96 Instruction Manual
rogramming keypad
P
1 2 3
Programming WM5 96
11
The “S” (1) key is used to enter programming, confirm the selected or
odified data and go back to the previous data if pressed for at least
m 2 seconds. The “ ” keys at N. 6 are used to scroll the menus and modify the selected values/data. To modify the values proceed as fol­lows: the “ (2) key increases in a cyclical way the selected digit
rom 0 to 9, to move to the next digit, use the “ (3) key, also in this
f case the system is cyclical.
Example, enter the value 213: starting situation 000, press “ ” twice to increase the number up to 2, then press “ ” to go on to the next digit (on the right); the instrument displays 200, press “ ” once to increase the number to 1, then press “ ” to go on to the next digit (on the right); the instrument displays 210, press “ ” three times to increase the digit to 3, press “S” to confirm the value. If you do not want to confirm the value, but only to modifiy it, press the “ ” key.
Resets
PASS ?
Access to programming
PASS ?
1
0000
123...
0000
Measure
> 2s
Measure
> 2s
By means of a single command and only from the “P00” page, you can carry out the reset of all the min. and max. values, the dmd and dmd max values, the reset of all the logged events and of the alarms with latch. To carry out this reset command you have to enter some numeric codes: press the “S” key, the instrument displays the “
PASS?
indication. Enter by means of the “ ” keys the following numeric codes depending on the reset you want to carry out.
5784= reset of all the minimum and maximum values;
5785= reset of all the “dmd” and “dmM” (dmd max) values;
7239= reset of all the “events”;
9288= reset of the alarms with latch.
To access the programming menus from the measuring and display phase, press the “S” key (only from the “P00” page), when the pass­word “PASS?” is requested, enter the correct PASSWORD value by means of the “ ” keys, then confirm by pressing the “S” key again. If the PASSWORD is correct (for all instruments the PASS value when they leave the factory is “0”), the instrument will enter the main menu. To cancel the choice and go back to the measuring mode, press the “S” key for at least 2 seconds.
**********MMAAIINN MMEENNUU*********
CHANGE P
MODULES SYSTEM
2
CT RATIO
RREEVV.. 11..44..22..66 ((AAVV55)
d
a
ASSWORD
b
e
Measure
*
c
)
> 2s
a
Menu title.
b
Menu.
c
Scroll bar.
d
Firmware revision of the instrument.
e
ype of measuring analogue output being mounted on the instru-
T ment.
Page 14
12
Programming WM5 96
WM5-96 Instruction Manual
Change password
******MMAAIINN MMEENNUU*****
CHANGE PASSWORD
MODULES SYSTEM
T RATIO
1
2
C
RREEVV.. 11..44..22..66 ((AAVV55)
******MMAAIINN MMEENNUU*****
HANGE PASSWORD
C
ODULES
M
NUOVA PASS ?
YSTEMA
S
m
in
CT RAGIO ORTO TA
max
RREEVV.. 11..44..22..66 ((AAVV55)
ma
Confer
Modules
******MMAAIINN MMEENNUU*****
CHANGE PASSWORD
MODULES
SYSTEM
1
2
CT RATIO
RREEVV.. 11..44..22..66 ((AAVV55)
MMOODDUULLEESS CCOONNFFIIGGUURRAATTIIOON
SLOT A
SLOT B SLOT C
AAOO22005500 ((AA00,, AA11) AANNAALLOOGG OOUUTTPPUUTT 00--2200mmA
This function allows the user to modify the PASS value with a new
alue (from 0 to 1000). Press the “S” key and when the instruments
Measure
*
v requires a new PASS (2) enter the desired value by means of the “
” keys and confirm the new value with the “S” key. The instrument
> 2s
goes back to the main menu.
)
a
The instrument shows the maximum and minimum limit of the
Go back
*
000
0
000
0
a
1000
)
123...
*
> 2s
Measure
value available for the password.
The WM5 96 does not support the automatic acknowledgment of the installed modules, therefore this information must be entered using
> 2s
the “MODULES” menu. Choose by means of the “ ” keys the modules menu, press “S” to confirm and then select by the “
)
keys the relevant slot A, B, C, D or E (where slot A is the first on the top right, watching it from the front of the instrument, and slot E refers to the central housing). Confirm the highlighted slot with the “S” key
Go back
N
)
A
and then select the code of the module (the code is written on the module itself): the instrument will also display a brief description of the selected code. The code of the module is placed on the side label
> 2s
of the module itself.
o confirm the code of the selected module,
T
press the “S” key.
A
N
Go back
> 2s
MMOODDUULLEESS CCOONNFFIIGGUURRAATTIIOON
AO1059 AO2051
AO2050
3
AAOO22005500 ((AA00,, AA11) AANNAALLOOGG OOUUTTPPUUTT 00--2200mmA
Confirm
)
Page 15
WM5-96 Instruction Manual
ystem
S
********** MMAAIINN MMEENNUU *********
HANGE PASSWORD
C MODULES
SYSTEM
T RATIO
1
C
RREEVV.. 11..44..22..66 ((AAVV55)
)
Measure
*
Programming WM5 96
his function allows the user to select the type of electrical system
T choosing among: 1- single phase 2 wires (1-PHASE 2-WIRE),
> 2s
2- dual phase 3 wires (2-PHASE 3-WIRE), 3- three phase 3 wires and 1 CT balanced load (3-PH 3W+1CT BAL.),
hree phase 3 wires unbalanced load (3-PH 3W UNBAL.),
4-t 5- three phase 4 wires unbalanced load (3-PH 4W UNBAL.).
13
2
CT ratio
1
2
Confir
********** MMAAIINN MMEENNUU *********
HANGE PASSWORD
C
1-PHASE 2-WIRE
MODULES
-PHASE 3-WIRE
2
SYSTEMA
-PH 3W+1CT BAL.
3
T RATIO TA
C
3-PH 3W UNBAL
RREEVV.. 11..44..22..66 ((AAVV55)
Confirm
******MMAAIINN MMEENNUU*****
HANGE PASSWORD
C
ODULES
M
YSTEM
S
CT RATIO
RREEVV.. 11..44..22..66 ((AAVV55)
******MMAAIINN MMEENNUU*****
CHANGE PASSWORD MODULES SYSTEMA
CT RATIOORTO TA
RREEVV.. 11..44..22..66 ((AAVV55)
m
RAPP min max
A ?
. T
1.000
0.100
a
6000
123...
Choose by means of the “ ” key the SYSTEM function (1), press the “S” key to confirm, then select by means of the “ ” keys the desired system (2) and confirm the selection with the “S” key.
> 2s
)
Go back
*
This function allows the user to select the value of the CT ratio (pri-
Measure
*
mary/secondary ratio of the current transformer being used). Example: if the CT primary (current transformer) has a current of 300A and the secondary a current of 5A, the CT ratio corresponds to 60 (obtained
> 2s
using the following calculation: 300/5). Choose the CT RATIO function
)
(1) by means of the “ ” keys and confirm the selection with the
“S” key. Then select the desired value by means of the “ ” keys
(2) and confirm the value with “S”.
Go back
*
> 2s
)
a
The instrument displays the maximum and the minimum value available for the CT ratio.
VT ratio
1
2
Confir
******MMAAIINN MMEENNUU*****
MODULES SYSTEM CT RATIO
VT-PT RATIO
RREEVV.. 11..44..22..66 ((AAVV55)
******MMAAIINN MMEENNUU*****
MODULES SYSTEM CT RATIO ORTO TA
VT-PT RATIOORTO TA
RREEVV.. 11..44..22..66 ((AAVV55)
m
VT RA min max
TIO ?
a
123...
*
*
1.000
1.000 6000
)
)
Measure
> 2s
Go back
> 2s
This function allows you to select the value of the VT-PT ratio (pri­mary/secondary ratio of the voltage transformer being used). Example: if the primary of the connected VT (voltage transformer) is 20kV and the secondary is 100V, then the VT-PT ratio corresponds to 200 (obtained carrying out the following calculation: 20000/100). Choose the VT-PT RATIO function (1) by means of the “ ” key and confirm the selection with “S”. Then select by means of the “ ” key the desired value (2) and confirm the value with the “S” key.
a
The instrument displays the maximum and minimum value avail­able for the VT-PT ratio.
Page 16
14
MD/AVG calculation
D
**********MMAAIINN MMEENNUU*********
SYSTEM
T RATIO
C VT-PT RATIO
MD/AVG. CALCUL.
1
2
3
D
RREEVV.. 11..44..22..66 ((AAVV55)
******DDEEMMAANNDD//AAVVEERRAAGGEE*****
******DDEEMMAANNDD//AAVVEERRAAGGEE*****
Confirm
Programming WM5 96
TYPE
IME
T SYNC
IPO
T
IXED
F
TEMPO
SLIDE
SINCRONISM
)
*
*
*
Measure
> 2s
Go back
> 2s
Go back
> 2s
WM5-96 Instruction Manual
his function allows the user to select the calculation method of the
T DMD/AVG value of the selected variable. To access these functions select DMD/AVG CALCUL. from the main menu by means of the “
” keys and confirm the selection with “S” (1).
TYPE (2): select the type of calculation mode to be used for the
MD/AVG calculation: FIXED or SLIDE (3). Select the desired type (3)
D by means of the “ ” keys and confirm with “S”.
IME(4): select the time interval for the DMD/AVG calculation (5, 10,
T
15, 20, 30, 60 minutes). Select the desired time (5) by means of the “ ” keys and confirm with “S”.
Where: Pmax is the maximum power, Pc is the contractual power, t1 is the selected time period for the calcula­tion of the A value.
VG/DMD
FIXED SELECTION: if, for example, a time interval of 15 minutes has been selected, the instrument will calculate the AVD/DMD value of the measured variable and updates its value every 15 minutes.
******DDEEMMAANNDD//AAVVEERRAAGGEE*****
TYPE
TIME
SYNC
4
*
Go back
> 2s
SLIDING SELECTION: if for example a time interval of 15 minutes has
******DDEEMMAANNDD//AAVVEERRAAGGEE*****
TIPO
5 min
TEMPO
10 min
SINCRONISMO
15 min
5
Confirm
20 min
*
Go back
been selected, the instrument calculates the AVG/DMD value and updates its value at the beginning after the first 15 values and then after every minute, thus generating a window whose width is of 15
> 2s
minutes and that moves forward every minute.
SYNCHRONISM (6): select the synchronization mode, that is the
method that controls the calculation method of the average/demand according to the selected time. Select the type of desired synchronism
(7) by means of the “ ” keys and confirm with “S”.
******DDEEMMAANNDD//AAVVEERRAAGGEE*****
TYPE TIME
SYNC
6
*
Go back
OFF (7)
instrument on.
CLOCK (7): the synchronization starts as soon as the first selected
> 2s
integration time multiple expir time at 10 minutes and the current time at 10:25, the synchronization
: the synchronization calculation starts when you switch the
es. Example: by setting the integration
will start at 10:30.
CONTACT (7): the synchronization starts at the status modification
(from ON to OFF or from OFF to ON) of one of the digital inputs pro-
******DDEEMMAANNDD//AAVVEERRAAGGEE*****
TIPO
OFF
TEMPO
CLOCK
SINCRONISMO
ACT
CONT
7
*
Go back
grammed such as SYNCHRONIZATION or TARIFF (see “DIGITAL INPUTS”). A consequent status change produces the reset and then
> 2s
again the starting up of the synchr
onization.
Confir
m
Page 17
WM5-96 Instruction Manual
nstalled power
I
Programming WM5 96
15
This menu allows you to set a power value (installed power) that, in the measuring phase, will represent 100% of the graph indicator “W%” present in the “P00” page (in the fourth quadrant).
> 2s
Select INS. POWER ? (1) by means of the “ ” keys and confirm with the “S” key. Then enter the full scale value by means of the “
)
*
Measure
********** MMAAIINN MMEENNUU *********
CT RATIO
T-PT RATIO
V DMD/AVG CALCUL.
NSTALLED POWER
1
I
RREEVV.. 11..44..22..66 ((AAVV55)
” keys and confirm with “S”.
a
The instrument shows the maximum and minimum limit of the value available for the full scale of the W% graph.
)
*
Go back
> 2s
********** MMAAIINN MMEENNUU *********
T RATIO
C
T-PT RATIO
V
I
NS. POWER ?
MD/AVG CALCUL.
D
m
in
INSTALLED POWER
m
2
RREEVV.. 11..44..22..66 ((AAVV55)
Confirm
a
ax
9999MW
123...
000 W
1
000 W
1
Page 18
16
eters Menu
M
1
Programming WM5 96
********** MMAAIINN MMEENNUU *********
T-PT RATIO
V
MD/AVG CALCUL.
D INSTALLED POWER
ETERS
M
RREEVV.. 11..44..22..66 ((AAVV55)
)
Measure
*
WM5-96 Instruction Manual
his menu allows the user to select/set all the parameters of the ener-
T gy meters. From the main menu, select METERS by means of the “
” keys and confirm with the “S” key to enter the METERS menu.
Then choose by means of the “ ” keys if entering the program-
> 2s
ming of the TARIFF (2) or entering the “ENERGY RESET” menu to
eset the meters (16); confirm the selection with S”.
r
ARIFF(3): the instrument identifies two operating modes.
T
**********MMEETTEERRSS MMEENNUU*********
TARIFF
NERGY RESET
E TOTALIZER RESET
2
**********MMEETTEERRSS MMEENNUU*********
TARIFF
Y CLOCK
B
NERGY RESET
E
BY DIG. INPUTS
TOTALIZER RESET
3
*
*
Go Back
> 2s
Go Back
> 2s
BY DIG. INPUTS (by digital inputs) (3): the instrument manages the tariff changes by means of the status changes of the digital inputs pro­grammed as TARIFF (see the section dedicated to the digital inputs). Select BY DIG. INPUTS by means of the “ ” keys and confirm with the “S” key.
BY CLOCK (3)
: the instrument manages the tariffs by means of the clock integrated in the RS232 serial module. Select BY CLOCK by means of the “ ” keys and confirm with the “S” key, then set the following additional parameters:
WEEK TYPE (4): set the WEEK TYPE where for each day you can select if it’s to be considered a working day or a holiday. The setting of the week type is very useful to simplify the following programming of the periods.
TTIIMMEE PPEERRIIOODD SSEELLEECCTTIIOON
WEEK TYPE
PERIODS DEFAULT TARIFF
4
N
Go Back
Select WEEK TYPE with the “ ” keys and confirm with the “S” key: you enter the window with the list of days (5), select the desired day with the “ ” keys and confirm with the “S” key, then choose
> 2s
by means of the “ ” keys if the day is to be considered as a (W) WORKING day or as a (H) HOLIDAY (6), and confirm with the “S” key. Proceed as above described with the remaining days. To conclude the operation and go back to step number (4), press the “S” for at least 2 seconds.
TTIIMMEE PPEERRIIOODD SSEELLEECCTTIIOON
SETTIMANA TIPO
MONDA
PERIODI
TUESDA
TARIFFA DEFAULT
WEDNESDAY (W)
5
THURSDAY (W)
(W)
Y
(W)
Y
N
Go Back
> 2s
TTIIMMEE PPEERRIIOODD SSEELLEECCTTIIOON
SETTIMANA TIPO
LUNEDI (W)
PERIODI
(W) WORKING
TEDI
MAR
TARIFFA DEFAULT
(H) HOLIDAY
MERCOLEDI (W)
6
Confirm
THURSDAY (W)
(W)
N
Go Back
> 2s
Warning: for the management of the tariffs with the “BY CLOCK” option, when counting the ener
gy with WM5, the AR1039 (RS232 + RTC) option is to be installed. The tariff management “BY CLOCK” is however possible also without this module, but in this case the “CLOCK” function, instead of the “METERS” function is to be select ed, and then the subfunction “WITHOUT BACKUP”: in this case,
-
remember that if, for any reason, the instrument is switched off, the current DATE and TIME will not be stored. These data will have to be set again every time the instrument is switched on.
Page 19
WM5-96 Instruction Manual
Programming WM5 96
17
TTIIMMEE PPEERRIIOODD SSEELLEECCTTIIOON
WEEK TYPE
ERIODS
P
DEFAULT TARIFF
7
N
Go back
PERIODS (7): set the periods to which the tariffs are to be coupled. It’s
ossible to divide each day into up to 24 periods. Up to 100 lines of
p period selection can be entered. Select PERIODS by means of the
> 2s
” keys and confirm with the “S” key: a table (8) will be dis­played which will be empty at the beginning and where the various programmed periods are to be entered. To set the START TIME of the
eriod (9), press the “S” key; then, by means of the “ ” keys,
p
time day
8
9
-- -- --/-- --/-- --
time day
-- -- --/-- --/-- --
START TIME ?
W H T
W H T
0
123...
0
Go back
Go back
select the time from 00 (00 AM) to 23 (11 PM); press “S” to confirm the
elected time and go to the setting of the “END TIME”: select the
s desired time by means of the “ ” keys, then press “S”.
> 2s
Afterwards, set the START DATE of the period by means of the “ ” keys and confirm with the “S” key, then set the END DATE of the period by means of the “S” keys and confirm with “S”. Once the time data have been set, the instrument will display a window (10) from
> 2s
which the period type can be selected, that is either WORKING, HOL­IDAY or ALL (the ALL selection will not make any difference between WORKING day and HOLIDA
Y, that is the default TARIFF will be
applied): select the desired mode by means of the “ ” keys and
END TIME ?
01
123...
confirm with “S”. As last setting, choose the TARIFF (11) to be coupled to the period, use the “ ” keys to select the desired tariff from 1 to 12 and confirm with the “S” key (up to 12 tariff if the management
S
TART TIME ?
0
123...
0/00
is carried out by means of internal clock, up to 4 tariffs if the manage­ment is carried out by means of the digital inputs). A new line (12) relevant to the period which has just been programmed
END TIME ?
time day
10
-- -- --/-- --/-- --
WORKING
HOLIDAY
ALL
00/00
123...
W H T
Go back
> 2s
will be added to the table: it’ll be possible to add up to 99 further lines. The following options (13) are also possible:
INSERT: enter a new line (new period); follow the instructions given from step (9) to step (11).
MODIFY: modify an existing period. Select the period to be modified by means of the “ ” keys, press the “S” key to display the rele-
Go back
time day
1
1
-- -- --/-- --/-- --
TARIFF ?
time day
00 01 01/01 01/01 01
W H T
1
123...
Go back
W H T
vant window (13), then choose MODIFY by means of the “ ” keys and confirm the choice with the “S” key. Afterwards follow the setting procedure as described from step (9) to step (11).
> 2s
DELETE: delete an existing period. Select the period to delete with the “ ” keys, press the “S” key to display the relevant window (13) then choose DELETE by means of the “ ” keys and confirm with the “S” key. Before deleting the period, the instrument displays the fol-
> 2s
lowing message: PAY ATTENTION, CURRENT LINE WILL BE DELET­ED!!!! WILL YOU CONTINUE? Choose YES to execute the command
12
line : 001
or NO to cancel the deleting.
DELETE ALL: delete all the existing periods. Press the “S” key to dis-
time day
INSERT
00 01 01/01 01/01 01
MODIFY DELETE
13
DELETE ALL
line : 001
W H T
Go back
play the r the “ ” and confirm with the “S” key. Before deleting all the peri­ods, the instrument displays the following message: PAY ATTENTION,
> 2s
ALL LINES WILL BE DELETED!!! WILL YOU CONTINUE? Choose YES to execute the command or NO to cancel the deleting.
elevant window (13), then choose DELETE ALL by means of
Page 20
18
Programming WM5 96
WM5-96 Instruction Manual
TTIIMMEE PPEERRIIOODD SSEELLEECCTTIIOON
EEK TYPE
W PERIODS
EFAULT TARIFF
D
14
N
Go back
DEFAULT TARIFF (14): it is the tariff (from 1 to 12) which will be cou-
led to the meters if no other programming is made or for the days
p (periods) set as non-working. Select DEFAULT TARIFF with the “
> 2s
” keys and confirm with the “S” key. Then set by means of the “ ” keys the tariff (15) that will be set as DEFAULT TARIFF, and confirm with “S”.
TTIIMMEE PPEERRIIOODD SSEELLEECCTTIIOON
EEK TYPE
W PERIODS
T
ARIFF ?
EFAULT TARIFF
D
1
N
Go back
ENERGY RESET (16): reset the ENERGY METERS choosing among:
OTAL, PARTIAL: resets all energy meters, both total and partial.
T TOTAL +: resets the total meters of imported energy.
> 2s
TOTAL -: resets the total meters of exported energy. PARTIAL +: resets the partial meters of imported energy.
15
Confirm
123...
PARTIAL -: resets the partial meters of exported energy. To reset the meters, select ENERGY RESET from the METERS MENU (16) by means of the “ ” keys and confirm with “S”; then select the type of RESET to be carried out by means of the “ ” keys and confirm
**********MMEETTEERRSS MMEENNUU*********
ARIFF
T
ENERGY RESET
OTALIZER RESET
T
6
1
********RREESSEETT MMEETTEERRSS*******
TOTAL, PARTIAL
TOTAL + TOTAL ­PARTIAL +
17
PARTIAL -
Confirm
*
*
Go back
> 2s
Go back
> 2s
with “S”. Befor
e carrying out the selected reset, the instrument displays the following message: PAY ATTENTION, THE SELECTED METERS WILL BE RESET! WILL YOU CONTINUE?: choose YES to proceed with the reset or NO to cancel it.
TOTALIZER RESET (18): reset the TOTALIZERS choosing among:
**********MMEETTEERRSS MMEENNUU*********
TARIFF ENERGY RESET
TOTALIZER RESET
*
Measure
> 2s
ALL TOTALIZER (19): reset all the totalizers. The other strings identify the totalizers set before that you want to reset individually (GAS, HOT H2O, COLD H2O, REMOTE HEATING). To reset the totalizers, select TOTALIZER RESET from the METERS
18
MENU (20) by means of the “ ” keys and confirm with “S”; then select the type of RESET to be carried out by means of the “
keys and confirm with “S”. Before carrying out the selected reset, the
******RREESSEETT TTOOTTAALLIIZZEERR*****
ALIZER
ALL TOT
(C1) REMOTE (C2) REMOTE
19
20
(C3) GAS
******RREESSEETT TTOOTTAALLIIZZEERR*****
PPAAYY AATTTTEENNTTIIOONN,, TTHHE
ALL TOTALIZER
SSEELLEECCTTEEDD TTOOTTAALLIIZZEERRS
(C1) REMOTE
WWIILLLL BBEE RREESSEETT!
(C2) REMOTE
WWIILLLL YYOOUU CCOONNTTIINNUUEE?
(C3) GAS
NNO
O
m
Confir
!
YYEES
Go back
*
Go back
*
E
S
?
S
instrument displays the following message: PAY ATTENTION, THE SELECTED TOT choose YES to proceed with the reset or NO to cancel it.
> 2s
> 2s
ALIZERS WILL BE RESET! WILL YOU CONTINUE?:
Page 21
WM5-96 Instruction Manual
rogramming example of the tariff
P parameters
In the “time periods” table you have at least 100 selection rows available.
Programming WM5 96
19
Let’s for example take into consideration the following tariff plan decided by the energy supplier for December. The working week is from Monday to Friday, while the weekend is made by Saturday and Sunday. During the working days there are two different tariffs with the following time periods: TARIFF 4 (T4) from 8
m to 5 pm, and TARIFF 3 from 5 pm to 8 am., while during the week
a end there is only one time period and therefore a single TARIFF (T2). The first thing to do is to program the week type (4 - 6), the working days will be indicated as follows: Monday (W), Tuesday (W), Wednesday (W), Thursday (W), Friday (W). While the weekend (Holidays) will be indicated as: Saturday (H), Sunday (H). As a conse­quence, the time periods relating to the working weeks will be divided into the following lines (8 - 13): First line: START TIME = 00, END TIME = 8, START DATE = 01/12, END DATE = 31/12, then select “WORKING” and TARIFF “3”. Second line: START TIME = 8, END TIME = 17, START DATE = 01/12, END DATE = 31/12, then select “WORKING” and TARIFF “4”. Third line: START TIME = 17, END TIME = 24, START DATE = 01/12, END DATE = 31/12, then select “WORKING” and TARIFF“3”. Moreover, a 4th line will be necessary to identify the NON-WORKING days other than the week end (Saturday and Sunday): for example December 25th and 26th. Fourth line: START TIME = 00, END TIME = 24, START DATE = 25/12, END DATE = 26/12, then select “HOLIDAY” and TARIFF “2”.
time day
00 08 01/12 31/12 03 08 17 01/12 31/12 04 17 24 01/12 31/12 03
00 24 25/12 26/12 02
line : 004
ENERGY SUPPLIER SETTING WM5
Month Week Time Type
om
fr
Y
MONDA
DECEMBER
to
FRIDAY
W H T
from 8
to 17
from 17
to 8
Working
The periods are now programmed and the instrument will display the window on the left.
The periods programmed as above only refer to the month of December, therefore it’s necessary to set a tariff referring to the remaining time period, which will be referred to as “T1”. In this case you can refer to the default tariff (14) set as TARIFF “1”, so that the instrument will automatically r gramming is pr
Selected
Tariff
4 00 08 01/12 31/12 W 3
3
Start time End time Start date End date Type Tariff
esent.
8 17 01/12 31/12 W 4
17 24 01/12 31/12 W 3
efer to TARIFF “T1” when no other pro-
Rest of the
year
fr
toSUNDA
All H24 -- 1 Default tarif
H24 Holiday 2 00 24 25/12 26/12 H 2
Y
Y
TURDA
om SA
f 1
Page 22
20
isplay Page
D
D I M
1
RREEVV.. 11..44..22..66 ((AAVV55)
Programming WM5 96
**********MMAAIINN MMEENNUU*********
MD/AVG CALCUL.
NSTALLED POWER
ETERS
DISPLAY PAGE
)
*
Measure
> 2s
WM5-96 Instruction Manual
This function allows the user to select the variables to be displayed on
age 00. Choose the DISPLAY PAGE function (1) by means of the “
p
keys and confirm with “S”, then by means of the “ ” keys
” select the display line where the selected variable (2) is to be displayed and press “S”. Afterwards select the variable to be displayed by means of the “ ” keys among those displayed by the instrument
(3) and confirm with “S”.
NOTE: only the page “00” can be free configured.
2dmd
*
*
Go back
> 2s
Go back
> 2s
************MMAAIINN PPAAGGEE***********
W
1dmd
2
************MMAAIINN PPAAGGEE***********
3
Confirm
w3dmd
W1dmd W2dmd
W3dmd
W
Page 23
WM5-96 Instruction Manual
igital Inputs
D
************MMAAIINN MMEENNUU***********
NSTALLED POWER
I
ETERS
M
ISPLAY PAGE
D
1
2
3
DIGITAL INPUTS
RREEVV.. 11..44..22..66 ((AAVV55)
******DDIIGGIITTAALL IINNPPUUTTSS*****
N C1
I
IN C2
N C3
I
E
RREEMMOOTTE
******DDIIGGIITTAALL IINNPPUUTTSS*****
NG. C1
I
REMOTE
ING. C2
SYNC
NG. C3
I
RIFF LSb
TA
O
TARIFF MSb
RREEMMOOTTO
OTALIZER
T
)
*
*
*
Measure
> 2s
Go back
> 2s
Go back
> 2s
Programming WM5 96
his function allows the operating modes selection of the digital
T
nputs. Choose the function DIGITAL INPUTS (1) by means of the “
i
21
” keys and confirm with the “S” key. (2) Select the digital input (ex.: A1=digital input 1 of slot A) by means of the “ ” keys, the instru­ment displays the function of the selected input (the inputs are set on REMOTE as default), press “S” to confirm the selected digital input. (3) From the list, select the operating mode to be coupled to the selected digital input by means of the “ ” keys, the functions are:
REMOTE (3): the status of the digital inputs is displayed from the
instrument and retrasmitted (if requested) by the serial communication. SYNC (SYNCHRONISM) (3): synchronises the calculation of the “dmd” power by means of an external signal. TARIFF LSb and MSb (3): This function allows the digital inputs to control the tariffs in the energy metering. The combination of the input status manages the tariff change as specified in the table (18-1). If only the “MSb” (one digital input only) is selected, then the energy meter­ing is possible with a maximum of 2 tariffs: the tariff change will only be possible between tariff 1 and tariff 3. If only the “LSb” is selected, then the tariff change occurs only between tariff 1 and tariff 2. The use of both digital inputs (MSb + LSb) allows the management of tariffs 1, 2, 3 and 4. Press “S” to confirm the selection. Proceed, if necessary, with the configuration of the other digital inputs.
4
GAS
4a
HOT H2O COLD H2O REMOTE HEAT
m3/PULSE
4c
min max
******TTOOTTAALLIIZZEERR MMEENNUU*****
TYPE
ENG. UNIT PULSES
4b
m3
Gal Lit dm3
Lit
10000s
0.001 1000s
s
dm3 Ft3 kWh
*
Go back
> 2s
MSb LSb
off off
off on
on off
on on
(18-1)
Tariff
1
2
3
4
TOTALIZERS (3): This function allows the digital inputs to totalize the
gas, water (hot or cold) and the remote heating. From the totalizer menu select: TYPE (4), to choose what the instrument has to totalize
(4a) GAS, HOT H2O (hot water), COLD H2O (cold water) or REMOTE
HEAT: in this case the WM5 counts the hot water by the digital input
ough an exter
thr
nal water counter but on the display the r
elevant value is reported in “kWh”. ENG. UNIT (4b), to choose the correct engeegniring unit (4b) to associate to the totalizer selected befor
e. PULSES (4c), to set the pulse weigth (4c), that is for each input pulse the totalizer increases in the chosen value.
Weight of pulse selected: 10.
Input pulse
0 10 20 30 ...
Page 24
22
igital outputs
D
************MMAAIINN MMEENNUU***********
ETERS
M
ISPLAY PAGE
D
IGITAL INPUTS
D
1
2
3
DIGITAL OUTPUTS
RREEVV.. 11..44..22..66 ((AAVV55)
******DDIIGGIITTAALL OOUUTTPPUUTTSS*****
AALLAARRMM
D
NND
******DDIIGGIITTAALL OOUUTTPPUUTTSS*****
AALLAARRMM
D
NND
Programming WM5 96
his function allows the selected function to be coupled to the select-
T ed digital output: pulse, alarm, remote control. Select the DIGITAL OUTPUT function (1) by means of the “ ” keys and confirm with “S”. (2) Select the digital output (ex.: D0=digital output 0 of slot D) by
> 2s
means of the “ ” keys, the instrument displays the function of the
elected output (a), press “S” to confirm the selected digital output (3)
s select the operating mode to be coupled to the selected digital output
y means of the “ ” keys. The functions are the following:
b
PULSE (3): the measured energy is retransmitted by the digital output
by means of pulses. Choose the PULSE function by means of the “
” keys and confirm with “S”. Then select the type of energy to be
retransmitted: select ENERGY TYPE by means of the “ ” keys
> 2s
and confirm with “S”, the instrument displays the list of energies (4a), (kWh +, kWh-, kvarh +, kvarh-): select the desired energy by means of the “ ” keys and confirm with the “S” key: for each energy it’s possible to r from T1 to T12 (4b), choose the desired energy detail by means of the “
” keys and confirm with the “S” key. At the end of the proce­dure, the value of the pulse associated to the energy measured by the instrument is to be decided, that is the number of pulses generated by
> 2s
the digital output for each measured kWh+ (or kWh- or kvarh+ or kvarh-). Choose PULSES (4) by means of the “ ” keys and con­firm with the “S” key, then (4c) set the desired value of the pulses by means of the “ ” keys, confirm the value with the “S” key.
UT D0
O
UT D1
O
UT D2
O
a
NG. C1
I
PULSE
ING. C2
LARM
A
ING. C3
EMOTE CONTR.
R
)
*
*
*
Measure
Go back
Go back
WM5-96 Instruction Manual
etransmit the total energy metering (TOT) or the tariffs
4
PULSE /kWh min max
4c
********PPUULLSSEE MMEENNUU*******
ENERGY TYPE
PULSES
.0000
1
0.0010 10000
123...
4a
4b
*
kWh +
kWh ­kvarh + kvarh -
TOT
L1 L2 L3
Page 25
WM5-96 Instruction Manual
******DDIIGGIITTAALL OOUUTTPPUUTTSS*****
NG. C1
I
ULSE
P
ING. C2
a
5
ALARM
ING. C3
EMOTE CONTR.
R
M
5
6
AALLAARRM
D
NND
******DDIIGGIITTAALL OOUUTTPPUUTTSS*****
UT D0
O
OUT D1 OUT D2
M
AALLAARRM
D
NND
Programming WM5 96
LARM(5): the digital output is enabled only if the expected alarm
A
23
status occurs. Select the ALARM function by means of the “
*
keys and confirm with the “S” keys (5a).Then select “ND” (normally de-energized relay) or “NE” (normally energized relay) (5b) by means of the “ ” keys. To program the values of the set-points refer to
he “alarm” menu.
t REMOTE CONTR. (5): the digital output can be enabled by means of a
ommand sent by means of serial port. Select the REMOTE CONTR. func-
c tion by means of the “ ” keys, then press the “S” key to confirm.
D
N
b
5
NE
Proceed, if necessary, with the configuration of the other digital out­puts (6).
*
The digital outputs highlighted in the below mentioned table, ar
e subject to
a self test that implies their brief acti­vation when the instrument is switched on. Their use as pulse outputs is ther
efore not
recommended.
Code Description Slot A Slot B Slot C Slot D
AO1058 1 r
elay output
A0 B0 C0 D0
AO1059 1 open coll. output A0 B0 C0 D0
AO1035 2 relay outputs A0 A1 B0 B1 C0 C1 D0 D1
AO1036 2 open coll. outputs A0 A1 B0 B1 C0 C1 D0 D1
AO1037 4 open coll. outputs A1 A2 A3 A4 B1 B2 B3 B4 C1 C2 C3 C4 D1 D2 D3 D4
(19-1)
Page 26
24
larm
A
**********MMAAIINN MMEENNUU*********
ISPLAY PAGE
D
IGITAL INPUTS
D DIGITAL OUTPUTS
LARMS
1
2
3
3a
4
4a 4c 4b
5
SET 2 ? min max
5a
A
RREEVV.. 11..44..22..66 ((AAVV55)
**********AALLAARRMMSS MMEENNUU*********
ALARM 01 (OFF)
A A ALARM 04 (ON) ALARM 05 (ON)
**********AALLAARRMMSS MMEENNUU*********
AALLAARRMM:: 001 EENNAABBLLEE:: OOFFF
OFF
ON
**********AALLAARRMMSS MMEENNUU*********
AALLAARRMM:: 001 LLAATTCCHH:: OOFFF
OFF
ON
**********AALLAARRMMSS MMEENNUU*********
AALLAARRMM:: 001 DDEELLAAYY OONN 22:: 00..00000
Programming WM5 96
LARM 02 (ON) LARM 03 (ON)
ENABLE
ARIABLE
V
YPE
T
UP
3b
DO IN OUT
LA
DISABLE SET 1
SET 1 ? min max
SET 2
OUT DELAY ON
0.0000.
-9999G 9999G
DELAY ON ? min max
1
TCH
1
F
1
a
F
3c
9.0000.
-9999G 9999G
0
5b
00000s 00000s 255
00s
*
)
*
*
V1
A1 W1 var1
*
*
OUT D0
OUT D3 NONE-FFT
5c
Measure
> 2s
Go back
> 2s
OFF
ON
WM5-96 Instruction Manual
his function allows you to set the alarm parameters. The instrument is
T able to manage up to 16 alarms (real or virtual). Select the ALARMS func­tion (1) by means of the “ ” keys and confirm with “S”. Then select the alarm to be programmed (2) by means of the “ ” keys and con­firm with the “S” key, then set the following parameters:
ENABLE (3): enable (ON) or disable (OFF) the alarm; the instrument dis-
lay the existing programming (a). Select the ENABLE function (3) by
p means of the “ ” keys and confirm with the “S” key. Then select ON
(3a) to enable or OFF (3a) to disable the alarm by means of the “
keys, then confirm with the “S” key.
VARIABLE (3): set the variable to be coupled to the alarm. Select the VARI-
ABLE submenu (3) by means of the “ ” keys and confirm with the “S” key. Then select the variable to be coupled to an alarm (the list of variables depends on the selected system) (3c) by means of the “ ” keys, then select with the “S” key
.
TYPE (3): set the operating mode of the alarm. UP: up alarm, the alarm
will be enabled if the measured value exceeds the set-point. DO: down alarm, the alarm will be enabled if the measured value goes below the set point. IN: internal window alarm, the alarm will be activated if the value is brtween the two set points. OUT: the alarm will be activated if the value is outside the two set points. See details on next page “Logic and alarm parameters”. Choose the TYPE (3) function by means of the “ ” and confirm with “S”. Then, select by means of the “ ” keys the desired alarm mode (3b) and confirm with “S”.
LATCH (4): set the latch function. The alarm will remain ON even if the
cause that has generated it is not present any more. The alarm can be reset only manually after the user has noticed it. Choose the LATCH func­tion by means of the “ ” keys and confirm with “S”. Then select ON
(4a) by means of the “ ” keys to enable the latch or OFF (4a) to dis-
able it, then confirm with “S”. DISABLE: set the DISABLE function. When the instrument is switched on, the first alarm condition will be ignored (useful also in case of DOWN alarm). Choose the DISABLE function by means of the “ ” keys and confirm with “S”. Then select ON
to enable it or OFF
(4b)
(4b) to disable
it by means of the “ ” keys and confirm with “S”. SET 1 (Set point 1) (4): set the first alarm set point of the variable. Choose SET 1 (4c) by means of the “ ” keys and confirm with “S”. Then set the value by means of the “ ” keys and confirm with “S”. SET 2 (5): set the second alarm set point of the variable. Choose SET 2
(5a) by means of the “ ” keys and confirm with “S”. Then set the
value by means of the “
” keys, then confirm with “S”.
OUT (5): select the output to be enabled in case of alarm. The list will dis-
play all the outputs that in the menu “Digital outputs” have been set as “alarm”. Choose the OUT function by means of the “
” keys and confirm with “S”. NONE-FFT, virtual alarm with trigger function for sig­nal FFT analysis. The analysis relevant to the 10s that precede the alarm event will be stored (FIFO) and available for the remote down­loading by means of a specific software PowerSoft.Then select by means of the “ ” keys the desired output (5b) then confirm with “S”. DELAY ON (5): set a delay on activation of the alarm. Choose the func­tion DELA
Y ON by means of the “ ” keys and confirm with “S”. Then set the value in seconds (5c) by means of the “ ” keys quindi confermare con “S”.
Y OFF
DELA
Y OFF function by means of the “ ” keys and confirm with “S”.
DELA Afterwar
ds set its value in seconds
: set a delay on deactivation of the alarm. Choose the
(6)
by means of the “
(6a)
” keys
then confirm it with “S”.
Page 27
WM5-96 Instruction Manual
8
8
**********AALLAARRMMSS MMEENNUU*********
ELAY ON
D
DELAY OFF
UNCTION
F
6
AALLAARRMM:: 001 DDEELLAAYY OOFFFF:: 00 s
DELAY OFF ?
6a
m
in
max
1
0000s
0
0000s
0
5500s
2
Logic and alarm parameters
Programming WM5 96
25
FUNCTION (6): set its OR or AND logic. Choose FUNCTION by means
f the “ ” keys and confirm with “S”. Afterwards select by the
*
o “ ” keys the logic function to be coupled to the OR (6b) or AND
(6b) alarm, then confirm with “S”. See details in “Logic and alarm
parameters”. Proceed, if necessary, with the configuration of the other
s
alarms (up to a maximum of 16 alarms).
To exit the menu “digital outputs” keep the “S” key pressed for at least 2
econds.
s
OR
6b
ND
A
Each symbol includes all the settings described in the “alarm” paragraph and listed on the right:
ET 1
S
SET 2
UP alarm
SET 1
SET 2
Example of AND/OR logic alarm:
AND OR
A: AND
B: AND
C: AND
- Enable.
- Variable
- Type
- Latch
- Disable
DOWN alarm
- Set 1
- Set 2
- OUT
- Delay on. Delay off.
- Function (and/or)
SET 1
SET 2
In-window alarm
Alarm is on when the value
SET 1 and SET 2
A: OR
B: OR
C: OR
is between
A, B, C... up to 16 locks to control parameters.
SET 1
SET 2
function enabling
m
Alar
Ext. window alarm with
disabling at power on
Alarm is on when value
exceeds SET 1
or goes below SET 2
A: OR
B: OR
C: AND
D: AND
OR+AND
A: AND
A: AND
B: AND
B: AND
C: AND
C: AND
A: OR
B: OR
C: OR
A: OR
B: OR
C: OR
A: OR
B: OR
C: AND
A: OR
B: OR
C: AND
D: AND
D: AND
Page 28
26
xample of alarm parameters
E
Programming WM5 96
programming
WM5-96 Instruction Manual
The enabling of an alarm is required when the system voltage VL1-N exceeds or is below the range 215V - 235VAC. The external window alarm is selected so that the output is enabled when the measured value exceeds 235V or is below 215V.
235V
15V
2
nabling
E
300kW
295kW
e suggest you to use a delay of a
W few seconds when disabling the alarms in or
der to avoid the conse quence of a switching on and off of the output and the subsequent damaging of the contacts due to fluctuations of the measur
ed signal with a value close to the selected alarm set­points.
Here below you will find the recommended programming:
Enabling of one of the 16 alarms (alarm example 01 ON)
-
- Choose the variable to be monitored: VL1-N= V1
- Choose the type of desired alarm: OUT
- Choose if the latch is to be enabled or not: OFF
- Choose if enabling or not the disabling of the first alarm status from the switching on of the instrument: ON.
- Set set-point 1: Set 1 = 235V
- Set set-point 2: Set 2 = 215V Choose to which digital output the alarm you are programming is to
­be addressed (the “C0” digital output is to be previously enabled to the “alarm” function; in the same menu it’s possible to select the desired type of output: “ND or NE”).
- Should a DELAY ON (delay on activation) be required, set the desired number of seconds: “5 seconds”.
- Should a DELAY OFF (delay on deactivation) be required, set the desired number of seconds: “5 seconds”
- Choose the kind of logic with which the alarm is to be treated: “OR” (see examples of logic alarm: AND/OR).
The disconnection of a load when a set value of absorbed power is required. For example when 300kW are exceeded, the alarm occurs and a set load is disconnected. An UP alarm is selected. Below you’ll find the recommended programming:
- Enabling of one of the 16 alarms (example: alarm 02 ON).
- Choose the variable to be monitored: W system (W)
ed alarm: “UP”
Choose the type of r
-
equir
- Choose if the latch is to be enabled or not:“OFF”
- Choose if the disabling of the first alarm status from the switching of the instrument is to be enabled or not: “OFF”
- Set set-point 1: Set 1 = 300kW
- Set set-point 2: Set 2 = 295kW
- Choose to which digital output the alarm you are programming is to be addr
essed: “D0” (the “D0” digital output is to be previously enabled to the “alarm” function; in the same menu it’s possible to select the desired type of output: “ND or NE”). Should a DELAY ON (delay on activation) be required, set the
­desired number of seconds: “5 seconds”.
- Should a DELAY OFF (delay on deactivation) be required, set the desired number of seconds: “5 seconds”
- Choose the kind of logic with which the alarm is to be treated: “OR”
­(see examples of logic alarm: AND/OR).
Page 29
WM5-96 Instruction Manual
Alarm 01
Programming WM5 96
The instrument is able to manage 2 types of alarms:
1. Real alarm = when the alarm is connected to a digital output (relay output or open collector output). The activation of a real alarm causes the switching at fixed light of the “AL” LED on the
ront of the instrument.
f
2. Virtual alarm = when the alarm is not connected to any output. The ctivation of a virtual alarm causes the switching at blinking light of the
a “AL” LED. In case a virtual alarm and a real alarm occur at the same time, the latter has the prevalence over the control of the above men­tioned LED (therefore it will be ON with a fixed light). By exploiting the OR and AND functions, it’s possible to connect together more than one virtual alarm and direct them together to one digital output (relay output or open collector output). In any case the limit of 16 program­mable alarms will still be valid.
27
Alarm 02
The above drawing explains the operation of the alarms described in the example. Having the “OR function been cho­sen for both the alarms, the operation of the digital output can be r
epresented by two contacts in parallel. The enabling of
only one of the two alarms will enable also the digital output.
Page 30
28
Events selection
************MMAAIINN MMEENNUU***********
IGITAL INPUTS
D
IGITAL OUTPUTS
D ALARMS
VENTS SELEC.
1
2
2a
2b
2c
E
RREEVV.. 11..44..22..66 ((AAVV55)
EEVVEENNTTSS SSEELLEECCTTIIOONN MMEENNU
ALARMS
IN
M
AX
M DMD MAX DIG. INPUTS
********AALLAARRMM EEVVEENNTTSS*******
ALARM 01 (OFF)
LARM 02 (ON)
A ALARM 03 (ON) ALARM 04 (ON) ALARM 05 (ON)
V1 (OFF)
v2 (ON) V3 (ON)
VL (ON) V12 (ON)
REMOTE (OFF)
SYNC (ON)
ARIFF (ON)
T
Programming WM5 96
his sub-menu allows the user to select which events are to be
T stamped. Choose the function EVENTS SELEC. (1) by means of the
)
*
U
*
Measure
Measure
Go back
Go back
Go back
” keys and confirm with “S”. Afterwards proceed as follows:
ALARMS (2): select if the alarm events are to be stamped. Choose
alarms by means of the “ ” keys and confirm with “S”, then ,
> 2s
larm by alarm by means of the “ ” keys, decide whether the
a stamping is to be enabled “ON” or disabled “OFF”, by pressing the
S” key.
MIN, MAX, DMD MAX (2): select whether to store the minimum, max­imum and dmd values for each variable measured by the instrument. Select the minimum, maximum or dmd maximum by means of the “
” keys and confirm with “S”, then select, variable by variable if the
> 2s
storing is to be carried out (2b), choose by means of the “ ” keys the desired variable and decide whether to enable the storing “ON” or disable the storing “OFF” by pressing the “S” key.
DIG. INPUTS (2)
: select whether to store the status variables of the digital inputs. Choose DIG. INPUTS by means of the “ ” keys and confirm with “S”, then select one by one the digital inputs of which the storing is required (2c), choose by means of the “ ” keys the type of digital inputs (remote, synchronism, tariff), then confirm whether to
> 2s
enable the storing “ON” or disable it “OFF” by pressing the “S” key. DIG. OUTPUTS (3): select whether to store the status variations of the digital outputs or not. Select the digital outputs by means of the “ ” keys and confirm with “S”, then confirm whether to enable the storing “ON” or disable it “OFF” by pressing the “S” key.
RESET (3): select whether the resets which have been carried out are
to be stored or not. Choose RESET by means of the “ ” keys, then select which resets are to be stored (3b), by means of the “
> 2s
keys and select whether to enable the storing “ON” or disable it “OFF” by pressing the “S” key.
> 2s
WM5-96 Instruction Manual
EEVVEENNTTSS SSEELLEECCTTIIOONN MMEENNU
MAX DMD MAX DIG. INPUTS
DIG. OUTPUTS
3
3a
3b
RESET
REMOTE
TOT. PAR. (OFF)
TOTAL + (ON) TOTAL - (ON) PARTIAL + (ON) PARTIAL - (ON)
PARTIAL + (OFF) PARTIAL - (ON) MIN - MAX DMD MAX LATCH (ON)
TOT
(OFF)
ALIZERS
(ON)
(ON)
(ON)
U
Go back
> 2s
Go back
> 2s
Go back
> 2s
Page 31
WM5-96 Instruction Manual
Analog outputs
************MMAAIINN MMEENNUU***********
IGITAL OUTPUTS
D ALARMS EVENTS SELEC.
NALOG OUTPUTS
1
2
3
V1
A1 W1 var1
A
RREEVV.. 11..44..22..66 ((AAVV55)
****AANNAALLOOGG OOUUTTPPUUTT MMEENNUU*
OUT A0
UT A1
O
UT B
O
OUT B1 OUT C0
****AANNAALLOOGG OOUUTTPPUUTT MMEENNUU*
V
MIN OUT MAX OUT
OOUUTT AA0 VVAARRIIAABBLLEE:: VV1
3a
MAX OUT ? min max
0
ARIABLE
0
3c
1
M
IN OUT ?
m
in
max
0.0000.
-9999G 9999G
Programming WM5 96
29
This submenu allows the programming of the analog outputs (0-20mA,
-10V, -5/+5mA). Select the function ANALOG OUTPUTS (1) by means
0
f the “ ” and confirm with “S”. Then select the output to be pro-
o
*
Measure
grammed by means of the “ ” keys (2) and confirm with “S”;
> 2s
afterwards set the following parameters:
VARIABLE (3): select the variable to be retransmitted by means of the
)
analog output. Choose the function “VARIABLE” (3) by means of the “ ” keys and confirm pressing the “S” key. Then select by the “
” keys the variable to be retransmitted (the list of the variables
*
depends on the selected system) (3a) then confirm with “S”. MIN OUT (3): set the value expressed as % of the output range (0­20mA, 0-10V, -5/+5mA) to be coupled to the minimum measured value. The instrument displays also the maximum and minimum value which can be set (min, max). Select the function “MIN OUT” by means of the “ ” keys and confirm with “S”. Then set the value (3b) by means of the “ ” keys and confirm with “S”.
MAX OUT (3): select
the value expressed as % of the output range (0-20mA, 0-10V, -
*
5/+5mA) to be coupled to the maximum measured value. The also instrument displays the maximum and minimum value which can be set (min, max). Select the MAX OUT function by means of the “ ” keys and confirm with “S”. Afterwards set the value (3c) by means of the
a
” keys and confirm with “S”. MIN IN (4): minimum value of the variable input range to which the “MIN OUT” value, retransmitted by the analog output, will be coupled. The instrument also displays the maximum and minimum value which
3b
0.0000.
-9999G 9999G
can be set (min, max) . Choose the function “MIN IN” by means of the “ ” keys and confirm with “S”. Then set the value (4a) by means of the “ ” keys, then confirm with “S”. MAX IN (4): maximum value of the variable input range to which the “MAX OUT” value, retransmitted by the analog output, will be coupled. The instrument also displays the maximum and minimum value which can be set (min, max). Choose the MIN IN function by means of the “
” and confirm with “S”. Then set the value (4b) by means of the
” keys and confirm with “S”.
4
MAX IN ? min max
****AANNAALLOOGG OOUUTTPPUUTT MMEENNUU*
MAX OUT MIN IN
MAX IN
0
OOUUTT AA0 MMAAXX IINN:: 00..00000
4b
0.0000.
-9999G 9999G
0
4a
MIN IN ? min max
a
*
0.0000.
-9999G 9999G
The instrument can manage the fol­lowing analog outputs combinations:
Max n. 8 0-10VDC outputs; Max n.8 5/+5mADC outputs. Any combination of the two above mentioned types considering that each module manages up to 2 outputs. Max n.4 0/20mADC outputs. Max n.4 0/20mADC outputs + max n.4 0-10VDC outputs.
Proceed, if possible, with the configuration of the other analog outputs (up to a maximum of 8 outputs). To exit the menu “analog outputs”, keep the “S” key pressed for at least 2 seconds.
Page 32
30
Programming examples of the analog outputs
Programming WM5 96
WM5-96 Instruction Manual
ower retransmission by means of a 0-20mA analog output.
P
It’s necessary to measure a consumed power up to 100kW and retransmit this value by means of a signal from 4 to 20 mA: the mod­ule to be used is AO2050 (2x from 0 to 20mA), the instrument is to be programmed as follows:
VARIABLE: W(system active power). MIN OUT: 20.0% means 4 mA the calculation to be carried out is the fol-
lowing: (100*minimum output) / fullscale output =100*4mA/ 20mA=20%. MAX OUT: 100.0% means 20mA, the calculation to be carried out is: (100*maximum output) / fullscale output = 100*20mA/20mA= 100. MIN IN: 0,0k; the multiple k,M,G can be selected on the instrument according to the chosen VT and CT values. MAX IN: 100.0k; the k, M, G multiples can be selected on the instru­ment according to the selected VT and CT values.
Transmission of active consumed and generated power by means of -5/+5mA analog output.
It’s necessary to measure both the consumed active power up to 100kW and the generated active power up to -100kW and retransmit this value by means of a signal from -5 to +5mA: the module to be used is AO2052 (2x from -5/+5mA), the instrument is to be pro­grammed as follows:
VARIABLE: W(system active power). MIN OUT: -100% means -5mA, the calculation to be carried out is:
(100*minimum output)/ fullscale output= 100*5mA/-5mA=-100%. MAX OUT: 100,0% means 20mA, the calculation to be carried out is: (100*max output)/ fullscale output= 100*5mA/5mA= +100%. MIN IN: -100,0k; the multiples k,M and G can be selected on the instrument on the basis of the VT and CT values. MAX IN: 100,0k; the multiples K, M and G can be selected on the instrument depending on the VT and CT values being selected.
Retransmission of the POWER FACTOR (PF) by means of the 0-20mA analog output.
It’s necessary to retransmit the whole range of the admitted values for the PF with a signal from 0 to 20mA. Particular attention must be paid to the value of the PF variable which may vary from C0,001 and L0,000 (for each phase): these values will be retransmitted and will then cor­respond to 0 and 20 mA. When the PF will have a value equal to 1, being in the middle between C0,001 and L0,000, the value of the out­put will correspond to the middle of the scale, that is 10mA. As a con­sequence, the instrument will have to be programmed as follows:
VARIABLE: PF L1 (or L2 or L3 or PF). MIN OUT:0,0%. MAX OUT: 100,0%. MIN IN: C0,001 (the C symbol shows a CAP MAX IN: L0,001 (the L symbol shows an INDUCTIVE value). L0,001
has been chosen as minimum value to be set in order to avoid any undesirable swifting of the r
epeated outputs.
ACITIVE value).
Page 33
WM5-96 Instruction Manual
erial outputs
S
************MMAAIINN MMEENNUU***********
LARMS
A EVENTS SELEC.
NALOG OUTPUTS
A
ERIAL OUTPUT
1
2
S
RREEVV.. 11..44..22..66 ((AAVV55)
************MMAAIINN MMEENNUU***********
S232
R
RS485
PTICAL
O
MMOODDBBUUSS:: 99660000,,88,,NN,,1
PARITY
2a
O PARITY
N
2b
ODD PA EVEN PA
RITY
RITY
*
Measure
Programming WM5 96
his function allows the user to set the RS232 and RS485 serial ports
T as well as the optical port. Choose the SERIAL OUTPUT function (1)
31
by means of the “ ” keys and confirm with “S”. Then choose the serial port to be set and the instrument will synthetically display the
> 2s
previously set parameters (a).
S232(2): set the parameters of the RS232 serial port (AR1039 mod-
)
R
ule), if present, then select RS232 by means of the “ ” keys and
onfirm with “S”, select the type of parity to be used (2a, 2b) with the
c “ ” keys and confirm the selection with the “S” key.
RS485 (2): set the communication parameters of the RS485 serial port
*
Go back
(AR1034 module, max 9600 bps, AR2040 max, 115,2kbps), if present;
> 2s
select it by means of the “ ” keys and confirm with the “S” key.
1
a
Then select the following:
ADDRESS (3): select the serial address to be given to the instrument
(from 1 to 247). Enter the desired address (3a) by means of the “ ” keys and confirm with the “S” key
.
BAUD RATE (3): select the baud rate of the serial port. Select the
desired baud rate value (3b) by means of the “ ” keys and con­firm with the “S” key.
PARITY (3): select the type of parity of the serial port (3c) by means of
the “ ” keys and confirm with the “S” key.
OPTICAL (2): set the parameters of the front optical port (see next
paragraph “Optical port”).
3
ADDRESS ? min max
3a
ADDRESS
BAUD RATE PARITY
1.0000.
00000
1 247000
123...
NO PARITY
3c
ODD PARITY EVEN PARITY
3b
To exit the various menus and submenus, keep the “S” key pressed for at least 2 seconds.
9600
19200 38400 115200
Page 34
32
ARMPowered
PRG
measures
S
S
R
!
R
A
L
WM5-96
SmartPowerQuality Analyzer
Class0.2 A/V
ptical port
O
Programming WM5 96
WM5-96 Instruction Manual
he serial optical port (d) transmits the data by means of an infrared
T device (e-b). The connection (c) to the PC through the relevant mag­netic connector (a) is made by means of a USB port (f) or 232 serial port (depending on the model being used): the user will then be able
a
b
c
d
e
to read, transmit or receive information on the WM5 programming (the
M5Soft software can be purchased on request).
W The optical port makes it also possible to retransmit the energy meter-
ng by means of pulses . In this case the light generated by the opti-
i cal port is not visible because it’s within the infrared spectrum. Set the desired type of use of the optical port (energy transmission by means of pulses or MODBUS or ANSI serial communication) before using it.
Follow steps
f
(2) by means of the “ ” keys and confirm with “S”. Afterwards, select the type of protocol/communication mode (4) to be used:
(1) and (2) on the previous page, then select “OPTICAL”
ANSI (4): standard ANSI serial communication protocol. Choose the
ANSI function by means of the “ ” keys and confirm with “S”.
MODBUS (4): MODBUS serial communication protocol. Choose the
MODBUS function by means of the “ ” keys and confirm with “S”. Then set the relevant parameters such as baud rate and parity, see steps (3b) and (3c) on the previous page.
PULSE (4): the port will be enabled for the energy retransmission.
Select the PULSE function by means of the “ ” keys and confirm with “S”. Then select the energy type to be retransmitted (5), select ENERGY TYPE by means of the “ ” keys and confirm with “S”.
********SSEERRIIAALL MMEENNUU*******
RS232
ANSI
RS485
MODBUS
OTTICA
PULSE
I
4
AANNSSI
*
The instrument will display the list of available energies (5a) (kWh+, kWh-, kvarh+, kvarh-) (5a); select the desired energy by means of the “ ” keys and confirm with the “S” key: for each energy it’s possi­ble to choose if r
etransmitting the total ener
gy metering (TOT) or by tariff from T1 to T12 (5b), choose the desired energy detail by means of the “ ” keys and confirm with “S”. Finally, set the value of the pulse related to the measured energy, that is the number of pulses generated for each measured kWh+ (or kWh- or kvarh+ or kvarh-).
****OOPPTTIICCAALL OOUUTT MMEENNUU***
ENERGY TYPE
PULSES
*
Choose PULSE (5) by means of the “ ” keys and confirm with “S”, then (5c) set the desired pulse value and confirm it by means of the “S” key
.
5
PULSE/kWh min max
5c
123...
1.0000
0.001 10000
0
Note: set the optical port in the MODBUS mode before using it relat­ed to the serial communication softwar
kWh +
kWh ­kvarh +
5a
kvarh -
TOT
L1 L2
5b
L3
e “Wm5Soft”.
Page 35
WM5-96 Instruction Manual
Ethernet Port
**********MMAAIINN MMEENNUU*********
VENTS SELEC.
E ANALOG OUTPUTS
ERIAL OUTPUT
S
THERNET PORT
1
2
E
RREEVV.. 11..44..22..66 ((AAVV55)
********EETTHHEERRNNEETT MMEENNUU*******
P ADDRESS
I
ATEWAY
G MODBUS TCP/IP
I
P
92. 168. 2. 122
2a
1
MASK
255. 255. 255. 0
123...
)
*
*
Measure
> 2s
Go back
> 2s
Go back
> 2s
Programming WM5 96
his function allows the user to set the Ethernet port. Choose the
T
33
ETHERNET PORT function (1) by means of the “ ” keys and con­firm with “S”. IP ADDRESS (2): set the IP parameters. Select IP ADDRESS by means of the “ ” keys and confirm with “S”. Then select IP (IP
ddress) or MASK (netmask) by means of the “ ” keys and con-
a firm with “S”. Set the value (2a) by means of the “ ” keys, where
he “ ” key increases the value and the “ ” key switches to the fol-
t lowing number. Confirm with “S”.
GATEWAY (2): set the IP gateway parameters. Select GATEWAY by
means of the “ ” keys and confirm with “S”. Set the value (2b) by means of the “ ” keys, where the “ ” key increases the value and the “ ” key switches to the following number. Confirm with “S”. MODBUS TCP/IP (2): to modify the default port of the Ethernet port (502), select MODBUS TCP/IP by means of the “ ” keys and con­firm with “S”. Set the value
(2b), from 1 to 65535, by means of the
” keys and confirm with “S”.
To exit the various menus and submenus, keep the “S” key pressed for at least 2 seconds.
PORT ? m
c
2
m
GW IP
2b
000. 000. 0. 000
Digital filter
**********MMAAIINN MMEENNUU*********
EVENT SELEC. ANALOG OUTPUTS SERIAL OUTPUT
1
2
FIL
RREEVV.. 11..44..22..66 ((AAVV55)
**********MMAAIINN MMEENNUU*********
RANGE
COEFFICIENT
in ax
TER
123...
123...
0.
5020 100000 655350
Go back
> 2s
*
)
*
Measure
> 2s
Go back
> 2s
Thanks to the digital filter, it’s possible to stabilize the measurements which ar
e too instable when displaying the r
elevant values. By means of the “ ” keys, select FILTER and confirm with “S”. Select the function to be set from the various submenus with the “ ” keys and confirm with “S”:
RANGE (3): set the operating range of the digital filter. The value is
expressed as a %. By means of the “ ” keys select RANGE and confirm with “S”, then set the value as a percentage by means of the “
” keys and confirm the value with the “S” key.
COEFFICIENT (4): set the filtering coefficient of the instantaneous
measures. By increasing the value, also the stability and the settling time of the measur
es are increased. Select COEFFICIENT by means of the “ ” keys and confirm with the “S” key, then set the value by means of the “ ” keys and confirm the new value with the “S” key.
RANGE
3
Confirm
4
Confir
min max
COEFFICIENT min max
m
1,000%
0.100%
100.0%
123...
4,000%
1.100%
255.0%
123...
Page 36
34
igital filter programming
D Examples
Programming WM5 96
WM5-96 Instruction Manual
Example 1
How to stabilize the value of the VL-N variable displayed on the
isplay, fluctuating from 222V and 228V.
d
The parameters of the digital filter have to be programmed as follows:
ANGE:the variable has fluctuations within the mean value whose
R
amplitude is equal to ±0,75% of the full scale rated value of the vari­able itself (obtained by the following calculation: (228-222)/ 2= ±3V, then ±3*100/400V= ±0,75% where 400V is the phase-neutral rated value of an AV5 input). The “range” parameter, representing the action range of the digital filter, is to be programmed to a value which must be slightly higher than the percentage amplitude of the fluctuation: ex.
1.0%.
COEFFICIENT:if the new value measur
the action range of the filter, the new displayed value is obtained by adding algebrically the previous value to the variation divided by the filtering coefficient. As a consequence, a value higher than this coeffi­cient implies a longer settling time and therefore a better stability. You generally obtain the best result by setting the filtering coefficient to a value equal to at least 10 times the range parameter value. In the following example: 1,0*10=10, the stability of the filtering coef­ficient can be improved by increasing the filtering coefficient, the admitted values are included within 1 and 255.
ed by the instrument is within
Example 2 How to stabilize the value of the displayed System Active Power
), fluctuating between 300kW and 320kW (the load is connect-
(W ed to the instrument by means of a 300/5A CT and a direct meas­ure of the voltage).
The parameters of the digital filter must be programmed as follows: RANGE: the variable has fluctuations within the mean value whose
2,78% of the full scale rated value of this vari
amplitude is equal to able. This value is obtained by the following calculation: (320-300)/ 2= ±10kW, then ±10*100/360kW= ±2,78%, where 360kW is the rated value of the System Active Power of an AV5 input, at the above men­tioned CT and VT ratios and obtained by means of the following for­mula: (320-300)/ 2= ±10kW, then ±10*100/360kW= ±2,78%, where 360kW is the rated value of the System Active Power of an AV5 input at the above mentioned CT and VT ratios and obtained by means of the following formula: “VLN * VT * IN * CT * 3” where VLN = rated input voltage (400V for the AV5 input), VT= primary/secondary ratio of the voltage transformer being used, IN = rated curr input), CT = primary/secondary ratio of the voltage transformer being used (in this example “400*1*5*60*3=360kW). The RANGE parameter, representing the digital filtering coefficient action range, is to be programmed to a value which must be slightly higher than the percentage of the fluctuation: ex. 3.0%. COEFFICIENT: if the new value acquired by the instrument is within the filter­ing action range, the new displayed value is obtained by adding algebrically the previous value to the variation divided by the filtering coefficient. As a con­sequence, a value higher than this coefficient implies an higher settling time
e a better stability
and ther setting the filtering coef the range parameters. In the example: 3.0*10=30. In or bility you can incr ed within 1 and 255.
efor
±
ent (5A for the AV5 type
Generally speaking the best r
.
ficient to a value equal to at least 10 times the value of
ease the filtering coef
ficient, the admitted values ar
esult is obtained
der to impr
ove the sta
e includ
-
-
-
Page 37
WM5-96 Instruction Manual
Programming WM5 96
Example 3. It’s necessary to stabilize the value of the displayed variable AL 1
phase current 1), fluctuating within 470V and 486V.
(
To be able to manage the alarm function and following activation and
eactivation of the relay, this value is not to be subject to continuous
d fluctuations. In this example we have considered using a 500/5A CT. The paramters of the digital filter is to be programmed as follows: RANGE: the variable has fluctuations within the mean value whose amplitude is equal to ±1,60% of the full scale rated value of this vari­able (obtained by means of the calculation: (486-470)/ 2= ±8A, then ±8*100/500V= ±1,60% where 500A is the value referred to the primary of the transformer being used). The “range” parameter, which repre­sents the action range of the digital filter value slightly higher than the pourcentage amplitude of the fluctuation: for example 2.0%. COEFFICIENT: if the new value acquired by the instrument is within the filtering action range, the new displayed value is calculated alge­brically adding to the previous value the variation divided by the filter­ing coefficient. As a consequence, a higher value of this coefficient implies a higher settling time and therefore a better stability. Generally speaking, the best result is obtained setting the filtering coefficient at a value equal to at least 10 times the value of the range parameter. In the example: 2.0*10=20. To improve the stability you can increase the filtering coefficient, the admitted values are within 1 and 255.
, is to be programmed to a
35
Page 38
lock
C
1
2
3
DATE
3d
36
Programming WM5 96
************MMAAIINN MMEENNUU***********
ANALOG OUTPUTS
ERIAL OUTPUT
S
ILTER
F
LOCK
C
3a
)
EUROPE
USA
RREEVV.. 11..44..22..66 ((AAVV55)
**********CCLLOOCCKK MMEENNUU*********
ONE
N
WITHOUT BACKUP
ITH BACKUP
W
********CCLLOOCCKK AADDJJUUSSTT*******
URO/USA
E TIME ZONE
DATE
TIME
2200//0066//0066 0088::2288::2266 AAM
12/31/06
123...
AFRICA
AMERICA
3b
ANTARCTICA
ARCTIC
ADAK
ANCHORAGE
3c
ANGUILLA
ANTIGUA
*
*
*
M
Measure
> 2s
Go back
> 2s
WM5-96 Instruction Manual
In the clock menu the clock functions can be set. Choose the CLOCK (1) func-
ion by means of the “ ” keys and confirm with “S”. The instrument dis-
t plays three functioning modes.
NONE (2): WM5 is not supplied with a clock, therefore it will not be possible
to manage the tariffs by means of TIME/CALENDAR (all functions which are linked to the presence of the clock will be deactivated). Choose the function NONE by means of the “ “ keys and press the “S” key to confirm.
ITHOUT BACKUP(2): WM5 will be able to manage date and time but, in
W
case of power supply failure, it will not store neither time nor date. When the
nstrument will be switched on again, the correct date and time will have to be
i set again. Select the WITHOUT BACKUP function by means of the “
eys and confirm with “S”.
k WITH BACKUP (2): WM5 is supplied with a clock and the correct date and time will be stored also in case of power failure. Select the function “WITH BACKUP” by means of the “ ” keys and confirm with “S”. The function “WITH BACKUP” can be selected only if the AR1039 module is used. Then all the information necessary to the clock are to be set.
EURO/USA (3)
: set the time format as 24h european (00:00) or 12h american (12:00 AM). Select the EURO/USA function by means of the “ ” keys and confirm with “S”. Select EUROPE (3a), to set the time format as 24h, or choose USA (3a), to set the time format as 12h. TIME ZONE (3): set the time zone related to the place where WM5 is installed, choose the correct time zone (3b) by means of the “ ” zone and confirm with “S”. Then select by means of the “ ” keys the correct country (3c) and confirm with “S”.
DATE (3): set the current date, choose the DATE function by means of the “
” keys and confirm with “S”. Then select the date by means of the “ ” keys , set the date (3d) and confirm with “S”.
TIME (4): set the current time, select the TIME function by means of the “
keys and confirm with “S”. Then set the time (4a) by means of the “ ” keys and confirm with the “S” key.
SYNC (4): enables or disables the synchronism of the clock with the digital
inputs enabled as SYNC, the clock at each synchronization pulse will set to the multiple of the integration time nearer to the current time. Choose the SYNC function by means of the “ “ keys and confirm with “S”. Then, by means of the “ ” keys, choose OFF (4b) to disable the synchronism or
(4b) to enable it and confirm with “S”.
ON
******CCLLOOCCKK AADDJJUUSSTT*****
TIME ZONE
TE
DA
TIME
SYNC
4
4a
2200//0066//0066 0088::2288::2266 AAM
TIME
10:04 PM
123...
*
M
OFF
4b
ON
The backup of the time is only possible with the AR1039 module being installed.
Page 39
WM5-96 Instruction Manual
anguage
L
**********MMAAIINN MMEENNUU*********
SERIAL OUTPUT
ILTER
F CLOCK
ANGUAGE
1
L
RREEVV.. 11..44..22..66 ((AAVV55)
)
*
Measure
> 2s
Programming WM5 96
37
This function allows to set the language to be displayed by WM5. Choose the
ANGUAGE function (1) by means of the “ ” keys and confirm with “S”.
L Then select the desired language by means of the “ ” keys and confirm
ith “S”.
w
**********MMAAIINN MMEENNUU*********
USCITA SERIALE
NGLISH
E
ILTRO
F
ITALIANO
LOCK
C
EUTSCH
D
LANGUAGE
RANCAIS
2
Confirm
F
RREEVV.. 11..44..22..66 ((AAVV55)
Serial number and Meter Info.
**********MMAAIINN MMEENNUU*********
LOCK
C
ANGUAGE
L
SERIAL NUMBER
1
2
METER INFO.
RREEVV.. 11..44..22..66 ((AAVV55)
527 70000 T 0017
EN
CAN ANSI
)
)
*
*
Go back
2s
>
Measure
> 2s
Selecting the SERIAL NUMBER function, the instrument displays the seri­al number of the instrument. Select the function SERIAL NUMBER (1) by means of the “ ” keys and confirm with “S”.
Selecting the METER INFORMATION function, the instrument displays the information relevant to the reference standard used by the instru­ment. Select the function METER INFORMATION (1) by means of the “ ” keys and confirm with “S”. Choose the desired standard by means of the “ ” keys and confirm with “S”. Note: the selected standard is displayed in the Info pages too.
Keypad lock, lock of access to pro­gramming and reset menu
a
By turning the screwdriver anticlockwise to the end-of-stroke of the relevant trimmer, as shown in the figure on the left (a), it’s possible to lock the access to the programming of the instrument both from key­pad and from serial communication. Moreover, after locking the key­pad it will not be possible to carry out the reset commands any more. However, it will be still possible to scroll all the display pages and the relevant details.
Page 40
38
Description Ch Part No. Legend
Model
Layout and list of modules
WM5-96 Instruction Manual
WM5 96 AV53 H XX XX XX XX XX
WM5-96 with optical port
AD2001
ANSI C12.18 type
1 3 3 4 2 1 1 2 2 2
3 3 4 2 1 1 2 2 2 1 1
AD2000
AQ2030 AQ2031
AP1021 AP1020
AR1061 AQ1038 AQ1042 AO1037 AO1036 AO1059 AO1058 AO2050 AO2051 AO2052
AQ1038 AQ1042 AO1037 AO1036 AO1059 AO1058 AO2050 AO2051 AO2052 AR1034 AR2040
WM5-96 without optical port
(Range code + system) measuring inputs
400/690VL-L 1/5A (10A) 120/208VL-L 1/5A (10A)
Power supply
18-60VAC/DC Power supply 90-260V
None Ethernet/Internet port Digital inputs Digital inputs + aux Open collector output Open collector output Open collector output Relay output Analogue output 20mADC Analogue output 10VDC Analogue output +/-5mA
None Digital inputs Digital inputs + aux Open collector output Open collector output Open collector output Relay output Analogue output 20mADC Analogue output 10VDC Analogue output +/-5mA RS485 9600bps RS485 115200bps
AC/DC Power supply
SLOT
A
B
SLOT
WM5 96
WM5 96
AV5.3 AV6.3
H
XX E2 D1
D2 O4 O2 O1 R1 B1
W1
B2
XX D1 D2 O4 O2 O1 R1 B1
W1
B2
S1
S2
L
Examples of which modules to order for: WM5-96 AV53 H B1 S1 R2 O2 SX
Material Code WM5 96
AV53 measuring inputs (400/690VL-L) 90-260VAC/DC power supply 20mA analogue output (2 channels) RS485 serial port at 9600 bps Relay output (2 channels) Open collector output (2 channels) RS232 serial port +RTC
AD2001 AQ2030 AP1020 AO2050 AR1034 AO1035 AO1036 AR1039
None Digital inputs Digital inputs + aux Open collector output Open collector output Open collector output Relay output Relay output Analogue output 10VDC Analogue output +/-5mA
None Digital inputs Digital inputs + aux Open collector output Open collector output Open collector output Relay output Relay output Analogue output 10VDC Analogue output +/-5mA
Utility grade with optical port RS232 + RTC (utility grade) Revenue approval. An “instru­ment setting” form must be pr erly filled in by the user RS232+RTC + “XU” option
RS232 + R out optical port.
TC (utility grade) with-
.
op-
3 3 4 2 1 1 2 2 2
3 3 4 2 1 1 2 2 2
1
1
1
AQ1038 AQ1042 AO1037 AO1036 AO1059 AO1058 AO1035 AO2051 AO2052
AQ1038 AQ1042 AO1037 AO1036 AO1059 AO1058 AO1035 AO2051 AO2052
AR1039
AR1039
AR1039
T O
SL
C
OT
SL
D
OPTIONS
T O
SL
E
XX D1 D2 O4 O2 O1 R1 R2
W1
B2
XX D1 D2 O4 O2 O1 R1 R2
W1
B2
XX SX XU
SU YY
SY
ower supply
P
ABCD
E
Measuring input (range+ system)
Page 41
Installing
WM5-96 Instruction Manual
Panel cut-out and dimensions
5mm
24mm
Fixing, removing the modules and mounting the instrument to the panel
39
2
1
3
4
1
5
a
c
b
b
Mounting and removing the modules: the different modules (input, output and power supply) have been designed to be plugged in the available slots. T
o know in which slots they are to be mounted, r
efer to the table on the left. For a correct assembling of the instrument, first plug the modules into the relevant slots, then as last operation plug in the central module which can also be a blind module or an RS232 serial communication module. This module also locks
emove the modules use a screwdriver with a flattened tip and move the
the other modules in the r
elevant slots. T
o r
two fixing tabs to the side (1); then remove the central module from its slot by pressing your thumb towards points (2)
. Finally extract the central module.
(5)
and Panel mounting: insert the instrument (holding its front side) into the panel and fasten it (from the back) by fixing the two lateral brackets being supplied (a) to the appropriate location (b), and then locking them by means of the two screws supplied with the instrument (c).
Sealing the instrument
c
a b
The special sealing kit, made by two plastic “cuneiform” devices (c) is supplied with the instrument; these two devices will have to be insert­ed as shown in the figur
e on the left
as indicated by the pointer
(b)
.
(a), then the seal is to be placed
Page 42
40
L1 L2
L3
2
9
10
1
8
L1L1
NN
L3L3
L2L2
U
1
1
3
3 55
22
44 66
I
I
L1L1
L2L2
L3L3
8
8 77 1010
99
1
692
345
87
10
L1 N
271
8
10
9
L1 L2
N
271
8
4
3
9
Wiring diagrams
1-phase, 2-wire input connections, (1P)
WM5-96 Instruction Manual
Fig. 1 Fig. 2
Direct connection CT connection
2-phase, 3-wire input connections (2P)
Fig. 4 Fig. 5
Fig. 3
CT and VT connection
Direct connection
CT connection
3-phase, 3 and 4-wire input connections - Balanced load (3P)
Fig. 6 Fig. 7
ect connection (3-wir
Dir
e system)
CT and VT connection (3-wire system)
Fig. 8
CT and VT connection (4-wire system)
Page 43
L1 L2
L3
2
9
10
1
3
4
5
6
8
L1L1
NN
L3L3
L2L2
U
11
33 55
22
44 66
II
L1L1
L2L2
L3L3
88 77 1010
99
1
692
345
87
10
L1 L2
L3 N
2
9
10
7
1
3
4
5
6
8
Wiring diagrams
WM5-96 Instruction Manual
3-phase, 3-wire input connections - unbalanced load (3P)
41
Fig. 9 Fig. 11
Direct connection (3-wire system)
CT connection (3-wire system)
3-phase, 3-wire ARON input connections (3P)
Fig. 12 Fig. 13
Fig. 10
CT and VT connection (3-wire system)
ARON connection (3-wir
e system)
ARON and VT connection (3-wire system)
3-phase, 4-wire input connections - unbalanced load (3P+N)
Fig. 14
Direct connection (4-wire system)
CT connection (4-wire system)
Fig. 15 Fig. 16
CT and VT connection (4-wir
e system)
Page 44
42
L
N
-
+
F
Wiring diagrams
Power supply, analogue and digital output modules
WM5-96 Instruction Manual
A2
A1
AO1058.
1 relay output
Digital inputs
AP1020. AP1021.
Power supply.
F= 1.25 A T
AO1035.
2 relay outputs
AO2050.
2 analogue outputs
0-20mA DC.
AO2052.
2 analogue outputs
-5/+5mA DC.
AO2051.
2 analogue outputs
10V DC.
AO1037. 4 open collector outputs. This wiring diagram is also valid for the open collector mod­ules with 1 or 2 outputs. The load resistances (Rc) must be designed so that the close contact current is lower than 100mA; the VDC voltage must be lower than or equal to 30VDC.
AQ1042.
Connection by means of
NPN transistors.
Connection by means of
AQ1042.
PNP transistors.
RS485 serial port (AR1034 9600bps, AR2040 115200bps)
b
T
a
TER
CONVER
SERIAL
e connection.
2-wir
nected in parallel
Termination of the serial port (T): it’s carried out only on the last instrument of the network, by means of a jumper between (Rx+) and (T).
b
T
a
TER
SERIAL
CONVER
4-wire connection. Other WM5-96 (a) lected in parallel (b). Serial RS485-RS232 converter.
erminatio of the serial port
T
instrument of the network, by means of a jumper between (Rx+) and (T).
AQ1042.
Connection by means of
contacts.
Other WM5-96 . Serial RS485-RS232 converter.
(b)
(a) provided with RS485 are con-
ovided with RS485 ar
pr
(T): it is carried out
AQ1038.
Connection by
means of contacts.
e col
only on the last
-
Page 45
WM5-96 Instruction Manual
Input Specifications
Technical specifications
43
Number of analogue inputs
Current 1 (1-phase; system code: 3)
3 (3-phase; system code: 3)
oltage 1 (1-phase; system code: 3)
V
4 (3-phase; system code: 3)
Digital inputs (on request) Up to 12
Q1038 No. of inputs: 3 (voltage-free)
A
Purpose “dmd” measurements
synchronisation. Tariff selection: energy. Contact status reading.
lock synchronisation.
Contact measuring current
AQ1042
C <8mA/ 17.5 to 25VDC
Number of inputs: 3 + excitation
output
Purpose “dmd” measurements
synchronisation. Tariff selection: energy. Contact status reading.
Clock synchronisation. Excitation output 16V<+Aux<24VDC Max15mA Contact measuring current
15mA
Common characteristics
Close contact resistance Max 1k Open contact resistance Min 100k Insulation see “Insulation between
inputs and outputs” table
Accuracy
(display, RS232, RS485)
In: 5A, If.s.: 10A
Un: see voltage ranges below
Current (AL1, AL2, AL3) from 0.05In to Imax:
(@20°C ±5°C, R.H. 75%)
±(0.2%RDG+2DGT)
from 0.01In to 0.05In:
±(0.5%RDG+2DGT)
Current (A
oltage
V
) ±0.5% RDG (0.2 to 2 In)
n
V5:
A
range
range AV6: 120/208V
from 40 to 100 Hz
5°C, R.H.75%)
±
C
°
(@20
400/690V
V
: from 185 V to 460 V
L-N
V
: from 320 V to 800 V
L-L
±(0.2%RDG+1DGT)
V
: fr
L-N
V
: from 78 V to 250 V
L-L
±(0.2%RDG+1DGT)
AC
L-L
AC
L-L
om 45 V to 145 V
Includes also: fr
equency, power supply and output load influences.
Frequency
±
440 Hz)
(40 to
RDG
0.1% Active power and 0.05In to Imax, PF 1: apparent power ±(0.5%RDG+1DGT)
(@ 20°C ± 5°C, R.H. 75%) 0.01In to 0.05In, PF 1:
±(1%RDG+1DGT)
0.1In to Imax, PF 0.5L, PF 0.8C:
(0.6%RDG+1DGT)
±
0.02In to 0.1In, PF 0.5L, (1%RDG+1DGT)
PF 0.8C:
±
Reactive power
(@ 20°C ± 5°C, R.H. ≤ 75%) 0.1In to Imax, sen
ϕ 0.5L/C:
±(2%RDG+1DGT)
0.05In ÷ 0.1In, sen
ϕ 0.5L/C:
±(2.5%RDG+1DGT)
Imax, sen
÷
0.05In
ϕ 1:
±(2%RDG+1DGT) 0.02In to
0.05In, sen
ϕ 1:
±(2.5%RDG+1DGT)
nergies
E
@ 20°C ± 5°C, R.H. 75%) Active: class 0.5 according
(
to
EN62053-22, ANSI C12.20 Reactive: class 2 according to EN62053-23, ANSI C12.1 In: 5A, Imax: 10A
0.1In: 500mA, Start-up current: 5mA Un: 400/690V
n:
U
Harmonic distortion 1% FS (FS: 100%)
(@ 20°C ± 5°C, R.H. 75%)
fase: ±2°; Imin: 5mA Imax: 15Ap; Umin: 30V Umax: 500Vp
emperature drift
T
200ppm/°C (A/V), 300ppm/°C
20/208V
1
L
-L
(AV5)
L-L
AV6)
(
RMS
RMS
(all other measurements)
Sampling rate
6400 7680
samples/s samples/s
@ 50Hz @ 60Hz
Display Graph LCD, backlighted
(128x64 dots). Re a d-out f or instantaneous variables: 4x4 digit Total energies: 4x9 digit; Partial energies: 4x9 digit
Display refresh time 100ms Max. and min. indication Max. 9999 (999,999,999),
Min. -9999 (–999,999,999)
Front LED Red
Blinking light in case of vir­tual alarm. Fixed light in case of digital output acti­vation (alarm)
Measurements Current, voltage, power,
gy, power factor
ener
, fre­quency, harmonic distortion (see “Display pages”). TRMS measurement of a distorted wave (voltage/current).
Coupling type Direct
est factor < 3, max 10A peak
Cr
Input impedance
400/690V 120/208V Current 0.01
(AV5) 1.77 M±5%
L-L
L-L
V6)
(A
885 k
±
5%
Frequency 40 to 440 Hz Overload (maximum values)
Permanent: voltage/current AV5: 460V
AV6: 145V
For 500ms: voltage/current AV5: 800V
AV6: 240V
, 800VLL/10A
LN
, 250VLL/10A
LN
, 1380VLL/36A
LN
, 416VLL/36A
LN
;
;
Page 46
44
Technical specifications
Output specifications
WM5-96 Instruction Manual
Analogue outputs (on request)
Number of outputs Fino a 8 (max 4 x 20mA + 4
Accu racy (@ 25°C ±5°C, R.H. ≤60%) ±0.1%FS (20mA o 10VCC)
ange 0 ÷ 20mA o 0 ÷ 10 VCC o
R
Scaling factor: Programmable within the
Response time
Ripple
Total temperature drift 500 ppm/°C Load : 20 mADC 350
Insulation see “insulation between
Optical communication port According to ANSI C12.18 RS422/RS485 port (on request) Multidrop
Connections 2 or 4 wires, max. distance
Protocol MODBUS RTU /JBUS, Data (bidir
Stored energy (EEPROM)
Data format
Baud rate
Insulation see “Insulation between
RS232 port (on request) Bidirectional (static and
Connections 3 wires, max. distance 15m, Data format
Baud-rate 9.6k bit/s Protocol MODBUS RTU /JBUS Other data as for RS422/485
10 VDC 10k ±5 mA 1400
ectional)
Dynamic (reading only) All display variables (see
Static (writing only)
x 10VDC o 4 x 20mA o 8 x 10VDC o 8 x ±5mA)
±0.3%FS (±5mA), FS=10mA
5mA
±
whole range of retransmis­sion; it allows the retrans­mission management of all values from: 0 and 20 mA, 0 and 10VDC, or -5mA and +5mA 400 ms typical (filter excluded)
1% (according to
IEC 60688-1,
inputs and outputs” table
bidirectional (static and dynamic variables)
1000m, termination directly on the module . Addresses: from 1 to 247, selectable by key-pad.
also the table “List of the variables that can be con nected to”...) All configuration parameters, reset of energy, activation of digital output
kWh/kvarh 1-start bit, 8-data bit, no parity/even parity, odd parity
9.6k, 19.2k, 38.4k, 115.2k bit/s selectable bauds
inputs and outputs” table
dynamic variables)
1-start bit, 8-data bit, no parity, even parity, odd parity, 1 stop bit
EN 60688-1)
max. 999.999.999
, 1 stop bit
Ethernet/Internet port
Protocols Modbus TCP IP configuration Static IP TCP port Selectable (default 502) Client connections Max 5 simultaneously Connections RJ45 10/100 BaseTX
igital outputs (on request)
D
ulse type
P
Number of outputs Up to 16 Type
Pulse duration
Alarm type
Number of outputs Alarm modes Up alarm, down alarm, in
Set-point adjustment from 0 to 100% of the
Hysteresis from 0 to full scale On-time delay 0 to 255s
Output status Selectable; normally
esponse time
Min. r
Note The 16 digital outputs
-
Static (digital) outputs (on request)
Purpose For pulse outputs or for
Signal V
Relay (digital) outputs (on request)
Purpose
Output type Relay SPDT
Insulation
Electrical life 10
Mechanical life 30x10
100ms, < 120msec (ON),
Programmable from 0.001 to 1000 pulses per kWh/kvarh (total and partial)
utputs connectable to the
O total and/or partial energy meters (Wh/varh)
100ms (OFF) according to EN62053-31
up to 16, independent
window alarm, out window alarm. All of them can be used with start up deactiva­tion function and/or latch. All the alarms can be con­nected to all variables (see the table “List of the vari­ables that can be connect­ed to”).
electrical scale
de-energised and normally energised 200ms, filters excluded, Set-point on-time delay: “0 s”
can also work as combination of pulse outputs and alarm outputs.
alarm outputs
1.2 VDC/ max. 100 mA
ON
V
30 VDC max.
OFF
Insulation between inputs and out­puts” table
For alarm outputs or for pulse outputs
AC 1-8A, 250VAC DC 12-5A, 24VDC AC 15-2.5A, 250VAC DC 13-2.5A, 24VDC
Insulation between
see “ inputs and outputs
5
operations (@ 8A,
, PF1)
250V
6
operations
see “
Insulation
” table
Page 47
WM5-96 Instruction Manual
Software functions
Technical Specifications
45
Password Numeric code of max 4
1st level Password “0”: no protection 2nd level Password from 1 to 1000:
System selection
ystem 1 1-phase (2 wires)
S System 2, unbalanced 2-phase (3 wires) System 3, balanced 3-phase (3 wires+1CT) System 3, unbalanced 3-phase (3 wires)
Transformer ratio CT up to 60 kA (6000 max)
Filters
Filter operating range 0.1 to 100% of the input
Filtering coefficient 1 to 255 Filter action
Alarms
Working mode “OR” or “AND” or
digits from 0 to 1000; 2 protection levels of the programming data
ll data are protected.
a
3-phase (4 wires)
VT (PT) up to 600 kV (6000 max)
electrical scale.
Display, alarms, serial outputs (fundamental variables: V, A, W and their derived ones).
“OR+AND” functions (see “Alarm parameter and logic” page). Freely programmable on up to 16 alarms. The alarms can be connected to any variables available in the table “List of the variables that can be connected to”
Reset By means of the key-pad or
ata stamping
D
ype of data Alarm, min, max, digital
T
Number of events Up to 10.000 Data management type FIFO Data storage type Data flash
Displaying 4 variables per page
Energy meters Up to 28 pages depending
Stored events 10,000 events.
Display language Selectable:
of the configuration software, it is possible to reset the following data:
- all the min, max, dmd,
md-max values.
d
total and partial counters.
-
latch alarms.
-
- all the events.
input status, digital output status as remote control, resets. All events are stored with date (dd:MM:yy) and hour (hh:mm:ss) reference
1 page that can be laid out by the user Up to 36 pages
on the selected tariff mode. Displaying of the consumed energy of the previous 12 months.
English, Italian, French, German, Spanish
Wm5Soft parameter programming and variable reading software
Wm5Soft Multi-language software
(Italian, English, Fr German, Spanish) for variable reading, instrument calibration and parameters programming The program runs under Windows XP/Vista Working mode Three different working modes can be selected:
- management of local
ench,
.
Data Storing In pre-formatted XLS files
ransfer
Data T
RS232 (MODBUS);
- management of local optical port (MODBUS);
- management of a local RS485 network (MODBUS);
(Excel data base). Manual or automatic at programmable timings.
Page 48
46
Technical specifications
Time period management
eters
M
otal 4 (9-digit)
T
artial and multitariff 48 (9-digit)
P
Tariffs Up to 12
ime periods
T
umber of periods U p to 24 per da y
N
ulse output
P
Energy meter recording Consumption history by
Up to 100 different days per year
onnectable to total and/or
C
artial meters (multitariff)
p
recording of the monthly energy meters (12 previous months). Recording of total and partial energy meters. Energy meter recording (EEPROM) Max.999.999.999kWh/kvarh.
WM5-96 Instruction Manual
Energy Meters
Total energy meters 4 (+kWh, +kvarh, -kWh, -kvarh)
Monthly energy meters 48 (energy meters for
Partial energy meters 16 (using digital inputs:
It is possible to divide each
nergy meter here above listed
e
n 3 additional energy meters
i
1 for each phase “L1-L2-L3”)
(
each month: “+kWh, +kvarh,
-kWh, -kvarh”)
max 4 tariffs). 48 (using the internal clock:
ax 12 tariffs)
m
Harmonic distortion analysis
Analysis principle FFT
Harmonic measurement
Current Up to the 63rdharmonic Voltage Up to the 63
Type of harmonics THD (VL1 and VL1-N)
Harmonic phase angle The instrument measur
THD odd (V THD even (V The same for the other phases: L2, L3. THD (A THD odd (A THD even (A The same for the other phases: L2, L3.
angle between the single har­monic of “V” and the single harmonic of “I” of the same
. Accor
der
or of the electrical angle, it is possible to know if the distor­tion is absorbed or generated.
d
r
harmonic
L1 and VL1-N)
and V
L1
L1)
L1)
L1)
ding to the value
L1-N)
es the
Note: if the system has 3 wires the angle cannot be measured.
Harmonic details The harmonic contents is dis-
played as a graph showing the whole harmonic spec­trum. This value is also given as a numerical information: THD % / RMS value THD even % / RMS value THD odd% / RMS value single harmonics in % / RMS value
System The harmonic distortion
can be measured in single­phase, 3-wire or 4-wire systems. Tw: 0.02 sec@50Hz with­out filter
Page 49
WM5-96 Instruction Manual
General Specifications
Technical Specifications
47
perating
O temperature (R.H. < 90% non-condensing)
Limit range of operating temp. -20° to +55°C (-4° to 131°F)
Storage -30° to +60°C (-22° to 140°F) temperature (R.H. < 90% non-condensing)
Installation category III Pollution degree 2 Altitude up to 2000m (6560 feet) above
nsulation reference voltage
I
Dielectric str Noise Rejection
CMRR 100 dB, 48 to 62 Hz
EMC
Emissions EN61000-6-3, EN60688
ength
10° to +45°C (14° to 113°F)
-
(R.H. <90% non-condensing)
ea-level
s
00 V
3
4k
residential environment, commerce and light industry ANSI/IEEE C37.90-1989 (surge, withstand and fast transient test)
o ground (AV5 input)
R
t
MS
VAC
RMS
(for 1 min)
Supply Specifications
ulse voltage (1.2/50µs)
P Safety standards IEC60664, IEC61010-1
Measurement standards IEC60688, EN60688,
ANSI C12.20, ANSI C12.1
Approvals CE, cURus and CSA Connections 5(6) A Screw-type
Housing
Dimensions 96x96x140 mm Material ABS,
Protection degree Front: IP65 / NEMA 4x
Weight Approx. 600 g
N61000-4-5
E
N60664, EN61010-1
E
EN62053-22, EN62053-23,
s
max. 2.5 mm2 wir (2x 1.5mm max screw tightening torque: 0.5Nm
self-extinguishing: UL 94 V-0
(packing included)
e
2
);
AC/DC voltage 90 to 260V (standard)
18 to 60V (on request)
Power consumption 30VA/12W (90 to 260V)
20VA/12W (18 to 60V)
Revenue Approval Settings
The access to the programming parameters via front key pad and/or serial communication ports is locked.
The front key pad (up and down keys) allows the displaying of the variables only, while the communication ports allows the
transmission of the variables only
A proper “instrument settings” form must be filled up by the user before equipment supplying.
WM5-96 is supplied with the desired modules plugged and sealed in the proper slots.
WM5-96 fulfils: the ANSI/IEEE C12.20-1998 requirements;
the CAN3-C17-M84 requir
and can be certified according to:
C12.20-1998, class 0.5 (independent labs); AE-0924 Industry Canada Approval.
.
ements;
Page 50
48
Technical Specifications
Function Description
Input and output scaling capability.
orking of the analogue outputs (y) versus input variables (x)
W
WM5-96 Instruction Manual
Figure A
The sign of measured quant­ity and output quantity re-
ains the same. The output
m quantity is proportional to the measured quantity.
Figure B
The sign of measured quant­ity and output quantity changes simultaneously. The output quantity is proportion­al to the measur
ed quantity.
Figure C
The sign of measured quant­ity and output quantity re­mains the same. On the range X0...X1, the output quantity is zero. The range X1...X2 is delineated on the entire output range Y0 = Y1...Y2 and thus pr
esented in
strongly expanded form.
0 50 A 100 A
0 10 mA 20 mA
-100 kW 0 100 kW
-5 mA 0 + 5 mA
80 V 100 V 120 V
0
mA
5
10 mA
Figure D
The sign of measured quant­ity and output quantity re­mains the same. With the measured quantity being zero, the output quantity already has the value Y1 = 0.2 Y2.
ive zero output.
L
Figure E
The sign of the measured quantity changes but the one of the output quantity remains the same. The out­put quantity steadily increas­es from value X1 to value X2 of the measured quantity.
Figure F
The sign of the measured quantity remains the same, the one of the output quantity changes as the measured quantity leaves range X0...X1 and passes to range X1...X2 and vice versa.
0 50 A 100 A
4 12 mA 20 mA
-100 kW 0 100 kW
0 10 mA 20 mA
0 50 A 100 A
- 5 mA 0 + 5 mA
Insulation between inputs and outputs
Meas. /digital
inputs
Meas. /digital
inputs
Relay output
Open collector
Analogue out.
10V, 20mA
Analogue out.
±5mA
AR1034
AR2040
AR1039
90-260VAC/DC
18-60VAC/DC
- 4kV 4kV 2kV 2kV 4kV 2kV 4kV 4kV 4kV
4kV 4kV (*) 4kV 4kV 4kV 4kV 4kV 4kV 4kV 4kV
4kV 4kV 4kV (*) 4kV 4kV 4kV 4kV 4kV 4kV 4kV
2kV 4kV 4kV 4kV (*) 4kV 4kV 4kV 4kV 4kV 4kV
2kV 4kV 4kV 4kV 200V (**) 4kV 4kV 4kV 4kV 4kV
4kV 4kV 4kV 4kV 4kV - - 4kV 4kV 4kV
2kV 4kV 4kV 4kV 4kV - - 4kV 4kV 4kV
4kV 4kV 4kV 4kV 4kV 4kV 4kV - 4kV 4kV
4kV 4kV 4kV 4kV 4kV 4kV 4kV 4kV - -
4kV 4kV 4kV 4kV 4kV 4kV 4kV 4kV - -
NOTE: In case of fault of first insulation the current from the measuring inputs to the ground is lower than 2 mA. (*) The given insulation is granted among outputs plugged in dif
e non insulation among the outputs.
efor
ther
Relay
output
Open
colllector
Analogue out.
10V, 20mA
Analogue out.
±5mA
fer
AR1034 AR2040 AR1039 90-260V
ent slots. The modules equipped with two or four outputs have
(**) Insulation between the 2 outputs of the same module is 200V for 1 min.
AC/DC
18-60V
AC/DC
Page 51
WM5-96 Instruction Manual
å
-
=
LLL L
LL
V
VV
ASY
)(
minm ax
å
-
=
LN
LNLN
LN
V
VV
ASY
)(
minmax
Accuracy
Technical Specifications
49
0.5%
Error
or C0.8
+1%
PF=1 PF=L0.5
+1.0% +
Class 0.5
1%
-0.5%
-1.0%
PF=1
PF=L0.5
or C0.8
sinϕ=1
+2.5%
2%
+
Error
0.05A
(0.01In)
0.1A
(0.02In)
0.25A (0.05In)
0.5A (0.1In)
5A (In)
5A (In)
-
10A (Imax)
10A (Imax)
Class 2
-2%
-2.5%
kWh, accuracy (RDG) depending on the current
lass 0.5 accuracy limits (active energy) EN62053-21
C
5(10A) start-up current: 5mA.
kvarh, accuracy (RDG) depending on the curr
ent
Class 2 accuracy limits (reactive energy) EN62053-23
5(10A) start-up current: 5mA.
ϕ=1
sin
sin
ϕ=0.5
0.1A
(0.02In)
0.25A
(0.05In)
0.25A (0.05In)
0.5A (0.1In)
Used calculation formulas
Phase variables
Instantaneous effective voltage
Instantaneous active power
Instantaneous power factor
Instantaneous effective current
Instantaneous apparent power
Instantaneous r
eactive power
5A (In)
5A (In)
System variables
Equivalent thr
Voltage asymmetry
ee-phase r
Thr
Neutral current
An =
ee-phase active power
Thr
ee-phase appar
Thr
10A (Imax)
10A (Imax)
ee-phase voltage
eactive power
A
L1+AL2+AL3
ent power
Three-phase power factor
(TPF)
Energy metering
Where:
i= considered phase (L1, L2 or L3) P= active power; Q= reactive power; t1, t2=starting and ending time points
of consumption recording; n= time unit;t= time interval between two successive power measurements;
n1, n2= starting and ending discrete
time points of power recording
Page 52
50
Technical Specifications
WM5-96 Instruction Manual
List of the variables that can be connected to:
Analogue outputs (all listed variables with the only exception of energies), alarm outputs (all listed variables with the only exception of energies), pulse outputs (only energies), communication (all listed variables).
No. 1-phase 2-ph. 3-wire 3-ph. 4-wire 3-ph. 3-wire 3-ph. 4-wire Notes
1 V L1 x x x o x 2 V L2 o x x o x 3 V L3 o o x o x 4 V L-N sys o x x o x Sys = system = 5 V L1-2 o x x x x 6 V L2-3 o o x x x 7 V L3-1 o o x x x 8 V L-L sys o o x x x Sys = system =
9A L1xxxxx 10 A L2 o x x x x 11 A L3 o o x x x 12 An o x x o x An=neutral current 13 W L1 x x x x x 14 W L2 o x x x x 15 W L3 o o x x x
W sys o x x x x
16 17 var L1 x x x x x 18 var L2 o x x x x 19 var L3 o o x x x 20 var sys o x x x x Sys = system = 21 VA L1 x x x x x 22 VA L2 oxxxx 23 VA L3 o o x x x 24 VA sys o x x x x Sys = system = 25 PF L1 x x x x x 26 PF L2 o x x x x 27 PF L3 o o x x x 28 PF s ys o x x x x Sys = system= 29 Hz x x x xx 30 ASY VL-N o x x o x Asymmetry of phase-neutral 31 ASY VL-L o o x x x Asymmetry of phase-phase 32 THD V1 xxxox 33 THD V2 o x x o x 34 THD V3 o o x o x 35 THD V1-2 o x x x x 36 THD V2-3 o o x x x
THD V3-1
37 38 THD A1 x x x x x 39 THD A2 o x x x x 40 THD A3 o o x x x 41 THDo V1 x x x o x 42 THDo V2 43 THDo V3 o o x o x
THDo V1-2 o
44 45 THDo V2-3 o o x x x 46 THDo V3-1 47 THDo A1 x x x x x 48 THDo A2 o x x x x 49 THDo A3 o o x x x 50 THDe V1 x x x o x 51 THDe V2 o x x o x
THDe V3
52 53 THDe V1-2 o x x x x 54 THDe V2-3 o 55
THDe V3-1 o o x x x
56 THDe A1 x x x x x
THDe A2
57 58 THDe A3 o o x x x
Phase seq.
59
system system bal. (1 CT) unbal. sys. unbal. sys.
o
o
o
ooxox
oxxxx
o
o
xxox
x
o
o
o x x x Phase sequence
xxx
xxx
x
xxx
x
x
Σ
Σ
Σ
Σ
Σ
Σ
Σ
Σ
Σ
Σ
(x) = available (o) = not available
Page 53
A
RMPowered
PRG
m
easures
SS
R
!
R
AL
WM5-96
SmartPowerQualityAnalyzer
Class0.2A/V
®
WM5-96, ARM
Powered
Customer’s data:
Serial number WM5 96:
Note:
WM5-96 User’s Page
Slot A
Empty AQ1038 AQ1042 AO1037 AO1036 AO1059 AO1058 AO2050 AO2051 AO2052
Slot B
Empty AQ1038 AQ1042 AO1037 AO1036 AO1059 AO1058 AO2050 AO2051 AO2052
AR1034 AR2040
Slot C
Empty AQ1038 AQ1042 AO1037 AO1036 AO1059 AO1058 AO1035 AO2051 AO2052
Slot D
Empty AQ1038 AQ1042 AO1037 AO1036 AO1059 AO1058 AO1035 AO2051 AO2052
Slot E
Empty
AR1039
Power supply
AP1021 AP1020
Measuring inputs
AQ2030 AQ2031
Page 54
Page 55
OUR SALES NETWORK IN EUROPE
A
USTRIA - Carlo Gavazzi GmbH
K
etzergasse 374, A-1230 Wien
T
el: +43 1 888 4112 Fax: +43 1 889 10 53 o
ffice@carlogavazzi.at
BELGIUM - Carlo Gavazzi NV/SA Schaarbeeklei 213/3, B-1800 Vilvoorde T
el: +32 2 257 4120 F
ax: +32 2 257 41 25 sales@carlogavazzi.be
D
ENMARK - Carlo Gavazzi Handel A/S
Over Hadstenvej 42, DK-8370 Hadsten Tel: +45 89 60 6100 F
ax: +45 86 98 15 30 handel@gavazzi.dk
FINLAND - Carlo Gavazzi OY AB Petaksentie 2-4, FI-00630 Helsinki T
el: +358 9 756 2000 F
ax: +358 9 756 20010 m
yynti@carlogavazzi.fi
F
RANCE - Carlo Gavazzi Sarl
Z
ac de Paris Nord II, 69, rue de la Belle
E
toile, F-95956 Roissy CDG Cedex Tel: +33 1 49 38 98 60 F
ax: +33 1 48 63 27 43 f
rench.team@carlogavazzi.fr
GERMANY - Carlo Gavazzi GmbH R
udolf-Diesel-Strasse 23,
D
-64331 Weiterstadt Tel: +49 6151 81000 F
ax: +49 6151 81 00 40
k
ontakt@carlogavazzi.de
GREAT BRITAIN - Carlo Gavazzi UK Ltd 7
Springlakes Industrial Estate,
D
eadbrook Lane, Hants GU12 4UH, GB-Aldershot T
el: +44 1 252 339600
F
ax: +44 1 252 326 799
sales@carlogavazzi.co.uk
OUR SALES NETWORK IN NORTH AMERICA
U
SA - Carlo Gavazzi Inc.
7
50 Hastings Lane, USA-Buffalo Grove, IL 60089 Tel: +1 847 465 6100 Fax: +1 847 465 7373 s
ales@carlogavazzi.com
,
C
ANADA - Carlo Gavazzi Inc.
2
660 Meadowvale Boulevard, CDN-Mississauga Ontario L5N 6M6, Tel: +1 905 542 0979 Fax: +1 905 542 22 48 g
avazzi@carlogavazzi.com
I
TALY - Carlo Gavazzi SpA
V
ia Milano 13, I-20020 Lainate
T
el: +39 02 931 761 Fax: +39 02 931 763 01 i
nfo@gavazziacbu.it
NETHERLANDS - Carlo Gavazzi BV Wijkermeerweg 23, N
L-1948 NT Beverwijk
T
el: +31 251 22 9345 Fax: +31 251 22 60 55 i
nfo@carlogavazzi.nl
NORWAY - Carlo Gavazzi AS Melkeveien 13, N-3919 Porsgrunn T
el: +47 35 93 0800 F
ax: +47 35 93 08 01 gavazzi@carlogavazzi.no
P
ORTUGAL - Carlo Gavazzi Lda
Rua dos Jerónimos 38-B, P
-1400-212 Lisboa
T
el: +351 21 361 7060 Fax: +351 21 362 13 73 carlogavazzi@carlogavazzi.pt
C
ANADA - Carlo Gavazzi LTEE
3
777 Boulevard du Tricentenaire Montreal, Quebec H1B 5W3 Tel: +1 514 644 2544 Fax: +1 514 644 2808 g
avazzi@carlogavazzi.com
S
PAIN - Carlo Gavazzi SA
A
vda. Iparraguirre, 80-82,
E
-48940 Leioa (Bizkaia) Tel: +34 94 480 4037 F
ax: +34 94 480 10 61
g
avazzi@carlogavazzi-sa.es
SWEDEN - Carlo Gavazzi AB N
attvindsgatan 1, S-65221 Karlstad
T
el: +46 54 85 1125 Fax: +46 54 85 11 77 g
avazzi@carlogavazzi.se
SWITZERLAND - Carlo Gavazzi AG Verkauf Schweiz/Vente Suisse S
umpfstrasse 32,
C
H-632 Steinhausen Tel: +41 41 747 4535 F
ax: +41 41 740 45 40
v
erkauf_vente@carlogavazzi.ch
OUR SALES NETWORK IN ASIA AND PACIFIC
SINGAPORE - Carlo Gavazzi Automation Singapore Pte. Ltd. N
o. 178 Paya Lebar Road #04-01/05 409030 Singapore Tel: +65 67 466 990 Fax: +65 67 461 980
MALAYSIA - Gavazzi Automation Sdn Bhd. 5
4, Jalan Rugbi 13/30, Tadisma Business Park Seksyen13 40100 Shah Alam, Selangor Tel: +60 3 55 121162 Fax: + 60 3 55 126098
OUR PRODUCTION SITES
Carlo Gavazzi Industri A/S Hadsten -
DENMARK
Tel: +45 89 60 6100
Carlo Gavazzi Industri Kaunas -
LITHUANIA
el: +370 3732 8227
T
Carlo Gavazzi Ltd Zejtun - MALTA Tel: +356 23 601 100
Carlo Gavazzi Automation (Kunshan) Co., Ltd. Kunshan -
CHINA
Tel: +86 512 5763 9300
HEADQUARTERS
Carlo Gavazzi Automation SpA Via Milano, 13 - I-20020 Lainate (MI) ­Tel: +39 02 93 76 info@gavazzi-automation.com www
ITALY
.carlogavazzi.com/ac
CARLO GAVAZZI
Automation Components
CHINA - Carlo Gavazzi Automation (China) Co. Ltd. N
o. 1001 Shangbu Middle Road, Futian, Shenzhen Tel: +86 755 83699500 Fax: +86 755 83699300
Carlo Gavazzi Contr
ITALY
Belluno ­Tel: +39 0437 931 000
ols SpA
HONG KONG - Carlo Gavazzi Automation Hong Kong Ltd. U
nit 3 12/F Crown Industrial Bldg., 106 How Ming St., Kowloon, Hong Kong Tel: +852 23041228 Fax: +852 23443689
SAIET Elettr Castel Maggiore (BO) ­Tel: +39 051 417 8811
onica SpA
Further information on www.carlogavazzi.com/ac
ITALY
Loading...