CareFusion Micro Diary Operating Manual

Page 1
Micro Diary
Operating Manual
Page 2
1
INTRODUCTION ........................................................... 3
INDICATIONS FOR USE................................................. 5
TARGET POPULATION .................................................. 5
ENVIRONMENT OF USE ............................................... 5
ENVIROMENT CONDITION OF USE .............................. 6
CONTRAINDICATION, WARNINGS AND CAUTIONS. ..... 6
CONTRAINDICATIONS ..................................................... 6
CAUTI ON: ................................................................... 7
PA CKA GE CO NT ENTS ................................................... 8
CONFIGURATION ......................................................... 9
OPERATION ............................................................... 11
SP I R O M E T R Y RE C O R D I N G ............................................... 11
SYMPTOM SC O R E E N T R Y ................................................ 14
MENU OPTIONS .......................................................... 18
1 PC CONNECTION ...................................................... 19
2 MESSAGE LOG ......................................................... 19
3 TRAINING BLOW ....................................................... 19
4 SE R I A L NU M B E R ....................................................... 19
5 UNIT ID ................................................................. 19
6 RE C O R D CO U N T ....................................................... 19
7 CA L I B R A T I O N CHECK ................................................. 20
8 TELEPHONE LO G ....................................................... 20
MAINTENANCE .......................................................... 20
BATTE RY MANAGEMENT ................................................. 20
BATTERY REPLACEMENT ................................................. 21
CALIBRATION CHECK .................................................... 21
CLEANING THE TRANSDUCER .......................................... 22
SERVICING ................................................................. 23
TROUBLE SHOOTING INFORM AT I ON......................... 23
ELECTROMAGNETIC COMPATIBILITY (EMC) .............. 24
Page 3
2
CONSUMABLES & SUPPORT I N G P R O D U C T S .............. 24
SPECIFICATION OF THE MICRO DIARY ....................... 25
SYMBOLS ................................................................... 26
CUSTOMER CONTACT INFORMATION ....................... 27
UK CUSTOMERS ONLY ................................................... 27
INTERNATIONAL CUSTOMERS ONLY ................................... 27
Page 4
3
Introduction
The Micro Diary spirometer is a compact, battery operated, and fully portable data recording spirometer designed specifically to collect FEV1, PEF, FVC or F E V6, and symptom scores in epidemiological studies and drug trials.
The Micro Diary connects to a PC for configuration and uploading of stored data using the Micro Diary System Windows™ based software. This software allows the unit to be easily customised for a particular trial and includes many built in safeguards to ensure the integrity of the accumulated data.
The unit can be configured to take spirometry readings and symptom scores, entered either numerically, on an analogue scale, or from a scroll down list of responses. The results are stored together with the time and date of testing. Various identifiers such as trial centre, unit number, and description are also stored.
Different sequences of questions and spirometry tests may be configured for 4 individual time zones. Each sequence is activated by the subject turning the unit on, within the relevant time zone, or by answering an option al alar m. After completing a sequence the unit turns off automatically. Having completed a time zone sequence, a supplementary sequence will be activated if the unit is turned on again within the time zone or between zones. Note that once a time zone sequence has been completed,
Page 5
4
the sequence cannot be repeated until the following day. During a sequence the last spirometry test or questionnaire response can be erased by pressing ‘del’.
The recorded tests can be uploaded to a computer for future analysis or printing, via a USB interface.
The spirometer uses the C a r e F u s i o n Digital Volume Transducer, an extremely stable form of volume transducer, which measures expired air directly at B.T.P.S (Body Temperature and Pressure with Saturated water vapour) thus avoiding the inaccuracies of temperature corrections. This transducer is insensitive to the effects of condensation and temperature and avoids the need for individual calibration prior to performing a test.
Page 6
5
Indications for Use
The Micro Diary Spirometer is used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs. Specifically, the Micro Diary Spirometer measures the following lung function parameters: FEV1, FVC, FEV6, an d PEF. The device also records the test data for later review and has the ability to transfer these records to a compatible computer.
Target Population
The device can be used on patients who require lung function measurements. It can be utilized for patients from 4 years and older, providing that they are able to follow the medical practitioner’s instructions.
Environment of Use
The environment of use is the hospital or a doctor’s or medical practitioner’s office or clinic or the patient’s home.
Page 7
6
Environment
Please observe the following precautions:
Avoid exposing the Micro Diary to direct sunlight.
Avoid operating the spirometer in dusty conditions or
near to heating appliances or radiators
Do not keep the spirometer in a damp or expose it to extreme temperatures
Operate the equipment under normal enviromental temperatures which are defined in this manual (see section “Specification of the Micro Diary”)
The transducer should not be hold towards a strong light source during operation of the spirometer.
Contraindications, Warnings and Cautions.
Contraindications
Acute disorders affecting test performance (e.g
Vomiting , nausea, vertigo)
Recent eye surgery (increases in intraocular
pressure during spirometry)
Oral or facial pain exacerbated by a mouthpiece
Page 8
7
Caution:
CAUTION: Read the manual before use
WARNING: The instrument is not suitable for use in
the presence of explosive or flammable gases,
flammable anaesthetic mixtures or in oxygen rich
environments.
CAUTION: Mouthpieces are single patient use. If used
on more than one patient there is a risk of cross-
infection. Repeat use may degrade materials and lead
to an incorrect measurement.
PLEASE NOTE: The product you have
purchased should not be disposed of as
unsorted waste. Please utilise your local
WEEE collection facilities for the disposal of this
product.
Page 9
8
Package Contents
The Micro Diary is packaged in a convenient carrying case and comes complete with the following items (Fig.1):
1. Micro Diary microcomputer unit.
2. Ca reFu sion Digital Volume Transducer.
3. Lithium coin cell, CR2450
4. USB cabl e
Figure 1: The Micro Diary
Delete
Enter
On
Scroll down
Scroll up
Page 10
9
Configuration
Before using the Micro Diary the required sequence of questions and spirometry tests must be defined using the PC software (Please see Help files on software for guidance ). Micro Diary PC Software may be downloaded using the link:
http://www.micromedical.co.uk/update/download.asp
(The PC Software is intended for use by Clinicians and Trial co-ordinators only).
Once defined, the sequence can be loaded into the Micro Diary. This can be done with a direct connection to a USB port of the PC: plug the unit into a USB port on the PC, using the cable provided, and turn the unit on with the ‘enter’ key depressed to display the operations menu. Once the operations menu is displayed, release the ‘enter’ key. Scroll to option 1, PC Connection, using the up and down keys and then press ‘enter’ (the other options are described in section Menu Options). From the PC software, download the program to the unit and switch off by pressing ‘del’.
The Micro Diary is now ready to perform the defined sequence of questions when the unit is turned on. The subject is instructed to complete the sequence of questions and spirometry tests. Once complete, the unit will turn off automatically.
If the subject turns the unit on but leaves the unit before completing the sequence, the unit will automatically turn
Page 11
10
off after several minutes. Data collected on a partially completed sequence will not be stored and the subject must turn the unit on again and complete the whole sequence before the results are stored.
If an alarm is set then the unit will turn on automatically at the required time, sound an audible alarm, and display:
The subject should then press ‘enter’ to initiate the sequence. If no response is given the unit will turn itself off after a few minutes (configurable) and repeat the alarm every 11 minutes (configurable) until the end of the time zone.
When turned on, the subject will be requested to perform a spirometry manoeuvre or to enter the answer to questions in the order specified in the program.
In order to familiarise the subject with the spirometry manoeuvre a test may be performed without the results being stored. To do this, turn the unit on in the ‘options’ menu, as described above. Then scroll to option 3, training Blow and press ‘enter’. A spirometry manoeuvre will be requested as outlined in the next section. Pressing ‘del’ will return the unit to the options menu.
Page 12
11
Operation
Spirometry Recording
When taking a spirometry measurement the following will be di splay e d :
The subject should then perform a spirometry manoeuvre using a clean mouthpiece connected to the transducer. After completion the display will show the results together with a spirometry manoeuvre quality check:
Page 13
12
The results and quality checks displayed to the user may be configured with the PC software. The software may also be configured to automatically reject any manoeuvre that fails any of the quality checks. The unit will ‘beep’ once at the end of the manoeuvre or three times if an alarm has been set and the result is below the acceptable level. The arrow at the bottom right of the screen indicates that the ‘down’ arrow key may be used to obtain more results. Using the ‘down’ arrow key will di spl a y t h e flow/volume curve, if configured by the PC software:
Page 14
13
The Quality checks can result either in discarding the blow or not. See the Quality Assurance Options page from the help files.
When an acceptable test has been performed, pressing ‘enter’ will move on to the next test or symptom score entry and save the spirometry results. When performing a series of tests it is possible to configure the unit to record only the best blow.
The unit may also be configured to alarm when the spirometry manoeuvre is below a set value or below a set percentage of the recorded best effort. When a manoeuvre is below the required level an alarm will sound continuously and the display will flash the reading, which has triggered the alarm.
If for any reason the subject is not satisfied with their effort, pressing ‘del’ will cause the test to be repeated and the results will not be stored until ‘enter’ is pressed.
Repeatedly pressing ‘del’ will erase, in turn, all the tests performed and may be used to erase a whole sequence if, for example, the subject decides that they are not able to complete a set of tests at that time.
Please note that the results to be displayed, and stored, are individually selected by the program. It is therefore possible, for example, to store FEV1 and FVC but display no results if a blind study is to be undertaken. In this case the Micro Diary will move on to the next question or spirometry test immediately after the test is performed unless the effort is below the minimum requirement in
Page 15
14
which case the alarm will sound and the display will flash the reading that triggered the alarm.
Symptom Score entry
Numerical data and responses to questions may be entered in five different ways.
Question with Numeric Answer:
The question is displayed at the top of the screen and the default response is shown flashing below. Note that when an up or down key is first used, the default number stops flashing. Subsequent use increments the number up or down.
In order to avoid the possibility of the subject si m pl y pressing ‘enter’ repeatedly to accept all default values in a series of questions, the ‘enter’ key is only recognised after an ‘up’ or ‘down’ key is used at least once.
Question with Decimal Numeric Answer:
Page 16
15
The display shows the question number with the default response (set by the PC software) to one decimal place. The subject adjusts the response, as described above, and then presses ‘enter’. Note that it is not possible to enter the default value whist the display is flashing. Pressing the up or down once will stop the flashing and subsequent use will increment or decrement the value. This feature ensures that the subject actively reviews the value and does not repeatedly press ‘enter’ for a sequence of questions in order to save time.
Question with Short Written Answer:
When this type of question is displayed the user uses the ‘up anddown arrow keys to select the correct response and then presses ‘enter’. If there are more answers than can be displayed on the screen, an arrow w i l l b e displayed to indicate that further answers are available. As before,
Page 17
16
the default response flashes until the ‘up’ or ‘down’ arrow keys are first used.
Question with Long Written Answer:
Page 18
17
When the answers are too long to fit on one line, paged responses may be used. Using the up and down keys will scroll between the programmed responses. Pressing the ‘enter’ key will select the required response.
Question with Analogue Scale Response:
Question requiring an answer on an analogue displ ay c a n also be displayed. The user uses the up and down arrows to move the solid bar to the right and left. When the required level is displayed press ‘enter’ to store the result.
Page 19
18
Menu Options
To obtain the options menu hold down the enter key, turn the unit on, and when the sounder is heard release the ‘enter’ key.
Use the up and down arrow keys to select the required option and then press enter.
Select option 9, Exit, when you have finished using the options.
Page 20
19
1 PC Connection
When this option is used connect the Diary unit to a PC running the software supplied to a USB port using the cable supplied. The Diary is then controlled by the PC in order to upload saved data or to reprogram the unit.
2 Message Log
The feature is not used in the Diary system.
3 Training Blow
Use this option to allow the user to familiarise themselves with performing a spirometry test. None of the tests performed in this mode of operation will be stored.
4 Serial Number
Use this option to identify the serial number of the unit.
5 Unit ID
Use this option to display the programmed Unit ID and the Full ID. These identifiers, loaded into the Diary at the time of programming.
6 Record Count
A record count and the percentage of memory used is av ail a ble with this option.
Page 21
20
7 Calibration Check
Use this option to perform a calibration check. A calibration syringe will be required.
8 Telephone Log
This function is not used in the Micro Diary.
Maintenance
Battery Management
The Diary is powered by a 3 Volt Lithium coin cell accessed via a slide cover on the rear of the instrument. When the battery is low a warning icon will be displayed and the battery should be replaced:
The recorded data is stored in non-volatile memory and no data will be lost when the battery expires. It is recommended that the battery is replaced at the commencement of each new study.
Page 22
21
Battery replacement
For access to the battery first remove the battery cover by sliding sideways as indicated
Disassembly Prise out the old coin cell
using a suitable, non­metallic implement
Assembly Replace the battery paying particular attention to its
orientation. The part number etched on its outer face should be uppermost and visible when the battery is inserted.
Press until clicked into pl a ce Slide the battery cover back into place.
Calibration Check
The Micro Diary is calibrated to read in litres at body temperature, barometric pressure saturated with water vapour (BTPS).
The calibration should remain stable indefinitely, unless the transducer is physically damaged, and the unit should not require re-calibration. However, to ensure the correct functioning of the unit we do recommend that a calibration check is performed periodically.
Page 23
22
To perform a calibration check turn the unit on with the ‘enter’ key pressed and select ‘Calibration Check’ from the menu options, see section Menu Options
Connect a 3L syringe to the Micro Diary with the minimum of adapters and the syringe volume should be injected into the transducer evenly, without pausing.
A 3 Litre Syringe is available from Care Fusi on as Cat.No.36-SM2125
If, after checking, the unit appears to require re­calibration then please refer to your supplier for service.
Cleaning the Transducer
The transducer requires no routine maintenance or servicing. However, if you wish to sterilise or clean the transducer it may be removed by means of the following procedure:
1. Remove the transducer by gently pushing out from the rear of the instrument.
2. The transducer may now be immersed in warm soapy water for routine cleaning or immersed in cold sterilising solutions e.g. Perasafe for a period not exceeding 15 minutes. (Alcohol and chloride solutions should be avoided.) After cleaning/sterilising, the transducer should be rinsed in distilled water and dried.
3. Re-assemble the transducer.
Page 24
23
Servicing
If your unit requires service or repair please contact your nearest service dept. See the Customer Contact Information section for more details.
Trouble Shooting Information
Should you encounter problems operating your spirometer, please consult the table below:
Problem
Possible Cause
Solution
Micro Diary cannot be switched on
Batteries are flat
Change the batteries
Every time you switch the instrument on the time is shown as 00:00
The internal battery of Micro Diary is defective
Contact your dealer.
Micro Diary is outside of % when conducting the calibration check
There are leaks in the syringe or connections
Check the syringe and connections for leaks
Page 25
24
Electromagnetic Compatibility (EMC)
Changes or modifications to the Micro Diary spirometer that are not expressly approved by Carefusion can cause EMC issues with this or other equipment. The device is designed and tested to comply with applicable regulation regarding EMC and needs to be installed and put into service as per interference exceeding levels specified in EN 50082-1:1992
WARNING: use of portable phones or other radio frequency (RF) emitting equipment near the system may cause unexpected or adverse operation
WARNING: No modification to this device is allowed.
Consumables & Supporting Products
Cat. No.
Description
36-CAB1098
USB A cable for MD03 PC connection.
36-PSA2200
Mouthpieces
36-SSC5000A
PeraSafe Sterilising Powder 81g
(to make up 5 litres of solution)
36-VOL2104
Nose C l i ps (pack of 5)
36-SM2125
3 Litre Calibration Syringe
36-TDX1050
CareFusion Digital Volume Transducer
Page 26
25
Specification of the Micro Diary
Measurements:
Forced Vital Capacity (FVC) or Forced Expired Volume in 6 seconds (FEV6)
Forced Expired Volume in 1 second (FEV1)
Peak Expiratory Flow rate (PEF)
Accuracy:
Meets the accuracy requirements of the ATS/ERS Standardisation of Lung Function Testing: Standardisation of Spirometry - 2005
Storage Capacity:
Typically > 1 year of data recording.
Display:
64 x 100 pixel LCD
Transducer Type:
CareFusion Uni-Directional Digital Volume.
Resolution:
10ml
Volume Range:
0 to 8 litres.
Power supply:
CR2450 Lithium coin cell.
Backup battery life:
>10 years
Main Battery Life:
Approximately 24 hours of continuous use.
Dimension s :
130 X 57 X 37mm.
Weight:
100g
Operating Temperature/ Humi dity:
0 to +40 °C / 30% to 90%
RH
Storage Temperature/ Humi dity:
-20 to + 70 °C / 10% to 90% RH
Page 27
26
Symbols
Type B device
0086
In accordance with Directive 93/42/EEC
Disposal in compliance with WEEE
Consult the instructions for use Date of manufacture
Manufacturer
Serial Number
(Rx only) : Federal U.S. law restricts this device to sale by or on the order of a physician
Loading...