Cardioline ECG100S, ECG200S User Manual

4 (2)

ECG100S and ECG200S

User manual

Rev.5 – 26.02.2018 1936

ECG100S and ECG200S

All rights reserved © Cardioline SpA.

CARDIOLINE® is a registered trademark of Cardioline SpA.

This publication may not be reproduced, in whole or in part, in any form or manner, without prior written authorisation by:

Cardioline Spa

Via Linz, 151

38121 Trento

Italy

ECG100S and ECG200S

Table of Contents

1.

GENERAL INFORMATION .........................................................................................................................

1

 

1.1.

Other important information...........................................................................................................

1

2.

SAFETY INFORMATION.............................................................................................................................

3

 

2.1.

Wireless Data Transmission .............................................................................................................

8

 

2.2.

Wirless Option .................................................................................................................................

8

3.

ELECTROMAGNETIC COMPATIBILITY (EMC).............................................................................................

11

 

3.1. Guidance and Manufacturer's declaration - Electromagnetic emissions.........................................

12

 

3.2. Guidance and Manufacturer's declaration - Electromagnetic immunity .........................................

13

 

3.3. Guidance and Manufacturer's declaration - Electromagnetic immunity .........................................

13

3.4.Recommended separation distances between portable and mobile RF communications

 

equipment and the ECG100S/ECG200S .......................................................................................................

15

 

3.5.

Transmission frequencies ................................................................................................................

15

4.

SYMBOLS AND LABEL ...............................................................................................................................

17

 

4.1.

Explanation of the symbols..............................................................................................................

17

 

4.2.

Device label......................................................................................................................................

18

5.

INTRODUCTION........................................................................................................................................

19

 

5.1.

Purpose of the manual.....................................................................................................................

19

 

5.2.

Recipients.........................................................................................................................................

19

 

5.3.

Intended use ....................................................................................................................................

19

 

5.4.

Description of the device .................................................................................................................

20

 

5.4.1.

General overview .....................................................................................................................

21

 

5.4.2.

Keyboard..................................................................................................................................

24

 

5.4.3.

Display......................................................................................................................................

25

6.

PREPARATION FOR USE............................................................................................................................

27

 

6.1.

Initial startup....................................................................................................................................

27

 

6.2.

Patient cable connection .................................................................................................................

27

 

6.3.

Loading the paper............................................................................................................................

28

 

6.4.

Power connection ............................................................................................................................

29

 

6.5.

Battery operation.............................................................................................................................

30

7. EXECUTION OF AN EXAM.........................................................................................................................

31

 

7.1.

General procedure...........................................................................................................................

31

 

7.2.

Before acquisition ............................................................................................................................

31

ECG100S and ECG200S

7.2.1.

Preparing the patient...............................................................................................................

31

7.2.2.

Connecting the patient ............................................................................................................

32

7.3.

Viewing the ECG...............................................................................................................................

34

7.3.1.

Disconnection of the leads.......................................................................................................

35

7.4.

Entering patient data .......................................................................................................................

36

7.4.1.

Manually entering the patient's data.......................................................................................

36

7.4.2.

Automatically entering the patient's data................................................................................

37

7.5.

Acquisition of an ECG.......................................................................................................................

38

7.5.1.

Automatic acquisition of an ECG (AUTO) .................................................................................

38

7.5.2.

Manual acquisition of an ECG (MANUAL) ................................................................................

40

7.6.

Identifying urgent ECGs (STAT) ........................................................................................................

41

7.7.

Printing an ECG ................................................................................................................................

42

7.7.1.

Printing formats .......................................................................................................................

42

7.8.

Storing an ECG .................................................................................................................................

43

7.9.

Transmitting an ECG ........................................................................................................................

43

8. MENU AND DEVICE SETTINGS..................................................................................................................

45

8.1.

Main menu.......................................................................................................................................

45

8.2.

Work lists .........................................................................................................................................

46

8.3.

ECG archive......................................................................................................................................

47

8.3.1.

Archive full ...............................................................................................................................

48

8.4.

Date and time ..................................................................................................................................

48

8.5.

Print configuration ...........................................................................................................................

49

8.6.

Settings ............................................................................................................................................

50

8.6.1.

General settings (page 1) .........................................................................................................

50

8.6.2.

Default ECG settings (page 2)...................................................................................................

51

8.6.3.

Manual printing settings (page 3) ............................................................................................

52

8.6.4.

Automatic printing settings (page 4)........................................................................................

52

8.6.5.

Connectivity Settings (page 5)..................................................................................................

53

8.6.6.

LAN Settings (page 6) ...............................................................................................................

53

8.6.7.

WLAN Settings (page 7) ...........................................................................................................

54

8.6.8.

WiFi security Settings (page 8).................................................................................................

55

8.7.

Protection of the Settings ................................................................................................................

55

9. CONNECTIVITY, RECEIVING WORK LISTS AND TRANSMITTING ECG EXAMS ............................................

57

9.1.

General ............................................................................................................................................

57

9.1.1.

Configuring LAN connectivity...................................................................................................

57

ECG100S and ECG200S

 

9.1.2.

Configuring WLAN connectivity ...............................................................................................

58

9.2.

USB connectivity ..............................................................................................................................

58

 

9.2.1.

Connection to a computer .......................................................................................................

59

 

9.2.2.

Transfer data to a USB memory stick.......................................................................................

59

9.3.

Connectivity LAN/WLAN ..................................................................................................................

60

10. UPGRADING YOUR DEVICE OPTIONS .......................................................................................................

61

11.

MAINTENANCE AND TROUBLESHOOTING ...............................................................................................

63

11.1.

Precautions ......................................................................................................................................

63

11.2.

Switching off the device...................................................................................................................

63

11.3.

Regular maintenance .......................................................................................................................

63

 

11.3.1.

Functional check ......................................................................................................................

63

 

11.3.2.

Patient cable cleaning ..............................................................................................................

64

 

11.3.3.

Device cleaning ........................................................................................................................

64

 

11.3.4.

Operation check.......................................................................................................................

64

11.4.

Recommendations ...........................................................................................................................

65

11.5.

Battery maintenance .......................................................................................................................

65

11.6.

Cleaning the thermal printer............................................................................................................

67

 

11.6.1.

Cleaning the printer .................................................................................................................

67

 

11.6.2.

Cleaning the thermal head of the printer ................................................................................

67

11.7.

Troubleshooting table......................................................................................................................

67

11.8.

Messages and solutions table ..........................................................................................................

68

12.

TECHNICAL SPECIFICATIONS ....................................................................................................................

71

12.1.

Harmonised standards applied ........................................................................................................

72

12.2.

Accessories.......................................................................................................................................

73

13.

WARRANTY ..............................................................................................................................................

75

14.

DISPOSAL

.................................................................................................................................................

77

ECG100S and ECG200S

ECG100S and ECG200S

1. GENERAL INFORMATION

1.GENERAL INFORMATION

This manual is an integral part of the device and should always be available as support material to the clinical practitioner or the operator. Strict compliance with the information contained in this manual is an essential prerequisite for a proper and reliable use of the device.

Have the operator read the manual thoroughly as the information related to the different chapters is only described once.

1.1.Other important information

This manual was written with the utmost care. Should you find any details which do not correspond to those contained in this manual, please inform Cardioline SpA who will correct such inconsistencies as soon as possible.

The information contained in this manual is subject to change without notice.

All changes will be in compliance with the regulations governing the manufacturing of medical equipment.

All trademarks mentioned in this document are property of their respective owners. Their protection is guaranteed.

No part of this manual may be reprinted, translated or reproduced without the manufacturer's written authorisation.

The code relating to this manual is listed below.

Language

Code

 

 

ENGLISH

36510209_ENG

 

 

1

ECG100S and ECG200S

1. GENERAL INFORMATION

2

ECG100S and ECG200S

2. SAFETY INFORMATION

2.SAFETY INFORMATION

Cardioline SpA will be held responsible for the safety, reliability and functionality of the devices only if:

1.the assembly operations, modifications or repairs are carried out by Cardioline SpA or by its Authorised Service Centre;

2.The device is used in compliance with the instructions provided in the use manual.

Always contact Cardioline SpA should you wish to connect any devices not mentioned in this manual.

Warnings

This manual provides important information on proper use and safety of the device. Failure to comply with the described operating procedure, improper use of the device, ignoring the specifications and recommendations supplied, may cause severe physical injuries to the operators, patients and bystanders, or may damage the device.

No modification of this equipment is allowed.

The device captures and presents the data that reflects the physiological condition of the patient; this information can be examined by specialist medical staff and will be useful in providing an accurate diagnosis. In any event, the data cannot be used as the only means to make an accurate diagnosis of the patient.

The operators for whom this device is intended must have the required competence regarding medical procedures and the treatment of patients. They must also be sufficiently trained in using the device. Have the operator carefully read and understand the contents of the operator manual and the other annexed documents before using the device for clinical applications. Inadequate knowledge or training could be at a greater risk for the physical safety of operators, patients and bystanders, or could damage the device. In case the operators are not trained in using the device, we recommend to contact Cardioline or its Authorised Distributors to plan adequate training courses.

The device and its power supply are classified as ME Equipment, since the power supply is considered an integral part of the device.

US Federal law restricts this device to sale by or on the order of a Physician.

To avoid the risk of electric shock, this equipment must only be connected to a supply main with protective earth.

If in doubt regarding the integrity of the external earth conductor, use the device with its inner battery.

The positioning of the device must be such as to don't make difficult the operation of disconnection from main supply when an external power supply is used. The plug of the main supply is the main

3

ECG100S and ECG200S

2. SAFETY INFORMATION

switch used to disconnect the device from the main supply. Please, be sure to keep it near the device.

All input and output signal connectors (I/O) are intended to be used only for connection to appropriate devices which comply with IEC 60601-1 standards or further IEC standards (e.g. IEC 60950). Connecting additional equipment to the device could increase leakage current to the chassis and/or patient. To avoid endangering the safety of the operator and patient, keep in mind the requirements of IEC 60601-1:2005+A1 clause 16 and measure the leakage current to confirm that there is no risk of electric shock.

For the correct operation of the device and for the safety of the operators, patients and bystanders, the device and the accessories must be exclusively connected as outlined in this manual. Do not connect telephone lines to the LAN connector.

To maintain immunity from potential interference of electromagnetic signals, a system must be used with shielded cables when connecting the device to the mains.

To guarantee the safety of the operator and of the patient, the equipment connected to the same line as the device must comply with IEC 60950 or IEC 60601-1 standards.

The power cable shielding (when present) must be connected to an earthing system appropriate for the area where the device is used. This will avoid electric shocks caused by different earth potentials which could exist between the various points of an electricity distribution system, or else by failures of the external equipment connected to the mains.

The safety of the patient and the operator is guaranteed if the peripheral units and the accessories that can come into direct contact with the patient comply with the UL 60601-1, IEC 60601-1 and IEC 60601-2-25 standards. Only use spare parts and accessories supplied with the device and available from Cardioline SpA. Refer to section 12.2 for the list approved accessories.

The patient cables to be used with the device are defibrillation-proof. Check the patient cables for ruptures or cracks before use.

Conductive parts of the patient cable, electrodes and associated connections of type CF applied parts, including the neutral conductor of the patient cable and electrode, should not come into contact with other conductive parts, including earth ground.

Defibrillation protection of the ECG relies on the use of the provided ECG cable and the use of any other ECG cable may impair the safe use of the equipment leading to electric shock for the patient and operator. Refer to section 12.2 for the list approved accessories.

To prevent death or any serious personal injuries during defibrillation, avoid contact with the device or patient cables. It is furthermore necessary to properly position the defibrillation pads with respect to the electrodes in order to minimize patient skin burns.

This device is designed to be used only with the electrodes specified in this manual. Strictly follow the correct clinical procedures to prepare the skin before the application of the electrodes and monitor the patient in order to avoid any irritation, inflammation or other skin reactions. The electrodes are designed for short-term applications and must be promptly removed once the examination is complete. Refer to section 12.2 for the list approved accessories.

The ECG electrodes may cause skin irritation; check the skin for any irritations or inflammations.

To prevent any infections, use the disposable components (e.g. the electrodes) only once. To ensure safety and use efficiency, do not use electrodes after their expiration date.

4

ECG100S and ECG200S

2. SAFETY INFORMATION

The quality of the signal produced by the electrocardiograph may be adversely affected by the use of other medical equipment such as defibrillators and ultrasound machines.

The device is intended for external use and is not intended for direct cardiac application.

There is a risk of explosion. Do not use the device in the presence of flammable anaesthetics.

There is no safety hazard if other equipment, such as pacemakers or other stimulators, is used simultaneously with the device; however, disturbance to the signal may occur.

The device is not designed for use with high-frequency (HF) surgical equipment, and does not provide any protective means against hazards to the patient.

The operation may be adversely affected by the presence of strong magnetic fields such as those produced by electrosurgery equipment.

The use of the device is not recommended in the presence of medical diagnostic imaging equipment such as the Magnetic Resonance Imaging (MRI) or Computerised Axial Tomography (CAT) in the same environment.

The following equipment may cause interference with the RF channel: microwave ovens, diathermy units with LANs (spread spectrum), amateur radios, and government radar.

Only use the recommended batteries. Using other types of batteries may cause danger of fire or explosion.

The internal rechargeable battery is hermetically sealed NiMH and requires no maintenance. Should the battery be faulty, contact Cardioline's technical assistance service.

The low battery warning is designed for the recommended batteries only. Using other types of batteries may lead to a lack of indication resulting in device failure. If the battery is low, connect the device to the electrical mains.

This device can be equipped with a wireless LAN (WLAN) module for transmitting ECG records. Device labeling will indicate if the device is equipped with such a module. If so equipped, the WLAN identification can be found on a label on the bottom of the device:

Shenzen Kexian Technology KX-WUL150 (model subject to change without notice)

Use of the WLAN module may interfere with other equipment operating in the vicinity. Check with local authorities or spectrum management officials in your facility to determine if restrictions apply to the use of this feature in your area.

The WLAN modules complies with applicable RF safety standards including standards and recommendations for the protection of public exposure to RF electromagnetic energy that have been established by governmental bodies and other qualified organizations, such as the following:

Directives of the European Community

Directorate General V in Matters of Radio Frequency Electromagnetic Energy

To ensure compliance with current regulations limiting both maximum RF output power and human exposure to radio frequency radiation, a separation distance of at least 20 cm must be maintained between the device's antenna and the head and body of the user and any nearby persons at all times. To help prevent degradation of RF signal and to avoid excess RF energy absorption, do not touch the device during data transmission.

5

ECG100S and ECG200S

2. SAFETY INFORMATION

The device is not intended as a general purpose storage device, thus no files should be stored except from those automatically created by the device itself. Using the electrocardiograph as a general purpose storage device may results in unwanted radio frequency emissions.

Do not clean the device or the patient cables by submersing them in liquid, autoclaving, or steam cleaning. This may cause serious damage to equipment or reduce its lifespan. Using non-specific detergents/disinfectants, failure to comply with the recommended procedures or contact with nonspecific materials may cause additional risks to operators, patients or bystanders or may damage the device. Do not sterilise the device or the patient cable with ethylene oxide gas (EO). Refer to Section 11 for instructions on proper cleaning and disinfection.

Do not leave the patient cable unattended in the presence of children as they could be accidentally strangled.

Do not leave the electrodes unattended in the presence of children as they could cause suffocation if accidentally swallowed.

Attention

To prevent any damage to the keyboard do not use sharp or heavy objects to press the keys, only use your fingertips.

The device and the patient cable should be cleaned before use. Check the connections for any damage or excessive wear before each use. Replace the patient cable should it present any damage or be excessively worn.

Do not pull or stretch patient cables as this could result in mechanical and/or electrical failures. Patient cables should be stored after forming them into a loose loop.

There are no user-serviceable parts inside the device. The device can only be dismantled by qualified service personnel. Any malfunctioning or defective device must be excluded from use and be checked/repaired by qualified service personnel before being reused.

The device does not require any calibration or special instrumentation for correct use and maintenance.

When it is necessary to dispose of the device, its components and accessories (e.g.: batteries, cables, electrodes) and/or packaging material, comply with local standards for waste disposal.

Notes

The movements of the patient may generate excessive noise and affect the quality of the ECG tracing or the correct analysis of the device.

An appropriate preparation of the patient is important in order to guarantee a proper application of the ECG electrodes and the correct operation of the device.

The incorrect positioning of the electrodes for the detection of the algorithm depends on the normal physiology and on the order of the ECG leads and tries to identify the most likely exchange.

6

ECG100S and ECG200S

2. SAFETY INFORMATION

However, it is recommended to check the positioning of the electrodes of the same group (limbs or chest).

If the electrodes are not properly connected to the patient, or one or more patient leads are damaged, the display will indicate a lead failure. When the ECG is printed, these leads will come out on paper as a square wave.

As defined by the IEC 60601-1 and IEC 60601-2-25 safety standards, the device is classified as follows:

IP class equipment (internal power ME) - class I on external AC/DC power supply

Defibrillation-proof Type CF applied parts.

Ordinary equipment.

Not suitable for use in the presence of flammable anaesthetics.

Continuous operation

NOTE: From a safety view point, the external power supply is declared “Class I” based on IEC 60601-1 standard. A three-pole plug is used to guarantee earthing together with the power lines. The earth terminal of the power cable is the only point where the unit is earthed. Exposed metal parts which are accessible during standard operation have a double insulation from the power lines. The internal earth connections are a functional earthing.

Accuracy of measurements taken with the device is compliant with IEC 60601-2-25.

Power Supply has the following characteristics:

Model: AFM60US18

Manufacturer: XP Power Limited

Rated Input: 100-240 VAC, 50-60 Hz, 1.5-0.9 A

Rated Output: 60 W, 18 V, 3.34 A

Protection Class: I

Degree of Protection: IP20

The device is a Class IIa in compliance with Directive 93/42/EEC.

The device is a prescription device according to FDA regulation

In order to prevent damage to the device during transportation and storage (when still in its original packaging), comply with the following environmental conditions:

Ambient temperature ..................... 5°C to 40°C Relative humidity ............................ 20% to 90%

Atmospheric pressure ..................... 700 mBar to 1060 mBar

The device is intended for use in hospitals or doctor's offices and should comply with the following environmental requirements:

7

ECG100S and ECG200S

2. SAFETY INFORMATION

Ambient temperature .....................

+10°C to 40°C

Relative humidity ............................

50% to 90%

Atmospheric pressure .....................

700 hPa to 1060 hPa

After using the device on battery power, always reconnect the power cable. This will guarantee that the batteries are recharged automatically the next time the device is used.

2.1.Wireless Data Transmission

The device can be equipped with an optional wireless data transmission module (WLAN mobile), that uses radios to transmit data to an application that can receive them.

Due to the nature of radio transmissions, it is possible that, due to the characteristics of the environment where the device is located, some other RF sources may interfere with the transmission generated by the device.

Cardioline SpA has tested the coexistence of the device with other devices that can interfere such as devices using WLAN, Bluetooth radio and/or cell phones. Although the current technology allows a very successful rate of transmission, it is possible that in some rare occurrences, the system may not perform at its best resulting in a “failed transmission”. When this occurs, patient data will not be erased from the device nor stored in the receiving application, ensuring that partial or corrupted data are not made available to the receiving station. If the failure mode persists the user should move to a position where the RF signals may propagate better and allow successful transmissions.

2.2.Wirless Option

Wireless options transmit at 2.4 GHz. Other nearby wireless devices may cause interference. If possible, move or turn off other devices to minimize potential interference.

The following table shows the channels allocated in different geographic areas in the world. Please consult with your IT personnel in order to set the device on the proper channels.

Specification

Description

Technology

Compliant withIEEE 802.11n, IEEE 802.11g and IEEE 802.11b

 

standards

Interface Type

USB2.0/1.1

 

 

Security

64/128bit WEP encryption

 

WPA/WPA2 and WPA -PSK/WPA2-PSK encrytions

 

Wi-fi Protected setup(WPS)

RF Frequency

2,412 to 2,462MHz (North America)

 

2,412 to 2,472MHz (Europe)

 

2,412 to 2,484MHz (Japan)

Operating Frequency

FCC 2412-2462MHz (Ch1-Ch11)

 

ETSI 2412-2472MHz (Ch1-Ch13)

8

 

ECG100S and ECG200S

2. SAFETY INFORMATION

RF Power

11n mode: 13dBm

 

11g mode: 15dBm

 

11b mode: 18dBm

In order to achieve the best transmission rate, it is necessary that the facility where the device is operated can provide good area coverage. Please consult the IT personnel of the facility to verify the proper WLAN availability in the area where the device will be used.

RF wave propagation may be blocked or reduced by the environment where the device is used. Most common areas where this may occur are: shielded rooms, elevators, underground rooms. In all the above situations, it is recommended to move the device to a proper location and verify with the IT personnel of the facility the areas where the WLAN signals are available.

9

ECG100S and ECG200S

2. SAFETY INFORMATION

10

ECG100S and ECG200S

3. ELECTROMAGNETIC COMPABILITY (EMC)

3.ELECTROMAGNETIC COMPATIBILITY (EMC)

This device requires particular precautions regarding Electromagnetic Compatibility. It must therefore be installed and commissioned in compliance with the information on Electromagnetic Compatibility contained in this manual.

Portable and mobile radio communication equipment can affect operation of the device.

Using accessories, transducers or cables different than those specified in par. 12.2 can increase the emissions or decrease the immunity of the appliance.

Warnings

This device is only intended to be used by professional healthcare personnel. This device could generate radio interference or disturb operation of the equipment in the vicinity. Therefore it could be necessary to take measures to mitigate these effects, such as re-directing or repositioning the device or shielding the room.

The use of accessories and cables other than those recommended by Cardioline may cause an increase in emissions or a lowering in the protection of the system.

The device must not be used near or superimposed to other equipment. If necessary, check that the device works according to its standard operation.

Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.

Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.

Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the ME EQUIPMENT, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. Where ME EQUIPMENT stands for the respective models, ECG100S and ECG200S accordingly.

Notes

The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment.

Electromagnetic compatibility during the use of the device is required with the surrounding devices.

An electronic device can generate or receive electromagnetic interference. The electromagnetic compatibility test (EMC) has been carried out on the electrocardiograph in compliance with the international

11

ECG100S and ECG200S

3. ELECTROMAGNETIC COMPABILITY (EMC)

EMC directive for medical equipment (IEC 60601-1-2). This IEC standard has been adopted as a European standard (EN 60601-1-2).

Fixed, portable and mobile equipment for RF communication may affect the protection of the medical equipment. See par. 3.4 for the recommended separation distance between the radio equipment and the device.

The purpose of the device is the acquisition of ECG signals and the presentation of ECG reports for diagnostic purposes, as defined in IEC 60601-2-25.

Electromagnetic disturbances can cause disturb or degradation of the acquired ECG signal, resulting in misdiagnosis or delayed treatment.

3.1.Guidance and Manufacturer's declaration - Electromagnetic emissions

The ECG100S is intended to operate in the electromagnetic environment specified below. The customer or the user of the ECG100S must guarantee that it is used in this environment.

Emission test

 

Compliance

 

Electromagnetic environment – guidance

 

 

 

 

 

Radio frequency (RF) emissions

 

 

 

The ECG100S uses RF energy only for its internal function. Therefore, its

 

Group 1

 

RF emissions are very low and are not likely to cause any interference in

CISPR 11

 

 

 

 

 

nearby electronic equipment.

 

 

 

 

Radio frequency (RF) emissions

 

Class A

 

 

CISPR 11

 

 

The ECG100S is suitable for use in all establishments, including domestic

 

 

 

Harmonics emissions

 

Class A

 

establishments and those directly connected to the public low voltage

IEC 61000-3-2

 

 

power supply network that supplies buildings used for domestic

 

 

 

Voltage fluctuations/

 

 

 

purposes, because the emission level is low enough to not affect nearby

flicker emissions

 

Complies

 

equipment.

IEC 61000-3-3

 

 

 

 

The ECG200S is intended to operate in the electromagnetic environment specified below. The customer or the user of the ECG200S must guarantee that it is used in this environment.

Emission test

 

Compliance

 

Electromagnetic environment – guidance

 

 

 

 

 

Radio frequency (RF) emissions

 

 

 

The ECG200S uses RF energy only for its internal function. Therefore, its

 

Group 1

 

RF emissions are very low and are not likely to cause any interference in

CISPR 11

 

 

 

 

 

nearby electronic equipment.

 

 

 

 

Radio frequency (RF) emissions

 

Class A

 

 

CISPR 11

 

 

The ECG200S is suitable for use in all establishments, including domestic

 

 

 

Harmonics emissions

 

Class A

 

establishments and those directly connected to the public low voltage

IEC 61000-3-2

 

 

power supply network that supplies buildings used for domestic

 

 

 

Voltage fluctuations/

 

 

 

purposes, because the emission level is low enough to not affect nearby

flicker emissions

 

Complies

 

equipment.

IEC 61000-3-3

 

 

 

 

12

ECG100S and ECG200S

3. ELECTROMAGNETIC COMPABILITY (EMC)

3.2.Guidance and Manufacturer's declaration - Electromagnetic immunity

The ECG100S/ECG200S is intended for use in the electromagnetic environment specified below. The customer or the user of the ECG100S/ECG200S should assure that it is used in such an environment.

Immunity test

 

Conformity

 

Compliance level

 

Electromagnetic Environmental Information

 

 

 

 

 

 

 

Electrostatic discharge

 

+/- 6 kV contact

 

+/- 6 kV contact

 

Floors should be wood, concrete or ceramic tile. If

(ESD)

 

 

 

floors are covered with synthetic material, the

 

+/- 8 kV air

 

+/- 8 kV air

 

IEC 61000-4-2

 

 

 

relative humidity should be at least 30 %.

 

 

 

 

 

Electrical Fast Transient/

 

+/- 2 kV for power

 

+/- 2 kV for power

 

 

 

supply lines

 

supply lines

 

Mains power quality should be that of a typical

Burst

 

 

 

 

+/- 1 kV for

 

+/- 1 kV for

 

commercial or hospital environment.

IEC 61000-4-4

 

 

 

 

input/output lines

 

input/output lines

 

 

 

 

 

 

 

 

 

+/- 1 kV line(s) to

 

+/- 1 kV line(s) to

 

 

Surge

 

line(s)

 

line(s)

 

Mains power quality should be that of a typical

IEC 61000-4-5

 

+/- 2 kV line(s) to

 

+/- 2 kV line(s) to

 

commercial or hospital environment.

 

 

earth

 

earth

 

 

 

 

<5% UT

 

<5% UT

 

 

 

 

(>95% dip in UT)

 

(>95% dip in UT)

 

 

 

 

for 0.5 cycle

 

for 0.5 cycle

 

 

Voltage dips,

 

40% UT

 

40% UT

 

Mains power quality should be that of a typical

short interruption and

 

 

 

 

(60% dip in UT)

 

(60% dip in UT)

 

commercial or hospital environment. If the user of

voltage variations

 

 

 

 

for 5 cycles

 

for 5 cycles

 

the ECG100S requires continued operation during

on power supply

 

 

 

 

 

 

 

 

 

70% UT

 

70% UT

 

power mains interruptions, it is recommended that

input

 

 

 

 

 

 

ECG100S be powered from an uninterruptible power

lines

 

(60% dip in UT)

 

(60% dip in UT)

 

 

 

 

supply or a battery.

IEC 61000-4-11

 

for 25 cycles

 

for 25 cycles

 

 

 

 

 

 

 

 

 

 

 

 

 

<5% UT

 

<5% UT

 

 

 

 

(>95% dip in UT)

 

(>95% dip in UT)

 

 

 

 

for 5 s

 

for 5 s

 

 

Power frequency and

 

 

 

 

 

Power frequency magnetic fields should be at levels

magnetic field

 

3 A/m

 

3 A/m

 

characteristic of a typical location in a typical

(50/60 Hz)

 

 

 

 

 

commercial or hospital environment.

NOTE: UT is the AC mains voltage prior to the application of the test level.

3.3.Guidance and Manufacturer's declaration - Electromagnetic immunity

The ECG100S/ECG200S is intended for use in the electromagnetic environment specified below. The customer or the user of the ECG100S/ECG200S should assure that it is used in such an environment.

Emission test

 

IEC 60601 test level

 

Compliance level

 

Electromagnetic Environmental Information

 

 

 

 

 

 

 

Conducted RF

 

3 V rms

 

 

 

Portable and mobile RF communications equipment

 

From 150 kHz to 80

 

3 V

 

should be used no closer to any part of the ECG100S,

IEC 61000-4-6

 

 

 

 

MHz

 

 

 

including cables, than the recommended separation

 

 

 

 

 

Radiated RF

 

3 V/m

 

3 V/m

 

distance calculated from the equation applicable to

IEC 61000-4-3

 

From 80 MHz to 2.5

 

 

the frequency of the transmitter.

 

 

 

 

13

 

 

 

ECG100S and ECG200S

 

 

 

3. ELECTROMAGNETIC COMPABILITY (EMC)

 

 

 

 

 

 

 

 

 

GHz

 

Recommended separation distance:

 

 

 

From 80 MHz to 800 MHz

 

 

 

From 800 MHz to 2.5 GHz

 

 

 

where P is the maximum output power rating of the

 

 

 

transmitter in watts (W) according to the transmitter

 

 

 

manufacturer and d is the recommended separation

 

 

 

distance in metres (m).

 

 

 

Field strengths from fixed RF transmitters, as

 

 

 

determined by an electromagnetic site survey (a),

 

 

 

should be less than the compliance level in each

 

 

 

frequency range (b).

 

 

 

Interference may occur in the vicinity of equipment

 

 

 

marked with the following symbol:

 

 

 

 

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a)Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ECG100S/ECG200S is used exceeds the applicable RF compliance level above, the ECG100S/ECG200S should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ECG100S/ECG200S.

b)Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

14

ECG100S and ECG200S

3. ELECTROMAGNETIC COMPABILITY (EMC)

3.4.Recommended separation distances between portable and mobile RF communications equipment and the ECG100S/ECG200S

The ECG100S/ECG200S is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ECG100S/ECG200S can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ECG100S/ECG200S as recommended below, according to the maximum output power of the communications equipment.

 

Maximum rated power output of the

 

 

Separation distance according to frequency of transmitter (m)

 

 

 

 

 

 

 

 

 

 

 

 

 

transmitter (W)

 

 

150 KHz to 800 MHz

 

 

800 MHz to 2.5 GHz

 

 

800 MHz to 2.5 GHz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.01

 

0.12

 

0.12

 

0.23

 

 

 

 

 

 

 

 

 

0.1

 

0.37

 

0.37

 

0.74

 

 

 

 

 

 

 

 

 

1

 

1.17

 

1.17

 

2.33

 

 

 

 

 

 

 

 

 

10

 

3.69

 

3.69

 

7.38

 

 

 

 

 

 

 

 

 

100

 

11.67

 

11.67

 

23.33

 

 

 

 

 

 

 

 

 

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

3.5.Transmission frequencies

 

Frequency

 

 

 

 

 

EIRP

 

 

EIRP

 

 

ERP =

 

 

ERP =

 

 

 

 

 

 

 

 

 

 

Channel

 

 

Power

 

 

Measured

 

 

EIRP/1.64

 

 

EIRP/1.64

 

 

Standard

 

 

Modulation

 

 

[MHz]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

[mW]

 

 

[dBm]

 

 

[Mw]

 

 

[dBm]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.412

 

1

 

8.43

 

9.26

 

5.14

 

5.64

 

 

802.11B

 

CCK, OFDM, BPSK, GPSK,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16-QAM, 64Q-AM

2.437

 

6

 

13.30

 

11.24

 

8.11

 

6.85

 

 

802.11B

 

CCK, OFDM, BPSK, GPSK,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16-QAM, 64Q-AM

2.480

 

13

 

11.67

 

10.67

 

7.12

 

6.50

 

 

802.11B

 

CCK, OFDM, BPSK, GPSK,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16-QAM, 64Q-AM

2.412

 

1

 

6.20

 

7.93

 

3.78

 

4.83

 

 

802.11G

 

CCK, OFDM, BPSK, GPSK,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16-QAM, 64Q-AM

2.437

 

6

 

9.27

 

9.67

 

5.65

 

5.89

 

 

802.11G

 

CCK, OFDM, BPSK, GPSK,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

15

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ECG100S and ECG200S

3. ELECTROMAGNETIC COMPABILITY (EMC)

 

 

 

 

 

 

 

16-QAM, 64Q-AM

 

 

 

 

 

 

 

 

2.480

13

9.64

9.84

5.87

6

802.11G

CCK, OFDM, BPSK, GPSK,

 

 

 

 

 

 

 

16-QAM, 64Q-AM

2.412

1

11.53

10.62

7.03

6.47

802.11N

CCK, OFDM, BPSK, GPSK,

 

 

 

 

 

 

 

16-QAM, 64Q-AM

2.437

6

4,48

6.51

2.73

3.96

802.11N

CCK, OFDM, BPSK, GPSK,

 

 

 

 

 

 

 

16-QAM, 64Q-AM

2.480

13

3.31

5.20

2.02

3.17

802.11N

CCK, OFDM, BPSK, GPSK,

 

 

 

 

 

 

 

16-QAM, 64Q-AM

16

ECG100S and ECG200S

4. SYMBOLS AND LABEL

4.SYMBOLS AND LABEL

4.1.Explanation of the symbols

Symbol

Description

 

 

 

Comply with the instructions in the use manual

 

 

 

CE marking – compliance with the European Union directives

 

 

 

Manufacturer

 

 

 

Reference number (product code)

 

 

 

Serial Number

 

 

 

Lot number

 

 

 

Year of manufacture

 

 

 

RF transmitter (if present)

 

 

 

Type CF equipment

 

 

 

Separate collection of electrical waste and electronic equipment

 

 

 

Read instruction for use

 

 

 

Consult instructions for use

 

 

 

Temperature variation

 

 

 

Humidity variation

 

 

 

No latex

 

 

 

Keep dry

 

 

17

Cardioline ECG100S, ECG200S User Manual

ECG100S and ECG200S

4. SYMBOLS AND LABEL

4.2.Device label

ECG100S

ECG200S

RF Symbol beside the connectors in the rear part of the device (see par. 5.4.1).

18

ECG100S and ECG200S

5. INTRODUCTION

5.INTRODUCTION

5.1.Purpose of the manual

This manual deals with the ECG100S/ECG200S devices.

The manual represents a guide for the execution of the following operations:

Reasonable use of the electrocardiograph, of the function keys and of the sequence of menus.

Preparation of the device for use. (Section 6)

Acquisition, printing and storage of ECG tracing. (Section 7)

System settings. (Section 8)

Connectivity and transmission of ECG tracings. (Section 9)

Device upgrading (Section 10)

Troubleshooting and electrocardiograph maintenance. (Section 11)

5.2.Recipients

This manual is intended for professional healthcare operators. They are therefore presumed to have specific knowledge of medical procedures and terminology, as required by clinical practice.

5.3.Intended use

ECG100S/ECG200S is a high-performance, multi-channel, interpretative resting electrocardiograph.

The ECG signal is acquired with a 10-wires patient cable and is displayed in real time on a LCD screen integrated in the device. The electrocardiograph can analyse and store the ECG traces, send them to an external peripheral via Internet or via USB, print the 12 lead ECG in automatic or manual mode by means of a thermal printer.

ECG100S/ECG200S is intended for assessment and diagnosis of cardiac functions. In any case the results of analysis performed by the electrocardiograph must be validated by a Physician.

ECG100S/ECG200S is intended for use in hospitals, in medical clinics and doctor’s offices of any size.

The device is indicated for use to acquire, analyse, display and print electrocardiograms.

The device is intended to provide the physician with an automatic interpretation of the ECG to be reviewed by a physician.

The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.

19

ECG100S and ECG200S

5. INTRODUCTION

The interpretations of ECG offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.

The device is indicated for use on adult and pediatric populations.

The device is not intended to be used as a vital signs physiological monitor.

5.4.Description of the device

The device is a 12-lead diagnostic electrocardiograph which displays, acquires, prints and stores ECG tracings for adults and children. It also calculates the main overall ECG parameters.

The device can be supplied with the optional 12-lead Glasgow resting ECG interpretation algorithm, with specific criteria by age, sex and race. If this option is enabled, the algorithm can provide an over-reading physician with a second opinion generating diagnostic messages in the ECG report.

For further information on the resting ECG interpretation algorithm, see the Guidance for the physician on the application on adults and children (see accessories list)

The device can be configured with a larger memory and with bidirectional connectivity (LAN/WiFi).

The device can be powered by battery or the electrical mains.

The printing formats supported include: standard or Cabrera 3, 3+1, 3+3, 6 or 12 (only ECG200S) channels in automatic mode and 3, 6 or 12 (only for ECG200S) channels rhythm strip printing.

The device includes:

1.Patient cable

2.Power supply

3.Paper

4.Pack of electrodes

5.Banana/clip adapter set

6.Guidance for the physician on the application on adults and children (with interpretative key)

7.User manual

20

Loading...