Cannon D155 User Manual

Page 1
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 2
This page left intentionally blank
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 3
Table of contents Page
1. Introduction ............................................................................................................................................................... 2
1-1. Overview of the instrument ........................................................................................................................... 2
1-2. About the manual .......................................................................................................................................... 3
1-3. Safety symbols ..................................................................................................................................... 4
2. Preparations for measurement ................................................................................................................................... 7
2-1. Supplied parts with D155 .............................................................................................................................. 7
2-2. Installation and start-up ................................................................................................................................. 9
2-2-1. Power cable ........................................................................................................................................ 9
2-2-2. Connecting the tube lines – D155 with Peristaltic Pump ................................................................. 10
2-2-3. Connecting the tube lines – D155 without Pump ............................................................................ 10
2-3. Parts name and functions ............................................................................................................................. 11
2-3-1. Outward appearance and configuration ........................................................................................... 11
2-3-2. Key functions (General description) ................................................................................................ 13
2-3-3. LCD Display Layout ........................................................................................................................ 14
2-4. Display during initialization ........................................................................................................................ 14
2-4-1. The initial display upon power up ................................................................................................... 14
2-4-2. Display during warm up period ....................................................................................................... 14
2-4-3. Display after warm-up ..................................................................................................................... 14
2-5. Preparation before measurements ................................................................................................................ 15
3. Basic procedure ....................................................................................................................................................... 16
3-1. Flow chart for the basic measuring procedure ............................................................................................. 16
3-2. Factor calibration with pure water ............................................................................................................... 17
3-3. Measuring procedure ................................................................................................................................... 20
3-4. Clean measuring cell after measurements ................................................................................................... 22
3-5. Procedure after a day’s measurements ........................................................................................................ 23
4. Menu Navigation ..................................................................................................................................................... 24
4-1. Menu:1 View Stored Results ....................................................................................................................... 24
4-2. Menu 2: Set Sampling Time ........................................................................................................................ 25
4-3. Menu 3: Select Oil Table ............................................................................................................................. 25
4-4. Menu 4: Set Test Temperature .................................................................................................................... 26
4-5. Menu 5: Set Date and Time ......................................................................................................................... 26
4-6. Menu 6: Set LCD Contrast .......................................................................................................................... 27
4-7. Menu 7: Reset Test Number ........................................................................................................................ 27
4-8. Menu 8: Clear Stored Results ...................................................................................................................... 27
4-9. Menu 9: Export Stored Data to USB Drive ................................................................................................. 28
4-10. Menu 10: View Serial No. & Firmware Ver. ............................................................................................ 29
4-11. Parameters and default .............................................................................................................................. 30
4-12. How to connect the optional printer or a personal computer (PC) ............................................................ 31
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 4
5. Maintenance ............................................................................................................................................................ 32
5-1. Daily check .................................................................................................................................................. 32
5-1-1. Cleaning the measuring cell ............................................................................................................. 32
5-2. Other maintenance ....................................................................................................................................... 33
5-2-1. Replacement of the power fuse ........................................................................................................ 33
5-2-2. Adjustment of Injection Adapter (and tightening of tube nozzles) .................................................. 33
5-2-3. Replacement of optional peristaltic pump tubing ............................................................................. 35
6. Troubleshooting ....................................................................................................................................................... 37
6-1. Error messages and remedies ....................................................................................................................... 37
6-2. Symptoms for error numbers and associated remedies ................................................................................ 37
6-3. Troubleshooting ........................................................................................................................................... 38
6-3-1. Cannot turn on the main unit with the power switch ........................................................................ 38
6-3-2. “Ready to Test” message does not appear and/or cell temperature is unstable. ............................... 39
6-3-3. The unit does not work even though power is on. ............................................................................ 39
6-3-4. LCD does not function ..................................................................................................................... 39
6-3-5. Key entry does not work .................................................................................................................. 40
6-3-6. Air bubbles are easily trapped in the sample and visible in the measurement cell ........................... 40
6-3-7. There is poor repeatability on measured values ............................................................................... 41
6-3-8. Sampling time is too long (optional peristaltic pump) ..................................................................... 42
6-3-9. Measuring time is too long ............................................................................................................... 43
6-3-10. Data cannot be stored in the memory ............................................................................................. 43
7. Others ...................................................................................................................................................................... 44
7-1. Basic specifications ...................................................................................................................................... 44
7-2. Principle of measurement ............................................................................................................................ 45
7-2-1. Oscillating U-tube Cell .................................................................................................................... 45
7-2-2. Density Tables .................................................................................................................................. 46
7-3. Warranty/Return Information ...................................................................................................................... 48
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 5

1. Introduction

1-1. Overview of the instrument

Please read this manual thoroughly in advance for the best performance of the equipment.
Thank you for your patronage of CANNON products. The D155 Density Meter you have purchased is
an oscillation-type density/specific gravity meter which can measure density, specific gravity and API
degrees. This meter is equipped with thermal control, which enables you to set a measurement
temperature. Also an optional peristaltic pump is available (factory installed) for sampling, thus
measurement is made easier by just pressing a button.
Key Features
1) Density, specific gravity and API degrees can be easily measured.
2) Measurement can be started simply by pressing a button.
3) The built-in thermal control enables you to set a measurement temperature in the range between
15°C to 25°C.
4) Measurement results on the display can be selected from density, specific gravity or API degrees.
5) The meter can be calibrated on one-point with pure water.
6) An optional printer can be connected, or data can be exported as an MS-Excel file and saved to an
external USB storage device for analysis on a personal computer.
2
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 6
There exists the danger of physical injury or even possible death if the
There exists the danger of property damage if the instruction is

1-2. About the manual

Please keep this manual near your system so that you can easily access to the necessary information
you are looking for while operating or preparing for measurement.
The examples below represent the three types of messages (warning, caution, and note) encountered
throughout the manual.
1. Where there exists a danger of physical injury or even possible death:
Warning!
instruction is disregarded.
2. Where there exists a danger of property damage:
3
Caution!
ignored.
3. When there exists a possibility of failure of instrument performance:
Note:
There exists the possibility of failure of instrument performance. If ignored, warranty may not be covered.
* It is prohibited to copy or reproduce in part or in whole this manual without authorization by copyright.
* If you should find any part in this manual not clear to understand or missing article, contact your local
dealer or sales representative.
* Manufacturer will not be liable for any loss or damage directly or indirectly caused by use of the
instrument or its consequences.
* This manual pertains directly to the D155. For details relating to other accessories or equipment
please refer to the appropriate manufacturers supplied documentation..
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 7

1-3. Safety symbols

Always observe these signs and instructions.
You must observe cautionary messages and warnings in order to protect yourself as well as prevent
others from physical injury or property damages.
4
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 8
5
WARNING!
Wear safety glasses, gloves or protective mask if necessary,and well ventilate the room.
Danger of injury on your skin or in the eyes by splashing chemical. Also your windpipe may get hurt if toxic gas is breathed in.
Danger of fire, electric shock or damage to the instrument.
Do not operate in a way other than specified in the manual.
CAUTION!
CAUTION!
Do not open housing case or overhaul the unit for repair except by an authorized service person.
Danger of fire, shock or malfunctioning of the unit.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 9
About place for installation
Avoid the use of this instrument under the environment described below.
(Failure can lead to the degradation of performance and reliability of the system.)
Operation of devices with strong electric motors using common power source
Near strong magnetic/electric field
Use of power source with too variable load
Location of strong vibration
Exposure to direct sunlight
Location with large temperature difference
Exposure to corrosive gas
Exposure to extreme heat (Operation temperature: 5 to 35°C (41 to 95°F))
Exposure to high humidity (Over 85%RH)
About power source
Power for this instrument is 100-240V AC ±10%and 50/60Hz.
Supply power direct from the outlet, and do not share power from a tap,
extension cord, or power outlet strip.
About place for storage
6
If the instrument is not operated for a long period of time (more than several
For storage, avoid a place with temperature extremes, high humidities and
Other caution
Do not use such a solvent as alcohol, acetone, thinner or the like for cleaning the
external parts or surfaces of this instrument. Doing so may adversely affect the
instrument, e.g. deformation, discoloration or cracks. When cleaning this instrument,
wipe it with a soft cloth or tissue paper, after applying mild detergent diluted with water to
the soft cloth or tissue paper and adequately wringing out excess water in order not to
allow water to enter the instrument.
Prior to the actual sample measurement, check the meter calibration with pure distilled
water. Additionally, certified density reference oils are available from Cannon Instrument
Company and can be used to verify the performance of the D155.
weeks), rinse the measurement cell with a suitable solvent followed by pure
water and a drying solvent, and then pack the whole unit into the carton box
which contained the instrument when first delivered.
dust.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 10
Part name
CANNON Part Number
Qty
Sketch
w/ ground
Operation
manual

2. Preparations for measurement

2-1. Supplied parts with D155

Check the supplied parts referring to the following parts list. If any missing or broken parts including main unit, accessories or manual are found, contact your sales representative or local dealer.
9726-P10
D155 unit
(with peristaltic pump)
115V AC Power cord
Connection
Tube with
Union
1 unit
9726-P15
74.2110
1 pc.
95.0254 1 pc.
7
Red Tubing,
1.5 mm OD,
95.0297 1 pc.
PTFE
Wrench,
open-ended
95.0298 1 pc.
95.0238
Syringe 2mL
Operation
(re-order in pack of 100)
95.0250
manual
5 pc.
1
copy
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 11
Part name
CANNON Part Number
Qty
Sketch
(L=2000mm)
Part name
CANNON Part Number
Qty
Sketch
Additional Parts supplied with optional peristaltic pump
Peristaltic
Pump
(includes all
tubing below)
Peristaltic
Pump Tube
Gore400
Connection
Tube with
Union
95.0255 1 pc.
95.0256 1 pc.
95.0254 1 pc.
8
Drain Tube 95.0253 1 pc.
Additional Parts supplied with optional printer
Dot Matrix
Printer Kit
(includes
cable and
spare ribbon)
Null-Modem
RS-232
Cable
DB9F-DB25M
95.0239 1 pc.
95.0240 1 pc.
Ribbon, Black
Theramark
P5570
82.0059 1 pc.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 12
grounds to the AC Mains earth.

2-2. Installation and start-up

2-2-1. Power cable

1) Make sure the power switch is in the OFF position.
2) Plug in the supplied power cable on the back of unit.
9
3) Connect the power cable to the power outlet.
< 3-pin plug >
The 3 pins plug has an earth terminal and
Warning! The earth wire must be grounded. If not, there exists a danger of electric shock.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 13
OUT
IN
2
1
2
Syringe

2-2-2. Connecting the tube lines – D155 with Peristaltic Pump

1) The sampling inlet and outlet ports (1) and (2) are Luer friction-type fittings. See image below for
connecting the tubing.
2) Insert and press the first Connection Tube with Union into the sample inlet Port (1) of the D155 as
shown below. The free end of this tube will later be inserted into the sample container.
3) Connect the second Connection Tube with Union to the sample outlet Port (2). The free end of
this tubing is then connected to the peristaltic pump IN (upper port) fitting as shown below. Connect
the Drain Tube to the peristaltic pump OUT fitting. The free end of the Drain Tube should then be
placed in a suitable waste container such as a beaker or bottle.
10

2-2-3. Connecting the tube lines – D155 without Pump

1) The sampling inlet and outlet ports (1) and (2) are Luer friction-type fittings. See image below for
connecting the tubing.
2) Use a syringe with friction Luer fitting to inject sample into the sample inlet Port (1) noted as “syringe”
of the D155 as shown below.
3) Insert and press the Connection Tube with Union into the sample outlet Port (2) on the D155 and
place the other end of the tube in a suitable waste container such as a beaker or bottle.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 14
2
3 4 5

2-3. Parts name and functions

2-3-1. Outward appearance and configuration

< Front and Right side panel >
1
11
6
1) Measuring cell unit
Internal to the instrument, this unit contains the oscillating U-tube measuring cell for sample
measurement.
2) Cell viewport
The sample can be seen through this window to check if the liquid contains air. The cell is
illuminated for easy viewing of the sample in the glass U-tube.
3) LCD display
The display of 20 digits by 4 lines shows measurement results, measured temperature, peristaltic
pumping condition (if installed) and instrument prompts.
4) Operating panel
This keypad contains 9 soft-touch buttons for controlling the density meter for setup, measurement
start, and function selection.
5) Sample inlet
Here is where the sample or cleaning solvent is injected either by syringe or the sampling tube is
connected for units which have the peristaltic pump installed.
6) Sample outlet
This port is connected to either a suitable waste container or the peristaltic pump “IN” port for units
with the peristaltic pump.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 15
12
10
9
8
7
11
13
14
< Rear panel >
12
7) Peristaltic pump “IN” (if equipped)
The sample outlet is connected here.
8) Peristaltic pump “OUT” (if equipped)
The tube for draining out the sample or solvent can be connected here.
9) Peristaltic pump (if equipped)
This pump transfers sample liquid by peristaltic movement.
10) RS-232C port
An optional printer or a personal computer (PC) can be connected here.
11) LINE
The AC Mains power cord is connected with this receptacle.
12) Power switch
Turn on or off the unit with this switch.
13) Fuse holder
A miniature 5x20mm cartridge fuse is housed in here. Only replace the fuse with like rating fuses
(T3.15A 250V) as noted on the rear label of the instrument
14) USB port
Use for connecting an external USB storage device for exporting stored data from the density meter.
This port is also used to uploaded firmware to the density meter.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 16

2-3-2. Key functions (General description)

Here is a brief description of function of each key:
[][][][] key : Key to select menu options, settings, or enter numeric values.
[MEAS] key : Key to start density measurement.
[PUMP] key : Key to turn or off the optional peristaltic sampling pump. This key can be
used during both calibration and measurement.
[STOP] key : Key to abort density measurement or calibration and also escape or return
back to standby mode (the main display screen) or previous menu selection.
[ENTER] key : Key to confirm the menu selection or numeric data entry.
[MENU] key : Key to select the menu selection screen. Pressing this key and then the
Up/Down arrow keys will cycle through the following menu options:
13
Menu:0 Calibrate Cell
Calibrate with pure water standard
Menu:1 View Stored Results Display measurement results
Menu:2 Set Sampling time Select sampling time (amount of time the peristaltic pump runs)
Menu:3 Select Oil-Table Select API temperature compensation table for oil
Menu:4 Set Test Temperature Set measurement temperature
Menu:5 Set Date & Time Select date and time for measurements
Menu:6 Set LCD Contrast Adjust brightness of display screen
Menu:7 Reset Test Number Reset Sample test number to “01”
Menu:8 Clear Stored Results Clear the memory of stored measurement results
Menu:9 Export Stored Data to USB Drive Export data to external USB storage device (MS-Excel format)
Menu:10 View Serial No. & Firmware Ver. Check instrument serial number and firmware version
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 17
Highlighted Text Denotes Blinking Cursor or Characters
1 0 / 3 0 / 1 2 0 9 : 5 0 D e n s i t y 0 . 9 9 8 5 g / c m ³ T e m p 2 0 . 0 0 ° C T e s t 0 1 P r e s s E N T E R t o P r i n t
C A N N O N D 1 5 5 D e n s i t y M e t e r V e r 2 . 0 0
1 0 / 3 0 / 1 2 0 9 : 2 9
1 0 / 3 0 / 1 2 0 9 : 3 0
E Q U I L I B R A T I N G T a r g e t 2 0 . 0 ° C < < P l e a s e W a i t > >
1 0 / 3 0 / 1 2 0 9 : 4 0 R e a d y t o T e s t 2 0 . 0 ° C
P l e a s e L o a d S a m p l e P r e s s M E A S t o S t a r t

2-3-3. LCD Display Layout

2-4. Display during initialization

2-4-1. The initial display upon power up

The following display appears during power up.
14

2-4-2. Display during warm up period

The instrument requires a warm up period to ensure an accurate measurement. Normally, the
warm-up period is complete within 30 minutes this will vary room temperature and test temperature.
The following display occurs during warm up period.

2-4-3. Display after warm-up

The following display occurs after the instrument has stabilized. The instrument is now ready to
measure a sample.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 18

2-5. Preparation before measurements

1) Tools required for sampling
2 to 5 mL Syringe (for manual injection) and a beaker or bottle to hold the sample
2) Solvent for cleaning the cell
Solvent (e.g. water for aqueous solution or heptane for oil) is necessary to clean the cell. Note,
user is responsible for determining the appropriate solvent for the sample type being measured.
3) Lint-free towels or tissues
Used for wiping the outside of the sampling tube or cell ports.
15
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 19
Set sample and press [Meas.] button
Drain out the sample and
clean the measuring cell
Measurement
Turn off the unit
Measurement completed?
No
Yes
Fill the measuring cell with pure water
No
Yes
Perform factor calibration with pure water
Power on
Ready mode?
Calibration
Calibrated properly?
Yes
No
[MEAS]

3. Basic procedure

3-1. Flow chart for the basic measuring procedure

16
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 20
must be taken to avoid air bubbles or other foreign

3-2. Factor calibration with pure water

For correct measurement, the meter needs to be checked and adjusted to the reference using the
density standard of water.
Prior to each day’s work, measure the water by pressing [MEAS] key to make sure the result of specific
gravity (SG) shows a value within ± 0.0005 of the nominal value of pure water at the test temperature. If
the measuring cell and tube lines were filled with ethanol (for short-term storage), wash out the ethanol
with as much pure water before measurement is started. If the result shows a SG (t/t) value outside of
the ± 0.0005 range, repeat the cell cleaning and re-measure the water standard. If the measurement
results for pure water continue to fall outside of this range, the meter should be factor calibrated as
follows:
Note:
The measuring cell must be cleaned with water before calibration. Care
materials/contaminates in the water in the cell during measurement to avoid measurement error.
17
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 21
18
set temperature and that
the cell well. For effective
thanol with as much pure water before
↓ [MENU]
3) Press [MENU] key
↓ [ENTER]
5) Press [ENTER] key.
If using a syringe, inject 2 mL taking care not to introduce any air
standard.
1 0 / 3 0 / 1 2
0 9 : 4 0
R e a d y t o T e s
t 2 0 . 0 ° C
P l e a s e L o
a d S a m p l e
P r e s s
M E A S t o S t a r t
M
e n u : 0
C a l
i b r a t e C e l l
P r e s s E N T E R o r U p / D n
M
e n u : 0
C a l i b r a t e
C e l l
P l e a s e L o a d
W a t e r
P r e
s s E N T E R t o S t a r
t
1) Make sure that the cell temperature is stable at the
“Ready to Test” appears on the main display as shown below.
D155 with pump
D155 with syringe
2) Clean the cell with pure water
a. Put the tube into water as shown on the left or inject the water with a
10 mL syringe. If using a syringe, inject two full syringes so that a
total of 20 mL is flushed through the cell.
b. If the optional peristaltic pump is installed, press [PUMP] key to flow
in more than 20 mL water, to flush
cleaning, lift up the tip of tube from the water a few times to allow
some air to be pulled in.
Note:
When the measuring cell and tube lines were filled with ethanol, rinse out e measurement is started.
4) Move the cursor with [] or [] key to “Menu:0 Calibrate Cell”
6) Inject the cell with a pure water standard
a.
bubbles in the measurement cell. Visually inspect the cell through
the cell window to ensure that no bubbles are present in the cell.
b. If the optional peristaltic pump is installed, place the tube in the water
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 22
19
the Menu:2 Sampling Time.
present in the cell.
M e n u
: 0
C a l i b r a t e C e l l
S A M P L I N G :
1 0 s
< < P l e a s e
W a i t > >
M e n u : 0
C a l i b r a t e C e l l C A L I B
R A
T I N G
< <
P l e a s e W a i t > >
M
e n u : 0
C a l i b r a
t e C e l l
O K
P r e s s S T O P o r M E N
U
[ENTER]
Press [ENTER] key.
If the peristaltic pump is installed, the pump will run for the amount of seconds set in
8) When calibration is normally completed, the display shows:
The above screen will only appear when using the optional peristaltic pump and
the Sampling Time (Menu:2 Sampling Time) is set to a value greater than zero.
7) When sampling is complete, the display shows:
Visually inspect the cell through the cell window to ensure that no bubbles are
When “NG” (No Good) appears instead of “OK”, press [STOP] and then repeat
the calibration starting with step 3 above.
9) If the optional peristaltic pump is installed, press [PUMP] key to drain out water.
Otherwise use an empty syringe to push out the water with air.
10) Blot the sample drops on the sampling tube with tissue paper.
Note:
If calibration fails, previously calibrated values will be printed out.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 23
measure the water
standard. If the measurement results for pure water continue to fall
be factor calibrated as per
that “Ready to Test” appears on the main display as shown above.
2 to 5 mL of the sample to be
measured.
will be
are present in the cell
1 0
/ 3 0 / 1 2
0 9
: 4 0
R e a d y t o T e
s
t
2 0 . 0 ° C
P l e a s e
L
o a d
S a m p l e
P r e s s M E A S t o
S t
a r t
1 0 / 3 0
/ 1 2 0 9 : 4 5
S A M P L I N G : 1 0 s
< < P l e a s e W a
i t > >
1 0 /
3 0 / 1 2 0 9 : 4 6
M E A S U R I N G
< < P l e a s e W a i t > >

3-3. Measuring procedure

Note:
Prior to each day’s work, measure pure water by pressing [MEAS] key to make sure the result of specific gravity (SG) shows a value within ± 0.0005 of the nominal value of pure water at the test temperature. If the measuring cell and tube lines were filled with ethanol (for short-term storage), wash out the ethanol with as much pure water before measurement is started. If the result shows a SG (t/t) value outside of the ± 0.0005 range, repeat the cell cleaning and re-
outside of this range, the meter should Section “3-2 Factor calibration with pure water.”
20
D155 with pump
D155 with syringe
1) Make sure that the cell temperature is stable at the set temperature and
2) Inject the sample to be measured into a clean cell:
a. If using the optional peristaltic pump, put the tube into the
sample liquid as shown on the left.
b. If using a syringe, inject
3) Press [MEAS] key. If using the peristaltic pump, the sample
automatically injected for the Sampling Time (set in Menu:2.) as shown
below:
4) When measurement starts, the display will show:
Visually inspect the cell through the cell window to ensure that no bubbles
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 24
re are four results display windows. To change between these
displays.
push out the sample with air.
after measurements”.
When you have changed the measurement temperature, make sure to
The D155
1 0 / 3 0 / 1 2
0 9 : 5 0
D e n s
i t y 0 . 9 9 8 5 g / c m
³
T e m
p 2 0 . 0 0 ° C T e s t
0 1
P r e s s U p / D n U
n i t s
1 0 /
3
0 /
1 2
0 9
: 5
0
A
P I D e g r e e 5 D :
2 4 . 1
T e m p 2 0 . 0 0 ° C
T
e s t 0 1
P r
e s
s U p / D n U n i t s
1 0 / 3 0 / 1 2
0 9 : 5 0
1 5 ° C D e n
5 3 D : 0 . 9 9 9 5
T e m p 2
0 . 0 0 ° C T e s t 0 1
P r e s s U p / D n
U n i t s
1 0 / 3 0 /
1 2 0 9 : 5 0
S G ( t / t ) :
0 . 9 9 9 0
T e m p 2 0 . 0
0 ° C T e s t 0 1
P r e s s U p / D n U n i t s
21
5) After measurements are completed, the results will appear on the screen
display.
If a printer is connected, the results will automatically print.
The
windows, press the [] [] keys. Below are the three additional results
7) Clean the measuring cell according to Section “3-4. Clean measuring cell
6) When continuing measurement on the same sample, return to Step 1).
When the intended number of measurements is completed, remove the
sample from the cell. If the optional peristaltic pump is installed, press
[PUMP] key to drain out the sample. Otherwise use an empty syringe to
Note:
perform calibration before measuring at the new temperature. will not allow a measurement to be made at a temperature which has not been calibrated.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 25
measurement cell

3-4. Clean measuring cell after measurements

The measuring cell must be cleaned when a different type of sample is to be measured or when a series
of measurement is complete. Follow the following steps:
Instrument with optional peristaltic pump
1) Remove the sampling tube from the sample and wipe with tissue paper.
2) Put the sampling tube into the cleaning solution/solvent and draw the solution/solvent with [PUMP]
key. For more effective cleaning, we suggest that you pull out the tube from the solution/solvent
momentarily (while the pump is running) - several times to draw in some air. Continue pulling
solution/solvent until roughly 10 mL have been pulled through the cell. Some samples may require
more than 10 mL of cleaning solution/solvent to effectively remove the sample from the cell (see
note below).
3) Take the sampling tube out of the cleaning solution/solvent. When the solution/solvent is clear from
the sample tube, press [PUMP] key to stop the pump.
4) Remove the sampling tube from the cleaning solution/solvent and wipe with tissue paper.
Instrument without pump
1) Remove the sample syringe from the sample inlet and wipe the inlet with tissue paper.
2) Ensure the sampling tube (outlet) on the D155 (upper port) is placed in a suitable waste container
such as a beaker or bottle.
3) Fill a 10 mL syringe with cleaning solution/solvent and inject the solution into the cell. Periodically
remove the syringe and re-insert to allow air to be introduced into the cell for more effective cleaning.
Repeat this step if necessary. Some samples may require more than 10 mL of cleaning
solution/solvent to effectively remove the sample from the cell (see note below).
Note:
When viscous samples are measured, rinse the thoroughly with the cleaning solvent/solution to remove the sample. Correct measurement results cannot be expected if any residue of the previous sample remains in the measuring cell.
22
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 26
ibration or

3-5. Procedure after a day’s measurements

When the day’s work of measurement is completed, the measuring cell needs to be cleaned with the
cleaning solution/solvent and be filled with pure water in order to avoid contamination on the glass
measurement cell walls:
1) Clean the measuring cell according to the steps in Section 3-4.
2) When certain oily substances or deposits are adhered to the cell, follow the steps in Section 5-1-1.
3) Fill the measuring cell with pure water:
a. If using the optional peristaltic pump, put the sampling tube into pure water and press [PUMP]
key to draw water into the cell. When the cell is filled with pure water, press [PUMP] key to
stop the pump.
a. If using a syringe, inject two full 10mL syringes so that a total of 20 mL is flushed through the
cell.
When the unit is not used longer than one week, the cell should be filled with 99% or higher
concentration of Ethanol instead of pure water in order to avoid the growth of mold inside the cell.
23
Note:
When the cell is filled with Ethanol for storage, flush the inside cell with at least 20mL pure water using either a syringe or the [PUMP] key if using the optional peristaltic pump before starting next cal measurement. If the residue of Ethanol remains in the cell or tube lines, correct calibration or measurement cannot be performed.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 27
M e n u : 1
V i e w S t o r e d R e s u l t s
P r e s s E N T E R o r U p / D n
1 0 / 3 0 / 1 2 0 9 : 5 0 D e n s i t y 0 . 9 9 8 5 g / c m ³ T e m p 2 0 . 0 0 ° C T e s t 0 1 P r e s s E N T E R t o P r i n t
1 0 / 3 0 / 1 2 0 9 : 5 0 A P I D e g r e e 5 D : 2 4 . 1 T e m p 2 0 . 0 0 ° C T e s t 0 1 P r e s s E N T E R t o P r i n t
1 0 / 3 0 / 1 2 0 9 : 5 0 1 5 ° C D e n 5 3 D : 0 . 9 9 9 5 T e m p 2 0 . 0 0 ° C T e s t 0 1 P r e s s E N T E R t o P r i n t
1 0 / 3 0 / 1 2 0 9 : 5 0
S G ( t / t ) : 0 . 9 9 9 0 T e m p 2 0 . 0 0 ° C T e s t 0 1 P r e s s E N T E R t o P r i n t

4. Menu Navigation

4-1. Menu:1 View Stored Results

1) Press [MENU] then press [] or [] to select Menu 1 as shown below
2) Press [ENTER]
3) Press [] or [] key to select the test result (numbered)
4) Press [] or [] to select the preferred data result screen. There are four results display windows.
To change between these windows, press the [] [] keys.
24
5) Press [ENTER] to print test result (if optional printer is connected to the D155)
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 28
23A
Crude Oils 60/60°F
23B
Refined Products 60/60°F
23D
Lube Oils 60/60°F
53A
Crude Oils 15°C
53B
Refined Products 15°C
M e n u : 2
S e t S a m p l i n g T i m e
P r e s s E N T E R o r U p / D n
M e n u : 2
S a m p l i n g T i m e : 1 0 s A r r o w K e y s t o C h a n g e
E N T E R t o A c c e p t
M e n u : 3
S e l e c t O i l T a b l e
P r e s s E N T E R o r U p / D n
M e n u : 3
O i l T a b l e : 5 3 A A r r o w K e y s t o C h a n g e
E N T E R t o A c c e p t

4-2. Menu 2: Set Sampling Time

Select sampling time adequate for the sample to be injected. The default is ‘10’ (seconds).
1) Press [MENU] then press [] or [] to select Menu 2 as shown below
2) Press [ENTER]
3) Press [] or [] key to select the digit to change
4) Press [] or [] to increase or decrease the number. The D155 allows any value from 00 (pump
off or disabled) to 99 seconds. The default setting for instruments with the pump installed is 10.
5) Press [ENTER] to accept the value entered

4-3. Menu 3: Select Oil Table

25
The API temperature compensation table for oil can be selected as follows. It depends on the samples
you wish to measure.
1) Press [MENU] then press [] or [] to select Menu 3 as shown below
2) Press [ENTER]
3) Press [] or [] to select the appropriate API temperature compensation table. The available
tables are:
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 29
53D
Lube Oils 15°C
59A
Crude Oils 20°C
59B
Refined Products 20°C
59D
Lube Oils 20°C
M e n u : 4 S e t T e s t T e m p e r a t u r e
P r e s s E N T E R o r U p / D n
M e n u : 4
T e s t T e m p . : 2 0 . 0 ° C A r r o w K e y s t o C h a n g e
E N T E R t o A c c e p t
M e n u : 5
S e t D a t e & T i m e
P r e s s E N T E R o r U p / D n
M e n u : 5
1 0 / 3 0 / 1 2 0 9 : 5 0 A r r o w K e y s t o C h a n g e
E N T E R t o A c c e p t
4) Press [ENTER] to accept the selected table.

4-4. Menu 4: Set Test Temperature

Measurement temperature can be set as follows:
1) Press [MENU] then press [] or [] to select Menu 4 as shown below
2) Press [ENTER]
26
3) Press [] or [] key to select the digit to change
4) Press [] or [] to increase or decrease the number. The D155 allows any temperature between
15.0 and 25.0 °C
5) Press [ENTER] to accept the value entered

4-5. Menu 5: Set Date and Time

Enter the present date and time so that they are stored/printed with each measurement result.
1) Press [MENU] then press [] or [] to select Menu 5 as shown below
2) Press [ENTER]
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 30
M e n u : 6
S e t L C D C o n t r a s t
P r e s s E N T E R o r U p / D n
M e n u : 6
L C D C o n t r a s t : 7 5 A r r o w K e y s t o C h a n g e
E N T E R t o A c c e p t
M e n u : 7
R e s e t T e s t N u m b e r
P r e s s E N T E R o r U p / D n
M e n u : 7
R e s e t T e s t N u m b e r ?
E N T E R t o A c c e p t S T O P o r M E N U t o Q u i t
3) Press [] or [] key to select the digit to change
4) Press [] or [] to increase or decrease the number.
5) Press [ENTER] to accept the date and time entered

4-6. Menu 6: Set LCD Contrast

The contrast of the LCD display can be selected as follows:
1) Press [MENU] then press [] or [] to select Menu 6 as shown below
2) Press [ENTER]
27
3) Press [] or [] to increase or decrease the LCD display contrast.
4) Press [ENTER] to accept the setting

4-7. Menu 7: Reset Test Number

The sample number will be output together with measurement results. The number is incremented each
time a new measurement is made up to 99. This resets the numbering to start again at “01”.
1) Press [MENU] then press [] or [] to select Menu 7 as shown below
2) Press [ENTER]
3) Press [ENTER] to accept the reset, or [STOP] or [MENU] to quit.

4-8. Menu 8: Clear Stored Results

CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 31
M e n u : 8 C l e a r S t o r e d R e s u l t s
P r e s s E N T E R o r U p / D n
M e n u : 8
C l e a r A L L R e s u l t s ?
E N T E R t o A c c e p t S T O P o r M E N U t o Q u i t
Date & Ti me
'07/13/ 13 04:31 Sample No. 2
Meas. Temp. 20.000000
Density 0. 1035 SG(t/t) 0.1037
Oil-Table 53A
15C De n 0. 1035 API Degre e
**.*
Date & Ti me
'07/13/ 13 04:31 Sample No.
3 Meas. Temp. 20.000000 Density 0.1035 SG(t/t)
0.1037 Oil-Table
53A 15C De n 0. 1035 API Degre e **.*
Date & Ti me '07/13/ 13 04:32 Sample No. 4 Meas. Temp.
20.000000 Density 0.1035
SG(t/t) 0.1037 Oil-Table 53A 15C Den 0. 1035
API Degre e **.*
Date & Ti me '07/13/ 13 04:33 Sample No.
5 Meas. Temp. 20.000000 Density
0.1035 SG(t/t) 0.1037 Oil-Table 53A 15C Den
0.1035 API Degree **.*
Date & Ti me
'07/13/ 13 04:34 Sample No. 6 Meas. Temp. 20.000000
Density 0. 1035
SG(t/t) 0.1037 Oil-Table 53A 15C Den
0.1035 API Degree **.*
Date & Ti me '07/13/ 13 04:35 Sample No. 7 Meas. Temp.
20.000000 Density 0.1035 SG(t/t) 0.1037 Oil-Table
53A 15C De n
0.1035 API Degree **.*
Date & Ti me '07/13/ 13 04:36 Sample No.
8 Meas. Temp. 20.000000
Density 0. 1035 SG(t/t) 0.1037 Oil-Table 53A 15C Den
0.1035 API Degree **.*
Date & Ti me '07/13/ 13 04:37 Sample No. 9 Meas. Temp. 20.000000 Density
0.1035 SG(t/t) 0.1037 Oil-Table 53A 15C Den
0.1035 API Degree **.*
Date & Ti me '07/13/ 13 04:38 Sample No. 10
Meas. Temp. 20.000000 Densi ty 0.1035 SG(t/t)
0.1037 Oil-Table 53A 15C De n 0. 1035 API Degre e **.*
Date & Ti me '07/13/ 13 04:44 Sample No. 11 Meas. Temp.
20.000000 Density 0.1035 SG(t/t)
0.1037 Oil-Table 53A 15C De n 0. 1035 API Degre e
**.*
M e n u : 9
E x p o r t S t o r e d D a t a
t o U S B D r i v e P r e s s E N T E R o r U p / D n
The last 10 measurement results are stored in memory. Subsequent data measurements will cause the
D155 to delete the oldest stored result first. These data can be cleared using this function. Caution,
data cannot be restored once it is cleared.
1) Press [MENU] then press [] or [] to select Menu 8 as shown below
2) Press [ENTER]
3) Press [ENTER] to clear all 10 stored results, or [STOP] or [MENU] to quit.

4-9. Menu 9: Export Stored Data to USB Drive

28
The D155 memory only stores the last 10 measurement results. Subsequent data measurements will
cause the D155 to delete the oldest stored result first. This function allows the data to be exported to
an external USB storage device in an MS-Excel file format. The D155 will create a folder on the USB
drive named X:/D155/Parameter, where ‘X:’ is the USB drive designation. If this folder already exists,
then it will continue to use this folder. The D155 will create a new file each time data is exported. The
filename will be created using the date and time of the export. Below is an example of the MS-Excel
file format for exported data:
2) Press [MENU] then press [] or [] to select Menu 9 as shown below
4) Press [ENTER]
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 32
M e n u : 9
E x p o r t A L L D a t a ?
E N T E R t o A c c e p t S T O P o r M E N U t o Q u i t
M e n u : 1 0
V i e w S e r i a l N o . &
F i r m w a r e V e r . P r e s s E N T E R o r U p / D n
M e n u : 1 0
S / N : J J A 0 8 B 4 3
V e r : 2 . 0 0 A 2
E N T E R t o C o n t i n u e
Press [ENTER] to export stored results to USB device, or [STOP] or [MENU] to quit.

4-10. Menu 10: View Serial No. & Firmware Ver.

You can check the serial number and firmware version of the unit. Such information will be necessary
to identify your density meter when calling for help.
1) Press [MENU] then press [] or [] to select Menu 10 as shown below
2) Press [ENTER]
29
3) Press [ENTER] to continue back to the menu selection screen.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 33
Menu/Parameter
Description
Setting range
Default
0: Calibrate Cell
Calibration with pure water
Results
Time
59D: Lube Oils 20°C
Temperature
Time
Time format: HH:MM (24 hour format)
12/31/99 23:59
Contrast
Number
Results
stores the last 10 measurement results.
Drive
Firmware

4-11. Parameters and default

30
1: View Stored
2: Set Sampling
3: Select Oil
Table
4: Set Test
5: Set Date &
Display measurement results
Select sampling time 00 to 99s 10s
Select temperature compensation table for oil
23A: Crude Oils 60/60°F
23B: Refined Products 60/60°F
23D: Lube Oils 60/60°F
53A : Crude Oils 15°C
53B : Refined Products 15°C
23A/23B/23D/
53A/53B/53D/
59A/59B/59D
53A
53D : Lube Oils 15°C
59A: Crude Oils 20°C
59B: Refined Products 20°C
Set measurement temperature 15.0 to 25.0°C 20.0°C
Enter date and time. Date format: MM/DD/YY
01/01/00 00:00 to
6: Set LCD
7: Reset Test
8: Clear Stored
9: Export Stored
Data to USB
10: Serial No. &
Adjust LCD brightness 0/25/50/75/100 50
Reset Sample Test No. to ‘01’ 01
Erase the memory of stored results. The D155 only
Export Stored Data from Memory to an MS-Excel
formatted file.
View D155 Serial Number and Firmware Version
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 34
line voltage. Make
the outlet.
Do not place the printer on the top of the main unit in order to avoid the
is configured
as shown below. When you have changed this switch configuration,
1 2 3 4 5 6 7 8
Dip SW-1
OFF
OFF
ON
OFF
OFF
OFF
OFF
OFF

4-12. How to connect the optional printer or a personal computer (PC)

The optional printer can be connected to the D155 unit so that the measurement and calibration results
together with the date and time of measurement can be printed out. Connect the supplied printer cable
to the RS-232C port (DB-9M) on the back of D155 unit as shown.
A personal computer (PC) can be connected instead of the printer so that data can be exported. A
data collection software package – SOFT-CAP Data Capture Software is available as an option with the
D155. Please consult the separate SOFT-CAP Operation Manual for more details on installation and
use.
31
Warning! Use correct AC power adapter suitable for the AC sure to use the proper rated voltage before plugging it into Failure may damage the printer or even cause the adapter itself to catch fire.
Note:
influence magnetic field of the printer on the measuring cell. The DIP switch on the Dot Matrix printer kit (P/N: 95.0239)
refer to the Instruction manual for the printer and set it as shown below:
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 35

5. Maintenance

5-1. Daily check

5-1-1. Cleaning the measuring cell

When the repeatability of measurements has degraded or measuring time is prolonged, the measuring
cell is considered to be contaminated. Also, when the cell is apparently contaminated with a sample of
such nature, the measuring cell must be cleaned with an appropriate solvent like household chlorine
bleach, acid or alkali. When particular samples containing fatty acids or protein (such as biodiesel) are
measured, the cell must be cleaned once a week as described below.
The following example describes how to rinse out fats and oils with Sodium hydroxide:
1) For units with the optional peristaltic pump installed
a. Insert the sampling tube into 0.10.5mol/L Sodium hydroxide and press [PUMP] key to inject.
b. When the cell becomes filled with Sodium hydroxide, press [PUMP] key to stop filling.
c. Leave the sodium hydroxide in the cell for about 15 minutes.
d. Drain out Sodium hydroxide solution with [PUMP] key.
2) For units without the pump
a. Inject 0.10.5mol/L Sodium hydroxide using a suitable 2-5 mL syringe.
b. Leave the sodium hydroxide in the cell for about 15 minutes.
c. Push out the sodium hydroxide from the cell using an empty syringe filled with air.
3) Clean the measuring cell with pure water according to the Step “3-4”.
32
Warning! The alkali may etch the glass of the measuring cell. Avoid rinsing the cell with strong alkali for hours. Always rinse the cell with plenty of pure water after cleaning the cell with alkali cleaners.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 36

5-2. Other maintenance

5-2-1. Replacement of the power fuse

If the power fuse should blow, replace it with a like size and power rating. Note the fuse size and rating
is listed on the rear of the instrument immediately above the fuse holder. As shown below, turn the
fuse holder cover by 90 degrees counterclockwise using a flat screw driver. After the fuse is replaced,
turn the fuse holder cover back 90 degrees clockwise.
33
Warning! Be aware of the danger of electric shock. Turn off the main switch and remove the AC mains power cord before replacing the power fuse.

5-2-2. Adjustment of Injection Adapter (and tightening of tube nozzles)

If leakage from the Injection Adapter (Sample inlet and outlet port nozzles) is observed, they need to be
re-tightened
1) Twist to remove the Connection Tubes with Unions from the Injection Adapter Nozzles.
2) Remove the black protective foam insert
3) Insert the red PTFE tubing into the tube nozzles as shown
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 37
4) Loosen the screw with the provided wrench
5) Tighten the tube nozzles clockwise – hand-tighten only
34
6) Follow steps 1 thru 4 in reverse to tighten the Injection Adapter nut and remove the red PTFE tubing
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 38

5-2-3. Replacement of optional peristaltic pump tubing

If the pump tube is deteriorated, the tubing may fail causing sample liquid to leak. This may eventually
lead to pump failure. Replace pump tubing as follows:
1) Disconnect the tubes connected to the pump tube cassette.
2) Turn the cassette counterclockwise to
remove it.
3) Remove the tube joint from the cassette by
pulling one of the ends, and then pull out the
tube.
35
4) Apply grease (such as petroleum jelly) over
the tube and the axle of pump cassette as
illustrated.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 39
Rotor
Tube joint
Pump tube
Pull out
Axle
Match
Pump tube Fix the cassette by turning clockwise until it stops.
Replace the
5) Fit in the new pump tube around the rotor as
shown by pulling it out a little bit.
Fix the tube joint securely to the groove as
illustrated.
6) Match the axle and rotor, and fix the
cassette by turning clockwise until it stops.
36
Caution! Some samples may shorten the life of the pump tubing. pump tubing when appropriate.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 40
Error message
Description
Remedy
or your local dealer.
message
2, 3, 4
The thermistor is defective (broken).
Contact CANNON or your local dealer.
wiring to the cell is damaged.
local dealer.
stored data/calibrations cleared.

6. Troubleshooting

6-1. Error messages and remedies

When the main unit is operated improperly, the following error messages will appear with a
corresponding beep sound.
Check the window to see if air
bubbles are trapped in the measuring
Time Over Measuring time exceeds 10 minutes.

6-2. Symptoms for error numbers and associated remedies

When a fault occurs within the D155 unit, the following warning number will appear with a beep sound.
cell.
If this message persists while nothing
wrong is observed, Contact CANNON
37
Error
5 The measuring cell is defective or the
6 Temperature is not stabilized in an
hour.
7 The voltage level of the back-up
battery becomes low. This will
cause memory to be initialized and all
Description Remedies
Contact CANNON or your local dealer.
After the following two steps are confirmed,
turn on the power again.
The ambient temperature for D155 is
535°C. Check to see if the ambient
temperature does not exceed this range
during operations.
Make sure that the air vents at the front,
rear, and bottom of unit are not clogged
with dust or blocked in any way. If the
error persists, Contact CANNON or your
Contact CANNON or your local dealer.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 41
START
Is the line voltage
conforming?
Repairing by the
manufacturer
Is everything O.K.?
END
U
se the correct line voltage
Yes
Yes
No
No
Voltage
100-240V
30 Watts
T3.15A/250V
50/60Hz

6-3. Troubleshooting

6-3-1. Cannot turn on the main unit with the power switch

38
(Note 1) Specifications on power source
AC Mains
Wattage Fuse Frequency
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 42
START
Check the air-intake at the
bottom to see if clogged
Check ambient
temperature (5
°C~+35
°
C)
Operate the unit within the range of the specified temperature
No
Yes
Check the air-intake at the bottom to see if clogged
No
Yes
END
Repairing by manufacturer
Is everything O.K.?
No
Yes

6-3-2. “Ready to Test” message does not appear and/or cell temperature is unstable.

Follow the following steps if “Ready to Test” message does not appear after more than 1 hour or the cell
temperature appears unstable after having reached the preset measurement temperature.
39

6-3-3. The unit does not work even though power is on.

If the “Ready to Test” message does not appear after the D155 has been powered on for more than 30
minutes, check on the line voltage. If correct power voltage is supplied, Contact CANNON or your local
dealer.

6-3-4. LCD does not function

If the LCD display is found to be defective (no message appears on screen) or the brightness of display
cannot be adjusted, Contact CANNON or your local dealer.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 43
START
Is the system leaking?
Check tube lines on any loosened joint. Also ensure the cross section of tube is upright.
Yes
No
Contact our local dealer.
Is everything O.K.?
END
Yes
No
Yes
Degas the sample (boil, stir or else)
No
Can the sample be
degassed?

6-3-5. Key entry does not work

If beep sound is not heard each time any keys are pressed or key entry itself does not function, Contact
CANNON or your local dealer.

6-3-6. Air bubbles are easily trapped in the sample and visible in the measurement cell

40
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 44
Contact our local dealer to check the cell.
START
Is the sample homogeneous?
Is there
any repeatability for the
measurement with
pure water?
Is the work bench vibrating?
Any sample residue
in joint portions (tube nozzle A,
joint B or the like)?
Are air bubbles found
in the sample?
Is the measuring cell dirty?
Is the temperature of
measuring cell stable?
Repeat sampling. (When air bubbles
appear again, refer to Section 6-3-6.)
Stir to make it homogeneous. (No repeatability when the sample is unevenly mixed.)
Overhaul the joints and clean them.
Rinse the cell thoroughly
.
Measure after temperature becomes stabilized.
Yes
No
Contact our local dealer.
Is everything O.K.?
END
Yes
No
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
No
No
Measure the sample on a stable bench.
No

6-3-7. There is poor repeatability on measured values

41
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 45
START
Is the tube peristaltic
pump deteriorated?
Replace the tube.
Contact our local dealer.
Is everything O.K.?
END
Yes
No
Yes
No

6-3-8. Sampling time is too long (optional peristaltic pump)

42
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 46
START
Are air bubbles
found in the sample?
Is cell temperature stable?
Measure after the cell temperature becomes stabilized.
Yes
No
Contact our local dealer.
Is everything O.K.?
END
Yes
No
Yes
No
Yes
Repeat sampling and measure again. (If air bubbles appear again, refer to Section 6-3-6.)

6-3-9. Measuring time is too long

43

6-3-10. Data cannot be stored in the memory

The back-up battery needs to be replaced when the clock function does not work or preset parameters
cannot be saved. Contact CANNON or your local dealer.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 47
Type and Model
D155 Density Meter
Discipline
Oscillating Glass U-tube
Measuring Range
0 to 2g/cm3
Measuring Temperature Range
15 to 25℃
Density
±0.001g/cm
3
Temperature
±0.1℃
Repeatability
Density
SD 0.0005g/cm3
Resolution
Density
0.0001g/cm3 step
Temperature
0.05℃ step (Setting Resolution: 0.1℃ step)
Specific Gravity
0.0001 step
Temperature
Parameter
API Degree
0.1 step
Measurement Time
2 to 4 min
Viscosity Correction
None
20℃Density)/ API Degree/ Other messages
(optional peristaltic pump for automatic sampling)
Min. required Sample Amount
Syringe: Approx. 1mL, Pump: 8mL
Maximum Sample Viscosity
Approximately 1000 mPa⋅s when using optional peristaltic pump
Calibration
Pure Water (factory/service calibration of air)
Storage of Measurement Result
Last 10 measurement results stored in internal memory
Temperature Compensation Table for Petroleum Products
ASTM D1250 Table 23(A,B,D), 53(A,B,D), 59(A,B,D) Interface
RS-232C×1ch, USB×1ch
2) Humidity: Below 85%RH (No condensation)
Power Source
AC100~240V 50/60Hz
Power Consumption
Approx. 30W
Dimension
251mm (W) × 408mm (D) × 165mm (H) [10” x 16” x 6.5”)
Weight
11.6kg (25.64 lbs)
- Tube 0.5x1.5 L=400 PTFE: Qty 1
- Soft-Cap Data Capture Software

7. Others

7-1. Basic specifications

Precision
44
Compensated
Display 1) Screen: 4×20 LCD backlight
Sampling Method Manual sampling by syringe
Ambient Conditions for unit 1) Temperature: 5 to 35
0.0001 step
2) Displayed Parameter: Temperature (℃)/ Density/ Specific Gravity Number of temperature compensation tables for petroleum products Temperature Compensated Parameter (60Specific Gravity/ 15
Density/
Standard Supplied Accessory
- Connection Tube: Qty 1
- Syringe 2ml: Qty 5
- AC Power Cord: Qty 1
- Wrench: Qty 1
Optional Accessories
- Dot matrix printer
- Peristaltic pump
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 48
T
dVc Mc
K
=+2π
d
K
Vc
T
Mc
Vc
=
4
2
2
π
F
K
Vc
d d
T T
a
w
a w
= =
4
2 2 2
π
d
d F T T
a a
=
( )
2
2

7-2. Principle of measurement

7-2-1. Oscillating U-tube Cell

When a measurement cell filled with liquid or gas is oscillated spontaneously, resultant oscillating cycle
varies with the density of the sample in the measurement cell. The oscillating cycle T is expressed
follows.
where,
d = Density of sample in measurement cell
Vc = Volume of sample in measurement cell, or measurement cell internal volume
Mc = Mass of measurement cell
K = Constant
45
Supposing that the oscillating cycles Ta and Tw are obtained through measurement of reference
materials of known density (density da and dw), the facto value F is calculated as follows.
Accordingly, the density “d” of unknown sample can be calculated according to the following formula by
measuring its oscillating cycle T.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 49
25.1013
)(
)(00367.01
0012932.0
)/(
3
hPaP
Ct
cmgd ×
×+
=
Temp.
(°C)
Density
(g/cm3)
Temp.
(°C)
Density (g/cm3)
Temp.
(°C)
Density
(g/cm3)
Temp.
(°C)
Density
(g/cm3)
0
0.00129
25
0.00118
50
0.00109
75
0.00101
1
0.00129
26
0.00118
51
0.00109
76
0.00101
2
0.00128
27
0.00118
52
0.00109
77
0.00101
3
0.00128
28
0.00117
53
0.00108
78
0.00101
4
0.00127
29
0.00117
54
0.00108
79
0.00100
5
0.00127
30
0.00116
55
0.00108
80
0.00100
6
0.00127
31
0.00116
56
0.00107
81
0.00100
7
0.00126
32
0.00116
57
0.00107
82
0.00099
8
0.00126
33
0.00115
58
0.00107
83
0.00099
9
0.00125
34
0.00115
59
0.00106
84
0.00099
10
0.00125
35
0.00115
60
0.00106
85
0.00099
11
0.00124
36
0.00114
61
0.00106
86
0.00098
12
0.00124
37
0.00114
62
0.00105
87
0.00098
13
0.00123
38
0.00113
63
0.00105
88
0.00098
14
0.00123
39
0.00113
64
0.00105
89
0.00097
15
0.00123
40
0.00113
65
0.00104
90
0.00097
16
0.00122
41
0.00112
66
0.00104
17
0.00122
42
0.00112
67
0.00104
18
0.00121
43
0.00112
68
0.00103
19
0.00121
44
0.00111
69
0.00103
20
0.00120
45
0.00111
70
0.00103
21
0.00120
46
0.00111
71
0.00103
22
0.00120
47
0.00110
72
0.00102
23
0.00119
48
0.00110
73
0.00102
24
0.00119
49
0.00110
74
0.00102

7-2-2. Density Tables

Dried air density
Density of dried air at t°C and 1013.25hPa is calculated according to the following formula.
Density of air at 1013.25hPa of air pressure;
46
Chemical Handbook Fundamental Version, Rev. 3, Table 5・1
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 50
(°C)
(g/cm3)
(°C)
(g/cm3)
(°C)
(g/cm3)
0
0.99984
25
0.99705
50
0.98805
1
0.99990
26
0.99679
55
0.98570
2
0.99994
27
0.99652
60
0.98321
3
0.99996
28
0.99624
65
0.98057
4
0.99997
29
0.99595
70
0.97779
5
0.99996
30
0.99565
75
0.97486
6
0.99994
31
0.99534
80
0.97183
7
0.99990
32
0.99503
85
0.96862
8
0.99985
33
0.99471
90
0.96532
9
0.99978
34
0.99438
10
0.99970
35
0.99404
11
0.99961
36
0.99369
12
0.99950
37
0.99333
13
0.99938
38
0.99297
14
0.99925
39
0.99260
15
0.99910
40
0.99222
16
0.99894
41
0.99183
17
0.99878
42
0.99144
18
0.99860
43
0.99104
19
0.99841
44
0.99033
20
0.99821
45
0.99022
21
0.99799
46
0.98980
22
0.99777
47
0.98937
23
0.99754
48
0.98894
24
0.99730
49
0.98849
Water density
Temp.
Density
Temp.
Density
Temp.
47
Density
Chemical Handbook Fundamental Version, Rev. 3, Table 5・2
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Page 51

7-3. Warranty/Return Information

1. Products limited warranty
In addition to other manufacturers’ warrantees, CANNON® Instrument Company (“the Company”) warrants all products (other than reagents and chemicals) delivered to and retained by their original purchasers to be free from defect in material and workmanship for one year from the date of the Company’s invoice to the purchaser. For a period of one year from the date of such invoice, the Company will correct, either by repair or replacement at the Company’s sole election, any defect in material or workmanship (not including defects due to misuse, abuse, abnormal conditions or operation, accident or acts of God, or to service or modification of the product without prior authorization of the Company) without charge for parts and labor. The determination of whether any product has been subject to misuse or abuse will be made solely by the Company.
The Company shall not be liable for any special, incidental, or consequential damages, or any damage to plant, personnel, equipment or products, directly or indirectly resulting from the use or misuse of any product. Representations and warranties made by any person, including dealers and representatives of the Company, which are inconsistent, in conflict with, or in excess of the terms of this warranty shall not be binding upon the Company unless placed in writing and approved by an officer of the Company.
2. Reagent and chemical warranty
48
CANNON® Instrument Company (“the Company”) warrants all reagents and chemicals sold by the Company and delivered to and retained by their original purchasers to conform to the weight, specifications and standards stated on the package. The Company will, at its sole option, either replace or refund the price (net of freight, handling charges and taxes), of any reagent or chemical sold by the Company which does not conform to such weight, specifications and standards upon the prompt return of the unused portion. Except for replacement or refund of the net price, the Company shall not be liable for any damages occurring as a consequence of the failure of any reagent or chemical sold by the Company to conform to the weight, specifications and standards stated on the package.
CANNON® D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA
Loading...