Cannon CT-1000 User Manual

CONTENTS
i
INTRODUCTION 1
Overview ............................................................................................................................. 1
Safety features ...................................................................................................................... 3
Operator Safety .................................................................................................................. 3
UNPACKING AND ASSEMBLY 5
Unpacking the CT -1000 ....................................................................................................... 5
Assembly procedure ............................................................................................................. 5
Glass jar installation .............................................................................................................. 6
Motor-stirrer ............................................................................................................ 8
Inserting viscometer tubes/thermometers................................................................................ 9
Filling the bath .................................................................................................................... 10
Draining the Bath ................................................................................................................ 11
BATH OPERATION 13
Turning on the CT -1000...................................................................................................... 13
Setting the bath temperature ................................................................................................ 14
Fixed temperature adjustments ................................................................................ 14
V ariable temperature adjustments ............................................................................ 14
Setting the Limit Control ..................................................................................................... 15
A
B
C
D
APPENDIX A CT-1000 PROBLEM ANALYSIS 17
APPENDIX B CORRECTING SHAFT & IMPELLER MISALIGNMENT 19
Shaft run-out correction ...................................................................................................... 19
APPENDIX C CHOOSING A TEMPERATURE BATH LIQUID 21
APPENDIX D ADJUSTING TRIMPOTS (POTENTIOMETERS) 23
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
ii
E
F
APPENDIX E CT-1000 SPARE PARTS LIST 25
WARRANTY/RETURN INFORMATION 27
Products limited warranty.................................................................................................... 27
Reagent and chemical warranty ........................................................................................... 27
Returning a product to CANNON®.................................................................................... 28
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
1
1
Overview
CT-1000 function
Temperature range
Manual/video
INTRODUCTION
The CANNON® CT-1000 Constant Temperature Bath is designed to maintain precise temperatures at a wide range of settings for accurate viscosity measurements. Because of its temperature stability and ease of use, it is also suitable for any other application where temperatures must be maintained within close tolerances.
The CANNON® CT-1000 will maintain temperatures of 20°C to 100°C within 0.01°C and temperatures of 101°C to 150°C within 0.03°C.
This manual is intended to provide information on the installation, characteristics and operation of the CANNON® CT-1000 Constant T emperature Bath. The manual materials are supplemented and enhanced
CANNON
Figure 1: The CT-1000
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
2
by an instructional video shipped with the bath.
Temperature selection
Bath description
T en of the most commonly-used temperatures for kinematic viscosity measurement can be set by using a switch on the bath front panel. The bath will equilibrate within a fraction of one degree of the desired temperature. A fine-tuning control permits further temperature adjust­ments. By switching to the variable setting on the front panel, the opera­tor can set any temperature within the operating range of the instrument.
The bath chamber is a cylindrical clear Pyrex® vessel 300 mm (12 inches) in diameter and 300 mm (12 inches) high. A stainless steel baf fle is located in the center of the bath and provides a convenient backdrop for viewing viscometers placed in the bath. The top cover contains seven round holes 51 mm (two inches) in diameter for insertion of viscometer holders. T wo additional holes are provided for thermometers. Twin fluorescent lamps provide glare-free illumination of the bath.
A solid-state control circuit equipped with a stainless steel-encased thermistor provides proportional temperature control. A motor-driven stirrer ensures that a uniform temperature is maintained throughout the bath.
The bath housing is fabricated from heavy aluminum and coated with a corrosion-resistant epoxy . The top cover consists of three layers; a stainless steel top surface, an insulating layer, and a bottom stainless steel heat reflector.
Temperature stability
The temperature control of ± 0.01°C below 100°C, and ± 0.03°C from 100 to 150°C, provides the sensitivity required by ASTM D 445 for kinematic viscosity measurements. A built-in cooling coil, when con­nected to tap water or a cooling system, permits operation below or slightly above ambient temperature.
snoitacificepS0001-TC
snoisnemiD
)sehcni32x52.81x52.71(
hgihmm485xpeedmm464xediwmm834
yticapaC )lag5.4(L71
thgieW )bl59(gk34
thgieWgnippihS )bl321(gk8.55
sttaw0041,zH06/05,CA)ledomtlov-021(02A-6279
rebmuNgolataC
sttaw0041,zH06/05,CA)ledomtlov-042(52A-6279
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
Safety features
3
Overheat thermistor
Thermistor detection/cutoff
Liquid-level control float
Operator Safety
A thermistor in the bath senses any over-temperature fault condition. If such a condition occurs, all power is removed from the bath until an operator resets the over-temperature limit circuit.
If the control thermistor is disconnected, all power to the bath heaters is cut off.
Operation of the bath is not possible unless it is filled with liquid to a safe operating level. A liquid-level control float prevents the control circuit from heating the bath until the safe operating level is attained. The bath heaters are automatically turned off if the bath liquid drops below the minimum safe level.
All technicians who use the CT-1000 should follow these basic safety procedures:
Do not place the CT-1000 system on an unstable cart or stand. The
CT-1000 should be placed on a stable laboratory table or bench.
If any liquids are spilled into the electronic components of the CT-
1000, contact CANNON
®
Instrument Company immediately .
Do not position power cords so that they are likely to be walked on
or pinched by items placed on or against them. Keep all connections as neat as possible.
If the CT-1000 will not be used for an extended period of time,
unplug the power cord from the wall outlet. To disconnect the power cord, pull it out by the plug. Never pull the cord itself.
CAUTION
Do not attempt to service the CT-1000 system by removing panels and trying to effect repairs. Contact CANNON® Instrument Company regard­ing service and repair needs.
CAUTION The bath fluid used with the CT-1000 may be dangerous. Use the proper
safety precautions when handling the bath fluid in use (refer to the Material Safety Data Sheet included with the bath fluid for more detail.)
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
4
This page intentionally left blank.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
5
2
UNPACKING AND ASSEMBLY
This chapter of the manual provides assistance in unpacking and assem­bling the CT-1000 Constant Temperature Bath. For additional assistance, consult the instructional video.
Unpacking the CT-1000
The CANNON® CT-1000 Constant Temperature Bath is shipped in two boxes:
Box 1 contains the glass jar, front glass panel pieces, jar gasket top,
trimpot adjust screwdriver, Allen wrench, seven hole covers, rubber thermometer holder and instruction manual.
Box 2 contains the bath housing, including the electronics drawer.
Box 2 also contains a smaller box with the motor and stirrer, includ­ing the impeller and mounting plate.
The bath unit housing is shipped completely assembled. However, the glass jar, the glass panels, and the motor and stirrer must be installed in it. T o allow this, some disassembly of the bath unit housing is required. The tools required are a utility knife, phillips screwdriver, and a 1/8" Allen wrench. The Allen wrench is included with the bath.
Assembly procedure
1. Unpack the bath unit housing.
CANNON
Figure 2: Primary bath components after unpacking
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
6
2. Move the bath unit housing to its permanent loca­tion on the laboratory bench.
3. Remove all eight screws from the stainless steel top covers (see Figure
3).
4. Disconnect all probes, heaters, and float switch from the rear panel.
5. Remove the front top cover and rear top cover (see Figure 4 and 5). Use caution when removing the rear top cover because the tempera­ture control probes and heating elements are attached to it.
Figure 3: Removing screws from top cover
Figure 4: Removing front top cover
Glass jar installation
1. Remove the glass jar from its
WARNING
CANNON
The glass jar is heavy. Use caution when lifting it.
2. Make sure the rubber support
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
box.
ring is seated properly around the bottom jar opening in the bath unit (see Figure 6).
Figure 5: Removing rear top cover
Figure 6: Checking support ring
3. Place the rubber gasket around the top rim of the jar (see Figure 7). The rubber may have to be trimmed slightly to allow the ends of the rubber gasket to meet with no gap when placed around the rim.
4. Remove the large piece of foam packing from the inside of the cabinet. Also remove the small piece of foam from the float level,
Figure 7: Placing gasket
located on the upper left­hand corner of the inside of the cabinet.
5. Lower the glass jar into the cabinet so it seats evenly on the rubber support ring (see Figure 8).
7
CANNON
Figure 8: Installing bath jar Figure 9: Placing inner glass plate
6. Unwrap the two glass panels. Place the thinner of the two piece of glass in the slot closest to the jar (see Figure 9).
7. Place the wider (tempered) piece of glass in the front slot furthest away from the jar. The middle slot is left empty as a vapor barrier.
8. Replace the rear top cover. Align the four holes, then insert and tighten the screws.
9. Replace the front top cover. Line up the four holes, then insert and tighten the screws.
10. T o ensure that the gasket forms a tight seal with the top covers of the bath, proceed as follows:
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
8
11. Pull out the drawer using the handles provided, disconnecting the two connection cables as you do so.
CAUTION
You will need to pull the AC line (power) cord through the rear panel opening as you remove the drawer—you should not try to remove the cord until you loosen the screw securing the power cord to its connection on the back of the drawer).
12. Press down or pull up on the plastic release bars on either side of the drawer track to release the drawer, then pull the drawer completely free of the unit and set it aside.
13. When the drawer is removed, locate the four 1/4-20 set screws visible at the top of the drawer opening underneath the bath.
14. Turn the set screws clockwise with the Allen wrench (in­cluded with the bath) until the top of the jar forms a tight seal with the covers (see Figure 10). Make sure you tighten the set screws uniformly so the jar remains level.
FIgure 10: Adjusting seal with set screws
Motor-stirrer
NOTE
15. Run the AC cord through the rear panel opening
16. Replace the drawer in the slide tracks and push the drawer back into its opening. Insert the two plugs into the rear of the drawer assembly .
Complete the assembly of the bath, including the motor-stirrer installa­tion, per the instructions below:
1. T ake the motor/stirrer from its box. Remove the two screws on the top heater housing and lift off the housing (see Figure 11).
2. After checking the impeller blades to make sure that the flat sections all lie in the same plane (see Appendix B), insert the motor stirrer in the opening provided (see Figure 12, next page).
To keep from accidentally bending the motor shaft, do not hold the motor assembly by the shaft. Use care when inserting the motor shaft and impeller to prevent damage to delicate components.
Figure 11: Removing housing
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
9
NOTE
Two screws located on either side of the opening for the motor stirrer serve as locating pins for the motor support pad. Do not remove these screws; the holes in the pad fit loosely over their heads. The motor line cord should point toward the rear of the bath (offset slightly to the right or left). The motor stirrer should now lie flat on the top of the bath.
3. Reattach the top heater
Figure 12: Installing the motor
housing , making sure that the heater, motor, and fan cords pass through the left-hand opening (as viewed from the rear) and that the control probe, over-temperature probe, and level switch cords exit from the right opening (as viewed from the rear). The back lip on the rear top cover fits into the slot on the top heater housing. Line up the holes, insert screws, and tighten.
4. Connect all plugs and probes to the correspondingly labeled sockets at the rear of the CT-1000 bath unit.
5. Adjust the four feet on the bottom of the bath housing to level the bath. This should be done before filling the bath with fluid.
Inserting viscometer tubes/thermometers
The top cover of the CT-1000 contains seven apertures, 51 mm (2”) in diameter, for the insertion of viscometer tube holders. Two additional holes are provided for insertion of thermometers.
Inserting viscometer tubes
NOTE
Thermometer immersion
If necessary , remove the viscometer tube hole cover(s) from the top of the bath and carefully place the viscometer tube(s), with the proper holder attached, into the bath through the aperture(s) in the top cover .
After filling the bath with fluid, adjust the height of the viscometer tube(s) to ensure that the liquid under test and/or any timing marks on the tube are a minimum of 6 mm (¼") below the top level of the liquid.
Proper thermometer immersion is critical for viscosity measurements. Even a calibrated thermometer will read incorrectly if is it improperly immersed in the bath. “T otal immersion” kinematic viscosity thermom­eters should be used with the bulb and only the mercury column beneath the surface of the liquid, but with the emergent stem above the surface at ambient temperatures.
NOTE
CANNON
Different thermometers have different immersion requirements. Refer to the information included with the thermometer in use for specific instruc­tions.
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
10
Filling the bath
After CT-1000 assembly is complete, you are ready to fill the bath.
CAUTION
Make sure that the bath is placed in its intended final position before adding bath fluid. The CT-1000 should not be moved with bath fluid in the bath jar. NEVER USE FLAMMABLE BATH LIQUIDS.
1. Make sure that the instrument power is OFF and select a bath liquid appropri-
Figure 13: Filling the CT-1000 bath
ate to your operating temperature range (see APPENDIX C).
2. Fill the jar with bath liquid at ambient temperature to a level suffi­cient to engage the float switch. This float permits bath operation when the minimum amount of fluid has been added to the bath jar.
3. Continue to add fluid until the bath liquid level has risen to approxi­mately 40 mm (1.5") of the top of the jar.
4. Turn the instrument power ON and incrementally heat the bath to desired control temperature while monitoring the bath liquid level carefully . The bath level must be 15-20 mm (approximately ½" to ¾") from the top of the jar at the control temperature. If it becomes apparent that this liquid level will not be achieved, return the bath to within 10°C of ambient, turn the instrument power OFF and add or remove liquid as necessary .
5. Repeat step four until you have attained the proper bath liquid level at the desired control temperature.
CAUTION
Different bath fluids expand at different rates. Do not overfill the bath!
WARNING Monitor the level of bath liquid closely when operating the CT-1000 at
higher temperatures (100-150°C). The bath liquid will expand as the temperature increases. The CT-1000 bath jar is not designed to contain liquid under pressure. If the bath is overfilled, liquid may overflow.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
Draining the Bath
11
If it becomes necessary to drain the liquid from the bath:
Obtain a suitable container to hold all of the liquid drained from the bath (approximately 20-22 liters or 4.5 - 5 gallons). Make sure that the bath liquid is within 10°C of ambient temperature. Then insert a tube into the bath chamber from the top opening and siphon the liquid from the bath into a container positioned lower than the bath.
WARNING
Always use a rubber bulb or similar device to apply suction to a tube containing bath liquids.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
12
This page intentionally left blank.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
13
3
NOTE
Turning on the CT-1000
The various functions controlling the bath are printed at the bottom of the front panel of the CT-1000. These functions, reading from left to right, are: 1) Power, 2) Temperature, 3) Temperature Adjust, 4) Preheat, and 5) Limit Control. In this manual the five functions are printed in boldface upper- and lowercase letters (to match their typeface on the panel). The switches and dials controlling each function are also printed in boldface type, but in capital letters (to match their typeface on the panel).
1. Turn the Power BATH switch on.
BATH OPERATION
NOTE
The bath heaters will not be activated if there is no liquid in the bath.
2. Set the Limit Control TEMPERATURE ADJUST turn the dial clockwise to its furthest setting to enable the bath to heat properly . Proper technique for resetting the dial after the bath reaches operat­ing temperature is described in the instructional video and in the following section of the manual
3. Turn on the Power LIGHT switch to illuminate the interior of the bath.
4. Move the Temperature Adjust SELECT switch to the left.
5. After releasing its lock, move the T emperatur e Adjust FIXED dial to 50. Relock the dial.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
14
NOTE
All the dials except Temperature SELECT are equipped with a dial lock to prevent accidental changing of dial settings. Push the lock upward to release the dial. When the dial has been set to the proper position, push the lock downward to relock the dial.
Setting the bath temperature
There are two ways to set the bath temperature, depending on whether or not the desired temperature is on the Temperature SELECT dial.
Fixed temperature adjustments
T o set the bath to a temperature on the Temperature SELECT switch:
1. Choose a temperature using the Temperature SELECT switch.
NOTE
The Preheat switch activates an auxiliary heater for the CT-1000. You may turn it ON to heat up the bath more rapidly. You should turn the Preheat switch OFF when the desired temperature is reached if the operational temperature is under 120°C. For bath temperatures in excess of 120°C, the Preheat switch must remain ON to maintain temperature control. After the Preheat switch has been turned ON, the Preheat HEAT light will glow continuously until the bath temperature is approximately that selected with the Temperature SELECT switch, when the light will begin to blink.
The light above the T emperature Adjust control will glow steadily when the bath is heating. The light will start to blink when the bath temperature is approximately that selected with the Temperature SELECT switch.
2. T o adjust the bath to the exact temperature, read from the kinematic viscosity thermometer (see Appendix F), then release the dial lock on the T emperature Adjust FIXED dial and turn the dial to incremen­tally adjust the temperature. To decrease the temperature, turn the dial counterclockwise; to increase the temperature, turn the dial clockwise. Relock the T emperature Adjust FIXED dial after each adjustment and allow several minutes for the bath to equilibrate.
NOTE
If the desired temperature cannot be obtained by turning the Tempera­ture Adjust FIXED dial, it may be necessary to adjust the small trimpots
on the circuit board in the electronics drawer (see Appendix D).
Variable temperature adjustments
T o set the bath to a temperature that is not on the Temperature SELECT switch you should use the following procedure:
1. Turn the Temperature SELECT switch to the temperature closest to that desired.
CANNON
2. Move the Temperature Adjust SELECT switch to the right.
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
15
NOTE
NOTE
The Preheat switch activates an auxiliary heater for the CT-1000. You may turn it ON to heat up the bath more rapidly. You should turn the Preheat switch OFF when the desired temperature is reached if the operational temperature is under 120°C. For bath temperatures in excess of 120°C, the Preheat switch must remain ON to maintain temperature control. After the Preheat switch has been turned ON, the Preheat HEAT light will glow continuously until the bath temperature is approximately that selected with the Temperature SELECT switch, when the light will begin to blink.
The light above the T emperature Adjust control will start to blink when the bath temperature has approximated the setting on the Temperature SELECT switch.
3. T o adjust the bath to the exact temperature, read from the kinematic viscosity thermometer (see Appendix F), then release the dial lock on the T emperatur e Adjust V ARIABLE dial and turn it. To incremen­tally decrease the temperature, turn the dial counterclockwise; to increase the temperature, turn the dial clockwise. Relock the Tem- perature Adjust V ARIABLE dial after each adjustment and allow several minutes for the bath to equilibrate.
If a desired bath temperature is within two or three degrees of one on the Temperature SELECT dial, it is sometimes possible to attain the tem­perature by using the Temperature Adjust FIXED dial. Because this dial has a finer adjustment than the Temperature Adjust VARIABLE dial, the final temperature may be reached more quickly. If you wish to try this alternate method, follow the procedure for fixed temperature adjustments given on the previous page.
NOTE
To attain temperatures below ambient temperature (20-25°C), a coolant must be pumped through the bath’s cooling coil. The coil is accessed through the rear of the bath. Coolant temperature should be approxi­mately 3°C below the desired bath temperature. Temperature stability below ambient temperature may be limited by the flow rate of the cool­ant.
Adjusting the High Temperature Limit Control
Using a screwdriver, slowly turn the LC- TEMPERA TURE ADJUST control counterclockwise until the OVERTEMP message lights up on the LC PUSH TO RESET button.
Turn the control clockwise approximately 1/4 to 1/2 turn. Push the LC- PUSH TO RESET button. If the bath doesn’t recover, repeat step 2 and try again.
NOTE
The temperature may change slightly when you set the Limit Control, but the bath temperature will quickly stabilize.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
16
This page intentionally left blank.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
17
A
Problem
Bath does not appear to have power.
Bath illumination not functioning.
Bath liquid not agitated.
APPENDIX A CT-1000
PROBLEM ANALYSIS
Probable Cause
• Power cable not connected to outlet or rear of electrical drawer.
• Over-temperature control set too low.
• Power out on mains.
• Separate switch for LIGHT must be on.
• fluorescent lamps may be defective.
• Lamps may be out of sockets.
• Lamp ballast may be defective.
• Power supply to ballast may not be working properly.
• Check connection of stirring motor on rear panel.
• Check for working motor by connecting to outlet with the proper voltage (see motor label for correct voltage).
Bath does not heat.
Bath control outside of specific limits.
Bath top surface temperature too high.
Air bubbles in bath fluid.
• Check connections for sensors on rear panel.
• Check temperature setting — it must be above existing bath temperature for heat to go on.
• Bath fluid level may be too low.
• Limit Control Temperature Adjust dial may be set too low (turn completely clockwise to 999, then press PUSH TO RESET button).
• Bath fluid viscosity may be too high (if the fluid is too viscous at the desired temperature the stirring will be inadequate, resulting in poor control).
• Check for normal function of stirring motor and motor impeller.
• Remove control thermistor probe plug and check resistance with an ohmmeter (call CANNON temperature).
• Check cooling fan on rear panel.
• Check cooling fan in top assembly.
• Level of fluid may be too low.
• Stirring impeller may be on shaft with the wrong orientation.
• Bath fluid may be too viscous for operation at this temperature.
®
for proper resistance at bath
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
18
This page intentionally left blank.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
19
B
NOTE
Set Screw Alignment
APPENDIX B CORRECTING
SHAFT & IMPELLER
MISALIGNMENT
Motor-stirrer units supplied by CANNON® Instrument Company are checked after assembly to ensure minimum run-out (deviation from concentric rotation) at the impeller end of the shaft. The following instructions are intended to assist those who may be experiencing exces­sive vibration in the motor-stirrer because of shaft run-out or misalign­ment of the impeller blades.
See the instructional video on the CT-1000 for a detailed description of the procedure described below.
The motor shaft contains a flat area. The coupling should be oriented in such a way that the set screw is aligned with the flat. The stirrer shaft and motor shaft should be inserted to approximately the same length in the coupling.
You can test for shaft run-out and impeller blade alignment by placing the motor on a soft surface with impeller facing up. Connect the motor to the appropriate AC power source and observe the impeller and shaft. If run-out is severe, the motor will vibrate and the shaft will flutter notice­ably . The impeller will appear blurred if the impeller blades are not in the same plane.
If necessary you can adjust the impeller blades by bending the large horizontal sections of each of the four segments so that they lie in the same plane. Since these segments are easily bent, check alignment whenever the impeller is bumped, or if there is any suspicion that the blade segments may have become misaligned.
Shaft run-out correction
The following procedure is suggested to correct shaft run-out:
Grasp the motor firmly with the left hand and touch the tip of a soft crayon (a glass marking pencil, for example) momentarily to the side of the shaft near the impeller end while the motor is running. It is advisable to remove the impeller while carrying out this procedure. The mark produced by the crayon will be evident as a line part way around the shaft.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
20
Disconnect the motor, hold the motor tightly against the body, and grasp the shaft in the right hand. Bend the shaft away from the line segment on the shaft.
Repeat the test procedure, marking the shaft at a location slightly re­moved from the first crayon mark.
If the mark appears at the same side of a shaft as the first mark and is about the same length, the bending was not great enough to alter the condition and should be repeated with more force applied. If the mark has shifted to the opposite side, the force was too great and the shaft must be bent back with less force to correct the situation.
This is a trial-and-error process which usually must be repeated several times. When the crayon makes a line at least two-thirds of the way around the shaft, the run-out has diminished to an acceptable level. A uniform line completely around the shaft would indicate no run-out detectable by this procedure.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
21
C
APPENDIX C CHOOSING A
TEMPERATURE BATH LIQUID
The ideal bath liquid would possess low viscosity , high heat capacity, and low vapor pressure over a wide range of temperatures. In addition, the liquid should have a very high flash point and be relatively low in cost. If the fluid is to be used in a kinematic viscosity bath where it is necessary to view the instruments through the bath liquid, then it is important for the liquid to be clear and without color. Unfortunately, no single fluid meets all these requirements.
Because no single fluid can be used at all possible bath temperatures, the choice of a suitable fluid must begin by establishing the temperature range over which the bath will be operated. The following table provides a list of operating temperature ranges and some liquids suitable for use in these ranges:
Methyl alcohol
)C°002+otC°001-(SNOITPODIULFHTAB
erutarepmeT
)C°(egnaR
C°01+otC°001-lohoclAlyhteM
C°02+otC°01-senociliSytisocsiV-woL,lohoclAlyporposI
C°5+
Methyl alcohol is relatively volatile, has a low flash point, and has a degree of toxicity . The only reason for using it is that there seems to be no reasonable substitute. There are some very expensive halogen-based fluids which might be considered, but these also have a high volatility and may be somewhat toxic.
C°06+ot
C°001+otC°06+
C°531+otC°001+)tSc02,diulf002woD(senociliS C°002+otC°531+)tSc05,diulf002woD(se
)tSc1,diulf002woD(
)tSc01,diulf002woD(
nociliS
sdiuqiLhtaBelbatiuS
senociliS,sliOytisocsiVwoL,retaW
senociliS,rotibihnInoitadixOhtiwsliOetihW
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
22
Isopropyl alcohol
Silicone fluids
Water
Refined white oils
Isopropyl alcohol is less toxic than methyl alcohol and somewhat less volatile. However, it becomes very viscous at low temperatures and is therefore unsuitable for use at very low temperatures.
Silicone fluids are available in a wide range of viscosities and can be used over a wide range of temperatures if the proper selection of viscos­ity is made for the temperature range of interest. Silicones are also relatively expensive liquids. However, a bath containing silicones requires extra care when used for capillary viscometry . If silicones are introduced into a viscometer capillary , its calibration factor will be altered by a significant amount.
Water is almost the ideal fluid in the temperature range in which it can be used. Because in some cases there is a tendency for algae formation, some degree of water treatment may be necessary . Water can be used at temperatures close to the boiling point, but water replenishment to offset evaporation becomes a nuisance and the hot vapor can make working above the bath uncomfortable. Also, it may be dif ficult to establish optimum control at elevated temperatures because of the rapid cooling resulting from surface evaporation.
Refined white oils (paraffin oils) of relatively low viscosity can be used at temperatures above the level at which water becomes unsatisfactory . Because these oils will turn faintly yellow and continue to darken with prolonged exposure to heat, we recommend adding an oxidation inhibitor to retard discoloration. The addition of an inhibitor will prolong the useful life of the oil, but it will eventually become as dark as untreated oil.
The search for more suitable bath oils is unending. Hydrogenated veg­etable oils, coconut oil, synthetic oils, and certain chemical compounds have been used with some success at various temperatures.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
23
D
APPENDIX D ADJUSTING
TRIMPOTS (POTENTIOMETERS)
The ten trimpots located immediately behind the front control panel can be accessed by pulling the drawer forward a few inches. An opening in the drawer cover allows direct access without the necessity of removing the cover. The trimpots are adjusted with a small screwdriver (see Figure
15).
There is one trimpot corre­sponding to each of the set temperatures on the dial. These are normally set so that the
T emperature Adjust FIXED
dial will be at the midpoint (500 on the ten-turn dial) of range when at the control temperature. Setting the T emperature Adjust FIXED dial at the midpoint ensures maximum adjustment in either direction to establish the desired temperature if additional adjustment is necessary .
Figure 15: Adjusting potentiometers
CANNON
If a temperature other than one that is available on the T emperatur e SE­LECT switch is to be accessed frequently , it may be possible to obtain this temperature by adjusting the trimpot that coincides with the Temperature SELECT temperature that is close to the desired temperature.
For example, if 100°F is a temperature that is often needed, and 40°C is not used, set the Temperature SELECT switch to 40°C and adjust the trimpot until the kinematic viscosity thermometer indicates the bath is controlling at 100°F. It is necessary to note for future reference that the 40°C position on the select dial is now 100°F. In some cases it may also be necessary to set the T emperature Adjust FIXED dial to a position other than the midpoint in order to obtain the desired temperature.
While the method described above will permit the Temperature SE- LECT setting to be altered, it may not be possible to obtain settings for temperatures that are widely different from those on the Temperature SELECT dial. In this case, use the T emperature Adjust V ARIABLE dial to obtain the desired temperature.
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
24
This page intentionally left blank.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
25
E
APPENDIX E CT-1000 SPARE
PARTS LIST
Following is a list of parts for the CT-1000 which may be reordered from
CANNON® Instrument Company .
Part Number Description
P20.1 PYREX JAR 12 X 12 P20.22 THERMOMETER HOLDER (RUBBER) P22.26 JAR TOP GASKET P22.39 HOLE COVERS & THERM HOLDER SET P22.40 JAR SUPPORT GASKET P27.1260 SOCKET LAMP P27.1300 FLOURESCENT LAMP P27.1310 LAMINA TED SAFETY GLASS P27.1320 PLATE GLASS P27.2190 LIGHTED SWITCH ROCKER 120V P27.2220 P ANEL MOUNT LED P27.2230 TEMPERA TURE RESET SWITCH P27.2230.11 LAMP RESET SWITCH P27.2250 SOLID ST A TE RELA Y 25A P27.2280 BALLAST SOILD ST A TE P27.2330 15 AMP CIRCUIT BREAKER 120V P27.2331 8 AMP CIRCUIT BREAKER 240V P27.3700 SCREWDRIVER TRIMPOT ADJ P27.5110 FRONT BA TH COVER W/ROUND HOLES P27.5250 BAFFLE, WHITE P27.5260 COOLING COIL (SS) P27.5270 HEATER 700W 120V P27.5290 HEATER 400W 120V P27.5291 HEATER 400W 240V P27.5320 TEMPERA TURE CONTROL PROBE P27.5330 TEMPERA TURE PROBE OVER P27.5410.1 LEVEL SWITCH P27.6100 CT1000 MOTOR STIRRER 120V P27.6101 CT1000 MOTOR STIRRER 240V P27.6121 SPONGE SUPPORT PA D P50.82 10 TURN DIAL
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
This page intentionally left blank.
F
Products limited warranty
In addition to other manufacturers’ warranties, CANNON® Instrument Company (“the Company”) warrants all products (other than reagents and chemicals) delivered to and retained by their original purchasers to be free from defect in material and workmanship for one year from the date of the Company’s invoice to the purchaser. For a period of one year from the date of such invoice, the Company will correct, either by repair or replacement at the Company’s sole election, any defect in material or workmanship (not including defects due to misuse, abuse, abnormal conditions or operation, accident or acts of God, or to service or modifica­tion of the product without prior authorization of the Company) without charge for parts and labor. The determination of whether any product has been subject to misuse or abuse will be made solely by the Company .
WARRANTY/RETURN
INFORMATION
The Company shall not be liable for any special, incidental, or consequen­tial damages, or any damage to plant, personnel, equipment or products, directly or indirectly resulting from the use or misuse of any product sold by the Company except as set forth in and limited by the foregoing warranties. Representations and warranties made by any person, includ­ing dealers and representatives of the Company , which are inconsistent, in conflict with, or in excess of the terms of this warranty shall not be binding upon the Company unless placed in writing and approved by an officer of the Company.
Reagent and chemical warranty
CANNON® Instrument Company (“the Company”) warrants all re­agents and chemicals sold by the Company and delivered to and retained by their original purchasers to conform to the weight, specifications and standards stated on the package. The Company will, at its sole option, either replace or refund the price (net of freight, handling charges and taxes), of any reagent or chemical sold by the Company which does not conform to such weight, specifications and standards upon the prompt return of the unused portion. Except for replacement or refund of the net price, the Company shall not be liable for any damages occurring as a consequence of the failure of any reagent or chemical sold by the Com­pany to conform to the weight, specifications and standards stated on the package.
28
Returning a product to
Procedure
Before returning a CANNON® product for repair or service, make every attempt to identify the problem. If, after careful checking, the problem remains unidentified or unsolved, telephone CANNON® Instrument Company (or the local service agent) to consult with a product specialist. If the specialist cannot recommend a simple solution or repair, CAN- NON® will authorize the return of the product through the issuance of a Return Authorization number (RA).
Products returned to CANNON to the following address:
Please include the following:
CANNON
CANNON
®
®
Telephone Number 814-353-8000
CANNON® Fax Number 814-353-8007
®
must be carefully packed. Ship prepaid
CANNON Instrument Company A TTN: Return Authorization # __________ 2139 High Tech Road State College, P A 16803 USA
Required information
Hazardous materials
Shipping notification
• The Return Authorization number (RA).
• The name and telephone number of the person at your company to contact regarding the product.
• Shipping and billing instructions for the return of the product to your location.
• A detailed explanation of the reason for the return.
If the product is not covered by warranty, the customer will be provided with an estimate of the repair costs and asked for approval before any repairs are made. The customer will be required to issue a purchase order for the cost of the repairs.
Stringent government regulations restrict the shipment of mercury. Please contact CANNON® before returning a product that could possibly contain mercury.
Products returned without prior notification (by either telephone or fax), or without Cannon’s authorization, will not be accepted.
The customer may be billed a testing fee if a product is returned to
CANNON® and found to be working properly.
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011; CANNON® Instrument Company
2139 High Tech Road • State College, PA • USA 16803
Loading...