2139 High Tech Road • State College, PA • 16803 • USA
CMRV-4500
Instruction & Operation Manual
®
Instrument Company
vi
This page intentionally left blank.
CANNON
®
Mini-Rotary Viscometer
Version 1.1e—October, 2011 • CANNON
CMRV-4500
Instruction & Operation Manual
®
Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
CHAPTER
1
1
Purpose of the manual
Instrument utility
NOTE
INTRODUCTION
This manual has been written to provide the information necessary for
proper installation, operation, and maintenance of the CANNON
Rotary Viscometer (CMRV-4500).
®
The CANNON
viscosity and yield stress of engine oils and drive line lubricants within
the temperature range of -10
4684, D 3829, D6821 and D 6896. ASTM test method D 4684 is required
by the Society of Automotive Engineers (SAE) Engine Oil Viscosity
Classification SAE J300.
The CMRV-4500 capabilities have been tailored to the requirements of
the current ASTM methods cited above. For this reason, the instrument
may not be suitable for some general-purpose viscometry applications.
Consult with CANNON® Customer Service before testing with materials
and/or methodology at significant variance with ASTM D 4684, ASTM D
3829 or ASTM D 6821.
Mini-Rotary Viscometer is used to measure the apparent
°
C to -40°C using ASTM test methods D
®
Mini-
Improvements
CMRV-4500 with pulley-wheel assembly
The CMRV-4500 is a state-of-the-art system offering many superior
features including:
New Method compatibility
The CMRV-4500 is fully compatible with ASTM D 6821,
the new Drive Line Lubricants test method. The CMRV-4500
is fully compatible with ASTM D 6896, a new method for
testing used diesel oils.
Improved insulation for temperature control
The CMRV-4500 instrument features a redesigned housing
and accessory rotor caps to enhance insulating characteristics
and precision.
Software library of temperature profiles
(cooling profiles)
Predefined cooling profiles that comply with methods ASTM
D 3829, ASTM D 4684, ASTM D 6821 and ASTM D 6896
are supplied with the CMRV software. The user may also
create custom cooling profiles with the Profile Designer.
VISCPRO
®
The VISCPRO
Profile Designer for creating unique cooling profiles.
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
2
Networking capability for multiple instruments
®
The VISCPRO
software can control/monitor up to four CMRV instruments with one computer via RS-485 serial connections. See APPENDIX
D for more information.
Instrument overview
The CANNON® Mini-Rotary Viscometer is designed for precision
control of temperature over time, enabling accurate yield stress and
viscosity measurement of oil samples in conformance with ASTM D
3829, ASTM D 4684, ASTM D 6821 and ASTM D 6896 test methods.
Test procedure
Oil samples placed in the CMRV-4500 viscometric cells are heated and
cooled at a predetermined rate according to a user-specified cooling
profile. The cooling profile parameters are downloaded to the CMRV4500 onboard memory via a serial interface with the host computer. The
CMRV-4500 then uses this profile to control the rate of temperature
change, independent of the host computer.
All three methods specify that the samples be initially heated to ensure
that all components of the sample are released into solution. The samples
are then slowly cooled to the test temperature using the user-selected
temperature profile and maintained at test temperature for a specified
soak period. Following the soak period, apparent viscosity (and yield
stress if applicable) for each sample is determined by applying a constant
torque to, and measuring the rotational speed of, a cylindrical rotor which
has been immersed in the sample.
The time required for completion of the ASTM D 3829 test cycle is
approximately 19 hours. The time required for completion of the ASTM
D 4684 (TP-1, two-day) test is 45½ to 53½ hours. The time required for
completion of the ASTM D 6821 test cycle is approximately 18 hours.
The time required for completion of the ASTM D 6986 test cycle is
approximately 43½ to 53½ hours.
NOTE
CANNON
See Chapter 10 for additional information on the ASTM procedures.
Please observe the following safety procedures and notices for proper
operation of your CMRV-4500 unit:
•Make sure that your unit is operated only by qualified personnel
•Make sure that you read and understand all operating instructions
and safety precautions listed in this manual before installing or
operating your unit. If you have questions regarding instrument
operation or documentation, contact CANNON
®
Instrument Company.
•Deviation from the installation, operation or maintenance procedures
described in this manual may result in a hazardous situation and may
void the manufacturer's warranty .
•Transport the unit with care. Sudden jolts or drops may cause dam-
age to components.
•Observe all warning labels.
•Never remove warning labels.
•Never operate damaged or leaking equipment.
•Always turn off the unit and disconnect the mains cable from the
power source before performing service or maintenance procedures,
or before moving the unit.
•Always remove sample from the cells before moving the unit.
•Never operate the equipment with damaged mains power cables.
2139 High Tech Road • State College, PA • 16803 • USA
4
•Refer all service and repairs to qualified personnel.
In addition to the warnings previously listed, additional cautions are
posted throughout the manual. These warnings may be designated by an
appropriate symbol inside an equilateral triangle. General cautions are
General Caution
indicated with an exclamation point (see diagram, left). Read and follow
these important instructions. Failure to observe these instructions can
result in permanent damage to the unit, significant property damage, and
personal injury.
Hot surface cautions (see diagram, left) may be attached on or near hot
surfaces of the CMRV-4500. Avoid touching these surfaces when running
Hot Surface Caution
profiles at temperatures above 50°C.
The Protective Conductor Terminal symbol is used to indicate required
ground connections for your instrument electrical supply .
Protective Conductor
WARNING
AC Power Input Symbol
( O )
Supply OFF Symbol
When supplying power to this instrument, connect the protective ground
(earth) terminals of the instrument to the protective conductor of the
(supplied) line (MAINS) power cord. The main plug for the power cord
should only be inserted in a socket outlet (receptacle) provided with a
protective ground (earth) contact.
Do not use an extension cord (power
cable) without a protective conductor (grounding).
The ~MAINS symbol indicates instructions or connections for the AC
power supply . The AC Power input must match the electrical specifications listed on the label on the rear panel of the instrument. The supplied
AC Mains power cord must be attached to the connector labelled
~MAINS. This connection serves as a means of disconnect and should be
readily accessible.
The (O) symbol indicates the OFF position for the electrical switches for
your unit (AC Mains or accessories).
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
CHAPTER
5
2
UNPACKING AND ASSEMBLY
Unpacking the CMRV-4500
CAUTION
Some CMRV components, including the Air/Water Heat Exchanger, are
quite heavy. To avoid injury, obtain necessary assistance when lifting
and moving shipping cartons and heavier unpacked components.
Remove all components from the shipping container(s).
Remove any and all packing materials (styrofoam, etc.) from the
components.
V erify reception of shipped materials by comparing equipment items
with packing/parts list(s). Report missing items to CANNON
ment Company immediately.
®
Instru-
Damaged items
System components
Inspect each component for signs for damage. Report damages to the
shipper and to the CANNON
Retain all packing materials until the instrument is connected and functioning properly. If any component(s) must be returned to
CANNON
aged in the original shipping container. Refer to Chapter 19 of this
manual for instructions on returning defective equipment. Customers
outside the United States should contact the local CANNON® agent for
procedures on returning products to CANNON® .
Before beginning assembly , please verify that all components listed on
the packing slip are present, including:
Instrument Company , the damaged item(s) should be repack-
®
Instrument Company immediately .
Main power cord
Pulley-wheel assembly
9 Matched rotors
9 Rings and threads
RS-232 interface cable
Small uncalibrated thermometer (0°C to +105°C)
Large calibrated ASTM thermometer (-46°C to +30°C)
3k-ohm temperature probe
CD-ROM with VISCPRO
The user must supply an electrical power source matching the electrical
requirements indicated on the rear panel of the CMRV-4500 model.
Computer
Cleaning supplies
®
An IBM-compatible computer with the W indows
XP® operating system
(see computer specifications sheet included with your instrument) and
printer are also required.
The following items are required for regular cleaning of the viscometer
cells between tests:
oil solvent
acetone
suitable solvent-resistant container for placing/cleaning rotors
vacuum source with trap for solvent and oil
flexible tube, about 150 mm long and 3-5 mm in diameter, connected
to the vacuum trap
two plastic squeeze bottles, each with an extension long enough to
direct oil solvent and acetone directly into the viscometer cells
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
CHAPTER
7
3
CMRV-4500 APPARATUS
CMRV-4500 unit/accessories
The CMRV-4500 unit contains an aluminum block with
a heater for warming the
block. Nine viscometric cells
are closely fitted into nine
holes in the block (see photo).
There are also two thermometer wells in the block.
Lower rotor bearing
Each viscometric cell
consists of an aluminum
block aperture with a
rotor resting in a stainless
steel cup at the base. The
rotor is attached to a rotor
shaft with a pivot point at
the bottom. The pivot
point fits a mating
conical depression at the
bottom of the stainless
steel cup; this mechanism
serves as the lower
bearing. The standard rotor composition is hardened stainless steel. The
drive line rotor composition is Delrin
(NOTE: The drive line rotors/pins/
weights must be purchased separately
from CANNON®.)
DESCRIPTION
CMRV-4 unit with rotors inserted
and thread wound on pulley-wheel
Rotor for oil testing (top) and
drive line lubricant testing
®
Upper rotor bearing
CANNON
The upper bearing consists of a brass
insert at the top of the rotor shaft with
a 1.2 mm hole on the shaft axis. A
cylindrical rotor pin is inserted
through the upper bearing plate about
one or two millimeters into the hole
on the shaft axis.
Upper bearing assembly with
rotor and rotor locking pin in
place
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
8
Rotor crossbar
Rotor locking pins
Pulley-wheel assembly
The rotor crossbar is
used to hold the loop
at the end of the
thread. It also serves
as an indicator for
(optional) manual
timing of rotor
rotation.
The rotor locking pins
String on rotor and pulley-wheel assembly
are used to prevent
unwanted rotor rotation. When the locking pin is lowered over the rotor
crossbar (see photo), rotation is prevented. When the pins are in the
raised (detent) position, the rotors are free to rotate.
The pulley-wheel
assembly consists
of a V-channel
timing wheel with
a digital-optical
sensor permitting
precise measurement of the wheel
rotation. The
pulley-wheel base
is designed to be
fitted to the
CMR V-4500 slide
track adjacent to
the viscometric
cells. The probe
Slide track with pulley-wheel assembly
cable is connected to the jack on the front of the CMRV-4500 housing.
Winding thread
Slide track
CANNON
The thread used for CMRV-4500 testing is a single, nonelastic strand of
70 cm (28") winding thread of silk, cotton, or similar material (Coats
North America or comparable brand, 0.1 mm radius) with a loop on one
end. One end is wound around the CMRV-4500 rotor. The other end of
the thread is tied to a small rigid plastic ring from which the test weight
may be suspended.
The pulley-wheel assembly is designed to interlock with the pulleywheel slide track and move laterally so it can be aligned with the scored
marks on the track opposite each of the nine rotor shafts.
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
9
Hook-weights
Temperature probe
Thermometers
Hook-weights are hung on the plastic ring at
the free end of the thread to apply the required
force to the rotor during yield stress and
viscosity measurements. Nine disk weights,
one hook-cage, and one larger hook-weight are
provided. Weights vary depending on the
ASTM Method used for testing.
CMRV-4500 block temperature is detected by
a 3k-ohm temperature probe, which must be
seated securely in the thermistor aperture at the
rear of the CMRV housing and plugged into
the electronic chassis.
Hook-weights
The CMRV-4500 unit is shipped from CAN-
®
with two thermometers to be used to check the temperature of the
NON
block.
The long thermometer used for the probe and cell calibrations is a
PRINCO mercury-thallium calibrated thermometer with a range of
-46°C to +30°C in .2°C increments and an accuracy of ± 0.01°C
below 20°C.
NOTE
If a thermometer is removed from the CMRV-4500 when the unit is
cooling, plug the thermometer aperture in the Plexiglas® cover to prevent
ice formation around the rotors.
The shorter, high-temperature uncalibrated thermometer has a range
of 0°C to 105°C in 1°C increments with an accuracy of ± 1°C.
Air/Water Heat Exchanger
CMRV-4500 cooling is
accomplished with internal
thermoelectric cells and an
external Air/W ater Heat
Exchanger circulating a
mixture of water and
antifreeze to remove heat
from the thermoelectric
modules.
The Air/Water Heat Exchanger contains the electronics necessary to monitor
the coolant flow rate and regulate the action of the circulating pump
during the running of ASTM-specified temperature profiles. The Heat
Exchanger also supplies DC power for the operation of the CMRV
thermoelectric modules.
Air/Water Heat Exchanger
(see chapter 4 for Setup diagram)
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
10
Front panel operations
CMRV front panel
Control features
CMRV front panel control features are simple and functional. The central
switch on the front panel is the power switch for the unit. The lightemitting diodes (LEDs) above the switch indicate function of the heating
and cooling systems. The Sensor connection on the right side of the
panel mates with the cable from the pulleywheel optical sensor. The StartTest button on the left side of the panel is used to initiate CMRVcontrolled testing and calibration routines. The Yield and Viscosity
LEDs light during yield stress or viscosity testing.
For additional details on front panel operation, refer to the calibration and
testing chapters of this manual.
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
CHAPTER
11
4
CAUTION
Physical placement
CAUTION
Control unit placement
SETUP
The instructions in this chapter are for setting up a single CMRV-4500
with the Air/Water Heat Exchanger. For additional information on multiunit configurations, see APPENDIX D.
Do not apply power to the CMRV-4500 or Heat Exchanger until all
connections are secure.
Some CMRV components, including the Air/Water Heat Exchanger, are
quite heavy. To avoid injury, obtain necessary assistance when lifting
and moving these components.
Place the CMRV-4500 on a stable laboratory bench or table top within 10
mm of the front edge. This will allow the weight suspended from the
pulley-wheel assembly to clear the edge of the table during viscosity and
yield stress tests. Allow 30.5 cm (12") of clearance to the back and sides
of the unit.
A/W Heat Exchanger
CAUTION
Thermometer placement
NOTE
Place the Air/Water Heat Exchanger below the table top on which the
CMRV-4500 is placed. The unit may be located within the parameters
permitted by the length of the hoses and cables shipped with the unit.
Do not obstruct the front or rear cooling vents on the Air/Water Heat
Exchanger.
When in use, the large thermometer (-46°C to +30°C) is placed in the
thermometer well on the left side of the CMR V-4500 unit (seen from the
front).
When in use, the small thermometer (0°C to 105°C) is placed in the thermometer well on the right side of the CMR V-4500 unit (as seen from the
front).
If the mercury column in a thermometer has separated during shipment,
read the instructions included in APPENDIX C. These instructions offer
several methods for joining separated mercury columns.
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
12
Tubing connections
There are two tubing connections on the rear of the CMRV-4500 housing
(see diagram, next page). The fitting on the left is the coolant INLET (for
introducing coolant to the CMRV-4500) and the fitting on the right of the
rear panel is the coolant OUTLET (for returning coolant to the Ex
changer). Clamp the open ends of both hoses to these fittings, then secure
the hoses to the appropriate bulkhead fittings on the Air/Water Heat
Exchanger using the bulkhead inserts provided with the tubing (see
Figure). The fittings will snap into place. To release and remove the
tubing from the Air/Water Heat Exchanger, push down on the button on
the bulkhead fittings to release the connector locking clips.
CMRV-4500 tubing (solid lines) and electrical (dotted lines) controller/AWHE connections
Filling the Air/Water Heat Exchanger
Obtain a supply of quality automotive antifreeze (ethylene glycol) and
mix it with water in a ratio of 30 percent antifreeze to 70 percent water.
Do not mix antifreeze types in the Air-Water Heat Exchanger.
After you have secured the Exchanger tubing connections (see previous
section) pour this antifreeze/water mixture into the reservoir opening on the
top of the Air/Water Heat Exchanger until it is full (approximately 4 liters).
CAUTION
NOTE
CANNON
Ethylene glycol is a toxic substance. Use proper safety precautions when
handling. Follow appropriate MSDS instructions.
Depending on the amount of fluid displacement in the coolant lines, it
may be necessary to add additional antifreeze/water mixture to the
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
Exchanger when the unit commences operation. Y ou will be able to
observe the liquid circulating inside the reservoir from the opening at the
top of the Exchanger during normal operation. If air bubbles are consistently visible in the coolant lines, add additional mixture until the Exchanger is full.
13
CAUTION
The water-antifreeze mix should be replaced annually for reliable performance and to prevent corrosion of internal components (see Flushingand draining the Air/Water Heat Exchanger, this chapter).
Electrical /serial connections
A/W Heat Exchanger
Plug the male end of the DC power cable into the DC POWER OUT
fitting on the rear panel of the Air/Water Heat Exchanger and turn the
locking clip clockwise to secure the connection. Attach the other end of
the cable to the receptacle labelled DC POWER at the back of the
CMRV-4500 unit (see figure, previous page). This connection provides
power to the thermoelectric cooling modules.
Insert the power cable for the Air/Water Heat Exchanger into the receptacle on the Exchanger rear panel. Secure the power cable in place by
tightening the Phillips screw on the cable clamp.
Make sure the Air/Water Heat Exchanger power switch is in the OFF
position; then insert the power line cord from the Exchanger into an
appropriate power source for your unit.
Before providing mains power to the unit, check the label on the rear
panel of the Exchanger to verify that the electrical specifications for the
unit match those of the power supply.
Thermistor
Figure 11: CMRV rear panel
Attach the M-F three-prong connector cable from the Exchanger PUMP
outlet to the PUMP receptacle on the CMRV-4500 controller (see figures on
pages 12-13). This power connection permits CMRV-4500 control of pump
operation.
Insert the Lemo
TEMP PROBE on rear of the CMRV controller (see Figure 11). Insert
the probe tip as far as it will go into the hole marked THERMISTOR on
the rear of the CMRV-4500 chassis.
2139 High Tech Road • State College, PA • 16803 • USA
®
plug from the temperature probe into the jack marked
®
Instrument Company
14
Heater
CMRV-4500 power cord
Serial connections
NOTES
Insert the heater plug from the upper section of the CMRV-4500 housing
into the HEATER receptacle on the rear panel of the CMRV controller.
Make sure the CMRV-4500 power switch is in the OFF position. Then
insert the power line cord from the rear panel of the CMRV controller
into an appropriate power source for your unit
Before providing mains power to the unit, check the label on the rear
panel of the Exchanger to verify that the electrical specifications for the
unit match those of the power supply. Use only the supplied, approved
appliance cords for the CMRV.
T o connect a single CMRV -4500 instrument to the host computer,
connect the computer cable to the RS-232, DB-25-pin socket at the rear
of the CMRV-4500 controller and secure the cable connection with the
two small screws on the ears of the plug. Attach the other end of the cable
to the RS-232 port at the rear of your computer.
COM 2 and COM 4 use the same IRQ settings on most computers,
meaning that they cannot be used simultaneously. The COM 1 and
COM 3 ports have the same problem. Do not try to use a device on COM
4 if you are using COM 2 for the CMRV instrument.
Some display adaptors (in particular, S3, 8514A and ATI mach 8) have
an address conflict with COM 4 ports. If this is the case, you may need to
use another COM port or replace your current display adaptor.
RS-485 serial connections
T o install multiple CMRV units using RS-485 serial cable connections,
see the multi-unit configuration instructions in APPENDIX D.
Pulley-wheel installation
T o install the pulley-wheel
assembly atop the CMRV:
1. Remove the left screw
from the front of the
upper panel (see photo).
This will provide
clearance to install the
assembly on the slide
track.
CAUTION
Use care in handling the pulley-wheel assembly to avoid damage to the
wheel or the movement sensor.
Removing screw to install pulley wheel
CANNON
2. Install the pulley-wheel assembly by sliding it onto the left end of the
slide track with the pulley-wheel facing out (see photo).
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
15
CAUTION
NOTE
Installation notes
Download button
Make sure that the
assembly is seated
securely on the
track.
3. Replace the
left front
screw . This
Pulley-wheel assembly mounted on slide track
will prevent
the assembly from sliding off the track.
4. Plug the free end of the pulley-wheel sensor wire into the jack
labelled WHEEL on the front of the CMRV-4500 controller.
To disconnect the sensor, pull it out by the knurled portion of the plug.
The CMRV-4500 is capable of receiving firmware updates from the
computer via the DOWNLOAD button on the rear of the electronic
®
chassis. Should such an update be required, CANNON
will provide
detailed instructions regarding the download procedure for updating the
firmware.
NOTE
If the DOWNLOAD button is pressed inadvertently, switch off the CMRV4500 power for at least four seconds, then
restore power to the unit.
Setting the CMRV instrument address
When installing/connecting a new CMR V
instrument, you must set the instrument
address using the SELECT ADDRESS
dial on the rear of the CMRV controller. This dial offers 16 settings (0-9,
A-F).
Procedure
CAUTION
T o set the address, rotate the dial (see photo) to a setting not currently in
use by other CMRV instruments.
The MRVW software will not function correctly unless each networked
CMRV instrument is set to a different address (see APPENDIX D for
multi-unit configuration).
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
16
Cleaning the instrument housing
CAUTION
Before cleaning the CMRV housing, turn off the instrument and unplug
the power cord. Do not clean the instrument unless the cell temperature
is within 10°C of ambient.
Periodically clean the outside of the unit with a damp cloth moistened
with water and/or a mild detergent solution.
Flushing and draining the Air/Water Heat Exchanger
The water-antifreeze mixture should be replaced annually for reliable
performance and to prevent corrosion of internal components. Do not
mix antifreeze types in the Air-Water Heat Exchanger.
T o flush fluid from the Exchanger and CMRV -4500, obtain replacement
antifreeze, a funnel, a drain bucket and a supply of warm water. Then
complete the following procedures:
Flushing fluid from the Exchanger
1. Turn off the Heat Exchanger power switch
to remove power from the Exchanger.
2. Detach the quick-connect fitting securing the
tubing from the CMRV-4500 to the IN
connection on the Exchanger (see photo) by
pushing down on the release button while
pulling the hose connection out. Place the
tube end in the drain bucket and pull the
hose off of the ribbed section of the fitting.
Fluid will begin draining from the CMRV4500 and Exchanger.
Removing the Exchanger tubing connection
4. Turn on the Exchanger power switch to engage the Exchanger
circulating pump. Fluid from the Exchanger and CMRV-4500 will
flow into the drain bucket.
5. Immediately begin adding warm water to the Exchanger reservoir
through the funnel and continue as it is pumped out into the drain
bucket until the water exiting the drain hose into the drain bucket
flows clear.
6. Stop adding water to the reservoir and immediately turn off the Heat
Exchanger power switch.
CAUTION
CANNON
Do NOT permit the Exchanger pump to operate without fluid in the
system. Doing so will damage the pump.
2139 High Tech Road • State College, PA • 16803 • USA
3. Open the lid of the Exchanger reservoir and
place the funnel in the opening.
®
Instrument Company
7. Replace the connector on the Exchanger hose, and reattach the
connector to the fitting on the Exchanger (see photo).
Draining fluid from the Exchanger
After the antifreeze mixture has been flushed from the Air/Water Heat
Exchanger per the preceding procedure, drain the system as follows:
1. Make certain that the Heat Exchanger power is OFF, then remove the
external housing from the Air/Water Heat Exchanger by removing
the six screws securing the housing to the sides of the frame (three
screws on each side of the unit) and the six screws on the top and
sides of the rear panel.
2. Locate the twin stopcocks at the top and bottom of the radiator (see
photo, next page). Attach a drain hose to the bottom stopcock and
place the opposite end of the hose in a drain bucket.
3. Open the top and bottom stopcocks by turning them parallel to the
nozzle and permit water to drain from the radiator. Then close the
stopcocks and remove the drain hose from the bottom stopcock.
17
NOTE
A small amount of water/antifreeze mixture may remain in the tubing
leading from the external connector to the reservoir. If desired, this fluid
may be drained by pulling the hose off of the ribbed section of the quickconnect fitting previously installed. Reattach the tubing before completing
the remainder of this procedure.
CANNON
Heat Exchanger with housing removed and stopcocks highlighted
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
18
4. Replace the external housing on the Air/Water Heat Exchanger and
secure it with the six screws previously removed. If necessary, reseat
the gasket around the reservoir opening with a small screwdriver.
5. Add water-antifreeze mixture to the Exchanger per the earlier
instructions in this chapter to complete the water-antifreeze flush and
drain procedure.
FHP MOTOR
NOTES
For rountine maintenance and improved bearing life expectancy , a few
drops of non detergent “twenty weight” oil can be added every 12 months.
Fans should be cleared of dust/other material on a regular basis.
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
CHAPTER
19
5
VISCPRO® for Windows® XP
VISCPRO® is a powerful new software product providing a generic
instrument interface for controlling and operating your CANNON
instrument(s) via computer. VISCPRO® also includes reporting/analysis
modules for processing and displaying sample data.
VISCPRO® INSTALLATION
®
Installing VISCPRO® software
T o install the VISCPRO® software, follow the instructions below in the
sequence presented. Make certain that you complete the sections on
checking instrument settings and calibration data. If you encounter
difficulties at any stage in the installation process, call CANNON
service at 814-353-8000.
Computer requirements
IBM-compatible computer with pentium processor and Windows® XP
installed and operating normally • SVGA compatible display •
32 MB RAM • 200 MB hard disk space • CD-ROM drive • serial port
®
®
®
Windows® XP
®
installation
Installation actions
1. Turn on your computer. Wait for the Windows® software to load.
2. From the Windows
the first VISCPRO® installation CD-ROM into the disk drive.
3. Double-click the Add/Remove Programs icon and follow the
Windows prompts to complete the installation procedure. The
executable file for VISCPRO
The installation program will:
®
Start Bar click Settings/Control Panel. Insert
®
software installation is SETUP.EXE.
create a directory for your data files. The default directory is
C:\Program Files\Cannon Instrument\VISCPRO).
write SETUP information to the Windows
®
registry.
copy the software executable file and other necessary files to the
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
20
Running the software
Loading software
Make certain that your CMRV instrument is properly connected to your
computer and the CMRV power switch is ON. Then start the VISCPRO
software by double-clicking the VISCPRO® icon on your W indows
desktop. Or click Start/Programs/VISCPRO/VISCPRO.EXE).
Right now , your computer monitor should look like this:
®
®
The VISCPRO® primary display
®
The VISCPRO
primary display window is framed on the top by the
VISCPRO® title bar and menu bar, and on the bottom by the VISCPRO
status bar.
Displaying the Instrument View window
The application window can be configured to display child windows, such as
the Instrument View window, which describes your CANNON® instrument and provides controls for running tests:
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
®
21
The Instrument View window
NOTE
Instrument ID/T ype
Instrument Configuration
To display the Instrument View window, click View Instrument from the
Main menu. The View Instrument window will appear. Then click the
MRV instrument ID to display the list of available configurations. For
now, select the default configuration and click OK.
If the Available Instruments list box is blank, your instrument(s) may not
be on-line. Check cable connections and make certain the instrument
power switch is ON.
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
22
Viewing report data
After you have completed CMRV calibration and testing (see next
chapters), you will be able to display a report window with CMRV-4500
test results. To access the database and display test data, click Analyses/View Analysis. Then select MRV Data Table from the list of available
analyses:
Report T ype
Report Configuration
The Choose Analysis window
Choose the desired report configuration from the list of available configurations and then click OK to open the Analysis Configuration
window. Then select report configuration options (see Chapter 13 for
more information) and click OK to generate the analysis from existing
sample data. If the report window is blank, you may have to change the
Date Filter options to include the desired range of samples from the
database.
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
23
Checking Configuration data
Configuration protection
Logging in
Follow the procedures in the next several sections of this chapter to select
and verify the instrument and calibration settings to ensure that they
®
conform to the actual characteristics of your CANNON
instrument.
T o check the configuration settings for your instrument(s), you must log
in to the security system as a manager. The software is installed with a
default Manager account. This account has no password, allowing any
operator access to manager-level software functions as long as the
password is not activated/changed. If you would like to engage the fullrelease security options, see Security Options, this chapter, for instructions.
1. Use your mouse to click Main from the VISCPRO® menu bar.
2. Click Log In from the Main menu options.
3. Click on the
(arrow) on the right side of the User Name: list box
to display the list of registered users.
4. Click Manager. Do NOT enter a password!
5. Click OK. The Log In window will close automatically and you will
be logged in as management personnel.
Checking Instrument Settings
1. Use your mouse to click (select) Configure from the VISCPRO
menu bar.
2. Select your instrument from the list of available instruments (there
may be only one instrument in the list).
3. Select Instrument Settings from the list of configuration options.
The Instrument Settings window will appear.
You will use the Instrument Settings window (see graphic following) to
describe and control MRV instrument operational features. These settings
affect the instrument as a whole. Check the instrument settings for your
instrument per the instructions, and make any necessary changes:
2139 High Tech Road • State College, PA • 16803 • USA
®
Instrument Company
®
24
The Instrument Settings window
4. Use the ID field to input instrument identification information using
up to 30 alphanumeric characters.
Checking initial calibration
The remaining fields in the Instrument Settings window are noneditable (information is obtained via serial communication with the
CMRV instrument). The Model: field will indicate the model of your
instrument. The S/N: field indicates the serial number from the label
on the rear service panel. The Firmware: field indicates the version
for the current instrument firmware. The Port: field indicates the
current communications port for the RS-232 cable connecting your
computer to the CMRV-4500 instrument.
5. When you have entered all settings, click OK.
After you have completed the CMRV instrument calibration for temperature and cell constants (see following chapters), follow the procedure
below to verify the initial calibration settings.
1. From the VISCPRO primary window, click Configure/MRV/View
Cell Constants. The View Cell Constants window will open.
CANNON
2. Ensure that Calibration data, including constants and viscosity, is
available for each cell. Then click OK to close the View Cell Con-stants window.