Cannon CCS-2100, CCS-2100LT Instruction & Operation Manual

CCS-2100/
2100LT
Automatic
Cold-Cranking
Simulator
Instruction
& Operation
Manual
CONTENTS
1
INTRODUCTION/INSTALLATION 1
Overview ............................................................................................................................ 1
Equipment........................................................................................................................... 2
Required accessories .......................................................................................................... 3
Utilities ............................................................................................................................... 4
Safety cautions.................................................................................................................... 4
Unpacking........................................................................................................................... 5
Positioning CCS components ............................................................................................. 6
Computer connections ........................................................................................................ 6
Controller power distribution ............................................................................................. 7
Optional connections for non-thermoelectric operation......................................... 8
Simulator Mechanism/Controller cable connections.......................................................... 9
Tubing connections............................................................................................................. 9
Waste system connections ...................................................................................... 9
Coolant connections ............................................................................................. 10
Filling the refrigerated chiller........................................................................................... 11
Installing VISCPRO® software......................................................................................... 12
Windows® installation .......................................................................................... 12
Installation actions................................................................................................ 12
Specifications ................................................................................................................... 13
2
Configuring the CCS 15
Checking/editing configuration settings in VISCPRO..................................................... 15
Running the software............................................................................................ 15
Logging in ............................................................................................................ 15
Checking/editing Instrument Settings .................................................................. 16
Viewing/editing other setup information .............................................................. 17
VISCPRO software interface elements ............................................................................ 18
Initial Stepper Motor Service (CCS-2100 only)............................................................... 20
Setting the CCS Set Motor Current .................................................................................. 21
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
ii
3
4
CALIBRATING THE CCS 23
Calibration procedure ....................................................................................................... 23
Preparing/updating the list of calibration standards ............................................. 23
Testing calibration standards ................................................................................ 25
Performing the calibration .................................................................................... 27
Checking calibration accuracy ......................................................................................... 28
Viewing current calibration constants .................................................................. 30
Saving a calibration .............................................................................................. 30
TESTING SAMPLES WITH THE CCS 31
Preparing the CCS and samples ....................................................................................... 31
Test options....................................................................................................................... 35
Running a test ....................................................................................................... 35
Pausing a test ........................................................................................................ 35
Resuming a test..................................................................................................... 35
Aborting a test ...................................................................................................... 35
Concluding a test .................................................................................................. 35
Viewing results .................................................................................................................36
Creating an analysis.......................................................................................................... 36
5
USING THE CCS SOFTWARE 39
VISCPRO® generic instrument interface.......................................................................... 39
Main options......................................................................................................... 39
Security options .................................................................................................... 42
Initial security setup ............................................................................................. 43
Print/Print setup options ....................................................................................... 44
Analyses options............................................................................................................... 44
Analysis types....................................................................................................... 45
Analyses menu options......................................................................................... 45
Window options................................................................................................................ 46
CCS module menu options ............................................................................................... 47
Configure options ................................................................................................. 47
Print Instrument and Tray (Test) Settings............................................................. 48
Instrument Settings ............................................................................................... 48
Probe constants..................................................................................................... 48
Tray (Test) Settings............................................................................................... 49
Saving a configuration.......................................................................................... 50
Restoring instrument settings from a saved configuration ................................... 51
Calibration ............................................................................................................ 51
CCS Calibration Standards................................................................................... 53
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
CCS Choose Calibration Standards...................................................................... 54
Current Cell Calibration Standards ...................................................................... 55
CCS Setup ............................................................................................................ 56
Service menu options ....................................................................................................... 56
Testing samples—software options .................................................................................. 57
Entering sample ID information ........................................................................... 57
Selecting sample actions....................................................................................... 58
Copy & Paste Sample ID data entry options ........................................................ 61
Inserting/deleting a sample ID in the test sequence ............................................. 61
Analysis modules.................................................................................................. 62
Status bar CCS error tracking........................................................................................... 62
iii
6
7
MAINTAINING AND SERVICING THE CCS 63
CCS components .............................................................................................................. 63
CCS Unit .............................................................................................................. 63
Waste system ........................................................................................................ 64
Emptying the waste container .......................................................................................... 64
Repairs/diagnostics........................................................................................................... 64
Rotor/stator assembly ........................................................................................... 65
Checking/reseating the thermistor probe.............................................................. 65
Replacing the vacuum pump diaphragm .............................................................. 66
Gapping the rotor in the stator well...................................................................... 67
Preventive maintenance................................................................................................... 68
Biannual maintenance .......................................................................................... 68
Cleaning the instrument housing ...................................................................................... 68
ANALYSIS CONFIGURATION OPTIONS 69
Creating an analysis.......................................................................................................... 69
Sorting analysis data......................................................................................................... 71
Using the date filter .......................................................................................................... 71
Using the sample/error filter............................................................................................. 72
Using the report/port output filter..................................................................................... 73
Reconfiguring a displayed analysis .................................................................................. 73
Resizing table columns..................................................................................................... 74
Saving a current analysis .................................................................................................. 74
Deleting an analysis configuration ................................................................................... 75
Printing an analysis........................................................................................................... 75
Keystrokes for selecting data for printing ............................................................ 75
Exporting analysis data..................................................................................................... 76
8
CCS DATA TABLE 77
Configuring the CCS Data Table.......................................................................... 78
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
iv
9
10
11
12
ERROR LOG TABLE ANALYSIS 81
Configuring the Error Log analysis ...................................................................... 81
EXPORT ANALYSES 85
Configuring the Port Export analyses................................................................... 86
USING THE DATABASE MANAGER 91
Archiving old data ................................................................................................ 91
Changing the database directory .......................................................................... 92
Importing archived data........................................................................................ 92
Repairing/compacting the database...................................................................... 93
Exit ....................................................................................................................... 93
REPLACEMENT PARTS LIST 95
13
A
B
I
WARRANTY/RETURN INFORMATION 97
Products limited warranty................................................................................................. 97
Reagent and chemical warranty........................................................................................ 97
Returning a product to CANNON®.................................................................................. 98
APPENDIX AWEISSENBERG EFFECT 99
APPENDIX BSAE VISCOSITY CLASSIFICATIONS
FROM -10°C to -35°C101
INDEX 103
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
CHAPTER
1
Overview
1
INTRODUCTION/INSTALLATION
Manual
The CCS instruments
This manual is intended for use with both the CCS-2100 and the CCS­2100L T Cold-Cranking Simulator (CCS) models from CANNON Instrument Company . It provides information about:
®
Installation and operation of the CCS-2100 instrumentsStandard ASTM testing and methods related to the CCSCalibration of the CCS for the customer’s specific needsUse of the CCS computer softwareMaintenance and repair of the CCS
The CANNON® Cold-Cranking Simulator (CCS-2100) is a solid-state, thermoelectrically-cooled, fully-automatic test instrument for the determi­nation of cold-cranking viscosity of engine lubricants. The CCS-2100 operator initially loads oil samples into the Sample Tray and enters sample identification via the computer keyboard or with a simple software
procedure restores a previously saved test configuration. Without further operator intervention, the CCS-2100 tests and determines the apparent viscosity for up to 30 test samples at specified sub-ambient test temperatures, and records the results.
New features The CCS 2000 Series instruments offer several advantages over previous CCS models, including improved temperature management and elimination of the required warm-up sample at the beginning of the test cycle. The thermoelectric system also enhances cell cleaning with a warming cycle, improving measurement precision. An attractive cabinet combines the simulator mechanism and the modular CCS Series II Control­ler.
Applications
The CCS-2100 is used to determine the apparent viscosity of engine oils at low temperatures at shear rates similar to those at starting conditions of cold engines. CCS instruments are found in quality control oil testing laboratories, major oil corporation laboratories, independent test facilities, blending facilities, and automotive corporation laboratories.
CANNON
Version 2.0a— November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
2
Precision
Temperature range
Viscosity range
Calibration constants
Equipment
The CCS-2100 is designed to produce measurements that meet or exceed the accuracy and precision dictated in ASTM test method D 5293.
The CCS-2100/2100LT is capable of measuring the apparent viscosity of oils at test temperatures ranging from -5°C to -40°C in increments of 5°C. The refrigerated water chiller must be set to +5°C to facilitate operation of the thermoelectric cooling system.
The CCS-2100 is capable of calculating viscosity values ranging from 900 to 25,000 cP at test temperatures. Alternative ranges are possible. Contact CANNON® technical services for more information.
The CCS-2100 will automatically calculate the appropriate calibration constants (refer to Chapter 3 for more information). After calibration, the CCS-2100 will use those constants for future tests without the necessity of operator intervention.
The CCS instrument includes a Simulator Mechanism and the integral CCS Series II Controller. The peripheral waste system includes a vacuum pump (for evacuation of the sample) and a waste container.
Simulator Mechanism
CCS Series II Controller
Waste system
The Simulator Mechanism contains a temperature-controlled rotor/stator test mechanism and a constant metering, positive displacement piston pump which transfers the oil samples from the sample bottles to the rotor/ stator assembly via an injection tube. The Simulator Mechanism also includes a 30-position automatic X-Y Sample Table. The Sample Table can process up to 30 sequential samples. The CCS-2100 automatically accesses a new sample when the previous test is complete.
The CCS Series II Controller translates input from the CCS sensors and transfers the data to the com­puter for analysis. The CCS Series II Con­troller also provides power and relays computer commands to the Simulator Mechanism.
The waste system is designed to eliminate the need for solvent cleaning. Used sample is flushed from the
CCS rotor/stator assembly
rotor/stator with a
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
portion of the sample to be tested. The waste system includes the tubing and connections from the CCS to the waste container and vacuum pump. The pump is encased in a protective housing that includes a mounting bracket for the large polypropylene waste container. The container is capable of holding waste from approxi­mately 45 sample tests.
3
CAUTION
The liquid level in the waste container should be checked frequently and the container should be emptied prior to multi-sample tests.
Required accessories
The CCS requires a refrigerated chiller capable of circulating a distilled water mixture with 10 percent antifreeze to the CCS thermoelectric system at a controlled temperature of +5°C at a rate of at least 1-2 liters/minute (see refriger­ated chiller specifications, this chapter). CANNON® recommends and supplies a Julabo chiller for this purpose. The CCS also requires an IBM or IBM-compatible computer and printer (request current specifi­cations from CANNON) to run the CCS software and produce printouts of the data collected by the CCS.
The CCS-2100 Series II Controller
The CCS waste system
CANNON
Version 2.0a — November 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
4
Utilities
The customer must be able to provide the appropriate power and ground­ing for the CCS.
AC power input
CAUTION
Grounding
Safety cautions
Specifications of the voltage and frequency of the AC power source for the CCS are provided by the customer when placing an order. If you have any questions regarding the correct voltage for your CCS model, call
®
CANNON
Only attach the power cord from the rear of the CCS Series II Controller to an AC mains with the power requirements specified on the label on the rear of the Controller.
for assistance.
It is very important that the power source for the CCS has an electrical ground. The CCS is equipped with an IEC 320 socket which must be plugged into a grounded outlet.
Please observe the following safety procedures and notices for proper operation of the CCS:
Make sure that your unit is operated only by qualified personnel.Make sure that you read and understand all operating instructions
and safety precautions listed in this manual before installing or operating your unit. If you have questions regarding instrument operation or documentation, contact CANNON
®
Instrument Company .
Do not deviate from the installation, operation or maintenance
procedures described in this manual. Improper use of the CCS instrument may result in a hazardous situation and may void the manufacturer’s warranty.
Handle and transport the unit with care. Sudden jolts or impacts may
cause damage to components.
Observe all warning labels.Never remove warning labels.Never operate damaged or leaking equipment.Unless procedures specify otherwise, always turn off the unit and
disconnect the mains cable from the power source before performing service or maintenance procedures, or before moving the unit.
Never operate the equipment with damaged mains power cables.Refer all service and repairs to qualified personnel.
General Caution
In addition to the cautionary statements listed previously , additional cautions may be posted throughout this manual. These cautions, identi­fied by the caution symbol (see left) indicate important operational procedures. Read and follow these important instructions. Failure to observe these instructions may void warranties, compromise operator safety , and/or result in damage to the CCS unit.
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
Protective Conductor
5
The Protective Conductor T erminal symbol is used to indicate required ground connections for your instrument electrical supply .
WARNING
MAINS
~
AC Power Input Symbol
( O )
Supply OFF Symbol
Hazardous materials
When supplying power to this instrument, ensure that the protective ground (earth) terminals of the instrument are connected to the protective conductor of the (supplied) line (MAINS) power cord. Use only the manufacturer-supplied power cord, which should be inserted in a socket outlet (receptacle) which is also provided with a protective ground (earth) contact.
Do not use an extension cord (power cable) without a protective
conductor (grounding). The ~MAINS symbol indicates instructions or connections for the AC
power supply . The AC Power input must match the electrical specifica­tions listed on the label on the rear panel of the instrument. The supplied AC Mains power cord must be attached to the connector labelled ~MAINS. This connection serves as a means of disconnect and should be readily accessible.
The (O) symbol indicates the OFF position for the electrical switches for your unit (AC Mains or accessories).
Routine CCS operation may require the use and handling of hazardous chemicals and solutions. CANNON® Instrument Company strongly urges the operators and technicians working with the CCS to take proper safety precautions when working with these materials. These safety procedures can be found in the Material Safety Data Sheets which accompany the solutions.
Unpacking
NOTE
CAUTION
CANNON
Version 2.0a — November 2010; CANNON
Some CCS components are quite heavy. To avoid injury, obtain neces­sary assistance when lifting and moving shipping cartons and heavier unpacked components.
1. Carefully remove all components from their shipping cartons.
2. Consult the packing list to ensure that all items listed have been received. The CCS is usually shipped in three packing crates (includ­ing the refrigerated chiller unit, if it is ordered from CANNON
3. Notify CANNON
®
Instrument Company if any items are missing.
4. Inspect all components for damage. Report any damage to the shipping company and to CANNON
®
Instrument Company immedi­ately . Remove any visible packing materials (packing foam, etc.) from the components.
You should retain all packing materials until the CCS is connected and functioning properly. If any component is returned to CANNON® Instru-
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
®
).
6
ment Company , it should be packed in it s original shipping container . You must call CANNON® prior to returning any products (see warranty and return information at the conclusion of this manual).
Positioning CCS components
1. Place the CCS on a stable laboratory bench or table.
2. Place the computer adjacent to the CCS.
3. Use the key provided to open the hinged door protecting the Simula­tor Mechanism.
4. Remove any pieces of foam packing that may have been included to prevent shipping damage to internal moveable components.
5. Close the hinged door.
Sample Tray
6. Place the CCS-2100 Sample Tray on the Simulator Mechanism Sample Table so that the arrow on the tray is aligned with the arrow on the left guide of the Sample T able (see photo). The gear on the underside of the Sample T ray must engage the gear in the center of the base of the instrument. When the align­ment is correct, the CCS sampler mechanism at its default position (farthest left) will be centered over sample #1 on the Sample Tray.
Computer connections
Ensure that the computer specifications meet the minimum requirements for your CCS instrument. These requirements are listed on the Computer Specifications sheet provided with your Instruction & Operation Manual.
RS-232 serial connections
Locate the RS-232 (25-pin) CCS Control Cable. Connect one end to the COM port on the rear of the computer . Connect the other end of this cable to the socket on the rear of the CCS Series II Controller (note that the connectors are “D” shaped, and fit only one way).
Sample Tray (top view) correctly aligned on Sample Table (guides
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
Controller power distribution
7
Receptacle connections
2 3 4 6 7 8
1
9 10
There are five AC power receptacles and one DC receptacle on the rear of the CCS Series II Controller. The AC receptacles provide AC power matching the AC mains input for the various CCS AC devices. The Motor receptacle supplies regulated DC voltage for the rotor motor .
Power for the CCS Simulator Mechanism is supplied via the CCS Series II Controller. AC Power cords from the CCS are labelled to match their receptacles. Connect the power cords as follows:
5
15
14
1211 13
CCS Series II Controller rear panel with connections marked (see list below)
1. Mains power (to AC power source), ON/OFF switch and fuses
2. CCS sampler mechanism power connection
3. Injection pump power connection
4. Coolant valve power connection (unused)
5. CCS rotor motor power connection
6. CCS sampler control
7. Rotor speed sensor connection
8. CCS temperature probe connection
9. Thermoelectric cooling power connection to internal power
supply
10. Vacuum pump power connection
11. Thermoelectric cooling control (from stator assembly)
12. RS-485 connection (for multiple CCS instruments)
13. Address select (unused)
14. RS-232 serial connection (to computer COM port]
15. Download button (for future firmware upgrades—not used for installation)
Mains
Make certain that the instrument power switch on the Controller rear panel above the Mains power cord connection is turned off. Then attach the Mains power cord from the Mains receptacle to the power supply matching the label on the rear of the CCS unit.
CANNON
Version 2.0a — November 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
8
NOTES
Sampler
Injection pump
Cool
Vacuum pump
NOTE
Motor
Use only the CANNON®-approved supplied power cables for the CCS unit. Do not use serial or power cables exceeding 3 meters in length. Do not turn on the CCS power switch unless the chiller is attached and operating.
Connect the Sampler power cord from the Simulator Mechanism to the SAMPLER receptacle on the CCS Series II Controller.
Connect the power cord from the INJECTION pump receptacle to the Simulator Mechanism, at the receptacle labeled INJECTION.
Connect the power cord from the thermoelectric power supply to the receptacle labelled COOL on the rear panel of the CCS Series II Control- ler. This receptacle is energized whenever the controller power is ON.
Connect the power cord from the waste container vacuum pump to the VACUUM pump receptacle on the Controller.
If an in-house vacuum is to be used to remove oil from the CCS, you may omit the use of the waste container vacuum pump connection.
Connect the Simulator rotor motor power cord to the MOTOR recep- tacle.
Control
RS-232
Valve
CAUTION
Do not connect any other cable to the MOTOR receptacle. The power draw will damage the motor control circuit.
If it is not already attached, connect the control cable from the stator thermoelectric cooling assembly to the CONTROL receptacle.
Connect the RS-232 serial cable from the computer COM port to the RS­232 connector.
Optional connections for non-thermoelectric operation
The Valve connection (see photo image, page 7) is for optional non­thermoelectric cooling only—it is not used for normal thermoelectric operation. The non-thermoelectric version CCS (earlier version) required a custom-built rotor/stator assembly. If using a non-thermoelectric (older) instrument, connect the power cord from the optional external solenoid valve to the VALVE receptacle on the CCS Series II Controller. The solenoid valve regulates coolant flow from the optional non-thermoelec­tric cooling device through the CCS cooling system.
CAUTION
Valve (or pump) voltage must match the CCS mains input voltage and frequency. Use only CANNON®-approved supplied accessories in the VALVE outlet. Do not circulate methanol refrigerants in thermoelectric instruments!
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
9
NOTES
The solenoid valve may be obtained from CANNON® and must be installed to the coolant return line by the customer. (In some cases, a solenoid valve may already be included in the customer-supplied chiller.) If the refrigerated chiller has a constant-running internal pump, it may not be necessary to connect the pump to the CCS Controller.
Simulator Mechanism/Controller cable connections
Three cables from the Simulator Mechanism must be connected to the matching receptacle on the rear panel of the CCS Series II Controller:
1. Connect the grey cable from the Simulator Mechanism speed sensor to the SPEED jack on the CCS Series II Controller.
2. Connect the white temperature probe cable with the metal LEMO plug to the matching TEMP PROBE jack on the Controller.
3. Attach the seven-pin connector cable to the seven-pin SAMPLER jack on the Controller (see diagram, page 10). Connect the other end of this seven-pin cable to the corresponding jack in the lower left corner of the Simulator Mechanism rear panel.
4. Attach the four-pin connector cable from the thermoelectric cooling unit to the CONTROL jack on the Controller (see diagram).
5. Set the Instrument Address on the rear panel of the CCS instru-
Tubing connections
Waste system connections
A short 1/4" yellow plastic tube links the sample/solvent waste container with the waste container vacuum pump. This tube is taped to the vacuum pump housing for shipping. Remove the tape and make certain that there are no kinks in the vacuum hose. The longer tube should be attached from the waste container to the CCS (see instructions below).
CAUTION
NOTE
The severely damage the vacuum pump.
If an in-house vacuum is to be used to remove oil from the CCS, you may omit the use of the waste container vacuum pump.
ment. Ordinarily the address should be set to "0". If other CANNON instruments are to be controlled from the same computer, each instrument must have a unique address. Each CCS requires a dedi­cated COM port.
waste container
must be connected properly or waste liquid may
Connection procedure
1. If the shorter length of yellow tubing is not already attached to the vacuum pump, it will be necessary to access the pump to attach it to the Waste Vacuum connector by removing the six screws securing the pump housing to the waste system assembly . After the housing has
CANNON
Version 2.0a — November 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
10
been removed, connect one end of the yellow tubing to the fitting on the vacuum pump. Attach the other end to the remain­ing fitting on the waste container (see photo, above).
2. Connect the longer length of 1/4" yellow plastic tubing to the other fitting on the lid (this fitting includes a hose extension which projects into the bottle).
3. Connect the other end of this longer tube to the vacuum fitting mounted on the right rear panel of the Simulator Mechanism. This
"T" connection from viscometric cell to rear panel waste container fitting
to
vacuum
pump
to
CCS unit
Waste container lid
fitting completes the vacuum connection to the CCS viscometric cell (see photo).
Coolant connections
There are two tubing connections on the rear of the CCS housing. The fitting on the left is the coolant INLET (for introducing the water/antifreeze mix­ture from the refrigerated chiller to the CCS) and the fitting on the right of the rear panel is the coolant OUTLET (for returning coolant to the refrigerated chiller). Secure the ends of both hoses to these fittings, then secure both hoses to the appropriate bulkhead fittings on the refrigerated chiller (see photo) using any connection hardware provided with the tubing (if applicable). Route the hose from the CCS OUTLET to the refrigerated chiller INLET, and the hose from the CCS INLET
to the refrigerated chiller OUTLET.
Refrigerated chiller coolant connections
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
Filling the refrigerated chiller
Check the operations manual for the refrigerated chiller to determine its capacity . Then obtain the necessary volume of water and high-quality automotive antifreeze (ethylene glycol) necessary to provide a mix with a ratio of 10 percent antifreeze to 90 percent water. After you have secured the refrigerated chiller tubing connections to the CCS instrument (see previous section) add the necessary quantity of water/antifreeze mixture to the refrigerated chiller. The modified Julabo refrigerated chiller includes a sight glass for monitoring the liquid level. the temperature for the circulating coolant should be set at +5°C.
11
For CCS operation,
CAUTION
NOTE
Ethylene glycol is a toxic substance. Use proper safety precautions when handling. Follow appropriate MSDS procedures.
Ensure that the chiller is connected to the CCS and functioning properly before providing power to the CCS.
The water-antifreeze mix should be replaced annually for reliable perfor­mance and to prevent corrosion of internal components.
Do not set the chiller temperature below the freezing point of water (0°C).
Depending on the amount of fluid displacement in the coolant lines, it may be necessary to add additional antifreeze/water mixture to the refrigerated chiller when the unit commences operation. Monitor the liquid level in the refrigerated chiller during normal operation. If air bubbles are consistently visible in the coolant lines, add additional mixture until the refrigerated chiller is full.
EPYT obaluJA.erusserpcirehpsomtaotnepo,)delaestontub(riovreserdesolC
SNOITACIFICEPSRELLIHCDETAREGIRFERDERIUQER
sirellihcdeifidom
.NONNACmorfelbaliavadnadednemmocer
MUIDEMGNILOOC ezeerfitnatnecrep01htiwretawdellitsidfoerutxiM
YTICAPAC
ETARWOLFPMUP 1ta.nim/)strauq1.2-1(sretil2-1tsaeltAdaehfo)sehcni93(retem
SENILGNILOOC ni)sehcni5.-52.0(mm21-6,htgnelni)sehcni93(retem1yletamixorppA
YTICAPACLAVOMERTAEH C°5+tasttaw002tsaeltA
ERUTAREPMETWOL
C°5+taniatniamdnalortnoC
retemaid
CANNON
Version 2.0a — November 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
12
Installing VISCPRO® software
VISCPRO® is a powerful new software product providing a generic instrument interface for controlling and operating your CANNON instrument via computer. VISCPRO® also includes reporting/analysis modules for processing and displaying sample data.
T o install the VISCPRO sequence presented. Make certain that you complete the sections on checking instrument settings and calibration data. If you encounter difficulties at any stage in the installation process, call CANNON T echnical Services at 814-353-8000.
Windows® installation
1. Turn on your computer. Wait for the Windows® software to load.
2. From the Windows
3. Double-click the Add/Remove Programs icon and follow the
Installation actions
The installation program will:
create a directory for your data files. The default directory is
write SETUP information to the Windowscopy the software executable file and other necessary files to the
update other files in your Windows
place a shortcut icon for the VISCPRO
®
®
software, follow the instructions below in the
®
®
Start Bar click Settings/Control Panel. Insert
the first VISCPRO® installation disk or CD-ROM into the disk drive.
Windows prompts to complete the installation procedure. The executable file for VISCPRO
®
software installation is SETUP.EXE.
C:\Program Files\Cannon Instrument\VISCPRO).
®
registry.
directory you specify.
®
directories to versions fully
compatible with the current VISCPRO® software.
®
executable file on your
Windows
®
desktop.
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
Specifications
SNOISNEMID mm001tsaeltadda(peed)"62(mm066×ediw)"8/1-31(mm333×hgih)"82(mm117
THGIEW fi(tinugniloocobaluJrof).sbl09(gk14/tinuyramirp0012-SCCrof).sbl531(gk16
THGIEWGNIPPIHS 521(gk75;seirosseccadnatinuyramirp0012-SCCrof).sbl462(gk021yletamixorppA
morfdesahcrup NONNAC
)noitallatsnirofhtpedot)"4(
®
(tinugniloocobaluJrof).sbl NONNAC
morfdesahcrupfi
13
SNOITACIFICEPS0012-SCC
seirosseccarof).sbl56(gk03;)
®
)
LACIRTCELE W0001,zH06/05,%01±CAstlov511/001:0012-SCC#ledoM
GNITAREPO
SNOITIDNOC
GNITARESUF mm02x5,A4V052M:tinutlov032;mm02x5,A8V052M:tinutlov511/001:rellortnoC
ECNAILPMOC ,CDV0091(TOP-IH,)CEE/32/37(evitceridegatlovwoL;)CEE/633/98(evitceridCME
W0001,zH06/05,%01±CAstlov032:F0012-SCC#ledoM
2eergednoitulloP,IIyrogetaCnoitallatsnI,gnisnednoc-nonHR%09-%01,C°0351
).ces06
.tinuruoyhtiwdeilppusdrocrewopdevorppaehtylnoesU
CANNON
Version 2.0a — November 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
14
This page intentionally left blank.
CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
Version 2.0a — November, 2010; CANNON
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
CHAPTER
Configuring the CCS
2
This chapter contains information on configuring the CCS instrument for successful operation. Additional configuration information may be found in a separate document, CCS Installation Notes, provided with your CCS instrument or software upgrade.
Checking/editing configuration settings in VISCPRO
15
Original installation
CAUTION
Configuration protection
Your original software installation package may have included preset instrument configuration data on an accompanying floppy disk. This information should be copied to the VISCPRO® directory on your computer immediately following initial installation of the VISCPRO software. Consult the CCS Installation Notes for further information.
Copying or recopying floppy disk information to your VISCPRO® direc­tory will overwrite existing sample data. If you wish to retain a record of previously-saved configurations, make certain to archive your existing sample data before doing so (see Chapter 11 for information on using the Database Manager software).
Follow the procedures in this chapter and the chapter on CCS calibration to verify/edit the instrument and calibration settings to ensure that they conform to the actual characteristics of your CANNON® instrument.
T o check the configuration settings for your instrument(s), you must log in to the security system as a manager. The software is installed with a default Manager account. This account has no password, allowing any operator access to manager-level software functions as long as the password is not activated/changed. If you would like to engage the full­release security options, see Security Options in Chapter 5 for instruc­tions.
®
Running the software
Logging in
CANNON
Version 2.0a— November, 2010; CANNON
2139 High Tech Road • State College, PA 16803 • USA
T o start the VISCPRO® software program, double-click the VISCPRO icon on your Windows® desktop (Windows® NT® users can click Start/
Programs/VISCPRO/VISCPRO.EXE).
1. Use your mouse to click Main from the VISCPRO® menu bar.
2. Click Log In from the Main menu options.
3. Click on the to display the list of registered users.
4. Click Manager. It is not necessary to enter a password.
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
(arrow) on the right side of the User Name: list box
®
Instrument Company
®
16
5. Click OK to complete the Log In procedure.
NOTE
For more information on VISCPRO security options, see page 40.
Checking/editing Instrument Settings
1. Use your mouse to click (select) Configure from the VISCPRO menu bar (see graphic, next section).
2. Select your instrument from the list of available instruments (there may be only one instrument in the list).
3. Select option 2, Instrument Settings, from the list of configuration options. The Instrument Settings window will appear.
You will use the Instrument Settings window (see below) to describe and control CCS instrument operational features. These settings affect the instrument as a whole. Check the instrument settings for your CCS per the instructions below , and make any necessary changes:
®
The Instrument Settings window
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
Use the ID field to input instrument identification information using up to 30 alphanumeric characters.
The S/N: field should correspond to the four-digit serial number from the label on the CCS rear service panel.
The probe constant values (a0, a1, and a2) must be identical to the values on the calibration certificate provided with your temperature probe. Verify that the values correspond.
Note that firmware and communication port settings are automatically determined. These values cannot be altered from this window.
When you have verified all settings for all editable fields, click OK.
Viewing/editing other setup information
If your instrument has already been set up by a technician, you can use the instructions in this section of the manual to check or, if necessary, change the instrument settings.
17
1. Click Configure from the VISCPRO® menu bar .
2. Select your instrument from the list of available instruments.
3. Select Tray (Test) Settings from the list of configuration options. The Tray (Test) Settings window will appear.
The Tray (Test) Settings window
CANNON
Version 2.0a — November 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
18
The T ray (Test) Settings window contains setup information for your instrument. Default test settings (which may be restored at any time by clicking the Defaults button) are specified in ASTM D 5293. In certain circumstances, the user may wish to alter these settings. For more infor­mation, see Tray (Test) Settings in Chapter 5.
VISCPRO software interface elements
Title Bar Menu Bar
Instrument View window
The VISCPRO® primary display window is framed on the top by the VISCPRO
®
title bar and menu bar, and on the bottom by the VISCPRO
status bar. The application window is usually preconfigured to include two child windows. The first is an Instrument View window, which
®
describes your CANNON
instrument and provides controls for CCS testing, and the second is an analysis window that displays data from CCS tests.
®
Analysis window
Status bar
The VISCPRO® primary display
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
19
NOTE
If the Instrument View window does not appear, click View Instrument from the Main menu:
The View Instrument window
Then click the desired instrument from the list of available instruments and click OK. If the Available Instruments list box is blank, your CCS instrument may not be on-line. Check cable connections and make certain the instrument is ON.
The Instrument View window
CANNON
Version 2.0a — November 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
20
The CCS Data Table window
NOTE
To change the Tray (Test) Settings for your instrument(s), you must log in to the security system as a manager.
4. Edit the values for Tray (Test) Settings as necessary. When done,
click OK to save current values and exit the Tray (Test) Settings window.
Initial Stepper Motor Service
Prior to performing instrument actions with the CCS-2100 Automatic Cold-Cranking Simulator, you must first home the Sample Tray mecha­nism as follows:
1. Make certain that the chiller is connected and operating properly .
Then turn the CCS instrument is ON.
2. Load the VISCPRO software by double-clicking the VISCPRO
on your W indows® desktop (Windows® NT® users can click Start/
Programs/VISCPRO/VISCPRO.EXE).
3. Click Service from the menu bar and select your CCS instrument
from the list.
4. Select Stepper Motor Service from the Service options.
®
icon
CCS sampler mechanism with tube in home position (raised)
CANNON
Version 2.0a — November, 2010; CANNON
5. Click on the Tube Up
button until the tube has reached the top of its travel.
6. Ensure that the arrow on
the Sample Tray is aligned with the arrow on the Sample T able (Table Forward/Table Back buttons).
7. Click on the Trolley Left
button until the sampler has reached the limit of its travel. If the Sample Tray has been aligned properly , the sampler should now be
CCS-2100 with tray in home posi-
centered over the #1 sample position.
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
8. Click the Set Home Flag button. Note that the button will disappear
when the home flag has been set successfully . Then click close the window .
Setting the CCS Set Motor Current
The CCS Motor Current window in VISCPRO® provides the software interface for adjustment of the CCS motor current. To set the CCS motor current:
1. Select the Set Motor Current option from the Service menu. The
Set Motor Current Service window will appear.
2. Prepare a 60 ml bottle of CANNON
place it in position 1 of the Sample Table. Then press the Start button. The CCS will begin the current adjustment procedure.
3. When the rotor/stator has reached the correct temperature, the
software will prompt you to adjust the CCS motor current using the dial on the front panel of the Control Unit. Adjust the current until the value in the Speed: field of the Motor Current Service window indicates a rotor speed of 240 rpm.
to
®
CL25 viscosity standard and
21
The Set Motor Current Service window will continue to display cell temperature and cell motor speed. When you have completed the rotor speed adjustment procedure, press the Close button to exit the Set
Motor Current Service window. To terminate the Set Motor Current Service process at any time, click Finished.
CANNON
Version 2.0a — November 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
22
This page intentionally left blank.
CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
Version 2.0a — November, 2010; CANNON
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
CHAPTER
3
Calibration is essential for the proper operation of the CCS. The CCS is calibrated before it is shipped to the customer; however, the CCS should be recalibrated after installation. Frequency of recalibration is at the discretion of the user.
The CCS calibration procedure ensures that the mathematical constants used in calculation of viscosity from CCS measurements accurately reflect the unique properties of the CCS rotor/stator mechanism.
Calibration procedure
During calibration, the software uses calibration standard data from CCS tests of viscosity standards to perform the mathematical calculations necessary to derive three calibration constants– b0, b1 and b2. Then the program displays the new constants and recomputes a viscosity (in cSt) for each of the calibration standards. The resulting viscosity calculation is compared with the known viscosity of the standard and displayed on the computer screen.
23
CALIBRATING THE CCS
NOTES
The software allows the operator to manually change the constants if necessary to more closely match actual and calculated viscosities.
A calibration must be performed for each temperature at which samples will be tested.
Preparing/updating the list of calibration standards
During installation of your VISCPRO® software package, a list of current CANNON CL standards (see table, next page) is automatically loaded into the VISCPRO® database. Before testing calibration standards, it will be necessary for you to enter the viscosity data from the standard bottles into the VISCPRO® program using the following procedure:
1. Click Configure from the VISCPRO® primary display and select
your CCS instrument.
2. Click CCS Calibration Standards from the Configure menu
options. The CCS Calibration Standards window will appear (see graphic, following).
3. Assemble your calibration standards and input the viscosity data (in
mPa·s) from the standard bottles for each CL standard and tempera­ture in the appropriate fields. If you have additional standards, including user-defined standards, you can enter these by right-
CANNON
Version 2.0a— November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
24
clicking your mouse in the Sample ID field and selecting the Insert New option . To delete an unused standard, right-click your mouse in the Sample ID field and select Delete.
®
NONNAC
-itarbilaC
liOno
C°5- C°01- C°51- C°02- C°52- C°03- C°53-
080LC 009
090LC 0021
001LC 0090071
011LC 05510052
021LC 00800610023
031LC 005100920584
041LC 008006105230007
051LC 0071007200640508
061LC 0052005500011
071LC 054105220073003600311
091LC 00810053004700071
002LC 776105620034055700731
022LC 00310052001500111
042LC 0522006300060070100891
052LC 00810053004700271
062LC 057100720044005700431
082LC 0031005200050039
003LC 0042057300160050100391
023LC 00810053003700951
043LC 00720024000749121
083LC 0092008500031
024LC 0025005850441
084LC 00320054005900012
035LC 0006348918861
006LC 0073003700651
086LC 0559
047LC 000600021
erutarepmeT
SDRADNATSNOITARBILAC
detacidnItaytisocsiVlanimoNetamixorppA
NOTE
A minimum of five oils are required for calibration at each temperature in accordance with ASTM D 5293- 08.
CANNON
Version 2.0a — November, 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
Instrument Company
25
CAUTION
Make certain to verify the Known Viscosity (calibration standard viscosity in mPa·s) in the calibration list BEFORE testing selected standards (see next sections). Incorrect or outdated viscosity data may skew future calibration results.
4. After you have entered all ID, temperature and viscosity information
into the CCS Calibration Standards list (see graphic), click the SAVE button to save the data and exit the window.
Testing calibration standards
Selecting standards
The CCS Calibration Standards list window
T o calibrate the CCS instrument, it is first necessary to test samples of known viscosity at the desired calibration temperatures.
1. Assemble a group of CANNON® standard oils with known viscosi-
ties for calibration. Refer to the table on the previous page when choosing the appropriate CL standards.
2. Click Configure from the VISCPRO
your CCS instrument.
3. Click CCS Choose Calibration Standards from the Configure
menu options. The CCS Choose Calibration Standards window (see graphic, next page) will appear.
4. Click the desired standards to select them for calibration. A check
mark indicates that the standard has been selected.
CANNON
Version 2.0a — November 2010; CANNON
®
CCS-2100 with VISCPRO® Instruction & Operation Manual
2139 High Tech Road • State College, PA 16803 • USA
®
®
primary display and select
Instrument Company
Loading...
+ 81 hidden pages