Braun ThermoScan Service manual

Welch
Allyn
Service Documentation
Rev. J
Welch Allyn
701627 REV. J Page 1 of 9
This documentation provides only technical information - technical data, calibration instruction, general information and regulations for medical products.
The function of the thermometer is described in the Operating Instructions, Welch Allyn P/N 701622
Technical data: Display: LCD-Display.
Display resolution: 0.1 °C / °F.
Displayed temperature range: 20.0 °C - 42.2 °C (68.0 °F - 108.0 °F). Work place temperature range: 10.0 °C - 40.0 °C (50.0 °F - 104.0 °F). Work place humidity: up to 95 %. In the range between: 35.5 °C - 42.0 °C (95.9 °F - 107.6 °F).
the accuracy of allowable tolerance with probe cover is: ± 0.2 °C (± 0.4 °F).
Outside the range from: 35.5 °C - 42.0 °C (95.9 °F - 107.6 °F). the accuracy of allowable tolerance with probe cover is: ± 0.3 °C (± 0.5 °F).
Probe tip: heated up to approximately 37 C. Power supply: rechargeable Braun battery-pack with
base-station and smart plug
or 2 1.5 V type AA (LR 6) batteries. Smart plug: 100-240 V - input / 12 V – output. Base-station: 12 V - input / approx. 2.4 V – output. Rechargeable Battery-life: 3 years / 1000 charging cycles Rechargeable battery capacity: 1 month standby / 500 measuring cycles Rechargeable battery charging: Thermometer stored in basis-station. Rechargeable battery charging time: 14 hours. Base-station function indications: 1 LED, green for power connection
1 LED, yellow for charging indication. Electronic anti-theft *: for Pro 4000 appliances only and with
the base-station in use. Automatic turn off: If no button is pressed within 60 seconds, the
appliance will turn off automatically. Manual turn off: Push - button for >3 seconds.
Welch
Allyn
Service Documentation
Rev. J
Welch Allyn
701627 REV. J Page 2 of 9
Attention:
Do not dispose of empty batteries or defective rechargeable batteries in household waste. Disposal can take place at appropriate collection sites provided in your country
Hint:
Type BF equipment This appliance conforms to the following standards: DIN EN 60601-1
<<Medical electrical equipment>>-Part 1: General requirements for safety.
Hint:
0297
This product conforms to the provisions of the EC directive 93/42/EEC (Medical Device Directive).
General hints
Production code Near the battery storage LOT 4042.
4 = Year / 042 = Day.
Labour Group: 03
Welch
Allyn
Service Documentation
Rev. J
Welch Allyn
701627 REV. J Page 3 of 9
Special tools: The accuracy of infrared thermometers can only be tested with a blackbody device
(BB-3200 Series, no longer available for purchase) or the 9600 PLUS Calibration Tester.
The Model 9600 PLUS Calibration Tester can be ordered through Welch Allyn Customer Service (see page 8) for Welch Allyn contact information.
Model Welch Allyn P/N Braun P/N 9600 PLUS Calibration Tester 01802-110 N/A
The following additional data will be necessary:
Bill to address Ship to address Quantity Billing contact person with fon/fax extension Shipping contact person with fon/fax extension Shipping method Purchase Order / Credit Card
The blackbody has to be calibrated every 12 months by the manufacturer. Contact Welch Allyn Customer Service to get a Return Material Authorization number to have your unit calibrated. Refer to Customer Service Listing (see page 8) for Welch Allyn contact information.
Please direct all requirements and questions concerning blackbodies directly to the Welch Allyn Service Center. Refer to Customer Service Listing (see page 8) for Welch Allyn contact information.