BMI Jolly Plus Technical Manual

Jolly Plus Series
Manual code: DQB26-EN [Rev. 02]
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 01] Page 2
JOLLY PLUS Series TECHNICAL MANUAL
GENERAL INDEX PAGE
CHAPTER 1 - Equipment Description
1.1 Graphic symbols
1.2 Utilization purposes and equipment operating life
1.3 Technical features
1.3.1 Electrical features
1.3.2 Operation features
1.3.3 Environmental data
1.3.4 Radiographic features
1.3.5 Mechanical features
1.3.6 Collimator
1.3.7 Monobloc
1.3.8 Inverter
1.4 Equipment description
1.4.1 Main elements
1.4.2 Control console
1.5 Description of electrical and electronic main elements
1.5.1 Control console assembly
1.5.2 Electronic box
1.5.3 Inverter
1.5.4 Capacitors box
1.5.5 Monobloc
1.6 Label
1.7 Dimension
1.8 Reference standard
1.9 Safety
1.10 Electromagnetic compatibility (EMC) 30
1.11 Declaration of responsibility 34
CHAPTER 2 - Installation
2.1 Introduction
2.2 Unpacking and verification of the equipment and accessories integrity
2.3 Electric installation
2.4 Setting at work
2.4 1 Electric inspection
2.4.2 Mechanical inspection
CHAPTER 3 - Utilization
3.1 Transport of the equipment
3.2 Equipment’s switching ON
3.3 Arm positioning
3.4 Monobloc and Collimator displacement
3.5 Collimator
3.6 Radiographic techniques
3.6.1 «Two-Point» manual technique radiographic exposure
3.6.2 «Three-Point» manual technique radiographic exposure
3.6.3 APR Radiographic exposure
3.6.4 APR - anatomical programming chart
3.6.5 Making changes to pre-programmed APR techniques
3.7 Taking an radiographic exposure
3.8 End of operation
9 10 11 11 11 12 13 15 15 16 16 17 17 18 20 21 22 24 25 27 27 28 28 29
37 37 38 38 38 38
41 42 43 44 45 46 46 46 47 48 49 50 51
Manual code: DQB26-EN [Rev. 02] Page 3
JOLLY PLUS Series TECHNICAL MANUAL
GENERALE INDEX PAGE
CHAPTER 4 – Adjustments and Set Up
4.1 Adjustments
4.1.1 X-ray tube formation
4.1.2 X-ray film centering
4.1.3 kV values adjustment
4.1.4 mA values adjustment
4.2 Set Up and Calibration
4.2.1 Language preselection
4.2.2 Options preselection
4.2.3 Counter
4.2.4 kV values calibration
4.2.5 mA values calibration
4.2.6 Touch screen calibration
4.2.7 Touch screen failure
CHAPTER 5 – Maintenance
5.1 General recommendations
5.2 Ordinary maintenance
5.3 Mechanical adjustment
5.3.1 Brake block adjustment
5.3.2 Spring adjustment
5.3.3 Arm clutch adjustment
5.3.4 Fork joke rotation clutch adjustment
5.3.5 Monobloc rotation clutch adjustment
5.3.6 Collimator lamp replacement
5.3.7 Steel–clad ropes and the mechanical equalizer replacement
5.4 Cleaning and disinfection
5.5 Disposal
CHAPTER 6 – Troubleshooting and spare parts
6.1 Troubleshooting
6.1.1 Introduction
6.1.2 Warning messages
6.2 Spare parts list
6.3 Fuses list
CHAPTER 7 – Schematics
7.1 Electronic boards topographic layout
7.2 Schematics
55 55 57 59 60 62 63 63 65 65 67 71 72
75 75 80 80 81 85 87 88 89 89 89 89
93 93 94 96 98
101 109
Manual code: DQB26-EN [Rev. 02] Page 4
JOLLY PLUS Series TECHNICAL MANUAL
THE TECHNICAL MANUAL SHOULD BE READ BEFORE STARTING USING THE EQUIPMENT. IT IS RECOMMENDED TO FOLLOW STRICTLY THE INSTRUCTIONS OF THE PRESENT IN ORDER TO AVOID ANY ILLEGITIMATE USE OF THE UNIT THE INSTALLATION AND MAINTENANCE MUST BE CARRIED OUT BY PEOPLE AUTHORIZED BY THE MANUFACTURER.
. THE UNIT IS SUPPOSED TO BE USED BY QUALIFIED AND AUTHORIZED PEOPLE ONLY.
0051
REVISIONS CHRONOLOGY
REVISION DATE MODIFICATIONS
00 13.04.2010 First issue 01 01.04.2011 Updated chapters nr. 1-2-3-4-5-6 02 21.09.2011 Added paragraph 1.10 and changed paragraph 1.10 in 1.11.
Updated Paragraphs nr. 1.1 and 2.2
Manual code: DQB26-EN [Rev. 02] Page 5
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 6
JOLLY PLUS Series TECHNICAL MANUAL
CHAPTER 1
EQUIPMENT DESCRIPTION
Manual code: DQB26-EN [Rev. 02] Page 7
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 8
JOLLY PLUS Series TECHNICAL MANUAL
T
T
1.1 GRAPHIC SYMBOLS
The following symbols are available in the technical manual, in the user manual and on the labels applied on the equipment:
Compliance with directive 93/42/CEE medical devices
Equipment manufacturing data (month and year)
Protection degree against electric hazard (CEI EN 60601-1; Type B)
Ionogenic radiation health hazard
Attention: consult the annex documentation
Caution: read the instructions
Attention: electrocution risk
X-Ray model insert in the monobloc
otal filtering of the X-ray tube / Monobloc assembly
X-Ray tube small focus
~
Protection earth
X-Ray tube large focus
Equipment stand by
Single phase power supply
his side up. Symbol in according to ISO 780.
Fragile. Symbol in according to ISO 780.
Sensitive to moisture, keep away from the rain and snow.
Symbol in according to ISO 780.
Do not use hand hooks. Symbol in according to ISO 780.
Equipment subject to separate disposal.
Manual code: DQB26-EN [Rev. 02] Page 9
JOLLY PLUS Series TECHNICAL MANUAL
1.2 UTILIZATION PURPOSES AND EQUIPMENT OPERATING LIFE UTILIZATION
PURPOSES
Jolly Plus series is composed by four models depending on power:
- Jolly 4 Plus - fixed anode 4 kW
- Jolly 15 Plus - rotating anode 15 kW
- Jolly 16 Plus - rotating anode 16 kW
- Jolly 30 Plus - rotating anode 30 kW
Jolly Plus mobile equipment is suitable for many medical applications and is supposed to be used by qualified people only, duly trained and informed about the risks deriving from Ionogenic radiations. This system is not included in the list of devices requiring continuity of operation. This system does not come into contact with the patient, only some parts of the device may accidentally touch the patient. The contact with the operator is contemplated during the equipment operation. The contact with the patient is non-invasive. The contact with the operator is required for reasons depending on the equipment utilization. The equipment has been specifically designed for radiological examinations and diagnostic investigations carried out in Hospitals and Medical Centers. It is suitable for operation in:
Respiratory System investigations
Skeleton investigations
Sport Medicine
Paediatrics Orthopaedics
First-Aid Stations
Plaster Room
Operating Theatre
Traumatology
EQUIPMENT LIFE
The average life of this equipment depends on the life of the X-ray tube installed in the monobloc. The compliance with the instructions of this Manual will make it possible to get the maximum life duration of the X-ray tube. It is not always possible to know in advance when the X-ray tube’s operation life is coming to an end. This component could even stop operating during the examination of a patient. For this reason it is recommended to check the focal spot yearly, in order to find if there is any degradation of the anode, indicated by the increase of the focal spot dimension and consequent worse image resolution.
The main causes of early wear and damages to the X-ray tube are:
Filament prolonged heating
Discharges in the tube
Tube over current
Cold anode overcharge, caused by excessive loads after a long period of inactivity, with consequent cratering of the
track.
Here below how to limit untimely wear and damages to the monobloc:
1. The X-ray tube preheating cycle avoids the overheating since the tube reaches the operating temperature slowly.
IF A HIGH VALUES EXPOSURE IS STARTED WHEN THE X-RAY TUBE IS STILL COLD, THE ANODE COULD BE IRREPARABLY DAMAGED
.
THE TUBE FORMATION PROCEDURE MAKES IT POSSIBLE TO REDUCE THE PROBABILITIES OF DISCHARGE
2. If a high values exposure is started when the device has not been used for a long time, the X-ray tube could be
irreparably damaged.
3. The X-ray tube overcharge is avoided by the software control on the preset kVp and mAs values. The system
prevents the exposure if these values exceed the maximum values provided for the X-ray tube.
4. Filament temperature preparation. It is recommended to effect the preparation only when the patient is positioned
for the examination.
Manual code: DQB26-EN [Rev. 02] Page 10
JOLLY PLUS Series TECHNICAL MANUAL
1.3 TECHNICAL FEATURES
1.3.1 ELECTRICAL FEATURES
Jolly 4 Plus, Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus have the same electrical data:
Description Value
Main power supply Voltage: 230 Vac single phase
Frequency: 50 Hz
Current consumption: 16 A maximum Line compensation ± 10% automatic Total line resistance 0,4 total Supply socket (standard) 16 A @ 230 Vac Insulating class EN60601-1 Class I with applied parts type B Utilization conditions Continuous operation at intermitted load
Classification Directive 93/42/CEE according to annex IX: Class II b
1.3.2 OPERATION FEATURES
Jolly 4 Plus, Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus have the same operation features:
Description Value
Control console Microprocessor controlled console
5.7” Colour TFT LCD touch screen display
Resolution 320 x 240 pixel
Acoustic and luminous X-ray emission alarm
Languages available:
- Italian
- English
- Spanish
- French
- Russian
- German
- Bulgarian
- Vietnamese
- Turkish
X-ray exposure control
Console control by means of two push buttons
st
- I
time – preparation and anode starting
nd
- II
time - X-ray exposure radiographic exposure
Remote control by wire
- two-step X-ray push button control with spiral cable extendable
up to 3.8 meter length
Preselected technique radiography 3-point technique (kV, mA and sec.)
Free technique radiography (APR)
2-point technique (kV and mAs)
Safety and protections Capacitors voltage
Inverter over current
mA min. and max safety device
Monobloc working temperature
X-ray filament safety device
X-ray over load safety device
X-ray tube anode rotation
X-ray tube Hu (thermal unit)
Manual code: DQB26-EN [Rev. 02] Page 11
JOLLY PLUS Series TECHNICAL MANUAL
1.3.3 ENVIRONMENTAL DATA
Jolly 4 Plus, Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus have the same environmental data:
Description Transport and Storage Operating ambient
Maximum temperature From -15 °C to +70 °C From +10 °C to +40 °C
Relative humidity non condensing From 5 %to 95 % From 20 %to 80 %
Altitude pressure range 3000 meters maximum
700 hPa / 525 mm Hg
LONG-TERM STORAGE OVER 40 °C WILL REDUCE THE SERVICE LIFE OF ELECTROLYTIC CAPACITORS.
RANSPORT AND STORAGE IS LIMITED TO A MAXIMUM DURATION OF 120 HOURS BETWEEN 50 AND 70 °C.
T
3000 meters maximum 700 hPa / 525 mm Hg
Manual code: DQB26-EN [Rev. 02] Page 12
JOLLY PLUS Series TECHNICAL MANUAL
1.3.4 RADIOGRAPHIC FEATURES
Jolly 4 Plus
Description Values
Max power 4 kW (40 mA - 100 kV @ 100 msec)
H.T. control H. F. Inverter
Frequency 40 KHz
Max ripple
kV value increasing time
2% @ maximum power
3 msec
Exposure time 2 msec ÷ 6000 msec (34 steps)
Range: 2 • 3 • 6 • 8 • 10 • 12 • 15 • 20 • 25 • 30 • 40 • 50 • 63 • 80 • 100 • 125 • 150
• 200 • 250 • 300 • 350 • 400 • 500 • 630 • 800 • 1000 • 1250 • 1500 • 2000 • 2500
• 3000 • 4000 • 5000 • 6000 msec
kV value variation range 40 ÷ 110 kV (step 1 kV)
mA Range 25 ÷ 80 mA (5 steps)
Range: 25 • 40 • 50 • 63 • 80 mA
mAs range
2 point technique
0,5 ÷ 200 mAs (24 steps) Range: 0,5 • 0,8 • 1 • 2 • 2,5 • 3 • 4 • 5 • 6,3 • 8 • 10 • 12,5 • 15 • 20 • 25 • 30 • 40
• 50 • 63 • 80 • 100 • 125 • 150 • 200 mAs
Jolly 15 Plus
Description Values
Max power 15 kW (150 mA - 100 kV @ 100 msec)
H.T. control H. F. Inverter
Frequency 40 KHz
Max ripple
2% @ maximum power
kV value increasing time
3 msec
Exposure time 2 msec ÷ 6000 msec (34 steps)
Range: 2 • 3 • 6 • 8 • 10 • 12 • 15 • 20 • 25 • 30 • 40 • 50 • 63 • 80 • 100 • 125 • 150 • 200
• 250 • 300 • 350 • 400 • 500 • 630 • 800 • 1000 • 1250 • 1500 • 2000 • 2500 • 3000
• 4000 • 5000 • 6000 msec
kV value variation range 40 ÷ 125 (step 1 kV)
mA Range 25 ÷ 200 mA (9 steps)
Range: 25 • 40 • 50 • 63 • 80 • 100 • 125 • 150 • 200 mA
mAs range
2 point technique
0,5 ÷ 250 mAs (25 steps) Range: 0,5 • 0,8 • 1 • 2 • 2,5 • 3 • 4 • 5 • 6,3 • 8 • 10 • 12,5 • 15 • 20 • 25 • 30 • 40
• 50 • 63 • 80 • 100 • 125 • 150 • 200 • 250 mAs
Manual code: DQB26-EN [Rev. 02] Page 13
JOLLY PLUS Series TECHNICAL MANUAL
Jolly 16 Plus
Description Values
Max power 16 kW (160 mA - 100 kV @ 100 msec)
H.T. control H. F. Inverter
Frequency 40 KHz
Max ripple
kV value increasing time
2% @ maximum power
3 msec
Exposure time 2 msec ÷ 6000 msec (34 steps)
Range: 2 • 3 • 6 • 8 • 10 • 12 • 15 • 20 • 25 • 30 • 40 • 50 • 63 • 80 • 100 • 125 • 150
• 200 • 250 • 300 • 350 • 400 • 500 • 630 • 800 • 1000 • 1250 • 1500 • 2000 • 2500
• 3000 • 4000 • 5000 • 6000 msec
kV value variation range 40 ÷ 125 (step 1 kV)
mA Range 25 ÷ 200 mA (9 steps)
Range: 25 • 40 • 50 • 63 • 80 • 100 • 125 • 160 • 200 mA
mAs range
2 point technique
0,5 ÷ 250 mAs (25 steps) Range: 0,5 • 0,8 • 1 • 2 • 2,5 • 3 • 4 • 5 • 6,3 • 8 • 10 • 12,5 • 15 • 20 • 25 • 30 • 40
• 50 • 63 • 80 • 100 • 125 • 150 • 200 • 250 mAs
Jolly 30 Plus
Description Values
Max power 30 kW (300 mA - 100 kV @ 100 msec)
H.T. control H. F. Inverter
Frequency 40 KHz
Max ripple
2% @ maximum power
kV value increasing time
3 msec
Exposure time 2 msec ÷ 6000 msec (34 steps)
Range: 2 • 3 • 6 • 8 • 10 • 12 • 15 • 20 • 25 • 30 • 40 • 50 • 63 • 80 • 100 • 125 • 150
• 200 • 250 • 300 • 350 • 400 • 500 • 630 • 800 • 1000 • 1250 • 1500 • 2000 • 2500
• 3000 • 4000 • 5000 • 6000 msec
kV value variation range 40 ÷ 125 (step 1 kV)
mA Range 50 ÷ 400 mA (11 steps)
Range: 50 • 63 • 80 • 100 • 125 • 150 • 200 • 250 • 300 • 350 • 400 mA
mAs range
2 point technique
0.5 ÷ 300 mAs (26 steps) Range: 0,5 • 0,8 • 1 • 2 • 2,5 • 3 • 4 • 5 • 6,3 • 8 • 10 • 12,5 • 15 • 20 • 25 • 30 • 40
• 50 • 63 • 80 • 100 • 125 • 150 • 200 • 250 • 300 mAs
Manual code: DQB26-EN [Rev. 02] Page 14
JOLLY PLUS Series TECHNICAL MANUAL
1.3.5 MECHANICAL FEATURES
Description Jolly 4 Plus Jolly 15 Plus / 16 Plus / 30 Plus
Width 680 mm 680 mm
Length 1062 mm 1062 mm
Height with locked arm 1580 mm 1580 mm
Focus-floor distance - variable from 410 to 2010 mm from 400 to 2000 mm
Antistatic pivoting wheels diameter 80 mm 80 mm
Rear wheels diameter 250 mm 250 mm
Weight - net ~165 Kg ~165 Kg
Unit displacement Manual Manual
Radiographic cassettes holder Up to nr. 6 radiographic cassettes format 35 x 43 cm (14 x 17 inches)
1.3.6 COLLIMATOR
Jolly 4 Plus, Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus are equipped with same X-ray collimator:
Description Data
Mobile blades Square field - single layer
Blades displacement Manual
Size of irradiated square radiographic field Minimum 0x0 cm / Maximum 43x43 cm Focus / film distance indication Retractable tape measure
Collimator lamp 24V - 100W. Automatic switch Off after 30 sec
Light intensity (lux) @ 24V supply 160 Lux
Inherent filtration 2.0 mm Al @ 75 kV
Collimator rotation angle ± 90°
Manual code: DQB26-EN [Rev. 02] Page 15
JOLLY PLUS Series TECHNICAL MANUAL
1.3.7 MONOBLOC
Monobloc Jolly 4 Plus Jolly 15 Plus / 16 Plus / 30 Plus
Anode type Fix Rotating 3000 r.p.m.
Focus small and large 0.5 - 1.5 mm
0.6 - 1.3 mm
Focus used: 1.5 mm
Anode target angle 12° 15°
Anode material Tungsten RTM
Anode maximum heat storage capacity 30 kJ (40 kHU) 80 kJ (107 kHU)
Anode maximum continuous heat dissipation 270 W 300 W
Monobloc maximum power 4 kW Small focus 11 kW
Large Focus 32 kW
kV - monobloc maximum voltage 110 kV 130 kV
mA - monobloc maximum anodic current 80 mA @ 40 kV Small focus 220 mA @ 50kV
Large focus 640 mA @ 50 kV
Monobloc thermal capacity 500 kJ (600 kHU) 600 kJ (800 kHU)
Monobloc continuous thermal capacity 55 W 55 W
Monobloc Inherent filtration 1.8 mm Al @ 75 kV 1.1 mm Al @ 75 kV
FFUURRTTHHEER MMAANNUUAAL
R
TTEECCHHNNIICCAAL
L
SSUUPPPPLLIIEED
L
IINNFFOORRMMAATTIIOON
D
TTOOGGEETTHHEER
R
WWIITTH
N
AABBOOUUT
H
TTHHE
T
E
TTHHE
E
EEQQUUIIPPMMEENNT
MMOONNOOBBLLOOC
T
C
CCOOUULLD
D
E
BBE
FFOOUUNND
IIN
N
TTHHE
E
CCOOMMPPOONNEENNT
D
1.3.8 INVERTER
Inverter Jolly
4 Plus
Jolly
15 Plus
Jolly
16 Plus
Jolly
30 Plus
T
TTEECCHHNNIICCAAL
L
Voltage output (switching) to monobloc ~ 320 Vdc ~ 420 Vdc ~ 420 Vdc ~ 420 Vdc
Digital outputs Yes Yes Yes Yes
Digital inputs Yes Yes Yes Yes
Circuit main power supply voltages Logic: +5Vdc
Analogue: +12Vdc / -12Vdc / +15Vdc / -15Vdc
Net weight ~ 5 Kg ~ 5 Kg ~ 5 Kg ~ 5 Kg
Manual code: DQB26-EN [Rev. 02] Page 16
JOLLY PLUS Series TECHNICAL MANUAL
1.4 EQUIPMENT DESCRIPTION
1.4.1 MAIN ELEMENTS
11
1
2
3
4
5
16
9
10
8
7
12
13
14
15
Legend
1. Control console
2. Transport handles and parking brakes control
3. Supply cable support
4. Radiographic cassettes holder
5. Bracket to lean the unit for overcoming little obstacles
6. Rear wheels diameter 250 mm
7. Pivoting front wheels
8. Column
9. Arm
6
17
10. Arm retainer
11. Arm handles
12. Monobloc’s covers
13. Goniometer
14. X-Ray collimator
15. Handgrip
16. Hand switch for radiographic exposure
17. Automatic circuit breaker (Power ON/OFF)
Manual code: DQB26-EN [Rev. 02] Page 17
JOLLY PLUS Series TECHNICAL MANUAL
1.4.2 CONTROL CONSOLE
The figure below describes the Jolly 4 Plus, Jolly 15 plus Jolly 16 Plus and Jolly 30 Plus control console:
1
2
3 4
Legend:
1. Control console
2. Back lighted colour touch screen display size 5,7”
3. Console push button for X-ray emission: I
4. Console push button for X-ray emission: II
st
Time preparation and anode starting
nd
Time radiographic exposure
The icons below illustrate the symbols appearing on the touch screen display, with relevant explanation of their meaning:
ICON DESCRIPTION
Increment control of the main radiographic parameters: kV, mAs, mA and seconds
Decrement control of the main radiographic parameters: kV, mAs, ma and time
Display window of the preset kV value
Display window of the preset mA value
Display window of the preset time value
Display window of the preset mAs value
Selection window of radiographic parameters manual selection mode
Selection window of radiographic parameters automatic selection mode
Selection window of ionization chamber (optional)
Alarm indicating the wrong automatic presetting of the radiographic parameters or the pause time before one radiographic exposure and the following
Manual code: DQB26-EN [Rev. 02] Page 18
JOLLY PLUS Series TECHNICAL MANUAL
ICON DESCRIPTION
X-rays preparation luminous signal
X-ray control luminous signal
Thermal capacity value - luminous signal
Indication of the automatically preset focus as a function of the preselected radiographic parameters
Potter Bucky selection window (optional)
Light indicator of radiation field Power ON
Manual code: DQB26-EN [Rev. 02] Page 19
JOLLY PLUS Series TECHNICAL MANUAL
1.5 DESCRIPTION OF MAIN ELECTRICAL AND ELECTRONIC COMPONENTS
The below reported picture shows the area where are located inside the equipment the main electrical and electronic components.
Monobloc 1.5.5
Control console assembly 1.5.1
Electronic box 1.5.2
Inverter 1.5.3
Capacitors box 1.5.4
Manual code: DQB26-EN [Rev. 02] Page 20
JOLLY PLUS Series TECHNICAL MANUAL
1.5.1 CONTROL CONSOLE ASSEMBLY
Nr. Description Q.ty
1 Thermoformed control desk 01 2 «BMI» silk-screen printed panel 01 3 Electronic board «CPU board» 01 4 Electronic board «display holder» 01 5 Ist and IInd time exposure push button 02
1
5
5
External view - operator side
2
5
4
3
5
Internal view - operator side
Manual code: DQB26-EN [Rev. 02] Page 21
JOLLY PLUS Series TECHNICAL MANUAL
1.5.2 ELECTRONIC BOX
Jolly 4 Plus:
Nr. Description Q.ty
1 Electronic board «filament» 01 2 Electronic board «main CPU» 01 3 Electronic board «main board» 01 4 Single phase EMC line filter 01 5 Solid state relay 02 6 Toroidal transformer single phase 120VA 01 7 Jolly Plus ground copper bar 01 8 Socket CN 1 - Capacitors box main power supply 01
9 Main power supply cable 01 10 External X-ray hand switch terminal block 01 11 Automatic single phase circuit breaker 250V 16A 01
6
5
5
9
4 8 7 3 2 1
Electronic box - operator side view
10 11
Electronic box – back side view
Manual code: DQB26-EN [Rev. 02] Page 22
JOLLY PLUS Series TECHNICAL MANUAL
Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus:
Nr. Description Q.ty
1 Electronic board «filament» 01
2 Electronic board «main CPU» 01
3 Electronic board «main board» 01
4 Electronic board «starter board»
5 Single phase EMC line filter 01
6 Solid state relay 02
7 Toroidal transformer single phase 120VA 01
8 Jolly Plus ground copper bar 01
9 Socket CN 1 - Capacitors box main power supply 01 10 Main power supply cable 01 11 External X-ray hand switch terminal block 01 12 Automatic single phase circuit breaker 250V 16A 01
7
6
6
10
5 9 8 3 2 1
Electronic box - operator side view
4
11 12
Electronic box – back side view
Manual code: DQB26-EN [Rev. 02] Page 23
JOLLY PLUS Series TECHNICAL MANUAL
1.5.3 INVERTER
Jolly 4 Plus
Code Description
S03S008 BMI H.F. Inverter for fixed anode monobloc
Inverter
Electronic board «control logic»
Equipment back side view
Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus:
Code Description
S03S004 BMI H.F. Inverter for rotating anode monobloc
Inverter
Electronic board «control logic»
Equipment back side view
Manual code: DQB26-EN [Rev. 02] Page 24
JOLLY PLUS Series TECHNICAL MANUAL
1.5.4 CAPACITORS BOX
Jolly 4 Plus
Nr. Description Q.ty
1 Discharge capacitors power charge push button 01
2 Fuse 690V 80A 01
3 Electrolytic capacitor 6800 µF 02
4 Rectifier bridge module 02
5 Resistance 50W 470R 01
1
4
5
Back side Front side
2
3
Manual code: DQB26-EN [Rev. 02] Page 25
JOLLY PLUS Series TECHNICAL MANUAL
Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus
Nr. Description Q.ty
1 Toroidal transformer single phase 1500 VA 01
2 Resistance 50W 470R 06
3 Rectifier bridge module 01
4 Electrolytic capacitor 6800 µF 16
5 Solid state relay 01
6 Discharge capacitors power charge push button 01
7 Panel fuse holder 10x38 01
8 Fuse 10x38F 32A 400V 01
9 Fuse 140A 660V 01
5
2
3
6
7-8
9
4
1
Back side Front side
Manual code: DQB26-EN [Rev. 02] Page 26
JOLLY PLUS Series TECHNICAL MANUAL
1.5.5 MONOBLOC
Nr. Code Description
1 E21A001 Monobloc for rotating anode X-ray tube 1 E21A002 Monobloc for fixed anode X-ray tube
1.6 LABEL
The following figure illustrates the position of the equipment’s adhesive label.
Jolly Plus Series
Equipment’s label
Manual code: DQB26-EN [Rev. 02] Page 27
JOLLY PLUS Series TECHNICAL MANUAL
1.7 DIMENSION
Here below the Jolly Plus, all models, overall dimensions:
Jolly 4 Plus Jolly 15 Plus/ 16 Plus / 30 Plus
1.8 REFERENCE STANDARD
The equipment complies with the following rules and directives:
EUROPEAN DIRECTIVE 93/42/EEC and 2007/47CE
CEI EN 60601-1
CEI EN 60601-1-2:
CEI EN 60601-1-3
CEI EN 60601-1-4
CEI EN 60601-2-7
CEI EN 60601-2-28
Medical electrical equipment. General requirements for safety
Electromagnetic compatibility - Requirements and tests General requirements for radiation protection in diagnostic X-ray equipment Programmable electrical medical systems
Particular requirements for the safety of high-voltage generators of diagnostic X-ray generators
Particular Requirements For The Safety Of X-Ray Source Assemblies And X-Ray Tube
Assemblies For Medical Diagnosis
CEI EN 60601-2-32
EN ISO 14971
Particular Requirements For The Safety Of Associated Equipment Of X-Ray Equipment
Application of risks management to medical devices
Directive concerning medical devices
Manual code: DQB26-EN [Rev. 02] Page 28
JOLLY PLUS Series TECHNICAL MANUAL
1.9 SAFETY
The equipments designed and manufactured by BMI Biomedical International are equipped with advanced radiological systems to safeguard safety and environment. The Jolly Plus equipments are not dangerous to people and environment if used in the way described in this Manual. Regarding the disposal of substances like the dielectric oil contained in the monobloc, the anti-X protective lead, the electronic boards and the electronic components, the provisions of the law in force both on the national territory and on the territory where the equipment is installed have to be complied with. BMI is a leading company in its field and constantly upgrades the equipment according to the progressive technological knowledge. Since this equipment emits X-rays, it must be exclusively used complying with the safety instructions included in this Manual and must not be used for other purposes than the diagnostic ones.
THE JURIDICAL BINDING VERSION OF THIS DOCUMENT IS THE ITALIAN LANGUAGE ONE. ANY TRANSLATION, EVEN IF VERIFIED AND APPROVED BY THE MANUFACTURER, IS ONLY SUPPLIED TO FACILITA TE THE USER.
SINCE THE EQUIPMENT EMITS X-RAYS, IT CAN ENDANGER THE OPERATOR AND THE PATIENT, IF THE EXPOSURE PARAMETERS ARE NOT RESPECTED
DANGER OF ELECTRIC SHOCK. ALWAYS CHECK THE CONDITION OF THE DIFFERENTIAL SWITCH AND OF THE EQUIPMENT SUPPLY CABLE
.
IT IS NECESSARY TO BE FAMILIAR WITH THE EQUIPMENT TO BE AWARE OF POSSIBLE DANGEROUS ANOMALIES. IN CASE OF ANOMALOUS OPERATION OR IF THERE IS ANY SAFETY PROBLEM UTILIZATION AND DO NOT USE IT UNTIL THE PROBLEM HAS BEEN SOLVED BY QUALIFIED PERSONNEL
THIS DEVICE IS CONNECTED WITH A PROTECTIVE GROUND DURING OPERATION. FOR THE PATIENTS AND THE OPERATORS SAFETY
, ONLY DEVICES COMPLYING WITH THE SAFETY STANDARDS, SUITABLE FOR MEDICAL DEVICES IN PERFECT OPERATING
CONDITIONS
, MUST BE USED
THIS EQUIPMENT MUST NOT BE USED IF INFLAMMABLE GASES ARE PRESENT.
SHOULD ANY ELECTROMAGNETIC INTERFERENCE INFLUENCE THE EQUIPMENT OR BE CAUSED BY IT, ASK FOR ASSISTANCE. THE PORTABLE RF EQUIPMENTS MAY ALTER THE ELECTROMEDICAL EQUIPMENTS
THIS SYSTEM HAS NOT BEEN DESIGNED TO WORK NEAR A DEFIBRILLATOR NEVER USE A DEFIBRILLATOR ON A PATIENT IN CONTACT WITH THE MOBILE RADI OGRAPHIC EQUIPMENT. THIS SYSTEM MUST BE HANDLED AS A CONDUCTIVE SURFACE AND POSITIONED AS FAR AS POSSIBLE FROM THE PATIENT BEFORE EFFECTING A DEFIBRILLATION TENSION CAN GO THROUGH THE PATIENT CONTACT WITH THE SYSTEM AND ALSO DAMAGE THE EQUIPMENT DEFIBRILLATOR USED ON A PATIENT BEING RADIOGRAPHED WITH THIS DEVI CE
. IF ANY PART OF THE DEVICE IS IN CONTACT WITH THE PATIENT WHEN USING THE DEFIBRILLATOR, THE
S BODY AND REACH THE SYSTEM. THIS CAN ENDANGER ANYBODY COMING INTO
, ISOLATE THE EQUIPMENT TO AVOID ANY NOT AUTHORI ZED
.
. ALWAYS REFER TO THE OPERATION INSTRUCTIONS FOR ANY
.
ONLY DOCTORS OR PEOPLE HAVING AN AUTHORIZATION ACCORDING TO THE LAW ARE AUTHORIZED BY THE LAW TO USE THIS EQUIPMENT
.
THERE ARE TWO KINDS OF EXPOSURE TO RADIATIONS WHEN USIN G THIS DEVICE: PROFESSIONAL, FOR THE OPERATORS, AND DIAGNOSTIC
, FOR THE PATIENTS.
Manual code: DQB26-EN [Rev. 02] Page 29
JOLLY PLUS Series TECHNICAL MANUAL
IT IS NECESSARY TO COMPLY WITH ALL SAFETY PRECAUTIONS AND RADIATIONS PROTECTION STANDARDS OF EACH SINGLE COUNTRY
During the performance of diagnostic examinations scattered radiations are always present. For this reason it is
, SINCE THE RADIATIONS HAVE A HARMFUL EFFECT ON THE HUMAN BODY:
absolutely necessary to use an X-ray protection. The operator is supposed to inquire if the patient is pregnant. Should the presence of a third person be required, this must also wear the anti-X protection, like the operator and the patient.
It is possible to have an additional protection against radiations if using a two-step push-button connected to the
control console by means of an extensible cable, featuring an extension of ~ 4 meters.
The incorrect utilization of Jolly Plus equipment can cause temporary or permanent injuries.
As a consequence, these instructions are to be carefully read and understood before putting the equipment at
work.
The manufacturer is ready to provide the after-sale servicing and, if required, to take care of the installation.
Even if this equipment gives a high level of protection against X-rays, no device can assure a complete
protection.
It is the operator’s task to take all the safety measures required to avoid any personal and other people’s risk,
coming from a wrong or excessive exposure to Ionogenic radiations
All the operators must receive a sufficient training and take the suitable safety measures to avoid any injury.
The equipment is sold with the following clause: the manufacturer, its agents and representatives decline all
responsibility for any damage or injury deriving from the wrong utilization of this equipment or due to lack of maintenance or to the utilization of not original spare parts.
It is highly recommended to use all the different protection devices available on the market.
The X-ray beam can go beyond the main protective shielding. Use radiographic cassettes or digital receptors with
leaded support.
Do not stand behind the receptor or on the X-ray beam line.
Only the patient can stay in the examination room during the X-ray exposure. If the operator or another person
presence is required, they must wear a leaded apron according to the safety rules
Use the greatest possible distance between the focal spot and the skin to give the patient the minimum dose.
Electrical safety: the buyer is supposed to use only qualified people (specialized engineers of the manufacturer or
equivalent personnel of a servicing company) to carry out any maintenance intervention on the equipment. The connections among the different parts of the device, the calibration and the testing must be carried out by qualified people, authorized by the manufacturer.
Only duly trained and qualified people have access to the internal parts of this device or to the disassembled
parts.
1.10 ELECTROMAGNETIC COMPATIBILITY (EMC)
This electronic equipment complies with EMC 89/336 Directive, defining the maximum emission levels allowed for the electronic equipment and the required immunity against electromagnetic field. This however does not exclude that signals coming from transmitters, like mobile phones or similar mobile radio transmitters, even if complying with EMC standards, can influence the correct operation of the Electro medical equipment if used nearby and with a quite high transmitting power. It is therefore necessary to avoid using these radio transmitters near electronically controlled medical systems, to exclude any interference. The electronic unit complying with EMC standards has been designed to avoid any risk of malfunctioning caused by electromagnetic interferences, when working in standard conditions. However, the radio signals coming from high frequency transmitters at a quite high transmitting power, if used near the electronic equipment, may cause electromagnetic incompatibility. Any transmission with mobile radio transmitter must be avoided. The mobile phones must be off in the area near the unit. These rules must be complied with when the unit is working. In the following pages, please find the EMC Certificate issued by the Italian notified body IMQ.
Manual code: DQB26-EN [Rev. 02] Page 30
JOLLY PLUS Series TECHNICAL MANUAL
Manual code: DQB26-EN [Rev. 02] Page 31
JOLLY PLUS Series TECHNICAL MANUAL
Manual code: DQB26-EN [Rev. 02] Page 32
JOLLY PLUS Series TECHNICAL MANUAL
Manual code: DQB26-EN [Rev. 02] Page 33
JOLLY PLUS Series TECHNICAL MANUAL
1.11 DECLARATION OF RESPONSIBILITY
BMI BIOMEDICAL INTERNATIONAL s.r.l. is responsible for the safety of its products only if their maintenance, repair
and/or modification are carried out by its own personnel or by expressly authorized personnel.
BMI BIOMEDICAL INTERNATIONAL does not assume any responsibility for any malfunctioning, damage and/or
hazard coming from an illegitimate use of the system or non compliance with the maintenance standards.
The PERSON RESPONSIBLE for the installation in which the system is working has the task to make sure that the
installation is used only by duly trained and qualified operators
CONFORMITY AND REFERENCE ADDRESS
Jolly 4 Plus, Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus equipments
complies with medical devices 93/42/EEC directive and 2007/47CE
The original copy of this declaration, as required by the law, is available at the manufacturer.
BMI BIOMEDICAL INTERNATIONAL s.r.l.
Via Enrico Fermi 52 Q/R
24035 Curno (Bergamo) Italy
Telephone: +39 035 4376381
Telefax: +39 035 4376401
Web site: www.bmibiomedical.it
E-Mail: info@bmibiomedical.it
Manual code: DQB26-EN [Rev. 02] Page 34
JOLLY PLUS Series TECHNICAL MANUAL
CHAPTER 2
INSTALLATION
Manual code: DQB26-EN [Rev. 02] Page 35
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 36
JOLLY PLUS Series TECHNICAL MANUAL
T
g
2.1 INTRODUCTION
The installation of the equipment must be carried out in three phases:
1. Unpacking and verification of the equipment and accessories integrity
2. Electric installation
3. Setting at work
The three phases are described in the following paragraphs.
THE EQUIPMENT IS CONFIGURATED AND TESTED IN THE FACTORY. SOME ADDITIONAL ADJUSTMENT COULD HOWEVER BE REQUIRED ON THE SPOT MANUAL
CHAPTER 4.
. THE WHOLE PARAMETERS ADJUSTMENT PROCEDURE IS DESCRIBED IN DETAIL IN THE TECH NICAL
2.2 UNPACKING AND VERIFICATION OF THE EQUIPMENT AND ACCESSORIES INTEGRITY
Preliminary Remarks
The equipment is wrapped in cellophane and then put inside a wooden or plywood case. Any special packing request must be made with the order. The equipment is secured by means of wooden crosspieces to avoid any movement during the transport. On the external side you can find the handling and storing instructions, the shipping address, the equipment model and serial number. Any particular request regarding the packing and the labelling must be specified with the order. The final customer is suggested to ask for special shockwatch labels to be applied on the external side of the packing case. When the customer receives the equipment, if the shockwatch labels have become red, this will mean that the equipment has been incorrectly handled during the transport or the storing, and has probably been damaged. For these reasons the customer must accept the equipment on condition, writing “…acceptance withheld for damaged packing (or similar words)…” on all copies of the transport documents accompanying the goods. Later the customer will open the case to check the equipment integrity and will inform the manufacturer about what happened and about the equipment condition. The customer must inform of:
Number of shipping document
Equipment model and serial number.
He will supply a detailed report with photographs illustrating the packing situation and the damages and/or
malfunctioning of the mechanical, electrical and electronic parts of the equipment, if any.
This communication must be sent within 2 days from the date of receipt of the equipment. After this delay the
supplier will not accept any responsibility for damages or defects caused by the transport.
Any request of returning the product back will have to be agreed up on with the manufacturer who must authorize it
in writing.
Unpacking of the equipment
Remove the case side marked «Open Here»
Take our the ramp from the case and position in on the ground
Remove the wooden crosspieces securing the equipment
Withdraw the equipment from the case using the above described board
Remove the cellophane
EXTERNALLY TO THE JOLLY PLUS CRATE BMI STICK ON THE DETAILED PROCEDURE TO CORRECTLY UNPACK THE EQUIPMENT.
Pay attention to the symbols printed outside the cardboard of the package:
his side up. Symbol in according to ISO 780.
Sensitive to moisture, keep away from the rain
and snow. Symbol in according to ISO 780.
ile. Symbol in according to ISO 780.
Fra
Do not use hand hooks. Symbol in according to
ISO 780.
Manual code: DQB26-EN [Rev. 02] Page 37
JOLLY PLUS Series TECHNICAL MANUAL
2.3 ELECTRIC INSTALLATION
Jolly Plus power supply is 230 Vac single-phase (± 10%) - 50 Hz - 16 A. The maximum current absorbed by the equipment is indicated both on the equipment label and on the technical data reported in the manuals supplied with the equipment. Make sure that the supply socket to which the supply cable is connected has 2 poles and the ground conductor and that it is homologated to deliver 230 Vac @ 16 A.
2.4 SETTING AT WORK
The setting at work is divided in three phases:
Electric inspection
Mechanical inspection
Operating test
The following three paragraphs give more details about the three above mentioned inspections:
2.4.1 ELECTRIC INSPECTION
Check visually each cable and each connector installed in the equipment to make sure that there is no damage that could jeopardize the correct operation of the equipment and the operator’s safety. Following components must be inspected:
Equipment supply cable and plug
X-ray control cable
Equipment / monobloc connecting cable
X-ray collimator supply cable
DAP interface cable (optional)
2.4.2 MECHANICAL INSPECTION
Check whether all the main parts of the equipment work properly, to avoid any defect that could cause malfunctioning or endanger the user or the patient. The mechanical inspections include the following operations:
Displacement of the equipment mobile parts in order to verify if the movements occur smoothly and noiseless.
Parking brakes testing to check if the equipment braking is quick and efficient.
Check if there is any break and /or dents and if any screw has become loose thus jeopardizing the equipment
operating and safety conditions.
The main controls are:
Rolling of the wheels to verify if there is any warpage or damage
X-ray equipment parking brake to check if the braking occurs properly.
Check the integrity of the control console and verify if the touch screen display has no dents or scratches
Integrity and fastening of the guards to verify if all parts screwed to the equipment structure are firmly fixed. Check
the arm balancing to verify that it keeps the required position when lifted (for example)
Manual code: DQB26-EN [Rev. 02] Page 38
JOLLY PLUS Series TECHNICAL MANUAL
CHAPTER 3
Utilization
Manual code: DQB26-EN [Rev. 02] Page 39
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 40
JOLLY PLUS Series TECHNICAL MANUAL
3.1 TRANSPORT OF THE EQUIPMENT
To displace the equipment the operator must:
Disconnect the unit supply cable from the socket and wind it on its support (see part c on the picture below)
Put the arm in parking position and make sure that it is correctly hooked to its retainer (see part f on the picture
below)
Push downwards the transport handles (see part b on the picture below) to unlock the parking brakes (see graphic
symbol on the picture below). To move the equipment keep the handle turned clockwise. The equipment is very handy, thanks to its low barycentre, its little weight and its pivoting front wheels. Should the operator accidentally let go the transport handle the provided dead-man mechanical brakes will immediately stop the unit. To overcome small obstacles easily (like for instance the lift door), hold tightly the transportation handle pressing your foot on the bracket
(see part d on the picture below) available on the rear side of the unit basement. The unit front side will lift, thus
permitting to overcome the obstacle
This picture shows how to lift up Jolly Plus to overcome little obstacles
ATTENTION! DO NOT PUSH THE EQUIPMENT FRONT SIDE AGAINST DOORS TO OPEN THEM
Following pictures illustrates Jolly Plus in transport position:
a
b
c
e
f
d
LEGEND
a. Rotation direction of the transportation handle to release the brakes b. Transport handle for parking brakes (dead-man brakes) control c. Supply cable support d. Bracket to lean the unit for overcoming little obstacles e. Arm f. Arm retainer
Manual code: DQB26-EN [Rev. 02] Page 41
JOLLY PLUS Series TECHNICAL MANUAL
3.2 EQUIPMENT’S SWITCHING ON
Connect the equipment supply cable cord plug to the socket.
Switch ON the equipment through the automatic circuit breaker (Pos. I), indicated with nr. 1 on the below reported picture.
Check on the console touch screen display if the first “Initializing” mask, indicated on the below reported picture, has
been visualized.
Note:
The mask is displayed for approximately 5 seconds.
The « INIZIALIZING…..» message is displayed only when the software language is set in English, otherwise this
message is written in one of the nine languages available.
When the « INIZIALIZING…..» message is over, the software automatically turns on “operator mode”.
Only in this situation it is possible to take the radiographic exposures, presetting the exposures parameters according to the procedures described in the following paragraphs.
1
ON (I)
OFF (0)
Automatic circuit breaker
Automatic circuit breaker
Ist “Initializing” mask II
nd
“Operator mode” mask
AFTER SWITCHING OFF, WAIT AT LEAST 30 SECONDS BEFORE SWITCHING ON THE EQUIPMENT
Manual code: DQB26-EN [Rev. 02] Page 42
JOLLY PLUS Series TECHNICAL MANUAL
3.3 ARM POSITIONING
Release the arm retainer (see part a on the picture below) and keep hold of the arm using the two available handles (see part b on the picture below) to prevent it from moving.
RELEASE THE ARM STANDING LATERALLY, IN ORDER TO AVOID GETTING HURT.
a
b
The picture below illustrates how to release the arm correctly:
1st step - release the arm, standing laterally 2nd step – using the handle, lift the arm
b
When the handle is released, the arm will remain in the acquired position thanks to a clutch device integrated in the
equipment.
Manual code: DQB26-EN [Rev. 02] Page 43
JOLLY PLUS Series TECHNICAL MANUAL
3.4 MONOBLOC AND COLLIMATOR DISPLACEMENT
The monobloc with the X-ray collimator assembly can rotate as described below:
Monobloc yoke support rotation: +90° / -90°
Monobloc rotation in the yoke: -57° / +167°
To position the monobloc / X-ray collimator assembly, use the handgrip (see part a on the picture below). The frontal inclination of the monobloc and the X-ray collimator is indicated by the goniometer placed on the front guard of the monobloc (see part b on the picture below).
b
a
Manual code: DQB26-EN [Rev. 02] Page 44
JOLLY PLUS Series TECHNICAL MANUAL
3.5 COLLIMATOR
OREWORD
F
Jolly Plus is supplied with a manual, square-field, single-layer, lightweight, compact X-ray collimator. The X-ray field is defined by two pairs of lead-lined shutters located near the exit window of the X-ray beam and by a cone (near the focus) limiting the X-ray beam to the larger field size. The two pairs of shutters move perpendicularly in the X-ray field while the shutter movements are manually controlled by two knobs available on the collimator front panel. The X-ray collimator can be rotated ± 90° around its vertical axis, just gripping the handle. The collimator rotation travel is stopped by a mechanical retainer. Before making a radiographic exposure set the shutters in the correct position, using the X-ray field light beam on the anatomic body part to be examined. Comply with the maximum admitted distance value between the monobloc and the patient’s body.
ETTING THE X-RAY FIELD ON THE COLLIMATOR FRONT PANEL
S
Press the X-ray field light beam lamp push- button (see part 1 on the picture below) to switch ON the halogen lamp.
The lamp automatically switches OFF after 30 seconds. To avoid the lamp overheating press again the lamp push button after 5 minutes.
Position the collimator /monobloc assembly above the patient.
Check if the X-ray light beam crosses is centered on the patient’s area to be examined.
Adjust the X-ray field size using the two knobs available on the front part of the X-ray collimator (see part 3 and part 4
on the picture below). The right knob adjusts the longitudinal shutters; the left one adjusts the transversal shutters. The knobs make it possible to the operator to set the X-ray field according to the radiographic cassette size. The collimator field size can vary from 0×0 cm minimum to 43×43 cm (17”x17”) maximum, at a focus film distance of 100 cm (40”).
The scale under each knob indicates the longitudinal and transversal field size at 100 cm (40”) distance.
Check the FFD using the retractable tape measure.
FOR FURTHER TECHNICAL INFORMATION PLEASE REFER TO THE COLLIMATOR MANUAL, ALWAYS SUPPLIED WITH THE EQUIPMENT
8
9
2
1
4 5 7
LEGEND
1. X-ray field light beam lamp push-button
2. Longitudinal diaphragm adjustment knob
3. Transversal diaphragm adjustment knob
4. Rails guides for accessories
5. DAP - Kermax plus IDP (optional)
6. Retractable tape measure for FFD measuring
3
4
6
7. Handgrip for ± 90° rotation on its vertical axis
8. Goniometer
9. Monobloc focus external reference point
Manual code: DQB26-EN [Rev. 02] Page 45
JOLLY PLUS Series TECHNICAL MANUAL
3.6 RADIOGRAPHIC TECHNIQUES
3.6.1 TWO-POINT MANUAL TECHNIQUE RADIOGRAPHIC EXPOSURE
On the “operator mode” mask touch the icon «MAN», detail nr 1 on the below reported picture.
On the “operator mode” mask touch exclusively the fields «kV» and «mAs», respectively detail nr 2 and detail nr 3 on
the below reported picture.
To adjust the kV or the mAs values touch the arrow icon, detail nr 4 on the below reported picture.
After having adjusted the kV or the mAs values the equipment microprocessor’s software verifies if the values inserted
are compatible with the equipment monobloc model performances, to avoid any failure.
If the kV or the mAs value inserted results not be compatible, on the “operator mode” lights up the warning icon
“X-ray tube alarm”, detail nr 5 on the below reported picture.
To take a radiographic exposure please refer to § 3.7
5
1
2
4
3
“Operator mode” mask
3.6.2 THREE-POINT MANUAL TECHNIQUE RADIOGRAPHIC EXPOSURE
On the “operator mode” mask touch the icon «MAN», detail nr 1 on the below reported picture.
On the “operator mode” mask touch exclusively the fields «kV», «mA» and «ms», detail nr 2, nr 3 and nr 4 on the
below reported picture.
To adjust the kV, the mA or the time (msec) value touch the arrow icon detail nr 5 on the below reported picture.
After having adjusted the «kV», «mA» and «ms» values the equipment microprocessor’s software verifies if the values
inserted are compatible with the equipment monobloc model performances, to avoid any failure.
If the kV, the mA or the or the time (msec) value inserted results not be compatible, on the “operator mode” lights up
the warning icon “X-ray tube alarm”, detail nr 6 on the below reported picture.
To take a radiographic exposure please refer to § 3.7
6
1
2
3
5
4
“Operator mode” mask
Manual code: DQB26-EN [Rev. 02] Page 46
JOLLY PLUS Series TECHNICAL MANUAL
3.6.3 APR RADIOGRAPHIC EXPOUSURE
The default APR techniques, associated with the anatomical parameters in the APR menu, are 170. For each technique the software works on kV and mAs values. These values must be considered as reference values to be verified before taking radiographic exposures.
The operator can also adjust the parameters stored in the APR, editing the values as described in § 3.6.5.
To select the desired APR anatomical, proceed as following:
«APR» icon to enable the technique (detail nr 1 on below reported left side mask)
«Anatomic Parts» icon to select the patient’s anatomical part to be examined, scrolling the drop-down menu (detail nr
2 on below reported left side mask)
«Thickness» icon to preset the patient’s size: small, medium, large, lateral and neonatal (detail nr 3 on below
reported left side mask) On § 3.6.4 has been reported the APR anatomical programming techniques chart. Now the equipment is ready to take a radiographic exposure, as described on § 3.7.
1 2
3
“APR mode” mask
Manual code: DQB26-EN [Rev. 02] Page 47
JOLLY PLUS Series TECHNICAL MANUAL
3.6.4 APR - ANATOMICAL PROGRAMMING CHART
ANATOMIC TECHNIQUES THICKNESS F.F.D.
SMALL MEDIUM LARGE LATERAL NEONATAL
HEAD
kV mAs kV mAs kV mAs kV mAs kV
mAs
Skull 74 12,5 77 12,5 80 12,5 85 20 64 12,5 1 m / 40”
Paranasal sinuses 62 8 65 8 68 8 - - 52 8 1 m / 40”
Jaw - - - - - - 60 12,5 - - 1 m / 40” Nasal bones - - - - - - 55 2 - - 1 m / 40”
CHEST
Lung 69 12,5 73 12,5 77 12,5 85 15 58 12,5 1,8 m / 72” Hemithorax ribs 57 25 60 25 63 25 - - 48 25 1,8 m / 72” Sternum 57 25 60 25 63 25 85 20 48 25 1 m / 40”
HAND
Wrist 52 5 55 5 58 5 65 5 42 5 1 m / 40” Finger 52 3 55 3 58 3 60 4 42 3 1 m / 40”
KNEE
Patella 63 6,3 67 6,3 70 6,3 65 5 52 6,3 1 m / 40” Intercondylar fossa 63 5 65 5 68 5 - - 52 5 1 m / 40”
FOOT
Heel bone 66 4 70 4 74 4 65 5 55 4 1 m / 40” Heel bone axial projection 66 6,3 70 6,3 74 6,3 - - 55 6,3 1 m / 40” Toes 52 3 55 3 58 3 60 5 42 3 1 m / 40” Toes oblique 52 5 55 5 57 5 - - 42 5 1 m / 40”
FOREARM
Radius 57 6,3 60 6,3 63 6,3 65 6,3 46 6,3 1m / 40” Ulna 57 6,3 60 6,3 63 6,3 65 6,3 46 6,3 1m / 40”
LEG
Femur 59 30 62 30 65 30 80 20 48 30 1m / 40” Tibia - fibula 73 5 77 5 81 5 65 5 62 5 1 m / 40” Ankle 75 3 79 3 83 3 60 5 63 3 1 m / 40”
PELVIS
Sacroiliac 64 10 67 10 70 10 - - 52 10 1m / 40” Sacroiliac oblique 63 30 65 30 68 30 - - 51 30 1m / 40” Pubic symphysis 62 20 65 20 68 20 - - 52 20 1m / 40” Hip – neck femur 59 30 62 30 65 30 - - 48 30 1m / 40” Axial Hip – neck femur 63 25 65 25 68 25 - - 51 25 1m / 40”
UPPER LIMB
Elbow 60 6,3 63 6,3 66 6,3 70 6,3 50 6,3 1m / 40” Humerus 60 12,5 63 12,5 66 12,5 - - 50 12,5 1m / 40”
VERTEBRAL SPINE
Cervical vertebrae 57 25 60 25 63 25 85 10 46 25 Axis 57 10 60 10 63 10 - - 46 10 Thoracic vertebrae 71 30 75 30 79 30 85 20 60 30 Lumbar vertebrae 62 25 65 25 68 25 85 30 50 25 Sacrum coccyx 68 30 72 30 76 30 85 30 56 30
SHOULDER
1,8 m / 72” 1,8 m / 72” 1,8 m / 72” 1,8 m / 72” 1,8 m / 72”
Clavicle 60 20 63 20 66 20 - - 50 20 1 m / 40” Shoulder blade 67 8 71 8 75 8 80 15 56 8 1 m / 40”
Manual code: DQB26-EN [Rev. 02] Page 48
JOLLY PLUS Series TECHNICAL MANUAL
3.6.5 MAKING CHANGES TO PRE-PROGRAMMED APR TECHNIQUES
The default APR techniques associated with the anatomical parameters in the APR display window could be adjusted by the operator, as follows:
Switch ON the equipment.
Enable on the operator mode the “APR mode”, touching the icon «APR» (detail nr 1 on below reported mask).
Select from Anatomic Parts drop-down menu (detail nr 2 on below reported mask) the anatomic techniques to be
adjusted.
Select from Thickness drop-down menu (detail nr 3 on below reported mask) the patient’s body size to be adjusted.
The sizes available are: small, medium, large, lateral and neonatal.
Adjust the kV value (detail nr 4 on below reported mask), increasing or decreasing the data visualized, touching one
of two arrow icons, respectively marked with nr 6 - to increase and nr. 7 to decrease the data, on below reported
mask.
To store the new adjusted value touch the icon «Save» (detail nr 8 on below reported mask).
Adjust the mAs value (detail nr. 5 on below reported mask), increasing or decreasing the data visualized, touching one
of two arrow icons, respectively marked with nr 6 - to increase and nr. 7 to decrease the data, on below reported
mask
To store the new adjusted value touch the icon «Save» (detail nr 8 on below reported mask).
Switch OFF the equipment.
Switch ON the equipment, after at least 30 seconds.
1
2
3
4
5
6 7 8
“APR mode” mask
Manual code: DQB26-EN [Rev. 02] Page 49
JOLLY PLUS Series TECHNICAL MANUAL
3.7 TAKING AN RADIOGRAPHIC EXPOSURE
Place Jolly Plus collimator on the patient’s area to be exanimate, placing the device respecting the F.F.D..
Place the radiographic cassette below the patient’s area to be delaminated.
Set the collimator X-ray field area (additional information are reported on the device manual) using its light beam, that
allows the operator to obtain on the patient body a the field marker. To switch ON the light beam depress the ON/OFF
button, which is located on the control desk panel and on the collimator itself. Refer also to § 1.4.2 and § 3.5.
Adjust the size of the field as described on § 3.5.
Set the exposure parameters to take the radiography exposure, as described in chapter 3 - § 3.6.1, § 3.6.2 and §
3.6.3.
The operator can take radiographic exposures from the Jolly Plus console, by means of two buttons or by means the
external hand switch.
To take a radiographic exposure from the control console:
Depress keeping hold the preparation button (1
On the operator mode mask the icon
Depress the exposure button (2
nd
time) in order to take the radiographic exposure
lights up immediately
st
time) in order to spin the X-ray tube rotor.
On the operator mode mask the icon
lights up immediately emitting also a warning sound (buzzer sound) to
advice the operator that the radiographic exposure has been taken.
To take a radiographic exposure using the hand switch remote control:
Depress halfway the preparation button (1
On the operator mode mask the icon
Depress completely the button (2
nd
time) in order to take the radiographic exposure. See figure below reported.
st
time) in order to spin the X-ray tube rotor. See figure below reported.
lights up immediately
Control console
Exposure button
IInd time - X-ray exposure
Preparation button
Ist time - preparation
On the operator mode mask the icon
lights up immediately emitting also a warning sound (buzzer sound) to
advice the operator that the radiographic exposure has been taken.
Hand switch remote control
Manual code: DQB26-EN [Rev. 02] Page 50
JOLLY PLUS Series TECHNICAL MANUAL
When the diagnostic examination is over, switch OFF Jolly Plus by means of the automatic circuit breaker (refer to § 3.2) and then disconnect the equipment supply cable cord from socket. Set Jolly Plus in transport position (refer to § 3.1) and park it in a suitable and safety place.
BEFORE TAKING ANY RADIOGRAPHIC EXPOSURE MAKES SURE THAT ALL NECESSARY PRECAUTIONS AGAINST IONOGENIC RADIATIONS HAVE BEEN TAKEN IMPORTANT TO MAKE THE FORMATION OF THE X THE X-RAY TUBE FORMATION PROCEDURE IS DESCRIBED IN PARAGRAPH 4.1.1.
. AFTER A LONG PERIOD OF INACTIVITY OF THE EQUIPMENT (TWO MONTHS) IT IS VERY
-RAY TUBE, TO AVOID ANY H.T. DISCHARGE THAT COULD DAMAGE IT.
WHEN THE X-RAY EMISSION IS OVER, THE RELEVANT WARNING LIGHT ON THE CONTROL CONSOLE DISPLAY GOES OUT AND AT THE SAME TIME THE SOUND SIGNAL STOPS AT THIS POINT THE TWO-STEP X-RAY CONTROL PUSH-BUTTON CAN BE RELEASED OR, IF THE EXPOSURE HAS BEEN CONTROLLED BY THE CONSOLE PUSH
BEFORE TAKING AN X-RAY EXPOSURE PROTECT THE OPERATOR AND THE PATIENT FROM X-RAYS USING THE SUITABLE LEAD PROTECTIONS
.
3.8 END OF OPERATION
Switch OFF Jolly Plus by means of automatic circuit breaker.
Disconnect the supply cable cord from the socket and wind it on its support.
Position the equipment arm in transport position and hook the arm to the stand by means of the arm retainer.
DO NOT SWITCH OFF THE EQUIPMENT BY REMOVING THE SUPPLY CABLE FROM THE SOCKET BEFORE DISCONNEC TING THE AUTOMATIC CIRCUIT BREAKER
.
.
-BUTTONS, BOTH OF THEM CAN BE RELEASED.
Manual code: DQB26-EN [Rev. 02] Page 51
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 52
JOLLY PLUS Series TECHNICAL MANUAL
CHAPTER 4
Adjustments & Setup
Manual code: DQB26-EN [Rev. 02] Page 53
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 54
JOLLY PLUS Series TECHNICAL MANUAL
4.1 ADJUSTMENTS
4.1.1 X-RAY TUBE FORMATION
This procedure is required only when the Jolly Plus has not been not used continuously for 2 months after the installation date or if the equipment has not been used continuously for more than 1 month in the Hospital department. Forming the monobloc’s X-ray tube prevents it from damages. The X-ray tube formation procedure consists in making radiographic exposures, according to below reported chart. For each Jolly lus model all the radiographic exposures should be taken using only the X-ray tube large focus. BMI suggest taking all the radiographic exposures closing the entire X-ray collimator leaded shutter. The X-ray tube formation procedure will take approximately one hour.
THE IONOGENIC RADIATIONS PRODUCED BY THIS EQUIPMENT CAN BE DANGEROUS TO THE PATIENT AND THE OPERATOR. STRICTLY COMPLY WITH THE SAFETY RULES IN FORCE
REMEMBER TO PRESET THE LARGE FOCUS FOR ALL RADIOGRAPHIC EXPOSURES AND TO SHUT THE X-RAY COLLIMATOR LEAD BLADES TO INCREASE THE SAFETY LEVEL AGAINST IONOGENIC RADIATIO NS
JOLLY 4 PLUS
PARAMETER NUMBER OF
kV mAs
EXPOSURES
INTERVAL TIME
BETWEEN EXPOSURES
40 6.3 3 20 sec
Pause 120 sec
50 6.3 3 20 sec
Pause 120 sec
60 6.3 3 20 sec
Pause 120 sec
70 6.3 3 20 sec
Pause 120 sec
80 6.3 3 20 sec
Pause 120 sec
90 6.3 3 20 sec
Pause 120 sec
100 6.3 3 20 sec
Pause 120 sec
110 6.3 3 120 sec
Pause 120 sec
40 20 1 120 sec 50 20 1 120 sec 60 20 1 120 sec 70 20 1 120 sec 80 20 1 120 sec
Manual code: DQB26-EN [Rev. 02] Page 55
JOLLY PLUS Series TECHNICAL MANUAL
JOLLY 15 PLUS / 16 PLUS / 30 PLUS
PARAMETER NUMBER OF
kV mAs
EXPOSURES
INTERVAL TIME
BETWEEN EXPOSURES
40 6.3 3 20 sec
Pause 120 sec
50 6.3 3 20 sec
Pause 120 sec
60 6.3 3 20 sec
Pause 120 sec
70 6.3 3 20 sec
Pause 120 sec
80 6.3 3 20 sec
Pause 120 sec
90 6.3 3 20 sec
Pause 120 sec
100 6.3 3 20 sec
Pause 120 sec
110 6.3 3 20 sec
Pause 120 sec
120 6.3 3 120 sec
Pause 120 sec
125 6.3 3 120 sec
Pause 120 sec
40 20 1 120 sec 50 20 1 120 sec 60 20 1 120 sec 70 20 1 120 sec 80 20 1 120 sec
If no anomalies are found, Jolly Plus is now ready to be used.
Manual code: DQB26-EN [Rev. 02] Page 56
JOLLY PLUS Series TECHNICAL MANUAL
4.1.2 X-RAY FIELD CENTERING
Place a radiographic cassette size 35x43 cm (14”x17”) on a radiographic table or any horizontal plane surface.
Position the monobloc/X-ray collimator assembly at 1 m distance above the cassette with the X-ray beam perpen-
dicular to it. The distance can be measured by means of the retractable meter mounted in the collimator.
Do not use the system indicators as reference, but measure the actual distance from the focus to the cassette surface.
Center the cassette in the X-ray field using the collimator light.
The switching on key is available both on the unit console touch screen display and on the collimator front panel. See
picture A, detail 1.
Mark the position of the cassette with adhesive tape so that it can be removed and placed again in the same position:
Put a white paper on the cassette to get the luminous field greatest contrast.
Open the collimator blades to get a 30x30 cm (14”x14”) field at 1 m distance, using the two knobs shown in
picture A, details 2a and 2B.
Switch on the light field and use it to position 18 coins as illustrated in the following drawing.
Position each pair of coins in such a way as they touch each other; the coin inside the luminous field must be
lighted as much as possible while the coin outside the luminous field must be lighted as little as possible. The tangential points define the sides of the luminous field. The other two coins will be positioned one in a corner to orientate the film and the other in the point of intersection of the diagonals, to define the film center.
Coin
Preset approx. 50 kVp and 5 mAs on the unit console touch screen display
Make a radiographic exposure and after developing the film, use the test to check the alignments described
below
ALIGNMENT OF THE COLLIMATOR WITH THE FOCAL POINT (PRIMARY BLADES CUT
Observe the projections of the four collimator blades composing the radiographic field edge. If the edge is not well
)
defined, this means that the radiographic field is cut by the primary blade, near the focal spot, and not by the external one. To modify this condition, use the security dowels displacing the collimator towards the side presenting a not defined edge. After the correction, repeat the exposure to check the condition.
ALLIGNMENT OF THE LUMINOUS FIELD WITH THE RADIOGRAPHIC FIELD
(according to standard EN 60601-13 paragraph 29.202.9)
The max. allowed tolerance between the luminous and the radiographic field, in both sides, must not exceed 2% of the FOCUS/FILM DISTANCE. In the case under consideration it would be less than 20 mm. If the test film shows the light field congruence (indicated by the coins shadows) with the radiographic field (given by the collimator blades shadows) within the diameter of a coin having a diameter smaller than 20 mm, the resulting alignment complies with the standards. As a matter of fact, it is possible to obtain a greater congruence degree. The max. Recommended deviation is one fourth. It is very important to adjust the field in order to get the best possible congruence degree. If the radiographic field, on the four sides, is not correct, check if the number of spacers is correct, that is, the distance between the focal point and the collimator mounting surface must be 80mm (± 1 mm). Should this distance require any adjustment, repeat the test exposure after the adjustment. Should the distance be correct but anyhow requiring an adjustment, do not move the collimator with respect to the radiographic tube, but act as follows: Position the test film on a white paper and put it on the cassette surface. Place the cassette in the position previously marked. Check the film orientation using the coins shadows.
Using the collimator blades images as radiographic field shape and dimension reference, adjust the light field to get the congruence with the radiographic field. Should a field adjustment be required, work on the transversal, vertical and longitudinal displacement of the light field.
Manual code: DQB26-EN [Rev. 02] Page 57
JOLLY PLUS Series TECHNICAL MANUAL
ALIGNMENT OF THE LUMINOUS FIELD WITH THE RADIOGRAPHIC FIELD
(According to standard EN 60601-1-3 par 29.202.9)
The max. allowed tolerance between the luminous and the radiographic field must not exceed 2% of the focus/film distance. In this case it would be less than 20 mm. If the test film shows the luminous field congruence (indicated by the coins shadows) with the radiographic field (given by the collimator blades shadow) within the diameter of a coin having a diameter smaller than 20 mm, the resulting alignment complies with the standards. As a matter of fact it is possible to get a much better congruence degree. The max. Recommended deviation is one fourth. It is very important to adjust the field in order to get the best con­gruence degree. If the radiation field, on the four sides, is not correct, check if the number of spacers is correct, that is, if the distance between the focal point and the collimator mounting surface is 80 mm, (± 1 mm). Should this distance require any adjustment, repeat the test exposure after the adjustment. If the distance is correct but an adjustment is required all the same, repeat the test exposure after adjustment. If the distance is correct but an adjustment is however required, do not displace the collimator with respect to the radiographic tube but proceed as follows:
Position the test film on a white paper and put it on the cassette surface.
Place the cassette in the previously marked position.
Check the film orientation using the coins shadows.
Using the collimator blades images as radiographic field shape and dimension reference, adjust the luminous field to
get it congruent to the radiographic field.
If a field adjustment is required, work on the transversal, vertical and longitudinal displacement of the luminous field.
S
HOULD IT BE NECESSARY TO MOVE THE LUMINOUS FIELD OF THE TRANSVERSAL OR VERTICAL COLLIMATOR, PLEASE REFER TO THE
MANUAL SUPPLIED WITH THE COLLIMATOR
2b 2a
1
Picture A
LEGEND
1 X-ray field light beam lamp push-button 2a Right knob to open or close the transversal shutter 2b Left knob to open or close the longitudinal shutter
Manual code: DQB26-EN [Rev. 02] Page 58
JOLLY PLUS Series TECHNICAL MANUAL
4.1.3 kV VALUES ADJUSTMENT
DO NOT CARRY OUT THE ADJUSTMENT WHEN THE M O NOB L OC IS COLD.
EMEMBER TO EFFECT A PREHEATING CYCLE MAKING AT LEAST TWO OR BETTER FIVE EXPOSURES, PRESETTING ON OPERATOR
R
MODE THE MINIMUM
mA AND kV VALUES
Measuring instrument to be used: non-invasive kilovoltmeter. The manufacturer recommends: RTI - PMX-I/R PIRANHA 507. Procedure:
Place the non-invasive kilovometer instrument on a flat and horizontal surface under the X-ray collimator.
Position the non-invasive kilovometer instrument at 50 cm from the X-ray collimator, using its retractable meter.
Align the collimator perpendicular to the non-invasive kilovometer instrument measuring surface using the collimator
light beam.
Take some radiographic exposures, setting in the “operator mode” the following parameters, according to the Jolly
Plus model to be adjusted:
Jolly 4 Plus 100 kV / 40 mA @ 100 msec
Jolly 15 Plus 100 kV / 100 mA @ 100 msec
Jolly 16 Plus 100 kV / 100 mA @ 100 msec
Jolly 30 Plus 100 kV / 200 mA @ 100 msec
For each radiographic exposure taken, verify that the kV value measured by the non-invasive instrument corresponds
to the reference value above written. The admitted kV measuring allowance is ± 5%.
To eventually adjust the kV value measured, the technician can adjust on the inverter electronic board «control logic
board» the trimmer nr. R117.
R117
Inverter electronic board «control logic»
Manual code: DQB26-EN [Rev. 02] Page 59
JOLLY PLUS Series TECHNICAL MANUAL
T
4.1.4 mA VALUES ADJUSTMENT
DO NOT CARRY OUT THE ADJUSTMENT WHEN THE M O NOB L OC IS COLD.
EMEMBER TO EFFECT A PREHEATING CYCLE MAKING AT LEAST TWO OR BETTER FIVE EXPOSURES, PRESETTING ON OPERATOR
R
MODE THE MINIMUM
mA AND kV VALUES
Measuring instrument to be used: Digital oscilloscope. mA value adjustment procedure:
Switch OFF the equipment.
On the Inverter electronic board «control logic» connect the oscilloscope’s probe (channel 01), in order to measure
the mA feedback signal wave form, to: TP18 signal / TP1 Ground.
P 1(Ground)
TP 18
Set on the oscilloscope the signal acquisition scale in order to measure correctly the time base:
Inverter electronic board «control logic»
20 msec / div., amplitude 1V / div.
Switch ON the equipment, entering in “operator mode”.
Manual code: DQB26-EN [Rev. 02] Page 60
JOLLY PLUS Series TECHNICAL MANUAL
Take 6 radiographic exposures presetting the parameters, according to the Jolly Plus to be adjusted, as written in the
below reported charts:
Jolly 4 Plus
kV mA Exposures time
60 25 100 msec 60 50 100 msec 80 25 100 msec 80 40 100 msec 100 25 100 msec 100 40 100 msec
Jolly 15 Plus Jolly 16 Plus
kV mA Exposures time kV mA Exposures time
60 50 100 msec 60 50 100 msec 60 200 100 msec 60 200 100 msec 80 50 100 msec 80 50 100 msec
80 150 100 msec 80 160 100 msec 100 50 100 msec 100 50 100 msec 100 100 100 msec 100 100 100 msec
Jolly 30 Plus
kV mA Exposures time
60 50 100 msec
60 250 100 msec
80 50 100 msec
80 200 100 msec 100 50 100 msec 100 160 100 msec
After having taken each single radiographic exposure, verify on the oscilloscope display that the mA wave form is
displayed as below reported, for example:
Channel 01
mA feedback wave form
(Measurement on Jolly 30 Plus @ 200mA)
The ratio between the measured mA on the oscilloscope and the Volt is:
Jolly 4 Plus 1 Volt = 20 mA
Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus 1 Volt = 100 mA
The mA value measured by the oscilloscope should be comprised in an allowance range of ± 10%, compared to ones
reported in the above reported charts (equipment model / kV / mA / exposure time). In case of mA value out of the allowance the technician should calibrate the mA range, as described in § 4.2.5.
ONLY WHEN THE MONOBLOC HAS BEEN WARMED UP, THE MA VALUES MEASURED SHOULD BE CONSIDERABLE RELIABLE.
Manual code: DQB26-EN [Rev. 02] Page 61
JOLLY PLUS Series TECHNICAL MANUAL
4.2 SET UP AND CALIBRATION
The equipment Set Up menu it is composed by four sub-menus:
st
Mask: Language and options preselecting. Refer respectively to § 4.2.1 and § 4.2.2.
1
nd
Mask: Counter. Refer to § 4.2.3.
2
rd
Mask: kV values calibration and the small / large focus pre-lighting calibration. Refer to § 4.2.4.
3
th
Mask: mA values calibration. Refer to § 4.2.5.
4
To enter in the Set Up menu, proceed as follows:
Switch ON the equipment
Press «Setup» button, indicated by letter A on the mask below reported, on the console touch screen display
A
Starting mask
Type the password
(TO BE REQUESTED TO BMI)
on the console touch screen display and then press the icon
«Ent», marked with letter B in the mask below reported, to confirm the data
B
Password mask
To enter in the sub menu of Set Up press the two arrows icons, pointed out in pictured below reported with letter C.
Left arrow icon is used to go back in the sub-menu meanwhile the right arrow icon is used to go forward in the sub-menu.
C
st
mask - «Set Up» menu
I
Manual code: DQB26-EN [Rev. 02] Page 62
JOLLY PLUS Series TECHNICAL MANUAL
4.2.1 LANGUAGE PRESELECTION
Select from the «Language» drop-down menu the language to be set, pressing the icon indicated by letter A on the
mask below reported
To confirm the preselection and to exit from this menu press the icon «Finish», indicated by letter B on the mask
below reported
Wait till the equipment return again on the operator modality
Switch OFF the equipment
Switch ON the equipment
A
B
st
mask - «Set Up» menu
I
4.2.2 OPTIONS PRESELECTION
In the first Set Up mask the operator can enable or disable 6 options. Four options, Potter, Starter, AEC and Real mAs are represented by icons in which an illuminable spot indicates the preselection status. Here below the option explanation:
1. Touch Calib.: It allows the technician to re-calibrate the touch screen display.
Refer to picture below listed - icon nr. 1.
2. Reboot: It allows the rebooting of the Jolly Plus software. It should be used only in case of software updating
Refer to picture below listed - icon nr. 2.
3. Starter: It enables the X-ray tube anode rotation.
Refer to picture below listed - icon nr. 3.
4. Real mAs button: It enables on operator mode the visualizing of real mAs values.
Refer to picture below listed - icon nr. 4.
5. Potter: It enables the potter bucky interface (optional).
Refer to picture below listed - icon nr. 5.
6. AEC: It enables the AEC interface - automatic exposure control (optional).
Refer to picture below listed - icon nr. 6.
5
3
4
To confirm the preselection and to exit from this menu press the icon «Finish», indicated by letter A on the mask
6
1
2
A
above reported
Wait till the equipment return again on the “operator mode”.
Switch OFF the equipment.
Switch ON the equipment.
Manual code: DQB26-EN [Rev. 02] Page 63
JOLLY PLUS Series TECHNICAL MANUAL
4.2.3 COUNTER
This mask visualizes the total and partial numbers of radiographic exposures done and the heating unit thermal capacity generated by the monobloc. It is possible to reset the partial counters in case of monobloc replacement.
Manual code: DQB26-EN [Rev. 02] Page 64
JOLLY PLUS Series TECHNICAL MANUAL
T
4.2.4 kV VALUES CALIBRATION
DO NOT CARRY OUT THE ADJUSTMENT WHEN THE M O NOB L OC IS COLD.
EMEMBER TO EFFECT A PREHEATING CYCLE MAKING AT LEAST TWO OR BETTER FIVE EXPOSURES, PRESETTING ON OPERATOR
R
MODE THE MINIMUM
mA AND kV VALUES
Measuring instrument to be used: Digital multimeter. kV value adjustment procedure:
Switch OFF the equipment.
On the Inverter electronic board «control logic» using the digital multimeter’s hook probes, measure the «kV Vref»
between the test points TP28 (Vdc signal) and TP1 (Ground).
P 1(Ground)
TP 28
Switch ON the equipment, entering in “Set Up mode” as described in the § 4.2.
Enter in the 3
rd
Set Up sub-menu. Here below is shown the mask:
Inverter electronic board «control logic»
Manual code: DQB26-EN [Rev. 02] Page 65
JOLLY PLUS Series TECHNICAL MANUAL
With reference to the Set Up mask, shown below:
Select the field «I’ STEP KV Calib.» pressing in it.
Adjust the value displayed in the field, in order to measure, with the digital multimeter, 2.05 Vdc.
To increase or to decrease the value press the two arrows icons pointed out in the below reported mask.
Value to be adjusted
Arrows icons
Select the field «II’ STEP KV Calib.» pressing in it.
Adjust the value displayed in the field, in order to measure, with the digital multimeter, 5.5 Vdc.
To increase or to decrease the value press the two arrows icons pointed out in the below reported mask .
Value to be adjusted
Arrows icons
«WARM SF CALIB» AND «WARM LF» FIELD PARAMETER IS SET IN FACTORY.
HE DEFAULT VALUE IS 1050.
T
To terminate the calibration procedure press the icon «Finish» pointed out in the below reported mask
Wait till the software returns in “operator mode”.
Switch OFF the equipment and after at least 30 second switch it ON.
Finish icon
Manual code: DQB26-EN [Rev. 02] Page 66
JOLLY PLUS Series TECHNICAL MANUAL
T
4.2.5 mA VALUES CALIBRATION
The mA values calibration is requested only when:
Jolly Plus has been installed on site for the first time.
Every 12 months as ordinary maintenance.
The monobloc has been replaced.
DO NOT CARRY OUT THE mA VALUES CALIBRATION WHEN THE MONOBLOC IS COLD.
EMEMBER TO EFFECT A PREHEATING CYCLE MAKING AT LEAST TWO OR BETTER FIVE EXPOSURES, PRESETTING ON OPERATOR
R
MODE THE MINIMUM
mA AND kV EXPOSURE VALUES.
Measuring instrument to be used: Digital oscilloscope. mA value adjustment procedure:
Switch OFF the equipment.
On the Inverter electronic board «control logic» connect to the below listed test points two oscilloscope’s probes, in
order to measure each signal wave form.
- Probe nr. 01 (channel 01): mA feedback signal TP18 signal / TP1 Ground.
- Probe nr. 02 (channel 02): kV feedback signal TP19 signal / TP1 Ground.
P 1(Ground)
TP 19
TP 18
Set on the oscilloscope the signal acquisition scale in order to measure correctly the time base:
Inverter electronic board «control logic»
10 msec / div., amplitude 1V / div.
Switch ON the equipment, entering in “Set Up mode” as described in the § 4.2.
Manual code: DQB26-EN [Rev. 02] Page 67
JOLLY PLUS Series TECHNICAL MANUAL
Enter in the menu «Calibration Gen…», here below is represented the menu mask:
The mA values to be calibrated are listed on field 1, marked in the below reported picture.
mA value to set
Field 1
This button, drop-down menu, does not work
On next page has been reported the reference mA measures calibration values numbers charts for each Jolly Plus
model.
To set the mA measure number, touch the related measures list displayed on field 1, marked in the below reported
pictures. To change the meausure number uses the two arrow keys, marked with nr. 2 on the below reported picture. To confirm the selection press the «Enter», marked with nr. 3 on the below reported picture.
1
This button, drop-down menu, does not work
2
Note:
In the «Calibration Gen…» menu the initials SF and LF means respectively the small and the large focus of the X-ray tube.
3
Manual code: DQB26-EN [Rev. 02] Page 68
JOLLY PLUS Series TECHNICAL MANUAL
mA reference values calibration charts for all Jolly Plus models:
JOLLY 4 PLUS
kV mA
40 25 40 40 40 63
40 80 100 25 100 40
1.5 mm
LARGE FOCUS
Measure 1 Measure 2 Measure 3 Measure 4 Measure 5 Measure 6
JOLLY 15 PLUS JOLLY 15 PLUS
kV mA kV mA
Measure 1 60 25 Measure 2 60 50 Measure 8 Measure 3 60 100 Measure 9 Measure 4 60 200 Measure 10
LARGE
FOCUS
SMALL
FOCUS
Measure 5 110 25 Measure 11 Measure 6 110 50 Measure 12
Measure 7
60 25 60 50
60 100 110 25 110 50 110 100
JOLLY 16 PLUS JOLLY 16 PLUS
kV mA kV mA
Measure 1 Measure 2 Measure 3 Measure 4
LARGE
FOCUS
Measure 5 Measure 6
Measure 13
60 25 60 50 60 100
60 200 110 25 110 50
Measure 7
Measure 8
SMALL
FOCUS
Measure 9 Measure 10 Measure 11 Measure 12
60 60 60
60 110 110 110
25
50 100 200
25
50 100
JOLLY 30 PLUS JOLLY 30 PLUS kV mA kV mA
Measure 1 Measure 2 Measure 3 Measure 4 Measure 5
LARGE
FOCUS
Measure 6 Measure 7 Measure 8
60 50 60 100 60 200 60 300
60 350 110 50 110 100 110 200
Measure 10 Measure 11 Measure 12
SMALL
FOCUS
Measure 9
Measure 13
60 50 60 100
60 200 110 50 110 100
Manual code: DQB26-EN [Rev. 02] Page 69
JOLLY PLUS Series TECHNICAL MANUAL
T
Make a radiographic exposure, respecting the three sequential commands are visualized on Status field of the
«Calibration Gen…» menu:
Pause the equipment software does not allow to make any radiographic exposures.
Not ready depress the X-ray push button 1
Ready depress the X-ray push button 2
After having taken each radiographic exposure, verify on the oscilloscope display that both kV and mA wave forms are
st
time – preparation.
nd
time - exposure.
display as below reported, for example:
Channel 2: 40 mA
Channel 1: 100 kV
Verify «Calibration Gen…» menu that the mA value displayed on the field «Max», marked with nr 1 on the below
reported picture, corresponds to the reference value. The allowance admitted is ± 2%.
To eventually adjust the mA value touch the field «Set Point», marked with nr 2 on the below reported picture.
Increase or decrease the Set Point value using the two arrow keys, marked with nr 3 on the below reported picture. To confirm touch the «Enter» button, marked with nr 4 on the below reported picture.
Take at least two radiographic exposures, as above mentioned, in order to verify if the mA value displayed is correct.
To terminate the calibration procedure press the icon «Finish» marked with nr 5.
Wait till the software returns in “operator mode”.
Switch OFF the equipment and after at least 30 second switch it ON.
mA reference value to set
his button, drop-down menu, does not work
1
2
3
4
5
Manual code: DQB26-EN [Rev. 02] Page 70
JOLLY PLUS Series TECHNICAL MANUAL
4.2.6 TOUCH SCREEN CALIBRATION
The touch screen calibration is requested only when:
The display LCD touch screen electronic boards group has been replaced.
Touching the display there is not possibility to interact whit the icons or with pointer.
With reference to the § 4.2 and 4.2.2, the touch screen calibration option is on the Set Up menu. Here below has been pointed out the Set Up menu mask and the calibration icon position.
Touch Calibration icon
After having pressed the icon touch calibration icon, a dark window with has visualized. In this windows on left-up side has been reported a cross pointer which allows the technician to re-calibrate the touch screen display. Here below is shown the dark window where has been pointed out the cross pointer.
Cross pointer
The cross pointer moves automatically on clockwise direction, reaching step by step 5 positions in total. For each position the technician should press the cross pointer, then the cross pointer moves to the next position automatically, till ending the procedure. Here below is shown the position of the 5 cross pointers on the display.
Pos. 1 Pos. 2
Pos. 5
Pos. 4 Pos. 3
Touch screen schematic area
th
After having pressed the cross pointer 5
position the software closes automatically the calibration window, returning on the Set Up menu mask. If the touch re-calibration procedure does not solve the malfunction, please get in contact with the BMI Service department.
Manual code: DQB26-EN [Rev. 02] Page 71
JOLLY PLUS Series TECHNICAL MANUAL
4.2.7 TOUCH SCREEN FAILURE
In case of the touch screen does not work for a failure there is also the possibility to connect to the display CPU electronic board a PS/2 mouse, in order to continue to use the equipment. To reach the display CPU board the technician should unscrew all the screws on the lower control console perimeter and then detach it. The electronic board is mounted below the touch screen display. In the below reported picture, has been pointed out where the PS/2 mouse connector is located.
BEFORE CONNECTING THE PS/2 MOUSE SWITCH OFF THE EQUIPMENT.
PS/2 Mouse connector
Manual code: DQB26-EN [Rev. 02] Page 72
JOLLY PLUS Series TECHNICAL MANUAL
CHAPTER 5
Maintenance
Manual code: DQB26-EN [Rev. 02] Page 73
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 74
JOLLY PLUS Series TECHNICAL MANUAL
5.1 GENERAL RECOMMENDATIONS
ORDINARY AND EXTRAORDINARY MAINTENANCE MUST BE CARRIED OUT BY DULY TRAINED TECHNICAL PERSONN EL AUTHORIZED BY THE MANUFACTURER
.
The equipment requires periodical checking and maintenance to assure correct operation and safety standards. Following information and instructions are therefore to be strictly complied with:
A prolonged utilization of the equipment may affect the system safety due to the quick wear of the mechanical and
electronic components.
The correct set up of the electronic and mechanical components and of the software can prevent failures that could
affect the safety and the irradiation exposure level to which the doctors, the paramedics and the patients are exposed.
The owner of the equipment is responsible for the maintenance correct execution and is expected to have preventive
checking and technical interventions carried out by expressly authorized technical personnel.
ONLY USE ORIGINAL SPARE PARTS SUPPLIED BY THE MANUFACTURER.
5.2 ORDINARY MAINTENANCE
The below reported chart illustrates the scheduled inspections to be carried out:
PERIOD: CHECKING: DAILY
CHECKING
WEEKLY CHECKING
SIX-MONTHLY CHECKING
Check the correct functionally of the two exposure buttons on control console
Check the correct functionally of the hand switch remote control
Check the correct functionally of the collimator
Check the wearing status of the supply cable and its plug
Check the parking brakes breaks functionality
Check if during the X-ray emission no noise is coming out from the monobloc
Check the functioning of the optional installed (e.g. the DAP measuring chamber)
Check if the arm retainer works properly
Carry out the periodic electric safety tests, according to the rules in force
Check the fastening and the functioning (dust and corrosion) of the electronics boards
Check if there is any dielectric oil leakage coming out from the monobloc
(see note A here below)
(see note B here
below)
YEARLY CHECKING
FIVE-YEARLY CHECKING
Check the X-ray tube wear
Check the good operation of the equipment’s parking brakes
Check the alignment of the luminous field with the radiographic field (
Check the balancing of the arm
Check the balancing of the fork-monobloc assembly
Check the kV values really emitted during X-ray exposure
Check the mA values really emitted during X-ray exposure
Check the wearing status of the interface cables between the equipment and the monobloc
Check the wearing status of the frontal and rear wheels
Check visually the wearing condition of the steel-clad ropes, their pulleys and the related
mechanical elements
Lubricate the steel-clad ropes with mineral grease
Check the integrity of the warning labels
Replace the steel-clad ropes and the mechanical equalizer
Replace the clutch and the pulleys
Replace the front and rear wheels
(see note C here below)
(see note E here below)
(see note G here below)
(see note D here below)
(see note F here below)
(see paragraph 4.1.3)
(see paragraph 4.1.4)
see paragraph 4.1.2)
Before proceeding to replace one of the above listed components get in contact with BMI - Service Department, in order to receive the related technical information together with the spare parts.
On next page the notes list of the ordinary maintenance interventions, requiring additional technical information:
Manual code: DQB26-EN [Rev. 02] Page 75
JOLLY PLUS Series TECHNICAL MANUAL
NOTE A - CHECKING OF CORRECT OPERATION OF THE ARM RETAINER
Check whether the hooking system keeps the arm firmly locked and in sequence hook an unfasten the arm verifying the correct insertion of the hook in its seat. Move the equipment with the arm locked to verify if it keeps its position. In the
negative, replace the “arm retainer assembly”.
B DIELECTRIC OIL LEAKAGE FROM THE MONOBLOC
NOTE
Switch OFF the equipment
Remove the X-ray collimator as follows:
Lift the equipment arm to the highest position.
Detach the rear cover (panel code RS895 – please refer to the collimator manual supplied with the equipment)
unscrewing the fastening screw.
Unwire the collimator lamp power supply wires from the terminal board (please refer to the collimator manual
supplied with the equipment).
Loosen the 4 fixing dowels positioned on the upper side of the collimator (please refer to the collimator manual
supplied with the equipment) keeping the collimator in order to not to drop down the device.
Switch ON the equipment.
Enter in “operator mode”.
Depress the first time – preparation button in order to light ON the X-ray focus.
Manually tilt softly the monobloc in the two directions across its rotation axis.
Look through the monobloc “X-ray windows” verifying if there is/are any air bubble/s inside the dielectric oil.
“X-ray window”
Check if there are any dielectric oil spots on the “X-ray window” and on the collimator plane.
Reassemble the item previously detached or unwired.
If the result of the test is negative send the monobloc to the manufacturer to be repaired or replaced.
Manual code: DQB26-EN [Rev. 02] Page 76
JOLLY PLUS Series TECHNICAL MANUAL
NOTE C - CHECKING OF X-RAY TUBE WEAR
Switch OFF the equipment.
Remove the X-ray collimator as follows:
Lift the equipment arm to the highest position.
Detach the rear cover (panel code RS895 – please refer to the collimator manual supplied with the equipment)
unscrewing the fastening screw.
Unwire the collimator lamp power supply wires from the terminal board (please refer to the collimator manual
supplied with the equipment).
Loosen the 4 fixing dowels positioned on the upper side of the collimator (please refer to the collimator manual
supplied with the equipment) keeping the collimator in order to not to drop down the device.
Switch ON the equipment.
Enter in “operator mode”.
Depress the first time – preparation button in order to light ON the X-ray focus.
Manually tilt softly the monobloc in the two directions across its rotation axis.
Look through the monobloc “X-ray windows” verifying if there is/are any air bubble/s inside the dielectric oil.
“X-ray window”
Check through the “X-ray window” the wearing status of the X-ray tube anode.
Reassemble the item previously detached or unwired.
If the result of the test is negative send the monobloc to the manufacturer to be repaired or replaced.
NOTE D - CHECK THE GOOD OPERATION OF THE PARKING BRAKES
Position the equipment on a flat floor and lock its arm.
Place a dynamometer in the middle of the column (
Verify that the equipment with the brakes locked does not move applying a pushing force of ~220 N [± 5 N].
Verify that the equipment starts to move with the brakes released applying a pushing force of ~20 N [± 5 N].
If the test result is negative, adjust the pair of blocks as indicated in § 5.3.1.
see drawing A – point of application A
).
NOTE E - CHECK THE ARM BALANCING Check using a dynamometer if the pushing and pulling force required lifting Up and Down the arm is ~50 N [± 10 N].
Position the arm at 90°, parallel to the floor, with the collimator perfectly perpendicular to the floor.
Check if the arm, without keeping in position, maintains this position.
Place the dynamometer below the arm, near the handles, before the fork rotating pin (
point B).
Push with the dynamometer the arm verifying that to lift up, is it necessary to apply a force of ~50 N [± 10 N]
Lift down, manually, the arm at 90°, parallel to the floor, with the collimator perfectly perpendicular to the floor.
Place the dynamometer on the arm, near the handles, before the fork rotating pin (
C).
Push with the dynamometer the arm verifying that to lift down, is it necessary to apply a force of ~50 N [± 10 N].
If the test result is negative, adjust the spring as described in § 5.3.2.
see drawing A – application
see drawing A – application point
Manual code: DQB26-EN [Rev. 02] Page 77
JOLLY PLUS Series TECHNICAL MANUAL
NOTE F - CHECK THE BALANCING OF THE FORK-MONOBLOC ASSEMBLY
Check using a dynamometer if:
The force to be applied on the monobloc fork to rotate the monobloc (± 90°) is ~40 N [± 5 N]
see drawing A – application point D
(
).
If the test result is negative, adjust the fork joke rotation clutch as described in § 5.3.4.
The force to be applied on the collimator’s handle to tilt the monobloc (-57°/+167°) is ~40 N [± 5 N]
see drawing A – application point E
(
).
If the test result is negative, adjust the monobloc rotation clutch as described in § 5.3.5.
NOTE G - CHECK THE STEEL-CLAD ROPES CONSUMPTION
Lift Up the arm.
Detach the column front guard, shown below with nr 1.
1
Verify on both two housings, using a torch/lamp,
that the pulleys, ropes and mechanical equaliser
does not damage them, prejudicing the
equipment safety.
During the inspection move continuously up and
down the arm in order to better evaluates the
wearing status of all the mechanical components.
Reassemble the column front guard
Close the arm
Manual code: DQB26-EN [Rev. 02] Page 78
JOLLY PLUS Series TECHNICAL MANUAL
DRAWING
A DYNAMOMETER APPLICATION POINTS
Point D
Point E
Point C
Point B
Point A
Manual code: DQB26-EN [Rev. 02] Page 79
JOLLY PLUS Series TECHNICAL MANUAL
5.3 MECHANIC ADJUSTMENTS
5.3.1 BRAKE BLOCK ADJUSTMENT
Switch OFF the equipment
Detach the cassette holder rear guard, unscrewing the later screws.
Cassette holder rear guard
Adjust at the same time the brakes stretch screws (see illustration below – detail A) screwing them half turn per time,
them in the same way.
A A Screw the bolt half turn per time
Grip the transportation handle and release two times the stationary brakes in order to evaluate if the brakes adjusting
has been done correctly.
Without releasing the stationary brakes, push manually the equipment verifying that it does not move.
Verify that each brake sliding block is placed 1 mm away from the wheel.
Brake block 1 mm distance
Reassemble the cassette holder rear guard.
Manual code: DQB26-EN [Rev. 02] Page 80
JOLLY PLUS Series TECHNICAL MANUAL
5.3.2 SPRING ADJUSTMENT
THE ADJUSTMENT OF THE SPRING MUST BE CARRIED OUT ONLY BY TECHNICIANS EXPRESSLY AUTHORIZED BY THE MANUFACTURER AND DULY TRAINED THE GOOD OPERATION OF THE SPRING IS VERY IMPORTANT FOR THE OPERATORS AND THE PATIENTS SAFETY.
.
Close the arm and hook it to its retainer as explained on § 3.3.
Detach the column’s upper cover guard (see picture 1, detail A), as follows:
Unscrew manually the two black taps (see picture 2, detail B)
Loosen the two ring nuts using a pin wrench (see picture 3, detail C)
Pull out the column’s upper cover guard as indicated by an arrow in picture 3
A
Picture 1
B
Picture 2
C
Pulling direction to remove the guard
Picture 3
Manual code: DQB26-EN [Rev. 02] Page 81
JOLLY PLUS Series TECHNICAL MANUAL
After having detached the column’s upper cover guard, loosen with a pin wrench the left and right ring nuts
(see picture 4).
Left ring nut
Clutch sector
Right ring nut
Picture 4
Lift Up the arm.
Detach the column front guard, shown below in picture 5 with nr. 1.
1
Spring adjusting housing
Picture 5
Manual code: DQB26-EN [Rev. 02] Page 82
JOLLY PLUS Series TECHNICAL MANUAL
Unloose the two bolts, marked with nr 1 and nr 2 on picture nr 6 below reported, shifting them to the midcourse of
the threaded bar, as shown in picture 7
Threaded bar
1
2
Midcourse
Picture 6
Bolts
Midcourse
Picture 7
Adjust the collimator height from the floor at 200 cm, using the collimator retractable meter, as shown on picture 8:
Keep the monobloc perpendicular to the floor!
Collimator
200 cm
Picture 8
Manual code: DQB26-EN [Rev. 02] Page 83
JOLLY PLUS Series TECHNICAL MANUAL
When the technician has verified that the F.F.D. is 200 cm, screw both the bolts attached to the “threaded bumper”,
as shown on picture 9:
Bolts
“Threaded bumper”
Picture 9
Grip both bolts to the “threaded bumper”, keeping hold the lower bolt and screw the upper bolt, as shown on picture
10. This procedure does not allow at the arm loosing the maximum’s arm FFD position.
Lower bolt
“Threaded bumper”
Upper bolt
Picture 10
Pull down the arm till reaching 90°. In this position the arm should keep the position due to a correct spring
balancing. In case the arm drops down, increase the spring traction force, screwing both bolts, vice versa reduce the spring traction force, unscrewing both bolt. On picture 11 has been shown the lower bolt to be adjusted.
Bolts
Picture 11
Reassemble the column front guard and the column’s upper cover guard
Manual code: DQB26-EN [Rev. 02] Page 84
JOLLY PLUS Series TECHNICAL MANUAL
5.3.3 ARM CLUTCH ADJUSTMENT
Close the arm and hook it to its retainer as explained on § 3.3.
Detach the column’s upper cover guard (see picture 1, detail A), as follows:
Unscrew manually the two black taps (see picture 2, detail B)
Loosen the two ring nuts using a pin wrench (see picture 3, detail C)
Pull out the column’s upper cover guard as indicated by an arrow in picture 3
A
Picture 1
B
Picture 2
C
Pulling direction to remove the guard
Picture 3
Manual code: DQB26-EN [Rev. 02] Page 85
JOLLY PLUS Series TECHNICAL MANUAL
After having detached the column’s upper cover guard, loosen with a pin wrench the left and right ring nuts
(see picture 4).
Left ring nut
Clutch sector
Right ring nut
Picture 4
Release the arm from its retainer and displace it at 90°.
Grip the two ring nuts to increase the clutch force.
Check that the arm keeps at 90°.
Note: If the arm lifts Up/Down, grip or loosen both ring nuts.
This adjustment should be done being sure that the spring adjustment has been done (§ 5.3.2).
Reassemble the column front guard and the column’s upper cover guard
Manual code: DQB26-EN [Rev. 02] Page 86
JOLLY PLUS Series TECHNICAL MANUAL
5.3.4 FORK JOKE ROTATION CLUTCH ADJUSTMENT
Detach the cover disk of the rotation pin.
Rotation pin
Fork
Cover disk
M12 bolt
Rotate the monobloc ± 90° around the pin, checking that it keeps the position on both positions. If it moves, grip the
M12.
DO NOT UNSCREW M12 BOLT OR THE FORK WILL COME OFF THE ARM
Reassemble the cover disk.
Manual code: DQB26-EN [Rev. 02] Page 87
JOLLY PLUS Series TECHNICAL MANUAL
5.3.5 MONOBLOC ROTATION CLUTCH ADJUSTMENT
Detach the two ABS cover guards of the monobloc.
Rotate the monobloc longitudinally to check if it keeps the position.
Adjust the clutch operating on the ring nut after having removed the nut retainer.
Position of the clutch
Ring nut
Nut retainer
Reassemble the ABS covers.
Manual code: DQB26-EN [Rev. 02] Page 88
JOLLY PLUS Series TECHNICAL MANUAL
5.3.6 COLLIMATOR LAMP REPLACEMENT
Before removing the lamp, switch OFF the equipment for approximately 3 minutes in order to let it cool.
Proceed as follows:
Switch OFF the equipment.
Remove the rear cover of the X-ray collimator (see manual supplied with the component).
Remove the thermal dissipater (see Manual supplied with the component).
Remove the broken lamp avoiding touching the glass with your hand.
Insert the new lamp that must be technically identical to the broken one.
Insert the lamp with care.
Reassemble the details you have previously removed.
5.3.7 STEEL–CLAD ROPES AND THE MECHANICAL EQUALIZER REPLACEMENT
The pair of steel ropes and the spring equalizing rocker arm have to be replaced every 5 years. The customer must inform BMI in advance and BMI will authorize the replacement and supply the spare parts together with the instructions for the replacement.
5.4 CLEANING AND DISINFECTION
Do not use any corrosive or abrasive product or solvent to clean the unit surfaces. To disinfect the system always comply with the rules in force regarding disinfection procedures and protection against explosion. For cleaning and disinfection operations, following precautions have to be complied with:
Switch OFF the equipment and disconnect the supply cable plug from the socket.
Be sure that no liquid is poured inside the equipment in order to prevent short circuits and the corrosion of the
electric and electromechanical components.
DO NOT SWITCH ON THE EQUIPMENT WHILE USING ANAESTHETICS, INFLAMMABLE DISINFECTION AND /OR CLEANING PRODUCT. IF USING ANY PRODUCT PRODUCING EXPLOSIVE GASEOUS MIXTURE, MAKE SURE THAT THE GASES DISSIPATE BEFORE SWITCHING ON THE EQUIPMENT AGAIN
.
5.5 DISPOSAL
This product is electronic equipment subject to Decree Law 151/05, according to 2002/95/CE (also known as RoHS) – 2002/96/CE (even known as WEEE) – 2003/108/CE. For this reason in the European Community territory it cannot be disposed of as urban waste but it must be returned to the manufacturer or the Distributor where it has been bought. This disposal operation does not involve any additional cost to the user.
Graphic symbol:
Concerning the markets outside the EEC, the laws in force must be complied with, considering that the product contains:
- Silver
- Electronic components
- Alloy iron, Al and copper
- Molybdenum
- Nickel
- Dielectric oil
- Lead
- Copper
- Rhenium
- Tungsten
- Glass
Manual code: DQB26-EN [Rev. 02] Page 89
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 90
JOLLY PLUS Series TECHNICAL MANUAL
CHAPTER 6
TROUBLESHOOTING AND SPARE PARTS
Manual code: DQB26-EN [Rev. 02] Page 91
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 92
JOLLY PLUS Series TECHNICAL MANUAL
1
6.1 TROUBLESHOOTING
6.1.1 INTRODUCTION
The Jolly Plus series operator’s console will display status messages on the touch screen display during normal and abnormal operation of the equipment. This chapter contains tables of those messages and suggests actions to be taken by service personnel to correct any malfunctions that may occur.
BEFORE OPERATING ON THE EQUIPMENT FOR MAINTENANCE IT IS MANDATORY TO DISCHARGE IN ADVANCE THE CAPACITORS
BOX CHARGE
!.
Discharging capacitors box procedure:
Switch OFF the equipment.
Disconnect the equipment power supply cord from the socket.
Remove the protection guard marked with nr. 1 in the photo below reported, by unscrewing its lateral screws.
Jolly Plus - operator’s side view
Discharge the capacitors box charge keeping depressed continuously (without using a sharpened tool or something
similar) for at least 100 seconds the red head discharge push button, marked with nr. 2 in the photo below reported.
ATTENTION!
EEP DEPRESSED CONTINUOUSLY THE DISCHARGE PUSH-BUTTON FOR AT LEAST 100 SECONDS IN
K
EELLEECCTTRRIICCAALL SSPPAARRKKSS IINN TTHHEE PPUUSSH
-
H
BBUUTTTTOONN CCOONNTTAACCTT TTEERRMMIINNAALL BBLLOOCCK
-
.
K
.
2
OORRDDEERR TTOO AAVVOOIIDD
Manual code: DQB26-EN [Rev. 02] Page 93
JOLLY PLUS Series TECHNICAL MANUAL
6.1.2 WARNING MESSAGES
These messages indicate the status of the Jolly Plus.
DISPLAYED ALARM DESCRIPTION SOLUTION
1st time PB released The 1st time button has been released,
before performing the 2
nd
time button.
2nd time PB released The 2nd time button has been released,
before ending the exposure.
mA values too low The mA values are lower than 50% of set
value.
Repeat the exposure. If the message persists, contact the Manufacturer Service department. Repeat the exposure. If the message persists, contact the Manufacturer Service department. Check if the led DL1 and DL3 are lighted ON, during the X-ray exposure. Perform the mA calibration (refer to § 4.2.5).
mA values too high The mA values are higher than 50% of
set value.
Check if the led DL1 is lighted ON, during the X-ray exposure. Perform the mA calibration (refer to § 4.2.5).
kV values 85% lower The kV values are lower than 85% of set
value.
Check if the led DL1, DL2 and DL3 are lighted ON, during the X-ray exposure. Verify fuse F1 and F2 (only F1 for 4kW model), verify the inverter main power supply voltage (Vdc), verify if the feedback cable on the electronic board “control_logic” (MR6) is broken. If the message persists, contact the Manufacturer Service department.
Capacitors under charge The capacitors are under charging.
Wait the end of the charging.
Check if the led DL3 is lighted OFF, after 60 seconds from unit Power ON. If the led DL3 does not light OFF check fuse F1 and F2 (only F1 for 4kW model). If the message persists, contact the Manufacturer Service department.
System not configurated The equipment’s software should having
lost the configuration parameters.
Re-configure the equipment. If the message persists, contact the Manufacturer Service department.
Potter contact not available The Potter Bucky does not interface
properly with the equipment’s electronic.
Check the connections of the potter bucky socket’s cable to the main board. Check the connections of the interface cable from the Potter Bucky to the socket. Check in Set Up menu that this option has been set (refer to § 4.2). If the message persists, contact the Manufacturer Service department.
Maximum load working The kV, mA values or the exposure time
are too long. Starter contact not available
The X-ray tube anode does not work
correctly.
The signal “Start ready” to spin the anode
is not present.
Adjust the exposure parameters, reducing each value. Check the anode rotates (refer to § 5.2 – Note C). Verify that the fuse F1 and F2 on the electronic board starter. Verify the connection of the cable from the electronic board Starter to the monobloc. Check in Set Up menu if the anode rotation is enabled, (refer to § 4.2). If the message persists, contact the
Manufacturer Service department. Carry out the setup Request of Set Up calibration. Re-configure the equipment Set Up. X-ray signal prevented No possibility to take a radiographic
exposure due to the wrong exposure
Insert correctly exposure parameters or wait
the pause time.
parameters inserted. Or wait the pause time from one exposure and the following. The output X-ray signal of the CPU is prevented.
Manual code: DQB26-EN [Rev. 02] Page 94
JOLLY PLUS Series TECHNICAL MANUAL
DISPLAYED ALARM DESCRIPTION SOLUTION
OVL signal working Capacitors box voltage is lower than
220Vdc.
Check if the led DL3 is lighted ON, during the
X-ray exposure. Verify fuse F1 and F2 (only F1
for 4kW model), verify the inverter main power
supply voltage (Vdc), verify if the feedback
cable on the electronic board “control_logic”
(MR6) is broken.
If the message persists, contact the
Manufacturer Service department. OVC signal working Over-current in the monobloc primary. Check if the led DL1 is lighted ON, during the
X-ray exposure.
Perform the X-ray tube formation (refer to §
4.1.1). Check the mA values (refer to § 4.1.4).
If the result of the checking is negative re-
calibrate mA (refer to § 4.2.5).
Check if oil leakage/spots are coming out from
the monobloc and the wear status of the X-ray
tube (refer to § 5.2 – Note B and C).
If the message persists, contact the
Manufacturer Service department. Setup not completed Set Up procedure not terminate correctly. Terminate the Set Up correctly.
NOTE:
Here below the inverter «Control_Logic» LED:
DL1: Over-current
DL2: kV Unbalanced
DL3: Over-load
Manual code: DQB26-EN [Rev. 02] Page 95
JOLLY PLUS Series TECHNICAL MANUAL
6.2 SPARE PARTS LIST
Should some part of the equipment be replaced BMI suggest using only the original spare parts in order to avoid any malfunctioning of the equipment and the consequent loss of warranty. The manufacturer does not assume any responsibility for damages to things or people or to the equipment caused by the utilization of spare parts that have not been previously approved or purchased by BMI.
SHOULD YOU NEED ANY SPARE PART, GET IN TOUCH WITH BMI SERVICE DEPARTMENT. IT IS RECOMMENDED TO COMMUNICATE THE EQUIPMENT SERIAL NUMBER AND MODEL TOGETHER WITH THE SPARE PART CODE.
DISPLAY CONSOLE ASSEMBLY FOR BOTH JOLLY PLUS MODELS
Code Description
E01S006 Display LCD touch screen complete of «display holder» and «display CPU board» electronic
board E10A009 Exposure console push button - Ist time preparation E10A010 Exposure console push button - IInd time preparation BMH Exposure two-step hand switch complete of extensible cable
ELECTRONICS BOX FOR JOLLY 4 PLUS
Code Description
E01S004 «filament» electronic board E01S005 «main board» and «main CPU» electronic board E12S001 Single phase line filter 28A @ 230V E19A001 Static relay GQ5048D13 50A @ 480V E11S002 Toroidal transformer 120VA single phase. Primary 2x115V / Secondary 1x18V. E19A009 Automatic circuit breaker single phase [16A – 250V] Only for 230 Vac main power supply
ELECTRONICS BOX FOR JOLLY 15 PLUS JOLLY 16 PLUS AND JOLLY 30 PLUS
Code Description
E01S003 «starter» electronic board E01S004 «filament» electronic board E01S005 «main board» and «main CPU» electronic board E12S001 Single phase line filter SR28A [28A - 230V] E19A001 Static relay GQ5048D13 50A @ 480V E11S002 Toroidal transformer 120VA single phase. Primary 2x115V / Secondary 1x18V. E19A009 Automatic circuit breaker single phase [16A – 250V] Only for 230 Vac main power supply
CAPACITORS BOX FOR JOLLY 4 PLUS
Code Description
E13A001 Capacitor 6800 µf @ 450V E19A001 Static relay GQ5048D13 50A @ 480V E10A001 Capacitors box discharge push button YW1B-M1E20R E22A006 Fuse Bussmann BSFE080 BS88 FE 80A 690V
CAPACITORS BOX FOR JOLLY 15 PLUS JOLLY 16 PLUS AND JOLLY 30 PLUS
Code Description
E11S001 Toroidal transformer 1500VA single phase. Primary 2x120V / Secondary 1x300V. E13A001 Capacitor 6800 µf @ 450V E19A001 Static relay GQ5048D13 50A @ 480V E10A001 Capacitors box discharge push button YW1B-M1E20R E22A004 Fuse GI138332 - 10×38F gG CH 32A 400V E22A003 Fuse Bussmann BSFEE140 BS88 FEE 140A 690V
Manual code: DQB26-EN [Rev. 02] Page 96
JOLLY PLUS Series TECHNICAL MANUAL
INVERTER FOR JOLLY 4 PLUS
Code Description
S03S008 BMI H.F. Inverter for fixed anode monobloc
INVERTER FOR JOLLY 15 PLUS JOLLY 16 PLUS AND JOLLY 30 PLUS
Code Description
S03S004 BMI H.F. Inverter for rotating anode monobloc
MONOBLOC FOR JOLLY 4 PLUS
Code Description
E21A002 Monobloc for fixed anode X-ray tube
INVERTER FOR JOLLY 15 PLUS JOLLY 16 PLUS AND JOLLY 30 PLUS
Code Description
E21A001 Monobloc for rotating anode X-ray tube
MECHANICS AND VARIOUS FOR JOLLY 4 PLUS, JOLLY 15 PLUS, JOLLY 16 PLUS AND JOLLY 30 PLUS
Code Description
S01S004 Monobloc ABS front and rear cover kit complete of goniometer and fixing items M01S003 Console without electric and electronic items M03S009 Silkscreen panel «BMI» without electric and electronic items M02A002 Rear wheel Ø 250 mm complete of ball bearing (one piece) M02A003 Pivoting antistatic front wheel Ø 80 mm (one piece) E16A18 Halogen Lamp 24V @ 100W for the X-ray collimator
S01S003 Block for brake of rear wheel E23S007 Monobloc - equipment connecting cables
R 104/163A Manual collimator square field single layer R 104/163/A E23S049 Main power supply cord with socket - 6 meters length Set of fuses type 5×10T for all electronics boards (nr. 40 @ 4A - nr. 10 @ 6,3A - nr. 10 @ 10A)
Manual code: DQB26-EN [Rev. 02] Page 97
JOLLY PLUS Series TECHNICAL MANUAL
6.3 FUSES LIST
Code: M283V1/07 Description: Main board electronic board
POSITION AMPERE VOLT TYPE NOTE
F1 10 A 250 V 5×20T Potter bucky main power supply F2 6,3 A 250 V 5×20T Collimator lamp main power supply F3 6,3 A 250 V 5×20T Collimator lamp main power supply F4 10 A 250 V 5×20T AC/DC Power supply F5 10 A 250 V 5×20T AC/DC Power supply F6 10 A 250 V 5×20T Electronic board ~ 230V main power supply F7 10 A 250 V 5×20T Electronic board ~ 230V main power supply
Code: E01S001 Description: Inverter electronic board (mounted inside the inverter)
POSITION AMPERE VOLT TYPE NOTE
F1 4 A 250 V 5×20T Transformer TRS1 ~ 230V main power supply
Code: M283V1/07 (Electronic board installed only inside the Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus electronic box) Description: Starter electronic board
POSITION AMPERE VOLT TYPE NOTE
F1 10 A 250 V 5×20T Electronic board ~ 230V main power supply F2 10 A 250 V 5×20T Electronic board ~ 230V main power supply
Code: E01S004 Description: Filament electronic board
POSITION AMPERE VOLT TYPE NOTE
F1 10 A 250 V 5×20T Electronic board ~ 230V main power supply F2 10 A 250 V 5×20T Electronic board ~ 230V main power supply
Code: S03S007 Description: Capacitors box assembled for Jolly 4 Plus
POSITION AMPERE VOLT TYPE NOTE
F1 80 A 690 V BSFEE80 Inverter main power supply
Code: S03S003 Description: Capacitor box assembled for Jolly 15 Plus, Jolly 16 Plus and Jolly 30 Plus
POSITION AMPERE VOLT TYPE NOTE
F1 140 A 690 V BSFEE140 Inverter main power supply F2 32 A 400 V 10×38F Rectifier bridge ~ 230V Phase
Manual code: DQB26-EN [Rev. 02] Page 98
JOLLY PLUS Series TECHNICAL MANUAL
CHAPTER 7
Schematics
Manual code: DQB26-EN [Rev. 02] Page 99
JOLLY PLUS Series TECHNICAL MANUAL
THIS PAGE IS LEFT INTENTIONALLY BLANK
Manual code: DQB26-EN [Rev. 02] Page 100
Loading...