Bluelab pH Controller User Manual

Care and use guide
*6 months
for pH probe
1 Bluelab
Connect Stick
required to
data log
(sold seperately)
Contents
page page
Features 3
Bluelab pH Controller Connect 3
What’s in the box? 4
IMPORTANT - Bluelab pH Probe care 5
Set up the Bluelab Connect So ware 6
Installing the pH controller connect 6
Connect the Bluelab pH Probe 6
Connect the Bluelab Temperature Probe 6
SAFETY - Handling pH up or pH down solution 6
Set up inlet dosing tube 7
Set up outlet dosing tube 7
Connect power adaptor 8
Display menu and buttons 8
Factory settings 9
Set the dose direction ­for pH up or down solution 12
Set the dosing OnTime and O Time 13
Set alarms (optional) 14
View current settings 14
Priming the pump 15
Set the mode 15
Change the screen backlight and/or contrast 16
Using the connect so ware 16
Hydrating the pH probe 17
Cleaning the pH probe 17
Troubleshooting guide 18
Frequently Asked Questions (FAQs) 19
Technical speci cations 20
Bluelab pH Controller Connect accessories and spare parts 21
Change the display language 9
pH calibration 10
Placement of the probes 12
Set the required pH 12
2
Limited product guarantee 23
Limitation of liability 23
Contact details 23
Features
ONE Bluelab Connect Stick is required, sold seperately
Flashing high and low alarm with safety control lockouts
Local and Cloud data logging capability (download Free Bluelab Connect Software)
‘Plant-safe’ green back lit LCD with adjustable brightness and contrast
Adjustment of control settings from a local PC via connect software
Auto resume dosing on restart after power loss
Option to view data and current status remotely via Google Docs™*
Water resistant, wall mount design
Automatic control and monitoring of system pH with reservoirs up to 200 US Gallons / 760 Liters
13 foot (4 meter) acid/alkaline resistant tubing supplied
Large, easy to read display International power supply
Simple push button pH calibration with on screen instructions
Separate Bluelab Temperature Probe (for pH ATC)
Easy to navigate menu to program and adjust settings Replaceable double junction pH probe
Dosing lockouts to protect from over-dosing Replaceable peristaltic pump and tubing
What is Plant Safe? Green lights are safe for continued growth during a plant’s fruiting stage when hours of darkness are required.
What are Dosing Lockouts for? They are a built-in safety feature that stops pH dosing if an error is detected in the system (see FAQs for more detail).
Bluelab pH Controller Connect
pH display successful pH calibration symbol
Alarm on
symbol
Control dosing
up or down
3
symbols
pH calibration
button
pH controller
connect status
Peristaltic
pump with
cover
Inlet tube
Outlet tube
Power input
ATC symbol
ATTENTION
If it dries, it dies!
Bluelab Temperature Probe (for pH ATC)
pH probe BNC connector
Temperature probe input
Keep your pH probe tip wet
at all times to avoid permanent damage
Enter button
Menu button
Up button
Down button
Bluelab pH Probe
pH probe storage cap
PLEASE FOLLOW THE MANUAL STEP BY STEP BEFORE FIRST USE BEGINS
1.0 What’s in the box?
Please verify the box contents from the information below.
6 7 8 9
5
1
2
14
11
10
1
Bluelab pH Controller Connect
2
Bluelab pH Probe with storage cap
3
Bluelab Temperature Probe
4
Bluelab pH Probe holder with suction cup
5
24V DC 0.4Amp power supply
6
Europe plug adaptor
7
UK plug adaptor
8
4
North American plug adaptor
9
10
resistant dosing tube with connectors
11
12
calibration solution sachets
13
14
4
NZ / Australia plug adaptor
13 foot (4 meter) Acid/Alkali
4 x mounting fasteners
20ml pH 7.0 and pH 4.0 single-use
Inlet tube cap for pH stock solution
Replaceable peristaltic pump with cover
3
13
12
2.0 IMPORTANT - Bluelab pH Probe care
pH probes DO NOT last forever. They age through normal use and will eventually fail. The life time of a pH probe depends on the environment it is used in and the way that it is treated. To receive a long life from your Bluelab pH Probe, please ensure you follow the guide below.
pH probes contain glass and are therefore FRAGILE. With good care, they will give a long service life.
Bluelab pH Probe
pH probe holder suction cup
BNC connector
pH probe storage cap
pH probe tip
pH probe KCl storage solution
Glass tube inside plastic barrel
Glass bulb
DO NOT let the pH probe tip dry. IF IT DRIES IT DIES!
DO NOT bend the probe; this will break its internal glass tube.
DO NOT knock the probe; this will break its internal glass tube or external glass bulb.
DO NOT plunge a cold pH probe into a hot liquid, or a hot probe into cold liquid. Sudden
temperature changes can crack the glass and permanently damage the probe.
DO NOT immerse in oils, proteins or suspended solids that will leave a coating on the glass bulb.
DO NOT ‘kink’ or bend the lead sharply.
DO NOT attempt to lengthen the lead on the pH probe.
DO NOT wet the BNC connector at the end of the lead.
Always remove pH probe storage cap before use
1. Grip the top of the cap and gently twist the base one rotation clockwise to loosen slightly.
2. Next slowly slide the cap off the pH probe. DO NOT completely remove the base of the cap from the top of the cap.
3. Store the storage cap in a safe place.
Storing the pH probe
When storing the pH probe, the pH probe tip must be kept moist.
To prepare the pH probe for storage, add enough Bluelab pH Probe KCl Storage Solution to the storage cap so the probe tip is covered. Then replace the cap and store in a secure place. DO NOT use RO (Reverse Osmosis), Distilled or De-ionized water. Pure water changes the chemistry in the reference, causing the probe to die.
Removing pH probe
storage cap
If the pH probe has been accidentally allowed to dry out;
The pH probe must be ‘hydrated’ for 24 hours in KCl storage solution (never use
5
RO, Distilled or De-ionized water). Following this; carry out a calibration to check if the probe has already suffered permanent damage.
Ensure probe tip is covered by
the KCl storage solution in cap
6.0 Connect the Bluelab Temperature Probe
Insert the temperature probe connector to the base of the pH controller connect where ‘ATC’ is labelled. Ensure the connector is fully inserted.
4.0 Installing the pH controller connect
Select a suitable location that is:
• Less than 5.65 feet / 2 meters from your reservoir. The probes must be immersed in solution at all times.
• Less than 4.9 feet / 1.5 meters from an electrical mains outlet.
• Less than 5.56 feet / 2 meters from pH Up or Down solution. The inlet tube must reach the bottom of the container.
• At a suitable height to see the display and for easy operation (recommend slightly below eye level when standing).
NOTE: Avoid placing the pH controller connect where it can be damaged by direct sunlight, water, nutrient salts or pH adjuster.
Fix the fasteners through the desired mounting holes in the top and bottom of the case.
5.0 Connect the Bluelab pH Probe
Connect the pH probe to the pH controller connect by lining up the lugs of the BNC ttings.
Fasten securely by pushing the pH probe connector on and twisting one quarter turn.
3.0 Set up the Bluelab Connect So ware
Note the 4 character KEY CODE on the rear of the Bluelab pH Controller Connect.
Follow instructions in the Bluelab Connect Stick ‘Getting Started’ guide to install the connect stick and connect software. Add the pH controller connect to the software.
7.0 SAFETY - Handling pH up and down solutions
Bluelab pH Up or pH Down Solutions are available in North America. They can be used undiluted with the pH controller connect. Always follow the manufacturer’s instructions for use and handling.
CAUTION: Wear eye protection and gloves when handling this product. Harmful if swallowed. Keep out of the reach of children. Store under cool and dry conditions. Not for food, drug or household use. Read MSDS before use.
1
2
1
Mounting holes at top
2
1
2
Inserting Twisting Attached
6
1
1
Mounting holes at bottom
temperature probe connection
9.0 Set up outlet dosing tube
Insert the remaining dosing tube with connector into the RIGHT peristaltic pump tube. This is the ‘outlet dosing tube’.
Cut the outlet dosing tube so that the tube sits above the maximum solution level of the reservoir/tank. The pH stock solution must drip into the nutrient reservoir/tank.
Note: If the outlet dosing tube is immersed in reservoir/tank solution, it could create a syphon. This would cause the pH stock solution to overfl ow.
8.0 Set up inlet dosing tube
To straighten tube:
Bend tube in the opposite direction and straighten by running through closed fi ngers.
Insert one end of the dosing tube with connector into the LEFT peristaltic pump tube. This is the ‘inlet dosing tube’.
Ensure the inlet dosing tube will reach the bottom of the pH stock solution container.
There also needs to be enough tube left for the outlet dosing tube to go into the reservoir/tank.
Cut the inlet dosing tube to the desired length. Use inlet tube cap supplied to place on
Bluelab Up or Down 500ml or 1 Liter bottles. This will hold the tube in position. Insert the inlet dosing tube through the hole in the cap and pierce the induction seal on the bottle.
1
2
3
1
Inlet ‘’ arrow on pump cover
Left pump tube
Connector
Inlet dosing tube
pH stock solution
Inlet tube cap
Dosing tube reaches bottom of container
Outlet ‘’ arrow on pump cover
Right pump tube
Connector
2
7
Inlet dosing tube
pH stock solution
Outlet dosing tube
Nutrient solution reservoir / tank
Loading...
+ 16 hidden pages