Bluelab EC Pen User Manual

Care and use guide
Guía de uso y cuidados
Notice d’utilisation et d’entretien
Gebruikershandleiding
Bedienungs- und Pegeanleitung
www.getbluelab.com
1.0 To operate
Turn pen on
Press power button. The last measurement is recalled for 3 seconds.
To turn pen off
Press and hold the power button until OFF is displayed.
NOTE: The pen will automatically turn off after 4 minutes to conserve battery power.
Measure EC
Place probe in solution and wait for reading to stabilize.
To hold a reading
If you want to “hold” the reading on the screen, short press the power button. To exit the hold function press the power button again.
To change units
Hold down the units button for 3 seconds until the conductivity and temperature units start flashing. Short press units button again to cycle between unit combinations. To exit this mode don’t press anything for 3 seconds.
NOTE: You can change units while in hold mode by holding down the units button.
Rinse conductivity probe
To reduce the build-up of nutrient salts, rinse under running water after each use.
The probe needs to be cleaned once every two weeks to ensure accurate readings. To clean the probe follow the cleaning instructions in section 2.0.
Features
Backlit LCD display Calibration optional Hold reading function Fully waterproof Fully guaranteed for 1 year Auto off function Low battery warning Automatic Temperature Compensation (ATC) Successful calibration indicator Selectable units for conductivity and temperature
English
1
Quick guide
2
Appears when batteries are low.
Power button / hold
Short press to turn on. Short press to hold reading. Long press to turn off.
Calibrate button
See calibration section.
Units button
Hold until units ash then short press to
change units. Screen will change back when no buttons.
have been pressed for 3 seconds.
Shroud
Attention:
The conductivity probe must be cleaned regularly to remove built up nutrient salts to ensure an accurate reading (See cleaning instructions).
“The instrument is only as accurate as the probe is clean.”
Tick to indicate successful calibration
3
1 second alternating displays
4
5
2
4.0 Calibration
Calibration of conductivity is not required for this unit as it is factory calibrated. However; if you wish to calibrate the unit follow the instructions below.
YOU MUST CLEAN THE PROBE BEFORE CALIBRATING.
See section 2.0.
Rinse probe in fresh water and place it in a known standard solution. See chart below for the correct solution.
Wait for reading to stabilize.
Hold down the cal button for 3 seconds until CAL appears.
Release button and CAL P should be displayed. If Err is displayed check the probe is clean and that the calibration solution is fresh and uncontaminated.
A tick will appear on the screen to indicate that the calibration was successful. The tick will disappear after 30 days. To reset back to factory default remove/replace battery.
NOTE: If you need to test or calibrate in a 1500 ppm solution, you MUST set the pen to EC, then multiply your result by 540. If calibrating, multiply 2.8 by 540 (2.8 x 540 = 1512).
This unit DOES NOT measure in the 540 ppm scale.
1
EC ppm 500
(TDS)
ppm 700
(EC x 700)
Solution value
2.77 1385 1940
Displayed value
2.8 1390 1940
5.0 Error messages
The following error messages appear for the following reasons.
2.0 Cleaning and maintenance
Cleaning the EC pen probe periodically ensures accurate readings. The probe is cleaned using the Bluelab Conductivity Probe Cleaner, or “Jif” a trade name for a liquid scourer cream used in home bathrooms and kitchens. Similar products are called “Liquid Vim”, “Soft Scrub”, “Cif cream”, or “Viss”. Never use scented varieties as they contain oils that contaminate the probe. Follow the steps below to clean the probe.
Remove shroud
Hold the body and pull the shroud off. Holding your hand around the shroud for a few seconds will help with removal.
Clean probe face
Place one or two drops of probe cleaner onto the probe face and rub with your finger or Bluelab Chamois firmly and vigorously, to clean the probe face.
If a heavy build up occurs around the temperature sensor clean with a soft toothbrush to remove contamination.
Rinse probe
Rinse off all traces of cleaner under running water using the same finger or other side of Bluelab Chamois.
Check that the water forms a film on the probe face with no “beads” of water. If beading is present repeat the cleaning process.
Replace shroud and test in a known solution to ensure the unit has been adequately cleaned.
3.0 Battery replacement
The EC pen is powered with 1 x AAA alkaline battery. Do not use rechargeable batteries. A low battery warning is indicated by a battery symbol appearing on the screen. Only remove the battery cap when the batteries require changing. Battery life is expected to be 350 hours. Follow these steps to replace the battery.
Remove old battery
Undo battery cap fasteners. Remove battery cap and tip out the old battery.
Check for corrosion
Flat batteries may leak and cause corrosion. Check battery contacts and the battery for any sign of corrosion. Battery contacts should be cleaned first if corrosion is found before proceeding to step 3.
Fit new battery
Insert the new batteries positive (+) end down into the body.
Ensure waterproof battery cap seal is clean
Waterproof seal will fail if any dirt is around the seal and body where the cap seals.
Replace battery cap
Tighten fasteners on battery cap until there is no space left between the cap and body. This ensures the unit remains 100% waterproof.
1
2
3
4
1
2
3
4
5
Waterproof seal
Pull
2
3
4
Temperature under range
Temperature over range
1385 ppm
Hold for 3
seconds
Temperature error
Hardware error
EC over range
3
Loading...
+ 4 hidden pages