Bluelab Combo Meter User Manual

Care and use guide
Contents page
Features 2
Bluelab Combo Meter 2
Introduction 3
Preparing for use 3
Calibration 4
Changing nutrient and temperature display units 6
Measuring hydroponic elements 7
Battery replacement 7
Cleaning the conductivity/ temperature probe 8
Cleaning the pH probe 9
Hydrating the pH probe 10
Storing the Bluelab Combo Meter 10
Error messages 11
Bluelab pH Probe replacement 13
Limitation of liability 14
English
Features
Measures pH, conductivity / nutrient (EC, CF, ppm 500 and ppm 700) and temperature (°C, °F)
Lightweight and portable Replaceable double junction pH probe Large easy to read display Over range and under range indicators Simple push button pH calibration Low battery indicator Successful pH calibration indicator 2 x AAA alkaline batteries included
No calibration required for conductivity and temperature
Auto off function
pH probe
connection
socket
(BNC fi tting)
Display
Bluelab
Conductivity/
Temperature
Probe
pH/calibrate
button
Bluelab Combo Meter
Bluelab pH Probe
Nutrient/units
button
Conductivity/
temperature
probe shroud
pH probe storage cap
The pH probe tip must not be allowed to dry out. Always place the storage cap back onto the pH probe after each use. Ensure the cap contains enough Bluelab pH Probe KCl Storage Solution to cover the probe tip.
Nutrient units
Low battery
indicator
pH unit
Temperature units Successful pH
calibration indicators
2
ATTENTION
If it dries, it dies!
Keep your pH probe tip wet
at all times to avoid permanent damage
Temp/units button
English
1.0 Introduction
The Bluelab Combo Meter has three press buttons; ‘pH / calibrate’, ‘nutrient / units’ and ‘temp / units’. The buttons have a short press and long press function. A short press means a button is released in about one second. The long press is a button being held for at least three seconds and released when the display starts fl ashing.
Turning the Combo Meter on
A short press of any button turns the Combo Meter on. The Combo Meter automatically turns off after four minutes if no buttons are pressed. If the Combo Meter turns off before the reading is taken, a short press of any button will turn the Combo Meter on again.
2.0 Preparing for use
The following tasks must be performed before the Bluelab Combo Meter is used for the fi rst time.
Insert batteries.
See section 6.0.
Connect pH probe
Connect the pH probe to the pH meter by lining up the lugs of the BNC fi ttings. Fasten securely by pushing the pH probe connector on and twisting one quarter turn.
Remove the storage cap
a) Remove the pH probe storage cap by
gripping the top of the cap and gently twisting the base one rotation clockwise to loosen slightly. Next slowly slide the cap off the pH probe. DO NOT completely remove the base of the cap from the top of the cap.
CAUTION: When the pH probe is not in use, add enough Bluelab pH Probe KCl Storage Solution to the storage cap so the probe tip is covered. Then replace the cap and store in a secure place.
DO NOT use RO (Reverse Osmosis), Distilled or De-ionized water. Pure water changes the chemistry in the reference, causing the probe to die.
Calibrate the pH
Calibrate the combo meter by following the instructions in section 3.0 of this manual.
This must be done before the Combo Meter is used for the fi rst time.
1
1
2
3
Inserting Twisting Attached
Attaching the
Bluelab pH Probe to the Meter
Removing pH probe
storage cap
3
4
Ensure probe tip is covered by
the KCl storage solution in cap
See section 3.0 for calibration steps
3.0 Calibration
pH calibration is required before fi rst use to ensure that the fi rst reading is accurate.
• The Bluelab Conductivity/Temperature Probe DOES NOT require calibration.
• The pH of the Bluelab Combo Meter DOES require calibration.
For accurate pH readings the pH probe is cleaned and recalibrated when:
• The reading is different to what you were expecting.
• The batteries have been removed or changed.
• The pH probe is replaced with a new one or is disconnected from the Combo Meter.
• The pH calibration indicators have disappeared.
When calibrating the pH after fi rst use the pH probe needs to be cleaned. See pH probe cleaning in section 8.0.
For best pH calibration
pH reading accuracy is dependant on the accuracy and age of the calibration solutions used, and use and cleanliness of the pH probe tip.
• Ensure the pH probe has been cleaned and rinse the pH probe with clean water between calibration solutions to reduce contamination of the pH solutions.
• Only fresh uncontaminated solutions should be used.
• Calibrate the pH at the same temperature as the solution to be measured.
• ALWAYS calibrate the pH probe with pH 7.0 then pH 4.0 or pH 10.0.
NOTE: The conductivity/temperature probe does not need to be calibrated, but must be cleaned to remove any build up of nutrient salts. See Section 7.0.
The pH calibration involves cleaning the pH probe tip and then calibrating in TWO SOLUTIONS.
If a reading below pH 7.0 is expected, use pH 7.0 and pH 4.0 calibration solutions. If a reading above pH 7.0 is expected, use pH 7.0 and pH 10.0 calibration solutions.
Follow the steps on the following page for Combo Meter pH calibration.
Storage and use of calibration solutions
• Always place the lid back onto the bottle after use or evaporation will occur rendering the solution useless.
• Store in a cool place.
• DO NOT measure directly into the bottle. Tip a small amount into a clean container and discard after use.
• Never add water to solutions.
pH reading accuracy is dependant on the accuracy and age of the calibration solutions used, and use and cleanliness of the pH probe tip.
English
4
3.0 Calibration cont.
To calibrate the pH
Clean pH probe tip.
See Section 8.0 (the pH probe does not require cleaning before the fi rst use).
pH 7.0 calibration
a) Turn pH meter on. Rinse pH probe tip in fresh
water, shake off excess water and place in a pH 7.0 calibration solution. Wait for at least one minute or longer if required for reading to stablize to a constant value.
b) Long press the ‘calibrate’ button. When the
display shows CAL release button.
pH 7 indicator is displayed indicating
a successful pH 7 calibration.
The pH 4 indicator will now fl ash
indicating pH 4.0 or pH 10.0 calibration is now required.
c) If Err appears during the calibration process see
Section 11.0.
d) The combo meter must be calibrated to two
points. If after an hour the combo meter has not been calibrated with a second calibration point the calibration indicators disappear and the combo meter reverts to an uncalibrated state. Calibration is required.
pH 4.0 / 10.0 calibration
a) Rinse the pH probe tip in fresh water, shake
off excess water and place the pH probe tip in either pH 4.0 or pH 10.0 calibration solution.
b) Wait for at least one minute for reading to
stabilize to a constant value.
c) Long press the ‘calibrate’ button. When the
display shows CAL release button.
pH 7 / pH 4 is displayed
or pH 7 / pH 10 is displayed
d) The pH meter is now calibrated and ready for use. e) After 30 days, the calibration indicators
disappear to let you know calibration is required.
Store the pH probe
Add enough Bluelab pH Probe KCl Storage Solution into the probe storage cap to fully submerge the pH probe tip. Place storage cap on probe.
English
1
2
3
clean the pH probe
pH 7.0 calibration solution
rinse in tap water
pH 4.0 or pH 10.0 calibration solution
4
5
Successful pH 7
and pH 4 calibration
Ensure probe tip is covered by
the KCl storage solution in cap
Loading...
+ 9 hidden pages