BK Precision 9832, 9833 User Manual

9830 Series
Model: 9832, 9833
High Power Programmable AC Power Source
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 Safety Summary
The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
Read and understand the safety and operational information in this manual.
Apply all the listed safety precautions.
Verify that the voltage selector at the line power cord input is set to the correct line
voltage. Operating the instrument at an incorrect line voltage will void the warranty.
Make all connections to the instrument before applying power.
Do not operate the instrument in ways not specified by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply with these requirements.
Category rating
The IEC 61010 standard defines safety category ratings that specify the amount of electrical energy available and the voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested, which defines the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the mains supply. The voltages in the environment are typically derived from a limited-energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains supply at a standard wall outlet or similar sources. Example measurement environments are portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains installation of a building. Examples are measurements inside a building's circuit breaker panel or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary power entering a building or other outdoor wiring.
i
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Do not use this instrument in an electrical environment with a higher category rating than what is specified in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument's category rating to maintain the instrument's category rating. Failure to do so will lower the category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be between 190 V to 250 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts (including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument's rear panel is for a reference of potential only and is not to be used as a safety ground.
Do not operate in an explosive or flammable atmosphere
ii
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finely­divided particulates.
The instrument is designed to be used in office-type indoor environments. Do not operate the instrument
In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or
finely-divided particulates.
In relative humidity conditions outside the instrument's specifications.
In environments where there is a danger of any liquid being spilled on the instrument or
where any liquid can condense on the instrument.
In air temperatures exceeding the specified operating temperatures.
In atmospheric pressures outside the specified altitude limits or where the surrounding
gas is not air.
In environments with restricted cooling air flow, even if the air temperatures are within
specifications.
In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is 0 °C to 40 °C and 20% to 80% relative humidity, with no condensation allowed.
Measurements made by this instrument may be outside specifications if the instrument is used in non-office-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric fields, or strong magnetic fields.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument's power cord, remove the instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.
iii
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual.
Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualified service-trained maintenance personnel who are aware of the hazards involved when the instrument's covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt any service or adjustment unless another person capable of rendering first aid and resuscitation is present.
Do not insert any object into an instrument's ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists.
Fuse replacement
iv
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Fuse replacement must be done by qualified service-trained maintenance personnel who are aware of the instrument's fuse requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specified in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a fire. Failure to use the specified fuses will void the warranty.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
Do not place heavy objects on the instrument.
Do not obstruct cooling air flow to the instrument.
Do not place a hot soldering iron on the instrument.
Do not pull the instrument with the power cord, connected probe, or connected test
lead.
Do not move the instrument when a probe is connected to a circuit being tested.
v
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements.
Compliance Statements
Disposal of Old Electrical & Electronic Equipment (Applicable in the European
Union and other European countries with separate collection systems)
CE Declaration of Conformity
This instrument meets the requirements of EMC Directive 2014/30/EU Electromagnetic Compatibility Directive and the following standards.
EMC Directive
- EN 61326-1:2013 Class A
- EN 61326-2-1:2013
- EN 61000-3-12:2011
- EN 61000-3-11:2000
- EN 61326-1:2013 (industrial locations)
- EN 610000-4-2:2009
- EN 610000-4-3:2006+A1:2008+2010
- EN 610000-4-4:2012
- EN 610000-4-5:2014
Find Quality Products Online at: sales@GlobalTestSupply.com
- EN 610000-4-6:2010
- EN 610000-4-8:2010
- EN 610000-4-34:2007+A1:2009
vi
www.GlobalTestSupply.com
Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument.
Electric Shock hazard
Hot surface
On (Power). Press the “I” at the top of the power switch to turn the instrument ON.
Off (Power). Press the “O” at the bottom of the power switch to turn the instrument off.
Direct current (DC)
Alternating current (AC)
Direct and alternating current (DC + AC)
Chassis (earth ground) symbol
Ground terminal
Protective earth ground
Fuse Symbol
CAT I (1000V)
IEC Measurement Category I. Inputs may be connected to mains (up to 1000 VAC) under Category I overvoltage conditions.
CAT II (300V)
IEC Measurement Category II. Inputs may be connected to mains (up to 300 VAC) under Category II overvoltage conditions.
Safety Symbols
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
vii
CAUTION indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
WARNING indicates a hazardous situation which, if not avoided, could result in death or serious injury
DANGER indicates a hazardous situation which, if not avoided, will result in death or serious injury.
NOTICE is used to address practices not related to physical injury.
Safety instructions (or equivalent) signs indicate specific safety-related instructions or procedures
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
viii
2 Notations
TEXT – Denotes a softkey.
TEXT – Denotes a front panel key.
ix
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Table of Contents
1 Safety Summary ............................................................................................................... i
Compliance Statements ............................................................................................................... vi
Safety Symbol ............................................................................................................................. vii
2 Notations ........................................................................................................................ ix
3 General Information ........................................................................................................ 1
3.1 Package Contents .............................................................................................................. 1
3.2 Product Dimensions ........................................................................................................... 2
3.3 Installation ......................................................................................................................... 3
3.4 Front Panel......................................................................................................................... 4
3.5 Rear Panel .......................................................................................................................... 6
4 Display Overview ............................................................................................................ 7
4.1 Display Description ............................................................................................................ 7
5 Getting Started.............................................................................................................. 10
5.1 Input Power Requirements ............................................................................................. 10
5.1.1 North American Split Phase Power .................................................................... 10
5.1.2 Fuse .................................................................................................................... 12
5.2 Preliminary Check ............................................................................................................ 12
5.2.1 Power-on Procedure .......................................................................................... 12
5.2.2 Warm-up Time ................................................................................................... 13
5.2.3 Power-off Procedure .......................................................................................... 13
5.3 Sense Lines ....................................................................................................................... 13
6 Menu Tree .................................................................................................................... 15
7 Front Panel Operation ................................................................................................... 16
7.1 Configure Voltage and Frequency Output....................................................................... 16
7.1.1 Setting voltage .................................................................................................... 16
7.1.2 Setting frequency ............................................................................................... 16
7.1.3 Setting Voltage ................................................................................................... 17
7.2 Program Settings ............................................................................................................. 18
7.2.1 Step Mode .......................................................................................................... 18
7.2.2 List Mode ............................................................................................................ 20
7.2.3 Pulse Mode ......................................................................................................... 25
8 Configure Menu ............................................................................................................ 27
x
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8.1 Config 1 ............................................................................................................................ 27
8.2 Config 2 ............................................................................................................................ 29
8.3 Limits ................................................................................................................................ 31
9 System Settings ............................................................................................................. 33
9.1 System Setup ................................................................................................................... 33
9.2 Communication Setup ..................................................................................................... 33
9.3 System Error .................................................................................................................... 34
9.4 System Next ..................................................................................................................... 35
9.5 Recall Default ................................................................................................................... 36
10 Save .............................................................................................................................. 38
10.1 Save Config ...................................................................................................................... 38
10.2 Save Screen ...................................................................................................................... 39
10.3 Recall Configuration ........................................................................................................ 39
11 Remote Interface Operation .......................................................................................... 41
11.1 Interface Connection ....................................................................................................... 41
11.1.1 RS232 ................................................................................................................. 41
11.1.2 USBVCP (virtual COM) ....................................................................................... 41
11.1.3 USBTMC ............................................................................................................. 42
11.1.4 GPIB ................................................................................................................... 42
11.1.5 LAN (Ethernet) ................................................................................................... 42
11.1.6 Web server ......................................................................................................... 42
12 Digital I/O ..................................................................................................................... 46
12.1 External Voltage Control.................................................................................................. 47
12.1.1 AC Couple........................................................................................................... 48
12.1.2 DC Couple .......................................................................................................... 48
12.2 Trigger In .......................................................................................................................... 48
12.3 15 VDC ............................................................................................................................. 49
12.4 Output Status Detection .................................................................................................. 49
12.4.1 /SYNC ................................................................................................................. 50
12.4.2 /Fault_out .......................................................................................................... 50
12.4.3 /Transient .......................................................................................................... 50
12.5 Remote_Inhibit ................................................................................................................ 51
12.6 Tx / Rx .............................................................................................................................. 52
12.7 Event_SW ......................................................................................................................... 52
12.8 Analog input (BNC) .......................................................................................................... 53
xi
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
13 Build-in Harmonic Wave ................................................................................................ 55
14 Calibration .................................................................................................................... 70
14.1 AC Voltage Calibration ..................................................................................................... 70
14.1.1 300V range calibration ...................................................................................... 71
14.1.2 150V range calibration ...................................................................................... 72
14.1.3 DC Voltage Calibration ....................................................................................... 72
14.1.4 AC Current Calibration ....................................................................................... 73
14.1.5 Restore to Factory Default (RECALL DATA) ....................................................... 74
14.1.6 External Voltage Calibration .............................................................................. 75
15 Specifications ................................................................................................................ 78
16 LIMITED THREE-YEAR WARRANTY ................................................................................. 81
xii
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 General Information
The B&K Precision 9830 series are low distortion single phase AC power sources delivering a maximum of 3000 VA, 300 Vrms, 30 Arms /97.5 Apk. The Output frequency is adjustable from 45 Hz to 1200 Hz. All models are capable of outputting AC, DC or AC+DC. Predefined waveforms include sine, square, clipped sine and THD waveforms. Standard remote interface include USBVCP and USBTMC-compliant. RS232, LAN, and GPIB interfaces are available to provide flexibility for remote operation.
Features
Output AC, DC or AC+DC
Built-in power factor correction (PFC) circuit at the AC input
Low harmonic distortion
Power Line Disturbance simulation
Step, pulse and list modes
Adjustable phase angle control
Built-in and user definable waveforms
Digital I/O port for external trigger, remote inhibit, command completion, failure status,
Analog input for external control
Comprehensive protection modes including OVP, OCP, OTP, fan failure, key lock
3.1 Package Contents
Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items from the shipping carton, and check for any obvious signs of physical damage that may have occurred during transportation. Report any damage to the shipping agent immediately. Save the original packing carton for possible future reshipment. Every instrument is shipped with the following contents:
1 x 9833 or 9832 AC Power Source
1 x AC input power cord
1 x Certificate of calibration
1 x Test report
Verify that all items above are included in the shipping container. If anything is missing, please contact B&K Precision.
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Note: Check that you have the most current User manual.
3.2 Product Dimensions
Figure 1 - Dimensions
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1
Step 1
2
3
4
Step 2
5
Step 3Step 4
6
3.3 Installation
The following diagram shows how to install the optional rack mount kit.
Figure 2 - Rack Mount Kit
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Power ON/OFF
USB disk port
VFD
Softkeys
Function keys & Indicator LEDs
Number keys
Function keys
Rotary knob
Direction keys
3.4 Front Panel
Figure 3 - Front Panel
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Key
Definition
Frequency key, press to set the output frequency
Voltage key, press to set the output voltage
Press to lock the keyboard, (Keyboard is locked when LED is illuminated).
Output On/Off key, press to enable/disable voltage output, (Output is On when LED is illuminated).
to
Number key 1 to 9 for direct numeric entry.
Set the number entered to a positive or negative value.
Main function: decimal point Secondary function: switch from remote control to local control
Escape key, press to escape the settings menu or cancel the current settings
Save key, store configurations and settings into the internal memory or a USB disk
Recall key, recall data from internal storage or external USB disk
Enter key, press to confirm the settings change
Rotary, to adjust value or make up/down selection (press it for ENTER function)
Up/Down/Left/Right direction keys Confirm key, press to confirm the settings
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5
AC output terminal
Remote sense terminal
ANALOG connector
DIGITAL Input /Output
LAN port
USB port (USBVCP or USBTMC)
GPIB port
Ground wire
AC input terminal
3.5 Rear Panel
Figure 4 - Rear Panel
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 Display Overview
The power on screen shows the model number and self-test status. The Interface, and Power module should indicate OK. The (Real Time Clock), RTC will show the battery voltage. When the self-test completes, the instrument will advance to the Settings and Measure screen. This process takes approximately 7 seconds. If an error does occur, the instrument must be returned for service.
Figure 5 - Power on screen
4.1 Display Description
At the top right of the screen the remote and output status are displayed. Each of the blue softkeys will have a corresponding label to the left. These labels will change on each screen to show different options. Settings Fields have a black background. Settings Fields can be selected by using arrow keys to move to the field and press enter to select and modify the value in the field. Press the Enter key again to confirm the change or press the Esc key to cancel the change.
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Parameter
Description
Volts (rms)
AC output Vrms setting
Volts (DC)
DC output setting
Freq (Hz)
Output frequency setting
Sync source
Select Phase or Immed (immediate)
Phase (deg)
Phase angle setting
Figure 6 - Display Description
The instrument provides 3 display screen options for accessing commonly used measurements and settings. Each of these screens can be selected by pressing the Display x of x softkey repeatedly.
Figure 7 - Display 1 of 3
Table 1 – Display 1 of 3 Settings
Display 2 of 3
This display shows all 12 power measurements and an output timer.
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Figure 8 - Display 2 of 3
Display 3 of 3
This displays shows a graphical representation of the settings, output measurements and waveforms.
Figure 9 - Display 3 of 3
Press the keys to select and view V (voltage waveform) or A (current waveform). Then press the key to change the scale until the complete waveform is visible.
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Connection of the 9832 or 9833 to an AC power source should be made by a qualified electrician or other qualified personnel. Incorrect wiring may damage the source or cause a fire hazard.
The included AC power cord is safety certified for this instrument operating in the rated range. If an extension cable is added, be sure that it can meet the required power ratings for this instrument. Any misuse with wrong or unsafe cables will void the warranty.
5 Getting Started
Before connecting and powering up the instrument, please review and go through the instructions in this chapter.
5.1 Input Power Requirements
The AC input accepts line voltage input within:
Voltage: 190 V – 250 V
Frequency: 47 Hz – 63 Hz
Maximum power consumption: 9832: 2500 VA; 9833: 3800 VA
Before connecting to an AC outlet or external power source, make sure that the power switch is in the OFF position and verify that the AC power cord, including the extension line, is compatible with the rated voltage/current and that there is sufficient circuit capacity for the power supply.
Follow the illustrations below to connect the new AC power cord to the AC input of the source in the rear panel.
5.1.1 North American Split Phase Power
Split phase is commonly used in North American to provide 208 to 240 Volts. This requires two wires that are both hot. For example 110 V and, 110 V resulting in 220 V across them. The included cable may be used in most regions by adding a band of blue electrical tape on each end of the white wire to identify both the black and white with blue band as hot. The green trace yellow wire is always used for ground and must be connected properly. Check with your local Authority Having Jurisdiction for clarification.
The power cable that is shipped with the instrument is preassembled from the factory. The following instruction will provide information for changing the power cable for split phase.
1. Verify that the source power in your facility can provide the minimum voltage and
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
10
current required to operate the instrument.
2. Using the power cable provided, identify the end that plugs into the instrument.
Before disassembling the black hooded connector, make a note of how the unit is assembled from the factory so it can be correctly reassembled.
3. Remove the 4 screws from the bottom of the hooded connector as shown and place the upper and the upper and lower hood pieces and screws to the side.
The black, white, and green wires should now be clearly visible. Use a screwdriver to unscrew and release the white wire only.
Add a band of blue electrical tape or heat shrink to each end of the white wire.
1. Return the white wire into the green connecter and tighten the screw. Verify all
three wires are secured in the connector.
2. Closely inspect each of the wires at the enter point of the green connector to
verify none of the copper strands are bridging over to the wrong terminal causing a short.
3. Carefully reassemble the hood so the Latch Fingers and Caution Label are facing
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Any disassembling of the case or fuse changes not performed by an authorized service technician will void the warranty of the instrument.
up and reinstall the 4 screws into the hood. The green connector should be locked in place.
4. Using the (2) screws provided with the cable, secure the assembled connector to
the back of the instrument as shown in Figure 10 – Securing. Note: Use only the screws provided.
The green (trace yellow) wire connects to electrical ground (required). The black wire connects to L (Line 1) and the white with blue band connects to N (Line 2).
5.1.2 Fuse
The factory installed fuse will meet the requirements when the instrument is operated with the specified Input Power requirements. This source is a switching mode power supply. The fuse installed inside should not fail under normal operation. If the fuse has blown, it may be an indication of a more serious malfunction inside of the source. In this event, contact B&K Precision.
5.2 Preliminary Check
5.2.1 Power-on Procedure
1. Complete the following steps to verify that the source is ready for use.
2. Verify AC Input Voltage
Figure 10 - Securing the power cable
3. Verify and check to make sure proper AC voltages are available to power the
12
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Do not connect multiple power supplies in series or parallel as this may cause damage or a malfunction.
When the AC input voltage is lower than 190 VAC, the source will activate an inner over temperature protector and cut off the output in response to the condition. To ensure that the entire test process can be completed smoothly, confirm that the input AC voltage is within the specified range.
When the AC input frequency is outside the range of 47 to 63 Hz, the source will beep. To ensure the normal operation of the source, please make sure the AC input frequency is within the required range.
instrument. The AC voltage range must meet per the acceptable specifications.
4. Connect Power
5. Connect specified AC power cord and verify the hood is in place and correctly secured to rear panel.
6. Press the power switch to the | ON position to turn ON the instrument. It will take a few seconds before the fans power on and the self-test begins.
5.2.2 Warm-up Time
The 9830 series is fully operable upon switching the power on. However, to reach the specified equipment accuracy, please allow the source to warm up for at least 15 minutes.
5.2.3 Power-off Procedure
When the instrument is not in use, make sure to set the front panel power switch to the OFF position. After the power switch is turned off the internal fans will continue to run for approximately 5-10 seconds to discharge the internal capacitors per safety requirements. Once the discharge process is complete, the instrument will carry out an automatic shutdown process of approximately 2-8 seconds. Do not turn the power back on until the instrument has completed a full shut down cycle.
5.3 Sense Lines
The source can be configured as local sense or remote sense to compensate the voltage drop of the test leads. The following diagram shows how to connect the load by local sense or remote sense configurations.
When local sense is selected, the sense end L is connected to the output end L and sense end N is connected to the output end N, whereas the output end L is connected to the load end L and output end N is connected to the load end N. When this sensing mode is selected, the wires connecting between the sense ends to the load end must be as short as possible. The local sense is the default configuration. When remote sense is selected, both
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
13
the sense end L and output end L are connected to the load end L, whereas both sense end N and output end N are connected to the load end N.
Figure - 11 Sense Lines
14
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 Menu Tree
15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...
+ 65 hidden pages