The following safety precautions apply to both operating and maintenance personnel and must be followed during all
phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
• Read and understand the safety and operational information in this manual.
• Apply all the listed safety precautions.
• Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument
at an incorrect line voltage will void the warranty.
• Make all connections to the instrument before applying power.
• Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design,
manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply
with these requirements.
2
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available and the
voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is
a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested,
which denes the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the
mains supply. The voltages in the environment are typically derived from a limited-energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains
supply at a standard wall outlet or similar sources. Example measurement environments are portable
tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains
installation of a building. Examples are measurements inside a building’s circuit breaker panel
or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary
power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specied in this manual
for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the
instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating
of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be
115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your
country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This
instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power
cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power
cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts
(including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the
recommended three-conductor AC line power cable may result in injury or death.
3
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only
and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
• In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided particulates.
• In relative humidity conditions outside the instrument’s specications.
• In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense
on the instrument.
• In air temperatures exceeding the specied operating temperatures.
• In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
• In environments with restricted cooling air ow, even if the air temperatures are within specications.
• In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is
0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by this instrument
may be outside specications if the instrument is used in non-oce-type environments. Such environments may include
rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise,
strong electric elds, or strong magnetic elds.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the
instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated,
and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the
instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases,
or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not
for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support
device or system.
4
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must
be made by qualied service-trained maintenance personnel who are aware of the hazards involved when the instrument’s
covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may
exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect
all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there
are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device
before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the
measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt
any service or adjustment unless another person capable of rendering rst aid and resuscitation is present.
Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit
exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the instrument’s fuse
requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses.
Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specied in this manual or on
the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a re. Failure
to use the specied fuses will void the warranty.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to
B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
• Do not place heavy objects on the instrument.
• Do not obstruct cooling air ow to the instrument.
• Do not place a hot soldering iron on the instrument.
• Do not pull the instrument with the power cord, connected probe, or connected test lead.
• Do not move the instrument when a probe is connected to a circuit being tested.
Safety Symbols
SymbolDescription
indicates a hazardous situation which, if not avoided, will result in death or serious injury.
5
indicates a hazardous situation which, if not avoided, could result in death or serious injury
indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
This is the In position of the power switch when instrument is ON.
This is the Out position of the power switch when instrument is OFF.
is used to address practices not related to physical injury.
Contents
1Quick Start10
1.1Front Panel10
1.2Rear Panel10
1.3Touch Screen Display10
1.4Waveform Selection and Setup10
1.5Create a simple sine wave11
1.5.1Frequency and Phase12
1.5.2Amplitude, and Oset12
1.6To Turn On/O Output13
1.6.1Function Keys13
2Sine Waveform14
2.1Frequency/Period14
2.2Amplitude14
2.3Oset15
2.4Phase16
2.5Harmonics16
2.5.1Harmonic Type17
2.5.2Harmonic Order18
2.5.3Harmonic Amplitude18
2.5.4Harmonic Phase18
3Square Wave19
3.1Duty Cycle19
3.1Oset, High/Low Levels19
3.2Phase20
4Ramp Wave23
4.1Frequency, Amplitude, Oset, High/Low level, and Phase23
4.2Symmetry24
5Pulse Wave27
5.1Frequency, Amplitude, Oset, High/Low level, and Phase27
5.2Pulse Width and DutyCycle27
5.3Rising and Falling Edges28
5.4Delay29
6Noise Wave32
6.1Standard Deviation32
6.2Mean33
6.3Bandwidth33
7DC Wave35
8Arbitrary Wave36
8.1DDS36
8.2TrueArb36
8.2.1Sampling Rate37
8.3Waveform Selection37
8.4Selecting a Built-in Waveform38
8.5Stored Waveform38
9Modulation Function40
9.1AM40
9.1.1To Select Modulation Source41
9.1.2Internal Source41
9.1.3External Source41
9.1.4Modulation Depth41
9.1.5Modulation Frequency41
9.1.5.1DSB-AM42
9.1.5.2FM42
9.1.5.3Frequency Deviation43
9.1.5.4PM43
9.1.5.5Phase Deviation44
9.1.5.6FSK44
9.1.6Key Frequency44
9.1.7Hop Frequency45
9.1.7.1ASK45
9.1.7.2PSK46
9.1.7.3PWM46
9.1.8Pulse Width/Duty Deviation46
10 Sweep Function49
10.1Sweep Frequency49
10.2Start Frequency and Stop Frequency49
10.3Center Frequency and Frequency Span50
10.4Sweep Type50
10.5Linear Sweep50
10.6Log Sweep50
10.7Sweep Trigger Source50
10.8Internal Trigger51
10.9External Trigger51
10.10Manual Trigger51
7
11 Burst Function52
11.1Burst Type52
11.1.1N-Cycle52
11.1.2Innite53
11.1.3Gated53
11.1.4Start Phase54
11.1.5Burst Period54
11.1.6Cycles/Innite54
11.1.7Delay54
11.1.8Burst Trigger Source55
11.1.9Internal Trigger55
11.1.10 External Trigger55
11.1.11 Manual Trigger55
12 Storage System56
12.1To Store and Recall56
12.1.1USB Device (0:)56
12.1.2Browse57
12.1.3File Type57
12.1.4State File58
12.1.5Data File58
12.2File Operation58
12.2.1To Save the Instrument State58
12.2.2Select the character59
12.2.3Delete the character59
12.2.4Save the le.59
12.2.5To Recall State File or Data File59
12.2.6To Delete File59
12.2.7To Copy and Paste File59
13 Utility Function60
13.1System Settings61
13.2Number Format62
13.3Language Setup62
13.4Power On62
13.5Set to Default63
13.6Beep63
13.7Screen Saver63
13.8System Info63
13.9Software Update63
13.10 Built-in Help64
13.11 Test/Cal65
13.11.1 Self Test65
13.11.2 Screen Test66
13.11.3 Key Test66
13.11.4 LED Test67
13.11.5 Board Test67
13.11.6 Touch Adjust67
13.12 Frequency Counter68
13.12.1 Counter Setup69
13.13 Parameters to be measured69
13.14 Reference Frequency69
13.15 Trigger Level69
13.16 Coupling Mode69
13.17 High Frequency Rejection70
13.18 Output70
13.19 Load70
13.19.1 Steps for setting the load:70
13.20 Polarity70
13.21 EqPhase71
13.22 Waveforms Combination Mode71
13.23 CH Copy/Coupling71
13.23.1 Channel Coupling71
13.23.2 Frequency Coupling72
13.23.3 Amplitude Coupling73
13.23.4 Phase Coupling73
13.24 Channel Tracking74
8
14 Output Synchronization75
14.1Sync Signals of Dierent Waveforms75
14.1.1Modulated Waveform75
14.1.2Sweep and Burst Waveform75
15 Clock Source76
15.1Sync methods for two or more instruments:76
15.2Synchronization among multiple instruments76
16 Channel Phase Mode77
17 Overvoltage Protection78
18 Remote Interface79
18.1User-dened programming79
18.2Remote Control via USB79
18.3Remote Control via GPIB80
18.4Remote Control via LAN80
19 Specications81
20 Appendix: Waveforms82
20.1Common Waveforms82
20.1Math Waveforms82
20.2Engine Waveforms83
20.3Window Waveforms84
20.4Trigonometric Waveforms85
20.5Square Waveforms85
20.6Medical Waveforms85
20.7Modulated Waveforms86
20.8Filter Waveforms86
20.9Demo Waveforms87
21 Daily Maintenance88
21.1Cleaning88
22 LIMITED THREE-YEAR WARRANTY89
23 Service Information90
9
Quick Start
1.1 Front Panel
4060B Series front panel includes a touch screen, menu softkeys, numeric keyboard, knob, function keys, arrow keys, and
channel control area as shown in Figure 1.1.
Item Description
1Power Switch
2USB Port (load waveforms, store settings, etc...)
3Touchscreen
4Soft Keys
5Number Pad
6Function and Channel Keys
7Knob and Selection Button
8Arrow Keys
9Channel Outputs
Figure 1.1 Front Panel View
1.2 Rear Panel
The rear panel shown in Figure 1.2 provides multiple interfaces, including Counter, 10MHz In/Out, Aux In/Out, LAN,
USB Device, Earth Terminal and AC Power Supply Input.
1.3 Touch Screen Display
The screen displays the parameters and a the waveform for a single channel. Most of the elds can be accessed either
through the function keys below the screen or by pressing their value on screen. Figure 1.3 shows a view of Channel
1 with a simple waveform (sine wave). The values and parameters available on-screen change depending on the mode,
options and waveform selected.
1.4 Waveform Selection and Setup
Press the “Waveform” button to open the waveform setup menu shown in Figure 1.4. The softkeys show the available
waveforms for that page. The “Page 1/2” button accesses the other page of waveforms shown in the second image in
Figure 1.4.
Quick Start11
Item Description
1Power Input
2USB
3Ethernet
4Counter Input
5Auxillary Input/Output
6Clock Input/Output (10 MHz)
7Ground Connection
Figure 1.2 Rear Panel of 4060B Series
Item Description
1Channel Tab
2Waveform
30 V reference
4Parameters
5Output Settings
6Status indicators (clock, lock and network)
7Settings menu
Figure 1.3 Touch Screen Display
1.5 Create a simple sine wave
By default, the generator starts congure with a 1 kHz, 4 V peak to peak, symmetric to the ground reference waveform
starting at 0 degrees. All of these parameters are changeable.
Quick Start12
Page 1
Page 2
Figure 1.4 Waveform Menu
1.5.1 Frequency and Phase
To set the frequency, press the rst soft-key, the frequence value to the right of the waveform, or the word “frequency”
above the rst soft-key. The eld, when ready for editing, is highlighted in blue as shown in Figure 1.5. Use the knob,
arrow keys, and the number pad to set the desired frequency. As the value is changed the output will follow the changes
as they occur if the output is on. Phase is modied in the same manner.
Figure 1.5 Frequency Setting
1.5.2 Amplitude, and Oset
Like setting the frequency, select the parameter to modify and use the knob and keys to change its value. The oset and
amplitude are related through the eective maximum and minimum voltages. The waveform output is fully dened by
either setting the amplitude and oset, or the high and low level. The high and low level dene the peak values of the
waveform.
Quick Start13
1.6 To Turn On/O Output
The two keys on the right side of the operation panel above each channel output are used to enable and disable the
output. When enabled, the key will light up. When enabled and lit, pressing the button again disables the respective
channel. Each key can also change the load impedance value by pressing and holding the key for 2 seconds.
Note: Load impedance is only modies the signal voltage setting. HiZ expects an impedance much larger than 50,
so the output voltage is the same as the internal driving source voltage. When set to 50, the displayed voltage is half
of this value. If the wrong setting is chosen, the output voltage can by up to double, or down to half of the expected
value. See Section 13.19 for more.
1.6.1 Function Keys
Access to most of the conguration and machine setup is through the function keys. See Figure 1.6.
Figure 1.6 Function Keys
Mod. This is the waveform “Modulation” setup and enable key. When pressed, the modulation set-
tings are applied to the current waveform. See Section 9 for details.
Sweep This is the waveform “Sweep” frequency setup and enable key. This key also applies sweep set-
tings to the current waveform when pressed. See Section 10 for details.
Burst This is the waveform “Burst” setup and enable key. This key also applies burst settings to the
current waveform when pressed. See Section 11 for details.
Parameter This key returns the menu system to the current waveform parameter settings. This is the menu
set that opens when selecting a “waveform”.
Utility This key opens the main menu for conguring the system settings of the generator. Generator
conguration such as the remote interface, clocking, synchronization, and other functions are accessed in this menu.
Store/Recall The le browser is opened by this key giving access to open, save and manage system setups.
Waveforms The set of waveform types is accessed in this menu.
Ch1/Ch2 This key, when pressed, changes the active waveform menu. See the color of the waveform and
the tab brought to the foreground to verify which channel is selected.
Sine Waveform
Press the “Waveforms” key and then press the Sine softkey. The screen shown in Figure 2.1 will open. The parameters
available for sine waveforms include frequency, period, amplitude, high level, low level, oset and phase.
ParameterDescription
Frequency/PeriodSet the signal frequency or period; The current parameter will be switched at a second press.
Amplitude/HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second press.
Oset/LowLevelSet the signal oset or low level; The current parameter will be switched at a second press.
PhaseSet the phase of the signal.
Figure 2.1 Sine Wave Parameters
2.1 Frequency/Period
Frequency is one of the most important parameters of basic waveforms. For dierent instrument models and waveforms,
the available ranges of frequency are dierent. For detailed information, please refer to “4060B Series Datasheet”. The
default frequency is 1 kHz.
1. Press Waveforms → Sine → Frequency, to set the frequency parameter. The frequency shown on the screen when the
instrument is powered on is the default value or the set value of last power down. If Period (rather than Frequency) is
the desired parameter, press Frequency/Period again to enter the Period mode. The current value for the waveform’s
period is now displayed in inverse color. Press the Frequency/Period key once again to return to the Frequency entry
mode.
2. Input the desired frequency.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
Note: When using the numeric keyboard to enter the value, the left arrow key can be used to move the cursor
backward and delete the value of the previous digit.
2.2 Amplitude
The amplitude setting range is limited by the “Load” and “Frequency/Period” settings. For detailed information, please
refer to “4060B Series Datasheet”.
Sine Waveform15
Figure 2.2 Set Frequency
1. Press Waveforms → Sine → Amplitude, to set the amplitude. The amplitude shown on the screen when the instrument
is powered on is the default value or the set value of last power down. If setting the waveform’s high level is desired,
press the Amplitude/HighLevel key again to switch into the high level parameter (the current operation is displayed
in inverse color).
2. Input the desired amplitude.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
Figure 2.3 Set Amplitude
2.3 Oset
The oset setting range is limited by the “Load” and “Amplitude/HighLevel” settings. For detailed information, please
refer to “4060B Series Datasheet”. The default value is 0Vdc.
1. Press Waveforms → Sine → Oset, to set the oset. The oset shown on the screen when the instrument is powered
on is the default value or the set value of last power down. If you want to set the waveform by low level, press the
Oset/LowLevel key again, to switch into the low level parameter (the current operation is displayed in inverse color).
2. Input the desired oset.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
Sine Waveform16
Figure 2.4 Set Oset
2.4 Phase
1. Press Waveforms → Sine → Phase, to set the phase. The Phase shown on the screen when the instrument is powered
on is the default value or the set value of last power down.
2. Input the desired phase.
Use the numeric keyboard to input the parameter value directly and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
Figure 2.5 Set Phase
Note When the independent mode is enabled, the phase parameter cannot be modied
2.5 Harmonics
The 4060B Series can be used as a harmonic generator to output harmonics with specied order, amplitude and phase.
According to the Fourier transform, a periodic time domain waveform is the superposition of a series of sine waveforms
as shown in the equation below:
Sine Waveform17
Figure 2.6
Press Waveforms → Sine → Harmonic and choose “On”, then press
ItemDescription
TypeSet the harmonic type to “odd”, “ever” or “all”.
OrderSet the order of the harmonic.
Harmonic Ampl Set the amplitude of the harmonic.
Harmonic Phase Set the phase of the harmonic.
CancelReturn to the sine parameters menu.
Figure 2.7Harmonic Interface
2.5.1 Harmonic Type
The 4060B Series can output odd harmonics, ever harmonics and user-dened orders of harmonics. After entering the
harmonic setting menu, press Type to select the desired harmonic type.
1. Press Even, the instrument will output fundamental waveform and even harmonics.
2. Press Odd, the instrument will output fundamental waveform and odd harmonics.
3. Press All, the instrument will output fundamental waveform and all the user-dened orders of harmonics.
Sine Waveform18
2.5.2 Harmonic Order
After entering the harmonic setting menu, press Order, the use the numeric keyboard or knob to input the desired value.
• The range is limited by the maximum output frequency of the instrument and current fundamental waveform frequency.
• Range: 2 to maximum output frequency of the instrument ÷ current fundamental waveform frequency The maximum
is 10.
2.5.3 Harmonic Amplitude
After entering the harmonic setting menu, press Harmonic Ampl to set the harmonic amplitude of each order.
1. Press Order to select the sequence number of the harmonic to be set.
2. Press Harmonic Ampl to set the amplitude of the harmonic selected. Use the arrow keys and knob to change the
value. Or use the numeric keyboard to input the amplitude value and then select the desired unit from the pop-up
menu. The units available are Vpp, mVpp and dBc.
2.5.4 Harmonic Phase
After entering the harmonic setting menu, press Harmonic Phase to set the harmonic phase of each order.
1. Press Order to select the sequence number of the harmonic to be set.
2. Press Harmonic Phase to set the phase of the harmonic selected. Use the arrow keys and knob to change the value.
Or use the numeric keyboard to input the phase value and then select the unit °.
Square Wave
Press Waveforms key to select the waveform function, and press the Square softkey. The square waveform parameters
are set by using the Square operation menu.
The parameters of square waveforms include frequency/period, amplitude/high level, oset/low level, phase and duty. As
shown in Figure 3.1, select DutyCycle. The duty cycle parameter area is highlighted in the parameter display window,
and users can set the duty cycle value here.
FunctionDescription
Frequency/ PeriodSet the signal frequency or period; The current parameter will be switched at a second press.
Amplitude/ HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second
press.
Oset/ LowLevelSet the signal oset or low level; The current parameter will be switched at a second press.
PhaseSet the phase of the signal.
DutyCycleSet the duty cycle for square waveform.
Figure 3.1 Square Wave Parameters
1 Duty Cycle
The ratio of the amount of time the pulse is in the high state and the waveform’s period.
The duty cycle setting range is limited by the “Frequency/Period” setting. For detailed information, please refer to
“4060B Series Datasheet”. The default value is 50%.
1. Press Waveforms → Square → DutyCycle, to set the duty cycle. The duty cycle shown on the screen when the
instrument is powered on is the default value or the set value of last power down.
2. Input the desired Duty Cycle. Use the numeric keyboard to input the parameter value directly and press the corresponding key to select the parameter unit. Or use the arrow keys to select the digit to edit, and then use the knob
to change its value. The generator will change the waveform immediately.
Note: The methods of setting other parameters of square signal are similar to sine waveform function.
3.1 Oset, High/Low Levels
Square Wave20
3.2 Phase
Figure 3.2 Set Duty Cycle
Figure 3.3 Set Duty Cycle
Figure 3.4 Set Oset
Square Wave21
Figure 3.5 Set Hi Level
Figure 3.6 Set Low Level
Figure 3.7 Set Phase
Square Wave22
Figure 3.8 Phase at 180 degrees
Ramp Wave
Press Waveforms key to select the waveform function, and press the Ramp softkey. The ramp waveform parameters are
set by using the ramp operation menu.
The parameters for ramp waveforms include frequency/period, amplitude/high level, oset/low level, phase and symmetry.
As shown in Figure 4.1, in the soft key menu, select Symmetry. The symmetry parameter area is highlighted in the
parameter display window, and users can set the symmetry value here.
Figure 4.1 Ramp Waveform Main Screen
Frequency and Period Set the signal frequency or period; The current parameter will be switched at a second press.
Amplitude/ HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second press.
Oset/ LowLevel Set the signal oset or low level; The current parameter will be switched at a second press.
Phase Set the phase of the signal.
Symmetry Set the symmetry for ramp waveform.
4.1 Frequency, Amplitude, Oset, High/Low level, and Phase
Set the frequency, amplitude, high level, low level, oset and phase as shown in Figures 4.2, 4.3, 4.4, 4.5, 4.6.
Figure 4.2 Set Frequency
Ramp Wave24
Figure 4.3 Set Amplitude
Figure 4.4 Set Oset
4.2 Symmetry
The percentage that the rising period takes up the whole Period.
Input Range: 0~100% Default Value: 50%
1. Press Waveforms → Ramp → Symmetry, to set the symmetry. The symmetry shown on the screen when the
instrument is powered on is the default value or the set value of last power down.
2. Input the desired Symmetry.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value. The generator will
change the waveform immediately.
Ramp Wave25
Figure 4.5 Set High Level
Figure 4.6 Set Low Level
Figure 4.7 Set Phase
Ramp Wave26
Figure 4.8 Set Symmetry 15%
Figure 4.9 Set Symmetry 90%
Pulse Wave
Press Waveforms key to select the waveform function, and press the Pulse softkey. The pulse waveform parameters are
set by using the pulse operation menu.
The parameters for pulse waveforms include frequency/period, amplitude/high level, oset/low level, width, rise/fall and
delay. As shown in Figure 5.1, in the soft key menu, select PulWidth. The pulse width parameter area is highlighted in
the parameter display window, and users can set the pulse width value here.
Figure 5.1 Pulse Waveform Main Screen
Frequency/ Period Set the signal frequency or period; The current parameter will be switched at a second press.
Amplitude/ HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second press.
Oset/ LowLevel Set the signal oset or low level; The current parameter will be switched at a second press.
PulWidth/ DutyCycle Set the signal pulse width or duty cycle; The current parameter will be switched at a second press.
Rise/ Fall Setting the rise edge or fall edge for pulse waveform. The current parameter will be switched at
a second press.
Delay Setting the delay for pulse waveform.
5.1 Frequency, Amplitude, Oset, High/Low level, and Phase
Set the frequency, amplitude, high level, low level, and oset as shown in Figures 4.2, 4.3, 4.4, 4.5, 4.6.
5.2 Pulse Width and DutyCycle
Pulse width is dened as the time from the 50% threshold of a rising edge amplitude to the 50% threshold of the next
falling edge amplitude (as shown in the gure below). The pulse width setting range is limited by the “Minimum Pulse
Width” and “Pulse Period” setting. For detailed information, please refer to “4060B Series Datasheet”. The default
value is 200s.
Pulse duty cycle is dened as the percentage that the pulse width takes up in the whole period. Pulse duty cycle and
pulse width are correlative. Once a parameter is changed, the other will be automatically changed.
1. Press Waveforms→ Pulse → PulWidth, to set the pulse width. The pulse width shown on the screen when the
instrument is powered on is the default value or the set value of last power down. If you want to set the waveform by
duty, press the PulWidth/DutyCycle key again, to switch into the duty parameter (the current operation is displayed
in inverse color).
Pulse Wave28
Figure 5.2 Set Frequency
Figure 5.3 Set Amplitude
2. Input the desired Pulse Width.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value. The generator will
change the waveform immediately.
5.3 Rising and Falling Edges
Rise edge time is dened as the duration of the pulse amplitude rising from 10% to 90% threshold, while fall edge time is
dened as duration of the pulse amplitude moving down from 90% to 10% threshold. The setting of rise/fall edge time
is limited by the currently specied pulse width limit. Users can set rise edge and fall edge independently.
1. Press Waveforms → Pulse → Rise to set the rise edge.
The rise edge shown on the screen when the instrument is powered on is the default value or the set value of last
power down. If you want to set the waveform by fall edge, press the Rise/Fall key again, to switch into the fall edge
parameter (the current operation is displayed in inverse color).
2. Input the desired rise edge.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value. The generator will
change the waveform immediately.
Pulse Wave29
5.4 Delay
Figure 5.4 Set Oset
Figure 5.5 Set High Level
Figure 5.6 Set Low Level
Pulse Wave30
Figure 5.7 Set Pulse Width
Figure 5.8 Set Duty Cycle
Figure 5.9 Setting the Risng Edge
Pulse Wave31
Figure 5.10 Pulse Delay
Noise Wave
Press Waveforms key to select the waveform function, and press the Noise softkey. The noise parameters are set by using
the noise operation menu.
The parameters for noise include stdev, mean and bandwidth. As shown in Figure 6.1, in the soft key menu, select
Stdev, The stdev parameter area is highlighted in the parameter display window, and users can set the stdev value here.
Noise is non-periodic signal which has no frequency or period.
BandSet Turn on/o the bandwidth setting.
Stdev Setting the stdev for noise waveform.
Mean Setting the mean for noise waveform.
Bandwidth Setting the bandwidth for noise waveform.
6.1 Standard Deviation
Figure 6.1Noise Parameters Display Interface
Figure 6.2Setting the Stdev
1. Press Waveforms → Noise → Stdev, to set the stdev. The stdev shown on the screen when the instrument is powered
on is the default value or the set value of last power down.
Noise Wave33
2. Input the desired stdev.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
6.2 Mean
Figure 6.3Setting the Mean
1. Press Waveforms → Noise → Mean, to set the mean.
The mean shown on the screen when the instrument is powered on is the default value or the set value of last power
down.
2. Input the desired mean.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
6.3 Bandwidth
Figure 6.4Setting the Bandwidth
Noise Wave34
Figure 6.5Setting the Bandwidth
1. Press Waveforms → Noise → BandSet and choose “On” to set the bandwidth.
The bandwidth shown on the screen when the instrument is powered on is the default value or the set value of last
power on. When changing the function, if the current value is valid for the new waveform, it will be used sequentially.
2. Input the desired bandwidth.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or you can use the arrow keys to select the digit you want to edit, and then use the knob to change its value.
DC Wave
Press Waveforms → Page 1/2 → DC, to enter the following interface. Please note that there is a ‘DC oset’ parameter
at the middle of the screen.
Figure 7.1DC Setting Interface
Arbitrary Wave
The Arb signal consists of two types: the system’s built-in waveforms and the user-dened waveforms. Built-in waveforms
are stored in the internal non-volatile memory. Users may also edit the arbitrary waveform with 8 to 8M data points,
namely 8pts to 8Mpts.
8.1 DDS
Choose Waveforms →Page 1/2 →Arb →Arb Mode and select the “DDS” output mode. The parameters include frequency/period, amplitude/high level, oset/low level and phase.
Figure 8.1 DDS
Frequency/Period Set the signal frequency or period; The current parameter will be switched at a second press.
Amplitude/HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second press.
Oset/LowLevel Set the signal oset or low level; The current parameter will be switched at a second press.
Phase Set the phase of the signal.
In DDS output mode, users can set the frequency or period of the arbitrary waveform. The instrument outputs an
arbitrary waveform which is made up of certain points according to the current frequency
8.2 TrueArb
Choose Waveforms →Page 1/2 →Arb →Arb Mode and select the “TrueArb” output mode. The parameters include
sampling rate/frequency, amplitude/high level, oset/ low level and phase.
SRate/ Frequency Set the signal sampling rate or frequency; The current parameter will be switched at a second
press.
Amplitude/ HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second press.
Oset/ LowLevel Set the signal oset or low level; The current parameter will be switched at a second press.
Phase Set the phase of the signal.
In TrueArb output mode, users can set the sampling rate (the output points per second) or frequency of the arbitrary
waveform. The instrument outputs an arbitrary waveform point by point according to the current sampling rate.
Arbitrary Wave37
Figure 8.2 TrueArb
8.2.1 Sampling Rate
1. Press Waveforms →Page 1/2 →Arb →TureArb →Srate, to set the sampling rate parameter.
The sampling rate shown on the screen when the instrument is powered on is the default value or the set value of last
power on. When setting the function, if the current value is valid for the new waveform, it will be used sequentially.
If you want to set the frequency for the waveform, press SRate/Frequency key again, to switch to the frequency
parameter (the current operation is displayed in inverse color).
2. Input the desired sampling rate.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or you can use the arrow keys to select the digit you want to edit, and then use the knob to change its value.
Figure 8.3 Set the Sampling Rate
Note: The methods of setting the parameters of arbitrary signal are similar to sine waveform function.
8.3 Waveform Selection
There are numerous built-in Arbitrary Waveforms and there is storage for user-dened Arbitrary Waveforms inside the
generator.
Arbitrary Wave38
8.4 Selecting a Built-in Waveform
Choose Waveforms →Page 1/2 →Arb →Arb Type →Built-In to enter the following interface, as shown in Figure 8.4.
See Chapter 20 for the set of available built-in waveforms.
Figure 8.4 Built-in Arbitrary Waveforms
Press Common, Math, Engine, Window, Trigo or other menus to switch to the desired category (the selected category in
the menu bar is highlighted), then rotate the knob or click the touch screen to choose the desired waveform (the selected
waveform is highlighted). Select Accept or press the knob to recall the corresponding waveform.
8.5 Stored Waveform
Choose Waveforms →Page 1/2 →Arb →Arb Type →Stored Waveforms to enter the following interface, as shown in
Figures 8.5, 8.6.
Figure 8.5 Stored Waveform Display Interface
Rotate the knob or touch the screen to choose the desired waveform. Then select Recall or press the knob to recall the
corresponding waveform.
Arbitrary Wave39
Figure 8.6 Stored Waveform Display Interface
Modulation Function
Use the Mod key to generate modulated waveforms. The 4060B Series can generate AM, FM, ASK, FSK, PSK, PM,
PWM and DSB-AM modulated waveforms. Modulating parameters vary with the types of the modulation. In AM,
users can set the source (internal/external), depth, modulating frequency, modulating waveform and carrier. In DSBAM, users can set the source (internal/external), modulating frequency, modulating waveform and carrier. In FM, users
can set the source (internal/external), modulating frequency, frequency deviation, modulating waveform and carrier.
In PM, users can set the source (internal/external), phase deviation, modulating frequency, modulating waveform and
carrier. In ASK, users can set the source (internal/external), key frequency and carrier. In FSK, users can set the source
(internal/external), key frequency, hop frequency and carrier. In PSK, users can set the source (internal/external), key
frequency, polarity and carrier. In PWM, users can set the source (internal/external), modulating frequency, width/duty
cycle deviation, modulating waveform and carrier.
We will introduce how to set these parameters in details according to the modulation types.
9.1 AM
The modulated waveform consists of two parts: the carrier and the modulating waveform. In AM, the amplitude of the
carrier varies with the instantaneous voltage of the modulating waveform.
Press Mod → Type → AM, the parameters of AM modulation are shown in Figure 9.1.
Function Explanation
TypeAMAmplitude modulation
SourceInternalThe source is internal
ExternalThe source is external. Use the [Aux In/Out] connector at the rear panel.
AM DepthSet the modulation depth.
ShapeSineChoose the modulating waveform.
Square
Triangle
UpRamp
DnRamp
Noise
Arb
AM FreqSet the modulating waveform frequency. Frequency range: 1mHz~1MHz (internal source
only).
Figure 9.1Setting Interface of AM Modulation
Modulation Function41
9.1.1 To Select Modulation Source
The 4060B Series can accept modulating signal from an internal or external modulation source. Press Mod → AM →
Source to select “Internal” or “External” modulation source. The default is “Internal”.
9.1.2 Internal Source
When internal AM modulation source is selected, press Shape to select Sine, Square, Triangle, UpRamp, DnRamp, Noise
or Arb as modulating waveform.
• Square: 50% duty cycle
• Triangle: 50% symmetry
• UpRamp: 100% symmetry
• DnRamp: 0% symmetry
• Arb: the arbitrary waveform selected of the current channel
Note: Noise can be used as modulating waveform but cannot be used as the carrier.
9.1.3 External Source
When external AM modulation source is selected, the generator accepts external modulating signal from the [Aux In/Out]
connector at the rear panel. At this time, the amplitude of the modulated waveform is controlled by the signal level
applied to the connector. For example, if the modulation depth is set to 100%, the output amplitude will be the maximum
when the modulating signal is +6V and the minimum when the modulating signal is -6V.
The 4060B Series can use one channel as a modulating source for the other channel. The following example takes the
output signal of CH2 as the modulating waveform.
1. Connect the CH2 output terminal to [Aux In/Out] connector on the rear panel using a dual BNC cable.
2. Select CH1 and press Mod to select the desired modulation type as well as set the corresponding parameters, and
then select external modulation source.
3. Select CH2 and select the desired modulating waveform and set the corresponding parameters.
4. Press Output to enable the output of CH1.
9.1.4 Modulation Depth
Modulation depth expressed as a percentage indicates the amplitude variation degree. AM modulation depth varies from
1% to 120%. Press AM Depth to set the parameter.
• In the 0% modulation, the output amplitude is the half of the carrier’s amplitude.
• In the 120 modulation, the output amplitude is the same with the carrier’s amplitude.
• For an external source, the depth of AM is controlled by the voltage level on the connector connected to the [Aux
In/Out]. ±6V correspond to 100% depth.
• When external modulation source is selected, this menu is hidden.
9.1.5 Modulation Frequency
When internal modulation source is selected, press AM Freq to highlight the parameter, then use the numeric keyboard
or arrow keys and knob to input the desired value.
Modulation Function42
• The modulation frequency ranges from 1mHz to 1MHz.
• When external modulation source is selected, this menu is hidden.
9.1.5.1 DSB-AM
DSB-AM is an abbreviation for Double-Sideband Suppressed Carrier –
Amplitude Modulation. Press Mod → Type → DSB-AM. The parameters of DSB-AM modulation are shown in Fig-
ure 9.2.
Function Description
TypeDSB-AMDSB Amplitude modulation.
SourceInternalThe source is internal.
ExternalThe source is external. Use the [Aux In/Out] connector at the rear panel.
DSB FreqSet the modulating waveform frequency. Frequency range: 1mHz~1MHz (internal source
only).
ShapeSineChoose the modulating waveform.
Square
Triangle
UpRamp
DnRamp
Noise
Arb
Figure 9.2Setting Interface of DSB-AM Modulation
Note: The methods of setting the parameters of DSB-AM are similar to AM.
9.1.5.2 FM
The modulated waveform consists of two parts: the carrier and the modulating waveform. In FM, the frequency of the
carrier varies with the instantaneous voltage of the modulating waveform.
Press Mod → Type → FM, the parameters of FM modulation are shown in Figure 9.3.
Modulation Function43
Function Explanation
TypeFMFrequency modulation
SourceInternalThe source is internal
ExternalThe source is external. Use the [Aux In/Out] connector at the rear panel.
Freq DevSet the frequency deviation
ShapeSineChoose the modulating waveform.
Square
Triangle
UpRamp
DnRamp
Noise
Arb
FM FreqSet the modulating waveform frequency. Frequency range 1mHz~1MHz (internal source).
Figure 9.3 Setting Interface of FM Modulation
9.1.5.3 Frequency Deviation
Press FM Dev to highlight the parameter, and then use the numeric keyboard or arrow keys and knob to input the desired
value.
• The deviation should be equal to or less than the carrier frequency.
• The sum of the deviation and the carrier frequency should be equal to or less than maximum frequency of the selected
carrier waveform.
Note: The methods of setting other parameters of FM are similar to AM.
9.1.5.4 PM
The modulated waveform consists of two parts: the carrier and the modulating waveform. In PM, the phase of the carrier
varies with the instantaneous voltage level of the modulating waveform.
Press Mod → Type → PM, the parameters of PM modulation are shown in Figure 9.4.
Modulation Function44
Function Explanation
TypePMPhase modulation
SourceInternalThe source is internal
ExternalThe source is external. Use the [Aux In/Out] connector at the rear panel.
Phase DevPhase deviation ranges from 0° ~ 360°.
ShapeSineChoose the modulating waveform.
Square
Triangle
UpRamp
DnRamp
Noise
Arb
PM FreqSet the modulating waveform frequency. Frequency range: 1mHz~1MHz.
Figure 9.4Setting Interface of PM Modulation
9.1.5.5 Phase Deviation
Press Phase Dev to highlight the parameter, and then use the numeric keyboard or arrow keys and knob to input the
desired value.
• Use the numeric keyboard or arrow keys and knob to input the desired value.
• The range of phase deviation is from 0° to 360° and the default value is 100°.
Note: The methods of setting other parameters of PM are similar to AM.
9.1.5.6 FSK
The FSK is Frequency Shift Keying, the output frequency of which switches between two preset frequencies (carrier
frequency and the hop frequency or sometimes known as mark frequency (1) and space frequency (0)).
Press Mod → Type → FSK, the parameters of FSK modulation are shown in Figure 9.5.
9.1.6 Key Frequency
When internal modulation source is selected, press Key Freq to set the rate at which the output frequency shifts between
“carrier frequency” and “hop frequency”.
Modulation Function45
Function Explanation
TypeFSKFrequency shift keying modulation.
SourceInternalThe source is internal.
ExternalThe source is external. Use the [Aux In/Out] connector at the rear panel.
Key FreqSet the frequency at which the output frequency shifts between the carrier frequency and the
hop frequency (internal modulation only): 1mHz~1MHz.
Hop FreqSet the hop frequency.
Figure 9.5 Setting Interface of FSK Modulation
• Use the numeric keyboard or arrow keys and knob to input the desired value.
• The key frequency ranges from 1mHz to 1MHz.
• When external modulation source is selected, this menu is hidden.
9.1.7 Hop Frequency
The range of the hop frequency depends on the carrier frequency currently selected. Press Hop Freq to highlight the
parameter, and then use the numeric keyboard or arrow keys and knob to input the desired value.
• Sine: 1uHz~120MHz
• Square: 1uHz~25MHz
• Ramp: 1uHz~1MHz
• Arb: 1uHz~20MHz
Note: The methods of setting other parameters of FSK are similar to AM. In addition, the external modulating signal
of FSK must be Square which complies with the CMOS level specication.
9.1.7.1 ASK
When using ASK (Amplitude Shift Keying), the carrier frequency and key frequency will need to be set. The key frequency
is the shift rate of modulated waveform amplitude.
Press Mod → Type → ASK, the parameters of ASK modulation are shown in Figure 9.6.
Modulation Function46
Function Explanation
TypeASKAmplitude shift keying modulation.
SourceInternalThe source is internal.
ExternalThe source is external. Use the [Aux In/Out] connector at the rear panel.
Key FreqSet the frequency at which the output amplitude shifts between the carrier amplitude and zero
(internal modulation only): 1mHz~1MHz.
Figure 9.6Setting Interface of ASK Modulation
Note: The methods for setting the parameters of ASK are similar to AM. In addition, the external modulating signal
of ASK must be Square which complies with the CMOS level specication.
9.1.7.2 PSK
When using PSK (Phase Shift Keying), congure the generator to “shift” its output phase between two preset phase
values (carrier phase and modulating phase). The default modulating phase is 180°.
Press Mod → Type → PSK, the parameters of PSK modulation are shown in Figure 9.7.
Note: The methods of setting the parameters of PSK are similar to AM. In addition, the external modulating signal
of PSK must be Square which complies with the CMOS level specication.
9.1.7.3 PWM
For only the “Pulse” a PWM (Pulse Width Modulation) is available. The pulse width of the pulse varies with the voltage
of the modulating waveform. Especially when using an ARB waveform for modulation, a wide range of waveforms is
possible.
Press Waveforms → Pulse → Mod, the parameters of PWM modulation are shown in Figure 9.8.
9.1.8 Pulse Width/Duty Deviation
Width Deviation represents the variation of the modulated waveform pulse width relative to the original pulse width.
Press Width Dev to highlight the parameter, and use the numeric keyboard or arrow keys and knob to input the desired
value.
• The width deviation cannot exceed the current pulse width.
• The width deviation is limited by the minimum pulse width and current edge time setting.
Modulation Function47
Function Explanation
TypePSKPhase shift keying modulation.
SourceInternalThe source is internal.
ExternalThe source is external. Use the [Aux In/Out] connector at the rear panel.
Key FreqSet the frequency at which the output phase shifts between the carrier phase and 180°
(internal modulation only): 1mHz~1MHz.
PolarityPositiveSet the modulating polarity.
Negative
Figure 9.7 Setting Interface of PSK Modulation
Duty Deviation represents the variation (%) of the modulated waveform duty relative to the original duty. Press Duty
Dev to highlight the parameter, and then use the numeric keyboard or arrow keys and knob to input the desired value,
as shown in the Figure 2-33.
• The duty deviation cannot exceed the current pulse duty cycle.
• The duty deviation is limited by the minimum duty cycle and current edge time setting.
• Duty deviation and width deviation are correlative. Once a parameter is changed, the other will be automatically
changed.
Note: The methods of setting other parameters of PWM are similar to AM.
Modulation Function48
Function Description
TypePWMPulse width modulation. The carrier is pulse.
SourceInternalThe source is internal.
ExternalThe source is external. Use the [Aux In/Out] connector at the rear panel.
Width DevSet the width deviation.
Duty DevSet the duty deviation.
ShapeSineChoose the modulating waveform.
Square
Triangle
UpRamp
DnRamp
Noise
Arb
PWM FreqSet the modulating waveform frequency. Frequency range: 1mHz~1MHz (internal source
only).
Figure 9.8 Setting Interface of PWM Modulation
Sweep Function
In the sweep mode, the generator steps from the start frequency to the stop frequency in the sweep time specied by the
user. The waveforms that support sweep include sine, square, ramp and arbitrary.
Press Sweep key to enter the following menu. Set the waveform parameters by using the operation menu.
FunctionExplanation
Sweep timeSet the time span of the sweep in which the frequency changes from the start
frequency to stop frequency.
Start Freq Mid FreqSet the start frequency of the sweep; Set the center frequency of the sweep.
Stop Freq Freq SpanSet the stop frequency of the sweep; Set the frequency span of the sweep.
SourceInternalChoose internal source as a trigger.
ExternalChoose external source as a trigger. Use the [Aux In/Out] connector at the rear
panel.
ManualTrigger a sweep by manual.
Trig OutODisable trigger out.
OnEnable trigger out.
Page 1/2Enter the next page.
Figure 10.1 Sweep Screen (Page 1/2)
1 Sweep Frequency
Use start freq and stop freq or center freq and freq span to set the range of the frequency sweep. Press the key again to
switch between the two sweep range modes.
2 Start Frequency and Stop Frequency
Start Frequency and Stop Frequency are the lower and upper limits of the frequency for sweep. Start Frequency Stop
Frequency.
• Choose Direction → Up, the generator will sweep from Start frequency to Stop frequency.
• Choose Direction → Down, the generator will sweep from Stop frequency to Start frequency.
Sweep Function50
Function Explanation
TypeLinearSet the sweep with linear prole.
LogSet the sweep with logarithmic prole.
Direction UpSweep upward.
DownSweep downward.
Page 2/2Return to the previous page.
Figure 10.2 Setting Interface of Sweep (Page 2/2)
3 Center Frequency and Frequency Span
Center Frequency = (|Start Frequency + Stop Frequency|)/2
Frequency Span = Stop Frequency – Start Frequency
4 Sweep Type
4060B Series provides “Linear” and “Log” sweep proles and the default is “Linear”.
5 Linear Sweep
In linear sweep, the output frequency of the instrument varies linearly in the way of “a number of Hertz per second”.
Choose Sweep → Page 1/2 → Type → Linear, there is a straight line displayed on the waveform on the screen, indicating
that the output frequency varies linearly.
6 Log Sweep
In log sweep, the output frequency of the instrument varies in a logarithmic fashion, that is, the output frequency changes
in the way of “decade per second”. Choose Sweep → Page 1/2 → Type → Log, there is an exponential function curve
displayed on the waveform on the screen, indicating that the output frequency changes in a logarithmic mode.
7 Sweep Trigger Source
The sweep trigger source can be internal, external or manual. The generator will generate a sweep output when a trigger
signal is received and then wait for the next trigger source.
Sweep Function51
Figure 10.3 Linear Sweep Interface
Figure 10.4Log Sweep Interface
8 Internal Trigger
Choose Source → Internal, the generator outputs continuous sweep waveform when internal trigger is selected. The
default is “Internal”. Choose Trig Out → On, the [Aux In/Out] connector at the rear panel will output the trigger signal.
9 External Trigger
Choose Source → External, the generator accepts the trigger signal inputted from the [Aux In/Out] connector at the
rear panel when external trigger is selected. A sweep will be generated once the connector receives a CMOS pulse with
specied polarity. To set the CMOS pulse polarity, choose Edge to select “Up” or “Down”.
10 Manual Trigger
Choose Source → Manual, a sweep will be generated from the corresponding channel when the Trigger softkey is pressed
when manual trigger is selected. Choose Trig Out → On, the [Aux In/Out] connector at the rear panel will output the
trigger signal.
Burst Function
The Burst function can generate versatile waveforms in n this mode. Burst times can last a specic number of waveform
cycles (N-Cycle mode), or when an external gated signals (Gated mode) is applied. Any waveform (except DC) may be
used as the carrier, but noise can only be used in Gated mode.
11.1 Burst Type
4060B Series provides three burst types including N-Cycle, Innite and Gated.
The default is N-Cycle.
Table 11.1 2-20 Relations among burst type, trigger source and carrier
11.1.1 N-Cycle
In N-Cycle mode, the generator will output waveform with a specied number of cycles after receiving the trigger signal.
Waveforms that support N-Cycle burst include sine, square, ramp, pulse and arbitrary.
Press Burst → NCycle → Cycles, and use the numeric keyboard or arrow keys and knob to input the desired cycles. Set
the waveform parameters by using the operation menu, as shown in Figure 2-38 and Figure 2-39.
FunctionExplanation
NCycleUse the N-Cycle mode.
Cycles InniteSet the number of the bursts in N-Cycle. Set the number of the bursts in N-Cycle to be
innite.
Start PhaseSet the start phase of the burst.
Burst PeriodSet the burst period.
SourceInternalChoose internal source as a trigger.
ExternalChoose external source as a trigger. Use the [Aux In/Out] connector at the rear panel.
ManualTrigger a burst by manual.
Page 1/2Enter the next page.
Figure 11.1 N-Cycle Burst Interface (Page 1/2)
Burst Function53
Function Explanation
DelaySet the delay time before the burst starts.
Trig Out ODisable trigger out.
OnEnable trigger out.
Page 2/2Return to the previous page.
Figure 11.2N-Cycle Burst Interface (Page 2/2)
11.1.2 Innite
In innite mode, the cycle number of the waveform is set as an innite value.
The generator outputs a continuous waveform after receiving the trigger signal. Waveforms that support innite mode
include sine, square, ramp, pulse and arbitrary.
Press Burst → NCycle → Innite, and set the trigger source to “external” or
“manual”. The screen will display an innite cycle burst, as shown in Figure 11.3.
Figure 11.3 Innite Burst Interface
11.1.3 Gated
In gated mode, the generator controls the waveform output according to the gate signal level. When the gated signal
is “true”, the generator outputs a continuous waveform. When the gated signal is “false”, the generator rst completes
the output of the current period and then stops. Waveforms that support gated burst include sine, square, ramp, pulse,
noise and arbitrary.
Burst Function54
Press Burst → Gated, to enter the interface in Figure 11.4.
FunctionExplanation
GatedUse the gated mode.
PolarityPositiveSet the polarity for the gated signal.
Negative
Start PhaseSet the start phase of the burst.
Burst PeriodSet the burst Period.
SourceInternalChoose internal source as a trigger.
ExternalChoose external source as a trigger. Use the [Aux In/Out] connector at the rear panel.
Figure 11.4 Gated Burst Interface
11.1.4 Start Phase
Dene the start point in a waveform. The phase varies from 0° to 360°, and the default setting is 0°. For an Arbitrary
Waveform, 0° is the rst waveform point.
11.1.5 Burst Period
Burst Period is only available when the trigger source is internal. It is dened as the time from the start of a burst to the
start of the next one. Choose Burst Period and use the numeric keyboard or arrow keys and knob to input the desired
value.
• Burst Period 0.99s + carrier period × burst number
• If the current burst period set is too short, the generator will increase this value automatically to allow outputting
the specied number of cycles.
11.1.6 Cycles/Innite
Set the number of waveform cycle in an N-Cycle (1 to 50,000 or Innite). If Innite is chosen, then a continuous waveform
will be generated once a trigger occurs.
11.1.7 Delay
Set the time delay between the trigger input and the start of the N-Cycle burst.
Burst Function55
11.1.8 Burst Trigger Source
The burst trigger source can be internal, external or manual. The generator will generate a burst output when a trigger
signal is received and then wait for the next trigger source.
11.1.9 Internal Trigger
Choose Source → Internal, the generator outputs continuous burst waveform when internal trigger is selected. Choose
Trig Out as “Up” or “Down”, the [Aux In/Out] connector at the rear panel will output a trigger signal with specied
edge.
11.1.10 External Trigger
Choose Source → External, the generator accepts the trigger signal inputted from the [Aux In/Out] connector at the rear
panel when external trigger is selected. A burst will be generated once the connector gets a CMOS pulse with specied
polarity. To set the CMOS pulse polarity, choose Edge to select “Up” or “Down”.
11.1.11 Manual Trigger
Choose Source → Manual, a burst will be generated from the corresponding channel when the Trigger softkey is pressed
when manual trigger is selected.
Storage System
12.1 To Store and Recall
4060B Series can store the current instrument state and user-dened arbitrary waveform data in internal or external
memory and recall them when needed.
Press Store/Recall to enter the interface shown in Figures 12.1, 12.2.
Function Description
File Type StateThe setting of the generator;
DataArbitrary waveform le
BrowseView the current directory.
SaveSave the waveform to the specied path.
RecallRecall the waveform or setting information in the specic position of the memory.
DeleteDelete the selected le.
Page 1/2Enter the next page.
Figure 12.1Store/Recall Interface (Page 1/2)
The 4060B Series provides an internal non-volatile memory (C Disk) and a USB Host interface for external memory.
1. Local (C:)
Users can store instrument states and arbitrary waveform les to C Disk.
12.1.1 USB Device (0:)
There is a USB Host interface located on the left side of the front panel which permits users to store/recall waveforms
or update the rmware version by U-Disk. When the generator detects a USB storage device, the screen will show the
drive letter “USB Device (0:)” and display a prompt message “USB device connected.”, as shown in Figure 12.3. After
removing the U-Disk, the screen will display a prompt message “USB device removed.” And “USB Device (0:)” in the
storage menu will disappear.
Note: The 4060B Series can only identify les of which lenames consist of English letters, number and underscore.
If other characters are used, the name may be displayed in the store and recall interface abnormally.
Storage System57
Function Description
CopyCopy the selected le.
PastePaste the selected le.
CancelExit the Store/Recall interface.
Page 2/2Return to the previous page.
Figure 12.2 Store/Recall Interface (Page 2/2)
Figure 12.3Storage System
12.1.2 Browse
• Use the knob to shift between the directories or click the corresponding location on the screen to choose Local (C:)
or USB Device (0:). Choose Browse, press the knob or click the selected folder to open the current directory.
• Use the knob to switch between folder and les under the current directory. Choose Browse, press the knob or click
the selected folder to open the subdirectory. Choose <up>, then choose Brower or press the knob to return to the
upper level directory.
12.1.3 File Type
Choose Store/Recall → File Type to select the desired le type. Available le types are State File and Data File.
Storage System58
12.1.4 State File
Store the instrument state in internal or external memory in “*.xml” format. The state le stored includes waveform
parameters and modulation, sweep, burst parameters of two channels and utility parameters.
12.1.5 Data File
The 4060B Series can recall the data les in “*.csv” or “*.dat” format from the external memory and transfer them into
“*.bin” format then store them in the internal memory. When it is done, the generator will enter the arbitrary waveform
interface automatically.
In addition, users can edit arbitrary waveforms with PC software — EasyWave, download them to the internal memory
through remote interface and store them (in “*.bin” format) in the internal memory.
12.2 File Operation
12.2.1 To Save the Instrument State
Users can store the current instrument state in internal and external memories. The storage will save the selected function
(including the basic waveform parameters, modulation parameters and other utility settings used.)
To save the instrument state, the procedures are given as followed:
1. Choose the le type to store.
Press Store/Recall → File Type → State, and choose state as the storage type.
2. Choose the location of the le.
Choose a desired location by rotating the knob or clicking the corresponding location on the touch screen.
3. Name the le.
Press Save, to enter the following interface.
Function Description
UpCursor upward to select.
DownCursor downward to select.
SelectSelect the current character.
DeleteDelete the current character.
SaveStore the le with the current name.
CancelReturn to the store/Recall interface.
Figure 12.4Filename Input Interface
Storage System59
12.2.2 Select the character
Users can select the desired character from the virtual soft keyboard by using the knob or Up and Down menus. Or touch
the location of the character on the screen directly. Then choose Select to display the character selected in the lename
area.
12.2.3 Delete the character
Use the left and right arrow keys to move the cursor in the le name. Then choose Delete to delete the corresponding
character.
12.2.4 Save the le.
After nishing inputting lename, press Save. The generator will save the le under the currently selected directory with
the specied lename.
12.2.5 To Recall State File or Data File
To recall the instrument state or arbitrary waveform data, the procedures are as follows:
1. Choose the le type.
Press Store/Recall → File Type, and choose state or data as the storage type.
2. Choose the le to be recalled.
Rotate the knob or click the touch screen to select the le you want to recall.
3. Recall the le.
Choose Recall, press the knob or click the location of the le on the screen, the generator will recall the selected le
and display corresponding prompt message when the le is read successfully.
12.2.6 To Delete File
To delete the instrument state or arbitrary waveform data, the procedures are as follows:
1. Choose the le.
Rotate the knob or click the touch screen to select the le you want to delete.
2. Delete the le.
Choose Delete, the generator will display prompt message ‘Delete the le?’ Then press Accept, the generator will delete
the currently selected le.
12.2.7 To Copy and Paste File
4060B Series supports the internal and external storage to copy les from each other. For example, copy an arbitrary
wave le in the U-disk to the instrument, the procedure is as follows:
1. Choose the le type.
Press Store/Recall → File Type, and choose “Data” as the storage type.
2. Choose the le to be copied.
Rotate the knob to select USB Device (0:) and press the knob to open its directory. Then rotate the knob to select the
le you want to copy and press Page 1/2 → Copy.
3. Paste the le.
Rotate the knob to select Local (C:) and press the knob to open its directory.
Then press Paste.
Utility Function
With the Utility function, the user can set the parameters of the generator such as Sync, Interface, System Setting, Self
Test and Frequency Counter, etc. Press Utility to enter the utility menu, as shown in Figure 13.1.
Figure 13.1 Utility Menu
Utility Function61
System Set the system conguration.
Test/Cal Test and calibrate the instrument.
Counter Frequency counter setting.
Output Setup Set the output parameters of CH1 and CH2.
CH Copy Coupling Set the track, channel coupling or channel copy function.
Interface Set the parameters of remote interfaces.
Sync Set the sync output.
CLKSource Choose the system clock source, internal or external.
Help View the help information.
OverVoltage Protection Turn on/o the overvoltage protection function.
13.1 System Settings
Press Utility → System, to enter the following interface.
Figure 13.2 System Menu
Number format Set the number format.
Utility Function62
Language Set the language by pressing the button to toggle between English and Chinese.
PowerOn Set the power on behavior to load the defaults, or last used settings.
Set to Default Set all the settings to default values.
Beeper Enable or Disable the beep.
ScrnSvr Enable or disable the screen saver and set the time.
System Info View the system information
Firmware Update Update the rmware by the U-disk.
Bandwidth Update Update the bandwidth of the generator.
Done Save the current settings and return to the Utility menu.
13.2 Number Format
Press Utility → System → Number Format, to enter the following interface.
Figure 13.3 Set the Number Format
Point Select either “.” or “,” to represent a decimal point.
Separator Select either a space, comma or no separator to show in large numbers.
Done Save the current settings and return to the System menu.
According to the dierent choices of the decimal point and the separator, the format can have various forms.
13.3 Language Setup
The generator oers two languages (English and Chinese). Press Utility → System → Language, to select the desired
language. This setting is stored in non-volatile memory and will not be inuenced by the Set To Default operation.
13.4 Power On
Choose the 4060B Series’s setting to load when the generator is powered on. Two choices are available: the default
setting and the last settings set when the unit was last powered down. Once selected, the setting will be applied when
Utility Function63
Figure 13.4 Chinese Interface
the instrument is powered on. This setting is stored in non-volatile memory and will not be inuenced by the Set To
Default operation.
Last includes all system parameters and states, except channel output state.
Default denotes the factory defaults except certain parameters (such as Language).
13.5 Set to Default
Press Utility → System → Set To Default, to set the system to the default setting. The default settings of the system
are as shown in Table 13.1.
13.6 Beep
Enable or disable the beeper. Press Utility →System → Beeper to select “On” or “O” and the default is “On”.
13.7 Screen Saver
Enable or disable screen saver. Press Utility → System → Page 1/2 → ScrnSvr to select “On” or “O” and the default
is “O”. Screen saver will be on if no action is taken within the time that you have selected. Click the touch screen or
Press any key to resume.
13.8 System Info
Select the System Info option of the utility menu to view the generator’s system information, including startup times,
software version, hardware version, model and serial number.
13.9 Software Update
The software version and conguration le of the generator can be updated directly via U-disk. Follow the steps below:
1. Insert U-disk with rmware update le (.ADS) and conguration le (.CFG) to USB host interface on the front panel
of the generator.
3. Select the rmware le (.ADS) and choose Recall to update the system software.
4. After the updating is nished, the generator will restart automatically.
Note: Powering o the generator during a rmware update can corrupt the unit.
Note: A confmguration le (.CFG) may or may not be included with a given rmware update. If a CFG le is not
included with a rmware update then it will not be required for that update.
13.10 Built-in Help
The 4060B Series provides a built-in help system, by which users can view the help information at any time when operating
the instrument. Press Utility → System → Page 1/2 → Help to enter the following interface.
UP Cursor upward to select.
Utility Function65
Figure 13.5 System Information Interface
Figure 13.6 Help Menu
Down Cursor downward to select.
Select Read the currently selected help information.
Cancel Exit the built-in help system.
13.11 Test/Cal
Choose Utility → Test/Cal, to enter the following interface.
SelfTest Perform a system self-test.
TouchCal Do a touch screen calibration.
Return Return to the Utility menu.
13.11.1 Self Test
Press Utility → Test/Cal → SelfTest, to enter the following menu.
ScrTest Run screen test program.
Utility Function66
Figure 13.7 Test/Cal function Menu
Figure 13.8Self Test Interface
KeyTest Run keyboard test program.
LEDTest Run key indicator lights test program.
BoardTest Run hardware circuit self-test program.
Cancel Return to the Test/Cal menu.
13.11.2 Screen Test
Select “ScrTest” to enter the screen test interface. The prompt message ‘Please press ‘7’ key to continue, press ‘8’ key
to exit.’ is displayed. Press the ‘7’ key for test and observe if there is any serious color deviation, bad pixel or display
error.
13.11.3 Key Test
Select KeyTest to enter the keyboard test interface, the on-screen white rectangle shapes represent the front panel keys.
The circle between two arrows represents the knob. Test all keys and knob and also verify that all the backlight keys
illuminate correctly.
The corresponding area of tested keys or knob would display in blue color.
The top of the screen displays ‘Please press ‘8’ key three times to exit.’
Utility Function67
13.11.4 LED Test
Select LEDTest to enter the LED test interface, the on-screen white rectangle shapes represent the front panel keys. The
prompt message ‘Please press ‘7’
Key to continue, press ‘8’ Key to exit.’ is displayed. Press the ‘7’ key continuously for testing and when a key is lighted,
the corresponding area on the screen will display in blue color.
Figure 13.9 LED Test Interface
13.11.5 Board Test
Select BoardTest to enter the following interface.
Figure 13.10Board Test Interface
13.11.6 Touch Adjust
Use the function regularly to calibrate the touch screen, which makes it more accurate when the nger or touch pen
touches the screen and avoids any misoperation.
Press Utility → Test/Cal → TouchCal, to enter the following interface.
According to the messages on screen, click the red circle on screen in sequence. After touch calibration is done, the
system will display the following tip. Then press any key or touch the screen to exit the current interface.
Utility Function68
13.12 Frequency Counter
The 4060B Series provides a frequency counter which could measure frequencies between 100mHz to 200MHz. The dual
channels can still output normally when the counter is enabled. Press Utility → Counter, to enter the following interface.
Period Mode
Frequency Mode
Figure 13.11 Frequency Counter Interface
State Enable or Disable the counter.
Frequency Measured frequency.
Period Measured period.
PWidth Measured positive width.
NWidth Measured negative width.
RefFreq Set the reference frequency. System will calculate the deviation between the measured frequency
and the reference frequency automatically.
TrigLev Set the trigger level voltage.
Duty Measured duty cycle.
Setup Set the counter conguration.
Utility Function69
Cancel Exit the frequency counter.
13.12.1 Counter Setup
Figure 13.12Counter Setup Interface
Mode Set the coupling mode to DC or AC
HFR High frequency rejection lter.
Default Set the frequency counter settings to default.
Done Save the current settings and return to the previous menu.
13.13 Parameters to be measured
The frequency counter on the 4060B Series can measure parameters including frequency, period, duty, positive pulse
width and negative pulse width.
13.14 Reference Frequency
System will calculate the deviation between the measured frequency and the reference frequency automatically.
13.15 Trigger Level
Sets the trigger level of the measurement system. The system triggers and obtains the measurement readings when the
input signal reaches the specied trigger level. The default is 0V and the available range is from -3V to 1.5V. Choose
TrigLev and use the numeric keyboard to input the desired value and select the unit (V or mV) from the pop-up menu.
Or use the knob and arrow keys to change the parameter value.
13.16 Coupling Mode
Sets the coupling model of the input signal to “AC” or “DC”. The default is “AC”.
Utility Function70
13.17 High Frequency Rejection
High Frequency Rejection can be used to lter out the high-frequency components of a measured signal and improve the
measurement accuracy in low-frequency signal measurement. Press HFR to enable or disable this function. The default
is “O”.
• Enable High Frequency Rejection when low-frequency signal with lower than a 250kHz frequency is measured to lter
out the high-frequency noise interference.
• Disable High Frequency Rejection when a signal with a frequency higher than 250 KHz is measured. The maximum
frequency that can be counted is 200 MHz.
13.18 Output
Press Utility → Output to enter the following interface.
Figure 13.13Output Setup Interface
13.19 Load
For the [CH1] and [CH2] connectors on the front panel, the generator has an output impedance of 50Ω. If the actual
load does not match the set load, the displayed voltage will not be the same as the output voltage. This function is used
to match the displayed voltage with the expected one. This setting does not actually change the output impedance to
any other value.
13.19.1 Steps for setting the load:
Press Utility → Output Setup → Load, to set the output load. The load parameter shown on the down bottom is the
default setting when the power is on or the pre-set load value.
High Impedance: displayed as HiZ;
Load: the default is 50Ωand the range is 50Ωto 100kΩ.
Note: Continue pressing the corresponding output key for two seconds to switch between High Impedance and 50Ω.
13.20 Polarity
Press Utility → Output Setup → Polarity to set the output signal as normal or inverted. The waveform’s inversion is
relative to the oset voltage, as shown in the following gure.
Utility Function71
Note: The Sync signal related to the waveform is not inverted when the waveform is inverted.
13.21 EqPhase
Press Utility → Output Setup → EqPhase to align the phases of CH1 and CH2.
Choosing the menu will re-congure two channels and enable the generator to output with specied frequency and start
phase. For two signals whose frequencies are the same or a multiple thereof, this operation will align their phases.
13.22 Waveforms Combination Mode
The CH1 output port of the 4060B Series outputs the waveform of CH1 in the general mode, while the waveform of
CH1+CH2 can be output in the combined mode. Similarly, the CH2 output port of 4060B Series outputs the waveform
of CH2 in the general mode while the waveform of CH1+CH2 can be output in the combined mode.
Press Utility → Output Setup → Wave Combine to enter the waveforms combining interface, as shown in the following
gure.
CH1 Switch Output the waveform of CH1.
CH1+CH2 Output the waveform of CH1+CH2.
CH2 Switch Output the waveform of CH2.
CH1+CH2 Output the waveform of CH1+CH2.
Return Save the current operation and exit the current interface.
Note: When the waveforms combining function is enabled, the load of two channels will be set to the same automat-
ically, default using the load value of the currently operated channel.
13.23 CH Copy/Coupling
The 4060B Series supports state and waveform copy function between its two channels. That is to say, it copies all
parameters and states (including the channel output state) and arbitrary waveform data of one channel to the other one.
Press Utility → CH Copy Coupling → Channel Copy, to enter the following interface.
CH1=>CH2 Copy all parameters and states of CH1 to CH2.
CH2=>CH1 Copy all parameters and states of CH2 to CH1.
Accept Perform the current selection and return to the Utility menu.
Cancel Give up the current selection and return to the Utility menu.
Note: Channel coupling or track function and channel copy function are mutually exclusive. When channel coupling
or track function is enabled, the menu Channel Copy is hidden.
13.23.1 Channel Coupling
The 4060B Series supports frequency, amplitude and phase coupling. Users can set the frequency deviation/ratio,
amplitude deviation/ratio or phase deviation /ratio of the two channels. When coupling is enabled, CH1 and CH2 can
be modied simultaneously. When the frequency, amplitude or phase of one channel (as the reference) is changed, the
corresponding parameter of the other channel will be changed automatically and always keeps the specied frequency
deviation/ratio, amplitude deviation/ratio or phase deviation /ratio relative to the base channel.
Press Utility → CH Copy Coupling → Channel Coupling, to menu.
Utility Function72
Figure 13.14 Waveforms Combining Interface
13.23.2 Frequency Coupling
1. To Enable Frequency Coupling Function: Press FreqCoup to turn frequency coupling “On” or “O”. The default is
“O”.
2. To Select Frequency Coupling Mode: Press FreqMode to choose “Deviation” or “Ratio”, and then use the numeric
keyboard or knob and arrow keys to input the desired value.
Utility Function73
Figure 13.15 Channel Copy Interface
Deviation the frequency deviation between CH1 and CH2. 𝐹 𝑟𝑒𝑞
𝐹 𝑟𝑒𝑞
Ratio The frequency ratio of CH1 and CH2.
𝐹 𝑟𝑒𝑞
𝐶𝐻2
= 𝐹 𝑟𝑒𝑞𝑅𝑎𝑡𝑖𝑜.
𝐶𝐻1
𝐶𝐻2
− 𝐹 𝑟𝑒𝑞
= 𝐹 𝑟𝑒𝑞𝐷𝑒𝑣.
𝐶𝐻1
13.23.3 Amplitude Coupling
1. To Enable Amplitude Coupling Function: Press AmplCoup to turn amplitude coupling “On” or “O”. The default is
“O”.
2. To Select Amplitude Coupling Mode: Press AmplMode to choose “Deviation” or “Ratio”, and then use the numeric
keyboard or knob and arrow keys to input the desired value.
Deviation The amplitude deviation between CH1 and CH2. 𝐴𝑚𝑝𝑙
𝐴𝑚𝑝𝑙
Ratio the amplitude ratio of CH1 and CH2.
𝐴𝑚𝑝𝑙
𝐶𝐻2
= 𝐴𝑚𝑝𝑙𝑅𝑎𝑡𝑖𝑜.
𝐶𝐻1
𝐶𝐻2
− 𝐴𝑚𝑝𝑙
= 𝐴𝑚𝑝𝑙𝐷𝑒𝑣.
𝐶𝐻1
13.23.4 Phase Coupling
1. To Enable Phase Coupling Function: Press PhaseCoup to turn phase coupling “On” or “O”. The default is “O”.
2. To Select Phase Coupling Mode: Press PhaseMode to choose “Deviation” or “Ratio”, and then use the numeric
keyboard or knob and arrow keys to input the desired value.
𝑃 ℎ𝑎𝑠𝑒
Deviation the phase deviation between CH1 and CH2.
𝑃 ℎ𝑎𝑠𝑒
Ratio the phase ratio of CH1 and CH2.
𝑃 ℎ𝑎𝑠𝑒
𝐶𝐻2
𝐶𝐻1
𝐶𝐻2
𝑃 ℎ𝑎𝑠𝑒
= 𝑃 ℎ𝑎𝑠𝑒𝐷𝑒𝑣
𝐶𝐻1
= 𝑃 ℎ𝑎𝑠𝑒𝑅𝑎𝑡𝑖𝑜.
• Channel coupling is only available when both the waveforms of the two channels are basic waveforms including Sine,
Square, Ramp and Arbitrary.
• When the Phase Coupling function is enabled, if the phase of one channel is changed, the phase of the other channel
will be changed accordingly. At this point, aligning phase between the two channels can be achieved without executing
the Eqphase operation.
• Channel coupling and channel function are mutually exclusive. When channel coupling is enabled, the menu Channel
Copy is hidden.
Utility Function74
13.24 Channel Tracking
When the track function is enabled, by changing the parameters or states of CH1, the corresponding parameters or states
of CH2 will be adjusted to the same values or states automatically. At this point, the dual channels can output the same
signal.
Choose Utility → CH Copy Coupling → Track to enable or disable the track function. When the track function is enabled,
channel copy and coupling functions are disabled; the user interface is switched to CH1 and cannot be switched to CH2,
as shown in the following gure.
Figure 13.16 Track Interface
Press PhaseDev to enter the following interface. Then use the numeric keyboard or knob and arrow keys to input the
desired value for the phase deviation between CH1 and CH2. The resulting signal is represented by:
Phase
CH2
-Phase
=PhaseDev.
CH1
Figure 13.17 Phase Deviation Interface
Output Synchronization
The generator provides Sync output through the [Aux In/Out] connector on the rear panel. When the synchronization is
on, the port can output a CMOS signal with the same frequency as basic waveforms (except Noise and DC), arbitrary
waveforms, and modulated waveforms (except external modulation).
Function Description
StateOClose the sync output
OnOpen the sync output
Channel CH1Set the sync signal of CH1.
CH2Set the sync signal of CH2.
AcceptSave the current settings and return to the Utility menu.
CancelGive up the current settings and return to the Utility menu.
Figure 14.1Sync Output Interface
14.1 Sync Signals of Dierent Waveforms
1. When the frequency of the waveform is less than or equal to 10MHz, the sync signal is a Pulse with 50ns pulse width
and the same frequency as the waveform.
2. When the frequency of the waveform is greater than 10MHz, there is no sync signal output.
3. Noise and DC: there is no sync signal output.
14.1.1 Modulated Waveform
• When internal modulation is selected, the sync signal is a Pulse with 50ns pulse width.
For AM, FM, PM and PWM, the frequency of the sync signal is the modulating frequency.
For ASK, FSK and PSK, the frequency of the sync signal is the key frequency.
• When external modulation is selected, there is no sync signal output, for the [Aux In/Out] connector on the rear panel
is used to input external modulating signal.
14.1.2 Sweep and Burst Waveform
When Sweep or Burst function is turned on, there is no sync signal output and the Sync menu is hidden.
Clock Source
The 4060B Series provides an internal 10MHz clock source. It also can accept external clock source form the [10 MHz
In/Out] connector at the rear panel. It can also output the clock source from the [10 MHz In/Out] connector for other
devices.
Press Utility → Page 1/2 → Clock → Source to select “Internal” or “External”.
If “External” is selected, the instrument will detect whether a valid external clock signal is input from the [10MHz In/Out]
connector at the rear panel. If not, the prompt message “No external clock source!” would be displayed and the clock
source would be switched to “Internal”.
15.1 Sync methods for two or more instruments:
Connect the [10MHz In/Out] connector of generator A (using internal clock) to the [10MHz In/Out] connector of
generator B (using external clock) and set the output frequencies of A and B as a same value to realize synchronization.
15.2 Synchronization among multiple instruments
Divide the 10MHz clock source of a generator (using internal clock) into multiple channels, and then connect them to
the [10MHz In/Out] connectors of other generators (using External clock), and nally set the output frequencies of all
the generators as a same value to realize synchronization.
Channel Phase Mode
Press Utility → Page 1/2 → Mode to enter the mode setup Interface, as shown in Figure 16.1.
Figure 16.1Mode Setup Interface
Phase-locked Mode When changing the frequency, the DDSs of both channels reset, and the phase deviation between
CH1 and CH2 is maintained.
Independent Mode When changing the frequency, neither channels’ DDS resets and the phase deviation between CH1
and CH2 changes at random. When the independent mode is enabled, the phase parameter cannot be modied and the menu Phase is hidden.
Overvoltage Protection
Choose Utility → Page 1/2 → OverVoltage Protection to turn on or o the function.
If the state is set to ON, overvoltage protection of CH1 and CH2 will take eect once any of the following conditions is
met. When overvoltage protection occurs, a message will be displayed and the output is disabled.
• The absolute value of input voltage is higher than 11V±0.5V when the amplitude of the generator is higher than or
equal to 3.2Vpp or the DC oset is higher than or equal to |2VDC|.
• The absolute value of input voltage is higher than 4V±0.5V when the amplitude of the generator is lower than 3.2Vpp
or the DC oset is lower than |2VDC|.
Remote Interface
The 4060B Series can be controlled remotely via USB, LAN and GPIB (option) interfaces. Users can set the corresponding
interface according to their needs.
Press Utility → Page 1/2 → Interface to open the following menu. The user can set LAN parameters or GPIB address.
Figure 18.1 Interface Settings
GPIB Set the GPIB address.
LAN State Turn on LAN.
LAN Setup Set the IP address, subnet mask, gateway and DHCP.
Accept Save the current settings and return to the Utility menu.
The 4060B Series can be controlled remotely via the following two methods:
18.1 User-dened programming
Users can program and control the instrument by using the SCPI commands (Standard Commands for Programmable
Instruments). For more information about the commands and programming, please refer to “Remote Control Manual”.
18.2 Remote Control via USB
The 4060B Series can communicate with a PC through the USBTMC protocol.
You are suggested to do as the following steps.
1. Connect the device: Connect the USB Device interface at the rear panel of 4060B Series with the PC via a USB
cable.
2. Install the USB driver: NI Visa is recommended.
3. Communicate with a remote PC: Open Measurement & Automation Explorer of NI and choose the corresponding
resource name. Then click “Open VISA Test Panel” to turn on the remote command control panel through which
you can send commands and read data.
Remote Interface80
18.3 Remote Control via GPIB
Each device connected to GPIB interface must have a unique address. The default value is 18 and values range from 1
to 30. The selected address is stored in non-volatile memory.
Choose Utility → Page 1/2 → Interface → GPIB to enter the following interface.
18.4 Remote Control via LAN
The 4060B Series can communicate with a PC through the LAN interface. Users can view and modify the LAN parameters.
Choose Utility → Page 1/2 → Interface → LAN State to turn on LAN. Then choose LAN Setup to enter the interface
conguration values. If available, DHCP (Dynamic Host Conguration Protocol), can set the appropriate values for
networks that provide a DHCP server.
Specications
Note: All specications apply to the unit after:
1. A temperature stabilization time of 15 minutes over an ambient temperature range of 23∘C±5∘C.
2. Short correction operation performed before making measurement.
Specications are subject to change without notice.
(4) VOH = 3.8 V (IOH = -8 mA), VOL = 0.44 V (IOL = 8 mA)
1 Common Waveforms
StairUp Stair waveform, rising
StairDn Stair waveform, falling
StairUD Stair waveform, rising and falling
Trapezia Trapezoidal waveform
Ppulse Positive pulse
Npulse Negative pulse
UpRamp UpRamp waveform
DnRamp DnRamp waveform
SineTra Sine-Tra waveform
SineVer Sine-Ver waveform
Appendix: Waveforms
20.1 Math Waveforms
ExpFall Exponential Decay function
ExpRise Exponential Rise function
LogFall Logarhythmic Fall function
LogRise Logarhythmic Rise function
Sqrt Square Root function
Root3 Root3 function
X^2 X2function
X^3 X3function
Airy Airy function
Besselj Bessel I function
Bessely Bessel II function
Dirichlet Dirichlet function
Erf Error function
Erfc Complementary error function
ErfcInv Inverted complementary error function
ErfInv Inverted error function
Laguerre 4-times Laguerre polynomial
Appendix: Waveforms83
Legend 5-times Legend polynomial
Versiera Versiera
Sinc Sinc function
Gaussian Gaussian function
Dlorentz Dlorentz function
Haversine Haversine function
Lorentz Lorentz function
Gauspuls Gauspuls signal
Gmonopuls Gmonopuls signal
Tripuls Tripuls signal
Weibull Weibull distribution
LogNormal LogNormal Gaussian distribution
Laplace Laplace distribution
Maxwell Maxwell distribution
Rayleigh Rayleigh distribution
Cauchy Cauchy distribution
20.2 Engine Waveforms
Cardiac Cardiac signal
Quake Analog quake waveform
Chirp Chirp signal
TwoTone TwoTone signal
SNR SNR signal
AmpALT Gain oscillation curve
AttALT Attenuation oscillation curve
RoundHalf RoundHalf Waveform
RoundsPM RoundsPM Waveform
BlaseiWave Time-velocity curve of explosive oscillation
DampedOsc Time-displacement curve of damped oscillation
SwingOsc Kinetic energy – time curve of swing oscillation
Discharge Discharge curve of NI-MH battery
Pahcur Current waveform of DC brushless motor
Combin Combination function
Appendix: Waveforms84
SCR SCR ring prole
TV TV signal
Voice Voice signal
Surge Surge signal
Radar Analog radar signal
Ripple Ripple wave of battery
Gamma Gamma signal
StepResp Step-response signal
BandLimited Bandwidth-limited signal
CPulse C-Pulse
CWPulse CW pulse
GateVibr Gate self-oscillation signal
LFMPulse Linear FM pulse
MCNoise Mechanical construction noise
20.3 Window Waveforms
Hamming Hamming window
Hanning Hanning window
Kaiser Kaiser window
Blackman Blackman window
GaussiWin GaussiWin window
Triangle Triangle window (Fejer window)
BlackmanH BlackmanH window
Bartlett-Hann Bartlett-Hann window
Bartlett Bartlett window
BarthannWin Modied Bartlett-Hann window
BohmanWin BohmanWin window
ChebWin ChebWin window
FlattopWin Flat top weighted window
ParzenWin ParzenWin window
TaylorWin TaylorWin window
TukeyWin TukeyWin (tapered cosine) window
Appendix: Waveforms85
20.4 Trigonometric Waveforms
Tan Tangent
Cot Cotangent
Sec Secant
Csc Cosecant
Asin Arc sine
Acos Arc cosine
Atan Arc tangent
ACot Arc cotangent
CosH Hyperbolic cosine
CosInt Integral cosine
Coth Hyperbolic cotangent
Csch Hyperbolic cosecant
SecH Hyperbolic secant
SinH Hyperbolic sine
SinInt Integral sine
TanH Hyperbolic tangent
ACosH Arc hyperbolic cosine
ASecH Arc hyperbolic secant
ASinH Arc hyperbolic sine
ATanH Arc hyperbolic tangent
ACsch Arc hyperbolic cosecant
ACoth Arc hyperbolic cotangent
20.5 Square Waveforms
Square waveforms with duty cycle percentages from 1 to 99% are shown in a table after pressing the “Square” button.
They are organized in 3 pages. If the desired percentage is higher than shown on screen, select the last eld in the table
“pageX” and press the knob to navigate to the next page. To return to a lower percentage than shown, do the same by
selecting the rst cell in the table showing “pageX” in the upper left corner of the table.
SquareDutyXX Is the name of the waveform with XX dening the percentage duty cycle. “XX” is replaced by
the desired percentage. For example, 10% is “SquareWave10”.
20.6 Medical Waveforms
Waveform Description
EOG Electro-Oculogram
Appendix: Waveforms86
EEG Electroencephalogram
EMG Electromyogram
Pulseilogram Pulseilogram
ResSpeed Speed curve of the respiration
ECG1 Electrocardiogram 1
ECG2 Electrocardiogram 2
ECG3 Electrocardiogram 3
ECG4 Electrocardiogram 4
ECG5 Electrocardiogram 5
ECG6 Electrocardiogram 6
ECG7 Electrocardiogram 7
ECG8 Electrocardiogram 8
ECG9 Electrocardiogram 9
ECG10 Electrocardiogram 10
ECG11 Electrocardiogram 11
ECG12 Electrocardiogram 12
ECG13 Electrocardiogram 13
ECG14 Electrocardiogram 14
ECG15 Electrocardiogram 15
LFPulse Waveform of the low frequency pulse electrotherapy
Tens1 Waveform 1 of the nerve stimulation electrotherapy
Tens2 Waveform 2 of the nerve stimulation electrotherapy
Tens3 Waveform 3 of the nerve stimulation electrotherapy
20.7 Modulated Waveforms
AM Sectional sine AM signal
FM Sectional sine FM signal
PFM Sectional pulse FM signal
PM Sectional sine PM signal l
PWM Sectional PWM signal
20.8 Filter Waveforms
Butterworth Butterworth lter
Chebyshev1 Chebyshev1 lter
Appendix: Waveforms87
Chebyshev2 Chebyshev2 lter
20.9 Demo Waveforms
demo1_375pts TureArb waveform 1 375 pts
demo1_16kpts TureArb waveform 1 16384 pts
demo2_3kpts TureArb waveform 2 3000 pts
demo2_16kpts TureArb waveform 2 16384 pts
Daily Maintenance
Do not store or leave the instrument in where the display screen will be exposed to direct sunlight for a long period of
time.
CAUTION: To avoid damage to the instrument, do not expose it to spray, liquid, or solvent.
21.1 Cleaning
If the instrument requires cleaning, disconnect it from all power sources and clean it with a mild detergent and water.
Make sure the instrument is completely dry before reconnecting it to a power source.
To clean the exterior surface, perform the following steps:
1. Remove loose dust on the outside of the instrument with a lint-free cloth. When cleaning the touch screen, be careful
to avoid scratching the transparent plastic protective screen.
2. Use a soft cloth dampened with water to clean the instrument.
WARNING: To avoid any damage to the surface of the instrument, do not use any abrasive or chemical cleaning agents.
LIMITED THREE-YEAR WARRANTY
B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free
from defects in workmanship and materials for a period of three years from date of purchase.
B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned
product must be accompanied by proof of the purchase date in the form of a sales receipt.
To help us better serve you, please complete the warranty registration for your new instrument via our website www.bkprecision.com
Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of
unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or removed.
B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting
from loss of use. Some states do not allow limitations of incidental or consequential damages. So the above limitation
or exclusion may not apply to you.
This warranty gives you specic rights and you may have other rights, which vary from state-to-state.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
www.bkprecision.com
714-921-9095
Service Information
Warranty Service
Please go to the support and service section on our website at bkprecision.com to obtain an RMA #. Return the product
in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem
and return any leads, probes, connectors and accessories that you are using with the device.
Non-Warranty Service
Please go to the support and service section on our website at bkprecision.com to obtain an RMA #. Return the product
in the original packaging to the address below. Clearly state on the RMA the performance problem and return any leads,
probes, connectors and accessories that you are using with the device. Customers not on an open account must include
payment in the form of a money order or credit card. For the most current repair charges please refer to the service and
support section on our website.
Return all merchandise to B&K Precision Corp. with prepaid shipping. The at-rate repair charge for Non-Warranty
Service does not include return shipping. Return shipping to locations in North America is included for Warranty Service.
For overnight shipments and non-North American shipping fees please contact B&K Precision Corp.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
http://bkprecision.com
714-921-9095
Include with the returned instrument your complete return shipping address, contact name, phone number and description
of problem.
Version: December 19, 2019
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.