The following safety precautions apply to both operating and maintenance personnel and must be followed during all
phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
• Read and understand the safety and operational information in this manual.
• Apply all the listed safety precautions.
• Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument
at an incorrect line voltage will void the warranty.
• Make all connections to the instrument before applying power.
• Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design,
manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply
with these requirements.
2
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available and the
voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is
a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested,
which denes the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the
mains supply. The voltages in the environment are typically derived from a limited-energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains
supply at a standard wall outlet or similar sources. Example measurement environments are portable
tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains
installation of a building. Examples are measurements inside a building’s circuit breaker panel
or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary
power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specied in this manual
for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the
instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating
of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be
115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your
country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This
instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power
cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power
cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts
(including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the
recommended three-conductor AC line power cable may result in injury or death.
3
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only
and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
• In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided particulates.
• In relative humidity conditions outside the instrument’s specications.
• In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense
on the instrument.
• In air temperatures exceeding the specied operating temperatures.
• In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
• In environments with restricted cooling air ow, even if the air temperatures are within specications.
• In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is
0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by this instrument
may be outside specications if the instrument is used in non-oce-type environments. Such environments may include
rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise,
strong electric elds, or strong magnetic elds.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the
instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated,
and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the
instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases,
or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not
for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support
device or system.
4
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must
be made by qualied service-trained maintenance personnel who are aware of the hazards involved when the instrument’s
covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may
exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect
all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there
are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device
before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the
measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt
any service or adjustment unless another person capable of rendering rst aid and resuscitation is present.
Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit
exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the instrument’s fuse
requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses.
Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specied in this manual or on
the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a re. Failure
to use the specied fuses will void the warranty.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to
B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
• Do not place heavy objects on the instrument.
• Do not obstruct cooling air ow to the instrument.
• Do not place a hot soldering iron on the instrument.
• Do not pull the instrument with the power cord, connected probe, or connected test lead.
• Do not move the instrument when a probe is connected to a circuit being tested.
Safety Symbols
SymbolDescription
indicates a hazardous situation which, if not avoided, will result in death or serious injury.
5
indicates a hazardous situation which, if not avoided, could result in death or serious injury
indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
This is the In position of the power switch when instrument is ON.
This is the Out position of the power switch when instrument is OFF.
is used to address practices not related to physical injury.
Contents
1Quick Start10
1.1Front Panel10
1.2Rear Panel10
1.3Touch Screen Display10
1.4Waveform Selection and Setup10
1.5Create a simple sine wave11
1.5.1Frequency and Phase12
1.5.2Amplitude, and Oset12
1.6To Turn On/O Output13
1.6.1Function Keys13
2Sine Waveform14
2.1Frequency/Period14
2.2Amplitude14
2.3Oset15
2.4Phase16
2.5Harmonics16
2.5.1Harmonic Type17
2.5.2Harmonic Order18
2.5.3Harmonic Amplitude18
2.5.4Harmonic Phase18
3Square Wave19
3.1Duty Cycle19
3.1Oset, High/Low Levels19
3.2Phase20
4Ramp Wave23
4.1Frequency, Amplitude, Oset, High/Low level, and Phase23
4.2Symmetry24
5Pulse Wave27
5.1Frequency, Amplitude, Oset, High/Low level, and Phase27
5.2Pulse Width and DutyCycle27
5.3Rising and Falling Edges28
5.4Delay29
6Noise Wave32
6.1Standard Deviation32
6.2Mean33
6.3Bandwidth33
7DC Wave35
8Arbitrary Wave36
8.1DDS36
8.2TrueArb36
8.2.1Sampling Rate37
8.3Waveform Selection37
8.4Selecting a Built-in Waveform38
8.5Stored Waveform38
9Modulation Function40
9.1AM40
9.1.1To Select Modulation Source41
9.1.2Internal Source41
9.1.3External Source41
9.1.4Modulation Depth41
9.1.5Modulation Frequency41
9.1.5.1DSB-AM42
9.1.5.2FM42
9.1.5.3Frequency Deviation43
9.1.5.4PM43
9.1.5.5Phase Deviation44
9.1.5.6FSK44
9.1.6Key Frequency44
9.1.7Hop Frequency45
9.1.7.1ASK45
9.1.7.2PSK46
9.1.7.3PWM46
9.1.8Pulse Width/Duty Deviation46
10 Sweep Function49
10.1Sweep Frequency49
10.2Start Frequency and Stop Frequency49
10.3Center Frequency and Frequency Span50
10.4Sweep Type50
10.5Linear Sweep50
10.6Log Sweep50
10.7Sweep Trigger Source50
10.8Internal Trigger51
10.9External Trigger51
10.10Manual Trigger51
7
11 Burst Function52
11.1Burst Type52
11.1.1N-Cycle52
11.1.2Innite53
11.1.3Gated53
11.1.4Start Phase54
11.1.5Burst Period54
11.1.6Cycles/Innite54
11.1.7Delay54
11.1.8Burst Trigger Source55
11.1.9Internal Trigger55
11.1.10 External Trigger55
11.1.11 Manual Trigger55
12 Storage System56
12.1To Store and Recall56
12.1.1USB Device (0:)56
12.1.2Browse57
12.1.3File Type57
12.1.4State File58
12.1.5Data File58
12.2File Operation58
12.2.1To Save the Instrument State58
12.2.2Select the character59
12.2.3Delete the character59
12.2.4Save the le.59
12.2.5To Recall State File or Data File59
12.2.6To Delete File59
12.2.7To Copy and Paste File59
13 Utility Function60
13.1System Settings61
13.2Number Format62
13.3Language Setup62
13.4Power On62
13.5Set to Default63
13.6Beep63
13.7Screen Saver63
13.8System Info63
13.9Software Update63
13.10 Built-in Help64
13.11 Test/Cal65
13.11.1 Self Test65
13.11.2 Screen Test66
13.11.3 Key Test66
13.11.4 LED Test67
13.11.5 Board Test67
13.11.6 Touch Adjust67
13.12 Frequency Counter68
13.12.1 Counter Setup69
13.13 Parameters to be measured69
13.14 Reference Frequency69
13.15 Trigger Level69
13.16 Coupling Mode69
13.17 High Frequency Rejection70
13.18 Output70
13.19 Load70
13.19.1 Steps for setting the load:70
13.20 Polarity70
13.21 EqPhase71
13.22 Waveforms Combination Mode71
13.23 CH Copy/Coupling71
13.23.1 Channel Coupling71
13.23.2 Frequency Coupling72
13.23.3 Amplitude Coupling73
13.23.4 Phase Coupling73
13.24 Channel Tracking74
8
14 Output Synchronization75
14.1Sync Signals of Dierent Waveforms75
14.1.1Modulated Waveform75
14.1.2Sweep and Burst Waveform75
15 Clock Source76
15.1Sync methods for two or more instruments:76
15.2Synchronization among multiple instruments76
16 Channel Phase Mode77
17 Overvoltage Protection78
18 Remote Interface79
18.1User-dened programming79
18.2Remote Control via USB79
18.3Remote Control via GPIB80
18.4Remote Control via LAN80
19 Specications81
20 Appendix: Waveforms82
20.1Common Waveforms82
20.1Math Waveforms82
20.2Engine Waveforms83
20.3Window Waveforms84
20.4Trigonometric Waveforms85
20.5Square Waveforms85
20.6Medical Waveforms85
20.7Modulated Waveforms86
20.8Filter Waveforms86
20.9Demo Waveforms87
21 Daily Maintenance88
21.1Cleaning88
22 LIMITED THREE-YEAR WARRANTY89
23 Service Information90
9
Quick Start
1.1 Front Panel
4060B Series front panel includes a touch screen, menu softkeys, numeric keyboard, knob, function keys, arrow keys, and
channel control area as shown in Figure 1.1.
Item Description
1Power Switch
2USB Port (load waveforms, store settings, etc...)
3Touchscreen
4Soft Keys
5Number Pad
6Function and Channel Keys
7Knob and Selection Button
8Arrow Keys
9Channel Outputs
Figure 1.1 Front Panel View
1.2 Rear Panel
The rear panel shown in Figure 1.2 provides multiple interfaces, including Counter, 10MHz In/Out, Aux In/Out, LAN,
USB Device, Earth Terminal and AC Power Supply Input.
1.3 Touch Screen Display
The screen displays the parameters and a the waveform for a single channel. Most of the elds can be accessed either
through the function keys below the screen or by pressing their value on screen. Figure 1.3 shows a view of Channel
1 with a simple waveform (sine wave). The values and parameters available on-screen change depending on the mode,
options and waveform selected.
1.4 Waveform Selection and Setup
Press the “Waveform” button to open the waveform setup menu shown in Figure 1.4. The softkeys show the available
waveforms for that page. The “Page 1/2” button accesses the other page of waveforms shown in the second image in
Figure 1.4.
Quick Start11
Item Description
1Power Input
2USB
3Ethernet
4Counter Input
5Auxillary Input/Output
6Clock Input/Output (10 MHz)
7Ground Connection
Figure 1.2 Rear Panel of 4060B Series
Item Description
1Channel Tab
2Waveform
30 V reference
4Parameters
5Output Settings
6Status indicators (clock, lock and network)
7Settings menu
Figure 1.3 Touch Screen Display
1.5 Create a simple sine wave
By default, the generator starts congure with a 1 kHz, 4 V peak to peak, symmetric to the ground reference waveform
starting at 0 degrees. All of these parameters are changeable.
Quick Start12
Page 1
Page 2
Figure 1.4 Waveform Menu
1.5.1 Frequency and Phase
To set the frequency, press the rst soft-key, the frequence value to the right of the waveform, or the word “frequency”
above the rst soft-key. The eld, when ready for editing, is highlighted in blue as shown in Figure 1.5. Use the knob,
arrow keys, and the number pad to set the desired frequency. As the value is changed the output will follow the changes
as they occur if the output is on. Phase is modied in the same manner.
Figure 1.5 Frequency Setting
1.5.2 Amplitude, and Oset
Like setting the frequency, select the parameter to modify and use the knob and keys to change its value. The oset and
amplitude are related through the eective maximum and minimum voltages. The waveform output is fully dened by
either setting the amplitude and oset, or the high and low level. The high and low level dene the peak values of the
waveform.
Quick Start13
1.6 To Turn On/O Output
The two keys on the right side of the operation panel above each channel output are used to enable and disable the
output. When enabled, the key will light up. When enabled and lit, pressing the button again disables the respective
channel. Each key can also change the load impedance value by pressing and holding the key for 2 seconds.
Note: Load impedance is only modies the signal voltage setting. HiZ expects an impedance much larger than 50,
so the output voltage is the same as the internal driving source voltage. When set to 50, the displayed voltage is half
of this value. If the wrong setting is chosen, the output voltage can by up to double, or down to half of the expected
value. See Section 13.19 for more.
1.6.1 Function Keys
Access to most of the conguration and machine setup is through the function keys. See Figure 1.6.
Figure 1.6 Function Keys
Mod. This is the waveform “Modulation” setup and enable key. When pressed, the modulation set-
tings are applied to the current waveform. See Section 9 for details.
Sweep This is the waveform “Sweep” frequency setup and enable key. This key also applies sweep set-
tings to the current waveform when pressed. See Section 10 for details.
Burst This is the waveform “Burst” setup and enable key. This key also applies burst settings to the
current waveform when pressed. See Section 11 for details.
Parameter This key returns the menu system to the current waveform parameter settings. This is the menu
set that opens when selecting a “waveform”.
Utility This key opens the main menu for conguring the system settings of the generator. Generator
conguration such as the remote interface, clocking, synchronization, and other functions are accessed in this menu.
Store/Recall The le browser is opened by this key giving access to open, save and manage system setups.
Waveforms The set of waveform types is accessed in this menu.
Ch1/Ch2 This key, when pressed, changes the active waveform menu. See the color of the waveform and
the tab brought to the foreground to verify which channel is selected.
Sine Waveform
Press the “Waveforms” key and then press the Sine softkey. The screen shown in Figure 2.1 will open. The parameters
available for sine waveforms include frequency, period, amplitude, high level, low level, oset and phase.
ParameterDescription
Frequency/PeriodSet the signal frequency or period; The current parameter will be switched at a second press.
Amplitude/HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second press.
Oset/LowLevelSet the signal oset or low level; The current parameter will be switched at a second press.
PhaseSet the phase of the signal.
Figure 2.1 Sine Wave Parameters
2.1 Frequency/Period
Frequency is one of the most important parameters of basic waveforms. For dierent instrument models and waveforms,
the available ranges of frequency are dierent. For detailed information, please refer to “4060B Series Datasheet”. The
default frequency is 1 kHz.
1. Press Waveforms → Sine → Frequency, to set the frequency parameter. The frequency shown on the screen when the
instrument is powered on is the default value or the set value of last power down. If Period (rather than Frequency) is
the desired parameter, press Frequency/Period again to enter the Period mode. The current value for the waveform’s
period is now displayed in inverse color. Press the Frequency/Period key once again to return to the Frequency entry
mode.
2. Input the desired frequency.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
Note: When using the numeric keyboard to enter the value, the left arrow key can be used to move the cursor
backward and delete the value of the previous digit.
2.2 Amplitude
The amplitude setting range is limited by the “Load” and “Frequency/Period” settings. For detailed information, please
refer to “4060B Series Datasheet”.
Sine Waveform15
Figure 2.2 Set Frequency
1. Press Waveforms → Sine → Amplitude, to set the amplitude. The amplitude shown on the screen when the instrument
is powered on is the default value or the set value of last power down. If setting the waveform’s high level is desired,
press the Amplitude/HighLevel key again to switch into the high level parameter (the current operation is displayed
in inverse color).
2. Input the desired amplitude.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
Figure 2.3 Set Amplitude
2.3 Oset
The oset setting range is limited by the “Load” and “Amplitude/HighLevel” settings. For detailed information, please
refer to “4060B Series Datasheet”. The default value is 0Vdc.
1. Press Waveforms → Sine → Oset, to set the oset. The oset shown on the screen when the instrument is powered
on is the default value or the set value of last power down. If you want to set the waveform by low level, press the
Oset/LowLevel key again, to switch into the low level parameter (the current operation is displayed in inverse color).
2. Input the desired oset.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
Sine Waveform16
Figure 2.4 Set Oset
2.4 Phase
1. Press Waveforms → Sine → Phase, to set the phase. The Phase shown on the screen when the instrument is powered
on is the default value or the set value of last power down.
2. Input the desired phase.
Use the numeric keyboard to input the parameter value directly and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value.
Figure 2.5 Set Phase
Note When the independent mode is enabled, the phase parameter cannot be modied
2.5 Harmonics
The 4060B Series can be used as a harmonic generator to output harmonics with specied order, amplitude and phase.
According to the Fourier transform, a periodic time domain waveform is the superposition of a series of sine waveforms
as shown in the equation below:
Sine Waveform17
Figure 2.6
Press Waveforms → Sine → Harmonic and choose “On”, then press
ItemDescription
TypeSet the harmonic type to “odd”, “ever” or “all”.
OrderSet the order of the harmonic.
Harmonic Ampl Set the amplitude of the harmonic.
Harmonic Phase Set the phase of the harmonic.
CancelReturn to the sine parameters menu.
Figure 2.7Harmonic Interface
2.5.1 Harmonic Type
The 4060B Series can output odd harmonics, ever harmonics and user-dened orders of harmonics. After entering the
harmonic setting menu, press Type to select the desired harmonic type.
1. Press Even, the instrument will output fundamental waveform and even harmonics.
2. Press Odd, the instrument will output fundamental waveform and odd harmonics.
3. Press All, the instrument will output fundamental waveform and all the user-dened orders of harmonics.
Sine Waveform18
2.5.2 Harmonic Order
After entering the harmonic setting menu, press Order, the use the numeric keyboard or knob to input the desired value.
• The range is limited by the maximum output frequency of the instrument and current fundamental waveform frequency.
• Range: 2 to maximum output frequency of the instrument ÷ current fundamental waveform frequency The maximum
is 10.
2.5.3 Harmonic Amplitude
After entering the harmonic setting menu, press Harmonic Ampl to set the harmonic amplitude of each order.
1. Press Order to select the sequence number of the harmonic to be set.
2. Press Harmonic Ampl to set the amplitude of the harmonic selected. Use the arrow keys and knob to change the
value. Or use the numeric keyboard to input the amplitude value and then select the desired unit from the pop-up
menu. The units available are Vpp, mVpp and dBc.
2.5.4 Harmonic Phase
After entering the harmonic setting menu, press Harmonic Phase to set the harmonic phase of each order.
1. Press Order to select the sequence number of the harmonic to be set.
2. Press Harmonic Phase to set the phase of the harmonic selected. Use the arrow keys and knob to change the value.
Or use the numeric keyboard to input the phase value and then select the unit °.
Square Wave
Press Waveforms key to select the waveform function, and press the Square softkey. The square waveform parameters
are set by using the Square operation menu.
The parameters of square waveforms include frequency/period, amplitude/high level, oset/low level, phase and duty. As
shown in Figure 3.1, select DutyCycle. The duty cycle parameter area is highlighted in the parameter display window,
and users can set the duty cycle value here.
FunctionDescription
Frequency/ PeriodSet the signal frequency or period; The current parameter will be switched at a second press.
Amplitude/ HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second
press.
Oset/ LowLevelSet the signal oset or low level; The current parameter will be switched at a second press.
PhaseSet the phase of the signal.
DutyCycleSet the duty cycle for square waveform.
Figure 3.1 Square Wave Parameters
1 Duty Cycle
The ratio of the amount of time the pulse is in the high state and the waveform’s period.
The duty cycle setting range is limited by the “Frequency/Period” setting. For detailed information, please refer to
“4060B Series Datasheet”. The default value is 50%.
1. Press Waveforms → Square → DutyCycle, to set the duty cycle. The duty cycle shown on the screen when the
instrument is powered on is the default value or the set value of last power down.
2. Input the desired Duty Cycle. Use the numeric keyboard to input the parameter value directly and press the corresponding key to select the parameter unit. Or use the arrow keys to select the digit to edit, and then use the knob
to change its value. The generator will change the waveform immediately.
Note: The methods of setting other parameters of square signal are similar to sine waveform function.
3.1 Oset, High/Low Levels
Square Wave20
3.2 Phase
Figure 3.2 Set Duty Cycle
Figure 3.3 Set Duty Cycle
Figure 3.4 Set Oset
Square Wave21
Figure 3.5 Set Hi Level
Figure 3.6 Set Low Level
Figure 3.7 Set Phase
Square Wave22
Figure 3.8 Phase at 180 degrees
Ramp Wave
Press Waveforms key to select the waveform function, and press the Ramp softkey. The ramp waveform parameters are
set by using the ramp operation menu.
The parameters for ramp waveforms include frequency/period, amplitude/high level, oset/low level, phase and symmetry.
As shown in Figure 4.1, in the soft key menu, select Symmetry. The symmetry parameter area is highlighted in the
parameter display window, and users can set the symmetry value here.
Figure 4.1 Ramp Waveform Main Screen
Frequency and Period Set the signal frequency or period; The current parameter will be switched at a second press.
Amplitude/ HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second press.
Oset/ LowLevel Set the signal oset or low level; The current parameter will be switched at a second press.
Phase Set the phase of the signal.
Symmetry Set the symmetry for ramp waveform.
4.1 Frequency, Amplitude, Oset, High/Low level, and Phase
Set the frequency, amplitude, high level, low level, oset and phase as shown in Figures 4.2, 4.3, 4.4, 4.5, 4.6.
Figure 4.2 Set Frequency
Ramp Wave24
Figure 4.3 Set Amplitude
Figure 4.4 Set Oset
4.2 Symmetry
The percentage that the rising period takes up the whole Period.
Input Range: 0~100% Default Value: 50%
1. Press Waveforms → Ramp → Symmetry, to set the symmetry. The symmetry shown on the screen when the
instrument is powered on is the default value or the set value of last power down.
2. Input the desired Symmetry.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value. The generator will
change the waveform immediately.
Ramp Wave25
Figure 4.5 Set High Level
Figure 4.6 Set Low Level
Figure 4.7 Set Phase
Ramp Wave26
Figure 4.8 Set Symmetry 15%
Figure 4.9 Set Symmetry 90%
Pulse Wave
Press Waveforms key to select the waveform function, and press the Pulse softkey. The pulse waveform parameters are
set by using the pulse operation menu.
The parameters for pulse waveforms include frequency/period, amplitude/high level, oset/low level, width, rise/fall and
delay. As shown in Figure 5.1, in the soft key menu, select PulWidth. The pulse width parameter area is highlighted in
the parameter display window, and users can set the pulse width value here.
Figure 5.1 Pulse Waveform Main Screen
Frequency/ Period Set the signal frequency or period; The current parameter will be switched at a second press.
Amplitude/ HighLevel Set the signal amplitude or high level; The current parameter will be switched at a second press.
Oset/ LowLevel Set the signal oset or low level; The current parameter will be switched at a second press.
PulWidth/ DutyCycle Set the signal pulse width or duty cycle; The current parameter will be switched at a second press.
Rise/ Fall Setting the rise edge or fall edge for pulse waveform. The current parameter will be switched at
a second press.
Delay Setting the delay for pulse waveform.
5.1 Frequency, Amplitude, Oset, High/Low level, and Phase
Set the frequency, amplitude, high level, low level, and oset as shown in Figures 4.2, 4.3, 4.4, 4.5, 4.6.
5.2 Pulse Width and DutyCycle
Pulse width is dened as the time from the 50% threshold of a rising edge amplitude to the 50% threshold of the next
falling edge amplitude (as shown in the gure below). The pulse width setting range is limited by the “Minimum Pulse
Width” and “Pulse Period” setting. For detailed information, please refer to “4060B Series Datasheet”. The default
value is 200s.
Pulse duty cycle is dened as the percentage that the pulse width takes up in the whole period. Pulse duty cycle and
pulse width are correlative. Once a parameter is changed, the other will be automatically changed.
1. Press Waveforms→ Pulse → PulWidth, to set the pulse width. The pulse width shown on the screen when the
instrument is powered on is the default value or the set value of last power down. If you want to set the waveform by
duty, press the PulWidth/DutyCycle key again, to switch into the duty parameter (the current operation is displayed
in inverse color).
Pulse Wave28
Figure 5.2 Set Frequency
Figure 5.3 Set Amplitude
2. Input the desired Pulse Width.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value. The generator will
change the waveform immediately.
5.3 Rising and Falling Edges
Rise edge time is dened as the duration of the pulse amplitude rising from 10% to 90% threshold, while fall edge time is
dened as duration of the pulse amplitude moving down from 90% to 10% threshold. The setting of rise/fall edge time
is limited by the currently specied pulse width limit. Users can set rise edge and fall edge independently.
1. Press Waveforms → Pulse → Rise to set the rise edge.
The rise edge shown on the screen when the instrument is powered on is the default value or the set value of last
power down. If you want to set the waveform by fall edge, press the Rise/Fall key again, to switch into the fall edge
parameter (the current operation is displayed in inverse color).
2. Input the desired rise edge.
Use the numeric keyboard to input the parameter value directly, and press the corresponding key to select the parameter
unit. Or use the arrow keys to select the digit to edit, and then use the knob to change its value. The generator will
change the waveform immediately.
Loading...
+ 64 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.