The following safety precautions apply to both operating and maintenance personnel and must
be followed during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
Read and understand the safety and operational information in this manual.
Apply all the listed safety precautions.
Verify that the voltage selector at the line power cord input is set to the correct line
voltage. Operating the instrument at an incorrect line voltage will void the warranty.
Make all connections to the instrument before applying power.
Do not operate the instrument in ways not specified by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the
safety standards of design, manufacture, and intended use of the instrument. B&K Precision
assumes no liability for a customer’s failure to comply with these requirements.
Category rating
The IEC 61010 standard defines safety category ratings that specify the amount of electrical
energy available and the voltage impulses that may occur on electrical conductors associated
with these category ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating
is also accompanied by a maximum voltage of the circuit to be tested, which defines the voltage
impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to
be connected to the mains supply. The voltages in the environment are typically derived from a
limited-energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be
connected to the mains supply at a standard wall outlet or similar sources. Example
measurement environments are portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be
connected to the mains installation of a building. Examples are measurements inside a
building's circuit breaker panel or the wiring of permanently-installed motors.
i
www..com
information@itm.com1.800.561.8187
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be
connected to the primary power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what
is specified in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal
to or higher than the instrument's category rating to maintain the instrument's category rating.
Failure to do so will lower the category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The
mains power should be 120 V RMS or 240 V RMS. Use only the power cord supplied with the
instrument and ensure it is appropriate for your country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an
electrical safety ground. This instrument is grounded through the ground conductor of the
supplied, three-conductor AC line power cable. The power cable must be plugged into an
approved three-conductor electrical outlet. The power jack and mating plug of the power cable
meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all
accessible conductive parts (including control knobs) may provide an electric shock. Failure to
use a properly-grounded approved outlet and the recommended three-conductor AC line
power cable may result in injury or death.
ii
www..com
information@itm.com1.800.561.8187
Unless otherwise stated, a ground connection on the instrument's front or rear panel is for a
reference of potential only and is not to be used as a safety ground.
Do not operate in an explosive or flammable atmosphere
Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finelydivided particulates.
The instrument is designed to be used in office-type indoor environments. Do not operate the
instrument
In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or
finely-divided particulates.
In relative humidity conditions outside the instrument's specifications.
In environments where there is a danger of any liquid being spilled on the instrument or
where any liquid can condense on the instrument.
In air temperatures exceeding the specified operating temperatures.
In atmospheric pressures outside the specified altitude limits or where the surrounding
gas is not air.
In environments with restricted cooling air flow, even if the air temperatures are within
specifications.
In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The
operating temperature range is 0 °C to 40 °C and the operating humidity is ≤ 90 % relative humidity at < 35 °C and ≤ 60 % relative humidity at 35 °C – 40 °C, with no condensation allowed.
iii
www..com
information@itm.com1.800.561.8187
Measurements made by this instrument may be outside specifications if the instrument is used
in non-office-type environments. Such environments may include rapid temperature or
humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise,
strong electric fields, or strong magnetic fields.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other
material gets on or inside the instrument, remove the instrument's power cord, remove the
instrument from service, label it as not to be operated, and return the instrument to B&K
Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives,
lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a
clean dry lint-free cloth or as instructed in this manual.
Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a
component in a life-support device or system.
iv
www..com
information@itm.com1.800.561.8187
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and
internal adjustments must be made by qualified service-trained maintenance personnel who
are aware of the hazards involved when the instrument's covers and shields are removed.
Under certain conditions, even with the power cord removed, dangerous voltages may exist
when the covers are removed. To avoid injuries, always disconnect the power cord from the
instrument, disconnect all other connections (for example, test leads, computer interface
cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any
conductors by measurements with a properly-operating voltage-sensing device before touching
any internal parts. Verify the voltage-sensing device is working properly before and after
making the measurements by testing with known-operating voltage sources and test for both
DC and AC voltages. Do not attempt any service or adjustment unless another person capable of
rendering first aid and resuscitation is present.
Do not insert any object into an instrument's ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a
fault condition in the circuit exists.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument.
Return the instrument to B&K Precision for service and repair to ensure that safety and
performance features are maintained.
v
www..com
information@itm.com1.800.561.8187
Cooling fans
This instrument contains one or more cooling fans. For continued safe operation of the
instrument, the air inlet and exhaust openings for these fans must not be blocked nor must
accumulated dust or other debris be allowed to reduce air flow. Maintain at least 25 mm
clearance around the sides of the instrument that contain air inlet and exhaust ports. If
mounted in a rack, position power devices in the rack above the instrument to minimize
instrument heating while rack mounted. Do not continue to operate the instrument if you
cannot verify the fan is operating (note some fans may have intermittent duty cycles). Do not
insert any object into the fan's inlet or outlet.
For continued safe use of the instrument
Do not place heavy objects on the instrument.
Do not obstruct cooling air flow to the instrument.
Do not place a hot soldering iron on the instrument.
Do not pull the instrument with the power cord, connected probe, or connected test
lead.
Do not move the instrument when a probe is connected to a circuit being tested.
vi
www..com
information@itm.com1.800.561.8187
Compliance Statements
This product is subject to Directive 2002/96/EC of the European
Parliament and the Council of the European Union on waste
electrical and electronic equipment (WEEE), and in jurisdictions
adopting that Directive, is marked as being put on the market after
August 13, 2005, and should not be disposed of as unsorted
municipal waste. Please utilize your local WEEE collection facilities
in the disposition of this product and otherwise observe all
applicable requirements.
Disposal of Old Electrical & Electronic Equipment (Applicable in the European
Union and other European countries with separate collection systems)
www..com
vii
information@itm.com1.800.561.8187
CE Declaration of Conformity
This instrument meets the requirements of 2006/95/EC Low Voltage Directive and 2004/108/EC
Electromagnetic Compatibility Directive with the following standards.
Low Voltage Directive
o EN61010-1: 2001
EMC Directive
o EN 61000-3-2: 2006
o EN 61000-3-3: 1995+A1: 2001+A2: 2005
o EN 61000-4-2 / -3 / -4 / -5 / -6 / -11
o EN 61326-1: 2006
viii
www..com
information@itm.com1.800.561.8187
Safety Symbols
Refer to the user manual for warning information
to avoid hazard or personal injury and prevent
damage to instrument.
Electric Shock hazard
Alternating current (AC)
Chassis (earth ground) symbol.
Ground terminal
On (Power). This is the In position of the power
switch when instrument is ON.
Off (Power). This is the Out position of the power
switch when instrument is OFF.
Power Switch (On/Off). This is the power switch
located in front of the instrument.
CAUTION indicates a hazardous situation which, if
not avoided, will result in minor or moderate
injury.
WARNING indicates a hazardous situation which, if
not avoided, could result in death or serious injury.
DANGER indicates a hazardous situation which, if
not avoided, will result in death or serious injury.
Notations
TEXT – Denotes a softkey.
TEXT – Denotes a front panel button.
[TEXT] – Denotes a physical connector in the front or rear panel.
www..com
ix
information@itm.com1.800.561.8187
Table of Contents
SAFETY SUMMARY .................................................................................................................................................. I
COMPLIANCE STATEMENTS ............................................................................................................................................ VII
SAFETY SYMBOLS .......................................................................................................................................................... IX
NOTATIONS ................................................................................................................................................................. IX
1 GENERAL INFORMATION .......................................................................................................................... 1
Front Panel Description ................................................................................................................................... 3
2 GETTING STARTED .................................................................................................................................... 6
2.1INPUT POWER REQUIREMENTS ............................................................................................................................... 6
Input Power ..................................................................................................................................................... 6
3.2SELECTING A CHANNEL ........................................................................................................................................ 14
Configure Frequency ...................................................................................................................................... 17
User-Defined High and Low Level ............................................................................................................................. 19
AM Modulation ............................................................................................................................................. 41
To Set Modulation Depth .......................................................................................................................................... 44
To Set Modulation Frequency ................................................................................................................................... 44
FM Modulation .............................................................................................................................................. 46
To Set Frequency Deviation ...................................................................................................................................... 47
To Set Phase Deviation .............................................................................................................................................. 48
To Set Key Frequency ................................................................................................................................................ 49
To Set Hop Frequency ............................................................................................................................................... 50
To Set Pulse Width/Duty Deviation ........................................................................................................................... 53
Sweep time .................................................................................................................................................... 56
Sweep Frequency ........................................................................................................................................... 56
Trig out .......................................................................................................................................................... 57
Sweep Type .................................................................................................................................................... 58
Direction ........................................................................................................................................................ 59
Burst Period .............................................................................................................................................................. 63
Language Setup ............................................................................................................................................. 68
Power On ....................................................................................................................................................... 68
Set to Default ................................................................................................................................................. 69
System Info .................................................................................................................................................... 70
Built-in Help ................................................................................................................................................... 72
Key Test ......................................................................................................................................................... 74
LED Test ......................................................................................................................................................... 74
Board Test ..................................................................................................................................................... 75
Frequency Coupling .................................................................................................................................................. 82
USB Interface ................................................................................................................................................. 86
LAN ................................................................................................................................................................ 88
To Set IP Address....................................................................................................................................................... 88
To Set Subnet Mask .................................................................................................................................................. 89
To Set Gateway ......................................................................................................................................................... 89
Sync methods for two or more instruments: ................................................................................................. 92
Synchronization between two instruments .............................................................................................................. 92
Synchronization among multiple instruments .......................................................................................................... 92
4.10OVER VOLTAGE PROTECTION ................................................................................................................................ 94
5 STORE AND RECALL ................................................................................................................................ 95
5.1STORAGE SYSTEM ............................................................................................................................................... 96
Local (C:) ........................................................................................................................................................ 96
USB Device (0:) .............................................................................................................................................. 96
5.2FILE TYPE .......................................................................................................................................................... 97
5.3SAVE THE INSTRUMENT STATE .............................................................................................................................. 98
5.4RECALL STATE FILE OR DATA FILE ........................................................................................................................ 100
SERVICE INFORMATION ..................................................................................................................................... 125
BK Precision’s 4050B series are dual-channel function/arbitrary waveform generators, capable
of generating sine and square waves of up to 10 MHz, 30 MHz or 60 MHz (depending on
model). They have an informative easy-to-read color display, user-friendly controls and a
numeric keypad which allows users to easily configure waveform properties. These waveform
generators can output square waves with frequencies up to 60 MHz (depending on model) and
less than 300 ps + 0.05 ppm of period of jitter. In addition, they feature non-volatile built-in
memory to create, store, and recall arbitrary waveforms up to 16K points with a 150 MSa/s
sampling rate, and 14-bit vertical resolution. 196 predefined arbitrary waveforms are also
available for output. Having USBTMC, LAN, and GPIB (optional adapter) interfaces, these
generators allow users to easily interface with application software to create and load arbitrary
waveforms into the instrument.
Features:
Dual-channel output with bandwidth up to 10 MHz, 30 MHz or 60 MHz
Amplitude up to 10 Vpp into 50 ohms.
150 MSa/s sampling rate, 14-bit vertical resolution, and 16K points waveform length
Capable of generating low jitter pulse waveforms
Square waves with low jitter and frequencies up to 60 MHz (depending on model)
AM, DSB-AM, FM, PM, FSK, ASK, PSK and PWM modulation functions
Sweep and burst functions
Harmonic waveforms generator
Waveform combining function
High-precision frequency counter
196 built-in arbitrary waveforms
4.3” TFT-LCD display, 480 x 272 resolution
1.2 Package Contents
Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items
from the shipping carton, and check for any obvious signs of physical damage that may have
occurred during transportation. Report any damage to the shipping agent immediately. Save
the original packing carton for possible future reshipment. Every instrument is shipped with the
following contents:
1
www..com
information@itm.com1.800.561.8187
1 x 4050B series function/arbitrary waveform generator
1 x Getting started manual (printed)
1 x AC power cord
1 x USB type A to Type B cable
1 x Certificate of calibration
Verify that all items above are included in the shipping container. If anything is missing, please
contact B&K Precision.
An optional USB to GPIB adapter model AK40G is also available.
2
www..com
information@itm.com1.800.561.8187
1.3 Front Panel Overview
Power On/Off switch
Menu softkeys
Menu buttons
Channel 1 output terminal
Channel 2 output terminal
Channel 1 and 2 Output On/Off buttons
Arrow keys
Rotary dial knob
Numeric keypad
TFT LCD color display
USB host port/*USB-to-GPIB adapter interface
Accepts USB flash drive to save/recall instrument settings and waveforms.
*This port can be used for connecting the USB-to-GPIB adapter (AK40G)
accessory. It can also be used for connecting an external USB flash drive.
1 2 3 4 5 6 7 8 9
10
11
1
10
2 3 5
6 7 8 9 11
4
Front Panel Description
www..com
3
information@itm.com1.800.561.8187
1.4 Rear Panel Overview
LAN Interface
USB interface
AC Power Input
Chassis ground
10MHz Out BNC connector
Aux In/Out BNC connector
Counter BNC Connector
1 2 3
4
5
6
7
123
4 6 7
5
Rear Panel Description
4
www..com
information@itm.com1.800.561.8187
1.5 Display Overview
Waveform display
Channel status bar
Waveform parameters display
Waveform mode and output impedance indicator
LAN status indicator
Mode phase locked/free running indicator
Menu display
Waveform parameters display
1 2 3 4 5 6 7
8
1 7 3
2
5
4
8
Display Description
www..com
5
information@itm.com1.800.561.8187
2 Getting Started
The included AC power cord is safety certified for this instrument operating in rated
range. To change a cable or add an extension cable, be sure that it can meet the
required power ratings for this instrument. Any misuse with wrong or unsafe cables
will void the warranty.
Before connecting and powering up the instrument, please review and go through the
instructions in this chapter.
2.1 Input Power Requirements
Input Power
The supply has a universal AC input that accepts line voltage and frequency input within:
100 – 240 V (+/- 10%), 50 – 60 Hz (+/- 5%)
100 – 127 V, 45 – 440 Hz
Before connecting to an AC outlet or external power source, be sure that the power switch is in
the OFF position and verify that the AC power cord, including the extension line, is compatible
with the rated voltage/current and that there is sufficient circuit capacity for the power supply.
Once verified, connect the cable firmly.
2.2 Output Connections
The waveform generator output circuits operate as a 50 Ω voltage source working into a 50 Ω
load. At higher frequencies, a non-terminated or improperly terminated output may cause
aberrations on the output waveform. In addition, loads with an impedance less than 50 Ω will
reduce the waveform amplitude, while loads with an impedance greater than 50 Ω will increase
waveform amplitude.
Excessive distortion or aberrations caused by improper termination are less noticeable at lower
frequencies, especially with sine and triangle waveforms. To ensure waveform integrity, follow
these precautions:
www..com
6
information@itm.com1.800.561.8187
1. Use good quality 50 Ω coaxial cable and connectors.
2. Make all connections tight and as short as possible.
3. Use good quality attenuators, if it is necessary to reduce waveform amplitudes applied
to sensitive circuits.
4. Use termination or impedance-matching devices to avoid reflections.
5. Ensure that attenuators and terminations have adequate power handling capabilities.
If there is a DC voltage across the output load, use a coupling capacitor in series with the load.
The time constant of the coupling capacitor and load must be long enough to maintain pulse
flatness.
Impedance Matching
If the waveform generator is driving a high impedance, such as a 1 MΩ input impedance
(paralleled by a stated capacitance) of an oscilloscope vertical input, connect the transmission
line to a 50 Ω attenuator, a 50 Ω termination and to the oscilloscope input. The attenuator
isolates the input capacitance of the device and terminates the waveform generator properly.
2.3 Preliminary Check
Complete the following steps to verify that the generator is ready for use.
1.Verify AC Input Voltage
Verify and check to make sure proper AC voltages are available to power the
instrument. The AC voltage range must meet the acceptable specification as explained
in section 2.1.
2.Connect Power
Connect AC power cord to the AC receptacle in the rear panel and press the power
switch to the ON position to turn ON the instrument. The instrument will have a boot
screen while loading, after which the main screen will be displayed.
3.Self Test
Press Utility, and select Test/Cal option. Then, select SelfTest option. The instrument
has 4 self-test options: screen test, key test, LED test, and board test.
7
www..com
information@itm.com1.800.561.8187
Figure 1 - Preliminary Self-Test
Figure 2 - Self-Test Options
Note: Refer to TEST/CAL section for more information about the Self-Test function.
Check Model and Firmware Version
The model and firmware version can be verified from within the menu system.
Press Utility, press System select and press Page1/2 from the menu to enter the second menu
page, and select System Info option. The software/firmware version, hardware version, model,
and serial number will be displayed. Press any Menu Softkey key to exit.
8
www..com
information@itm.com1.800.561.8187
Output Check
Output
Default
Function
Sine Wave
Frequency
1 kHz
Amplitude/Offset
4 Vpp/0 Vdc
Phase
0°
Load
High Z
Modulation
Default
Carrier
1 kHz Sine wave
Modulating
100 Hz Sine wave
AM Depth
100%
FM Deviation
100 Hz
ASK Key Frequency
100 Hz
FSK Key Frequency
100 Hz
FSK Hop Frequency
1 MHz
PSK Key Frequency
10 0Hz
PM Phase Deviation
100°
PWM Width Dev
190 μs
Sweep
Default
Start/Stop Frequency
500 Hz/1.5 kHz
Sweep Time
1 s
Trig Out
Off
Mode
Linear
Direction
↑
Burst
Default
Burst Period
10 ms
Follow the steps below to do a quick check of the settings and waveform output.
1. Turn on the instrument and set the instrument to default settings. To set to default,
press Utility → System→ Set To Default to set the system to the default setting. The
instrument will set both channels with the following parameters:
9
www..com
information@itm.com1.800.561.8187
Start Phase
0°
Cycles
1 Cycle
Trig Out
Off
Delay
521 ns
Trigger
Default
Source
Internal
Table 1 - Default Settings
1. Connect the BNC output of CH1 (yellow) into an oscilloscope.
2. Press the Output button on top of CH1 output BNC to turn on the output and observe a
sine wave with the parameters above.
3. Press the Parameter button.
4. Press Freq or Period option in the menu and use the rotary knob or the numeric keypad
to change frequency. Observe the changes on the oscilloscope display.
5. Press the Amplitude option in the menu and use the rotary knob or the numeric keypad
to change the amplitude. Observe the changes on the oscilloscope display.
Press the Offset option in the menu and use the rotary knob or the numeric keypad to
change the DC offset. With the oscilloscope set for DC coupling, observe the changes on
the display.
Connect the BNC output of CH2 (blue) into an oscilloscope and follow steps 3 to 6 to
check its output. Use the Ch1/Ch2 to toggle between channels.
10
www..com
information@itm.com1.800.561.8187
3 Front Panel Operation
Softkey will toggle
selection between
Frequency or Period.
Softkey will toggle
selection
between
Amplitude or
HighLevel.
Softkey will toggle
selection
between Offset or
LowLevel.
Softkey will select
Phase.
Softkey will enable
or disable the
Harmonic
Generation
Function.
Waveforms
Parameter
Function
Sine
Frequency/Period
Configures the frequency or period of the waveform.
Amplitude/HighLevel
Configures the amplitude or the high level of the
waveform.
Offset/LowLevel
Configures the DC offset or the low level of the waveform.
Phase
Configures the phase relative to the other channel.
Harmonic/Off
Turns Harmonics Generator on or off. See Harmonics
Generator section.
Square
Frequency/Period
Configures the frequency or period of the waveform.
Amplitude/HighLevel
Configures the amplitude or the high level of the
waveform.
Table 2 - Menu Options Softkeys
3.1 Menu Options
All settings and parameters can be configured from the menu system of the instrument. The
channel specific menu options are the same for both channel 1 and channel 2. Use the
Ch1/Ch2 button keys to toggle the channel selection. The selected option will be highlighted in
blue. Some settings are common for most waveforms (i.e. frequency, amplitude, offset, etc.)
and some are specific to each type of waveform (i.e. Duty cycle only available for square and
pulse waveforms). Section 3.3 provides instructions on how to configure these settings.
Many options are grouped in pairs and can be selected by toggling their corresponding menu
function keys. For example:
The menu system is organized as follows:
www..com
11
information@itm.com1.800.561.8187
Offset/LowLevel
Configures the DC offset or the low level of the waveform.
Phase
Configures the phase relative to the other channel.
DutyCycle
Configures the duty cycle of the waveform.
Ramp
Frequency/Period
Configures the frequency or period of the waveform.
Amplitude/HighLevel
Configures the amplitude or the high level of the
waveform.
Offset/LowLevel
Configures the DC offset or the low level of the waveform.
Phase
Configures the phase relative to the other channel.
Symmetry
Configures the symmetry of the waveform.
Pulse
Frequency/Period
Configures the frequency or period of the waveform.
Amplitude/HighLevel
Configures the amplitude or the high level of the
waveform.
Offset/LowLevel
Configures the DC offset or the low level of the waveform.
PulWidth/DutyCycle
Configures the pulse width or the duty cycle of the pulse.
Rise/Fall
Configures the rise or fall time of the pulse.
Delay
Configures the delay of the pulse waveform.
Noise
Stdev
Configures the standard deviation of the noise waveform.
Mean
Configures the mean value of the noise waveform.
DC
DC Offset
Configures the DC offset of the DC waveform.
Arb
Frequency/Period
Configures the frequency or period of the waveform.
Amplitude/HighLevel
Configures the amplitude or the high level of the
waveform.
Offset/LowLevel
Configures the DC offset or the low level of the waveform.
Phase
Configures the phase relative to the other channel.
Arb Type
Access selectable built-in arbitrary waveforms and userdefined arbitrary waveform stored in the generator.
Mod
Type
Configures the type of modulation. Configures parameters
for AM, FM, PM, ASK, FSK, DSB-AM, or PWM modulation.
www..com
12
information@itm.com1.800.561.8187
Source
Selects modulating source.
AM Depth
Set the modulation depth.
Shape
Configures the modulating waveform shape.
AM Freq
Set the modulating waveform frequency. Frequency range:
1 mHz to 20 kHz (internal source only).
Sweep
Sweep Time
Configures the sweep time for sweep output.
StartFreq/CenterFReq
Configures the sweep start frequency or the center
frequency.
StopFreq/FreqSpan
Configures the sweep stop frequency or the frequency
span of the sweep output.
Source
Selects the sweep source: Internal, external or manual.
Trig Out
Enable/disable trigger out.
Type
Selects between linear or logarithmic sweep operations.
Direction
Selects the sweep direction.
Burst
Ncycle/Gated
Selects burst by number of cycles or selects external gated
burst.
Cycles/Infinite
Configures number of cycles for burst or infinite burst
(external source only).
Configures signal triggering on rising or falling edge.
Utility
System
Access system settings.
Test/Cal
Configure synchronization output with respect to either
channel 1 or 2.
Counter
Selects built-in counter function.
www..com
13
information@itm.com1.800.561.8187
Output Setup
Set the output parameters of CH1 and CH2.
CH Copy Coupling
Set the track, channel coupling or channel copy function.
Interface
Select and set parameters of remote communication.
Clock
Choose the system clock source: internal or external.
Mode
Choose Phase-locked or independent mode.
OverVoltage Protection
Turn on/off the overvoltage protection function.
CHCopy
Copy channel settings between channels.
Table 3 - Menu System Organization
Figure 3 – Channel 1 Selected
3.2 Selecting a Channel
The 4050B series function/arbitrary waveform generators have dual channel outputs. They can
be operated independently or in sync with each other. To select between channel 1 and 2 and
view/change their parameters, press the Ch1/Ch2 key.
When Channel 1 is selected, the display will look like the following:
When Channel 2 is selected, the display will look like the following:
www..com
14
information@itm.com1.800.561.8187
Waveforms
Sine
Square
Ramp
Pulse
Noise
DC
Arbitrary
Figure 4 - Channel 2 Selected
Table 4 - Waveforms
3.3 Configure Waveform Output
Configure Waveform Shape
The instrument can generate many standard as well as arbitrary waveforms. There is a
dedicated waveform key on the front panel that will allow the user to select between different
waveform shapes to output, as listed in Table 4.
15
www..com
information@itm.com1.800.561.8187
Press Waveforms to select a waveform. Menu options relevant to the selected waveform
Sine Waveform
Square Waveform
Ramp Waveform
Pulse Waveform
Noise Waveform
DC Waveform
Arbitrary Waveform
shape will display at the bottom of the screen. The screenshots below illustrates the menu
options for each of the waveform types.
16
www..com
information@itm.com1.800.561.8187
Loading...
+ 108 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.