The information contained in this document is subject to change without notice.
Safety Notice Supplement
As described in the International Electrotechnical Commission (IEC) Standard IEC 664, digital
multimeter measuring circuits (e.g., B&K Models 2831E and 5491B) and the USB terminal are
Installation Category II (CATII). The AC input terminal is rated CAT I.
This equipment is POLLUTION DEGREE 2, INDOOR USE product.
Safety Summary
The following safety precautions apply to both operating and maintenance personnel and
must be observed during all phases of operation, service, and repair of this instrument.
Before applying power, follow the installation instructions and become familiar with the
operating instructions for this instrument.
GROUND THE INSTRUMENT
To minimize shock hazard, the instrument chassis and cabinet must be connected to an
electrical ground. This instrument is grounded through the ground conductor of the
supplied, three-conductor ac power cable. The power cable must be plugged into an
approved three-conductor electrical outlet. Do not alter the ground connection. Without the
protective ground connection, all accessible conductive parts (including control knobs) can
render an electric shock. The power jack and mating plug of the power cable meet IEC
safety standards.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes. Operation of
any electrical instrument in such an environment constitutes a definite safety hazard.
KEEP AWAY FROM LIVE CIRCUITS
Instrument covers must not be removed by operating personnel. Component replacement
and internal adjustments must be made by qualified maintenance personnel. Disconnect the
power cord before removing the instrument covers and replacing components. Under certain
2
conditions, even with the power cable removed, dangerous voltages may exist. To avoid
accessible conductive parts (including control knobs) can render an electric shock.
To avoid electrical shock hazard, disconnect power cord before removing covers.
Before connecting the line cord to the AC mains, check the rear panel AC line
lying a line voltage other than the indicated voltage can
destroy the AC line fuses. For continued fire protection, replace fuses only with
static discharge
sensitive
injuries, always disconnect power and discharge circuits before touching them.
DO NOT SERVICE OR ADJUST ALONE
Do not attempt any internal service or adjustment unless another person, capable of
rendering first aid and resuscitation, is present.
DO NOT SUBSTITUTE PARTS OR MODIFY THE INSTRUMENT
Do not install substitute parts or perform any unauthorized modifications to this instrument.
Return the instrument to B&K Precision for service and repair to ensure that safety features
are maintained.
WARNINGS AND CAUTIONS
WARNING and CAUTION statements, such as the following examples, denote a hazard
and appear throughout this manual. Follow all instructions contained in these statements.
A WARNING statement calls attention to an operating procedure, practice, or condition,
which, if not followed correctly, could result in injury or death to personnel.
A CAUTION statement calls attention to an operating procedure, practice, or condition,
which, if not followed correctly, could result in damage to or destruction of parts or the
entire product.
WARNING: Do not alter the ground connection. Without the protective ground connection, all
The power jack and mating plug of the power cable meet IEC safety standards.
WARNING:
Refer servicing to qualified personnel.
CAUTION:
voltage indicator. App
those of the specified voltage and current ratings.
CAUTION:
This product uses components which can be damaged by electro(ESD). To avoid damage, be sure to fo llow proper procedures for handling, storing
and transporting parts and subassemblies which contain ESDcomponents.
3
SAFETY SYMBOL
This product is subject to Directive 2002/96/EC of the European
Parliament and the Council of the European Union on waste
adopting that Directive, is marked as being put on the market after
August 13, 2005, and should not be disposed of as unsorted
This symbol on an instrument indicates that the user should refer to the operating
instructions located in the manual.
Electrical Shock hazard.
Chassis ground symbol.
CAT I
(1000V)
CAT II
(300V)
IEC Measurement Category II.
Inputs may be connected to
mains (up to 300 VAC) under
Category II overvoltage conditions.
IEC Measurement Category II.
Inputs may be connected to
mains (up to 300 VAC) under
Category II overvoltage conditions.
Compliance Statements
Disposal of Old Electrical & Electronic Equipment (Applicable in the European
Union and other European countries with separate collection systems)
electrical and electronic equipment (WEEE) , and in jurisdictions
municipal waste. Please utilize your local WEEE collection facilities
in the disposition of this product and otherwise observe all applicable
requirements.
2.1 Front Panel Overview .................................................................................................................................. 9
2.2 Annunciators on Screen ............................................................................................................................. 10
2.3 Front Panel Menu Reference ......................................................................................................................11
2.4 Front Panel Menu Overview ...................................................................................................................... 12
2.6 Power up .................................................................................................................................................... 14
2.6.1 Power Line Connection .............................................................................................................. 14
3.2 Measuring Voltage ..................................................................................................................................... 18
3.3 Measuring Current ..................................................................................................................................... 19
3.5 Measuring Frequency and Period .............................................................................................................. 24
3.5.1 Trigger Level a nd Measurement Errors ................................................................................... 24
3.5.2 Gate Time ..................................................................................................................................... 24
3.8.2 Using the 2nd parameter display ............................................................................................... 28
3.9 Math Functions .......................................................................................................................................... 29
4.1.1 Range ........................................................................................................................................... 33
4.2.2 Reading Hold ............................................................................................................................... 36
4.3 MAX / MIN ............................................................................................................................................... 36
4.5 System Operations ..................................................................................................................................... 37
4.5.1 Beeper Control ............................................................................................................................ 38
5.3 Data Format ............................................................................................................................................... 42
The 2831E 4½ digit and 5491B 50,000 count multimeters provide measurements with high accuracy,
great stability and fast measurement rates. The meters provide a maximum measurement rate of 25
readings/sec and a 0.02% and 0.03% DC voltage basic accuracy for model 5491B and 2831E
respectively. Both meters have broad measurement ranges:
DC voltage from 50 μV to 1000 V
AC (RMS) voltage from 1 mV to 750 V
DC current from 100 nA to 20 A
AC (RMS) current from 100n A to 20 A
Two -wire resistance from 10 mΩ to 20 MΩ (50 MΩ for mode l 549 1B)
Frequency from 5 Hz to 1 MHz
Some additional cap abilities:
Full range of functions: In addition t o tho se list ed ab ov e, the meter functions includ e per iod, d B,
dBm, continuity, diode testing, max, min and percent.
Programming languages and remote control interfaces: The meters are programmable via US B
(Virtual COM) interface using SCPI commands
1.2 Incoming Inspection
Please inspect the instrument mechanically and electrically upon receiving it. Unpack all item s f rom the
shipping carton, and check for any obvious signs of physical damage that may have occurred during
transportation. Report any damage to the shipping agent immediately. Save the original packing carton
for possible future reshipment. The followin g items are included with ev er y order:
Multimeter
Test leads
Power cord
Sp ar e f use
Operation Manual
Calibration certificate and test report
Verify that all items above are included in the shipping container. If anything is missing, please contact
B&K Precision.
8
Chapter 2 Overview
2.1 Front Panel Overview
The front panel of the multimeter is shown in Figure 2-1. This figure includes s ome import ant abbrevi ated
information that should be reviewed before operatin g t he instr ument.
1. Function keys
Input terminals
Current
Input fuse
Primary
Display
Secondary
Display
On /off
Switch
2. Math function key
5. Range
1. Measurement Function keys (shifted and unshifted)
Select measurement function: DC voltage and current, AC voltage and current, resistance,
continuity, frequency, period, dB, dBm, True RMS AC+DC and diode test .
2. Math function keys
Select the math function: Rel, %, Max/Min, Comp and Hold.
nd
3. 2
4. Menu operation keys
Display and speed key
Changes reading rate: Fast, Medium and Slow.
→ turns on/ off the 2nd parameter display.
→ Open/Close menu
Move through selections w it hin me nu level, command level or parameter level
Move through selections w ithin menu level, command level or parameter leve l.
Move up a level.
Selection
4. Menu
Operation
Figure 2-1 Front Panel
nd
3. 2
Display
And speed key
7. Shift/local
Key
6. Hold/Trig
Key
Move down a level.
(ENTER) Save the change made on “parameter” level, and retur n t o the “command” level.
Cancel the change m ade on “parameter” level, and return to the “command” level.
9
5. Range and Combinat ion func tion se lecting keys
Trig
Trig
Select a 2nd display parameter
Select a 2nd display parameter
Select a higher range and disable auto ranging.
Select a lower range and disable auto ranging.
Toggle between auto ranging and manu al ranging.
6. Trig/ Hold Key
Trigger a me asur ement from the front p anel.
Hold a reading on the display.
7. Shift/Local keys
Used to access shifted keys.
(LOCAL) Cancel remote control mode and back to the LO CA L mode.
2.2 Annunciators on Screen
Figure 2-2 Display A nnunciat ors
FAST Fast reading rate
MED Medium reading rate
SLOW Slow reading rate
TRIG Indicates trigger (front panel, bus) selected.
HOLD Reading HOLD function is enabled
REL Relative reading displayed
MATH A math operation is enabled (%, dB, dBm).
(Speaker) Beeper on for continuity testing function
(Diode) Instrument is in diode test ing function
DC DC operation is enabled
AC AC operation is enabled
COMP Limit test ing function is enabled
10
HI/IN/LO Indicates t he li m it t esting results
RMT Multimeter is in remote control mode
AUTO Auto ranging enabled
Max/Min MAX / MIN operation is enabled
ERR Hardware or remote control err or det ec t ed
SHIFT Accessing shifted keys
2.3 Front Panel Menu Reference
A:MATH MEN U
1: HI LIMIT → 2: LO LIMIT → 3: PERC REF → 4: dB REF→ 5: dBm REF
1. HI LIMIT Set the high limit for limit testing.
2. LO LIMIT Set the low limit for limit testing.
3. PERC REF Set the reference value for PERCENT function
4. dB REF Set the dB reference voltage value.
5. dBm REF Set the dBm reference impedance value.
B:TRIGGER MENU
1: TRIG MOD
1. TRIG MOD Select IMMediate, Manual or Bus trigger source mode.
C:SYSTEM MENU
1: BEEP STA → 2: BAUD RAT → 3: TX TERM → 4: RETURN → 5: KEY SONG → 6: REVISION
1. BEEP STA Enable or disable the beeper funct io n
2. BAUD RA T Select the baud rate for USB (virtual COM) or RS232 (Model 5491B only)
operation.
3. TX TERM Set the terminal character for USB (virtual COM) or RS232 (Model 5491B
only) operation, which identifies the end of a command string
4. RETURN Sets whether to enable Echo of the “sent” SCPI commands in return string.
(i.e. if ON, sending *IDN? will return with *IDN? and model information on
two separate lines).
5. KEY SONG Enable or disable the key sound when you press a key.
6. REVISION Shows the model number and firmware versio n of the instrument.
11
2.4 Front Panel Menu Overview
The menu is organized in a top-down tree structure with three levels (menus, commands and
parameters) as shown in Figure 2-3. You can use down (
one level to another. Each of the thr ee levels has se veral horizont a l choices which y ou can view by using
left (
To turn on the menu, press
To turn off the menu, press
To confirm a change on the “parameter” level, pres s
To cancel a change on the “parameter” level, pr es s
) or right ().
Menus
Commands
Parameters
Figure 2-3 Menu Tree
→(Menu).
→(Menu), or press any of the function or math keys on
the front panel.
) or up () to move menu tree from
(ENTER).
(Menu).
Note: If you press on the “menu” level, this is the top level of the menu an d you ca nnot
go any higher; similarly if you press
the menu and you cannot go any lower.
on the “parameter” level, this is the bottom level of
12
2.5 Rear Panel Summary
1
2
4
The rear panel of the multimeter is shown in Figure 2-4. This section includes important information that
should be reviewed before oper ating the instrument.
3
Figure 2-4 Rear Panel
1. USB Device Port
Connection port for USB rem ot e c ont r ol.
2. Grounding
Chassis Grounding terminal
3. Power-Line Fuse-Holder Assembly
The multimeter can be configured for line voltage of 110/220 V ± 10% AC at line frequency of 50/60
Hz ± 5%.
Power-line fuse is used for instrument prot ection. (220 V/500 mA or 110 V/1 A)
Note: Please use the same-type fuse. To verify and replace the fuse, remove the power
cable and pull out the fuse holder. To select bet w een line volt age operations of 110 V
or 220 V, turn the grey colored fuse holder until the correct line voltage label is shown
on the outer win dow of t he fus e holder. It should say “110” or “220 ” de pen din g w hic h
way you turn it.
4. RS232 Communication Port (Model 5491B only)
Connection port for remote control via RS232.
13
2.6 Power up
2.6.1 Power Line Connection
Follow the procedure below to connect the multimeter to line power and turn on t he instr ument.
1. Check to make sure that the line voltage is in the correct range of 110 V or 220 V ± 10% (198 V to
242 V) and line frequency is in the range of 60 Hz or 50 Hz ± 5% (or 47.5 to 52.5 Hz) and that the
line voltage switch is in the correct position before connecting the power cord. To che ck , f i rs t l ook
at the label on the fuse co m partment to verify if it is shown as “110” or “220”. If it is not switched to
the correct one, open up the fuse compartment and turn the grey fuse holder until it shows the
correct voltage label (110 or 220) on the outer part of the fuse compartment. Then, check to make
sure correct fuse value is used for the selected volt age operation.
CAUTION: Operating the instrument on an incorrect voltage may cause damage to the
instrument, possibly voidi ng the warranty.
2. Before plugging in t he power cord, make sure t hat the front panel pow er sw it ch is in the off posit ion.
3. Connect the female end of t he supplie d power c ord to t he AC receptacle on t he rear p anel. Co nnect
the other end of the power cord t o a gr ounded AC outlet.
WARNING: The power cord supplie d w ith t he multimeter contains a separate groun d wi re for use
with grounded outlets. When proper connections are made, instrument chassis is
connected to pow er line groun d throug h the gro und w ire in t he power co rd. Fai lure to
use a grounded outlet may r esult in personal injury or death due to electri c shoc k.
4. Turn on the instrument by pressing the front panel power switch and instrument is ready for
measuring.
2.6.2 Input Terminals
The input terminals are shown in Figure 2-5. The multimeter is protected against overloads up to the
limits as shown in table 2-1. Exceeding these limits may result in a hazard to both the multimeter and
operator.
14
2 mA-200 mA for 2831E, 5 mA-500 mA for model 5491B
range DC/AC current input terminal for DC/AC current
Input-High terminal for Volts, ohms, Hz, period, diode and
continuity measurements
Common terminal
2 A (5 A for model 5491B) and 20 A range current input
terminal for DC/AC curren t meas ur ement
T1AL/250 V type fuse for the mA measurement range
Figure 2-5 Input terminals
15
Table 2-1 Input protection Limits
,
Function Input Terminals Maximum Allowable Input
DCV
ACV,HZ
mA, HZ
20A,HZ
Ω
All functions Any terminals to earth 1000V DC or 1000V peak AC
to COM
to COM
500mA to COM
20A to COM
to COM
to COM
1010V DC
757.5V AC RMS,1000V Peak
200mA (Model 5491B: 500 mA) DC
or AC RMS
20A DC or AC RMS
500V DC or AC RMS
500V DC or AC RMS
2.6.3 Power-up S e que nce
On power-up, the multimeter performs self-tests on its EPROM and RAM and lights all segments and
annunciators for about 1 second. If a failure is detected, the instrument momentarily displays an error
message and the ERR annunciator turns on.
If the instrument passes self-tests, the fir mwar e r evision will be displayed momentarily.
2.6.4 High Energy Circuit Safety Precautions
To optimize safety when measuring voltage in high energy distribution circuits, read and use the
directions in the following.
When making measurements in high energy circuits, use test leads and accessories that meet the
following requirements:
Test leads and accessories must be fu lly insulated.
Only use test leads that can be connected t o the circuit (e.g., alligator clip s, sp ade lugs, etc.) for
hands-off measureme nt s.
Do not use test leads or accessories that decrease voltage spacing. This diminishes arc
protection and creates a hazardous condition.
Use the following sequenc e w hen measuring high energy circ uits:
1. De-energize the circuit using a regular installed connect-disconnect device, such as a circuit
breaker , main switch, etc.
2. Attach the test leads to the circuit under t es t . Use appropriate safety rat ed t est leads .
3. Set the multimeter to the pr oper measurement funct ion and range.
4. Energize the circuit using the installed connect-disconnect device and make measurements
without disconnecting the multimeter.
5. De-energize t he c ircuit using the installe d connect-disconnect device.
6. Disconnect the test leads from the circuit under t est .
WARNING: The maximum common-mode voltage (voltage between COM and the chassis
ground) is 500V peak. Exceeding this value may cause a breakdown in insulation,
creating a shock hazard.
16
2.6.5 Power-on Defaults
2831E and 5491B have a factor y default setting for the pow er -on setting.
Since the basic measurement procedures in this manual assume the factory defaults, reset the
instrument to the factory settings when following step-by-step procedures. Table 2-2 lists the factory
default settings.
Table 2-2 Factory Default Settings
SettingFactory Default
Function
Range
Rate
Remote/Local
Trigger Mo de
Relative Mode
Compare Mode
HI Limit
Lo Limit
Percent Mode
Reference
Max/Min Mode
Reading Hold
Secondary Display Mode
Cal Mode
DCV
AUTO
Medium
Local
Immediate
OFF
OFF
+1
-1
OFF
+1
OFF
OFF
OFF
OFF
2.6.6 Warm-up time
The multimeter is ready for use as soon as the power-up sequence has completed. H ow ever, to achieve
rated accuracy and stability, allow the instrument to warm up for half an hour. If the instrument has been
subjected to extreme te m per at ur es, allow additional time for internal temperatures to s tabilize.
2.7 Display
The display of the multimeter is primarily used to display readings, along with the units and type of
measurement. Annunciators located on the left, right and bottom indicate various states of operation.
See section 2.2 for a complete listing of annunciators.
17
Chapter 3 Basic Measurements
ACV
DCV
ACV
3.1 Preparation
One of the first things you would like to do with your multimeter is to become acquainted with its front
panel. We have provided some exercises in foregoing chapters about preparations for use and
operations of front pa nel.
The front panel has six rows of keys to select v arious fu nctio ns a nd o perati ons. Most keys hav e a sh i fte d
function printed in blue above the key. To perform a shifted function, press
will turn on). Then, press the key that has the desired label above it. For example, to select the AC
current function, press
If you accidentally press
then press (ACI).
, just press it again to turn off the Shift annunciator.
3.2 Measuring Voltage
Voltage ranges: 200 mV, 2 V, 20 V, 200 V, 1000 V (750 VAC) (model 5491B: 500 mV, 5 V, 50 V, 500 V,
1000 V (750 VAC))
Maximum resolution: 50 μV for DC and 1 mV for AC (on 200 mV range (model 2831E) and on 500 m V
range (model 5491B))
AC technique: true RMS, AC-coupled, 1000 V Peak AC
(the Shift annunciator
3.2.1 Connections
Assuming the multimeter is under factor y default conditions, the basic procedure is as follows:
1. Connect test leads to
2. Select DC or AC voltage me as urement by pressing or .
3. Press
displayed when in auto ranging mode. If you want manual ranging, use the RANGE
4. Connect test leads to the sources as shown in Figure 3-1.
CAUTION: Do not apply more than 1000V peakto the input or i ns t r ument damages may occur.
5. In manual range, if the “OVL.D” message is displayed, press the up
range until a desired reading is displayed (or press
possible range for the best r esolut i on.
6. Press
to toggle between auto and manual ranging. Notice that the AUTO annunciator is
keys to select a measurement r ange consistent with expected voltage.
+ to turn on the secondary display, Use or key to choose the
and COM terminals.
and
key to se lect a hi gher
key for auto ranging). Use the lowest
18
function for secondary dis play.
AC VOLTAGE SOURCE
DC VOLTAGE SOURCE
7. Ta ke rea dings from the display.
CAUTION: Maximum Input = 1010 V peak
Input Resistance = 10 MΩ
CAUTION: Maximum Input = 750 V RMS or 1000 V peak, 3x10
Model 2831E current measurement range: 2 mA, 20 mA, 200 mA, *2 A, *20 A
Model 5491B current mea s urement range: 5 mA, 50 mA, 500 mA, *5 A, *20 A
*Indicates range is availab le f or manual range only. Auto range will not work when reading requires
Input Impedance = 1.1 MΩ and 100 pF
Figure 3-1 DC and AC Voltage Measurement Connections
3.3 Measuring Current
19
7
V-Hz
setting to this range.
DCV
ACV
Maximum resolution: 100 nA ( on 2 mA range (on 5 mA range for model 549 1B))
Current measurement is a vailable in Auto range for the lower current measurement s only.
Note:
For 2831E, readings that are below 200 mA using the l ow current measure m ent in put can be set in auto
range. This means it will auto range bet w een 2 mA, 20 mA, and 200 mA range. For readings above
200 mA, auto range is
available. Manual range must be used to read any readings higher than 200
not
mA. This means, for ranges 2 A and 20 A, only m anual range is available. Additionally, the high
current measurement inp ut must be used to obtain meas urements in these two hig her r anges.
For 5491B, readings that are below 500 mA usin g the low current measurem ent input (fus e protected u p
to 500 mA max.) can be set in auto range. This means it will auto range between 5 mA, 50 mA, and
500 mA range. For readings above 500 mA, auto range is
not available. Manual range must be used
to read any readings higher t han 500 mA. This means, for ranges 5 A and 20 A, only manual range is
available. Additionally, the high current measurement input mu st be used to obtain measureme nt s in
these two higher ranges.
3.3.1 Connections
Assuming the multimeter is under factory default conditions, the basic procedure to measure current is
as follows (users must use manual range for measurement, following step 3):
1. Connect test leads between 500 mA terminal (for lower current meas urement s) and COM terminal or
between 20 A terminal (for higher current measure m ents) and COM terminal.
2. Select DCI or ACI measurem ent f unct ion by pressing
3. Press
toggles auto ranging. Note that auto range is only available for readings that are
below 200 mA (for 2831E) and 500 mA (for 5491B) and when using the 500 mA max. terminal and
COM terminal connections. For readings above these limits, only manual range is available and
the 20 A terminal and COM terminal must be used to obtain measurement results. If instrument
shows the AUTO annunciator on display, press again to turn off auto ranging mode and go into
manual range mode. Use the RANGE and
consistent with expected current.
4. Connect test leads to the source as shown in Figure 3-2.
CAUTION: Do not apply more than 1 A, 250 V to the 500 mA input terminal or the fuse will be
blown. For measuring current higher than 500 mA(or 200 mA for model 2831E), use the 20 A
terminal instead.
→or →.
keys to select a measurement range
5. In manual range, if the “OVL.D” message is displayed, press up
key to select a higher range
until a desired reading is displayed. U se the lowest possible range for the best resolution. Press
20
+ to turn on the 2nd parameter display, Use or key to select function for
AC CURRENT SOURCE
DC CURRENT SOURCE
nd
display.
2
6. Ta ke rea dings from the display.
(Model 2831E) DC Current measurement on Range: 2 mA, 20 mA, 200 mA
(Model 5491B) DC Current measurement on Range: 5 mA, 50 mA, 500 mA
(Model 2831E) AC Current measurement on Range: 2 mA, 20 mA, 200 mA
(Model 5491B) AC Current measurement on Range: 5 mA, 50 mA, 500 mA
21
(Model 2831E)DC Current measurement on Range: 2 A, 20 A
AC CURRENT SOURCE
DC CURRENT SOURCE
(Model 5491B) DC Current measurement on Range: 5 A, 20 A
(Model 2831E) AC Current measurement on Range: 2 A, 20 A
(Model 5491B) AC Current measurement on Range: 5 A, 20 A
CAUTION: Maximum Input = 20 A DC or RMS Maximum test times : < 20 s
Figure 3-2 DC and AC Current Measurements
3.3.2 Front Panel Fuse Replacement
WARNING: Make sure the instrument is disconnected from the power line and other equipment
1. Turn off the power and disconn ec t t he pow er line and test leads.
2. From the front panel, use a screwdriver to rotate the fuse carrier several turns counter-clockwise.
Take the fuse carrier out of the socket.
3. Remove the fuse and replace it with the same type ( T1AL, 250V, 5×20 mm ).
before replacing the AMPS fuse.
22
Loading...
+ 51 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.