Bizerba iS10, iS20 Operating Instructions Manual

0 (0)
Terminal iS10 / iS20
Operating instructions
as of program version 1.01
38019417000 en
Bizerba Service
www.bizerba.com
Alle Rechte vorbehalten All rights reserved Tous droits réservés Reservados todos los derechos Tutti i diritti riservati
© 07/2015
Bizerba GmbH & Co. KG, 72336 Balingen
P.O. Box 10 01 64 72301 Balingen, Germany Telephone (+49 7433) 12-0, Fax (+49 7433) 12-2696 Email: Internet: www.bizerba.com
marketing@bizerba.com
Translation of original operating instructions 38019417000 en

Operating instructions iS10 / iS20 Contents

Contents Page
1 About these instructions .............................................. 5
1.1 Content ........................................................ 5
1.2 Safe-keeping ................................................... 5
1.3 Target group ................................................... 5
1.4 Symbols used .................................................. 5
1.4.1 How notes and information are depicted ......................... 5
1.4.2 Explanation of warnings ......................................... 6
2 About the device ...................................................... 7
2.1 Device types ................................................... 7
2.2 Scope of delivery ............................................... 7
2.3 Device view .................................................... 7
2.3.1 Device configuration ............................................ 7
2.4 Identification plates ............................................. 9
2.4.1 Position of identification plates ................................... 10
2.5 Intended use ................................................... 10
2.6 Protective mark locations ........................................ 10
2.6.1 Markings and labels ............................................. 10
2.6.2 Safety plaques ................................................. 10
2.7 Sight window for internal safety mark ............................. 11
2.8 Verification-related notes ........................................ 11
2.9 Metrologically approved data storage ............................. 12
2.10 Regulations of approval for weighing data memory ................ 12
2.11 Operating conditions ............................................ 12
2.12 Software ....................................................... 13
2.13 Load receptor (optional) ......................................... 13
2.14 Warranty ....................................................... 13
2.14.1 Different warranty conditions for products acquired at the Bizerba on-
line shop ....................................................... 14
2.15 Disposal/Environmental protection ............................... 14
3 Safety instructions .................................................... 15
3.1 Requirements for operating personnel ............................ 15
3.2 Supply circuit disconnect ........................................ 15
3.2.1 Removable power cable (optional) ............................... 15
38019417000 en 1
Contents Operating instructions iS10 / iS20
3.3 System-related sources of hazards ............................... 16
3.3.1 Mains supply ................................................... 16
4 Installation ............................................................ 17
4.1 Transport and storage ........................................... 17
4.2 Check the electric connection .................................... 17
4.3 Battery (option) ................................................. 18
4.4 Removable power cable (optional) ............................... 19
4.4.1 Connect power cable ........................................... 21
5 Operation ............................................................. 22
5.1 Display and operating field ...................................... 22
5.2 Switch device on / reset ......................................... 24
5.3 Switching device off ............................................. 24
5.4 Function assignment of key F1 ................................... 25
5.5 Functions of menu key F2 ....................................... 26
5.6 Change the tens digit for input of numbers ........................ 27
5.7 Display metrological data ........................................ 27
5.8 Call up logbook ................................................. 30
5.9 Call up metrologically approved data storage ...................... 31
5.10 Call up tendency control ......................................... 33
5.10.1 Working with tendency control ................................... 35
5.10.2 Enter target value and tolerance limits ............................ 36
5.11 Battery pack operation (optional) ................................. 38
5.12 Weighing operation ............................................. 39
5.12.1 Weight display .................................................. 39
5.12.2 Zero setting of scale ............................................ 39
5.12.3 Net weighing with tare balance ................................... 40
5.12.4 Deleting tare ................................................... 40
5.12.5 Counting operation .............................................. 40
5.12.6 Weight display with increased resolution (not metrologically ap-
proved) ........................................................ 42
5.12.7 Total display .................................................... 43
5.12.8 Net weighing with fixed tare value ................................ 43
5.12.9 Display tare value ............................................... 44
5.12.10 Display gross weight ............................................ 44
5.12.11 Display supply / battery voltage .................................. 44
5.12.12 Display tendency control ........................................ 45
2 38019417000 en
Operating instructions iS10 / iS20 Contents
5.13 Recording operation ............................................ 45
5.13.1 Item recording, adding .......................................... 45
5.13.2 Total recording ................................................. 47
6 PC/EDP and printer interface .......................................... 50
6.1 PC/EDP interface ............................................... 50
6.1.1 PC/EDP standard dialog ........................................ 50
6.1.2 PC/EDP – continuous sending of data ............................ 55
6.1.3 PC/EDP interface parameters .................................... 56
6.2 Printer interface ................................................ 61
6.2.1 Printer interface parameters ..................................... 61
7 IXNET dialogs ......................................................... 66
7.1 The dialog frame of the transport protocol ......................... 66
7.2 Data dialog IXNET interface ..................................... 67
7.2.1 Device functions ................................................ 68
8 Ethernet interface ..................................................... 75
8.1 Performance of Ethernet interface ................................ 75
8.2 Ethernet interface parameters ................................... 75
9 Service menu ......................................................... 78
9.1 General parameters ............................................. 78
9.2 Scale parameter ................................................ 83
9.3 Scale calibration menu .......................................... 91
9.4 Switch settings of hardware seal ................................. 93
9.5 Hardware seal settings .......................................... 93
10 Examples of operating procedures .................................... 94
10.1 Weighing, recording, totaling ..................................... 94
10.2 Counting, recording, totaling ..................................... 96
11 Troubleshooting ...................................................... 98
11.1 Periodic faults .................................................. 98
11.2 Bring-in service ................................................. 98
38019417000 en 3
Contents Operating instructions iS10 / iS20
11.3 Fault descriptions ............................................... 99
12 Maintenance .......................................................... 101
12.1 Cleaning ....................................................... 101
12.1.1 Display and operating unit ....................................... 101
12.1.2 Stainless steel surfaces ......................................... 101
13 Technical data ........................................................ 103
13.1 Dimensions .................................................... 103
13.2 Housing ........................................................ 103
13.3 Display and operating unit ....................................... 103
13.4 Surrounding temperature range .................................. 103
13.5 Power supply ................................................... 103
13.6 Protection class ................................................ 104
13.7 Interfaces ...................................................... 104
13.7.1 Parallel inputs and outputs ...................................... 104
13.8 Connections .................................................... 105
Appendix
CE declaration of conformity
4 38019417000 en
Operating instructions iS10 / iS20 About these instructions
1

1 About these instructions

Read these instructions before switching on the device. These instructions include basic information on how to operate the device and how to avoid dangers.
These instructions do not distinguish between standard equipment and options. Please, contact your Bizerba specialist regarding available variants.
Our products undergo continuous further development and are subject to different country­specific regulations. Examples of pictures and graphics included in these instructions may vary from the version you have received.

1.1 Content

These instructions describe the operation of the industrial terminals iS10 and iS20, called device in the following.

1.2 Safe-keeping

This technical document is an essential component of the device. It must be stored close to the device and easily accessible for everybody.
1
If the device is resold, these complete instructions must be provided with it.

1.3 Target group

The device may only be operated by trained personnel. The operator must be familiar with the contents of these instructions. Providing no alternative agreement has been made, in­stallation, maintenance and repair works must be performed by Bizerba authorized spe­cialists only.
You may also perform the installation yourself if the device is offered by Bizerba in a Bizer­ba online shop and if you have acquired it directly at a Bizerba online shop operated by Bizerba.

1.4 Symbols used

The following symbols can be found in the manual:
Text with arrow prompts you to carry out an action. Position number in figure.
<OK> Text inside a < > refers to a key or soft key. "Display" Text inside a " " refers to display text.
Prerequirements are displayed with a gray background.
1.4.1
Notes and information are depicted as follows:
38019417000 en 5

How notes and information are depicted

Observance of these notes is mandatory.
1
About these instructions Operating instructions iS10 / iS20
Additional information for better understanding.

1.4.2 Explanation of warnings

The signal word above the symbol indicates the risk level.
DANGER Source of danger with high risk of imminent danger to persons!
This may lead to life threatening injuries and serious health dam­age.
Measurements for prevention of danger are specified.
WARNING Source of danger with medium risk of possible danger to persons!
This may lead to serious injuries, health damage or serious dam­age to property.
Measurements for prevention of danger are specified.
CAUTION Source of danger with low risk of possible danger to persons!
This may lead to injuries or damage to property.
Measurements for prevention of danger are specified.
CAUTION Source of danger, improper use!
Damage to property can result.
Measurements for prevention of danger are specified.
6 38019417000 en
Operating instructions iS10 / iS20 About the device
1
2
3
4
5
6

2 About the device

2.1 Device types

The device is available in many different versions.
Terminal iS10: The device is an industrial terminal w/o interfaces. Terminal iS20: The device is an industrial terminal with additional interfaces
and LEDs.

2.2 Scope of delivery

Industrial terminal (remote, table top display, column or wall-mounted) – Operating instructions

2.3 Device view

2
Fig. 1: Table-top installation
Control panel
Identification plate (max, min, e)
7-segment display
Standby/reset key
LED display (only with iS20)
Viewing window for control mark

2.3.1 Device configuration

The interfaces mentioned in the following are only available on terminal iS20. Terminal iS10 has no interfaces.
38019417000 en 7
2
1
2
3
4
About the device Operating instructions iS10 / iS20
Fig. 2: Device configuration
GLP 58
CITIZEN / EPSON or compatible printer
PC/EDP
Load receptors
COM1 = printer
COM2 = EDP
Ethernet = EDP
8 38019417000 en
Operating instructions iS10 / iS20 About the device
1
2
3
4
5
6
7
8
9
10
11
12
13

2.4 Identification plates

Main identification plate
The main identification plate may differ depending on device and country of installation.
Fig. 3: Main identification plate (example)
Device number
Device number of the connected load receptor
Weighing range and verification value
Zone (DEU, AUT, GBR, ITA, HUN)
Protection type
2
Electrical data
No. of EC type approval
Field for notified body during conformity valuation by Bizerba
Year of manufacture (2-digit)
Manufacturer
Country of manufacture
Type Designation
Scale accuracy class
Identification plate with max, min and e
The identification plate may differ in design depending on the weighing range.
Fig. 4: Identification plate with max, min and e
38019417000 en 9
About the device Operating instructions iS10 / iS20
1
2
2

2.4.1 Position of identification plates

The main identification plate in accordance with OIML R76-1 is located on the right side of the housing. The metrological information is located on an additional inserted plate which is near the window close to the display.
Main identification plate in accordance with OIML R76-1
Identification plate with max, min and e
Fig. 5: Position of identification plates

2.5 Intended use

This device is intended for use as a terminal in an industrial environment.
The device may only be used in industrial or commercial fields. – The device may not be used in potentially explosive areas. – The device must only be transported using appropriate means of transportation and if
it was secured accordingly. – No unauthorized constructional modifications must be applied to the device. – The device may only be used for verifiable transactions if the software is operating
correctly and is in a secure state. The software IDs must be correctly displayed and
checked by the operator daily. The internal safety mark must not be damaged.

2.6 Protective mark locations

2.6.1 Markings and labels

All markings and labels that are required according to the valid EC directives are located on the identification plate. The symbol for conformity evaluation (initial verification) is af­fixed to the measuring device.

2.6.2 Safety plaques

The scale adjustment data are protected by a safety clip under the lead seal. The connec­tor for the load receptor is also located under the sealing cover. The fixing screw of the lead seal is secured against removal by a yellow seal mark.
For load receptors with analog load cells, the cable connection cabinet is also provided with a seal mark.
10 38019417000 en
Operating instructions iS10 / iS20 About the device

2.7 Sight window for internal safety mark

The safety mark is visible through the sight window. The safety mark is located on the pro­tective cover of the load receptor connection. In order to see the safety mark in the dark housing, there is a lamp inside the sight window. The sight window is located at the front of the device, see page 7.

2.8 Verification-related notes

Remark referring to scales which have been used in Germany since 01 Jan 2015.
If new or refurbished devices [...] are used, they need to be registered at the notified authority according to state law, at latest six weeks after setting into operation.
This, however, is not mandatory for accessory components.
Metrological / verification information for EC countries
On the identification plate of the scale there is the green (metrology mark). This type of scale can be commissioned and used in-situ, providing that it is not connected to an addi­tional set-up (e.g. printer).
2
Scales that are connected on site to an additional device must have the conformity valua­tion carried out either by the responsible verification office or by a Bizerba verification engi­neer. Scales as well as the additional device must be tested and verified. If the scale with add. device meets the verification-related requirements, it can be used in operations that are subject to verification. In the case of extensions at a later date, once the scales have already been put into operation, the relevant verification office must be informed.
Scales and additional devices not bearing the EC mark and without any mark of conformity may not be used in transactions subject to verification.
Scales that have been adjusted to a certain gravity zone (gravitational acceleration) (see data on identification plate), must not be applied in other gravity zones without being ad­justed and verified accordingly.
According to statutory regulations, the user of a scale must ensure that it is used for its intended purpose. This includes observation of the verification guidelines, in particular amendments, supplements and additions.
Information on verification (early re-verification)
The verification of a scale is executed according to the respective statutory country regula­tions. For example, the verification validity period for non-automatic scales is generally 2 years in Germany. The verification validity period begins when the scales are put into cir­culation (installation and commissioning).
38019417000 en 11
About the device Operating instructions iS10 / iS20
2
A yellow mark of conformity is affixed to the load receptor of the scale, with the inscription "Gesichert von BIZERBA" (for scales with a yellow mark of conformity, the valuation of conformity has been carried out by Bizerba). Verifications must be carried out by the operator of the scale according to the legal requirements.
Verification-related notes for non-EC countries
The statutory regulations of the specific countries must be observed.

2.9 Metrologically approved data storage

The verifiable data memory serves to save verifiable measuring data. The recorded weigh­ing results are saved (non-volatile) with a consecutive number.

2.10 Regulations of approval for weighing data memory

The points listed below are prescribed by the EC type test certificate; these must be abso­lutely observed by the user. The following conditions must be fulfilled in relation to the long-term storage of weighing results:
Weighing results must be stored together with an identification, so that each weighing
process of each weighing result can be easily assigned and verified if this should be
necessary. These identifications must also be specified for documents that have been
created with additional devices. – If taring was performed, tare and gross values have to be stored. – Contracting parties must be able to verify the stored weighing results.
Business documents that are created by an additional device not subject to statutory met­rological inspection in accordance with the preliminary comment to Appendix 1 of directive 2009/23/EC, must contain the following information:
Identification for each weighing conveyor. – When used in Germany, an additional note is required stating that, based on the iden-
tification information, the weighing results can be compared to the stored weighing re-
sults in compliance with the guidelines.

2.11 Operating conditions

Please contact us or our customer service departments if you have any doubts concerning the practical application of these conditions.
Protection type, temperature and air humidity
The permitted values and the protection type can be found on the device identification plate and in the Technical data.
Air convection
In order to prevent unacceptable heating, there must be free air convection around the de­vice.
12 38019417000 en
Operating instructions iS10 / iS20 About the device
supply voltage
The permitted values can be found on the device identification plate and in the Technical data.
Written permission for changes
Modifications to the devices require our prior written consent.
Repairs on the device may only be performed by the manufacturer or authorized specialist workshops.

2.12 Software

The device has a software download.
A software must only be loaded if approved by the owner of the meas­uring device.
2

2.13 Load receptor (optional)

Free-standing, unfixed and mobile scales are equipped with a level. After each change of location check if the load receptor is horizontally aligned.
Load receptors may only be loaded up to the maximum approved load. The maximum load can be found on the type plate of the load receptor.

2.14 Warranty

Installation, putting into operation as well as initial instruction regarding operation is carried out by Bizerba Customer Service, Bizerba specialists or companies commissioned by Bi­zerba.
Warranty for defects shall not exist in particular if and insofar as the defects occurred to the delivered goods or to assets of the ordering party are attributable to the following rea­sons:
Non-observance of operating conditions – Non-adherence to the technical documentation – Defective electrical installation by the customer – Structural modifications to the equipment – Incorrect programming and operation – Missing backup – Natural wear and tear – Wrong cleaning agent
38019417000 en 13
2
About the device Operating instructions iS10 / iS20
The warranty will lapse if defects/damage that occur are caused by persons not authorized by Bizerba. The warranty will lapse if any spare parts or operating material other than origi­nal Bizerba ones are used. If you have any questions on warranty, spare parts or operat­ing material, please contact your Bizerba consultant.
If you reset or program devices, check the new setting by means of a test run and test printout. This helps to avoid incorrect results.
Check that the Bizerba products are handled correctly and repeat train­ing if necessary.
2.14.1 Different warranty conditions for products acquired at the Bi-
zerba online shop
At the Bizerba online shop only products are available, for which the installation and set-up can be performed by the customer himself.
You may also perform the installation and set-up yourself if the device is offered by Bizer­ba in a Bizerba online shop and if you have acquired it directly at a Bizerba online shop operated by Bizerba. Warranty will lapse if defects or damages that occur are caused through improper installation, set-up or instructions.
The other conditions of defects warranty apply unchanged.

2.15 Disposal/Environmental protection

The legal disposal/environmental protection terms of the respective countries apply.
Only valid for Germany!
Batteries are fitted in this device which are subject to the Battery Ordinance on the return and disposal of used batteries and accumulators.
This ordinance requires you, as the end user, to return the device to the manufacturer or dealer, at the end of its proper use, for the purposes of recycling and correct disposal.
If your device contains a rechargeable battery, you are required by the Battery Ordinance to return the used battery to the dealer or to a public return point provided for this purpose.
14 38019417000 en
Operating instructions iS10 / iS20 Safety instructions

3 Safety instructions

The safety instructions must be observed throughout the installation, operation, cleaning and maintenance procedures.

3.1 Requirements for operating personnel

The device must only be operated by personnel trained in the operation of this equipment. The operating personnel must be familiar with the safety features of the device and must have read and understood the safety instructions included in these operating instructions.
Only qualified trade personnel or Bizerba service technicians may open the device components.
Even though this equipment includes all required safety devices, disregard the safety in­structions could result in injuries to the operating personnel or damage to property.
The minimum age of the operating personnel by German law is 14. Different requirements may apply in your country.

3.2 Supply circuit disconnect

3
Mains supply with power plugs
The device comes standard with power plugs. The power plugs are the supply circuit dis­connect. Note the following requirements:
To de-energize the device, you must disconnect the mains plug. – Please, note that devices with battery pack are still internally supplied with approx.
12 V. – Customer provided power sockets must be easily accessible. – The distance between the mains power sockets and the device must be shorter than
3.6 m (12 ft).
Mains supply without power plugs
The user of the system is responsible for the safety relevant execution of the mains supply and the supply circuit disconnect in accordance with national regulations.

3.2.1 Removable power cable (optional)

The device can be optionally supplied with a removable power cable.
DANGER
Electrical voltage on the plug-in connector with mains plug inser­ted!
Danger to life from electric shock.
Do not apply voltage to or separate removable power cable.
38019417000 en 15
Safety instructions Operating instructions iS10 / iS20

3.3 System-related sources of hazards

3
DANGER Electrical voltage in device with power plug inserted!
Danger to life due to electrical shock.
Pull power plug before working inside the device. – Work inside the device must be performed by qualified personnel
only.

3.3.1 Mains supply

WARNING Incorrect supply voltage!
Fire hazard!
Destruction of electric components.
Compare the data of the supply voltage with the specifications on
the device identification plate.
Do not connect the device to the electric mains if the data of the
supply voltage do not match the connection values of the device.
16 38019417000 en
Operating instructions iS10 / iS20 Installation

4 Installation

4.1 Transport and storage

Always transport and store the device in its original packaging. If the device was stored in an area with low temperatures before being unpacked, the device could condensate in a room with normal temperatures after being unpacked.
Switch on device only after it has reached room temperature and no longer condensates. In order to accelerate this process, remove load platter, if necessary.

4.2 Check the electric connection

Installation of power supply provided by customer
The installation of the power supply for connecting to our equipment must be carried out in accordance with the international specifications and the regulations derived from them. These essentially include the recommendations of at least one of the following commis­sions:
International Electro-technical Commission (IEC) – European Committee for Electro-technical Standardization (CENELEC) – German Association for Electrical, Electronic and Information Technologies (VDE)
In case of a strongly contaminated network interference suppressing measures need to be taken, e. g:
Provide separate power line to our devices.
4
Include a capacitive decoupled isolation transformer or some other interference sup-
pressing device into the feeding lines to our devices.
The provided network cable must be suitable for the local supply volt­age.
It must show no external damage. – It must not get in contact with liquids. – It must comply with the requirements of the Safety Extra Low Volt-
age as per EC Directive.
Removable power cables must not be damaged on both ends of the plug-in connections.
Both ends of the plug-in connections must be connected according
to the coding.
The connections must be firmly screwed to each other.
38019417000 en 17
4
Installation Operating instructions iS10 / iS20
Check if the power supply data of the identification plate matches the power supply data of the power connection prior to connecting the de­vice.
WARNING Incorrect supply voltage!
Destruction of electric components, fire hazard.
Compare the data of the supply voltage with the specifications on
the device identification plate.
Do not connect device to power supply system if the data of the
supply voltage does not match the values of the device.

4.3 Battery (option)

The devices are available with and without batteries. The device can be operated without a mains power supply if a battery is installed.
The installation or removal of the battery may only be carried out by our customer services and/or a company or person assigned by us.
The battery pack must be fully charged prior to first start-up. In order to fully charge the battery pack, the device must be operated with mains supply for about 11 hours.
WARNING Incorrect handling of lithium ion battery pack!
The consequences can be:
Risk of explosion Leaking of corrosive solutions Escaping of harmful vapors Severe bodily injuries and damage to property
Do not short-circuit, reverse polarity of, open or solder the lithium
ion battery pack. – Keep away from heat and corrosive liquids. – Please follow manufacturer's instructions when disposing of used
batteries.
18 38019417000 en
Operating instructions iS10 / iS20 Installation
WARNING Risk of explosion due to incorrect lithium ion battery pack!
This may result in serious injuries and damage to property.
Always use the prescribed battery type for the lithium ion battery
pack. – Please follow manufacturer's instructions when disposing of used
batteries.

4.4 Removable power cable (optional)

The option "Removable power cable" has not been approved and can­not be ordered for devices with UL approval.
Devices with battery pack will be supplied with a removable power cable. In order for the screwable plug-in connection to comply with the specified IP protection type of the termi­nal, the following requirements must be met:
4
In battery operation the protective cap must be firmly screwed to the device plug-in
connector.
In mains operation the cable plug-in connector must be firmly screwed to the device
plug-in connector.
Removable power cables must not be damaged on both ends of the plug-in connections. Both ends of the plug-in connections must be con­nected according to the coding.
Device plug-in connector (housing variant: table top display)
The device plug-in connector for the power cable is located on the right side of the de­vice between housing and load receptor.
Fig. 6: Device plug-in connector
38019417000 en 19
4
Installation Operating instructions iS10 / iS20
Device plug-in connector (housing variant: column, wall-mounted, remote)
The device plug-in connector for the power cable is located on the back of the device.
Fig. 7: Device plug-in connector
Protective cap
The protective cap 1 must be mounted to the device plug-in connector if no power ca-
ble is connected to the device.
Fig. 8: Firmly screwed together protective cap
20 38019417000 en
Operating instructions iS10 / iS20 Installation
1
2
1
2

4.4.1 Connect power cable

DANGER Electrical voltage on the plug-in connector with mains plug inser-
ted!
Danger to life due to electrical shock.
Do not apply voltage to or separate removable power cable.
4
Fig. 9: Cable plug-in connector
Coding of cable plug-in connector
Coding of device plug-in connector
Attach cable plug-in connector to device plug-in connector according to coding.
Fig. 10: Screw connection
Secure screw connection
Unsecure screw connection
Firmly screw together screw connection between cable plug-in connector and device plug-in connector.
In order to ensure a secure connection there must be no gap between the connecting pieces.
38019417000 en 21
5
1
2
3
4
5
6
7
Operation Operating instructions iS10 / iS20

5 Operation

5.1 Display and operating field

Fig. 11: Display and operating unit
Standby/reset key
Status symbols
Weight, tare and application values
Bar segments
Unit of measured value (g, kg, t, lb, oz, pcs)
LED display (only with iS20) Tolerance control:
left LED lights up red: Under weight – LED in the middle lights up green: Good weight – right LED lights up yellow: Overweight
Registration: – all LEDs light up green: Registration process running Control panel
Key functions
Key Function in weighing operation Menu functions
Briefly press key: Save data and restart.
Press and hold key for more than 3 s: Standby mode.
Configurable function key, see
Return to weighing mode. <ESC> key.
page 25.
After a restart: Select menu, see page 26.
In Menus, scroll upwards or increase value.
22 38019417000 en
Operating instructions iS10 / iS20 Operation
Key Function in weighing operation Menu functions
Zero setting To the left to change decade to be edi-
ted.
Deleting tare In Menus, scroll downwards or reduce
value.
taring To the right to change decade to be edi-
Function in weighing operation: Item recording adding.
Menu functions: Confirm input, <Enter> key.
Special characters in the display
This symbol signalizes that the displayed measured value is a net value. Tar­ing was performed.
Operating mode tendency control
In this operating mode the weight is determined in reference to a specified tar­get weight. The absolute or relative difference as compared to the target weight will be displayed.
Operating mode totals
In this operating mode the weight values of several items from the total memo­ry are displayed.
Operating mode number of pieces
In this operating mode the number of pieces of several products to be weighed is established and displayed after the piece weight was weighed.
ted.
5
Display tare value
This symbol signalizes that the display shows a tare value.
T: Weighed value (tare balancing)
PT: Manual tare value (EDP)
Battery operation
This symbol shows the battery pack operation in scales with integrated battery pack. If the battery pack power falls below 50%, the symbol flashes.
Registration
This symbol shows that a recording process is running (data transfer to printer or EDP interface).
Unit symbol
This field shows the unit of the measured value (g, kg, t, lb, oz, pcs).
38019417000 en 23
5
Operation Operating instructions iS10 / iS20
Symbol for approved or non-approved operation
For scales subject to legal control only: In approved operation this symbol must not flash. If the symbol flashes, metrologically relevant data is not protec­ted.
Multi-range scale: effective range is "1", "2" or "3".

5.2 Switch device on / reset

Plug in power pack The scale is supplied via the network.
Key Display presentation Process description
All display segments will be activated.
The program number is briefly displayed.
Middle segments are briefly displayed.
Press key to save inputs or to restart the device.
The device automatically switches to weighing mode. The scale is ready for operation.

5.3 Switching device off

After a brief warm-up period, the maximum weighing accuracy is reached. It is recommended to leave the weighing terminal connected to the mains voltage during the entire day. This will ensure a constant operating temperature and maximum weighing accuracy.
Key Process description
Press and hold key for more than 3 s.
The device changes to stand-by mode.
Unplug to separate device from power supply.
24 38019417000 en
Operating instructions iS10 / iS20 Operation

5.4 Function assignment of key F1

Key Display presentation Process description
All display segments will be activated.
Press and hold while middle segments are being displayed.
The function selection appears.
Counting operation
With selection call up reference quantity.
Possible selection: 5 / 10 / 15 / 20 / 25 / 50 / 75 / 100 / 125
Fine division
Total display
Call-up: Fixed tare value
Change fixed tare value: Place weight on scale and confirm with
Display: Tare value
.
5
Display: Gross value
Display of supply / battery voltage in volts
Tolerance check
Incl. saving: Press key.
38019417000 en 25
Operation Operating instructions iS10 / iS20

5.5 Functions of menu key F2

In order to select a function under <F2> proceed as follows:
Key Display presentation Process description
Restart device.
5
All display segments will be activated.
Press and hold while middle segments are being displayed.
The menu selection appears.
Service parameters, see page 27
Ethernet interface parameters, see page 75
Total display, see page 43
PC/EDP interface parameters, see page 50
Printer interface, see page 61
Tolerance control, see page 33
General parameters, see page 78
Confirm selection of desired function with .
26 38019417000 en
Operating instructions iS10 / iS20 Operation

5.6 Change the tens digit for input of numbers

In order to change from a 1-digital to a 10-digital decade in case of two-digital menu num­bers, proceed as follows:
Key Display presentation Process description
or
or
or
Example: PC/EDP interface parameter is called up.
The first parameter is displayed.
The tens digit is selected.
A bar above the tens digit marks the selection.
Select desired parameter number.
The ones digit is selected.
A bar above the ones digit marks the selection.
Select desired parameter number.
5
Confirm input with .

5.7 Display metrological data

The verification status displays if there were verification relevant parameter changes. Fur­thermore, the identification of the connected load cell will be displayed.
Key Display presentation Process description
Weight symbol flashing: Calibration switch is not secured.
Weight symbol static: Calibration switch is in se­cured status.
38019417000 en 27
Operation Operating instructions iS10 / iS20
Display metrological data
Key Display presentation Process description
Restart device.
Press and hold while middle segments are being displayed.
5
The menu selection appears.
Select service parameter menus and confirm with
Select verification status and confirm with .
Call up service parameters
Key Display presentation Process description
A service parameter menu is called up.
Select scale parameters.
Start calibration.
.
Call up metrologically approved data storage ("alibi memory"), see page 31.
Select verification status.
Call up logbook, see page 30.
Return to weighing mode.