Information and descriptions contained in this manual are the property of Bistos Co., Ltd. and may
not be copied, reproduced, disseminated, or distributed without express written permission from
Bistos Co., Ltd.
Information furnished by Bistos Co., Ltd is believed to be accurate and reliable. However, no
responsibility is assumed by Bistos for its use, or any infringements of patents or other rights of third
parties that may result from its use. No license is granted by implication or otherwise under any patent
or patent rights of Bistos Co., Ltd.
The information contained herein is subjects to change without notice.
Prepared by:
Bistos Co., Ltd.
th
7
302, Galmachi-ro, Jungwon-gu, Seongnam-si,
Gyeonggi-do, Korea
15.2 Recommended separation distances between portable and mobile RF communications
equipment and BT- 720 -----------------------------------------------3 4
Used to identify safety information.
Be well-known this information thoroughly before using BT-720.
Used to identify safety information.
Indicates the protection level against the ingress of liquid.
It correspond the device, patient monitor.
Indicates the protection level against the ingress of liquid.
It correspond the accessories for SpO2.
Indicates DC power supply.
Indicates the device is in the battery operation mode.
0 Safety information
Before using BT-720 Patient monitor, read this entire manual and be fully understood the following
safety information to prevent injury of patient and user.
Symbols Used
The following symbols identify all instructions that are important to safety. Failure to follow these
instructions can lead to injury or damage to the patient monitor. When used in conjunction with the
following words, the symbols indicate:
Can lead to serious injury or death.
Can lead to minor injury or product/property damage
The following symbols are placed on product, label, packaging and this manual in order to stand for
the information about:
Be well-known this information thoroughly before using BT-720
IPX1
IPX2
IPX1 is protection against some water drops falling vertically.
IPX2 is protection from some water drops when the device is tilted up to
and including 15°.
Refer to operation manual. Read manual before placing the device.
Indicates nurse call interface.
Indicates network interface.
Indicates USB interface.
Indicates power adapter polarity.
Indicates the production date.
Indicates the manufacturer.
Indicates the serial number of the device.
Indicates the authorized representative in the European Community of
manufacturer.
Indicates a defibrillation-proof type BF applied part.
Indicates a defibrillation-proof type CF applied part.
Indicates CLASS II equipment.(Adapter)
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Indicates the date after which the medical device is not to be used.
Page 6
BT-720 Operation manual
5
Indicates to not dispose the device together with unsorted municipal
put on the market after 13 August 2005.
Indicates to keep the device dry.
Indicates the medical device that can be broken or damaged if not handled
carefully.
Indicates to keep upright
Indicates the maximum stacking limit.
Indicates the temperature limitation for operation, transport and storage.
Indicates the humidity limitation for operation, transport and storage.
Indicates the range of atmospheric pressure to which the medical device
can be safely exposed.
Indicates the device contains natural rubber latex.(Accessories)
Indicates the packing material is recyclable.
waste(for EU only). The solid bar symbol indicates that mains adapter is
0.1 General precautions, warnings and cautions
• Examine the patient monitor and any accessories periodically to ensure that the cables,
adapter cords and instruments do not have visible evidence of damage that may affect
patient safety or performance. The recommended inspection interval is once per week or
less. Do not use the patient monitor if there is any visible sign of damage.
• Only the DC power adapter supplied with the BT-720 is approved for use with the device.
• Do not attempt to service the BT-720 patient monitor. Only qualified service personnel by
Bistos Co. Ltd. should attempt any needed internal servicing.
• Perform periodic safety testing to insure proper patient safety. This should include
leakage current measurement and insulation testing. The recommended testing interval is
once per year.
• If the hospital or healthcare institutions using this device fail to implement a satisfactory
maintenance schedule, it will result in device failure and may endanger the patient’s
safety.
• Use the patient monitor under the conditions specified in this operation manual. Beyond
the conditions, the patient monitor may not function properly and the measurement
results may not accurate and may result in device failure or endangering the patient’s
safety.
• Do not operate the BT-720 patient monitor if it fails to pass the power on self-test
procedure.
• During the operation, do not disconnect any cable.
• The BT-720 patient monitor is intended to be used by clinical professionals or trained
doctors, nurses or laboratory assistant.
• Do not service and maintain or clean the device including accessories while in use with a
patient.
• Using the device to one patient at a time.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 7
BT-720 Operation manual
6
WARNING
An operator may only perform maintenance procedures specifically described in
•Thoroughly read and understand the manual prior to use of the BT-720. Failure to
do so could result in personal injury or equipment damage.
• The device is intended for clinical patient monitoring, and only trained and
qualified doctors and nurses should use the device.
• The alarm volume, upper and lower alarm limits should be set according to the
actual situation of the using environment. Do not just rely on audio alarm system
while monitoring the patient, because too low alarm volume or muted alarm may
result in notice failure of alarm situation and endanger the patient’s safety. Please
pay close attention to the actual clinical status of the patient.
• Use only the power adapter supplied with monitor.
• Position the monitor where it is easy to de-energize the monitor when needed.
• Do not open the enclosure to avoid an electric shock. Any repair and upgrade of
monitor should be done by service personnel trained and authorized by Bistos. Co.,
Ltd.
• When handling packaging materials, abide by local laws and regulations or hospital
waste disposal regulations. Keep the packaging materials away from children.
• Do not use in the presence of flammable anesthetics to prevent explosion or fire.
• Install the power lines and cables of accessories carefully to avoid patient
entanglement or suffocation, cables tangled or electrical interference.
• When the monitor is used together with electrosurgical devices, the user (a doctor
or a nurse) should ensure the safety of the patient and instrument.
• The physiological wave, physiological parameters and alarm information displayed
on the monitor are only for the doctor’s reference and should not be directly used
as the basis for clinical treatment.
• This is not a therapeutic device.
• For patients with pacemakers, the cardio tachometer may count the pacemaker
pulse in case of a cardiac arrest or arrhythmias. Never rely solely on the cardio
tachometer alarm. Closely monitor the patients with pacemaker. For the inhibition
of the device on pacemaker, refers to this manual.
• Use of accessories other than those listed and approved for use with this product
may result in increased emissions or decreased immunity.
• Medical electrical equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in
this manual. In addition, portable and mobile RF communications equipment can
affect medical electrical equipment.
• The equipment shall not be used adjacent to other devices unless verification of
normal operation in the configuration in which it is to be used can be achieved.
• Keep matches, and all other sources of ignition, out of the room in which the
patient monitor is located. Textiles, oils, and other combustibles are easily ignited
and burn with great intensity in air enriched with oxygen. Personal injury or
equipment damage could occur.
• A fire and explosion hazard exists when performing cleaning or maintenance
procedures in an oxygen-enriched environment.
• The patient monitor has been validated with the accessories and options listed in
this manual and found to comply with all relevant safety and performance
requirements applicable to the device. It is therefore the responsibility of the
person or organization who makes an unauthorized modification, or incorporates
an unapproved attachment to the device.
•
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 8
BT-720 Operation manual
7
this manual.
or replace components.
CAUTION
Please install or carry the instrument properly to prevent damage due to falling,
depends on use of proper cables.
WARNING
Unplug the monitor from its power source prior to cleaning or maintenance to prevent
a build-up of conductive dust or dirt. Do not allow cleaning agents to contact electrical
• Do not remove the covers of a BT-720 yourself to avoid damage to the equipment
and unexpected electrical shock. Only qualified Bistos service engineer must repair
•
collision, strong vibration or other mechanical force.
• Avoid instrument splashed by water.
• Avoid high temperatures, the instrument should be used within a temperature
range of 5 ℃ ~ 40 ℃。
• Avoid using instrument in the environment such as pressure is too high, poor
ventilation, dusty, or contain salt, sulfur gas and chemical.
• Before using the monitor, check the monitor and accessories if there is damage
that may affect patient safety. If there is obvious damage or aging, replace the
parts before use. The replacement should be made with same parts of original
parts.
• Before powering on the device, make sure that the power used by the device
complies with the supply voltage and frequency requirements on the equipment
label or in the Operator’s Manual.
• Equipment should be tested at least once a year, the test should be done and
recorded by trained, have security testing knowledge and experienced personnel.
If there are any problems in the tests, they must be repaired.
• When the instrument and accessories are about to exceed the useful life
(expected service life: 5 years), it must be treated in accordance with relevant
local laws and regulations or the hospital's rules and regulations.
• Do not connect to other equipment or network which not specified in the
instruction for use, in risk of external high voltage.
• Do not connect any equipment or accessories that are not approved by the
manufacturer or according to IEC 60601-1 to the monitor. The operation or use of
non-approved equipment or accessories with the monitor is not tested or
supported, and monitor operation and safety are not guaranteed in such a case.
• Any non-medical equipment (such as the external printer) is not allowed to be
used within the patient vicinity (1.5m/6ft.).
• Parts and accessories used must meet the requirements of the applicable safety
standards, and/or the system configuration must meet the requirements of the
medical electrical systems standard.
• Ensure that the conductive parts of electrodes and associated connectors,
including neutral electrodes, do not come in contact with earth or any other
conducting objects.
• Protection of ME EQUIPMENT against effects of discharge of a cardiac defibrillator
0.2 Shock hazards
personal injury or equipment damage.
Some chemical cleaning agents may be conductive and leave a residue that may permit
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 9
BT-720 Operation manual
8
components and do not spray cleaning solutions onto any of these surfaces. Personal
service documentation should service the monitor.
WARNING
Improper operation may cause the internal lithium ion battery to be hot, ignited or
Properly dispose of or recycle the depleted battery according to local regulations.
Avoid placing in an area where
30 % ~ 85 %.
Power off when the
could be damaged.
injury or equipment damage could occur.
Do not expose the unit to excessive moisture that would allow for liquid pooling.
Personal injury or equipment damage could occur.
Do not touch the patient and signal input/output parts simultaneously
Due to the risk of electrical shock hazard, only qualified personnel with appropriate
0.3 Battery warnings
exploded, and it may lead to the decrease of the battery capacity. It is necessary to read
the operation manual carefully and pay more attention to warning message.
Do not open the battery compartment. Only the qualified service personnel authorized
by the manufacturer can open the battery compartment and replace the battery, and
batteries of same model and specification should be replaced.
Be careful when connecting the battery with polarity.
Do not use the battery near fire or environmental temperature exceeds 60 ℃. Do not
heat or splash the battery or throw it into fire or water.
Do not destroy the battery. Do not pierce battery with a sharp object such as a needle.
Do not hit with a hammer, step on or throw or drop the battery. Do not disassemble or
modify the battery. The battery can heat, smoke, deformation or burning.
When leakage or foul smell is found, stop using the battery immediately. If your skin or
cloth comes into contact with leaked liquid, cleanse it with clean water at once. If the
leaked liquid splashes into your eyes, do not wipe them. Irrigate them with clean water
first and go to see a doctor immediately.
0.4 General precautions on environment
Do not keep or operate the equipment under the environment listed below.
Avoid placing in an area exposed
to moisture. Do not touch the
equipment with wet hand.
Avoid exposure to direct
sunlight
high variation of temperature
exists. Operating temperature
ranges from 5
Operating humidity ranges from
℃ ~ 40℃.
Avoid placing in an area where
there is an excessive humidity
rise or ventilation problem.
Avoid placing in an area where
chemicals are stored or where
there is in danger of gas leakage.
Do not disjoint or disassemble
the device. Bistos Co., Ltd. does
not have liability of it.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Avoid in the vicinity of
electric heater.
Avoid placing in an area
where there is an excessive
shock or vibration.
Avoid dust and especially
metal material enter into
the equipment
equipment is not fully
ready to operate.
Otherwise, the equipment
Page 10
BT-720 Operation manual
9
Blood oxygen saturation (SpO2)
Indication:
in pulmonary
rehabilitation
Contraindications
be hypoxic
Source: NHS. “Clinical Procedure_ Procedure for Pulse Oximetry/SPO2”. Wirral Community NHS
Trust. Sep, 2013
Non-invasive blood pressure (NIBP)
Indication:
- Assessing a person’s suitability for a spot or certain occupations
1 System basics
1.1 Intended use
The BT-720 Patient Monitors acquire the physiological signals for non-invasive blood pressure (NIBP),
pulse rate (PR) and blood oxygen saturation (SpO
processed, examines the data for alarm conditions and display the data. The monitor also provides
operating control for the user. The patient monitor intend to use in hospital clinical area such as
intensive care units, operating room, emergency department, to provide additional information to the
medical and nursing staff about the physiological condition of the patient. Th e BT-720 patient monitors
are intended to be used only under regular supervision of clinical personnel. It is suitable for adult and
pediatric, neonate. The intended locations of use are hospitals and clinics.
1) Intended patient population
- Adult (>18 years adults) and Pediatrics (30 days < and <18 years) and Neonate (0 days<
and <30days)
2) Intended user profile
- Doctor, physicians or nursing staff who is qualified personnel
- Basic experiences or knowledge on medical field, especially on patient monitoring
- Trained or requested to read IFU before use
3) Environment of use
- Hospital and clinic
- Requirements: Stable power source
4) Scope of application
This monitor is suitable for bedside monitoring of patient. This monitor enables blood oxygen
saturation (SpO
), pulse rate (PR) and monitoring. It is equipped with a replaceable built-in battery to
2
provide convenience for the patient movement in hospital.
5) Indications and contraindications
- Monitoring effectives of oxygen therapy
- A reading is needed to facilitate the completion of an early warning score to
inform clinical assessment
- Sedation or anesthesia
- Transport of patients who are unwell and require oxygenation assessment
- Haemodynamic instability (e.g. cardiac failure or Myocardial Infarction)
- Respiratory illness e.g. asthma, chronic obstructive pulmonary disease
- Monitoring during administration of respiratory depressant drugs, e.g. opiate
epidural or patient-controlled analgesia.
- Assessing oxygen saturation during physical activity e.g.
). The signals are converted into digital data and
2
- Pulse oximetry does not give an indication of haemoglobin so if the patient is
profoundly anaemic then their oxygen saturation may by normal but they may still
- To determine a patient’s blood pressure
- Screen for hypertension
- Following the effect of anti-hypertensive treatments in a patient to optimize their
management
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 11
BT-720 Operation manual
10
- Estimation of cardiovascular risk
- Figuring out whether a patient is clinically deteriorating or is at risk.
- Determining for the risk of various medical procedure
- Oscillometric blood pressure devices may not be accurate in patients with weak or
thready pulse
- In patients with heart beats below 50 beats/minutes, even if the rhythm is regular,
some of the semi-automatic devices are unable to reduce their deflation rate
sufficiently so that too rapid a falling in cuff pressure results in underestimation of
systolic blood pressure and overestimation of diastolic blood pressure.
- Do not apply to limb with AV fistula, significant injury or burn, or lymph node
Community NHS Trust. Dec, 2013
[2] Clinical Quality& Patient Safety Unit, QAS. Clinical Practice Procedures:
Assessment/Non-invasive blood pressure
1.2 Operating principle
Refer to the chapters for every physiological parameter from chapter 5 to chapter 6.
1.3 System configurations
Basic configuration of BT-720
• Main body with 4.3” touch screen and built-in lithium-ion battery
• Adult SpO2 probe and extension cable
• AC/DC adapter
Options of BT-720
•Non-invasive blood pressure cuff
Adult SpO2 sensor
(standard)
SpO2 extension cord
(Standard)
Adult NIBP cuff
(Option)
NIBP extension tube
(Option)
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Adapter
(Standard)
SpO2 sensor for adult 1ea
Cord to connect the SpO2 sensor
and main body
Measures NIBP for adult 1ea
Tube to connect the NIBP cuff and
main body
For power supply 1ea
1ea
1ea
Page 12
BT-720 Operation manual
11
Name
Description
NIBP tube interface. This is option. When you purchase the monitor
only with SpO2, this interface will be blocked.
2
SpO2
SpO2 cable interface
3
Display area
Display the waveform and measured value
Indicates the priority of physiological alarm and technical alarms in
- Low priority: Yellow, constant on
[Setting]
[NIBP]
invasive blood pressure measurement manually.
[Alarm reset]
1.4 Product outlook
Figure1-1: Front view
1.5 Description of monitor
1 NIBP
4 Alarm indicator
5
different colors and flashing frequencies.
- High priority: Red, fast flashing (1.4 ~ 2.8 Hz)
- Medium priority: Yellow, slow flashing (0.4 ~ 0.8 Hz)
Enter to the setting mode. Press again to close the setting mode.
Figure1-3: Front view
Figure1-2 : Rear view
6
7
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Start and stop the nonThis is option. When you purchase only with SpO2 this button will be
deleted.
To reset the alarm condition.
Page 13
BT-720 Operation manual
12
DC power
Battery indicator
- On: The battery is being charged or has been fully charged.
- Flashing: The monitor is being powered by the battery.
- Power On: Press down the key more than 2 seconds.
system will display the alarm message “The system will shut down
3 seconds”.
Name
Description
1
Network port
For CMS
3
Type B USB
Reserved for data transfer
4
Power adapter
15V, 2.4A adapter
Description
Current alarm message. When an alarm occurs, this area will displayed yellow or white
depending on the alarm type.
2
SpO2 value. Display the measured SpO2 value.
3
SpO2 upper alarm limit. Display the user set upper alarm limit
4
SpO2 waveform. Display the measured SpO2 waveform.
5
SpO2 lower alarm limit. Display the user set lower alarm limit
8
indicator
9
10
2
1.6 Understanding the display
[Power]
Type A USB For software upgrade
Turned on when the monitor is being powered by the adapter.
- Off: The battery has not been installed.
- Power Off: Press down the keys more than 2 seconds and the
Figure1-4: Rear view
Figure 1-5: Standard display
1
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 14
BT-720 Operation manual
13
6
Perfusion Index. Display the measured perfusion index.
7
Sweep speed. Display the user set SpO2 waveform speed.
8
RTIP. Display the current measured cuff pressure value.
9
MAP. Display the average pressure value.
10
Diastolic blood pressure value.
11
Systolic blood pressure value.
12
The unit of NIBP.
13
Battery status.
14
Patient type.
15
Display current time.
16
The unit of SpO2.
WARNING
All analog and digital devices connected to the monitor must be certified by IEC
. Without permission, any
hich a signal
input/output part may be connected.
1.7 Essential performance
This device Patient Monitor provides various patient vital signs such as pulse rate, blood oxygen
saturation and non-invasive blood pressure by placing the sensors to the appropriate site of patient.
The device is composed with display, control circuit and panel, and input part for various sensors. It
detects SpO2, PR and NIBP using specific sensors and cuff. The detected analog signal amplifies and
converted to digital. This concerted data feed to the CPU and converted to the display format as
number and waveform. This device is incorporated with alarm system. The alarm generated when the
detected signal range is beyond the user set alarm limits.
2 Preparing for operations
2.1 Installation
To ensure normal working of the monitor, read this chapter before use, and install as required.
standards (e.g. IEC 60950 Data processing equipment standard and IEC 60601-1 Medical
equipment standard). Furthermore, all configurations shall comply with valid version of
IEC 60601-1 standard. The personnel connecting additional devices to the input / output
signal ports are responsible for the compliance with IEC 60601-1 standard. If there is any
question, please contact Bistos.
If the patient cable interface and network interface are connected with multiple devices,
the total electric leakage current cannot exceed the allowable value.
The copyright of monitor software belongs to Bistos
organization or individual shall not interpolate, copy or exchange by any means or form.
When the monitor is combined with other devices, it must comply with IEC 60601-
1:2005 + A1:2012, and should not be connected with multi-socket outlet or extension
cord.
Do not connect the device on other equipment or network, to w
Prior to installation, the operator must ensure that the following space, power, environmental
requirements are met.
2.1.1 Unpack and check
BT-720 patient monitor was inspected rigorously at the factory before delivery, in order to avoid being
hit when transported, carried out careful packaging. Before unpacking, carefully inspect the package. If
any damage, please immediately contact the Bistos. Unpack in the correct way, carefully remove the
monitor and accessories from the box and check with the packing list. Check if there is any mechanical
damage, the all listed are completely packed. If you have questions, please contact the marketing
department of Bistos or agency.
Please keep the packing box and materials for use in future transporting or storage.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 15
BT-720 Operation manual
14
Ambient temperature
5℃ ~ 40 ℃
Relative humidity
30 % ~ 85 % (Non-condensing)
Atmospheric pressure
700 ~ 1060 mbar (hPa)
Transportation
Prevent severe shock, vibration, rain and snow splashing during transport.
The packaged monitor should be stored in well-ventilated room with
corrosive gases.
WARNING
Ensure that the monitor is used under specified environment. Fail to do this, the
damage to equipment and other unforeseen consequences.
WARNING
Do not try to open the monitor when the power is connecting.
During the operation, do not disconnect any cable.
2.1.2 Placement requirements
Equipment installation must meet:
- Ensure that the operating floor and the monitor have enough space for connecting the
accessory wires.
2.1.3 Power requirements
- DC power supply adapter
Input: A.C. 100 V ~ 240 V, 50/60 Hz
Output: D.C. 15 V, 2.5 A
- Built-in rechargeable lithium-ion battery: D.C. 11.1 V, 4400 mAh
2.1.4 Environmental requirements
The storage, transport and use of the monitor must meet the following environmental requirements.
Operating
environment
Storage
ambient temperature -20 ℃ ~ 60℃, relative humidity 0 ~ 95 % (Noncondensing), atmospheric pressure 700 ~ 1060 mbar(hPa), and without
The operating environment of the monitor should avoid noise, vibration, dust, corrosive or flammable
and explosive materials. In order to allow air flowing smoothly and achieve good heat dissipation, at
least 2 inches (5cm) clearance should be kept around the device.
When the device is moved from one environment to another, the device may have condensation due
to the differences in temperature or humidity. In this case, wait until the condensation disappears
before using the device.
technical specifications declared in this manual may not be met and it may result in
2.2 Connecting to power
Connect to power adapter in the following steps:
- Make sure that the AC power supply meets the following specifications: a.c.100V-240V,
50/60Hz.
- Use the power adapter provided with the monitor. Plug the power adapter into the
power connector of the monitor, and plug the other end of the power adapter into the
mains (low voltage power supply network facilities) power outlet.
3 Basic operations
3.1 Turn on
3.1.1 Check the monitor
- Before turn on the monitor, check whether there is mechanical damage to the monitor, and
whether the external cables and accessories are connected correctly.
- Plug the power adapter into the AC power outlet. If using battery power, make sure the battery
is fully charged.
- Check all the functions required for patient monitoring to make sure that the monitor operates
properly.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 16
BT-720 Operation manual
15
WARNING
If the monitor is damaged, or fails to work normally, do not use it for patient monitoring.
Please contact the maintenance personnel or Bistos immediately.
CAUTION
recommended.
3.1.2 Start the monitor
If finish to check the monitor, it is ready to start the monitor.
Press the [Power] key and the system enters the main interface within seconds.
- If any fatal error occurs during self-test, the system will alarm. If this case persists, please
stop to using the monitor and contact the maintenance personnel or Bistos.
- Check all available monitor functions to ensure that the monitor operate properly.
- If the monitor equipped with a battery, charge the battery after each use to ensure
sufficient power.
- After unpacking, when use the monitor first time, the monitor should be powered with
3.1.3 Connect the sensors
Connect the required sensor to the monitor and the monitoring site of patient.
3.1.4 Start monitoring
Start monitoring in the following steps:
3.2 Turn off
Turn off the monitor in the following steps
adapter.
- Check if the patient cable and the sensor are connected properly.
- Check if the settings of the monitor are corrects, such as patient type.
- For the details of parameter measurement or monitoring, see the appropriate section.
- The operator can operate according to their own habits, standing in front, left or right of
the monitor, easy to observe and operate the monitor.
- Disconnect the cables and sensors connected to the patient.
- Press and hold the [Power] key for 2 seconds to pop up the 3 seconds countdown
window, and the monitor turns off in 3 seconds.
• If the monitor is not turned off properly, you can simply disconnect the power to
shutdown forcibly. But the forced shutdown may cause data loss, and it is not
4 Setup the monitor
In the main screen, Press the [Setting] key enter the setup menu.
Select “Patient Type”, you can choose “Adult” or “Pediatric” or “Neonate”.
Select “Alarm Delay”, you can choose off, 1s, 2s, 3s, 4s, 45s, 5s, 6s, 7s or 8s.
Select “Alarm Volume”, and enter value(Range: 0-9).
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 17
BT-720 Operation manual
16
Select “Backlight”, and enter value(Range: 0-5), 5 is the brightest.
Select “Alarm Reminder Signal”, you can choose “On”, or “Off”.
Select “Alarm Reminder Interval”, you can choose “1min”, “2min”or “3min”.
Select “Date Format”, and set the date format in accordance with custom
“YYYY-MM-DD”: Year- Month-Day.
“MM-DD-YYYY”: Month -Day-Ye a r.
“DD-MM-YYYY”: Day-Month-Ye ar.
“Date (YMD)”: Set the year, month, and day.
“Time (24H)”: Set the hour, minute and second.
Select “Screen Setup”, set “SpO2+NIBP”.
Select “Language”, and select the option as needed:
“English”: The interface language of the monitor is English.
“Türkçe”: The interface language of the pulse oximeter is Turkish.
“Español”: The interface language of the pulse oximeter is Spanish.
“Français”: The interface language of the pulse oximeter is French.
“Polski”: The interface language of the pulse oximeter is Polish.
“Italiano”: The interface language of the monitor is Italian.
“Deutsch”: The interface language of the monitor is German.
“Magyar”: The interface language of the monitor is Hungary.
“Portuguese”: The interface language of the monitor is Portugal.
Select “Network”, you can choose “On”, or “Off”.
Select “Default”, back to the initial state.
Select “Network Setup”, set up the network.
Select “Passwords”, set Passwords.
Select “SpO2”, set “S-089”.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 18
BT-720 Operation manual
17
WARNING
Please use SpO2 sensor supplied from Bistos, operate in accordance with the Manual,
SpO2 low alarm limit cannot be less than 85.
NOTE
Do not put the oxygen probe and blood pressure cuff on the same limb, because blood
The monitor cannot be used to verify the accuracy of SpO2 probe and SpO2 equipment.
5 SpO
2
5.1 Overview
Blood oxygen saturation (SpO2) is the percentage of oxyhemoglobin (HbO2) capacity bound by oxygen
in the blood in the total hemoglobin (Hb) capacity that can be combined, that is, the concentration of
oxygen in the blood.
The principle for monitoring the pulse SpO
use the finger (or toe) as a transparent container for hemoglobin, use 660nm wavelength red light and
950nm near-infrared light as the incident light, maximum output power is 300 mW, measure the light
transmission intensity through the tissue bed, and calculate the concentration of hemoglobin and SpO
The passing lights depend on a variety of factors, most of which are constant. However, one of these
factors, the arterial blood flow, changes with time, as it is pulsating. By measuring the light absorbed
during pulsating, it is possible to obtain the arterial blood SpO
“plethysmography” wave and pulse rate signal.
The main screen displays “SpO
” value and “plethysmography” wave.
2
This monitor applies to measure SpO
neonate (<30 days). Contact SpO
“plethysmography” wave.
function of this monitor has been calibrated in factory.
SpO
2
The monitor is defibrillation proof, so the monitor operates normally after defibrillation.
5.2 Safety information
and observe all warnings and precautions.
Before monitoring, check whether the sensor cable is normal. When SpO
unplugged from the socket, the screen will display “SpO2 Sensor Off” error message,
and trigger an audible and visual alarm simultaneously.
If the sensor or sensor packaging has signs of damage, do not use this SpO
return it to the manufacturer.
If there is carboxyhemoglobin, methemoglobin or dye diluted chemical, the SpO
will have deviation.
When the patient has a tendency to hypoxia, use the oximeter to analyze blood samples
in order to fully grasp the patient's condition.
Do not put the sensor on limbs with arterial duct or intravenous tube.
Do not intertwine electrosurgical equipment cable with the sensor cable.
Avoid using the monitor and sensors while using the NMR equipment, in order to avoid
severe burns to the patient as a result of induced currents.
During long time continuous monitoring of a patient, check the position of SpO
once every 2 hours, and move properly when the skin changes or every four hours.
Some patients may require more frequent inspection, such as patients with perfusion
disorders or sensitive skin, because persistent and prolonged monitoring may increase
unpredictable skin changes, such as allergies, redness, blistering or pressure necrosis.
Before using, verify compatibility between the monitor, probe and cable, otherwise it
may cause injury to the patient.
Functional testers cannot be used to evaluate the accuracy of pulse oximetry and pulse
oximetry.
is to fix the probe fingerstall on the patient’s finger or toe,
2
. Detection pulsation can give a
2
of adults (>18 years) and pediatric (<18 years,>30 days),
2
probe to Patient’s finger (or toe) to get “SpO2” value and
2
sensor cable is
2
sensor;
2
sensor
2
value
2
.
2
flow occlusion during blood pressure measurement will affect the SpO2 readings.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 19
BT-720 Operation manual
18
NOTE
Turn on the monitor, plug in SpO2 probe and connect patient’s finger (or toe), monitor
new sensor.
5.3 Monitoring steps
1. Select the appropriate SpO
2. Turn on the monitor, and connect the SpO
3. Clean the measurement site, such as finger with nail polish.
4. Put the SpO
5. Select the appropriate alarm settings.
6. Start monitoring.
5.4 Setting SpO2
Select SpO2 parameter area to enter the SpO2 set interface
sensor probe on the patient’s finger or toe.
2
displays SpO2 wave, “SpO2 Pulse Search” displayed in the technical alarm area until the
monitor measured SpO
technical alarm area until the monitor measured pulse rate. Check the sensor mounting
position, whether the sensor is damaged or sensor type. Reconnect the sensor or use
sensor according to the patient.
2
value and pulse rate. “SpO2 Search Timeout” displayed in the
2
lead wire to the monitor.
2
Select “Wave Speed” and set wave speed to “12.5mm/s” or “25mm/s”. The faster speed, the
smoother wave.
Select “Wave Mode”, and set the wave drawing mode to “Scan” or “Fill”.
Select “Average Time”, and set the average time to “2-4s”, “4-8s”, “8-16s”.
The user can set the pulse volume. The pulse volume can be set to 0, 1, 2, 3, 4, 5, 6, 7, 8, or 9.
Select key to display the following interface.
Select “SpO2 Low Limit”, and enter value(Range: 0-99), Adult/Pediatric/Neonate Default: 90.
Select “SpO2 High Limit”, and enter value(Range: 1-100), Adult/Pediatric/Neonate Default:
100/100/95.
Select “Alarm Level”, you can choose “Mid” or “High”.
Select “PR Low Limit”, and enter value(Adult Range: 15-299, Pediatric/Neonate Range: 15-
Alarm ,Select “Default”, the alarm parameter is set to the default value
Select “PI Low Limit”, and enter value(Range: 0.00-19.90). Default: 0.00.
Select “PI High Limit”, and enter value(Range: 0.10-20.00). Default: 20.00.
5.5 Measuring influencing factors
During operation, the following factors can affect the accuracy of SpO2 measurement:
High-frequency radio wave interference, such as interference generated by the host system or
interference from electrosurgery instrument connected to the system.
Intravenous dye.
Too frequent movement of the patient.
External light radiation.
Sensor is improperly installed or improperly in contact with the patient.
Sensor temperature.
The sensor is placed on limbs with blood pressure cuff, arterial duct or lumen tube.
Concentration of non-functional hemoglobin such as carboxyhemoglobin (COHb) and
methemoglobin (MetHb).
Shock, anemia, hypothermia, and the application of vasoconstrictor drugs may reduce the
arterial blood flow to a level that cannot be measured.
The measurement also depends on the absorption of specific wavelengths of light by
oxyhemoglobin and reduced hemoglobin. If there is any other substance that absorbs the same
wavelength, the measurement may have false or low SpO
methemoglobin, methylene blue, and indigo carmine.
Accessories have passed the biocompatibility test and meet the requirements of ISO 10993-1.
Fluke's index 2XL Oxygen Analyzer can be used to check the function of the monitor and can be
used to assess the accuracy of the pulse rate but cannot be used to assess the accuracy of blood
oxygen.
Functional testers cannot be used to evaluate the accuracy of pulse oximetry probe and pulse
oximetry.
Measure the maximum temperature between the oxygen probe and the tissue contact
surface:Measured as described in Annex BB of ISO 80601-2-61,the temperature is less than 41 °
C.
6 NIBP (Option)
6.1 Overview
The monitor uses oscillometric method to measure noninvasive blood pressure (NIBP).
The oscillometric method for measuring blood pressure is to inflate a cuff with a certain amount of
pressure until the arterial blood flow has been completely blocked. As applied pressure decreases, the
arterial blood flow which was completely occluded gradually opened, and completely opened. Then,
the pulsation of the arterial vascular wall will generate a shock wave in the cuff. SBP, MAP, and DBP
are obtained by measuring and analyzing cuff pressure oscillations when deflating.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 21
BT-720 Operation manual
20
WARNING
Do not carry out non-invasive blood pressure measurement on patients with sickle cell
measurements and compare, if necessary, contact the Equipment Division.
Produce first most clear signal - reflect SBP
Oscillation amplitude reaches the peak - reflect MAP
When the cuff pressure is suddenly lowered - reflect DBP
Measuring mode: manual mode and automatic mode. Each mode shows systolic, mean and diastolic
blood pressure.
Manual mode
Using Manual mode start to measures by hand
Automatic mode measures
Use manual mode to open automatic mode, then the measure will automatically turn to
automatic mode after a certain time. During measurement, any error will stop the current
automatic measurement, but not affect next automatic measurement unless the time interval
less than 30s. If the time interval less than 30s, should delay the next automatic measurement,
keep the interval more than 30s.
The time interval can be choose In Automatic mode as 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180,
240, 480 minutes.
The monitor is defibrillation proof, so the monitor operates normally after defibrillation.
6.2 Safety information
disease and skin damage or any expected damage. Do not measure NIBP on traumatic
body part. This may cause further injury.
When pediatric and neonate patients are measured, in order to ensure the cuff pressure
does not exceed its maximum measurement range of patient types (Adult mode:
300mmHg and Pediatric mode: 240mmHg, Neonate mode: 150mmHg), you must ensure
that you have selected the correct patient type (see patient information menu settings).
Using the wrong type of pattern is likely to endanger the patient to patient safety, as
higher blood pressure levels for adults does not apply to pediatric and neonate.
For patients with severe coagulation disorder, determine if the automatic blood pressure
measurement is carried out according to the clinical evaluation, since the friction of body
and cuff may produce hematoma.
Do not install a cuff on the limbs with intravenous infusion or duct, because it may lead
to tissue damage around the duct when the cuff is inflated and makes the infusion slow
down or be blocked.
The inflatable tube connecting the blood pressure cuff and the monitor should be
smooth without entanglement. The pressure generated by being kinked connection
tubing may cause blood flow interference.
For patients with severe thrombotic disorders, determine whether to carry out
automatic blood pressure measurement according to the clinical situations, since the
limb bundled with a cuff may produce hematoma.
Measure blood pressure frequently will affect the distribution of blood flow, May
endanger the safety of patients.
Check the patient's physiological condition before measure blood pressure, in order to
ensure that long time measure will not damage the circulation of patients
For mastectomy patients, applying the NIBP cuff on the surgery side arm can cause
lymphedema. Measure blood pressure on opposite side arm.
Pressurization of the cuff can temporarily cause loss of function of simultaneously used
monitoring device on the same limb.
Measurement results may be affected by posture and mental state of the patient.
If there are doubts on the measurement results, please use other blood pressure
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 22
BT-720 Operation manual
21
6.3 Measurement limits
According to the patient's condition, the oscillometric method has some limitations. This
measurement is to look for the regular pulse waves generated by arterial pressure. If the patient's
condition makes this detection method difficult, the measured value becomes unreliable, and pressure
measurement time increases. The user should be aware that the following conditions may interfere
with measurement method, making the pressure measurement unreliable or extend the time. In this
case, the patient's condition does not allow measurement.
Patient movement
If the patient is talking, moving, shaking or cramping, the measurement will be unreliable or
even impossible, as these may interfere with the detection of arterial pressure pulse, and
extend the pressure measurement time.
Arrhythmia
If the patient shows arrhythmia which results in irregular heartbeat, the measurement will be
unreliable and even cannot be done, and the pressure measurement time will be extended.
Use of an artificial heart-lung machine
If a patient is connected to an artificial heart-lung machine, the measurement will be impossible.
Pressure changes
If the arterial pressure pulse is being analyzed to obtain a measured value at a certain time and
the blood pressure of the patient changes rapidly, the measurement will be unreliable or
impossible.
Severe shock
If the patient is in severe shock or hypothermia, the pressure measurement will not be reliable,
because the decrease of blood flow to the periphery would cause decrease in arterial pulsation.
Limit heart rate
If the heart rate is below 40bpm (beats / min) or above 240bpm (beats / min), the blood
pressure measurement is impossible.
Obese patients
A thick layer of fat around a limb blocks the arterial oscillation so that it cannot reach the cuff.
The accuracy is lower than normal.
Environmental Requirements
Measuring blood pressure should meet the environment range as follow:
ambient humidity 30%~85%, no condensing,
ambient temperature 5~40℃,
Atmospheric pressure: 700hPa~1060hPa.
NIBP performance and measurement accuracy will be affected beyond the range.
6.4 Measurement procedure
6.4.1 Prepare the measurement
1. Turn on the monitor, and check if it works properly.
2. Verify the patient category, and make changes if improper. Depending on the current patient
type, the patient type is selected in the patient information interface.
3. Connect the blood pressure cuff extension tube to the monitor.
4. Select the cuff in accordance with the following method, make sure that the cuff is completely
deflated, and then tie it to the upper arm or thigh of the patient.
Determine the limb circumference of the patient.
Select the appropriate cuff (marked with appropriate limb circumference). Cuff width
should be 40% of the limb circumference (50% for neonate) or 2/3 of the upper arm
length. The length of the inflated part of the cuff should be sufficient for 50%~80%
around the limb.
Place the cuff on the upper arm or thigh of the patient, and ensure that the marking
“ARTERIA” is located just above the appropriate artery. Make sure that the cuff does not
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 23
BT-720 Operation manual
22
NOTE
When have doubt about blood pressure measuring result, re-measure after the
sphygmomanometer.
6.4.2 Patient posture requirements during measurement
1. Sit comfortable or lie down relaxedly.
2. No crossing legs.
3. Back and elbow should be supported.
4. The center of NIBP cuff and the right atrium are at in the same level.
5. Remind patients, no talking during measurement and try to relax.
6.4.3 Start/stop measurement
Use the [NIBP start/stop] key on the monitor panel to start / stop the blood pressure
measurement.
6.4.4 Correcting measurement results
The position of limb blood pressure measurement should be in the same horizontal position of the
patient's heart. Otherwise, correct the measurement results with the following correction method.
wrap too tight around the limb, or it may cause distal discoloration or even ischemia.
patient sit-in about 5 minutes. If still have doubt, replace the blood pressure
measuring equipment and measure again.
The operator should be in the position where he/she can readily operate the
If the cuff is above the heart level position, increase 0.75mmHg (0.10kPa) per centimeter of gap
to the measured results.
If the cuff is below the heart level position, subtract 0.75mmHg (0.10kPa) per centimeter of gap
from the measured results.
If the patient is obese or clothes are too thick, subtract 5mmHg ~ 10mmHg (0.65kPa ~ 1.3kPa)
from the measured results.
6.5 Setting NIBP
Select “Patient Type”, you can choose “Adult” or “Pediatric” or “neonate”.
Select “Interval”, you can choose Manual, 1min, 2min, 3min, 4min, 5min, 10min, 15min, 30min,
60min, 90min, 120min, 180min, 240min or 480min.
Select “Unit”, and select the unit “mmHg” or “kPa”.
Select “Initial Pressure”, and set the appropriate cuff pressure value. When the patient is adult,
the pressure can be select from “140”, ”160”, ”180”. The default cuff pressure value is “160”.
Select “Initial pressure”, and set the appropriate cuff pressure value. When the patient is
pediatric, the pressure can be select from “140”, ”160”. The default cuff pressure value is “140”.
Select “Initial pressure”, and set the appropriate cuff pressure value. When the patient is
neonate, the pressure can be select from “100”, ”120”. The default cuff pressure value is “100”.
Select “Continue Measure”, continued measure pressure.
Select “Reset”, and restore the inflation pressure of the blood pressure pump to currently
configured initial settings. When the blood pressure pump is not working properly, but no
warning is given, you can reset the blood pressure pump, and automatically restores the blood
pressure pump.
Select key to display the following interface.
Select “SYS Low Limit”, and enter value (Adult Range: 30-279, Pediatric Range: 30-229, Neonate
Select “Alarm Level”, you can choose “Mid” or “High”.
Select “Default”, the alarm parameter is set to the default value.
Select “Calibration”, Users can not calibrate NIBP. If calibration is required, please contact your
service representative. Cuff pressure sensor should be checked and calibrated at least once
every two years by qualified professional service personnel.
Select “Leak Test”, the purpose of leakage test is to detect if the sealing of the air passage is in
good condition. If the leakage test passes, the alarm area displays “Leakage test Stopped”. If not
passed, the alarm area displays “Cuff leak” message. NIBP leakage test shall be at least once
every two years or when you think that the reading is not accurate.
6.6 Clean and disinfection method of NIBP cuff
If necessary, NIBP cuff and NIBP extension tube can be cleaned and disinfected together without
separated
6.6.1 Cleaning method
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 25
BT-720 Operation manual
24
NOTE
Please be especially careful to clean the air ball and control valve of whole air system.
the cuff and extension tube.
1. Prepare enzyme cleaning agent, distilled water and 10% solvent, respectively in different spray
bottle.
2. Sprinkle cleaning agent on NIBP cuff, connector and extension tube, keep 1 minute for the dry
stains.
3. Use a soft cloth to wipe smooth face. Use soft hair brush to brush visible stain and irregular
surface
4. Rinsed with copious amounts of distilled water.
Do not allow any liquid entering into reversing valve and saturated valve.
Don’t use a soft cotton ball and fiber to clean this accessory because they will stick on
6.6.2 Disinfection method
1. Sprinkle bleach solution (Formula: the proportion of water and bleaching powder to 1:10) then
keep 5 minutes
2. Wipe off excess bleach solution and elute with distilled water again
3. Natural dry cuff
7 Review
Use the Trend screen to recall all the historical patient data in a list, including monitoring time (in 1minute intervals), and SpO
the top of the list.
, PR, SYS, DIA and MAP values. The most recent measurements display at
2
To view historical data:
Figure 7-1: SpO2 trend (1)
Press the [Setting] key twice to display the Trend screen.
Select or key to page up and down to view patient data.
Press key to exit the Trend screen and display the Main screen or press the [Setting] key
to see the trend graph.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Figure 7-2: SpO2 trend (2)
Page 26
BT-720 Operation manual
25
Source
Alarm message
SpO2 Too High
SpO2 Too Low
PR Too High
PR Too Low
Systolic Too High
Systolic Too Low
Mean pressure Too High
Mean pressure Too Low
Diastolic Too High
Diastolic Too Low
Source
Alarm message
SpO2 Communication Stop
SpO2 Communication Error
SpO2 No Sensor
SpO2 Sensor Off
SpO2 Search Timeout
SpO2 Search Pulse
NIBP Communication Stop
NIBP self-check error
Cuff type error
Cuff loose or no cuff
Cuff leak
Air pressure error
Select ‘SpO2’, ‘PR’ or ‘PI’ to see.
Press key to exit the Trend screen and display the Main screen or press the [Setting]
key to see the NIBP Trend.
Select or key to page up and down to view patient data.
8 Alarm
Select button, and you can temporarily turn off the alarm sound of currently occurring
physiological alarms of the monitor, but the alarm information is still retained. For technical alarms,
clear the alarm state, display alarm prompt information, the alarm state icon area displays the
icon. When a new physiological alarm or technical alarm occurs, the alarm reset is automatically
canceled.
Physiological alarm information.
Technical alarm information.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
SpO2
NIBP
SpO2
NIBP
Page 27
BT-720 Operation manual
26
NIBP signal weak
NIBP over range
NIBP signal unstable
NIBP over pressure
NIBP signal saturated
NIBP system error
Measurement timeout
WARNING
Keep the battery out of the reach of children. Use only the designated battery.
hand directly.
9 Battery
9.1 Overview
The monitor has a built-in rechargeable battery to ensure that the monitor can also be used normally
in case of patient transfer or power failure. When the monitor is connected to an DC power source, it
will charge the battery no matter whether the monitor is turned on or not. In the case of power failure,
the system will automatically use the battery to power the monitor to avoid interrupting the monitor
working.
The battery icon on the screen indicates the battery status:
Battery is working properly and is fully charged.
Battery is working properly and the green part indicates the battery power.
Battery power is low, and requires charging immediately, or else the pulse
oximeter will turn off automatically.
Battery is not installed.
Battery is properly installed and being charged.
The battery power can only maintain for some time. Low battery voltage will trigger a high level
technical alarm “Battery Low”; in this case, connect the monitor to DC power and charge the battery.
9.2 Battery usage guide
Battery life depends on the frequency and time of use. If the battery maintenance and storage are
proper, the lithium battery life is three years. If you do not use the battery properly, its life may be
shortened. It is recommended to replace the lithium battery once every three years.
In order to ensure the maximum capacity of the battery, please note the following usage guide:
Do not drop the battery.
Check the battery performance once every two years. Before servicing the monitor or you
suspect that the battery is the fault source, also check the battery performance.
•
• If the battery is damaged or leaks, replace it immediately. Do not use a defective battery
for the monitor.
• Do not open the battery compartment. Only the qualified service personnel authorized
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
by the manufacturer can open the battery compartment and replace or check the
battery, and when it needs replacement, the replacement should be same model with
established battery.
• Do not disassemble the battery, throw it in fire, or short-circuit it. Battery fire, explosion
and leakage may lead to personal injury. Do not touch the leaking battery with bare
Page 28
BT-720 Operation manual
27
WARNING
Do not open the battery compartment. Only the qualified service personnel authorized
established battery.
WARNING
Do not disassemble the battery, throw it in fire, or short-circuit it. Battery fire, explosion
hand directly.
9.3 Checking battery performance
Please refer to the following steps to check the battery performance:
Disconnect the monitor from the patient and stop all monitoring or measurement.
Connect DC power to the monitor, and charge battery for more than 4 hours uninterruptedly.
Disconnect the DC power and power the monitor with battery until the monitor is turned off.
Battery duration reflects the battery performance.
If the battery operating time is significantly shorter than the time stated in specifications, please
contact our service personnel for replacing the battery.
•
by the manufacturer can open the battery compartment and replace or check the
battery, and when it needs replacement, the replacement should be same model with
9.4 Battery recycling
If the battery has visible damage or cannot store power, it should be replaced and recycled properly.
Follow the appropriate regulations to dispose of used batteries.
•
and leakage may lead to personal injury. Do not touch the leaking battery with naked
10 Caring and cleaning
10.1 Overview
In the using process, please make sure that there is no dust on or near your device. To prevent
damage, please use the diluted detergents and disinfectants specified in this Manual, and use the
lowest possible concentration. For the damage or accident caused by using other materials or
methods, our company does not assume any responsibility.
10.2 Cleaning
The device should be cleaned regularly. In the heavily polluted environment, increase the frequency of
cleaning. Before cleaning, please consult the hospital about device cleaning requirements.
Below are available cleaning agents:
Turn off the monitor and disconnect the power.
Use a soft cotton ball to adsorb appropriate amount of cleaning agent and wipe the
display screen.
Use a soft lint-free cloth to adsorb appropriate amount of cleaning agent and wipe the
surface of the device.
If necessary, use a clean, dry, lint-free cloth to remove any excess detergent.
Dry the device naturally in a ventilated cool environment.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 29
BT-720 Operation manual
28
WARNING
Before cleaning the monitor or sensor, turn off the power and disconnect the DC power.
soap and water.
CAUTION
To avoid damaging the monitor:
Do not leave any cleaning solution on the surface of any part of the device.
NOTE
Wipe the monitor and sensor surface with medical alcohol, dry it naturally or with clean,
for advice.
CAUTION
To prevent damage to the monitor, do not disinfect the monitor with gas (EtO) or
formaldehyde.
WARNING
If the hospitals or institutions using this instrument can’t implement a satisfactory
maintenance schedule, it will result in device failure and may endanger human health.
•
• The monitor should be kept clean. It is recommended to regularly clean the enclosure
surface and the display screen. Cleaning the enclosure with non-etching cleaner such as
•
Do not use strong solvents such as acetone.
Most cleaners must be diluted before use. Diluting should be according to the
manufacturer's instructions.
Do not use abrasive materials (such as steel wool).
Do not allow any liquid entering into the enclosure, and never immerse any part of
the device into liquid.
dry, lint-free cloth.
Bistos is not liable for effectiveness of using these chemicals for infectious disease
control. Please consult the infectious disease control officers or experts of the hospital
10.3 Disinfection
In order to avoid damage to the product, we recommend that the product is disinfected only when it is
deemed necessary by the hospital maintenance procedures. We also recommend that the instrument
to be disinfected must first be cleaned.
•
11 Maintenance
•
11.1 Checking
Check the following basic items before using the monitor:
Check for any mechanical damage.
Check all exposed wires, insertions and accessories.
Check all instrument functions that may be used for patient monitoring and ensure that the
instrument is in good working condition.
If the instrument function has any sign of damage, do not use this monitor for any patient monitoring.
Please contact the hospital's professional maintenance personnel or our customer service personnel.
Every 6-12 months or after each repair, a comprehensive examination must be performed by trained
and qualified technical service personnel, including functional safety checks; the specific inspection
items are as follows:
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 30
BT-720 Operation manual
29
Inspection / Maintenance Item
Frequency
Check the safety according to IEC
60601-1
At least once every two years, after replacing the
power supply or the monitor falls down.
Check all monitoring or measuring
functions not listed
At least once every two years, or when you suspect
that the measured value is not accurate.
At least once every two years, or follow hospital
regulations
At least once every two years, or follow hospital
regulations
WARNING
Use the accessories specified in this manual. Using other accessories may damage the
replace in time.
No.
Description
QTY
Type-number
Manufacturer:
Unimed Medical Supplies,Inc
2
SpO2 extension cable
U403-01
Manufacturer:
CO., LTD
UE36LCP1-150240SPA
Environment and power meet the requirements.
Device and accessories have no mechanical damage.
The power supply has no wear, and the insulation is good.
Specified accessories are used.
Alarm system is functioning correctly.
Battery performance meets the requirements.
Monitoring functions are in good working condition.
Ground impedance and leakage current meet the requirements.
If the instrument function has any sign of damage, do not use this monitor for any patient monitoring.
Please contact the hospital's professional maintenance personnel or our customer service personnel.
All checks that require disassembling the instrument must be performed by qualified service personnel.
Safety and maintenance checks may also be carried out by the Company's personnel.
11.2 Maintenance plan
The following tasks can only be done by qualified service personnel of Bistos. When the following
maintenance is needed, please contact your service representative. Before testing or maintenance,
clean and disinfect the device.
NIBP leakage test
NIBP calibration
12 Accessories
•
monitor, or cannot reach the safety and performance claimed in this manual.
• The operating and storage environment of the monitor should meet the requirements of
the accessories. Please refer to the manual of the accessories for these requirements.
• Disposable accessories can only be used once, because repeated use can cause
performance degradation.
• If the packaging or accessories have any sign of damage, do not use such accessories.
• For SpO
Standard accessories are as follows:
1 Adult Finger Clip SpO2 Sensor
3 Power Adapter 1
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Sensor and Blood Pressure Cuff, the normal life time is two years. Please
device must be powered by the internal power supply (battery).
Electric shock protection
grade
Type CF applied part (defibrillation proof)
Explosion protection
grade
Common equipment, no explosion protection
Liquid inlet protection
grade
IPX1
Operating mode
Continuous mode
Movement
Portable equipment
Power
Input: AC 100 ~ 240V (50/60 Hz)
Output: DC 15V / 2.4A
11.1V Li-ion battery 4400 mA
Operating Time(When it fully charged): 5 hours
Charging Time(Fully): 4 hours
Physical Characteristics
Dimensions
Main Unit: 254(W) X 90(H) X 185(D)
Weight
< 1.5 Kg
Display
Type
Color TFT touch screen LCD
Size
4.3”, 480 x 272 pixels
LED
Alarm Indicator
Yellow & Red
Adapter power indicator
1 green
Optional accessories list is as follows:
Adult Non-Invasive
1
blood pressure cuff
1
2 NIBP extension tube 1
XIAMEN CONJOIN ELECTRONICS TECHNOLOGY CO.,
13 Specifications
13.1 Safety specifications
13.1.1 Product category
In accordance with classification specified in the European Medical Device Directive 93/42/EEC, this
monitor is Class IIb device. The monitor is classified as follows in accordance with IEC 60601-1:
Category Name Specification
protection
13.1.2 Power
Adapter
When you question the integrity of the external protective earthing
or protective ground conductor parameter of the equipment, the
Rechargeable Battery
13.2 Hardware specifications
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 32
BT-720 Operation manual
31
Battery status indicator
1 green
Audio
Alarm sound (45 ~ 85 dB), key pressing sound
PR sound
Alarm sound meet the IEC 60601-1-8 standard requirements
Alarm signal
Alarm delay
Off, 1s, 2s, 3s, 4s, 5s, 6s, 7s, 8s, depending on the setup
Data storage
Trend
168 hours. Resolution: 1 min
Environment
Operating
Transport and storage
Temperature
5℃~40℃(41°F~104°F)
-20℃~60℃(-4°F~140°F)
Relative Humidity
30~85% RH, Non-condensing
0 ~ 95 % RH, Non-condensing
Atmospheric
pressure
70kPa~106kPa
70kPa~106kPa
SpO2
Standards compliant
ISO 80601-2-61:2011
Display range
0% ~ 100%
SpO2 display resolution
1%
SaO2 accuracy
±2% (70%~100%) (adult/pediatric mode);
not define when lower than 70% ;
Upper alarm limit
1%~100%
Lower alarm limit
0%~99%
SpO2 alerting signal generates a
delay
No delay
SpO2 value refresh period
1s/time
Low sensitivity
6~8s
Intermediate sensitivity
4~6s
Advanced sensitivity
2~4s
Low sensitivity
8s
Intermediate sensitivity
6s
Advanced sensitivity
2s
Alarm sign generates delay
period
0s
PR
Measuring range
25~250bpm
Resolution
1 bpm
Accuracy
±2% or ±2bpm,whichever is greater
NIBP
Speaker
13.3 Functional specifications
13.3.1 SpO2
±3% (70%~100%) (neonate mode);
SpO2 alarm limit range
Average period
Alarm condition delay period
13.3.2 NIBP
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
<
<
<
Page 33
BT-720 Operation manual
32
Standards compliant
IEC 80601-2-30:2009
Measurement method
Automatic oscillometric method
Operating mode
Manual, automatic
Useful life
100, 000 times
Measurement interval
in automatic mode
1/2/3/4/5/10/15/30/60/90/120/180/240/480min
Typical measurement
time
20~40s
Adult
Pediatric
Neonate
Systolic blood pressure
40-270
40-200
40-130
Mean blood pressure
20-230
20-175
20-100
Diastolic blood pressure
10-210
10-162
10-90
Measurement
accuracy
Maximum average error: ±5mmHg
Maximum standard deviation: 8mmHg
Resolution
1mmHg
Default
Pressure setting range
Adult
160mmHg
140mmHg, 160mmHg, 180mmHg
Pediatric
140mmHg
140mmHg, 160mmHg,
Neonate
100mmHg
100mmHg, 120mmHg,
Adult: 300mmHg
Pediatric: 240mmHg
Neonate: 150mmHg
Adult: 320~330mmHg
Pediatric: 265~275mmHg
Neonate: 160~165mmHg
Static Pressure
accuracy
±3mmHg
Electrical characteristics
Supply voltage
10V~14V DC
Maximum power consumption
3.6w
Quiescent current
50mA
Maximum current during
measurement
180mA
Maximum current during inflation
300mA
Faults
Solution
Not power on
Check the battery. If the battery is low, please contact to Bistos.
Blank Screen
Check the screen and screen line.
The system time is not
1. Set up error, can be reset through the system User
contact to Bistos.
No SpO2 waveform or
1. Is the red light on the finger sensor flashing? If not there might
3. Is the environmental temperature too low? Never expose the
Normal mode
measuring range
(mmHg)
Initial inflation
pressure
Overpressure
protection point
(software)
Overpressure
protection point
(hardware)
14 Common faults and maintenance
The following table shows the common faults on the operation, and the solution.
correct
value
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Maintenance menu.
2. The button battery on main control board is run out, please
be poor contact. Check the extension cable and the connector.
2. Is the patient’s arm under pressure? Never take blood pressure
and SpO2 measurements on the same arm.
Page 34
BT-720 Operation manual
33
patient’s arm to cold air since this can affect the readings.
removed?
SpO2 value turn on
monitoring
During long term monitoring, patient movement might result in
due to patient hand motion are normal.
Blood pressure
1
Check whether the pump is broken.
3
Check whether the blood pressure plate is normal.
Blood pressure
1
Check whether the blood pressure cuff is leak.
4
Check whether the pressure sensor is normal.
NOTE
Using unqualified accessories, sensors and cables will increase the electromagnetic
device.
The BT-720 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-720 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
The BT-720 uses RF energy only for its internal function.
to cause any interference in nearby electronic equipment.
The BT-720 is suitable for use in all establishments other
4. Has all patient nail polish, especially blue or purple been
and off during
SpO2 interruption. Keep the patient stabilized. SpO2 interruptions
.
measurement does
not start
2. Check whether the trachea is broken.
.
.
started, but couldn't
measure the value
2. Check whether the NIBP extension tube and machine connect
is well.
3. Check whether the deflating valve on blood pressure plate is
normal.
.
If the above doesn't solve the problem, please contact Bistos after-sales department or dealers.
15 Manufacturer’s declaration on EMC
BT-720 needs special precautions regarding EMC (Electromagnetic compatibility) and needs to be used
according to the EMC information provided in this user manual. Wireless communications equipment
such as wireless home network devices, mobile phones, cordless telephones and their base stations,
walkie-talkies can affect the BT-720 and should be kept at least 1 m away from the equipment.
emission and reduce the electromagnetic immunity of the device.
Do not put the device close to other devices or stack together. When necessary,
observe the device closely to ensure that it runs normally in the environment.
The device requires special EMC protection, and it is necessary to install and maintain it
in the environment that meets the following EMC information.
Even if other devices comply with CISPR emission requirements, they may also cause
interference to this device.
When the input signal amplitude is smaller than the minimum amplitude specified in
the technical specifications, it may result in inaccurate measurements.
Mobile communication devices or wireless network devices may have an impact on the
15.1 Electromagnetic emissions
RF emissions
CISPR 11
RF emissions
CISPR 11
Group 1
Class A
Therefore, its RF emissions are very low and are not likely
than domestic, and may be used in domestic
establishments and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 35
BT-720 Operation manual
34
warning is heeded:
The BT-720 is intended for use in an electromagnetic environment in which radiated RF
according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
[m]
�
�
[
3
]�
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is
Harmonic
emissions
IEC 61000-3-2
Class A
Warning: This BT-720 is intended for use by healthcare
professionals only. This equipment/ system may cause
radio interference or may disrupt the operation of nearby
Voltage
fluctuations /
flicker
Complies
equipment. It may be necessary to take mitigation
measures, such as re-orienting or relocating the BT-720 or
shielding the location.
emissions
IEC 61000-3-3
15.2 Recommended separation distances between portable and mobile RF communications
equipment and BT-720
disturbances are controlled. The customer or the user of the BT-720 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the BT-720 as recommended below,
Rated
maximum
output power
of transmitter
[W]
150 kHz to 80 MHz
d=
3.5
80 MHz to 800 MHz
d=
3.5
800 MHz to 2.5
GHz
7
d=
0.35 0.35 0.23
1.11 1.11 0.74
3.5 3.5 2.34
11.07 11.07 7.38
35 35 23.24
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
affec ted by absorption and reflection from structures, objects and people.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 36
BT-720 Operation manual
35
Electrostatic
±8 kV Contact
±8 kV Contact
Floors should be wood,
be at least 30 %.
Electrical fast
±2 kV for power
±2 kV for power
Mains power quality
Surge
±1 kV differential
±1 kV differential mode
Mains power quality
Voltage dips, short
< 5 % Uт (> 95 %
dip in Uт ) for 5 s
< 5 % Uт (> 95 % dip in
Mains power quality
power supply.
Power frequency
3 A/m
3 A/m
Power frequency
environment.
15.3 Electromagnetic immunity
The BT-720 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-720 should assure that it is used in such an environment.
Immunity test
discharge (ESD)
IEC 61000-42:2009
transient/burst
IEC 61000-44:2004
IEC 61000-45:2006
interruptions and
voltage variations
on power supply
input lines
IEC 61000-411:2004
IEC 60601
Test level
±15 kV air
supply lines
±1 kV for
input/output
lines (>3m)
mode
±2 kV common
mode
dip in Uт) for 0.5
cycles
40 % Uт (60 % dip
in Uт ) for 5 cycles
70 % Uт (30 % dip
in Uт) for 25
cycles
<5 % Uт (> 95 %
Compliance level
±15 kV air
supply lines
±1 kV for input/output
lines (>3m)
±2 kV common mode
Uт) for 0.5 cycle
40 % Uт (60 % dip in Uт )
for 5 cycles
70 % Uт (30 % dip in Uт)
for 25 cycles
<5 % Uт (> 95 % dip in
Uт ) for 5 s
Electromagnetic
environment -guidance
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
should be that of a
typical commercial or
hospital environment.
should be that of a
typical commercial or
hospital environment.
should be that of a
typical commercial or
hospital environment. If
the user of the BT-720
image intensifier
requires continued
operation during power
mains interruptions, it is
recommended that the
BT-720 be powered from
an uninterruptible
(50 Hz and 60 Hz)
magnetic field
IEC 61000-48:2010
NOTE Uт is the a.c. mains voltage prior to application of the test level.
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
magnetic fields should
be at levels
characteristic of a typical
location in a typical
commercial or hospital
Page 37
BT-720 Operation manual
36
Conducted RF
3 Vrms
3 Vrms
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Electromagnetic environment - guidance
ers, an electromagnetic site survey should be
The BT-720 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-720 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
IEC 61000-4-6:2009
Portable mobile RF communications equipment should be used no closer to any part of the
BT-720, including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d-
1.2� (d-3.5�)
d-1.2� (Resp: d-3.5�) 80 to 800MHz
d-1.2� 800M to 2.5GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as deter-mined by an electromagnetic site survey
should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol :
NOTE 1) At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitt
considered. If the measured field strength in the location in which the BT-720 is used exceeds
the applicable RF compliance level above, the BT-720 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the BT-720.
b
Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.
150 kHz to 80 MHz
a
b
,
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Page 38
BT-720 Operation manual
37
Product Name
Patient Monitor
Model Name
BT-720
Serial No.
Warranty Period
2 Ye ars
Date of Purchase
Hospital:
Telephone:
Product Warranty
Customer
Sales Agency
Manufacture Bistos Co., Ltd.
※ Thank you for purchasing BT-720.
※ This product is manufactured and passed through strict quality control and inspection.
※ Compensation standard concerning repair, replacement, refund of the product complies with
“Framework Act on Consumers” noticed by Fair Trade Commission of Republic of Korea.
Address:
Name:
Service Telephone and Fax. Numbers
Telephone: +82 31 750 0340
Fax: +82 31 750 0344
th
FL., A Bldg., Woolim Lions Valley 5-cha, 302,
7
Galmachi-ro, Jungwon-gu, Seongnam-si,
Bd. Général Wahis 53 1030 Brussels / BELGIUM
Bistos Co., Ltd.
Gyeonggi-do, Korea
bistos@bistos.co.kr
Telephone: +32 2. 732.59.54
Fax: +49 2. 732.60.03
www.bistos.co.kr
Obelis s.a
P/N:720-ENG-OPM-ENG-R03 Bistos Co., Ltd. 2020.10
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.