Information and descriptions contained in this manual are the property of Bistos Co., Ltd. and
may not be copied, reproduced, disseminated, or distributed without express written
permission from Bistos Co., Ltd.
Information furnished by Bistos Co., Ltd is believed to be accurate and reliable. However, no
responsibility is assumed by Bistos for its use, or any infringements of patents or other rights of
third parties that may result from its use. No license is granted by implication or otherwise
under any patent or patent rights of Bistos Co., Ltd.
The information contained herein is subjects to change without notice.
Prepared by:
Bistos Co., Ltd.
th
7
FL., A Bldg., Woolim Lions Valley 5-cha,
302, Galmachi-ro, Jungwon-gu, Seongnam-si,
Gyeonggi-do, Korea
13.2 Recommended separation distances between portable and mobile RF communications
equipment and BT-710 ····································································································· 50
Figure 1-1: Front view ············································································································ 12
Figure 1-2: Rear view ············································································································· 13
Figure 1-3: Top view ·············································································································· 13
Figure 1-4: Bottom view ········································································································ 13
Figure 1-5: Front view ············································································································ 14
Figure 1-6: Top view ·············································································································· 15
Figure 1-7: Bottom view ········································································································· 15
Figure 1-8: Standard display ·································································································· 16
BT-710 Operation manual
4
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
CAUTION
Indicates the protection level against the ingress of liquid.
It correspond the device and the accessory for SpO2.
manufacturer.
0 Safety informatio
Before using BT-710 Pulse oximeter, read this entire manual and be fully understood the
following safety information to prevent injury of patient and user.
Symbols Used
The following symbols identify all instructions that are important to safety. Failure to follow
these instructions can lead to injury or damage to the pulse oximeter. When used in
conjunction with the following words, the symbols indicate:
Can lead to serious injury or death.
Can lead to minor injury or product/property damage
The following symbols are placed on product, label, packaging and this manual in order to
stand for the information about:
IPX2
Used to identify safety information.
Be well-known this information thoroughly before using BT-710.
Used to identify safety information.
Be well-known this information thoroughly before using BT-710
IPX2 is protection from some water drops when the device is tilted up
to and including 15°.
Refer to operation manual. Read manual before placing the device.
Indicates the production date.
Indicates the manufacturer.
Indicates the serial number of the device.
Indicates the authorized representative in the European Community of
Indicates a BF applied part.
Indicates CLASS II equipment.(Adapter)
BT-710 Operation manual
5
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Indicates the medical device that can be broken or damaged if not
on for operation, transport and
Indicates the date after which the medical device is not to be used.
Indicates to keep the device dry.
handled carefully.
Indicates to keep upright
Indicates the temperature limitati
storage.
Indicates the humidity limitation for operation, transport and storage.
Indicates the packing material is recyclable.
Indicates to not dispose the device together with unsorted municipal
waste (for EU only). The solid bar symbol indicates that mains adapter
is put on the market after 13 August 2005.
0.1 General precautions, warnings and cautions
•Examine the pulse oximeter and any accessories periodically to ensure that the cables
including adapter cords and instruments do not have visible evidence of damage that
may affect patient safety or performance. The recommended inspection interval is
once per week or less. Do not use the pulse oximeter if there is any visible sign of
damage.
•Only the DC power adapter supplied with the BT-710 is approved for use with the
device.
•Do not attempt to service the BT-710 pulse oximeter. Only qualified service personnel
by Bistos Co. Ltd. should attempt any needed internal servicing. There is no user
serviceable part.
•Perform periodic safety testing to insure proper patient safety. This should include
leakage current measurement and insulation testing. The recommended testing
interval is once per year.
•If the hospital or healthcare institutions using this device fail to implement a
satisfactory maintenance schedule, it will result in device failure and may endanger
the patient’s safety.
•Use the pulse oximeter under the conditions specified in this operation manual.
BT-710 Operation manual
6
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
•Thoroughly read and understand the manual prior to use of the BT-710. Failure
regulations or hospital waste disposal regulations. Keep the away from
to prevent explosion or
cables of accessories carefully to avoid patient
•The physiological wave, physiological parameters and alarm information
Beyond the conditions, the pulse oximeter may not function properly and the
measurement results may not accurate and may result in device failure or
endangering the patient’s safety.
• During the operation, do not disconnect any cable.
• Do not operate the BT-710 pulse oximeter if it fails to pass the power on self-test
procedure.
•The BT-710 pulse oximeter is intended to be used by clinical professionals or trained
doctors, nurses or laboratory assistant.
•Do not service and maintain or clean the device including accessories while in use
with a patient.
•Using the device to one patient at a time.
to do so could result in personal injury or equipment damage.
•The device is intended for measure the clinical blood oxygen saturation via
pulse oximeter, and only trained and qualified doctors and nurses should use
the device.
•The alarm volume, upper and lower alarm limits should be set according to the
actual situation of the using environment. Do not just rely on audio alarm
system while monitoring the patient, because too low alarm volume or muted
alarm may result in notice failure of alarm situation and endanger the patient’s
safety. Please pay close attention to the actual clinical status of the patient.
• Use only the power adapter supplied with pulse oximeter.
• Do not open the enclosure to avoid an electric shock. Any repair and upgrade
of pulse oximeter should be done by service personnel trained and authorized
by Bistos. Co., Ltd.
•When handling packaging materials and the device, abide by local laws and
children.
•Do not use in the presence of flammable anesthetics
fire.
•Install the power lines and
entanglement or suffocation, cables tangled or electrical interference.
•When the pulse oximeter is used together with electrosurgical devices, the user
(a doctor or a nurse) should ensure the safety of the patient and instrument.
BT-710 Operation manual
7
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
displayed on the pulse oximeter are only for the doctor’s reference and should
Medical electrical equipment needs special precautions regarding EMC and
The equipment shall not be used adjacent to other devices unless verification
Keep matches, and all other sources of ignition, out of the room in which the
is located. Textiles, oils, and other combustibles are easily
n this
manual and found to comply with all relevant safety and performance
person or organization who makes an unauthorized modification, or
yourself to avoid damage to the
equipment and unexpected electrical shock. Only qualified Bistos service
engineer must repair or replace components.
CAUTION
Avoid using pulse oximeter in the environment such as pressure is too high,
not be directly used as the basis for clinical treatment.
• This is not a therapeutic device.
• Use of accessories other than approved for use with this product may result in
increased emissions or decreased immunity.
•
needs to be installed and put into service according to the EMC information
provided in this manual. In addition, portable and mobile RF communications
equipment can affect medical electrical equipment.
•
of normal operation in the configuration in which it is to be used can be
achieved.
•
pulse oximeter
ignited and burn with great intensity in oxygen rich environment. Personal
injury or equipment damage could occur.
•A fire and explosion hazard exists when performing cleaning or maintenance
procedures in an oxygen rich environment.
•The pulse oximeter has been validated with the accessories listed i
requirements applicable to the device. It is therefore the responsibility of the
incorporates an unapproved attachment to the device.
•An operator may only perform maintenance procedures specifically described
in this manual.
• Do not remove the covers of a BT-710
•Please install or carry the pulse oximeter properly to prevent damage due to
falling, collision, strong vibration or other mechanical force.
•Avoid instrument splashed by water.
range of 5 ℃ ~ 40 ℃。
• Avoid high temperatures, the instrument should be used within a temperature
•
BT-710 Operation manual
8
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
poor ventilation, dusty, or contain salt, sulfur gas and chemical.
Before powering on the device, make sure that the power used by the device
oltage and frequency requirements on the
Equipment should be tested at least once a year, the test should be done and
testing knowledge and experienced
When the instrument and accessories are about to exceed the useful life
Do not connect any equipment or accessories that are not approved by the
operation and safety are not
Parts and accessories used must meet the requirements of the applicable
safety standards, and/or the system configuration must meet the
Ensure that the conductive parts of electrodes and associated connectors,
including neutral electrodes, do not come in contact with earth or any other
WARNING
Unplug the pulse oximeter from its power source prior to cleaning or maintenance
ome chemical cleaning agents may be conductive and leave a residue that may
permit a build-up of conductive dust or dirt. Do not allow cleaning agents to
•Before using the pulse oximeter, check the pulse oximeter and accessories if
there is damage that may affect patient safety. If there is obvious damage or
aging, replace the parts before use. The replacement should be made with
same parts of original parts.
•
complies with the supply v
equipment label or in this manual.
•
recorded by trained, have safety
personnel. If there are any problems in the tests, they must be repaired.
•
(expected service life: 5 years), it must be treated in accordance with relevant
local laws and regulations or the hospital's rules and regulations.
•Do not connect to other equipment or network which not specified in the
instruction for use, in risk of external high voltage.
•
manufacturer or according to IEC 60601-1 to the pulse oximeter. The operation
or use of non-approved equipment or accessories with the pulse oximeter is
not tested or supported, and pulse oximeter
guaranteed in such a case.
•Any non-medical equipment (such as the external printer) is not allowed to be
used within the patient vicinity (1.5m/6ft.).
•
requirements of the medical electrical systems standard.
•
conducting objects.
0.2 Shock hazards
to prevent personal injury or equipment damage.
S
BT-710 Operation manual
9
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
contact electrical components and do not spray cleaning solutions onto any of
WARNING
Improper operation may cause the internal lithium ion battery to be hot, ignited or
nly the qualified service personnel
Do not destroy the battery. Do not pierce battery with a sharp object such as a
disassemble or modify the battery. The battery can heat, smoke, deformation or
Properly dispose of or recycle the depleted battery according to local regulations.
these surfaces. Personal injury or equipment damage could occur.
Do not expose the pulse oximeter to excessive moisture that would allow for liquid
pooling. Personal injury or equipment damage could occur.
Do not touch the patient and signal input/output parts simultaneously
Due to the risk of electrical shock hazard, only qualified personnel with appropriate
service documentation should service the pulse oximeter.
0.3 Battery warnings
exploded, and it may lead to the decrease of the battery capacity. It is necessary to
read the operation manual carefully and pay more attention to warning message.
Do not open the battery compartment. O
authorized by the manufacturer can open the battery compartment and replace
the battery, and batteries of same model and specification should be replaced.
Be careful when connecting the battery with polarity.
Do not use the battery near fire or environmental temperature exceeds 60 ℃. Do
not heat or splash the battery or throw it into fire or water.
needle. Do not hit with a hammer, step on or throw or drop the battery. Do not
burning.
When leakage or foul smell is found, stop using the battery immediately. If your
skin or cloth comes into contact with leaked liquid, cleanse it with clean water at
once. If the leaked liquid splashes into your eyes, do not wipe them. Irrigate them
with clean water first and go to see a doctor immediately.
BT-710 Operation manual
10
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Avoid placing in an area
Power off when the pulse
0.4 General precautions on environment
Do not keep or operate the pulse oximeter under the environment listed below.
Avoid placing in an area
exposed to moisture. Do
not touch the pulse
oximeter with wet hand.
where high variation of
temperature exists.
Operating temperature
ranges from 5
Operating humidity
℃ ~ 40℃.
ranges from 30 % ~ 85 %.
Avoid exposure to direct
sunlight
Avoid in the vicinity of
electric heater.
Avoid placing in an area
where there is an
excessive humidity rise
or ventilation problem.
Avoid placing in an area
where chemicals are
stored or where there is
in danger of gas leakage.
Do not disjoint or
disassemble the pulse
oximeter. Bistos Co., Ltd.
does not have liability of
it.
Avoid placing in an area
where there is an
excessive shock or
vibration.
Avoid dust and especially
metal material enters
into the pulse oximeter.
oximeter is not fully
ready to operate.
Otherwise, the pulse
oximeter could be
damaged.
BT-710 Operation manual
11
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
1 System basics
1.1 Intended use
The BT-710 pulse oximeters acquire the blood oxygen saturation (SpO2) and pulse rate (PR).
The signals are converted into digital data and processed, examines the data for alarm
conditions and display the data. The pulse oximeter also provides operating control for the
user. The pulse oximeter intend to use in hospital clinical area such as general ward, to provide
additional information to the medical and nursing staff about the blood oxygen saturation of
the patient. The BT-710 pulse oximeters are intended to be used only under regular
supervision of clinical personnel. It is suitable for adult and pediatric, neonate. The intended
locations of use are hospitals and clinics.
1) Intended patient population
- Adult (>18 years adults) and Pediatrics (30 days < and <18 years) and Neonate (0
days< and <30days)
2) Intended user profile
- Doctor, physicians or nursing staff who is qualified personnel
- Basic experiences or knowledge on medical field, especially on patient monitoring
- Trained or requested to read IFU before use
3) Environment of use
- Hospital and clinic
- Requirements: Stable power source
4) Scope of application
This pulse oximeter is suitable for bedside monitoring of patient. This pulse oximeter enables
the monitoring of blood oxygen saturation (SpO
) and pulse rate (PR). It is equipped with a
2
replaceable built-in battery to provide convenience for the patient movement in hospital.
5) Indications and contraindications
Blood oxygen saturation (SpO
)
2
Indication:
- Monitoring effectives of oxygen therapy
- A reading is needed to facilitate the completion of an early warning score to inform
clinical assessment
- Sedation or anesthesia
- Transport of patients who are unwell and require oxygenation assessment
- Haemodynamic instability (e.g. cardiac failure or Myocardial Infarction)
- Respiratory illness e.g. asthma, chronic obstructive pulmonary disease
BT-710 Operation manual
12
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
- Monitoring during administration of respiratory depressant drugs, e.g. opiate
epidural or patient-controlled analgesia.
- Assessing oxygen saturation during physical activity e.g. in pulmonary rehabilitation
Contraindications
- Pulse oximetry does not give an indication of haemoglobin so if the patient is
profoundly anaemic then their oxygen saturation may by normal but they may still be
hypoxic
Source: NHS. “Clinical Procedure_ Procedure for Pulse Oximetry/SPO2”. Wirral Community NHS Trust. Sep, 2013
1.2 Operating principle
Refer to the chapters for every physiological parameter from chapter 5.
1.3 System configurations
Basic configuration of BT-710
• Main body with 4.3” touch screen and built-in lithium-ion battery
• Adult SpO
• AC/DC adapter
sensor probe
2
1.4 Product outlook
Figure1-1: Front view
BT-710 Operation manual
13
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Figure1-2 : Rear view
Figure1-3 : Top view
Figure1-4 : Bottom view
BT-710 Operation manual
14
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Turned on when the pulse oximeter is being powered by
- Power On: Press down the key more than 2 seconds.
s
The
system will shut down 3 seconds”.
[Alarm reset]
[Setting]
Enter to the setting mode. Press again to close the setting
1.5 Description of pulse oximeter
1
2
3
Name Description
DC power indicator
Display area Display the waveform and measured value
[Power]
4
5
Figure1-5: Front view
the adapter.
- Power Off: Press down the keys more than 2 second
and the system will display the alarm message “
To reset the alarm condition.
mode.
BT-710 Operation manual
15
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Figure1-6: Top view
6
Name Description
SpO2 SpO2 sensor probe interface
Figure1-7: Bottom view
7
8
9
Name Description
SD card interface For software upgrade
Power adapter 5V, 2A adapter
Lanyard eyelet
For convenient hand held
BT-710 Operation manual
16
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Current alarm message. When an alarm occurs, this area will displayed yellow
or red depending on the alarm type.
SpO2 waveform. Display the measured SpO2 waveform. The waveform is not
Sleep mode. Touch this area makes the pulse oximeter to enter the sleep
1.6 Understanding the display
Figure 1-8: Standard display
Description
1
2 SpO2 value. Display the measured SpO2 value.
3 SpO2 upper alarm limit. Display the user set upper alarm limit
4 SpO2 lower alarm limit. Display the user set lower alarm limit
5
normalized.
6 Pulse rate value. Display the measured pulse rate per minute.
7
mode. To exit "sleep mode", press [Power] or [Alarm reset] or [Setting]
8 Perfusion Index. Display the measured perfusion index.
9 Pulse rhythm strength.
10 Battery Status
11 Patient type.
12 The unit of SpO2.
BT-710 Operation manual
17
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
1.7 Essential performance
This device Pulse oximeter provides patient vital signs such as pulse rate, blood oxygen
saturation and perfusion index by placing the sensors to the appropriate site of patient. The
device is composed with display, control circuit and panel, and input part for SpO
detects SpO2 and PR using specific sensors. The detected analog signal amplified and
converted to digital. This converted data feed to the CPU and converted to the display format
as number and waveform. This device is incorporated with alarm system. The alarm generated
when the detected signal range is beyond the user set alarm limits.
sensors. It
2
BT-710 Operation manual
18
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
All analog and digital devices connected to the pulse oximeter must be certified by
ions shall comply with
to the input / output signal ports are responsible for the compliance with IEC
extension cord.
2 Preparing for operations
2.1 Installation
To ensure normal working of the pulse oximeter, read this chapter before use, and install as
required.
IEC standards (e.g. IEC 60950 Data processing equipment standard and IEC 60601-1
Medical equipment standard). Furthermore, all configurat
valid version of IEC 60601-1 standard. The personnel connecting additional devices
60601-1 standard. If there is any question, please contact Bistos.
The copyright of pulse oximeter software belongs to Bistos. Without permission,
any organization or individual shall not interpolate, copy or exchange by any means
or form.
When the pulse oximeter is combined with other devices, it must comply with IEC
60601-1:2005 + A1:2012, and should not be connected with multi-socket outlet or
Prior to installation, the operator must ensure that the following space, power, environmental
requirements are met.
2.1.1 Unpack and check
BT-710 pulse oximeter was inspected rigorously at the factory before delivery, in order to avoid
being hit when transported, carried out careful packaging. Before unpacking, carefully inspect
the package. If any damage, please immediately contact the Bistos. Unpack in the correct way,
carefully remove the pulse oximeter and accessories from the box and check with the packing
list. Check if there is any mechanical damage, and the all listed components are completely
packed. If you have questions, please contact the marketing department of Bistos or agency.
Please keep the packing box and materials for use in future transporting or storage.
2.1.2 Power requirements
- DC power supply adapter (Model: UE10WCP1-050200SPA)
Prevent severe shock, vibration, rain and snow splashing during transport.
The packaged pulse oximeter should be stored in well-ventilated room
corrosive gases.
WARNING
Ensure that the pulse oximeter is used under specified environment. Fail to do this,
WARNING
Do not try to open the pulse oximeter when the power is connecting.
NOTE
The operation of pulse oximeter after the supply mains has been interrupted
2.1.3 Environmental requirements
The storage, transport and use of the pulse oximeter must meet the following environmental
requirements.
Operating
Ambient temperature
5℃ ~ 40 ℃
environment
Storage
with ambient temperature -20 ℃ ~ 60℃, relative humidity 0 ~ 95 % (Noncondensing), atmospheric pressure 700 ~ 1060 mbar(hPa), and without
The operating environment of the pulse oximeter should avoid noise, vibration, dust, corrosive
or flammable and explosive materials. In order to allow air flowing smoothly and achieve good
heat dissipation, at least 2 inches (5cm) clearance should be kept around the device.
When the device is moved from one environment to another, the device may have
condensation due to the differences in temperature or humidity. In this case, wait until the
condensation disappears before using the device.
the technical specifications declared in this manual may not be met and it may
result in damage to equipment and other unforeseen consequences.
2.2 Connecting to power
Connect to power adapter in the following steps:
- Make sure that the AC power supply meets the following specifications: A.C. 100V-
240V, 50/60Hz.
- Use the power adapter provided with the pulse oximeter. Plug the power adapter
into the power connector of the pulse oximeter, and plug the other end of the power
adapter into the mains (low voltage power supply network facilities) power outlet.
and is restored after a period of time that is longer than 30s.
BT-710 Operation manual
20
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
If the pulse oximeter is damaged, or fails to work normally, do not use it. Please
contact the maintenance personnel or Bistos immediately.
3 Basic operations
3.1 Turn on
3.1.1 Check the pulse oximeter
- Before turn on the pulse oximeter, check whether there is mechanical damage to the
pulse oximeter, and whether the external cables and accessories are connected correctly.
- Plug the power adapter into the AC power outlet. If using battery power, make sure the
battery is fully charged.
- Check all the functions required to make sure that the pulse oximeter operates properly.
3.1.2 Start the pulse oximeter
If finish to check the pulse oximeter, it is ready to start the pulse oximeter.
Press the [Power] key and the system enters the main interface within seconds.
- Once the power is supplied, the system performs a power-on self test to check the
functions before start-up. If any fatal error occurs during boot up, the system will
alarm. If this case persists, please stop to using the pulse oximeter and contact the
maintenance personnel or Bistos.
- Check all available pulse oximeter functions to ensure that the pulse oximeter
operate properly.
- If the pulse oximeter equipped with a battery, charge the battery after each use to
ensure sufficient power.
- After unpacking and use the pulse oximeter at first time, the pulse oximeter should
be powered with adapter.
3.1.3 Connect the sensors
Connect the SpO2 sensor probe to the pulse oximeter and the monitoring site of patient.
BT-710 Operation manual
21
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
CAUTION
it is not recommended.
3.1.4 Start monitoring
Start monitoring in the following steps:
- Check if the patient cable is connected properly.
- Check if the settings of the pulse oximeter are corrects, such as patient type.
- For the details of parameter measurement or monitoring, see the appropriate
section.
- The operator can operate according to their own habits, standing in front, left or
right of the monitor, easy to observe and operate the monitor.
3.2 Turn off
Turn off the pulse oximeter in the following steps
- Disconnect the sensor probe connected to the patient.
- Press and hold the [Power] key for 2 seconds to pop up the 3 seconds countdown
window, and the pulse oximeter turns off in 3 seconds.
• If the pulse oximeter is not turned off properly, you can simply disconnect the
power to shutdown forcibly. But the forced shutdown may cause data loss, and
BT-710 Operation manual
22
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
4 Setup the pulse oximeter
In the main screen, Press the key enter the setup menu.
Select “Patient Type”, you can choose “Adult” or “Pediatric” or “Neonate”. According to
patient type, the settable PR alarm limit and default are changed.
Select “Alarm Delay”, you can choose “off”, “1s”, “2s”, “3s”, “4s”, “5s”, “6s”, “7s” or “8s”.
It sets the time from the occurrence of a triggering event either in the patient, for
physiological alarm conditions to when the alarm system determines that an alarm
condition exists. Technical alarm conditions are not delayed.
Select “Backlight”, and enter value(Range: 0-5). It can set brightness of the background
according to needs of use. “5” is the brightest.
Select “Alarm Volume”, which is alarm tone volume and settable among(Range: 0-9).
“0” is alarm sound off.
Select “Alarm Reminder Signal”, you can choose “On”, or “Off”. It is periodic signal that
reminds the operator that the alarm system is in an alarm signal inactivation for both
physiological alarm conditions and technical alarm conditions.
Select “Alarm Reminder Interval”, when the “Alarm Reminder Signal” sets “On”, you can
choose alarm interval to “1min”, “2min”or “3min”.
Select key to display the following interface.
BT-710 Operation manual
23
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Select “Year ”, “Month”, “Day”, “Hour”, “Minute” and “Second”, and set the current date.
Select “Date Format”, and set among the below date formats in accordance with custom.
“YYYY-MM-DD”: Ye a r- Month-Day.
“MM-DD-YYYY”: Month -Day-Ye a r.
“DD-MM-YYYY”: Day-Month-Ye ar.
Select key to display the following interface.
BT-710 Operation manual
24
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Select “Displays”, set “SpO2”.
Select “Language”, and select the option as needed:
“English”: The interface language of the pulse oximeter is English.
“Türkçe”: The interface language of the pulse oximeter is Turkish.
“Español”: The interface language of the pulse oximeter is Spanish.
“Français”: The interface language of the pulse oximeter is French.
“Polski”: The interface language of the pulse oximeter is Polish.
“Italiano”: The interface language of the monitor is Italian.
“Deutsch”: The interface language of the monitor is German.
Select “Default”, clears all the setting except the date and time and go back to the initial
factory setup.
Select key to display the following interface.
BT-710 Operation manual
25
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
NOTE
Information about wavelength range can be especially useful to clinicians.
WARNING
Please use SpO2 sensor supplied from Bistos, operate in accordance with the
error message, and trigger an audible and visual alarm
If there is carboxyhemoglobin, methemoglobin or dye diluted chemical, the SpO2
5 SpO
2
5.1 Overview
Blood oxygen saturation (SpO2) is the percentage of oxyhemoglobin (HbO2) capacity bound by
oxygen in the blood in the total hemoglobin (Hb) capacity that can be combined, that is, the
concentration of oxygen in the blood.
The principle for monitoring the pulse SpO
use the finger as a transparent container for hemoglobin, use 660nm wavelength red light and
905nm near-infrared light as the incident light, maximum output power is 300 mW, measure
the light transmission intensity through the tissue bed, and calculate the concentration of
hemoglobin and SpO
.
2
The passing lights depend on a variety of factors, most of which are constant. However, one of
these factors, the arterial blood flow, changes with time, as it is pulsating. By measuring the
light absorbed during pulsating, it is possible to obtain the arterial blood SpO
pulsation can give a “plethysmography” wave and pulse rate signal.
The main screen displays “SpO
” value and “plethysmography” wave. While the monitor gets
2
and displays the physiological signal from a patient, the pulse tone which is the auditory
information signal. The volume of this pitch tone is user adjustable.
is to fix the probe fingerstall on the patient’s finger,
2
. Detection
2
This pulse oximeter applies to measure SpO
days), neonate (<30 days). Contact SpO
of adults (>18 years), pediatric (<18 years,>30
2
probe to Patient’s finger to get “SpO2” value and
2
“plethysmography” wave.
SpO
function of this pulse oximeter has been calibrated in factory.
2
5.2 Safety information
Manual, and observe all warnings and precautions.
Before monitoring, check whether the sensor probe is normal. When SpO
sensor probe is unplugged from the socket, the screen will display “SpO2 No
Sensor connected”
simultaneously.
2
If the sensor or sensor packaging has signs of damage, do not use this SpO
sensor; return it to the manufacturer.
2
BT-710 Operation manual
26
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
value will have deviation.
four hours. Some patients may require more frequent inspection, such as
patients with perfusion disorders or sensitive skin, because persistent and
ch as
Functional testers cannot be used to evaluate the accuracy of pulse oximetry.
NOTE
Do not put the oxygen probe and blood pressure cuff on the same limb, because
SpO2 equipment.
NOTE
When you turn on the pulse oximeter, plug in SpO2 probe and connect patient’s
value and pulse rate.
“SpO2 Search Timeout” displayed in the technical alarm area until the pulse
When the patient has a tendency to hypoxia, use the oximeter to analyze blood
samples in order to fully grasp the patient's condition.
Do not put the sensor on limbs with arterial duct or intravenous tube.
Do not intertwine electrosurgical equipment cable with the patient cable.
Avoid using the pulse oximeter and sensors while using the NMR equipment, in
order to avoid severe burns to the patient as a result of induced currents.
During long time continuous monitoring of a patient, check the position of SpO
2
sensor once every 2 hours, and move properly when the skin changes or every
prolonged monitoring may increase unpredictable skin changes, su
allergies, redness, blistering or pressure necrosis.
Before using, verify compatibility between the monitor and probe, otherwise it
may cause injury to the patient.
blood flow occlusion during blood pressure measurement will affect the SpO2
readings.
The pulse oximeter cannot be used to verify the accuracy of SpO
probe and
2
5.3 Monitoring steps
1. Select the appropriate SpO
2. Turn on the pulse oximeter, and connect the SpO
3. Clean the measurement site, such as finger with nail polish.
4. Put the SpO
sensor probe on the patient’s finger.
2
5. Select the appropriate alarm settings.
6. Start monitoring.
finger (or toe), monitor displays SpO2 wave, “SpO2 Pulse Search” displayed in the
technical alarm area until the monitor measured SpO
sensor according to the patient.
2
sensor probe to the pulse oximeter.
2
2
BT-710 Operation manual
27
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
oximeter measured pulse rate. Check the sensor mounting position, whether the
are meaning that the value of SpO2 or PR might be inaccurate.
sensor is damaged or sensor type. Reconnect the sensor or use new sensor.
The alarm message for signal inadequacy such as “SpO2 No Sensor connected”,
“SpO2 Sensor Off”, “SpO2 Search Timeout”, “SpO2 Signal Unstable”, “SpO2
Sensor Error” and low priority alarm and indication for signal instability like “---”
5.4 Setting SpO2
Select the main screen to enter the SpO2 set interface
Select “Wave Speed” and set wave speed to “12.5mm/s” or “25mm/s”. The faster speed,
the smoother wave.
Select “Wave Mode”, and set the wave drawing mode to “Scan” or “Fill”. This displays a
wave as line graph or filled line graph.
Select “Average Time”, and set the average time to “2-4s”, “4-8s”, “8-16s”. According to
the selected option, the module displays the average of data collected within that time.
The shorter the averaging time is, the quicker the monitor responds to changes in the
patient’s oxygen saturation level.
Select “Pulse Volume”, and set the volume for auditory information signal for pulse rate
( Range: 0-9 ).
Select “Sensitivity”, and you can choose “1”, “2”, “3”. It is responding speed for changes of
parameter setup. It is faster when setting value is higher.
BT-710 Operation manual
28
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
NOTE
Setting values including alarm settings are stored and does not changed even
when the power supply interrupted.
Select key to display the following interface.
Select “SpO2 Low Limit”, and set the SpO2 alarm low limit value (Range: 0-99),
Adult/Pediatric/Neonate default: 90. The input cannot enter higher than the SpO2 alarm
upper limit.
Select “SpO2 High Limit”, and set the SpO2 alarm upper limit value (Range: 1-100),
Adult/Pediatric/Neonate default: 100/100/95. The input cannot enter less than the SpO2
alarm low limit.
Select “PR Low Limit”, and set the RP alarm low limit value (Adult Range: 15-299,
Pediatric/Neonate Range: 15-349), Adult/Pediatric/Neonate default: 50/75/100. The input
cannot enter higher than the PR alarm upper limit.
Select “PR High Limit”, and set the PR alarm upper limit value (Adult Range: 16-300,
Pediatric/Neonate Range: 16-350), Adult/Pediatric/Neonate default: 120/160/200. The
input cannot enter less than the PR alarm low limit.
Select “PI Low Limit”, and set the PI alarm low limit value (Range: 0.00-19.90). Default:
0.00.
Select “PI High Limit”, and set the PI alarm upper limit value (Range: 0.10-20.00). Default:
20.00.
Select “Alarm Level”, you can choose “Mid” or “High” of priority level for the physiological
alarm.
Alarm, Select “Default”, the alarm parameter is set to the default value.
BT-710 Operation manual
29
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
5.5 Measuring influencing factors
During operation, the following factors can affect the accuracy of SpO2 measurement:
High-frequency radio wave interference, such as interference generated by the host
system or interference from electrosurgery instrument connected to the system.
Intravenous dye.
Too frequent movement of the patient.
External light radiation.
Sensor is improperly installed or improperly in contact with the patient.
Sensor temperature.
The sensor is placed on limbs with blood pressure cuff, arterial duct or lumen tube.
Concentration of non-functional hemoglobin such as carboxyhemoglobin (COHb) and
methemoglobin (MetHb).
Shock, anemia, hypothermia, and the application of vasoconstrictor drugs may reduce the
arterial blood flow to a level that cannot be measured.
The measurement also depends on the absorption of specific wavelengths of light by
oxyhemoglobin and reduced hemoglobin. If there is any other substance that absorbs the
same wavelength, the measurement may have false or low SpO
values, such as: carbon
2
hemoglobin, methemoglobin, methylene blue, and indigo carmine.
SpO
Operating environment limit: Operating temperature range: 5~40℃, Humidity range:
The SpO2 sensor probe (specified in Chapter 11) material which contacts patients or other
staff has passed the biocompatibility test and meet the requirements of ISO 10993-1.
Fluke's index 2XL Oxygen Analyzer can be used to check the function of the pulse oximeter
and can be used to assess the accuracy of the pulse rate but cannot be used to assess the
accuracy of blood oxygen.
Functional testers cannot be used to evaluate the accuracy of pulse oximetry probe and
pulse oximetry.
Measure the maximum temperature between the oxygen probe and the tissue contact
surface:Measured as described in Annex BB of ISO 80601-2-61,the temperature is less
than 41 ° C.
BT-710 Operation manual
30
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
The PR accuracy shall be stated as the root-mean-square (rms) difference between paired
pulse rate data recorded pulse oximeter equipment and with a reference method (E.G, an
electronic pulse simulator, ECG heart rate, palpated pulse, thoracic auscultator or a
second pulse equipment which has been qualified by comparison to one of these
references).
BT-710 Operation manual
31
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
6 Review
Use the trend screen to recall all the historical patient data in a list, including monitoring time
(in 1-minute intervals), and SpO
top of the list.
and PR values. The most recent measurements display at the
2
Figure 6-1: SpO
Trend (1)
2
To view historical data:
Press the [Setting] key twice to display the Trend screen.
Select or key to page up and down to view patient data.
Press key to exit the Trend screen and display the Main screen or press the
[Setting] key to see the trend graph.
BT-710 Operation manual
32
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Figure 6-2: SpO
Select ‘SpO2’, ‘PR’ or ‘PI’ to see.
Press key to exit the Trend screen and display the Main screen.
Trend (2)
2
BT-710 Operation manual
33
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
In any single area (e.g. intensive care unit), it can be a potential danger if there
are the same or similar devices using different alarm preset.
7 Alarm
Alarm means that the pulse oximeter prompts the medical staff through sound and light when
the abnormal changes in vital signs are monitored or the pulse oximeter has a failure or is
unable to monitor the patient successfully. For better viewing those alarms, it is recommended
to look straight at the display when using the monitor.
After setting, the alarm and other parameters of the monitor won’t be lost when the system is
power off, unless modified manually. Connect the power again and turn on the monitor, it will
resume normal working, and the alarm and other parameters remain unchanged.
7.1 Alarm types
According to the nature of the alarm, the alarms of the monitor can be divided into
physiological alarms, technical alarms, alarm reminder signal and prompt messages.
Physiological alarms
A physiological alarm is usually triggered when a physiological parameter of the patient
exceeds the alarm limit or the patient has physiological abnormalities. The information of
physiological alarm is displayed in the physiological alarm area on top of the screen.
Technical alarms
Technical alarm is also known as a system error message, which is caused by improper
operation or system failure resulting in system malfunction or monitoring result distorted.
The information of technical alarm is displayed in the technical alarm area on top of the
screen.
Alarm Reminder Signal
Alarm Reminder Signal is periodic signal that reminds the operator that the alarm system
is in an alarm signal inactivation for both physiological alarm conditions and technical
alarm conditions.
Prompt messages
Strictly speaking, the prompt messages are not alarms. The monitor also will display some
information associated with system status in addition to the physiological alarms and
technical alarms, and generally such information do not involve the patient's vital signs.
The prompt messages generally appear in the technical alarm area and parameters area.
7.2 Alarm condition priorities
According to the severity of the alarm conditions, the physiological alarms of the monitor can
be divided into high priority, medium priority.
BT-710 Operation manual
34
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
High priority alarms
The patient is in critical condition that is life-threatening, and should be immediately
rescued, or the monitor has a serious mechanical failure or malfunction, causing it unable
to detect the patient's critical state and endangering the patient’s life.
Medium priority alarms
The patient's physiological signs are abnormal and appropriate measures or treatment
should be taken immediately, or although it won’t endanger the patient’s life, the
mechanical failure or deactivation of the monitor will affect the normal monitoring of key
physiological parameters.
Low priority alarms
The certain monitoring function is invalid due to mechanical failure or deactivation, but it
won’t endanger the patient’s life.
The levels of some physiological alarms can be modified. But, the priority of all technical alarms
and some physiological alarms have been set in the monitor at the factory and cannot be
modified by the user.
7.3 Alarm mode
When an alarm occurs, the monitor uses the following audible or visual alarm to prompt the
user:
Visual alarm
Audible alarm
Alarm info
Parameter
Of which, the visual alarm, audible alarm, and alarm information distinguish the alarm levels in
a different manner respectively.
7.3.1 Visual alarm
When an alarm occurs, the alarm indicator will flash in different colors and frequencies to
prompt the alarm priority.
High priority alarm: Red
flashing
Medium and Low priority alarm: Yellow
7.3.2 Audible alarm
An audible alarm is that the monitor prompts the alarm priorities with alarm tone
characteristics when an alarm occurs.
BT-710 Operation manual
35
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Audible alarm pattern:
High priority alarm: Beep-beep-beep--beep-beep----beep-beep-beep--beep-beep
Medium priority alarm: Beep-beep-beep
Low priority alarm: Beep-
7.3.3 Alarm information
Alarm information displayed on the physiological or technical alarm area of the monitor
indicates the corresponding alarm information when an alarm occurs. The system will
distinguish the alarm priority with different background colors:
High priority alarm: Red
Medium priority alarm: Yellow
Low priority alarm: Yellow
The following flags in front of physiological alarms are used to distinguish the alarm priorities.
High priority alarm: ***
Medium priority alarm: **
Low priority alarm: *
7.3.4 Parameter flashing
When the physiological parameter values in the parameter area will flash once per second, and
the indicator for the upper limit and lower limit of the parameter area will also flash at the
same frequency, it indicating that the parameter exceeds the upper limit or lower limit.
7.4 Alarm state
7.4.1 Alarm reset
Select button, and you can temporarily turn off the alarm sound of currently occurring
physiological alarms of the pulse oximeter, but the alarm information is still retained. For
technical alarms, clear the alarm state, display alarm prompt information, the alarm state icon
area displays the icon. When a new physiological alarm or technical alarm occurs, the
alarm reset is automatically canceled.
BT-710 Operation manual
36
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
When the alarm is off, and the alarm reminder signal is on, the system will have
alarm reminder tone.
Source
SpO2
Default
Causes and countermeasures
Value
Neonate)
SpO2 value is higher than the
are appropriate for the patient.
PR value is higher than the settled
are appropriate for the patient.
PI value is higher than the settled
7.4.2 Alarm sound off
When “Alarm Volume” sets “0”, the alarm state area on the screen shows the icon.
If “Alarm Volume” sets bigger than “0”, the system will cancel alarm sound off state.
7.5 Alarm information
This chapter lists some important physiological and technical alarm information, and some
alarms are not necessarily listed.
Note that in this chapter: P column indicates the default alarm priority: H indicates high
priority, M indicates middle priority, L indicates low priority, and “*” indicates configurable
priority by the user.
Corresponding countermeasures are listed for each alarm message. If you operate in
accordance with the countermeasures but the problem persists, contact your service
personnel.
Physiological alarm information.
Parameter Alarm message
SpO2 High
Limit
SpO
Too High
2
(Adult/
Pediatric/
100/100/
95
P
settled upper alarm limit or lower
than the settled lower alarm limit.
Check the patient's physiological
SpO2 Low
Limit
PR High
Limit
SpO
Too Low 90
2
PR Too High
120/160/
200
M*
condition, and check if the patient
category and alarm limit settings
upper alarm limit or lower than
the settled lower alarm limit.
Check the patient's physiological
PR Low
Limit
PR Too Low
50/75/
100
condition, and check if the patient
category and alarm limit settings
PI High
Limit
PI Too High
0.00
upper alarm limit or lower than
the settled lower alarm limit.
Check the patient's physiological
BT-710 Operation manual
37
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
condition, and when the PI alerts
monitoring site.
The physiological signal shows a
patient's physiological condition.
Source
Alarm message
P
Causes and countermeasures
SpO2 No Sensor
Indicates the condition that SpO2
socket.
SpO2 Search Timeout
When you turn on the monitor, plug
the sensor or use new sensor.
SpO2 Sensor Off
SpO2 sensor falls off from the patient
the sensor or use new sensor.
Connect to DC power supply, and
charged.
Press down the keys more than 2
shut down 3 seconds”
PI Low
limit
PI Too Low 20.00
- No Pulse - H
Technical alarm information.
connectedL
SpO2
the clinician to consider another
potentially life-threatening drop
in oxygen saturation. Check the
sensor cable is unplugged from the
in SpO2 probe and connect patient’s
finger (or toe), monitor displays SpO
wave, “SpO2 Pulse Search” displayed
in the technical alarm area until the
monitor measured SpO
pulse rate. “SpO
L
2
2
Search Timeout”
displayed in the technical alarm area
until the monitor measured pulse
rate. Check the sensor mounting
position, whether the sensor is
damaged or sensor type. Reconnect
2
value and
Battery Low M
Battery
System will shutdown N/A
or monitor, malfunctions, or sensor
other than specified in this Manual is
used. Check the sensor mounting
L
position, whether the sensor is
damaged or sensor type. Reconnect
charge the battery, and power with
the battery as needed after fully
seconds and the system will display
the alarm message “The system will
BT-710 Operation manual
38
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
8 Battery
8.1 Overview
The pulse oximeter has a built-in rechargeable battery to ensure that the pulse oximeter can
also be used normally in case of patient transfer or power failure. When the oximeter is
connected to a DC power source, it will charge the battery no matter whether the pulse
oximeter is turned on or not. In the case of power failure, the system will automatically use the
battery to power the pulse oximeter to avoid interrupting the pulse oximeter working.
The battery icon on the screen indicates the battery status:
Battery is working properly and is fully charged.
Battery is working properly and the green part indicates the battery power.
Battery power is low, and requires charging immediately, or else the pulse oximeter will
turn off automatically.
Battery is not installed.
Battery is properly installed and being charged.
The battery power can only maintain for some time. Low battery voltage will trigger a high
level technical alarm “Battery Low”; in this case, connect the pulse oximeter to DC power and
charge the battery.
8.2 Battery usage guide
Battery life depends on the frequency and time of use. If the battery maintenance and storage
are proper, the lithium battery life is three years. If you do not use the battery properly, its life
may be shortened. It is recommended to replace the lithium battery once every three years.
In order to ensure the maximum capacity of the battery, please note the following usage guide:
Before using the battery, please read this manual and labels on the battery surface
carefully.
Do not drop the battery.
BT-710 Operation manual
39
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
Keep the battery out of the reach of children.
artment. Only the qualified service personnel
WARNING
Do not open the battery compartment. Only the qualified service personnel
If it won’t be used for a long time (over three months), please store the battery properly.
Charge the battery to 50%, and wrap the battery with non-conductive material in order to
avoid direct contact with metal, resulting in damage. Keep the battery in a cool dry place.
Check the battery performance once every two years. Before servicing the pulse oximeter
or you suspect that the battery is the fault, also check the battery performance.
•
• Use only the designated battery.
• If the battery is damaged or leaks, replace it immediately. Do not use a defective
battery for the monitor.
• Do not open the battery comp
authorized by the manufacturer can open the battery compartment and replace
or check the battery, and when it needs replacement, the replacement should be
same model with established battery.
Do not disassemble the battery, throw it in fire, or short-circuit it. Battery fire,
explosion and leakage may lead to personal injury. Do not touch the leaking
battery with bare hand directly.
8.3 Checking battery performance
Please refer to the following steps to check the battery performance:
Disconnect the pulse oximeter from the patient and stop the monitoring or measurement.
Connect DC power to the monitor, and charge the battery for more than 4 hours
uninterruptedly.
Disconnect the DC power and power the pulse oximeter with battery until the pulse
oximeter is turned off.
Battery duration reflects the battery performance.
If the battery operating time is significantly shorter than the time stated in specifications,
please contact our service personnel for replacing the battery.
•
authorized by the manufacturer can open the battery compartment and replace
or check the battery, and when it needs replacement, the replacement should be
same model with established battery.
BT-710 Operation manual
40
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
Do not disassemble the battery, throw it in fire, or short-circuit it. Battery fire,
battery with naked hand directly.
8.4 Battery recycling
If the battery has visible damage or cannot store power, it should be replaced and recycled
properly. Follow the appropriate regulations to dispose of used batteries.
•
explosion and leakage may lead to personal injury. Do not touch the leaking
BT-710 Operation manual
41
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
Before cleaning the pulse oximeter or sensor, turn off the power and disconnect
9 Caring and cleaning
9.1 Overview
In the using process, please make sure that there is no dust on or near your device. To prevent
damage, please use the diluted detergents and disinfectants specified in this manual, and use
the lowest possible concentration. For the damage or accident caused by using other materials
or methods, our company does not assume any responsibility.
9.2 Cleaning
The device should be cleaned regularly. In the heavily polluted environment, increase the
frequency of cleaning. Before cleaning, please consult the hospital about device cleaning
requirements.
Below are available cleaning agents:
Diluted ammonia
Diluted sodium hypochlorite (washing bleach)
Diluted formaldehyde
Hydrogen peroxide (3%)
Ethanol (70%)
Isopropanol (70%)
Before cleaning:
Turn off the pulse oximeter, disconnect the power cord.
Use a soft cotton ball to adsorb appropriate amount of cleaning agent and wipe the
display screen.
Use a soft lint-free cloth to adsorb appropriate amount of cleaning agent and wipe
the surface of the device.
If necessary, use a clean, dry, lint-free cloth to remove any excess detergent.
Dry the device naturally in a ventilated cool environment.
•
the DC power.
BT-710 Operation manual
42
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
The pulse oximeter should be kept clean. It is recommended to regularly clean
etching cleaner such as soap and water.
CAUTION
Do not leave any cleaning solution on the surface of any part of the device.
NOTE
Wipe the pulse oximeter and sensor surface with medical alcohol, dry it naturally
lease consult the infectious disease control officers or experts of the
hospital for advice.
CAUTION
with gas (EtO) or formaldehyde.
•
the enclosure surface and the display screen. Cleaning the enclosure with non-
•To avoid damaging the pulse oximeter:
Do not use strong solvents such as acetone.
Most cleaners must be diluted before use. Diluting should be according to
the manufacturer's instructions.
Do not use abrasive materials (such as steel wool).
Do not allow any liquid entering into the enclosure, and never immerse any
part of the device into liquid.
or with clean, dry, lint-free cloth.
Bistos is not liable for effectiveness of using these chemicals for infectious disease
control. P
9.3 Disinfection
In order to avoid damage to the product, we recommend that the product is disinfected only
when it is deemed necessary by the hospital maintenance procedures. We also recommend
that the instrument to be disinfected must first be cleaned.
• To prevent damage to the pulse oximeter, do not disinfect the pulse oximeter
BT-710 Operation manual
43
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
If the hospitals or institutions using this instrument can’t implement a
satisfactory maintenance schedule, it will result in device failure and may
endanger human health.
10 Maintenance
•
10.1 Checking
Check the following basic items before using the pulse oximeter:
Check for any mechanical damage.
Check all wires and accessories.
Check all instrument functions that may be used for pulse oximeter and ensure that the
instrument is in good working condition.
If the instrument function has any sign of damage, do not use this pulse oximeter. Please
contact the hospital's professional maintenance personnel or our customer service personnel.
Every 6-12 months or after each repair, a comprehensive examination must be performed by
trained and qualified technical service personnel, including functional safety checks. The
specific inspection items are as follows:
Environment and power meet the requirements.
Device and accessories have no mechanical damage.
The power supply has no wear, and the insulation is good.
Specified accessories are used.
Alarm system is functioning correctly.
Battery performance meets the requirements.
Monitoring functions are in good working condition.
Ground impedance and leakage current meet the requirements.
All checks that require disassembling the instrument must be performed by qualified service
personnel. Safety and maintenance checks may also be carried out by the Bistos personnel.
BT-710 Operation manual
44
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Problem
Solution
Check the battery. If the battery is low, it should be
promptly charged.
Is the red light on the finger sensor flashing? If not,
and the connector.
Is the patient’s arm under pressure? Never take
same arm
Is the environmental temperature too low? Never
affect readings.
Has all patient nail polish, especially blue or purple,
been removed?
During long term monitoring, patient movement
motion are normal.
Inspection / Maintenance Item
Frequency
Check the safety according to IEC
60601-1
At least once every two years, after replacing the
power supply or the pulse oximeter falls down.
Check all monitoring or measuring
functions not listed
At least once every two years, or when you suspect
that the measured value is not accurate.
10.2 Trouble shootings
Device not power on
there might be poor contact. Check the patient cable
SpO2 waveforms or values do not
displayed on the screen
SpO2 values turn on and off during
SpO
monitoring
2
blood pressure and SpO2 measurements on the
expose the patient’s arm to cold air since this can
might result in SpO2 interruptions. Keep the patient
stabilized. SpO
interruptions due to patient hand
2
10.3 Maintenance plan
The following tasks can only be done by qualified service personnel of Bistos. When the
following maintenance is needed, please contact your service representative. Before testing or
maintenance, clean and disinfect the device.
BT-710 Operation manual
45
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
WARNING
Use the accessories specified in this manual. Using other accessories may
onment of the monitor should meet the
requirements of the accessories. Please refer to the manual of the accessories
If the packaging or accessories have any sign of damage, do not use such
For SpO2 Sensor, the normal life time is two years. Please replace in time.
No.
Description
QTY
Type-number
Manufacturer:
Unimed Medical Supplies,Inc
U403-01
11 Accessories
•
damage the monitor, or cannot reach the safety and performance claimed in this
manual.
• The operating and storage envir
for these requirements.
•
accessories.
•
Standard accessories are as follows:
Adult Finger Clip SpO
1
Sensor Probe
2
1
BT-710 Operation manual
46
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Category Name
Specification
Type of electric shock
protection
Class II and internally powered equipment
Electric shock protection
Type BF applied part
Explosion protection
Common equipment, no explosion protection
Liquid inlet protection
IPX2
Operating mode
Continuous mode
Movement
Hand-held equipment
Power
Input: AC 100 ~ 240V (50/60 Hz)
Output: DC 5V / 2.0A
3.7V Li-ion battery 3000 mA
Operating Time(When it fully charged): 5 hours
Charging Time(Fully): 4 hours
Physical Characteristics
Dimensions
Main Unit: 84(W) X 158.5(H) X 34.5(D)
Weight
< 1.5 Kg for standard configuration
Display
Type
Color TFT touch screen LCD
Size
4.3”
12 Specifications
12.1 Safety specifications
12.1.1 Product category
In accordance with classification specified in the European Medical Device Directive 93/42/EEC,
this pulse oximeter is Class IIb device. The pulse oximeter is classified as follows in accordance
with IEC 60601-1:
grade
grade
grade
12.1.2 Power
Adapter
Rechargeable Battery
12.2 Hardware specifications
BT-710 Operation manual
47
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Audio
Alarm tone (45 ~ 85 dB)
Pulse tone
Alarm sound meet the IEC 60601-1-8 standard requirements
SpO2
Standards compliant
ISO 80601-2-61:2011
Display range
0% ~ 100%
SpO2 display resolution
1%
SpO2 accuracy
±2% (at the range 70%~100%)(adult/pediatric mode)
not define when lower than 70% ;
Measurement accuracy verification
Among them medium skin are 4 subjects, light skin are 5
Select “Alarm Delay”, you can choose “off”, “1s”, “2s”, “3s”,
or “8s”.
PR
Display range
25~250bpm
Resolution
±1 bpm
Accuracy
±2% or ±2bpm,whichever is greater
Speaker
12.3 Functional specifications
±3% (at the range 70%~100%) (neonate mode)
The SpO2 accuracy has been verified in human experiments by
comparing with arterial blood sample reference measured
with a CO-oximeter. Pulse oximeter measurements are
statistically distributed and about two-thirds of the
measurements are expected to come within the specified
accuracy range compared to CO- oximeter measurements.
The accuracy of the oximeter has been validated by a clinical
trial involving 12 healthy adult subjects - 4 women and 8 men.
generation delay
Data averaging
Alarm condition delay
subjects, dark skin are 3 subjects, the age from 21 to 28.
Over the range of 70% to 100%, overall accuracy was determined
by calculating the root mean square error across all samples
If the “Alarm Delay” sets more than “1s”, the alarm generation
“4s”, “5s”, “6s”, “7s”
BT-710 Operation manual
48
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
NOTE
Using unqualified accessories, sensors and cables will increase the
electromagnetic emission and reduce the electromagnetic immunity of the
Do not put the device close to other devices or stack together. When necessary,
The device requires special EMC protection, and it is necessary to install and
When the input signal amplitude is smaller than the minimum amplitude
specified in the technical specifications, it may result in inaccurate
on the device.
13 Manufacturer’s declaration on EMC
BT-710 needs special precautions regarding EMC (Electromagnetic compatibility) and needs
to be used according to the EMC information provided in this user manual. Wireless
communications equipment such as wireless home network devices, mobile phones, cordless
telephones and their base stations, walkie-talkies can affect the BT-710 and should be kept at
least 1 m away from the equipment.
device.
observe the device closely to ensure that it runs normally in the environment.
maintain it in the environment that meets the following EMC information.
Even if other devices comply with CISPR emission requirements, they may also
cause interference to this device.
measurements.
Mobile communication devices or wireless network devices may have an impact
BT-710 Operation manual
49
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
The BT-710 is suitable for use in all
13.1 Electromagnetic emissions
The BT-710 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-710 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
The BT-710 uses RF energy only for its internal
RF emissions
CISPR 11
Group 1
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Class A
Class A
Voltage fluctuations /
flicker emissions
Complies
IEC 61000-3-3
establishments other than domestic, and may
be used in domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes, provided the
following warning is heeded:
Warning: This BT-710 is intended for use by
healthcare professionals only. This equipment/
system may cause radio interference or may
disrupt the operation of nearby equipment. It
may be necessary to take mitigation measures,
such as re-orienting or relocating the BT-710 or
shielding the location.
BT-710 Operation manual
50
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
The BT-710 is intended for use in an electromagnetic environment in which radiated RF
can help prevent
maintaining a minimum distance between portable and
as recommended
Separation distance according to frequency of transmitter
�
�
[
7
3
]
�
0.01
0.1
1
10
100
r transmitters rated at a maximum output power not listed above, the recommended
maximum output power rating of the
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is
13.2 Recommended separation distances between portable and
mobile RF communications equipment and BT-710
disturbances are controlled. The customer or the user of the BT-710
electromagnetic interference by
mobile RF communications equipment (transmitters) and the BT-710
below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
[W]
Fo
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
150 kHz to 80 MHz
d=3.5
0.35 0.35 0.23
1.11 1.11 0.74
3.5 3.5 2.34
11.07 11.07 7.38
35 35 23.24
80 MHz to 800 MHz
[m]
d=3.5
800 MHz to 2.5 GHz
d=
affected by absorption and reflection from structures, objects and people.
BT-710 Operation manual
51
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
IEC 60601
Electromagnetic
Electrostatic discharge
±8 kV Contact
±8 kV Contact
Floors should be wood,
Electrical fast
±2 kV for power
±2 kV for power
Mains power quality
Surge
±1 kV differential
±1 kV differential
Mains power quality
Voltage dips, short
< 5 % Uт (> 95 %
< 5 % Uт (> 95 %
Mains power quality
Power frequency (50
3 A/m
3 A/m
Power frequency
13.3 Electromagnetic immunity
The BT-710 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-710 should assure that it is used in such an environment.
Immunity test
(ESD)
IEC 61000-4-2:2009
transient/burst
IEC 61000-4-4:2004
IEC 61000-4-5:2006
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11:2004
Test level
±15 kV air
supply lines
±1 kV for
input/output lines
(>3m)
mode
±2 kV common
mode
dip in Uт) for 0.5
cycles
40 % Uт (60 % dip
in Uт ) for 5 cycles
70 % Uт (30 % dip
in Uт) for 25
cycles
<5 % Uт (> 95 %
dip in Uт ) for 5 s
Compliance level
±15 kV air
supply lines
±1 kV for
input/output lines
(>3m)
mode
±2 kV common
mode
dip in Uт) for 0.5
cycle
40 % Uт (60 % dip
in Uт ) for 5 cycles
70 % Uт (30 % dip
in Uт) for 25
cycles
<5 % Uт (> 95 %
dip in Uт ) for 5 s
environment -guidance
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
should be that of a typical
commercial or hospital
environment.
should be that of a typical
commercial or hospital
environment.
should be that of a typical
commercial or hospital
environment. If the user
of the BT-550 image
intensifier requires
continued operation
during power mains
interruptions, it is
recommended that the
BT-710 be powered from
an uninterruptible power
supply.
Hz and 60 Hz)
magnetic field
IEC 61000-4-8:2010
NOTE Uт is the a.c. mains voltage prior to application of the test level.
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
BT-710 Operation manual
52
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Conducted RF
IEC 61000-4-6:2009
3 Vrms
150 kHz to 80 MHz
3 Vrms
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Electromagnetic environment - guidance
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
mobile radios, amateur radio, AM and FM radio broadcast and TV
environment due to fixed RF transmitters, an electromagnetic site survey should be
is used
normal operation. If abnormal performance is observed, additional measures may be
Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.
The BT-710 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-710 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Portable mobile RF communications equipment should be used no closer to any part of the
BT-550, including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d-
1.2� (d-3.5�)
d-1.2� (Resp: d-3.5�) 80 to 800MHz
d-1.2� 800M to 2.5GHz
where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as deter-mined by an electromagnetic site survey
a
, should be less than the compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked with the following symbol :
NOTE 1) At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is
telephones and land
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
considered. If the measured field strength in the location in which the BT-710
exceeds the applicable RF compliance level above, the BT-550 should be observed to verify
necessary, such as re-orienting or relocating the BT-710.
b
BT-710 Operation manual
53
P/N : 710-ENG-OPM-EUR-R01
Bistos Co., Ltd.
2019.06
Product Name
Pulse oximeter
Model Name
BT-710
Serial No.
Warranty Period
2 Years
Date of Purchase
Hospital:
Sales Agency
Manufacture
Bistos Co., Ltd.
Product Warranty
Address:
Customer
※ Thank you for purchasing BT-710.
※ This product is manufactured and passed through strict quality control and inspection.
※ Compensation standard concerning repair, replacement, refund of the product complies
with “Framework Act on Consumers” noticed by Fair Trade Commission of Republic of
Korea.
Name:
Telephone:
Service Telephone and Fax. Numbers
Telephone: +82 31 750 0340
Fax: +82 31 750 0344
Bistos Co., Ltd.
th
7
FL., A Bldg., Woolim Lions Valley 5-cha, 302,
Galmachi-ro, Jungwon-gu, Seongnam-si,
Gyeonggi-do, Korea
www.bistos.co.kr
bistos@bistos.co.kr
Obelis s.a
Bd. Général Wahis 53 1030 Brussels /
BELGIUM
Telephone: +32 2. 732.59.54
Fax: +49 2. 732.60.03
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.