Bistos BT-250 User Manual

BT-250
BT-250
Fetal Doppler
Operation Manual
Keep this manual for future reference
BT-250 Operation Manual
1
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Proprietary Material
Information and descriptions contained in this manual are the property of Bistos Co., Ltd. and may not be copied, reproduced, disseminated, or distributed without express written permission from Bistos Co., Ltd.
Information furnished by Bistos Co., Ltd is believed to be accurate and reliable. However, no responsibility is assumed by Bistos for its use, or any infringements of patents or other rights of third parties that may result from its use. No license is granted by implication or otherwise under any patent or patent rights of Bistos Co., Ltd.
The information contained herein is subjects to change without notice.
Prepared by: Bistos Co., LTD.
th
7
FL., A Bldg., Woolim Lions Valley 5-cha, 302,
Galmachi-ro, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea
Telephone: +82 31 750 0340 Fax: +82 31 750 0344
Revision 07
Printed in Korea Copyright © Bistos Corporation 2018. All rights reserved.
BT-250 Operation Manual
2
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Table of Contents
1. · System basics ············································································· 3
1.1 Device description and Intended Use ···························································· 3
1.2 Options and Accessories ················································································ 3
1.3 Description of Front Panel ··········································································· 4
1.4 Description of Side Panel ············································································· 4
1.5 Description of Rear Panel ············································································ 5
2. · Preparing to use ······································································· 6
2.1 Place to use ····································································································· 6
2.2 AC/DC adapter and Transducer connection ················································· 6
2.3 Factory Default Setting ················································································· 7
2.4 Understanding the BT-250 Display Screen ···················································· 7
2.5 Button description ························································································ 9
2.6 BT-250 Control Knob ····················································································· 10
2.7 Data Saving ··································································································· 11
2.8 Trend Mode (Data Tracing Mode) ································································ 11
3. · Monitoring fetal heart rate ························································ 13
4. · Cleaning and disinfection ························································· 15
4.1 Monitor ········································································································· 15
4.2 Transducer ···································································································· 15
5. · Trouble shooting and maintenance ·········································· 17
5.1 Transducer ···································································································· 17
5.2 Battery Disposal and Handling ······································································ 17
5.3 Maintenance ································································································· 17
6. · Safety and regulatory information ············································ 18
6.1 Warnings ······································································································· 18
6.2 Cautions ········································································································ 20
6.3 General precaution on environment ···························································· 22
6.4 Symbols ········································································································· 23
6.5 Compliance to the standards ········································································ 24
6.6 Guidance and manufacturer’s declaration – Electromagnetic emissions ······· 25
6.7 Guidance and manufacturer’s declaration – electromagnetic immunity ····· 25
6.8 Recommended separation distances between portable and mobile RF communications equipment and the BT-250 ················································ 28
7. · Specifications ············································································· 29
Product Guarantee ······································································· 30
BT-250 Operation Manual
3
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
(1 meter of water for up to 30 minutes.)
1 System basics
1.1 Device description & Intended Use
The BT-250 is a desktop fetal Doppler that measures the fetal heart rate, which is displayed on a LCD, and outputs the fetal heart sound through built-in speaker. This device is for use only by trained medical personnel.
1.2 Options and Accessories
Accessory Name Description
Doppler
transducer
(1ea)
Ultrasound transducer for measuring
FHR
IPX7 : Waterproof
Power cord
(1ea)
Power adaptor
(1ea)
Ultrasound gel
(1ea)
Table 1.1. BT-250 Accessories
Input : AC100~240 V[50/60 Hz]
AC power cord
AC/DC power adaptor
Output : DC 9V, 2.0A
Ultrasound transmission gel
BT-250 Operation Manual
4
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
① ②
④ ③
⑤ ⑥ ⑦
1.3 Description of the Front Panel
Fig. 1.1 BT-250 Front Panel
Power Indicating LED ( AC: Green / Battery: Orange) Power On/Off Button Control knob Mode change button Save On/Off button Trend mode Button Event mark TFT-Color LCD
1.4 Description of the Side Panel
Fig. 1.2 Left Panel Fig. 1.3 Right Panel
Earphone jack connector Doppler transducer connector Doppler transducer holder
BT-250 Operation Manual
5
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
① ②
1.5 Description of the Rear Panel
Fig. 1.4 Rear Panel
Grip Battery compartment AC/DC adaptor connector RS-232C port USB port Built-in speaker Doppler transducer holder
BT-250 Operation Manual
6
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
2 Preparing to use
2.1 Place to use
Certain strong electromagnetic fields can interfere with the ultrasound transducer and cause a false heart rate reading. This interference is rare, and usually found in the vicinity of large machinery. In order to avoid the possibility of these interfering signals being analyzed as fetal heart rates, the following procedure should be followed whenever the monitor is to be used in a new location, or if it is known that electrical machinery is being operated in the vicinity.
After connecting the ultrasound Doppler transducer, turn on the monitor and observe the heart rate indications on the screen for 30 seconds while no signal input applied to the transducer surface. Intermittent display of random heart rates is acceptable. However, if there is a constant display of a heart rate lasting more than 5 seconds, this is an indication that there is a source of electromagnetic interference in the vicinity. The following steps should be taken to determine if it is possible to use the monitor in this environment.
Move all line cords and line-powered equipment at least 6 feet (1.8 meter)
away from the BT-250. Check for extension cords running behind or under the bed and equipment in adjacent rooms. If the artifact heart rate indication ceases, the monitor may be used normally.
Remove the line cord from the monitor’s power supply. If the artifact heart
rate indication ceases, the monitor may be used normally.
If these measures do not result in cessation of the artifact heart rate, the monitor cannot be safely used in this environment.
2.2 AC/DC adaptor & Transducer Connection
Connect the power cord which supplied by Bistos Co., Ltd. to power outlet and power adapter receptacle. Connect the adapter plug to the BT-250 AC/DC adaptor connector as shown in Figure 2.1. Turn on the BT-250 by pressing down the power ON/OFF button about 2 seconds.
Connect the Doppler transducer cable to BT-250 as shown in the figure below.
BT-250 Operation Manual
7
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Setting parameter
Factory default value
Graph Area
30~240
Auto Shunt Down
5
Language
English
Start Volume
3
Status frame
FHR graph
frame
Patient ID frame
Mode frame
FHR frame
Date/Time
frame
Fig. 2.1 Power adaptor and Transducer Connection
2.3 Factory Default Setting
To enter the factory setting mode, turn on BT-250 by press down the power ON/OFF button for about 2 seconds while press down the control knob simultaneously. In factory setting mode, all configuration parameters are initialized to factory setting value. The initial factory setting values of each parameter are as below;
Monitoring Mode Number Mode
2.4 Understanding the BT-250 Display Screen
Fig. 2.2 Main Monitoring Screen – Graph Mode
BT-250 Operation Manual
8
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Mode frame
FHR frame
Date/Time
frame
Patient ID frame
Status frame
Fig. 2.3 Main Monitoring Screen – Number Mode
To change the monitoring mode between “Number Mode” and “Graph Mode”, press the [MODE] button.
2.4.1 Mode frame
The mode frame shows the current mode. There are monitoring mode
(Number and Graph), setup mode and trend mode.
2.4.2 Heart Rate frame (FHR frame)
The heart rate (FHR) frame displays the fetal heart rate with a heart icon and current speaker volume setting. The heart icon blinks at the measured heart rate interval. The solid heart icon blinks when a valid rate is present and only the outline of heart icon blinks when a measured rate is unstable or weak.
The volume icon indicates the current speaker volume setting for the fetal echo sounds.
2.4.3 Heart Rate Graph frame (FHR Graph frame)
The Heart Rate (FHR) Graph frame displays a graphical representation of the fetal heart rate. The vertical scale is labeled corresponding to the recorder paper (30 to 240 BMP).
2.4.4 Status frame
This frame shows battery status and data saving status.
Symbol Name Description
Battery Status Icon Indicates the battery residual quantity
Save Icon Indicates the data saving status
BT-250 Operation Manual
9
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Low Battery Alarm
Icon
Blinking until AC/DC adapter is
connected.
CAUTION
Never use sharp or pointed objects to operate the front-
To enter into or exit from Trend mode. The trend frames shows historical patient data and the control knob provides navigation capability.
When alarm occurs, alarm status is shown below.
Symbol Name Description
2.4.5 Patient ID frame
This flame displays the patient identification number. The BT-250 uses time and date information to generate the part of ID number (6 digits). The last 3 digits used for the individual patient ID. Default ID number is YYMMDD001 when YYMMDD is the current date information. To change the individual ID number (3 digits) enter [Setup mode] by pressing control knob button. (Refer to ‘4.4 BT­250 Control Knob’ section)
2.4.6 Time and Date
This frame shows the current time and date saved. These settings can be changed as needed. (Refer to ‘4.4 BT-250 Control Knob’ section)
2.5 Button description
There are 5 buttons located on the front panel. The operation of the buttons is summarized below.
panel buttons
Symbol Description
Turns power on or off.
MODE Display mode Change [ Graph Mode Number Mode]
SAVE Start and stop the save function.
TREND
E.MARK Marking event
BT-250 Operation Manual
10
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
CAUTION
Pressing the knob on ‘Delete All Memory’ item makes the all
ID (Last 3 digits)
001 ~ 999
Graph Area
30~240/100~180
Auto Shut Down (minutes)
5/10/15/30/off
Time
HH:MM:SS
Language
English/Spanish
Bright
Start Volume
2.6 BT-250 Control Knob
In monitoring mode, the control knob decrease and increase the fetal heart audio volume. In Trend mode, use Control Knob to search the stored data to recall. After selecting the stored data, press down the Knob to see the data. In the Setup mode, use Control Knob to adjust parameters. Press down the control knob in monitoring mode to enter Setup mode. Rotate the knob to select the item and press down the knob for editing. Rotate the knob again to change the value and press knob after change.
saved data is deleted.
To save the changed value and exit from Setup mode, select ‘ESC’ and press the knob. The BT-250 will return to monitoring mode after storing the changes values.
Fig. 2.4 System setup menu
Configuration parameter Available List
Date YY/MM/DD
1~5
1~7
BT-250 Operation Manual
11
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Saved data
frame
Data Searching
Frame
FHR graph
frame
Patient ID frame
2.7 Data Saving
BT-250 has a data saving function. It can save up to 4 hours (10 minutes for one time, total 25 times).
Press [SAVE] button to start saving. When the function is stared, the save icon [ ] is activated and rotated. Press [SAVE]button again to stop saving.
2.8 Trend Mode (Data Tracing Mode)
Press [TREND] button to enter the Trend mode. In trend mode, the saved data is displayed.
Fig. 2.5 Trend Mode display
Rotate the control knob to select the saved data and press it to see the saved data.
Mode frame
FHR frame
Start Time
Fig. 2.6 Saved Data Tracing Mode display
BT-250 Operation Manual
12
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Button
Description
2.8.1 Saved Data Start Time Frame
This frame shows the date and start time of data saving.
2.8.2 Patient ID Frame
This frame shows the saved patient ID.
2.8.3 Data Searching Frame
This frame is consisted of control buttons for searching saved data. The description of each button is shown below:
Searching for saved data in previous page.
Searching for saved data in next page.
Tracing the saved data
BT-250 Operation Manual
13
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
3 Monitoring fetal heart rate
Fetal heart rate is measured by placing an ultrasound Doppler transducer on the maternal abdomen and by analyzing the echo signal to calculate a heart rate and an audible sound.
Step 1: Preparing the Monitor
Turn the monitor on and verify that the normal monitoring screen appears on the display. Do not use the BT-250 if an error occurs.
Check whether the monitor is powered from the internal battery or AC power. If operating on the internal battery, check the power status frame on the display to determine whether the battery has sufficient charge to complete the monitoring session. Use the AC power if the battery is too low.

Step 2: Acquiring the Fetal Heart Signal

Determine the location of the fetal heart using palpation or a fetoscope. Place the transducer on the maternal abdomen and listen for the fetal heart signal. Reposition the transducer for the loudest fetal heart signal and verify the heart icon on the screen is blinking at the fetal heart rate.
Secure the ultrasound transducer. Make sure the transducer is still positioned for the loudest fetal heart signal.
Verify the monitor is displaying fetal heart rate values and that the heart icon on the screen is blinking at the measured heart rate.

Step 3: Monitor Adjustments

Readjust the volume settings for the desired loudness.
BT-250 Operation Manual
14
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06

Detail Procedure

Explain procedure to the patient.
Turn the monitor power on. The power button is located on the front panel.
Determine the position of the fetus using Leopold’s maneuvers. The
strongest fetal heart tones are heard through the fetal back.
Plug the ultrasound transducer cable into the connector labeled “DOP.”
Apply a small amount of ultrasonic transmission gel to the face of the
transducer.
Place the transducer face down on the maternal abdomen over the area
determined to be the fetal back.
Volume Up/Down may be used to adjust the volume.
Reposition the transducer as necessary until the clearest heart sound is
heard. Three to five seconds after a clear heart beat sound is heard, the
heart icon will flash synchronously with the sound. This indicates that the
received signal is stable.
Ultrasound transducer
Fig. 3.1 Direction of Doppler Transducer
BT-250 Operation Manual
15
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
WARNING
Unplug the monitor from the AC power source and detach
the water or allow liquids to enter the case.
CAUTION
Take extra care when cleaning display surface, which are
with a soft, dry cloth.
WARNING
To avoid electrical shock and damage to the monitor,
disconnect the transducer prior to cleaning and disinfecting.
CAUTION
Transducers are sensitive instruments – irreparable damage
Do not autoclave or EO gas sterilize.
4 Cleaning and disinfection
The BT-250 requires proper care and preventive maintenance. This ensures consistent operation and maintains the high level of performance necessary in monitoring procedures. All exterior parts and transducer should be cleaned and/or disinfected as necessary or between uses. Clean each compartment to remove any surface particles.
4.1 Monitor
Turn off the monitor prior to clean.
Use a clean gauze pad or lint-free cloth, lightly moistened with a mild detergent, to wipe the surface of monitor. Ensure that cleaning solution does not seep into the monitor. Be careful not to power on the monitor during cleaning. After cleaning, use a clean, lint-free cloth to dry the surface.
4.2 Transducers
all accessories before cleaning. Do not immerse the unit in
sensitive to rough handling. Rub the lens that covers them
may occur if they are dropped, knocked against other objects, cut, or punctured. Do not attempt to repair to alter any part of a transducer.
BT-250 Operation Manual
16
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
To clean a transducer,
1. Disconnect the transducer from monitor.
2. Moisten a clean gauze pad with water and wipe the transducer to
remove any gel or particles remaining on the transducer. If water is not effective, then you can use an approved pre-cleaner or low-level disinfectant.
3. Carefully wipe the entire transducer, including the cable and connector.
4. After cleaning, use a clean cloth to dry the transducer.
To disinfect a transducer,
1. Disconnect the transducer from the monitor.
2. Thoroughly clean, rinse, and dry the transducer.
3. Take care to keep the cable strain relief and the connector of the
transducer dry while immersing the transducer in an approved disinfectant.
4. Carefully follow the disinfectant manufacturer’s instructions for
disinfections or high-level disinfection.
5. After disinfecting or high-level disinfecting, use a clean cloth to dry the
transducer.
The following high-level disinfectant agents have been approved for use with transducers.
Cidex OPA Cidex Plus
BT-250 Operation Manual
17
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
5 Troubleshooting and maintenance
5.1 Transducer
To test an ultrasound transducer:
1. Properly connect the transducer to the side of the monitor.
2. Turn on the monitor.
3. Adjust the speaker volume to an audible level.
4. Hold the transducer on one hand and tap on the transducer face with the
other hand. The tapping sound should be heard from the monitor speaker if the transducer operating properly. The transducer is operating properly if you can hear noise from the speaker. Do not use the transducer if no sound is heard or until the proper cause is identified and repaired.
5.2 Battery Disposal and Handling
Be sure to follow the applicable regulations and/or laws regarding recycling when dispose the internal Li-ion battery. Avoid storing battery above 60°C (140 °F). If cloth or skin comes in contact with material from inside the battery, immediately wash with plenty of clean water.
5.3 Maintenance
To maintain the safety and functionality of the BT-250, maintenance must be performed every 12 months. Electrical safety tests must also be performed at regular intervals as specified by local safety regulations.
BT-250 Operation Manual
18
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
WARNING
A WARNING notice indicates a hazard. You need to observe an
injury or death.
CAUTION
A CAUTION notice indicates a hazard. You need to observe an
the system or less of important data.
WARNING
EXPLOSION HAZARD — Do not use the BT-250 in an
Before cleaning up and disinfecting the monitor, always
6 Safety and regulatory information
You should make sure to comply with the following safety precautions during all phase of operation. If you fail to comply with these safety precautions or specific warnings in this manual, you violate safety standards in terms of design, manufacture, and intended use of this monitor. Bistos Co., Ltd. does not have liability for your failure to comply with these requirements.
Safety Notice
operating procedure, practice, or conducts like that. If you do not correctly perform this notice, it could result in personal
operating procedure, practice, or conducts like that. If you do not correctly perform this notice, it could result in damage to
6.1 Warnings
explosive atmosphere.
SHOCK HAZARD —Do NOT use BT-250 with RF Surgical
equipment.
SHOCK HAZARD — Do not attempt to connect or
disconnect a power cord with wet hands.
Always use accessories and cables supplied or appointed by
Bistos Co., Ltd.
Do not contact RS-232C port and patient at the same time.
Use only AC/DC Adaptor supplied or appointed by Bistos
Co., Ltd.
Do not use stacking and location close to other equipment. Do not remove the covers of a monitor yourself to avoid
damage to the monitor and unexpected electrical shock. Only qualified Bistos service engineer must repair or replace components.
BT-250 Operation Manual
19
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
make sure turn off the monitor and unplug the power cord
secure in transducer holder when you do not use it.
from the power outlet.
Do not allow water or liquids on or above the monitor.
Dripping water or liquids into the monitor may cause electrical shock and damage to the monitor.
Always use accessories approved by Bistos. You must
secure connect the accessories to the monitor.
Do not modify the monitor such as components, or
software. When you modify the monitor, it may cause safety hazards. Only qualified Bistos service engineer must modify the monitor.
Always use the monitor properly to avoid serious injury.
Before using the monitor, you must read the instructions for use carefully.
Always use transducers approved or supplied by Bistos.
When you observe that the monitor causes any
malfunction, you must stop operating the monitor and contact to Bistos service engineer.
Always make sure that you do not use the monitor in an
explosive atmosphere.
You should use the monitor after few hours when the
monitor is in a high humidity place.
Using spray cleaners on the monitor drips cleaning fluid
into the monitor. It damages component in the monitor.
Do not use aerosol spray cleaners on the monitor to
prevent electrical shock and damage to the monitor.
Do not use damaged or defective transducer to prevent
monitor damage and serious patient injury.
Use only approved or supplied ultrasound transmission gel.
Using unapproved gels may damage the transducer and void the warranty.
Do not drop the transducer. Always keep the transducer
BT-250 Operation Manual
20
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
CAUTION
The relevant law restricts this device to sale by or on the
Only the AC line cord supplied with the BT-250, or its
6.2 Cautions
order of a physician.
Keep the operating environment free of dust, vibrations,
corrosive or flammable materials and extremes of temperature and humidity. The unit should be kept clean and free of transducer gel and other substances.
When installing the unit into a cabinet, allow for adequate
ventilation, accessibility for servicing, and room for adequate visualization and operation.
Do not operate the unit if it is damp or wet because of
condensation or spills. Avoid using the equipment immediately after moving it from a cold environment to a warm, humid location
Never use sharp or pointed objects to operate the front-
panel switches.
Always connect the certified equipment with relevant
International standards (e.g. IEC 60950 for IT equipment and IEC 60601-1 for medical equipment) to RS-232C and USB connector.
Do not autoclave or gas sterilize the BT-250 or any
accessories. Follow cleaning and disinfection instructions in Section 6 of this manual.
Do not immerse transducers in liquid. Follow cleaning and
disinfection instructions in Section 6 of this manual.
When washing the transducer belts, the water temperature
must not exceed 60°C (140°F).
When operating the device by external power supply, noise
can be appeared on the sound output due to the instability of power supply. In this case, operate the device by internal battery.
Examine the device main body and any accessories
periodically to ensure that the cables, line cords, transducers, and instruments do not have visible evidence of damage that may affect patient safety or monitoring performance. The recommended inspection interval is once per week or often. Do not use the device if there is any visible sign of damage.
BT-250 Operation Manual
21
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
equivalent, is approved for use with the device.
Do not tamper the BT-250. Only qualified service personnel
should attempt any needed internal servicing.
The BT-250 is not specified or intended to use with
defibrillators or during defibrillator discharge.
The BT-250 is not specified or intended to use with
electrosurgical equipment.
The BT-250 is not specified or intended to use with any
other type of monitoring equipment except the specific devices that have been identified for use in this Operator’s Manual.
Perform periodic safety testing to insure proper patient
safety. This should include leakage current measurement and insulation testing. The recommended testing interval is once per year.
BT-250 Operation Manual
22
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
30 % to 85 %.
6.3 General precaution on environment
Do not keep or operate the equipment under the environment listed below.
Avoid placing in an area exposed to moisture. Do not touch the equipment with wet hand.
Avoid placing in an area where high variation of temperature exists. Operating temperature ranges from 10 ℃ to 40
. Operating
humidity ranges from
Avoid exposure to direct sunlight
Avoid in the vicinity of electric heater.
Avoid placing in an area where there is an excessive humidity rise or ventilation problem.
Avoid placing in an area where chemicals are stored or where there is in danger of gas leakage.
Do not disjoint or disassemble the device. Bistos Co., Ltd. does not have liability of it.
Avoid placing in an area where there is an excessive shock or vibration.
Avoid dust and especially metal material enter into the equipment
Power off when the equipment is not fully installed. Otherwise, the equipment could be damaged.
BT-250 Operation Manual
23
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Safety Symbol

Description

This symbol identifies a safety note. Ensure you understand Patient applied part meets the isolation requirements for the
6.4 Symbols
Power On/Off Button
the function of this control before using it. Control function is described in the operating manual.
External Signal IN/OUT Port
type BF equipment
IPX7
Protection against ingress of water and particulates IPX7 Waterproof
BT-250 Operation Manual
24
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
6.5 Compliance to the standards
The BT-250 ultrasound Doppler system conforms to the following classifications, in accordance with the EN 60601-1
Type of protection against electrical shock : Class II, internally powered
equipment
Degree of protection against electrical shock (Patient connection ) : Type BF
equipment
Degree of protection against harmful ingress of water : ultrasound transducer
IPX7
Mode of operation : Continuous operation
The BT-250 ultrasound Doppler system conforms to the following standards;
EN 60601-1:2006 + A11:2011
EN 60601-1-2:2007/AC:2010
EN 60601-2-37:2008+A11:2011
EN 60601-1-6:2010
EN 62304:2006
EN 62366:2008
IEC 60529:2013
EN 1041:2008
EN ISO 14971:2012
EN 980:2008
EN ISO 10993-1:2009
EN ISO 10993-5:2009
EN ISO 10993-10:2013
BT-250 Operation Manual
25
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
The BT-250 is intended for use in the electromagnetic environment specified
such an environment.
Electromagnetic environment ­guidance
The BT-250 uses RF energy only for its
electronic equipment.
RF emissions CISPR 11
The BT-250 is suitable for use in all
Harmonic emissions IEC 61000-3-2
Voltage fluctuations /
IEC 61000-3-3
The BT-250 is intended for use in the electromagnetic environment specified
such an environment.
Electromagnetic
guidance
Floors should be
30 %.
± 1kV for input /
Mains power quality should be that of a
environment.
6.6 Guidance and manufacturer’s declaration - Electromagnetic emissions
This system is suitable for use in the following environment. The user must assure that it is used only in the electromagnetic environment as specified.
below. The customer of the user of the BT-250 should assure that it is used in
Emission test Compliance
RF emissions CISPR 11
flicker emissions
Group 1
Class B
Class A
Complies
internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby
establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
6.7 Guidance and manufacturer’s declaration - electromagnetic immunity
below. The customer or the user of the BT-250 should assure that it is used in
Immunity Test
Electrostatic discharge (ESD) IEC 61000-4-2
IEC 60601 Test level
± 6 kV contact ± 8kV air
Compliance level
± 6 kV contact ± 8kV air
environment -
wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least
Electrical fast transient/ burst IEC 61000-4-4
± 2 kV for power supply lines
output lines
± 2 kV for power supply lines ± 1 kV for input / output lines
typical commercial or hospital
BT-250 Operation Manual
26
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
± 1 kV Line(s) to
± 2kV Line(s) to
Mains power quality should be that of a
environment.
Voltage dips,
and voltage variations on power supply
<5% U
for 5s
<5% U
for 5s
Power frequency
environment.
Note UT is the a.c. mains voltage prior to application of the test level.
IEC 60601 test level
Compliance level
Electromagnetic environment ­guidance
80 MHz to
Portable and mobile RF
Surge
IEC 61000-4-5
short interruptions
input lines IEC 61000-4­11
Power frequency (50/60Hz) magnetic field IEC 61000-4-8
± 1 kV Line(s) to
line(s)
line(s) ± 2kV Line(s) to
earth
T
(>95% dip in UT) for 0.5 cycle
40% U
T
(60% dip in U
T
5 cycles
70% U
T
(30% dip in U
T
25 cycles
<5% U
T
(>95% dip in UT)
earth
(>95% dip in UT) for 0.5 cycle
40% U
) for
(60% dip in U for 5 cycles
70% U
) for
(30% dip in U for 25 cycles
<5% U (>95% dip in UT)
3 A/m 3 A/m
typical commercial or hospital
T
Mains power quality
T
should be that of a typical commercial or
)
T
hospital environment. It is
T
recommended to be powered by the
)
T
internal battery if it is needs to operate when the supply of
T
main power is cut.
magnetic fields should be meet the level of a typical commercial area or hospital
Immunity test
Conducted RF IEC 61000-4-6
Radiated RF
3 Vrms 150 kHz to 80 MHz
3 V/m
IEC 61000-4-3
2.5 GHz
3 Vrms 150 kHz to 80 MHz
3 V/ms 80 MHz to
2.5 GHz
communications equipment should be used no closer to any part of the BT-250, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance.
BT-250 Operation Manual
27
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
= 1.2√
NOTE 1) At 80 MHz and 800 MHz, the higher frequency range applies.
and people.
a
Field strengths from fixed transmitters, such as base stations for radio
than 3 V/m.
= 1.2√ 80 MHz to 800 MHz
= 2.3√ 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol :
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the EUT is used exceeds the applicable RF compliance level above, the EUT should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the EUT.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
BT-250 Operation Manual
28
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Rated maximum
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
= 1.2√
80 MHz to 800 MHz
= 1.2√
800 MHz to 2.5 GHz
= 2.3√
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73 1 1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
NOTE: The application range of high frequency is from 80 MHz and 800 MHz.
absorption and reflection from structures, objects and people.
6.8 Recommended separation distances between portable and mobile RF communications equipment and the BT-250
The BT-250 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the BT-250 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the BT-250 as recommended below, according to the maximum output power of the communications equipment
output power of transmitter [W]
NOTE: However, it does not mean that these guidelines are applied to every situation without an exception; electromagnetic waves are affected by
BT-250 Operation Manual
29
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Physical Characteristics
Main body
90 mm(H) x 250 mm(L) x 118 mm(D)
Transducer
29.7mm(Ø) x 145 mm(H)
Weight
approx. 1.5 kg
Power
Internal
3.7V, Li-ion, rechargeable
4 hours Fast Charge
Continuously 5 hours operating
External
AC/DC Adaptor
Input : AC100~240 V[50/60 Hz]
Output : DC 9 V, 2.0 A
Power dissipation
18VA, maximum
Environmental
Operating Temperature
10°C to 40°C (50°F to 104°F)
Transfer & Storage Temperature
–20°C to 60°C (–4°F to 140°F)
Relative Humidity
20% to 90% non-condensing
Altitude
0 - 3048m (0 -10,000 ft)
Doppler Ultrasound FHR Monitoring
I Entrance beam dimensions
20 mm, circular
Ultrasonic frequency
2 MHz
BPM Range
30-240 BPM
Accuracy
±2% of range
Leakage
<10 µA @ 264 VAC applied to transducer
Isolation
>4 kV RMS, Type BF applied part
7 Specifications
spta
< 94 /
BT-250 Operation Manual
30
P/N: 250-ENG-OPM-EUR-R07
Bistos Co., Ltd.
2018,06
Product Name
Fetal Doppler
Model Name
BT-250
Approval No.
Approval Date
Serial No.
Warranty Period
2 Years (Transducer excluded)
Date of Purchase
Hospital:
Telephone:
Sales Agency
Manufacture
Bistos Co., Ltd.
Product Warranty
Customer
Address: Name:
Thank you for purchasing BT-250. This product is manufactured and passed through strict quality control
and inspection.
Compensation standard concerning repair, replacement, refund of the
product complies with “Framework Act on Consumers” noticed by Fair Trade Commission of Republic of Korea.
Service Telephone and Fax. Numbers
Telephone: +82 31 750 0340
Fax: +82 31 750 0344
th
FL., A Bldg., Woolim Lions Valley 5-cha, 302, Galmachi-ro,
7
Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea
Bistos Co., Ltd.
www.bistos.co.kr
bistos@bistos.co.kr
Obelis s.a
Bd. Général Wahis 53
1030 Brussels, BELGIUM
Telephone: + (32) 2. 732.59.54
Fax.: + (32) 2.732.60.03
Loading...