Indicates the protection level against the ingress of solid object and
IP2X is protection against solid foreign object like a finger.
Intended use/population/user
BT-150 electric breast pump is intended to be used by a lactating woman for expressing and collecting
milk from her breasts.
Product description
As the same principle with the baby sucking the mother's milk, the breast pump electronically operates
the vacuum pump and applies vacuum pressure generated at that time to the breast of the mother in
order to express the breast milk from the breasts to outside.
BT-150 is an electric breast pump used in the home environment, which service massage and expression
mode and is capable of single and double pumping.
The operating life of this breast pump is expected to be approximately 150 minutes sessions per day, for
one year.
Contraindications
There are no known contraindications for the breast pump.
1. Safety Information
Before using BT-150 breast pump, read this entire manual and be fully understood and follow
instructions and safety information to prevent injury.
Warning symbols
The following symbols identify all instructions that are important to safety. Failure to follow these
instructions can lead to injury or damage to the breast pump. When used in conjunction with the
following words, the symbols indicate:
The following symbols are placed on product, label, packing and this manual in order to stand for the
information about:
Used to identify safety information for warning.
Be well-known this information thoroughly before using BT-150.
Used to identify safety information for caution.
Be well-known this information thoroughly before using BT-150
IP21
liquid.
IPX1 is protection against some falling water drops vertically.
Refer to operation manual. Read manual before placing the device.
This symbol indicates alternating current.
BT-150 Opeartion Manual
5
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Indicates the BF applied part.
Diaphragm bottom and Bottle.
Indicates the temperature limitation for operation, transport and
range of atmospheric pressure to which the medical
(Recognized Component Mark for Canada and the United States)
Indicates DC power supply.
Indicates the production date.
Indicates the manufacturer.
Indicates the serial number of the device.
Indicates a reference number.
This symbol indicates that the power adaptor is a class II device.
It is applicable to Funnel, Funnel block, Air tube, Diaphragm Top,
Indicates to keep the device dry.
Indicates the medical device that can be broken or damaged if not
handled carefully.
Symbols on the adaptor
This symbol indicates that the power adaptor is a class II device.
Indicates to keep upright
Indicates to keep the device away from sunlight.
storage.
Indicates the humidity limitation for operation, transport and storage.
Indicates the
device can be safely exposed.
Indicates the packing material is recyclable.
This symbol indicates that the power adaptor is for indoor use only.
This symbol indicates compliance with both Canadian and U.S. component requirements.
BT-150 Opeartion Manual
6
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
WARNING
Do not touch or operate BT-150 with wet hands to avoid electric shock.
Do not reach for the device if it has fallen into water. Unplug the power adapter
The breast pump should be hygienic. Using second hand or rental product can
, temperature and atmospheric
pump should be observed to verify normal operation in the configuration in which
Only the qualified service personnel who
battery replacement could be caused danger such as excessive
immediately from wall socket outlet.
Only use the BT-150 breast pump for its intended use as described in this manual.
Never use while pregnant.
cause cross contamination.
Do not use the device while sleeping or overly drowsy.
Keep away the main body from direct sunlight.
BT-150 and accessories are not heat-resistant. Keep away from radiators, open
frame and heated surface.
BT-150 is a personal care device. Do not use the device by more than one person.
Supervision is necessary when breast pump used in the vicinity of children or pets.
And keep all components not in use out of the reach of them.
Do not use the damaged components including power adapter.
Do not use in the out of range for humidity
pressure environment than indicated in this manual.
The BT-150 electronic breast pump should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the BT-150 electric breast
it will be used.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the BT-150, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Do not use the device as a general suction device for other purposes except the
expression of breast milk.
Do not disassemble the device.
authorized by the manufacturer can open the battery compartment and replace
the battery, and the battery of same model and specification should be replaced.
The incorrect
temperatures, fire or explosion.
If the device is not used for long time, it should be stored that battery is removed.
Charge the battery and store it at room temperature in low humidity.
Use only AC/DC Adaptor supplied or appointed by Bistos Co., Ltd.
BT-150 Opeartion Manual
7
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
CAUTION
If you are a mother who is infected like Hepatitis B, Hepatitis C or Human
Immunodeficiency Virus (HIV), pumping breastmilk will not reduce or remove the
efore each use visually inspect the individual components for cracks, chips, tears,
n the event that damage to the device is observed,
st or nipple, turn the pump off or slide a finger
some inconvenience in the early stage of usage. If the
For expressing breastmilk, make yourself comfortable and do not incline the
For operating pump, connect diaphragm assay tightly to prevent the expressed
Clean and sanitize all compartments that come into contact with breast and milk
isassemble and wash all parts that come in contact with the breast and breast
e sure that the expressed milk does not flow into the main body by checking air
ilk flow in reverse direction, turn the power off, remove the milk from
in reverse direction to the main body, stop using the product
auxiliary tubing port with port cover to prevent
emove the power adapter from breast pump immediately when charging
If breast pump remain unused for a longer period of time, make sure that the
risk of transmitting the virus to your baby through your breastmilk.
B
discoloration or deterioration. I
please discontinue use until the parts have been replaced.
If you feel pain in your brea
between the funnel and your breast to break the suction.
You can experience
inconvenience is persistent or injury or bleeding around the nipple appears,
contact a healthcare professional or breastfeeding specialist immediately.
bottle.
milk flow into the main body.
according to this manual before and after every use.
D
milk immediately after use to avoid dry up of milk residue and to prevent the
growth of bacteria.
Only use drinking-quality water for washing.
Keep dry the all components after clean and sanitize.
B
tube. If m
air tube or diaphragm and replace or clean as instructed this user manual.
If milk follow
immediately and call your nearest service enter.
Do not fill up the bottle to avoid backflow.
Make sure that the air tubes not to get tangled while double pumping.
When single pumping cover the
the air flow into the breast pump.
R
finished.
The fuel level displayed in LCD can differ from actual battery capacity. Use breast
pump with full charged battery when possible.
Charge the battery immediately when [Lo Bt] displayed.
battery full charged before use.
Only authorized service personnel shall repair or replace of BT-150.
BT-150 Opeartion Manual
8
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Never use damaged device. Inspect all appropriate components before each use.
Do not use outdoors, or operate where aerosol (spray) products are being used or
Mobile and portable radio frequency communications equipment can affect the
of burning because of too hot milk and microwave can also change the
hawed breastmilk is safe in the
lace the sealed bottle in a bowl of warm water for 20 minutes to bring it to body
where oxygen is being administered.
If air tube becomes moldy, discontinue use and replace it.
operation of breast pump.
Breast pump, plastic bottles and component parts may break when dropped.
Do not pumping while bathing or showering.
Do not thaw frozen breastmilk in a microwave or in a pan of boiling water (danger
composition of breastmilk)
Thaw breastmilk overnight in the refrigerator. T
refrigerator for 24 hours. Do not refreeze thawed breastmilk.
Quickly thaw breastmilk by holding the bottle under warn running water.
P
temperature.
If you are adding expressed breastmilk to a container of already frozen breastmilk,
make sure to add a lesser amount than the already frozen amount.
When disposing of the BT-150, adhere to all applicable laws regarding recycling.
When handling package materials, abide by local waste disposal laws and
regulations. Keep the packaging materials away from children.
Properly dispose or recycle the depleted battery according to local regulations.
Always use accessories supplied by Bistos Co., Ltd.
BT-150 Opeartion Manual
9
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Name
User manual
Power adapter
Main Body
Nipple
Qty. 1 1 1 1
Name
Diaphragm Top
Diaphragm
Diaphragm Bottom
Air tube connector
Qty. 1 1 1 2
Name
Funnel
Funnel Block
Funnel cap
Air tube/80cm
Qty. 1 1 1 1
Name
Bottle/160 ml
Bottle cap
Bottle cover
Bottle disc
Qty. 1 1 1 1
CAUTION
Use Bistos original accessories only.
To help maintain optimum performance, it is recommended to replace the
2. Product configuration
BT-150 Electric breast pump is utilizes the vacuum to express the breastmilk. It is classified as a
medium vacuum and intermittent suction device.
When unpacking the box, check all the following components are enclosed. All components except main body and adapter are consumables. The standard configuration is composed.
Shape
Shape
Shape
Shape
The color of main body or components can be changed without notice but the performance characteristics
are not influenced by these changes.
Check consumable set for wear or damage before use and replace if necessary.
Especially, as for the nipple, inspect before each use and pull in all directions.
Throw away at the first signs of damage or weakness.
The period of use for consumables excluding battery is recommended 1 to 3
months, but it might be different for the use environment.
Do not leave a nipple in direct sunlight or heat or disinfectant (“sterilizing
solution”) for longer than recommended, as this may weaken the part.
consumables periodically.
BT-150 Opeartion Manual
10
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
NOTE
Basic tubing port:
Power adapter port
LCD Display
Buttons
Remove the protective film from the LCD window before use.
①
②
③
④
①
②
①
Tubing port for single pumping
Auxiliary tubing port:
②
Tubing port for double pumping
Port plug
③
④
BT-150 Opeartion Manual
11
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
3. Prepare to use
3.1 Air tube and connector assembling
Insert the metal part of air tube connector into the end tip of air
tube as shown picture left.
3.2 Assemble and disassemble
⑤
⑥
① Place the diaphragm in the diaphragm bottom in the direction that the cross lined surface facing to
bottom. Cover the diaphragm top and turn it to lock.
② Put the funnel to the funnel block.
③ Put the funnel cap beneath the funnel block. Screw bottle into the funnel assembly.
④ Connect the air hole of diaphragm bottom to the funnel assembly.
⑤ Connects the air tube to basic tubing port on main body and diaphragm top
⑥ Connects the air tube to the bottle block as shown in the picture.
⑦ To clean or storing after use, disassemble the bottle assembly in the reverse order.
BT-150 Opeartion Manual
12
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Fig.
Name
Status
Description
Off
SINGLE expression pressure range
Off
Power adapter is not connected.
automatically when fully discharged.
Express
Mode
The pressure level is displayed from 1 to 16, but the
cycle.
Express
Mode
Massage
Mode
On
Operating in expression mode.
3.3 Display
Pressure
Range
On DOUBLE expression pressure range
Adapter
connection
Battery Fuel level
On Power adapter is connected.
Full charged
75 ~ 100 %
50 ~ 75 %
25 ~ 50 %
0 ~ 25 %
Full discharge
Flashing means charging needed. Shut down
Program
number
On Displays the currently specified program number
Pressure
level
Massage
Mode
Cycle
level
Expression
mode
Flashing Operating in stored expression mode sequence.
Display the pressure level from 1 to 16
maximum pressure level is adjusted according to the
Display the cycle level from 1 to 6
Display the cycle level from 1 to 3
BT-150 Opeartion Manual
13
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Operating time after power on. After 30 minutes, the
displayed.
On
Operating in massage mode
On
There is an executable program sequence.
Program
running
Program
stop
Operating
time
Massage
mode
Program
status
sequence
Program
sequence skip
sequence
On
main body will turned off automatically.
When the battery needs to be charged, [Lo bt] will be
Flashing Operating in stored massage mode sequence.
Flashing Currently, the program sequence is operating.
On Currently, the program sequence is playing
On
Skips the currently playing program sequence one step.
For 2 seconds
On Indicates that program sequence playback is stopped.
BT-150 Opeartion Manual
14
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Basic function
Program function
Short Press
Press & Long Hold
Power On/Off button
Memory start/stop button
Lamp On/Off button
Turn on, change the lamp brightness and
off when you press the third time.
Change Program ID button
Start/stop program storage button
Press shortly to change the active
program ID which is able to check through
The device stores the 8 different program
Program IDs from 1 to 4 are provided by
the user. Program IDs from 5 to 8 are able
Button for entering the mode for setting
Once all of above the operation sequences
have been saved, and then press and hold
this button for more than 2 seconds to
See the next (2) How to use the program
3.4 Operating functions
BT-150 has two operating functions: basic massage/express function and program function to save
the operating sequence. Using program function, comfortable and optimum massage/express mode,
level of speed/vacuum strength/cycle, massage/express time can be saved and restored anytime.
(1) Buttons
Several buttons have two functions. Pressing shortly will make it perform basic functions and
press longer more than 2 seconds will make it perform program functions.
Button
Turn on and off the power.
turn off the lamp
When purchasing a product with a lamp
option, Press the lamp button on the
power-off pump to turn on the lamp. Each
time the button is pressed again, the
brightness of the lamp becomes bright.
Press the button fourth times to turn off
the lamp.
If you purchase a product without a lamp
option, When you press the lamp button
on the power-off pump, the LCD back light
turns on. After that, the LCD back light
brightness becomes bright each time you
press it again, and the LCD back light turns
the PGM screen.
IDs.
the default setting and unchangeable by
to set by the user.
the program sequence of a Program.
Pressing this button for more than 2
seconds starts to save the operation
sequence to the Program ID number from
5 to 8 currently displayed on the LCD
screen.
One program is composed of 8 program
sequences and each program sequence
consists the operation mode, pressure
level, cycle level and running time.
store as the assigned Program ID.
BT-150 Opeartion Manual
15
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
function for more details of the setting.
Mode change button
SINGLE/DOUBLE pressure change button
Toggle between massage and express
express mode.
Toggle between SINGLE and DOUBLE
Vacuum decrease button
Start button
Decrease the vacuum in the operating
value of the pressure level and minute.
Vacuum increase button
Skip button
It is possible to skip a specific operation
sequence during program operation
If this button is pressed for longer than 2
seconds during program playback, the
current program sequence step is ignored
Cycle decrease button
Stop button
Increase the speed the operating mode.
value of the cycle level and second.
Cycle increase button
Increase the speed in the operating mode.
value of the pressure level and minute.
mode.
In the program sequence input mode, the
button is used for selecting the mode of a
program sequence between massage and
pressure mode
mode.
In the program sequence input mode, the
button is used for setting the decreasing
Increase the vacuum in the operating
mode. (Level 1 ~ 16)
In the program sequence input mode, the
button is used for setting the increasing
value of the pressure level and minute.
(Level 1~6 for express mode, Level 1~3 for
massage mode)
In the program sequence input mode, the
button is used for setting the decreasing
Start to operate as saved sequence.
No response when no sequence saved.
sequence reproduction.
Stop to operate as saved sequence.
No response when not operating
according to saved sequence.
(Level 1 ~ 6 for express mode, Level1~3 for
massage mode)
In the program sequence input mode, the
button is used for setting the increasing
BT-150 Opeartion Manual
16
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
(2) How to use the program function
You can save and restore the convenient and optimum massage/express mode, speed/vacuum
level and operating time.
① Able to use the below Program IDs provided as the default setting.
- Program ID number 1 considering the single pumping for mothers with newborn
- Program ID number 2 considering the dual pumping for mothers with newborn
- Program ID number 3 considering the single pumping for mothers with infants
- Program ID number 4 considering the dual pumping for mothers with infants.
Able to set the desired operation sequence to the 5 to 8 Program IDs by the below procedure.
② Select the proper Program ID number () by pressing the Program ID button () shortly.
③ Press the Program ID button () for long to start a mode for setting the operating
sequence.
④ Able to identify the possible state for input by the blinked box of pressure level (), cycle
level () and operating time(). Entering the next input is made by pressing the
Program ID button () shortly.
⑤ While the box of pressure level () and cycle level () is blinking, start to set each
operation sequence for a program sequence. The operation mode can select through the
Mode change button (). The Vacuum decrease button () and Vacuum increase
button () is used to set a decrease and increase of the pressure level. The Cycle decrease
button () and Cycle increase button () is used to set a decrease and increase of the
cycle level.
⑥ While the box of operating time () is blinking, start to set the operation time for
previously set operation mode, pressure level and cycle level. The Vacuum decrease button
() and Vacuum increase button () is used to set a decrease and increase of the
minute. And the Cycle decrease button () and Cycle increase button () is used to set
a decrease and increase of the second.
⑦ If you want to add another program sequence(a next program sequence) within the
Program, press the Program ID button () shortly and repeat the above instruction from
④ to ⑥.
⑧ When you finish all set for the program sequences, press the Program ID button() for
long to exit and store the program sequence input mode.
Operate the device the following instruction:
⑨ The program ID 5 to 8 which have been set up the program sequences through the ② to
BT-150 Opeartion Manual
17
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
CAUTION
Increasing the Frequency might be accompanied by an increase in noise and
⑧ instruction or the program ID 1 to 4 provided as the default setting are able to play by
pressing the Start button () for long.
⑩ To skip a certain program sequence, press the Skip button () for long.
⑪ Anytime, to exit the playing Program, press the Stop button() for long.
⑫ In the operation mode, pressure level and cycle level can change even while playing the
program.
vibration.
3.5 Essential performance
BT-150 specifies to limit the vacuum pressure as - 290 mmHg.
BT-150 Opeartion Manual
18
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
CAUTION
Before each use visually inspect the individual components for cracks, chips, tears,
tments that come into contact with breast and milk
device with a position that air tube or other lines are possible to
4. Pumping
discoloration or deterioration. In the event that damage to the device is observed,
please discontinue use until the parts have been replaced.
Clean and sanitize all compar
according to this manual before first use and after every use.
Do not use the components if dirt, mold or other residual is visible when they do
not be gone even after clean and sanitizing.
Do not use the
be entangled.
4.1 Prepare to pump
(1) Wash hands thoroughly (at least one minute) with soap and water before touching the breast
and pump parts.
(2) Check pump set components for wear or damage before use and replace if necessary.
(3) Make sure that the components that come into contact with breastmilk are cleaned and
sanitized appropriately. Assemble the funnel, diaphragm and bottle according to this manual and
connect to main body via air tube. As to funnel, choose the proper size by individuals.
(4) Check the air tube connected tightly to prevent the vacuum leakage.
(5) Check the battery capacity and connect the power adapter necessary.
(6) Massage the breast for about 5 minutes with warm towel to promote expression.
4.2 Pumping
(1) To prevent the backflow, always pumping while seated.
(O) (X) (X) (X)
(2)
Place the funnel on your breast so that your nipple is centered.
(3) Turn the breast pump on to pumping.
(4) Adjust the speed/vacuum to a comfortable level. Increase speed/vacuum level until pumping
feels slightly uncomfortable (not painful), then decrease it slightly to find the comfortable level.
(5) Normal pumping times are from 15 minutes to 30 minutes but it can vary individually.
(6) When the pumping session is over, turn off the power and slide a finger between the funnel and
your breast to separate the funnel from your breast.
BT-150 Opeartion Manual
19
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Main body
towel after each pumping session
Air tube
or dirty tubing with a new set
CAUTION
Disassemble and wash all parts that come in contact with the breast and
breastmilk immediately after use to avoid dry up of milk residue and to prevent
5. Wash and Sanitize
(1) Clean all parts before using BT-150 for the first time. Especially, all parts that come in contact
with breast milk including the bottle set, before the first use, must be place in boiling water for 5
minutes following the ‘5.2 sanitize’ instruction. This is to ensure hygiene.
(2) Clean the device (main body) after each use with clean and dry towel.
5.1 Wash
(1) Disassemble the bottle block for washing.
(2) Rinse all separated parts that came in contact with breast and breast milk in cool water in order
to remove breast milk residue.
(3) Wash in sink, or
• Soak all separated parts in warm, soapy water for 5 minutes.
• Wash each part with a clean dish-cloth or soft brush.
• Rinse all separate parte with clear water.
(4) Wash in dishwasher
• Wash all separated parts on top rack of dishwasher.
• Allow all breast pump parts to air dry in a clean area.
(5) Place parts on a clean surface and/or towel.
(6) Allow all parts to air dry.
(7) Store dry parts in a clean, cool place when not in use. Do not store wet or damp parts.
when dirty wipe with a clean dry
Do not clean tubing, replace damaged
the growth of bacteria.
Only use drinking-quality water for washing.
Do not use strong detergent to clean.
BT-150 Opeartion Manual
20
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
(1) Disassemble all parts that come in contact with breastmilk.
Boiling water
5.2 Sanitize
(2) Fill a pot with enough water to cover all parts.
(3) Boil the water for 5 minutes while parts in the water.
(4) Allow water to cool and gently remove parts from water with
tongs.
(5) Place parts on a clean surface and/or towel.
(6) Allow all parts to air dry.
(7) Store dry parts in a clean, cool place when not in use.
Do not store wet or damp parts.
6. Storage of breastmilk
The information below is referenced from website of Office of Women’s Health.
Store your breastmilk in clean glass or hard BPA-free plastic bottles with tight-fitting lids. You can
also use milk storage bags, which are made for freezing human milk. Do not use disposable bottle
liners or other plastic bags to store breastmilk.
After each pumping
Label the date on the storage container. Include your child’s name if you are giving the milk to
a child care provider.
Gently swirl the container to mix the cream part of the breastmilk that may rise to the top
back into the rest of the milk. Do not shake the milk. This can cause some of the milk’s
valuable parts to break down.
Refrigerate or chill milk right after it is expressed. You can put it in the refrigerator, place it is a
cooler or insulated cooler pack, or freeze it in small (2 to 4 ounce/60 to 110 g) batches for
later feeding.
Tips for freezing milk
Wait to tighten bottle caps or lids until milk is completely frozen.
Try to leave an inch or so from the milk to the top of the container because it will expand
when freezing.
Store milk in the back of the freezer, not on the shelf of the freezer door.
Tips for thawing and warming up milk
Clearly label milk containers with the date the milk was expressed. Use the oldest stored milk
first.
Breastmilk does not need to be warmed. Some moms prefer to take the chill off and serve at
room temperature. Some moms serve it cold.
Thaw the bottle or bag of frozen milk (1) by putting it in the refrigerator overnight, (2) by
holding it under warm running water, or (3) by setting it is a container of warm water.
Never put a bottle or bag of breastmilk in the microwave. Microwaving creates hot spots that
could burn your baby and damage the milk.
Swirl the milk, and test the temperature by dropping some on your wrist. The milk should be
comfortably warm, not hot.
Use thawed breast milk within 24 hours. Do not refreeze thawed breastmilk.
BT-150 Opeartion Manual
21
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Place
Temperature
How long
Things to Know
Room temperature
(60℉ to 85℉/16℃ to 29℃)
Refrigerator
(39℉/4℃ or colder)
Thawed
breastmilk
Up to 1 to 2 hours is best. Up to 3 to 4
hours is okay.
Do not
refreeze.
Guide to storing fresh breastmilk for use with healthy full-term infants
Countertop,
table
Refrigerator
Freezer
Room temp
(up to 77℉ (25℃))
39℉(4℃) or colder
0℉ (-18℃) or
colder
Up to 3 to 4 hours is
best.
Up to 6 to 8 hours is okay
for very clean expressed
milk.
Up to 3 days is best.
Up to 5 days is okay for
very clean expressed
milk.
Up to 3-6 months is best.
Up to 9 months is okay
for very clean expressed
milk.
Containers should be covered
and kept as cool as possible.
Cover the container with a clean
cool towel may keep milk cooler.
Throw out any leftover milk
within 1 to 2 hours after the baby
is finished feeding.
Store milk in the back of the main
body of the refrigerator. When at
work, you can place your
expressed milk in the refrigerator.
Use a canvas or insulated bag
that blends in with your
coworkers’ lunch bags, and place
it at the back of refrigerator.
Store milk toward the back of the
freezer where the temperature is
most constant. Milk stored at 0℉
or colder is safe for longer
durations, but the quality of the
milk might not be as high.
Deep
freezer
-4℉(-20℃) or
colder
Guide to storing thawed breastmilk
Up to 6 months.
Up to 12 months is okay
for very clean expressed
milk.
24 hours
Store milk toward the back of the
freezer where the temperature is
most constant. Milk stored at 0℉
or colder is safe for longer
durations, but the quality of the
milk might not be as high.
Any freezers
BT-150 Opeartion Manual
22
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
If you still have a problem with charging, replace the adapter with new
Wipe main body with dry cloth and store the main body in a warm, dry
7. Trouble shooting
No or low
expression
Battery not
charged
• Check the power button.
• Ensure the funnel, diaphragm assembly and bottle assembled correctly.
• Inspect all components for tear, cracks or bending.
• Check the adapter connected correctly.
•
one.
Does not power on • Inspect the power on/off button and case for any foreign materials.
• Unplug the power adapter from the socket.
Main body
becomes wet
• Turn off the main body.
•
place more than 12 hours.
• Turn off the main body.
Breastmilk in tube
• Disconnect the air tube from diaphragm top.
• Remove the milk from tube and diaphragm.
Milk flow in
reverse direction
up to Main Body
※
If you have not resolved the problem with breast pump or you have further questions, please contact Bistos
customer center.
• Stop using the product immediately..
• Contact service center.
BT-150 Opeartion Manual
23
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
etic environment specified
should assure that it is used in
Emissions test
Compliance
Electromagnetic environment-guidance
The electric breast pump BT-150 uses RF energy only
RF emissions
CISPR 11
Harmonic emission
IEC61000-3-2
The electric breast pump BT-150 should not be used adjacent to or stacked with other equipment.
150 electric breast pump should be observed to
Portable RF communications equipment (including peripherals such as antenna cables and external
, including
he manufacturer. Otherwise, degradation of the performance of this
or provided by the manufacturer of this equipment could
results in increased electromagnetic emissions or decreased electromagnetic immunity of this
8. Manufacturer’s declaration on EMC
The electric breast pump Hi bebe
(Electromagnetic compatibility) and needs to be used according to the EMC information provided in
this user manual. Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect the electric breast pump
and should be kept at least 1 m away from the equipment. And it does not suitable for use in an MRI
environment.
The set pressure value and operating mode should be unchanged during EMC testing. It is the essential
performance of BT-150 and checked during EMC testing.
super
BT-150 needs special precautions regarding EMC
8.1 Electromagnetic emissions
The electric breast pump BT-150 is intended for use in the electromagn
below. The customer or the user of the electric breast pump BT-150
such an environment.
RF emissions
CISPR 11
Group 1
for its internal function. Therefore, its RF emissions
are very low and are not likely to cause interference
in nearby electronic equipment.
Class B
The electric breast pump BT-150 is suitable for use in
Class A
all establishments, including domestic and those
directly connected to the public low-voltage power
Voltage fulctuations /flicker
emissions
Complies
supply network that supplies buildings used for
domestic purposes.
IEC61000-3-3
WARNING
If adjacent or stacked use is necessary, the BTverify normal operation in the configuration in which it will be used.
antennas) should be used no closer than 30 cm (12 inches) to any part of the BT-150
cables specified by t
equipment could result.
Use of adapter other than those specified
equipment and result in improper operation.
BT-150 Opeartion Manual
24
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
can help prevent electromagnetic interference by maintaining a minimum distance between portable
150 as recommended
150 kHz ~ 80 MHz
d = 1.2P
80 MHz ~ 800 MHz
d = 1.2P
800 MHz ~ 2.5 GHz
d = 2.3P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters at a maximum output power not listed above, the recommended separation
etic environment specified
should assure that it is used in
Electromagnetic
environment-guidance
contact discharge
±8kV, ±15kV
contact discharge
±8kV, ±15kV
Floors should be wood,
least 30%.
Mains power quality should
environment.
Mains power quality should
environment.
8.2 Recommended separation distances between portable and mobile RF
communications equipment and the electric breast pump BT-150
The electric breast pump BT-150 is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the electric breast pump BT-150
and mobile RF communications equipment and the electric breast pump BTbelow, according to the maximum output power of the communications equipment.
Rated maximum output
Separation distance according to frequency of transmitter [m]
power of transmitter [W]
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where p is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
8.3 Electromagnetic immunity
The electric breast pump BT-150 is intended for use in the electromagn
below. The customer or the user of the electric breast pump BT-150
such an environment.
Immunity test Test level Compliance level
Electrostatic
discharge(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
±8kV
air discharge:
±2kV, ±4kV,
± 2 kV
100 kHz repetition
frequency
± 0,5 kV , ± 1 kV
Line-to-line
±8kV
air discharge:
±2kV, ±4kV,
± 2 kV
100 kHz repetition
frequency
± 0,5 kV , ± 1 kV
Line-to-line
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
be that of a typical
commercial or domestic
be that of a typical
commercial or domestic
BT-150 Opeartion Manual
25
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
0 % UT ; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°, and
315°
0% UT ; 1 cycle and
70% UT ; 25/30 cycles
Single Phase; at 0°
0% UT ; 250/300 cycles
Single Phase; at 0°
0 % UT ; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°, and
315°
0% UT ; 1 cycle and
70% UT ; 25/30 cycles
Single Phase; at 0°
0% UT ; 250/300 cycles
Single Phase; at 0°
Mains power quality should
be that of a typical
commercial or domestic
environment.
If the user of the electric
breast pump BT-150 requires
continued operation during
power mains interruptions, it
is recommended that the
electric breast pump BT-150 is
powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60 ㎐)
30 A/m, 50 or 60 Hz 30 A/m, 50 or 60 Hz
magnetic field
IEC61000-4-8
Power frequency magnetic
fields should be at levels
characteristic of a typical
commercial or domestic
BT-150 Opeartion Manual
26
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Immunity test
Test level
Compliance level
Electromagnetic environment-guidance
Portable and mobile RF communications
MHz
MHz
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Excessive ambient EM (Electromagnetic) DISTURBANCES can cause the pressure of the unit to be
3 Vrms
150 kHz to
80 MHz
Conducted RF
IEC 61000-4-6
6 Vrms in ISM
bands between
0.15 MHz and
80 MHz
10 V/m
80 MHz to
Radiated RF
IEC 61000-4-3
2.7 GHz 80%,
1 kHz AM
RF Wireless Com
m.
3 Vrms
150 kHz to
80 MHz
6 Vrms in ISM
bands between
0.15 MHz and
80 MHz
10 V/m
80 MHz to
2.7 GHz 80%,
1 kHz AM
RF Wireless Com
m.
equipment should be used no closer to any
part of the electric breast pump BT-150,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1.2P
d = 1.2P 80 MHz ~ 800 MHz
d = 2.3P 800 MHz ~ 2.7 MHz
where p is the maximum output power rating
of the transmitter (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters(m)
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
, should be less than the compliance level in
each frequency range b.
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1) At 80
NOTE 2)These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the electric breast
pump BT-150 is used exceeds the applicable RF compliance level above, the electric breast pump BT-150
should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the electric breast pump BT-150.
b
and 800
, the higher frequency range applies.
WARNING
temporarily excessively high or low. Please use in environments below the above test standards.
BT-150 Opeartion Manual
27
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Functional Characteristics
Type
6 cm diameter Mono LCD
Pressure Level
Numeric Display
Operating Time
Numeric Display (min:sec)
Battery Level
3 levels
expression mode
massage mode
Auto Shut Off
30 minutes
8 Program IDs.
Backflow protection
Power
The certified adaptor in accordance with IEC
should be used for electric
Output: DC 12V/2.0A
Main body
Input: DC12V/1.3A
7.4 V Li-ion Polymer 2200 mA
Operating Time: 120 min
Charging Time: 150 min (100%)
Maximum charge-discharge cycles: 300 cycles
Standard Configuration
Operation manual
1ea
Power Adaptor
1ea
Funnel & Bottle Set
1ea(default: 24 mm)
Equipment Description : Bluetooth LE Module
Effective Radiated Power (ERP) : 4.06 dBm EIRP
9. Technical specifications
Type Electric Breast Pump
Pumping pressure(Vacuum) MAX. - 290mmHg
Display
Mode
- 16 pressure levels
- 6 cycle levels
Function
Program
- 16 pressure levels
- 3 cycle levels
8 Program Sequences for each IDs
60601-1 or IEC 60950-1
Adaptor
Rechargeable Battery
safety.
Input: AC 100 ~ 240V (50/60 Hz)
Options
Nursing Night Lamp
Wireless Interface
3 step brightness control
Frequency Band of Transmission :
2402~2480MHz(2MHz Separation, 40 Channel)
Frequency Characteristics of the Modulation : GFSK
BT-150 Opeartion Manual
28
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Available devices : CISPR 32 Class-A or higher certified
RF RSE : EN 300 328 v2.1.1
Carrying Bag
1ea
Funnel
21 mm/24mm/27 mm/32mm
Funnel & Bottle Set (for Dual)
Default 24mm funnel
Physical Characteristics
Main Unit
190(W) x 190(D) x 79(H)mm
Packing
238(W) x 156(D) x 181(H)mm
Carton Box(10ea)
805(W) x 255(D) x 385(H)mm
Main Unit
1.1Kg
Packing
1.7Kg
Carton Box(10ea)
10Kg
Atmospheric Pressure for
Storage Temperature
–20 to 60°C (–4 ~ 140°F)
Storage Humidity
0 ~ 95% non-condensing
Atmospheric Pressure for
Performance
Characteristics
Suction cycle
(cycles/min)
Vacuum level (negative mmHg)
Single pumping
Double pumping
Massage mode
(at level 1)
Massage mode
(at level 16)
Expression mode
(at level 1)
High expression
(at level 16)
Dimension
Weight
smart phone
Certification : RF-EMC : EN 301 489-1, 17
RED : TEC
FCC/IC : FCC Part 15 Subpart B
Environmental Conditions
Operating Temperature 10 to 40°C (50 ~ 104°F)
Operating Humidity 5 ~ 85% non-condensing
operating
storage
70/80/90 50 30
70/80/90 230 150
35/40/45/50/55/60 50 30
80kPa ~ 106kpa
70kPa ~ 106kpa
35/40/45/50/55/60 250 200
BT-150 Opeartion Manual
29
P/N: 150-ENG-OPM-EUR-R00
Bistos Co., Ltd.
2019,04
Product Name
Electric Breast Pump
Brand name
Hi bebe
super
Model Name
BT-150
Serial No.
Main body and Adapter: 1 year from the date of purchase
(excluding batteries consumed over time)
Date of Purchase
Name:
Telephone:
Sales Agency
Manufacture
Bistos Co., Ltd.
Product Warranty
Warranty Period
Customer
Battery: 6 months from the date of purchase
Address:
※Thank you for purchasing Hi bebe
super
BT-150.
※ This product is manufactured and passed through strict quality control and inspection.
※ Compensation standard concerning repair, replacement, refund of the product complies with
“Framework Act on Consumers” noticed by Fair Trade Commission of Republic of Korea.
Service Telephone and Fax. Numbers
Telephone: +82 31 750 0340
Fax.: +82 31 750 0344
th
7