Bird Technologies 73-67-25 User Manual

Y
OU'RE HEARD, LOUD AND CLEAR.
Installation and Operation Manual for
T-Pass® Transmit Multicouplers
73-67-11 Series and
73-67-25/25 (High Power) Series
Manual Part Number
7-9120
Warranty
This warranty applies for five years from shipping date.
TX RX Systems Inc. warrants its products to be free from defect in material and workmanship at the time of shipment.
Our obligation under warranty is limited to replacement or repair, at our option, of any such products that shall have been defective at the time of manufacture. TX RX Systems Inc. reserves the right to replace with merchandise of equal performance although not identical in every way to that originally sold. TX RX Systems Inc. is not liable for dam- age caused by lightning or other natural disasters. No product will be accepted for repair or replacement without our prior written approval. The purchaser must prepay all shipping charges on returned products. TX RX Systems Inc. shall in no event be liable for consequential damages, installation costs or expense of any nature resulting from the purchase or use of products, whether or not they are used in accordance with instructions. This warranty is in lieu of all other warranties, either expressed or implied, including any implied warranty or merchantability of fitness. No repre­sentative is authorized to assume for TX RX Systems Inc. any other liability or warranty than set forth above in con­nection with our products or services.
TERMS AND CONDITIONS OF SALE
PRICES AND TERMS:
Prices are FOB seller’s plant in Angola, NY domestic packaging only, and are subject to change without notice. Fed­eral, State and local sales or excise taxes are not included in prices. When Net 30 terms are applicable, payment is due within 30 days of invoice date. All orders are subject to a $100.00 net minimum.
QUOTATIONS:
Only written quotations are valid.
ACCEPTANCE OF ORDERS:
Acceptance of orders is valid only when so acknowledged in writing by the seller.
SHIPPING:
Unless otherwise agreed at the time the order is placed, seller reserves the right to make partial shipments for which payment shall be made in accordance with seller’s stated terms. Shipments are made with transportation charges col­lect unless otherwise specified by the buyer. Seller’s best judgement will be used in routing, except that buyer’s routing is used where practicable. The seller is not responsible for selection of most economical or timeliest routing.
CLAIMS:
All claims for damage or loss in transit must be made promptly by the buyer against the carrier. All claims for shortages must be made within 30 days after date of shipment of material from the seller’s plant.
SPECIFICATION CHANGES OR MODIFICATIONS:
All designs and specifications of seller’s products are subject to change without notice provided the changes or modifi­cations do not affect performance.
RETURN MATERIAL:
Product or material may be returned for credit only after written authorization from the seller, as to which seller shall have sole discretion. In the event of such authorization, credit given shall not exceed 80 percent of the original pur­chase. In no case will Seller authorize return of material more than 90 days after shipment from Seller’s plant. Credit for returned material is issued by the Seller only to the original purchaser.
ORDER CANCELLATION OR ALTERATION:
Cancellation or alteration of acknowledged orders by the buyer will be accepted only on terms that protect the seller against loss.
NON WARRANTY REPAIRS AND RETURN WORK:
Consult seller’s plant for pricing. Buyer must prepay all transportation charges to seller’s plant. Standard shipping pol­icy set forth above shall apply with respect to return shipment from TX RX Systems Inc. to buyer.
DISCLAIMER
Product part numbering in photographs and drawings is accurate at time of printing. Part number labels on TX RX products supersede part numbers given within this manual. Information is subject to change without notice.
Bird Technologies Group TX RX Systems Inc.
Manual Part Number 7-9120
Copyright © 2010 TX RX Systems, Inc.
First Printing: January 1994
Version Number Version Date
1 04/01/94
2 08/02/94
3 03/27/96
4 11/30/06
5 07/26/10
Symbols Commonly Used
WARNING
CAUTION or ATTENTION
High Voltage
Heavy Lifting
ESD Electrostatic Discharge
Hot Surface
Electrical Shock Hazard
NOTE
Important Information
Bird Technologies Group TX RX Systems Inc.
Changes to this Manual
We have made every effort to ensure this manual is accurate. If you discover any errors, or if you have suggestions for improving this manual, please send your comments to our Angola, New York facility to the attention of the Technical Publications Department. This manual may be periodically updated. When inquiring about updates to this manual refer to the manual part number and revision number on the revision page following the front cover.
Contact Information
Sales Support at 716-217-3113
Customer Service at 716-217-3144
Technical Publications at 716-549-4700 extension 5019
Bird Technologies Group TX RX Systems Inc.

Table of Contents

General Description ........................................................................................... 1
Unpacking ............................................................................................................ 4
Installation Overview........................................................................................... 5
RF Cables and Connectors ................................................................................. 5
Intermodulation Considerations........................................................................... 5
General Installation Procedure ............................................................................ 6
Transmitter Combiner Checkout........................................................................ 7
Required Equipment ............................................................................................ 7
Procedure .......................................................................................................... 7
Measurement Accuracy..................................................................................... 7
General Tuning Information ............................................................................... 7
Tuning Specifics .................................................................................................. 8
Fine Cavity Tuning............................................................................................. 8
Procedure ........................................................................................................ 8
Cavity Tuning Tip............................................................................................. 9
Coarse Cavity Tuning ........................................................................................ 9
Procedure ...................................................................................................... 10
Retuning System to all new Frequencies .......................................................... 10
Combiner Expansion .........................................................................................11
Typical Expansion Channel Installation ............................................................. 11
Peg Rack Procedure (6.625” and 10” Cavities) ............................................... 11
Relay Rack Procedure (10” Cavities) .............................................................. 12
Relay Rack Procedure (6.625” Cavities) ......................................................... 13
Setting Cavity Insertion Loss ............................................................................. 14
Cavity Loss Setting Procedure 1 ....................................................................... 15
Required Test Equipment................................................................................ 15
Procedure for T-Pass Loop ............................................................................. 16
Procedure for BandPass Loop ........................................................................ 17
Cavity Loss Setting Procedure 2 ....................................................................... 18
Required Test Equipment................................................................................ 18
Procedure for T-Pass Loop ............................................................................. 18
Procedure for BandPass Loop ........................................................................ 20
Maintenance ....................................................................................................... 21
Isolators.............................................................................................................. 21
Table of Contents Manual 7-9120-5 07/26/10
Figures and Tables
Figure 1: Block diagram of typical system ............................................................ 1
Figure 2: Noise suppression graph for 6.625” cavities .......................................... 2
Figure 3: Noise suppression graph for 10” cavities ............................................... 3
Figure 4: Typical Peg rack model .......................................................................... 4
Figure 5: Typical 19” Relay rack model ................................................................. 4
Figure 6: Typical combiner installation .................................................................. 5
Figure 7: Measuring T-Pass channel performance ............................................... 6
Figure 8: T-Pass cavity fine tuning ........................................................................ 8
Figure 9: T-Pass cavity tuning controls .................................................................9
Figure 10: Coarse tuning a T-Pass cavity ........................................................... 10
Figure 11: Peg rack mounting details .................................................................. 12
Figure 12: Relay rack mounting details ............................................................... 13
Figure 13: Top view of T-Pass cavity .................................................................. 14
Figure 14: Setting loop adjustment reference...................................................... 16
Figure 15: Setting T-Pass loop using step attenuators........................................ 17
Figure 16: Setting BandPass loop using step attenuators................................... 18
Figure 17: Setting T-Pass loop insertion loss ...................................................... 19
Figure 18: Setting Bandpass loop insertion loss .................................................20
Table 1: General specifications 6.625” cavity systems.......................................... 2
Table 2: General specifications for 10” cavity systems ......................................... 3
Table 3: Cavity insertion loss reference loop settings ......................................... 15
Appendix A
UHF Isolators (Compact style)
General Description .......................................................................................... 22
Installation.......................................................................................................... 22
Verifying Isolator Functionality ........................................................................ 22
Recommended Test Equipment ...................................................................... 22
Measuring Reverse Isolation (S12) ................................................................... 22
Measuring Insertion loss (S21) .......................................................................... 23
Figure A1: Verifying Reverse Isolation ...............................................................23
Figure A2: Verifying Insertion Loss .................................................................... 23
Figure A3: Typical Reverse Isolation Waveform ................................................ 24
Figure A4: Typical Insertion Loss Waveform...................................................... 24
Table A1: Specifications .................................................................................... 25
Table of Contents Manual 7-9120-5 07/26/10

GENERAL DESCRIPTION

T
T
T
T
T
The model 73-67-11/25-XX-NN Series T-Pass Transmit Combiners are designed to connect multi­ple transmitters to a common antenna. They use three-port bandpass filters (called T-Pass cavities) and ferrite isolators to provide low channel inser­tion loss, high isolation between transmitters, high antenna-to-transmitter isolation, high intermodula­tion suppression, and excellent transmitter noise suppression. T-Pass transmit combiners are broadband and easily adaptable to the most diffi­cult duplex system design requirements.
Transmitter Combiner (T-Pass)
X5
The block diagram of a typical transmit combiner is shown in Figure 1. The T-Pass filter passes one
X4
narrow band of frequencies and attenuates all oth­ers with increasing attenuation above and below the pass frequency. The T-Pass filter has a “dual­port” output loop plate which allows the filter to be easily connected to other T-Pass filters. Connec­tions between the filters are made with a “thru-line” cable that behaves like a low loss 50 Ohm trans­mission line. The thru-line cables are individually optimized to their own channel frequency. No com-
X3
promises are necessary to accommodate other channel frequencies. Each channel can therefore be anywhere in a very broad frequency range.
An isolator is added at the input to each T-pass channel to increase channel isolation. The ferrite isolators will isolate the transmitter from unwanted signals that enter the system via the antenna. The
X2
transmitter sees an excellent impedance match on its output, because the isolator absorbs reflected power that would otherwise enter the transmitters output stage. This improves the stability, spectral purity and long-term reliability of the transmitter.
The model 73-67-11/25-XX-NN Series T-Pass transmit combiners are available with either 6.625” or 10” cavities. TX combiners constructed with
X1
6.625” cavities are ideal for operation at channel separations of 115 KHz or more, with 110 to 150 Watt transmitters. These models are suitable for 19” relay rack mounting or TX RX peg rack mount­ing. TX combiners constructed with high perfor-
S
mance 10” diameter cavities, which have higher selectivity and power handling capabilities, allow operation at 75 KHz minimum separation with 125 to 150 Watt transmitters. In addition, High power 10” diameter models are also available which con­tain 250 Watt dual isolators with 100 or 250 Watt loads.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 1
Figure 1: Block diagram of a typical TX T-Pass
combiner. Typical five channel system shown as an example.
Frequency Range (MHz) 406 - 512 MHz
Cavity Type and Diameter 3/4 wave, 6.625” (168 mm)
Maximum Continuous Transmit Power 150 Watts
Isolator Load Power (Continuous) 73-67-11-2B-nn: 5W/25W [Note 3] ; 73-67-11-2D-nn: 5W/100W
Minimum TX-TX Separation at Cavity Loss 215 KHz @ -1.5 dB ; 115 KHz @ -2.5 dB
Typical TX-TX Isolation at Minimum Separation (dB) 80 dB
Typical Antenna - TX Isolation (dB) 70 dB
Typical TX Noise Suppression depends on cavity loss
Nominal Impedance (Ohms) 50
Maximum Input Return Loss (VSWR) -20 dB (1.22:1)
Temperature Range (°C) -30 to +60
Connectors, Input and Antenna N
Mounting Peg Rack ®
Mounting Options
Maximum Channels / Rack 15 [Note 4]
Dimensions 65.25” x 24” x 36” (H x W x D) [Note 6] ; (1659 x 610 x 914 mm)
Weight - Basic - Single Channel [lb (Kg)] 73-67-11-2B-nn: 36 (16.3) ; 73-67-11-2D-nn: 37 (16.7)
Weight - Expansion Channel Assembly [lb (Kg)] 73-67-11-2B-nn: 15 (6.8) ; 73-67-11-2D-nn: 16 (7.2)
Note 1: -nn in model number represents the number of channels.
Note 3: Models available with 5W/60W loads. same specifications as 25W and 100W models, except load power. Note 4: -MC option reduces maximum number of channels to ten 10-inch or twelve 6.625-inch channels per rack.
Note 5: -LR systems are tuned and tested on customer frequencies, then disassembled for shipping.
Note 6: Rack depth with cavity tuning rods at maximum frequency. Rod travel is approximately 5.1” (130 mm).
Note 2: These specifications are applicable to 406 - 512 MHz models.
MC: 19” rackmount adapter plates, 17.5” high
LR: System supplied without Peg Rack ® [Notes 4,5]
Tab l e 1 : General specifications for 6.625” cavity system.
73-67-11-Series Systems
0
-5
-10
-15
-20
6.625" Diameter 3/4-Wave, Fo = 460 MHz
-25
-30
Attenuation (dB)
-35
IL = 1.0 dB IL = 1.5 dB IL = 2.0 dB
-40
IL = 2.5 dB IL = 3.0 dB
-45
-50
0.01 0.1 1
10
Offset from Fo (MHz)
Figure 2: Typical transmitter noise suppression using 6.625” cavities.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 2
Frequency Range (MHz) 406 - 512 MHz
Cavity Type and Diameter 3/4-wave, 10” (254 mm)
Maximum Continuous Transmit Power 150 Watt
Isolator Load Power (Continuous) 5W/25W [Note 3]
Minimum TX-TX Separation at Cavity Loss 150 KHz @ -1.5 dB ; 75 KHz @ -2.5 dB
Typical TX-TX Isolation at Minimum Separation (dB) 80 dB
Typical Antenna - TX Isolation (dB) 70 dB
Typical TX Noise Suppression Depends on cavity loss
Nominal Impedance (Ohms) 50
Maximum Input Return Loss (VSWR) -20 dB (1.22 : 1)
Temperature Range (°C) -30 to +60
Connectors, Input and Antenna N
Mounting Peg Rack ®
Mounting Options
Maximum Channels / Rack 12 [Note 4]
Dimensions 79.5” x 24” x 36” (H x W x D) [Note 6] ; (2019 x 610 x 914 mm)
Weight - Basic Single-Channel [lb (Kg)] 14 (18.6)
MC: 19” rackmount adapter plates, 17.5 “ high
LR: System supplied without Peg Rack ® [Notes 4,5]
Weight - Expansion Channel Assembly [lb (Kg)] 19 (8.6)
Note 1: -nn in model number represents the number of channels.
Note 3: Models available with 5W/60W loads. same specifications as 25W and 100W models, except load power. Note 4: -MC option reduces maximum number of channels to ten 10-inch or twelve 6.625-inch channels per rack.
Note 5: -LR systems are tuned and tested on customer frequencies, then disassembled for shipping.
Note 6: Rack depth with cavity tuning rods at maximum frequency. Rod travel is approximately 5.1” (130 mm).
Note 2: These specifications are applicable to 406 - 512 MHz models.
Table 2 : General specifications for 10” cavity systems.
73-67-25-Series Systems
0
-5
-10
-15
-20
-25
-30
Attenuation (dB)
-35
-40
-45
10" Diameter 3/4-Wave, Fo = 460 MHz
IL = 1.0 dB IL = 1.5 dB IL = 2.0 dB IL = 2.5 dB IL = 3.0 dB
-50
0.01 0.1 1
10
Offset from Fo (MHz)
Figure 3: Typical transmitter noise suppression using 10” cavities.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 3
The TX combiners can be expanded one channel at a time with factory-tuned, easy-to-install expan­sion channel assemblies. Expansion is usually accomplished without modifications to the existing system, and usually amounts to nothing more than placing a new channel assembly, or several, on top of the existing system. New channel frequencies can be above, below, or between existing channel frequencies.
The number of channels in the combiner is indi­cated by the last two digits of the model number in place of the NN designation. All of the information for both installation and expansion is included in this manual. The combiner is easy to install and has been factory tuned in most cases so that no
adjustments are necessary. The specifications for the 73-67-11/25-XX-NN family of T-Pass combin­ers are listed in Tables 1 and 2 for the 6.625” and 10” cavities respectively. The curves shown in Fig- ures 2 and 3 show the typical transmit noise sup­pression for the 6.625” and 10” cavity systems respectively. Noise suppression depends on the cavity’s loss setting. Figure 4 shows a typical Peg rack model and Figure 5 shows a typical 19” rack mount model.

UNPACKING

It is important to visually inspect the system com­ponents for any shipping damages as soon as pos­sible after taking delivery. It is the customers responsibility to file any necessary damage claims with the carrier.
Figure 4: Example of typical Peg-rack model using 10” cavities.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 4
Figure 5: Example of typical Relay-rack model
using 10” cavities.
The transmit combiner is a very rugged device and is well packed for damage-free shipping to any place in the world. However, a high impact during shipping can have a detrimental affect. A damaged shipping container is a sure sign of rough handling. The most easily damaged parts of the combiner are the tuning rods. These rods are marked where they exit from the locking nut with a dab of red var­nish or other color/type of paint. If this seal appears to be broken it may indicate that the system has been detuned in transit.

INSTALLATION OVERVIEW

The combiner should be located in a dry and level area, indoors. It is best if all transmitters are as equal in distance as possible from the combiner so that cable losses are the same for all channels. Figure 6 shows a suggested orientation for the equipment. Two points are important. First, a work area space should be left as illustrated so that the tuning controls are easy to access. This will facili­tate tuning when channel frequencies are changed. Secondly, space is needed when adding expansion channels. If there is a lack of space to access the side of the combiner, then plan to allow the rack to be moved into the indicated work area to facilitate adding channels. This will require some slack in the cables that connect to the station trans­mitters. Each transmitter connects to its respective channel through an ‘N style’ female connector on the isolator. We recommend using a high quality double shielded or semi flexible cable.
This system is designed for use with separate transmit and receive antennas. For best operation, the transmit and receive antennas should be sepa­rated vertically by 20 feet with little or no horizontal offset between antennas. Lesser separations can be used but with an increased risk of harmful inter­ference between radio systems. In most cases, it will be desirable to mount the receive antenna higher than the transmit antenna to maximize the talk-back range of low power portable radios.

RF Cables and Connectors

All connections to and from the combiner system should be made with double-shielded or semi-rigid heliax cable. High quality 'N' connectors that use either silver or gold plated contacts should be used.

Intermodulation Considerations

Following the previously mentioned antenna spac­ing recommendations will go a long way toward minimizing or eliminating intermodulation (IM) interference. IM is the result of a frequency mixing process that occurs when two or more RF signals are present simultaneously in the same circuitry where nonlinearities occur. Product frequencies generated have frequencies that are determined by relatively simple mathematical relationships such as F(im) = 2F1-F2 and are normally determined by doing a computer intermodulation analysis for the antenna site. These products can be generated in a corroded tower joint, metal-roofing, transmitter final amplifier or the receiver front-end.
T-Pass
Transmitter
Combiner
Radio
Cabinet
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 5
Radio
Cabinet
Work
Area
Figure 6: Typical combiner installation.
Radio
Cabinet
Radio
Cabinet
Both cavity filters and ferrite isolators isolate the transmitters connected to the combiner from one­another thus reducing intermodulation interference. However in all transmitter combiners, intermodula­tion products are reduced in strength but never completely eliminated. They have to be reduced by an amount to meet the federal Communications Commission,
43 + 10 Log(Power Out)
rule for spu­rious output reduction. Because of the limitations imposed by the tension and friction joints in con­nectors, IM products will be down 100 to 120 dB below carrier so they are still strong enough to cause problems if they fall on a near-by receiver frequency.
Transmitter
To avoid transmitter generated IM problems, do not put two channels on the same combiner that your IM software predicts will cause interference by generating either 3rd or 5th order IM products. Having at least two transmitter combiners allows maximum flexibility in dealing with transmitter gen­erated IM.

General Installation Procedure

1) Install the peg rack or relay rack in the radio equipment room.
2) Connect the transmitters and the transmitting antenna to the appropriate connectors on the cavities.
Wattmeter 2
T-Pass
Cavity Filter
50 Ohm
Load
Transmitter
Wattmeter 1
UG57
Male-Male
Transmitter
Single
or Dual
Section
Isolators
Single
or Dual
Section
Isolators
Single
or Dual
Section
Isolators
Channel 3
Channel 2
Channel 1
UG27 Elbow Connector
& UG57 Male-Male
Adaptor
This T-Pass Loop
requires a 3-1268
short circuit
connector
Figure 7: Equipment hookup for measuring T-Pass channel performance.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 6
3) Verify proper operation of each channel by measuring the power output for each individual channel.

TRANSMITTER COMBINER CHECKOUT

It is recommended that the performance of the transmitter combiner be checked initially and data recorded for future reference. This is done by mea­suring the input and output power of each channel and recording the data. Figure 7 shows the equip­ment hook up.

Required Equipment

If a power monitoring system is not installed along with the combiner, two Bird Model 43 thruline watt­meters or equivalent can be used. They should be equipped with elements for the frequency band of interest and rated for the expected transmitter power output. The use of two wattmeters elimi­nates errors that can occur from changing cable lengths. The measurements should only be done one channel at a time because most wattmeters cannot accurately measure the total power of two or more transmitters simultaneously. A pocket cal­culator with Log functions makes for easy calcula­tion of power loss in dB using this measured data.

PROCEDURE

Start with channel 1 and proceed to the next higher channels. The two wattmeters should be con­nected to the equipment as shown in figure 7. Note that the use of the elbow and/or male-male con­nectors allows the shortest connections and negli­gible hook up loss. Longer cable lengths will tend to increase measurement error.
It is important that the same wattmeters and watt­meter elements be used in the same position throughout the tests. The serial numbers of the wattmeters should be recorded for future refer­ence. Wattmeter elements may not have serial numbers so they need to be labeled or otherwise keyed to a specific wattmeter to assure repeatabil­ity of the measurements.

MEASUREMENT ACCURACY

The Bird thruline wattmeter has a measurement accuracy of +/- 5% of full scale. When using a 100 watt element in this meter, the measurement error can be as great as + or - 5 watts.
As an example of the actual dB loss readings that might be produced using the wattmeter method, consider a T-Pass channel that has a factory mea-
sured loss of 3.0 dB. We would expect that a 100 watt transmitter would produce 50 watts out of this channel but the actual wattmeter reading for the input power could measure as low as 95 watts to as high as 105 watts. The measured output power could run from 45 to 55 watts. It is possible that the output reading may be 5 watts low while the input reading is 5 watts high or just the opposite. These two extremes would yield the following dB loss val­ues:
For a Power Out (PO) of 45 watts and a Power Input (PI) of 105 watts: Loss (dB) = 10 Log10 (45/105) Loss (dB) = -3.7
For a PO of 55 watts and PI of 95 watts: Loss (dB) = 10 Log10 (55/95) Loss (dB) = -2.4
So the calculated loss for this channel can run from
-2.4 to -3.7dB and be acceptable considering the measurement error factor. The actual error could be much greater if a 250 watt element was used; the measured values could vary by as much as +/- 12.5 watts so using a wattmeter element with the smallest possible rating is important for accu­racy. Use of between series adapters or UHF type connectors for making connections to the wattme­ters, device under test or loads could make this error even worse due to the additional impedance mismatch that these connectors can cause.

GENERAL TUNING INFORMATION

T-Pass transmitter combiners are pre-tuned at the factory and usually require no adjustment. T-Pass expansion channels are also pretuned but may require fine tuning after being installed in an exist­ing system. Channels that are close in frequency (adjacent channels in the multicoupler) to the expansion channel may also benefit from fine tun­ing due to the slight interaction that occurs with the new channel. The procedures that follow may be used at any time to verify that any or all channels are properly tuned or to correct any misalignments. It is interesting to note that T-Pass filters, bandpass filters and cavity filters in general can act as imped­ance transformers as well as filters. It is for this reason that many field service personnel claim that they can always tune a filter or duplexer better than the factory. What isn't generally realized is that their tuning efforts are usually producing better impedance matching between transmitter and antenna which can be improved by the transform-
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 7
ing action of filters. Since the filters are usually
r s
r s
tuned using laboratory grade 50 ohm loads, the tuning adjustment that produces this improved match will be slightly different than the factory adjustment. While this tuning may produce slightly greater power output readings, it will rarely pro­duce any discernible change in system perfor­mance and may detune any notching circuitry contained in the cavities.
It is our recommendation that channel tuning only be attempted under the previously mentioned con­ditions or when it is suspected that the combiner has been tampered with or subjected to extreme shock in shipping or installation. This condition is indicated when the channel loss is in excess of that expected from actual measurement of power input and output.

Tuning Specifics

Tuning of the combiner consists of tuning the indi­vidual T-Pass channels. T-Pass channel tuning involves cavity filter tuning. The isolators are
broad band and do not require adjustments.
The procedures for tuning cavities follows.

FINE CAVITY TUNING

Figure 8 shows hookups which are suitable for fine tuning any channel under power while installed in
the combiner. The term fine tuning here refers to cavities that have already been tuned to frequency and may only require adjustment of the fine tuning control (+/- 50 KHz). The transmitter is used as a signal source and the cavity is adjusted for mini­mum reflected power.
Procedure
With the transmitter keyed, the cavity fine tuning control is adjusted (pushed in or out) to obtain a minimum meter reading. See Figure 9 for a detail of the cavity tuning controls. If a minimum meter reading is obtained with the fine tuning rod fully in or completely out, do the following:
1) Set the fine tuning rod so that about 1/2 of its length is inserted into the cavity.
2) Loosen the coarse tuning rod locking screw (5/ 32”/4mm Allen/Hex-key wrench required) and move the rod in or out slightly to obtain mini­mum meter reading. Small movements of the coarse tuning rod are facilitated by tapping the rod with the handle end of a screw driver while gently pushing or pulling the main tuning rod. Tighten the coarse tuning locking screw.
3) Adjust the fine tuning control for a minimum meter reading.
Fine
Tuning
Two Single Section
or One Dual Isolators
Transmitter
Input
50 Ohm
Termination
Figure 8: Using a wattmeter for T-Pass cavity fine tuning.
Output
Coarse
Tuning
Wattmeter
Output Section Termination
T-Pass
Cavity Filter
To Othe
Channel
To Othe
Channel
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 8
Coarse Tuning Rod
Fine Tuning Rod
Coarse Tuning Lock 10-32 Cap Screw
Cavity Resonator
Input/Output Port
Loop Plate Hold Down Screws
Loop Plate Assembly
Calibration Index
Figure 9: T-Pass cavity tuning controls details.
4) Tighten the fine tuning locking mechanism.

CAVITY TUNING TIP

When tuning a cavity that has been in service for some time it is not unusual to find the main tuning rod hard to move in or out. This occurs because TX RX uses techniques borrowed from microwave technology to provide large area contact surfaces on our tuning plungers. These silver plated sur­faces actually form a pressure weld that maintains excellent conductivity. This pressure weld devel­ops over time and must be broken to move the main tuning rod. This is easily accomplished by gently tapping the tuning rod with a plastic screw­driver handle or small hammer so that it moves into the cavity. The weld will be broken with no damage to the cavity.
When adjusting the coarse tuning rod, it is easy to put the cavity far off resonance and cause most of the transmitter power to be reflected back into the isolator output section load. This load should be capable of dissipating this power or damage could result. If in doubt about the loads capability, follow the coarse tuning procedure outlined below. It is based on the use of a tracking generator which avoids the need to consider power levels.
Input/Output Port
Loop Plate Assembly
Calibration Mark
Calibration Index
Fine Tuning Lock
Knurled Thumb Nut

COARSE CAVITY TUNING

When a T-Pass cavity frequency has to be changed by over 100 KHz, adjustment of the main tuning rod is required. Large frequency changes are more easily observed when using a tracking generator and a return loss bridge to give a swept display of the return loss curve. The return loss curve is a very precise indicator of T-Pass cavity tuning. The test equipment hookup for doing this is illustrated in Figure 10 and uses the following equipment or its equivalent;
1) Spectrum Analyzer that covers the frequencies of interest such as the Bird Technologies “Sig­nal Hawk ™”.
2) Signal generator capable of producing the fre­quencies of interest.
3) Eagle Return Loss Bridge (35 dB directivity). Model RLB150N3A.
4) Double shielded coaxial cable test leads (RG142 B\U or RG223/U).
5) 50 Ohm load with at least -35 dB return loss (1.10:1 VSWR).
6) 3-1268 short circuit connector.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 9
Spectrum Analyzer
Bird SignalHawk
Signal Generator
RLB - 150 Bridge
LOAD

PROCEDURE

1) Set the spectrum analyzer for the desired cen­ter frequency (display center) and vertical scale of 10 dB/div. Set the signal generator for the desired center frequency.
2) Connect the return loss bridge to the spectrum analyzer and signal generator as shown in fig­ure 10 but do not connect it to the cavity. Leave the test port (called the load port) on the bridge open.
3) Set up a 0 dB return loss reference display on the spectrum analyzer. Then connect the return loss bridge.
4) Loosen the fine tuning rod locking nut and set the fine tuning rod so that 1/2 its length is inserted into the cavity.
5) Loosen the main tuning rod locking screw and move the main tuning rod in or out to obtain maximum return loss at the desired frequency. Small movements of the main tuning rod are facilitated by tapping the rod with the handle end of a screw driver while gently pushing or pulling the main tuning rod.
6) Lock the Main and Fine tuning rods and recon­nect the cavity into the system. Use the previ­ously outlined fine tuning procedure to verify proper tuning under power.

Retuning System To All New Frequencies

50 Ohm Load
T-Pass
Cavity Filter
When retuning the combiner to all new frequencies perform the following procedure in a step-by-step fashion;
1) Determine new thruline cable lengths for the new channels and the specific stacking order in the rack. TX RX Systems Sales engineers will assist by making the calculations using their design software. Due to variations in coaxial cable characteristics and assembly techniques, factory supplied cables are recommended.
2) Use the coarse tuning procedure to tune each
3-1268 Short Circuit Connector
cavity channel to the new transmitter frequen­cies.
3) Connect the channels according to the new thru-line cable chart.
Figure 10: Coarse tuning a T-Pass cavity.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 10
4) Fine tune each channel using the fine tuning procedure starting with channel 1 and proceed­ing to the next higher channel. After tuning all channels, repeat this step a second time to ver­ify that their is no more channel interaction.
(1) Cavity and isolator mounting hardware.
(1) T-Pass thru-line cable.
(1) T-Pass thruline chart.
5) Verify channel losses if desired using the checkout procedure outlined previously.

Combiner Expansion

Expansion channels for your combiner may be ordered directly from TX RX Systems or its autho­rized representative. If you wish, a TX RX Systems engineer will help you select the right model and any required options.
The expansion channel and options are shipped with mounting instructions and a new T-Pass Thruline cable sheet which shows the exact mount­ing location of the new channel in the existing sys­tem. In most cases, this channel will be added directly to the next topmost position in the rack and the antenna connection will then move to this cav­ity. A new thruline cable will connect this channel to the existing cavities.
The system engineer may also advise that the cav­ity insertion loss on some of the existing channels needs to be changed in order to accommodate a new channel. This can be necessary when the new channel is much closer in frequency separation to existing channels than was previously encoun­tered. This usually means increasing the cavity loss for all close spaced channels which provides the increased selectivity required. Cavity insertion loss values are shown on the T-Pass Thruline cable sheet.

Typical Expansion Channel installation

The following text is a procedure for adding expan­sion channel components to a typical T-Pass Transmit Combiner system. Please keep in mind that if instructions are shipped with the expansion components they would supersede these proce­dures.
Typical Parts Included: (Quantity and Description)
(1) T-Pass Cavity Assembly.
(1) Dual or single isolator with load(s).
PEG RACK PROCEDURE (6.625” AND 10” CAVITIES)
1) Determine the location of the Expansion Chan­nel in the rack by consulting the new THRULINE cable chart.
2) Mount the cavity in the peg rack using two (2) stainless band clamps, refer to Figure 11.
3) Rotate the cavity body so that the connectors are oriented the same as those on the other cavities and that no cavity-end cap screws are preventing a flush fit with a mounting peg.
4) Tighten the cavity mounting clamps.
5) Attach the isolator mounting plate to the cavity using two (2) band clamps. Clamp screws should be positioned as shown in figure 11. Do not tighten the clamps.
6) Rotate the isolator mounting bracket so that the isolator is in the vertical plane as illustrated, forming a smooth line in relation to the other channels in the rack.
7) Due to the limited space, tightening may require the use of a 5/16" open end wrench. Tighten both clamps securely.
8) Connect the black isolator-to-cavity cable using a pair of cable pliers to tighten-up the connec­tors.
9) Connect the new channel to the combiner using the proper length T-Pass Thruline cable. Use a pair of cable pliers to tighten these connections.
The required length thruline cable and
NOTE
new cabling chart has either been fac­tory supplied or is to be determined and fabricated by the customer as determined at the time of order. Use T-Pass thruline design sheets sup­plied by the factory.
(1) Isolator to cavity interconnect cable.
10) If necessary, reset the cavity insertion loss of
adjacent channels as noted on the Thru-line
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 11
Figure 11: Peg-rack mounting details.
cable sheet. Follow the procedure outlined below under Setting Cavity Insertion Loss.
11) Fine tune the T-Pass cavity of the expansion channel according to the fine tuning procedure outlined earlier.
RELAY RACK PROCEDURE (10” CAVITIES)
Because of their width, 10” cavities
NOTE
are mounted in relay racks with a ver­tical orientation as shown in Figure
12.
1) Install the expansion cavities to the rack above
existing channels using four mounting screws. Make sure you leave sufficient space between the upper and lower cavity groups so that the tuning rods and interconnect cables do not interfere.
2) Connect the black isolator-to-cavity cable using a pair of cable pliers to tighten-up the connec­tors.
3) Connect the new channel to the combiner using the proper length T-Pass Thruline cable. Use a pair of cable pliers to tighten these con­nections.
The required length thruline cable and
NOTE
new cabling chart has either been fac­tory supplied or is to be determined and fabricated by the customer as determined at the time of order. Use T-Pass thruline design sheets sup­plied by the factory.
4) If necessary, reset the cavity insertion loss of adjacent channels as noted on the Thru-line cable sheet. Follow the procedure outlined below under Setting Cavity Insertion Loss
.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 12
Screw
Screw
Screw
Screw
Figure 12: Relay rack mounting details. Ten-inch cavities shown as an example.
5) Fine tune the T-Pass cavity of the expansion channel according to the fine tuning procedure outlined earlier.
RELAY RACK PROCEDURE (6.625” CAVITIES)
Because of their width, 6.625” cavities
NOTE
are mounted on relay racks in a hori-
1) Determine the location of the expansion chan­nel in the rack by consulting the new THRULINE cable chart.
2) If necessary install an empty cavity deck in the rack using 4 Phillips screws. If there is room on an already existing cavity deck then skip this step of the procedure.
zontal orientation on cavity deck plates.
3) Mount the cavity on the deck by laying the cav­ity onto the “V shaped” cavity bracket.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 13
4) Rotate the cavity body so that the connectors are oriented the same as those on the other cavities in the system. Secure the new cavity to the brackets using (2) stainless band clamps.
5) Tighten the cavity mounting band clamps.
6) Connect the black isolator-to-cavity cable using a pair of cable pliers to tighten-up the connec­tors.
7) Connect the new channel to the combiner using the proper length T-Pass Thruline cable. Use a pair of cable pliers to tighten these con­nections.
The required length thruline cable and
NOTE
new cabling chart has either been fac­tory supplied or is to be determined and fabricated by the customer as determined at the time of order. Use T-Pass thruline design sheets sup­plied by the factory.
8) If necessary, reset the cavity insertion loss of adjacent channels as noted on the Thru-line cable sheet. Follow the procedure outlined below under Setting Cavity Insertion Loss
.
9) Fine tune the T-Pass cavity of the expansion channel according to the fine tuning procedure outlined earlier.

Setting Cavity Insertion Loss

It is sometimes necessary to reset the insertion loss of a T-Pass cavity filter in order to change its selectivity. Increasing the loss will increase the cavity selectivity which may be necessary to accommodate more closely spaced channels.
observed across the tee. The depth of the rejection notch is directly related to the loop's coefficient of coupling.
The first procedure uses precision rotary attenua­tors, a signal generator and a RF millivolt meter to provide very accurate results. The actual loss set­ting obtained when this procedure is carefully fol­lowed will be within one tenth of a dB of the desired value and the return loss will be 20 dB (1.25:1) or better.
The second procedure uses a spectrum analyzer and a signal generator and produces slightly less accurate results. When this procedure is carefully followed, the loss settings will be within two tenths of a dB of the desired value and the return loss will usually be -15 dB (1.5:1 VSWR) or better. The advantage of this procedure is that it is much faster to do, does not require precision attenuators and will yield acceptable results in most cases.
Table 3 is a reference chart for setting T-Pass cav­ity loss with either procedure. The chart shows the desired cavity loss settings and the reference set­ting for both the T-Pass and bandpass loop assem­bly. The reference notch depth for a given loss is that which can be observed across a tee connector connected to either loop assembly.
Calibration
Mark
Loop Locking
Screws (6 places)
Changing the loss is accomplished by rotating the coupling loops to change the coefficient of cou­pling. Both loops are normally adjusted for a given insertion loss setting. Most T-Pass cavities have a calibration index label beside both loops that gives a relative indication of their settings, see Figure
13. In actual practice, these marks are not accurate
enough for setting different loss values consis­tently.
Two procedures are offered for setting the cavity
Bandpass
Loop
T-Pass
Loop
loss. Both procedures take advantage of the fact that when a tee connector is placed on a single
Figure 13: Top view of T-Pass cavity.
bandpass or T-Pass loop, a rejection notch can be
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 14
Cavity Loss (dB) Coupling Loop Type TXRX Part # Reference Notch Depth
1.0
1.5
2.0
2.5
3.0
T-Pass 3-3724 -9.2
Bandpass 2-0675 -12
T-Pass 3-3724 -7.4
Bandpass 2-0675 -10.2
T-Pass 3-3724 -5.6
Bandpass 2-0675 -8.8
T-Pass 3-3724 -4.4
Bandpass 2-0675 -8.0
T-Pass 3-3724 -3.6
Bandpass 2-0675 -7.2
Table 3: Cavity insertion loss reference loop settings.

Cavity Loss Setting Procedure 1

This procedure uses precision rotary attenuators, a signal generator and an RF millivolt meter.

REQUIRED TEST EQUIPMENT

1) Signal generator capable of producing a CW signal level of at least -10 dBm with variable output level capability at the frequency of inter­est.
2) An RF voltmeter with a 0.001 V (-50 dBm) scale and a 50 ohm input adapter. Helper Instruments RF millivolter used for this exam­ple.
3) Rotary Attenuators, 1@ 0-1 dB in 0.1 dB incre­ments. 1@ 0-10 dB in 1.0 dB increments. 1@ 0-70 dB in 10 dB increments. JFW Industries model 50BR-017.
4) Two 10 dB fixed attenuator pads with BNC connectors. JFW Industries model 50F-010.
5) UG-914/U, BNC(F) to BNC(F) union. TX RX Systems' part # 8-5805.
6) UG-28A/U, N(F), N(F), N(F) tee.
7) UG-57B/U, N(M)-N(M) coupling.
8) Two, UG-201A/U BNC(F)-N(M) adapter. TX RX Systems' part # 8-5814.
9) 50 ohm coaxial cable test leads with BNC male connectors (high quality cable).
A spectrum analyzer may be used in place of the RF voltmeter. However, the personnel doing the work should fully understand the procedure and understand the use of the analyzer for this applica­tion.
We have found it convenient to use test cables with BNC connectors. They allow for more convenient connection to test equipment and to small attenua-
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 15
Modulated
Signal Source RF Voltmeter
2
0 0 000 0 00
All cables are 50 Ohm
ZERO SET
coaxial. Double shielded
cables preferred.
1
4567
3
50 Ohm Adaptor
3
8
9
10
ZERO
SET
UG914/U
0.1 dB/Div. 1.0 dB/Div. 10 dB/Div.
Female-Female Connector
10 dB Attenuator Pads
Figure 14: Setting loop adjustment reference level.
tor pads. UG-201 BNC to N adapters are used when connections to N connectors are needed.

PROCEDURE FOR T-PASS LOOP

1) Set the signal generator at the desired fre­quency (within 1 MHz of operating frequency) and an output level of approximately -10 dBm. Set the rotary attenuators for the Reference Notch Depth Value shown in table 3 for the desired insertion loss.
2) Connect the test leads together through the female union, as shown in Figure 14, and adjust the range switch and the zero set on the voltmeter for a convenient reference level (
level of 2 on the 0 to 3 scale for example
) on the meter. The generator output level may also be adjusted slightly if necessary.
3) Remove the bandpass loop from the cavity and reinsert it, connector end first, back into the cav­ity and tighten all 3 screws securely. See Fig- ure 15.
4) Set all three attenuators for 0 dB but leave them in the circuit.
5) Connect a UG-28A/U Tee connector and UG­57B/U coupling to the T-Pass loop as shown in figure 15. Then connect the test leads as
Rotary Attenuators
Set to Loop Reference Settings
shown. Make sure to install the 3-1268 short cir­cuit connector from the top of the T-Pass rack.
6) Loosen the main tuning rod locking screw and slowly slide the tuning rod in or out to obtain a dip (minimum voltage) in the meter reading which indicates cavity resonance. Use the fine tuning control to maximize the dip (the fine tun­ing rod should not be full in or out which would indicate that slight adjustment of the main tun­ing rod is necessary). Note the meter reading.
7) If the meter reading is greater or less than the reference level from step 2, the T-Pass loop
A
rotation will have to be adjusted. If the meter reading is greater than the reference level, the loop will have to be rotated so that the calibra­tion mark on the loop points to a slightly higher number on the calibration index label. Con­versely, if the meter reading is less than the ref­erence, the loop will have to be rotated so that the index mark points to a slightly lower number on the calibration index. Loosen the three loop locking screws and rotate the loop so that the index mark is moved to the next higher or lower calibration tag number as needed and tighten the 3 locking screws. Note that tight screws are necessary for accuracy.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 16
Modulated
Signal Source RF Voltmeter
Short Circuit Connector 3-1268 from top of rack
10 dB Pad 10 dB Pad
Figure 15: Setting the T-Pass loop using step attenuators.
UG-28A/U
UG-57B/U
T-Pass Loop
50 Ohm Adaptor
0.1 dB/Div. 1.0 dB/Div. 10 dB/Div.
Rotary Attenuators
Set to Loop Reference Settings
Bandpass Loop turned upside down with connector inserted into cavity. Loop visible and screws tight.
8) Repeat steps 6 and 7 until the minimum meter reading is equal to the reference level from step
2. Rotation of loops will change the cavity fre­quency slightly.
9) The Bandpass loop should be reinstalled with the connector facing upward and the ground point circle oriented toward the center of the cavity as shown in Figure 16.
10) Remove the short circuit connector from the T-
Pass loop.

PROCEDURE FOR BANDPASS LOOP

1) Maintain the previous signal generator settings and set the rotary attenuators for the proper set­ting as shown in table 3 for the Bandpass Loop.
2) Connect the test leads together through the female union and adjust the range switch and the zero set on the voltmeter for a reference level (
A level of 2 on the 0 to 3 scale is conve-
nient
) on the meter. See figure 14. The genera­tor output level may also be adjusted slightly if convenient.
3) Set all three attenuators for 0 dB but leave them in the circuit.
4) Connect a UG-107 Tee and the UG-57B/U to the Bandpass loop as shown on figure 16. Then connect the test leads as shown. Make sure the short circuit connector has been removed from the T-Pass loop.
5) Loosen the main tuning rod locking screw and slowly slide tuning rod in or out to obtain a dip (minimum voltage) in the meter reading which indicates cavity resonance. Use the fine tuning control to maximize the dip (the fine tuning rod should not be full in or out which would indicate that slight adjustment of the main tuning is nec­essary). Note the meter reading.
6) If the meter reading is greater or less than the reference level from step 2, the bandpass loop rotation will have to be adjusted. If the meter reading is greater than the reference level, the loop will have to be rotated so that the calibra­tion mark on the loop, points to a slightly higher number on the calibration index label. Con­versely, if the meter reading is less than the ref erence, the loop will have to be rotated so that
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 17
Modulated
Signal Source RF Voltmeter
UG-28A/U
UG-57B/U
Bandpass Loop
50 Ohm Adaptor
10 dB Pad 10 dB Pad
Figure 16: Setting the bandpass loop using step attenuators.
the index mark points to a slightly lower number on the calibration index. Loosen the three loop locking screws and rotate the loop so that the index mark is moved to the next higher or lower calibration tag number as needed and tighten the 3 locking screws. Note that tight screws are necessary for accuracy.
7) Repeat steps 5 and 6 until the minimum meter reading is equal to the reference level from step
2. Rotation of loops will change the cavity fre­quency slightly.
0.1 dB/Div. 1.0 dB/Div. 10 dB/Div.
Rotary Attenuators
Set to Loop Reference Settings
Small Circle on Bandpass Loop indicates ground end of loop and should be oriented as shown.
Previously calibrated T-Pass Loop 3-1268 short circuit removed.
3) UG-914/U, BNC(F) to BNC(F) union. TX RX Systems' part # 8-5805.
4) UG-28A/U, N(F), N(F), N(F) tee.
5) UG-57B/U, N(M)-N(M) coupling.
6) Two, UG-201A/U BNC(F)-N(M) adapter. TX RX Systems' part # 8-5814.
7) 50 ohm coaxial cable test leads with BNC male connectors (high quality cable).
8) Make sure that all the loop locking screws are tight. The cavity loops are now set and the cav­ity should now be tuned to the desired fre­quency.
We have found it convenient to use test cables with BNC connectors. They allow for a more convenient connection to test equipment and small attenuator pads. UG-201 BNC to N adapters are used when connections to N connectors are needed.

Cavity Loss Setting Procedure 2

This procedure uses a spectrum analyzer and sig­nal generator.

PROCEDURE FOR T-PASS LOOP

1) Remove the screws that hold in the bandpass loop assembly; remove the assembly; invert it

REQUIRED TEST EQUIPMENT

1) Spectrum Analyzer and a signal generator.
and place it back into the cavity (see Figure
17). The coupling loop will be visible. Install and tighten the three locking screws.
2) Two 10 dB fixed attenuator pads with BNC connectors. JFW Industries model 50F-010.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 18
UG-28A/U
n
UG-57B/U
T-Pass Loop
Connector 3-1268
Spectrum Analyzer
Bird SignalHawk
Signal Generator
10 dB Pad10 dB Pad
Short Circuit
from top of rack
Bandpass Loop turned upside dow with connector inserted into cavity. Loop visible and screws tight.
Figure 17: Setting a T-Pass loop for specific cavity insertion loss.
2) Connect the test leads to the spectrum ana­lyzer; turn it on and let it warm up for at least 30 minutes.
3) Connect the 10 dB attenuator pads to the test leads. They will remain connected for all subse­quent measurements.
4) Note the Reference Notch Depth value for the T-Pass loop assembly to be adjusted from table
3.
5) Set the spectrum analyzer for the frequency of the channel of interest (within 1 MHz of actual operating frequency).
6) If the Reference Notch Depth is 8 dB or less then set the display for a vertical range of 2dB/ div otherwise set it for 10dB/div.
7) Temporarily connect the test leads from the spectrum analyzer together through a UG-914 BNC union to set the zero reference.
8) Connect a UG-28 tee and a UG-57 coupling to the T-Pass loop as shown in figure 17.
9) Connect the test leads from the spectrum ana­lyzer to the tee connector as shown in figure 17.
10) Adjust the cavities main tuning rod so that a
rejection notch appears in the center of the dis­play.
11) Loosen the three loop locking screws and
rotate the loop to obtain the reference notch depth from step 4. Tighten the T-Pass loop locking screws only. Note that the tightness of the locking screws affects the depth of the rejection notch slightly. It is usually necessary to rotate the loop for a notch depth that is
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 19
slightly less than the reference. The Notch
.
d
depth will tend to increase slightly as all three locking screws are tightened.
5) If the Reference Notch Depth is 8 dB or less then set the display for a vertical range of 2dB/ div otherwise set it for 10dB/div.
12) Remove the bandpass loop and place it back into the cavity with the connector-end up.

PROCEDURE FOR BANDPASS LOOP

1) The Bandpass loop should be installed with the
connector up and the ground point circle ori­ented toward the center of the cavity as shown in Figure 18.
2) Connect the test leads, with 10 dB pads
attached, to the spectrum analyzer; turn it on and let it warm up for at least 30 minutes if this has not been done.
3) Note the Reference Notch Depth value for the
Bandpass loop assembly to be adjusted from table 3.
4) Set The spectrum analyzer for the frequency of
the channel of interest (within 5 MHz of actual operating frequency).
UG-28A/U
6) Temporarily connect the test leads from the spectrum analyzer together through a UG-914 BNC union to set the zero reference. Make sure to use the 10 dB pads which should remain on the test cables for all measurements.
7) Connect a UG-28 tee and a UG-57 coupling to the bandpass loop as shown in figure 18.
8) Connect the test leads from the spectrum ana­lyzer to the tee connector as shown in figure 18.
9) Adjust the cavities main tuning rod so that a rejection notch appears in the center of the dis­play.
10) Loosen the three loop locking screws and
rotate the loop assembly to obtain the refer­ence notch depth from step 3. Note that the tightness of the locking screws affects the depth of the rejection notch slightly, it is usually
Signal Generator
UG-57B/U
Bandpass Loop
Spectrum Analyzer
Bird SignalHawk
Figure 18: Setting a bandpass loop for specific cavity insertion loss.
10 dB Pad 10 dB Pad
Small Circle on Bandpass Loop indicates ground end of loop and should be oriente as shown.
Previously calibrated T-Pass Loop 3-1268 short circuit removed
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 20
necessary to rotate the loop for a notch depth that is slightly less than the reference. The Notch depth will tend to increase slightly as all three locking screws are tightened.
11) Tighten all loop locking screws. The cavity loss is now set. The cavity will have to be tuned to its operating frequency following the proce­dures outlined earlier in this manual.

MAINTENANCE

Because T-Pass transmit combiners are com­posed of passive components, they will continue to operate without any maintenance for years and there is no recommended maintenance period. However, we do feel that it is wise to check com­biner performance by measuring channel loss peri­odically and this may be done at any convenient time along with other radio system maintenance.

ISOLATORS

Isolators perform two important functions. Their pri­mary function is to keep unwanted RF frequencies out of the transmitter so that intermodulation prod­ucts cannot be generated. Isolators have a sub-
stantial amount of reverse isolation. They also ensure that the transmitter never sees any signifi­cant reflected power so it will always operate with maximum stability at full-power output. Isolators prevent energy from getting into the transmitters output by dumping any RF energy entering the out­put of the isolator into a dummy load. The model 73-67-11/25-XX-NN series of T-pass transmit com­biners will use either single section or dual section isolators at the input to each T-pass channel.
Single-section isolators have one load port. A prop­erly sized load capable of dissipating the maximum expected reflected power that might be encoun­tered should be used. Dual section isolators have two load ports, one for each section. Although loads of equal power rating may be used for both ports, it is customary to use an output load capable of dissipating the maximum expected reflected power that might be encountered. A small load (5 watts) is usually factory installed on the first section of the isolator where high reflected power is not a factor. Refer to Appendix A for a further discus­sion of isolators.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 21
Appendix A
UHF Isolators (Compact Style)

GENERAL DESCRIPTION

Isolators perform two important functions. Their pri­mary function is to keep other RF frequencies out of the transmitter so that intermodulation products cannot be generated. Isolators have a substantial amount of reverse isolation. They also insure that the transmitter never sees any significant reflected power so it will always operate with maximum sta­bility at full-power output. Isolators prevent energy from getting into the transmitters output stage by dumping reflected RF energy entering the output of the isolator into a dummy load.
The UHF (compact style) isolators available from TXRX Systems are broad-band and do not require tuning. The isolators are available as either single section or dual section models. Dual section mod­els consist of two single sections mounted in the same case with a load permanently attached to the load port of the first section. Ta b l e A 1 lists the UHF isolators available from TX RX Systems along with their performance specifications.

INSTALLATION

The isolators can be mounted on most types of surfaces but should not be physically located where they will be exposed to moisture or very high humidity. TXRX Systems isolators are well shielded magnetically and may be mounted on steel cabinets or panels.
The isolators can get quite hot during operation. This can occur when an antenna system compo­nent fails causing high reflected power which is then dissipated by the isolator load. These loads can get hot enough to burn skin so use caution when servicing these systems.

VERIFYING ISOLATOR FUNCTIONALITY

If you suspect there may be a problem with an iso­lator you can verify the functionality of the device by measuring its reverse isolation and insertion
loss. It is important to electrically remove the isola­tor from the system before testing. This is easily accomplished by disconnecting the input and out­put cables.
WARNING: Do not make or break
cable connections to the isolator while the circuit is under transmit power. Shut down the transmitter before servicing.

RECOMMENDED TEST EQUIPMENT

The following equipment or it’s equivalent is rec­ommended when verifying isolator functionality.
1) Spectrum Analyzer. Bird Technologies Signal Hawk.
2) A pair of double shielded coaxial cable test leads (RG142 B/U or RG223/U).
3) 50 Ohm load with at least -35 dB return loss (1.10 : 1) VSWR.

Measuring Reverse Isolation (S12)

The reverse isolation of your isolator can be veri­fied by performing the following procedure in a step-by-step fashion.
1) Make sure the transmitter associated with the isolator is turned off.
2) Disconnect the input and output cable to the isolator.
3) Connect a spectrum analyzer and tracking gen­erator to the input and output ports of the isola­tor respectively, as shown in Figure A1.
4) Make sure that a 50 Ohm load is connected to the load port of the isolator. If you are testing the isolator on the bench make sure you con­nect a load. If you are testing the isolator while it is still mounted on the system rack/cabinet leave the existing load connected.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 22
5) Inject a test signal (-10 dBm) from the tracking generator into the output port of the isolator. The test signal should sweep across the operat­ing bandwidth of the isolator.
Measuring Insertion Loss (S21)
The insertion loss of your isolator can be verified by performing the following procedure in a step-by­step fashion.
6) Compare your displayed waveform against the example shown in Figure A3 as well as the specification listed in table A1.
Spectrum Analyzer
Bird SignalHawk
Tracking Generator
1) Make sure the transmitter associated with the isolator is turned off.
Spectrum Analyzer
Bird SignalHawk
Tracking Generator
50 Load
Figure A1: Verifying Reverse Isolation.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 23
Figure A2: Verifying Insertion Loss.
50 Load
2) Disconnect the input and output cable to the isolator.
3) Connect a tracking generator and spectrum analyzer to the input and output ports of the iso­lator respectively, as shown in Figure A2.
4) Make sure that a 50 Ohm load is connected to the load port of the isolator. If you are testing the isolator on the bench make sure you con­nect a load. If you are testing the isolator while it is still mounted on the system rack/cabinet leave the existing load connected.
5) Inject a test signal into the input of the isolator from the tracking generator which will sweep across the operating bandwidth of the isolator. The strength of the test signal should be -10 dBm.
6) Compare your displayed waveform against the example shown in Figure A4 and the specifica­tion listed in table A1.
Figure A3: Typical reverse isolation waveform.
Figure A4: Typical insertion loss waveform.
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 24
TXRX
Systems
Part #
3-8195P 295 - 400 18 0.60
3-23018P 390 - 420 25 0.25
3-23269P 380 - 400 25 0.25
3-8191P 406 - 430 25 0.35
3-22401P 430 - 450 25 0.35
3-8192P 450 - 470 25 0.35
3-8193P 470 - 490 25 0.35
3-8194P 490 - 512 25 0.35
3-8195-1P 510 - 530 25 0.35
3-8195PA 295 - 400 18 0.50
3-8191PA 406 - 430 25 0.25
3-22401PA 430 - 450 25 0.25
3-8192PA 450 - 470 25 0.25
3-8193PA 470 - 490 25 0.25
3-8194PA 490 - 512 25 0.25
3-8195-1PA 510 - 530 25 0.25
3-8195PB 295 - 400 18 0.60
3-8191PB 406 - 430 25 0.35
3-22401PB 430 - 450 25 0.35
3-8192PB 450 - 470 25 0.35
3-8193PB 470 - 490 25 0.35
3-8194PB 490 - 512 25 0.35
3-8195-1PB 510 - 530 25 0.35
3-20721P 300 - 350 46 0.44
3-10010P 350 - 400 46 0.44
3-23425P 380 - 400 50 0.44
3-8196P 406 - 430 50 0.44
3-8201P 430 - 450 50 0.44
3-8197P 450 - 470 50 0.44
3-8198P 470 - 490 50 0.44
3-8199P 490 - 512 50 0.44
3-8200P 512 - 530 50 0.44
Freq
Range
(MHz)
Isolation
(dB) (min)
Insertion
Loss
(dB) (max)
TXRX
Systems
Part #
3-20721PL 300 - 350 23 0.22
3-10010PL 350 - 400 23 0.22
3-23425PL 380 - 400 25 0.22
3-8196PL 406 - 430 25 0.22
3-8201PL 430 - 450 25 0.22
3-8197PL 450 - 470 25 0.22
3-8198PL 470 - 490 25 0.22
3-8199PL 490 - 512 25 0.22
3-8200PL 512 - 530 25 0.22
3-20721PLA 300 - 350 23 0.22
3-10010PLA 350 - 400 23 0.22
3-8196PLA 406 - 430 25 0.22
3-8201PLA 430 - 450 25 0.22
3-8197PLA 450 - 470 25 0.22
3-8198PLA 470 - 490 25 0.22
3-8199PLA 490 - 512 25 0.22
3-8200PLA 512 - 530 25 0.22
3-20721PLB 300 - 350 23 0.22
3-10010PLB 350 - 400 23 0.22
3-8196PLB 406 - 430 25 0.22
3-8201PLB 430 - 450 25 0.22
3-8197PLB 450 - 470 25 0.22
3-8198PLB 470 - 490 25 0.22
3-8199PLB 490 - 512 25 0.22
3-8200PLB 512 - 530 25 0.22
3-20721PLC 300 - 350 46 0.44
3-10010PLC 350 - 400 46 0.44
3-8196PLC 406 - 430 50 0.44
3-8201PLC 430 - 450 50 0.44
3-8197PLC 450 - 470 50 0.44
3-8198PLC 470 - 490 50 0.44
3-8199PLC 490 - 512 50 0.44
Freq
Range
(MHz)
Isolation
(dB) (min)
Insertion
Loss
(dB) (max)
3-8200PLC 512 - 530 50 0.44
Tabl e A 1: Specification for UHF Isolators (Compact Style).
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 25
TX RX Systems Inc. Manual 7-9120-5 07/26/10 Page 26
8625 Industrial Parkway, Angola, NY 14006 Tel: 716-549-4700 Fax: 716-549-4772 sales@birdrf.com www.bird-technologies.com
Loading...