The following are general safety precautions that are not necessarily related to
any specific part or procedure, and do not necessarily appear elsewhere in this
publication. These precautions must be thoroughly understood and apply to all
phases of operation and maintenance.
WARNING
Keep Away From Live Circuits
Operating Personnel must at all times observe general safety precautions.
Do not replace components or make adjustments to the inside of the test
equipment with the high voltage supply turned on. To avoid casualties,
always remove power.
WARNING
Shock Hazard
Do not attempt to remove the RF transmission line while RF power is present.
WARNING
Do Not Service Or Adjust Alone
Under no circumstances should any person reach into an enclosure for the
purpose of service or adjustment of equipment except in the presence of
someone who is capable of rendering aid.
WARNING
Safety Earth Ground
An uniterruptible earth safety ground must be supplied from the main
power source to test instruments. Grounding one conductor of a two
conductor power cable is not sufficient protection. Serious injury or death
can occur if this grounding is not properly supplied.
Safety Symbols
WARNING
Chemical Hazard
Dry cleaning solvents for cleaning parts may be potentially dangerous. Avoid
inhalation of fumes or prolonged contact with skin.
WARNING
Resuscitation
Personnel working with or near high voltages should be familiar with
modern methods of resuscitation.
WARNING
Remove Power
Observe general safety precautions. Do not open the instrument with the power on.
WARNING
Warning notes call attention to a procedure which, if not correctly
performed, could result in personal injury.
CAUTION
Caution notes call attention to a procedure which, if not correctly performed,
could result in damage to the instrument.
The caution symbol appears on the equipment indicating there is
important information in the instruction manual regarding that
particular area.
Note: Calls attention to supplemental information.
iii
Warning Statements
The following safety warnings appear in the text where there is danger to
operating and maintenance personnel and are repeated here for emphasis.
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied. Leaking RF energy is a potential health hazard.
Refer to page 5, 35, and 41.
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the
line voltage setting.
Refer to page 35, 38, 39, and 41.
WARNING
Use only Nickel Metal Hydride (NiMH) batteries that have a minimum
capacity of 4500 milliamper hours (mAh). Do not install batteries that are
not Nickel Metal Hydride. Do not install NiMH batteries that have less than
4500 mAh capacity. Failure to comply may result in damage to the batteries,
damage to the instrument, and injury to personnel from battery chemicals.
Refer to page 40.
WARNING
Heavy load. Do not attempt to lift unaided.
Caution Statements
Refer to page 41.
The following equipment cautions appear in the text whenever the equipment
is in danger of damage and are repeated here for emphasis.
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.
Refer to page 2, 41, 43, 45, and 49.
CAUTION
The Bird 4421 must be powered off when connecting or disconnecting the
power sensor from the power meter.
Refer to page 5 and 41.
CAUTION
Do not use the power sensor with a load VSWR greater then 2:1. Damage to
the power meter, power sensor, or the RF power source could occur.
Refer to page 5.
iv
CAUTION
Long-term storage of this instrument can affect battery performance and
reduce battery life. Do not store the instrument for long periods of time
without recharging the batteries (refer to page 40). Failure to comply may
result in reduced battery charge and shortened battery life.
Refer to page 7
CAUTION
During remote operation, periodically monitor the bus service request line.
Failure to detect a service request could result in equipment damage.
Refer to page 11 and 16.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
maintenance should not be attempted. Removal or disturbance of the power
sensor cover can result in cancellation of lifetime warranty.
Refer to page 35 and 41.
CAUTION
Failure to install the properly rated fuse may result in equipment damage or
nuisance failures.
Refer to page 39.
v
Safety Statements
USAGE
ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED
BY THE MANUFACTURER MAY IMPAIR THE INSTRUMENT’S
SAFETY PROTECTION.
USO
EL USO DE ESTE INSTRUMENTO DE MANERA NO
ESPECIFICADA POR EL FABRICANTE, PUEDE ANULAR LA
PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO.
BENUTZUNG
WIRD DAS GERÄT AUF ANDERE WEISE VERWENDET ALS VOM
HERSTELLER BESCHRIEBEN, KANN DIE GERÄTESICHERHEIT
BEEINTRÄCHTIGT WERDEN.
UTILISATION
TOUTE UTILISATION DE CET INSTRUMENT QUI N’EST PAS
EXPLICITEMENT PRÉVUE PAR LE FABRICANT PEUT
ENDOMMAGER LE DISPOSITIF DE PROTECTION DE
L’INSTRUMENT.
IMPRIEGO
QUALORA QUESTO STRUMENTO VENISSE UTILIZZATO IN
MODO DIVERSO DA COME SPECIFICATO DAL PRODUTTORE LA
PROZIONE DI SICUREZZA POTREBBE VENIRNE
COMPROMESSA.
vi
SERVICE
SERVICING INSTRUCTIONS ARE FOR USE BY SERVICETRAINED PERSONNEL ONLY. TO AVOID DANGEROUS
ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLESS
QUALIFIED TO DO SO.
SERVICIO
LAS INSTRUCCIONES DE SERVICIO SON PARA USO
EXCLUSIVO DEL PERSONAL DE SERVICIO CAPACITADO. PARA
EVITAR EL PELIGRO DE DESCARGAS ELÉCTRICAS, NO
REALICE NINGÚN SERVICIO A MENOS QUE ESTÉ CAPACITADO
PARA HACERIO.
WARTUNG
ANWEISUNGEN FÜR DIE WARTUNG DES GERÄTES GELTEN
NUR FÜR GESCHULTES FACHPERSONAL.
ZUR VERMEIDUNG GEFÄHRLICHE, ELEKTRISCHE SCHOCKS,
SID WARTUNGSARBEITEN AUSSCHLIEßLICH VON
QUALIFIZIERTEM SERVICEPERSONAL DURCHZUFÜHREN.
ENTRENTIEN
L’EMPLOI DES INSTRUCTIONS D’ENTRETIEN DOIT ÊTRE
RÉSERVÉ AU PERSONNEL FORMÉ AUX OPÉRATIONS
D’ENTRETIEN. POUR PRÉVENIR UN CHOC ÉLECTRIQUE
DANGEREUX, NE PAS EFFECTUER D’ENTRETIEN SI L’ON N’A
PAS ÉTÉ QUALIFIÉ POUR CE FAIRE.
ASSISTENZA TECNICA
LE ISTRUZIONI RELATIVE ALL’ASSISTENZA SONO PREVISTE
ESCLUSIVAMENTE PER IL PERSONALE OPPORTUNAMENTE
ADDESTRATO. PER EVITARE PERICOLOSE SCOSSE
ELETTRICHE NON EFFETTUARRE ALCUNA RIPARAZIONE A
MENO CHE QUALIFICATI A FARLA.
vii
UNITS ARE EQUIPPED WITH RECHAREABLE BATTERIES. THESE ARE TO
BE REPLACED BY AUTHORIZED SERVICE PERSONNEL ONLY!!!
LAS UNIDADES VIENEN EQUIPADAS CON BATERIAS RECARGABLES. ¡¡¡Y
SOLAMENTE EL PERSONAL DE SERVICIO AUTORIZADO PUEDE
REEMPLAZARLAS!!!
GERÄTE SIND MIT WIEDER AUFLADBAREN BATTERIEN BESTÜCKT.
BATTERIEN SIND NUR VON QUALIFIZIERTEM SERICE PERSONAL
AUSZUWECHSELN!!!
CES DISPOSITIFS SONT ÉQUIPÉS DE BATTERIES RECHARGEABLES.
SEUL LE PERSONNEL D’ENTRETIEN AUTORISÉ EST HABILITÉ À LES
REMPLACER !
LE UNITÀ SONO DOTATE DI BATTERIE RICARICABILI, CHE DEVONO DA
COME SPECIFICATO DAL PRODUTTORE LA PROTEZIONE DI SICUREZZA
POTREBBE VENIRNE COMPROMESSA.
viii
BE SURE THE 115/230V AC VOLTAGE SELECTOR IS SET TO THE PROPER
LINE VOLTAGE, AND THE CORRECT AC LINE FUSE IS INSTALLED BEFORE
AC POWER IS APPLIED.
S’ASSURER QUE LE SÉLECTEUR DE TENSION 115/230V C.A. EST BIEN
RÉGLÉ POUR LA TENSION DU RÉSEAU ET QUE LE FUSIBLE DE LIGNE
C.A. CORRECT EST EN PLACE AVANT DE METTRE SOUS TENSION C.A.
CERCIORESE QUE EL SELECTOR DE VOLTAJE DE 115/230V CA ESTE
COLOCADO A LA LINEA DE VOLTAJE APROPIADA Y QUE EL FUSIBLE
ESTE INSTALADO A LA LINEA CA ANTES DE APLICAR LA CORRIENTE
ALTERNA.
VOR EINSCHALTEN DER WECHSELSTROMZUFUHR SICHERSTELLEN,
DASS DER 115/230V WECHSELSPANNUNGS-SELEKTOR AUF DIE
VORSCHRIFTSMÄSSIGE LEITUNGSSPANNUNG EINGESTELLT UND DIE
RICHTIGE WECHSELSTROM-HAUPTSICHERUNG EINGESETZT IST.
PRIMA DI EROGARE CORRENTE, ASSICURARSI CHE IL SELETTORE DI
VOLTAGGIO 115/230 V.C.A. SIA REGOLATO CORRETTAMENTE E CHE IL
FUSIBLE ADATTO ALLA LINEA DI ALIMENTAZIONE C.A. SIA INSTALLATO.
ix
About This Manual
This manual covers the Bird 4421 RF Power Meter and the its sensors. This manual covers the operating and maintenance instructions for the following models:
We have made every effort to ensure this manual is accurate. If you should
discover any errors, or if you have suggestions for improving this manual,
please send your comments to our factory. This manual may be periodically
updated. When inquiring about updates to this manual refer to the part number and revision level on the title page.
Contents
Chapter Layout
Introduction - Describes the purpose and function of the Power Meter as well
as a general overview of the product.
Installation - Describes the installation instructions for the power meter models
covered in this manual.
Operating Instructions - Describes the features of the power meters and pro-
vides power-up instructions.
IEEE-488 GPIB Interface - Describes the features of the IEEE-488 GPIB Inter-
face.
RS-232 Interface - Describes the features of the RS-232 Interface.
Maintenance - Contains preventative maintenance information, troubleshoot-
ing section, and a list of replacement parts with part numbers.
VOLTAGE 115/230 V
FUSE RATING T630mA/T315mA
FREQUENCY 50/60 Hz
MAXIMUM POWER 60W
AC SUPPLY
ON
O OFF
I
POWER
SENSOR
REAR
6
7
8
9
101312
11
12-1/4” (312 mm)
4
5
SIDE
10-1/8” (257 mm)
4-1/4”
(108 mm)
12
3
FRONT
MODEL 4421 RF POWER METER
AUTO
RANGE
UP
DOWN
SWRRFLFWDMIN MAX dBm LIGHTON / OFF
1. Operating Push
Buttons
2. LCD
3. ON/OFF Push Button
4. Handle
5. Central Button
6. Fuse Access Door
7. AC Line Module
8. Master ON/OFF
Switch
9. GPIB connector
10. GPIB DIP switches
11. RS-232 connector
12. RS-232 DIP switches
13. Power Sensor Socket
Power Meter
The Bird 4421 RF Power Meter is one component of a complete RF power measurement system. An RF power sensor such as a Bird 4021 is also required.
The system can be controlled with the front panel buttons, or remotely
through an RS-232 connection or a GPIB-488 connection.
Items Supplied
z Bird 4421 RF Power Meter
z AC Power Cord
z Sensor Cable
z Instruction Manual
Optional Accessories
Panel Mount Kit (P/N 4421-250) - Allows the Bird 4421 to be installed in a
standard 19" panel for rack mount applications.
Null Modem Kit (P/N 4380-250) - Contains the hardware necessary to allow
the 4421 to be remotely controlled by controllers with different wiring
arrangements. Requires an RS-232 interface module.
Figure 1 Bird 4421 Meter Outline Drawing
1
Bird 4421 RF Power Meter
DIRECTIONAL POWER SENSOR
MODEL 4022
0.3 - 1000 WATTS
25 - 1000 MHz
3-3/4”
(95 mm)
3/4”
(19 mm)
1-3/8”
(35 mm)
2-1/2”
(64 mm)
3-1/4”
(83 mm)
-CAUTION-
SOURCELOAD
SENSOR PLUG
SOCKET
Power Sensors
4020 Series
4027A Series
Power sensors are available with a variety of connectors; see "Available Connectors" on page 49 for a complete list. Since the accuracy is critically dependent on the connectors used at calibration, do not remove or change the
connectors.
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.
Bird 4020 Series Power Sensors are designed for lab or field use and are accurate to within ±3%(1σ) of reading.
Bird 4027A Series Power Sensors are designed for use in semiconductor processing and calibration applications. Stringent calibration provides long-term
unit-to-unit repeatability, allowing consistent amounts of RF energy to be
applied to the etch process over many etch cycles. 4027A Sensors are accurate
to ±1%(1σ) at specified calibration frequencies and power levels.
4027F Series
4028 Series
Bird 4027F Series Power Sensors are similar to the 4027A series. However,
additional filtering allows the 4027F to ignore harmonics of the signal being
measured. The 4027F is also less sensitive to AM components of the signal.
4027F Sensors are accurate to ±1%(2σ) at specified calibration frequencies
and power levels.
Figure 2 Power Sensor Outline Drawing, 4020, 4027A, 4027F, & 4028A Series
Bird 4028 Series Power Sensors are high power sensors otherwise similar to
the 4027A series. 4028A sensors are based on a 7/8” line, 4028B sensors use a
1-5/8” line, and 4028C sensors use a 3-1/8” line. 4028 Sensors are accurate to
±2% at specified calibration frequencies and power levels.
2
Introduction
SENSOR PLUG
SOCKET
3.75"
(96 mm)
4.75"
(121 mm)
3.5"
(89 mm)
6.75"
(172 mm)
SOURCELOAD
Figure 3 Power Sensor Outline Drawing 4028B Series only
Figure 4 Power Sensor Outline Drawing 4028C Series only
This chapter provides information about preparing the Bird 4421 for use.
Sensor Connection
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.
CAUTION
The Bird 4421 must be powered off when connecting or disconnecting the
power sensor from the power meter.
1. Turn OFF the ON/OFF rocker switch on the meter’s rear panel.
2. Align the latch on the cable with the notch of the “Power Sensor” socket on
the power meters rear panel.
3. Insert the cable until it clicks into place.
4. Connect the other end of the sensor cable to the sensor.
RF Line Connection
Never attempt to connect or disconnect RF equipment from the transmission
line while RF power is being applied.
Leaking RF energy is a potential health hazard.
Do not use the power sensor with a load VSWR greater then 2:1. Damage to
the power meter, power sensor, or the RF power source could occur.
Connect the end of the power sensor labeled “SOURCE” to the RF source. Connect the end labeled “LOAD” to the load or antenna. Reversing these connections will cause measurement errors.
Panel Mounting the 4421 Power Meter
You can install the 4421 Power Meter in an equipment rack if you have the
optional panel mount kit (refer to Optional Accessories, page 1). The panel
mount kit includes complete installation instructions. Figure shows the overall dimensions and mounting points for a 4421 Power Meter installed in a
panel mount kit.
Note: The power supply interrupt switch for the 4421 Power Meter
is located on the rear panel. When you install a unit in a panel mount
kit, you need to provide a means to interrupt the power supply that is
easily accessible to the user (such as a switch mounted in the panel).
WARNING
CAUTION
5
Bird 4421 RF Power Meter
RF POWER METER
dBmMAX LIGHTSWRMINRFL
4421
FWD
SENSOR
POWER
178 - 264 VAC
VOLTAGE
100 - 132 VAC
AT 47.5 - 66 Hz
MAXIMUM POWER 15W
REMOVE AC POWER BEFORE ATTEMPTING
TO SERVICE THIS INSTRUMENT.
W A R N I N G
T .125A, 250V
T .250A, 250V
FUSE
AC SUPPLY
O OFF
FUSE
I ON
POWER
SENSOR
Figure 5 Panel Mounting Dimensions
Handle Operation
The handle on the Bird 4421 can be set to four different positions (see
Figure 6). To adjust the handle, press the center buttons on both sides.
Releasing the buttons will lock the handle into position.
Figure 6 Handle Positions
115/230V AC Input Power
The internal power supply can operate at 115 VAC or 230 VAC. There is no
user setting to select the input voltage because the power supply automatically senses the input voltage. However, you must make sure that you have
the correct power cord plug and fuses installed for the line voltage you are
using (refer to Replacing Fuses, page 39).
AC Line Connectors
To make the AC line cord compatible with non-U.S. voltages, users must
install the appropriate connector on the power cord.
6
Installation
Batteries
The Bird 4421 RF Power Meter is completely portable and is powered from
internal rechargable nickel metal hydride batteries.
The batteries are shipped in a low charge state. It is, therefore, recommended
that you chrage the instrument for approximately 28 hours before using it for
continous operation.
Note: Battery charging at temperatures greater than 45° C (113°
F) can result in reduced operational time.
CAUTION
Long-term storage of this instrument can affect battery performance and
reduce battery life. Do not store the instrument for long periods of time
without recharging the batteries (refer to page 40). Failure to comply may
result in reduced battery charge and shortened battery life.
7
Bird 4421 RF Power Meter
8