651 South Main Street
Middletown, CT 06457
Tel (860) 344-1079 (800) 711-6776
Fax (860) 344-1068
Version 2.3 30SEPT2005 p/n 13-063
Page 2
THE ZO NEGUARD SINGLE CHANNEL FIXED GAS DETECTION SYSTEM HAS
BEEN DESIGNED FOR THE DETECTION AND MEASUREMENT OF
POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS
IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF
POTENTIALLY DANGEROUS CONDITIONS, IT IS ESSENTIAL THAT THE
INSTRUCTIONS IN THIS MANUAL BE READ, FULLY UNDERSTOOD, AND
FOLLOWED.
ZoneGuard Operat ion Manual
Version 2.3
Part Number 13-063
Copyright 2005
by
Biosystems
651 South Main Street
Middletown, Connecticut 06457
All rights reserved.
No page or part of this operation manual may be reproduced in any form
without written permission of the copyright owner shown above.
Page 1
Page 3
Table of Contents
SIGNAL WORDS .............................................................................................................4
APPENDIX A ZONEGUARD FIXED GAS DE TECT ION SYSTEM ORDERING INFORMATION----------------------------------------20
APPENDIX B ZONEGUARD FIXED GAS DE TECT ION SYSTEM TECHNICAL SPECIFICATIONS------------------------------------21
APPENDIX C ZONEGUARD DEFAULT ALARM SETTINGS ---------------------------------------------------------------------------22
APPENDIX D ZONEGUARD SENSOR RANGES ---------------------------------------------------------------------------------------22
APPENDIX E ZONEGUARD TOXIC SENSOR CROSS SENSITIVITY DATA-----------------------------------------------------------22
APPENDIX F ZONEGUARD DIMENSIONAL DRAWINGS, ELECTRICAL SCHEMATICS, AND WIRING DIAGRAMS ----------------23
APPENDIX G CALIBRATION FREQUENCY --------------------------------------------------------------------------------------------26
BIOSYSTEMS STANDARD WARRANTY GAS DETECTION PRODUCTS....................27
Page 3
Page 5
SIGNAL WORDS
The following signal words, as defined by ANSI Z535.4-1998, are used in the Zoneguard Reference Manual.
indicates an imminently hazardous situation which, if not avoided, will result in
death or serious injury.
indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
indicates a potentially hazardous situation, which if not avoided, may result in
moderate or minor injury .
CAUTION used without the safety alert symbol indicates a potentially hazardous situation which, if
not avoided, may result in property damage.
WARNINGS
1.
dangerous atmospheric cond itions. An alarm condition indicates th e presence of a potentially life-threatening
hazard and sh ould be taken very seriously. Failure to immediately leave the area during an alarm condition may
result in serious injury or death.
2.
course of action is to immediately leave th e affected area, and to return only after further testing determines that the
area is once again safe for entry. Failure to immediately leave the area during an alarm condition may result in
serious injury or death.
3.
calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings.
4.
to contaminants by testing with known concentration test gas before further use. Failure to check accuracy can
lead to inaccurate an d potentially dangerous readings.
5.
immediately. An instrument that fails calibration may not be used until testing with kn own concentration test gas
determines that accuracy has been restored, and the instrument is once again fit for use.
6.
concent ration that differs from the one that is normally supplied by Biosystems for use in calibrating the
Zoneguard. Use of inappropriate calibration gas may lead to in accurate and potentially dan g erous readings.
Customers are strongly urged to use only Biosystems calibration materials when calibrating the Zoneguard. Use of
non-standard calibration gas and/or calibration kit components can lead to dangerously inaccurate readings and
may void the stan dard Biosystems warranty.
7.
Zoneguard can lead to inaccurate and potentially dangerous readings and may void the standard Biosystems
warranty.
Biosystems offers calibration kits and long-lasting cylinders of test gas specifically developed for easy Zoneguard
calibration. Customers are strongly urged to use only Biosystems calibration materials when calibrating the
Zoneguard.
8.
9.
Read and un d erstand this reference manual before operating or servicing the Zoneguard.
10.
combustible gas concen tration that exceeds the Zoneguard’s zero to 100 percent LEL detectionrange. The safest
course of action is to immediately leave th e affected area, and to return only after further testing determines that the
area is once again safe for entry.
The Zoneguard personal, portable gas detector has been designed for the detection of
In the event of an alarm condition it is important to follow established procedures. The safest
The accuracy of the Zoneguard should be checked periodically with known concentration
The accuracy of the Zoneguard should be checked immediately following any known exposu re
A sensor that cann ot be calibrated or is found to be out of tolerance should be replaced
Do not reset the calibration gas concentration unless you are using a calibration gas
Use of non-standard calibration gas and/or calibration kit components when calibrating the
Substitutio n of components may impair intrinsic safety.
For safety reasons this equipment must be operated and serviced by qualified personnel only.
A rapid up-scale reading followed by a declining or erratic reading may indicate a hazardous
Page 4
Page 6
CHAPTER 1. DESCRIPTION
1.1 ZoneGuard capabilities
The ZoneGuard is a microprocessor based fixed gas
detection system designed to accept the input from a single
oxygen, combustible gas or toxic sensor. ZoneGuard
systems are designed to provide continuous round-the-clock
monitoring for a variety of common atmospheric hazards
including ox ygen deficiency and/or enrichment, combustible
gases and vapors, and specific toxic contaminan ts such as
carbon monoxide (CO), hydrogen sulfide (H2S), sulfur
dioxide (SO2), chlorine (Cl2), chlorine dioxide (ClO2),
phosphine (PH3), and nitrogen dioxide (NO2). Sensors may
be integrally mounted to the ZoneGuard controller housing,
or located up to 50 feet away from the controller. Remote
detector assemblies are available in explosion-proof or nonexplosion-proof versions. (Combustible sensors may onl y
be mounted in explosion-proof sensor housings.) The unit is
designed for operation via 120VAC 60 Hz. external power
source.
Standard features in clude two independently adjustable
alarm setpoints , built-in loud audibl e alarm (85 dB at 3-feet),
easily -read, backlit LCD display, and bright LED alarm light
indicators for OK / power on (green), low alarm (amber) and
high alarm (red) conditions. Stan dard features also in clude
DPDT relay contacts (rated 5A at 250V non- inductiv e load)
for low and high alarm conditions . Optional accessories
include externally mounted high intensity strobe lights and
high intensity horn.
The ZoneGuard offers a choice of two modes of operation,
Text Only (“OK”), and Basic (numeric readings) to provide
the right amount of information for users with different skill
levels and monitoring requirements. Biosystems’ “One
Button” operating logic allows al l procedures necessary for
normal day-to-day operation, including fully automatic
calibration adjustment, to be made through the use of the
single On/Off “Mode” button.
ZoneGuard sensors which are installed in explosion-proof
remote detector housings are classified by Underwriters
Laboratories, Inc. and the Canadian Standards Association
as Explosion-Proof for use in Cl ass I, Division 1, Grou ps A,
B, C, D, E, F and G hazardous locations.
Each ZoneGuard system includes the controller with all
standard alarms, relays and electron ics installed in a NEMA
4X fiberglass housing; s ensor, mounting ears, calibration
adaptor, Reference Manual, Quick Reference Card, and
other standard accessories. Remote detector assemblies
include the remote sensor housing, the appropriate amplifier
board (installed in the remote detector housing), and sensor.
Note: Conduit, wire and other components necessary
to connect the remote detector assembly with the
instrument controller are not included and must be
ordered separately.
The ZoneGuard design includes many user selectable
features and is designed to be eas ily customized to meet
specific customer requi rements. This chapter discusses
some of these special features as well as basic capabilities
used on a day-to-day basis.
1.2 Method of sampling
Once it has been connected to a power source and placed
in normal operation, the ZoneGuard system continuously
monitors for the hazard the sen sor is designed to detect.
ZoneGuard sensors may be either integrally mounted to the
controller housing, or remotely located u p to 50 feet away
from the controller. The atmosphere being measured
reaches the sensor by diffusing through a protective filter
into th e sensor. Normal air movements are enough to car r y
the sample to the sensor. The sensor r e acts to changes in
the concentration of the hazard being measured. Values
are constantly updated and di splayed on the ZoneGuard
controller LCD.
1.2.1 Wiring recommendations
For oxygen remote detectors use (minimum) 22 gauge
AWG, 6 conductor twisted-pair shielded with drain. For toxic
remote detectors use (minimum) 22 gauge AWG, 4
conductor twisted-pair shielded with drain. For combustible
remote detectors use 18 gauge AWG, 4 conductor twistedpair shielded with drain.
The maximum distance a ZoneGuard remote detector may
be located from the controller is 50 feet.
1.3 Sensor options
ZoneGuard systems can be configured to monitor for a
number of different atmospheric hazards. The hazard the
system is configured to detect is determined by the type of
sensor installed. The types of sensors currently available
include a galvanic oxygen (O2) sensor, a catalytic-bead
pellistor for LEL combustible gas, and substance-specific
electrochemical sensors for carbon monoxide (CO),
hydrogen sulfide (H2S), sulfur dioxide (SO2), chlorine (Cl2),
chlorine dioxide (ClO2), phosphine (PH3), and nitrogen
dioxide (NO2). Oxygen readings are given in percent-byvolume, combustible gas readings in percent LEL, and toxic
gas readings in PPM (parts per million) or 0.1 PPM
concentrations.
1.4 Calibration
ZoneGuard systems have been designed for easy
verification of accuracy .
The accuracy of the Zoneguard
should be checked periodically with known
concen t ration calibration gas. Failure to check
accuracy can lead to in accurate and potentially
dangerous readings.
1.4.1 One Button Auto Calibration
Biosystems’ “One Button” logic means that all normal dayto-day operating procedures, including automatic calibration
adjustment, may be made through the use of the single
On/Off Mode button.
Calibration procedures are discussed in detail in
Chapter 3.
Page 5
Page 7
1.5 Displays and controls
ZoneGuard systems include an 8-character LCD readout
located at the controller which is used to display gas
readings and other messages. Built-in LED alarm lights are
used to indicate “OK” / Power-on (green), “Low Alarm”
(amber), and “High Alarm” (red) conditions.
The On/Off “Mode” button is to initia te the automatic
calibration sequence and to control most other day-to-day
procedures. The button is l ocated on the exterior of the
controller housing, on the underside of the case. Four
additional push-bu tton controls (“-”, “+”, “Alarm” and “Cal”)
are located on the inside of the controller housing door.
These buttons are used for advanced procedures such as
changing operating modes, changing alarm setpoints,
adjusting calibration gas values, making optional setup
choices, an d other advanced procedures.
1.6 Alarm Logic
The Zoneguard personal, portable
gas detector has been design ed for the detection of
dangerous atmospheric condit ions. An alarm condition
indicates the presen ce of a potentially life-threatening
hazard and sh ould be taken very seriously. Failure to
immediately leave the area during an alarm condition
may result in serious injury or death.
In the event of an alarm condition it is important to
follow established procedures. The safest course of
action is to immediately leave the affected area, and
return only after further testing together with other
appropriate safety procedures determine that the area
is once again safe for en try.
ZoneGuard alarms are user adjustable and may be set
anywhere within the range of the specific sensor. When an
alarm set point i s exceeded the audible alarm sounds, the
appropriate bright LED alarm light indicator are turned-on,
and (if equipped) the optional external strobe light(s) and/or
remote high intensity audible alarm are activated. (Any
extern al alarm devices controlled by means of the
ZoneGuard alarm relay contacts will also be activated.)
ZoneGuard alarms are normally of the self-resetting type.
When readings drop back below the pre-set alarm levels,
the instrument returns to normal operation, and the visual
and audible alarms cease.
It is possible to set ZoneGuard alarms so that they "latch".
In the latched condition, once an alarm occurs both visual
and audible alarms will continue to sound even after the
atmospheric hazard has cleared. The instrument must be
manually reset before the alarms are silenced.
Procedures for latching ZoneGuard alarms are given in
Chapter 4.
ZoneGuard systems equipped with toxic or combustible gas
sensors are provided with two levels of gas alarm: “Low”
(warning) and “High” (danger). Toxic and combustible
alarms are configured as “ascending”, meaning that they are
triggered when gas concentrations exceed the al arm
setpoints.
ZoneGuard systems equipped with oxygen sensors have
four levels of gas alarm, two “ascending” alarms (“Warning”
and “Danger”) for oxy gen enrichment, an d two “descending”
(“Warning” and “Danger”) alarms for oxygen deficiencies.
Descending alarms are activated when oxygen
concentrations drop below the alarm setpoints for oxygen
deficiency.
Note: “Warning” level alarms may be manually
acknowledged and silenced for a period of 10 minutes
by pressing the Mode button. If the alarm condition
continues to exist after 10 minutes h as elapsed, the
“Warning” alarms will be reactivated. If t he alarm
condition reaches the “Danger” level during this 10
minute period, the “Danger” level alarms will
immediately be activated regardless of whether or not
the “Warning” alarm has been silenced. “Danger” level
alarms may not be manually silenced.
The procedures for adjusting alarm setpoints are given
in Chapter 4.
1.7 Special microprocessor features
Several automatic programs prevent tampering and misuse
of the ZoneGuard by unauthorized persons. Each time the
detector is turned on, an electronic self-test is performed
that assures the user of proper performance.
Optional setup choices include latching alarms, changing the
concentrati o n of the gas that is used during Auto Calibra tion
adjustment, as well as changing or restoring the factory
alarm settings by using the 4 pus h-buttons on the instrument
keypad. It is also possible to enable or dis able use of the
extern al Mode button.
ZoneGuard software also incl udes an “Alarm Test” function
that can be used to automatically “ramp” readings
throughout the full range of the sensor. The “Alarm Test”
function is used to verify alarm setpoints as well as the
proper activation of the audible alarm, LED alarm light
indicators, optional s tr obe lights or horn, as well as other
alarm devices or peripheral equipment controlled by means
of the ZoneGuar d r e lay contacts.
Chapters 4 and 5 describe how to use these advanced
technical features and setup choices in greater detail.
1.8 Relays and outputs
See Appendix F for ZoneGuard electrical schematics
and wiring diagrams.
1.8.1 VAC relay contacts
ZoneGuard systems include DPDT relay contacts for
warning (low) and danger (high) gas alarm conditions.
ZoneGuard rela y contacts are rated at 5 Amp at 250 VAC,
non-i nductive l oad.
1.8.2 VDC outputs
ZoneGuard systems also include +12 VDC outputs rated at
0.5 Amp for warning (low) and danger (hi gh) gas alarm
conditions.
Page 6
Page 8
CAUTION: ZoneGuard +12 VDC outputs may only be
used to provide power to remote alarm devices which
have been approved by Biosystems for use with the
ZoneGuard system. Use of non-approved devices may
cause damage to the ZoneGuard system, and void the
Biosystems Standard Warranty for Gas Detection
Products. Contact Biosystems Technical Service
Department at (860) 344-1079 for additional information
on approved remote alarm devices.
1.8.3 Output operation
ZoneGuard alarms are normally of the “self-resetting” type,
and will cease when readings drop back below the pre-set
alarm levels. ZoneGuard alarms may als o be configured to
“latch” so that the alarms remain activated even after the
alarm condition has ceased. Once activated, a latched
alarm will continue to sound until the instrument is manually
reset by pressing the mode button.
The procedure for changing the alarm latch setting is
given in Chapter 4
If the gas concentration causes the activation of the low
(warning level) alarm, the audible alarm is activ ated, the low
level (amber) alarm LED illuminates, and the low alarm relay
is activated. Warning level alarms may be silenced and the
relay deactivated for a period of 10 minutes by pressing the
Mode button.
If the gas concentration causes the activation of the high
(danger level) alarm, the audible alarm is activated, the high
level (red) alarm LED illuminates, and the high alarm relay is
activated. High level alarms may not be manually silenced
by pressing the Mode button.
ZoneGuard systems are supplied with relays and outputs
preset with the following standard factory setup:
ZoneGuard remote detector assembli es are available in
both explosion-proof and non-expl osion-proof versions.
Explosion-proof version remote detector assemblies have
been su bmitted to Underwriters Laboratories, Inc. for
Classificati on per ULC as Explosion-Proof for use in Class I,
Division 1, Groups A, B, C, D, E, F, and G explosive
atmospheres.
1.10.1 ZoneGuard controller
The ZoneGuard controller unit is enclosed in a durable,
gasketed, fiberglass NEMA 4X case. The On/Off Mode
button i s located on the exterior of the controller housing, on
the underside of the case. Four additional push-button
controls (“-”, “+”, “Alarm” and “Cal”) are located inside the
door of the controller housing.
1.10.1.1 LED alarm light indicators
ZoneGuard controllers include 3 bright system status and
alarm condition LED indicators; green for “OK” / Power On,
amber for “Warning”, and red for “Danger”.
1.10.1.2 Audible alarm
A loud (85 dB at 3 feet) built-in audible alarm is provided to
warn us ers of hazardous conditions. Optional remote
audible alarms and high intensity horns are also available.
1.10.1.3 Detector display
An 8-character LCD readout is used to displ ay gas readings
and other information.
Readings, messages, and modes of operation are
discussed in Chapter 2.
1.10.2 Remote detector assemblies
Remote detector assemblies include sensor, sensor specific
amplifier board (installed in the remote detector housing),
and remote detector hous ing. Oxygen and toxic gas remote
detector assemblies are available in both “standard” (nonexplosion-proof) and “ex plosion-proof” versions. Remote
detector assemblies for combustible gas are only available
in explosion-proof housings.
Installation procedures are discussed in Chapter 2.
1.10.2.1 Standard Detector housing
“Standard” remote detector assemblies are housed in a
heavy-duty, ni ckel-coated, molded junction box. Not for use
in hazardous locations.
1.10.2 . 2 Explosion-proof detector housing
“Explosion-proof” remote detector assemblies are housed in
a heavy duty steel explosion-proof (Class I, Divi sion 1,
Groups A, B, C, D, E, F, G) junction box with threaded lid.
1.10 ZoneGuard design components
See Appendix F for dimensional drawings, mounting
schematics, and wiring diagrams for the ZoneGuard
controller and remote detector assemblies.
Page 7
Page 9
CHAPTER 2. INSTALLATION AND
BASIC OPERATION
2.1 Installation overview
Biosystems gas detection systems are fully tested and
calibrated before they leave the factory. However, after
installation we strongly recommend that full testing and
verification of proper operation is carried out before the
system is placed in normal service. Make sure all
ZoneGuard system elements and/or peripheral alarm,
notification, or con trol devices are properly installed
before connecting the system to a power source!
Make sure that th e ZoneGuard system is disconnected
from the external po wer source before installing,
replacing, or handling ZoneGuard system components!
Repair of system components damaged as a result of
improper installation can be expensive! ZoneGuard
installation, initial setup, and / or system modification
should only be undertaken by individuals who are
qualified and authorized to do so. Call Biosystems’
Service Department at (860) 344-1079 if you need help
or have any questions.
2.2 ZoneGuard controller unit
The ZoneGuard controller unit is designed for wall mounting
in any convenient non-hazardous location where a suitable
power supply is available. ZoneGuard control ler units are
designed for use with 120VAC, 60Hz. external power.
Note: ZoneGuard controller housing accessories
include a set of 4 external mounting brackets. Use of
these mounting brackets is optional. The controller
hous ing may be mounted directly to the wall or other
mounting surface by use of screws inserted throu gh
apertures in 4 mounting wells located on the back of the
controller case, or by means of the external mounting
brackets, whichev e r is more convenient.
ZoneGuard controller housing accessories also include a ½”
NPT galvanized cable grab “gland”. If the cable connecting
the remote detector housing with the controller is run through
conduit, it will be necessary to remove the cable grab gland
and replace it with a threaded ½” NPT conduit bushing (not
provided). When making use of ZoneGuard relay contacts
to control or actuate external alarm devices it will be
necessary to remove the “hole-seal” from the bottom of the
controller case, and install the appropriate glands or
bush ings. To prevent dust and water ingress all cabl e
entries and knockouts must be checked and sealed as
necessary.
2.2.1 Controller PCBs
The ZoneGuard controller includes a displ ay PCB located
underneath the hinged controller cas e cover, and the relay
PCB located on a mountin g plate inside the controller case.
In the case of ZoneGuard systems with integrally mounted
sensors, a sensor specific amplifier board is mounted onto
the relay PCB. In the case of ZoneGuard systems with
remote sensors, the amplifier board is located in the remote
detector housing.
The four push-buttons (“+”,“-”, “Alarm”, and “Cal”) us ed to
make use of advanced features and setup choices are
located on the display PCB on the inside of the con troller
hous ing door.
2.3 Remote detector assemblies
Oxygen and toxic gas remote detector assemblies are
available in both “standard” (non-explosion-proof) and
“explosion-proof” housi ngs. Remote detector assemblies for
combustible gas are only available in explosion-proof
housings.
ZoneGuard re mote detector as semblies may be located up
to 50 feet away from the controller. Power is provided to
the remote detector by the Zo neGuard controller.
The ZoneGuard remote detector assembly contains a
factory installed, sensor specific amplifier PCB. The remote
detector assembly is labeled as to the type of hazard it is
designed to measure (i.e. H2S, CO, O2, etc.).
2.3.1 Standard remote detector housing
Sensors for the measu r emen t of oxygen and toxic gas may
be installed either in explosion-proof or “standard” remote
detector housings.
Sensors that are hou sed in
“standard” remote detector housings may not be
installed in hazardous locations requiring an
“Explosion-Proof” classification.
2.3.2 Explosion-proof remote detector
housing
Explosion-proof remote detector housin gs can be installed in
Class I, Division 1 Groups A thro ugh G hazardous locations.
Explosion-proof remote detectors are installed in a heavy
duty stainless steel explosion-proof (Class I, Division 1,
Groups A through G) junction box with threaded lid.
ZoneGuard sensors for the measurement of
combustible gas may only b e installed in ex plosionproof remote detector housings.
2.4 Siting of remote detectors
The specific placement of remote detector assemblies
should be a function of the type of area being monitored, the
type and source of atmospheric hazard being measured,
prevailing wind patterns, and other information.
Call Biosystems Technical Service Department at (860)
344-1079 for additional advice.
In general, for gases lighter than air, the detector should be
placed at a level slightly above the area where leaks are
likely to occur. For gases which are heavier than air the
detectors should be located close to floor level or in
inspection pits or ducts into which heavier than air gases
may collect or flo w.
There are many circumstances which may modify this
general advice. For example, the molecular weight of
nitrogen (MW ≅ 28) is very close to that of air (MW ≅ 29).
Page 8
Page 10
When nitrogen is at the same temperature as the air into
which it is introduced, it mixes readily, and tends to spread
evenly through the affected atmosphere. On the other hand,
if the nitrogen is under pressure, and then suddenly released
into the atmosphere, as the gas expands (going from higher
pressure to a lower pressure) it cools. Because the cooler
gas is now denser than the air in to which it is being
introdu ced, it no longer mixes in as readily. Instead, the
nitrogen tends to fall to floor level, then spread l aterally. In
this case remote detectors being used to monitor for oxygen
deficiency should be located n ear floor level in order to
detect the deficiency as quickly as possible.
The nearer in density to air a gas is, the more easily it will
flow with air due to draughts, ventilation etc. A compromise
approach for placement of detectors used to measure
gases whi ch are only s lighter lighter (such as carbon
monoxide) or heavier (such as hydrogen sulfide) than air is
to mount the detectors at a height as close as possible to
the breathing area of personnel being protected.
When installing detector assemblies it is important to ensure
that the sensor is not exposed to liquid or du st contamination
which would interfere with th e passage of gas through the
protective filter into the sensor. Remote detector
assemblies shoul d be placed so that the axis of the sensor
points straight downward. Splash deflectors should be used
when water or other liquid is chronically present in the area
where the detectors have been located.
2.5 Wiring recommendations
For oxygen remote detectors use (minimum) 22 gauge
AWG, 6 conductor twisted-pair shielded with drain. For toxic
remote detectors use (mini mu m) 22 gau ge AWG, 4
conductor twisted-pair shielded with drain. For combustible
remote detectors use 18 gauge AWG, 4 conductor twistedpair shielded with drain.
The maximum distance a ZoneGuard remote detector may
be located from the controller is 50 feet.
Due to the relatively low signal levels carried by wiring
between gas detectors and their control unit, it is essential
not to run wire near high power electrical equip men t. When
ZoneGuard systems are installed in environments which
contain high power electrical equipment it is usually best to
install the wire in conduit.
2.5.1 Grounding
Wire used to connect ZoneGuard remote detector
assemblies with the controller shou ld always be with drain
lead. In order to reduce the chances for electromagnetic
interference (EMI), the screen (drain) of the wire should be
connected to the ground plate of the detector head (l ong
reach glands and/or the screen terminals should be us ed for
this purpose). Ground loops must be avoided! Grounding is
done through the normal circuit ground of the system.
2.6 Connecting remote alarm and
detector assemblies with the
ZoneGuard controller
2.6.1 Cautions
CAUTION: Make sure that the ZoneGuard system is
disconnected from the external power source before
installing, replacing, or handling installed ZoneGuard
system components!
CAUTION: Make sure the remote detector housing is
properly connected to the correct controller terminal
block connector as discu ssed below.
CAUTION: +12 VDC outputs may only be used to
provide power to remote alarm devices which have
been approved by Biosystems for use with the
ZoneGuard system. Use of non-approved devices may
cause damage to the ZoneGuard system, and void the
Biosystems Standard Warranty for Gas Detection
Products. Contact Biosystems Technical Service
Department at (860) 344-1079 for additional information
on approved remote alarm devices.
2.6.2 ZoneGuard controller connections
Electrical connections and wiring diagrams for the
ZoneGuard controller are shown in Appendix F.
Full wiring sch ematics are available on the Biosystems
website or by contacting Biosystems directly.
Connectio ns are by screw terminal blocks located on the
mounting plate and relay PCB located in the ZoneGuard
controller. Warning (Low) and Danger (Hi gh) VAC output
relays are DPDT, are isolated from the system power
supply, and are rated at 5 Amp, 250VAC, non-inductive.
Warning (Low) and Danger (High) VDC outputs (+12 VDC
at 0.5 Amp) may only be us ed to provide power to remote
alarm devices which have been approved by Biosystems for
use with ZoneGuard systems.
2.7 Initial setup
1. Make sure the controller is not con nected to any other
peripheral or external equipment or alarm devices.
2. Disconn ect the built-in “loud” 85 dB audible alarm.
3. Make sure any other alarm or notification devices
controlled by mean s of the ZoneGuard relay contacts
are disconnected or turned off.
4. Connect mains power to the controller.
5. Connecting the ZoneGuard system with mains power
will cause the detector to turn on in normal operation.
The green “OK / Power On” LED indicator should be lit
and the LCD disp lay should be on. The LCD disp lay
shou ld show a sequence of start-up screens as the
detector goes through a self test sequ ence, followed by
the curren t gas level readings. The low or high LED
alarm indicators, optional strobe lights, remote audible
alarms, and other devices connected to may be
activated while the sensor is stabilizing.
Page 9
Page 11
Note: The ZoneGuard system is designed so that in
the event of a power interruption, the instrument will
automatically turn itself back on. To turn the system
off it will be necessary to hold down the Mode button
until the display goes blank. When the Mode button
is released the display will ind icate “OF F”.
6. Allow the readings to stabilize fully.
Note: Combustible and oxygen sensors may take up
to an hour after power is applied to the system to
stabilize fully. Carbon monoxide, hydrogen sulfide,
sulfur dioxide and phosphine may take up to 2
hou rs. Chlorine, chlorine dioxide, and nitrogen
dioxide sensors may take up to 12 hours.
7. If readings are still in alarm, the ZoneGuard should be
fresh air zeroed. Since the amount of adjustment
required at the time of initial setup may be large, it is
usually better to make a “manual” fresh air zero
adjustment. To do a manual fresh air zero, make sure
the sensor is currently located in fresh air, then press
the “Cal” button to enter the “Zero Adjust Mode”. The
display should show the message “Zero Gas ” . Press
the Cal button once more to automaticall y make a fresh
air zero adjustment. (The complete procedures for
zero and span calibration are discussed in Chapter
3.)
8. Reconnect the 85dB audible alarm and any other alarm
devices which were turned off or disconn ected prior to
“powering up” the ZoneGuard system.
The ZoneGuard system must be properly calibrated as
discussed in Chapter 3 at the time of initial installation
before the system is put into normal service.
2.8 Normal operation
2.8.1 Overview of ZoneGuard operation
The ZoneGuard offers a choice of two modes of operation,
Text Only, (“OK”) and Basic, (numeric readings). Mode
selection should be determined based on how much
information is required, the skill level of the user, and the
nature of the environment in which the ZoneGuard system is
installed.
2.8.2 Turning the ZoneGuard on
The black switch located on the bottom of the exterior of the
ZoneGuard con troller housing is cal led the Mode button. It
is used to turn the ZoneGuard on and off, as well as to
control most other operations of the instrument. Press and
hold th e Mode button for one second to turn the ZoneGuard
on.
2.8.3 Turning the ZoneGuard off
Turn the ZoneGuard off by pressing and holding the Mode
button until the LCD goes blank. When the button is
released the LCD will display the message “OFF”.
2.8.4 Start-up sequence
After the ZoneGuard has been turned on, it will automatically
go through an electronic self test and start-up sequence that
takes approximately twenty seconds. During start-up the
LCD display will show several messages or "screens" in
sequence. The first screen displays the software version
number: The second shows the type of sensor installed.
Additional screens will briefly appear indicatin g the current
low (warning) and high (danger) alarm setpoints. For
instance, a ZoneGuard for carbon monoxide being operated
in the Basic mode would show the following sequ ence of
factory installed (default) alarm set points when the
instrument is first turned on. (Note: The software version
installed in your ZoneGuar d may differ with that shown
below.)
→
→
→
→
At the conclusion of the startup sequence, the screen will
cycle back and forth between the current gas readings
screen and the sensor identification screen.
↔
Note: LCD screens indicating readings or alarm
setpoints include the measurement u nits being used. In
the case of toxic gas the last three ch aracters will
indicate “PPM” (parts-per-million). In the case of
combustible gas (LEL) and oxygen the last three
display characters will indicate “PCT” (percen t ) .
The alarm set-point screens are omitted when the
instrument is tu r ned on while in the “Text Only” mode.
→
→
→
2.9 Operating modes
Two operating modes provide ZoneGuard users with the
right amount of information to meet their monitoring needs.
The ZoneGuard is designed to be as simple or as
sophisticated as the job that needs to be done.
2.9.1 Text Only mode
The simplest mode of operation i s the “Text Only” mode. In
this mode, during normal operation the LCD screen dis plays
an “OK” message rather than numerical readings as long as
no alarms are present. The “OK” screen alternates with th e
sens or identification screen.
↔
During an alarm condition (when the sensor readings exceed
a pre-set alarm level), the ZoneGuard will show readings as
a numerical value, activate the appropriate LED alarm light,
soun d the audible alarm, and (if equipped) activate the
optional high intensity strobe light(s), remote audi ble alarm,
Page 10
Page 12
and/or other alarm or notification devices controlled by the
ZoneGuard rela y contacts.
↔
ZoneGuard alarms are normally self-resetti ng. When
readings drop back below the pre-set alarm levels, the
screen returns to the "OK" in dication, and visual and audible
alarms cease.
2.9.2 Basic mode
Basic mode provides numerical readings during normal
operation.
↔
2.9.3 Changing operating modes
To change operating modes, with the instrument running
open the door of the con troller housing to access the four
mini-push-buttons, and simultaneously press and hold the
"+" and "-" buttons until the word “MODE” is displayed.
Each time the operatin g mode is changed the display screen
will briefly indicate which operating mode “TEXT” or “BSIC”
(Basic) has been selected.
Note: Shifting modes or otherwise reprogramming the
instrument should only be done by employees who are
authorized to do so.
2.10 Enabling or disabling use of the
Mode button
Most day-to-day operati ng procedures (such as automatic
calibration adjustment) are normally made by using the
On/Off Mode button. It is possible to disable the u se of the
Mode button. Once disabled, the Mode button may no
longer used for any purpose. To res tore normal functions,
the Mode button must be re-enabled.
To enable or disable use of the Mode button press and hold
down the “+” key while the ZoneGuard system is in normal
operation. The display will show the message “MODE OFF”
or “MODE ON” to indicate when the procedure of enabling
or disabli ng the Mode button is complete.
CHAPTER 3 CA LIBRATION
3.1 Warnings and cautions concerning
verification of accuracy
The accuracy of the Zoneguard
should be checked immediately following any known
exposure to contaminants by testing with known
concentration test gas before further use. Failure to
check accuracy can lead to inaccurate and potentially
dangerous readings.
A sensor that cannot be calibrated
or is found t o be out of tolerance should be replaced
immediately. An instrument that fails calibration may
not be used until testing with known concentration test
gas determines that accuracy has been restored, and
the instrument is once again fit for use.
Do not reset the calibration gas
concentration unless you are using a calibration gas
concentration that differs from the one that is normally
supplied by Biosystems for use in calibrating the
Zoneguard. Use of inappropriate calibration gas may
lead to in accurate an d potentially dangerous readings.
Customers are strongly urged to use only Biosystems
calibration materials when calibrating the Zoneguard.
Use of non-standard calibration gas and/or calibration
kit components can lead to dangerously inaccurate
readings and may void the standard Biosystems
warranty.
Use of non - st andard calibration gas
and/or calibration kit components wh en calibrating th e
Zoneguard can lead to inaccurate and potentially
dangerous readings and may void the standard
Biosystems warranty.
Biosystems offers calibration kits and long-lasting
cylinders of test gas specifically developed for easy
Zoneguard calibration. Customers are strongly urged
to use only Biosystems calibration materials when
calibrating the Zoneguard.
3.2 Calibration overview
The ZoneGuard gas detection system has been designed
for easy calibration. A single button, th e on / off Mode
button, is used to initiate the “Au to-Calibration” sequence,
and automatical ly make all calibra tion adjustments.
It is also possible to manually calibrate the instrument by
using simple pus h-button controls located on the inside of
the door of the ZoneGuard controller housing.
“One-Button Auto-Cal” procedures are discussed in
Section 3.6. Manual calibration procedures are
discussed in Section 3.7.
3.3 Verification of accuracy
To verify accuracy, first place the Zoneguard in Basic Mode
so that readings are given in numbers. See section 2.9.3 for instructions on changing operating modes.
Page 11
Page 13
Verification of accuracy for a Zoneguard equipped with an
O2 sensor is a one-step procedure. Si mply provide the
sensor with a known source of fresh air and veri fy that the
display reads 20.9% oxygen. If the display shows anything
besides 20.9% while exposed to a known source of fresh
air, it must be fresh air calibrated as discussed below in
section 3.6.1.
Verification of accuracy for a Zoneguard equ ipped with
either a combustible gas / LEL sensor or any tox ic sensor is
a two-step procedure.
Step one is to provide the sensor with a known source of
fresh air and verify that the display shows “0 PCT” for LEL
units or “0 PPM” for toxic units. If an instrument equipped
with an LEL or toxic sens or shows anything besides “0” while
exposed to a known source of fresh air, it must be fresh air
calibrated as discussed bel ow in section 3.6.1. If the
ZoneGuard is located in an area in which the air may not be
fresh, see section 3.3.1 below for instructions on fresh ai r
“zero” calibrations in contamin ated air.
Step two is to test sensor response by exposing the sensor
to appropriate test gas of known concentration. This is
known as a functional (bump) test. Readings are
considered to be accu rate when the display shows a reading
between 90% and 120% of the expected values as gi ven on
the calibration gas cyli nder. If readings are accurate, there
is no need to adjust your gas detector. See section 3.5 for
explicit functional/bump test instructions.
If the readings are inaccurate, th e instrument must be
calibrated before further use as discussed in section 3.6.
Biosystems offers calibration kits and long lasting cylinders
of test gas specifically developed for easy ZoneGu ard
calibration . Customers are urged to use Biosystems
calibration materials when calibratin g their ZoneGuard
detectors.
3.3.1 Fresh air "zero" in contaminated air
The fresh air "zero" for all types of sensors must be done
while the sensor i s located in fresh, uncontaminated air. In
this procedure the instrument automatically adjusts its
readings to match the concentrations present i n fresh air.
Fresh air should contain 20.9% oxygen, 0% LEL combustible
gas, and 0 parts-per-million toxic gas.
If the atmosphere in which the ZoneGuard sensor is located
is not fresh, or if it is not certain whether or not the air is
contaminated, speci al procedures are required. In this case,
the zero adjustment must be made while the sensor is
exposed to contaminant free atmosphere by means of a
cylin der of “Zero Air” which contains 20.9% oxygen, 79.1%
nitrogen and no contaminants. This gas is supplied to the
sensor and the instrument may then be fresh air “zero”
calibrated as discussed i n section 3.6.1.
3.4 Effect of contaminants on ZoneGuard
sensors
The atmosphere in which th e ZoneGuard monitor is being
used can have an effect on the sensor. Sensors may be
poisoned or suffer degraded performance if exposed to
certain substances.
Caution: The accuracy of the ZoneGuard should be
checked immediately following any known exposu re to
contaminan ts by testing with known concentration test
gas before further use.
The ZoneGuard is avail able with three basic types of
sensors: oxygen, combustible gas (LEL), and
electrochemical toxic. Each type of sens or uses a slightly
different detection principle. The kinds of conditions that
affect the accuracy of the sensors vary from one type of
sens or to the next.
3.4.1 Effects of contaminants on oxygen
sensors
Oxygen sensors may be affected by prolonged exposure to
"acid" gases such as carbon dioxide. The oxygen sensors
used in Biosystems instruments are not recommended for
continuous use in atmospheres which contain more than 25
% CO2.
3.4.2 Effects of contaminants on
combustible sensors
Combustible sensors may be affected by exposure to
substances containing silicone (found in many lubricants and
hydrau lic fluids), the tetra-ethyl-lead in "leaded" gasoline,
and halogenated hydrocarbons (Freons®, or solvents such
as trichloroethylene and methylene chloride). High
concentrations of hydrogen s ulfide may also damage the
sensor.
For a comprehensive list of known LEL sensor poisons,
see the last paragraph of the Biosystems Standard
Warranty at the end of this manual.
If sensitivity of the combustible sensor is lost due to
poisoning, it tends to be lost first with regards to methane. A
partially poisoned sensor might still respond accurately to
propane while showing a dangerously reduced response to
methane.
Biosystems’ “Equivalent” calibration gas mixtures have been
developed to eli minate this potentially dangerous source of
calibration error. Biosystems’ “Equivalent” mixtures are
based on methane, so any loss of sensi tivity to methane is
detected (and can be corrected) immediately. Biosystems
offers several different “Equivalent” combustible calibration
gas mixtures for use in various monitoring applications.
Please contact Biosystems’ Service Department at (800)
711-6776 for help in deciding which calibration gas mixture
to specify for use.
Using Biosystems brand calibration gas insures that
proper sensitivity is maintained over the life of the
sensor.
3.4.2.1 Effects of high concentrations of
combustible gas on the combustible
sensor
The accuracy of combustible sensors may also be affected
by exposure to high concentrations of combustible gas. To
minimize the chance for damage or loss of sensitivity to the
combustible sensor, the ZoneGuard system is designed to
Page 12
Page 14
"alarm latch" whenever the concen tration of combustible gas
exceeds 100 percent LEL. Under these conditions the
combustible gas reading will show an “X” to indicate an overlimit condition. The audible and visual alarms will sound
continuously until the instrument is manually reset by turning
it off, then turning the instrument back on.
A combustible sensor overrange
alarm indicates a potentially explosive atmosphere.
Failure to leave th e area immediately may result in
serious injury or death!
Note: If the over limit condition still exists at the time
the instru ment is turned back on, the ZoneGuard
system will revert to the over-limit alarm-latch
condition.
truly cleared before attempting to make an auto-zero
adjustment. Auto-zero adjustments may only be made
when the ZoneGuard sensor is located in air that is
known to be fresh . Accuracy of the combustible gas
sensor should be verified by exposure to known
concentration test gas before further use.
Note: The combustible sensor used in the ZoneGuard
design requires the presence of oxygen in order to
detect combustible gas. The accuracy of the
combustible sensor may be affected if the instrument is
used in oxygen deficient atmospheres.
by a rapid declining or erratic reading may indicate a
gas concentration beyond the u pper scale limit which
may be hazardous.
Make sure the alarm condition has
Any rapid up-scale reading followed
3.4.3 Effects of contaminants on toxic gas
sensors
Biosystems “substance-specific” electrochemical sensors
used to measu r e toxic gases such as CO, H2S, SO2, Cl2,
ClO2, PH3, and NO2 have been carefully designed to
minimize the effects of common interfering gases.
“Substance-speci fic” sensors are designed to respond only
to the gases they are supposed to measure. The higher the
specificity of the sensor the less likely the sensor will be
affected by exposure to oth er gases which may be
incidentally present. For instance, a “substance-specific”
carbon monoxide sensor is deliberately designed not to
respond to oth er gases which may be present at the same
time, such as hydrogen sulfide or methane.
Even though care has been taken to reduce cross sensitivity, some interfering gases may still have an effect on
toxic sensor readin gs. In some cases the interfering effect
may be “positive” and result in readings wh ich are higher
than actual. In some cases the interference may be
negative and produce readings which are lower than actual.
Cross sensitivity of ZoneGuard toxic sensors to
common interfering gases is listed in Appendix E.
toxic gas sensor. For further details on verifying the
accuracy of an instrument with an O2 sensor, see
section 3.3.
To perform a Functional (bump) test,
1. Place the Zoneguard in Basic Mode so that readings are
given in numbers.
2. Make sure the sensor is located in fresh air. If you are
not certain the air is fresh, use a cylinder of Zero Air to
supply atmosphere to the sensor.
3. Verify that the readings match the con centrations
present in fresh air. Instruments equipped with an LEL
sensor or any of the toxic gas sensors should read “0”.
If necessary, fresh air “zero” calibrate the ins trument
using the procedures discussed in Section 3.6.1.
4. Screw the regulator into the cylinder and connect it to
the detector using the short section of tubing and the
calibration adapter. The regulator will automatically
begin flowing calibration gas at the correct flow rate as
soon as it is screwed into the cylinder of gas.
5. The instrument will display the sensor readings as they
rise. When they have stabilized at their highest value,
note the readings. Readings are considered to be
accurate when the display is between 90% and 120% of
the ex pected values as given on the calibration gas
cylinder. If readings are accurate, there is no need to
span calibrate the detector.
Readings that are considered inaccurate indicate the
instrument must be adjusted using the “span”
calibration procedures discussed in Section 3.6 before
further use.
3.6 Auto-calibration
Biosystems “One-Button Auto-Calibration” mode may be
used to verify accuracy any time du ring normal operation
while the instrument is being used in the Basic operating
mode. Press the Mode button 3 times in rapid sequence to
initiate the “Auto-Calibration” sequence. Sensor
adjustments are made automatically in Auto Calibration
mode.
Auto-calibration is a one step procedure for instruments
equipped with an oxygen sensor. If the detector doesn’t
read 20.9 PCT in a known fresh air environment, perform
the fresh air “zero” calibration as described in section 3.6.1.
Auto-calibration is a two step procedure for instruments
equipped with either a combustible gas / LEL sensor or a
toxic gas sensor. In the first step the sensor is ex posed to
fresh air and a "zero" adjus tmen t is made automatically as
discussed below in section 3.6.1. The second step is th e
sensor response or "span" calibration adjustment. In this
step the accuracy of the ZoneGuard sen sor is established by
exposing it to known concentration calibration gas as
discussed below in section 3.6.2. The sensitivity or “span” is
automatically adjusted.
3.5 Functional (bump) test
Note: The functional (bu mp) test is only used on
instruments equipped with a combustible gas / LEL or
Page 13
Page 15
3.6.1 Fresh air "zero" auto-calibration
sequence
The “One Button” fresh air zero procedure may only be done
while the ZoneGuard system is being operated in the Basic
operating mode.
1. Make sure gas readings are given in numbers. If the
display shows “OK”, then the ZoneGuard is currently
being operated in the “Text Only” mode. To change
over to Basic mode simul taneously hold down th e "+"
and"-" keys as discussed in section 2.9.3.
2. Make sure the ZoneGuard system has been turned on
and operating for a sufficient period of time for the
sensor readings to have stabilized fully before initiating
auto-calibrati on procedures.
If the sensor has just been replaced, the ZoneGu ard
should be turned on and allowed to stabilize for the
following period of time prior to performing any of the
calibration options.
Sensor Stabilization Period
Oxygen 1 hour
LEL 5 minutes
All Toxic sensors
available for ZoneGuard
3. Make sure the sensor is located in an area where the air
is known to be fresh or provide fresh air to the sensor
with a cylinder of “Zero Air”. See section 3.3.1 for more
details on using a “Zero Air” cylinder during the Fresh Air
“zero” calibration.
4. Press the Mode button 3 times within two seconds. This
will "wake up" the instrument from normal operation,
and put it into the “Auto-Calibration” mode. A screen
will briefly display the message “ZERO MODE”.
5. Press the Mode button withi n five seconds to initiate the
fresh air adjustment. (If the Mode button is not push ed
within 5 seconds, the instrument will return to normal
operation.)
6. The screen will show the message “ZERO WAIT” while
the adjustment is being completed.
7. After successful completion of the “zero” auto-
calibration, the display will prompt you with the message
“SPAN MODE” to cont inue with a span calibration
adjustment.
Span calibration is covered in detail below in section
3.6.2.
3.6.1.1 Reading “Too High” or “Too Low” for
zero adjust
To reduce the chances of the ZoneGuard being
inadvertently zeroed in contaminated air, only small
15 minutes
↔
↔
adjustments are allowed through the use of the “One Button”
auto-zero sequence. If the necess ary adjustments are too
large the display will indicate that the ZoneGuard “CANT
ZERO”. In this case the instrument mu st be fresh air zeroed
using the “Cal” button on the instrument keypad and
procedures discussed in Section 3.7.1.
Once the instrument has been successfully zeroed using the
“Cal” button, subsequent calibration adjustments may be
made using the Mode button and “One Button Auto Cal”
logic discussed in this section.
3.6.2 “Span” auto-calibration sequence
After successful comple tion of the FRESH “zero” autocalibration adjustment, instruments equipped with either a
combustible gas / LEL sensor or a toxic gas sensor will
automatically move on to “SPAN MODE”.
↔
1. Press the Mode button within 5 seconds to initiate
“span” auto-calibration. The screen will show “FLOW
GAS” and will continue to display this screen until it
determines that calibration gas has been applied. If the
Mode button is not pushed within five seconds a span
calibration adjustment will not be made, and the
instrument will be returned to normal operation.
Note: It is possible to exit t he auto-calibration mode at
any time prior to completion by pressing and holding
down the Mode button to turn th e instrument off.
↔
2. Screw the regulator into the cylinder and connect it to
the detector using the short section of tubing and the
calibration adapter. The regulator will automatically
begin flowing calibration gas at the correct flow rate as
soon as it is screwed into the cylinder of gas.
Make sure the regulator, cylinder
seating surfaces and threads are clean and dry before
attaching the regulator to the cylinder of gas.
Introduction of contaminants th rough the regulator
fittings may alter or degrade the concentration of the
gas contained in the cylinder and may lead to
inaccurate and potentially dangerous gas readings.
3. The instrument will display the sensor readings as they
rise. When they have stabilized at their highest value,
the instrument will note the response and, if necessary,
adjust the readings to match the concentration of gas
being used to calibrate the detector . (The instrument
will beep as the adjustment is being made.)
and/or calibration kit components wh en calibrating th e
ZoneGuard may lead to dangerously inaccurate
readings and may void the standard Biosystems
warranty.
4. When span adjustment has been completed, the di splay
will show the message “DONE”.
Use of non-standard calibration gas
Page 14
Page 16
Note: Make sure the calibration adaptor and all other
fittings have been removed from the detector before it
is returned to service.
3.7 Manual calibration procedure
It is also possible to calibrate the ZoneGuard manually using
the four buttons on the instrument keypad.
3.7.1 Manual Fresh air "zero" through
keypad buttons
1. Make sure gas readings are given in numbers. If the
display shows “OK”, then the ZoneGuard is currently
being operated in the “Text Only” mode. To change
over to Basic mode, simultaneously hold down the "+"
and"-" keys as discussed in section 2.9.3.
2. Make sure the ZoneGuard system has been turned on
and operating for a sufficient period of time for the
sensor readings to have stabilized fully before initiating
manual calibration procedures.
If the sensor has just been replaced, the ZoneGu ard
should be turned on and allowed to stabilize for the
following period of time prior to performing any of the
calibration options.
Sensor Stabilization Period
Oxygen 1 hour
LEL 5 minutes
All Toxic sensors
available for ZoneGuard
3. Make sure the sensor is located in fresh, contaminant
free atmosphere or use a cyl inder of Zero Air to supply
atmosphere to the sensor.
4. Press the keypad button marked "Cal". The “GAS
ZERO” calibration message will appear on the
instrumen t LCD.
5. Press “Cal” to automatically zero the instrument. The
LCD will show the message “ZERO WAIT” while the
adjustment is being made.
6. After completion of the zero adjustment the ZoneGuard
will automatically return to the current gas readings
screen.
7. The manual fresh air zero adjustment can be canceled
at any time prior to completi on by pressing the Mode
button. Pressing the Mode button causes the calibration
values in the memory to remain unchanged from the
last time a fresh air adjustment was made. An
information screen is displayed briefly which verifies that
↔
↔
15 minutes
the zero values have not been changed. The instrument
will then return to normal operation.
↔
3.7.2 “Manual” span calibration through
keypad buttons
Span calibration procedures using buttons on the instrument
keypad are only necessary when the adjustment necessary
to restore accuracy is outside the permissible window of
adjustment for “One Button Auto Calibration” pr ocedures.
Span calibration using the keypad buttons may onl y be done
while in the Basic operating mode.
1. Make sure gas readings are given in numbers. If the
display shows “OK”, then the ZoneGuard is currently
being operated in the “Text Only” mode. To change
over to Basic mode, simultaneousl y hold down the "+"
and"-" keys as discussed in section 2.9.3.
2. Make sure the ZoneGuard system has been turned on
and operating for a sufficient period of time for the
sensor readings to have stabilized fully before initiating
manual calibration procedures .
If the sen sor has just been replaced, the ZoneGuard
should be turned on and allowed to stabilize for the
following period of time prior to performing any of the
calibration options.
Sensor Stabilization Period
Oxygen 1 hour
LEL 5 minutes
All Toxic sensors
available for ZoneGuard
3. Turn the instrument off by pressing and holding the
Mode button.
4. While holding down the "Cal" bu tton, press the Mode
button to turn the ZoneGuard back on. A screen
sh owing the message “SPAN” will be displayed which
alternates wi th the sen sor readings.
Press the Mode button at any time to cancel the calibration.
5. Screw the regulator into the cylinder and connect it to
the detector using the short section of tubing and the
calibration adapter. The regulator will automatically
begin flowing calibration gas at the correct flow rate as
soon as it is screwed into the cylinder of gas.
6. Allow the readings to stabilize. When the readings
stabilize, use the "+" and "-" keys to raise or lower the
readings to match the concentration printed on the
calibration cylinder label.
7. Make sure the correct cylinder of gas is attached before
attempting to adjust the span! If the concentration of
gas reachi ng the sensor is too low to al low the
instrument to be adjusted, or if the wrong type of gas i s
applied to the sensor being adjusted, the alarms will be
↔
15 minutes
Page 15
Page 17
activated and a screen will be displayed indicating that
the ins trument “CANT SPAN”. Pressing the Mode
button cl ears the “CANT SPAN” message and returns
the instrument to normal operation.
↔
In the event t hat a “CANT SPAN” message is displayed,
verify the type and concentration of the gas being used is
correct for the sensor being calibrated, and replace the
cylinder if necessary. Verify that the flow rate of the
regulator is 1.0 lpm. If the "CANT SPAN" message is
repeated it may indicate that the sensor is unresponsive,
or dead, and requir es replacement.
8. When span calibration has been completed, press and
hold th e "Cal" button down until the LCD s creen
indicates that calibration is complete by showing the
message “DONE”. The ZoneGuard will then turn itself
off.
Caution: The “Cal” button must be held down until
the screen in dicates that span calibration has been
successfully completed.
If the button is released before this message is displayed,
span values will not be updated, and remain unchanged
from the last time a span calibration was successfully
completed.
Pressing the Mode button at any time cancels the manual
span calibration procedure. A screen will announce “NO
SAVE” after which the instrument will turn itself off.
↔
A sen sor that cannot be calibrated
or is found to be ou t of tolerance should be replaced
immediately. An instrument that fails calibration may
not be used until test ing with known concentration test
gas determines that accuracy has been restored, and
the instru ment is once again fit for use.
Pressing the Mode button at any time during the span
adjustment cancels the procedure, and causes the
calibration values in memory to remain unchanged from the
last time a span calibration adjustment was made.
3.7.3 Manual true zero span calibration for
oxygen sensors
Note: During normal operation of a ZoneGuard system
for the detection of oxygen, the only calibration
adjustment necessary is the automatic fresh air “zero”
procedure as discussed in Section 3.6.1.
Following reinitialization of the instrument, or in cases where
the ZoneGuard is being used to monitor oxygen lev e ls that
are close to 0.0%/volume and increased accuracy near 0 is
required, th e ZoneGuard may be manually span adjusted. A
cylinder of calibration gas containing 100% nitrogen and no
oxygen is ne cessary for this calibration routine.
If the ZoneGuard is not span adjusted following
reiniti alization, message screens will indicate “NEED CAL”
when the instrument is next turned on.
↔
1. Make sure gas readings are given in numbers. If the
display shows “OK”, then the ZoneGuard is currently
being operated in the “Text Only” mode. To change
over to Basic mode, si mu ltaneously hold down the "+"
and"-" keys as discussed in section 2.9.3.
2. Make sure the ZoneGuard system has been turned on
and operating for a sufficient period of time for the
sensor readings to have stabilized fully before initiating
manual calibration procedures . If the oxygen sensor
has just been replaced, the ZoneGu ard for oxygen
should be turned on and allowed to stabilize for one
hour pri o r to performing any of the calibration options.
3. Turn the instrument off by pressing and holding the
Mode button until Off is displayed.
4. While holding down the "Cal" bu tton, press the Mode
button to turn the ZoneGuard back on. Release the
button when the word “ SPAN” appears. The display
sh ould alternate between “SPAN” and the c urrent
oxygen reading (20.9 in fresh air). Pressing the Mode
button at any time cancels the span calibration.
↔
5. Screw the regulator into the cylinder and connect it to
the detector using the short section of tubing and the
calibration adapter. The regulator will automatically
begin flowing calibration gas at the correct flow rate as
soon as it is screwed into the cylinder of gas.
6. Allow the reading to stabilize. If the reading is a positive
number, use the “+” and “-“ keys to bring the value to
zero.
7. When finished, press and hold down the “Cal” key until
the display reads “DONE”.
8. Perform a fresh air zero adjustment as discussed in
section 3.6.1 when the instrument is next turned on.
Page 16
Page 18
CHAPTER 4. SETTING ALARM
LEVELS
ZoneGuard alarms are user adjustable and may be set
anywhere within the range of the sensor channel. When an
alarm set point i s exceeded a loud audible alarm sounds, the
appropriate LED is activated, and (if equipped) the optional
strobe light(s) and/or remote audible alarms are activated.
In addition, any other alarm devices controlled by means of
the ZoneGuard relays will also be activated.
ZoneGuard alarms are normally self resetting. When
readings drop below the alarm setpoint, the alarm ceases. It
is possible, if desired, to se t ZoneGuard alarms so that they
“latch”. In the latched condition, once an alarm occurs both
visual and audi ble alarms continue to sound even after the
atmospheric hazard has cleared. The instrument must be
manually reset by pressing the Mode button. Pressing the
Mode button s ilences th e alarms and restores n ormal
operation. Procedures for latching ZoneGuard alarms are
given in the next section.
6. Press the “Cal” button again to advance the display to
the next available alarm adjustment option, enabling or
disabling the warning level (Low) audible alarm. The
display alternates between a screen which shows the
message “AUD” and the status of the warnin g level
alarm, either “ON” or OFF”. Use the “+” or “-” buttons to
enable or disable this feature.
↔
7. Press the “Cal” button again to advance the display to
the next available alarm adjustment option, “latching”
the warning alarm. In the latched condition, the warning
alarms will continue to sound even after the alarm
condition has cleared. Latched alarms will continue to
sound until they are manually silenced by pressing the
Mode button. The display alternates between a screen
which shows the mess age “ L OCK” and the status of the
alarm latch condition, either “ON” or OFF”. Use the “+”
or “-” buttons to enable or disable th is feature.
4.1 Alarm adjustment sequence
It is necessary to be in the "Alarm Adjust" mode to change
alarm settings.
ZoneGuard default alarm settings are listed in Appendix
C.
Note: The order of these screens may be different
depending on software version and sensor type.
Instruments equipped with a toxic or combustible gas
sensor are equipped with a low “warning” alarm and a
high “danger” alarm. Instruments equipped with an
oxygen sensor include ascending and descending low
“warning” alarms and high “danger” alarms (a total of
four alarm settings).
To enter the alarm adjust mode:
1. Make sure gas readings are given in numbers. If the
display shows “OK”, then the ZoneGuard is currently
being operated in the “Text Only” mode. To change
over to Basic mode, simultaneously hold down the "+"
and"-" keys as discussed in section 2.9.3.
2. Turn the ZoneGuard off by pressing and holding the
Mode button until the LCD goes blank. When the button
is released the LCD will display the message “OFF”.
3. With the unit turned off, press and h o ld down the
"Alarm" button.
4. While holding down the "Alarm" button, press the Mode
button to turn the ZoneGuard back on.
5. A screen will appear showing the first alarm point to be
adjusted, the “Low” (warning level) al arm set point. Use
the "+" and "-" keys to raise or lower the Low al arm set
point.
↔
↔
→
↔
8. Press the “Cal” button again to advance the display to
the next availabl e alarm adjustment option, the high
level (Danger) alarm setpoint. Use the “+” and “-” keys
to raise or lower the High alarm set point.
↔
↔
9. Press “Cal” again to advance to the enable/disable
option for the Danger level alarm. Set as discussed in
Step 6.
10. Press “Cal” agai n to advance to the alarm latching
option for the “Danger” (high) alarm. Latching the
Danger level alarm is in the same way as latching the
warning level alarm (see step 7).
11. When all alarm adjustments have been completed,
press and hold the "Alarm" button down until an
information screen indicates that alarm adjustment is
DONE.
Caution: The “Alarm” button must be h e ld down until
the screen in dicates that alarm adjustment has been
successfully completed.
Failur e to pr ess and hold the “Alarm” button down until the
“DONE” message is displayed will cause the alarm settings
to revert to the way they were the last time the adjustment
was successfully completed.
Pressing the Mode button at any time cancels the alarm
adjustment mode, and returns the instrument to the gas
reading screen. A screen will announce “NO SAVE” after
which the instrument will turn itself off.
↓
→
Page 17
Page 19
↔
4.2 Restoring the factory default alarm
settings
ZoneGuard al arm settings are set conservatively at the
factory. (See Appendix C) It is possible to restore these
default settings at any time by doing the following.
1. Make sure gas readings are given in numbers. If the
display shows “OK”, then the ZoneGuard is currently
being operated in the “Text Only” mode. To change
over to Basic mode, simultaneously hold down the "+"
and"-" keys as discussed in section 2.9.3.
2. Press the "Alarm" button on the instrument key-pad.
The display will briefly show the default alarm message,
and the ty pe of sensor currentl y installed.
audible alarm, LED alarm light indicators, optional strobe
lights or horn, as well as other alarm devices or peripheral
equipment controlled by means of the ZoneGuard relay
contacts.
To use the “Alarm Test” function:
1. Make sure gas readings are given in numbers. If the
display shows “OK”, then the ZoneGuard is currently
being operated in the “Text Only” mode. To change
over to Basic mode, simultaneousl y hold down the "+"
and"-" keys as discussed in section 2.9.3.
2. Simultaneously press the “Alarm” and “Cal” buttons to
begin the test. The displ a y will show the message
screens, “ALRM TEST”, followed by numerical
readings.
→
→
→
→
3. This screen will be followed by several more showing
the specific default alarm settings. These screens will
be shown in continuous r o tation. In the case of a
ZoneGuard with a hydrogen sulfide sensor installed, the
following sequence of screens would be shown.
→
→
→
4. Push and hold the "Alarm" button to restore the factory
default alarm setting s. The display indicates when the
settings have been restored:
5. If the Mode button is pushed, the current alarm settings
remain unchanged, and the display shows the following
screen:
→
→
3. The readings shown on the ZoneGuard display will
slowly “ramp” up from zero to 10% above the high
alarm setpoint, then back down again. The audible an d
visual alarms will be activated as the readings exceed
the pre-set alarm levels. (In the case of ZoneGuard
systems equipped with an oxygen sensor, readings will
both ramp up through the “ascending” warning and
danger alarms for oxygen enrichment, as well as ramp
“down” in order to test the “descending” warning and
danger alarms for oxygen deficiency.)
4. The ZoneGuard will automatically return to normal
operation when the Alarm Test function has been
completed.
5. The Alarm Test can be terminated at any time by
pressing the Mode button.
↔
6. After selection of the default or current alarm settings
has been made, the display reverts to the current gas
reading screen.
4.3 Alarm Test function
ZoneGuard s oftware in cludes an “Alarm Test” function that
can be used to automatically “ramp” readings throughout the
full range of the s ensor. The “Alarm Test” function is used to
verify alarm setpoints as well as the proper activation of the
Page 18
Page 20
CHAPTER 5. ADVANCED
FUNCTIONS
5.1 ZoneGuard Advanced features
overview
The ZoneGuard microprocessor circuitry makes possible a
number of advanced features and capabilities. The four
push buttons located on the inside of the controller housin g
door are used to program advanced ZoneGuard functions.
The techn ique for usin g these buttons requires several
buttons to be pressed and h e ld at the same time.
Caution: Do not use this technique with any other
combination of buttons than those listed below. Doing
so may result in unintended changes to the instrument
setup.
5.2 Re-setting the calibration gas
concentration
During auto calibration adjustment the ZoneGuard adjusts
the sensitivity of the sensor to match the concentration of the
gas used to calibrate the sen sor. The value the instrument
uses to adjust the reading is the calibration gas
concentration stored in the instrumen t memory.
For instance, Biosystems normally supplies 50 ppm carbon
monoxide calibration gas for use in calibrating CO
ZoneGuard detectors. This is the default calibration gas
concentration stored in the instrumen t memory. When gas
is appli ed to the sensor during auto calibration, the
instrument automaticall y adjusts the calibration constants to
produce readings which match th is 50 ppm concentration.
SensorGas us ed
to calibrate
sensor
OxygenZero Air20.9%20.9%
Combustible
gas
Carbon
monoxide
Hydrogen
sulfide
Sulfur
dioxide
ChlorineCl25 ppm5 ppm
Chlorine
dioxide
Phosphi nePH35 ppm5 ppm
Nitrogen
dioxide
Table 5.2. Default calibration gas concentrations used
It is necessary to r e-set the calibra tion gas concentration
stored in instrument memory before making use of a
different calibration gas concentration.
Propane
Equivalent
CO50 ppm50 ppm
H2S25 ppm25 ppm
SO210 ppm10 ppm
Cl25 ppm1.7 ppm
NO25 ppm5 ppm
during auto-calibration adjustment
True cal
gas concentration
50% LEL50% LEL
Value us ed by
ZoneGuard to
adjust readings
Do not reset the calibration gas
concentration unless you are using a calibration gas
concen t ration which dif f e rs from that normally supplied
by Biosystems for use in calibrating your instrument.
Non-matching calibration gas and calibration gas value
settings will lead to inaccurate and potentially
dangerous readings.
Customers are strongly urged to use only Biosystems
calibration materials when calibrating their ZoneGuard
detector. Use of non-standard calibration gas and/or
calibration kit components can lead to dangerously
inaccurate readings and may void the standard
Biosystems warranty.
Use the following procedure to re-set the calibration gas
concentration:
1. Make sure gas readings are given in numbers. If the
display shows “OK”, then the ZoneGuard is currently
being operated in the “Text Only” mode. To change
over to Basic mode, simultaneousl y hold down the "+"
and"-" keys as discussed in section 2.9.3. 2. Turn the
instrument off by pressing the Mode button until the
display blanks.
3. While holding down the “+” button, press the Mode
button to turn the ZoneGuard back on.
4. The ZoneGuard LCD will display alternating “CAL” and
“GAS” messages for 4 seconds. The display will then
alternate between the current concentration stored in
memory and the “GAS” message screens.
5. Use the “+” and “-” buttons to change the concentration.
6. Once the proper new concentration has been selected,
press and hold down the “Alarm” button for three
seconds . Do not release the “ Alarm” button until the
LCD displays the message “DONE” to indicate that the
new value has been successfully entered.
7. Pressing the Mode button at any time cancels the
procedure. The LCD will show the message “NO
SAVE” to i ndicate that the calibration gas concentration
values have not been changed.
Turn on Press Mode.
Turn off Press and hold Mode until ZoneGuard shuts off.
Change operating mode Turn Zon eGuard on. Simultaneou sly press “+” and “-” to switch between
TEXT ONLY and BASIC operating modes.
Enable/disable use of Mode button Turn ZoneGuard on. Press and hold “+” to disable the Mode button. Press
and hold “+” to re-enable the Mode button.
Auto calibration Turn ZoneGuard on. In fresh air, press Mode 3 times to enter “Auto
Calibration Mode”. Press Mode to do Auto-Zero. When complete press
Mode to do Auto Span. Attach calibration adapter. Apply span gas. Remove
fittings when Auto Cal complete.
Manual fresh air calibration Turn ZoneGuard on. In fresh air, pres s CAL to enter fresh air “zero” mode.
Press CAL to do the fresh air zero; or press Mode to exit without doin g “zero”.
Manual span calibration Hold CAL down and tur n ZoneGuard on. Attach calibration adapter. Apply
span gas. Wait for the ZoneGu ard’s reading to stabilize. Use “+” and “-” to
adjust the reading to the value printed on the span gas con tainer. Press an d
hold CAL to save the span adjustment; or pr ess Mode to exit the manual span
without saving.
Set default alarms Turn ZoneGuard on. Press ALARM to enter “Default Alarm Mode”.
ZoneGuard will display the default alarm settings. Press and hold ALARM to
restore the factory defaul t alarm settings; or press Mode to exit without setting
the alarm levels to the default values.
Alarm adjustments Hold ALARM down and turn ZoneGuard on. Press CAL to switch to the alarm
option to be changed. Use “+” and “-” to adjust the alarm option to the desi r ed
value. Press and hold ALARM to save the changes to the ZoneGuard; or
press Mode to exit alarm adjustment mode without saving any changes.
Alarm test Turn ZoneGuard on. Press ALARM and CAL simultaneously to do an alarm
test. Press Mode to return ZoneGuard to normal operation.
APPENDICES
Appendix A ZoneGuard fixed gas detection system ordering information
54-17-A0101 Single integrally mounted external strobe (red)
54-17-A0102 Dual integrall y mounted external strobes (red and amber)
54-17-A0201 Remote programmable (volume adjustable) audible alarm
54-17-A0202 Remote high intensity audi ble alarm
54-17-A0301 Splash guard for sensors inst al led in explosion-proof remote housings
54-17-A0302 Splash guard for sensors installed in standard housings
54-17-A0401 Wire, 22 gauge, 6 conductor, twisted pair, shielded with drain, used to connect oxygen remote detectors
with ZoneGuard controller, 50 feet
54-17-A0402 Wire, 22 gauge, 4 conductor, twisted pair, shielded with drain, used to connect toxic remote detectors
with ZoneGuard controller, 50 feet
54-17-A0403 Wire, 18 gauge, 4 conductor, twisted pair , shi elded with drain, used to connect combustibl e remot e
detectors with ZoneGuard controller, 50 feet
Calibration Materials
54-17-K0301 Calibrati on ki t for ZoneGuard fixed detection systems. Includes foam l ined carrying case with room f or
up to 3 cylinders of gas, one standard 1. 0 LPM fi xed flow rate regulator , calibration adapt or s, and 3 f eet
of tubing. Kit does not include gas.
54-17-A0303 Calibrati on adapt or f or sensors i nst al led in standard housings
54-17-A0304 Calibrati on adapt or for sensors installed in explosion-proof housi ngs
54-9031 Methane 50% LEL in air, 103 liters
54-9032E Propane 50% LEL equivalent in air, 103 liters
54-9033 Carbon monoxide (CO), 50 ppm in air, 103 l iters
54-9034 Hydrogen sulfide (H2S), 25 ppm i n ni trogen, 58 liters
54-9037 Sulfur dioxide (SO2), 10 ppm i n nitrogen, 58 liters
54-9052 Chlorine (Cl2), 5 ppm i n nitrogen, 58 liters
54-9055 Nitric oxide (NO), 30 ppm in nit r ogen, 58 l i t ers
54-9056 Nitrogen dioxide (NO2), 5 ppm i n nitrogen, 58 liters
54-9065 Phosphine (PH3), 5 ppm i n nitrogen, 58 liters
54-9039 Zero Air, 20.9% oxygen in nitrogen, 103 liters
Appendix B ZoneGuard fixed gas detection system technical specifications
ZoneGuard VAC Outputs Warning (Low) and Danger (high) alarm, DPDT relay - contacts rated 5 Amp at 250 VAC, non-
inductive load
ZoneGuard VDC Outputs Warning (Low) and Danger (high) alarm +12 VDC at 0.5 Amp
ZoneGuard Display 8 character 14-segment backlit LCD for gas r eadings and other messages
ZoneGuard Indicators
(LED)
Audible alarm Built-i n audi ble alarm (85 dB at 3 feet)
Power Requirements 110 VAC - 50/60 Hz
Temperature Range
Humidit y Range 0 to 100% RH
Classification for use in
hazardous locations
Weather Protection Contr oller housing: NEMA 4X enclosure (IP rating of 654)
Dimensions ZoneGuard controller: 10” (25 cm) x 7.5” ( 19 cm) x 5.25” (13 cm )
Wi r i ng: For oxygen remote detectors use (minimum) 22 gauge AWG, 6 conductor twisted-pair shielded
Power on: green; low alarm: amber; high alarm: red
System components designed to function over range of -40° F to 110° F (-40° C to 43° C)
(Type of sensor installed may modify range.)
Sensors installed in explosion-proof rem ot e housings: Submitted for Classi fication by ULCas
Explosion-Proof for Class 1, Di vision 1, Groups A, B, C, D, E, F, anf G hazardous locations.
with drain. For t oxic use (minim um) 22 gauge AWG, 4 conductor twisted-pair shielded with
drain. For combustible use 18 gauge AWG, 4 conduct or t wi sted-pair shielded with drain.
Page 21
Page 23
Appendix C ZoneGuard default alarm settings
Gas Warning Level Danger Level
Combustible gas 10% LEL 20% LEL
Oxygen Low: 19.5%
High:22.0%
Carbon monox ide 35 PPM 100 PPM
Hydrogen sulfide 10 PPM 20 PPM
Sulf ur dioxi de 2.0 PPM 5.0 PPM
Nitrogen di oxide 1.0 PPM 5.0 PPM
Chlorine 0.5 PP M 1.0 PPM
Chlorine dioxide 0.3 PPM 1.0 PPM
Phosphine 0.3 PPM 1.0 PPM
Appendix D ZoneGuard sensor ranges
Gas Range
Combustible gas 0-100% LE L
Oxygen 0-25%
Carbon monox ide 0-1000 PPM
Hydrogen sulfide 0-200 PPM
Sulf ur dioxi de 0-100 PPM
Nitrogen di oxide 0-20 PPM
Chlorine 0-20 PPM
Chlorine dioxide 0-10 PPM
Phosphine 0-10 PPM
Low: 18.0%
High:23.5%
Appendix E ZoneGuard toxic sensor cross sensitivity data
The table below lists the cross sensitivity of electrochemical t oxic sensors used in Biosystems por t abl e gas detectors to gases
other than their target gas. Depending on the nature of the reaction each gas has with the sensor, the effect can either
decrease the signal (negative cross sensitivit y) or i ncr ease the signal; (posit i ve cross sensitiv i t y) . Each figure represents the
reaction of the sensor to 100 ppm of gas, t hus pr oviding a percentage sensitivity to that gas relat i ve to its target gas.
Appendix F ZoneGuard dimensional drawings, electrical schematics, and wiring
diagrams
Dimensional drawing of ZoneGuard controller
Page 23
Page 25
Dimensional drawing of ZoneGuard explosion-proof remote detector assembly
Page 24
Page 26
Dimensional drawing of ZoneGuard standard (non-explosion-proof) remote detector assembly
Page 25
Page 27
Appendix G Calibration
Frequency
One of the most common questions
that we are asked at Biosystems is:
“How often sh ould I calibr ate my g as
detector?”
Sensor Reliability and Accuracy
Today’s sensors are designed to
provide years of reliable service. In
fact, m any sensors are designed so that
with normal use they will only lose 5%
of their sensitivit y per year or 10% over
a two-year period. Given this, it should
be possible to use a sensor for up to
two full years without any significant
loss of sensitivity.
A lot of sensors indeed do last that long
with only minimal loss of sensitivity.
However, there are a number of
reasons why a sensor may
unexpectedly lose additional sensiti v i t y
or even fail to respond to gas. Such
reasons include desiccation, poisoning,
physical restriction of airfl ow,
overexposure, leakage, and mechanical
damage due to dropping or immersion.
Verification of Accuracy
Wi t h so many reasons why a sensor
can lose sensitivity and given the fact
that dependable sensors can be key to
survival in a hazardous environment,
frequent verification of sensor
performance is paramount.
There is only one sure way to verify that
a sensor can respond to the gas for
which it is designed. That i s t o expose it
to a known concentration of tar get gas
and compare the reading with the
concentration of the gas. This is
referred to as a “bump” test. This test is
very simple and takes only a few
seconds to accomplish. The saf est
course of action is to do a “bump”
test prior t o each day’s use. It is not
necessary to make a calibration
adjustment i f t he readings are between
90% and 120% of the expected value.
As an example, if a CO sensor is
checked using a gas concentration of
50 PPM it is not necessary to perform a
calibrati on unl ess the readings are
either below 45 PPM or above 60 PPM.
Lengthening the Intervals
between Verification of
Accuracy
We are often asked whether there are
any circumst ances in which the period
between accuracy checks may be
lengthened.
Biosystems is not t he only
manufactur er to be asked this
question! One of the professional
organizations to which Biosyst em s
belongs is the Industrial Saf ety
Equipm ent As s oc iation ( ISEA). The
“Instrument Products” group of this
organization has been very active in
developing a protocol to clari fy the
mini mum conditions under which the
interval between accuracy checks may
be lengthened.
A number of leading gas detection
equipment manufact ur ers have
participat ed in the development of the
ISEA guidelines concerning c alibrat ion
frequency. Biosystems procedures
cl osel y fo llow t hese guideli nes.
If your operating procedures do not
permit daily checking of the sensors,
Biosystems recommends the following
procedure to establish a safe and
prudent accuracy check schedule for
your Biosystems instr uments:
1. During a period of initial use of at
least 10 days in the intended
atmosphere, check the sensor
response daily to be sure there is
nothing in the atmosphere that is
poisoning the sensor(s). The
period of initi al use must be of
sufficient durat i on t o ensure that
the sensors are exposed to all
conditions that might have an
adverse effect on the sensors.
2. If t hese tests demonstrate that it
is not necessary to make
adjustments, the time between
checks may be lengthened. The
interval between accuracy
checking should not exceed 30
days.
3. When the interval has been
extended the toxic and
combustible gas sensors should
be replaced immediately upon
warranty expiration. This will
minimize the risk of failur e during
the interval between sensor
checks.
4. The history of t he inst r ument
response between verifications
should be kept. Any conditions,
incidents, experiences, or exposure
to contami nants that might have an
adverse effect on the calibration
state of the sensors should trigger
immediate re-verification of
accuracy before further use.
5. Any changes in the environment in
which the instrum ent is being used,
or changes in the work that is being
performed, should t r i gger a
resumption of dai ly checking.
6. If t here is any doubt at any t ime as
to the accuracy of the sensors,
verify the accuracy of the sensors
by exposing them to known
concentration test gas before
further use.
Gas detectors used for the detection of
oxygen deficiencies, fl ammable gases
and vapors, or toxic contaminants m ust
be maintained and operated properly to
do the job they were designed to do.
Always follow the guidelines provided
by the manufacturer for any gas
detection equipment you use!
If there is any doubt regarding your gas
detector's accuracy, do an accuracy
check! All it takes is a few mom ents to
verify whether or not your instruments
are safe to use.
One Button Auto Calibration
Whi l e it is only necessary to do a
“bump” test to ensure that the sensors
are working properly, all current
Biosystems gas detectors offer a one
button auto cali br ation feature. This
feature allows you to calibr at e a
Biosystems gas detector in about the
same time as it takes to complete a
“bump” t est. The use of autom atic
bump test and cali bration stations can
further simpl i f y t he tasks, whil e
automati c al ly maint ai ni ng records.
Don't take a chance
with your life.
Verify accuracy frequently!
Biosystems Applications Notes are
available at the Biosystems website at
http://www.biosystems.com
Page 28
BIOSYSTEMS STANDARD WARRANTY GAS DETECTION PRODUCTS
General
Biosystems LLC (hereafter Biosystems) warrants gas detectors, sensors and accessories manufactured and sold by
Biosystems, to be free from defects in materials and workmanship for the periods listed in the tables below.
Damages to any Biosystems products that result from abuse, alteration, power fluctuations including surges and lightning
strikes, incorrect voltage settings, incorrect batteries, or repair procedures not made in accordance with the Instrument’s
Reference Manual are not covered by the Biosystems standard warranty.
The obligation of Biosystems under this warranty is limited to the repair or replacement of components deemed by the
Biosystems Instrument Service Department to hav e been defective under the scope of this standard warranty. To receive
considerati on for warranty repair or replacement procedures , products must be returned with transportation and shipping
charges prepaid to Biosystems at its manufacturing location in Middletown, Connecticut, or to a Biosystems Authorized
Warranty Service Center. It is necessary to obtain a return authorization number from Biosystems prior to shipment.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESEN TATIONS,
EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF FITNESS FOR A PARTICULAR
PURPOSE. BIOSYSTEMS WILL NOT BE LIABLE FOR LOSS OR DAMAGE OF ANY KIND CONNECTED TO THE
USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION O R OPERATE PROPERLY.
Instrument & Accessory Warranty Periods
Product(s) Warranty Period
PhD5, PhD Lite, PhD Plus, PhD Ultra, Cannonball3, MultiVision,
Toxi, Toxi/Oxy Plus, Toxi/Oxy Ultra, ToxiVision, Ex Chek
ToxiPro®, MultiPro
ToxiLtd®
Mighty-Tox
Mighty- Tox 2
Prorated credit is given towards repair or purchase of a new unit of
the same type.
IQ Systems, Series 3000, Airpanel, T r avel panel, ZoneGuar d,
Gas9Chek1 and Gas9Chek4
Battery pac k s and charger s, sampling pumps and other
components, whic h by their design are consumed or depleted during
normal operation, or whic h may requir e per iodic replacement
As long as the instrument is in service
2 years from date of pur c hase
2 years after ac tivation or 2 year s after
the “Must Be Activated By” date,
whichever comes first
90 days after ac tivation or 90 days after
the “Must Be Activated By” date,
whichever comes first
0 – 6 months of use 100% c r edit
6 – 12 months of use 75% credit
12 – 18 months of use 50% credit
18 – 24 months of use 25% credit
ToxiVision, Toxi Pro
Toxi, Toxi/Oxy Plus, Toxi/Oxy Ultra
All O thers All Sensors 1 Year
** Damage to combustible gas sensors by acute or chronic exposure to known sensor poisons such as volat i l e lead
(aviation gasol i ne addi tive), hydride gases such as phosphine, and vol atile silicone gases emitted from silicone
caulks/sealants, silicone rubber molded product s, laboratory glassware greases, spray lubri cant s, heat transfer
fluids, waxes & poli shi ng compounds (neat or spray aerosols) , mold rel ease agent s for plastics injection mol ding
operations, water proofing form ul at ions, vinyl & leather preser vat ives, and hand lotions which may contain
ingredients l i st ed as cyclomethicone, di methicone and polym et hi cone (at the discretion of Biosystems Inst r ument
Service departm ent) void Biosystems’ St andar d Warrant y as it appl i es t o the replacement of combustible gas
sensors.
®
, Ex Chek
O2, LEL**, CO, CO+, H2S
& Duo-Tox
2 Years
All O ther Sensors 1 Year
CO, CO+, H2S 2 Years
All O ther Sensors 1 Year
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.