651 South Main Street
Middletown, CT 06457 USA
(800) 711-6776 (860) 344-1079
Fax (860) 344-1068
http://www.biosystems.com
Version 4.11 p/n 13-088
19APR2007
Page 2
1
Page 3
THE PhD LITE PERSONAL PORTABLE GAS DETECTOR HAS BEEN DESIGNED FOR
THE DETECTION OF DEFICIENCIES OF OXYGEN, ACCUMULATIONS OF FLAMMABLE
GASES AND VAPORS AND ACCUMULATIONS OF TOXIC VAPORS.
IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF
POTENTIALLY DANGEROUS ATMOSPHERIC CONDITIONS, IT IS
ESSENTIAL THA T THE INSTRUCTIONS IN THIS REFEREN CE MANUAL
BE READ, FULLY UNDERSTOOD, AND FOLLOWED.
PhD Lite Reference Manual
Version 4.11
Copyright 2007
by
Biosystems
Middletown, Connecticut 06457
All rights re served .
No page or part of this operation manual may be reproduced in any form
without written permission of the copyright owner shown above.
All rights reserved to correct typographical errors.
2. Time Weighted Average (TWA):------------------------------------------------------------------------------------------------------ 62
3. Short Term Exposure Limits (STEL): ------------------------------------------------------------------------------------------------62
Appendix B PhD Lite sensor ranges ----------------------------------------------------------------------------------------- 63
Appendix C Sensor Cross-Sensitivity Chart------------------------------------------------------------------------------- 64
Appendix D Calibration Frequency------------------------------------------------------------------------------------------- 65
Appendix E Biosystems Standard Warranty Gas Detection Products -------------------------------------------- 66
The following signal words, as defined by ANSI Z535.4-1998, are used in the PhD Lite
Reference Manual.
indicates an imminently hazardous situation which, if not avoided,
will result in death or serious injury.
indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
indicates a potentially hazardous situation, which if not avoided, may
result in moderat e or minor inju ry.
CAUTION used without the saf ety alert symbol indicates a pot entially hazardous
situation which, if not avoided, may result in property d amage.
Signal Words
5
Page 8
1. The PhD Lite personal, portable gas detector has been designed for the detection of
dangerous atmospheric conditions. An alarm condition indicates the presence of a potentially life-threatening
hazard and should be taken very seriously.
2.
safest course of action is to immediately leave the affected area, and to return only after further testing
determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious
injury or death.
3.
Radio Shack 23-874 (not ATEX approved) size AA 1.5V Alkaline batteries, Eveready CH15 (not ATEX approved) or
Radio Shack 23-149 (not ATEX approved) size AA 1.2V NiCad batteries, or Eveready L91 size AA 1.5V Lithium
batteries (not CSA or ATEX approved). Substitution of batteries may impair intrinsic safety.
4.
calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. (The
Canadian Standards Association (CSA) requires an accuracy check using known concentration calibration gas
prior to each day’s use.)
5.
to contaminants by testing with known concentration test gas before further use. Failure to check accuracy can
lead to inaccurate and potentially dangerous readings.
6.
immediately. An instrument that fails calibration may not be used until testing with known concentration test gas
determines that accuracy has been restored, and the instrument is once again fit for use.
7.
concentration that differs from the one that is normally supplied by Biosystems for use in calibrating the PhD Lite.
Customers are stro ngly urged to use only Biosyst ems calibration materials when calibrati ng t he PhD Lite. Use of
non-standard calibration gas and/or calibration kit components can lead to dangerously inaccurate readings and
may void the standard Biosystems warranty.
8. Use of non-standard calibration gas and/or calibration kit components when calibrating the
PhD Lite can lead to inaccurate and potentially dangerous readings and may void the standard Biosystems
warranty.
Biosystems offers calibration kits and long-lasting cylinders of test gas specifically developed for easy PhD Lite
calibrati on. Custome rs are str ongly urged to us e only Biosystems ca l ibratio n materials whe n calibrating t he PhD
Lite.
9.
10.
only. Read and understand this reference manual before operating or servicing the PhD Lite.
11.
combustible gas concentration that exceeds the PhD Lite’s zero to 100 percent LEL detection range.
In the event of an alarm condition it is important to follow established procedures. The
Use only Duracell MN1500 or Ultra MX1500, Eveready Energizer E91-LR6,
The accuracy of the PhD Lite should be checked periodically with known concentration
The accuracy of the PhD Lite should be checked immediately following any known exposure
A sensor that cannot be calibrated or is found to be out of tolerance should be replaced
Do not reset the calibration gas concentration unless you are using a calibration gas
Substitution of components may impair intrinsic safety.
For safety reasons this equipment must be operated and serviced by qualified personnel
A rapid up-scale reading followed by a declining or erratic reading may indicate a hazardous
Use only the following Biosystems chargers to recharge the PhD Lite:
Use only the following pumps with the PhD Lite:
Eveready EN91,
6
Page 9
Chapter 1. Description
1.1 PhD Lite capabilities
The PhD Lite gas detector can be configured to meet a
wide variety of requirements. This chapter provides an
overview of many of the features of the PhD Lite. More
detailed descriptions are contained in the subsequent
chapters of this manual.
1.2 Methods of sampling
The PhD Lite may be used in either diffusion or sampledraw mode. In either mode, the gas sample must reach
the sensors for the instrument to register a gas reading.
The sensors are located inside of the instrum ent.
In diffusion mode, the atmosphere being measured
reaches the sensors by diffusing through vents in the
sensor compartment cover. Normal air movements are
enough to carry the sample to the sensors. The sensors
react quickly to changes in the concentrations of the gases
being measured. Diffusion-style operation monitors only
the atmosphere that immediately surrounds the detector.
The PhD Lite can also be used to sample remote locations
with either the hand-aspirated sample-draw kit that is
included with every PhD Lite, or with a motorized
continuous sample draw pump that is available separately.
During remote sampling, the gas sample is drawn into the
sensor compartment through the probe assembly and a
length of tubing.
Use of the sample draw kits is covered in section
2.5.1.
A detailed description of the PhD Lite probe assembly
is given in section 2.5.3.
1.3 Multi-sensor capability
The PhD Lite can be configured to simultaneously monitor
oxygen, combustible gases and vapors and either one or
two toxic gases. Sensors can be added, removed,
changed, and replaced in the field. The PhD Lite
microprocessor and “Smart Sensor” circuitry eliminates
the need for laborious reconfiguration procedures.
Note: It is necessary to verify the accuracy of the PhD
Lite by calibration with known concentration test gas
whenever a change is made to the sensors installed in
the instrument.
Calibration procedures are discussed in detail in
Chapter 4.
The PhD Lite uses electrochemical toxic gas sensors that
have been designed to minimize the effects of common
interfering gases. These sensors provide accurate,
dependable readings for toxic gases commonly
encountered during confined space entry and other
industrial applications. Toxic sensors currently available
for use in the PhD Lite include hydrogen sulfide (H
carbon monoxide (CO), sulfur dioxide (SO
In addition to sensors designed to measure specific toxic
hazards, Biosystems also offers two different sensors
that allow for the simultaneous detection of both CO and
S.
H
2
The “Duo-Tox” sensor is a dual channel electrochemical
sensor designed to simultaneously detect both carbon
monoxide and hydrogen sulfide. The Duo-Tox allows
one sensor port to be used for the direct detection of both
carbon monoxide and hydrogen sulfide.
For more information on the Duo-Tox sensor see
section 4.3.1.1.
The “CO Plus” sensor is a single channel electrochemical
sensor. The CO Plus is ideal for situations requiring the
use of a single sensor to monitor for both Carbon
Monoxide and Hydrogen Sulfide, but where the user
does not need to know specifically which gas is present.
For more information on the CO Plus sensor see
section 4.3.1.2.
Different measurement units are used depending on the
gas being measured.
Type of Hazar d Measurement unit
Oxygen (O2) Percent age by volume
Combustible gas Percentage of lower
explosive limit (%LEL)
All toxic sensors Parts per million (PPM)
Table 1.3. PhD Lite Standar d Units of M eas ur e me n t
Sensor configuration procedures are discussed in
greater detail in Chapter 2.
1.4 Calibration
The PhD Lite detector features fully automatic fresh air
and span calibration.
The accuracy of the PhD Lite
should be checked periodically with known
concentration calibration gas. Failure to check
accuracy can lead to inaccurate and potentially
dangerous readings. (The Canadian Standards
Association (CSA) requires an accuracy check using
known concentration calibration gas prior to each
day’s use.)
Calibration procedures are discussed in detail in
Chapter 4.
Use of these procedures is reserved for authorized
personnel.
1.5 Alarm logic
PhD Lite gas alarms are user-adjustable and may be set
anywhere within the range of the specific sensor type.
When an alarm set point is exc eeded a loud audible
alarm sounds, and the bright red LED alarm light flashes.
The procedure for adjusting alarm settings is
covered in section 3.2.2.
1.5.1 Alarm latch
The PhD Lite includes a latching alarm feature that can
be enabled or disabled according to the needs of the
user. When the PhD Lite’s alarm latch is enabled, the
7
Page 10
audible and visible alarms will continue to sound even
after the atmospheric hazard has cleared. To turn the
alarm off, simply press the MODE button. If the PhD Lite’s
alarm latch is disabled and the alarm condition is no
longer present, the instrument will automatically return to
normal operation, and the visible and audible alarms will
cease without further input from the user.
The procedure for changing alarm and OK latch
settings is covered in section 3.2.3.
1.5.2 Atmospheric hazard alarms
The PhD Lite portable gas
detector has been designed for the detection of
deficiencies of oxygen, accumulations of flammable
gases and vapors, and accumulations of toxic vapors.
An alarm condition indicating the presence of one or
more of these potentially life-threatening hazards
should be taken very seriously.
In the event of an alarm condition
it is important to follow established procedures. The
safest course of action is to immediately leave the
affected area, and to return only after further testing
determ i nes that th e area i s once a gain safe f or entry.
Failure to immediately leave the area may result in
serious injury or death.
A rapid up-scale reading followed
by a declining or erratic reading may indicate a
hazardous combustible gas concentration that
exceeds the PhD Lite’s zero to 100 percent LEL
detection range.
The combustible gas alarm is activated when the percent
LEL (Lower Explosive Limit) gas concentration exceeds
the pre-set alarm point.
Two oxygen alarm set points have been provided; one for
low concentrations associated with oxygen deficiencies
and one for high concentrations associated with oxygen
enrichm ents.
Three alarm set points have been provided for each toxic
gas that is monitored: TWA (Time Weighted Average),
STEL (Short Term Exposure Limit), and Ceiling.
Appendix A discusses alarm levels and factory default
alarm settings.
The procedure for adjusting alarm settings is covered
in section 3.2.2.
1.5.3 Sensor overrange alarms.
The PhD Lite toxic gas channel will go into alarm if a
sensor is exposed to a concentration of gas that exceeds
its established range. If the peak alarm is enabled it will
go off before an overrange alarm provided a STEL or TWA
alarm is not activated first. If the peak alarm is disabled
and a toxic sensor goes into overrange alarm a “SENSOR
OUT OF RANGE” message will appear at the bottom of
the display while the audible and visible are activated.
The maximum range value will be displayed for the sensor
in alarm. If the LEL sensor goes into overrange alarm, the
message “LEL OVERRANGE” will intermittently appear,
both the audible alarm and the flashing LED alarms will be
activated and an “X” will appear on the LCD in the place
of the numeric reading for the LEL sensor.
In the event of an LEL overrange
alarm the PhD Lite must be turned off, brought to an
area that is known to be safe and then turned on
again to reset the alarm.
Note: The PhD Lite features automatic warning
against LEL sensor response failure due to lack of
oxygen. When oxygen levels fall below 10% of gross
volume, the PhD Lite will intermittently display a
message indicating that O
operate.
1.5.4 Low battery alarms
is too low for LEL to
2
Note: The voltage figures given below are for PhD
Lite instruments with firmware version 3.18 and may
be slig htly diff eren t for units wit h othe r vers ions of
instrument firmware.
The PhD Lite includes low battery alarms that are
activated whenever battery voltage approaches a level
that will soon lead to instrument shut down. When the
battery voltage is reduced to approximately 3.23 volts, an
audible alarm will sound and the display will indicate that
a low battery condition exists. At this stage, the low
battery alarms may be silenced for a fifteen-minute
period by pressing the MODE button. After the first low
battery alarm, the alarm will sound again every fifteen
minutes until the voltage drops to the “Very Low Battery”
level.
The “Very Low Battery” level occurs when the battery
voltage drops to 3.13 volts. Due to the risk of imminent
shut down, when the battery voltage reaches the “Very
Low Battery” level it is no longer possible to silence the
low battery alarms. At this point, it is necessary to
immediately leave the hazardous area in which the
instrum ent is being used.
When the voltage drops to 3.10 volts, the PhD Lite will
display a "Dead Battery" message to warn the user of
imminent shut down. The instrument will then
automatically turn itself off.
After any low battery alarm the batteries should be
replaced if the PhD Lite is equipped with alkaline
batteries or the battery should be recharged if the PhD
Lite is equipped with a NiMH rechargeable battery.
The PhD Lite must be located in a
non-hazardous location whenever the alkaline
batteries are removed from the instrument.
Removing the alkaline batteries from the instrument
in a hazardous location may impair intrinsic safety.
Use only Duracell MN1500 or
Ultra MX1500, Eveready Energizer E91-LR6, Eveready
EN91, Radio Shack 23-874 (not ATEX approved) size
AA 1.5V Alkaline batteries, Eveready CH15 (not ATEX
approved) or Radio Shack 23-149 (not ATEX
approved) size AA 1.2V NiCad batteries, or Eveready
L91 size AA 1.5V Lithium batteries (not CSA or ATEX
approved). Substitution of batteries may impair
intrinsic safety.
8
Page 11
1.5.5 Other alarms and special microprocessor
features
PhD Lite software includes a number of additional alarms
designed to ensure the proper operation of the instrument.
When the PhD Lite detects that an electronic fault or
failure condition has occurred, the proper audible and
visible alarms are activated and an explanatory message
is displayed.
The PhD Lite is designed to detect
potentially life threatening atmospheric conditions.
Any alarm condition should be taken seriously. The
safest course of action is to immediately leave the
affected area, and return only after further testing
determ i nes that th e area i s once a gain safe f or entry.
1.6 Other electronic safeguards
Several automatic programs prevent tampering and
misuse of the PhD Lite by unauthorized persons. Each
time the detector is turned on, the PhD Lite automatically
tests the LED alarm lights and audible alarm. The battery
is monitored continuously for proper voltage. The PhD
Lite also monitors the connection of sensors that are
currently installed. The detection of any electronic faults
causes the activation of the audible and visible alarms and
causes the display of the appropriate explanatory
message.
1.6.1 Security beep
The PhD Lite offers a security beep that may be
configured to “beep” at defined intervals to indicate that
the instrument is turned on.
Optional set-up choices, including security beep
settings, are accessed through the Options Menu.
See section 3.4.3 for details on the security beep.
1.7 Classification for intrinsic safety
The PhD Lite has the following certifications for intrinsic
safety:
UL Class I, Division 1, Groups A, B, C, and D
UL Class ll, Division 1, Groups E, F, and G
CSA Class I, Division 1, Gr oups A, B, C, and D
ATEX Certific ation:
II 2G EEx ia d IIC T4 (Alkalin e ) / EEx ia d IIC T3 (NiMH)
UL International DEMKO A/S 03 ATEX 130960X
The PhD Lite is also CE-Marked (Conformité
Européenne). The UL, CSA, and ATEX cert ification
marks indicate compliance with United States,
Canadian, and European Community requirements.
These classifications extend to instrument when
operated with the motorized pump .
1.8 Sensors
The PhD Lite can be configured to simultaneously monitor
Oxygen, combustible gases and vapors and either one or
two toxic gases. Up to three sensors can be installed in
the PhD Lite. With the “Duo Tox” dual channel CO/H
S
2
sensor installed, the instrument is capable of displaying
readings for up to four different channels of detection: O
LEL, CO, and H
S. The sensor configuration of the PhD
2
2
Lite can be specified at the time of purchase, or changed
in the field by appropriately trained personnel.
A sensor that cannot be
calibrated or is found to be out of tolerance must be
replaced immediately. An instrument that fails
calibration may not be used until testing with known
concentration test gas determines that accuracy has
been restored, and the instrument is once again fit
for use.
Calibration procedures are discussed in detail in
Chapter 4.
1.9 Continuous sample draw pump
An optional slip-on, motorized sample-draw pump is
available for situations requiring continuous "hands free"
remote monitoring.
The pump contains a pressure sensor that detects
restrictions in airflow caused by water or other fluids
being drawn into the unit and immediately acts to turn the
pump off in order to protect the sensors, pump, and other
PhD Lite components from damage.
Pump status is continuously monitored by the PhD Lite
microprocessor. A spinning “
” icon located in the upper
left hand corner of the LCD display indicates that the
pump is attached and in normal operation. Low flow or
other pump fault conditions activate audible and visible
alarms and cause the display of the appropriate
explanatory message.
1.10 Black box data recorder
Starting with instrument firmware version 1.29, every
PhD Lite purchased without a datalogger has included a
“black box” data recorder that functions similarly to a
datalogger, with one important distinction. The data
stored in a PhD Lite with a datalogger option can be
accessed directly by the user with Biosystems’ Biotrak
Database Software. Instruments with the black box data
recorder must be sent back to Biosystems for data
retrieval.
If the data stored in a PhD Lite equipped with a black box
data recorder is needed, simply call Biosystems’
Instrument Service Department at (860) 344-1079 for a
return authorization number and send the instrument
back to Biosystems. Biosystems will extract the data
from the instrument and print an incident report at no
charge. You only pay for shipping.
Datalogging functions are discussed in detail in
Chapter 5.
1.11 Remote Vibrating Alarm
The PhD Lite can be equipped with an optional remote
vibrating alarm for use in noisy environments. The MHP1 or MHP-2 remote alarm is designed to be worn on a
belt or shirt collar or placed inside a pocket, and is
connected to the instrument via a 20" or 40" cable.
,
9
Page 12
During an alarm condition the MHP-1 will vibrate
continuously until the alarm condition is resolved.
Note: To use the MHP-1 or MHP-2 remote alarms, the
PhD Lite must be equipped with the remote alarm
connector.
1.12 PhD Lite design components
(1) Case: The instrument is enclosed in a solid, metal-
plated ABS case. A black PVC gasket between the
upper and lower sections of the case protects against
leakage or exposure to liquids.
(2) Front face: The front face of the instrument houses
the graphics capable LCD display, MODE button, four
navigation arrows, alarm light/IrDA port, and audible
alarm port.
(3) LCD display: A graphics capable liquid crystal
display (LCD) meter allows display of readings,
messages, menu and other information. An
automatically activated backlight allows the display to
be read even in low light conditions.
(4) Alarm lig h t / IrDA p o r t: Two bright red LED (light-
emitting diode) alarm lights provide a visual indication
of the alarm state. The lights are visible from the front
and top of the instrument. The light also functions as
the IrDA link to the unit for uploading to and
downloading from a PC.
(5) Control buttons: The large rectangular push-button
is called the MODE button. It is used to turn the PhD
Lite on and off, to confirm menu selections and to
access the automatic calibration sequences. The four
triangular navigation arrows are used to scroll or
navigate through additional menu choices and
screens of information.
(6) Sensor compartment cover: The sensors are
protected by a vented sensor compartment cover.
Each sensor is individually protected by its own
protective filter cap.
(7) Audible alarm: A cylindrical resonating chamber
extending past the sensor compartment contains the
loud audible alarm. Three small circular apertures in
the front face of the instrument cover the resonating
chamber.
(8) Battery compartment cover: For alkaline units, the
battery compartment cover provides access to the set
of three AA alkaline batteries. The cover is held in by
a knurled retention screw.
(9) Bottom surface: The belt clip attaches to the bottom
surface of the instrument.
1.13 PhD Lite standard accessories
Standard accessories included with every PhD Lite include
sample draw / calibration adapter, hand-aspirated sampledraw kit, 2 feet of additional tubing for use during
calibration, reference manual, quick reference card and
training CD-ROM.
The sample draw kit consists of a sample draw /
calibration adapter, squeeze bulb, sample probe,
replacement sample probe filters, and ten feet of tubing.
Standard configurations of the PhD Lite are delivered in a
foam-lined box.
1.13.1 Alkaline PhD Lite detectors
If the PhD Lite has been purchased as an alkaline
instrument, the standard accessories also include a set of
3 disposable AA alkaline batteries.
For domestic orders, the batteries will be pre-installed in
the PhD Lite at the time of delivery (unless otherwise
requested).
For some international orders, the batteries may be
removed from the instrum ent prior to shipment.
1.13.2 NiMH PhD Lite detectors
If the PhD Lite has been purchased as a “NiMH”
instrument, the standard accessories also include a slip
in PhD Lite fast charger.
1.14 PhD Lite kits
PhD Lite detectors may also be purchased as part of a
complete kit that includes calibration gas, calibration gas
regulator and a hard-shell carrying case.
1.14.1 PhD Lite Confined Space Kits
In addition to the standard accessories listed above,
Confined Space Kits also include calibration fittings,
fixed-flow regulator with pressure gauge, and appropriate
large cylinder(s) of calibration gas in a foam-lined,
waterproof hard-shell carrying case.
1.14.2 PhD Lite Value Packs
PhD Lite Value Packs include an alkaline PhD Lite, all
standard accessories, calibration fittings, small
cylinder(s) of calibration gas, and fixed flow rate regulator
in a foam-lined non-waterproof hard-shell carrying case.
Figure 1.14.2 Major PhD Lite Features (Top and Front
Surfaces)
10
Page 13
Chapter 2. Basic operation
This chapter will cover how to use the PhD Lite for safe
work in potentially hazardous atmospheres.
2.1 Operational warnings and cautions
1. The PhD Lite personal,
portable gas detector has been designed for the
detection of dangerous atmospheric conditions.
An alarm condition indicates the presence of a
potentially life-threatening hazard and should
be taken very seriously.
2.
In the event of an alarm
condition it is important to follow established
procedures. The safest course of action is to
immediately leave the affected area, and to
return only after further testing determines that
the area is once again safe for entry. Failure to
immediately leave the area may result in serious
injury or death.
3.
Use only Duracell MN1500
or Ultra MX1500, Eveready Energizer E91-LR6,
Eveready EN91, Radio Shack 23-874 (not ATEX
approved) size AA 1.5V Alkaline batteries,
Eveready CH15 (not ATEX approved) or Radio
Shack 23-149 (not ATEX approved) size AA 1.2V
NiC ad batteries, or Eveready L91 size AA 1.5V
Lithium batteries (not CSA or ATEX approved).
Substitution of batteries may impair intrinsic
safety.
4.
The accuracy of the PhD
Lite should be checked periodically with known
concentration calibration gas. Failure to check
accuracy can lead to inaccurate and potentially
dangerous readings. (The Canadian Standards
Association (CSA) requires an accuracy check
using known concentration calibration gas prior
to each day’s use.)
5.
The accuracy of the PhD
Lite should be checked immediately following
any known exposure to contaminants by testing
with known concentration test gas before
further use. Failure to check accuracy can lead
to inaccurate and potentially dangerous
readings.
6.
A sensor that cannot be
calibrated or is found to be out of tolerance
should be replaced immediately. An instrument
that fails calibration may not be used until
testing with known concentration test gas
determines that accuracy has been restored,
and the instrument is once again fit for use.
7.
Do not reset the calibration
gas concentration unless you are using a
calibration gas concentration that differs from
the one that is normally supplied by Biosystems
for use in calibrating the PhD Lite.
Customers are stro ngly urged to use only
Biosystems calibration materials when
calibrating the PhD Lite. Use of non-standard
calibration gas and/or calibrati on ki t components
can lead to dangerously inaccurate readings and
may void the standard Biosystems warranty.
8.
Use of non-standard
calibration gas and/or calibration kit
components when calibrating the PhD Lite can
lead to inaccurate and potentially dangerous
readings and may void the standard
Biosystems warranty.
Biosystems offers calibration kits and longlasting cylinders of test gas specifically
developed for easy PhD Lite calibration.
Customers are strongly urged to use only
Biosystems calibration materials when
calibrating the PhD Lite.
9.
Substitution of components
may impair intrinsic safety.
10.
For safety reasons this
equipment must be operated and serviced by
qualified personnel only. Read and understand
this reference manual before operating or
servicing the PhD Lite.
11.
A rapid up-scale reading
followed by a declining or erratic reading may
indicate a hazardous combustible gas
concentration that exceeds the PhD Lite’s zero
to 100 percent LEL detection range.
12.
Use only the following
chargers to recharge the PhD Lite:
U.S. & Canada: 54-26-0201, 54-26-0202,
54-26-0204
Europe: 54-26-0205, 54-26-0206, 54-26-0207
U.K.: 54-26-0208, 54-26-0209
13.
Use only the following
pumps with the PhD Lite:
54-26-0101 (Standard),
54-26-0102 (Plated)
2.2 On and off sequences
The rectangular push-button on the top of the PhD Lite
case is called the “MODE” button. It is used to turn the
PhD Lite on and off, to initiate the automatic calibration
sequence, and to confirm menu choices.
2.2.1 Start-up sequence
To turn the PhD Lite on:
1. Press the MODE button once and release it.
11
Page 14
Several screens will be shown as the instrument loads
and evaluates information from the “Smart Sensors” that
are currently installed.
biosystems
OTP V.1. 10
biosystems
PhD Lite V.3.18
Loadin g Sensors
O2
Loadin g Sensors
O2 LEL
S/N 1047 0
Black Bo x DL
Date 19 Sept 2003
Time 14:25
The PhD Lite will then automatically go through an
electronic self-test and start-up sequence that will take
approximately thirty seconds.
Performing Self
Test
During the self-test sequence, the display backlight will
momentarily turn on, the visual LED alarm light will
flash, and the audible alarm will sound. The PhD Lite
will also determine which “Smart Sensors” are currently
installed in the instrument, and whether there have been
any changes since the last time the instrument was
used. The instrument will then proceed through the
time/date and battery/temperature screens.
Batter y = 3.9V
Temp = 71F 22C
Note: The temperature shown is actually a reading
taken on the inside of the instrument case in the
area where the sensors are located and may not
correspond with ambient air temperatures. The PhD
Lite microprocessor uses these readings to
compensate for temperature changes in the sensor
compartment.
The instrument will proceed to display the current alarm
settings of the sensors it detects.
CURREN T ALARM
LEVELS
LOW ALAR M LEVEL
O2 = 19.5%
Other alarm level screens may be shown depending
upon the configuration of the instrument.
Note: PhD Lite alarm settings are adjustable by the
user and may be set anywhere within the detection
range of the specific sensor type. Factory default
settin gs may be easi ly restored at any time.
Sensor ranges for individual PhD Lite sensors are
given in Appendix C.
The procedure for changing alarm settings is
discussed in section 3.2.2.
Calibr ation Due
in 30 Days
After the self-test and start-up sequence is completed,
the current gas level screen will be shown. This screen
displays sensors currently installed and the current
readings. When the instrument is operated in “Basic”,
“Basic/Peak” or “Technician” mode, numerical readings
are shown.
O2 LEL CO H2S
20.9 0 0 0
Current gas l ev el scr e en in Basic, Basi c/Peak and
Technician modes with no alarms present.
If the instrument is operated in the “Text Only” mode an
“OK” message will be displayed as long as measured
concentrations are below the alarm set points. If a
reading exceeds an alarm level, the m essage for the
affected sensor channel will change from “OK” to a
numerical reading, the LED alarm light will flash, and the
audible alarm will sound.
O2 LEL CO H2S
OK OK OK OK
Current gas level screen in Text Only mode, no alarms
present.
2.2.1.1 Other start-up screens
Several additional screens may be shown under some
circumstances. Some screens may require the user to
acknowledge the message by pressing the MODE
button.
2.2.1.1.1 Non-standard alarms
"During the start up sequence, if the instrum ent detects
a low oxygen alarm setpoint of less than 18% or an LEL
ceiling alarm setpoint of greater than 30%, the LCD will
display a warning message indicating non-standard
alarms and will then display the non-standard alarm
setting(s) for the specific sensor(s)."
WARNIN G Alarms
Non-Standard
LEL Ceil ing=50%
MODE=Acknowledge
Press the MODE button to acknowledge and use the
non-standard settings.
Note: Factory default settings may be easily
restored at any time. The procedure for restoring
factory default alarm settings is covered in section
3.2.4.1.
12
Page 15
2.2.1.1.2 “Warning Sensor Needs Cal”
The PhD Lite will display the “Warning Sensor Needs
Cal” message for any of the following reasons:
1. The instrument’s sensor configuration has been
modified since the last time the instrum ent was
used.
2. The last calibration was not successfully completed.
3. The current date exceeds the calibration due date
that has been programmed into the instrument.
A sensor that cannot be
calibrated or is found to be out of tolerance should
be replaced immediately. An instrument that fails
calibration may not be used until testing with known
concentration test gas determines that accuracy
has been restored, and the instrument is once again
fit for use.
The “Needs Cal” warning message may be
acknowledged (and silenced) by pressing the MODE
button.
2.2.2 Shut-down sequence
To turn the PhD Lite off, hold the MODE button down
until the “Release Button” message appears. After the
MODE button is released the display will show the
message "Begin SHUTDOWN Please Wait”. The
shutdown sequence is complete when the display
blanks out.
Begin SH UTDOWN
Please Wait
Savin g Instrument
Data
The PhD Lite will then shut itself off.
2.3 Operating modes
The PhD Lite offers a choice of four modes of operation:
"Text Only”, "Basic”, “Basic/Peak” and "Technician”.
Mode selection should be based on how much
information is required, the skill level of the user, and
the nature of the job.
•Displays ‘OK’ for gas-level concentr ations unless an alarm
condition is pr esent.
•Upon alarm c ondition, actual gas-level concentrations will
be displayed.
•2 screens available (toggle by pressing the MODE button):
1. Current gas readings scr een (s ee below).
2. Information screen (see below).
• Gas-level concentrations always displayed.
• Access to calibration funct ions.
• 2 screens available (toggle by pressing the MODE button):
1. Current gas readings screen (s ee below).
2. Information screen (see below).
Text Only M o de:
Basic Mode:
• Gas-level concentrations always displayed.
• Access to calibration functions.
• 3 screens available (scroll by pressing the MODE
button):
1. Current gas readings screen (s ee below).
2. Peak readings screen (s ee below).
3. Information screen (see below).
• Gas-level concentrations always displayed.
• Access to calibration functions.
• 4 screens available (scroll by pressing the MODE
button):
1. Current gas readings screen (s ee below).
2. Peak readings screen (s ee below).
3. STEL/TWA/AVG readings screen (see below).
4. Information screen (see below).
Basic/Peak Mode
Technician Mode:
:
The INFORMATION screen (shown below) can be
accessed from any of the four operating modes by
pressing the MODE button.
01:15 15:08
3.9V 71F
Regardless of mode selection, the PhD Lite remembers
the peak readings of all gases measured, and is
calculating the average readings, Time Weighted
Average (TWA) and Short Term Exposure Limit (STEL)
for any toxic gas sensors installed. Regardless of mode
selection the PhD Lite will go into alarm whenever any
alarm set point is exceeded.
2.3.1 Text Only mode
The simplest mode of operation is the "Text Only" mode.
In Text Only mode during normal operation, the LCD
screen indicates "OK” unless an alarm condition is
present.
MENU
O2 LEL CO H2S
OK OK OK OK
Current gas level screen in Text Only mode, no alarms
present.
During an alarm condition the PhD Lite will no longer
display “OK” for the sensor that is in alarm. Instead, the
display will show the numerical gas reading value for the
sensor channel in flashing reverse text.
O2 LEL CO H2S
19.1 OK OK OK
Current gas level screen in Text Only mode, oxygen alarm
PhD Lite alarms are self-resetting unless the alarm latch
is enabled. With the PhD Lite’s alarm latch enabled, the
audible and visible alarms will continue to sound after
the atmospheric hazard has cleared. To reset the
alarms, simply press the MODE button. If the PhD
Lite’s alarm latch is disabled and the alarm condition is
no longer present, the instrument automatically returns
to normal operation, and the visible and audible alarms
cease without further input from the user.
13
Page 16
For more information on the alarm latch see section
3.2.3.1.
In Text Only mode, the information screen (see above)
is also available to the user. Press the MODE button to
toggle back and forth between the current gas level and
information screens.
Note: It is n o t possib le to in itiate th e a u tomatic
calibration sequence while the PhD Lite is in Text
Only mode. To initiate the “Auto-Calibration”
procedure, the PhD Lite must be in Basic,
Basic/Peak or Technician mode.
2.3.2 Basic mode
Basic mode is designed for users who require numerical
sensor readings at all times. In Basic mode numerical
gas level readings are always provided and it is possible
to initiate the automatic calibration sequence in order to
make fresh air and span calibration adjustments.
Calibration procedures are discussed in detail in
Chapter 4.
O2 LEL CO H2S
20.9 0 0 0
Current gas l ev el scr e en in Basic, Basi c/Peak and
Technician Mode s, no al ar m s pr e sen t .
An alarm condition occurs when one of the sensor
readings exceeds the pre-set alarm level. When an
alarm condition occurs, the display will show the
numerical gas reading value for the sensor channel in
flashing reverse text, the LED alarm lights will flash, and
the audible alarm will sound.
O
LEL CO H
2
2
S
19.1 0 0 0
Current gas l ev el scr e en in Basic, Basi c/Peak and
Technician Modes, alarm condition.
PhD Lite alarms are self-resetting unless the alarm latch
is enabled. When the PhD Lite’s alarm latch is enabled,
the audible and visible alarm will continue to sound after
the atmospheric hazard has cleared. To reset the
alarms after the atmospheric hazard has cleared, simply
press the MODE button. If the PhD Lite’s alarm latch is
disabled and the alarm condition is no longer present,
the instrument automatically returns to normal
operation, and the visible and audible alarms cease
without further input from the user.
For more information on the alarm latch see section
3.2.3.1.
In Basic mode, the information screen is also available
to the user. While in Basic mode the MODE button may
be used to toggle back and forth between the current
gas level and information screens.
2.3.3 Basic/Peak mode
The Basic/Peak mode of operation provides users with
numerical sensor readings and access to the peak
readings screen. In Basic/Peak mode it is possible
initiate the automatic calibration sequence in order to
make fresh air and span calibration adjustments.
Calibration procedures are discussed in detail in
Chapter 4.
2.3.3.1 Peak readings
Peak readings for the accumulations of combustible
gases and vapors, and for the accumulation of toxic
gases represent the highest values registered by the
instrument during the period of operation. Peak
readings are updated at a rate of once per second.
Peak readings for oxygen include both the highest and
lowest values registered by the instrument during the
period of operation.
PEAK LOW HIGH
O2 20.7 20.9
Peak oxygen readings screen in Basic/Peak and
Technician mode s.
PEAK LEL CO H2S
0 0 0
Peak LEL and toxic readings screen in Basic/Peak and
Technician mode s.
While in Basic/Peak mode the MODE button may be
used to scroll through the current gas level, peak
readings, and information screens.
2.3.3.2 To reset peak readings
Peak readings may be reset during any period of
operation. To reset the peak readings:
1. Press the MODE button until the peak oxygen
reading screen appears.
PEAK LOW HIGH
O
20.7
2
20.9
2. Hold the right navigation arrow down for
approximately one second or until the following
screen appears:
Reset Pe aks?
YES NO
3. With YES highlighted, press the MODE button to
reset the peak settings
Note: If peak readings are reset, the PhD Lite’s data
recorder automatically records that the peak
readings have been reset and the word “PEAK” in
the peak readings screen will be highlighted for the
remainder of the current operating session.
PEAK LOW HIGH
O
20.7
2.3.4 Technician mode
2
Technician mode provides access to all advanced
functions and displays of the PhD Lite.
20.9
14
Page 17
While in Technician mode the MODE button may be
used to scroll through the current gas level, peak
readings, STEL, TWA, AVG and information screens.
2.3.4.1 STEL
The STEL (Short Term Exposure Limit) for a particular
toxic gas is the maximum average concentration to
which an unprotected worker may be exposed in any 15
minute interval during the day. The STEL value
displayed by the PhD Lite is the average concentration
for the most recently completed 15 minutes of operation.
Note: For the first 15 minutes after the PhD Lite is
initially turned on the STEL reading is a projected
value. The PhD Lite will begin projecting a STEL
value after the first 30 seconds of operation. For the
first 30 seconds the STEL screen will show an "X"
where the reading should be.
The STEL reading is continuously updated. Audible and
visible alarms will be activated immediately any time the
most recent 15-minute average exceeds the STEL
alarm set point.
STEL CO H2S
0 0
STEL screen in Technician Mode:
Appendix A discusses Permissible Exposure Limit
alarm calculations in greater detail.
2.3.4.2 TWA readings
Time Weighted Average or TWA values are calculated
by taking the sum of exposure to a particular toxic gas in
the current operating session in terms of parts-permillion-hours and dividing by an eight-hour period.
TWA CO H2S
0 0
TWA screen in Technician Mode:
Note: It is not possible to calculate a toxic gas TWA
reading until the PhD Lite has been operating for 15
mi nutes. For the first 15 m inutes after start-up, t he
TWA screen will show an “X” in place of the
calculation. After 15 minutes, the TWA calculation
will be shown.
Appendix A discusses Permissible Exposure Limit
alarm calculations in greater detail.
2.3.4.3 Average readings
The average readings displayed by the PhD Lite are the
simple arithmetic averages registered by the instrument
during the current operating session.
AVERAGE O2 LEL
20.9 0
Average readings screen for O2 and LEL in Technician
Mode.
2.3.5 Changing operating modes
To change operating mode:
1. Press the MODE button until the information
screen is displayed.
01:15 15:08
3.9V 71FMENU
2. Hold down the left arrow key until EXIT appears and
is highlighted (approximately three seconds).
01:15 15:08
3.9V 71F
3. Press the down navigation arrow once to highlight
MENU and press the MODE button to enter the
Main Menu.
EXIT
MENU
01:15 15:08
3.9V 71F
ALARMS CALIBRATION
OPTIONS SCREEN
VIEW TIME
EXIT
4. Use the navigation arrows to highlight OPTIONS
and press the MODE button to enter the Options
Menu.
ALARMS CALIBRATION
OPTIONS
VIEW TIME
DISPLAY EXIT
USER_MODE SEC_BEEP
PASSCODE LANGUAGE
DECIMAL T_UNITS
EXIT
5. Once in the Options Menu, use the navigation
arrows to highlight USER MODE and press the
MODE button. The up and down navigation arrows
may then be used to toggle between operating
modes.
USER_MODE
PASSCODE LANGUAGE
DECIMAL T_UNITS
SCREEN
EXIT
MENU
SEC_BEEP
EXIT
Operat ing Mode
Text Only
↑↓
EXIT
Operat ing Mode
Basic
↑↓
EXIT
Operat ing Mode
Basic/Peak
↑↓
EXIT
Operat ing Mode
Technician
6. Once the preferred operating mode is shown,
confirm the selection by pressing the MODE button.
EXIT
15
Page 18
Save Ch anges?
YES NO CANCEL
7. Press the MODE button with YES highlighted to
save the new operating mode.
Note: Changing the operating mode or otherwise
reprog ramming the PhD Li te is reserved for
authorized employees.
2.4 Batteries
The PhD Lite can be equipped with either Alkaline of
NiMH rechargeable batteries
a non-hazardous location whenever the alkaline
batteries are removed from the instrument.
Removing the alkaline batteries from the instrument
in a hazardous location may impair intrinsic safety.
2.4.1 Alkaline batteries
The PhD Lite with alkaline batteries is designed to
provide up to 17 hours of continuous use with each set
of 3 fresh AA-cell disposable alkaline batteries.
Operation times may be reduced if the PhD Lite is used
with the motorized pump or in low light conditions where
the instrument backlight is on.
2.4.1.1 Replacing alkaline batteries
1. Check that the PhD Lite is not located in a
hazardous (potentially combustible) area.
2. Make sure that the PhD Lite is turned off.
3. Loosen the battery retention screw located at the
top of the instrument, remove the battery cap and
then remove the alkaline batteries.
4. Install new batteries in accordance with the diagram
underneath the battery cap and reinstall the battery
cap.
Not e: Always di spos e of alkaline batter ies in
accordance with local ordinances.
Note: The PhD Lite is designed to turn itself on
whenever the batteries are removed and then
replaced. This ensures that in the event of a power
interruption, the instrument automatically turns
it self back o n. Any time the batt eries are
momentarily removed or repl ace d it will be
necessary to manually turn the PhD Lite off if the
instrument is not going to be put into immediate
use.
Ultra MX1500, Eveready Energizer E91-LR6,
Eveready EN91, Radio Shack 23-874 (not ATEX
approved)size AA, 1.5V Alkaline batteries, Eveready
CH15 (not ATEX approved) or Radio Shack 23-149
(not ATEX approved) size AA 1.2V NiCad batteries,
or E veready L9 1 size AA 1.5V Lith ium ba t teries (not
CSA or ATEX approved). Substitution of batteries
may impair intrinsic safety.
The PhD Lite must be located in
Use only Duracell MN1500 or
2.4.2 NiMH rechargeable battery
The PhD Lite may be equipped with a rechargeable
NiMH (nickel metal hydride) battery that is designed to
provide up to 17 hours of continuous use between
charging cycles. The NiMH battery pack is contained
inside of the PhD Lite and should only be replaced by
appropriately trained personnel.
Operation tim es may be reduced if either alkaline or
NiMH PhD Lites are used with the motorized pump or in
low light conditions where the instrument backlight is on.
2.4.2.1 Storage guidelines for the NiMH battery.
Never store NiMH-version PhD Lite instruments at
temperatures above 30 degrees Celsius (86 degrees
Fahrenheit). Nickel Metal Hydride batteries may suffer
deterioration resulting in damage to the internal
components when stored at high temperatures. The
battery may be irretrievably damaged resulting in
reduced battery capacity and voltage.
Biosystems recommends leaving PhD Lite instruments
with NiMH rechargeable batteries on the charger when
not in use.
2.4.2.2 Charging guidelines for NiMH battery
The NiMH battery in the PhD Lite should never be
charged at temperatures lower than 5 degrees Celsius
(40 degrees Fahrenheit) or higher than 30 degrees
Celsius (86 degrees Fahrenheit. Charging at
temperature extremes can cause permanent damage
the PhD Lite NiMH battery.
The PhD Lite must be located in
a non-hazardous location during the charging cycle.
Charging the PhD Lite in a hazardous location may
impair intrin sic safety.
Use only the following
Biosy s tems charg ers to ch a rge the P h D L ite:
U.S. & Canada: 54-26-0201, 54-26-0202, 54-26-0204.
Europe: 54-26-0205, 54-26-0206, 54-26-0207.
U.K.: 54-26-0208, 54-26-0209
2.4.2.3 Charging procedure for NiMH battery
1. Verify that the instrument is turned off. (If it is not,
press the MODE button for three seconds until the
message "Release Button" appears.)
2. Plug the power supply in and check to see that the
“POWER” indicator LED on the charging cradle is
lit.
3. Insert the PhD Lite into the charging cradle. While
the battery is charging the red ‘FAST’ indicator LED
will be lit on the charger.
Note: The “Char g e” L ED indicator will initially light
up and remain lit for the first 15 minutes regardless
of battery pack voltage.
4. Charging will be completed in 2 hours or less, and
will be indicated by the green ‘’TRICKLE’ indicator
16
Page 19
LED. Charging is complete any time after the
“TRICKLE” indicator is lit.
CAUTION To achieve optimal charge and ensure
long battery life of the NIMH battery, make sure that
charging takes place in an area where the ambient
air temperature is between 40 and 86 degrees
Fahrenheit (5 and 30 degrees Celsius). Charging
the battery in temperatures above or below this
range can damage the battery and will drastically
affect battery life.
2.4.3 Low battery alarms
Note: The voltage figures given below are for PhD
Lite instruments with firmware version 3.18 and may
be slig htly diff eren t for units wit h othe r vers ions of
instrument firmware.
The PhD Lite includes low battery alarms that are
activated whenever battery voltage approaches a level
that will soon lead to instrument shut down. When the
battery voltage is reduced to approximately 3.23 volts,
an audible alarm will sound and the display will indicate
that a low battery condition exists. At this stage, the low
battery alarms may be silenced for a fifteen-minute
period by pressing the MODE button. After the first low
battery alarm, the alarm will sound again every fifteen
minutes until the voltage drops to the “Very Low Battery”
level.
The “Very Low Battery” level occurs when the battery
voltage drops to 3.13 volts. Due to the risk of imminent
shut down, when the battery voltage reaches the “Very
Low Battery” level it is no longer possible to silence the
low battery alarms. At this point, it is necessary to
immediately leave the hazardous area in which the
instrum ent is being used.
When the voltage drops to 3.10 volts, the PhD Lite will
display a "Dead Battery" message to warn the user of
imminent shut down. The instrument will then
automatically turn itself off.
After any low battery alarm the batteries should be
replaced if the PhD Lite is equipped with alkaline
batteries or the battery should be recharged if the PhD
Lite is equipped with a NiMH rechargeable battery.
Ultra MX1500, Eveready Energizer E91-LR6,
Eveready EN91, Radio Shack 23-874 (not ATEX
approved)size AA, 1.5V Alkaline batteries, Eveready
CH15 (not ATEX approved) or Radio Shack 23-149
(not ATEX approved) size AA 1.2V NiCad batteries,
or E veready L9 1 size AA 1.5V Lith ium ba t teries (not
CSA or ATEX approved). Substitution of batteries
may impair intrinsic safety.
2.5 Methods of sampling
The PhD Lite may be used to sample the immediate
vicinity in diffusion mode, or it may be used to sample
remote locations. Sampling remote locations requires
the use of either the manual sample draw kit that is
included with the PhD Lite at purchase, or the PhD Lite
motorized sample pump that is available separately.
Use only Duracell MN1500 or
In normal operation, the PhD Lite functions as a
diffusion-style gas detector. The atmosphere reaches
the sensors by diffusing through the vents in the sensor
compartment cover. Normal air movements are enough
to carry the sample to the sensors. Once turned on, the
PhD Lite monitors continuously. The sensors react
quickly to changes in the concentrations of the gases
being measured. This type of "diffusion" operation
monitors only the atmosphere immediately surrounding
the detector.
To sample more remote locations. It is necessary to use
one of the two sample-draw kits that are available for
the PhD Lite. A manual sample draw kit comes as a
standard accessory with every PhD Lite and includes a
sample draw probe, squeeze bulb and ten feet of tubing.
A motorized sample draw pump is also available. In
either case the sample is drawn in through a sample
draw probe, and then proceeds through a length of hose
back to the sensor compartment.
2.5.1 Using the hand-aspirated sample draw kit
1. Connect the end of the hose that is closer to the
squeeze bulb to the sample draw adapter. Then
connect the other end of the hose to the sample
probe as shown in Figure 2.5.1.1.
2. Attach the sample draw adapter to the PhD Lite as
shown in Fi gure 2.5.1.2.
3. Cover the end of the sample draw probe assembly
with a finger, and squeeze the aspirator bulb. If
there are no leaks in the sample draw kit
components, the bulb should stay deflated for a few
seconds.
4. Insert the end of the sample probe into the location
to be sampled.
5. Squeeze the aspirator bulb several times to draw
the sample from the remote location to the sensor
compartment. Allow one squeeze of the bulb for
every one foot of sampling hose for the sample to
reach the sensors. Continue to squeeze the bulb
for an additional 45 seconds or until readings
stabilize.
6. Note the gas measurement readings.
Figure 2.5.1.1. Hand-aspirated sample draw kit
17
Page 20
Figure 2.5.1.2. PhD Lite with hand-aspirated
sample draw kit attached
CAUTION: Hand aspirated remote sampling only
provides continuous gas readings for the area in
which the probe is located when the bulb is being
continuously squeezed.
Note: Each time a reading is desired, it is necessary
to squeeze the bulb a sufficient number of times to
bring a fresh sample to the sensor compartment.
2.5.2. Motorized sample draw pump
Use of the motorized sample draw pump allows the PhD
Lite to continuously monitor remote locations. The
pump is powered by the PhD Lite battery. When the
pump is attached and functioning normally, an animated
” icon will appear in the upper left hand corner of the
“
LCD display.
O2 LEL CO H2S
20.9 0 0 0
CAUTION: Never perform remote sampling with
the PhD Lite without the sample probe assembly.
The sample probe handle contains replaceable
filters designed to block moisture and remove
particulate contaminants. If the pump is operated
without the probe assembly in place, contaminants
may cause damage to the pump, sensors and
internal components of the PhD Lite.
The sample draw pump includes a pressure sensor
designed to protect the PhD Lite from exposure to water
or other liquids. If there is a change in pressure in the
sample draw assembly due to fluid intake or other
blockage, the pump immediately shuts down. After a
few seconds audible and visible alarms indicating a low
flow condition will also be activated.
CAUTION: Insertion of the sample draw tube
into a fluid horizontally or at a low angle may lead to
water ingress and may cause damage to the PhD
Lite.
The pressure sensor in the sample draw pump is
designed to detect pressure changes while the sampledraw probe is being held in a vertical position. If the
probe is held horizontally or at a low angle while
inserted into a fluid, a pressure drop sufficient to cause
the pump to shut down may not be generated, and
water could be drawn into the pump assembly causing
damage to the pump, sensors and internal components
of the PhD Lite.
To avoid potential damage, care must be taken to keep
the probe vertical whenever fluids may be present.
with the PhD Lite:
54-26-0101 (Standard), 54-26-0102 (Plated)
2.5.2.1 Using the continuous sample draw pump
1. Connect the pump to the hose and probe assembly
as shown below in Figure 2.5.2.1.
2. Slide the pump onto the PhD Lite as shown in
Figure 2.5.2.2.
3. Make sure the pump is securely attached. Tighten
the captive screw.
4. Turn on the PhD Lite. The pump will be activated
automatically when properly attached to the PhD
Lite. The display will indicate that the pump has
been detected.
5. When prompted cover the end of the sample draw
probe assembly with a finger. If there are no leaks
in the sample draw components, a message will be
displayed indicating that the pump test passed.
6. Insert the end of the sample probe into the location
to be sampled.
7. Wait long enough for the pump to have drawn the
sample through the entire length of hose, and for
the sensors to have stabilized. (Allow one second
for each foot or three seconds for each meter of
sample hose for the sample to reach the sensors)
8. Note the gas measurement readings
Use only the following pumps
Pump De tected
Pump Testing
Block Sa mple
Pump Test Passed
Press Mo de
18
Page 21
Figure 2.5.2.1. Motorized sample draw pump and
probe assembly
2.5.2.2 Protective “low flow” shut-downs
If a protective pump shut-down occurs, the following
steps should be taken before the instrument is put back
into use:
1. Turn off the PhD Lite detector and disconnect the
sample draw pump.
2. Rem ove the sample draw assembly from the area
being monitored. Be careful to keep the sample
draw probe in a vertical position.
3. Examine the sample draw probe and hose to make
sure no fluids remain trapped.
4. Allow any trapped fluids to completely drain. (It
may be necessary to disconnect the hose or
sample draw probe before drainage can occur.)
5. Replace the filters in the sample draw probe as
necessary.
6. Re-attach the pump in fresh air and wait for
readings to stabilize.
7. Resume sampling.
Figure 2.5.2.2. PhD Lite with motorized sample draw
pump attached
2.5.2.3 Resuming diffusion monitoring
To resume diffusion monitoring, simply disconnect the
pump assembly from the PhD Lite. The audible and
visual alarms will be activated and the LCD will display
the message “Pump Disconnect”. Press the MODE
button to acknowledge the disconnection and resume
normal diffusion operation.
Pump Di sconnected
Press Mo de
2.5.3 Sample probe assembly
The sample probe handle contains moisture barrier and
particulate filters designed to remove contaminants that
might otherwise harm the instrum ent.
CAUTION: Never perform remote sampling
without the sample probe and hose assembly. The
sample probe handle contains replaceable filters
designed to block moisture and remove particulate
contaminants. If the pump is operated without the
probe assembly in place, contaminants may cause
damage to the pump, sensors and internal
components of the PhD Lite.
Particulate contaminants are removed by means of a
cellulose filter. The hydrophobic filter includes a 0.1 µm
Teflon™ barrier which blocks the flow of moisture as well
as any remaining particulate contaminants.
Sample probe filters should be replaced whenever
visibly discolored due to contamination. A spare filter
replacement kit (Biosystems part number 54-05-K0401)
is included with every PhD Lite.
19
Page 22
Figure 2.5.3. PhD Lite sample probe assembly
2.5.3.1 Changing sample probe filters
The threaded sample probe handle is unscrewed (as
shown in Figure 2.5.3.) to provide access to the filters.
The particulate filter is held in place by means of a clear
filter cup. To replace the particulate filter, remove the
old filter and cup, insert a new filter into the cup, and
slide the cup back into place in the probe handle. The
hydrophobic barrier filter fits into a socket in the rear
section of the probe handle. (The narrow end of the
hydrophobic barrier filter is inserted towards the rear of
the handle.)
2.5.3.2 Changing sample probe tubes
The standard 11.5” long butyrate probe tube is held in
place by means of a hex-nut compression fitting and
compression sleeve. The standard probe tube is
designed to be easily interchangeable with other custom
length sections of 1/4” OD tubing, or probe tubes made
of other materials (such as stainless steel).
Probe tubes are exchanged by loosening the hex-nut
compression fitting, removing the old tube, sliding the
compression sleeve into place around the new tube,
inserting the new tube into the probe handle, then
replacing and tightening the hex-nut.
Note: The sample probe must be checked for
leakage (as discussed in Section 2.5.1.) whenever
filters or probe tubes are exchanged or replaced
before being put back into service.
2.6 Biosystems EEPROM equipped “Smart
Sensors”
Each sensor installed in a PhD Lite detector is equipped
with its own non-volatile memory storage device or
“EEPROM”. The contents of the sensor’s memory
device are designed to be read and updated directly by
the PhD Lite. The fact that each sensor is capable of
remembering and communicating important information
about itself to the instrument allows for a number of
important PhD Lite operating benefits.
2.6.1 Identification of type of sensor by
instrument
Sensors installed in the PhD Lite automatically identify
themselves to the instrument microprocessor. The PhD
Lite automatically displays the recognized sensors on
the liquid crystal display (LCD) and assigns the correct
alarm settings.
2.6.2 Ot her inform ation stored wi th the sen sor
EEPROM
The PhD Lite automatically uploads the sensor serial
number, the most recent calibration settings,
temperature compensation curves, and current alarm
settings whenever the instrument is turned on,
whenever a change is made during operation, and
whenev er the instrument is turned off.
If a sensor is changed or replaced the PhD Lite
recognizes that a change has occurred, displays a
“Needs Cal” message the next time the instrument is
turned back on, and identifies the affected sensors.
Even if the change is only to replace one sensor with
another of the same kind, the PhD Lite will still note the
change in serial numbers of the sensors installed, and
display the “Needs Cal” message.
The accuracy of the PhD Lite
should be checked periodically with known
concentration calibration gas. Failure to check
accuracy can lead to inaccurate and potentially
dangerous readings. (The Canadian Standards
Association (CSA) requires an accuracy check
using known concentration calibration gas prior to
each day’s use.)
The accuracy of sensors
identified as “Needing Calibration” must be verified
by exposure to known concentration calibration gas
before the PhD Lite is put back into service. Failure
to do so may result in inaccurate and potentially
dangerous readings.
2.6.3 Sensor removal and replacement
Note: The PhD Lite must be turned off to replace
sensors.
1. Remove the three philips screws and take off the
sensor cover from the front of the instrument.
20
Page 23
For replacement of an existing sensor perform
steps 2a and 3a then proceed to step 4a or 4b.
2a. From the outer surface of the sensor cover, use a
flat blade screwdriver to gently push out the metal
screen (if present), gasket & filter ring assembly.
Discard old gasket, filter ring assembly & metal
screen. The metal screen is not to be reused and
its absence will not affect sensor performance.
3a. Remove any remaining traces of adhesive from the
recessed hole in the sensor cover.
For new sensor installation perform steps 2b and 3b
then proceed to step 4a or 4b.
2b. From the outer surface of the sensor cover, push
out yellow dust cap with a blunt tool.
3b. Remove sensor blank from the sensor
compartm ent.
For Sensors O
, LEL, CO, CO Plus, CO-H, H2S, NO,
2
DUO-TOX perform step 4a then proceed to step 5.
4a. Firmly press the new filter ring assembly into the
recessed hole in the sensor cover with ridge side
up.
For Reactive Gas Sensors: SO
perform step 4b then proceed to step 5.
ClO
2
, NO2, PH3, HCN, Cl2,
2
4b. Firmly press the new teflon spacer into the
recessed hole in the sensor cover. For optimal
sensor response, no external filter element is used
with these sensors.
CAUTION: Since the reactive gas sensor face
is openly exposed to the ambient environment,
extra care should be taken to prevent water or any
other liquids from coming into contact and
remaining on the sensor face.
5. Peel the backing paper from the new rubber gasket
and center it over the newly mounted filter ring
assembly with the adhesive side down.
6. Remount sensor cover and secure with three
philips screws. DO NOT overtighten.
7. Install functional battery pack or batteries in the
detector.
8. New sensors must be allowed a stabilizing time –
with detector powered off and with functional
batteries or battery pack installed, according to the
following schedule.
Sensor Stabilization Period
Oxygen (54-25-90) 1 hour
LEL (54-25-80A) 5 minutes
All Toxic sensors except
those shown below
54-25-04 NH3 Sensor
54-25-06 NO Sensor
9. The PhD Lite will automatically recognize the
changes that have been made upon turn on and
display the “Warning Needs Cal” message.
10. Calibrate the PhD Lite with calibration gas
appropriate for the new sensor before the
instrument is put back into service.
11. If newly installed sensor(s) do not appear during
power-up, or if the message "not supported"
appears, contact Biosystems Instrument Service
Dept. for further guidance.
15 minutes
24 hours
PhD Lite programming includes safeguards to recognize
maladjusted sensors. If the settings on the new sensor
are significantly different from those of the sensor that is
being replaced, a message indicating that the sensor is
reading “Too Low” or “Too High” for Auto-Calibration
may follow the first attempt at one-button calibration.
Note: The first fresh air calibration adjustment after
installation of a new sensor should be done using
the “manual” calibration procedure as discussed in
section 4.5 of this manual
Once the new sensor has been fresh-air calibrated
using the “manual” calibration procedure, subsequent
fresh air and span calibrations can be made by using
the MODE button and One-Button Auto-Calibration
procedures.
2.6.4 Missing sensor
The PhD Lite is able to recognize when a sensor is
removed or becomes disconnected while the instrument
is in normal operation. If a sensor becomes
disconnected while the instrument is turned on, the PhD
Lite will trigger a missing sensor alarm, the
corresponding channel of the display will show an “X” in
place of the normal sensor readings and the audible and
visible alarms will be activated.
2.6.5 “Sensor not found”
If the PhD Lite is suddenly unable to read the EEPROM
of a smart sensor currently installed the corresponding
sensor channel will show an “X” and the audible and
visible alarms for the affected sensor channel will be
activated.
If a smart sensor is removed while the instrument is
turned off without being replaced with another sensor, a
message will be displayed during the start-up sequence
indicating that the sensor is missing. Pressing the
MODE button acknowledges the condition, and allows
the use of the instrument for those sensors that have
been successfully detected.
NO TOX1 SENSOR
MODE=Acknowledge
21
Page 24
Chapter 3. Advanced Functions
The microprocessor circuitry in the PhD Lite makes
modifications to the operational set-up parameters easy.
By using the MODE button and four navigation arrows
located on the instrument keypad, the user can access
the Main Menu, which provides access to the
operational set-up choices for the instrument.
Note: This chapter covers advanced functions in
units without the datalogger upgrade. Some menu
items may be slightly different for instruments with
the datalogger upgrade. Chapter 5 covers options
unique to instruments with the datalogger upgrade.
Reprogramming the PhD Lite is reserved for
authorized personnel.
3.1 The Main Menu
The Main Menu provides access to all user-configurable
options of the PhD Lite.
3.1.1 Entering the Main Me n u
1. Turn the instrument on and wait until the gas
readings screen appears.
2. Press the MODE button until the information screen
is displayed.
01:15 15:08
3.9V 71F
3. Hold down the left navigation arrow for 3 seconds
or until EXIT appears and is highlighted.
MENU
01:15 15:08
3.9V 71F
4. Press the down navigation arrow once to highlight
MENU and press the MODE button. The Main
Menu will then be displayed.
3.2 The Alarm Menu
PhD Lite gas level alarms are user-adjustable and may
be set anywhere within the range of the sensor channel.
When the alarm set point for a specific sensor channel
is exceeded, the audible alarm sounds, the bright red
LED alarm light blinks, and the alarm channel is
displayed in flashing reverse text.
PhD Lite alarms are self-resetting unless the alarm latch
is enabled. With the alarm latch disabled, visible and
audible alarms cease when gas readings drop back
below the pre-set alarm levels. With the alarm latch
enabled, visible and audible alarms continue to sound
after the atmospheric hazard has cleared. The user
must then manually reset the alarms by pressing the
MODE button.
Alarm latch settings are discussed in section 3.2.3.
ALARMS CALIBRATION
OPTIONS SCREEN
VIEW TIME
EXIT
EXIT
MENU
Factory default settings can be restored at any time
during normal operation by using the procedures
discussed in section 3.2.4.
3.2.1 Entering the Alarm Menu
1. Enter the Main Menu as described above in section
3.1.1. Use the navigation arrows to highlight
ALARMS.
ALARMS
OPTIONS SCREEN
VIEW TIME
CALIBRATION
EXIT
2. Press the MODE button to enter the Alarms Menu.
The Alarms Menu will then be displayed.
3.2.2 Custom alarm settings
CUSTOM LATCH
DEFAULT TEMPERATURE
WARNINGS
EXIT
PhD Lite gas alarms are user-adjustable and may set
anywhere within the range of the sensor channel.
Sensor ranges for individual PhD Lite sensors are
given in Appendix C.
To enter custom alarm settings:
1. Enter the Alarm Menu as described above in
section 3.2.1. Use the navigation arrows to
highlight CUSTOM.
CUSTOM
DEFAULT TEMPERATURE
WARNINGS EXIT
LATCH
2. Press the MODE button. “ALARM ADJUSTMENT”
will be shown followed by the current alarm
settings.
ALARM AD JUSTMENT
LOW ALAR M ADJUST
O2 = 19.5 %
3. Use the right and left navigation arrows to scroll
through the sensors and their alarm settings until
the alarm that requires adjustment is displayed
EXIT
LOW ALAR M ADJUST
O2 = 19.5 %
→
EXIT
HIGH ALARM ADJUST
O2 = 23.5 %
→
EXIT
HIGH ALARM ADJUST
LEL = 10 %
4. Once the alarm that requires adjustment is
reached, use the up and down navigation arrows to
adjust the alarm value.
EXIT
22
Page 25
HIGH ALARM ADJUST
LEL = 10 %
↓
EXIT
HIGH ALARM ADJUST
LEL = 9 %
5. Once the desired alarm setting is reached, press
the MODE button to confirm the selection or move
to another sensor by following steps 3 and 4 above.
EXIT
Save Ch anges?
YES NO CANCEL
6. Press the MODE button with YES highlighted to
save the new alarm settings. The instrument will
then return to the Main Menu.
Alarm Le vels
Saved
3.2.3 Alarm and OK latches
The PhD Lite includes alarm latch and OK latch
functions that can be enabled or disabled according to
the user’s requirements.
3.2.3.1 Alarm latch settings
When the PhD Lite’s alarm latch is enabled, the audible
and visible alarms will continue to sound after the
atmospheric hazard has cleared. To reset the alarm
after the hazard has cleared, simply press the MODE
button.
When the PhD Lite’s alarm latch is disabled and the
alarm condition is no longer present, the instrument
automatically returns to normal operation, and the
visible and audible alarms cease without further input
from the user.
To adjust the alarm latch settings:
1. Enter the Alarm Menu as described above in
section 3.2.1 and use the navigation arrows to
highlight LATCH.
CUSTOM
DEFAULT TEMPERATURE
WARNINGS EXIT
LATCH
2. Press the MODE button. The current latch settings
will then be shown.
ALARM LATC H
SETTINGS
ALARM LATC H
DISABLED
3. Once the alarm latch setting is shown, the up and
down navigation arrows may be used to toggle
between ENABLED and DISABLED.
ALARM LA TCH
DISABLED
↓↑
ALARM LA TCH
ENABLED
4. Once the desired setting is reached, press the
MODE button to confirm the setting.
Save Ch anges?
YES NO CANCEL
5. Press the MODE button with YES highlighted to
save the new latch settings. The instrument will
then return to the Main Menu.
Alarm La tch
Saved
3.2.3.2 OK latch settings
If an alarm condition occurs and clears while the PhD
Lite is operated in text-only mode with the OK latch
enabled, the instrument will continue to display numeric
readings for the sensor that was in alarm. This allows
the user to know that an alarm condition was present
during the current operating session.
If an alarm condition occurs and clears while the PhD
Lite is operated in text-only mode with the OK latch
disabled, the PhD Lite will again display OK for the
sensor that was in alarm.
To adjust the OK latch settings:
1. Enter the Alarm Menu as described above in
section 3.2.1 and use the navigation arrows to
highlight LATCH.
CUSTOM
DEFAULT TEMPERATURE
WARNINGS EXIT
LATCH
2. Press the MODE button. The current alarm latch
setting will then be shown.
ALARM LATC H
SETTINGS
ALARM LATC H
DISABLED
3. Press the right navigation arrow once to move from
the alarm latch setting to the OK latch setting
ALARM LATC H
DISABLED
→
OK LATCH
DISABLED
23
Page 26
4. Once the latch setting is shown, the up and down
navigation arrows may be used to toggle between
ENABLED and DISABLED.
OK LATCH
DISABLED
↓↑
OK LATCH
ENABLED
5. Once the desired setting is reached, press the
MODE button to confirm the selection.
Save Ch anges?
YES NO CANCEL
6. Press the MODE button with YES highlighted to
save the new latch settings. The instrument will
then return to the Main Menu screen.
Alarm La tch
Saved
3.2.4 Default alarm settings
PhD Lite alarm settings are set at the factory and may
be restored at any time by performing the following
procedure.
3.2.4.1 Restore factory default alarm settings
1. Enter the Alarm Menu as described above in
section 3.2.1 and use the navigation arrows to
highlight DEFAULT.
CUSTOM LATCH
DEFAULT
WARNINGS EXIT
TEMPERATURE
2. Press the MODE button. The default low alarm
setting for oxygen will then be displayed. If the
instrument does not have an oxygen sensor, then
the LEL or toxic alarm default setting will be shown.
DEFAUL T ALARM
LEVELS
DEF LOW AL ARM
O2 = 19.5 %
3. The right and left navigation arrows are used to
scroll through the various default alarm settings.
EXIT
DEF LOW AL ARM
O2 = 19.5 %
→
EXIT
DEF HIGH ALARM
O2 = 23.5 %
→
EXIT
DEF HIGH ALARM
LEL = 10 %
4. Press the MODE button at any time to exit.
EXIT
Set Def Al arms?
YES NO CANCEL
5. Press the MODE button with YES highlighted to
restore default alarms. The instrument will then
return to the Main Menu screen.
Alarm Le vels
Saved
3.2.5 Temperature alarm
The PhD Lite includes both high and low temperature
alarms for all sensors recognized by the instrument.
The alarm setpoints are pre-programm ed into the
individual sensor EE-proms and may not be modified in
any way, but the high or low temperature alarms for all
recognized sensors can be enabled or disabled
depending on the needs of the user.
3.2.5.1 Enable/disable temperature alarms
1. Enter the Alarm Menu as described above in
section 3.2.1 and use the navigation arrows to
highlight TEMPERATURE.
CUSTOM LATCH
DEFAULT
WARNINGS EXIT
TEMPERATURE
2. Press the MODE button. The enabled/disabled
setting for the low temperature alarm will be
displayed. The right and left navigation arrows are
used to toggle between the high and low
temperature alarm settings.
LOW TEMP ALARM
DISABLED
← →
EXIT
HIGH TEMP ALARM
DISABLED
3. Once the alarm that required adjustment is shown,
the up and down navigation arrows are used to
toggle between ENABLED and DISABLED.
EXIT
HIGH TEMP ALARM
DISABLED
↑↓
EXIT
HIGH TEMP ALARM
ENABLED
4. Once appropriate setting is shown, press the MODE
button to enter the new setting.
EXIT
24
Page 27
Save Ch anges?
YES NO CANCEL
5. Press the MODE button with YES highlighted to
save the new temperature alarm settings.
TemperatureAlarm
Saved
3.2.6 Warning alarms.
The PhD Lite includes adjustable warning alarms for the
toxic and combustible gas sensor channels that can be
enabled or disabled according to the needs of the user.
Warning alarms are normally set slightly lower than the
actual alarm to warn the user of potential danger.
3.2.6.1 Access warning alarm settings
1. Enter the Alarm Menu as described above in
section 3.2.1 and use the navigation arrows to
highlight WARNINGS.
CUSTOM LATCH
DEFAULT TEMPERATURE
WARNINGS
EXIT
2. Press the MODE button. The following screen will
then be displayed:
3.2.6.2 Enable/disable warning alarms
ENABLE
ADJUST
TIMEOUT
EXIT
1. Enter the warning alarm subdirectory of the Alarm
Menu as described above in section 3.2.6.1. Use
the navigation arrows to highlight ENABLE.
ENABLE
ADJUST
TIMEOUT
EXIT
2 Press the MODE button. The enabled or disabled
setting will then be shown. The up and down
navigation arrows may then be used to toggle
between enabled and disabled.
WARNINGS
ENABLED
↑↓
EXIT
WARNINGS
DISABLED
3. Once the appropriate setting is shown, press the
MODE button to enter the new setting.
EXIT
Save Ch anges?
YES NO CANCEL
4. Press the MODE button with YES highlighted to
save the new warning alarm setting.
3.2.6.3 Adjust warning alarm levels
1. Enter the warning alarm subdirectory of the Alarm
Menu as described above in section 3.2.6.1. Use
the navigation arrows to highlight ADJUST.
ENABLE
ADJUST
TIMEOUT
EXIT
2. Press the MODE button. The warning alarm setting
for either the LEL or toxic sensor will then be
shown.
Note: The warning alarm must be enabled for the
warning alarm level to be shown. Section 3.2.6.2
above covers enabling the warning alarms.
HIGH WA RNING ADJ.
LEL = 5 %
3. The right and left navigation arrows to scroll
through the various default alarm levels for the LEL
and toxic sensors.
EXIT
HIGH WA RNING ADJ.
LEL = 5 %
→
EXIT
HIGH WA RNING ADJ.
CO = 17ppm
4. To adjust a specific alarm level, use the up and
down navigation arrows to change the setting.
EXIT
HIGH WA RNING ADJ.
CO = 17ppm
↑
EXIT
HIGH WA RNING ADJ.
CO = 18ppm
5. Use the right and left arrow keys to reach the next
alarm for adjustment and repeat step 4 or press the
MODE button to confirm the new warning alarm
levels.
EXIT
Save Ch anges?
YES NO CANCEL
6. Press the MODE button with YES highlighted to
save the new warning alarm setting.
3.2.6.4 The warning alarm timeout function
The timeout feature is an interval setting that allows the
user to acknowledge the warning alarm and suspend it
for the length of time specified by the interval.
3.2.6.4.1 Adjust timeout interval
To adjust the timeout interval:
1. Enter the warning alarm subdirectory of the Alarm
Menu as described above in section 3.2.6.1. Use
the navigation arrows to highlight TIMEOUT.
25
Page 28
ENABLE
ADJUST
TIMEOUT
EXIT
2. Press the MODE button. The timeout interval
setting will then be shown. Use the up and down
navigation arrows to change the interval setting.
TIMEOU T ADJUST
TIME = 5 min
↓
EXIT
TIMEOU T ADJUST
TIME = 4 min
3. To confirm the new timeout interval setting, press
the MODE button.
EXIT
Save Ch anges?
YES NO CANCEL
4. Press the MODE button with YES highlighted to
save the new warning alarm setting.
3.2.6.4.2 Disabling the timeout interval
1. To disable the warning timeout function entirely,
follow the directions above in section 3.2.6.4.1 and
set the time interval to 0 minutes. The following
screen will then be shown.
TIMEOU T ADJUST
TIMEDISABLED
2. To confirm the new timeout interval setting, press
the MODE button.
EXIT
Save Ch anges?
YES NO CANCEL
3. Press the MODE button with YES highlighted to
save the new warning alarm setting.
3.3 The Calibration Menu
All calibration functions are covered in detail in Chapter
4.
3.4 The Options Menu
The Options Menu controls operating mode selection,
and security beep, language, passcode and toxic
decimal settings. Each of these options is described in
greater detail below.
3.4.1 Entering the Options Menu
1. Enter the Main Menu as described above in section
3.1.1. Use the navigation arrows to highlight
OPTIONS.
ALARMSCALIBRATION
OPTIONS
VIEW TIME
EXIT
2. Press the MODE button. The Options Menu will
then be shown.
SCREEN
USER_MODE SEC_BEEP
PASSCODE LANGUAGE
DECIMAL T_UNITS
3.4.2 User Modes
EXIT
The PhD Lite offers a choice of four modes of operation:
"Text Only”, "Basic”, “Basic/Peak” and "Technician”.
Mode selection should be based on how much
information is required, the skill level of the user, and
the nature of the job.
3.4.2.1 Overview of user modes
•Displays ‘OK’ for gas-level concentrations unless an alarm
condition is pr esent.
•Upon alarm condition, gas-level conc entrations will be
displayed.
•2 screens available (toggle by pressing the MODE
button):
1. Current gas readings scr een (s ee below).
2. Information screen (see below)
• Gas-level concentrations always displayed.
• Access to calibration functions.
• 2 screens available (toggle by pressing the MODE button):
1. Current gas readings screen (see below).
2. Information screen (see below).
• Gas-level concentrations always displayed.
• Access to calibration functions.
• 3 screens available (scroll by pressing the MODE
button):
1. Current gas readings screen (see below).
2. Peak readings screen (s ee below).
3. Information screen (see below).
• Gas-level concentrations always displayed.
• Access to all advanced functions.
• 4 screens available (scroll by pressing the MODE
button):
1. Current gas readings screen (see below).
2. Peak readings screen (s ee below)
3. STEL/TWA/AVG readings screen (see below)
4. Information screen (see below)
Text Only M o de:
Basic Mode
:
Basic/Peak Mode
Technician Mode:
:
Regardless of mode selection, whenever the PhD Lite is
in use it remembers the peak readings of all gases
measured, and is calculating both Time Weighted
Averages and Short Term Exposure Levels for any toxic
gas sensors installed. Regardless of mode selection,
the PhD Lite will go into alarm whenever any alarm set
point is exceeded.
3.4.2.2 Text Only Mode
The simplest mode of operation is “Text Only" mode. In
Text Only Mode during normal operation, the LCD
screen does not display numerical readings, only the
indication "OK” unless an alarm condition is present.
26
Page 29
O2 LEL CO H2S
OK OK OK OK
Current gas readings screen in Text Only Mode, no alarms
present.
If an alarm condition occurs the indication changes from
"OK" to the numerical value, the LED alarm lights flash,
and the audible alarm sounds.
O2 LEL CO H2S
19.1 OK OK OK
Current gas readings screen in Text Only Mode, oxygen
sensor in alarm.
In Text Only Mode, the information screen (see above)
is also available to the user. Press the MODE button to
toggle back and forth between the current gas level and
information screens.
Note: It is n o t poss ib le to in itiate th e “ A utoCalibration” procedure while the PhD Lite is
operated in Text Only Mode. To initiate the
automatic calibration sequence, the PhD Lite must
be in Basic, Basic/Pea k or Techni cian Mode.
3.4.2.3 Basic Mode
In Basic Mode, numerical gas level readings are always
provided and it is possible to initiate “Auto-Calibration” in
order to make fresh air and span calibration
adjustments.
Calibration procedures are discussed in detail in
Chapter 4.
O2 LEL CO H2S
20.9 0 0 0
Current gas readings screen in Basic, Basic/Peak and
Technician Mode s, no al ar m s pr e sen t .
An alarm condition occurs when one of the sensor
readings exceeds the pre-set alarm level. When an
alarm condition occurs, the numerical reading changes
to reflect the new value, the LED alarm lights flash, and
the audible alarm sounds.
O2 LEL CO H2S
20.9 0 0 0
Current gas readings screen in Basic, Basic/Peak and
Technician Mode s, O xy gen sensor in al ar m .
In Basic Mode, the information screen (see above) is
also available to the user. Press the MODE button to
toggle between the current gas and information screens.
3.4.2.4 Basic/Peak Mode
In Basic/Peak Mode, numerical gas level readings are
always provided and the peak readings screen is also
available. From the current gas readings screen, press
the MODE button to scroll through the peak readings
screens and the information screen.
As in Basic Mode, It is possible to initiate the automatic
calibration functions in order to make fresh air and span
calibration adjustments.
Calibration procedures are discussed in detail in
Chapter 4.
3.4.2.4.1 Peak readings
Peak readings represent the highest and lowest values
registered by the instrument during any period of
operation. Peak readings are updated once per second.
PEAK LOW HIGH
O2 20.7 20.9
Peak oxygen readings screen in Basic/Peak or Technician
Mode.
PEAK LEL CO H2S
0 0 0
Peak LEL and toxic readings screen in Basic/Peak or
Technician Mode.
In Basic/Peak Mode, the MODE button is used to scroll
through the current gas level, peak readings, and
information screens.
Note: The procedure for resetting peak readings is
covered in section 2.3.3.2.
3.4.2.5 Technician Mode
Technician Mode provides access to all advanced
functions and displays, including Auto-Calibration.
In Technician Mode, the MODE button is used to scroll
through the current gas level, peak readings, STEL,
TWA, average and information screens.
3.4.2.5.1 STEL
The STEL (Short Term Exposure Limit) for a particular
toxic gas is the maximum average concentration to
which an unprotected worker may be exposed during
any 15 minute interval. The STEL value displayed by
the PhD Lite is the average concentration for the most
recently completed 15 minutes of operation.
STEL CO H2S
0 0
STEL screen in Technician mode.
Note: For the first 15 minutes after the PhD Lite is
initially turned on the STEL reading is a projected
value. The PhD Lite will begin projecting a STEL
value after the first 30 seconds of operation. For the
first 30 seconds the STEL screen will show an "X"
where the reading would normally appear.
The STEL reading is updated twice per minute. Audible
and visible gas alarms will be activated immediately
whenev er the most recent 15-minute average exceeds
the STEL alarm set point.
3.4.2.5.2 TWA readings
Time Weighted Average or TWA values are calculated
by taking the sum of exposure to a particular toxic gas in
the current operating session in terms of parts-permillion-hours and dividing by an eight-hour period.
27
Page 30
TWA CO H2S
1:45 0 0
TWA screen in Technician Mode.
Note: It is not possible to calculate a toxic gas TWA
reading until the PhD Lite has been operating for 15
mi nutes. For the first 15 m inutes after start-up, t he
TWA screen will show an “X” in place of the
calculation. After 15 minutes, the TWA calculation
will be shown.
Appendix A discusses Permissible Exposure Limit
alarm calculations in greater detail.
3.4.2.5.3 Average readings
Average readings are calculated by taking the sum of
exposure to a particular toxic gas in the current
operating session in terms of parts-per-million-hours
and dividing by the amount of time that has passed in
the current operating session.
AVERAGE O2 LEL
20.9 0
Average readings screen (O2 & LEL) in Technician Mode.
3.4.2.6 Changing the user mode
1. Enter the Options Menu as discussed in section
3.4.1 and use the navigation arrows to highlight
USER_MODE.
USER_MODE
PASSCODE LANGUAGE
DECIMAL T_UNITS
SEC_BEEP
EXIT
2. Press the MODE button. The up and down arrows
may then be used to toggle between operating
modes.
Operat ing Mode
Text Only
↑↓
EXIT
Operat ing Mode
Basic
↑↓
EXIT
Operat ing Mode
Basic/ Peak
↑↓
EXIT
Operat ing Mode
Techni cian
3. Once the preferred operating mode is displayed,
press the MODE button.
EXIT
Save Ch anges?
YES NO CANCEL
4. Press the MODE button with YES highlighted to
save the new operating mode.
Note: Changing modes or otherwise
reprogramming the instrument is reserved for
authorized employees.
3.4.3 Security beep
The security beep is an audible alarm that “beeps” on a
regular basis while the PhD Lite is in normal operation.
The beep serves as a reminder that the instrument is
turned on.
3.4.3.1 Adjusting the security beep
1. Enter the Options Menu as discussed in section
3.7.1
USER_MODE
PASSCODE LANGUAGE
DECIMAL T_UNITS
SEC_BEEP
EXIT
2. From the Options Menu use the navigation arrows
to highlight SEC_BEEP. To enable or disable the
security beep, use the right and left navigation
arrows to highlight OFF (or ON). The up and down
navigation arrows may then be used to toggle
between security beep on and security beep off.
Security Beep
Off
↓↑
EXIT
Security Beep
On 01m 00 s
3. The interval setting will appear when the security
beep is turned ON (as above). Use the right and
left navigation arrows to highlight the interval if it
requires adjustment. Then use the up and down
navigation keys to adjust the interval.
EXIT
Security Beep
On 01m 00s
↑
EXIT
Security Beep
On 02m 00s
4. Once the desired interval is reached use the right
and left navigation arrows to highlight EXIT.
EXIT
Security Beep
On 02m 00s
5. Press the MODE button to enter the new setting.
EXIT
Save Ch anges?
YES NO CANCEL
6. Press the MODE button with YES highlighted to
save the new setting.
28
Page 31
3.4.4 Passcode
Access to the automatic span calibration and the
following subdirectories of the Main Menu can be
restricted through the PhD Lite’s passcode feature:
ALARMS, CALIBRATION, OPTIONS, and SCREEN.
The DATALOGGER subdirectory will also be restricted
on units with the datalogger upgrade.
To enter the Passcode subdirectory of the Options
Menu:
1. Enter the Options Menu as discussed in section
3.4.1 and use the navigation arrows to highlight
PASSCODE.
USER_MODESEC_BEEP
PASSCODE
DECIMAL T_UNITS
LANGUAGE
EXIT
2. Press the MODE button. The following screen will
then be shown.
CHANGE_PASSCODE
PASSCODE_ON/OFF
SPAN-LOCK
EXIT
3.4.4.1 Enable/Disable passcode
To enable or disable the passcode:
1. Enter the passcode subdirectory of the Options
Menu as discussed in section 3.4.4.
CHANGE_PASSCODE
PASSCODE_ON/OFF
SPAN-LOCK
EXIT
2. Use the navigation arrows to highlight
PASSCODE_ON/OFF and press the MODE button.
The up and down navigational arrows are then
used to toggle between ENABLED and DISABLED.
PassCode
DISABLED
↓↑
EXIT
PassCode
ENABLED
Note: If this is the first time that the passcode has
been enabled, the PhD Lite will automatically
prompt you to enter a new passcode. See section
3.4.4.2 below for more details.
3. Once the desired setting is displayed, press the
MODE button to enter the setting.
EXIT
Save Ch anges?
YES NO CANCEL
4. Press the MODE button with YES highlighted to
save the new setting.
3.4.4.2 Changing the passcode
To change the passcode:
1. Enter the passcode subdirectory of the Options
Menu as discussed in section 3.4.4.
CHANGE_PASSCODE
PASSCODE_ON/OFF
SPAN-LOCK
EXIT
2. Use the navigation arrows to highlight
CHANGE_PASSCODE and press the MODE
button. The new passcode screen will then be
shown. Use the right and left navigation arrows to
select the digit for change and then use the up and
down navigational arrows to change the digit itself.
New Pass code
0 0 0 0
→↑
ENTER
EXIT
New Pass code
0 1 0 0
3. Once the desired passcode is shown, press the
MODE button once to highlight ENTER.
ENTER
EXIT
New Pass code
0 1 0 0
4. Press the MODE button to enter the new passcode.
The PhD Lite will then proceed to the verification
screen. The new passcode will have to be reentered.
ENTER
EXIT
Verif y Passcode
0 0 0 0
→↑
ENTER
EXIT
New Pass code
0 1 0 0
5. Once the passcode has been re-entered, press the
MODE button. ENTER will then be highlighted.
New Pass code
0 1 0 0
6. Press the MODE button again to confirm the new
passcode.
ENTER
EXIT
ENTER
EXIT
New Pass Code
Saved
Note: Although 0 0 0 0 is a valid initial passcode
value, once the passcode has been changed, 0 0 0 0
becomes invalid and may never be used as a
passcode again.
3.4.4.3 Span Lock
With the Span Lock feature enabled, the instrument will
prompt for the passcode prior to entering the automatic
span calibration subroutine.
Note: The passcode feature must be enabled for the
Span Lock feature to function. See section 3.4.4.1
for instructions on enabling the passcode feature.
29
Page 32
To enable or disable the span lock feature:
1. Enter the passcode subdirectory of the Options
Menu as discussed in section 3.4.4.
CHANGE_PASSCODE
PASSCODE_ON/OFF
SPAN_LOCK
EXIT
2. Use the navigation arrows to highlight SPAN_LOCK
and press the MODE button. The up and down
navigational arrows are used to toggle between
ENABLED and DISABLED.
SPAN-LOCK
DISABLED
↓↑
EXIT
SPAN-LOCK
ENABLED
3.4.5 Language
The PhD Lite can be set to display messages and
readings in a variety of languages. The languages
currently available are Portuguese (Portugues), French
(Francais), Spanish (Espanol), Danish (Dansk), Italian
(Italiano), English, Polish (Polski), Norwegian (Norsk),
Dutch (Nederlands), Catalan (Catala), Lithuanian
(Lietuviskai), Swedish (Svenska), German (Deutsch),
Chinese and Finnish (Suomi).
3.4.5.1 Changing the readout language
1. Enter the Options Menu as discussed in section
3.4.1.
USER_MODE SEC_BEEP
PASSCODE
DECIMAL T_UNITS
2. From the Options Menu use the navigation arrows
to highlight LANGUAGE and press the MODE
button.
PORTUGUES FRANCAIS
ESPANOL DANSK
ITALIANO ENGLISH
POLSKI ---Æ
3. Use the navigation arrows to highlight the language
that is to be used or press the MODE button with
the -Æ highlighted to access more language
options.
PORTUGUES FRANCAIS
ESPANOL DANSK
ITALIANO ENGLISH
POLSKI ---Æ
4. Press the MODE button with the appropriate
language highlighted to enter the new language
setting.
LANGUAGE
EXIT
EXIT
Francais
YES NO
5. Press the MODE button with YES highlighted to
save the new language settings.
Langue
Enregistree
3.4.6 DECIMAL: Changing the precision of the
toxic sensor readout
Toxic gas readings may be given in full parts-per-million
(PPM) increments, or in tenths of parts-per-million (0.1PPM) increments for some sensors. If the decimal point
is enabled, 0.1-PPM increments will be shown during
normal operation of the toxic sensor. If the decimal
point is disabled readings will be shown in full part-permillion increments.
Note: The decimal point for the CO, CO Plus and
the CO side of the Duo-Tox sensor can not be
enabled. CO sensor readings are given only in full
parts-per-million (PPM).
1. Follow the instructions in section 3.4.1 to reach the
Options Menu and use the navigation arrows to
highlight DECIMAL.
USER_MODE SEC_BEEP
PASSCODE LANGUAGE
DECIMAL
2. Press the MODE button. The following screen will
be shown:
T_UNITS
EXIT
TOX1 de cimal pt.
DISABLED
↑↓
EXIT
TOX1 de cimal pt.
ENABLED
Note: If the Duo-Tox sensor is installed, the Tox2
decimal point setting will be shown, which
represents the decimal point setting for the H
sensor.
3. Once the appropriate decimal point setting is
shown, press the MODE button to enter the new
settings.
EXIT
S
2
Save Ch anges?
YES NO CANCEL
4. Press the MODE button with YES
highlighted to save the new decimal settings.
Decima l Points
Saved
3.4.7 Temperature units: Changing the
Temperature readings may be given in Fahrenheit or
Celsius. To change from one temperature scale to the
other, do the following:
1. Follow the instructions in 3.4.1 to reach the Options
temperature scale
Menu.
30
Page 33
USER_MODE SEC_BEEP
PASSCODE LANGUAGE
DECIMAL
T_UNITS
EXIT
2. Use the navigation arrows to highlight T_UNITS
and press the MODE button. The up and down
navigation arrows may then be used to toggle
between Celsius and Fahrenheit.
TEMPER ATURE UNIT S
FAHREN HEIT
EXIT
↑↓
TEMPER ATURE UNIT S
CELSIU S
EXIT
3. Once the appropriate setting is shown, press the
MODE button to enter the new setting
Save Ch anges?
YES NO CANCEL
4. Press the MODE button with YES highlighted to
save the new temperature setting.
Note: The temperature shown is actually a reading
taken on the inside of the instrument case in the
area where the sensors are located and may not
correspond with ambient air temperatures. The PhD
Lite microprocessor uses these readings to
compensate for temperature changes in the sensor
compartment.
3.5 The Screen Menu
The Screen Menu provides access to the PhD Lite’s
adjustable contrast and backlight settings.
3.5.1 Entering the Screen Menu
1. Enter the Main Menu as described above in section
3.1.1. Use the navigation arrows to highlight
SCREEN.
ALARMS CALIBRATION
OPTIONS
VIEW TIME
2. Press the MODE button to enter the Screen Menu.
CONTRAST
3.5.2 Adjusting the contrast
To adjust the screen contrast:
EXIT
1. Enter the Screen Menu as described above in
section 3.5.1. Use the navigation arrows to
highlight CONTRAST.
2. Press the MODE button.
SCREEN
EXIT
BACKLIGHT
ORIENTATION
CONTRAST
BACKLIGHT
ORIENTATION
EXIT
Contra st Adjust
↑ or ↓
3. Adjust the contrast with the up and down navigation
arrows. When the contrast reaches the desired
level press the MODE button.
EXIT
Save Ch anges?
YES NO CANCEL
4. Press the MODE button with YES highlighted to
save the new contrast settings.
3.5.3 Adjusting the backlight
Once the backlight is activated, it will remain on for the
length of time specified in the backlight setting. The
backlight may also be set to remain on always. To
adjust the backlight setting:
1. Enter the Screen Menu as described above in
section 3.5.1. Use the navigation arrows to
highlight BACKLIGHT.
2. Press the MODE button.
CONTRAST
BACKLIGHT
ORIENTATION
EXIT
Backlight
Offafter20s
The PhD Lite’s backlight may be configured to either be
“ON Always” or to automatically turn off after an interval
of between 5 and 90 seconds.
3.5.3.1 Backlight ON Always
To turn the PhD Lite’s backlight “ON Always” perform
the following steps:
1. Follow the instructions above in section 3.5.3 to
reach the Backlight settings and use the right
navigation arrow to highlight Off.
EXIT
Backlight
Offafter20s
2. The up and down navigation arrows may then be
used to toggle between “OFF after” to “ON always.
EXIT
Backlight
Offafter20s
↓↑
EXIT
Backlight
On -always
3. Once the appropriate backlight setting is reached,
use the navigation arrows to highlight EXIT and
press the MODE button to confirm the new setting.
EXIT
31
Page 34
Backlight
On -always
EXIT
Save Changes?
YES NO CANCEL
4. Press the mode button with YES highlighted to
save the new backlight setting.
3.5.3.2 Adjusting the backlight interval
1. Follow the instructions above in section 3.5.3 to
reach the backlight settings and use the navigation
arrows to highlight the time interval in seconds.
Backlight
Offafter20s
2. The up and down navigation arrows may then be
used to adjust the interval to any length of time
between 5 and 90 seconds.
EXIT
Backlight
Offafter20s
↑
EXIT
Backlight
Offafter21s
3. Once the appropriate interval is reached, use the
navigation arrows to highlight EXIT and press the
MODE button to enter the new setting.
EXIT
Save Ch anges?
YES NO CANCEL
4. Press the mode button with YES highlighted to
save the new interval setting.
3.5.4 Adjusting the screen orientation
The screen orientation and text properties are controlled
though the orientation submenu.
3. Press the MODE button when finished.
Save Ch anges?
YES NO CANCEL
4. Press the mode button with YES highlighted to
save the new display setting.
3.5.4.2 Invert text color
The Display may be configured to show in either
Standard (white background with black letters) or
Inverse (black background with white letters). The
Invert text color option is a hidden function that is not
shown on the display of the PhD Lite.
O2 LEL CO H2S
20.9 0 0 0
Standard display settings
O2 LEL CO H2S
20.9 0 0 0
Inverse display settings
To change the setting:
1. Enter the Screen Menu as described above in
section 3.5.1. Use the navigation arrows to
highlight ORIENTATION.
2. Press the MODE button. The “Flip Display” option
will be shown.
↑ or ↓ to Flip
Display
3. To invert the text color on the display, press the left
or right arrow once.
EXIT
↑ or ↓ to Flip
Display
EXIT
→
3.5.4.1 Flip Display
To rotate the screen orientation 180 degrees:
1. Enter the Screen Menu as described above in
section 3.5.1. Use the navigation arrows to
highlight ORIENTATION.
CONTRAST
EXIT
BACKLIGHT
ORIENTATION
2. Press the MODE button. Press the up or down
arrow to rotate the display.
↑ or ↓ to Flip
Display
↑↓
EXIT
↑ or ↓ to Flip
Display
3. Press the MODE button when finished.
EXIT
Save Ch anges?
YES NO CANCEL
4. Press the mode button with YES highlighted to
save the new display setting.
3.6 The View Menu
Information pertaining to alarm levels, current battery
voltage, service dates and software versions may be
viewed, but not altered through the View Menu.
32
Page 35
3.6.1 Entering the View Menu
1. Enter the Main Menu as described above in section
3.1.1. Use the navigation arrows to highlight VIEW.
2. Press the MODE button to enter the View Menu.
3.6.2 View alarm levels
ALARMS CALIBRATION
OPTIONS SCREEN
VIEW
DISPLAY
ALARM_LEVELS
BAT_VOLTAGE
SERVICE
VERSION
TIME
EXIT
EXIT
1. Enter the View Menu as discussed above in section
3.6.1. Use the navigation arrows to highlight
ALARM_LEVELS.
ALARM_LEVELS
BAT_VOLTAGE
SERVICE
VERSION EXIT
2. Press the MODE button. The current alarm
settings will then be shown.
CURREN T ALARM
LEVELS
LOW ALAR M LEVEL
O2 = 18.0 %
3. The right and left navigation arrows may then be
used to scroll through the various alarm settings.
EXIT
LOW ALAR M LEVEL
O2 = 18.0 %
→
EXIT
HIGH AL ARM LEVEL
O2 = 23.5 %
→
EXIT
HIGH AL ARM LEVEL
LEL = 10 %
4. Press the MODE button at any time to return to the
Main Menu.
3.6.3 View battery voltage
1. Enter the View Menu as discussed above in section
3.6.1. Use the navigation arrows to highlight
BAT_VOLTAGE.
ALARM_LEVELS
BAT_VOLTAGE
SERVICE
VERSION EXIT
2. Press the MODE button. The battery voltage
screen will be shown.
EXIT
Batter y Voltage
3.6 Volts
EXIT
3. Press the MODE button to return to the Main Menu.
3.6.4 View service information
1. Enter the View Menu as discussed above in section
3.6.1. Use the navigation arrows to highlight
SERVICE.
ALARM_LEVELS
BAT_VOLTAGE
SERVICE
VERSIONEXIT
2. Press the MODE button. The service information
screen will be shown.
For Serv ice Call
860-344-1079
3. Press the MODE button to return to the Main Menu.
3.6.5 View Software Version
The Version subdirectory of the View Menu contains
information on the current software version, current OTP
version and the date of the last update.
1. Enter the View Menu as discussed above in section
3.6.1. Use the navigation arrows to highlight
VERSION.
2. Press the MODE button. The software version will
then be shown. Use the right and left navigation
arrows to scroll through the three available screens.
ALARM_LEVELS
BAT_VOLTAGE
SERVICE
VERSION
EXIT
EXIT
biosystems
PhD Lite V.3.18
→
biosystems
OTP V01.10
→
2003/09/ 08
14:14:34
3. Press the MODE button to return to the Main Menu.
3.7 The Time Menu
For PhD Lite instruments that do not include the
datalogger upgrade, the Time Menu provides access to
date and time-related functions.
Note: PhD Lite instruments with instrument
firmware earlier than version 1.29 may not include
the Time menu option.
Note: If your PhD Lite shows DATALOGGER in
place of TIME in the Options Menu, proceed to
Chapter 5.
3.7.1 Entering the Time Menu
1. Enter the Main Menu as described above in section
3.1.1. Use the navigation arrows to highlight TIME.
33
Page 36
ALARMS CALIBRATION
OPTIONS SCREEN
VIEW
2. Press the MODE button to enter the Time Menu.
3.7.2 Time and date settings
To change the time and date:
TIME
EXIT
TIME/DATE
SERV_DATE
EXIT
1. Enter the Time Menu as described above in section
3.7.1. Use the navigation arrows to highlight
TIME/DATE.
2. Press the MODE button.
TIME/DATE
SERV_DATE
EXIT
Date 23 SEP 20 03
Time 14:10
3. The right and left navigation arrows are used to
move back and forth between the day, month, year,
hour and minute settings and the EXIT option.
Once the time setting that needs to be adjusted is
highlighted, the up and down navigation arrows are
used to make the adjustment.
EXIT
Date 23 SEP 20 03
Time 11:10
→↑
EXIT
Date 18 OCT 20 03
Time 11:10
4. The MODE button may be pressed at any time to
enter the new settings.
EXIT
Save Ch anges?
YES NO CANCEL
5. Press the MODE button with YES highlighted to
confirm the new time and date settings.
3.7.3 Service date settings
The PhD Lite can be programmed to automatically
remind the user to service the sensors in the unit. To
reach the service due date settings for the individual
sensors:
1. Enter the Time Menu as described above in section
3.7.1. Use the navigation arrows to highlight
SERV_DATE.
2. Press the MODE button. The following screen will
then be shown.
TIME/DATE
SERV_DATE
EXIT
CHANGE_SERV_DATE
SERVICE_ON/OFF
EXIT
3.7.3.1 Enable/Disable sensor service due dates
1. Enter the service date subdirectory of the Time
Menu as described above. Use the up navigation
arrow to highlight SERVICE_ON/OFF.
CHANGE_SERV_DATE
SERVICE_ON/OFF
EXIT
2. Press the MODE button. The following screens will
then be shown in succession.
SENSOR SERVICE
EXPIRATION ALARMS
O2 SERVIC E ALARM
DISABLED
3. Use the right and left navigation arrows to scroll
through the service setting date for each sensor.
EXIT
O2 SERVIC E ALARM
DISABLED
→
EXIT
LEL SERV ICE ALARM
DISABLED
4. Once the appropriate sensor is reached, use the up
and down navigation arrows to change the setting.
EXIT
LEL SERV ICE ALARM
DISABLED
↓↑
EXIT
LEL SERV ICE ALARM
ENABLED
5. Once the appropriate sensor due date settings are
shown, press the MODE button to enter the
settings.
EXIT
Save Ch anges?
YES NO CANCEL
6. Press the MODE button with YES highlighted to
confirm the new sensor service due dates.
3.7.3.2 Change sensor service due date settings
1. Enter the service date subdirectory of the Time
Menu as described above. Use the up navigation
arrow to highlight CHANGE_SERVICE_DATE.
CHANGE_SERV_DATE
SERVICE_ON/OFF
EXIT
34
Page 37
2. Press the MODE button. The following screen will
then be shown in succession.
SENSOR SERVICE
EXPIRATION DATES
NEXT O2 SERVIC E
28 OCT 2003
NEXT
EXIT
3. To advance to another sensor, press the MODE
button with NEXT highlighted.
NEXT O2 SERVIC E
28 OCT 2003
NEXT
EXIT
NEXT LE L SERVICE
28 OCT 2003
4. Once the appropriate sensor is reached, use the
right and left navigation arrows to highlight the day,
month or year that requires adjustment.
NEXT LE L SERVICE
28 OCT 2003
←←
NEXT
EXIT
NEXT
EXIT
NEXT LE L SERVICE
28 OCT 2003
5. Use the up and down navigation arrows to change
the setting.
NEXT
EXIT
NEXT LE L SERVICE
28 OCT 2003
↑
NEXT
EXIT
NEXT LE L SERVICE
28 NOV 2003
6. Once the sensor service dates have been changed
use the right and left navigation arrows to highlight
EXIT.
NEXT
EXIT
NEXT LE L SERVICE
28 NOV 2003
7. Press the MODE button to enter the new service
dates.
NEXT
EXIT
Save Ch anges?
YES NO CANCEL
8. Press the MODE button with YES highlighted to
confirm the new sensor service due dates.
35
Page 38
Chapter 4. Calibration
The PhD Lite multi-gas detector has been designed for
easy calibration. A single control, the on/off MODE
button, can be used to initiate the automatic calibration
sequence and to automatically make calibration
adjustments.
Manual and single-sensor calibration procedures can
also be initiated by using the navigation arrows located
on the instrument.
Note: If a sensor has just been replaced, it must be
allowed to stabilize prior to initiating any of the
calibration subroutines detailed below. See section
6.1.1 for further details concerning sensor
stabilization requirements.
The PhD Lite’s automatic calibration features are
described in section 4.4 below.
The PhD Lite’s manual calibration features are
described in section 4.5 below.
4.1 Verification of accuracy
Verification of accuracy is a two step procedure.
Step one is to take the PhD Lite to an area where the
atmosphere is known to be fresh and check the
readings. If the readings differ from those expected in
fresh air, then a fresh air calibration adjustment must be
made.
Step two is to make sure the sensors are accurate by
exposing them to a test gas of known concentration and
noting the sensor response. This procedure is known
as a functional (bump) test and is covered in section
4.6. Oxygen readings are considered to be accurate
when the display is within 0.5%/volume of the expected
concentration as given on the calibration gas cylinder.
LEL and toxic readings are considered accurate when
they are between 90%* and 120% of the expected value
as given on the calibration gas cylinder. If readings are
accurate, there is no need to adjust your gas detector.
If the readings are inaccurate, the instrument must be
span calibrated before further use.
* * The Canadian Standards Association (CSA)
requires the instrument to undergo calibration when
the displayed value during a bump test fails to fall
between 100% and 120% of the expected value for
the gas.
The accuracy of the PhD Lite
should be checked periodically with known
concentration calibration gas. Failure to check
accuracy can lead to inaccurate and potentially
dangerous readings. (The Canadian Standards
Association (CSA) requires an accuracy check
using known concentration calibration gas prior to
each day’s use.)
Always check the expiration
date on calibration gas cylinder(s) prior to use.
Expired calibration gas can lead to inaccurate and
potentially dangerous readings.
Biosystems offers calibration kits and long lasting
cylinders of test gas specifically developed for easy
PhD Lite calibration.
Use of non-standard calibration
gas and/or calibration kit components when
calibrating the PhD Lite can lead to inaccurate and
potentially dangerous readings, and may void the
standard Biosystems warranty.
Customers are strongly urged to use only
Biosystems calibration materials when calibrating
the PhD Lite.
4.2 Effect of contaminants on PhD Lite sensors
The atmosphere in which the PhD Lite is used can have
lasting effects on the sensors. Sensors may suffer
losses in sensitivity leading to degraded performance if
exposed to certain substances.
There are three basic types of sensors that may be
installed in the PhD Lite: galvanic oxygen, catalytic hotbead combustible gas, and electrochemical toxic. Each
type of sensor uses a slightly different detection
principle, so the kinds of conditions that affect the
accuracy of the sensors vary from one type of sensor to
the next.
4.2.1 Effects of contaminants on oxygen sensors
Oxygen sensors may be affected by prolonged
exposure to "acid" gases such as carbon dioxide. The
oxygen sensors used in Biosystems instruments are not
recommended for continuous use in atmospheres
containing more than 25% CO
.
2
See Appendix B for cross-sensitivity data for the
sensors used in the PhD Lite.
4.2.2 Effects of contaminants on combustible
sensors
Combustible sensors will be adversely affected by
exposure to substances containing volatile silicone,
which is found in many commercial formulations such as
spray lubricants, plastic mold(ing) release agents,
waterproofing agents, heat transfer fluids, and is
released during the cure of silicone-based caulks and
rubbers (RTV). Other combustible gas sensor poisons
and inhibitors include, but are not limited to: tetraethyl
lead as in "leaded" gasoline grades (aviation "low-lead"
fuel), halogenated hydrocarbons such as Freons
TM
,
other such refrigerants and solvents such as 1,1,1trichloroethane, perchloroethylene and methylene
chloride. Chronic exposures to high concentrations
(above human health and safety levels) of hydrogen
sulfide (H
combustible sensor performance.
S) and Phosphine (PH3) can also impair
2
Note: Damage to combustible gas sensors incurred
by exposure to known sensor poisons such as
silicones, tetra-ethyl lead, and/or other substances
may (at the discretion of Biosystems’ Instrument
Service Department) void Biosystems’ Standard
Warranty as it applies to the replacement of
36
Page 39
combustible gas sensors. For a more complete list
of known sensor poisons see Biosystems Standard
Warranty in Appendix E.
After any detector exposure to a suspected or known
poison/inhibitor source, combustible sensor accuracy
should be verified immediately by exposure to
calibration gas of known percent LEL concentration.
Note: If the combustible sensor in the PhD Lite
suffers a loss of sensitivity, it tends to be lost first
with regards to methane.
As described above, combustible gas sensors may
become desensitized if exposed to certain substances.
In some cases a desensitized combustible sensor may
still respond accurately to propane and other
hydrocarbons while showing a dangerously reduced
response to methane.
Biosystems’ “Propane Equivalent” calibration gas
mixtures have been developed to eliminate this
potentially dangerous source of calibration error.
Biosystems’ “Propane Equivalent” mixtures are based
on methane, so any loss of sensitivity to methane is
detected (and can be corrected) immediately.
Using Biosystems brand calibration gas and
regularly verifying accuracy ensures that proper
sensi tivity is main ta in ed for the life of the se nsor .
4.2.3 Effe cts of high concentrations of
combustible gas on the combustible sensor
The accuracy of combustible sensors may also be
affected by exposure to high concentrations of
combustible gas. To minimize the chance for damage
or loss of sensitivity to the combustible sensor, the PhD
Lite is designed to "alarm latch" whenever the
concentration of combustible gas exceeds 100 percent
LEL. Under these conditions an “X” will appear in place
of the combustible gas reading to indicate that an overlimit condition has occurred, and “LEL OVERRANGE”
will be displayed at the bottom of the LCD.
During an “LEL OVERRRANGE” condition, the power to
the LEL sensor is interrupted and the audible and visible
alarms are activated until the instrument is manually
reset by turning it off.
A combustible sensor
overrange alarm indicates a potentially explosive
atmosphere. Failure to leave the area immediately
may result in serious injury or death!
In the event of a combustible
sensor overrange alarm, the PhD Lite must be
turned off, brought to an area that is known to be
safe and then turned on again to reset the alarm.
Make sure that the PhD Lite is
located in fresh air before turning the instrument
back on after a combustible sensor alarm latch
condition has occurred. Fresh air calibration
adjustments may only be made w hen the PhD Lite is
locate d in air th at is kn o wn to be fr esh. After a
combustible sensor alarm-latch condition occurs,
the accuracy of the combustible gas sensor must be
verified by exposure to known percentage LEL
concentration test gas before further use.
Note: The combustible sensor used in the PhD Lite
requires a minimum of 10% oxygen by volume in
order to generate accurate combustible gas
readings. Combustible sensor accuracy may be
diminished if the instrument is used in oxygendeficient atmospheres.
A rapid up-scale reading
followed by a declining or erratic reading may
indicate a hazardous combustible gas concentration
that exceeds the PhD Lite’s zero to 100 percent LEL
detection range. Failure to leave the area
immediately may result in serious injury or death!
4.2.4 Effects of contaminants on toxic gas
sensors
Biosystems “substance-specific” electrochemical “smart
sensors” used to measure many toxic gases have been
carefully designed to minimize the effects of common
interfering gases. “Substance-specific” sensors are
designed to respond only to the gases that they are
supposed to measure. The higher the specificity of the
sensor, the less likely the sensor will react to other
gases, which may be incidentally present in the
environment. For instance, a “substance-specific”
carbon monoxide sensor is deliberately designed not to
respond to other gases that may be present at the same
time, such as hydrogen sulfide and methane.
Although great care has been taken to reduce crosssensitivity, some interfering gases may still have an
effect on toxic sensor readings. In some cases the
interfering effect may be positive and result in readings
that are higher than actual. In other cases the
interference may be negative and produce readings that
are lower than actual or even result in negative gas
readings.
4.3 Single sensors capable of monitoring for
two different gases
The OSHA standard for permit-required confined space
entry (29 CFR 1910.146) explicitly requires the use of a
direct-reading, substance-specific sensor whenever a
particular toxic hazard is likely to be present. For
exam ple, if hydrogen sulfide is likely to be present, one
of the toxic sensors selected should be specifically
designed for the direct detection of H
S.
2
37
Page 40
4.3.1 Using one sensor to monitor for Carbon
Monoxide and Hydrogen Sulfide
Carbon monoxide and hydrogen sulfide are the two
most common toxic gases associated with confined
space entry. Biosystems offers two different sensors,
the Duo-Tox and the CO Plus, which allow the user to
monitor for both of these hazards while utilizing only one
sensor port.
4.3.1.1 Biosystems Duo-Tox dual purpose Carbon
Monoxide/Hydrogen Sulfide sensor
The Duo-Tox sensor is a substance-specific, dualchannel, electrochemical sensor designed to directly
detect both carbon monoxide and hydrogen sulfide
without cross interference. When the Duo-Tox sensor is
installed in the PhD Lite, it provides two independent
channels of monitoring information while using only one
sensor port. This allows the instrument to provide four
channels of detection with only three sensors installed.
A PhD Lite equipped with a Duo-Tox sensor will show
both CO and H
S on the current gas readings screen.
2
O2 LEL CO H2S
20.9 0 0 0
Current gas readings screen with O2, LEL and Duo-Tox
sensor installed.
Calibration gas containing known concentrations of both
CO and H
Tox sensor.
S must be used to properly calibrate the Duo-
2
4.3.1.2 Biosystems “CO Plus” dual purpose carbon
monoxide/hydrogen sulfide sensor
The “CO Plus” sensor is ideal for situations requiring the
use of a single sensor to monitor simultaneously for
both CO and H
S, in which the user does not need to
2
definitively know which hazard is being encountered.
While the “CO Plus” sensor will simultaneously detect
both carbon monoxide and hydrogen sulfide, it is only
possible to directly monitor for one of these hazards.
The PhD Lite offers a choice of two different calibration
gas settings for the CO Plus sensor. Calibration gas
setting determines whether the PhD Lite is configured
for the direct detection of CO or H
S. When the
2
calibration gas value is set to CO, the PhD Lite will show
CO+ on the current gas readings screen. Alternately,
when the calibration gas value is set to H
Lite will show H
S+ on the current gas readings screen.
2
S, the PhD
2
O2 LEL CO+
20.9 0 0
Current gas re adings screens with CO Plus sensor;
calibration gas value set to CO.
O2 LEL H2S+
20.9 0 0
Current gas re adings screens with CO Plus sensor;
calibration gas value set to H
S.
2
Note: The procedure for changing the calibration
gas value setting for the CO Plus sensor is covered
below in section 4.7.2.2.
4.3.1.2.1 Relative response of the CO Plus sensor
to carbon monoxide and hydrogen sulfide
The “CO Plus” sensor will accurately and directly
measure the gas type to which it has been successfully
calibrated. OSHA has assigned an 8-hour TWA of 35
PPM as the permissible exposure limit for carbon
monoxide. If the “CO Plus” sensor is successfully
calibrated to carbon monoxide, and then exposed to 35
PPM carbon monoxide, the display will show 35 PPM.
Appendix A discusses Permissible Exposure Limit
alarm calculations in greater detail.
Regardless of the calibration gas selection, the “CO
Plus” sensor will also show a “relative response” to
other interfering gases. When calibrated with carbon
monoxide, the “CO Plus” sensor responds to hydrogen
sulfide in a ratio of about 3.5 to 1.0. This means a
concentration of about 10 PPM hydrogen sulfide would
produce a reading of 10 X 3.5 or 35 PPM in an
instrum ent that has been calibrated to CO.
This is a very convenient relative response. The 8-hour
TWA permissible exposure limit for hydrogen sulfide is
10 PPM. This means that even when configured for the
direct reading of CO, the CO Plus gas alarms will be
activated any time the concentration of hydrogen sulfide
exceeds the permissible exposure limit.
Note: The procedure for changing the calibration
gas value setting for the CO Plus sensor is covered
below in section 4.7.2.2.
4.3.2 Cl2 and ClO2 Non-Specific sensors
Four different sensors are currently available for the
PhD Lite for the detection of chlorine (Cl
dioxide (ClO
(part numbers 54-25-08 and 54-25-12) have built-
ClO
2
in cross-sensitivity as follows:
4.3.2.1 Cl2 Non-Specific
). The non-specific versions of the Cl2 and
2
) and chlorine
2
The 54-25-08 Cl2 non-specific sensor is cross sensitive
at a rate of 3.1 to 1. This means that when a
to ClO
2
properly calibrated PhD Lite with a Cl
sensor encounters 1PPM ClO
a reading of 3.1PPM Cl
4.3.2.2 ClO2 Non-Specific
2
.
2
non-specific
2
, it will respond by giving
The 54-25-12 ClO2 non-specific sensor is cross
sensitive to Cl
when a properly calibrated PhD Lite with a ClO
specific sensor encounters 1PPM Cl
giving a reading of approximately 0.3PPM ClO
at a rate of 1 to 3.1. This means that
2
, it will respond by
2
non-
2
.
2
Note: For more information on cross sensitivity for
any of the sensors available in the PhD Lite, see
Appendix B.
38
Page 41
4.4 Automatic calibration
Biosystems one-button auto-calibration procedure may
be used to calibrate the PhD Lite at any time during
normal operation in all modes except Text-Only Mode.
Auto-calibration is a two-step procedure. In the first
step the PhD Lite is taken to an area where the
atmosphere is known to be fresh and a fresh air
adjustment is made as detailed below.
The second step is the sensor response or "span"
calibration adjustment. In this step the accuracy of the
PhD Lite sensors is established by exposing them to
known concentration calibration gas(es). The sensitivity
or “span” is automatically adjusted by the instrument.
Note: If the Passcode and Span Lock functions are
enabled, the PhD Lite will prompt the user for the
passcode prior to entering the span calibration
subroutine. See section 3.4.4.3 for further details.
The automatic fresh air calibration procedure may only
be performed while the instrument is being operated in
Technician, Basic/Peak or Basic operating modes.
1. Turn the instrument on and make sure gas readings
are given in numbers. (This is an indication that the
PhD Lite is currently in Basic, Basic/Peak or
Technician Mode).
If sensor readings are displayed in the form of “OK”
text messages, the instrument is currently in TextOnly mode. It will be necessary to change the
operating mode to Basic, Basic/Peak or Technician
mode as explained in section 2.3.5 before
proceeding with the calibration.
2. Wait at least three minutes after turning the
instrument on to allow sensor readings to stabilize
fully before initiating any calibration procedures. If
a sensor has just been replaced, follow the sensor
stabilization schedule in section 6.1.1.
3. Make sure the instrument is located in an area
where the air is known to be fresh.
4. Press the MODE button three times within two
seconds. This will "wake up" the instrument from
normal operation, and initiate the auto-calibration
sequence. A screen will briefly display the
message “One Button Auto-Calibration”.
One Butt on Auto
Calibration
The instrument will then proceed to the following screen
and begin a 5-second countdown.
Zero Ca libration
Mode = Ad just 5
5. To initiate the fresh air/zero calibration, press the
MODE button before the unit finishes the
countdown.
Zero Ca libration
Please Wait
Zero Ca libration
Completed
Upon successful completion of the fresh air/zero
calibration, the instrument will automatically proceed to
the automatic span calibration sequence.
4.4.1.1 Reading “Too High” or “Too Low” for zero
adjust
To reduce the chances of the PhD Lite being
inadvertently fresh air/zero calibrated in contaminated
air, only small adjustments are allowed in the automatic
fresh air/zero calibration sequence. If the necessary
adjustm ents are too large, the display will indicate the
sensor (or sensors) affected, and a message screen will
indicate that the reading is “Too Low” or “Too High” for
zero adjustment. In this case the instrument must be
fresh air calibrated using the manual calibration
procedures as discussed in section 4.5.1 of this manual.
Oxygen Too Low
for Zero Cal
Once the instrument has been successfully fresh air
calibrated using the manual fresh air calibration
procedure, subsequent calibration adjustments may be
made using the MODE button and automatic calibration
logic discussed in this section.
4.4.2 Automatic span calibration sequence
Following the completion of the automatic fresh air
calibration, the PhD Lite will display the countdown for
the automatic span calibration:
Note: If the Passcode and Span Lock functions are
enabled, the PhD Lite will prompt the user for the
passcode prior to entering the span calibration
subroutine. See section 3.4.4.3 for further details.
Span Ca libration
Mode = Ad just 5
If span calibration is not required, allow the unit to
complete the 5-second countdown without pressing any
buttons.
Press MODE before the before the end of the 5-second
countdown to initiate the automatic span calibration
sequence. The PhD Lite will then display:
Apply Ga s
MODE = Ca ncel
Attach the cylinder of calibration gas to the PhD Lite
through the regulator, tubing and calibration adapter as
shown below in figure 4.4.2. Continue to flow gas
throughout the span calibration procedure.
Note: The PhD Lite requires a fixed flow rate of 1.0
liter/minute for all calibration procedures.
39
Page 42
Figure 4.4 . 2 T y p ical cal i bration set - up .
The PhD Lite automatically recognizes the type of gas
supplied and displays the current reading for each
sensor that may be calibrated using the current gas
mixture. The span adjustm ent process from this point
on is automatic and requires no user input.
Please Wait
If multi-component calibration gas is being used, the
following sequence will be shown.
Multi Ca l Gas
Detected
LEL Sensor
Reading 48
Once the readings stabilize, the PhD Lite will
automatically adjust itself so that the sensor readings
match the expected gas concentration values as they
appear in the gas values subdirectory of the calibration
menu.
LEL Set to 50
MAX ADJUSTMENT 171
Note: The max adjustment value shown is an
indication of the relative health of the sensor. As a
sensor loses sensitivity, the maximum adjustment
level will approach the calibration gas
concentration.
Once the LEL sensor is adjusted, the instrument will
move on to the next sensor and perform a similar
adjustment.
CO Sensor
Reading 52
CO Set to 50
MAX ADJUSTMENT 181
Once the span-calibration is successfully completed, the
instrument will show the following screen prior to turning
itself off.
Auto Ca libration
Completed
Begin Sh ut Down
Please Wait
After shut down, remove all gas fittings from the PhD
Lite, and disconnect the regulator from the calibration
gas cylinder. Then press the MODE button to turn the
instrument on and resume normal operation.
Note: If the MODE button is pressed at any time
prior to c ompletion of the calibra tion, th e c a libra t io n
procedur e will be cancelled an d the in strument will
return to no rmal ope ra tion.
Wait at least 3 minutes after
turning the instrument on to allow sensor readings
to stabilize fully before initiating any calibration
procedures. Failure to wait three minutes before
initiating calibration procedures may lead to
inaccurate and potentially dangerous readings.
Make sure the regulator,
cylinder seating surfaces and threads are clean and
dry before attaching the regulator to the cylinder of
gas. Introduction of contaminants through the
regulator fittings may alter or degrade the
concentration of the gas contained in the cylinder
and may lead to inaccurate and potentially
dangerous gas readings.
CAUTION: The use of regulators with on/off
valves may introduce contaminants into the gas
cylinder and may lead to the breakdown of reactive
gases prior to the cylinder expiration date.
Biosystems strongly recommends the use of
continuous fixed-flow regulators for all calibration
procedures.
gas and/or calibration kit components when
calibrating the PhD Lite may lead to dangerously
inaccurate readings and may void the standard
Biosystems warranty.
4.4.3 Automatic span calibration with more than
one gas cylinder
Depending on the sensor configuration of the PhD Lite,
it may be necessary to use more than one cylinder of
calibration gas to complete the calibration.
The PhD Lite will automatically recognize the type of
gas that is applied and proceed with the span calibration
of the sensor(s) that react to the gas. Once that/those
sensor(s) have been successfully calibrated, the PhD
Use of non-standard calibration
Apply Ga s
MODE = Ca ncel
40
Page 43
Lite will automatically request the calibration gas for the
next sensor to be calibrated.
Note: If multiple cylinders of calibration gas are
used during calibration, it will be necessary to
change cylinders between span adjustments. In
this case the display will indicate the type and
concentration of the next cylinder of calibration gas
to be applied.
Apply 10PPM SO2
Once the instrument detects the proper gas, the
instrum ent will proceed to automatically calibrate the
sensor channel.
SO2 Sensor
Reading 10
When all sensors currently recognized have been
successfully span-calibrated, the instrument will display
“Auto Calibration Completed” and proceed to shut itself
off.
After shut down, remove all gas fittings from the PhD
Lite, and disconnect the regulator from the calibration
gas cylinder. Then press the MODE button to turn the
instrument on and resume normal operation.
Note: It is possible to exit the automatic span
calibration sequence at any time prior to completion
by pressing and holding down the MODE button to
turn the instrument off. The instrument will retain
the updated settings for those sensors whose span
adjustments have been successfully completed.
Sensors that were not successfully adjusted at the
time the auto-calibration sequence was terminated
will trigger a “Needs Cal” message at the time the
instrument is next turned on. The accuracy of those
remaining sensors should be verified by exposure
to known concentration test gas before the
instrument is put back into service.
4.5 Manual calibration
It is possible to calibrate the PhD Lite manually by using
the navigation arrows to enter the Calibration Menu and
to then select the desired calibration procedure.
4.5.1 Manual fresh air calibration procedure
In some cases, it will be necessary to manually fresh air
calibrate the PhD Lite. Manual fresh air calibrations are
performed as follows:
1. Turn the instrument on and wait at least three
minutes to allow sensor readings to stabilize fully
before initiating any calibration procedures.
2. Make sure the instrument is located in an area
where the air is known to be fresh.
3. Enter the Main Menu as described in section 3.1
above.
ALARMS
OPTIONS SCREEN
VIEW TIME
CALIBRATION
EXIT
4. Use the navigation arrows to select CALIBRATION
and then press the MODE button.
FRESH_AIR
SPAN_CAL HISTORY
GAS_VALUES LAST_CAL
IQ LINK EXIT
REMIND
5. Use the navigation arrows to highlight FRESH_AIR
and press the MODE button.
Do Zero Ca l Now?
YES NO
6. Press the MODE button with YES highlighted to
initiate the zero calibration.
Zero Ca libration
Please Wait
Zero Ca libration
Completed
7. After the completion of the fresh air calibration the
PhD Lite will automatically return to the Main Menu.
ALARMS CALIBRATION
OPTIONS SCREEN
VIEW TIME
EXIT
8. Press the MODE button with EXIT highlighted to
return to the current gas readings screen.
4.5.1.1 Shortcuts to the manual fresh air
calibration procedures
There are two additional ways to reach the manual fresh
air calibration procedure detailed above.
4.5.1.1.1 Shortcut to fresh air calibration while in
normal operation
At the current gas readings screen in Basic, Basic/Peak
or Technician mode, press and hold the left navigation
arrow for six seconds until the PhD Lite instructs you to
release it.
Releas e Button
Do Zero Ca l Now?
YES NO
The instrument will then proceed directly to the manual
fresh air calibration detailed above in section 4.5.1.
Note: If no action is taken within fifteen seconds of
entering the “Do Zero Cal Now?” screen, the
instrument will automatically return to the current
gas readings screen.
41
Page 44
4.5.1.1.2 S h o r tc ut to Ma in Menu while turning the
PhD Lite on
Turn on the PhD Lite while holding down the left
navigation arrow. After the initial start up sequence the
Main Menu screen will be displayed.
Entering
Main Menu
To enter the fresh air calibration from the Main Menu,
follow the directions above in section 4.5.1 starting at
step 3.
4.5.2 Manual span calibration procedures
Manual span calibration procedures are useful when the
calibration of a single sensor is required or when nonstandard calibration gas is being used. Manual span
calibrations are performed as follows:
1. Turn the instrument on.
2. Wait at least three minutes after turning the
instrument on to allow sensor readings to stabilize
fully before initiating any calibration procedures. If
a sensor has just been replaced, follow the sensor
stabilization schedule in section 6.1.1.
3. Enter the Main Menu as described above in section
3.1. Use the navigation arrows to highlight
CALIBRATION.
ALARMS
OPTIONS SCREEN
VIEW TIME
CALIBRATION
EXIT
4. Press the MODE button. The calibration menu will
then be displayed
FRESH_AIRREMIND
SPAN_CAL
GAS_VALUES LAST_CAL
IQ LINK EXIT
HISTORY
5. Use the navigation arrows to highlight SPAN CAL
and press the MODE button.
Span-Cal
LEL = 0%
6. Use the right and left navigation arrows to scroll
through the available sensors until the sensor that
requires calibration is displayed.
EXIT
Span-Cal
LEL = 0%
→
EXIT
Span-Cal
CO = 0
7. Once the sensor is reached, apply the calibration
gas as shown above in figure 4.4.2. The reading
shown on the display should respond quickly.
date on calibration gas cylinder(s) prior to use.
Always check the expiration
EXIT
Expired calibration gas can lead to inaccurate and
potentially dangerous readings.
Span-Cal
CO = 48
8. Once the readings stabilize, use the up and down
navigation arrows to raise or lower the readings to
match the level of the gas concentration printed on
the calibration cylinder label.
EXIT
Span-Cal
CO = 52
↓↓
EXIT
Span-Cal
CO = 50
9. When the span calibration for a particular sensor is
completed, use the right and left navigation arrows
to advance to the next sensor channel to be span
adjusted.
10. When finished, press the MODE button to exit.
EXIT
Save Ch anges?
YES NO CANCEL
11. Press the MODE button with YES highlighted to
save the span calibration.
12. Remove all gas fittings from the PhD Lite, and
disconnect the regulator from the calibration gas
cylinder.
Note: If multiple cylinders of calibration gas are
necessary to complete the manual span calibration,
each calibration must be completed and
independently saved prior to changing cylinders.
4.6 Functional (bump) test
The accuracy of the PhD Lite may be verified at any
time by a simple functional (bump) test.
To perform a functional (bump) test, do the following:
1. Turn the PhD Lite on and wait at least three
minutes to allow the readings to fully stabilize.
2. Make sure the instrument is located in fresh air.
3. Verify that the current gas readings match the
concentrations present in fresh air. If the PhD Lite
is operating in Basic, Basic/Peak or Technician
operating mode the fresh air readings should equal
20.9 % O
any toxic sensors installed. If the instrument is
operated in the Text Only mode all readings should
indicate that conditions are “OK”. If necessary,
fresh air calibrate the instrument using the
procedures discussed in section 4.5.1 above.
4. Apply the calibration gas as shown above in figure
4.4.2.
5. Wait for the readings to stabilize. (Forty-five
seconds to one minute is usually sufficient.
Reactive gas sensors may take longer.)
, 0 % LEL or 0.0 % CH4, and 0 PPM for
2
42
Page 45
6. Note the readings. Oxygen readings are
considered accurate when the display is within
±0.5%/volume of the expected concentration as
given on the calibration gas cylinder. LEL and toxic
readings are considered accurate when they are
between 90%* and 120% of the expected value as
given on the calibration gas cylinder.
* The Canadian Standards Association (CSA)
requires the instrument to undergo calibration when
the displayed value during a bump test fails to fall
between 100% and 120% of the expected value for
the gas.
Note: If gas concentration readings are off by more
than 10 percent higher or lower than the expected
values during a functional (bump) test, the
instrument must be adjusted using the "span"
calibration procedures discussed in section 4.5.2
before further use.
Always check the expiration
date on calibration gas cylinder(s) prior to use.
Expired calibration gas can lead to inaccurate and
potentially dangerous readings.
4.7 The Calibration Menu
The Calibration Menu is an immediate subdirectory of
the Main Menu and provides access to all calibration
functions and controls.
4.7.1 Entering the Calibration Menu
To enter the Calibration Menu:
1. Turn the instrument on and wait until gas readings
appear.
2. Press the MODE button until the information screen
is displayed.
01:15 15:08
3.9V 71F
3. At the INFORMATION screen hold down the left
navigation arrow for 3 seconds or until EXIT
appears and is highlighted.
MENU
01:15 15:08
3.9V 71F
4. Press the down navigation arrow once to highlight
MENU and press the MODE button. The Main
Menu will then be displayed.
ALARMS CALIBRATION
OPTIONS SCREEN
VIEW TIME
5. Use the navigation arrows to highlight
CALIBRATION and press the MODE button. The
calibration menu will then be displayed.
EXIT
FRESH_AIR SPAN_CAL
GAS_VALUES REMIND
IQ LINK
CAL LOCK
EXIT
EXIT
MENU
4.7.2 Gas values
Calibration gas concentration values may be viewed
and adjusted through the gas values screen.
To change calibration gas values:
1. Enter the Calibration Menu as described above in
section 4.7.1.
FRESH_AIRSPAN_CAL
GAS_VALUES
IQ LINK EXIT
CAL LOCK
REMIND
2. Use the navigation arrows to highlight GAS
VALUES and press the MODE button.
CALIBR ATION GAS
LEL 50.0%
EXIT
Calibration values shown in the
calibration value table must match those appearing
on the calibration gas cylinder(s) that will be used to
calibrate the PhD Lite. Non-matching calibration
gas and calibration gas value settings will lead to
inaccurate and potentially dangerous readings.
3. Use the right and left navigation arrows to scroll
through the available calibration gas settings.
CALIBR ATION GAS
LEL 50.0%
EXIT
→
CALIBR ATION GAS
CO 50.0ppm
4. Once the appropriate sensor is reached, use the up
and down navigation arrows to modify the gas
setting.
CALIBR ATION GAS
CO 50.0ppm
↓
EXIT
EXIT
CALIBR ATION GAS
CO 49.9ppm
5. Once the concentration matches the value listed on
the calibration gas cylinder(s), press the MODE
button.
EXIT
Save Ch anges?
YES NO CANCEL
6. Press the MODE button with YES highlighted to
save the new settings.
Cal Gas Va lues
Saved
4.7.2.1 Changing the combustible gas readout
fr om LEL t o CH
The PhD Lite may be configured to show combustible
gas readings in terms of percent of LEL (Lower
4
43
Page 46
Explosive Limit) or in terms of the percent by volume of
methane (CH
).
4
With the PhD Lite configured to read in terms of percent
by volume of methane (CH
), the LEL sensor must be
4
calibrated to the actual percent by volume of methane
used in Biosystems calibration gas cylinders, not to the
%LEL value given on the label. The actual percentage
by volume of CH
will be stamped in indelible black ink
4
on the side of the cylinder body. For example,
Biosystems popular all-in-one mix, part number 549044E, with 50% LEL propane equivalent will list
±1.62% CH
case, the percent by volume CH
should be set to 1.62%.
on the side of the cylinder body. In this
4
calibration gas value
4
For easy reference, the actual percent by volume of CH4
for the following Biosystems LEL component mixtures is
listed in the following table.
LEL Component
Description
50% LEL Methane
50% LEL Propane
Volume %
Methane ( CH
±2.50
±1.62
)
4
Equivalent
50% LEL Pentane
±1.25
Equivalent
Table 4.7. 2 .1 P erc en t LE L v er su s percent by vo l ume of
methane for comm on Bi osys te ms c alibr ati on ga s
cylinders.
To change the calibration gas value setting from
percentage of LEL to percent by volume of methane or
vice versa:
1. Enter the Calibration Menu as described above in
section 4.7.1.
FRESH_AIRSPAN_CAL
GAS_VALUES
IQ LINK EXIT
CAL LOCK
REMIND
2. Use the navigation arrows to highlight GAS
VALUES and press the MODE button.
Depending on the existing calibration gas value setting,
the instrument will display the calibration value either as
percentage of LEL or as percent by volume of methane
).
(CH
4
CALIBR ATION GAS
LEL 50.0%
EXIT
or
CALIBR ATION GAS
CH4 2.50%
EXIT
3. Press the left navigation arrow once to highlight LEL
.
or CH
4
CALIBR ATION GAS
LEL 50.0%
or
CALIBR ATION GAS
CH4 2.50%
EXIT
EXIT
4. The up and down navigation arrows may be used to
toggle between the calibration gas settings (LEL or
). Once the appropriate setting is shown, press
CH
4
the right navigation arrow once to highlight the
calibration gas concentration in percent of LEL or
percent by volume (CH
). To change the calibration
4
gas concentration setting, use the up and down
navigation arrows to increase or decrease the
amount.
CALIBR ATION GAS
CH4 2.50%
↓
CALIBR ATION GAS
CH4 2.49%
5. Once the appropriate calibration gas type and
concentration are reached, press the MODE button
to confirm the setting.
EXIT
EXIT
Save Ch anges?
YES NO CANCEL
6. Press the MODE button with YES highlighted to
save the new settings.
Cal Gas Va lues
Saved
4.7.2.2 Changing the direct reading setting of the
CO Plus sensor from CO to H
S
2
Biosystems CO Plus sensor is designed for the
simultaneous detection of both carbon monoxide and
hydrogen sulfide, but it can only be calibrated for the
direct detection of one of these hazards.
For instruments with the CO Plus sensor, calibration gas
settings determine whether the PhD Lite is configured
for the direct detection of CO or H
S. When the
2
calibration gas selected is CO, the PhD Lite will show
CO+ on the current gas readings screen. Alternately,
when the calibration gas selected is H
will show H
S+ on the current gas readings screen.
2
S, the PhD Lite
2
O2 LEL CO+
20.9 0 0
Current gas re adings screens with CO Plus sensor;
calibration gas value set to CO.
O2 LEL H2S+
20.9 0 0
Current gas re adings screens with CO Plus sensor;
calibration gas value set to H
To change the direct reading selection of the CO Plus
sensor through the calibration gas value settings:
1. Enter the Calibration Menu as detailed above in
section 4.7.1.
S.
2
44
Page 47
FRESH_AIRSPAN_CAL
GAS_VALUES
IQ LINK EXIT
CAL LOCK
REMIND
2. Use the navigation arrows to highlight
GAS_VALUES and press the MODE button to enter
the gas values screen.
CALIBR ATION GAS
LEL 50.0%
3. If the LEL calibration gas setting screen is shown,
press the right navigation arrow once to access the
calibration gas setting for the CO Plus sensor.
Depending on the existing calibration gas setting for the
CO Plus sensor, the calibration gas value will appear as
either CO or H
S.
2
CALIBR ATION GAS
CO 50.0ppm
or
CALIBR ATION GAS
H2S 25.0pp m
4. The right and left navigation arrows are used to
move back and forth between the type and the
amount of the calibration gas. To change the type
of calibration gas, move the cursor to highlight CO
S.
or H
2
EXIT
EXIT
EXIT
CALIBR ATION GAS
CO 50.0ppm
Or
CALIBR ATION GAS
H2S 25.0pp m
5. The up and down navigation arrows may then be
used to toggle between CO and H
desired gas setting is shown, press the MODE
button to save the setting or move on to step 6 to
adjust the concentration of the calibration gas.
CALIBR ATION GAS
CO 50.0ppm
↑↓
EXIT
EXIT
S. Once the
2
EXIT
CALIBR ATION GAS
H2S 25.0ppm
6. To change the concentration of CO or H2S used in
calibration, use the right navigation arrow to
highlight the concentration in parts-per-million.
Then use the up and down navigation arrows to
adjust the gas concentration.
EXIT
CALIBR ATION GAS
CO 50.0ppm
↑
EXIT
CALIBR ATION GAS
CO 50.1ppm
EXIT
7. Press the MODE button when the appropriate gas
selection and concentration have been reached.
Save Ch anges?
YES NO CANCEL
8. Press the MODE button with YES highlighted to
save the new settings.
Cal Gas Va lues
Saved
calibration gas mixtures containing both carbon
monoxide and hydrogen sulfide when calibrating a
PhD Lite with a CO Plus sensor installed.
Calibration of the CO Plus sensor with multicomponent calibration gas mixtures containing both
CO and H
dangerous readings.
S may lead to inaccurate and potentially
2
Biosystems multi-component calibration gas
mixtures containing both carbon monoxide and
hydrogen sulfide are labeled as “Not for use with
CO Plus sensors”.
calibration gas setting determines whether the
instrument is configured for the direct reading of
CO, or for the direct reading of H
corresponding to the direct reading requirement
must be used in the calibration of the instrument. If
carbon monoxide is chosen in the calibration gas
setti n g option, the displ a y will s how CO+ and
carbon monoxide must be used to verify accuracy.
Similarly, if hydrogen sulfide is chosen in the
calibration gas setting option, the instrument will
display H
S+ and hydrogen sulfide must be used to
2
verify accuracy. Use of the incorrect calibration gas
may lead to inaccurate and potentially dangerous
readings.
4.7.3 Calibration reminder:
The calibration reminder feature reminds the user when
a scheduled calibration is due. The calibration reminder
can be set by the user to an interval between 1 and 180
days and is automatically reset by a successful
calibration.
4.7.3.1 Changing the calibration reminder
To change the calibration reminder:
1. Enter the Calibration Menu as described above in
section 4.7.1.
Do not use multi-component
With the CO Plus sensor, the
S. Calibration gas
2
FRESH_AIR SPAN_CAL
GAS_VALUES
IQ LINK EXIT
CAL LOCK
REMIND
45
Page 48
2. Use the navigation arrows to highlight REMIND and
press the MODE button to confirm the selection.
The calibration reminder interval will then be
shown.
Remind Cal
Every 30 Days
EXIT
3. To change the calibration reminder interval, press
the right navigation arrow once to highlight the
interval in number of days. Once the interval is
highlighted, use the up and down navigation arrows
to increase or decrease the interval.
Remind Cal
Every 30 Days
EXIT
↓
Remind Cal
Every 29 Days
4. Once the desired calibration reminder interval has
been reached, press the MODE button.
EXIT
Save Ch anges?
YES NO CANCEL
5. Press the MODE button with YES highlighted to
save the changes to the calibration reminder
interval.
Once the calibration reminder has been set, the
instrum ent will display the number of days until the next
required calibration during every subsequent instrument
start up.
Note: The calibration reminder interval has been
factory set to 30 days. If the calibration is past due
when the PhD Lite is turned on, the instrument will
display the message “Calibration Due” and the
alarms will be activated. To bypass the calibration
due alarm, press the MODE button.
4.7.3.2 Disabling the calibration reminder
To disable the calibration reminder, follow the directions
given in section 4.7.3.1 and decrease the interval to
zero days.
Remind Cal
Every day
EXIT
↓
Remind Cal
Never
4.7.4 Calibration history
The PhD Lite automatically remembers the latest
successful calibration dates for all sensors currently
recognized by the unit. To view the calibration history,
do the following:
1. Enter the Calibration Menu as described in above in
section 4.7.1.
EXIT
FRESH_AIR SPAN_CAL
GAS_VALUES
IQ LINK EXIT
CAL LOCK
REMIND
2. Use the navigation arrows to highlight HISTORY
and press the MODE button. The following two
screens will then be shown in succession:
←→
↓↑
Next Gas?
Cal. Dates
O2 ZERO CAL.
DATES
3. Use the right and left navigation arrows to scroll
through the information for the individual sensors.
O2 ZERO CAL.
DATES
→
LEL ZERO CAL .
DATES
→
LEL SPAN CAL .
DATES
→
CO ZERO CAL.
DATES
→
CO SPAN CAL.
DATES
4. Once the appropriate sensor is reached, use the
down navigation arrow to see the most recent span
or zero calibration for the given sensor.
LEL SPAN CAL .
DATES
↓
Date 08 OC T 2003
Time 09:41
EXIT
5. The down navigation arrow may be pressed again
to access the next most recent calibration record.
Date 08 OC T 2003
Time 09:41
↓
Date 02 OC T 2003
Time 13:13
6. Press the MODE button at any time to return to the
main menu.
EXIT
EXIT
46
Page 49
4.7.5 Last calibration
The details of both the last fresh air/zero and the last
span calibration can be viewed by accessing the
LAST_CAL option from the Calibration Menu.
To view the details of the last calibration:
1. Enter the Calibration Menu as described above in
section 4.7.1.
FRESH_AIR REMIND
SPAN_CAL HISTORY
GAS_VALUES
IQ LINK EXIT
LAST_CAL
2. Use the navigation arrows to highlight LAST_CAL
and press the MODE button to confirm the
selection. The details of the last calibration will
then be shown. The right and left navigation arrows
are then used to scroll through the last calibration
details for the various sensors.
O2 LAST CAL
ZERO 11OCT01
→
LEL LAST CAL
ZERO 11OCT01
SPAN 11OCT01 TO 50
MAX ADJUST 92 EXIT
4.7.6 IQ Link
The IQ Datalink System is a data management system
that downloads instrument readings, calibration
information and other data from Biosystems gas
detectors and automatically stores them in a userspecified database. With the IQ Link function enabled
the PhD Lite will automatically initiate IrDA
communications following a calibration. The instrument
is then placed in front of the PC’s IrDA port and the data
is downloaded into the IQ Database.
1. Enter the Calibration Menu as described above in
section 4.7.1.
FRESH_AIR REMIND
SPAN_CAL HISTORY
GAS_VALUES LAST_CAL
IQ LINK
EXIT
2. Use the navigation arrows to highlight IQ LINK and
press the MODE button to confirm the selection.
The IQ Link setting will then be shown. The up and
down navigation keys are used to change the
enabled/disabled setting.
CONNECT TO IQ LINK
AFTER AUTOCAL
ENABLE D
EXIT
↓↑
CONNECT TO IQ LINK
AFTER AUTOCAL
DISABL ED
EXIT
3. One the appropriate setting is shown, press the
MODE button to exit.
47
Page 50
Chapter 5. Record Keeping
5.1 Overview of record keeping options
Any version of the PhD Lite may be equipped with an
optional datalogger. Starting with instrument firmware
version 1.29, every PhD Lite not ordered as a
datalogger has been equipped with a “black box” data
recorder at no additional cost. Both the black box data
recorder and the datalogger store similar information
such as gas readings, turn-on times, turn-off times and
battery conditions.
Data in instruments with a fully enabled datalogger can
be downloaded to an IBM-compatible PC using Biotrak
software or the IQ System. Instruments with the black
box data recorder must be returned to Biosystems for
data retrieval. The datalogger also offers numerous
user-configurable options that are not available with the
black box data recorder.
Instruments equipped with the datalogger option may
also be equipped with Biosystems Touch ID interface for
easier entry of user and location ID information into the
session memory. Touch ID uses small, virtually
indestructible memory buttons to store user ID and
location ID information. The user simply touches the
Touch ID button to the ID reader on the PhD Lite to
enter a user or location setting into the session memory.
5.2 Black box data recorder
PhD Lite instruments that are ordered without the
datalogging option are equipped with a “black box” data
recorder at no charge. The “black box” is continually in
operation whether the user is aware of it or not. The
black box stores important information such as gas
readings, turn-on times, turn-off times, temperatures,
battery conditions, the 8 most recent calibration dates
and settings, types of sensors currently installed, sensor
serial numbers, warranty expiration and service due
dates, and current alarm settings.
There is a finite amount of memory storage available in
the black box data recorder. Once the memory is “full”,
the PhD Lite will begin to write the new data over the
oldest data. With a typical four-gas configuration, the
black box data recorder will store approximately 48
hours of data. Since the oldest data is overwritten first,
newer data is always conserved.
If the data in the black box is required for any reason,
the entire PhD Lite instrument must be returned to
Biosystems, where the data will be extracted and a
report will be generated. The unit and the report will
then be returned to the user. Simply call Biosystems at
(800) 711 6776 to obtain a return authorization number.
There is no charge for the downloading service, but the
user is responsible for any freight charges incurred.
Note: The PhD Lite black box data recorder can be
upgraded to a full user-configurable datalogger at
any time. Call Biosystems Technical Service
Department at 800 711 6776 x-509 for details.
5.3 Datalogger upgrade
The “full” datalogger option is available on the PhD Lite
for users who need imm ediate access to the data stored
in the instrument, or who require the ability to customize
their record-keeping process.
Datalogging is a "transparent" function that is continually
in operation. As long as the datalogger has not been
disabled, it is not necessary to do anything special to
begin logging data. Simply turning on the instrument
causes it to begin recording data. The information
stored by the datalogger may be downloaded to a
Windows-compatible PC to create a permanent record,
or directly displayed by the PhD Lite.
Upon turn on, PhD Lite instruments with an enabled
datalogger will display the sampling interval among the
self-test screens as shown below. In this case, the
default sampling interval of 1 minute is shown.
Sampling Interval
01m00s 48h41m
With the datalogging option, the PhD Lite will
automatically record gas readings, turn-on / turn-off
times, temperatures, battery conditions, the 8 most
recent calibration dates and settings, types of sensors
currently installed, sensor serial numbers, warranty
expiration and service due dates, temperature
compensation curves, and current alarm settings. The
PhD Lite automatically updates all of this information
whenev er the instrument is turned on, whenever a
change is made during operation, and again as the
instrum ent is turned off.
The datalogger in the PhD Lite can store the exposure
values for approximately 2900 datalogging intervals
when using a four channel configuration (O
S). This provides for storage of approximately 48
and H
2
hours of four-gas monitoring when a one-minute
datalogging interval is specified. Using a longer
datalogging interval increases the length of monitoring
time before the oldest data is overwritten by new data.
For instance, if a datalogging interval of two minutes is
selected, a little over 97 hours of monitoring data will be
stored before the oldest data is overwritten. Different
sensor configurations and/or datalogging interval
settings allow for more or less datalogging time before
the oldest data is overwritten.
5.3.1 BioTrak database software
Biotrak software is designed to facilitate both the
downloading of stored data and the uploading of new
instrument configurations. BioTrak allows information to
be retrieved from the PhD Lite and also allows the PhD
Lite to be programmed directly from a Windowscompatible PC.
It is important to note that it is not necessary to use the
BioTrak software to program the PhD Lite.
Configuration and setup options may be programmed
directly by using the navigation arrows on the instrument
keypad.
, LEL, CO,
2
48
Page 51
Manual programming procedures are covered in
detail be lo w in sec tion 5.4.
BioTrak software has been designed to simplify
programming, downloading, and data analysis. The
software allows optional instrument setups to be created
by "filling out" forms right on the computer screen.
Note: The PhD Lite is designed to download data
from the instrument to a PC using an IrDA compliant
infrared data transceiver. It may be necessary to
purchase an IrDA transceiver if your computer is
not equipped with this feature. Most new laptopstyle PCs are equipped with an IrDA port, but many
desktop and older laptop-style PC’s are not.
Consult your computer’s reference manual for
details.
Once information has been downloaded to the
computer, it may be used for a variety of purposes.
Data may be displayed and reviewed in detail through
the computer monitor screen, or used to generate and
print reports, tables and graphs of time history exposure
data. It is also possible to export records to other
software applications in the form of ASCII text or in a
spreadsheet format. Another option is to simply retain
downloaded records within the BioTrak program for
future use as needed.
Note: The remaining material in this chapter is
designed to acquaint the user with datalogging
features that can be programmed through the PhD
Lite instrument itself. Consult the BioTrak
Reference Manual for complete instructions on the
use of Bio Trak software.
5.3.2 IQ System
Biosystems Instrument Quality (IQ) System offers all of
the data tracking features of BioTrak, but also includes
many features that were not possible when BioTrak was
originally released. The IQ Controller can be used with
IQ Docking Stations to provide hands-free calibration of
up to 10 Biosystems gas detectors. The IQ System is
also e-mail capable and can automatically notify you
when an instrument fails calibration.
Call Biosystems for more information on the IQ
Systems, or visit the IQ System website at
http://www.bioIQsystems.com
5.4 Datalogger menu
.
It is possible to customize the way the PhD Lite records
data in a number of different ways. Options include
extended recording time, tagging the exposure data with
time and date information, or assigning location and ID
information.
PhD Lite instruments equipped with datalogging
capability will have an additional subdirectory in the
Main Menu called DATALOGGER. The Datalogger
Menu provides access to all user-configurable
datalogging options. To enter the Datalogger Menu:
1. Turn the instrument on and wait until the gas
readings screen appears.
2. Press the MODE button until you reach the
Information screen (shown below).
01:15 15:08
3.9V 71F
3. Hold down the left navigation arrow for 3 seconds
or until EXIT appears and is highlighted.
MENU
01:15 15:08
3.9V 71F
4. Press the down navigation arrow once to highlight
MENU and press the MODE button. The Main
Menu will then be displayed and EXIT will be
highlighted.
5. Use the navigation arrows to highlight
DATALOGGER and press the MODE button.
The datalogger menu will then be displayed.
5.4.1 Setting the datalogging interval
The datalogger samples continuously, so the data
stream must be broken into intervals to be recorded.
The datalogging interval defines the frequency of the
breaks in the data stream. The interval may be set
anywhere between one second and one hour by using
the navigation arrows as detailed below. The default
datalogging interval is 1 minute.
There is a finite amount of memory storage available in
the PhD Lite. Once the memory is “full”, the PhD Lite
will begin to write the new data over the oldest data. In
this way, the newest data is always conserved.
Different sensor configurations and/or datalogging
intervals may increase or decrease the length of time
before old data is overwritten. A longer sampling
interval will allow the retention of more hours of data
before old data is overwritten making the PhD Lite ideal
for long-term sampling projects.
The datalogger in the PhD Lite can store the exposure
values for approximately 2900 datalogging intervals
when using a four channel configuration of O
and H
2
hours of four-gas monitoring when a one-minute
datalogging interval is specified. Using a longer
datalogging interval increases the length of monitoring
time before the oldest data is overwritten by new data.
For instance, if a datalogging interval of two minutes is
selected, a little over 97 hours of monitoring data will be
stored before the oldest data is overwritten.
ALARMS CALIBRATION
OPTIONS SCREEN
VIEW DATALOGGER
TIME
ALARMS CALIBRATION
OPTIONS SCREEN
VIEW
TIME EXIT
INTERVAL CLEAR_DL
SESSIONS COMMN
ID_INFO
EXIT
DATALOGGER
EXIT
S. This provides for storage of approximately 48
EXIT
MENU
, LEL, CO,
2
49
Page 52
Note: Calculations that are made on an ongoing
basis (i.e. TWA, STEL, Ceilings, and Peak exposure
values) are updated at regular intervals by the PhD
Lite microprocessor. Adjustments to the
datalogging interval do not effect the way in which
TWA, STEL, Ceiling, and Peak exposure values are
calculated.
To adjust the sampling interval:
1. Enter the Datalogger Menu as described above in
section 5.4.
INTERVAL
SESSIONS COMMN
ID_INFO EXIT
CLEAR_DL
2. Use the navigation arrows to highlight INTERVAL
and press the MODE button.
SAMPLING INTERVAL
On 01:00
3. Use the right and left navigation arrows to highlight
the time interval. The interval is given in terms of
minutes and seconds.
48:41
EXIT
SAMPLING INTERVAL
On 01:00
As discussed above, there is a finite amount of memory
storage in the PhD Lite. As the sampling interval is
either lengthened or shortened, so is the time-estimated
figure, which is given in hours and minutes and appears
above the word EXIT on the right side of the screen.
This figure represents the length of time before the
oldest data will be overwritten.
4. With the interval highlighted, use the up and down
navigation arrows to adjust the interval
48:41
EXIT
SAMPLING INTERVAL
On 01:00
↑
48:41
EXIT
SAMPLING INTERVAL
On 02:00
5. Once the sampling interval is adjusted, use the
navigation arrows to highlight exit and press the
MODE button. The following two screens will be
shown in succession.
97:22
EXIT
Longes t Session
Estimate 97h22m
Save Chan ges?
YES NO CANCEL
6. Press the MODE button with YES highlighted to
save the changes.
Sampling interval
Saved
5.4.1.1 Enable/disable datalogger
To enable or disable the datalogger:
1. Enter the Datalogger Menu as described above in
section 5.4.
INTERVAL
SESSIONS COMMN
ID_INFO EXIT
CLEAR_DL
2. Use the navigation arrows to highlight INTERVAL
and press the MODE button. With On or Off
highlighted, the up and down navigational arrows
are used to toggle between ON and OFF.
SAMPLING INTERVAL
On 01:0 0
↑↓
48:41
EXIT
SAMPLING INTERVAL
Off
3. Once the appropriate setting is shown use the
navigation arrows to highlight exit and press the
MODE button to enter the new interval. The
following screen will be shown.
EXIT
Save Changes?
YES NO CANCEL
4. Press MODE with YES highlighted to save the new
sampling interval setting.
Sampling interval
Saved
5.4.2 Clearing the datalogger
The datalogger can be cleared of all information through
the CLEAR DL option.
To clear the datalogger, do the following:
1. Enter the Datalogger Menu as described above in
section 5.4.
INTERVAL
SESSIONS COMMN
ID_INFO EXIT
CLEAR_DL
2. Use the navigation arrows to highlight CLEAR DL
and press the MODE button. The following two
screens will then be shown in succession.
Warnin g Lose
All Sess ions Data
Clear Da talogger ?
YES NO
3. Press the MODE button with YES highlighted to
clear the datalogger memory.
DATALOGGER CLEAR
COMPLETED
50
Page 53
5.4.3 Sessions
Data recorded for all monitoring sessions currently in
the datalogger memory can be accessed through the
SESSIONS subdirectory of the Datalogging Menu. To
view the session memory:
1. Enter the Datalogging Menu as described above in
section 5.4.
INTERVALCLEAR_DL
SESSIONS
ID_INFO EXIT
COMMN
2. Use the navigation arrows to highlight SESSIONS
and press the MODE button. The information
gathered from the most recent operating session
will be displayed.
#1 08:14- 16:45
18 OCT 20 03
3. The right and left navigation arrows are used to
scroll through the data from individual sessions.
#1 8:14-16:45
18 OCT 20 03
→
#2 8:02-17:23
13 OCT 20 03
4. The up and down navigation arrows are used to
access specific data from the individual monitoring
session that is displayed.
#2 8:02-17:23
13 OCT 20 03
↓
PEAK LOW HIGH
O2 20.5 20.9
↓
PEAK LEL CO H2S
0 0 0
5. Press the MODE button at any time to return to the
Main Menu.
5.4.4 Communications mode
Communications mode enables the PhD Lite IrDA port
for downloading to or uploading from a personal
computer. To enter communications mode, do the
following:
1. Enter the Datalogger Menu as described above in
section 5.3.2.
2. Use the navigation arrows to highlight COMM and
press the MODE button.
INTERVAL CLEAR_DL
SESSIONS
ID_INFO EXIT
COMM
TO COMM MODE
YES NO
3. Press the MODE button with YES highlighted to
enter communication mode and enable the IrDA
port.
Note: If an IrDA link with a PC is not established
within 30 seconds of the initiation of
Communic atio n s Mode, the PhD Lite will shut itself
off.
5.4.5 ID information
The ID_INFO subdirectory of the Datalogger Menu
controls manual access to the location and user identity
functions. To enter the ID_INFO menu:
1. Enter the Datalogger Menu as described above in
section 5.4.
2. Use the navigation arrows to highlight ID_INFO and
press the MODE button.
5.4.5.1 User and location lists
The PhD Lite is capable of storing up to 15 different
user ID’s and 15 different location ID’s. User and
location lists may be manually entered into the PhD
Lite’s memory through the USER_LIST and LOC_LIST
options in the ID Information Menu. The list functions
control the entry of user and location names into the
datalogger memory. To access the existing list, or to
start a list:
1. Enter the ID_INFO option from the Datalogger
Menu as described above in section 5.4.5. Use the
navigation arrows to highlight USER_LIST or
LOC_LIST.
2. Press the MODE button. User #1 or Location #1
will then be shown.
3. Press the MODE button with NEXT highlighted to
scroll through the list of users or locations.
INTERVAL CLEAR_DL
SESSIONS COMM
ID_INFO
USER_LIST USER_ID
LOC_LIST LOC_ID
TOUCH_ID
USER_LIST
LOC_LIST LOC_ID
TOUCH_ID EXIT
USER_LIST USER_ID
LOC_LIST
TOUCH_ID EXIT
<BILL SAWKA >
User #1
<651 S Main St >
Location#1
EXIT
EXIT
USER_ID
LOC_ID
NEXT
EXIT
NEXT EXIT
51
Page 54
<BILL SAWKA >
User #1
<LAR S BOETTERN >
User #2
NEXT
EXIT
NEXT
EXIT
3. To enter a new name or location to the datalogger
memory, scroll through the names until the first
blank screen is reached.
< >
User #3
NEXT
EXIT
5. Press the right arrow key once to move the cursor
into the text line.
< >
User #3
NEXT
EXIT
6. Press the MODE button once. The select character
screen will then be shown.
S E L E C T
A B C D E F G H I J K L M N O P Q R S
T U V W X Y Z 1 2 3 4 5 6 7 8 9 0 ’ , . /
C H A R A C T E
R
7. Use the right and left navigation arrows to highlight
the character required.
8. Press the MODE button to enter the character.
S E L E C T
A B C D E F G H I J K L M N O P Q R S
T U V W X Y Z 1 2 3 4 5 6 7 8 9 0 ’ , . /
<P >
C H A R A C T E
User #3
R
NEXT
EXIT
9. Press the right navigation arrow once to move the
cursor to the next space in the text line.
<P >
User #3
NEXT
EXIT
10. Press the MODE button to return to the select
character screen and repeat steps 7 though 9 until
the new name or location appears in the text line.
<PAUL G >
User #3
NEXT
EXIT
11. To enter another name, press the right navigation
arrow until NEXT is highlighted and press the
MODE button. The next memory location will then
be shown. Repeat steps 6 through 10 to enter the
next name or location.
< >
User #4
NEXT
EXIT
12 To exit, use the right or left navigation arrows to
highlight NEXT and then press the down navigation
arrow once to highlight EXIT.
< >
User #4
NEXT
EXIT
13 Press the MODE button with EXIT highlighted to
leave the user or location list. The “save changes”
screen will then be displayed.
Save Chan ges?
YES NO CANCEL
14. Press the MODE button with YES highlighted to
confirm the changes.
Users Li st
Saved
Note: Names and locations may also be added
using a PC equipped with BioTrak software and an
IrDA port.
Note: The user and location ID is not retained if the
PhD Lite is turned off. The user and location ID
must be re-entered or selected from the list prior to
any new operating session.
5.4.5.2 User and location ID’s
User and location names may be manually entered into
the current session memory by selecting the name or
location ID while in the USER_ID or LOC_ID option of
the datalogger menu.
Note: If th e User or Locat io n ID is e n tered or
changed after the first two minutes of any session, a
new session is automatically created.
1. Enter the ID_INFO option from the Datalogger
Menu as described above in section 5.4.5. Use the
navigation arrows to highlight USER_ID or LOC_ID.
2. Press the MODE button. One of the following
screens will then be shown:
3. User and location ID’s can be manually entered by
pressing the right arrow key once to highlight the
first space in the location memory.
4. To manually enter the ID, follow the instructions
above in section 5.4.5.1 starting at instruction 6.
USER_LIST USER_ID
LOC_LIST LOC_ID
TOUCH_ID EXIT
USER_LIST USER_ID
LOC_LIST LOC_ID
TOUCH_ID EXIT
< >
User ID
< >
LOCATION:
< >
User ID:
LIST
EXIT
LIST
EXIT
LIST
EXIT
52
Page 55
5. To select a user or location ID from the pre-entered
list, press the MODE button with LIST highlighted.
<BILL SAWKA >
User #1
ENTER
EXIT
6. The right and left navigation arrows are then used
to scroll through the user or location lists.
<LAR S BOETTERN >
User #2
ENTER
EXIT
7. Once the correct user or location ID is displayed,
press the MODE button with ENTER highlighted.
<LAR S BOETTERN >
User #2
ENTER
EXIT
8. Press the down navigation arrow once to highlight
EXIT and press the MODE button.
Save Chan ges?
YES NO CANCEL
9. Press the MODE button with YES highlighted to
confirm the changes.
Curren t User ID
Saved
Note: The user and location ID is not retained if the
PhD Lite is turned off. The user and location ID
must be re-entered or selected from the list prior to
any new operating session.
5.4.5.3 Touch ID option
PhD Lite detectors with the “full” datalogging option may
also be equipped with Biosystems unique Touch ID
automatic memory download system. The system uses
small, virtually indestructible memory buttons to store
user ID and location ID information. User and Location
ID’s can be set by simply pressing a pre-programmed
“Touch Memory Button” to the “Touch Memory Module”
on the back of PhD Lite while the instrument is in normal
gas detection mode.
USER IS NOW
BILL SAWKA
Note: If th e User or Locat io n ID is e n tered or
changed after the first two minutes of any session, a
new session is automatically created.
5.4.5.3.1 Touch ID menu
In PhD Lite instruments with Touch ID, a TOUCH ID
option will be provided in the Datalogger Menu.
Touch ID buttons contain either a name or a location
and may be programmed with the instrument through
the Touch ID function in the Datalogger Menu.
To program a Touch ID button:
1. Enter a name or location into the session memory
as detailed in section 5.4.5.1.
2. Enter the Datalogger Menu as described above in
section 5.4.
USER_LIST USER_ID
LOC_LIST LOC_ID
TOUCH_ID
EXIT
3. Use the navigation arrows to highlight TOUCH ID
and press the MODE button.
5.4.5.3.1.1 Location memory
LOCATION_MEMORY
USER_ID_MEMORY
CLEAR_MEMORY
EXIT
To program a location from the instrument onto a Touch
ID button:
1. Set the location in the current session to the location
that will be written onto the Touch ID button.
Locations may be manually input into the PhD Lite
with the directions given in section 5.4.5.1.
2. Enter the Touch ID Menu as described above in
section 5.4.5.3.1.
LOCATION_MEMORY
USER_ID_MEMORY
CLEAR_MEMORY
EXIT
3. Use the up navigation arrow to highlight LOCATION
MEMORY and press the MODE button.
INSERT ID BUTTON
PLEASE
4. Press the Touch ID button to the Touch ID reader
on the back of the PhD Lite.
EXIT
TOUCH ID FOUND
WRITIN G COMPLETE D
5. The instrument will program the current location
setting onto the Touch ID button and return to the
Main Menu upon completion.
5.4.5.3.1.2 User ID memory
The User ID memory includes the individual’s name,
language preference and operating mode preference.
To program this data from the instrument onto a Touch
ID button:
1. Set the user ID in the current session to the ID that
should be written onto the Touch ID button. User
ID’s may be manually input into the PhD Lite with
the directions given in section 5.4.5.1.
2. Enter the Touch ID Menu as described above in
section 5.4.5.5.
LOCATION_MEMORY
USER_ID_MEMORY
CLEAR_MEMORY
EXIT
3. Use the up navigation arrow to highlight USER ID
MEMORY and press the MODE button.
INSERT ID BUTTON
PLEASE
EXIT
53
Page 56
4. Press the Touch ID button to the Touch ID reader
on the front of the PhD Lite.
TOUCH ID FOUND
WRITIN G COMPLETE D
5. The instrument will program the current user ID
setting onto the Touch ID button and return to the
Main Menu upon completion
5.4.5.3.1.3 Clear ID button
To clear a Touch ID button:
1. Enter the Touch ID Menu as described above in
section 5.4.1.
LOCATION_MEMORY
USER_ID_MEMORY
CLEAR_MEMORY
EXIT
2. Use the up navigation arrow to highlight CLEAR ID
BUTTON and press the MODE button.
INSERT ID BUTTON
PLEASE
3. Press the Touch ID button to the Touch ID reader
on the front of the PhD Lite.
EXIT
TOUCH ID FOUND
AND DELE TED
The instrument will delete the current data contained on
the Touch ID button and then return to the Main Menu
upon completion.
54
Page 57
Chapter 6. Basic maintenance
CAUTION: Maintenance of the PhD Lite should only
be performed by appropriately trained personnel.
6.1 Sensors
Figure 6.0 Ph D Lite ex t er n al f e atu r es
At instrument start up, the PhD Lite is designed to
recognize the “Smart Sensors” that are installed and
automatically set the appropriate alarm and display
readings. The PhD Lite automatically recognizes when
sensor changes have been made since the instrument
was last turned on.
Not e: Any changes made to t he sensors install ed,
even changing one sensor for another of the exact
same type, will trigger a “Needs Cal” message the
next time the instrument is turned on. The PhD Lite
must be calibrated before being returned to service
following any sensor changes.
Caution: The PhD Lite must be turned off prior to
removing or replacing sensors.
6.1.1 Sensor replacement
Note: The PhD Lite must be turned off to replace
sensors.
1. Remove the three philips screws and take off the
sensor cover from the front of the instrument.
For replacement of an existing sensor perform
steps 2a and 3a then proceed to step 4a or 4b.
2a. From the outer surface of the sensor cover, use a
flat blade screwdriver to gently push out the metal
screen (if present), gasket & filter ring assembly.
Discard old gasket, filter ring assembly & metal
screen. The metal screen is not to be reused and
its absence will not affect sensor performance.
3a. Remove any remaining traces of adhesive from the
recessed hole in the sensor cover.
For new sensor installation perform steps 2b and 3b
then proceed to step 4a or 4b.
2b. From the outer surface of the sensor cover, push
out yellow dust cap with a blunt tool.
3b. Remove sensor blank from the sensor
compartm ent.
For Sensors O
, LEL, CO, CO Plus, CO-H, H2S, NO,
2
DUO-TOX perform step 4a then proceed to step 5.
4a. Firmly press the new filter ring assembly into the
recessed hole in the sensor cover with ridge side
up.
For Reactive Gas Sensors: SO
perform step 4b then proceed to step 5.
ClO
2
, NO2, PH3, HCN, Cl2,
2
4b. Firmly press the new teflon spacer into the
recessed hole in the sensor cover. For optimal
sensor response, no external filter element is used
with these sensors.
CAUTION: Since the reactive gas sensor face
is openly exposed to the ambient environment,
extra care should be taken to prevent water or any
other liquids from coming into contact and
remaining on the sensor face.
5. Peel the backing paper from the new rubber gasket
and center it over the newly mounted filter ring
assembly with the adhesive side down.
6. Remount sensor cover and secure with three
philips screws. DO NOT overtighten.
7. Install functional battery pack or batteries in the
detector.
8. New sensors must be allowed a stabilizing time –
with detector powered off and with functional
batteries or battery pack installed, according to the
following schedule.
Sensor Stabilization Period
Oxygen (54-25-90) 1 hour
LEL (54-25-80A) 5 minutes
All Toxic sensors except
those shown below
54-25-04 NH3 Sensor
54-25-06 NO Sensor
9. The PhD Lite will automatically recognize the
changes that have been made upon turn on and
display the “Warning Needs Cal” message.
10. Calibrate the PhD Lite with calibration gas
appropriate for the new sensor before the
instrument is put back into service.
11. If newly installed sensor(s) do not appear during
power-up, or if the message "not supported"
15 minutes
24 hours
55
Page 58
appears, contact Biosystems Instrument Service
Dept. for further guidance.
PhD Lite programming includes safeguards to recognize
maladjusted sensors. If the settings on the new sensor
are significantly different from those of the sensor that is
being replaced, a message indicating that the sensor is
reading “Too Low” or “Too High” for Auto-Calibration
may follow the first attempt at one-button calibration.
Note: The first fresh air calibration adjustment after
installation of a new sensor should be done using
the “manual” calibration procedure as discussed in
section 4.5 of this manual
Once the new sensor has been fresh-air calibrated
using the “manual” calibration procedure, subsequent
fresh air and span calibrations can be made by using
the MODE button and One-Button Auto-Calibration
procedures.
6.1.2 New sensor releases
Biosystems occasionally releases a new type of toxic
sensor, or makes changes to existing sensors in order
to improve performance. In some cases it may be
necessary to update the PhD Lite’s internal software
before making use of the new sensor. If a sensor that is
incompatible with the current configuration of the
instrument’s internal software is installed, a message
stating that the sensor is “Not Supported” will be
displayed at the time the instrument is turned on.
For instructions on updating the internal instrument
software in your PhD Lite, see section 6.4 below.
6.1.3 Troubleshooting sensor problems
Many difficulties with sensors can be resolved easily.
Below are a few of the most common sensor-related
problems.
6.1.3.1 Can’t make automatic fresh air/zero
calibration adjustment
Possible causes and solutions
1. The atmosphere in which the instrument is located
is contaminated (or was contaminated at the time
the instrument was last fresh air calibrated.
2. A new sensor has just been installed.
3. Instrument has been dropped or banged since last
turned on.
4. There has been a significant change in temperature
or humidity since the instrum ent was last used.
Solution:
Take the instrument to fresh air and allow readings to
stabilize. Perform a manual fresh air/zero adjustment
as discussed in section 4.5.1. If the manual fresh
air/zero procedure fails to correct the problem, call
Biosystems Instrument Service Department at (860)
344-1079 or (800) 711-6776.
6.1.3.2 Can’t make span calibration adjustment.
Causes:
1. Instrument failed the last fresh air / zero calibration.
2. Empty calibration gas cylinder.
3. Expired calibration gas cylinder.
4. Calibration gas setting does not correspond to
calibration gas.
5. Sensor has drifted outside of the range for
automatic calibration.
6. LEL only: Type of calibration gas (standard) has
changed significantly.
7. Dead sensor.
8. Instrument problem.
Solutions:
1. Verify that the instrument can be fresh air / zero
calibrated by performing the procedure discussed
above in section 6.1.3.1.
2. Verify that the calibration gas is not empty.
3. Verify that the calibration gas is not expired.
4. Verify that the calibration gas is the correct type for
the sensor and that the calibration gas settings are
identical to those given on the cylinders. See
section 4.7.2. for instructions related to calibration
gas settings.
5. Span calibrate the instrument manually by following
the instructions given in section 4.5.2.
6. If the calibration gas standard has changed
significantly, span calibrate the instrument manually
be following the instructions given in section 4.5.2.
7. Replace sensor.
8. Send instrument back to Biosystems.
6.2 Motorized pump
A motorized sample pump is available for the PhD Lite
for remote sampling. The motorized pump is powered
by the PhD Lite battery. When the pump is attached
56
Page 59
and functioning normally, an animated “” icon will
appear in the upper left-hand corner of the LCD display.
O2 LEL CO H2S
20.9 0 0 0
CAUTION: Never perform remote sampling with
the PhD Lite without the sample probe assembly.
The sample probe handle contains replaceable
filters designed to block moisture and remove
particulate contaminants. If the pump is operated
without the probe assembly in place, particulate
contaminants may be drawn into the instrument
causing damage to the pump, sensors and internal
components of the PhD Lite.
The motorized sample pump includes a pressure sensor
designed to protect the PhD Lite from exposure to water
or other liquids. If there is a sufficient change in
pressure in the sample draw assembly due to fluid
intake or other blockage, the pump immediately shuts
down. After a few seconds audible and visible alarms
indicating a low flow condition will also be activated.
Pump Fault
Pump Lo w Flow
Remov e Blockage
Pump Lo w Flow
MODE=Acknowledge
CAUTION: Make sure that the sample draw tube
is not inserted into a fluid horizontally or at a low
angle.
The pressure sensor in the motorized pump is designed
to detect changes while the sample-draw probe is being
held in a vertical position. If the probe is held
horizontally or at a low angle while immersed in a fluid,
pressure sufficient to cause the pump to shut down may
not be generated, and water may be drawn into the
pump assembly causing damage to the pump, sensors
and internal components of the PhD Lite.
To avoid potential damage, care must be taken to keep
the probe vertical whenever fluids may be present.
As an additional safeguard, the pump contains an
internally-housed particulate filter. If the pump is
operated without the sample probe assembly in place,
or in a particularly dirty atmosphere, this internal filter
may become clogged and require replacement.
Standard accessories with every 54-26-0101 motorized
pump include a package of 10 replacement filters for the
pump (Biosystems part number 61-001).
6.2.1 Pump performance
The PhD Lite automatically verifies pump performance
upon start-up of the instrument, or whenever the pump
is attached during normal operation.
6.2.1.1 Verifying pump performance
Pump performance is automatically tested when the
pump is attached to the instrument. A similar test is
automatically initiated if the pump is attached prior to
turn-on.
1. Attach the pump to the PhD Lite. Make sure that
the sample probe and tubing assembly is attached
to the pump inlet.
Pump De tected
2. The PhD Lite will automatically prompt the user to
test the pump system.
Pump Testing
Block Sa mple
3. Verify that there are no leaks in the system by
covering the end of the sample probe tube with a
finger. If there are no leaks a message will be
displayed indicating that the pump test passed.
Pump Test Passed
Remov e Blockage
4. Press the MODE button to reset the pump. The
following screen will then be displayed. Notice the
animated “
” icon in the upper left corner of the LCD
display, which indicates that the pump is attached
and in normal operation.
O2 LEL CO H2S
20.9 0 0 0
6.2.1.2 Pump test failed
The pump includes a pressure sensor designed to
measure pump flow. If there is insufficient pump flow,
the pump test will fail and the PhD Lite will display the
following message.
Pump Test Failed
Remove Pump
Remove the pump. Since pump failure results from
insufficient pump flow, there is either a leak in the
sample draw system or the pump itself is not operating
properly.
To locate or rule out a leak in the sample draw system:
1. Remove the sample draw probe and tubing from
the pump.
2. Reattach the pump to the instrument without the
sample probe and tubing assembly. The
instrum ent will again proceed with the pump test.
3. Block the pump inlet with a finger. If the pump test
passes, then the leak is located in the probe and
tubing assembly. If the pump test fails, then the
failure is inside of the pump itself.
57
Page 60
If the leak is located in the sample probe and tubing
assembly, inspect it and look for the leak. See section
6.3 for instructions on the replacement of key probe
parts.
If the leak is located in the pump itself, the pump may be
disassembled. See section 6.2.2 for instructions on how
to open the pump assembly.
6.2.1.3 Low flow pump failure
If the instrument detects a pump fault during normal
operation, it will automatically indicate that a pump fault
exists on the display and the audible and visible alarms
will be activated.
Pump Fault
The instrument will then instruct the user to remove the
blockage.
Pump Lo w Flow
Remov e Blockage
Pump Lo w Flow
MODE=Acknowledge
In most cases, the cause of the blockage is easily
recognized. Remove the blockage, reattach the hose
and probe assembly and press the MODE button to
reset the pump. Once the pump starts up, retest the
pump seals by placing a finger over the end of the
sample probe.
6.2.2 Internal pum p f ilt er rep lacem ent
The following procedure describes how to replace the
internal pump filter:
Figure 6.2 . 2 Inter n al PhD Lite pum p di agram
1. Remove the four bottom screws from the pump
assembly and disassemble the upper and lower
sections of the pump case.
2. Rem ove the two pieces of tubing from the filter cap
by gently twisting and pulling until the ends are free.
3. Remove the filter cap and replace the used 61-001
filter.
4. Reattach the filter cap and tubing. Make sure the
tubing is secure to the filter cap before
reassembling the upper and lower pump cases.
5. Reassemble the upper and lower sections of the
pump case. Do not over-tighten the four screws!
Note: Proper operation must be verified before the
pump is put back into service. See section 6.2.1.1
above.
6.2.3 Specific problems with motorized pump
6.2.3.1 Pump will not turn on
Possible causes:
Pump is not properly attached to the instrument.
Instrument is not turned on. Instrument battery too low
for pump operation.
Solution(s):
Make sure pump is properly attached to instrument,
recharge or replace instrument batteries.
58
Page 61
6.2.3.2 Can’t resume normal operation after a “Low
Flow” shut down
Possible causes:
Sample probe or internal pump filters need replacement,
sample hose kinked, sample probe and probe assembly
contains fluids.
Solution(s):
Turn off PhD Lite, remove pump, disconnect sample
probe an hose assembly, allow any trapped fluids to
drain; replace filters as necessary, examine hose to
make sure there are no kinks blocking normal flow.
6.3 Sample probe assembly
The sample probe handle contains moisture barrier
(hydrophobic) and particulate filters designed to remove
contaminants that might otherwise harm the instrument.
CAUTION: Never operate the PhD Lite in remote
sampling mode without the sample probe and hose
assembly. The sample probe handle contains
replaceable filters designed to block moisture and
remove particulate contaminants. If the pump is
operated without the probe assembly in place
partic ulate contaminants may c ause d amage to the
pump, sensors and internal components of the PhD
Lite.
Particulate contaminants are removed by means of a
cellulose filter. The hydrophobic filter includes a 0.1 µm
Teflon™ barrier which blocks the flow of moisture as well
as any remaining particulate contaminants.
Sample probe filters should be replaced whenever
visibly discolored due to contamination. A spare filter
replacement kit (Biosystems part number 54-05-K0401)
is included with every PhD Lite.
6.3.1 Changing sample probe filters
Figure 6.3 PhD Lite sa mple dr a w prob e
The threaded sample probe handle is unscrewed (as
shown in Figure 6.3 above) to provide access to the
filters. The particulate filter is held in place by a clear
filter cup. To replace the particulate filter, remove the
old filter and cup, insert a new filter into the cup, and
slide the cup back into place in the probe handle. The
hydrophobic barrier filter fits into a socket in the rear
section of the probe handle. (The narrow end of the
hydrophobic barrier filter is inserted towards the rear of
the handle.)
6.3.2 Changing sample probe tubes
The standard 11.5” long probe tube is held in place by
means of a hex-nut compression fitting and
compression sleeve. The standard probe tube is
designed to be easily interchangeable with other custom
length sections of 1/4” OD tubing, or probe tubes made
of other materials (such as stainless steel).
To exchange probe tubes, loosen the hex-nut
compression fitting, remove the old tube, slide the
compression sleeve into place around the new tube,
insert the new tube into the probe handle, and replace
and tighten the hex-nut.
59
Page 62
6.4 Firmware upgrade from the Biosystems
website
The instrument firmware for the PhD Lite may be
upgraded directly from Biosystems website. The Flash
download is a two-step download process that will
prepare your PC to update the flash program of your
PhD Lite.
Note: An IrDA port is required to download the new
Flash Program to your PhD Lite.
To install the files:
1. Close all applications (except the internet
connection) that are currently running on your
computer.
2. Go to www.biosystems.com and select the
Technical / Software Downloads / PhD Lite from
control bar at the right of the screen.
3. Follow the directions given on the screen to
download the flash utility program and then the
flash file itself.
6.5. Returning your PhD Lite to Biosystems for
service or repair
Please contact the Biosystems Service Department at
(860) 344-1079 to obtain a “Return Authorization”
number prior to shipment. A Biosystems Instrument
Service Representative will record all relevant
information or special instructions at that time.
To ensure safe transport, whenever possible please use
the original PhD Lite packing materials when returning
instrum ents to Biosystems for service. If the original
packing materials are not available, please take
additional care to pack the instrument in packing
materials that will protect the instrument and
accessories during shipment.
Note: The return authorization number must be
clearly marked on the outside of the box.
Writing the return authorization number prominently on
the outside of the box ensures that the return will be
immediately identified and logged into our system at the
time it is received. Proper tracking helps avoid
unnecessary delays in completion of service
procedures.
Please contact the Biosystems Instrument
Service Department at (800) 711-6776 or
(860) 344-1079 if you require any
additional information.
Thank you for choosing the PhD Lite, and
thank you for choosing Biosystems.
12 34-046 Gasket Main Housing
13 35-0172 Lower Case Assembly
14 17-036 O-ring for Main Housing
15 05-1117 Main Housing Screw
16 31-255 IRDA overlay
17 10-332 IRDA Lens
18 35-0237 Alarm Board Phd Lite
Part
Number
35-0205
35-0241
Description
Alkaline Top Case Assembly
NiMH Top Case Assembly
housing
housing
Alkaline Main Board Phd
Lite
NiMH Main Board Phd Lite
screws
61
Page 64
Appendices
Appendix A Toxic gas measurement - Ceilings,
TWAs and STELs
Many toxic substances are c ommonly encountered in industry. The
presenc e of toxic substances may be due to materials being stored
or used, the work being performed, or may be generated by natural
process es. Exposure to toxic substances c an produce dis ease,
bodily injury, or death in unprotected workers.
It is important to determine the amounts of any toxic materials
potentially present in the workplace. The amounts of toxic materials
potentially present will aff ect the procedures and personal protective
equipment which must be used. The s afest c ourse of action is to
eliminate or permanently control hazards through engineering,
workplace controls, ventilation, or other safety procedures.
Unprotected workers may not be expos ed to levels of toxic
contaminants whic h exceed Permissible Exposure Limit (PEL)
concentrations. Ongoing monitoring is necess ary to insur e that
exposure levels have not changed in a way that requires the use of
different or more rigorous procedures or equipment.
Airborne toxic substances are typically classified on the basis of
their ability to produce physiologic al effects on exposed workers.
Toxic substances tend to produce symptoms in two time frames.
Higher levels of exposure tend to produc e immediate (acute) effects,
while lower levels of long-term (chronic) exposure may not produce
physiological s ymptoms for years.
Hydrogen sulfide (H2S) is a good example of an acutely toxic
substance which is immediately lethal at relativel y low
concentrations. Exposure to a 1,000 ppm (parts per million)
concentration of H2S in air produces rapid paralysis of the
respirat ory system, cardiac arrest, and death within minutes.
Carbon monoxide (CO) is a good example of a chronically toxic gas.
Carbon monoxide bonds to the hemoglobin molecules in red blood
cells. Red blood cells contaminated with CO are unable to transport
oxygen. Although very high concentrations of c arbon monoxide may
be acutely toxic, and lead to immediate respiratory arrest or death, it
is the long term physiological eff ects due to chr onic exp osure at
lower levels that take the greatest toll of affected workers. This is
the situation with regards to smokers, parking garage att endants, or
others chronic ally exposed to carbon monoxide in the workplace.
Exposure levels are too low to produce immediat e symptoms, but
small repeated doses reduc e the oxygen c arrying capacity of the
blood over time to dangerously low levels. This partial impairment of
the blood supply may lead over time to serious physiological
consequences.
Because prudent monitoring programs must take both time
frames into account, there are three independent exposure
measurements and alarm types built into the PhD5 design.
Time History Graph
Ceiling
2. Time Weighted Average (TWA):
The maximum average concentration to which an unprotected
worker may be expos ed over an eight hour working day. During this
time, STEL and ceiling concentration limits may not be exceeded.
Time History Graph
(8 hour)
Ceiling
TWA
3. Short Term Exposure Limits (STEL):
Toxic substances may have short term exposure limits which are
higher than the eight hour TW A. The STEL is the maximum
average concentration to which an unpr otected worker may be
exposed in any fifteen minute interval during the day. During this
time, neither the eight hour TWA or the ceiling concentration may
be exceeded.
Any fifteen minute periods in which the average STEL
concentration exceeds the permissible eight hour TWA must be
separated from each other by at least one hour. A maximum of
four of t hese periods are allowed per eight hour shift.
1. Ceiling level:
OSHA has assigned some, but not all, toxic substances with a
ceiling level. T his is the highest c oncentr ation of a toxic substance
to which an unprotect ed worker should ever be exposed, even for a
very short time. Nev er enter an envir o n m ent even mom entarily
when concentrations of toxic substances exceed the ceiling
level.
62
Time History Graph
15 Minutes
Ceiling
STEL
TWA
.
Page 65
Appendix B PhD Lite sensor ranges
Part No. Description Range Resolution
54-25-80 LEL Combustible gas 0 – 100% LEL 1% LEL
54-25-90 O
Special reactive gas calibration adapt er (54-26-0405) r equired for proper calibration.
The non-s pec ific Cl2 and ClO2 sensors can be used for the detection of both gases. The specific Cl2 and ClO2 sensors are designed for the
3
Phosphine 0 – 20 PPM 0.1 PPM
63
Page 66
64
Appendix C Sensor Cross-Sensitivity Chart
The table below provides the cross-sensitivity response of the PhD Lite toxic gas sensors to common interference gases. The values are express ed as a percentage
of the primar y sensitivity, or the reading of the s ensor when exposed to 100ppm of the inter fering gas at 20ºC. These values are approximate. The actual values
depend on the age and condition of the sensor. Sensors should always be calibrated to the primary gas type. Cross-sensitive gases should not be used as sensor
calibrati on surrogat es wit hout the express wri tten consent of Bio systems.
SENSOR
CO H2S SO2 NO NO2 Cl2 ClO2 H2 HCN HCl NH3C2H4C2H
2
Carbon Monoxide(CO)
10010 5 10 -15 -5 -15 50 15 3 0 75 250
Carbon Monoxide (CO+)
100 350 50 30 -60 -60 -120 50 n/d n/d 0 75 250
Carbon Monoxide (CO-H)
10010 5 n/d (-) (-) (-) 5 n/d n/d n/d (+) (+)
Hydrogen Sulfide (H2S)
0.5 100 20 2 -20 -20 -60 0.2 0 0 0 n/d n/d
Sulfur Dioxide (SO2)
1 1 100 1 -100-50 -150 0.2 n/d n/d 0 (+) (+)
Nitrogen Dioxide (NO2)
-5 -8 -1 0 100 90 270 0 n/d n/d 0 n/d n/d
Chlorine (Cl2) (nonspecific)
0 -3 -1 0 110 100 310 0 n/d n/d 0 n/d n/d
Chlorine (Cl2) (specific)
0 -3 0 n/d 12 100 0 0 0 2 n/d 0 0
Chlorine Dioxide (ClO2) (nonspecific)
0 -1 -0.3 0 40 33 100 0 n/d n/d 0 n/d n/d
Chlorine Dioxide (ClO2) (specific)
0 0 0 n/d n/d 0 100 0 0 0 n/d 0 0
Ammonia (NH3) –21
0 <5 0 n/d 0 0 n/d 0 0 0 1000 0
Ammonia (NH3) -04
0 130 70 15 -5 -50 -150 0 5 0 100 0 n/d
Phosphine (PH3)
0.5 25 20 n/d (-) (-) (-) 0.1 n/d n/d n/d 1 0.5
Hydrogen Cyanide (HCN)*
0.5 350 160 -5 -100-20 -60 0.1 100 65 -5 50 n/d
n/d = no data, (+) undetermined positive, (-) undetermined negative
O
2
Sensor/CO2 Cross Sensitivity: The output of the O
2
sensor used in the PhD Lite will be enhanced by approximately 0.3% of signal per 1% of CO2.
*Reduced sulfur gases (H
2
S and SO2) are considered to be HCN sensor poisons. HCN sensors are not recommended for use in areas prone to sulfur gas presence.
Page 67
Appendix D Calibration Frequency
One of the most common questions
that we are asked at Biosystems is:
“How often should I calibrate my
gas detector?”
Sensor Reliability and Accuracy
Today’s sensors are designed to provide years of reliable service. In fact,
many sensors are designed so that
with norm al use they will only lose 5%
of their sensitivity per year or 10%
over a two-year period. Given this, it
should be possible to use a sensor for
up to two full years without any
significant loss of sensitivity.
A lot of sensors indeed do last that
long with only minimal loss of
sensitivity. However, there are a
number of reasons why a sensor may
unexpectedly lose additional
sensitivity or even fail to respond to
gas. Such reasons include
desiccation, poisoning, physical
restriction of airflow, overexposure,
leakage, and mechanical damage due
to dropping or immersion.
Verification of Accuracy
With so many reasons why a sensor
can lose sensitivity and given the fact
that dependable sensors can be key
to survival in a hazardous
environment, frequent verification of
sensor performance is paramount.
There is only one sure way to verify
that a sensor can respond to the gas
for which it is designed. That is to
expose it to a known concentration of
target gas and compare the reading
with the concentration of the gas.
This is referred to as a “bump” test.
This test is very simple and takes only
a few seconds to accomplish. The
safest course of action is to do a
“bump” test prior to each day’s
use. It is not necessary to make a
calibration adjustment if the readings
are between 90%* and 120% of the
expected value. As an example, if a
CO sensor is checked using a gas
concentration of 50 PPM it is not
necessary to perform a calibration
unless the readings are either below
45 PPM or above 60 PPM.
** The Canadian Standards
Association (CSA) requires the
instrument to undergo calibration
when the displayed value during a
bump test fails to fall between
100% and 120% of the expected
value for the gas.
Lengthening the Intervals between
Verifications of Accuracy
We are often asked whether there are
any circumstances in which the period
between accuracy checks may be
lengthened.
Biosystems is not the only
manufacturer to be asked this
question! One of the professional
organizations to which Biosystems
belongs is the Industrial Safety
Equipment Association (ISEA). The
“Instrument Products” group of this
organization has been very active in
developing a protocol to clarify the
minimum conditions under which the
interval between accuracy checks
may be lengthened.
A number of leading gas detection
equipment manufacturers have
participated in the development of the
ISEA guidelines concerning
calibration frequency. Biosystems
procedures closely follow these
guidelines.
If your operating procedures do not
permit daily checking of the sensors,
Biosystems recommends the
following procedure to establish a
safe and prudent accuracy check
schedule for your Biosystems
instrum ents:
1. During a period of initial use of at
least 10 days in the intended
atmosphere, check the sensor
response daily to be sure there is
nothing in the atmosphere that is
poisoning the sensor(s). The period
of initial use must be of sufficient
duration to ensure that the sensors
are exposed to all conditions that
might have an adverse effect on the
sensors.
2. If these tests demonstrate that it
is not necessary to make
adjustm ents, the time between
checks may be lengthened. The
interval between accuracy checking
should not exceed 30 days.
3. When the interval has been
extended the toxic and combustible
gas sensors should be replaced
immediately upon warranty
expiration. This will minimize the
risk of failure during the interval
between sensor checks.
4. The history of the instrument
response between verifications
should be kept.
Any conditions, incidents,
experiences, or exposure to
contaminants that might have an
adverse effect on the calibration
state of the sensors should trigger
imm ediate re-verification of
accuracy before further use.
5. Any changes in the environment
in which the instrument is being
used, or changes in the work that is
being performed, should trigger a
resum ption of daily checking.
6. If there is any doubt at any time
as to the accuracy of the sensors,
verify the accuracy of the sensors
by exposing them to known
concentration test gas before
further use.
Gas detectors used for the detection
of oxygen deficiencies, flammable
gases and vapors, or toxic
contaminants must be maintained and
operated properly to do the job they
were designed to do. Always follow
the guidelines provided by the
manufacturer for any gas detection
equipment you use!
If there is any doubt regarding your
gas detector's accuracy, do an
accuracy check! All it takes is a few
moments to verify whether or not your
instrum ents are safe to use.
One Button Auto Calibration
While it is only necessary to do a
“bump” test to ensure that the sensors
are working properly, all current
Biosystems gas detectors offer a one
button auto calibration feature. This
feature allows you to calibrate a
Biosystems gas detector in about the
same time as it takes to complete a
“bump” test. The use of automatic
bump test and calibration stations can
further simplify the tasks, while
automatically maintaining records
Don't take a chance
with your life.
Verify accuracy frequently!
Please read also Biosystems’
application note: AN20010808 “Use of ‘equivalent’ calibration gas mixtures”.
This application note provides
procedures to ensure safe calibration
of LEL sensors that are subject to
silicone poisoning.
All of Biosystems Applications Notes
are located on the Biosystems website
at
http://www.biosystems.com
65
Page 68
Appendix E Biosystems Standard Warranty Gas Detection Products
General
Biosystems LLC (hereafter Biosystems) warrants gas detectors, sensors and accessories manufactured and sold by
Biosystems, to be free from defects in materials and workmanship for the periods listed in the tables below.
Damages to any Biosystems products that result from abuse, alteration, power fluctuations including surges and
lightning strikes, incorrect voltage settings, incorrect batteries, or repair procedures not made in accordance with the
Instrument’s Reference Manual are not covered by the Biosystems standard warranty.
The obligation of Biosystems under this warranty is limited to the repair or replacement of components deemed by the
Biosystems Instrument Service Department to have been defective under the scope of this standard warranty. To
receive consideration for warranty repair or replacement procedures, products must be returned with transportation and
shipping charges prepaid to Biosystems at its manufacturing location in Middletown, Connecticut, or to a Biosystems
Authorized Warranty Service Center. It is necessary to obtain a return authorization number from Biosystems prior to
shipment.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS,
EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF FITNESS FOR A PARTICULAR
PURPOSE. BIOSYSTEMS WILL NOT BE LIABLE FOR LOSS OR DAMAGE OF ANY KIND CONNECTED TO THE
USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
Toxi/Oxy Plus, Toxi/Oxy Ultra, ToxiVision, Ex Chek
As long as the instrument is in service
Toxi/Oxy Pro, MultiPro 2 years from date of purchase
2 years after activation or 2 years
Toxi Limited
after the “Must Be Activated By” date,
whichever comes first
3 years after activation or 3 years after the
Toxi3Ltd®
“Must Be Activated By” date, whichever
comes first
90 days after activation or 90 days
Mighty-Tox
after the “Must Be Activated By” date,
whichever comes first
Mighty-Tox 2
Prorated credit is given towards repair or purchase of a new unit of the
same type.
IQ Systems, Series 3000, Airpanel, Travelpanel, ZoneGuard,
Gas9Chek1 and Gas9Chek4
0 – 6 months of use 100% credit
6 – 12 months of use 75% credit
12 – 18 months of use 50% credit
18 – 24 months of use 25% credit
One year from the date of purchase
Battery packs and chargers, sampling pumps and other components,
which by their design are consumed or depleted during normal
One year from the date of purchase
operation, or which may require periodic replacement
Sensor Warranty Periods
Instrument(s) Sensor Type(s) Warranty Period
PhD Plus, PhD Ultra, PhD5, PhD Lite,
Cannonball3, MultiVision, MultiPro,
ToxiVision, ToxiPro, Ex Chek
Toxi, Toxi/Oxy Plus, Toxi/Oxy Ultra
All Others All Sensors 1 Year
** Damage to combustible gas sensors by acute or chronic exposure to known sensor poisons such as volatile lead
(aviation gasoline additive), hydride gases such as phosphine, and volatile silicone gases emitted from silicone
caulks/sealants, silicone rubber molded products, laboratory glassware greases, spray lubricants, heat transfer
fluids, waxes & polishing compounds (neat or spray aerosols), mold release agents for plastics injection molding
operations, waterproofing formulations, vinyl & leather preservatives, and hand lotions which may contain
ingredients listed as cyclomethicone, dimethicone and polymethicone (at the discretion of Biosystems Instrument
Service department) void Biosystems’ Standard Warranty as it applies to the replacement of combustible gas
sensors.
O2, LEL**, CO, CO+, H2S
& Duo-Tox
2 Years
All Other Sensors 1 Year
CO, CO+, H2S 2 Years
All Other Sensors 1 Year
66
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.