Bio-Rad Rat Diabetes Assays User Manual

Bio-Plex Pro
Diabetes Assays
Instruction Manual
For technical support, call your local Bio-Rad office, or in the U.S., call 1-800-424-6723. For research use only. Not for diagnostic procedures.
Table of Contents
Introduction 1
Principle 2
Kit Contents and Storage 4
Recommended Materials 5
Assay Workflow 6
Important Considerations 7
Detailed Instructions
1. Plan Plate Layout 8
2. Prepare Instrument 9
3. Prepare Wash Method 10
4. Prepare Standards 11
5. Prepare Samples 16
6. Prepare Coupled Beads 20
7. Run Assay 22
8. Read Plate 28
Troubleshooting Guide 35
Appendix: Protease Inhibitors 40
Plate Layout Template 41
Calculation Worksheet 42
Safety Considerations 46
Legal Notices 46
Ordering Information 47

Introduction

Bio-Plex Pro™ diabetes assays are magnetic bead–based multiplex assays designed to measure multiple diabetes-related biomarkers in a minimal volume of matrix such as serum, plasma, tissue culture supernatant, or other biological fluids. The biomarkers chosen for these assays are circulating proteins involved in the regulation of glucose metabolism.
These multiplex assays are configured for the detection of various metabolic markers in human, mouse, and rat samples. The multiplex assays for human matrices were tested and found to be cross-reactive to at least four common non-human primate (NHP) species, including rhesus, cynomolgus, baboon, and marmoset. The degree of cross-reactivity was profiled according to the ability of each assay to detect these metabolic markers in the sera and mitogen-stimulated peripheral blood mononuclear cell (PBMC) culture supernatant of these animals.
The high level of gene homology across the major NHP species may allow the NHP diabetes assays to be used in other species, such as chimpanzees, African green monkey, and pigtail macaques. However, assay performance has not been specifically evaluated in these animals.
Cross-Reactivity with Non-Human Primates
African green Pigtail Rhesus Cynomolgus Baboon Marmoset Chimpanzee monkey macaque
Tested Not tested
Bio-Plex Pro assays enable researchers to quantify multiple protein biomarkers in a single well of a 96-well plate in 3–4 hours. These robust immunoassays require as little as 12.5 μl serum or plasma or 50 μl cell culture supernatant or other biological fluid. The use of magnetic (MagPlex) beads allows researchers to automate wash steps on a Bio-Plex Pro (or similar) wash station. Magnetic separation offers greater convenience, productivity, and reproducibility compared to vacuum filtration.
For more information please visit www.bio-rad.com/bio-plex.
1

Principle

Technology
The Bio-Plex® multiplex system is built upon the three core elements of xMAP technology:
n
Fluorescently dyed microspheres (also called beads), each with a distinct
color code or spectral address to permit discrimination of individual tests within a multiplex suspension. This allows simultaneous detection of up to 500 different types of molecules in a single well of the 96-well microplate on the Bio-Plex molecules on the Bio-Plex of molecules on the Bio-Plex
n
On the Bio-Plex 200 and Bio-Plex 3D systems, a dedicated flow
cytometer with two lasers and associated optics to measure the different molecules bound to the surface of the beads. In the Bio-Plex MAGPIX, the entire sample load volume is injected into a chamber where the beads are imaged using LED and CCD technology
n
A high-speed digital signal processor that efficiently manages the
fluorescence data
Assay Format
Bio-Plex Pro™ assays are essentially immunoassays formatted on magnetic beads. The assay principle is similar to that of a sandwich ELISA (Figure 1). Capture antibodies directed against the desired biomarker are covalently coupled to the beads. Coupled beads react with the sample containing the biomarker of interest. After a series of washes to remove unbound protein, a biotinylated detection antibody is added to create a sandwich complex. The final detection complex is formed with the addition of streptavidin-phycoerythrin (SA-PE) conjugate. Phycoerythrin serves as a fluorescent indicator or reporter.
®
3D system, up to 100 different types of
®
200 system, and up to 50 different types
®
MAGPIX™ system
2
Biomarker of Interest
Streptavidin
Magnetic Bead
Capture
Antibody
Fig. 1. Bio-Plex sandwich immunoassay.
Biotinylated
Detection
Antibody
Phycoerythrin
Fluorescent
Reporter
Data Acquisition and Analysis
Data from the reactions are acquired using a Bio-Plex system or similar Luminex-based reader. When a multiplex assay suspension is drawn into the Bio-Plex 200 reader, for example, a red (635 nm) laser illuminates the fluorescent dyes within each bead to provide bead classification and thus assay identification. At the same time, a green (532 nm) laser excites PE to generate a reporter signal, which is detected by a photomultiplier tube (PMT). A high-speed digital processor manages data output, and Bio-Plex Manager
software presents data as median fluorescence intensity (MFI) as well as concentration (pg/ml). The concentration of analyte bound to each bead is proportional to the MFI of reporter signal.
Using Bio-Plex Data Pro
software, data from multiple instrument runs can be combined into a single project for easy data management, quick visualization of results, and simple statistical analysis.
3

Kit Contents and Storage

Reagents Supplied
Bio-Plex Pro™ diabetes assays are offered in a convenient kit format that includes assay, reagent, and diluent components in a single box (Table 1).
Table 1. Contents of Bio-Plex Pro diabetes assays.*
Component
Standard diluent (not included in adiponectin/adipsin assays)
Sample diluent (not included in adiponectin/adipsin assays)
Serum-based diluent (adiponectin/adipsin assays only)
Assay buffer
Wash buffer
Detection antibody diluent
Streptavidin-PE (100x)
Filter and/or flat bottom plate (96-well)
Sealing tape
Assay Quick Guide 1 booklet
Coupled magnetic beads (20x)
Detection antibodies (20x)
Standard
* Volumes shown are approximate. Quantities in Bio-Plex
1 x 96-Well
Format
10 ml
40 ml
70 ml
50 ml
200 ml
5 ml
1 tube
1 plate
1 pack of 4
1 tube
1 tube
1 vial
®
Express assays will vary.
Storage and Stability
Kit contents should be stored at 4°C and never frozen. Coupled magnetic beads and streptavidin-PE should be stored in the dark. All components are guaranteed for a minimum of six months from the date of purchase when stored as specified.
10 x 96-Well
Format
100 ml
80 ml
N/A
500 ml
1.5 L
50 ml
1 tube
10 plates
10 packs of 4
1 tube
1 tube
10 vials
4

Table 2. Recommended materials.

Item
Bio-Plex Pro Assays Quick Guide 2
Ordering Information
Bulletin #10024973 (download at www.bio-rad.com/bio-plex)
Bio-Plex 200 system or Luminex system with HTF
Bio-Plex validation kit
Bio-Rad catalog #171-000205
Bio-Rad catalog #171-203001 Run the validation kit monthly to ensure optimal performance of fluidics and optics systems
Bio-Plex calibration kit
Bio-Rad catalog #171-203060 Run the calibration kit daily to standardize fluorescence signal
Bio-Plex Pro wash station
Bio-Rad catalog #300-34376 For use with magnetic bead–based assays only
Bio-Plex Pro II wash station
Bio-Rad catalog #300-34377 For use with both polystyrene (nonmagnetic) and magnetic bead–based assays
Bio-Plex handheld magnetic washer
Bio-Rad catalog #170-20100 For use with magnetic bead–based assays only
Bio-Plex Pro flat bottom plates, 40 x 96-well
Bio-Rad catalog #171-025001 For magnetic separation on the Bio-Plex Pro wash station
Microtiter plate shaker
IKA MTS 2/4 shaker for 2 or 4 microplates
IKA catalog #320-8000 or Barnstead/Lab-Line Model 4625 plate
VWR catalog #57019-600 shaker (or equivalent capable of 300–1,100 rpm)
®
Aurum™ vacuum manifold
Bio-Rad
For vacuum filtration
BR-2000 vortexer
Reagent reservoirs, 25 ml
For capture beads and detection antibodies
Reagent reservoir, 50 ml (for reagents and buffers)
Pall Life Science Acrodisc, 25 mm PF syringe filter
(0.8/0.2 µm Supor membrane)
Filter plate, 1 x 96-well, clear plastic lid and tray
®
Titertube
Micro test tubes
For preparing replicate standards, samples, and controls
Bio-Rad catalog #732-6470
Bio-Rad catalog #166-0610
VistaLab catalog #3054-1002
or
VistaLab catalog #3054-1004
VistaLab catalog #3054-1006
Pall Life Sciences
catalog #4187
Bio-Rad catalog #171-304502
Bio-Rad catalog #223-9390 prior to loading the plate
Other: 15 ml polypropylene tubes for reagent dilutions, calibrated pipets, pipet tips, sterile distilled water, aluminum foil, absorbent paper towels, 1.5 or 2 ml microcentrifuge tubes, and standard flat bottom microplate (for calibrating vacuum manifold).
5
Assay Workflow
Prewet wells
(for lter plate only)
Add 50 μl 1x beads to wells
Wash 2 x 100 μl
Add 50 μl standards, blank, samples
incubate 1 hr at RT with shaking at 850 rpm
Wash 3 x 100 μl
Add 25 μl 1x detection antibody, incubate
30 min at RT with shaking at 850 rpm
Wash 3 x 100 μl
Add 50 μl 1x streptavidin-PE, incubate
10 min at RT with shaking at 850 rpm
Wash 3 x 100 μl
Resuspend in 125 μl assay buffer,
shake at 850 rpm for 30 sec
Read plate on Bio-Plex system
6

Important Considerations

Instruments and Software
The diabetes assays described in this manual are compatible with all currently available Luminex-based life science research instruments. Assays can be read and analyzed with either Bio-Plex Manager or Luminex xPONENT software.
Assay Procedures
Pay close attention to vortexing, shaking, and incubation times and to Bio-Plex
®
reader PMT (RP1) setting, as these have been optimized
specifically for each assay panel.
Assay Quick Guide
Each assay kit includes a printed Bio-Plex Pro Assay Quick Guide (bulletin #10024973), which can be used to prepare and run a full 1 x 96-well assay plate. Users can also download a copy at www.bio-rad.com/bio-plex.
Bead Regions
Bead regions for all analytes are listed in the Read Plate section.
Multiplexing Compatibility
For human and mouse, the maximum number of singleplex diabetes and cytokine analytes that may be mixed is limited by the 10x cytokine antibody stock concentrations as shown in the table below.
Table 3. Maximum number of singleplex diabetes and cytokine analytes that may be multiplexed.
Human, mouse, and rat diabetes
Mouse cytokine (group III) analytes (20x)
0 2 4 6 8 10
software
Human and mouse cytokine (groups I, II) analytes (10x)
10 9 8 7 6 5
7

1. Plan Plate Layout

Prior to running the assay, determine the total number of wells in the experiment using the Plate Layout Template on page 41 or the Plate Formatting tab in Bio-Plex Manager shown in Figure 2, with all conditions in duplicate.
1. Assign standards to columns 1 and 2, with the highest concentration in row A and the lowest concentration in row H.
2. Assign the blank to wells A3 and A4. The blank should consist of your chosen standard diluent and be processed in the same manner as sample and standard wells. Note that Bio-Plex Manager automatically subtracts the blank (B) MFI value from all other assay wells.
3. User-defined controls are assigned to wells in columns 3 and 4.
4. The remainder of the plate is available for samples.
5. Once the total number of wells is known, calculate the required volumes of beads, detection antibody, and streptavidin-PE needed. Use Tables 9–11, 14–16, and 17, respectively, or the Calculation Worksheet on page 43.
Legend
S Standard
software. A suggested plate layout is
B Blank
X Samples
C Controls
Fig. 2. Suggested plate layout. For detailed instructions on plate formatting in Bio-Plex Manager, see Section 8.
8

2. Prepare Instrument

Start up and calibrate the Bio-Plex® 100/200 or similar system with Bio-Plex Manager kit should be run daily or before each use of the instrument to standardize the fluorescent signal. To prepare either a Bio-Plex 3D or Bio-Plex MAGPIX
The validation kit should be run monthly to ensure performance of fluidics and optics systems. Refer to either the software manual or online Help for directions on how to conduct validation.
reader, consult its respective user manual.
software prior to setting up the assay. The calibration
®
Start Up System (Bio-Plex 100, 200, or Similar)
1. Empty the waste bottle and fill the sheath fluid bottle before starting if high throughput fluidics (HTF) are not present. This will prevent fluidic system backup and potential data loss.
2. Turn on the reader, XY platform, and HTF (if included). Allow the
system to warm up for 30 min (if not already done).
3. Select Start up
for 4 hr without acquiring data, the lasers will automatically turn off. To reset the 4-hr countdown, select Warm up lasers/optics to reach operational temperature.
and follow the instructions. If the system is idle
and wait for the
Calibrate System
1. Select Calibrate and confirm that the default values for CAL1
and CAL2 are the same as the values printed on the bottle of Bio-Plex calibration beads. Use the Bio-Plex system low RP1 target value even if assays will be run at high RP1.
2. Select OK and follow the software prompts for step-by-step
instructions for CAL1 and CAL2 calibration.
Note: In Bio-Plex Manager version 6.1 and higher, startup, warm up,
and calibration can be performed together by selecting the “Start up and calibrate”
icon.
9

3. Prepare Wash Method

Bio-Plex Pro™ assays are compatible with both magnetic separation and vacuum filtration methods. However, for best results, we recommend performing the assays in a flat bottom plate with magnetic separation.
Table 4. Summary of compatible wash stations and plate types.
Wash Method Wash Station Assay Plate
Magnetic separation Bio-Plex Pro Flat bottom plate Bio-Plex Pro II (use MAG programs) Bio-Plex
Vacuum filtration Bio-Plex Pro II (use VAC programs) Filter plate Vacuum manifold (manual)
Setting up the Bio-Plex Pro or Bio-Plex Pro II Wash Station
The wash station does not require calibration; however, it should be primed before use. For more information, refer to the Bio-Plex Pro and Pro II wash station quick guide (bulletin #5826).
1. Install the appropriate plate carrier on the wash station.
2. Use the prime procedure to prime channel 1 with wash buffer.
Setting Up the Bio-Plex Handheld Magnetic Washer
Place an empty flat bottom plate on the magnetic washer by sliding it under the retaining clips. Push the clips inward to secure the plate. Make sure the plate is held securely. If needed, the clips can be adjusted for height and tension. For detailed instructions, refer to the user guide (bulletin #10023087).
®
handheld magnetic washer
Setting up a Vacuum Manifold
Calibrate the vacuum manifold by placing a standard 96-well flat bottom plate on the unit and adjusting the pressure to –1 to –3" Hg. In general, 100 µl liquid should take 3–4 sec to clear the well. For more detailed instructions, refer to bulletin #10005042.
10

4. Prepare Standards

General Instructions
n
It is essential to prepare standards exactly as described in this section.
Incorrect preparation may lead to low signal or variable measurements from plate to plate
n
The peel-off sticker provided with the standards lists the most
concentrated point on the standard curve (S1). Enter this information into Bio-Plex Manager
n
For users who wish to mix assays from different panels, such as diabetes assays with group I cytokines, guidance is provided here for mixing 2 different lyophilized standards. Bead regions were chosen to avoid overlap whenever possible. However, performance of multiplexes containing assays from different groups have not been extensively validated. Therefore, users must confirm that the assay performance is
still fit for their purpose
Selecting a Diluent for Standards
Refer to Table 5 for recommended diluents based on different sample types.
As a general rule, reconstitute and dilute standards in a diluent similar to the final sample type or sample matrix.
Table 5. Summary of recommended diluents for standards.
Sample Type Diluent for Standards Add BSA
Serum and plasma Standard diluent None Culture media, with serum Culture media None Culture media, serum-free Culture media To 0.5% final
For Adiponectin and/or Adipsin Assays
Serum and plasma Serum-based diluent None
software as instructed in section 8
11
RP1 (PMT) Setting for Standard Curves
The Bio-Plex 200 and 3D systems have two RP1 (PMT or photomultiplier tube) setting options, while the Bio-Plex
®
MAGPIX™ has no PMT and therefore no PMT setting options. Instead, MAGPIX uses default instrument settings similar to low PMT on the Bio-Plex 200 (Table 6).
Table 6. Overview of PMT setting options on Bio-Plex systems.
Instrument RP1 (PMT)
Bio-Plex 100, 200* Low, high
Bio-Plex 3D* Standard, enhanced
Bio-Plex MAGPIX* N/A, use default instrument settings
* Or similar Luminex-based system.
The Bio-Plex Pro diabetes assays were developed on the high PMT setting using the Bio-Plex 200 system. Protocols using alternative standard dilution series or low PMT settings should be validated by the end user, for example when mixing diabetes assays with cytokine assays (Table 7).
Table 7. Settings for optimal sensitivity on the Bio-Plex 200 system*.
Assay Low RP1 (PMT) High RP1 (PMT)
Human/NHP* diabetes Mouse diabetes User validation required* Rat diabetes
Low RP1 (PMT), broad High RP1 (PMT), narrow Cross-panel mixing
Human diabetes + cytokines Mouse diabetes + cytokines User validation required* Rat diabetes + cytokines
* Contact Bio-Rad technical support for the most up-to-date recommendations on PMT
settings and cross-panel multiplexing compatibility.
range cytokine curve range cytokine curve
•
•
Reconstitute a Single Vial of Diabetes Standards
This procedure prepares enough material to run each dilution in duplicate.
1. Gently tap the vial containing the lyophilized standard.
2. Add 500 μl of the appropriate standard diluent. Do not use assay buffer or sample diluent to reconstitute the standards.
12
3. Gently vortex the reconstituted standard for 5 sec then incubate on
ice for 30 min. Be consistent with the incubation time in every assay to ensure best results.
4. During the incubation period, prepare the samples as instructed in the
Prepare Samples section.
Prepare Diabetes Standard Dilution Series from a Single Antigen Vial
The following procedure produces an eight-point standard curve with a fourfold dilution between each point. Pipet carefully using calibrated pipets and use new pipet tips for every volume transfer.
1. Label nine 1.5 ml polypropylene tubes S1 through S8 and Blank.
2. Add the specified volume of standard diluent to each tube (Figures 3 and 4).
3. Vortex the reconstituted standards gently for 5 sec before removing
any volume. Add 128 µl into the S1 tube containing 72 µl of standard diluent. Vortex at medium speed for 5 sec, then use a new pipet tip to transfer 50 µl from S1 tube to S2 tube.
4. Continue with 1:4 (fourfold) serial dilutions from tube S2 to S8 as
shown in Figure 3. Use reconstituted and diluted standards immediately. Do not freeze for future use.
Fig. 3. Preparing a fourfold dilution series of diabetes standards.
128 50 50 50 50 50 50 50
Transfer Volume, µl
Reconstituted
Diabetes Standard
72 150 150 150 150 150 150 150 150
S1 S2 S3 S4 S5 S6 S7 S8 Blank
13
Diluent, µl
Loading...
+ 35 hidden pages