The Bio-Rad Technical Support department is open Monday through Friday, 5:00 AM to
5:00 PM, Pacific time.
Phone: 1-800-424-6723, option 2
Email: Support@Bio-Rad.com (U.S./Canada only)
For technical assistance outside the U.S. and Canada, contact your local technical support
office or click the Contact us at www.bio-rad.com.
Legal Notices
No part of this publication may be reproduced or transmitted in any form or by any means,
electronic or mechanical, including photocopy, recording, or any information storage or
retrieval system, without permission in writing from Bio-Rad Laboratories.
Bio-Rad reserves the right to modify its products and services at any time. This user guide is
subject to change without notice.
Although prepared to ensure accuracy, Bio-Rad assumes no liability for errors, or for any
damage resulting from the application or use of this information.
Chemiluminescent substrate technology is protected by U.S. Patent Number 6,432,662.
Clarity Max Western ECL Substrate is manufactured by Cyanagen Srl and is the subject
of patent application numbers US7855287, EP1950207, US9040252, AU2011202658,
CA2742025, US8129136, and EP1962095, together with other equivalent granted patents
and patent applications in other countries like CN102313732.
Precision Plus Protein Standards are sold under license from Life Technologies Corporation,
Carlsbad, CA for use only by the buyer of the product. The buyer is not authorized to sell or
resell this product or its components.
Bio-Rad, Trans-Blot, and Mini-PROTEAN are trademarks of Bio-Rad Laboratories, Inc. in
certain jurisdictions. All trademarks used herein are the property of their respective owner.
Safety Warnings
Warning: Do not open or attempt to repair the Trans-Blot Turbo instrument or cassettes.
Doing so will void your warranties and can put you at risk for electric shock. Return the
Trans-Blot Turbo instrument or cassettes to the factory (U.S. customers) or to an authorized
distributor (all other customers) if repairs are needed.
Under normal operating conditions, Trans-Blot Turbo cassettes may become warm during
transfer.
Clean and inspect the instrument and cassette after each use for damage. Contact Bio-Rad
technical support if any damage is noted.
Install and operate the Trans-Blot Turbo Transfer System in a clean and dry location away from
excess moisture. During operation, avoid using excessive buffer in or around the instrument.
Trans-Blot Turbo Transfer System | i
Page 4
Safety Compliance
Safety Compliance
This instrument has been tested and found to be in compliance with all applicable
requirements of the following safety and electromagnetic standards:
■
EN61010-1 Electrical Equipment for Measurement, Control, and Laboratory Use
■
UL STD No. 61010A-1 Electrical Equipment for Measurement, Control, and Laboratory
Use, Part 1: General Requirements
■
CAN/CSA C22.2 No. 61010-1-04 Safety Requirements for Measurement, Control, and
Laboratory Use, Part 1: General Requirements (includes Amendment 1)
■
IEC 61010-1 Safety Requirements for Measurement, Control, and Laboratory Use, Part
1 General Requirements
Electromagnetic Compatibility (EMC)
EN61326 Class A Electrical Equipment for Measurement, Control, and Laboratory Use,
General Requirements
FCC Warning and Notes
Warning: Changes or modification to this unit not expressly approved by the party
responsible for compliance could void the user’s authority to operate the equipment.
This Bio-Rad instrument is designed and certified to meet EN61010* and the EMC
requirements of EN61326 (for class A) safety standards. Certified products are safe to use
when operated in accordance with the instruction manual. This instrument should not be
modified or altered in any way. Alteration of this instrument will:
■
Void the manufacturer’s warranty
■
Void the EN61010 safety certification
■
Create a potential safety hazard
* EN 61010 is an internationally accepted electrical safety standard for laboratory
instruments.
Bio-Rad Laboratories is not responsible for any injury or damage caused by the use of this
instrument for purposes other than those for which it is intended, or by modifications of the
instrument not performed by Bio-Rad Laboratories or an authorized agent.
We strongly recommend that you follow the safety specifications listed in this section and
throughout this manual. Use only the supplied power cord with the instrument, making sure
to choose the plug adaptor that corresponds to the electrical outlets in your region.
ii | Trans-Blot Turbo Transfer System
Page 5
Specifications
Trans-Blot Turbo Instrument
Dimensions: (L x W x H) 26.0 x 21.1 x 20.4 cm
Weight: 8 lb with cassettes, 4.5 lb without cassettes
Input power: 100–240 VAC, 276 VA, 50–60 Hz, 175 W max
Fuses: Two 6.3 A, 250 V, fuses located above the power connection
On/off switch: Yes, mains connected
USB port: Yes, input only, for firmware updates
Cooling fan: Yes
Output power: 0–26 VAC (1 V increments), 0–2.6 A DC (0.1 A increments) for each cassette
3.4.5 Disassembly and Removal of the Membrane .................................................................... 18
3.5 Setting Up a New Protocol .......................................................................................................... 19
3.6 Transfer Using Traditional Semi-Dry Consumables ......................................................................20
3.7 Optimizing Transfer Conditions ....................................................................................................21
Appendices Appendix A. Maintenance ..............................................................................................................23
Appendix B. Troubleshooting ........................................................................................................24
Overview of the Three Protocol Modes ........................................................................................26
Consumables and Related Products ............................................................................................29
Trans-Blot Turbo Transfer System | v
Page 8
vi | Trans-Blot Turbo Transfer System
Page 9
1 Introduction
Fig. 1. Trans-Blot Turbo Blotting System.
Semi-dry western blotting is a common technique in many research and diagnostic
laboratories. Conventional semi-dry blotting protocols are often cumbersome, requiring a
great deal of time-consuming reagent preparation and setup, followed by an electrophoretic
transfer that could take up to an hour or more. Bio-Rad’s Trans-Blot Turbo System
accelerates the semi-dry blotting process without sacrificing performance. With the
Trans-Blot Turbo System, transfer time is reduced to as little as 3 min, and the prepackaged
transfer packs provide excellent transfer efficiency and reproducibility while eliminating the
time and mess associated with traditional transfer methods.
The Trans-Blot Turbo System consists of the main instrument with two blotting cassettes
and single-use prepackaged transfer packs that contain buffer-saturated membranes and
ion reservoir stacks.
Each blotting cassette has a built-in pair of anode and cathode electrode plates and can
accommodate one or two mini format (7.0 x 8.5 cm) gels or a single midi format
(8.5 x 13.5 cm) gel. Closed cassettes lock to hold a gel in contact with a transfer membrane
placed between two ion reservoir stacks. The loaded cassettes are simply inserted into the
instrument to access the power supply for protein transfer. A preprogrammed user-defined
Trans-Blot Turbo Transfer System | 1
Page 10
Introduction
transfer protocol is easily selected from the firmware menus using the navigation and
selection buttons and the alphanumeric keypad. During the run, transfer conditions and run
progress are displayed on the LCD screen.
Selection
buttons
Navigation
buttons
Fig. 2.
Home screen of the Trans-Blot Turbo System.
The buffer-saturated ion reservoir stacks and membrane are available as ready-to-use
prepackaged, disposable, single-use Trans-Blot Turbo Transfer Packs in two sizes, for
transfer of single mini gels (7.0 x 8.5 cm) and a larger size (8.5 x 13.5 cm) for transfer of midi
gels or simultaneous transfer of two mini gels (7.0 x 8.5 cm). Both the mini and midi transfer
packs are available with either nitrocellulose or PVDF membranes. These prepackaged
transfer packs allow for efficient transfer in as little as 3 min. Gels can be transferred
immediately after electrophoresis without equilibration. The reagents and consumables
available in the transfer packs are also available in ready-to-assemble format, as part of the
Trans-Blot Turbo RTA Transfer Kits. The system will also accommodate traditional semi-dry
western blotting consumables such as filter paper and Towbin buffer with efficient transfer in
30–60 min.
Selection of transfer protocols using the navigation and selection buttons is simple and the
system is designed to handle multiple sequential runs with no intervals of delay except for
the reloading of the cassettes.
1.1 Unpacking and Setup Instructions
The Trans-Blot Turbo Starter Kits include the main unit, two cassettes (bases and lids),
blot roller, standard power cord, quick start guide, instruction manual, and your choice of
nitrocellulose or PVDF membranes in mini or midi sizes.
Place the system on a level surface with at least 6 cm of clearance in the back for proper
ventilation. Plug the power cord into the back of the instrument, and then connect it to a
standard grounded outlet. Remove the foam ring separating the instrument and cassettes.
Power on the instrument using the switch on the right side of the unit.
Note: Do not place heavy items on the top of the Trans-Blot Turbo System.
1.2 Chemical Compatibility
The Trans-Blot Turbo System and cassette components are not compatible with strong
acids or bases, chlorinated hydrocarbons (for example, chloroform), aromatic hydrocarbons
(for example, toluene, benzene), or acetone. The cassettes and instrument casing can be
cleaned with water and a mild detergent, but do not use abrasives or organic solvents. The
stainless steel cathode electrode, which is housed in the cassette lid, can be cleaned with
7% acetic acid and wiped down with water.
2 | Trans-Blot Turbo Transfer System
Page 11
1.3 Safety/Cautions/Warning
The following guidelines should be observed and followed when using the Trans-Blot Turbo
instrument.
The Trans-Blot Turbo instrument has been tested for operation at 15–31ºC ambient
temperature and 0–95% relative humidity (noncondensing). Operating the unit outside these
conditions is not recommended.
■
To ensure adequate cooling, be sure that there is at least 6 cm of clearance behind the
unit and that the fan vent at the rear of the unit is not blocked
■
Always connect the unit to a grounded AC outlet using the power cord provided
■
Use caution when removing a cassette from the unit after a transfer run. The cassette
may be warm to the touch
■
Do not operate in extreme humidity (>95%) or where condensation can affect the internal
electrical circuits of the unit
■
Operation of the Trans-Blot Turbo System at temperatures <15ºC is not recommended.
However, the unit can function in a cold environment and can be operated immediately.
When returning the unit to normal conditions, allow it to equilibrate to room temperature
before use
Introduction
■
For your safety and for the protection of your Trans-Blot Turbo instrument, Bio-Rad
suggests you routinely clean the instrument in accordance with the enclosed instructions
and that you routinely check the contact pins located at the back of the instrument
cavity, as outlined in Appendix A
■
Bio-Rad recommends that you rinse and dry each cassette, base, and lid in deionized
water to remove residual salts and to prevent salt buildup. Air-dry or use a paper towel
to dry the cassette
■
Do not add excess or additional buffer into the transfer cassette at any point prior to
running the instrument
■
Install and operate the Trans-Blot Turbo Transfer System in a clean and dry location
away from excess moisture. During operation, avoid using excessive buffer in or around
the instrument.
Important: This instrument is intended for laboratory use only.
This product conforms to the class A standards for electromagnetic emissions intended for
laboratory equipment applications. It is possible that emissions from this product may interfere
with some sensitive appliances when placed nearby or on the same circuit as those appliances.
The user should be aware of this potential and take appropriate measures to avoid interference.
This product is designed and certified to meet EN 61010* safety standards. Certified products
are safe to use when operated in accordance with the instruction manual. This safety certification
does not extend to accessories that are not EN 61010 certified, even when used with this unit.
This instrument should not be modified or altered in any way. Alteration of this instrument will
void the manufacturer’s warranty, void the EN 61010 certification, and create a potential safety
Trans-Blot Turbo Transfer System | 3
Page 12
Introduction
1.4 Warranty
hazard for the user. Bio-Rad is not responsible for any injury or damage caused by the use of
this instrument for purposes other than those for which it is intended, or by modifications of the
instrument not performed by Bio-Rad or an authorized agent.
* EN 61010 is an internationally accepted electrical safety standard for laboratory
instruments.
The Trans-Blot Turbo Transfer System and associated accessories are covered by
a standard Bio-Rad Laboratories warranty. Contact your local Bio-Rad Laboratories
representative for the details of the warranty.
4 | Trans-Blot Turbo Transfer System
Page 13
2 Equipment and Reagents
Overview
The Trans-Blot Turbo instrument and reagents are designed to provide fast, efficient, and
reproducible western blots of protein gels in as little as 3 min. Each convenient transfer
pack contains the necessary materials to efficiently transfer a single mini gel (7.0 x 8.5 cm)
or one midi or two mini gels (8.5 x 13.5 cm). The transfer packs contain two buffer-soaked
ion reservoir stacks along with either a prewetted nitrocellulose or PVDF membrane. The
membranes do not require any pretreatment before use.
The reagents and consumables available in the transfer packs are also available in a readyto-assemble format as part of the Trans-Blot Turbo RTA Transfer Kits. These kits provide the
same optimized buffers and consumables as the transfer packs, but require wetting of the
membranes and filters before use.
Note: Traditional semi-dry western blotting consumables can also be used with the system.
See section 3.7.
LCD display
Navigation
buttons
Alphanumeric
keypad
Cassette bay A
(upper)
Cassette bay B
(lower)
Fig. 3. Trans-Blot Turbo instrument.
Trans-Blot Turbo Transfer System | 5
Page 14
Equipment and Reagents Overview
Fig. 4. Cassette base and lid. Reversible lid is labeled with lock, unlock, and safety hazard symbols.
For transfer, a gel is sandwiched between two ion reservoir stacks in the cassette. Each
cassette can hold one or two mini format gels (7.0 x 8.5 cm) or one midi format gel
(8.5 x 13.5 cm). One or both cassettes can be used for a blotting run. If both cassettes are
run, they must use the same protocol and have the appropriate combinations of gels, as
shown in Table 1.
The cassettes can be run individually or simultaneously with independent start times
using the same protocol. This allows the user to perform multiple sequential runs without
interruption, except for reloading a cassette. At the end of each cassette run, the ion
reservoir stacks are discarded. Membranes and gel (if needed) can be used immediately for
downstream applications or stored for later use.
The cassette bays are labeled A and B for convenience, and the cassettes are freely
interchangeable between the bays. There is no preference as to which bay contains which
cassette when running the combination of one midi and two mini gels. Also, one bay (either
A or B) can remain empty during a run.
Table 1. Acceptable gel transfer combinations when both bays are in use.
Upper bay A1 mini format gel2 mini or 1 midi
Lower bay B1 mini format gel2 mini or 1 midi
2.1 User Interface
The user interface consists of an LCD menu screen on the front of the system. Three
buttons below the screen are used to navigate among menu screens. Three buttons to the
right of the screen are for user selections. A standard alphanumeric keypad is used for input
of text and numbers.
Acceptable Gel CombinationsUnacceptable Gel Combinations
1212
1 mini format gel2 mini or 1 midi
format gels
format gels
2 mini or 1 midi
format gels
format gels
1 mini format gel
6 | Trans-Blot Turbo Transfer System
Page 15
Equipment and Reagents Overview
Fig. 5. Trans-Blot Turbo System user interface.
From the Home menu, three modes are available:
■
NEW mode. Create, run, and save a new protocol for your protein of interest
■
LIST mode. Select from either a list of Bio-Rad preprogrammed protocols optimized for
a variety of protein and gel types or a list of user-defined protocols
■
TURBO mode. Quick access menu for transfers of mixed MW proteins (MW 5–150 kD).
Designed for efficient transfer of a wide variety of proteins over a broad range of
molecular weights
Trans-Blot Turbo Transfer System | 7
Page 16
Equipment and Reagents Overview
8 | Trans-Blot Turbo Transfer System
Page 17
3 Using the Trans-Blot Turbo
Transfer System
The Turbo protocol, when combined with Trans-Blot Transfer Packs, provides highly efficient
transfer for a wide variety of proteins. However, run conditions may need to be adjusted for
a particular protein of interest. Protocols optimized for different protein and gel types are
available in the Bio-Rad preprogrammed protocols in LIST mode.
The system may also be used with Trans-Blot Turbo RTA Transfer Kits, which offer the same
transfer speed and efficiency as the transfer packs, or superior performance to traditional
semi-dry blotting consumables such as extra-thick filter paper and Towbin transfer buffer
(speed and efficiency will be similar to standard semi-dry blotting techniques). Please refer
to the Bio-Rad Protein Blotting Guide, bulletin 2895, for more information on the methods of
electrophoretic transfer. The Protein Blotting Guide can be downloaded as a PDF from
bio-rad.com, or call your local Technical Support number listed at the end of this document
and request bulletin 2895.
3.1 Transfer Using Transfer Packs
Recommendations:
■
Wear gloves at all times during the blotting process to prevent contamination of the gels
or membranes
■
Transfer gels immediately after electrophoresis without an equilibration step
■
The transfer membranes and stacks of the transfer packs are prewetted and do not
require any further pretreatment. Open the transfer packs immediately before use to
avoid drying the membrane
■
Use only the appropriate combinations of gels, as shown in Table 1
■
If you need to move the membrane, carefully use flat tweezers or forceps
Select the appropriate transfer pack for your application according to Table 2. These packs
are stable for several months at room temperature. However, to extend their shelf life, store
Trans-Blot Turbo Tranfer System | 9
Page 18
Using the Trans-Blot Turbo Transfer System
them at 4ºC. The transfer membranes and stacks of the transfer packs are prewetted and
do not require any further pretreatment (for example, wetting of the PVDF membrane).
Table 2. Transfer pack options.
0.2 µm Nitrocellulose Membrane,
Gel Size, cm
For one mini format gel, 7.0 x 8.517041581704156
For one midi format gel or two min
format gels, 8.5 x 13.5
catalog number
17041591704157
0.2 µm PVDF Membrane,
catalog number
3.1.1 Mini Transfer Pack (Single Mini Gel) Setup
The packaging for the mini transfer pack has the two ion reservoir stacks in the left and
right wells of the tray. The right well contains the stack for the anode (below the gel) and is
layered with the transfer membrane (nitrocellulose or PVDF). The left well contains the stack
for the cathode (above the gel) (Figure 6). Use the finger wells of the tray to access the
stacks and easily lift them from the tray. The text above each finger well signifies the stack
location in the cassette. The stack and membrane in the right finger well, labeled Bottom (+)
is placed on the anode in the cassette base. The stack in the left finger well, labeled Top (–)
is placed on top of the transfer gel, closest to the cathode.
Fig. 6. The anode stack with the membrane is on the right, while the cathode stack is on the left.
3.1.2 Midi Transfer Pack (Single Midi Gel or Two Mini Gels) Setup
The packaging for the midi transfer pack has a tray containing the two ion reservoir stacks
placed on top of each other in the tray. The ion reservoir stack on the top has a tab on
the right side and is for the anode (below the gel). This stack is layered with the transfer
membrane. The ion reservoir stack for the cathode (above the gel) is located below the
membrane and anode stack and has a tab on the left side (Figure 7). Use the finger wells
and tabs to access the appropriate stack and lift it from the tray. As with the mini trays, the
text above the finger wells signifies the stack location in the cassette. The top stack and
membrane, with the tab on the right, is labeled Bottom (+) and is placed on the anode in
the cassette base. The stack below the anode stack, with the tab on the left, is labeled Top
(–) and is placed on top of the transfer gel, closest to the cathode. For transferring two mini
10 | Trans-Blot Turbo Transfer System
Page 19
Using the Trans-Blot Turbo Transfer System
gels simultaneously, arrange the gels so that the foot of the gels (low molecular weight side)
face each other on the membrane. The gels will have to be placed longitudinally on the
stack (Figure 12).
Fig. 7. The anode stack with the membrane is on the top while the cathode stack lies beneath the anode stack and membrane.
The transfer packs are vacuum sealed. Check that the membrane does not lift away with
the foil packaging lid. If necessary, hold the edge of the membrane and peel while opening
the pack.
Note: The base of the tray is PETG plastic and can be recycled according to local
regulations for class 1 plastics.
AB
Fig. 8. Open cassette. A, lid with cathode (–) on the underside; B, base with anode (+).
Trans-Blot Turbo Transfer System | 11
Page 20
Using the Trans-Blot Turbo Transfer System
3.2 Placing a Transfer Pack into the Cassette
See Figures 10 and 11 for step-by-step instructions on placing a transfer pack into a
cassette.
■
The membranes and stacks are designed to be lifted from the tray and placed in the
cassette. Do not invert the stacks when removing them from the tray
■
The assembled transfer packs should be reasonably centered and flat within the base
of the cassette (anode)
■
We do not recommend moving the stacks once they are placed, as this may introduce
air between the layers. If the stack must be moved after placement, use a blot roller to
expel any trapped air bubbles
■
Each stack uses multiple layers of material for maximum transfer efficiency. Removing
stack layers is not advised
■
Refrain from adding any extra transfer buffer to the cassette, as saturated transfer
stacks provide ample transfer buffer.
Figure 9 shows the proper final assembly of the blotting sandwich in the cassette and is
applicable to all transfer packs.
■
If the gel and ion reservoir stacks are layered out of order during assembly of the
blotting sandwich, carefully disassemble the components so as not to damage the
membrane. Reassemble the stack and use the blot roller to remove air bubbles
between the components
Top (–) cassette
electrode (cathode)
Top ion reservoir
stack
Gel
Blotting membrane
Bottom ion reservoir
stack
Bottom(+) cassette
Fig. 9. Proper layering of the assembled transfer pack.
12 | Trans-Blot Turbo Transfer System
Page 21
Using the Trans-Blot Turbo Transfer System
Lay the ion reservoir stack with
the membrane (anode stack) in
the center of the cassette base.
Ensure that the stack is not
overlapping the green rubber
molding in the base.
Carefully align the gel on the
membrane. If necessary, gently
use the blot roller to remove air
bubbles between the gel and
membrane. If transferring two mini
gels, place them on the membrane
so that the feet of the gels are
facing toward each other.
Mini Transfer Pack
(for one mini format gel)
Midi Transfer Pack
(for one midi format gel or
two mini format gels)
Gently place the second ion
reservoir stack (cathode stack)
on the gel.
Use the blot roller to remove any
air bubbles in the assembled
transfer pack and provide
consistent contact between
the layers.
Fig. 10. Assembling the mini format
transfer pack.
Fig. 11. Assembling the midi format
transfer pack.
Trans-Blot Turbo Transfer System | 13
Page 22
Using the Trans-Blot Turbo Transfer System
A B C
Fig. 12. Suggested placement of assembled transfer stacks in a cassette. A, midi stack and gel placement; B, mini stack and
gel placement; C, two mini gels on a midi stack.
Set the transfer stack close to the center of the cassette. A midi stack only fits along the
long axis of the cassette. It is not necessary to move an assembled transfer pack if it is only
slightly off center in the cassette.
Once the stacks are positioned in the cassette base, place the cassette lid on the base. The
lid is reversible, but ensure that the electrical contacts fit closely into the slots in the base.
Press the lid down firmly and turn the dial clockwise to engage the lid pins into the locking
slots.
Slide the cassette (with the dial facing up) into one of the Trans-Blot Turbo instrument bays
until it makes contact with the magnetic interlock in the back of the instrument tub and
you hear a click. The cassette can be inserted into the bays with or without power to the
system.
The bays are labeled A (top) and B (bottom), and cassettes can be inserted interchangeably
into the bays. There is no requirement for both cassettes to be inserted; one bay (either A or
B) can be left empty when a protocol is run.
The instrument is now ready to begin a transfer protocol. Refer to the following sections for
details on transfer using preprogrammed and user-defined protocols.
Note: A transfer for one cassette can be started while you are assembling a second
cassette. The second transfer can be started independently as long as the same protocol is
being used for both.
3.3 Setting Up Transfer Using RTA Transfer Kits
Trans-Blot Turbo RTA Transfer Kits provide the same reagents and consumables available in
the transfer packs, but in a ready-to-assemble format. The protocol for using them is similar
to that used for the transfer packs, except membranes and transfer stacks must be soaked
in Trans-Blot Turbo Transfer Buffer before use. Do not equilibrate the gel before transfer.
1. Prepare the Trans-Blot Turbo Transfer Buffer according to instructions on the bottle.
2. Wet and equilibrate the membrane and transfer stacks:
■
Nitrocellulose membrane — immerse in 30 ml of 1x transfer buffer and equilibrate for
2–3 min
■
PVDF and LF PVDF membranes — immerse in 100% methanol or ethanol (use
reagent grade ~85% or molecular biology grade ~95–98% purity) until the membrane
14 | Trans-Blot Turbo Transfer System
Page 23
is translucent. Transfer to a soaking tray containing 30 ml of 1x transfer buffer, ensure
the membrane is submerged, and equilibrate 2–3 min
■
Transfer stacks
− Midi stacks — immerse two stacks separated by blue sheets in two soaking trays,
each containing 50–70 ml of transfer buffer for 2–3 min
− Mini stacks — immerse two stacks separated by blue sheets side-by-side in a
soaking tray containing 50–70 ml of transfer buffer for 2–3 min
3. Follow the procedure described in Figures 9–10 above for assembling the transfer
sandwiches in transfer cassettes, followed by an additional step: once assembled,
remove excess transfer buffer by inverting the cassette base, with the assembled stack
carefully held in place, into a waste container. Once the excess buffer is removed, place
the cassette lid on the base and proceed with the transfer step (section 3.4).
3.4 Transfer Protocols
1. Turn on the Trans-Blot Turbo System using the switch located on the right side of the
instrument. After an initial Boot screen, the system will proceed to the Home screen.
Using the Trans-Blot Turbo Transfer System
2. On the Home screen you can use the navigation buttons to choose your protocol:
■
New — set up a new protocol (see section 3.5)
■
List — select a preprogrammed protocol (see section 3.4.1)
■
Turbo — select Turbo program (see section 3.4.2)
3. After selecting your protocol press the navigation button that corresponds to A:RUN for
the cassette in the upper bay or B:RUN for the cassette in the lower bay (Figure 13).
A beep will sound to signal the start of the transfer for the chosen cassette. If you are
running both cassettes, press the button to begin one transfer (either A:RUN or B:RUN),
and then press the other button at any time during the run to immediately start the
second cassette. Both cassettes must use the same protocol.
Gel formatProtocal name
Program settings
Star indicates an active run
Transfer conditions
during run (real time)
Fig. 13.
Transfer Turbo Run screen.
Trans-Blot Turbo Transfer System | 15
Page 24
Using the Trans-Blot Turbo Transfer System
4. The protocol will run automatically. The screen will display the conditions of the transfer
and the progress of the run.
Note: A run can be paused by pressing the corresponding A: STOP or B: STOP navigation
button during the run. The user has a choice to continue from the time point of the pause,
restart the run from the beginning, or terminate the run (Figure 14.)
Fig. 14. Trans-Blot Turbo System Terminate Run screen.
When the transfer protocol is complete, the screen will display RUN COMPLETE, and
an alarm will be heard (Figure 15). Skip to section 3.4.5, Disassembly and Removal
of Membrane.
Fig. 15. Trans-Blot Turbo Run Complete screen.
3.4.1 Transfer Using Preprogrammed Protocols
All steps are identical to those in section 3.4 except for step 2, which is described here.
On the Home screen press the navigation button that corresponds with LIST (Figure 12)
to choose between the Bio-Rad preprogrammed protocols and user-defined protocol lists
(Figure 17).
3.4.2 Transfer Using the Turbo Protocol
The Turbo button immediately accesses the MIXED MW program. It is designed for efficient
transfer of a wide variety of proteins over a broad range of molecular weights. The same
parameters can be also accessed in the Bio-Rad preprogrammed protocols under LIST >
BIO-RAD > MIXED MW.
1. Turn on the Trans-Blot Turbo System using the switch located on the right side of the
instrument. After an initial Boot screen, the system will proceed to the Home screen.
16 | Trans-Blot Turbo Transfer System
Page 25
Using the Trans-Blot Turbo Transfer System
2. On the Home screen you can use the navigation buttons to choose your protocol.
3. After selecting Turbo, there will be a selection screen for the number, size, and type of
gel to be transferred. Use the corresponding selection button to choose the option for
the combination of gels in the run. The option varies the current for the run (1.3 A for a
single mini format gel, 2.5 A for a single midi or two mini format gels). Mini-PROTEAN
TGX Gels can be transferred using the “1 Mini Gel” protocol or the “1 Mini-PROTEAN
TGX” protocol, which transfers a single Mini-PROTEAN TGX Gel in 3 min.
Follow steps 3 and 4 from section 3.4.
3.4.3 Preprogrammed Protocols
Pressing the Bio-Rad preprogrammed protocols button accesses the protocols described in
Table 3.
■
STANDARD SD provides typical semi-dry transfer conditions for use with conventional
semi-dry western blotting consumables (see section 3.7)
■
1.5 mm GEL uses a longer transfer time (10 min) for more efficient transfer when using
1.5 mm thick gels
■
HIGH MW is optimized for more efficient transfer of large proteins (>150 kD)
■
LOW MW is optimized for more efficient transfer of small proteins (<30 kD)
■
MIXED MW is for efficient transfer of proteins over a broad range of molecular weights
(5–150 kD). This protocol is also accessed via the Turbo navigation button
■
1 Mini-TGX is an ultrafast protocol that will transfer a single Mini-PROTEAN TGX Gel
with mixed MW proteins (5–150 kD) in 3 min with excellent efficiency
Table 3. Bio-Rad preprogrammed protocols.
2 Mini Format Gels or
Protocol NameMW, kDTime, min
Standard SDAny30Up to 1.0 A; 25 V
1.5 mm GELAny10
High MW>15010
Low MW<305
Mixed MW (Turbo)5–1507
1 Mini-TGX5–1503–2.5 A, up to 25 V
1 Midi Format Gel (per
cassette)
2.5 A, up to 25 V 1.3 A, up to 25 V
1 Mini Format Gel (per
cassette)
Trans-Blot Turbo Transfer System | 17
Page 26
Using the Trans-Blot Turbo Transfer System
3.4.4 User-Dened Protocols
Selecting USER-DEFINED protocols accesses the protocols that have been previously saved
by users (Figure 16). Use the top and bottom selection buttons to scroll to the appropriate
choice. Push RUN to go to the system run screen where a transfer can be started.
Fig. 16. User protocols screen.
3.4.5 Disassembly and Removal of the Membrane
Note: Use caution when removing a cassette from the unit after a transfer run. The cassette
may be warm.
1. Remove the cassette from the bay by pulling it straight out of the instrument. The
LCD menu screen will automatically return to the protocol screen that has just been
completed. The system is ready for another run. If a different protocol or combination
of gels will be blotted in the next run, refer to Table 1 for the appropriate combinations
of gels that can be used in a single run, and proceed to the Home menu to select the
correct protocol.
2. Unlock the cassette by turning the dial counterclockwise to the Unlock position.
3. Disassemble the blotting sandwich and place the membrane in a suitable container.
If you are using a PVDF membrane, place it immediately into a storage solution (for
example, deionized water or blocking or staining solution) as the membrane will quickly
dry out. If a PVDF membrane requires rewetting, dip it in methanol or ethanol until
uniformly opaque, then wash with deionized water.
4. Discard the ion reservoir stacks after one use; do not attempt to reuse them.
5. Empty residual liquid from the cassette. If no additional transfer will be performed
immediately, rinse the base and lid of the cassette with deionized water and dry them
with a paper towel.
18 | Trans-Blot Turbo Transfer System
Page 27
6. Turn off the Trans-Blot Turbo System with the power switch if it is no longer required.
Fig. 17. Protocol options.
3.5 Setting Up a New Protocol
Assembly of the blotting sandwich is described in sections 3.1 and 3.2. This section
describes setting up a new protocol.
On the Home screen (Figure 12), press the navigation button that corresponds with NEW.
Using the Trans-Blot Turbo Transfer System
Pressing NEW will bring up the Edit screen (Figure 18), where the user can set the three
parameters of current (maximum 2.5 A), voltage (maximum 25 V), and time (maximum
90 min) using the alphanumeric keypad. The user can toggle between setting a constant
voltage (CONST V) or a constant current (CONST A) using a navigation button. After the
settings have been determined, a prompt will appear to name and save the protocol for
subsequent use, if desired. A saved protocol can later be accessed under LIST > USERDEFINED protocols. Saved protocols can be deleted when no longer required.
Selection buttons for
voltage, current, and
time
Toggle between CONST A and CONST V
Fig. 18. Edit NEW protocol screen.
To prevent damage to the cassette or instrument, the Trans-Blot Turbo System will terminate
a transfer if 30 watts per hour is exceeded over the course of the run. Depending on the
initial run conditions when designing a protocol, the system may display a message warning
that the 30 watts/hour limit may be exceeded during the run.
Trans-Blot Turbo Transfer System | 19
Page 28
Using the Trans-Blot Turbo Transfer System
3.6 Transfer Using Traditional Semi-Dry Consumables
Typical procedure utilizing the Trans-Blot Turbo System and conventional semi-dry western
blotting consumables is detailed below:
1. Equilibrate the gel in Towbin transfer buffer (25 mM Tris, 192 mM glycine pH 8.3,
20% MeOH) for 10 min.
2. Soak two pieces of extra-thick (2.4 mm) filter paper in transfer buffer. Six pieces of thick
(0.8 mm) filter paper can be used if extra-thick paper is not available.
3. While the gel is equilibrating, prepare a transfer membrane. Wet a nitrocellulose
membrane briefly in transfer buffer or PVDF membrane in methanol or ethanol for
30 sec, then wash in water for 1–2 min, and equilibrate in transfer buffer for at least 10
min with agitation.
4. Assemble the transfer sandwich on the cassette base (anode) by placing one piece
of wet extra-thick or three pieces of wet thick filter paper on the bottom, then the
membrane, the gel, and finally, the remainder of the wet filter paper on top. Use the blot
roller to remove air from between the assembled layers (Figure 19).
5. Once the stacks are positioned in the cassette base, place the cassette lid on the base.
The lid is reversible, but ensure that the electrical contacts fit closely into the slots in the
base. Press the lid down firmly and turn the dial clockwise to engage the lid pins into the
locking slots.
6. Load a second cassette if desired. Refer to Table 1 for the appropriate combinations of
gels that can be used in a single run.
7. Slide the cassette (with the dial facing up) into the bay until it makes contact with the
magnetic interlock and you hear a click. Cassettes can be inserted into the bays in any
order, with or without power to the system.
8. Select the LIST button from the Home menu and the STANDARD SD transfer protocol
from the Bio-Rad preprogrammed protocols, or the user-defined protocol of choice.
9. To initiate the run, press the navigation button that corresponds to A:RUN for the
cassette in the upper bay or B:RUN for the cassette in the lower bay.
Top (–) cassette
electrode (cathode)
Top ion reservoir
Blotting membrane
Bottom ion reservoir
Bottom(+) cassette
electrode (anode)
stack
Gel
stack
Fig. 19. Proper assembly of blotting sandwich using traditional consumables.
20 | Trans-Blot Turbo Transfer System
Page 29
For further information, refer to the Bio-Rad Protein Blotting Guide, bulletin 2895, for
detailed information. Bulletin 2895 can be downloaded as a PDF from bio-rad.com, or call
your local Technical Support number listed at the end of this document.
3.7 Optimizing Transfer Conditions
The following techniques, alone or in combination, will increase transfer efficiency:
■
Use a low-percentage gel or a gradient gel. High-percentage gels retard protein transfer,
especially with large proteins
■
High molecular weight proteins may require increased transfer times, particularly when
using thick gels. Increase the transfer time or power conditions
■
With long transfer times or high power conditions, some very low molecular weight
proteins may transfer through the membrane to the lower ion reservoir stack. Use a
shorter transfer time or reduce power conditions for the most efficient transfers
■
Use the blot roller to remove any air bubbles when assembling the transfer sandwich.
Air bubbles between layers of the assembled sandwich will prevent protein transfer,
producing blank spots on the membrane.
Using the Trans-Blot Turbo Transfer System
Table 4. Recommended power conditions for transfer using the Trans-Blot Turbo System.
Single Mini GelTwo Mini Gels or One Midi Gel
With Turbo Transfer Packs25 V, 1.3 A, 7 min25 V, 2.5 A, 7 min
With Standard Semi Dry Components 25 V, 1.0 A, 30 min
Refer to the Bio-Rad Protein Blotting Guide, bulletin 2895, for more information on optimizing
electrophoretic transfer. The Protein Blotting Guide can be downloaded as a PDF from
bio-rad.com, or call your local Technical Support number listed at the end of this document.
Trans-Blot Turbo Transfer System | 21
Page 30
Using the Trans-Blot Turbo Transfer System
22 | Trans-Blot Turbo Transfer System
Page 31
4 Appendices
Appendix A. Maintenance
The Trans-Blot Turbo System requires little maintenance. After you finish using the TransBlot Turbo System, rinse each cassette, base, and lid in deionized water to remove residual
salts and to prevent salt buildup. Air-dry or use a paper towel to dry the cassette.
For each transfer cassette there are four connection contacts located at the back of the
instrument base cavity (Figure 20). Over time it is possible that buffer or other contaminants
have been exposed to these contacts causing them to stick in the depressed position.
With the unit unplugged, use isopropyl alcohol and a cotton swab or isopropyl wipes to
clean these contacts. Push on each contact to ensure that it depresses freely. The contacts
should be checked and cleaned monthly as part of standard equipment maintenance
procedures.
Fig. 20. Connection contacts inside the instrument base cavity. Your instrument will have either leaf spring contacts (left) or
pogo pin contacts (right).
Clean any accidental buffer spills in the instrument bays with a dry towel.
The cassettes and instrument casing can be cleaned with water and mild detergent; do not
use abrasives or organic solvents.
Trans-Blot Turbo Transfer System | 23
Page 32
Appendices
Periodically unplug the instrument and wipe the instrument casing with a moist cloth or
paper towel. Ensure that the electrode contacts in the bays are clean. Electrode plates
should be occasionally cleaned with deionized water or a mild detergent to reduce salt
buildup. If the cathode plate has excessive salt buildup, 7% acetic acid can be used to
remove the residue.
Occasionally check that the cooling fan vent is free of debris and dust.
Clean and inspect the instrument and cassette after each use for damage. Contact Bio-Rad
technical support if any damage is noted.
Appendix B. Troubleshooting
For more information on troubleshooting, see the Bio-Rad Protein Blotting Guide,
bulletin 2895, which can be downloaded as a PDF from bio-rad.com, or call your local
Technical Support number listed at the end of this document.
Symptom/ProblemSuggested Corrections
Poor or no electrophoretic transfer;
excessive protein remains in the gel
(as detected by staining the gel after
transfer)
Poor or no electrophoretic transfer;
little or no protein remains in the gel
(as detected by staining the gel after
transfer)
Blot sandwich exceeds 70ºC after
transfer
Transfer time may be too short. High molecular weight proteins
(>150 kD) may require increased transfer time.
Change the percentages of the components for handcast gels. The
Bio-Rad Protein Blotting Guide, bulletin 2895, has extensive information
on casting acrylamide gels.
The wrong protocol relative to the gel combination may have been used
for the transfer. Refer to Table 1 for acceptable combinations of gels
that can be used in a single run.
There may be an excessive amount of protein loaded on the gel.
Reduce the amount of protein on the gel.
While the preprogrammed protocols efficiently transfer most proteins,
some proteins may require further optimization. Increase the transfer
time in 1 min increments. Change voltage and current if required.
The blotting sandwich may be assembled in the wrong order. See
Figure 9 for the correct assembly.
The detection system may not be working or may not be sensitive
enough. Include proper antigen controls to test detection kit sensitivity
or use a total protein stain or stain-free technology to detect protein
transfer to the membrane prior to immunodetection.
An incorrect program for the gel combination was used. Refer to
Table 1 for the acceptable combinations of gels that can be used in a
single run.
Reduce run current, voltage, or time.
Cassette does not stay in place in the
instrument bay
POWER FAILURE! PRESS ANY KEY TO
CONTINUE is displayed on the screen
24 | Trans-Blot Turbo Transfer System
Make sure that the dial of the cassette lid is in the Lock position and the
lid pins are engaged. The dial should be facing up when the cassette is
inserted into the bay.
The magnetic interlock may be damaged or have failed. Contact
Technical Support for assistance.
Power to the instrument was interrupted during a run and the run was
aborted. It will need to be restarted. Shut off the power switch, correct
the power outage, and restart the instrument.
Page 33
Appendices
Swirls or missing bands; diffuse
transfers
NO LOAD DETECTED is displayed on
the Run screen
There may be poor contact between the membrane and the gel. Air
bubbles (or excess buffer when using traditional semi-dry consumables)
remain between the membrane and gel. Use the blot roller to ensure
that air bubbles or excess buffer are removed.
There may be poor contact between the assembled sandwich and the
electrodes (when conventional semi-dry consumables are used). Use
extra-thick filter paper or add additional sheets of thick buffer-wetted
filter paper to improve contact.
The membrane may have dried. White spots on a nitrocellulose or
PVDF membrane indicate dry areas where protein will not bind.
Power conditions may be too high for your particular gel. Adjust voltage
and current conditions.
The gel electrophoresis may be at fault. Artifacts of electrophoresis
may be produced by poor polymerization, inappropriate running
conditions, contaminated buffers, sample overload, etc. Consult your
electrophoresis manual for more details.
When using a single mini transfer pack, the assembled transfer pack
may have been placed at the extreme edge of the cassette bottom.
Ensure that assembled transfer packs are placed near the center of the
cassette base.
When two mini gels are transferred simultaneously on a midi format
stack, ensure the gels are arranged foot-to-foot (low molecular weight
portions of the gels are toward each other).
The unit is detecting the presence of a cassette, but current is not
passing between the electrode plates. Ensure that the blotting
assembly is positioned properly and is thick enough to contact both the
anode and cathode plates.
If using a custom buffer, the buffer may have an ion capacity that is
too low or too high for efficient transfer. Check the concentrations of
components in the buffer.
The electrical contacts on the cassette or instrument may be dirty. With
the instrument unplugged, clean the contacts with a cotton swab or
wipe moistened with isopropyl alcohol.
Trans-Blot Turbo Transfer System | 25
Page 34
Appendices
Overview of the Three Protocol Modes
TURBO: 1 or 2 mini format gels, or 1 midi format gel in 7 min;
1 Mini-PROTEAN TGX Gel in 3 min.
26 | Trans-Blot Turbo Transfer System
Page 35
LIST: Bio-Rad preprogrammed protocols or user-defined protocols.
Appendices
Trans-Blot Turbo Transfer System | 27
Page 36
Appendices
NEW: User can edit protocol parameters and name, save, and run the protocol.
28 | Trans-Blot Turbo Transfer System
Page 37
Consumables and Related Products
Consumables and Related Products
Catalog # Description
Trans-Blot Turbo Accessories
1704151 Trans-Blot Turbo Cassette, single
1704152 Trans-Blot Turbo Base
Trans-Blot Turbo Starter Kits
Starter kits include a Trans-Blot Turbo instrument, a Trans-Blot Turbo Transfer Pack corresponding to the listed modifier, Precision Plus Protein
WesternC Standards, and Precision Protein StrepTactin-HRP Conjugate.
1704156 Trans-Blot Turbo Transfer Pack, 0.2 µm PVDF membrane, mini format, pack of 10
1704157 Trans-Blot Turbo Transfer Pack, 0.2 µm PVDF membrane, midi format, pack of 10
1704158 Trans-Blot Turbo Transfer Pack, 0.2 µm nitrocellulose membrane, mini format, pack of 10
1704159 Trans-Blot Turbo Transfer Pack, 0.2 µm nitrocellulose membrane, midi format, pack of 10
Trans-Blot Turbo RTA Transfer Kits
1704270 Trans-Blot Turbo RTA Transfer Kit, nitrocellulose, mini format (7.0 x 8.5 cm), for 40 blots
1704271 Trans-Blot Turbo RTA Transfer Kit, nitrocellulose, midi format (8.5 x13.5 cm), for 40 blots
1704272 Trans-Blot Turbo RTA Transfer Kit, PVDF, mini format (7.0 x 8.5 cm), for 40 blots
1704273 Trans-Blot Turbo RTA Transfer Kit, PVDF, midi format (8.5 x 13.5 cm), for 40 blots
1704274 Trans-Blot Turbo RTA Transfer Kit, LF PVDF, mini format (7.0 x 8.5 cm), for 40 blots
1704275 Trans-Blot Turbo RTA Transfer Kit, LF PVDF, midi format (8.5 x 13.5 cm), for 40 blots
Related Products
1705060 Clarity Western ECL Substrate, 200 ml, includes 100 ml luminol/enhancer and 100 ml peroxide solution
1705061 Clarity Western ECL Substrate, 500 ml, includes 250 ml luminol/enhancer reagent and 250 ml peroxide solution
1705062 Clarity Max Western ECL Substrate, 100 ml, includes 50 ml luminol/enhancer reagent and 50 ml peroxide solution
1705047 Immun-Star Goat Anti-Mouse (GAM)-HRP Conjugate
1705046 Immun-Star Goat Anti-Rabbit (GAR)-HRP Conjugate
1705010 Immun-Star GAM-AP Detection Kit, includes substrate, enhancer, GAM antibody
1705011 Immun-Star GAR-AP Detection Kit, includes substrate, enhancer, GAR antibody
1708238 Amplified Opti-4CN Substrate Kit, includes substrate, diluent, blocker, PBS
1706515 Blotting-Grade Conjugate, goat anti-rabbit IgG (H+L)-HRP, 2 ml
1706516 Blotting-Grade Conjugate, goat anti-mouse IgG (H+L)-HRP, 2 ml
1706404 Blotting-Grade Blocker, nonfat dry milk, 300 g
Trans-Blot Turbo Transfer System | 29
Page 38
30 | Trans-Blot Turbo Transfer System
Page 39
Page 40
Bio-Rad
Laboratories, Inc.
Life Science
Group
10000071567 Ver
Web site bio-rad.com USA 1 800 424 6723 Australia 61 2 9914 2800 Austria 43 01 877 89019 Belgium 32 03 710 53 00
Brazil 55 11 3065 7550 Canada 1 905 364 3435 China 86 21 6169 8500 Czech Republic 36 01 459 6192 Denm ark 45 04 4 52 10 00
Finland 35 08 980 422 00 France 33 01 479 593 0 0 German y 49 089 3188 4393 Hong Kong 8 52 2789 3300 Hungar y 36 01 459 6190
India 91 124 4029300
The Netherlands 31 0 318 540 6 66 New Zealand 64 9 415 22 80 N orway 47 0 233 841 30 Poland 36 01 459 6191 Portugal 3 51 21 472 77 17
Russia 7 495 721 14 04 Singapore 65 6415 3188 South Africa 36 01 459 6193 Spain 34 091 49 0 6 580 Swed en 46 08 555 127 00
Switzerland 41 0617 17 9 555 Taiwan 886 2 2578 7189 Thailand 66 2 651 8311 United Arab Emirates 971 4 8187300
United Kingdom 44 01923 47 1301
H (10020688) US/EG
Israel 972 03 963 6050
Italy 39 02 494866 00 Japan 81 3 6361 7000 Kore a 82 2 3473 4460 Mexico 52 55 5 488 7670
18-0598 0718 Sig 0118
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.