Bio-Rad 10000070438 Console Operation & Maintenance Manual

Page 1
CFX96™ Dx and
CFX96 Deep Well Dx Systems
Operation Manual
1845097-IVD 1844095-IVD 1841000-IVD 12007917
Manual revision: January 2019 Software revision: 3.1
ETL LISTED
UL Std. 61010-1 UL Std. 61010-2-010 UL Std. 61010-2-101
3120330
UL Std. 61010-2-081
CERTIFIED TO
CAN/CSA Std. C22.2 NO. 61010-1-12 CAN/CSA Std. C22.2 NO. 61010-2-010 CAN/CSA Std. C22.2 NO. 61010-2-101 CAN/CSA Std. C22.2 NO. 61010-2-081:2015
Page 2
Page 3
CFX96™ Dx and CFX96 Deep Well Dx
Systems
Operation Manual
Version 3.1
Page 4
Bio-Rad Technical Support
For technical assistance, please contact your local Bio-Rad representative.
Notice
No part of this publication may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopy, recording, or any information storage or retrieval system, without permission in writing from Bio-Rad.
Bio-Rad reserves the right to modify its products and services at any time. This guide is subject to change without notice. Although prepared to ensure accuracy, Bio-Rad assumes no liability for errors or omissions, or for any damage resulting from the application or use of this information.
The following are trademarks of Bio-Rad Laboratories, Inc.: Bio-Rad, CFX Manager, CFX96, C1000.
Bio-Rad thermal cyclers and real-time systems are covered by one or more of the following U.S. patents or their foreign counterparts owned by Eppendorf AG: U.S. Patent Numbers 6,767,512 and 7,074,367.
Copyright © 2018 by Bio-Rad Laboratories, Inc. All rights reserved.
Page 5
Table of Contents
Safety and Regulatory Compliance 11
Safety Warning Labels 11
Safe Use Specifications and Compliance 12
Regulatory Compliance 12
Hazards 13
Biohazards 13
Chemical Hazards 15
Explosive or Flammability Hazards 15
Electrical Hazards 15
Transport 15
Battery 15
Disposal 16
Warranty 16
Chapter 1 Introduction 17
CFX Dx PCR Detection Systems 17
Intended Use 18
Finding Out More 18
Chapter 2 Setting Up the C1000 Dx Thermal Cycler 19
Site Requirements 19
Bench Space Requirements 19
Environment Requirements 20
Power Requirements 20
System Overview 21
Front View 21
Back View 22
Optical Reaction Modules 23
Recommended Sample Volumes 23
Installing the C1000 Dx Thermal Cycler 24
Operation Manual | iii
Page 6
Table of Contents
Unpacking and Setting Up the C1000 Dx Thermal Cycler 24
Attaching the Optical Reaction Module 25
Removing the Shipping Screw 26
Loading Sample Plates 27
Detecting Connected Instruments 29
Detaching the Reaction Module 30
Shutting Down the C1000 Dx Thermal Cycler 30
Chapter 3 Installing CFX Manager Dx Software 31
System Requirements 32
Installing CFX Manager Dx Software 33
Detecting Connected Instruments 33
Software Files 34
Recommended Cybersecurity Measures 34
Chapter 4 The Workspace 37
The Home Window 38
The Startup Wizard 39
The Protocol Editor Window 40
The Plate Editor Window 41
The Data Analysis Window 42
Chapter 5 The Home Window 43
The Home Window 44
File Menu Commands 45
View Menu Commands 45
User Menu Commands 46
Run Menu Commands 46
Tools Menu Commands 47
Help Menu Commands 47
Toolbar Commands 48
The Startup Wizard 49
Status Bar 49
Detected Instruments Pane 50
Viewing the Properties of an Instrument 53
Before You Begin 55
iv | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 7
Table of Contents
Setting User Preferences 55
Creating a Reaction Master Mix 69
Calibrating New Dyes 72
Chapter 6 Creating Protocols 75
Protocol Editor Window 76
File Menu Commands 76
Settings Menu Command 77
Tools Menu Commands 77
Toolbar Commands 77
Protocol Editing Controls 78
Creating a Protocol in the Protocol Editor 80
Opening a New Protocol File in the Protocol Editor 80
Opening an Existing Protocol in the Protocol Editor 81
Setting Up a New Protocol 82
Adding Steps to a Protocol 84
Inserting a Gradient Step 84
Inserting a GOTO Step 86
Inserting a Melt Curve Step 86
Adding or Removing a Plate Read Step 88
Changing Step Options 88
Deleting a Step 89
Copying, Exporting, or Printing a Protocol 89
Creating a Protocol with the Protocol AutoWriter 90
Using the Ta Calculator 92
About the Ta Calculator 92
Chapter 7 Preparing Plates 97
Plate Editor Window 98
File Menu Commands 99
Settings Menu Commands 99
Editing Tools Menu Commands 99
Toolbar Commands 100
Creating a Plate File Using the Plate Editor 101
Opening a New Plate File in the Plate Editor 101
Opening an Existing Plate File in the Plate Editor 103
Operation Manual | v
Page 8
Table of Contents
Setting Up a New Plate File 104
Assigning Optional Parameters to the Plate File 110
Assigning a Target to Wells 110
Assigning a Sample Name to Wells 112
Assigning Biological Sets to Wells 113
Assigning Replicate Numbers to Wells 115
Assigning a Dilution Series to Standard Sample Types 116
Copying Well Contents into Another Well 117
Adding a Note to a Well 117
Clearing Wells of All Content 118
Changing Experiment Settings 119
Creating Well Groups 121
Changing Trace Styles 124
Viewing the Plate in Spreadsheet Format 125
Creating a Plate Layout Using the Plate Setup Wizard 127
Using the Plate Setup Wizard 127
Chapter 8 Running Experiments 131
Accessing the Run Setup Window 131
The Run Setup Window 132
Protocol Tab 133
Plate Tab 135
Start Run Tab 137
Running an Experiment 138
Run Details Dialog Box 140
Run Status Tab 140
Real-time Status Tab 142
Time Status Tab 144
Performing PrimePCR Experiments 145
Chapter 9 Data Analysis Overview 147
Data Analysis Window 147
Data Analysis Toolbar 148
Data Analysis Menu Bar 149
Tab Details 152
Step Number Selector 152
vi | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 9
Table of Contents
Viewing Well Groups in Data Analysis 153
Changing Well Contents after a Run 153
Data Analysis Settings 154
Adjusting the Threshold 154
Baseline Settings 154
Analysis Mode 155
Cycles to Analyze 156
Well Selector 157
Well Selector Right-Click Menu Items 158
Temporarily Excluding Wells from Analysis 159
Charts 160
Common Right-Click Menu Items for Charts 160
Copying Chart Data to the Clipboard 160
Modifying the Baseline Threshold Settings 161
Sorting Target and Sample Data 162
Magnifying an Area in the Chart 163
Copying Charts into a Microsoft File 163
Spreadsheets 164
Common Right-Click Menu Items for Spreadsheets 164
Export 166
Exporting All Data Sheets 166
Creating a Custom Export File 167
Exporting to a LIMS Folder 168
Exporting Seegene-Formatted Data 168
Chapter 10 Data Analysis Details 169
Quantification Tab 170
Fluorophore Options 170
Trace Styles Dialog Box 171
Log Scale Option 172
Standard Curve Chart 173
Amplification Chart Menu Options 173
Quantification Tab Spreadsheet 174
Quantification Data Tab 175
Results Spreadsheet 175
Operation Manual | vii
Page 10
Table of Contents
Standard Curve Results Spreadsheet 177
Plate Spreadsheet 178
RFU Spreadsheet 178
Melt Curve Tab 179
Adjusting Melt Curve Data 181
Melt Curve Data Tab 182
Melt Peaks Spreadsheet 182
Plate Spreadsheet 183
RFU Spreadsheet 184
-d(RFU)/dT Spreadsheet 185
End Point Tab 186
Results Data 187
Adjusting the End Point Data Analysis 188
RFU Spreadsheet for End Point Analysis 188
Allelic Discrimination Tab 189
Adjusting Data for Allelic Discrimination 190
Chart Menu Options 191
Allelic Discrimination Spreadsheet 191
Custom Data View Tab 192
Creating a Custom Data View 193
QC Tab 194
Changing QC Criteria 194
Excluding Wells That Fail QC 194
Run Information Tab 195
Data Analysis Reports 196
Data Analysis Report Categories 197
Creating a Data Analysis Report 200
Creating Well Group Reports 201
Chapter 11 Gene Expression Analysis 203
Plate Setup for Gene Expression Analysis 203
Guided Plate Setup 204
Gene Expression Charts 205
Bar Chart 206
Sorting Target and Sample Data 208
viii | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 11
Table of Contents
Adjusting Gene Expression Data 208
Experiment Settings 210
Target Stability Value 212
Right-Click Menu Options 213
Data Spreadsheet 214
Show Details Option 214
Clustergram 217
Settings 217
Right-Click Menu Options 217
Data Spreadsheet 217
Scatter Plot 218
Settings 218
Right-Click Menu Options 218
Data Spreadsheet 218
Results 219
Gene Study 220
Inter-Run Calibration 220
Gene Study Dialog Box 220
Study Setup Tab 220
Preparing a Gene Study 221
Study Analysis Tab 222
Creating a Gene Study Report 223
Gene Study Report Categories 223
Appendix A Data Analysis Calculations 225
Reaction Efficiency 225
Relative Quantity 225
Relative Quantity When a Control Is Selected 226
Standard Deviation of Relative Quantity 226
Efficiency Corrected Cq (CqE) 227
Mean Efficiency Corrected Cq (MCqE) 227
Normalization Factor 227
Normalized Expression 228
Normalized Expression When a Control Is Selected 228
Standard Deviation for the Normalized Expression 229
Operation Manual | ix
Page 12
Table of Contents
Normalized Expression Scaled to Highest Expression Level 230
Normalized Expression Scaled to Lowest Expression Level 230
Normalized Expression Scaled to Average Expression Level 230
Standard Deviation for the Scaled Normalized Expression 231
Regulation 231
Corrected Values Formulas 232
Appendix B Managing CFX Manager Dx Users and Roles 233
Managing Users 233
Adding and Removing Users 233
Managing the Rights of Roles 235
Logging In to CFX Manager Dx Software 236
Changing Users 236
Changing User Passwords 237
Viewing Your Role and Rights 237
Appendix C LIMS Integration 239
Creating LIMS-Compatible Data Files 239
Setting Up LIMS Folder and Data Export Options 239
Creating a LIMS Protocol 240
Creating a LIMS File 241
Starting a LIMS Run 244
Exporting Data to a LIMS 244
Appendix D Troubleshooting CFX Manager Dx Software Connection Issues 245
Application Log 245
Troubleshooting 246
Power Failure 246
Retrieving Files to the CFX Manager Dx Computer 248
Installing CFX Manager Dx Software Manually 248
Reinstalling the Drivers 248
Appendix E References 249
x | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 13

Safety and Regulatory Compliance

For safe operation of the CFX96™ Dx System or CFX96 Deep Well Dx System with CFX Manager™
Dx Software, referred to as the CFX Dx system in this document, Bio-Rad strongly recommends that
you follow the safety specifications listed in this section and throughout this manual.
Important: The CFX96 Dx and CFX96 Deep Well Dx systems are approved for use as invitro
diagnostic (IVD) medical devices.

Safety Warning Labels

Warning labels posted on the instrument and in this manual warn you about sources of injury or
harm. Table 1 defines each safety warning label.
Table 1. Meaning of safety warning labels
Icon Meaning
Warning about risk of harm to body or equipment
Operating the CFX Dx system before reading this manual can constitute a personal
injury hazard. For safe use, do not operate this instrument in any manner unspecified in
this manual. Only qualified laboratory personnel trained in the safe use of electrical
equipment should operate this instrument. Always handle all components of the system
with care and with clean, dry hands.
Warning about handling biohazardous materials
When handling biohazardous samples, adhere to the recommended precautions and
guidelines and comply with any local guidelines specific to your laboratory and location.
Warning about risk of burning
A thermal cycler generates enough heat to cause serious burns. Wear safety goggles or
other eye protection at all times during operation. Always allow the sample block to
return to idle temperature before opening the lid and removing samples. Always allow
maximum clearance to avoid accidental skin burns.
Warning about risk of explosion
The sample blocks can become hot enough during the course of normal operation to
cause liquids to boil and explode.
Operation Manual | 11
Page 14
Safety and Regulatory Compliance

Safe Use Specifications and Compliance

Table 2 lists the safe use specifications for Bio-Rad’s CFX Dx real-time PCR detection systems.
The supplied shielded cables must be used with these instruments to ensure compliance with the
Class A FCC limits.
Table 2. Conditions for safe use
Usage Aspect Conditions for Safe Use
Rated input power 100–240 VAC, 50–60 Hz, 850 W max
Overvoltage category II
Fuses 10 A, 250 V, 5 x 20 mm, fast blow (qty. 2)
Environment Indoor use only
Usage temperature 15–31°C
Storage temperature –20 to 60°C
Relative humidity Up to 80% (non-condensing)
Altitude Up to 2,000 meters above sea level
Pollution degree 2

Regulatory Compliance

The CFX Dx real-time PCRdetection system has been tested and found to be in compliance with
all applicable requirements of the following safety and electromagnetic standards:
n IEC 61010-1:2010 (3rd ed.), EN61010-1:2010 (3rd ed.). Electrical equipment for
measurement, control, and laboratory use - Part 1: General requirements
n IEC 61010-2-010:2014, EN 61010-2-010:2014. Safety requirements for electrical equipment
for measurement, control, and laboratory use. Part 2-010: Particular requirements for
laboratory equipment for the heating of materials
n IEC 61010-2-081:2015, EN 61010-2-081:2015. Safety requirements for electrical equipment
for measurement, control, and laboratory use. Part 2-081: Particular requirements for
automatic and semi-automatic laboratory equipment for analysis and other purposes (includes
Amendment 1)
12 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 15
n IEC 61010-2-101:2015 (2nd ed.). Safety requirements for electrical equipment for
measurement, control, and laboratory use. Particular requirements for in vitro diagnostic (IVD)
medical equipment
n IEC 61326-1:2012 (Class A), EN 61326-1:2013 (Class A). Electrical equipment for
measurement, control, and laboratory use. EMC requirements, Part 1: General requirements
n IEC 61326-2-6:2012, EN 61326-2-6:2013 (Class A). Electrical equipment for measurement,
control, and laboratory use. EMC requirements. Particular requirements for in vitro diagnostic
(IVD) medical equipment
Important: This equipment generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the provided instructional documentation, may
cause harmful interference to radio communications. Operation of the systems in a residential
area is likely to cause harmful interference, in which case users will be required to correct the
interference at their own expense.

Hazards

Hazards
The CFX Dx real-time PCR detection system is designed to operate safely when used in the
manner prescribed by the manufacturer. If the CFX Dx real-time PCR detection system or any of its
associated components is used in a manner not specified by the manufacturer, the inherent
protection provided by the instrument may be impaired. Bio-Rad Laboratories, Inc. is not liable for
any injury or damage caused by the use of this equipment in any unspecified manner, or by
modifications to the instrument not performed by Bio-Rad or an authorized agent. Service of the
CFX Dx real-time PCR detection system should be performed only by trained Bio-Rad personnel.

Biohazards

The CFX Dx real-time PCR detection system is a laboratory product. However, if biohazardous
samples are present, adhere to the following guidelines and comply with any local guidelines
specific to your laboratory and location.
Note: No biohazardous substances are exhausted during normal operations of this
instrument.
General Precautions
n Always wear laboratory coat, laboratory gloves, and safety glasses with side shields or
goggles.
n Keep your hands away from your mouth, nose, and eyes.
n Completely protect any cut or abrasion before working with potentially infectious materials.
Operation Manual | 13
Page 16
Safety and Regulatory Compliance
n Wash your hands thoroughly with soap and water after working with any potentially infectious
material before leaving the laboratory.
n Remove wristwatches and jewelry before working at the bench.
n Store all infectious or potentially infectious material in unbreakable leak-proof containers.
n Before leaving the laboratory, remove protective clothing.
n Do not use a gloved hand to write, answer the telephone, turn on a light switch, or touch
anything that other people may touch without gloves.
n Change gloves frequently. Remove gloves immediately when they are visibly contaminated.
n Do not expose materials that cannot be properly decontaminated to potentially infectious
material.
n Upon completion of an operation involving biohazardous material, decontaminate the work
area with an appropriate disinfectant (for example, a 1:10 dilution of household bleach).
Specific IVD Precautions
n All patient samples are a potential biohazard and should be handled accordingly using
universal precautions.
n No biohazardous substances are exhausted during normal operation of this instrument.
Surface Decontamination
WARNING! To prevent electrical shock, always turn off and unplug the instrument
prior to performing decontamination procedures.
The following areas can be cleaned with any hospital-grade bactericide, virucide, or fungicide
disinfectant:
n Outer lid and chassis
n Inner reaction block surface and reaction block wells
n Control panel and display
To prepare and apply the disinfectant, refer to the instructions provided by the product
manufacturer. Always rinse the reaction block and reaction block wells several times with water
after applying a disinfectant. Thoroughly dry the reaction block and reaction block wells after
rinsing with water.
14 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 17
Important: Do not use abrasive or corrosive detergents or strong alkaline solutions. These
agents can scratch surfaces and damage the reaction block, resulting in loss of precise
thermal control.
Disposal of Biohazardous Material
Dispose of the following potentially contaminated materials in accordance with laboratory local,
regional, and national regulations:
n Clinical samples
n Reagents
n Used reaction vessels or other consumables that may be contaminated

Chemical Hazards

The CFX Dx real-time PCRdetection system contains no potentially hazardous chemical materials.
Hazards

Explosive or Flammability Hazards

The CFX Dx real-time PCRdetection system poses no uncommon hazard related to flammability or
explosion when used in a proper manner as specified by Bio-Rad Laboratories.

Electrical Hazards

The CFX Dxreal-time PCR detection system poses no uncommon electrical hazard to operators if
installed and operated properly without physical modification and connected to a power source of
proper specification.

Transport

Before moving or shipping the CFX Dx real-time PCR detection system or its optical reaction
module or its thermal cycler base, decontamination procedures must be performed. Always move
or ship the CFX Dx real-time PCR detection system and optical reaction modules in separate
containers with the supplied packaging materials, which will protect the instrument from damage. If
appropriate containers cannot be found, contact your local Bio-Rad office.

Battery

The CFX Dx system thermal cycler uses a 3 V lithium-metal coin cell battery and a 4.8 V nickel-
metal hydride rechargeable battery pack to maintain time settings and run data in the event of AC
power loss. If the time and/or run data do not remain set after the unit is turned off, it may be an
Operation Manual | 15
Page 18
Safety and Regulatory Compliance
indication that the batteries are getting weak. If this occurs, contact Bio-Rad Technical Support for
assistance.
Do not attempt to change the batteries. Contact Bio-Rad Technical Support.

Disposal

The CFX Dx real-time PCR detection system contains electrical materials; they should be disposed
of as unsorted waste and must be collected separately, according to European Union Directive
200296CE on waste and electronic equipment — WEEE Directive. Before disposal, contact your
local Bio-Rad representative for country-specific instructions.

Warranty

The CFX Dx real-time PCR detection system and its associated accessories are covered by a
standard Bio-Rad warranty. Contact your local Bio-Rad office for the details of the warranty.
16 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 19

Chapter 1 Introduction

Bio-Rad’s CFX Dx real-time PCR amplification systems for in vitro diagnostics (IVD) feature the
latest technological advances, allowing for PCRquantification with standard curve, gene
expression analysis, allelic discrimination, and end-point analysis.
The CFX Dx systems are comprised of two hardware modules and software:
n CFX96™ Dx or CFX96 Deep Well Dx Optical Reaction Module (ORM)
n C1000™ Dx Thermal Cycler
n CFX Manager™ Dx Software
When used with CFX Manager Dx software you can
n Generate immediate results with the Startup Wizard
n Enter or edit well information before, during, or after a run
n Interpret complex data and make sense of your gene expression study with tools such as
PrimePCR™ controls analysis, and the reference gene selector tool
n Prepare comprehensive reports of your real-time PCR data

CFX Dx PCR Detection Systems

Table 3 lists Bio-Rad’s IVD PCR products that ship with CFX Dx systems.
Note: A CFX Dx system ships with the CFX Manager Dx Software, the C1000 Dx Thermal
Cycler, and either the CFX96 Dx or CFX96 Deep Well Dx Optical Reaction Module.
Table 3. CFX IVD PCR Detection Systems
Catalog # Description
1845097-IVD CFX96 Dx ORM *
1844095-IVD CFX96 Deep Well Dx ORM
1841000-IVD C1000 Dx Thermal Cycler
12007917 CFX Manager Dx Software v3.1
* Optical Reaction Module
Operation Manual | 17
Page 20
Chapter 1 Introduction

Intended Use

CFX96 Dx System and CFX96 Deep Well Dx System with CFX Manager Dx Software are intended
to perform fluorescence-based PCR to detect and quantitate nucleic acid sequences. The systems
and software are intended for in vitro diagnostic use by trained laboratory technicians. The systems
are intended to be used with third party diagnostic nucleic acid tests which have been
manufactured and labeled for diagnostic purposes.

Finding Out More

This document explains how to safely set up and operate the CFX96 Dx and CFX96 Deep Well Dx
real-time PCR detection systems, which carry the CE-IVD mark. These systems are referred to as
the CFX Dx system in this document. In addition, this document explains how to use CFX Manager
Dx software with the CFX Dx system.
Tip: Click the Bio-Rad logo in the upper right corner of any CFX Manager Dx software window
to launch Bio-Rad’s website. This site includes links to technical notes, manuals, product
information, and technical support. This site also provides many technical resources on a wide
variety of methods and applications related to PCR, real-time PCR, and gene expression.
18 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 21
Chapter 2 Setting Up the C1000 Dx Thermal
Cycler
This chapter explains how to set up the CFX Dx system's C1000 Dx thermal cycler at your site.
Tip: Before setting up the thermal cycler, familiarize yourself with the thermal cycler and its
optical reaction module, ports, and accessories.

Site Requirements

The tables in this section list the room, environment, and power requirements necessary to
successfully install and use the CFX Dx system thermal cycler.
Note: Install your CFX Dx system thermal cycler on a flat, dry surface with sufficient cool
airflow for it to run properly.

Bench Space Requirements

Table 4. CFX Dx system thermal cycler bench space requirements
Item Specification
Input power Up to 850 W, maximum
Frequency 50–60 Hz, single phase
USB ports 5 A, 1 B
Dimensions W: 13 in; 33 cm
D: 18 in; 46 cm
H: 14 in; 36 cm
Weight 47 lb; 21 kg
Operation Manual | 19
Page 22
Chapter 2 Setting Up the C1000 Dx Ther mal Cycler

Environment Requirements

Table 5. CFX Dx system thermal cycler environment requirements
Parameter Range Humidity Range
Operating conditions 15–31°C
59–87.8° F
Storage conditions 15–31°C
59–87.8° F
0–80% RH, non-condensing
0–80% RH, non-condensing

Power Requirements

Power to the CFX Dx system thermal cycler must be stable and within specifications to ensure
proper operation. The power cable connected to the power inlet port must be rated for 7 amps or
more.
Table 6. CFX Dx system power requirements
Item Specification
Mains input voltage 100–240 VAC; 50–60 Hz, single phase
Maximum power usage <850 watts
Number of power sockets Aminimum of 2 power sockets:
n 1 socket for the thermal cycler
n 1 socket for the computer running CFX Manager Dx
software
20 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 23

System Overview

The illustrations in this section display the main components of the C1000 Dx thermal cycler base.

Front View

System Overview
LEGEND
1.
2.
3.
4.
5.
6.
7.
Optical reaction module
thermal cycler block. The CFX Dx real-time PCR detection systems support either a CFX96™
Dx or CFX96 Deep Well Dx module.
Status LED
Lid button
— indicates when the block is in use.
— opens or closes the optical reaction module lid and seals the reaction chamber.
C1000™ Dx thermal cycler base
accommodates CFX96 Dx and CFX96 Deep Well optical reaction modules.
Front panel display and buttons
— includes an optical system to collect fluorescent data and a
— provides system power and communi cation, and
— allows control of the system in stand-alone mode.
Important: To ensure IVD gene study data integrity, the CFX Manager Dx software does
not support data generated by the thermal cycler in stand-alone mode.
Heated inner lid
Sample/reaction block
— maintains the lid temperature to prevent condensation and evaporation.
— holds reaction vessel, including tubes and microplates.
Operation Manual | 21
Page 24
Chapter 2 Setting Up the C1000 Dx Ther mal Cycler

Back View

LEGEND
1.
2.
3.
4.
5.
6.
7.
8.
Ethernet port
USB Type B port
Manager Dx software.
USB Type A ports
— connects the C1000Dxthermal cycler to your network.
— connect the C1000Dxthermal cycler to a computer running CFX
— transfer data to and from a USB flash drive.
Important: To ensure IVD gene study data integrity, the CFX Manager Dx software does
not support data generated by the thermal cycler in stand-alone mode.
Serial test port
Cooling vents
— for service testing only.
— cools thermal cycler.
Important: Do not block the cooling vents. For optimal operation, ensure that air can
circulate behind the thermal cycler base.
Power input
Power switch
Fuses
specifications.
— AC mains power; use provided power cord.
— rocker switch to turn thermal cycler on and off.
— see
Safe Use Specifications and Compliance on page 12
for fuse
22 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 25

Optical Reaction Modules

The C1000 Dx thermal cycler is compatible with the following Bio-Rad optical reaction modules for
real-time PCR.
n CFX96 Dx optical reaction module
n CFX96 Deep Well Dx optical reaction module
The chosen CFX Dx optical reaction module and the thermal cycler are shipped in separate boxes.
CFX Manager Dx software is shipped with the optical reaction module.
Important: The optical reaction module is calibrated with the thermal cycler base with which it
ships. Therefore, do not use the optical reaction module with any other thermal cycler base, or
the thermal cycler base with any other optical reaction module.
Both optical reaction modules include a fully adjustable heated lid that is capable of running
reliably with a broad range of reaction vessels. Each optical reaction module contains cooling fans
for fast heating and cooling.
OpticalReaction Modules
Each CFX Dx optical reaction module comprises the following components:
n Heated inner lid — maintains the lid temperature to prevent condensation and evaporation.
n Sample/reaction block — holds reaction vessels, including tubes and microplates.
n Lid button — opens and closes the lid and seals the reaction.
n Status LED — when on, indicates that the block is in use.

Recommended Sample Volumes

When using the C1000 Dx thermal cycler, the maximum sample volume is determined by the type
of reaction module used. Table 7 lists the recommended volumes to use with each reaction
module.
Table 7. Size and volume limit for reaction modules
Recommended Sample Volume, µl
Number of Wells Number of Blocks
96-well 1 10–50
96–deep well 1 10–125
(Upper Limit)
Operation Manual | 23
Page 26
Chapter 2 Setting Up the C1000 Dx Ther mal Cycler

Installing the C1000 Dx Thermal Cycler

The C1000 Dx thermal cycler base ships in a separate box from the optical reaction module. The
package includes:
n C1000 Dx thermal cycler base
n Power cord
n 1 USB cable
To install the C1000 Dx thermal cycler:
1. Unpack and set up the C1000 Dx thermal cycler base.
2. Attach the reaction module to the base.
3. Remove the shipping screw.
This section explains these tasks in detail.

Unpacking and Setting Up the C1000 Dx Thermal Cycler

Important: Before operating the thermal cycler, read the information in Safety and Regulatory
Compliance on page 11 and Safety Warning Labels on page 11.
Tip: During setup, ensure that you have sufficient space near the thermal cycler for a computer
on which to run the CFX Manager Dx software.
To unpack and set up the thermal cycler base
1. Locate the package containing the thermal cycler base.
2. Remove the base from the packing material.
Tip: Store the packing material for future use. If any item is missing or damaged, contact your local Bio-Rad office.
3. Place the thermal cycler base on a flat, dry surface with sufficient cool airflow to run properly.
4. Locate the power cord in the shipping package and insert one end into the power inlet port on back of the thermal cycler.
Important: Do not power on the instrument at this time.
5. Attach the IVD reaction module to the base. Proceed to Attaching the Optical Reaction Module
on page 25.
24 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 27
Installing the C1000 Dx T hermal Cycler

Attaching the Optical Reaction Module

Bio-Rad ships the CFX96 Dx or CFX96 Deep Well optical reaction module with its C1000 Dx
thermal cycler base (but in a separate box). Carefully unpack the optical reaction module and
verify that the power and USB cables are included in the shipping container.
Important: Each optical reaction module is calibrated with the thermal cycler base with which
it ships. Therefore, do not use the optical reaction module with any other thermal cycler base.
Ensure that the C1000 Dx thermal cycler base rests on a flat, dry surface with sufficient cool airflow
to run properly.
To attach the reaction module to the thermal cycler base
1. Place the C1000 Dx thermal cycler in a suitable location with the locking bar down.
2. Lifting the optical reaction module using the handle indents above the side air vents, position
the module in the C1000 Dx reaction module bay, leaving about 2cm of space in the front.
When in the bay, the optical module should be covering the Bio-Rad logo in front of bay.
3. Pull up the locking bar until it is flush with the sides of the module bay. This action moves the
module forward, locking it into place.
Operation Manual | 25
Page 28
Chapter 2 Setting Up the C1000 Dx Ther mal Cycler
4. Check that the module is completely and evenly seated in the C1000 Dx thermal cycler base.
There should be no extra space between the module and the base.
5. Plug the power cord into the back of the C1000 Dx thermal cycler base and into an
appropriate electrical outlet, and then press the power switch on the back panel of the C1000
Dx thermal cycler to start the system.

Removing the Shipping Screw

Important: Bio-Rad’s optical reaction modules ship with a red shipping screw inserted in the
inner lid to stabilize the optical reaction module during shipping. You must remove the
shipping screw before you can operate the optical reaction module.
To remove the shipping screw
1. The C1000 Dx thermal cycler recognizes that the shipping screw is inserted in the optical
reaction module and displays a message instructing you to remove the screw.
26 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 29
Installing the C1000 Dx T hermal Cycler
2. Follow the instructions to remove the shipping screw. The following diagram shows the
location of the shipping screw.
Note: You must reinsert the shipping screw should you need to return the reaction
module for any reason. Save the screw in a safe and accessible place.

Loading Sample Plates

To ensure uniform heating and cooling of samples, plates must be in complete contact with the
reaction block. To ensure adequate contact, do the following:
n Confirm that the block is clean before loading samples.
n Firmly press the individual tubes, tube strips, or microplates into the block wells.
n When using one or a few tubes, use the tube frame (catalog #1849000 or #1849001) or load
at least one empty tube in each corner of the block to ensure the lid exerts even pressure on
individual tubes.
Operation Manual | 27
Page 30
Chapter 2 Setting Up the C1000 Dx Ther mal Cycler
Loading Plates into the Optical Reaction Module
Important: When running the CFX Dx system, always balance the tube strips or add tube caps
to the corner wells to ensure the heated lid applies even pressure across the block.
To load plates into the reaction module
1. To open the motorized lid, do one of the following:
n In the Detected Instruments pane in CFX Manager Dx software, click Open Lid.
n On the Start Run tab in the software, click Open Lid.
n Press the lid button on the front of the instrument.
2. Place the microplate, individual tubes, or tube strips with sealed lids in the block.
Important: Verify that the tubes are completely sealed to prevent leakage.
Tip: For optimal results, load sample volumes of 10–25 µl for the CFX Dx system.
3. For accurate data analysis, verify that the orientation of reactions in the block is exactly the
same as the orientation of the well contents in the Plate tab in CFX Manager Dx software.
Tip: You can edit the well contents using CFX Manager Dx software before, during, or
after the run.
4. To close the motorized lid, do one of the following:
n Press the lid button on the instrument.
n In the Detected Instruments pane in the software, click Close Lid.
n On the Start Run tab in the software, click Close Lid.
Important: Ensure that nothing blocks the lid when it closes. Although there is a safety
mechanism to prevent the lid from closing if it senses an obstruction, do not place
anything in the way of the lid before closing.
28 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 31
Installing the C1000 Dx T hermal Cycler
PCR Plastics and Reagents Consumables
To find and order recommended plastic consumables for the CFX Dx system go to the Bio-Rad
web site. You can access this site from Help>PCRPlasticConsumablesWebSite menu item in
the CFX Manager Dx software. Additionally, refer to the Plastics Selector and the Reagents
Selector to help you easily find and order plastic consumables and reagents for your specific
hardware and PCR needs.

Detecting Connected Instruments

During installation, the CFX Manager Dx software installer automatically installs the instrument
drivers onto the computer running the CFX Manager Dx software. CFX Manager Dx detects
connected instruments when you start the software.
Important: You must disconnect the C1000 Dx thermal cycler from the CFX Manager Dx
computer before you install the software. You do not need to turn off the thermal cycler during
the software installation.
To detect connected instruments
1. If you have not yet done so, insert the square (male) end of the supplied USB Type B cable
into the USBType B port located on the back of the base.
2. Insert the other (port) end into a USB port on the CFX Manager Dx computer.
3. If the thermal cycler is not already running, press the power switch on the back of the
instrument to turn it on.
4. Start CFX Manager Dx software.
The software automatically detects the connected instrument and displays its name in the Detected
Instruments pane in the Home window.
Note: If the instrument does not appear in the Detected Instruments pane, verify that the USB
cable is properly installed. To reinstall drivers, select Tools > Reinstall Instrument Drivers in
the Home window in CFX Manager Dx software.
Operation Manual | 29
Page 32
Chapter 2 Setting Up the C1000 Dx Ther mal Cycler

Detaching the Reaction Module

Important: Power off the C1000 Dx thermal cycler before detaching a reaction module (see
Shutting Down the C1000 Dx Thermal Cycler on page 30). Cooling fins within the reaction module
might be hot immediately after running a protocol or incubation. Ensure that the fins are cool
before detaching the reaction module.
To detach the optical reaction module from the thermal cycler base
1. On the back of the thermal cycler base, push the locking bar down to unlock and release the
optical reaction module.
2. Carefully lift the optical reaction module out of the bay using the handle indents on each side.
3. Set the optical reaction module on a clean, flat surface where it cannot get bumped, scraped,
or dropped.

Shutting Down the C1000 Dx Thermal Cycler

To shut down the thermal cycler
1. After a run, press the open lid button on the front of the CFX optical reaction module to access
the samples loaded in the block.
2. Remove the samples from the block and press the close lid button to close the lid.
3. Press the power switch on the back panel of the C1000 Dx thermal cycler to power down the
system.
30 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 33

Chapter 3 Installing CFX Manager Dx Software

This chapter explains how to install the CFX Manager™ Dx software.
CFX Manager Dx software is required to analyze real-time PCR data from the CFX96™ Dx and
CFX96 Deep Well Dx systems. You can also use this software to control these systems in software-
controlled mode.
For information about installing the CFX Dx system thermal cycler and optical reaction module, see
Setting Up the C1000 Dx Thermal Cycler on page 19.
Operation Manual | 31
Page 34
Chapter 3 Installing CFX Manager Dx Software

System Requirements

Table 8 lists the minimum and recommended system requirements for the computer running the
CFX Manager Dx software (known as the CFX Manager Dx computer).
Table 8. Computer requirements for CFX Manager Dx software
System Minimum Recommended
Operating system Microsoft Windows 7 SP1
Pro
Any of the following:
n Microsoft Windows 7 SP2 Pro (32- and 64-bit)
n Microsoft Windows 10 Pro (64-bit only)
n Microsoft Windows 10 Enterprise (64-bit only)
Important: Secure Boot must be disabled on both Microsoft Windows 10 Pro and Enterprise.
Ports 2 USB 2.0 High-speed ports 2 USB 2.0 High-speed ports
Hard disk space 128 GB 128 GB
Processor speed 2.4 GHz, Dual Core 2.4 GHz, Quad Core
RAM 4 GB RAM 8 GB RAM
Screen resolution 1024 x 768 with true-color
mode
PDF reader Adobe PDF Reader or Windows PDF Reader
1280 x 1024 with true-color mode
from one of the supported Microsoft Office
Suites:
n 2007
n 2010
n 2013
32 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 35

Installing CFX Manager Dx Software

Important: You must disconnect any connected instruments from the CFX Manager Dx
computer before you install or upgrade the software. You do not need to turn off the thermal
cycler during the software installation. Ensure that you have saved all runs and that no
experiments are running.
Note: If you are installing CFX Manager Dx software on Windows 10, verify that Secure Boot is
disabled before beginning the installation procedure.
To install CFX Manager Dx software
1. If necessary, disconnect any connected instruments from the computer.
Locate and disconnect the instrument’s USB cable on the CFX Manager Dx computer. The
end inserted in the instrument can remain in place.
2. Log in to the CFX Manager Dx computer with administrative privileges.
3. Place the CFX Manager Dx software CD in the computer’s CD drive.
Installing CFX Manager Dx Softwar e
4. The software launch page should appear automatically. Double-click Install Software on the
software launch page.
Note: If the launch page does not appear automatically, navigate to the CD drive and
open the CFX_Manager folder and then double-click setup.exe to start the software
installation wizard.
Tip: In the installation wizard, click the Documentation button to find searchable copies of
the release notes, instrument manuals, and other documentation.
5. Follow the on-screen instructions to complete installation. When completed, the CFX manager
software icon will appear on the desktop of the computer.
6. After the installation completes, you can safely eject the CD.

Detecting Connected Instruments

During installation, the CFX Manager Dx software installer automatically installs the instrument
drivers onto the CFX Manager Dx computer. CFX Manager Dx detects connected instruments
when you start the software.
To detect connected instruments
1. If you have not yet done so, insert the square (male) end of the supplied USB Type B cable
into the USBType B port located on the back of the instrument’s base.
2. Insert the other (port) end into a USB port on the CFX Manager Dx computer.
Operation Manual | 33
Page 36
Chapter 3 Installing CFX Manager Dx Software
3. If the instrument is not already running, press the power switch on the back of the instrument to
turn it on.
4. Start CFX Manager Dx.
The software automatically detects the connected instrument and displays its name in the Detected
Instruments pane on the Home window.
Note: If the instrument does not appear in the Detected Instruments pane, verify that the USB
cable is properly installed. To reinstall drivers, on the Home window in CFX Manager Dx
select Tools>Reinstall Instrument Drivers.

Software Files

Table 9 lists the CFX Manager Dx software file types.
Table 9. CFX Manager Dx software file types
File Type Extension Details
Protocol .prcl Contains protocol setup details to perform a PCR run.
Plate .pltd Contains plate setup details to perform a PCR run.
Data .pcrd Contains the results of an experiment run and PCR
analysis.
PrimePCR™ run .csv Contains protocol and plate layout for PrimePCR plates.
Gene study .mgxd Contains results of multiple PCR runs and gene
expression analyses.
LIMS .plrn Contains plate setup and protocol information required to
conduct a LIMS compatible run.

Recommended Cybersecurity Measures

Bio-Rad recommends you work with your IT department to implementcybersecurity measures for the computer used with theCFX96 Dx system. For example:
n Install and configureappropriate virus protectionand firewall applications.
Important: Configure the virus scan to occur during off-hours or when the instrument is
not actively running. If a virus scan is initiated while CFX Manager Dx is running an
experiment, the run may be canceled and the data lost.
34 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 37
Recommended CybersecurityMeasures
n CFX Manager Dx software does not have a user session inactivity timeout feature. Implement
Windows or third-parity user access security measures (for example, implement a screen
saver with log in required).
n Removable media security:
o Use passwords and encryption on your USB device to protect data.
o Disable the autorun and autoplay features for all removable media devices.
o Enforce USB scanning whenever a thumb drive is plugged in.
n Employ a backup utility to facilitate data recovery.
Operation Manual | 35
Page 38
Chapter 3 Installing CFX Manager Dx Software
36 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 39

Chapter 4 The Workspace

CFX Manager™ Dx software provides an interface for setting up plates, developing PCR protocols,
running them on CFX Dx instruments, and analyzing data from PCR runs.
CFX Manager Dx software presents five primary workspaces:
n The Home window
n The Startup Wizard
n The Protocol Editor window
n The Plate Editor window
n The Data Analysis window
Each workspace is shown and briefly described in this chapter.
Operation Manual | 37
Page 40
Chapter 4 The Workspace

The Home Window

CFX Manager Dx software opens to the Home window and displays the Startup Wizard, from which
you can set up an experiment, perform or repeat a run, or analyze an existing run. From the Home
window you can also view application and instrument logs, create and manage users, and access
multiple useful tools. For more information, see Chapter 5, The Home Window.
38 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 41

The Startup Wizard

Use the Startup Wizard to quickly set up and run user-defined experiments or select and run a
PrimePCR™ experiment. You can also use this wizard to repeat a run or analyze run data.
The Startup Wizard
Operation Manual | 39
Page 42
Chapter 4 The Workspace

The Protocol Editor Window

In the Protocol Editor you can create, open, review, and edit a protocol. You can also modify the lid
temperature for the open protocol. Protocol Editor functionality is detailed in Chapter 6, Creating
Protocols.
40 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 43

The Plate Editor Window

In the Plate Editor you can create, open, review, and edit a plate. Plate Editor functionality is
detailed in Chapter 7, Preparing Plates.
The Plate Editor Window
Operation Manual | 41
Page 44
Chapter 4 The Workspace

The Data Analysis Window

In the Data Analysis window you can view and compare run data, perform statistical analyses,
export data, and create publication-ready reports. Data Analysis functionality is detailed in Chapter
9, Data Analysis Overview. See also Chapter 10, Data Analysis Details.
42 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 45

Chapter 5 The Home Window

CFX Manager™ Dx software provides an interface for developing PCR protocols, running them on
CFX Dx systems, and analyzing PCR run data.
This chapter introduces the CFX Manager Dx software and describes the features accessible from
the Home window.
Operation Manual | 43
Page 46
Chapter 5 The Home Window

The Home Window

CFX Manager Dx opens to the Home window and displays the Startup Wizard, from which you can
set up a run, perform or repeat a run, or analyze an existing run. From the Home window you can
also view application and instrument logs, create and manage users, and access multiple useful
tools.
LEGEND
1. The software title bar displays the name of the software and the logged in user.
2. The menu bar provides quick access to File, View, Users, Run, Tools, Window, and Help menu
commands.
3. Toolbar commands provide quick access to menu options.
4. The left pane displays the instruments connected to the CFX Manager Dx computer and
provides buttons from which you can operate the lid and view the status of the instruments.
5. The main pane displays the working window. The default working window on the Home screen
is the Startup Wizard.
6. The status bar displays names of the connected instruments and the logged in user.
44 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 47
The Home Window

File Menu Commands

New — opens a dialog box from which you can choose to create a new protocol, plate, or gene study.
Open — opens a dialog box from which you can choose to navigate to and open an existing
protocol, plate, data file, gene study, LIMS file, or PrimePCR™ run file.
Recent Data Files — displays a list of recently opened PCR files.
Repeat a Run — opens Windows Explorer to the location of saved PCR files, in which you can
locate a run to repeat.
Exit — closes CFX Manager Dx.

View Menu Commands

Application Log — displays a software usage log from initial installation to the current day.
Run Reports — displays a list of run reports.
Startup Wizard — displays the Startup Wizard in the main pane.
Run Setup — displays the Run Setup window in the main pane.
Instrument Summary — displays the Instrument Summary window in the main pane.
Detected Instruments — toggles between displaying and not displaying connected instruments in
the left pane. By default, the software displays connected instruments in the left pane.
Toolbar — toggles between displaying and not displaying the toolbar on the top of the screen. By
default, the software displays the toolbar.
Status Bar — toggles between displaying and not displaying the status bar on the bottom of the
screen. By default, the software displays the status bar.
Show — opens a dialog box from which you can
n View or block the Status log.
n Open and view the CFX Manager Dx data folder.
n Open and view the user’s data folder.
n Open and view the LIMS file folder.
n Open and view the PrimePCR folder.
n View the run history.
n View the properties of all connected instruments.
Operation Manual | 45
Page 48
Chapter 5 The Home Window

User Menu Commands

Select User — opens the Login screen on which you can select a user from the User Name
dropdown list and log in to the application.
Change Password — opens the Change Password dialog box, in which users can change their
CFX Manager Dx software password.
User Preferences — opens the User Preferences dialog box, in which users can change the
default settings for
n Sending and receiving email notification upon run completion
n Saving data files
n Creating protocols via the Protocol Editor or the Protocol AutoWriter
n Creating plates
n Analyzing data
n Performing gene expression analysis
n Determining the quality of the data
n Exporting CFX Dx instrument data
User Administration — opens the User Administration dialog box, in which administrators can
create users, modify role permissions, and assign roles to users.
Bio-Rad Service Login — for Bio-Rad technical service staff use only. Do not select this command.

Run Menu Commands

User-defined Run — opens the Run Setup window, in which you can set up a user-defined
protocol and plate, and then run a PCR experiment on selected instruments.
PrimePCR Run — opens the Start Run tab in the Run Setup window with the default PrimePCR
protocol and plate layout loaded based on the selected instrument.
End-Point Only Run — opens the Start Run tab in the Run Setup window with the default end-
point protocol and plate layout loaded based on the selected instrument.
Qualification Run — opens the Start Run tab in the Run Setup window with the default Bio-Rad
qualification protocol and plate layout loaded for the selected instrument.
46 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 49
The Home Window

Tools Menu Commands

Master Mix Calculator — opens the Master Mix Calculator, in which you can create a reaction
mixture and print the calculations.
Protocol AutoWriter — opens the Protocol AutoWriter dialog box, in which you can easily create a
new protocol.
TaCalculator — opens the TaCalculator, in which you can easily calculate the annealing
temperature of primers.
Dye Calibration Wizard — opens the Dye Calibration wizard, in which you can calibrate an
instrument for a new fluorophore.
Reinstall Instrument Drivers — reinstalls the drivers that control communication with Bio-Rad’s
real-time PCR systems.
Zip Data and Log Files — opens a dialog box in which you can select files to condense and save
in a zipped file for storage or to email.
Batch Analysis — opens the Batch Analysis dialog box, in which you can set parameters for
analyzing more than one data file at a time.
Options — opens a dialog box in which you can
n Configure your email server settings.
n Configure export settings for LIMS and other data files.

Help Menu Commands

Tip: The Help menu is available on the menu bar in all CFX Manager Dx software windows.
Open Operation Manual — opens a PDF of this manual.
Gene Expression Gateway Web Site — opens the Bio-Rad home page for the CFX Dx system.
PCR Reagents Web Site — opens Bio-Rad’s PCR reagents web site, from which you can order
PCR reagents, supermixes, dyes, and kits.
PCR Plastic Consumables Web Site — opens Bio-Rad’s PCR Plastics and Consumables web
site, from which you can order PCR plates, plate seals, tubes and caps, and other plastics
accessories.
Software Web Site — opens Bio-Rad’s PCR Analysis Software web site, from which you can order
updated versions of Bio-Rad’s CFX Manager Dx software.
About — displays CFX Manager Dx copyright and version information.
Operation Manual | 47
Page 50
Chapter 5 The Home Window

Toolbar Commands

— opens Windows Explorer, in which you can navigate to and open a data file or a gene
study file.
— opens the Master Mix Calculator.
— opens the Run Setup window.
— opens the Run Setup window with the default PrimePCR protocol and plate layout loaded
based on the selected instrument.
— opens the Startup Wizard.
48 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 51
The Home Window

The Startup Wizard

When CFX Manager Dx starts, the working pane displays the Startup Wizard. From the Startup
Wizard you can
n Select an instrument from the detected instruments and set up a user-defined or PrimePCR
run.
n Open and repeat a run.
n Open a data file to analyze results from a single run or a gene study file for results from
multiple gene expression runs.
These tasks are explained in detail in the chapters that follow.

Status Bar

The left side of the status bar at the bottom of the main software window displays the current status
of the detected instruments. The right side of the status bar displays name of the current user and
the date and time.
Operation Manual | 49
Page 52
Chapter 5 The Home Window

Detected Instruments Pane

The Detected Instruments pane displays each instrument that is connected to the CFX Manager Dx
computer. By default, each instrument appears as an icon, and its serial number appears as its name.
For example, the following image displays four detected instruments:
n Two C1000™ thermal cyclers with CFX96™ Deep Well reaction modules (CFX96DW1 and
CFX96DW2)
n Two C1000™ thermal cyclers with CFX96™ reaction modules (CFX96l01 and CFX96l02)
From this pane you can do the following:
n View the properties and calibrated dyes for a selected instrument.
For information about instrument properties, see Viewing the Properties of an Instrument on
page 53.
n View the status of a connected instrument.
n Open the motorized lid on the selected instrument.
n Close the motorized lid on the selected instrument.
n View the status of all connected instruments.
To view the status of a connected instrument
u In the Detected Instruments pane, select the target instrument and do one of the following:
n Click View Status in the Selected Instrument section.
n Right-click and select View Status on the menu that appears.
50 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 53
The Run Details dialog box appears displaying the Run Status tab. The status of the selected
instrument appears below the run status pane, for example:
To open or close the lid of an instrument
u In the Detected Instruments pane, select target instrument and do one of the following:
n Click Open Lid or Close Lid in the Selected Instrument section.
The Home Window
n Right-click and select the appropriate action on the menu that appears.
n Open the Run Details dialog box, select Run Status tab, and click Open Lid or Close Lid.
To view the status of all detected instruments
u Do one of the following:
n In the All Instruments section in the Detected Instruments pane, click View Summary.
n On the menu bar, select View > Instrument Summary.
The Instrument Summary dialog box appears:
Tip: If the system detects only one connected instrument, the All Instruments section does
not appear in the Detected Instruments pane. To view the instrument summary for a single
instrument, select View > Instrument Summary.
Operation Manual | 51
Page 54
Chapter 5 The Home Window
Instrument Summary Toolbar Controls
Table 10 lists the controls and functions in the Instrument Summary toolbar.
Table 10. Instrument Summary Toolbar Controls
Button Button Name Function
Create a new Run Creates a run on the selected block by
opening the Run Setup window.
Stop Stops the current run on selected blocks.
Pause Pauses the current run on selected blocks.
Resume Resumes the run on selected blocks.
Flash Block Indicator Flashes the indicator LED on the lid of the
selected blocks.
Open Lid Opens the selected block’s motorized l id.
Close Lid Closes the selected block’s motorized lid.
Hide Selected Blocks Hide the selected blocks in the Instrument
Summary list
Show All Blocks Show the selected blocks in the Instrument
Summary list
Show Select which blocks to show in the list.
Select one of the options to show all
detected blocks, all idle blocks, all blocks
that are running with the current user, or all
running blocks
52 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 55
The Home Window

Viewing the Properties of an Instrument

From the Detected Instruments pane you can view details about a selected instrument, including its
properties, the status of its shipping screw, and a list of its calibrated dyes (fluorophores).
To view instrument properties
u In the Detected Instruments pane, right-click the target instrument and select Properties on the
menu that appears.
Properties Tab
The Properties tab lists technical details about the selected instrument including the model, serial
numbers of its components, and firmware versions. The default name of the instrument (its serial
number) appears in many locations, including the Detected Instruments pane and in the header
bar of the Instrument Properties dialog box. You can rename the instrument in order to more easily
identify it.
To rename an instrument
u In the Instrument Properties tab, type a name in the Rename box at the top of the Properties
tab and click Rename.
The new name appears in the Nickname row in the Properties tab as well as in the Instrument
Properties header bar and the Detected Instruments pane.
Operation Manual | 53
Page 56
Chapter 5 The Home Window
Shipping Screw Tab
The Shipping Screw tab displays the current status of the shipping screw for the selected
instrument (Removed or Installed). The tab also includes instructions for installing or removing the
red shipping screw.
Tip: If the software detects the shipping screw, the Instrument Properties dialog box
automatically displays the Shipping Screw tab. Follow the instructions to remove the screw.
Note: You must remove the shipping screw before you can use the instrument. For more
information, see Removing the Shipping Screw on page 26.
Calibrated Dyes Tab
The Calibrated Dyes tab displays the calibrated fluorophores and plates for the selected
instrument.
To see detailed information about a calibration, click its Info button in the Detail column.
54 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 57

Before You Begin

Setting User Preferences

Tip: It is not required to perform these tasks in order to use CFX Manager Dx software. You
can safely skip this section or perform these tasks at any time.
In CFX Manager Dx you can customize your working environment. If your administrator created
software users, each user can customize his or her working environment. If your administrator did
not create users, preference changes apply to everyone who logs in to CFX Manager Dx. (For
information about creating CFX Manager Dx users, see Appendix B, Managing CFX Manager Dx
Users and Roles.)
For example, in the Users > User Preferences menu, you can do the following:
n Set up email notification of run completion.
n Change the default settings for
Before You Begin
o The location in which to save files
o The run setup files
o The file naming prefix
n Set the default parameters to use when creating a new protocol and plate.
n Set the default data analysis and gene expression parameters.
n Customize the default quality control parameters.
n Customize data export data parameters.
In the Tools menu, you can do the following:
n Create a master mix.
n Calibrate dyes for a specific instrument.
Note: The master mix and dye calibration are available to anyone who logs in to CFX
Manager Dx.
This section explains how to perform these tasks in detail.
Setting Up Email Notification
You can connect CFX Manager Dx to your outgoing email server to send email notification of run
completion to a list of users. You can also choose to attach a data file and an analysis report to the
Operation Manual | 55
Page 58
Chapter 5 The Home Window
list of users. To set up the connection between CFX Manager Dx and your SMTP server, see
Connecting CFX Manager Dx to an SMTP Server on page 57.
Note: A user's ability to access email setup features depends on the user group and
permissions assigned by the administrator. For details on managing users and their roles, see
Managing Users on page 233.
To set up email notifications
1. Select Users > User Preferences to open the User Preferences dialog box.
The User Preferences dialog appears displaying the Email tab.
Note: You are informed if the system detects that you have not set up a valid SMTP server
for CFX Manager Dx. Click Configure Outgoing Email to open the Options dialog box and
configure the email SMTP server. For more information, see Connecting CFX Manager Dx
to an SMTP Server on page 57.
2. In the To text box, type the email address of each person who you plan to inform of run
completion. All recipients will receive email after the run completes.
Note: You must enter each email address on a separate line. Press Enter or Return after
each address.
3. (Optional) In the cc text box, type the email address of any recipient to whom you plan to send
a copy of each email notification.
4. (Optional) By default, all recipients receive a copy of the data file as an attachment. Clear this
checkbox if you do not want to attach a copy the data file.
5. (Optional) Select Attach Analysis Report to attach a PDF of the analysis report to the email.
6. Click OK to save the changes and close the User Preferences dialog box.
56 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 59
Before You Begin
To edit a recipient’s email address
u Modify the email address as necessary and click OK.
To remove an email recipient
1. Select the email recipient and press the Delete key.
2. Click OK to save the changes and close the dialog box.
Important: Clicking Restore Defaults in the User Preferences dialog box resets all
preferences on all tabs to the original factory settings. Take care when clicking this button.
Connecting CFX Manager Dx to an SMTP Server
Important: Some commercial webmail service providers (such as Yahoo! and Gmail) have
increased email security. If you use these accounts, you must enable the setting Allow less
secure apps in their account settings to enable CFX Manager Dx to send email. See the
security information for your webmail service provider for more information.
You must establish a connection from CFX Manager Dx to your email server before the software
can send email notification.
To connect CFX Manager Dx to an email server
1. Do one of the following:
n Select User > User Preferences and click Configure Outgoing Email on the Email tab.
n Select Tools > Options.
The Options dialog box appears displaying the Email tab.
Operation Manual | 57
Page 60
Chapter 5 The Home Window
2. Provide the following information for your company:
n SMTP Server Name — the name of the outgoing email server at your company.
n Port — the port number of your SMTP server. This is usually 25.
n Use SSL — Secure Sockets Layer (SSL) option. Some SMTP servers require this setting.
If it is not required at your company, clear this checkbox.
n Use Default “From” Address — the name of the email server at your company. Some
SMTP servers require all sent email to have a “from” address that is from a certain domain, for example, name@YourCompany.com. If that is the case, clear this checkbox and provide a valid email address.
n Authentication Required — if your site requires account authentication, verify that this
checkbox is selected.
n User Name — the name of the authenticated account. This is required only if
Authentication Required is selected.
n Password — the password for the authenticated account. This is required only if
Authentication Required is selected.
3. To verify that the SMTP server settings are correct, enter a valid email address in the Test
Email Address text box and click Test Email.
Note: Some SMTP servers do not allow attachments and others allow attachments only
up to a specific size. If you plan to email data files and/or reports using CFX Manager Dx,
select Test Attachment and set Attachment Size in MB to 5 megabytes (MB) or more.
4. Click OK to save the changes and close the dialog box.
Changing the Default File Settings
In the Files tab on the User Preference dialog box, you can change the following:
n The default location in which to save CFX Manager Dx files
n The default files for run setup
n The default file naming parameters
To change the default file settings
1. Select Users > User Preferences to open the User Preferences dialog box.
2. In the User Preferences dialog box, select the Files tab.
58 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 61
Before You Begin
3. In the Default Folder for File Creation section, navigate to and select a default folder in which
you want to save new files. You can select a different location for each file type:
n Protocol
n Plate
n Data File
n Gene Study
4. In the File Selection for Run Setup section, navigate to and select the target protocol and plate
files to appear when you open the Experiment Setup window.
5. In the Data File section, define the prefix and/or suffix for data files. For any part, select a new
value from its dropdown list. You can also provide custom prefix and suffix values in the Prefix
and Suffix text boxes.
CFX Manager Dx displays a preview of the file name below the selection boxes.
6. Click OK to save the changes and close the dialog box.
Important: Clicking Restore Defaults in the User Preferences dialog box resets all
preferences on all tabs to the original factory settings. Take care when clicking this button.
Operation Manual | 59
Page 62
Chapter 5 The Home Window
Setting the Default Protocol Parameters
To set default protocol parameters for the Protocol Editor and Protocol AutoWriter
1. Select Users > User Preferences to open the User Preferences dialog box.
2. In the User Preferences dialog box, select the Protocol tab.
3. In the Protocol Editor section, specify values for the following settings that appear in the
Protocol Editor:
n Sample volume — the volume of each sample in the wells (in µl).
n Lid Shutoff temperature — the temperature in ºC at which the lid heater turns off during a
run.
4. In the Protocol AutoWriter section, specify values for the following settings that appear in the
Protocol AutoWriter:
n Annealing temperature — the temperature in ºC for experiments that use iProof™ DNA
polymerase, iTaq™ DNA polymerase, or other polymerases.
n Amplicon length — the length of the amplicon in bp.
5. Click OK to save the changes and close the dialog box.
Important: Clicking Restore Defaults in the User Preferences dialog box resets all
preferences on all tabs to the original factory settings. Take care when clicking this button.
60 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 63
Before You Begin
Setting Default Plate Parameters
Changes that you make to the Plate tab are available to all users of the software. Changes that you
make during plate setup are available to users after you save and close the plate file.
In the User Preferences dialog box you can do the following:
n Set default plate parameters.
n Add new target and sample names to their respective libraries.
n Delete target and sample names from their respective libraries.
To set the default plate parameters
1. Select Users > User Preferences to open the User Preferences dialog box.
2. In the User Preferences dialog box, select the Plate tab.
3. Specify values for the followings settings for a new plate file. These values appear in the Plate
Editor window:
n Plate type
n Plate size
n Units — the concentration of the starting template for wells that contain standards.
CFX Manager Dx uses these units to create a standard curve in the Data Analysis
Quantification tab.
Operation Manual | 61
Page 64
Chapter 5 The Home Window
n Scientific notation — when selected, CFX Manager Dx displays the concentration units in
scientific notation.
n Scan mode — the number or type of channels to scan during a run.
n Fluorophores — the default fluorophores that appear in the Plate Editor well loading
controls.
n Libraries — the target and sample names that you typically use in your experiments:
o Target names — the names of target genes and sequences.
o Sample names — the names of experiment samples or an identifying characteristic
for the samples (for example, Mouse1, Mouse2, Mouse3).
4. Click OK to save the changes and close the dialog box.
To add a new target or sample name
u In the appropriate library box, type the name for the target or sample and click OK.
To delete a target or sample name
u In the appropriate library box, select the name and press the Delete key and then click OK.
Important: Names that you remove from the library are removed from the software and are no
longer available to users. To restore the default CFX Manager Dx names, click Restore
Defaults. Clicking Restore Defaults in the User Preferences dialog box resets all preferences
on all tabs to the original factory settings. Take care when deleting default CFX Manager Dx
names and when clicking this button.
62 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 65
Setting Default Data Analysis Parameters
To set default Data Analysis parameters
1. Select Users > User Preferences to open the User Preferences dialog box.
2. In the User Preferences dialog box, select the Data Analysis tab.
Before You Begin
3. In the Analysis Mode section, select the mode in which to analyze the data (either Fluorophore
or Target).
4. In the PCR Quantitation section, set default parameters for the following options:
n Baseline Setting — the baseline method for analysis mode.
n Cq Determination Mode — the mode in which Cqvalues are calculated for each
fluorescence trace (either regression or single threshold).
n Threshold Calc. Mode — the end-point target amount.
The default is Auto. That is, the software automatically calculates the end-point target. To
set a specific threshold, clear the Auto checkbox and enter your end-point amount,
calculated in relative fluorescence units (or RFU). The maximum value is 65000.00 RFUs.
Data files for subsequent runs will use this threshold setting.
Operation Manual | 63
Page 66
Chapter 5 The Home Window
n Baseline Calc. Mode — the baseline value for all traces.
The default is Auto. That is, the software automatically calculates the baseline for all
traces. To set a specific baseline value, clear the Auto checkbox and enter minimum and
maximum values for the cycle range (1 to 9999). Data files for subsequent runs will use
this cycle range.
n Log View — determines how the software displays the amplification data:
o On — the amplification data are displayed in a semilogarithmic graph.
o Off — (the default) the amplification data are displayed in a linear graph.
5. In the End Point section, select the number of end cycles to average when calculating the end-
point calculations:
n PCR — the number of end cycles to average for quantification data (default is 5).
n End Point Only run — the number of end cycles to average for end-point data (default is
2).
6. In the Melt Curve section, select the peak type to detect (either positive or negative).
7. In the Well Selector section, select how to display well labels (by sample type, target name, or
sample name).
8. Click OK to save the changes and close the dialog box.
Important: Clicking Restore Defaults in the User Preferences dialog box resets all
preferences on all tabs to the original factory settings. Take care when clicking this button.
64 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 67
Setting Default Gene Expression Data File Parameters
To set the default parameters for a new gene expression data file
1. Select Users > User Preferences to open the User Preferences dialog box.
2. In the User Preferences dialog box, select the Gene Expression tab.
Before You Begin
3. Specify the values for the following settings:
n Relative to — graphs the gene expression data relative either to a control (originating at
1) or to zero:
o Zero — the software ignores the control. This is the default when no control sample is
assigned in the Experiment Settings window.
o Control — the software calculates the data relative to the control sample assigned in
the Experiment Setup window.
n X-axis — graphs the sample or the target on the x-axis.
n Y-axis — graphs linear, log2, or log10 scale on the y-axis.
n Scaling — the scaling option for the graph (the default option is unscaled):
o Highest — the software scales the graph to the highest data point.
Operation Manual | 65
Page 68
Chapter 5 The Home Window
o Lowest — the software scales the graph to the lowest data point.
o Unscaled — the software presents the data unscaled in the graph.
n Mode — the analysis mode, either relative quantity (C
(∆∆C
).
q
) or normalized expression
q
n Error Bar — the data variability presented as either the standard deviation (Std. Dev.) or
the standard error of the mean (Std. Error Mean).
n Error Bar Multiplier — the standard deviation multiplier used to graph the error bars
(defaultis1).
You can increase the multiplier to either 2 or 3.
n Sample Types to Exclude — the sample types to exclude from the analysis.
You can select one or more sample to exclude from the analysis. To exclude all sample
types, clear the checkboxes of any selected sample types.
4. Click OK to save the changes and close the dialog box.
Important: Clicking Restore Defaults in the User Preferences dialog box resets all
preferences on all tabs to the original factory settings. Take care when clicking this button.
Customizing Quality Control Rules
In CFX Manager Dx, you can set quality control rules, which are applied to data in the Data
Analysis window. The software validates the data against the rules that you set.
Note: By default, all quality control rules are enabled.
Tip: You can easily exclude wells that fail a QC parameter from analysis in the QC module of
the Data Analysis window.
To customize quality control rules
1. Select Users > User Preferences to open the User Preferences dialog box.
2. In the User Preferences dialog box, select the QC tab.
66 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 69
where:
n NTC — no template control
Before You Begin
n NRT — no reverse transcriptase control
n Efficiency — reaction efficiency
n Std Curve R^2 — R square value for the standard curve
n Replicate group Cq Std Dev — standard deviation calculated for each replicate group
3. For each QC rule, do one of the following:
n To use its default value, do nothing.
n To change its value, click its Value text box, type a new value, and press the Enter key.
n To disable the rule, clear its Use checkbox.
4. Click OK to save the changes and close the dialog box.
Important: Clicking Restore Defaults in the User Preferences dialog box resets all
preferences on all tabs to the original factory settings. Take care when clicking this button.
Operation Manual | 67
Page 70
Chapter 5 The Home Window
Customizing Data Export Parameters
You can export CFX Manager Dx data in the following formats:
n Text (.txt)
n CSV (.csv)
n Excel 2007 (.xlsx)
n Excel 2003 (.xls)
n XML (.xml)
n HTML (.html)
You can specify the type of data to export and customize the output of the exported data.
To customize data export parameters
1. Select Users > User Preferences to open the User Preferences dialog box.
2. In the User Preferences dialog box, select the Custom Export tab.
3. On the Export Format dropdown list, select a format in which to export the data.
68 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 71
Before You Begin
4. In the Data to Export section, select or clear the checkboxes for the type of data to export. The
selected items appear in the Exported Columns list box.
Note: By default, the run information is included in the header. Clear this checkbox if you
do not want the run information included.
5. You can change the output display order of the selected items.
In the Exported Columns list box, highlight the item and then click the arrow buttons to the left
of the list to move it up or down.
6. Optionally, you can change the output column names of the selected items:
a. Click Customize Column Names.
The Column Name Customizer dialog box appears.
b. For each default column name that you want to change, type the new name in its Custom
Name field.
c. Do one of the following:
n Click OK to save the changes and return to the Custom Export tab. The new name
appears in parentheses beside the default column name in the Exported Columns list
box.
n Click Cancel to clear the changes and return to the Custom Export tab.
7. Click OK to save the changes and close the dialog box.
Important: Clicking Restore Defaults in the User Preferences dialog box resets all
preferences on all tabs to the original factory settings. Take care when clicking this button.

Creating a Reaction Master Mix

Using CFX Manager Dx’s Master Mix Calculator, you can easily calculate the required volume of
each component in your master mix. You can print the master mix calculation table to your default
printer, and save the calculations for each target for later use.
To create a reaction master mix using the Master Mix Calculator
1. To open the Master Mix Calculator, do one of the following:
n Select Tools > Master Mix Calculator.
n Click Master Mix Calculator on the toolbar.
The Master Mix Calculator appears.
Operation Manual | 69
Page 72
Chapter 5 The Home Window
2. In the Reaction section, select a detection method:
n SYBR®Green/EvaGreen
n Probes
3. To create a new target, in the Target section click Create New. A new target name appears in
the target dropdown list.
4. (Optional) To change the default target name:
a. Highlight the target’s name in the dropdown target list.
b. Type a new target name in the Target box.
c. Press the Enter key.
5. Adjust the starting and final concentrations for the forward and reverse primers and any
probes.
70 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 73
Before You Begin
6. In the Master Mix Setup section, adjust the values for
n Number of reactions to run
n Reaction volume per well
n Template volume per well
n Supermix concentration per well
n Excess reaction volume per well
7. (Optional) Perform steps 26 for as many targets as necessary.
8. In the Choose Target to Calculate section, select the target to calculate.
Tip: You can calculate only one or several or all targets at the same time.
The calculated volumes of the required components for each selected target appear in the
master mix table.
9. Click Set as Default to set the quantities input in the Target and Master Mix Setup sections as
new defaults.
10. Click OK to save the contents of the Master Mix Calculator dialog box.
To print the master mix calculations table
u To print a master mix calculations table, click Print.
The calculations table prints to your default printer.
To save the master mix calculations table as a PDF
u Change your default printer to a PDF driver and click Print on the Master Mix Calculator.
To delete targets
u Select the target using the dropdown target list and click Remove.
Important: Removing a target from the target list also removes it from any master mix
calculations it is used in. Take care when deleting a target.
Operation Manual | 71
Page 74
Chapter 5 The Home Window

Calibrating New Dyes

The CFX96™ Dx systems are factory calibrated for commonly used fluorophores in white-well and
clear-well plates. Table 11 lists the fluorophores and channel for which each instrument is calibrated.
Note: The CFX96 systems also include a channel dedicated to FRET chemistry. This channel
does not require calibration for specific dyes.
Table 11. Factory calibrated fluorophores and channels
Fluorophores Channel Excitation, nm
FAM, SYBR® Green I 1 450–490 515–530
Detection,
nm
VIC, HEX, CAL Fluor Gold 540, Cal Fluor
Orange 560
ROX, Texas Red, CALFluor Red 610, TEX
615
CY5, Quasar 670 4 620–650 675–690
Quasar 705, Cy5.5 5 672–684 705–730
2 515–535 560–580
3 560–590 610–650
To calibrate new dyes for CFX systems
1. In the Home window, select a target instrument in the Detected Instruments pane.
2. Select Tools > Calibration Wizard to open the Dye Calibration wizard.
Fluorophores already calibrated for the target instrument appear in the Calibrated
Fluorophores table.
72 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 75
Before You Begin
3. In the Calibrate New or Existing Fluorophores section, select the fluorophore to calibrate from
the dropdown list.
If the fluorophore name is not included in the list, type its name in the text box to add it to the
list.
4. Select the plate type for the fluorophore.
If the plate type is not included in the list, type the name in the text box to add it to the list.
5. Select a channel for the fluorophore.
6. Select a plate column for the fluorophore.
7. (Optional) Type a color to associate with the fluorophore.
8. Click Add to Plate to add the fluorophore.
9. (Optional) Repeat steps 38 to add each fluorophore you plan to calibrate for the plate.
10. When you finish adding fluorophores, click View Plate to open the Pure Dye Plate Display
window.
Use this window as a guide for loading dyes into the plate.
11. Prepare a 96-well plate for dye calibration:
a. Pipette dye solution into each well, following the pattern shown in the Pure Dye Plate
Display.
b. For each fluorophore, fill four wells with 50µl (96-well plate) of 300 nM dye solution.
Notice that at least half the plate contains blank wells.
c. Seal the plate using the sealing method you will use in your experiment.
12. Place the calibration plate in the block and close the lid.
13. In the Dye Calibration wizard, click Calibrate and then OK to confirm that the plate is in the
block.
14. When CFX Manager Dx software completes the calibration run, a dialog box appears. Click
Yes to finish calibration and open the Dye Calibration Viewer.
15. Click OK to close the window.
Operation Manual | 73
Page 76
Chapter 5 The Home Window
74 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 77

Chapter 6 Creating Protocols

A protocol is set of steps that are executed in a specific sequence. In CFX Manager™ Dx software,
all of the steps are associated with options on the instrument. For example, the steps instruct the
instrument to control block and lid temperature, apply a temperature difference across the block,
take a plate read, or perform a melt curve analysis. Each option is specified for different plate and
run types.
CFX Manager Dx provides two options for creating protocols: Protocol Editor and Protocol
AutoWriter.
The Protocol Editor features include the following:
n Standard protocol controls to quickly create protocols
n Ability to quickly calculate a gradient for the selected number of rows
n Ability to quickly calculate run time for the selected plate type
n Ability to edit protocol steps
n Ability to save protocols for reuse
n Ability to print the protocol to a default printer
The Protocol AutoWriter automatically generates a customized PCR protocol with hot start, initial
denaturation, annealing, and extension steps using parameters that you provide. You can then
view a graphical representation of the suggested protocol and edit, run, or save the protocol.
Operation Manual | 75
Page 78
Chapter 6 Cr eating Protocols

Protocol Editor Window

Use the Protocol Editor to create, open, review, and edit a protocol. By default, the Protocol Editor
opens displaying a generic real-time 2-step protocol for a 96-well plate.
LEGEND
1. The menu bar provides quick access to the File, Settings, and Tools menu commands.
2. The toolbar provides quick access to save and print the protocol, determine where to insert a
step, set sample volume, and view estimated protocol run time.
3. The main pane displays a graphical representation of the protocol.
4. The lower pane displays the protocol outline.
5. The left pane displays the protocol controls that you can add to customize the protocol.

File Menu Commands

Save — saves the current protocol.
Save As — saves the current protocol with a new name or in a new location.
Close — closes the Protocol Editor.
76 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 79
Protocol Editor Window

Settings Menu Command

Lid Settings — opens the Lid Setting dialog box in which you can change or set the lid
temperature.

Tools Menu Commands

Gradient Calculator — opens a dialog box from which you can select the block type for a gradient
step. The default is 96 wells.
Run time Calculator — opens a dialog box from which you can select the plate type and scan
mode in order to calculate the estimated run time in the Run Setup window. The default is 96 wells,
all channels.

Toolbar Commands

— saves the current protocol file.
— prints the selected window.
— use this command to select where to insert steps relative to the currently
selected step.
— use this command to enter a sample volume in µl. Sample volumes
differ depending on the type of block:
n For a 96–deep well block the range is 0–125 µl.
n For a 96–well block the range is 0–50 µl.
— displays the estimated run time based on the protocol steps, ramp rate,
and the type of block selected.
— displays Help information about protocols.
Operation Manual | 77
Page 80
Chapter 6 Cr eating Protocols

Protocol Editing Controls

The left pane of the Protocol Editor window comprises controls that you can use to create
protocols.
Each control consists of a set of parameters that represent a step in the protocol. You can modify
each parameter and add or remove them to customize your protocol. This section describes the
options in each control.
repetitions of steps 2–4. After the final repetition, the software will have performed steps 2–4 a total of 40 times. You can edit the return-to (GOTO) step and the number of cycles either in the graphical display or in the protocol outline.
n Insert Step — inserts a step before or after the selected step. You
can edit the temperature and hold time values either in the protocol’s graphical display or in the protocol outline.
n Insert Gradient — inserts a gradient step based on the type of well
block selected in the gradient calculator. You can edit the gradient range in the Gradient pane that appears when a gradient step is inserted.
n Insert GOTO — inserts a cycling (loop) step, which informs the
software to repeat specific steps in sequence for a specified number of cycles. The repetitions begin after the first cycle is complete. For example, you can inform the software to perform 39
n Insert Melt Curve — inserts a melt curve read step.
n Insert Plate Read to Step — adds a plate read command to the selected step. A plate read
measures the amount of fluorescence at the end of a cycle. The plate read step is generally the last step in a GOTO loop.
Tip: After you add a plate read command to a step, the button changes to Remove Plate
Read when you select the step.
n Remove Plate Read — removes a plate read command from the selected step.
Tip: After you remove a plate read command from a step, the button changes to Add Plate
Read to Step when you select the step.
n Step Options — opens the Step Options dialog box and displays the options available for the
selected step. See Step Options on page 79 for detailed information about the step options.
Tip: You can also access the Step Options by right-clicking the step in the graphic display.
n Delete Step — deletes the selected step from the protocol.
78 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 81
Protocol Editor Window
Step Options
Open the Step Options dialog box to view the options you can add, change, or remove from a step.
n Plate Read — when selected, adds a plate read to the step.
n Temperature — sets the target temperature for the selected step.
n Gradient — sets the gradient range for the step; the range is 1–24°C.
Note: A gradient runs with the lowest temperature in the front of the block (in this image,
row H) and the highest temperature in the back of the block (in this image, row A).
n Increment — the amount to increase (or decrease) the temperature of the selected step; this
value amount is added to the target temperature with each cycle. The range is ±0.1–10°C.
Note: To decrease the temperature, type a minus sign (–) before the numerical value
(forexample, –5°C).
n Ramp Rate — the ramp rate for the selected step; the range depends on the block size.
n Time — the hold time for the selected step.
n Extend — the amount of time (in sec) to extend or decrease the selected step; this option is
added to the hold time in each cycle; the range is 1–60 sec.
n Beep — when selected, a beep sounds at the end of the step.
Tip: When you enter a number that is outside the option range, the software changes the
number to the closest entry within the range.
Operation Manual | 79
Page 82
Chapter 6 Cr eating Protocols

Creating a Protocol in the Protocol Editor

Using the Protocol Editor, you can create custom protocol files. You can also edit and save
previously saved protocol files or sample protocol files shipped with CFX Manager Dx software.
To create a new protocol file, perform the following:
n Open a protocol file in the Protocol Editor.
Tip: You can open a new or existing protocol in the Protocol Editor.
n Set up the new protocol.
n Add steps to the protocol from the protocol controls pane.
n Edit the properties of the steps.
n Save the protocol.
Tip: To create a new protocol from a previously saved or sample protocol file, see Opening an
Existing Protocol in the Protocol Editor on page 81.

Opening a New Protocol File in the Protocol Editor

CFX Manager Dx offers multiple options to open a new protocol file:
n From the Home window
n From the Startup Wizard dialog box
n From the Run Setup dialog box
To open a new protocol file from the Home window
u Select File > New > Protocol.
The Protocol Editor window opens, displaying the default protocol file.
Tip: For information about setting your default protocol, see Changing the Default File
Settings on page 58.
To open a new protocol file from the Startup Wizard
1. In the Home window, do one of the following to open the Startup Wizard if it is not in view:
n Select View > Startup Wizard.
n Click Startup Wizard on the toolbar.
By default, the Startup Wizard displays the Run setup tab with the CFX96™ instrument type
selected.
80 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 83
Creating a Protocol in the ProtocolEditor
2. If necessary, select the instrument type from the dropdown list.
3. Click User-defined as the run type.
The Run Setup dialog box opens to the Protocol tab and displays the default protocol file.
4. Click Create New.
The Protocol Editor window opens, displaying the default real-time protocol.
To open a new protocol from the Run Setup dialog box
1. In the Home window, do one of the following to open the Run Setup dialog box:
n Select Run > User-defined Run.
n Click User-defined Run Setup on the toolbar.
The Run Setup dialog box opens to the Protocol tab and displays your default protocol file.
2. Click Create New.
The Protocol Editor window opens, displaying the default real-time protocol.

Opening an Existing Protocol in the Protocol Editor

CFX Manager Dx provides sample protocol files that you can edit and save as custom new
protocols. You can also create a new protocol from an existing custom protocol.
To open a sample protocol file
1. In the Home window, select File > Open > Protocol.
By default, Windows Explorer opens to the location of the CFX Manager Dx Sample files
folder.
2. Open the Sample files folder. You see the following folders:
n ConventionalProtocols — contains example protocol files for traditional PCR analysis.
n DataFiles — contains example data files that you can use to explore CFX Manager Dx’s
features.
n MeltCalibration — contains example protocol files for use with Bio-Rad’s Precision Melt
Analysis software.
n Plates — contains example plate files.
n RealTimeProtocols – contains example protocol files for real-time PCR analysis.
Operation Manual | 81
Page 84
Chapter 6 Cr eating Protocols
3. Open the protocol folder for the type of run you plan to perform, either ConventionalProtocols
or RealTimeProtocols.
4. Select the protocol of choice and click Open.
The sample protocol opens in the Protocol Editor window.
5. Select File > Save As and save the protocol with a new name or in a new folder.
To open an existing protocol
1. In the Home window, do one of the following:
n Select File > Open > Protocol, navigate to and select the target protocol, and click Open.
n Open the Startup Wizard and do one of the following:
o To edit the displayed protocol, click Edit Selected.
o To edit another existing protocol, click Select Existing and navigate to the target file.
The protocol opens in the Protocol Editor window.
2. Select File > Save As and save the protocol with a new name or in a new folder.

Setting Up a New Protocol

Tip: If your protocol file includes the required parameters (for example, if you are editing an
existing plate file) you can skip this section. Proceed to Adding Steps to a Protocol on page 84.
New protocol files require the following parameters:
n Block type
n Scan mode for the chosen block type
n Lid temperature
n Sample volume
Setting the Block Type
CFX Manager Dx automatically calculates temperature increments for gradient steps based on the
block type.
Note: The plate type set in the Protocol Editor must be the same as the plate in the reaction
module.
82 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 85
Creating a Protocol in the ProtocolEditor
To set the block type
u In the Protocol Editor window, select Tools > Gradient Calculator and choose the appropriate
plate type in the dropdown list that appears.
Selecting the Scan Mode for the Chosen Block Type
To determine the run time for the protocol, select the target block type and scan mode.
To select the block type and scan mode
u In the Protocol Editor window, select Tools > Run time Calculator and choose the appropriate
plate type and scan mode in the dropdown list that appears.
Adjusting the Lid Temperature
CFX Manager Dx sets the default lid temperature to 105.0°C.
You can change the default settings or turn off the lid heater as necessary for the protocol.
Tip: You can change the default lid temperature in the User Preferences dialog box. See
Setting the Default Protocol Parameters on page 60.
To adjust the lid temperature
1. In the Plate Editor window, select Settings > Lid Settings.
The Lid Settings dialog box appears.
2. Do one of the following:
n Select User Defined and enter a temperature value in the text box.
n Select Turn Off Lid Heater.
3. Click OK to accept the changes and close the dialog box
Setting the Sample Volume
By default, CFX Manager Dx sets the sample volume for each well to 25 µl. However, the CFX Dx
system range is 0–125 µl.
The instrument uses one of two temperature control modes to determine when the sample reaches
the target temperature in a protocol:
n Calculated mode — when the sample volume is set to a volume appropriate for the block, the
instrument calculates the sample temperature based on the sample volume. This is the
standard mode.
Operation Manual | 83
Page 86
Chapter 6 Cr eating Protocols
n Block mode — when the sample volume is set to zero (0) µl, the instrument records the
sample temperature as the same as the measured block temperature.
To set the sample volume for a specific block
u In the Plate Editor window, type the correct value in the Sample Volume text box on the
toolbar.
Tip: You can change the default sample volume in the User Preferences dialog box. See
Changing the Default File Settings on page 58.

Adding Steps to a Protocol

To add a step to a protocol
1. Open the protocol in the Protocol Editor window.
2. Determine where to insert the new step. On the toolbar, select Before or After in the Step
dropdown list.
3. On the graph, select the step before or after which you plan to insert the new step.
4. In the left pane, click Insert Step.
5. To change the temperature or hold time, click the default value on the graph or the protocol
outline and type a new value.
6. (Optional) In the left pane, click Step Options to display the Step Options dialog box and
modify the available options for the selected step.
Tip: You can access the Step Options dialog box on the right-click menu in either the
graph pane or the protocol outline pane.
7. Click OK and then click Yes to save changes to the protocol.
The Save As dialog box appears
8. In the Save As dialog box, type a name for the new protocol file and click Save.

Inserting a Gradient Step

To insert a gradient step
1. Verify that the plate size for the gradient is the same as the block type of the instrument, 96-
well.
2. If you have not yet done so, select the plate size for the gradient:
84 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 87
Creating a Protocol in the ProtocolEditor
Select Tools > Gradient Calculator and choose the appropriate well type from the dropdown
list.
3. On the toolbar, select either Before or After from the Insert Step dropdown list.
4. In the graph or outline pane, select the step before or after which you plan to insert the
gradient step.
5. In the left pane, click Insert Gradient. The new gradient step is highlighted in the graph and the
outline pane, for example:
The temperature of each row in the gradient appears in the Gradient table in the right pane.
6. To edit the gradient temperature range, do one of the following:
n Click the default temperature in the graph or outline pane and enter a new temperature.
n Click Step Options to enter the gradient range in the Step Options window.
n Change the Range value in the Gradient table.
7. To edit the hold time, click the default time in the graphic or text view and enter a new time.
8. Click OK and then Yes to save the changes.
Operation Manual | 85
Page 88
Chapter 6 Cr eating Protocols

Inserting a GOTO Step

Note: You cannot insert a GOTO step within a GOTO set; you cannot create nested GOTO
loops.
To insert a GOTO step
1. On the toolbar, select Before or After from the Insert Step dropdown list.
2. In the graph, select the step before or after which you plan to insert the GOTO step.
3. In the left pane, click Insert GOTO.
4. To edit the GOTO step number or number of GOTO repeats, select the default number in the
graph or outline pane and enter a new value.
5. Click OK and then Yes to save the changes.

Inserting a Melt Curve Step

Tip: You cannot insert a melt curve step inside a GOTO loop.
Note: The melt curve step includes a 30 sec hold at the beginning of the step that is not shown
in the protocol.
To insert a melt curve step
1. On the toolbar, select Before or After from the Insert Step dropdown list.
2. In the graph, select the step before or after which you plan to insert the melt curve step.
86 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 89
Creating a Protocol in the ProtocolEditor
3. In the left pane, click Insert Melt Curve. The new melt curve step is highlighted in the graph
and the outline pane, for example:
4. To edit the melt temperature range or increment time, select the default number in the graph or
outline pane and enter a new value.
5. Click OK and then Yes to save the changes.
Operation Manual | 87
Page 90
Chapter 6 Cr eating Protocols

Adding or Removing a Plate Read Step

Tip: After you add a plate read command to a step, the button changes to Remove Plate Read
when you select the step.
To add a plate read to a step
1. On the toolbar, select Before or After from the Insert Step dropdown list.
2. On the graph, select the step before or after which you plan to insert the plate read step.
3. In the left pane, click Add Plate Read to Step to add a plate read to the selected step.
4. Click OK and then Yes to save the changes.
To remove a plate read from a step
u On the graph, select the step that contains the plate read and click Remove Plate Read in the
left pane.

Changing Step Options

To change step options for a selected step
1. Select the target step in the graph or outline pane.
2. In the left pane, click Step Options to open the Step Options dialog box.
Alternatively, right-click the target step in either pane and select Step Options in the menu that
appears.
3. To add, modify, or remove options:
n Enter a value in the appropriate text box.
n Edit a value in the specific text box.
n Select or clear a checkbox.
4. Click OK to save the changes and close the Step Options dialog box.
5. Click OK and then Yes to save the protocol.
88 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 91

Deleting a Step

To delete a step in the protocol
1. Select the step in the graph or outline pane.
2. In the left pane, click Delete Step to delete the selected step.
3. Click OK and then Yes to save the protocol.

Copying, Exporting, or Printing a Protocol

To copy a protocol
u Right-click the protocol outline and select Copy Protocol.
You can paste the outline into a .txt, .xls, .doc, or .ppt file.
To export a protocol
1. Right-click the protocol outline and select Export Protocol.
Creating a Protocol in the ProtocolEditor
The Save As dialog box appears.
2. (Optional) In Windows Explorer, navigate to a folder in which to save the protocol file.
3. In File name, type a name for the exported protocol file.
4. Click Save.
To print a protocol
u Right-click the protocol outline and select Print.
You can print the protocol outline to your default printer.
Operation Manual | 89
Page 92
Chapter 6 Cr eating Protocols

Creating a Protocol with the Protocol AutoWriter

Important: Bio-Rad does not guarantee that running a protocol created with the Protocol
AutoWriter will always result in a real-time PCR product.
CFX Manager Dx’s Protocol AutoWriter automatically generates cycling protocols based on the
following input parameters:
n Amplicon length — the expected length of the PCR product
n Annealing temperature — the reaction Tafor the primers being used
If the Tais unknown, you can use the Tacalculator to automatically calculate it based on your
primer sequences.
Note: The Tais adjusted from the primer melting temperature (Tm) information that is
based on the selected enzyme and the protocol speed.
n Enzyme type — the DNA polymerase enzyme (iTaq™, iProof™ DNA polymerase, or Other)
If you use an enzyme other than iTaq or iProof DNA polymerase, you can enter additional
information, including the gradient range, hot-start activation time (in sec), and the final
extension time (in sec).
n Run speed — the reaction speed (standard, fast, or ultrafast)
The Protocol AutoWriter optimizes the protocol depending on the selected speed setting. The
total run time is determined by the number of steps and cycles, the incubation time at each
step, and the time it takes to reach uniformity at the target temperature.
Using parameters that you enter and standard PCR guidelines, the Protocol AutoWriter
automatically generates a customized PCR protocol with hot start, initial denaturation, annealing,
and extension steps. You can then view a graphical representation of the suggested protocol and
edit, run, or save the protocol.
90 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 93
Creating a Protocol with the Protocol AutoWriter
To create a new protocol using CFX Manager Dx’s Protocol AutoWriter
1. In the Home window, select Tools > Protocol AutoWriter.
The Protocol AutoWriter dialog box appears.
2. In the Enter Target Values/Enzyme section, do the following:
n Enter the annealing temperature (Ta) for the primers, if known.
Tip: Using the Ta Calculator on page 92 for more information.
Note: For information about the calculations used in the TaCalculator, see Breslauer
et al. 1986.
n Enter the amplicon length in base pairs (bp).
n Select an enzyme type from the list of options (iTaq™ DNA polymerase, iProof™ DNA
polymerase, or Other).
Tip: If you select Other as the enzyme type, the parameters in the Additional
Parameters (Optional) section become active.
Operation Manual | 91
Page 94
Chapter 6 Cr eating Protocols
3. If you selected Other as the enzyme type, you can add any or all of the following parameters to
the protocol:
n Gradient range
n Hot start activation temperature
n Final extension time
4. In the Type section, move the sliding bar to select a protocol speed (Standard, Fast, or
Ultrafast). CFX Manager Dx adjusts the total run time.
5. Select the type of PCR to perform (Real-time PCR is the default).
With real-time PCR, CFX Manager Dx adds a plate read step to collect fluorescence data.
6. In the Preview section, review the protocol. You can make changes as needed.
7. Do one of the following:
n Click OK to save the new protocol. After saving it, the protocol opens in the Startup
Wizard. Click Edit Selected to make any changes to the protocol. For example, you might
need to change the lid temperature and sample volume.
n Click Cancel to close the window without saving the protocol.

Using the TaCalculator

When the annealing temperature for the primer is unknown, you can use the TaCalculator to
calculate the value. You can use the value in the Protocol AutoWriter or in the Protocol Editor to
create your protocol.

About the TaCalculator

The TaCalculator calculates the Tmvalue for each primer as well as the Tavalue for the protocol at
standard speed.
The Tafor the protocol is based on the average primer Tmvalues with the following rules applied:
n If the difference between the primer Tmvalues is >4°C, the Ta= (lower of the two primer T
values + 2) – 4°C
n If the difference between the T
4°C
values is 4°C, the T
m
= (average of the primer Tmvalues) –
a
m
92 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 95
Using the Ta Calculator
Base Pair Counting Method
For each primer, the TaCalculator uses the base pair counting method for sequences of 14 base
pairs (bp) or fewer.
Tm= ((w*A + x*T) * 2) + ((y*G + z*C) * 4)
where w, x, y, and z are the numbers of the bases A, T, G, and C in the sequence, respectively.
Nearest Neighbor Method
For sequences longer than 14 bp, the nearest neighbor method is used. In the nearest neighbor
method, the melting temperature calculations are based on the thermodynamic relationship
between entropy (order or a measure of the randomness of the oligonucleotide), enthalpy (heat
released or absorbed by the oligonucleotide), free energy, and temperature.
H = G + T* S
where:
n H = Enthalpy value, Cal/Mole*K
n T = temperature, Kelvin
n S = Entropy value, Cal/Mole*K n G = Gibbs free energy in Cal/Mole*K
The change in entropy and enthalpy is directly calculated by summing the values for nucleotide
pairs shown in Table 12 (Breslauer et al. 1986).
The relationship between the free energy and the concentration of reactants and products at
equilibrium is given by:
G = R*T*ln ((DNA * Primer)/(DNA + Primer))
where R is the gas constant (1.986 Cal/Mole*K).
Substituting G in the two equations and solving for T gives
T = H/(S + R*Ln((DNA * Primer)/(DNA + Primer)))
assuming that the concentration of DNA and DNA-primer complex are equal.
It has been determined empirically that there is a 5 kcal free energy (3.4 kcal) (Sugimoto et al.
1996) change during the transition from single-stranded to B-form DNA. This is presumably helix
initiation energy. Finally, adding an adjustment for salt gives the equation that the Tacalculator
uses:
T = (H – 5(KCal/K*Mole))/(S + (R * ln(1/(primer)))) + 16.6 log
(SaltMolarity)
10
Operation Manual | 93
Page 96
Chapter 6 Cr eating Protocols
No adjustment constant for salt concentration is needed, since the various parameters were
determined at 1 M NaCl, and the log10of 1 is zero.
The thermodynamic calculations assume that annealing occurs at pH 7.0. The Tmcalculations
assume that the sequences are not symmetrical and contain at least one G or C.
The oligonucleotide sequence should be at least 14 bases long to give reasonable Tmvalues.
Less than 14 bases uses the base pair counting method (see Table 12 that follows).
Table 12. Breslauer interaction constants
Interaction
AA TT 9.1 24 1.5
AT TA 8.6 23.9 1.5
AC TG 6.5 17.3 1.3
AG TC 7.8 20.8 1.6
TA AT 6 16.9 0.9
TT AA 9.1 24 1.9
TC AG 5.6 13.5 1.6
TG AC 5.8 12.9 1.9
CA GT 5.8 12.9 1.9
CT GA 7.8 20.8 1.6
CC GG 11 26.6 3.1
CG GC 11.9 27.8 3.6
GA CT 5.6 13.5 1.6
H S G
GT CA 6.5 17.3 1.3
GC CG 11.1 26.7 3.1
GG CC 11 26.6 3.1
94 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 97
Using the TaCalculator
To use the TaCalculator
1. To open the TaCalculator, do one of the following:
n If you are currently in the Protocol AutoWriter, click TaCalculator.
n In the Home window, select Tools > TaCalculator.
The TaCalculator dialog box appears.
Using the Ta Calculator
2. In the Forward Primer text box, type or paste the forward primer sequence.
Tip: You can also use the A, T, G, C buttons on the left side of the dialog box to enter the
sequence.
3. Type or paste the reverse primer sequence in the Reverse Primer text box.
4. Click Calculate.
The TaCalculator calculates and displays the Tmof each primer and the average Tmand T
values, for example:
a
Operation Manual | 95
Page 98
Chapter 6 Cr eating Protocols
If the primer Tmvalues are more than 4°C apart, the Protocol AutoWriter uses the lower primer
Tmvalue + 2°C as a basis for calculating the Tavalue, which you can further modify by
changing the enzyme and reaction speed.
The TaCalculator generates an annealing temperature for standard speed with iTaq DNA
polymerase. When using a different enzyme, the speed settings automatically adjust the Ta.
5. Do one of the following:
n If you opened the TaCalculator from the Protocol AutoWriter, click OK. You return to the
Protocol AutoWriter. The annealing temperature is automatically modified.
n If you opened the TaCalculator from the Tools menu, record the calculations and click
Cancel to close the calculator.
96 | CFX96 Dx and CFX96 Deep Well Dx Systems
Page 99

Chapter 7 Preparing Plates

A plate file contains information about run parameters such as scan mode, fluorophores, and well
contents. After the run, CFX Manager™ Dx software links the well contents to the fluorescence data
collected during the run and applies the appropriate analysis in the Data Analysis window. For
example, wells loaded with standard sample type are used to generate a standard curve.
CFX Manager Dx software provides two options for creating plates: The Plate Editor for real-time
PCR runs and the Setup Wizard for normalized gene expression analysis.
The Plate Editor includes the following features:
n Standard fluorophores and sample types to assign to plate wells
n Ability to set reference target and control sample for gene expression analysis
n Ability to edit plate setup before, during, or after a run
n Ability to save plate files for reuse
n Ability to print the plate file to a default printer
The Setup Wizard guides you through creating a plate layout for normalized gene expression
analysis. You can use the Setup Wizard before, during, or after a run.
Operation Manual | 97
Page 100
Chapter 7 Preparing Plates

Plate Editor Window

You use the Plate Editor to create custom plates or modify existing plates.
LEGEND
1. The menu bar provides quick access to File and Settings menu commands as well as plate
editing tools options.
2. The toolbar provides quick access to important plate loading functions.
3. The main pane displays the plate outline and the plate options as you apply them.
4. The right pane displays options that you use to customize your plate.
5. The bottom pane displays the plate type and provides quick access to viewing options.
98 | CFX96 Dx and CFX96 Deep Well Dx Systems
Loading...