instructions for use
Codefree
BLOOD GLUCOSE MONITOR
Step by step
GL44
Beurer GmbH • Söflinger Straße 218 • 89077 Ulm, Germany www.beurer.com
germanlengineering
Page 2
Page 3
Contents
1 Getting to know your instrument ................................................................................ 4
1.1 Delivery scope, replacements and accessories ..................................................................... 5
1.2 Functions of the device ..........................................................................................................6
1.3 Signs and symbols ................................................................................................................. 7
2 Warnings and safety notes ............................................................................................8
3 Description of units and accessories ..................................................................... 11
3.1 Blood glucose monitor ......................................................................................................... 11
3.2 Lancing aid and lancet needles ...........................................................................................11
3.3 Display symbols ................................................................................................................... 12
3.4 Test strips ............................................................................................................................. 12
4 Initial use and basic settings ...................................................................................... 14
4.1 Removing the battery insulation strips, replacing the batteries ...........................................14
4.2 Making and changing basic settings ...................................................................................14
5 Measuring blood glucose values ..............................................................................16
5.1 Preparing to take a blood sample ........................................................................................16
5.2 Taking a blood sample ......................................................................................................... 17
5.3 Reading out result and highlighting measurements ............................................................20
5.4 Post-processing and disposal .............................................................................................21
5.5 Evaluating measured blood glucose values ........................................................................22
5.6 Functional check with control solution ................................................................................23
6 Measurement memory ................................................................................................... 26
6.1 Displaying individual values ................................................................................................. 26
6.2 Displaying average blood glucose values ............................................................................ 27
6.3 Displaying average blood glucose values for highlighted values ........................................27
6.4 Deleting individual measured values from the memory .......................................................29
6.5 Deleting the entire memory for measured values ................................................................29
6.6 Reset to basic settings ........................................................................................................ 29
6.7 Transferring measurements to a PC .....................................................................................30
7 Storing, maintaining and disinfecting the device .............................................. 31
7.1 Cleaning ............................................................................................................................... 31
7.2 Disinfection ..........................................................................................................................31
8 What if there are problems? ........................................................................................ 32
9 Technical specifications ................................................................................................ 33
10 Comparison of measured values with laboratory value ................................35
11 Usage limits for specialist personnel from the healthcare sector ...........36
12 Guarantee and customer service .............................................................................39
Beurer GL44 mg/dL 3
Page 4
1 GettinG to know your instrument
Dear Customer,
Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested products for applications in the areas of heat, weight, blood pressure, blood glucose measurement, body temperature, pulse, gentle therapy, massage and air. Please read these Instructions for Use carefully and keep them for later use, be sure to make them accessible to other users and observe the notes they contain.
With kind regards Your Beurer team.
Getting to know your instrument
The GL44 blood glucose monitor is intended for fast and simple blood glucose measurement of fresh capillary blood samples, either for self-testing or in a clinical environment by trained personnel. It enables quick and easy blood glucose measurement, measured values to be stored and the aver­age of all measured values to be displayed to ideally assist you in monitoring diabetes. The test is performed exclusively externally (IVD). The large backlit display shows measured values clearly. The user-friendly design with handy test strips and a reduction in the controls to a few buttons guarantee simple, yet reliable measurements. The device can be connected directly to a PC using the USB cable provided. You can evaluate the measured values on your PC using blood glucose diary software and use the results to monitor your blood glucose values. Beurer blood glucose software is available as a free download at www.beurer.com
4 Beurer GL44 mg/dL
Page 5
1.1 Delivery scope, replacements and accessories
Check that the set packaging has not been tampered with and check the contents for completeness. Before use, ensure that all packaging materials are removed and that there is no visible damage to the unit or accessories. When in doubt, do not use the unit and contact your dealer or the customer service address provided.
A 1 Blood glucose monitor B 1 Lancing device with AST cap to take blood samples from alternative body parts C 10 test strips D 10 sterile lancet needles E 2 button cell batteries, 3 V CR2032 (already inserted) F 1 practical case G 1 USB cable
These Instructions for use, additional information
If the packaging has sustained considerable damage or the contents are incomplete, please return
the system to your retailer.
The blood glucose monitor, test strips and additionally available control solutions have been specially
designed to complement each other. For this reason, use only test strips and control solutions that have been approved for this blood glucose monitor.
Note
Use original manufacturer accessories only.
Beurer GL44 mg/dL 5
Page 6
Follow-up purchases
You can also obtain test strips, control solution and lancet needles without a prescription.
Item REF
50 test strips REF 464.15 Control solution LEVEL 3 and 4 REF 464.16 100 lancet needles REF 457.01
1.2 Functions of the device
This device is intended for measuring the blood glucose content in human blood. It is also suitable for self-testing at home. The blood glucose monitor enables you to quickly and simply:
Measure your blood glucose level,
Display, highlight and save measured values,
Display the average measured blood glucose value from the last 7, 14, 30 and 90 days,
Display the average of the highlighted measured blood glucose values from the last 7, 14, 30 and 90 days,
Set the time and date,
Transfer stored measured values to a PC for evaluation (additional accessories required).
The blood glucose monitor also includes the following monitoring functions:
Warning in the event of unsuitable temperatures.
Low battery display.
Low test strips warning.
Warning
Do not use the device to diagnose diabetes; it is intended for regular monitoring only.
Please consult your GP with regard to insulin doses.
6 Beurer GL44 mg/dL
Page 7
1.3 Signs and symbols
The symbols on the packaging, type plate of the blood glucose monitor and accessories represent the following:
In vitro diagnostics Manufacturer
Serial number
Temperature limit +2 °C to +30 °C
Not for re-use/For single use only
Please observe the Instructions for Use
PCT: certification symbol for products that are exported to the Russian Federation and members of the CIS
Green dot (Der Grüne Punkt):
Use by
German dual waste collection system
Maximum shelf life after initial opening in months
Batch designation
Art.-Nr.
Sterilised by radiation (lancets)
Warning, please observe ac­companying documents
Contents sucient for <n> tests
/
Order number
Unit of measurement for blood glucose value
Biohazard, risk of infection
In the Instructions for Use, the symbols represent the following:
Warning
Warning indicating a risk of injury or risk to your health/your patients’ health.
Important
Safety note indicating possible damage to the unit/accessory.
Note
Note on important information
Beurer GL44 mg/dL 7
Page 8
2 warninGs and safety notes
Risk of infection
All components of the blood glucose monitor and its accessories may come into contact with human blood and are therefore a possible source of infections.
Warning
This blood glucose monitor must display the blood glucose content in mg/dL. The unit of meas-
urement mg/dL accompanies each blood glucose value. Please contact our customer service if the device does not display mg/dL. You risk damaging your health if you measure your blood glucose value using a unit of measurement with which you are not familiar, misinterpret the values and therefore take incorrect measures.
When using the blood glucose monitor for various persons, please observe the generally applicable
regulations regarding disinfection, safety and contamination.
Medical carers and others who use this system on several patients must be aware that all products
or objects that come into contact with human blood must be handled, even after cleaning, as though they could transfer pathogens.
The lancing device is suitable for self-testing. Do not share the lancing aid or the lancet needles with
others or amongst various patients (risk of infection!).
Use a new, sterile lancet needle for each blood sample (for single use only).
General notes
Warning
Do not use the device in the vicinity of strong electromagnetic fields and keep it away from radio systems or mobile telephones.
Measuring the blood glucose content
Warning
The measurements taken by you are for your information only – they are no substitute for a medical
examination! Regularly consult your GP regarding your measured values. Do not alter the procedures prescribed by your GP.
Despite the simple usage of the Beurer GL44 monitor for self-monitoring of blood glucose levels, you
may possibly need to obtain instructions for using the system from your healthcare professional (for example, your GP, chemist or diabetes consultant). Only proper use will guarantee exact measure­ments.
A lack of water, high fluid loss, for example perspiration, frequent passing of water, severe hypoten-
sion (low blood pressure), shock or hyperosmolar hyperglycaemic non-ketotic coma may lead to incorrect measured results.
A hematocrit value between 20% and 60% has no significant influence on the measurement results.
An excessively high or low hematocrit value (proportion of red blood cells) may lead to incorrect
measurements. In the event of a very high hematocrit value (above 60%), the displayed blood glu-
8 Beurer GL44 mg/dL
Page 9
cose value may be too low; in the event of a very low hematocrit value (below 20%), it may be too high. Please consult your GP if you do not know your hematocrit value.
Do not use the test strips to measure blood glucose values on newborns.
Do not use NaF or potassium oxalate anticoagulants to prepare for venous blood samples.
Do not test any severely ill patients using this device.
Use fresh capillary whole blood only. Do not use serum or plasma.
Use capillary blood without squeezing the penetration area. If the area is squeezed, the blood is di-
luted with tissue fluid and this may lead to an incorrect measuring result.
Do not use the test strips above an altitude of 7010 m.
Very high levels of humidity may influence the test results. Relative humidity of more than 90% may
lead to inexact results.
Note
The Beurer GL44 mg/dL measuring system is intended to measure capillary whole blood.
Storage and maintenance
Warning
Store the blood glucose monitor and its accessories out of the reach of small children and pets.
Small parts, such as lancet needles, parts of the lancing device, batteries or test strips may be life­threatening when swallowed. If swallowed, immediately seek medical attention.
The test strip box contains desiccant, which may irritate the skin or eyes when inhaled or swallowed.
Keep the box out of the reach of children.
The blood glucose monitor is made from precision and electronic components. The accuracy of the measurements and service life of the device depend on its careful handling:
Protect the device and its accessories from impacts, humidity, dirt, strong temperature fluctuations
and direct sunlight. Do not store the device, test strips and control solution in your vehicle, in the bathroom or in a cooling appliance!
Do not drop the device.
Batteries/Saving measured values
Warning
Keep batteries out of the reach of children. Children may put batteries in their mouth and swallow
them. This may severely damage their health. If aected, immediately seek medical attention!
Normal batteries must not be charged, heated or thrown into an open fire (risk of explosion!).
Important
Do not disassemble or short-circuit the batteries.
Always replace all batteries at the same time and use batteries of the same type.
Leaking batteries may damage the device. If you do not intend to use the device for longer periods,
remove the batteries from the battery compartment.
Attention!
Leaked or damaged batteries may burn the skin. In this case, use suitable protective gloves.
Beurer GL44 mg/dL 9
Page 10
Note
The stored blood glucose values are retained when the batteries are replaced. If applicable, the date
and time must be reset after replacing the batteries.
Use lithium-ion batteries only.
Repairs
Note
Do not open the device. Failure to comply will result in voiding of the warranty.
Do not repair the device. Faultless functionality can no longer be guaranteed in this case.
Do not dismantle the lancing device into individual parts, except in the steps described in these in-
structions.
Please contact our customer service for repairs.
Disposal
Warning
It is essential to comply with the generally applicable safety precautions for handling blood when
disposing of materials of the blood glucose monitor. Dispose of all blood samples and materials with which you or your patients come into contact correctly in order to prevent injury and infection of other persons.
After use, dispose of test strips and lancets in a puncture-proof container.
Note
Batteries must not be disposed of with the household waste. As a consumer, you are required by law to recycle used batteries. You can recycle your old batteries at public collection points in your community or wherever batteries of this type are sold.
The codes below are printed on batteries containing harmful substances:
Pb = Battery contains lead,
Cd = Battery contains cadmium,
Hg = Battery contains mercury.
For environmental reasons, do not dispose of the device with the household waste at the end of its service life. Please dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local au­thorities responsible for waste disposal.
10 Beurer GL44 mg/dL
Page 11
3 description of units and accessories
3.1 Blood glucose monitor
Front
1 Receptacle for test strips, illuminated 2 Display 3 Highlight button 4 On/O button 5 PC port 6 “+ –” rocker switch
6
1
2
3
4
5
Rear
1 Battery compartment
1
(bottom side)
2 Reset button
CR2032
CR2032
reset
2
3.2 Lancing aid and lancet needles
1 Cap 2 AST Cap 3 Protective lancet disc 4 Sterile lancet needles 5 Lancet holder
1
2
3 6 Safety switch 7 Trigger 8 Tensioning device
Beurer GL44 mg/dL 11
6
7
8
4
5
Page 12
3.3 Display symbols
1 Date 2 Symbols for highlighting measurements 3 Measured value display, HI, LO display, average
blood glucose value, Err 4 Blood droplet symbol 5 Speaker symbol 6 Hand symbol 7 Test strip symbol 8 Blood glucose unit mmol/L – not functional 9 Blood glucose unit mg/dL 10 Replace battery symbol 11 Memory symbol 12 Temperature symbol 13 Time
Note
The blood glucose monitor is supplied with the following basic settings:
Blood glucose unit: mg/dL
Acoustic signal on
Backlighting on
1
2
3
4 5
6
7
3.4 Test strips
Front Rear
1
2
13 12 11 10
9 8
3
1 Gap for blood input 2 Handle 3 Contacts
12 Beurer GL44 mg/dL
You can identify the rear by the contact tracks.
Page 13
Insert the test strip into the device so that the contacts are pointing inside the slot. Please ensure that the front of the test strip is facing you.
Note
Carefully read the following information on handling and storing your test strips. Please observe all notes to ensure that the test strips supply accurate measured results.
Warning
Use each test strip only once and for one patient only!
Handling test strips
Note
Immediately securely close the test strip box after taking out a test strip.
Do not use the test strips if they have expired. The use of expired test strips may lead to incorrect
measurements. The expiry date is printed on the box next to the hourglass symbol .
Test strips expire three months after the box is opened. Make a note of the expiry date (opening date
+ 3 months ) on the label. The shelf life is limited to the expiry date (see date next to the hourglass symbol ).
Do not use the test strips if one of the expiry dates has passed ( / ).
You can touch the test strip at any point with clean, dry hands.
Use the test strip for measurement immediately after removing it from the box.
Do not bend, cut or otherwise modify the test strips.
Do not use test strips that have come into contact with fluids.
Storing test strips
Note
Keep the test strips in a cool, dry place above +2 °C and below +30 °C. Do not expose the test strips
to direct sunlight or heat. Do not store in your vehicle, in the bathroom or in a cooling appliance.
Permitted relative air humidity below 90 %.
Keep the test strips in the original box only – never use other containers.
Beurer GL44 mg/dL 13
Page 14
4 initial use and basic settinGs
4.1 Removing the battery insulation strips, replacing the batteries
Note
Two batteries are included in the deliv-
ery scope of the blood glucose monitor. These have already been inserted into the battery compartment.
Remove the insulation strip before initial
use.
CR2032
Remove the battery compartment lid on the bottom side of the device.
1
Remove all batteries when replacing batteries. The device retains the date and time as long
2
as one battery is still inserted. If applicable, reset the date and time (see “Making and chang ing basic settings” on page 14).
Insert two new CR 2032 3 V batteries. Make sure that the batteries are inserted with the cor-
3
rect polarity, according to the label. Please observe the graphic in the battery compartment. Subsequently close the battery compartment lid carefully.
4
Note
The batteries are almost empty if the replace battery symbol appears. Replace both batteries
as soon as possible.
If “LP” appears on the display, the battery power level is so low that no more measurements are
possible.
4.2 Making and changing basic settings
Remove the batteries and reinsert them. Alternatively, press the “+” button
1
and the On/O button for a minimum of five seconds. An acoustic signal sounds. The year display flashes.
CR2032
reset
-
14 Beurer GL44 mg/dL
Page 15
Setting the date and time
2
Note
You must set the date and time. Otherwise, you will not be able to save your measured values
correctly with a date and time to retrieve them later.
• The time is displayed in the 24-hour format. Set the year (calendar to 2099) by pressing the “+” or “–” button. Confirm by pressing the On/
O button [4]. The day display flashes. Proceed as described above for the month, day, hour and minute. “ ” and “ ” will be displayed. The background of the display is simultaneously illumi­nated for a few seconds.
Switch backlighting on/o
3
To switch the blue backlighting o, press the “+” or “–” button. “ ” and “ ” will be displayed. Confirm by pressing the On/O button [4]. “  ”, “ ” and the speaker symbol are displayed.
Switching the acoustic signal on/o
4
To switch the acoustic signal o, press the “+” or “–” button. “
” and “ ” is displayed. The speaker symbol is no longer shown in the display. Confirm by pressing the On/O button [4].
Mem
” and “ ” are displayed.
Deleting stored values
5
Proceed as follows to delete stored values:
Press the
Confirm by pressing the On/O button
To irreversibly delete your values, press the On/O button again to confirm. “Mem”, “ ” and
“ ” are displayed.
If you do not wish to delete your values, press the “+” or “–” button. “ played again. Confirm by pressing the On/O button.
The blood glucose monitor is now ready for operation
6
“+”
or
“–”
button. “Mem” and “ ” are displayed
[4]
. “Mem” and “ ” flash in the display.
Mem
” and “ ” are dis-
Beurer GL44 mg/dL 15
Page 16
5 measurinG blood Glucose values
Warning
If the protective disc on a lancet needle has already been removed, do not use the lancet needle.
If you have dropped the lancing aid with an inserted lancet needle, carefully pick it up and dispose
of the lancet.
Important
Use the lancing aid only with lancet needles by the same manufacturer. The use of other lancet nee-
dles may impede the functionality of the lancing aid.
If you are using a third-party lancing aid, please read the respective Instructions for Use.
5.1 Preparing to take a blood sample
Choose a part of the body from which to take a blood sample. The lancing aid is intended
1
for taking blood samples from the fingertip or other body parts such as the palm of the hand, forearm or upper arm. We recommend taking blood samples from the fingertip. To make the procedure as painless as possible, do not take samples directly from the centre of the finger­tip, but slightly to either side.
Warning
In the event of suspected hypoglycaemia: take blood from the fingertip only.
Reason: changes to the blood glucose level can be determined quickly in blood samples taken from the fingertip.
Blood samples from the fingertip and samples taken from another body part (AST) may lead
to completely dierent measured results. It is essential to seek medical advice before taking measurements from other body parts.
Prepare the following items: blood glucose monitor, test strip box, lancing aid, sterile lancet
2
needle. You will also need the AST cap for measurements at other body parts. Wash your hands with soap and warm water before taking a blood sample. This not only en-
3
sures optimal hygiene but also encourages good blood circulation at the puncture area on the finger. Carefully dry your hands. Also ensure that the puncture area is hygienically clean when taking blood samples from other body parts (AST).
Warning
If you have used alcohol for cleaning, ensure that the area has fully dried prior to measuring.
16 Beurer GL44 mg/dL
Page 17
5.2 Taking a blood sample
Warning
Change the puncture area for each measurement, e.g. a dierent finger or the other hand. Repeat-
edly using the same area may cause inflammation, numbness or scarring.
Without the cap, there is a danger of injury from the exposed lancet.
Do not use the AST cap to take blood samples from the finger.
Do not squeeze your finger to obtain a larger drop of blood. If squeezed, the blood is diluted with
tissue fluid and this may lead to an incorrect measuring result.
Please note that insucient blood circulation at the puncture area, e.g. caused by cold temperatures
or illness, may lead to incorrect results.
Important
Do not apply any blood samples or control solutions to the test strips before placing these in the measuring device.
Remove the cap from the lancing device.
1
Place a sterile lancet needle into the lancing aid and secure the lancet.
2
Rotate the lancet to remove the protective disc while holding the shaft
3
of the lancet. Keep the protective disc to safely dispose of the used lancet needle after taking a blood sample.
Use various caps depending on the area from which you take the
4
sample:
Fingertip: Cap Other body parts: AST cap (transparent)
Place the selected cap on the lancing device.
Beurer GL44 mg/dL 17
Page 18
Setting the penetration depth
5
You can set seven dierent penetration depths on the lancing de­vice. The penetration depth is displayed by the markings in the cap.
soft or thin skin
normal skin
thick or callous skin Turn the moving upper part of the cap until the desired penetration depth is displayed.
Note
When using the AST cap, penetration depth does not need to be set. Pull the tensioning device back until it audibly engages. Release the
6
tensioning device. This automatically snaps back into the default po­sition. The lancing device is now ready for use.
Put the already prepared lancing aid to one side and prepare the device for measurement.
7
Take a test strip from the box and immediately close it again.
8
Hold the blood glucose monitor so that the display is facing you.
9
Firmly insert a test strip into the device with the contacts first. Please
10
make sure that the front is facing you. You can touch the test strip at any point with clean, dry hands. Use the test strip within three minutes of taking it from the box.
The device automatically switches on and shows the start display. The device is ready for
11
measurement as soon as the hand and the droplet symbol flash.
18 Beurer GL44 mg/dL
Page 19
You may now use the lancing aid to take a blood sample. Please en-
12
sure that the blood remains a droplet and is not spread. Immediately use the blood droplet to take a measurement.
Blood sample from the fingertip
The best penetration areas are the middle finger and the ring finger. Firmly position the lancing aid slightly to the side of the centre of the fingertip. Press the trigger. Remove the lancing aid from the finger. A round drop of blood of at least 0.6 microlitres (cor­responds to approx. 1.4 mm, original size: ) must have formed.
Blood samples from other body parts (AST)
Find a soft area away from bones, without any visible blood vessels and with as little hair as possible. Warm the penetration area to en­hance circulation by gently massaging the area. Press and hold the lancing aid against the penetration area for a few seconds and sub­sequently press the trigger. Continue to hold the lancing aid against the skin until a round drop of blood has formed under the cap. Maintain the pressure until the drop of blood reaches a size of at least 0.6 microlitres (corresponds to approx. 1.4 mm, original size: ) Carefully lift the lancing aid o the skin.
Note
Only take blood from other areas at the following times:
On an empty stomach (more than two hours after most recent meal).
At least two hours after administration of insulin.
At least two hours after physical activity.
Please also note the following:
If the blood glucose test results do not match how you feel, carry out another test using blood
from the fingertips.
DO NOT change your treatment purely on the basis of a measurement that was carried out
using blood taken from an alternative area. Carry out another test with blood from the finger­tips in order to confirm the test result.
If you often fail to notice that you have a low blood glucose level, carry out a test using blood
from the fingertips.
If there is an insucient amount of blood, repeat steps 1 to 12 with a greater penetration depth.
13
Note
When using the AST cap, repeat measurement at dierent parts of the body.
Beurer GL44 mg/dL 19
Page 20
14 Hold the blood input gap (at the tip of the test strip) to the drops
OK
of blood until the gap is completely filled and the measuring de­vice in the display starts counting backwards. Do not press the penetration area (fingertip or other body parts) to the test strip. The blood must not be spread. The blood is sucked into the gap.
Note
Error message “002” appears on the display if the gap was not correctly and suciently filled with blood. Repeat the measurement using a new test strip and a greater penetration depth.
Note
Do not apply blood to the sides of the test strips
Do not add blood later if the device does not start measurement. Remove the test strip and end this
test. Use a new test strip.
The device switches itself o if the test strip has already been inserted into the device but no blood
is added to the test strip within two minutes. Briefly remove the test strip and reinsert it so that the device automatically switches itself back on.
Contact our customer service if you are unable to fully cover the test strip in blood.
If you are measuring in a dark environment, press the On/O button to switch the device on. The test
strip light switches on and makes it easier to insert the test strip. The backlight is also switched on in the results display.
5.3 Reading out result and highlighting measurements
Evaluating result
As soon as the gap is suciently filled with blood, the device performs the blood glucose measure ment. The blood glucose monitor counts down for approx. five seconds. The measured result is subsequently shown on the display.
Read the measured values. For explanations and measures for the measured values, see chapter “5.5 Evaluating measured blood glucose values”. If an error message is displayed, read chapter “8. What if there are problems?”
Highlighting measured values
You have the following options to highlight measured values.
Before meals.
After meals.
General highlight (e.g. after exercising).
Highlighting measured values enables you, your GP or diabetes consultant to better check your blood glucose values. For example, you can display the average values of all measurements taken before meals.
-
20 Beurer GL44 mg/dL
Page 21
The measured value can be highlighted as soon as it is displayed. You cannot highlight it later.
1
2
For this purpose, briefly press the highlight button [3]. a) Pressing once highlights the value with b) Pressing again highlights the value with . c) Pressing yet again highlights the value with . d) Pressing once more removes the highlight.
The selected highlight is stored in the memory of the device when it is switched o.
.
5.4 Post-processing and disposal
Remove the test strip from the device and carefully dispose of it according to the currently
1
applicable regulations to avoid infecting others. Carefully undo the cap from the lancing device and remove it.
2
Place the stored protective disc flat on a hard surface. Stick the tip
3
of the needle into the protective disc so the needle is covered. Take care not to touch the used lancet.
Press and hold the trigger and pull the tensioning device with your
4
other hand. This causes the lancet to drop out of the holder. Carefully dispose of all blood samples and materials that came into contact with you or your patient. Dispose of the lancet in a puncture­proof container. This prevents injuries and the infection of others.
Replace the cap onto the lancing device.
5
Beurer GL44 mg/dL 21
Page 22
5.5 Evaluating measured blood glucose values
Your blood glucose monitor can process values between 20 and 630 mg/dL. The “ ” warning is displayed for measured results below 20 mg/dL. The “ ” warning is displayed for measured results above 630 mg/dL.
Warning
If you suspect that the blood glucose results are incorrect, first repeat the test and, if applicable,
perform a functional test using control solution. Seek medical advice if doubtful results persist.
Immediately seek medical attention if your symptoms are not in line with your measured blood glu-
cose values and you have observed all instructions for the Beurer GL44 blood glucose monitoring system.
Do not ignore symptoms of too high/low blood glucose levels. Seek medical attention.
Blood glucose
The following table provides a classification of blood glucose values according to the diabetes guide­lines of the Deutsche Diabetes Gesellschaft (DDG, German diabetes association).
Time of blood glucose measurement
With an empty stomach
Whole blood, capillary
(haemolysed)
Plasma venous
Two hours after a meal
Source: Deutsche Diabetes Gesellschaft (DDG) 2009
Normal blood
Suspicion Diabetes
glucose values
Below 90 mg/dL 90 –109 mg/dL ≥ 110 mg/dL Below 100 mg/dL 100 –125 mg/dL ≥ 126 mg/dL Below 140 mg/dL 140 –199 mg/dL ≥ 200 mg/dL
22 Beurer GL44 mg/dL
Page 23
Evaluating critical measured values
Display Blood glucose Action
Very low blood glucose level
Below 20 mg/dL Low blood glucose level
Below 70 mg/dL
Immediately seek medical attention.
Have a suitable snack. Follow your GP’s instructions.
High blood glucose level On an empty stomach above 100 mg/dL
Two hours after a meal Above 140 mg/dL
High blood glucose level, possibly ketones
Above 240 mg/dL Very high blood glucose level
Above 630 mg/dL
If this high value persists two hours af­ter your last meal, this may indicate hy­perglycaemia. Seek medical attention to coordinate any measures, if applicable.
Perform a ketone test. For this purpose, seek medical attention.
Take another measurement using a new test strip. If the result is identical: seek medical attention immediately.
5.6 Functional check with control solution
The control solution is used to test the entire blood glucose monitoring system. This helps to de­termine whether the monitor and the test strips are working optimally together and whether the test was performed correctly. Perform the control solution test if you suspect that the blood glucose monitor and/or the test strips could be faulty or if you have repeatedly measured unusual blood glucose values. Also test the blood glucose monitor if it has fallen down or is damaged. The control solution is available separately. Please observe the additional notes in the Instructions for Use of the control solution for the test.
Important
Do not use third-party control solution. Correct functioning of your measuring device can only be
tested using Beurer control solutions (LEVEL3 + LEVEL4).
Control solution measurements: When using the device, specialist personnel must follow statutory
guidelines.
Do not apply any blood samples or control solutions to the test strips before placing these in the
measuring device.
Beurer GL44 mg/dL 23
Page 24
Performing a functional test using control solution
Warning
To obtain correct results, the measuring device, the test strip and the control solution must be the same temperature. For the “Functional test using the control solution”, the temperature is to be be tween 20 °C and 26 °C.
Hold the blood glucose monitor so that the display is facing you.
1
Insert a test strip into the slot on the blood glucose monitor with the contacts first. Make sure
2
that the front of the test strip is facing you (see “Test strip” on page 12). The device automatically switches on and briefly shows the start display. The device is ready
3
for measurement as soon as the hand and the symbol flash. IMPORTANT: Control solutions and blood react to temperature influences in dierent ways.
It is therefore of vital importance that control solution measurement is always performed in control solution mode. If this mode is not used, results may be obtained that are outside the target range.
Note
Press the rocker switch “+” or “–” to change to the control mode. “ play. This means that the result value is not stored in the memory, therefore not influencing your measured value statistics. Press “+” or “–” again. “ ” disappears from the display and the value is stored normally in the memory.
A clean surface is required to perform a correct functional test.
4
Shake the control solution well before use. Undo the cap and press two drops next to each other on the clean surface without touching them. Use the second drop for the measurement.
Note
Do not apply the drop directly to the test strip to avoid contaminating the remaining control solution in the bottle by touching the test strip with the tip of the bottle.
Hold the input gap (at the tip of the test strip) to the drops of control solution until the gap is
5
completely filled and the measuring device in the display starts counting backwards. If the gap is suciently filled with solution, the device performs a measurement. The devices counts down for approx. five seconds. The measured result is subsequently shown on the display.
Check whether the result is within the specified range of results for the control solution. This
6
range of results is printed on the test strip box.
” is shown on the dis-
-
24 Beurer GL44 mg/dL
Page 25
Expected results
At room temperature, the measured values of the test using the control solution should be within the range printed on the test strip box in approx. 95 % of all tests.
Warning
The result range printed on the test strip box applies only to the control solution. This is not a rec ommended value for your blood glucose level.
If measured values are outside the specified range, check the following possible causes:
Cause Action
The first drop of control solution was not dis-
posed of.
• The tip of the bottle was not cleaned correctly.
The bottle was not shaken well enough.
Control solution and/or test strips have passed their expiry date or are contaminated.
The control solution, test strips or blood glucose monitor are too warm or too cold.
The test strips and control solution were kept at temperatures and in humidity outside the speci­fied range.
Damaged test strips. For example
Test strips that were exposed to fresh air for
too long.
The test strip box was not closed completely. Old test strips. Open a new box of test strips. Repeat the test. There is a problem with the blood glucose moni-
tor. Functional test was incorrectly performed. Repeat the test and follow the instructions.
Rectify the cause and repeat the test.
Repeat the test using a new bottle of control solution and/or new test strip.
Bring the control solution, test strips and blood glucose monitor to room temperature (+20 °C to +26 °C) and repeat the test.
Repeat the test using new/correctly stored test strips and control solution.
Repeat the test using a new test strip and/or a new box of correctly stored test strips.
Contact our customer service.
-
Warning
Do not use the system to measure your blood glucose level if you are repeatedly provided with
measured values outside the specified range when using control solution. Contact our customer service.
Beurer GL44 mg/dL 25
Page 26
6 measurement memory
For each measurement, your blood glucose value is automatically saved with the date and time un­less “ ” was activated for a blood glucose measurement using control solution. The memory can store a maximum of 480 measured values. If the memory is full, the oldest value is replaced by the most recent value. You can call up every individual measured blood glucose value. You can also calculate and display the average value for the last 7, 14, 30 and 90 days.
Note
If you have already saved measured values and you reset the date, the average values are calculated
as from the new period.
“---” indicates an empty memory for measured values. Press the On/O button to switch o the de-
vice.
6.1 Displaying individual values
Individual values of the last 480 measurements are displayed. The most recent measured value is dis­played first, and the oldest last. The blood glucose monitor simultaneously shows the date and time.
Switch the monitor on using the on/o button [4]. The initial display is shown briefly. Press
1
the “+” or “–” rocker switch [6]. “
Mem
2
” and the number of saved blood glucose tests are displayed briefly (Image 1). The display then changes to the saved values including the measurement unit, date, time, “
Mem
” and any possible highlighting. (Image 2).
Image 1 Image 2 Pressing the rocker switch “–” [6] will each time display the memory space number and then
3
the previous measured value. You can display a maximum of 480 previous measurements. You can cancel the process at any time. For this purpose, press the On/O button or wait
4
until the device switches itself o automatically after two minutes.
26 Beurer GL44 mg/dL
Page 27
6.2 Displaying average blood glucose values
You can display the average measured blood glucose value from the last 7, 14, 30 and 90 days.
Switch the monitor on using the on/o button [4]. The initial display is shown briefly. Press
1
the “+” rocker switch twice [6]. The measurement unit of the blood glucose value, “ ” and the average value are displayed.
Press “+” [6] repeatedly to display the average value for 7, 14, 30 and 90 days.
2
You can cancel the process at any time. For this purpose, press the On/O button or wait
3
until the device switches itself o automatically after two minutes.
Pos. Explanation
1 Average value 2 Time period, e.g. 7 3 Number of saved values used to calculate the aver
age
3
-
2
1
6.3 Displaying average blood glucose values for highlighted values
You can display the average measured blood glucose value for highlighted values from the last 7, 14, 30 and 90 days.
Switch the monitor on using the on/o button [4]. The initial display is shown briefly. Press
1
the “+” rocker switch twice [6]. The measurement unit of the blood glucose value urements are displayed.
and the average value of all meas-
Beurer GL44 mg/dL 27
Page 28
Press “+” [6] repeatedly to display the average value of all measured values for 14, 30 and
2
90 days. After the display of the 90-day average value of all measured
values
the seven-day average for values measured “before meals”,
the symbol,
the unit of measurement for blood glucose values and
are shown on the display.
Press “+” [6] repeatedly to display the average blood glucose level from the last 14, 30 and 90 days “before meals”
After displaying the average value for 90 days “before meals”
the seven-day average for values measured “after meals”,
the symbol,
the unit of measurement for blood glucose values and
are shown on the display.
Press “+” [6] repeatedly to display the average blood glucose level from the last 14, 30 and 90 days “after meals”
After displaying the average value for 90 days “after meals”
the average for the last seven days of values highlighted as “gen-
eral”,
the symbol,
the unit of measurement for blood glucose values and
are shown on the display.
28 Beurer GL44 mg/dL
.
.
Page 29
Press “+” [6] repeatedly to display the average blood glucose level from the last 14, 30 and
2
90 days for values highlighted as “general” . You can cancel the process at any time. For this purpose, press the On/O button or wait
3
until the device switches itself o automatically after two minutes.
Tip: Speed function
You are in the measurements memory. By pressing the highlight button [3], you can switch to the dierent 7-day average values. In this way, you can get to the desired average value more quickly. For example:
6.4 Deleting individual measured values from the memory
Press the buttons as described in 6.1 until the value appears that you would like to delete.
1
Press and hold the On/O button. Additionally press the “–” button for two seconds.
2
”, “Mem” and the number of the value to be deleted flash. Press the On/O button again. “ ”, the number of the value to be deleted, “Mem” and “ ” are displayed. If you do not wish to delete the value, briefly press the “+” or “–” button. You are taken back to the memory spaces.
The device subsequently displays that the following measured value has moved to the memory
3
space of the deleted measured value.
6.5 Deleting the entire memory for measured values
You have two options to delete the measured value memory. Option 1: Proceed as described in “4.2 Making and changing basic settings”. After explaining how to set the beep, the section explains how to delete the measured value memory. Option 2 is described in the following chapter.
6.6 Reset to basic settings
The blood glucose monitor must be switched o.
1
Remove the battery compartment lid.
2
Press the ‘RESET’ button for 1 second. All settings and measured values are subsequently
3
deleted.
Beurer GL44 mg/dL 29
Page 30
Close the battery compartment lid again.
4
The blood glucose monitor is now in the settings mode.
5
6.7 Transferring measurements to a PC
The GL44 blood glucose monitor includes a PC interface [5] that enables you to transfer measured values saved on the device to a PC (for the position of the port, see page 11). Beurer blood glucose software is available as a free download at www.beurer.com. This software enables you to evaluate your measured values, add insulin doses and print results. The software helps you and your GP to better monitor your blood glucose level. For more information, please read the instructions for use of the blood glucose diary software (also available to download). There you will find all the information required for the data transfer and a de tailed description of the software.
The GL44 monitor is also compatible with Diabass and SiDiary.
Note
An eective evaluation is only possible if you have correctly set the date and time (see page 14).
During data transfer, no measurements can be taken.
The measurements remain saved on the blood glucose monitor after they have been transferred to
the PC.
Important
Use only the USB cable provided for data transfer. Otherwise you may damage your blood glucose monitor or PC.
Preparation
Position the blood glucose monitor near to your PC.
Connect the GL44 monitor to your PC using the USB cable provided.
Install the blood glucose diary software on your PC as described in the instructions for use of the
software.
Transferring measurements
The blood glucose monitor must be switched o.
1
Insert the larger, flat USB connector of the connec­tion cable into a USB port on your PC. Insert the mini USB connector into the integrated PC interface socket on the GL44 monitor.
” is shown on the display of the blood glucose
2
monitor. The blood glucose monitor is now ready for the data transfer.
Follow the information on data transfer and evaluation provided in the software and the in-
3
structions for use of the software.
-
30 Beurer GL44 mg/dL
Page 31
7 storinG, maintaininG and disinfectinG the device
Storing
Keep the Beurer GL44 blood glucose monitor in the case supplied after each measurement and do not expose it to direct sunlight.
Note
Do not store the device, test strips and control solution in your vehicle, in the bathroom or in a cool­ing appliance!
Retain these Instructions for Use.
Remove the batteries if you do not intend to use the device for a prolonged period of time.
7.1 Cleaning
Device
Only clean the device when it is switched o. Clean the surface of the device using a soft, slightly damp cloth (water or a mild cleaning solution). Dry the device using a lint-free cloth. Make sure that moisture does not enter the test strip insertion slot. Do not spray cleaning agent di rectly on the device. Do not submerge the device in water or any other fluids and make sure that no fluids can enter the device.
Lancing device
Clean the surface of the lancing device using a soft, slightly damp cloth (water, a mild cleaning so lution or rubbing alcohol). The lancing device must not be immersed in water or other liquids or be cleaned in the dishwasher. Dry the lancing aid using a lint-free cloth.
7.2 Disinfection
Device
Please comply with the generally applicable guidelines on disinfection when using the device on dif ferent persons. Do not submerge the device in disinfection solutions or any other fluids and make sure that no fluids can enter the device.
Note
The blood glucose monitor is made of precision components. The accuracy of the measurements and service life of the device depend on its careful handling:
Protect the device from impacts and do not drop it.
Protect the device from damaging factors such as moisture, dirt, dust, blood, control solution or
water, strong temperature fluctuations, direct sunlight and extreme cold.
Do not use the device in the vicinity of strong electromagnetic fields, radio systems or mobile tel-
ephones.
-
-
-
Beurer GL44 mg/dL 31
Page 32
8 what if there are problems?
Display messages on batteries and blood glucose measurement
No. Cause Solution
LP Batteries empty. Replace all batteries.
Temperature of the measuring environ-
Ht
ment, blood glucose monitor or test strip above the permitted range.
Temperature of the measuring environ-
Lt
ment, blood glucose monitor or test strip below the permitted range.
Used or contaminated test strip inserted.• Insert unused test strip that has not ex-
Err
Err001
System error. Remove batteries, reinsert batteries. Contact
Err002
Insucient amount of blood on the test strip.
The hematocrit value is outside the range
Err003
20 to 60%
Err005
System error. Remove batteries, reinsert batteries. Contact
Unknown error messages. Remove batteries, reinsert batteries. Contact
Repeat the test using a new test strip as soon as the measuring environment, blood glucose monitor and test strips have reached room temperature (+20 °C to +26 °C).
Repeat the test using a new test strip as soon as the measuring environment, blood glucose monitor and test strips have reached room temperature (+20 °C to +26 °C).
pired.
Repeat the blood glucose measurement.
our customer service if the problem persists. Repeat the measurement using a new test
strip. Repeat the measurement using a new test
strip. Contact Customer Services if the prob­lem persists.
our customer service if the problem persists.
our customer service if the problem persists.
Problem: device does not switch on
Cause Solution
Batteries empty. Replace batteries. Incorrectly inserted or missing batteries.
32 Beurer GL44 mg/dL
Check whether the batteries have been inserted correctly (see “Inserting and replacing the bat­teries” page 14).
Page 33
Cause Solution
Test strip inserted incorrectly or not completely.
Device faulty. Contact customer service.
Problem: the test does not start after inserting the test strip into the device and applying blood
Cause Solution
Insucient amount of blood or test strip not filled correctly.
Faulty test strip. Repeat the test using a new test strip. Blood was applied while the device was switched
o. The basic settings of the device were changed
and the changes were not completed (see “Mak­ing and changing basic settings” on page 14).
Device faulty. Contact customer service.
Firmly insert the test strip into the slot on the device with the contacts first. Make sure that the front of the test strip is facing you (see “Test strip” on page 12).
Repeat test using a new test strip and a larger drop of blood. Please note the correct filling of the test strip (see page 20).
Repeat the test using a new test strip and only apply blood after and flash.
Remove the test strip and press the “On/O” button until “ ” is displayed. Repeat test.
9 technical specifications
Dimensions (W x H x D) 52 x 95 x 16 mm Weight 44 g incl. batteries Power supply 2 x 3 V CR 2032 button cell batteries
Battery life
Measured value memory 480 measured values with date/time
Average values for 7, 14, 30, 90 days Automatic switch-o Two minutes after last actuation Storage/
transport temperature Operating ranges Temperature: +10 °C – +40 °C
Measuring range, glucose Glucose: 20 – 630 mg/dL
500 measurements with backlighting 1000 measurements without backlighting
Data retained when batteries are changed
Temperature: +2 °C – +30 °C Relative humidity: < 90 %
Relative humidity: < 90 % non-condensing
Beurer GL44 mg/dL 33
Page 34
Blood sample capillary whole blood Amount of blood 0.6 microlitres Blood glucose measuring duration Approx. 5 seconds Calibration Plasma Test procedure Amperometric bio sensor Using the device Suitable for self-testing System function test Upon every switch-on
EMC
This device complies with the European standard EN 61326 and is subject to particular precautions with regard to electromagnetic compatibility. Please note that portable and mobile HF communica tion systems may interfere with this unit. For more details, please contact our customer service at the address indicated.
Test strip functionality
Test strips enable a quantitative measurement of the glucose level in fresh whole blood. When the gap for taking blood comes into contact with a drop of blood, it is automatically filled by simple capillary action. The blood is sucked into the absorbing gap on the test strip and the blood glucose monitor measures the blood glucose level in the blood. The test is based on the measurement of an electric current that is generated by the chemical reaction of the glucose with the enzyme glucose dehydrogenase (Aspergillus oryzae) on the strip. During the reaction, a mediator transports electrons through the electrode surface and so gener­ates a current. The blood glucose monitor analyses this current. The current flow is proportional to the glucose con­tent in the blood sample. The results are shown on the blood glucose monitor display. Only a small amount of blood is required (0.6 microlitres) and measurement duration is approx. five seconds. The test strip detects blood glucose values from 20 to 630 mg/dL.
Chemical components of the test strip sensor
FAD glucose dehydrogenase 6 %
Potassium ferricyanide 56 %
Non-reactive components 38 %
Control solution functionality
The control solution contains a fixed amount of glucose that reacts with the test strip. A test with con­trol solution is similar to a blood test. However, control solution is used instead of blood. The measured result using control solution must be within the result range. This range is printed on each test strip box.
Chemical composition of the control solution
The control solution is a red solution with the following D-glucose level (in percentage shares): Ingredients Control solution LEVEL 3 Control solution LEVEL 4
D-glucose 0.14 % 0.37 % Non-reactive components 99.86 % 99.63 %
-
34 Beurer GL44 mg/dL
Page 35
Standards
The Beurer GL44 monitor corresponds to European guidelines IVD (98/79/EC) and MDD (93/42/EC).
10 comparison of measured values with laboratory value
Precision
Three lots of the GL44 blood glucose test strips have been tested to assess the precision of the GL44 blood glucose measuring system. This includes a repeat assessment using venous blood and a labora­tory precision assessment using the control material. The blood glucose content of the venous blood samples ranges from 42.7 to 418.0 mg/dL and control material from three concentrations is used.
Results of the repeat precision measurements
Sample Venous blood
(mg/dL)
1 42,7 36,0 2,0 5,6 2 62,0 59,2 3,5 5,9 3 120,5 127,1 4,1 3,2 4 201,0 213,8 6,7 3,1 5 316,5 329,9 10,1 3,1 6 418,0 433,5 14,5 3,3
Results of the intermediate precision measurement
Sample Control material
(mg/dL)
1 70,0 71,3 1,0 1,4 2 135,6 136,3 1,4 1,1 3 351,5 350,8 2,8 0,8
Grand mean value (mg/dL)
Grand mean value (mg/dL)
Pooled standard de­viation
Pooled standard deviation
Pooled coecient ofvariation (%)
Pooled coecient of variation (%)
System accuracy
The GL44 blood glucose monitor in comparison with the YSI. Three lots of GL44 blood glucose test strips have been tested to assess the system accuracy of the GL44 blood glucose measuring system and to compare it with the reference method in which capil lary whole blood concentrations of 34.4 to 442.8 mg/dL have been used.
Beurer GL44 mg/dL 35
-
Page 36
Result of the system accuracy for glucose concentrations <100 mg/dL (<5.55 mmol/L)
Within ±5 mg/dL (Within ±0.28 mmol/L)
55/180 (30,6%) 111/180 (61,7%) 175/180 (97,2%)
Results of the system accuracy for glucose concentrations ≥100 mg/dL (≥5.55 mmol/L)
Within ±5% Within ±10% Within ±15%
220/438 (50,2%) 357/438 (81,5%) 422/438 (96,3%)
Results of the system accuracy for combined glucose concentrations between 34.4 mg/dL (1.9 mmol/L) and 442.8 mg/dL (24.6 mmol/L).
Within ±15 mg/dL or ±15% (Within ±0.83 mmol/L or ±15%)
597/618 (96,6%)
In comparison to the YSI, the GL44 met the EN ISO 15197:2013 standard, whereby 95% of the blood glucose values measured have to fall within the following zones: either ±0.83 mmol/L (±15 mg/dL) of the measured average value when using the reference measuring procedure for blood glucose con­centrations <100 mg/dL (<5.55 mmol/L) or ±15% for blood glucose concentrations of ≥100 mg/dL (≥5.55 mmol/L). 99% of the individual measured blood glucose values must fall within zones A and B of the Consensus Error Grid (CEG) for diabetes type 1.
Performance evaluation by the user
A study to assess the glucose values of blood samples of capillary blood from the fingertips, which were obtained from 103 individuals that had no special training, produced the following results:
96.7% within ±15 mg/dL (±0.83 mmol/L) and 95.9% within ±15% of the values obtained in the medi­cal laboratory with glucose concentrations of at least 100 mg/dL (5.55 mmol/L).
You will find further details and information regarding blood glucose results and various technologies in generally relevant specialist medical literature.
Within ±10 mg/dL (Within ±0.56 mmol/L)
Within ±15 mg/dL (Within ±0.83 mmol/L)
11 usaGe limits for specialist personnel from the healthcare
sector
1. If the patient shows the following symptoms, it may be the case that no correct values can be displayed:
• Acute dehydration
• Acute hypotension (low blood pressure)
• Shock
• Hyperosmolar hyperglycaemic condition (with or without ketosis)
36 Beurer GL44 mg/dL
Page 37
2. Lipaemic samples: cholesterol levels up to 500 mg/dL and triglycerid levels up to 1000 mg/dL do not influence the results. Severely lipaemic blood samples were not tested with the Beurer GL44 blood glucose monitor; therefore, using the device with these samples is not recommended.
3. In the case of severely ill patients, blood glucose monitors for home use should not be used.
4. The influence of interfering substances on the measurements depends on the concentration in the blood. The maximum concentrations of certain substances listed below do not significantly influence the measurements.
Influence
Blood glucose
Concentration of tested substances
Acetaminophen 7 mg/dL (0.46 mmol/L) 8.1 mg/dL
Ascorbic acid 4 mg/dL (0.26 mmol/L) 6.6 mg/dL
Bilirubin 1.2 mg/dL (0.02 mmol/L) 0.2 mg/dL
Cholesterol 500 mg/dL (12.9 mmol/L) 9.6 mg/dL
Creatinine 30 mg/dL ( 2.7 mmol/L) 1.3 mg/dL
Dopamine 2.2 mg/dL (0.14 mmol/L) 8.0 mg/dL
Galactose 20 mg/dL (1.11 mmol/L) 6.2 mg/dL
Gentisic acid 7 mg/dL (0.45 mmol/L) 9.8 mg/dL
Glutathione 1 mg/dL (0.03 mmol/L) 3.7 mg/dL
Haemoglobin 300 mg/dL (0.05 mmol/L) 3.8 mg/dL
Ibuprofen 50 mg/dL (2.43 mmol/L) 3.9 mg/dL
Icodextrin 5 mg/dL (0.003 mmol/L) 3.6 mg/dL
value
50-100 mg/dL (2.8-5.6 mmol/L)
250-350 mg/dL (13.9-19.4 mmol/L)
5.3%
(0,45 mmol/L)
5.8%
(0,37 mmol/L)
5.2%
(0,01 mmol/L)
7.2%
(0,53 mmol/L)
1.6%
(0,07 mmol/L)
3.2%
(0,44 mmol/L)
2.5%
(0,34 mmol/L)
3.6%
(0,54 mmol/L)
6.5%
(0,21 mmol/L)
5.2%
(0,21 mmol/L)
2.7%
(0,22 mmol/L)
1.4%
(0,20 mmol/L)
Beurer GL44 mg/dL 37
Page 38
Influence
Blood glucose
Concentration of tested substances
L-dopa 2 mg/dL (0.10 mmol/L) 10.0 mg/dL
Maltose 20 mg/dL (0.56 mmol/L) 6.5 mg/dL
Methyldopa 4 mg/dL (0.19 mmol/L) 9.0 mg/dL
Pralidoxime iodide 5 mg/dL (0.14 mmol/L) 2.8 mg/dL
Sodium salicylate 40 mg/dL (2.5 mmol/L) 4.3 mg/dL
Tolbutamide 100 mg/dL (3.70 mmol/L) 1.4 mg/dL
Tolazamide 2.5 mg/dL (0.08 mmol/L) 2.5 mg/dL
Triglyceride 1000 mg/dL (11.3 mmol/L) 4.0 mg/dL
Uric acid 20 mg/dL (1.2 mmol/L) 7.2 mg/dL
Xylose 9.5 mg/dL (0.63 mmol/L) 7.0 mg/dL
value
50-100 mg/dL (2.8-5.6 mmol/L)
(0,56 mmol/L)
(0,36 mmol/L)
(0,50 mmol/L)
(0,16 mmol/L)
(0,24 mmol/L)
(0,08 mmol/L)
(0,14 mmol/L)
(0,22 mmol/L)
(0,40 mmol/L)
(0,39 mmol/L)
250-350 mg/dL (13.9-19.4 mmol/L)
8.7%
4.2%
3.7%
3.3%
2.2%
2.3%
3.6%
7.4%
4.0%
7.5%
38 Beurer GL44 mg/dL
Page 39
12 Guarantee and customer service
Warranty
We oer a 3-year product warranty against material and production faults. The warranty does not cover:
Damage due to improper operation
Wearing parts
Deficiencies already known to the customer upon purchase
Cases of personal negligence on the part of the customer
Outside interference The warranty does not aect the customer’s statutory rights. To assert a warranty claim within the warranty period, the customer must provide proof of purchase. The warranty claim must be submit­ted within a period of 3 years from the date of purchase to BEURER GmbH, Söflinger Straße 218, D-89077 Ulm (Germany). In the event of a warranty claim, the customer reserves the right to have the goods repaired at our own workshop or at a workshop authorised by us. No further rights are granted to the customer (on the basis of the warranty).
Customer service address
Please contact our customer service if you have any questions: Please refer to the address list en closed for our customer service address.
OUR COMMITMENT TO YOU: We aim to satisfy you by providing high-quality medical products and the best customer service. Please contact our customer service if you are not entirely satisfied with the product.
-
Beurer GL44 mg/dL 39