Belimed WD-290 User manual

Page 1
Instruction Manual WD290
Article number: 10268-0068
Version: 001-03/08
Page 2
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Before you read on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Amendments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Symbols and references used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 For your safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Duty of care in handling the device. . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Non-intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.4 Instruction of personnel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.5 Fields of application for the device . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.6 Process validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3 Device description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1 Device unclean side (US) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Device clean side (CS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Controller unclean side (US) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4 Controller clean side (CS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4 Pre-treating medical devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.1 Responsibility for pre-treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.1.1 SOPs (Standard Operating Procedures) . . . . . . . . . . . . . . . . . . . . 10
4.2 Preparing medical devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.3 Pre-cleaning pre-treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.3.1 Avoiding subsequent cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5 Preparing the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
6 Self-disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.1 Why self-disinfection? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.2 Starting self-disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
7 User identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.1 Identification via operating panel . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.2 Identification via barcode reader . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
8 Loading and identifying racks . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
8.1 Loading racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
8.2 Rack identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
8.2.1 Rack identification via the operating panel . . . . . . . . . . . . . . . . . . . 17
8.2.2 Rack identification via barcode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
9 Identification of batch content. . . . . . . . . . . . . . . . . . . . . . . . . . . 18
9.1 Batch content identification via operating panel . . . . . . . . . . . . . . . 18
9.2 Batch content identification via barcode . . . . . . . . . . . . . . . . . . . . . 18
10 Loading from the unclean side . . . . . . . . . . . . . . . . . . . . . . . . . . 19
10.1 Manual loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
10.2 Automatic rack station (optional). . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2/39
11 Washing and disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
001 Instruction Manual WD290
10268 © Belimed
Page 3
11.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
11.2 Washing, disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
12 Unloading from the clean side. . . . . . . . . . . . . . . . . . . . . . . . . . . 23
13 Switching off the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
14 Daily servicing and cleaning work . . . . . . . . . . . . . . . . . . . . . . . 25
14.1 Servicing in general . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
14.2 Servicing the bottom washing arm . . . . . . . . . . . . . . . . . . . . . . . . . 25
14.3 Servicing the top washing arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
14.4 Cleaning the surface sieve and coarse sieve . . . . . . . . . . . . . . . . . 27
15 Device fails to clean properly . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
15.1 Checking the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
15.2 Checking dosing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
15.3 Checking rack loading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
16 Faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
16.1 Fault text without process interruption . . . . . . . . . . . . . . . . . . . . . . 29
16.2 Fault text with process interruption . . . . . . . . . . . . . . . . . . . . . . . . . 30
16.3 Device does not run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
17 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
17.1 Modem connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
17.1.1 Activating the connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
17.2 Independent measurement data acquisition IPD . . . . . . . . . . . . . . 31
17.3 Tank heating manually switchable . . . . . . . . . . . . . . . . . . . . . . . . . 32
17.3.1 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
17.3.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
17.4 Built-in printer CS / US . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
17.4.1 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
17.4.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
17.4.3 Changing paper rolls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
17.5 Emergency stop CS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
17.5.1 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
17.5.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
18 Conformity and certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
19 Glossary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
20 Organisation Belimed AG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
20.1 Manufacturer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
20.2 Subsidiaries, Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
21 Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Instruction Manual WD290 001 © Belimed 10268
3/39
Page 4
Introduction

1 Introduction

1.1 Before you read on

Your product meets high standards and is easy to operate. Nevertheless, please take time to read these instructions carefully. You will become familiarised with your product and be able to use it to its best.

1.2 Target group

These instructions are a component of the product and are intended for the own­er, users, operators, as well as servicing personnel. They must be accessible for this group of persons.

1.3 Amendments

The text, graphics and data correspond to the technical status of the product at the time of going to print. Amendments in the sense of technical development re­main reserved.

1.4 Symbols and references used

The following symbols and references to occupational safety used throughout the documentation are important to avoid harm to health and life.
DANGER! There is an imminent risk to the life and health of persons.
WARNING! There may be a risk to the life and health of persons.
CAUTION! A situation for which there is a warning of damage to property and equip­ment.
NOTE User tips and useful information on the best possible use of the equipment.
4/39
001 Instruction Manual WD290
10268 © Belimed
Page 5
For your safety

2 For your safety

With the EC Declaration of Conformity and the CE mark, we affiirm that this prod­uct complies with the basic health and safety requirements in accordance with Directive 93/42/EEC Annex II (see Chap. 18 "Conformity and certifications")
Hazards may still arise from the product if it is used incorrectly by inadequately trained personnel or not as intended.

2.1 Intended use

This product is exclusively approved for the uses stated in the instructions. Namely for central sterilisation, substerilisation in surgery, in hospitals, clinical laboratories and in industry.

2.2 Duty of care in handling the device

• Only use original racks, spare parts and accessories
• Load the racks as intended (see Chap. 8 "Loading and identifying racks")
• Daily servicing work on the device must be carried out regularly and in ac-
cordance with regulations (see Chap. 14 "Daily servicing and cleaning work")
• Validation of the programme parameters must be performed regularly (see
Chap. 2.6 "Process validation")
• Installation, deinstallation, servicing or modification must only be carried
out by persons authorised by Belimed
• In the case ofincorrectly installed, operated or maintained devices all
warranty claims are invalidated

2.3 Non-intended use

All other applications are considered as non-intended use.
Damages caused by operator error, non-intended use, failure to observe the in­structions, operation by untrained personnel, unauthorised modifications and conversions without the written consent of the manufacturer, are not permitted and preclude the manufacturer's liability for the ensuing damage to property and personal injury.

2.4 Instruction of personnel

This product must only be used, maintained and repaired by authorised, trained and briefed personnel. This assumes that these instructions are read and under­stood.
Responsibilities and competencies in operation, servicing and maintenance must be clearly defined and observed.
Instruction Manual WD290 001 © Belimed 10268
5/39
Page 6
For your safety

2.5 Fields of application for the device

Cleaning and conditioning of:
• Surgical instruments
• Minimal-invasive instruments
• Instruments for anaesthesia and intensive care
• Baby bottles and teats
• Containers
• OP shoes
• Laboratory instruments from research and production
• Rigid endoscopes
• Eye instruments
• Neurosurgery

2.6 Process validation

The aim of process validation is to achieve a high level of safety in the recondi­tioning of medical devices in order to afford the operators and patients the great­est possible protection.
Process validation consists of:
a) Type testing / factory testing
b) Process validation consisting of:
• IQ - Installation Qualification
• OQ - Operational Qualification
• PQ - Performance Qualification
c) Routine testing / Annual requalification
(refer to appendix EN ISO 15883-1 November 2001, Chap. 6, Pages 35-37)
6/39
NOTE Further information on process validation may be obtained from Belimed Customer Service.
CAUTION! Validation must only be carried out by authorised persons!
Devices must only be operated with processes validated in accordance with regulations! Only use components (items to be washed, racks, programmes, chemicals) which have been validated together.
The safety of operators and patients may be compromised if the devices used are not validated in accordance with regulations.
001 Instruction Manual WD290
10268 © Belimed
Page 7
Device description
1
3
4
5
2
8
7
9
10
Abb 201
6

3 Device description

3.1 Device unclean side (US)

1 Door drive (behind the panel) 2 Wash compartment doors made of double-layer safety glass 3 Operating panel 4 Main switch (with EMERGENCY OFF function) 5 Service panel 6 Location type plate 7 Transfer trolley 8 Rack 9 Automatic wash compartment doors 10 Door safety switching strip
Instruction Manual WD290 001 © Belimed 10268
7/39
Page 8
Device description
1
2
4
3
Abb 301

3.2 Device clean side (CS)

8/39
1 Door drive (behind the panel) 2 Operating panel with door button 3 Wash compartment doors made of double-layer safety glass 4 Service panel
001 Instruction Manual WD290
10268 © Belimed
Page 9
Device description
1
2
3
4
5
6
Program 1
Program 2
Program 3
Program 4
Program 5
Program 6
1
432
5
6
7
8
Abb 179
RUN
ERROR
1
2
Abb 178
4
3

3.3 Controller unclean side (US)

1) Display
– With screensaver "BELIMED INFEC-
TION CONTROL"; this means that this display automatically appears af­ter approx. 1 h. Press any button, dis­play"Program ready" appears again
2) Cursor left
– Print operating data such as program
formula and setup parameters – Self disinfection On/Off – IPD verification On/Off
3) Cursor right
– Acoustic signal On/Off
4) Cursor down
– Printer On/Off
5) Cursor up
– Shift button Programs P7 - P12
6) Program buttons
– Selection of Programs P1 - P6, with
Shift P7 - P12
7) Door button
– Opens/close door
8) On/Off button
– Display batch number (press 4 sec-
onds)

3.4 Controller clean side (CS)

Instruction Manual WD290 001 © Belimed 10268
1) Display
– With screensaverBELIMED INFEC-
TION CONTROL; this means that this
display automatically appears after
approx. 1 h. Press any button, dis-
play"Program ready" appears again
2) Door button
– Door opens/closes
3) RUN
LED lit = In progress
4) ERROR
LED lit = Fault (See Chap. 16 "Faults")
9/39
Page 10
Pre-treating medical devices

4 Pre-treating medical devices

4.1 Responsibility for pre-treatment

CAUTION! Always observe the specifications from the medical device manufacturer!
The owner is responsible for pre-treatment of medical devices. The best possible washing results are only to be achieved with correct pre-treatment as intended.
Various treatments fix proteins and may contribute to preserving prion infectivity.
CAUTION! The air bubbles in the foam prevent pressure building-up in the cleaning sys­tem and therefore the best possible contact between the cleaning agent and the items to be washed.

4.1.1 SOPs (Standard Operating Procedures)

The contents of the following criteria and specifications must be regulated:
• Product responsibility
• Transport routes and duration of waste disposal (time for soiling to dry in)
• Type of soiling (blood, ointments, bone meal...)
• Material properties and compatibility of the items to be washed (risk groups acc. to RKI Ordinance)
• Consideration of all operating instructions and reconditioning regulations for medical devices
• Necessary knowledge of the medical devices to be reconditioned
• Maintenance plan and regular inspections
Belimed recommends producing work instructions which describe the procedure within a working process.

4.2 Preparing medical devices

CAUTION! Not all medical devices are suitable for mechanical reconditioning (see Chap.
2.5 "Fields of application for the device")
All inner and outer surfaces must be accessible for cleaning (open valves, taps, joint instruments...) Special caution is required for the lumen. Disassemble MIC or other complex instruments according to the manufacturer's specifications.
10/39
001 Instruction Manual WD290
10268 © Belimed
Page 11
Pre-treating medical devices

4.3 Pre-cleaning pre-treatment

Remove coarse soiling immediately after use. Dried-on blood or tissue reduces the effectiveness of cleaning.

4.3.1 Avoiding subsequent cleaning

Various treatments fix proteins and may contribute to preserving prion infectivity.
The following pre-treatment methods may cause impairments in subsequent washing:
• Pre-treatment with aldehydic disinfectants
• Pre-treatment with alcohol solutions
• Pouring antiseptic solutions on the items to be rinsed
• Aldehyde and alcohol vapours
• Heat pre-treatment
Small quantities of foam are permissible with manual cleaning in the immersion bath or ultrasonic cleaning.
Instruction Manual WD290 001 © Belimed 10268
11/39
Page 12
Preparing the device
0 I
Abb 209

5 Preparing the device

WARNING! Only operate the device if it is in a technically faultless condition! Damaged or defected components must be reported to the technical specialist.
After a prolonged period of disuse (approx. 1 week), the daily servicing work must be carried out to prepare the device (see Chap.14 "Daily servicing and cleaning work").
Check the quantity of detergents (see Chap. 16 "Faults")
Switch the device on at the main switch (Fig. 209) with the button. Four
possible display texts may now appear
Performing self-disinfection
(See Chap. 6 "Self-disinfection")
Display text:
Self-disinfection
Start
Program ready
Continue with program flow (see Chap. 7 "User identification")
Display text:
Program ready
Fault without process interruption
Rectify fault (see Chap. 16 "Faults")
Example display text:
Dosing device
Empty
Fault with process interruption
Report to the technical specialist
Example display text:
No pressure
Fault Code 110
12/39
001 Instruction Manual WD290
10268 © Belimed
Page 13
Self-disinfection
1
6

6 Self-disinfection

6.1 Why self-disinfection?

Microorganisms form in the washing chamber, tubing and DI boiler after a pro­longed period of disuse. Following thermal disinfection step, residues of dead bacteria on the medical device may constitute a risk for patients.
In the case ofactive self-disinfection the owner is requested to perform self-dis­infection after a programmed period (24 hours as standard) following the last pro­gram run.
NOTE If there is no mains supply (mains switch OFF) the timer is not active, as the request for"Self Disinfection Start" always appears when active self-disinfec­tion is switched on.

6.2 Starting self-disinfection

Open the door with the button
Slide in the empty rack
CAUTION! Do not wash any items with the self-disinfection program! There is no cold pre-rinse. There is the risk of excessive foam formation and fixing proteins.
Close the wash compartment door with the button
User identification (see Chap. 7 "User identification")
WARNING! Do not touch the door or its surrounding paneling during the closing process. A crushing hazard exists.
If any objects are trapped between the panel and door, the technical person­nel must be informed to rectify the problem.
The device is ready to start
Display text:
Self Disinfection
Start
Press any program button to
Display text:
SD |_| 13min
Desinf A068 89°C
Legend:
SD |_| = program step status
13min = remaining run time
Desinf = program name
A068 = current A0 value
89°C = temperature of rinsing agent or drying air
Instruction Manual WD290 001 © Belimed 10268
13/39
Page 14
Self-disinfection
0 I
Procedure in case of faults:
If a fault occurs during operation, acknowledge with the buttonIf the fault persists (see Chap. 16 "Faults")If the fault cannot be rectified, the technical specialist must be informed
DANGER! If smoke is emitted or water escapes, immediately turn off at the main switch and isolate the device from the mains supply. Inform the technical specialist.
The end of the program is indicated with an acoustic signal
Display text:
Program name
Correctly Finished
NOTE For manual loading/unloading, unload the rack back to the US again at the end of the program. The CS door remains locked.
Self-disinfection is inactive for automatic loading/unloading.
14/39
001 Instruction Manual WD290
10268 © Belimed
Page 15
User identification

7 User identification

The user must logon each time before using the device. The password identifi­cation is forwarded to the printer or the digital documentation system and as­signed to the relevant batch.
Detection takes place via the operating panel or by means of barcode reader.

7.1 Identification via operating panel

User identification is requested once the device is switched on.
User Name o No.?
--
Inputting the identification number via the operating panel
NOTE For input via the operating panel we recommend a list of numbers with the corresponding names (Example 1).
The identification number must lie within the range 11-65. Only the digits 1-6
can be used for a number. (Not possible: 17, 18, 19, 27, 28, 29....)
For an input of No. 0 or 66 "No identification" is output.
Example 1:
No. Name
11 M. Smith
12 H. Jones
0 No Identification
66 No Identification
Device ready
Display text:
Program ready
Display text for a prolonged period of disuse (see Chap. 6 "Self-disinfection"):
Self Disinfection
Start
Instruction Manual WD290 001 © Belimed 10268
15/39
Page 16
User identification
Abb 182

7.2 Identification via barcode reader

User identification is requested once the device is switched on.
Display text:
User Name o No.?
--
Reading the barcode with the barcode reader
NOTE The length of the name input is limited to 20 characters. If the no. 0 or 66 is read"No identification" is output (see Example 2)
Barcodes for name labels can be produced oneself from the Internet (free of charge). (http://www.barcodemagic.com/barcodemagic.html).
Display text:
Example 2:
First name / last name
Heike Meyer
No Identification
0 / 66
Program ready
Display text:
Program ready
--.--.---- --:--
Display text for a prolonged period of disuse (see Chap. 6 "Self-disinfection"):
Self Disinfection
Start
16/39
001 Instruction Manual WD290
10268 © Belimed
Page 17
Loading and identifying racks

8 Loading and identifying racks

8.1 Loading racks

The loading of racks has a significant influence on the washing outcome. For correct and intended loading it is absolutely necessary to follow the in­structions supplied with the rack.
DANGER! Always wear protective goggles and gloves when handling the items to be washed. Soiled items to be washed may cause infections!

8.2 Rack identification

The rack identification is forwarded to the printer or the digital document system and assigned to the relevant batch.
Detection takes place via the operating panel or by means of barcode reader.

8.2.1 Rack identification via the operating panel

Display text:
Rack Name o No.?
--
Inputting the identification number via the operating panel
NOTE The identification number must lie within the range 11-65. Only the digits 1-6
can be used for a number. (Not possible: 17, 18, 19, 27, 28, 29....)
For an input of No. 0 or 66 "No identification" is output.

8.2.2 Rack identification via barcode

Display text:
Rack Name o No.?
--
Read barcode
NOTE The length of the name input is limited to 12 characters. If the no. 0 or 66 is read"No identification" is output (see Chap. 7.2 "Identification via barcode reader")
Barcodes can be produced oneself from the Internet (free of charge). (http:// www.barcodemagic.com/barcodemagic.html).
Instruction Manual WD290 001 © Belimed 10268
17/39
Page 18
Identification of batch content
0 I
Abb 227

9 Identification of batch content

The identification of batch content is forwarded to the printer or the digital docu­mentation system and assigned to the relevant batch.
Detection takes place via the operating panel or by means of barcode reader.

9.1 Batch content identification via operating panel

Display text:
Read Content No.
No=--?
Inputting the identification number via the operating panel
NOTE The identification number must lie within the range 11-65. Only the digits 1-6
can be used for a number. (Not possible: 17, 18, 19, 27, 28, 29....)
For an input No. 66 "No identification" is output.
A maximum of 18 objects can be identified. The same number cannot be
input twice. Always complete input with 66.

9.2 Batch content identification via barcode

Display text:
Quantity of Contents
--
Reading the batch content with the barcode reader
NOTE The reading process can be cancelled with the button. Always conclude with no. 66.
Reading is terminated after reaching a max. 18 batch contents.
NOTE The length of the name input is limited to 12 characters. If the no. 0 or 66 is read "No identification" is output (see Chap. 7.2 "Identification via barcode reader")
Barcodes can be produced oneself from the Internet (free of charge). (http:// www.barcodemagic.com/barcodemagic.html).
Example:
18/39
001 Instruction Manual WD290
10268 © Belimed
Page 19
Loading from the unclean side
1
3
2
4
Abb 203

10 Loading from the unclean side

DANGER! Always wear protective goggles and gloves when loading. Soiled items to
be washed may cause infections!

10.1 Manual loading

Attach the transfer trolley with the loaded rack to the device (Fig. 203 Pos. 1)
1 Attachment and detachment hook 2 Rack 3 Locking hook 4 Transfer trolley
Open the washing compartment with the button
Slide the loaded rack into the washing chamber (Fig. 203)
- Rack docking must coincide with the device docking
- The rack has to lock into the guide (Fig. 168)
CAUTION! Sharp or pointed objects may fall down when sliding the rack. This can cause injuries.
The rinsing arms must not be blocked by too high objects or objects protruding downwards. Check rotation by hand!
Instruction Manual WD290 001 © Belimed 10268
19/39
Page 20
Loading from the unclean side
Abb 168
1
2
1 Rack 2 Rack downholder
Close the washing compartment with the button
WARNING! Do not touch the door or its surrounding panel during the closing process. A crushing hazard exists.
If any objects are trapped between the panel and door, the technical person­nel must be informed to rectify the problem.
The device is ready to start.
Program ready
dd.mm.20yy hh:mm
NOTE Check that the date and time are correct. This is important for the batch doc­umentation. Report discrepancies to the technical specialist!
20/39
001 Instruction Manual WD290
10268 © Belimed
Page 21
Loading from the unclean side
Abb 206
1
5
2
3
4

10.2 Automatic rack station (optional)

For correct and intended operation and servicing it is absolutely necessary to fol­low the instructions supplied wit the automated rack station.
1 Rack 2 Rack station Emergency Stop 3 Rack station operating buttons 4 Automatically driven rollers 5 Rack station
Instruction Manual WD290 001 © Belimed 10268
21/39
Page 22
Washing and disinfection
1
6
0 I

11 Washing and disinfection

11.1 General information

WARNING! Only operate the device with the original metal paneling! Uncontrolled escape of water due to a burst pipe or split tubing can cause scalding. Only remove the metal panel once the device is isolated from the mains supply.

11.2 Washing, disinfecting

Press the required button to
This step proceeds automatically for automated program recognition (option­al)
NOTE In the case of "automatic program recognition" check the items to be washed against the program name!
Program flow
Display text:
P2|____| 13min
Wash 36°C
P2|____| = program step status
13min = remaining run time
Wash = current step
36°C = temperature of washing agent or drying air
Procedure in case of faults:
If a fault occurs during operation, acknowledge with the buttonIf the fault persists (see Chap. 16 "Faults")If the fault cannot be rectified, the technical specialist must be informed
DANGER! If smoke is emitted or water escapes, immediately turn off at the main switch and isolate the device from the mains supply. Inform the technical specialist.
The end of the program is indicated with an acoustic signal
Display text:
Program name
Correctly Finished
22/39
001 Instruction Manual WD290
10268 © Belimed
Page 23
Unloading from the clean side
1
2
3
4
5
Abb 204

12 Unloading from the clean side

Attach the transfer trolly to the device
Open the CS washing compartment door with the button.
Remove the washed items (Fig. 204)
CAUTION! The washed items and washing chamber are hot! Always wear protective goggles and gloves when unloading.
1 Rack 2 Transfer trolley 3 Unlocking lever 4 CS door button 5 Attachment and detachment hook
Open the CS washing compartment door with the button
On the display appears "Program ready"
Visually inspect the results of cleaning. In the case of washed items that are
unclean, repeat washing with the same program!
CAUTION! There must be no more soiling (incrustations, depositions) visible.
Detach the transfer trolley with the lever (Fig. 204 Pos. 3)
Washed items ready for further processing
Instruction Manual WD290 001 © Belimed 10268
23/39
Page 24
Switching off the device
0 I

13 Switching off the device

Switch off the device after use
CAUTION! Switch off the device with the button .
For servicing, switch off the device at the main switch (see Chap. 14.1 "Ser­vicing in general").
NOTE If the device is switched off at the main switch, the controller is also no longer active. Consequently, the water level is not checked and, in the case of a bro­ken valve, water can overflow.
24/39
001 Instruction Manual WD290
10268 © Belimed
Page 25
Daily servicing and cleaning work
2
1
5
4
3
6
Abb 177
7

14 Daily servicing and cleaning work

14.1 Servicing in general

The owner is responsible for carrying out servicing and cleaning work.
DANGER! Switch off at the main switch prior to all servicing work.
Only carry out servicing and cleaning work with safety goggles and gloves! Soiling residues can cause infections!
CAUTION! Never spray the device for cleaning. The device is not water-jet-proof.
Only use chrome steel cleaning agents or surface disinfectants for cleaning the outside of the device. No solvents!

14.2 Servicing the bottom washing arm

WARNING! Pointed or sharp objects can cause injuries when washing the device.
1 Clamping fixture 2 Direction for loosening 3 Top bearing yoke 4 Bottom washing arm 5 Bottom bearing yoke 6 Supporting bearing 7 Sealing O-ring for clamping fixture
Twist the clamping fixture (Fig. 177 Pos. 1) counter-clockwise and pull it up-
wards
Disassemble the remaining components acc. to Fig. 177
Check the seal on the clamping fixture (Fig. 188 Pos. 7) for damage
Instruction Manual WD290 001 © Belimed 10268
25/39
Page 26
Daily servicing and cleaning work
Abb 108
2
1
4
3
5
Check the washing arm (Fig. 177 Pos. 4) for clogged nozzles and clean as
required
Clean the bearing yokes (Fig. 177 Pos. 2+5), check for wear and replace as
required
Replacement criteria: Score marks or deep scratches
Reassemble components
CAUTION! Rinsing nozzles upwards!
Check the rotation of the washing arm

14.3 Servicing the top washing arm

The washing arm must be checked daily and clogged nozzles cleaned.
1 Washing arm 2 Top washing arm 3 Dome nut 4 Slide ring 5 Rotor bearing
DANGER! Switch off at the main switch prior to all servicing work.
Only carry out servicing and cleaning work with safety goggles and gloves! Soiling residues can cause infections!
Loosen dome nut (Fig. 108 Pos. 3)
NOTE Do not lose the washer (Fig. 108 Pos. 4)!
Disassemble components (Fig. 108)
Check the washing arm (Fig. 108 Pos. 2) for clogged nozzles and clean as
required
Check the rotor bearing (Fig. 108 Pos. 5) for wear and replace as required
26/39
001 Instruction Manual WD290
10268 © Belimed
Page 27
Daily servicing and cleaning work
Abb 96
3
4
1
2
Replacement criteria: Score marks or deep scratches
Reassemble components
Check the rotation of the washing arm

14.4 Cleaning the surface sieve and coarse sieve

DANGER! Switch off at the main switch prior to all servicing work.
Only carry out servicing and cleaning work with safety goggles and gloves! Soiling residues can cause infections!
Remove surface sieve (Fig. 96 Pos. 4) and clean as required
Remove coarse sieve (Fig. 96 Pos. 3) and clean as required
CAUTION! The tank heating elements can be very hot!
1 Tank heating elements 2 Level float 3 Coarse sieve 4 Surface sieve
Removing foreign bodies from the washing compartment
Reinsert the coarse sieve correctly
Reinsert the surface sieve correctly
CAUTION! Never rinse without the surface sieve and coarse sieve!
Instruction Manual WD290 001 © Belimed 10268
27/39
Page 28
Device fails to clean properly

15 Device fails to clean properly

15.1 Checking the device

• Is the right program selected for the items to be washed?
• Do the washing arms rotate? (See Chap. 14 "Daily servicing and cleaning work")
• Are the washing arms correctly installed? (See Chap. 14 "Daily servicing and cleaning work")
• Are the washing arm nozzles clogged? (See Chap. 14 "Daily servicing and cleaning work")
• Are the suction and intermediate sieves clogged or absent? (See Chap. 14.4 "Cleaning the surface sieve and coarse sieve")
• Does the trolly dock to the device properly? (See Chap. 14.2 "Servicing the bottom washing arm")
• Are the waste water and media connections correctly established? (See Chap. 5 "Preparing the device") The checks are to be performed by the
technical specialist!

15.2 Checking dosing

• Is there sufficient cleaning agent in the containers? (See Chap. 16.1 "Fault text without process interruption")

15.3 Checking rack loading

• Are the items to be washed properly loaded on the trolley provided for this purpose? (See Chap. 8 "Loading and identifying racks")
• Are the washing nozzles on the rack clogged? (See Chap. 8.1 "Loading racks")
• Is the rack inserted correctly? (See Chap. 10.1 "Manual loading")
• Have the items to be washed been proberly pre-treated? (See Chap. 4 "Pre­treating medical devices")
28/39
001 Instruction Manual WD290
10268 © Belimed
Page 29
Faults
0 I
0 I

16 Faults

16.1 Fault text without process interruption

Fault text display Cause Remedy
Doser _ _ _ _
Empty
Program Recognition
Fault Code 118
Door Interlock
Select Pr again
Probe Difference
Is too Large
Replace Filter
Periodical Service
Exhaust facility
defective
Peak Load Block
No Communication
Documentation system
Doser container is empty.
Empty indicator float incorrectly fit­ted or defective.
1) Fit a new canister
2) Pull the empty indicator lance out of the empty canister and correctly insert it into the new canister.
3) Activate the washing program with the button
Caution!
Always use protective goggles and gloves when refilling the detergents!
The rack position in the washing chamber is not correct.
Reed sensor for"automatic program
Pull out the rack and slide it in again.
Inform the technical specialist.
recognition" is defective.
If the program is interrupted with
New program can be selected. the button the washing com­partment door remains locked.
Difference between the monitoring probe and the regulating probe is too large (+- 2°C).
New program can be selected.
Inform the technical specialist
immediately. Recalibrate the probes
at least at 80°C.
The sterile filter for drying is clogged.
Servicing promt after the specified batch number. (Generally every 1500 batches)
Inform the technical specialist
immediately.
New program can be selected.
Initiate service call from contractual
partner.
Building exhaust facility failure. New program can be selected.
Inform the technical specialist
immediately.
The building peak load manage­ment has blocked the device as a load.
Connection to PC for batch record­ing is interrupted.
A new program cannot be selected.
Wait until it is enabled once again.
New program can be selected.
Re-establish the connection
between the device and PC for
batch recording.
Instruction Manual WD290 001 © Belimed 10268
29/39
Page 30
Faults
Abb 209

16.2 Fault text with process interruption

WARNING! Report fault text with process interruption to the technical specialist immedi­ately. Under no circumstances continue working with the device!
Fault texts with process interruption appear during the program flow. The fault code and the cause of the fault blink alternately on the display (exception Fault Code 101).
Example:
Leakage Pan
Fault Code 112

16.3 Device does not run

• Is the device switched on at the main switch? (Fig. 209)
• Are the fuses intact? (inspection by the technical specialist)
• Are the washing compartment doors correctly locked? (See Chap. 5 "Prepar­ing the device")
• Fault code on the display? (See Chap. 16 "Faults")
30/39
001 Instruction Manual WD290
10268 © Belimed
Page 31
Options
0 I

17 Options

17.1 Modem connection

If a modem is installed, the device can be linked with the Belimed Service Centre. Inform the Service Centre before activating the modem (see Chap. 20 "Organi­sation Belimed AG".

17.1.1 Activating the connection

Display text Description
Program ready
XX.XX.XXXX 12:00
Password ?
---
Remote Support
Initialize Modem
Remote Support
Not Connected
Remote Control
Connected
In standby mode press the button for 4 sec­onds
Press the button for 4 seconds
Modem connection is established
Modem connection could not be established
Modem connection is established. Device is remote controlled

17.2 Independent measurement data acquisition IPD

With automatic verification, the ongoing process is interrupted with the relevant error message. The function can be switched on and off from the operating panel. For function "On" a process interruption follows, for "Off" only a text on the dis­play.
Activation:
Device must be on the status
Program ready
dd.mm.20yy hh:mm
Press the button 3 times until"Verification On/Off" appears
Process Verification
On
Use the button to switch between "On" or "Off"
Instruction Manual WD290 001 © Belimed 10268
31/39
Page 32
Options
Abb 231
2
1
3

17.3 Tank heating manually switchable

17.3.1 Function

The selection switch (Fig. 231 Pos.3) is used to switch between electrical heating E (Fig. 231 Pos. 2) for the washing chamber and steam heating D (Fig. 231 Pos.1).
1 Steam heating 2 Electrical heating 3 Manual selection switch

17.3.2 Operation

Fault 161 or 162 appears
No heat
Fault Code 161
No heat
Fault Code 162
Acknowledge fault with door button
Manual switching with the selection switch (Fig. 231 Pos. 3) from electrical
heating (Fig. 231 Pos. 2) to steam heating (Fig. 231 Pos. 1)
Restart program
CAUTION! If the same fault still occurs after switching to steam, the technical specialist must be informed immediately (see Chap. 16.2 "Fault text with process inter­ruption")
32/39
001 Instruction Manual WD290
10268 © Belimed
Page 33
Options
Abb 232
2
4
5
1
3

17.4 Built-in printer CS / US

17.4.1 Function

The built-in printer allows operating data, such as process steps and error mes­sages, to be recorded and printed out.
NOTE The function of the built-in printer is the same for both built-in versions CS/ US.
1 Recess lid opens upwards 2 Paper slit 3 Costom built-in printer 4 Paper feed button 5 Print button (not active)
Instruction Manual WD290 001 © Belimed 10268
33/39
Page 34
Options
4
2
5
4
Abb 233

17.4.2 Operation

Display text "Program ready"
Activate printer with the button
Archiving printed data
Use the button and the button to print out operating data
Deactivate the printer with the button

17.4.3 Changing paper rolls

Open up the lid on the recess (Fig. 232 Pos. 1)
Replace roll acc. to brief instructions on the inside of the lid
Close lid. Feed paper through the paper slit (Fig. 232 Pos. 2)
Printer ready for operation again

17.5 Emergency stop CS

All program steps, faults and program interruptions are now recorded and printed out
Operating data = program formulas, setup data

17.5.1 Function

If necessary, an ongoing process can be interrupted on the clean side with the additional emergency stop.

17.5.2 Operation

If a problem occurs during a process, implement the following steps without de­lay:
Press the emergency stop (Fig. 233)
The ongoing process is interrupted immediately
Rectify fault (see Chap. 16 "Faults")
Restart program
34/39
001 Instruction Manual WD290
10268 © Belimed
Page 35
Conformity and certifications

18 Conformity and certifications

Instruction Manual WD290 001 © Belimed 10268
35/39
Page 36
Glossary

19 Glossary

Authorised persons
Approval for special persons, especially the issuance of rights, for third-party us­age as applicable.
Operators Persons who load the device and perform simple washing and servicing work.
Loading Collective term to describe all items, devices and materials that are inserted into
a washing/disinfection device for the purpose of treatment in a cycle.
Intended use Use of the device in accordance with the information provided in the user infor-
mation.
Boiler Closed vessel in which water is heated indirectly by the flow of a heated medium
by means of a heat exchanger at a pressure higher than atmospheric pressure.
Disinfection Reduction in the number of viable microorganisms on a product to a predefined
level suitable for its continued handling or use.
Performance qualification
System with which the process cycle can be interrupted or modified depending on requirements; illustrative and documenting evidence that the device, as it is installed and operated according to the operating procedures, works durably in accordance with the predefined criteria and therefore products are obtained that fulfil their definitions, i.e. the washing/disinfection device delivers products that are cleaned and disinfected according to the required standard.
Error, error message
State of a unit in which it is incapable of fulfilling the required function, whereby the functionality caused by servicing or other planned actions or the absence of external resources is excluded.
Hazard Potential source of damage.
Device Aggregate of parts or assemblies combined together, of which at least one is
movable with the corresponding drive elements, control and energy circuits joined together for a specific application, especially for processing, treatment, transporting or packaging of a material.
Device for continuous
Device that automatically transports for loading through each individual process cycle.
operation
Glove Alkali and acid resistant glove
Installation qualification
Illustrative and documenting evidence that the device has been equipped and in­stalled in accordance with the relevant definitions.
CW Cold water
Medical devices All individual or interconnected instruments, apparatuses, device materials or
other objects including the software used for he proper function of the medical device that are determined by the manufacturer for use on persons for the follow­ing purposes:
• identification, prevention, monitoring, treatment or alleviation of diseases;
• Identification, monitoring, treatment, alleviation or compensation of injuries or disabilities;
• examination, replacement or modification of the anatomical structure or of a physiological process;
• contraception,
and their proper main function in or on the human body is neither achieved by pharmacological, immunological nor metabolic means, but whose mode of action may be supported by these means.
Washing Removal of contamination from the surface to be washed with an aqueous me-
dium with or without process chemicals depending on requirements.
36/39
001 Instruction Manual WD290
10268 © Belimed
Page 37
Glossary
Washing/ disinfection device
Device for washing and disinfecting medical products and other objects used in the areas of medicine, dental medicine, pharmaceuticals and veterinary medi­cine.
Note: Device types are excludes that are designed specifically to wash cloths or other laundry. Definitions for devices intended for sterilisation or termed as "ster­ilisers" are specified in other standards , e.g. in EN 285.
Washing/ disinfection device for instruments
Two-door washing/
Washing/disinfection device intended for washing and disinfecting loads contain­ing surgical instruments, anaesthetic accessories, containers, devices, glass de­vices and similar objects.
Washing/disinfection device with separate doors for loading and unloading.
disinfection device
Risk Combination of the probability of occurrence of damage or the extent of damage.
RKI The task assigned to theRobert KochInstitute includes both the observation of
occurrence of diseases and the relevant health hazards for the population, as well as the deriving and scientifically establishing the required measures for ef­fective protection of the health of the population.
CS Clean side
Protective goggles Alkali and acid resistant fully closed protective goggles
Safety Term for a state free of unjustifiable risks or impairment or seen as being free of
hazards.
SOP Standard Operating Procedure is a document describing the approach within an
operational working process. Frequently recurring workflows are described in texts and provided for the person implementing the task.
Controller Installation that controls the washing/disinfection device step-by step through the
required phases of the process cycle(s) or the process according to the defined process parameters.
Fault Identification by the automated controller that the prescribed process variable for
the process cycle of the washing/disinfection device are not adhered to.
Tank Container permanently installed in the washing/disinfection device to store solu-
tions consumed during the process.
Technical specialist Technical specialist responsible for performing simple service and maintenance
work on the device.
Door Mechanism for closing and sealing the chamber.
US Unclean side
Validation, process validation
Documented procedure for achieving, recording and interpreting the required re­sults to demonstrate that a procedure always agrees wit the prescribed specifi­cations.
Lock, locked Mechanical, electrical or other type of mechanism having the purpose of prevent-
ing the execution of hazardous device functions under defined conditions.
Verification Affirmation by providing objective proof that the defined requirements have been
fulfilled [EN ISO 9000:2000]
DI water Deionised or demineralised or fully demineralised water (H2O) without the min-
erals (salts, ions) occurring in normal spring water and mains water.
Rack Mechanism for the correct loading and washing of items.
Washing chamber The part of the washing/disinfection device in which the load is treated.
Note: Steam generator, piping, e.g. drains and connections, from which the chamber is separated, are not included.
WW Warm water
Instruction Manual WD290 001 © Belimed 10268
37/39
Page 38
Organisation Belimed AG
Belgium
Belimed SA Rue de Clairvaux 8 1348 Louvain-La-Neuve Phone +32 10 42 02 40 Fax +32 10 42 02 49 [email protected]
Germany
Belimed Deutschland GmbH Edisonstrasse 7a 84453 Mühldorf am Inn Phone +49 8631 9896 0 Fax +49 8631 9896 300 [email protected]
Branch Office West Belimed Deutschland GmbH Emil-Hoffmann-Strasse 27 50996 Köln Phone +49 2236 9642 0 Fax +49 2236 9642 200 [email protected]
France
Belimed SAS Parc GIVIO 330 Allée des Hetres, Hall E 69760 Limonest Phone +33 4 37 41 63 03 Fax +33 4 37 41 63 04 [email protected]
Pharma Belimed SAS Parc Espale 1, av. Pierre Pflimlin 68390 Sausheim Phone +33 3 89 63 65 40 Fax +33 3 89 63 65 41 [email protected]
United Kingdom
Belimed Limited Unit 4 Newbuildings Place Dragons Green Road West Sussex, RH13 8GQ Phone +44 1403 738 811 Fax +44 1403 730 830 [email protected]
Netherlands
Belimed B.V. Energieweg 8 6658 AD Beneden-Leeuwen Phone +31 487 59 11 00 Fax +31 487 59 15 90 [email protected]
Austria
Belimed GmbH Grüne Lagune 8350 Fehring Phone +43 3155 40 6990 Fax +43 3155 40 699 10 [email protected]
Switzerland
Belimed Sauter AG Zelgstrasse 8 8583 Sulgen Phone +41 71 644 85 00 Fax + 41 71 644 86 00 [email protected]
Slowenia
Belimed d.o.o. Kosovelova cesta 2 1290 Grosuplje Phone +386 1 7866 010 Fax +386 1 7866 011 [email protected]
USA
Belimed Inc. 2284 Clements Ferry Road Charleston, SC 29492 Phone +001 843 216 7424 Fax +001 843 216 7707 [email protected]
Other Countries Medical
Belimed AG Dorfstrasse 4 6275 Ballwil Phone +41 41 449 78 88 Fax +41 41 449 78 89 [email protected]
Other Countries Pharma
Belimed Sauter AG Zelgstrasse 8 8583 Sulgen Phone +41 71 644 85 00 Fax + 41 71 644 86 00 [email protected]
Abb 150
China
Belimed Medical Equipment (Shanghai) Co. Ltd CaiLun Road 780 ZhangJiang Hi-Te ch Park 201203 Pudong, Shanghai Phone +86 21 513 709 98 Fax +86 21 513 709 96 [email protected]
Branch Office West Belimed Sauter AG Ehrlenauweg 17 Phone +41 31 720 44 55 Fax +41 31 720 44 50 [email protected]

20 Organisation Belimed AG

20.1 Manufacturer

Belimed AG, Dorfstrasse 4, CH-6275 Ballwil (Switzerland)
Phone +41 41 449 78 88
Fax +41 41 449 78 89

20.2 Subsidiaries, Customer Service

38/39
001 Instruction Manual WD290
10268 © Belimed
Page 39
Notes

21 Notes

Instruction Manual WD290 001 © Belimed 10268
39/39
Loading...